Home
        "service manual"
         Contents
1.                           c 4  Press the Pulse Rate Alarm Limit button twice to view the current  Pulse Rate lower alarm limit     32    Using the N 550    psa  d       t    Seo   A        5  Press the SatSeconds Alarm Limit button to view the current  SatSeconds setting     pa               Spaz  Kt        LA SatSeconds units are being reviewed  The SatSeconds  12    9 Note  The SatSeconds  12 O clock  indicator lights  indicating that  O clock  indicator lights for all SatSeconds except Off        Alarm Limits Changed Indicator    N 550    If an alarm limit is changed from the N 550 s power on default  setting  a decimal point appears after the applicable displayed value   during patient monitoring and when alarm limits are viewed  The  decimal point remains on the display until the N 550 is turned off or  the limit is returned to the default value         pa     e       8  E       E      990       33    Using the N 550    Alarm limits that have been changed from the default setting are  identified by a decimal point     after the displayed reading   SpO2  or Pulse Rate         Setting Alarm Limits       Discussion    Alarm limits determine the upper and lower points of patient data at  which the N 550 will sound an alarm        Procedure    With the N 550 in the normal monitoring mode     1  Press SpO2 Alarm Limit button to view the current  SpO2 upper  alarm limit        C  2  Press the Adjust Up button or Adjust Down button to increase or  G2 decrease the alarm limit set
2.            Figure 1  Front Panel Buttons and Symbols    1     SpO  Sensor Port    12     SatSeconds Alarm Limit  Button       2     Power On Off Button    13     SpO2 Alarm Limit Button       3     96SpO2 Display    14     Motion Indicator       4     Pulse Amplitude  Indicator    5     Pulse Rate Display    15     Sensor Off Indicator    16     Sensor Message Indicator       6     Alarm Silence Button    17     Pulse Search Indicator       7     Alarm Silence Indicator    18     Data In Sensor Indicator       8     Adjust Up Button    19     Low Battery Indicator       9     Adjust Down Button    20     AC Power Indicator       10     Pulse Rate Alarm Limit  Button    2      Speaker       11     SatSeconds    Display          Description of Controls  Indicators  and Symbols       Identification of Rear Panel Components       1     Data Port Connector   3     AC Power Connector       2     Foot Switch and Visual   4     Equipotential Connector  Alarm Connector  ground     Figure 2  Rear Panel Symbols       N 550 Symbols    The symbols that are located on the rear panel of the N 550 are as  follows              Data Interface 3     Foot Switch       2     Caution   Do not connect   4     Equipotential Terminal  while power is on  ground           Description of Controls  Indicators  and Symbols    The symbols that are located on the front panel of the N 550 are as  follows         s  9   gt        1     Type BF Applied Part     6     Pulse Search  Not Defibrillator Proof      
3.            seeeeeesse 27  Setting the Alarm Volume    oreet rnt itus 28  Setting Alarm Silence Duration                                  esses 28  DiscussiOn           cccccccececcecaesesssecececeececeusaaeseseeeeeeseaeaanees 28  Proced  tr   i  d RI eoErate Pn Perse 29  Disabling Audible Alarms                             eene 30  DISCUSSION  iiti REIHE EMI 30  Proced  re 5  nne eret iion tiras 30  Verify Patient Settings                         eeeseseseseseeeeeeee 31  Alarm Limits Changed Indicator                              sesssssssss 33  Setting Alarm Limits   cati steterit ta npe Etettdo net eret 34  DISCUSSION           c cccecececcecaesesssecececeeceseuaeeuseseseeeeseaeaagees 34  Procedure    n Lum 34  Setting SatSeconds Duration                             sss 36  DISCUSSIOTnL 2 pco teret eee odo eut Rad 36  Proced  l         tette thee dete Peru 36  Setting the Data Port Baud Rate                       sssssussss 37  DISCUSSION  ii DERI EMI MEN   37  PROCCQUNC EE 37  Setting the Data Port Protocol                      eeesssssssssss 39  Clearing Trend Information                          ssssssseeseeeeeesses 40  N S90 TODO uusderauesdeiteXxs qao FERREA T RCLARS DAVE ECT eens 43  Trend Data Operation                         sesessseeeseeeee 43  Trend  Data   iin erede etie tens 43   1  Trend Pr eio rorem Mae 44   42  Trend Cleat uh E MERE 44    32 Not Used         nn nein 44    4  Baud Rate           ccccccecccecccecsesesseeeeeecescaeseaeaseeeeeeeeeees 44    5  Da
4.        f  5    f 5  9  M        t  e       CJ  4  Press the Adjust Up button or Adjust Down button to select the     desired protocol  Possible settings are     e 1 ASCII    2   External equipment communications  Refer to the external  equipment manuals for the interfacing instructions        Clearing Trend Information    With the N 550 in the normal monitoring mode     1  Simultaneously press and hold the SpO2 Alarm Limit button and    Pulse Rate Alarm Limit button for at least 3 seconds  Option 1  will be displayed           2  Press the Adjust Up button to select Option 2     40    Using the N 550    som       TR 3  Press the SatSeconds Alarm Limit button to clear the trend data     The N 550 emits 3 beeps  indicating that data is cleared     N 550 41    Blank Page    N 550 Trend       Trend Data Operation    From the initial measurement of a patient  trend data  a data point  is  stored in memory every 4 seconds  Up to 50 alarm limit changes can  also be stored in trend data  The N 550 can store up to 24 hours of  trend data     The N 550 trend data will be lost if the coin cell battery fails or is  removed  The coin cell battery is located on the main circuit board     CAUTION  Changing alarm limit settings uses trend memory  space  Change alarm limits only as needed     data  with newly collected data overwriting the oldest data on a  rolling basis  The N 550 continues to record data points as long  as the N 550 is powered on and an initial patient measurement  has been
5.      e If the N 550 is operating on AC power  the fuse may be  blown  Notify service personnel to check and  if necessary   replace the fuse     e If the N 550 is on battery power  the battery fuse may require  replacement  the battery may be missing  or the battery may  be discharged  Charge the battery or notify service personnel  to replace the battery or the battery fuse  as required     One or more display segments or indicators do not light  during the power on self test     e Do not use the N 550  contact qualified service personnel or  your local Nellcor representative     The N 550 does not sound a tone indicating successful  completion of the Power On Self Test  POST      e The N 550 has failed the power on self test  Do not use the  N 550  This tone not only indicates the successful completion  of POST  but it confirms that the audible alarm is functional   Contact qualified service personnel or your local Nellcor  representative     The Pulse Search indicator is lit for more than 10 seconds  while the sensor is connected to the patient     e Check the sensor directions for use to determine if an  appropriate sensor is being used and if it 1s applied properly   Check sensor and sensor cable connections  Test the sensor  on someone else  Try another sensor or sensor cable     79    Troubleshooting    80    e Perfusion may be too low for the N 550 to track the pulse   Check the patient  Test the N 550 on someone else  Change  the sensor site  Try another type of sens
6.    Operation during  vibration    TEC 60068 2 6 and IEC 60068 2 34       Cleaning    IEC 60601 1  sub clause 44 7       Case surface made of  non toxic materials    Case resistant to heat and  fire    IEC 60601 1  sub clause 48    IEC 60601 1  sub clause 59 2 b        N 550 power entry  module fuse holder    IEC 60601 1  sub clause 59 3       N 550 exterior markings    Front panel and case  labeling    IEC 60601 1  sub clause 6 1  6 3  and  6 4  EN 865  clause 6    IEC 60878  EN 980  ISO 7000  EN  60417 1  EN 60417 2       N 550 button spacing    ISO 7250       Year of manufacture  symbol    EN 980       Electromagnetic  Compatibility    IEC 60601 1  sub clause 36  IEC EN  60601 1 2 1993       Radiated and conducted  emissions    EN 55011  Group 1  Class B       Harmonic emissions       IEC 61000 3 2    N 550    Item    Specifications    Compliant With       Voltage fluctuations   flicker emissions    IEC 61000 3 3       Electrostatic discharge  immunity    EN 61000 4 2  level 3 table top  equipment       Radiated radio frequency  electromagnetic field  immunity    IEC 61000 4 3  3 V m   1 kHz       Electrical fast transient   burst immunity    Surge immunity    IEC 61000 4 4   0 5 kV  signal  cable   1 kV  AC mains     IEC 61000 4 5  level 3       Conducted EMI  susceptibility    IEC 61000 4 6  3 Vrms   1 kHz       Power frequency  magnetic fields    IEC 61000 4 8  100 A m   50 Hz       Operation with line  voltage variations       IEC 61000 4 11       107    Blank Page    I
7.   Saturation       Without Motion   Adults     70 to 100   2 digits       Without Motion   Neonate     With Motion   Adults and  Neonates     70 to 100   3 digits    70 to 100   3 digits       Low Perfusion     70 to 100   2 digits             Pulse Rate  Without Motion  20 to 250 bpm  3 digits  With Motion normal physiological range     55   125 bpm   5 digits       Low Perfusion     N 550    20 to 250 bpm  3 digits       97    Specifications    98    Accuracy and Motion Tolerance         Adult specifications are shown for OxiMAx MAX A and MAX N  sensors with the N 550  Neonate specifications are shown for  OxiMAx MAX N sensors with the N 550  Saturation accuracy will  vary by sensor type  Refer to the Sensor Accuracy Grid    2 Applicability  Ox  Max MAX A  MAX AL  MAX P  MAX I  and  MAX N sensors     x Specification applies to N 550 performance        Table 6  Tone Definition                Audible Indicator Parameter Value  Alarm Silence Volume level Not changeable  Reminder   Pitch   30 Hz  784 Hz  Pulse width   20 msec  150 msec  Pulse repetition interval 150 msec      20 msec   double burst                 Repetitions 3  Confirmation of Volume level Not changeable  Button Pressed   Pitch   30 Hz  784 Hz   Pulse width   20 msec  150 msec   Pulse repetition interval 150 msec      20 msec   double burst              Repetitions 3    N 550    Audible Indicator    Table 6  Tone Definition    Parameter    Specifications    Value       High Priority Alarm    Volume level    Ad
8.   limits  If no data for a parameter is available  three dashes         are  displayed     PA represents pulse amplitude  The number can range from 0 to 254   There are no alarm parameters for this value  It can be used for  trending information as an indication of a change in pulse volume   relative pulse strength  or circulation     N 550 Version 1 00 00 CRC  XXXX SpO2 Limit  85 100  PR Limit  40 170 bpm  ADULT 0 SAT S    TIME 9 6SpO2 PR  bpm  PA Status  30 Sep 03 16 00 00 100 190  220 PH       The Status column indicates alarm conditions and operating status of  the N 550  In this example  Pulse High  PH  means that the pulse rate  upper alarm limit has been exceeded  The status codes are listed  below  As many as four codes can be displayed at one time in the  Status column     Code Meaning       AO Alarm Off    63    Printing    64    Code    Meaning       AS    Alarm Silence       LB    Low Battery       LM    Loss of Pulse w Motion       LP    Loss of Pulse       MO    Patient MOtion       PH    Pulse Rate Upper Limit Alarm       PL    Pulse Rate Lower Limit Alarm       PS    Pulse Search       SH    Saturation Upper Limit Alarm       SL    Saturation Lower Limit Alarm       SD    Sensor Disconnect       SO       Sensor Off                to be displayed in the patient data section of the printout     9 Note  A sensor disconnect and sensor off will also cause three dashes    Sensors and Accessories    Sensors and Accessories       Selecting a Sensor    N 550    A     g
9.   vi    Table 1   Table 2     Table 3   Table 4   Table 5   Table 6     Data Port Pinouts             ccccccescsssseseseeeeseseeeseeeeeeeees 48  Nellcor Oximetry Sensor Models   and Patient Weights                            eeeeeeeees 66  Operator s Menu 1ssioscecus cen tonc Sex PEE cv ooo a dpa ERO ERE 75  Error Codes  cette ERR IER RES 78  Factory Default Settings                                sssssses 88  Tone Definition sirsiran 98    Safety Information       Warnings    A  A    A    N 550    A    Warnings are identified by the WARNING symbol shown above     Warnings alert the user to potential serious outcomes  death  injury  or  adverse events  to the patient or user     WARNING  Explosion hazard  Do not use the N 550 pulse  oximeter in the presence of flammable anesthetics or gases     WARNING  Pulse oximetry readings and pulse signals can be  affected by certain ambient environmental conditions  sensor  application errors  and certain patient conditions  See the  appropriate sections of the manual for specific safety  information     WARNING  Ensure that the speaker is clear of any obstruction   Failure to do so could result in an inaudible alarm tone     Safety Information       Cautions              e    Cautions are identified by the CAUTION symbol shown above     Cautions alert the user to exercise care necessary for the safe and  effective use of the N 550     CAUTION  When connecting the N 550 to any instrument   verify proper operation before clinical use  Bo
