Home
        MAGLIFE light
         Contents
1.       Alarm field  mT   Shutdown field  mT     AA  DU  A       Alarm limits               Releases    Reserved SCHILLER    i i    Device OFF       Download  Connect the USB PEN DRIVE  to the DEVICE    DEVICE WILL SHUTDOWN  AT THE END OF THE PROCESS          Part number  0 48 0080 Page 41    MAGLIFE light    4 9 Alarms    4 9 1 Screen alarm symbols    S1    S2    S3    S4       Audible alarms enabled    Alarm sound disabled   Displayed when at least one monitoring threshold is disabled   OFF Technical alarms will go off all the same     X    Alarm sound disabled for 2 minutes   Displayed when you press key  9   for less than three  amin seconds   The remaining time is displayed under the symbol     1 44    y  a  mer  T    Alarm sound disabled permanently   Displayed when the alarm off key  9  is held down for more  than three seconds     3 DK    Alarm muting function    Key  9  is used to mute the alarms    e When this key is pressed briefly  alarms stay disabled for two  minutes and the symbol   3  shows the time remaining in Minutes up  the end of the disabling of the alarm    e  f you press key  9  for three seconds or more  alarm sounds remain  disabled permanently or until the key  9  is pressed again  Symbol   S4  is displayed  the symbol    co    flashes and a beep is emitted once  in every two minutes to remind the user     Audible alarm enabling disabling  The permanent disabling of audible alarms is not permitted in some  medical facilities  That is why that function
2.     3 1 MAGLIFE light symbols      Main device on off key    FX  Mains power on   Battery charge   Entry into main menu or exit from any menu  Audio alarm enable disable    Non invasive blood pressure start stop           CF type device  is protected against defibrillation shocks  device designed  for direct heart applications    A Caution  Read the instructions for use of the device        Electrical and electronic device identification symbol  The device  components must be disposed of separately and the relevant parts must be  sent to available recycle centres  Inappropriate disposal may be harmful to  the environment and to public health as a result of the presence of  hazardous materials in electrical and electronic devices       E Notified body for CE certification  G MED   0459    3 2 Power symbols       Power connected         Power input    Part number  0 48 0080 Page 10    MAGLIFE light    4  Use of MAGLIFE  ight    4 1 Installation    MAGLIFE light has been designed to operate at the patient s bedside  It is installed in the  Faraday cage  i e  in the room with the MRI system     Minimum distances must be kept in relation to the measuring tunnel entrances  These  depend on the magnetic field of the magnet and the type of magnet    As a result  it is imperative the MAGLIFE light be kept outside the area  around the magnet demarcated by the 40 mT  400 G  line  A magnetic field  detector sets off an alarm if that value is reached  Beyond the 40 mT limit  the  monitor 
3.   0 48 0080 Page 23    MAGLIFE light    e Use the Alarm limits tab to manage thresholds  For more details  refer  to section 4 7 1 3 Alarm limits and 4 9 Alarms    On   Off    Average    Patient type    Alarms limit    Exit       e SPO2 alarm limits setting  Select a limit by rotating the navigation knob  6   press to confirm    High SpO2  Low  p02  High Pulse    Low Pulse       Select the limit value by rotating the navigation knob  6  again  press to  confirm the selected value     High SpO2    Low SpO2    High Pulse    Low Pulse       Note  At the power up time  the alarm levels corresponds to the standard values  of the selected patient  The operator selection is done in general alarm limit  management window refer to section 4 7 1 3 Alarm limits    Part number  0 48 0080 Page 24    MAGLIFE light    4 7 1 3 Alarm limits menu    The Alarm limits menu is used to set the thresholds for triggering alarms    relating to the different parameters     Caution  The Thresholds menu only displays the thresholds for the    parameters present in the configuration     Caution  If one or more alarms have been disabled  an icon  with a crossed out bell and the word OFF is  displayed in the upper right hand corner of the    screen   The following submenus are available     Alarms limit Neonate    Spor Pulse 90    fo 85    NIBP SYS 49  DIA 30  MAP 39    Operator 1   Operator 2                  Default    200  100  140  100  120    p min            When the device is powered up and when it 