10.  2     Pulse Rate 7     Data In Sensor       3     Motion 8     Low Battery       4     Sensor Off 9     AC Power Battery Charge       5     Sensor Message          Description of Controls       result in either a valid or an invalid button tone  If the button  pressed fails to emit a tone  contact qualified service personnel     The Power On Off button is used to turn the N 550 on or off     9 Note  Each button press  except the Power On Off button  should    N 550    Description of Controls  Indicators  and Symbols    10        IN    2    is    The Alarm Silence button is used to silence current alarms for the  alarm silence duration period  When an alarm has been silenced   pressing the button again reactivates  or    unsilences    the alarm  It is  also used to view and adjust alarm silence duration and alarm volume     The Adjust Up button is used to increase alarm limit values  alarm  silence duration  pulse beep volume  alarm volume  and data port  baud rate  The Adjust Up button is used to select the communication  protocol and time settings  Press the Adjust Up button one time to  increase the display by one digit  Holding the Adjust Up button down  for more than one second cause the display digits to scroll     The Adjust Down button is used to decrease alarm limit values  alarm  silence duration  pulse beep volume  alarm volume  and data port  baud rate  The Adjust Down button is used to select the  communication protocol and time settings  Press the Adjust Down  b
11.  30 Hz  296 Hz to 662 Hz   varies with  saturation    Pulse width   20 msec  40 msec   Pulse repetition interval N A    msec  20 msec   double   burst    Repetitions 1          100       Electrical    N 550    Table 6  Tone Definition    Specifications             Audible Indicator Parameter Value  Valid Button Press Volume level Not changeable  Pitch   30 Hz  784 Hz  Pulse width   20 msec  30 msec       Pulse repetition interval    20 msec   double burst     N A       Repetitions    1       Volume Setting    Volume level    Adjustable alarm  volume                         Pitch   30 Hz  752 Hz   Pulse width   20 msec  Infinite   Pulse repetition interval N A     20 msec   double burst    Repetitions N A  Instrument             Power 100 to 240 volts AC  18 to 26 volt amps to  Requirements be compliant with IEC 60601 1 sub clause  10 2 2  Fuses qty 2  2 A  250 volts  slow blow  IEC  5 x  20 mm   Battery       The battery provides at least 1 5 to 2 hours of battery life when  new and fully charged with no alarms  no serial data  and no nurse  call outputs  while using a pulse simulator set for 200 bpm  high  light and low modulation        Type    Lead acid    101    Specifications          Battery  Voltage 12 Volts DC  1 2 AH  Recharge e 11 hours with N 550 turned off    e 12 hours with N 550 operating       Shelf Life e 2 months  new  fully charged battery    e After 2 months storage the N 550 will  run for 50  of stated battery life          Sensors    Wavelength The wavele
12.  Do not make any clinical judgments based on the  oximeter s measurements only     Intended Use for the N 550    N 550    The N 550 Pulse Oximeter is indicated for the continuous  noninvasive monitoring of functional oxygen saturation of arterial  hemoglobin  SpO2  and pulse rate  The N 550 is intended for use  with neonatal  pediatric  and adult patients during both no motion and  motion conditions and for patients who are well or poorly perfused  in  hospitals  hospital type facilities  intra hospital transport  and home  environments  For prescription use only     Note  Hospital use typically covers such areas as general care floors   operating rooms  special procedure areas  intensive and critical  care areas  within the hospital plus hospital type facilities   Hospital type facilities include physician office based  facilities  sleep labs  skilled nursing facilities  surgicenters  and  sub acute centers     Intra hospital transport includes transport of a patient within  the hospital or hospital type facility     Use with any particular patient requires the selection of an  appropriate oxygen transducer  sensor  as described in this Operator s  Manual     Motion performance claims are applicable to models MAX A   MAX AL  MAX P  MAX N  and MAX I Nellcor OxiMax     oximetry sensors     Blank Page    Description of Controls  Indicators     and Symbols       Identification of Front Panel Buttons and Symbols    N 550          N  t             san  aes  if     I  eo  Co   Co  
13.  N 550 will  produce no audible alarms     Visual indications of an alarm condition are not affected by disabling  the audible alarms     The ability to set the alarm silence duration to OFF can be enabled or  disabled by qualified service personnel as described in the service  manual     WARNING  Do not silence the audible alarm or decrease its  volume if patient safety could be compromised        Procedure    With the N 550 in the normal monitoring mode     30    Using the N 550    E 1  Press the Alarm Silence button until XX SEC is displayed                  2st890    2  While pressing the Alarm Silence button  press and hold the     Adjust Up button until OFF is displayed  Release buttons     m       gos  Rl       3  The Alarm Silence indicator will flash  indicating that the alarm  sounds are disabled  The N 550 will sound three beeps    approximately every three minutes to warn the user that the alarm  sound has been silenced        Verify Patient Settings    With the N 550 in the normal monitoring mode     1  Press the SpO2 Alarm Limit button to view the current  SpOz2  upper alarm limit     N 550 31    Using the N 550     2l  mt            f   gt     a     v          oy       2  Press the SpO2 Alarm Limit button twice to view the current   SpO2 lower alarm limit           2l   i     T  9I    I      990    3  Press the Pulse Rate Alarm Limit button to view the current Pulse  Rate upper alarm limit      2l   i            g   gt     f                    ooo              
14.  TXD   RS 422 negative output        13 NC  14 NC             15 Nurse Call Common       The pin layouts  as viewed from the rear panel of the N 550  are  illustrated in Figure 3  The conductive shell is connected to earth  ground when connected to a PC or printer         9 10 11 12 13 14 15  23 45 67 8           1       Figure 3  Data Port Pin Layout  Pins 2  3  5  and 11 provide data in RS 232 format   Pins 1  4  9  and 12 provide data in RS 422 format  TXD  and TXD     are the differential transmit data pair  RXD  and RXD  are the  differential receive pair     Use the Data Port Setup procedure to configure the N 550 data port  baud rate and protocol     49    Using the Data Port    Procedure    With the N 550 in the normal operating mode     1  Simultaneously press and hold the SpOz Alarm Limit button and    Pulse Rate Alarm Limit button for at least 3 seconds  Option 1  will be displayed     in  e      j          Just          2  Press the Adjust Up button until Option 4 is displayed                  na   i                     Hj  E        stS900        3  Press the SatSeconds Alarm Limit Button to select Option 4  The  selected baud rate will be displayed     50    Using the Data Port    pa  dt       t    Seo f         Note  The available selections are     e 24   2400 baud rate  e 96   9600 baud rate  e 192   19200 baud rate    4  Press the Adjust Up button or the Adjust Down button to select  the desired baud rate     rate     TR 5  Press the SatSeconds Alarm Limit bu
15.  as carboxyhemoglobin or methemoglobin  In  contrast  hemoximeters such as the IL482 report fractional saturation     oxygenated hemoglobin expressed as a percentage of all measured  hemoglobin  including measured dysfunctional hemoglobins  To  compare functional saturation measurements to those from an  instrument that measures fractional saturation  fractional  measurements must be converted as follows     fractional saturation    functional saturation          1hHJmMHH x1  100   96 carboxyhemoglobin     methemoglobin        Measured versus Calculated Saturation    When saturation is calculated from a blood gas partial pressure of  oxygen  PO2   the calculated value may differ from the SpO2  measurement of a pulse oximeter  This usually occurs because the  calculated saturation was not appropriately corrected for the effects of  variables that shift the relationship between PO2 and pH   temperature  the partial pressure of carbon dioxide  PCO2   2 3 DPG   and fetal hemoglobin  See Figure 7 on page 96     N 550 95    Principles of Operation    96    Saturation        100    50          t pH     Temperature    PCO2  7    2 3 DPG m  t Fetal Hb di         pH   t Temperature  t PCO2   t 2 3 DPG    50 100  PO   mmHg     Figure 7  Oxyhemoglobin Dissociation Curve    Specifications       Performance    Measurement Range       SpO2    Pulse Rate    1  to 100     0 and 20 beats per minute  bpm  to  250 bpm          Perfusion Range    0 03  to 20        Accuracy and Motion Tolerance  
16.  made  with    blank    data points collected if no sensor  is connected to the N 550 or patient     Blank    data will  overwrite older patient data if the memory becomes full   Therefore  if you want to save old patient data  it is important  that you turn your N 550 off when you are not monitoring a  patient  and that you download the trend memory before it fills  up and overwrites the old data with new data  or    blank    data      9 Note  Trend memory always contains the most recent 24 hours of       Trend Data    Trend data information may be retrieved or cleared through the N 550  data port using options available in a display menu     N 550 43    N 550 Trend    To access the menu options  simultaneously press the SpO2 Alarm    Limit and Pulse Rate Alarm Limit buttons until Option 1 appears on  the display        er con    25558    4  E           L           Then  using the Adjust Up button and Adjust Down button  you may  G2 scroll through the available menu options as follows         1  Trend Print    on a PC using the Hyper Terminal program  You must press the  SatSeconds Alarm Limit button to initiate printing  Refer to Trend  Data Printout on page 59     Le Allows printing of the existing trend data  The output may be viewed     2  Trend Clear    Clears the existing trend data  To activate  press the SatSeconds  Alarm Limit button         3  Not Used        4  Baud Rate    Allows the interconnection to various printers  Refer to Setting the  Data Port Baud Rate o
17.  sensor  single pateint use  adult 52 5 mW       OXIMAX adhesive sensor  single pateint use  adult  52 5 mW    longer cable 36 inches  91 44 cm        OXIMAX adhesive sensor  single pateint use  neonatal  52 5 mW    adult       OXIMAx adhesive sensor  single pateint use  pediatric 52 5 mW       OXIMAX adhesive sensor  single pateint use  infant 52 5 mW       103    Specifications       Ox1Max Sensor Power Dissipation                                     Physical Characteristics    104    OxiMAx Sensor Dissipation  OXIMAX adhesive sensor  single pateint use  adult nasal   52 5 mW  OXIMAX OxiCliq   nonadhesive sensor  single patient  52 5 mW  use  adult  reusable cable   OXIMAX OxiCliq nonadhesive sensor  single patient  52 5 mW  use  neonatal adult  reusable cable   OXIMAX OxiCliq nonadhesive sensor  single patient  52 5 mW  use  pediatric  reusable cable   OXIMAX OxiCliq nonadhesive sensor  single patient  52 5 mW  use  infant  reusable cable   OXIMAX Durasensor   finger clip sensor  reusable  adult 52 5 mW  OxIMAX Oxiband   sensor  reusable  neonatal adult 52 5 mW  OXIMAX Oxiband sensor  reusable  pediatric infant 52 5 mW  OXIMAX Dura Y   multisite sensor  reusable 52 5 mW  Weight 3 07 Ibs   1 39 kg  without pole mount       Dimensions 2 87 in  x 7 87 in  x 5 in   7 3 cm x 20 cm x    12 7cm              Compliance    N 550    Item    Specifications    Compliant With       Equipment classification    Safety Standards  IEC 60601 1  same  as EN60601 1   EN475  EN865  EN   IEC 60
18.  shorter     It is recommended that qualified service personnel replace the  internal battery every 24 months     CAUTION  If the N 550 is to be stored for 2 months or longer   notify service personnel to remove the battery from the N 550  prior to storage  Recharge the battery when it has not been  charged for 2 or more months     The Low Battery Indicator flashes and a high priority alarm begins to  sound when the battery reaches the lowest battery voltage at which an  N 550 can support normal operation  This alarm cannot be silenced  while running on battery power  If the N 550 is not connected to AC  power  the N 550 will shut off after 10 seconds     Using the N 550       Turning on the N 550       Discussion       Procedure    N 550    96    Before using the N 550 in a clinical setting  you must verify that the  N 550 is working properly and is safe to use  Proper working  condition can be verified by successful completion of the Power On  Self Test  POST   described in the following section     WARNING  The N 550 should not be used adjacent to or stacked  with other equipment  If adjacent or stacked use is necessary  the  N 550 should be observed to verify normal operation in the  configuration in which it is used     CAUTION  If any indicator or display element does not light  or  the speaker does not sound  do not use the N 550  Instead  contact  qualified service personnel  your local Nellcor representative  or   Nellcor s Technical Services Department  1 800 635 52
19.  silence  duration  the alarm silence duration is ended and the audible alarms  are re enabled     Visual indications of an alarm condition cannot be turned off  For  example  if the  SpOz2 lower alarm limit is exceeded  the alarm can  be silenced for the alarm silence duration  but the  SpO2 value will  continue to flash     If the alarm condition is still present when the alarm silence duration  has elapsed  the alarm will sound     The power on default setting for audible alarm silence duration is set  at the factory to 60 seconds  The default setting can be adjusted by  service personnel as described in the N 550 service manual        Procedure    With the N 550 in the normal monitoring mode       1  Press the Alarm Silence button until XX SEC is displayed   A Durations are  OFF  30  60  90  and 120 seconds          Note  For steps 2 and 3  the Adjust Up and Adjust Down buttons  RE must be pressed while pressing the Alarm Silence button     N 550 29    Using the N 550    2  While pressing the Alarm Silence button  press and hold the  Adjust Up button to increase alarm silence duration to Off  30   60  90  or 120 seconds     3  While pressing the Alarm Silence button  press and hold the  Adjust Down button to decrease alarm silence duration to OFF   30  60  90  or 120 seconds        Note  Releasing the Adjust Up or Adjust Down button sets the alarm  silence duration        Disabling Audible Alarms       Discussion    Setting the alarm silence duration to OFF means that the
20.  solution of 10  chlorine bleach in tap water     Before attempting to clean an SpO  sensor  read the directions for use  enclosed with the sensor  Each sensor model has cleaning instructions  specific to that sensor     86    Technical Information       Description of Alarms    The N 550 has three levels of audible alarms     1  High priority alarm  A high pitched  fast pulsating tone  indicating loss of pulse with no patient motion     2  Medium priority alarm  A medium pitched  normal pulsating  tone indicating an SpOz or pulse rate limit violation     During a medium priority alarm  the green display turns red and  flashes with the patient parameter that violated the limit   SpOz or  Pulse Rate   If the alarm is a SatSeconds alarm  the SatSeconds  indicator  clock  will flash     3  Low priority alarm  A low pitched  slow pulsating tone  indicating a sensor disconnect  low battery  or N 550 failure     During a low battery alarm  the Low Battery indicator illuminates  and the alarm tone sounds immediately  even if the alarms are  silenced or set to OFF     N 550 87    Technical Information       Factory Defaults    88    The N 550 is shipped with factory default settings  Authorized  technical personnel using the procedures described in the N 550  service manual can change default settings     Table 5  Factory Default Settings                            Factory  Parameter Range Default   Setting   SpO2 Upper Alarm Lower Alarm Limit 100   Limit plus 1 to 100    SpO2 Lower 
21.  the rear of the N 550     Pole mount  bracket       70    Performance Considerations    WARNING  Pulse oximetry readings and pulse signals can be   A affected by certain ambient environmental conditions  sensor  application errors  and certain patient conditions  See the  appropriate sections of the manual for specific safety  information        Performance Verification    The performance of the N 550 can be verified by following the  procedures outlined in the Performance Verification section of the  N 550 service manual  Qualified service personnel should perform  these procedures before using the N 550 for the first time in a clinical  setting        N 550 Performance Considerations    N 550    Certain patient conditions can affect the measurements of the N 550  and cause the loss of the pulse signal     Inaccurate measurements can be caused by     e prolonged patient movement    venous pulsations    e intravascular dyes  such as indocyanine green or methylene  blue      defibrillation    71    Performance Considerations    Dysfunctional Hemoglobins       Anemia       Saturation    Pulse rates       Dysfunctional hemoglobins such as carboxyhemoglobin   methemoglobin  and sulphemoglobin are unable to carry oxygen   SpO2 readings may appear normal  however  a patient may be  hypoxic because less hemoglobin is available to carry oxygen   Further assessment beyond pulse oximetry is recommended     Anemia causes decreased arterial oxygen content  Although SpO2  readings may ap
22.  your local Nellcor representative for technical assistance  Write  down the numbers and have them available prior to requesting  technical assistance     6  Ifthe N 550 detects an internal problem during the self test  an  alarm tone sounds and the N 550 displays an Error Code and the  corresponding number  See Troubleshooting on page 77     N 550 23    Using the N 550    Se K       7  Upon successful completion of the POST  the N 550 sounds a  one second tone indicating that the N 550 has passed the test     WARNING  If you do not hear the POST pass tone  do not use  the N 550     WARNING  Ensure that the speaker is clear of any obstruction   Failure to do so could result in an inaudible alarm tone     9 Note  In addition to serving as the POST pass verification  the POST    pass tone also functions as an audible confirmation that the  speaker is performing properly  If the speaker does not  function  the alarm sounds cannot be heard        Sensor Attached    The Pulse Search indicator and the Sensor Off indicator light and the    N 550 displays dashes in the  SpOz and Pulse Rate displays while  the N 550 is searching for a valid pulse     24    N 550    Using the N 550       1    Sensor Off   2     Pulse Search Indicator       When a valid pulse is detected  the N 550 enters the Monitoring  Mode and displays patient parameters  Look for movement of the blip  bar and  if the pulse beep volume is not 0  listen for pulse beeps to  verify that displayed measurement values are 
23. 0 bpm  ADULT 0 SAT S   TIME 96SpO2 PA Status   30 Sep 03 18 24 24         SD   30 Sep 03 18 24 26         SD   30 Sep 03 18 24 28 98   30 Sep 03 18 24 30 98   30 Sep 03 18 24 32 99       Figure 4  Real Time Data Printout       Trend Data Printout    The format of data displayed when a trend printout is requested is the  same as the real time data  The only differences are that    TREND    is  displayed in the top row instead of the    CRC  XXXX  software  verification number and there is no  Status  column     Readings are displayed in 4 second intervals  The values on each row  are an average for the 4 second period     At the end of the printout  an    Output Complete  line indicates that  the transmission was successful  If the    Output Complete  line is not  present  a corruption of the data may have been detected and the data  should be ignored     N 550 59    Printing    N 550 VERSION 1 00 00 TREND SpO2 Limit  85 100  PR Limit 40 170 bpm  ADULT 0 SAT S   TIME  SpO2 BPM PA   10 Nov 03 21 26 53 98 53   10 Nov 03 21 26 57 98 53    10 Nov 03 21 30 01 98 110 53   N 550 VERSION 1 00 00 TREND SpO2 Limit  80 100  PR Limit 40 170 bpm  ADULT 0 SAT S   TIME  SpO2 BPM PA    10 Nov 03 21 30 05 98 53  10 Nov 03 21 30 09 95 51  10 Nov 03 21 30 13 93 50  10 Nov 03 21 30 17 91 49  10 Nov 03 21 30 21 85 47  10 Nov 03 21 30 25 80 42  10 Nov 03 21 30 29 78   41  10 Nov 03 21 30 33 78   41  Output Complete       Figure 5  Trend Data Printout    Once a trend printout has begun  the N 550 mu
24. 550     2  Connect an SpO2 Sensor to the other end of the DOC 10 pulse  oximetry cable     3  Place the SpOz Sensor on the patient or yourself     Turn on the N 550     wn    Press and hold the Adjust Up or Adjust Down button to increase  or decrease the pulse beep tone volume     S    N 550 27    Using the N 550       Setting the Alarm Volume    With the N 550 in the normal monitoring mode             Press and hold the Alarm Silence button until the alarm volume  ANY display appears     pe  mt    Sea  Ki           2  While continuing to press the Alarm Silence button  press the  Adjust Up button or Adjust Down button to increase or decrease    GO the alarm volume        Setting Alarm Silence Duration    Discussion    WARNING  Do not silence an audible alarm or decrease its  volume if patient safety could be compromised     Alarms can be silenced for a preset period called alarm silence  duration  To view the current setting  press and hold the Alarm  Silence button until 30 SEC  60 SEC  120 SEC  or OFF is displayed   To adjust the settings  press and hold the Alarm Silence button until  30 SEC  60 SEC  120 SEC  or OFF is displayed  Continue holding the    28    Using the N 550    Alarm Silence button  and use the Adjust Up button or Adjust Down  button to increase or decrease the value  Possible values are 30  60   90  or 120 seconds  or OFF  The OFF selection is discussed under  Disabling Audible Alarms on page 30     If the Alarm Silence button is pressed during the alarm
25. 601 1 2 1993  CAN CAS  C22 2 No  601 1  UL 2601 1  ISO  10993 1       Type of protection    Class I  on AC power   Internally powered  on battery  power        Degree of protection    Type BF   Applied part       Mode of operation    Continuous       N 550 resistant to liquid  ingress    IEC 60601 1  sub clause 44 6 for  class IPX1 Drip Proof equipment       Degree of Safety in  presence of a flammable  anaesthetic    IEC 60601 1  Sub clause 37 5  Not  suitable       Applied sensor label to  indicate Type BF applied  part    IEC 60601 1 Symbol 2 of Table DII  of Appendix D       Equipotential lug symbol  to indicate a potential  equalization conductor    IEC 60601 1 Symbol 9 of Table DI of  Appendix D       Attention symbol  consult  accompanying  documentation    IEC 60601 1 Symbols 14 of Table DI  of Appendix D       External case made with  non conductive plastic    IEC 60601 1  sub clause 16 a        No holes in case top    IEC 60601 1  sub clause 16 b        Rigid case    IEC 60601 1  sub clause 21 a        Case mechanically strong       IEC 60601 1  sub clause 21 b     105    Specifications    106    Item    Compliant With       Case handle    IEC 60601 1  sub clause 21 c        N 550 resistant to rough  handling    IEC 60601 1  sub clause 21 6       N 550 resistant to liquid  ingress due to spills    Environmental    IEC 60601 1  sub clause 44 3 as  modified by EN 865  clause 4    IEC 60601 1  sub clause 44 5       Operation during physical  shock    IEC 60068 2 27    
26. 67     Note  The N 550 should complete the POST function within 12  seconds     1  Turn on the N 550 by pressing and holding the Power On Off  button for more than one second     21    Using the N 550    2  The N 550 displays sounds              Display Sound  5  in pulse rate left window  one beep tone  5  in pulse rate center window  one beep tone  0  in pulse rate right window  one beep tone  n  in SpO2 left window  none  n  in SpO2 center window  none             3  The N 550 automatically starts the Power On Self Test  POST    which tests N 550 circuitry and functions     CAUTION  During POST  immediately after power up   confirm  that all display segments and indicators light  and the speaker    sounds a 1 second pass tone     4  While performing POST  the self test display appears for  approximately 2 to 4 seconds  During this time     e All indicators illuminate  e All segments of all numeric digits light and change from red  to green    22    Using the N 550    e All segments of the Pulse Amplitude Display light      All segments of the SatSeconds indicator light       5  Once the display test portion of POST is complete  the N 550  software version is displayed for approximately 2 seconds     Sco   Kt          Note  The device version illustrated is only a sample  The device    version identifies the hardware configuration and the software  revision     Device version numbers are often needed when calling  Nellcor s Technical Services Department  1 800 635 5267  or 
27. Accessories e ien ett ei tii abuse et iria eap crate 68  germen                          69  Visual Alarm Indicator                               eseseeeeeseees  69  Pole Mount Bracket                ccccccsseeeeeseeeseeeseeenseesenensenes 70   Performance Considerations                          creer 71   Performance Verification              cccceceescccecesseeeetenseseersneeseeteaes 71   N 550 Performance Considerations                                 es 71  Dysfunctional Hemoglobins                               essssssssss 72  Anemia See a a NUMEN TII 72  SalbllatiOni enrera cere rdiet sert er ea do 72  Pulse rates                 sssesssseeeeseenereemneneennennnen nnn 72   Sensor Performance Considerations                                      72    N 550 iii    Contents    Operator S IIO 4 06 uade we ae rece eee RA dU Ra aon X V dr RE GA ean 75  TIOUDIBE BODIE uakaxadetnhauamendaERRR ER ER REqESAaE RE E Rad 77  Error GOd S   oet i tree ten iG erbe 77  Corrective ACIIOn      itte rete rese kel pP a bild Ide 78  EMI  Electro magnetic Interference                                        81  Obtaining Technical Assistance                        eseessssssss 82  PONCE ou uadit 3299 3 8 808 2 3 0 ardeo o C DR CRCRCRC RC E EU aen aO RR 85  Returning the N 550 Liu ea uode asit inn etri ere Ere sid ruere des 85  SEVICE ecd tee chere ie rec detente 85  Periodic Safety Checks             ccccsscseeceeeeeeeseneeeeeeeessessessneeeeees 86  CHI RP r              86  Technical Informall
28. Alarm   20  to Upper Alarm   85   Limit Limit minus 1  Pulse Rate Upper Lower Alarm Limit 170 bpm  Alarm Limit plus 1 to 250 bpm  Pulse Rate Lower 30 bpm to Upper 40 bpm  Alarm Limit Alarm Limit minus 1  Alarm Silence On or Off On  Reminder  Alarm Silence Off  30  60  90  120 60 seconds  Duration seconds  Alarm Silence e Audible reminder   Audible  Restrictions i reminder   e No audible  reminder  e Donot allow  alarms off   Alarm Sound Selector   1  2 3 1  Alarm Volume 1 to 10 4  Data Port Baud Rate 2400  9600  19200 19200  Data Port Protocol 1 2 1  ASCID          Technical Information    Table 5  Factory Default Settings                               Factory  Parameter Range Default  Setting  Display 0 1 0  On   In Sensor Trend Mode   0 1 2 0  Event SpO2   Language English English  Pulse Beep Volume 0 to 10 4  RS 232 Nurse Call Normally high  Normally low  Priority normally low  SatSeconds Off  10  25  50  100 Off  Silence Alarms 0  1 1  Off   SatSeconds Display    When the N 550 SatSeconds technology detects an SpO2 value  outside the alarm limit  the SatSeconds indicator LEDs begin to light   fill  clockwise  When the SpO2 value is within the set limits  the  SatSeconds indicator LEDs will extinguish counterclockwise     When all of the SatSecond indicator LEDs are lit  indicating that the  SatSeconds setting has been reached  an audible alarm sounds and the  SatSeconds indicator LEDs flash  As with traditional alarm  management  the audible alarm may be silenced by press