4.  Fichier 21 12 2005 18 25  E  update_files 1 Ko Fichier 21 12 2005 18 26    Insert the USB key in the connector  15  provided  select the update button  and loading will start automatically  Follow the instructions  At the end of  the loading process  the device will go off automatically     Download  Connect the USB PEN DRIVE  to the DEVICE    DEVICE WILL SHUTDOWN  AT THE END OF THE PROCESS       Note  The procedure has been described in detail in the Service  Manual of the device  For more information  please contact  the technical service network of Schiller Medical     Part number  0 48 0080 Page 37    MAGLIFE light    4 7 2 5 Version submenu    The Version tab will show detailed info about used software versions and  releases in MAGLIFE  ight       4 7 2 6 Reserved SCHILLER submenu    This submenu is only for technical purpose and is described in a separate  manual     Part number  0 48 0080 Page 38    MAGLIFE light    4 8 Structure of menus    4 8 1 User menus    NIBP Menu       Alarm limits menu  BO display    Trend Menu             Settings menu       138   82 102       59   as 136 78 un     1028   ro   se   438 77 os     40 28 6   oF 138 82 402     7030   o   m     yo   10 32 5 oF 142   88 404       ag   s   ss   1m oo      40 36   60 98 138  77 86    Part number  0 48 0080 Page 39    MAGLIFE light      NIBP Menu           16 39    poa    al    Part number  0 48 0080 Page 40    MAGLIFE light    4 8 2 Configuration Menus              Same for Child    Same for Adult  
5.  Special displays  ES    Pulse pimin Spo  4h NIBP mmHg    se a  a ons oo 1387 82 102    Technical alarm    w20  s   o8   138782 102  o 10 22 1601   o9   maz s8 10  nos   ten   o   138 77 9    mos 60   o   138 82 102    T ctl          4 7 1 6 Edit menu     To be written      Part number  0 48 0080 Page 31    MAGLIFE light    4 7 1 7 Settings menu    The Settings menu is used to set the general parameters of MAGLIFE  light and to access the demonstration mode        4 7 1 7 1 Sound submenu    The sound submenu is used to set the level of the sound for alarm signals   pulse bip or keyboard clics    Alarms sound  Pulse sound    Keyboard sound       Sounds may be High  Medium  Low or Off  with the exception of alarms   which can never be Off     Alarms sound High  Medium  Pulse sound Low    Keyboard sound       Part number  0 48 0080 Page 32    MAGLIFE light    4 7 1 7 2 Clock submenu    The various buttons are used to set the date and time  The daylight saving  option can be disabled  The clock change date is pre set     Day    el    Summerfainter time       Example of date setting  Only plausible values can be selected        Note  A time or date setting will erase all trends     4 7 1 7 3 Screen brightness setting  You can set the screen brightness from 50 to 100         Part number  0 48 0080 Page 33    MAGLIFE light    4 7 2 Device configuration menu    The configuration menus are hidden during normal operation  Access is  initiated by pressing the navigation button  6  when 
6.  can be configured   See  Permanent muting section 4 7 2 2      4 9 2 Physiological alarms    Part number  0 48 0080    If the measured value of a parameter exceeds a threshold for over three  seconds  an alarm goes off and    the display of the measured value flashes in red   an intermittent audible alarm sounds  sequence of four digital sounds that  is repeated    the window of the parameter over the threshold flashes on a red  background and the display colour are inversed   Depending on the choice made while configuring  See section 4 7 2 2  the  alarm stops      As soon as its cause stops  non latched      After the causes disappears and key  9  is pressed  latched     Page 42    MAGLIFE light    4 9 3 Technical alarms    When a technical alarm is triggered    an error message is displayed in the display field of the parameter in  question   an intermittent audible alarm goes off  sequence of two audible  signals that is repeated   it depends on the audible alarm  configuration criteria   a question mark       is displayed instead of the measured value     Note  going beyond of the measurement limits of the device    In this case no error message is displayed    e technical alarm sounds    e 3 fixed indents         are displayed in the place of the measured value     These alarms go off automatically when their causes disappear     5  Technical specifications    9 1    Manufacturer  Device name  Dimensions  main unit  power supply  Weight  main unit  power supply    Prot