29. Net  is for patients whose saturation levels  frequently go below the limit but do not stay below the limit long  enough for the SatSeconds setting to be reached  When 3 or more  limit violations occur in 60 seconds  an alarm will sound even if the  SatSeconds setting has not been reached     Principles of Operation       Oximetry Overview    N 550    The N 550 uses pulse oximetry to measure functional oxygen  saturation in the blood  Pulse oximetry works by applying a sensor to  a pulsating arteriolar vascular bed  such as a finger or toe  The sensor  contains a dual light source and a photo detector     Bone  tissue  pigmentation  and venous vessels normally absorb a  constant amount of light over time  The arteriolar bed normally  pulsates and absorbs variable amounts of light during the pulsations   The ratio of light absorbed is translated into a measurement of  functional oxygen saturation  SpO2      Because a measurement of SpO2 is dependent upon light from the  sensor  excessive ambient light can interfere with this measurement     Specific information about ambient conditions  sensor application   and patient conditions is contained throughout this manual     Pulse oximetry is based on two principles  that oxyhemoglobin and  deoxyhemoglobin differ in their absorption of red and infrared light   spectrophotometry   and that the volume of arterial blood in tissue   and hence  light absorption by that blood  changes during the pulse   plethysmography   A pulse oximeter 
30. Operator s Manual        D      09   E   gt  lt   O  oD   w  5  am                 00N RENES         This ISM device complies with Canadian ICES 001   Cet appareil ISM est conforme    la norme NMB 001 Canada     Nellcor Puritan Bennett Inc  is an affiliate of Tyco Healthcare  Nellcor  Oxiband  Durasensor   OxiCliq  OxiBand  Dura Y  MAX FAST  SatSeconds  PediCheck  Oxismart and OxiMax are  trademarks of Nellcor Puritan Bennett Inc     To obtain information about a warranty  if any  contact Nellcor s Technical Services Department   or your local representative     Purchase of this instrument confers no express or implied license under any Nellcor Puritan  Bennett patent to use the instrument with any sensor that is not manufactured or licensed by  Nellcor Puritan Bennett     Covered by one or more of the following U S  Patents and foreign equivalents  4 621 643   4 653 498  4 700 708  4 770 179  Re  35 122  4 802 486  4 869 254  4 928 692  4 934 372  5 078 136   5 351 685  5 368 026  5 485 847  5 533 507  5 662 106  5 853 364  and 6 083 172     Contents    COMENS ee ee ee er ene aE eee ee ree ee eT pee mee re en i  FR  MES ons ck heer dGane SNE EE E d dd WE RE RAE A Rd E d ES RR d VR wew es V  Tables ous daa d rr vi  Salely Informati  n cides ctu ra daAWARARRARRRRGRRERREEPAAE AE ERAT 1  Warnings  pieren n io i tos dedii pastu diateedt cuius 1    BUTS ie uidi Pon sab odoris medo aud ced coches that ach Donius 2   Plc EEUU EET 3   EDGE BON 3a Xdcedcari c Es dad AA A Ide ae dowd E
31. UER Crees wears 5  Intended Use for the N 550           nece hens eene 5   Description of Controls  Indicators  and Symbols                    7  Identification of Front Panel Buttons and Symbols                    7   Identification of Rear Panel Components                                  8   N 550 Symbols t                 8   Description of Controls                     seseseen 9   Description of Displays and Indicators                                    11   Description of Audible Indicators                               sessesssss 13   bU UR Ihe PB sid ka d A Ses RERO RAO OC ACAG  ROROR OR RR d e ded 15  List of Components  1  tn rx eth ii secon X FEE en oS 16   Connecting the N 550 to AC Power                     ssesseeees 17   Connecting a Sensor to the N 550                       essssss 17   Hallory DDSPAUOR iua nau wach ce cede d o RR ERROR OCA Urn e RD d RR 19  Operating the N 550 on Battery Power                             ssss 19   Low Battery Indicator 3  redis ker ento heit tege ipd c saeua unes 20   sind Th   N OOU  assured arie x RR ER ES dE CR n RrRR WARE es ce a ur CR wens 21  Turning on the N 550 irure iiii Eno teii oet Sc Rei oiu ans 21   Discussion                ssssssse mnes 21   Procedure iei iae xy x hia ed a en ava eA en eas 21   Sensor Attached 1st iter ebbe tin tides 24   No Sensor Attached                             sssseessssse 26    Sensor Message   2 eut co Ga ELI Deum 26    N 550      Contents    Setting the Pulse Beep Volume             
32. art number 1 747579 2   and compatible  pins  AMP part number 66570 2   The cable should be no more than  25 feet  7 6 meters  in length  The external ITE  Information  Technology Equipment  device must be certified to UL 1950 or  IEC 60950     47    Using the Data Port       The cable used must have a braided shield providing 100  coverage   such as a Belden cable  Belden part number 9609  or equivalent  The  shield must have a 360 degree connection to the metal shell on the  N 550 s DB 15 connector and to the connector on the PC or serial  printer  Do not create sharp bends in the cable  as this may tear or  break the shielding     No hardware flow control is used  However  in the ASCII mode   XON XOFF flow control is supported     Data Port Pinouts    48    TXD represents the Transmit Data line  and RXD is the Receive Data    line     The pinouts for the data port are listed in Table 1     Table 1  Data Port Pinouts                         Pin Signal Name  1 RXD   RS 422 positive input   2 RXD  RS 232 input   3 TXD  RS 232 output   4 TXD   RS 422 positive output    5 Signal Ground  isolated from Earth Ground   6 NC  No connection   7 Normally Open  relay closure nurse call  normally open   8 Normally Closed  relay closure nurse call  normally  closed   9 RXD   RS 422 negative output   10 Signal Ground  isolated from earth ground   11 Nurse Call  RS 232 level output           Data Port Setup       Using the Data Port    Table 1  Data Port Pinouts    Pin Signal Name       12
33. ator and the pulse search indicator are simultaneously lit  it is an  indication that the artifact is significant and or has been persistent     The Pulse Search Indicator  Lights continuously prior to initial  acquisition of a pulse signal and during prolonged and challenging  monitoring conditions  It flashes during a loss of pulse signal     The Sensor Off Indicator is lit when either the sensor is invalid  or no  longer on the patient s finger  It usually indicates that the sensor is not  on the patient     The Sensor Message Indicator  Lights when the N 550 cannot  determine an SpO2 level or a pulse rate  The Sensor Message  recommendations for improving the signal are      Reposition sensor   e Check or change adhesive wrap     Choose alternate site   e Warm site   e Cover sensor   e Use forehead  nasal  or ear sensor  adult patients only    e Use OxiMax adhesive sensor   e Secure cable   e Secure with headband  Max Fas7    e Remove nail polish   e Loosen sensor  too tight     e Isolate external interference  electrosurgical device  cell  phone     e Clean site  MAX R        Description of Controls  Indicators  and Symbols    The Data In Sensor Indicator  The indicator blinks for approximately  one minute when initially connected to the N 550 to indicate that the  attached OxiMAx sensor contains a patient sensor event record  The  indicator lights continuously to indicate that the attached sensor  memory is full  The indicator does not light when there is no data in  th
34. ause of the proliferation of radio frequency transmitting  equipment and other sources of electrical noise in health care  environments  for example  electrosurgical units  cellular phones   mobile two way radios  electrical appliances  and high definition  television   it is possible that high levels of such interference due to  close proximity or strength of a source may result in disruption of  performance of this device     81    Troubleshooting       The N 550 is designed for use in environments in which the pulse can  be obscured by electromagnetic interference  During such  interference  measurements may seem inappropriate or the N 550  may not seem to operate correctly     Disruption may be evidenced by erratic readings  cessation of  operation  or other incorrect functioning  If this occurs  the site of use  should be surveyed to determine the source of this disruption  and the  following actions taken to eliminate the source     e Turn equipment in the vicinity off and on to isolate the  offending equipment    e  Reorient or relocate the interfering equipment      Increase the separation between the interfering equipment    and this equipment     The N 550 generates  uses  and can radiate radio frequency energy  and  if not installed and used in accordance with these instructions   may cause harmful interference with other devices in the vicinity     If assistance is required  contact Nellcor s Technical Services  Department  1 800 635 5267  or your local Nellcor re
35. ble for use at any time     The N 550 cannot operate with a dead battery  even when plugged  in   Before attempting to turn on an N 550 with a depleted battery   first plug the N 550 into an AC outlet to allow the battery to charge  for a few minutes  The N 550 may then be powered on     To charge a low or dead battery  connect the N 550 to AC power  A  full charge of a dead battery takes 11 hours while the N 550 is turned  off or 12 hours while the N 550 is on     When all of the following conditions are present for 15 minutes  the  N 550 will automatically shut down    e N 550is running on battery power   e No buttons have been pressed    e No pulse has been detected  for example  when no patient is  connected to the sensor or the sensor is disconnected     19    Battery Operation       e No alarms are present  other than low battery or a  non correctable error     Low Battery Indicator    20    e E The Low Battery Indicator lights and a low priority alarm begins to    o        e c    sound when 15 minutes but not more than 20 minutes of monitoring  time remain on the existing battery charge  This alarm cannot be  silenced while running on battery power  Connecting the N 550 to  AC power will silence the alarm  If the N 550 is not connected to AC  power within approximately 15 minutes  the N 550 will shut off     Note  As the battery is used and recharged over a period of time  the  amount of time between the onset of the low battery alarm and  the N 550 shut off may become
36. bs   single pateint use  adult nasal  OXIMAX OxiCliq  nonadhesive sensor    OxiCliq A 230 kg  66 lbs   single patient use  adult  reusable cable  OXIMAX OxiCliq nonadhesive sensor  OxiCliq N  lt 3 kg or  gt 40 kg  single patient use  neonatal adult     6 6 lbs or  gt 88 Ibs   reusable cable  OXIMAX OxiCliq nonadhesive sensor  OxiCliq P 10 to 50 kg  22 to  single patient use  pediatric  reusable 110 Ibs   cable  OXIMAX OxiCliq nonadhesive sensor  OxiClig I 3 to 20 kg  6 6 to  single patient use  infant  reusable 44 1 lbs   cable  OXIMAX Durasensor   finger clip sensor  DS 100A  gt 40 kg  88 Ibs   reusable  adult  OXIMAx Oxiband   sensor  reusable  OXI A N  lt 3 kg or 240 kg  neonatal adult    6 6 lbs or  gt 88 Ibs   OXIMAX Oxiband sensor  reusable  OXI P I 3 kg to 40 kg  pediatric infant  6 6 Ibs to 88 Ibs     67    Sensors and Accessories    Table 2  Nellcor Oximetry Sensor Models and Patient Weights          Patient Size  OxiMAx Sensor Model   zgreater than    zless than  OXIMAX Dura Y   multisite sensor  D YS  gt l kg  22 2 lbs   reusable  For use with the Dura Y sensor   Ear clip  Reusable  nonsterile  D YSE 230 kg  66 Ibs   Pedi Check    pediatric spot check    i T 3 kg to 40 k  clip  Reusable  nonsterile  D YSPD  6  Pies io E Ibs              Biocompatibility Testing    Biocompatibility testing has been conducted on Nellcor sensors in  compliance with ISO 10993 1  Biological Evaluation of Medical  Devices  Part 1  Evaluation and Testing  The sensors have passed the  reco
37. ccuracy Grid posted on the Internet at     http   www mallinckrodt com respiratory resp Serv_Supp ProductManuals html    Follow the sensor sterilization procedures in the applicable sensor    directions for use  DFU      Table 2  Nellcor Oximetry Sensor Models and Patient Weights          Patient Size  Ox  Max Sensor Model  gt  greater than    lt  less than  OXIMAX MAX FAST adhesive forehead MAX FAST  gt 10 kg  22 lbs     sensor  single patient use       OXIMAX Softcare nonadhesive sensor  SC PR  single patient use  preteen infant     lt 1 5 kg  3 3 Ibs        OXIMAX Softcare nonadhesive sensor  SC NEO  single patient use  adult    1 5 to 5 kg  3 3 to  11 Ibs              OXIMAX Softcare nonadhesive sensor  SC A  gt 40 kg  88 Ibs   single patient use  preteen infant   OXIMAX adhesive sensor  MAX A  gt 30 kg  66 lbs   single pateint use  adult   OXIMAX adhesive sensor  MAX AL  gt 30 kg  66 Ibs   single pateint use  adult  longer cable   36 inches  91 44 cm    OxiMAx adhesive sensor  MAX N  lt 3 kg or  gt 40 kg  single pateint use  neonatal adult    6 6 lbs or  gt 88 lbs     Sensors and Accessories    Table 2  Nellcor Oximetry Sensor Models and Patient Weights                               Patient Size  OxiMax Sensor Model  gt  greater than   lt  less than  OXIMAx adhesive sensor  MAX P 10 to 50 kg  22 to  single pateint use  pediatric 110 lbs   OXIMAx adhesive sensor  MAX I 3 to 20 kg  6 6 to  single pateint use  infant 44 1 Ibs   OXIMAX adhesive sensor  MAX R  gt 50 kg  110 I
38. current and accurate        When a sensor is connected to the N 550 and is not connected to a  patient  the display reads dashes and the N 550 remains in the Pulse  Search mode            ra  mt       t              g  9  9h    I  2sESTD        25    Using the N 550    No Sensor Attached    x f es      ca         ww  I    o       Ld                   gt           gt        N       When the sensor is not attached the N 550 displays dashes       and  the Pulse Search indicator is not lit  indicating that the N 550 failed to  detect a sensor        Sensor Message    26    The Sensor Message feature is an indication that the sensor position   or site needs to be considered  The Sensor Message indicator lights   when the N 550 cannot determine an SpO  level or a pulse rate  The  e  w  Sensor Message recommendations for improving the signal are     Reposition sensor   Check or change adhesive wrap   Choose alternate site   Warm site   Cover sensor   Use forehead  nasal  or ear sensor  adult patients only   Use OxiMAx adhesive sensor   Secure cable   Secure with headband  Max Fasr    Remove nail polish    Loosen sensor  too tight     Using the N 550    e Isolate external interference  electrosurgical device  cell  phone     e Clean site  MAX R        Setting the Pulse Beep Volume    nri con       1     SpO2 Sensor Port       WARNING  Use only Nellcor approved sensors and sensor  A cables     1  Connect a DOC 10 pulse oximetry cable to the SpO2 Sensor Port   1  on the front of the N 
39. d into a properly  grounded AC outlet     ec   3  Verify that the AC POWER INDICATOR is lit        Connecting a Sensor to the N 550    N 550  Use of another sensor cable will have an adverse effect on  performance  Do not attach any cable that is intended for  computer use to the sensor port  Do not connect any device other  than a Nellcor approved sensor to the sensor connector       WARNING  Use only the DOC 10 pulse oximetry cable with the    N 550 17    Setting up the N 550       1     SpO2 Sensor Port       1  Connect a DOC 10 pulse oximetry cable to the SpO2 Sensor Port   1  on the front of the N 550     2  Connect an SpO2 Sensor to the other end of the DOC 10 pulse  oximetry cable  Plug the sensor connector firmly into the  DOC 10 pulse oximetry cable     18    Battery Operation    A       WARNING  Dispose of an old battery by following local  guidelines for disposal of lead acid batteries     Operating the N 550 on Battery Power    N 550    Q    The N 550 has an internal battery that may be used to power the  N 550 during transport or when AC power is not available  A new   fully charged battery will provide at least 1 5 to 2 hours of monitoring  time under the following conditions  no audible alarms sound and no  serial output devices are attached     Note  Whenever the N 550 is connected to AC power  the battery is  being charged  Therefore  it is recommended that the N 550  remain connected to AC power when not in use  This will  make a fully charged battery availa
40. determines SpO2 by passing  red and infrared light into an arteriolar bed and measuring changes in  light absorption during the pulsatile cycle  Red and infrared  low voltage light emitting diodes  LED  in the oximetry sensor serve  as light sources  a photo diode serves as the photo detector     Because oxyhemoglobin and deoxyhemoglobin differ in light  absorption  the amount of red and infrared light absorbed by blood is  related to hemoglobin oxygen saturation  To identify the oxygen  saturation of arterial hemoglobin  the N 550 uses the pulsatile nature  of arterial flow  During systole  a new pulse of arterial blood enters  the vascular bed  and blood volume and light absorption increase     93    Principles of Operation       During diastole  blood volume and light absorption reach their lowest  point  The N 550 bases its SpO2 measurements on the difference  between maximum and minimum absorption  measurements at  systole and diastole   By doing so  it focuses on light absorption by  pulsatile arterial blood  eliminating the effects of nonpulsatile  absorbers such as tissue  bone  and venous blood     There are various matrixes within the OxiMax algorithm  Some are  used to assess the severity of conditions presented to the N 550 in  measuring SpO2 and pulse rate  These individual matrixes or  combinations of these matrixes are used to drive the LED indicators  on the N 550 front panel     During challenging measurement conditions  which could be caused  by low perfusio
41. e sensor  even though a valid sensor is connected to the N 550     The SatSeconds    Indicator  Fills in clockwise as the SatSeconds  alarm management approaches the SatSeconds alarm limit threshold   All segments of the SatSeconds indicator flash during a SatSeconds  alarm  When a SatSeconds setting other than OFF is selected  the  green LED at the top of the SatSeconds indicator will light  The green  LED at the 12 o clock position indicates that SatSeconds alarm  management is engaged     Description of Audible Indicators    N 550    Following are descriptions of N 550 audible indicators     Power On Self Test A 1  second tone indicating that the N 550  Pass has been turned on and has successfully  completed the power on self test       Valid Button Press A short  medium pitched tone indicating  that an appropriate button has been  pressed       Invalid Button Press   A short  low pitched tone indicating that a  button has been pressed that is not  appropriate for the current state of the  N 550       High Priority Alarm   A high pitched  fast pulsating tone  indicating loss of pulse with no patient       motion  Medium Priority A medium pitched  normal pulsating tone  Alarm indicating an SpO2 or pulse rate limit  violation       13    Description of Controls  Indicators  and Symbols       Low Priority Alarm A low pitched  slow pulsating tone  indicating a sensor disconnect  low  battery  or N 550 failure       Alarm Silence Three beeps that sound approximately  Reminder e
42. ected  and observe all  warnings and cautions presented in the directions for use  accompanying the sensor  Clean and remove any substances such as  nail polish from the application site  Periodically check to ensure that  the sensor remains properly positioned on the patient     application or inappropriate duration of use of an SpO  sensor       WARNING  Tissue damage can be caused by incorrect  Inspect the sensor site as directed in the sensor directions for use     N 550 73    Performance Considerations    74    e    High ambient light sources such as surgical lights  especially those  with a xenon light source   bilirubin lamps  fluorescent lights   infrared heating lamps  and direct sunlight can interfere with the  performance of an SpO2 sensor  To prevent interference from ambient  light  ensure that the sensor is properly applied and cover the sensor  site with opaque material     Caution Failure to cover the sensor site with opaque material in  high ambient light conditions may result in inaccurate  measurements     If patient movement presents a problem  try one or more of the  following remedies to correct the problem    e verify that the sensor is properly and securely applied   e move the sensor to a less active site   e use an adhesive sensor that tolerates some patient motion   e use a new sensor with fresh adhesive backing  If poor perfusion affects performance  consider using the OXIMAX  MAX R sensor  it obtains measurements from the nasal septal  anterior et
43. he  display  If alarm limits have been changed from their power on  defaults  a decimal point     is displayed after the SpO2 value  100       The Pulse Rate Display  Shows the pulse rate in beats per minute  It  flashes zeros during loss of pulse alarms and flashes the beats per  minute value in red when the pulse rate is outside of the alarm limit   During pulse search  the N 550 continues to update the display  Pulse  rates outside of the pulse rate range  20 to 250 bpm  are displayed as  the closest value within the range  If alarm limits have been changed  from their power on defaults  a decimal point     is displayed after the  pulse rate value  112       The AC Power Indicator  Lights continuously when the N 550 is  connected to AC power  It also indicates that the battery is charging   It is off when the N 550 is being powered by its internal battery     The Low Battery Indicator  Lights continuously to indicate that 15 or    fewer minutes of battery capacity remain  The Low Battery indicator  flashes when the battery is critically low     11    Description of Controls  Indicators  and Symbols    12       The Alarm Silence Indicator  Lights continuously when an audible  alarm has been silenced  It flashes when the alarm silence duration  has been set to OFF     The Motion Indicator  Lights continuously whenever the OxiMAx   algorithm detects the presence of artifact independent of its severity  or the impact on the SpO2 or pulse rate values  Whenever the motion  indic
44. hmoid artery  an artery supplied by the internal carotid    artery  This sensor may obtain measurements when peripheral  perfusion is relatively poor     Operator   s Menu    N 550    This operator   s menu provides a quick reference to the functions on  the N 550  The functions that are Bold are the default settings     Table 3  Operator   s Menu                      Sub    Reference  Menu Function  Menu Paragraph  1   Trend print  tabular Trend Data Printout  N 550 trend only  on page 59  2   Trend clear Clearing Trend  Information on page  40  3   Language English    4   Baud rate  Setting the Data Port  Baud Rate on page 37    24 2400 baud      96 9600 baud      192 19200 baud    5   Data Port Protocol  Setting the Data Port  Protocol on page 39    1 ASCII      2 External equipment Refer to the external          communications        equipment manuals  for the interfacing  instructions        75    Blank Page    Troubleshooting    A    A  e       Error Codes    N 550    WARNING  If you are uncertain about the accuracy of any  measurement  check the patient s vital signs by alternate means   then make sure the N 550 is functioning correctly     WARNING  The cover should be removed only by qualified  service personnel  There are no user serviceable parts inside     CAUTION  Do not spray  pour  or spill any liquid on the N 550   its accessories  connectors  switches  or openings in the chassis     When the N 550 detects an error condition  it may display the letters   EEE  f
45. indicator 12    N    N 550 returning 85  Nurse Call  Relay Contacts 54  Using 53    O    Operating  relative humidity 102  temperature 102  Operator   s menu 75  OXIMAX technology 5  Oximetry overview 93    P    Patient motion 64   Patient settings  verifying 31  Performance considerations 71  Performance verification 71    110    PH 64   PL 64   Print  trend 44   Printing trend information 57  PS 64   Pulse amplitude indicator 11  Pulse beep volume  setting 27  Pulse rate display 11   Pulse rate high limit alarm 64  Pulse rate low limit alarm 64  Pulse search 64   Pulse search indicator 12    R    Rear panel 8  Returning the N 550 85    S    Safety checks 86   Safety information 1  SatSeconds duration  setting 36  Saturation high limit alarm 64  Saturation low limit alarm 64  Saturation  calculated 95  Saturation  fractional 95  Saturation  measured 95   SD 64   Selecting a sensor 65   Sen Err 77   Sensor disconnect 64   Sensor messages indicator 13  Sensor off 64   Sensor off indicator 12  Sensor selection 66   Sensor  selecting 65   Service 85   Setting pulse beep volume 27  SH 64   Shut down  automatic 19    SL 64   SO 64   Specifications 97  battery 101  compliance 105  electrical 101  electrical  instrument 101  environmental 102  performance 97  physical 104   Storage  relative humidity 103    N 550    Index    temperature 102  Symbols 8    T    Technical assistance 82  Transport  relative humidity 103  temperature 102  Trend clear 44  Trend data operation 43  Trend pr
46. ing  the N 550 must be turned  off and then turned back on     When connecting the N 550 to a printer or computer  verify proper  operation before clinical use  Both the N 550 and the printer or PC  must be connected to a grounded AC outlet     Any printer or computer connected to the N 550 s data port must be  certified according to IEC Standard 60950  All combinations of  equipment must be in compliance with IEC Standard 60601 1 1  systems requirements  Anyone who connects a printer or PC to the  data output port configures a medical system and is therefore  responsible that the system complies with the requirements of system  standard IEC Standard 60601 1 1 and the electromagnetic  compatibility system standard IEC Standard 60601 1 2     57    Printing       1     Data Port Connector       1  Connect the serial printer to the N 550 s Data Port connector  1    2  Turn on the printer   3  Connect the N 550 to an AC outlet     4  Turnon the N 550     The printer will start printing real time trend data or the PC will start  displaying real time data at a rate of one line every 2 seconds     58    Printing    N 550 VERSION 1 00 00 CRC XXXX SpO2 Limit  85 100  PR Limit 40 170 bpm  ADULT 0 SAT S   TIME  SpO2 BPM PA Status   30 Sep 03 16 00 00 100 120 220   30 Sep 03 16 00 02 100 124 220   30 Sep 03 16 00 04 100 170 220   30 Sep 03 16 00 06 120 220   30 Sep 03 18 00 43   Le e SD   30 Sep 03 18 00 45            sD   N 550 VERSION 1 00 00 CRC XXXX SpO2 Limit  80 100  PR Limit 40 17
47. ing the  Alarm Silence button     green LED at the top of the SatSeconds indicator will light  The    3 Note  When a SatSeconds setting other than OFF is selected  the  green LED indicates that SatSeconds have been engaged     N 550 89    Technical Information       Describing SatSeconds    90    With traditional alarm management  upper and lower alarm limits are  set for monitoring oxygen saturation  During monitoring  as soon as  an alarm limit is violated by as little as one percentage point  an  audible alarm immediately sounds  When the  SpOz2 level fluctuates  near an alarm limit  the alarm sounds each time the limit is violated   Such frequent alarms can be distracting     The N 550 utilizes Nellcor SatSeconds alarm management  With the  SatSeconds technique  upper and lower alarm limits are set in the  same way as traditional alarm management  However  the clinician  also sets a SatSeconds limit that allows monitoring of  SpO2 below  the selected lower alarm limit and above the selected upper alarm  limit for a period of time before an audible alarm sounds     The SatSeconds limit controls the time that the  SpOz level may fall  below the alarm limit before an audible alarm sounds     The method of calculation is as follows   The number of percentage points that the  SpO2z falls outside the  alarm limit is multiplied by the number of seconds that the  SpOz  level remains outside that limit  This can be stated as an equation   Points x Seconds   SatSeconds  Where     P
48. int 44    111    Blank Page    Blank Page     NELLCOR    Tyco Healthcare Group LP   Nellcor Puritan Bennett Division  4280 Hacienda Drive   Pleasanton  CA 94588 U S A   Telephone Toll Free 1 800 563 5267    Authorized Representative  Tyco Healthcare UK LTD    154 Fareham Road  Gosport PO13 0AS  UK        2003 Nellcor Puritan Bennett Inc   All rights reserved    Rx ONLY    CE    0123    067856B 1103    tyco    Healthcare    
49. justable alarm  volume       Pitch   30 Hz     932 Hz       Pulse width   20 msec     Nellcor   255  msec       IAW IEC 60601   1 8   120 msec       IAW EN 475    150 msec       Pulse repetition interval    Nellcor   320                              20 msec   double burst    msec  IAW IEC 60601   1 8   6940 msec  IAW EN 475    7500 msec   Repetitions Continually   Invalid Button Press Volume level Not changeable   Pitch   30 Hz  180 Hz   Pulse width   20 msec  70 msec   Pulse repetition interval N A     20 msec   double burst    Repetitions 1       Low Priority Alarm    Volume level    Adjustable alarm  volume                   Pitch   30 Hz  500 Hz  Pulse width   20 msec  200 msec  Pulse repetition interval 15000 msec    20 msec   double burst    Repetitions Continually       99    Specifications    Table 6  Tone Definition          Audible Indicator Parameter Value  Medium Priority Volume level Adjustable alarm  Alarm volume   Pitch   30 Hz  752 Hz       Pulse width   20 msec  Nellcor   N A       IAW IEC 60601   1 8   160 msec       IAW EN 475    200 msec       Pulse repetition interval Nellcor   N A    20 msec   double burst        IAW IEC 60601   1 8   7600 msec       IAW EN 475    20000 msec       Repetitions Continually       POST Pass Volume level Not changeable       Pitch   30 Hz  784 Hz       Pulse width   20 msec  1000 msec       Pulse repetition interval N A    20 msec   double burst                       Repetitions 1  Pulse Beep Volume level Adjustable   Pitch  
50. l for the procedure     Pins 7 and 15 provide a relay that closes when an alarm is sounding  on the N 550  Pins 8 and 15 provide a relay that opens when an alarm  is sounding  Pin 15 is a common lead for both relays     Note  When the relay is closed there is approximately 27 ohms of     resistance        Setting Nurse Call RS 232 Polarity    The nurse call polarity can be set to a high signal on an N 550 alarm  condition or a low signal on an N 550 alarm condition  Refer to the  N 550 service manual for setting the Nurse Call RS 232 polarity        Setting Nurse Call Relays Normally Open Closed    Data port pins 7 and 15 provide a relay that closes  nominally 27  ohms  when an alarm is sounding on the N 550  Pins 8 and 15    54    Using the Data Port    provide a relay that opens when an alarm is sounding  Pin 15 is a  common lead for both relays  The relay operates whether the N 550 is  operating on AC power or battery     N 550 55    Blank Page    Printing       Printing N 550 Real Time Data          N 550    Real time data is continuously sent to the data port on the back of the  N 550  Patient data can be obtained through the data port by  connecting to a computer or serial printer  When a real time printout  or display is being transmitted to a printer or computer  a new line is  printed displayed every 2 seconds  Column headings are printed   displayed every 25 lines  or if one of the values in the column heading  is changed     Note  If the data output stops transmitt
51. mmended biocompatibility testing and are therefore in  compliance with ISO 10993 1        Optional Accessories    Several optional accessories are offered with the N 550  Contact  Nellcor s Technical Services Department  1 800 635 5267  or your  local Nellcor representative for information about these accessories   e Foot switch  e Visual alarm indicator    e Pole mount bracket    68    Sensors and Accessories    Accessories for the N 550 are listed on the Internet at     http   www mallinckrodt com respiratory resp Serv_Supp Apartweb main   PartAcceMenu html       Foot Switch    An optional foot switch is available for the N 550  The foot switch    may be used to silence N 550 alarms  The foot switch is connected to  the rear of the N 550     CAUTION  The N 550 must be turned off when connecting or  disconnecting the foot switch        1     Foot switch connector          Visual Alarm Indicator    An optional visual alarm indicator is available for the N 550  The  visual alarm indicator mounted on the top of the N 550 with double   back tape  The visual alarm indicator is connected to the rear of the  N 550     CAUTION  The N 550 must be turned off when connecting or  disconnecting the visual alarm indicator     N 550 69    Sensors and Accessories       1     Visual alarm indicator       Pole Mount Bracket    An optional pole mount bracket is available for the N 550  The pole  mount bracket may be used to attach the N 550 to an IV pole  The  pole mount bracket is attached to
52. n  motion  external interference  like ambient light  or  a combination of these  the OxiMAx algorithm automatically extends  the amount of data required for measuring SpO2 and pulse rate  If the  resulting dynamic averaging time exceeds 20 seconds  the pulse  search indicator is lit solid and SpO2 and pulse rate will continue to be  updated every second  As these conditions become even more  challenging  the amount of data required continues to extend  If the  dynamic averaging time reaches 40 seconds  the pulse search  indicator begins flashing  the SpO2 and pulse rate displays flash zero  indicating a loss of pulse condition     Automatic Calibration    94    Because light absorption by hemoglobin is wavelength dependent and  because the mean wavelength of LEDs varies  an oximeter must  know the mean wavelength of the sensor s red LED to accurately  measure SpO2     During monitoring  the N 550 s software selects coefficients that are  appropriate for the wavelength of that individual sensor s red LED   these coefficients are then used to determine SpO2     Additionally  to compensate for differences in tissue thickness  the  light intensity of the sensor s LEDs is adjusted automatically     Principles of Operation       Functional versus Fractional Saturation    This N 550 measures functional saturation    oxygenated hemoglobin  expressed as a percentage of the hemoglobin that can transport  oxygen  It does not detect significant amounts of dysfunctional  hemoglobin  such
53. n page 37  The baud rate selections are     e 24  2 400 baud rate     44    N 550 Trend    e 96  9 600 baud rate   e 192 default  19 200 baud rate         5  Data Port Printout  Selections are as follows     Option 1  Printout in ASCII characters     Option 2    External equipment communications  Refer to the external equipment  manual for the interfacing instructions     N 550 45    Blank Page    Using the Data Port       Overview       Patient data can be obtained through the data port on the back of the  N 550 by connecting it to an attached PC or serial printer     When connecting the N 550 to a printer or PC  verify proper  operation before clinical use  Both the N 550 and the printer or PC  must be connected to a grounded AC outlet  The N 550 protocol  setting must be ASCII     Any printer or PC connected to the N 550 s data port must be certified  according to IEC Standard 60950  All combinations of equipment  must be in compliance with IEC Standard 60601 1 1 systems  requirements  Anyone who connects a printer or PC to the data output  port configures a medical system and is therefore responsible for  ensuring that the system complies with the requirements of system  standard IEC Standard 60601 1 1 and the electromagnetic  compatibility system standard IEC Standard 60601 1 2     Connecting to the Data Port    N 550    The N 550 data port may be connected to the printer or PC by using a  cable terminated with an AMP connector  AMP part number  747538 1   ferrule  AMP p
54. ndex    Symbols  96SpO2 display 11    A    AC power indicator 11  Adjust down button 10  Adjust up button 10   Alarm limits  setting 34  Alarm off 63   Alarm priorities 87   Alarm silence 64   Alarm silence button 10  Alarm silence duration  setting 28  Alarm silence indicator 12  Alarm volume  setting 28  Altitude 102  103   Anemia 72   AO 63   AS 64   Audible indicators 13  Audio alarms  disabling 30  Automatic calibration 94  Automatic shut down 19    B    Barametric pressure 102  103  Battery charging 19   Battery disposal 19   Battery use 19   Baud rate 44   Blipbar 11    C    Calculated saturation 95    N 550    Calibration  automatic 94  Cleaning 86   Components  list of 16  Connecting a sensor 17  Connecting to the data port 47  Corrective action 78    D    Data port pinouts 48   Data port printout  Option 1 45  Data port printout  Option 2 45  Data port  connecting to 47  Decimal point 34   Default  factory 88  Disinfecting 86   Disposal  battery 19  Dysfunctional hemoglobins 72    E    Electromagnetic interference 81  EMI 81  Error codes 77    F    Factory defaults 88   Fractional saturation 95   Front panel 7   Functional saturation 95   Functional versus fractional saturation 95    H    Hospital type environments 5    109    Index    L    LB 64   LM 64   Loss of pulse 64   Loss of pulse with motion 64  Low battery 64   Low battery indicator 11   LP 64    M    Measured saturation 95   MO 64   Monitor  accuracy and motion tolerance 97  measurement range 97   Motion 
55. ngth range of the light emitted are  near 660 nm and 890 nm           Environmental Conditions       Operation  Temperature 10   C to 45   C  50   F to 113   F   Altitude  390 m to 3 012 m      1 254 ft  to 9 882 ft      Barometric Pressure 70 kPa to 106 kPa   20 6 in  Hg to 31 3 in  Hg        Relative Humidity 10  to 95  non condensing to be  compliant with IEC 60601 1   sub clause 44 5          Transport and Storage  not in shipping container     Temperature  20   C to 60   C    4   F to 140   F         390 m to 5 574 m    1 254 ft  to 18 288 ft      Altitude       102    N 550    Specifications    Transport and Storage  not in shipping container        Barometric Pressure    50 kPa to 106 kPa   14 7 in  Hg to 31 3 in  Hg        Relative Humidity       15  to 95  non condensing       Transport and Storage  in shipping container        Temperature  20   C to 70   C    4   F to 158   F   Altitude  390 m to 5 574 m    Barometric Pressure      1 254 ft  to 18 288 ft      50 kPa to 106 kPa   14 7 in  Hg to 31 3 in  Hg        Relative Humidity       15  to 95  non condensing    Ox1Max Sensor Power Dissipation       OxiMAx Sensor    Dissipation       OXIMAX MAX FAST adhesive forehead sensor  single    52 5 mW    patient use    OXIMAX Softcare nonadhesive sensor  single patient  52 5 mW  use  preteen infant   OxiMAx Softcare nonadhesive sensor  single patient  52 5 mW  use  adult   OXIMAX Softcare nonadhesive sensor  single patient  52 5 mW    use  preteen infant    OXIMAX adhesive
56. oints   SpO2 percentage points outside of the limit    Seconds   number of seconds that SpO2 remains at that  point outside of the limit    The alarm response time  assuming a SatSeconds limit set at 50 and a  lower alarm limit set at 90  is described and illustrated below     Technical Information    In this example  the  SpOz level drops to 88  2 points  and remains  there for a period of 2 seconds  2 points x 2 seconds   4   The  SpO2  then drops to 86 for 3 seconds and then to 84 for 6 seconds  The  resulting SatSeconds are                     SpO2 Seconds SatSeconds  2x 2  4  4x 3  12  6x 6  36  Total SatSeconds   52          After approximately 10 9 seconds  the SatSeconds alarm will sound   because 50 SatSeconds will have been exceeded  See arrow  1  in                   Figure 6   90  88   SpO2  86  84      50 Second  SatSeconds  Alarm Point    0123 45 6 7 8 9 10 11  SECONDS    Figure 6  Alarm Response with SatSeconds    N 550 91    Technical Information       Saturation levels may fluctuate rather than remaining steady for a  period of several seconds  Often   SpO2z levels may fluctuate above  and below the alarm limit  reentering the non alarm range several  times     During such fluctuation  the N 550 integrates the number of  SpOz  points  both positive and negative  until either the SatSeconds limit   SatSeconds setting  is reached  or the  SpO2 level returns within a  normal range and remains there     SatSeconds    Safety Net     92    The SatSeconds    Safety 
57. ol iussa ko ORE CROR ORA RR CR ICH E c 87  Description of Alarms uisoccacipugsi ieu itededa e acaanineeddetes riot 87  Factory aU i d  quem  LT      88  SatSeconds Display M Mc n 89  Describing SatSeconds               cssseccceceeeesesseeeeeeeeeeeeseesnsaeees 90  SatSeconds    Safety Net     nennen 92  Principles of Operation    cere wckec rau rh REIR yuan 93  Oximetry OVEIVIEW saccsiissccuiisist essc teta qe k raqutaaaetea n tih tae 93  Automatic Calibration                         eeeeeee 94  Functional versus Fractional Saturation                                  95  Measured versus Calculated Saturation                                 95  SDOUIDONUDNS LaadeuadAEEARSASARERCARARRRRATRRGRRA Y A RE RR RRAA X AE 97  Performance  eet e bootie nets 97  Electrical                    eeeesssesssssseeeeeeenennennnnnnne nnne 101  Environmental Conditions                          eeeee 102  Physical Characteristics rrr oia 104  oum o V    nnne enn ene 105   pii EBERT ES TNT TET t ees 109    Figures    N 550    Figure 1   Figure 2   Figure 3   Figure 4   Figure 5   Figure 6   Figure 7     Contents    Front Panel Buttons and Symbols                           7  Rear Panel Symbols          eet rico ttn 8  Data Port Pin Layout    eeen 49  Real Time Data Printout                 ccccccceeceeteeeteees 59  Trend Data Printout                    sess 60  Alarm Response with SatSeconds                        91  Oxyhemoglobin Dissociation Curve                      96    Contents    Tables  
58. ollowed by an error code     Sea   f        When an error code is displayed  turn the N 550 off  wait 10 seconds   and turn the N 550 on  If the error code is listed in Table 4  follow the  action s  listed  If the action does not correct the error condition  notify service personnel  If the error code is not listed in Table 4   notify service personnel     When the N 550 detects a defective sensor connected to the N 550    the N 550 displays an error code of    Sen Err     The sensor should be  replaced and the N 550 power should be recycled     77    Troubleshooting    et     E  o              hj       Table 4  Error Codes    Error    Code Action       513 1     Charge battery  2     Notify service personnel        514 1     Restart the N 550   2     Notify service personnel        525 1     Restart the N 550   2     Notify service personnel        526 1     Restart the N 550   2     Notify service personnel              528 1     Restart the N 550   2     Notify service personnel                    Corrective Action    If you experience a problem while using the N 550 and are unable to  correct it  contact Nellcor   s Technical Services Department or your  local Nellcor representative  The N 550 service manual  which is for  use by qualified service personnel  provides additional  troubleshooting information     Following is a list of possible problems and suggestions for  correcting them     78    N 550    Troubleshooting    There is no response to the Power On Off button
59. or     e Excessive patient motion may be preventing the N 550 from  tracking the pulse  Keep the patient still  if possible  Verify  that the sensor is securely applied  and replace it if necessary   Change the sensor site  Use a type of sensor that tolerates  more patient movement  for example  an adhesive sensor      e The sensor may be too tight  there may be excessive ambient  light  or the sensor may be on an extremity with a blood  pressure cuff  arterial catheter  or intravascular line   Reposition the sensor  as necessary     e Excessive environmental motion or electromagnetic  interference may be preventing the N 550 from tracking the  pulse  Remove the source of interference or try to stabilize  the environment  or do both     The Pulse Search indicator lights after successful  measurements have been made       Check the patient     e Perfusion may be too low for the N 550 to track the pulse   Test the N 550 on someone else  Change the sensor site  Try  another type of sensor  Refer to Sensor Performance  Considerations on page 72     e Prolonged patient motion may be preventing the N 550 from  tracking the pulse  Verify that the sensor is securely applied  and replace it if necessary  Change the sensor site  Use a type  of sensor that tolerates more patient movement  for example   an adhesive sensor      e The sensor may be too tight  there may be excessive ambient  light  or the sensor may be on an extremity with a blood  pressure cuff  arterial catheter  or intra
60. pear normal  an anemic patient may be hypoxic   Correcting anemia can improve arterial oxygen content  The pulse  oximeter may fail to provide an SpO2 if hemoglobin levels fall below  5 gm dl     The N 550 will only measure saturation levels between 1 and 100      The N 550 will only measure pulse rates between 20 and 250 beats  per minute  Detected pulse rates outside the range of 20 to 250 beats  per minute are displayed as the closest value within the range     Sensor Performance Considerations    72    WARNING  Pulse oximetry readings and pulse signal can be  affected by certain ambient conditions  sensor application errors   and certain patient conditions     Performance Considerations    Inaccurate measurements can be caused by       incorrect application of the sensor    e placement of the sensor on an extremity with a blood  pressure cuff  arterial catheter  or intravascular line      ambient light      prolonged patient movement  Loss of pulse signal can occur for the following reasons     e the sensor is applied too tightly    e ablood pressure cuff is inflated on the same extremity as the  one with the sensor attached    e there is arterial occlusion proximal to the sensor    Use only Nellcor sensors and sensor cables     than those specified may result in increased emission and or  decreased immunity and inaccurate readings of the N 550 pulse  oximeter       WARNING  The use of accessories  sensors  and cables other    Select an appropriate sensor  apply it as dir
61. presentative     Obtaining Technical Assistance    82    For technical information and assistance  or to order parts or a service  manual  contact Nellcor s Technical Services Department   1 800 635 5267  or your local Nellcor representative  The service  manual includes block diagrams and a parts list required by qualified  personnel when servicing the N 550     When calling Nellcor s Technical Services Department   1 800 635 5267  or your local Nellcor representative  you may be  asked to tell the representative the software version number of your  N 550     The device version appears in the N 550 display each time the N 550  successfully completes the power on self test  Write the number  down and have it available whenever requesting technical assistance     Troubleshooting    The most recent revision of this manual and the service manual are  available on the Internet at     http   www mallinckrodt com respiratory resp Serv_Supp   ProductManuals html    N 550 83    Blank Page    Maintenance             CAUTION  Follow local governing ordinances and recycling  instructions regarding the disposal or recycling of device components  and accessories     Returning the N 550       Service    N 550    A    Contact Nellcor   s Technical Services Department  1 800 635 5267  or  your local Nellcor representative for shipping instructions including a  Returned Goods Authorization  RGA  number  Unless otherwise  instructed by Nellcor   s Technical Services Department  it is not  nece
62. red SatSeconds setting        Setting the Data Port Baud Rate          Discussion  The baud rate determines the speed at which the N 550 sends data to  the attached equipment  printer or portable computer   The baud rate  is determined by the capabilities of the attached equipment   Procedure    With the N 550 in the normal monitoring mode     1  Simultaneously  press the SpO2 Alarm Limit button and Pulse    Rate Alarm Limit button for 3 seconds  Menu option 1 will be  displayed     N 550 37    Using the N 550    38        pa  I         g  Lal    i   4 w   590       2  Press the Adjust Up button until menu item option 4 is displayed                 r   et            e         EID       5        3  Press the SatSeconds Alarm Limit button  Current baud rate is  displayed              ntiuscon         2l  mt       t        gt     PTS    t           C2 Press the Adjust Up button or Adjust Down button to select the     desired baud rate  Possible settings are    e 24  2 400 bps    e 96  9 600 bps    e 192 default  19 200 bps      Using the N 550       Setting the Data Port Protocol    With the N 550 in the normal monitoring mode     1  Simultaneously  press the SpO2 Alarm Limit button and Pulse    Rate Alarm Limit button for 3 seconds  Menu option 1 will be  displayed     ra  mt       t       E                     i  ez             e                       st990        3  Press the SatSeconds Alarm Limit button  Current protocol is  displayed     N 550 39    Using the N 550    d   
63. rm limit  is 100   Pulse Rate alarm limits are 40 to 170 bpm  SatSeconds  alarm limit is set to off     N 550 61    Printing    N 550 Mode    N 550 VERSION 1 00 00 CRC  XXXX SpO2 Limit  85 100  PR Limit  40 1708PM  ADULT 0 SAT S    TIME  SpO2 BPM PA Status       The N 550 mode is ADULT        Data Column Headings    N 550 VERSION 1 00 00 CRC  XXXX SpO2 Limit  85 100  PR Limit 40 170BPM    ADULT 0 SAT S  TIME 96SpO2 BPM PA Status       Actual column headings are in the third row of the column heading  line  Patient data that is presented in the chart  from left to right  are  the time that the patient data on the row was obtained  the current   SpO2 value being measured  the current Pulse Rate in beats per  minute  bpm   the current Pulse Amplitude  PA   and the operating  status of the N 550        Time    TIME  6SpO2 BPM PA Status  30 Sep 03 16 00 00 100 120 220    The Time column represents the N 550 real time clock  Refer to the  N 550 service manual to set the N 550 real time clock     62    Patient Data       Operating Status    N 550    Printing    N 550 Version 1 00 00 CRC  XXXX SpO2 Limit  85 100  PR Limit  40 170 bpm  ADULT 0 SAT S    TIME  SpO2 BPM PA Status  30 Sep 03 16 00 00 100 190  220 PH       Patient data is highlighted in the display above  Parameter values are  displayed directly beneath the heading for each parameter  In this  example  the  SpOz is 100  and the pulse rate is 190 beats per  minute  An asterisk indicates that the parameter is outside the set
64. ssary to return the sensor or other accessory items with the  N 550  Pack the N 550 in its original shipping carton  If the original  carton is not available  use a suitable carton with appropriate packing  material to protect it during shipping     Return the N 550 by any shipping method that provides proof of  delivery     WARNING  The cover should be removed only by qualified  service personnel  There are no user serviceable parts inside     The N 550 requires no routine service or calibration other than  changing the battery at least every 24 months  Refer to the N 550  service manual for the battery changing procedure     If service is necessary  contact qualified service personnel or your  local Nellcor representative     85    Maintenance       Periodic Safety Checks    It is recommended that the following checks be performed every 24  months     e Inspect the equipment for mechanical and functional damage     e Inspect the safety relevant labels for legibility        Cleaning    CAUTION  Do not spray  pour  or spill any liquid on the N 550   its accessories  connectors  switches  or openings in the chassis   For surface cleaning and disinfecting the N 550  follow your  institution s procedures or     e The N 550 may be surface cleaned by using a soft cloth  dampened with either a commercial  nonabrasive cleaner or a  solution of 70  alcohol in water  and lightly wiping the  surfaces of the N 550     e The N 550 may be disinfected using a soft cloth saturated    with a
65. st be turned off and  back on again before a new trend printout can begin        Column Headings    N 550 VERSION 1 00 00 CRC  XXXX SpO2 Limit  85 100  PR Limit  40 170BPM  ADULT 0 SAT S    TIME  SpO2 BPM PA Status       Column headings are displayed or printed after every 25 lines  or if  one of the values in the column heading changes     60    Printing    Data Source    N 550 VERSION 1 0 0 0 CRC  XXXX SpO2 Limit  85 100  PR Limit  40 160BPM    ADULT 0 SAT S  TIME  SpO2 BPM PA Status       Data in the highlighted box above represents the model number of the  N 550  in this case the N 550        Device Software Revision Level    N 550 VERSION 1 00 00 CRC  XXXX SpO2 Limit  85 100  PR Limit  40 170BPM  ADULT 0 SAT S    TIME  SpO2 BPM PA Status       The next data field tells the user the software level  Version 1 00 00   and a software verification number  CRC  XXXX   Neither of these  numbers should change during normal operation  The numbers may  change if the N 550 is serviced and receives a software upgrade     Alarm Limits    N 550 VERSION 1 00 00 CRC  XXXX SpO2 Limit  85 100  PR Limit 40 170BPM  ADULT 0 SAT S    TIME  SpO2 BPM PA Status       The last data field in the top line indicates the upper and lower alarm  limits for  SpO2 and for the pulse rate  PR   and the SatSeconds  alarm setting  OFF   The SatSeconds setting may be OFF  10  25  50   or 100 depending on the SatSeconds alarm setting  In the example  above  the low alarm limit for SpO2 is 85  and the upper ala
66. t   gt   gt  PP    WARNING  Before use  carefully read the sensor directions for  use  including all warnings  cautions  and instructions     WARNING  Do not use a damaged sensor or pulse oximetry  cable  Do not use a sensor with exposed optical components     WARNING  Use only Nellcor sensors and pulse oximetry cables  with the N 550  Other sensors or pulse oximetry cables may cause  improper N 550 performance     WARNING  Use only one pulse oximetry cable to increase the  length of the sensor  Use of more than one pulse oximetry cable  may have an adverse effect on performance  Do not attach any  cable that is intended for computer use to the sensor port     WARNING  Pulse oximetry readings and pulse signal can be  affected by certain ambient environmental conditions  sensor  application errors  and certain patient conditions     WARNING  Tissue damage can be caused by incorrect  application or duration of use of an SpO sensor  Inspect the  sensor site as directed in the sensor directions for use     65    Sensors and Accessories    66      WARNING  Do not immerse or wet the sensor     When selecting a sensor  consider the patient s weight and activity  level  the adequacy of perfusion and the available sensor sites  the  need for sterility  and the anticipated duration of monitoring  For  more information  refer to Table 2 or contact your local Nellcor    representative     For a complete and up to date listing of all sensors applicable to the  N 550  refer to the Sensor A
67. ta Port Printout                      eeseeeeeeerenenenn 45  RDO Mr                         45   el Mr              45   Using the Data POM 6 664 khao a OREOR CRM  ROORODR RR CIO RR RC RR ORO 47  e p                         47  Connecting to the Data Port                    sssssssssseeeees 47  Data Port Pino  ts                eese a eren 48    Contents    Data Port Setup eicere nere ases Ub tis tuccerneussseecvsgcants 49  DISCUSSION darti Et eb E eH ILE UIS 49  irepeieW e REERETTUUM 50   Nurse Call Interface                    sssssssee 53  Setting Nurse Call RS 232 Polarity                                54  Setting Nurse Call Relays Normally Open Closed         54   PIMOS Laxzdaus  sqadCRAaAPAdERAAR RR RAE PARRRCSAR RARE ELC eee gd d Rd 57   Printing N 550 Real Time Data                                    sssssss  57   Trend Data Printout                     eesseeeeesseeeennnrer 59  Column Headings d eeu E PE MERE be ER A aie 60  Data  SOUIce   enne epe o taies 61  Device Software Revision Level                                    61  Alarm DImlts sssoseibtt cin te Etat Iba tenta E este 61  N55 0  MOC MP 62  Data Column Headings                          seseseess 62  Bst e MU E 62  Patient  Data    cesarea sa inr re Ea ER eR 63  Operating STATUS  i  ttr e bei ae iin E 63   Sensors and Accessories           0000 eee e eee eee eee eee eee 65   Selecting a Sensor assaxitsiexantirisate o S tetra mp bc ERATES KE 65   Biocompatibility Testing   d reci dated Gide  68   Optional 
68. ted  from the patient  the N 550 remains in the Pulse Search Mode for  awhile  then the N 550 displays           3 dashes  in the  SpOz  and pulse rate display  One way to test the nurse call function is to  create an alarm condition  for example  sensor disconnect  and  verify that your facility s nurse call system is activated       WARNING  The nurse call feature is not functional whenever the    N 550 53    Using the Data Port    The nurse call feature of the N 550 is operational when the N 550 is  powered by AC power or battery power  The nurse call feature of the  N 550 works in conjunction with the nurse call system of your  institution when the N 550 sounds an audible alarm     The N 550 provides two different types of nurse call interfaces  an  RS 232 format and relay closure  Both interfaces function when the  N 550 is operating either on AC power or battery power     The remote location is signaled anytime there is an audible alarm  If  the audible alarm has been turned off or silenced  the nurse call  function is also turned off     Pin 11 on the data port is the RS 232 level nurse call signal and pin 5  or 10 is ground  see Table 1 on page 48   When there is no alarm  condition  the voltage between pins 10 and 11 is  5 to  12 VDC   Whenever the N 550 is in an alarm condition  the output between pins  10 and 11 is  5 to  12 VDC  This is the default condition  normally  low   There is a service menu to change the default condition  Refer  to the N 550 service manua
69. th the N 550 and  the instrument connected to it must be connected to a grounded  outlet  Accessory equipment connected to the N 550 s data  interface must be certified according to IEC Standard 60950 for  data processing equipment or IEC Standard 60601 1 for  electromedical equipment  All combinations of equipment must  be in compliance with IEC Standard 60601 1 1 systems  requirements  Anyone who connects additional equipment to the  signal input port or signal output port  N 550 data port  connector  configures a medical system and is therefore  responsible for ensuring that the system complies with the  requirements of system standard IEC Standard 60601 1 1 and  the electromagnetic compatibility system standard IEC Standard  60601 1 2  The N 550 accuracy may degrade if it is connected to  secondary I O devices when the instrument is not connected to  earth reference     CAUTION  Do not lift the N 550 by the sensor cable or power  cord because the cable or cord could disconnect from the N 550   causing damage to the N 550 or injuring the patient     CAUTION  Federal law  U S A   restricts this device to sale by or  on the order of a physician     Safety Information       Notes          Notes are identified by the Note symbol shown above     Notes provide additional useful information     N 550 3    Blank Page    Introduction    A       WARNING  The N 550 is intended only as an adjunct in patient  assessment  It must be used in conjunction with clinical signs and  symptoms 
70. ting       Note  When an alarm limit is changed from power on default  the  LA N 550 displays a decimal point     after the changed parameter     34    Using the N 550    pa  mi       t    Seo   Ri       3  Press the SpO2 Alarm Limit button twice to view the current    C        N 550     SpO2 lower alarm limit     pa  d    5   gt        4  Press the Adjust Up button or Adjust Down button to increase or  decrease the alarm limit setting     5  Press the Pulse Rate Alarm Limit button to view the current upper  alarm limit for Pulse Rate          pa               d      sts00    6  Press the Adjust Up button or Adjust Down button to increase or  decrease the alarm limit setting     35    Using the N 550    7  Press the Pulse Rate Alarm Limit button twice to view the current    C  lower alarm limit for Pulse Rate     D    g   9   BA          if    stSn         C  8  Press Adjust Up button or Adjust Down button to increase or  C2 decrease the alarm limit setting        Setting SatSeconds Duration       Discussion    Refer to Describing SatSeconds on page 90 for a description of the  SatSeconds function        Procedure    With the N 550 in the normal monitoring mode     1  Press the SatSeconds Alarm Limit button  The current SatSeconds  setting is displayed     36    Using the N 550    on  v  9  Lg   gt   y          A Note  The possible settings for SatSeconds are Off  10  25  50  and  LA 100 seconds        2  Press the Adjust Up button or Adjust Down button to select the  C2 desi
71. tton to save the desired baud       6  Simultaneously press and hold the SpO  Alarm Limit button and    Pulse Rate Alarm Limit button for at least 3 seconds  Option 1  will be displayed     N 550 51    Using the Data Port    E     e    g  2    44z600                     2l  d             Es        II  Anesa        selected protocol will be displayed     TR 8  Press the SatSeconds Alarm Limit Button to select Option 5  The          52        ra     e           ES    I   44208 H       Note  The available selections are     e     Real time ASCII    e 2   External equipment communications  Refer to the  external equipment manuals for the interfacing instructions     Using the Data Port       9  Press the Adjust Up button or the Adjust Down button to display    G2 the desired protocol     10  Press the SatSeconds Alarm Limit button to save the desired  protocol           Nurse Call Interface    WARNING  The nurse call feature should not be used as the  primary source of alarm notification  The audible and visual  alarms of the pulse oximeter  used in conjunction with clinical  signs and symptoms  are the primary sources for notifying  medical personnel that an alarm condition exists     pulse oximeter alarms are silenced     Caution  The nurse call function needs to be tested after it has  been set up in your facility  The nurse call feature should be  tested whenever setting up the N 550 pulse oximeter in a location  that uses nurse call  If an attached OxiMax sensor is disconnec
72. ubject the N 550 to extreme moisture  such as  direct exposure to rain  Such exposure may cause inaccurate  performance or device failure     15    Setting up the N 550     gt   gt  PP       WARNING  Do not use an N 550  sensor  cables  or connectors  that appear to be damaged     WARNING  The N 550 is not defibrillator proof  However  it  may remain attached to the patient during defibrillation or while  an electrosurgical unit is in use  but the readings may be  inaccurate during use and shortly thereafter     WARNING  In the USA  do not connect the N 550 to an electrical  outlet controlled by a wall switch because the N 550 may be  accidentally turned off     WARNING  Use only the DOC 10 pulse oximetry cable with the  N 550  Use of another sensor cable will have an adverse effect on  performance  Do not attach any cable that is intended for  computer use to the sensor port  Do not connect any device other  than a Nellcor approved sensor to the sensor connector     List of Components    1     N 550 Pulse Oximeter   1     Nellcor Sensor or Assortment Pack  1     DOC 10 Pulse Oximeter Cable   1     N 550 Operator s Manual    1     Hospital Grade Power Cord or power cord appropriate for  country of sale    1     Sensor Accuracy Grid    1     Quick Guide    Setting up the N 550       Connecting the N 550 to AC Power       1     Power Connector    1  Plug the female connector of the power cord into the N 550 AC  power connector  1      2  Plug the male connector of the power cor
73. utton one time to decrease the display by one digit  Holding the  Adjust Down button down for more than one second cause the  display digits to scroll     The SpO2 Alarm Limit button is used to view the SpO2 alarm limit   When the SpO  Alarm Limit button is pressed at the same time as the  Pulse Rate Alarm Limit button for approximately three seconds  the  menu options are enabled     The Pulse Rate Alarm Limit button is used to view the pulse rate  alarm limit  When the Pulse Rate Alarm Limit button is pressed at the  same time as the SpO2 Alarm Limit button for approximately three  seconds  the menu options are enabled     The SatSeconds Alarm Limit button is used to view the SatSeconds  alarm limit  When pressed  the menu options are enabled  The Adjust  Up and Adjust Down buttons are used to change the SatSeconds limit  settings        Description of Controls  Indicators  and Symbols    Description of Displays and Indicators    N 550    M     SpO2    O min    ecc    Sec  R        I        E  IE  I                         The Pulse Amplitude Indicator  blip bar   A 10 segment LED that  indicates pulse beat and shows the relative pulse amplitude  As the  detected pulse becomes stronger  more bars light with each pulse     The  SpO2 Display  Shows the saturation level of oxygenated  hemoglobin  The display value flashes zero during loss of pulse  alarms and flashes the SpO2 value when the SpOz is outside the alarm  limits  During pulse search  the N 550 continues to update t
74. vascular line   Reposition the sensor  as necessary     e Excessive environmental motion or electromagnetic  interference may be preventing the N 550 from tracking the       Troubleshooting    pulse  Remove the source of interference or try to stabilize  the environment  or do both     6  The letters EEE  followed by a number  appear on the display     e This is an error code  To confirm  press the Power On Off  button to turn the N 550 off  then press the button again to  turn it back on  If the display shows the error code once  again  record the number and provide that information to  qualified service personnel  or your local Nellcor  representative     e Error Code    EEE 513  is displayed when the battery is  discharged to a critically low level  Turn the N 550 off and let  it charge for about 10 minutes and then turn the N 550 back  on  If the error code is still present  turn the N 550 off and let  it continue to charge  If the N 550 has been charged for 30  minutes and the error code is still present  notify service  personnel     Refer to the N 550 service manual for a complete listing of error  codes     EMI  Electro magnetic Interference     N 550    e    CAUTION  This device has been tested and found to comply with  the limits for medical devices to the IEC 60601 1 2  EN60601 1 2   and the Medical Device Directive 93 42 EEC  These limits are  designed to provide reasonable protection against harmful  interference in a typical medical installation     However  bec
75. very 3 minutes when alarms are silenced  with the alarm silence duration set to OFF       Pulse Beep A single beep sounds for each detected  pulse  The pitch changes as monitored  SpO2 values increase or decrease        Volume Setting Tone   A continuous tone that is used to adjust  the alarm volume       Confirmation Tone Three beeps sound to indicate that default  settings have been saved or reset to  factory defaults or trend data has been  deleted          14    Setting up the N 550    N 550    P PPP PP    WARNING  Explosion hazard  Do not use the N 550 pulse  oximeter in the presence of flammable anesthetics or gases     WARNING  Pulse oximetry readings and pulse signals can be  affected by certain ambient environmental conditions  sensor  application errors  and certain patient conditions  See the  appropriate sections of the manual for specific safety  information     WARNING  To ensure patient safety  do not place the N 550 in  any position that might cause it to fall on the patient     WARNING  As with all medical equipment  carefully route  patient cabling to reduce the possibility of patient entanglement  or strangulation     WARNING  Disconnect the N 550 and Nellcor sensor from the  patient during magnetic resonance imaging  MRD scanning   Induced current could potentially cause burns  The N 550 may  affect the MRI image  the MRI unit may affect the accuracy of  oximeter measurements     WARNING  To ensure accurate performance and prevent device  failure  do not s
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
686bx user`s manual - kitamuracomputers.net  ウォームレット S/G 施工説明書  WorkCentre 5135-5150 Information Assurance Disclosure  EL-EVA1/2, KON1/2, SK1/2  Response NETWORK CONTROLLER User manual  Modbus Option Board  放射線測定器(γ) NaI シンチレーション式サーベイメーター AustralRad    Copyright © All rights reserved. 
   Failed to retrieve file