7.  connect a device that simulates a physiological signal to the device  e g  SPO     wait for the value to stabilise and take the reading   e select the corresponding parameter with button  6  and select the Thresholds  button   e modify one of the threshold values so that the value of the physiological parameter  read earlier exceeds it   e confirm by pressing button  6    After 3 seconds  an alarm will go off     Part number  0 48 0080 Page 48    MAGLIFE light    9  Additional accessories and indications    MAGMOVE light Amagnetic trolley with storage basket for MAGLIFE light    0 13 0010 Adult finger SpO  sensor  5 2 m   fibre optic  0 13 0011 Child finger SpO2 sensor  5 2 m   fibre optic  0 13 0012 Universal Y SpO2 sensor  5 2 m   fibre optic  0 13 0008 Adult finger SpO2 sensor  4 2 m   fibre optic  0 13 0002 Child finger SpO2 sensor  4 2 m   fibre optic  0 13 0001 Universal Y SpO2 sensor  4 2 m   fibre optic  0 13 0009 Adult finger SpO2 sensor  3 5 m   fibre optic  0 13 0004 Child finger SpO2 sensor  3 5 m   fibre optic  0 13 0003 Universal Y SpO2 sensor  3 5 m   fibre optic  0 22 0003 5 m cuff tube   W1404413 3 5m m cuff tube   U50143 2 5 cm neonate cuff   U50142 4 cm neonate cuff   U50130 6 cm child cuff   U50129 7 cm child cuff   U50140 9 cm child cuff   0 40 0002 12 cm adult cuff   U50128 14 cm adult cuff   U50141 16 cm adult cuff   3 10 0118 Shielded MAGLIFE  ight power cable   10  Disposal    10 1 Battery disposal  Important    In normal use  the battery requires no
8.  disconnect  turn the connector to the left and pull it out      3  Place the sensor over the nail of the index finger or equivalent site   place the cable on the top of the hand and fix it using Velcro    strap  or secure to the patient s wrist without tightening it     Warning  Be cautious in placing the cable correctly so as to prevent damage  or injury  tangling and or strangulation      Part number  0 48 0080 Page 14    MAGLIFE light    Warning     Warning     Caution     Caution   Warning     Warning     Warning     Warning     Warning     Caution     Part number  0 48 0080    Remove nail polish or false nails before placing the sensor on the  patients finger  as they could lead to inaccurate oximeter  measurements  Cut long nails  as they could hinder the placing of  the sensor     If the sensor is secured with tape  do not tighten the tape  excessively  If the tape is too tight  it may affect the measurement  accuracy of the device and blisters may form on the patient s skin   due to the lack of circulation in the skin and not because of a  source of heat      Do not place the sensor on an extremity with an invasive probe or  a blood pressure cuff     Do not place the cuff on the same limb that the oxygen saturation  is being measured  as it could disrupt the correct determination of  the data  resistance that hinders the blood flow      The oximeter measurement may be modified by the presence of  strong ambient light  Cover the sensor  e g  with a surgical sheet  if 
9.  maintenance   After five years  the battery must be replaced  whether or not the device has  been used     Caution  Explosion hazard  The battery may not be incinerated or disposed of    with household waste     Caution  Acid burn hazard  Never open or overheat the battery     In accordance with national law  the battery may only be disposed of in an approved  disposal facility or sent back to SCHILLER     10 2 Device disposal    The device components must be disposed of separately and the relevant parts must be  sent to appropriate recovery and disposal centres    If you do not know of such a recovery and disposal system  you may return the device  to the distributor or manufacturer  who will take charge of disposing of the device in  accordance with applicable regulations  In that way  you will contribute to the recycling  and recovery of old electrical and electronic devices and their reuse in other forms   Inappropriate disposal may be harmful to the environment and to public health as a  result of the presence of hazardous materials in electrical and electronic devices     Part number  0 48 0080 Page 49    11 Front    o     s 1 82        10    11    Part number  0 48 0080    Page 50    IEE       12    12 Rear and power supp     lt        Part number  0 48 0080 Page 51    
10.  required     The incorrect application or improper use of the sensor can lead to  measurement inaccuracies  as can the presence of significant  levels of dysfunctional haemoglobins  e g  carboxyhaemoglobin or  methaemoglobin  or intravascular dyes such as indocyanine green  or methylene blue  the exposure to excessive light such as with  surgical lamps  especially lamps with xenon sources   bilirubin  lights  fluorescent lamps  infrared heating lamps or direct sunlight   excessive movement by the patient  vein pulsations  the installation  of a sensor on a limb with a blood pressure cuff  an arterial probe  or an intravascular line     In some cases  when the perfusion and signal are weak  e g  with  patients with a thick or dark skin  the device may produce  abnormally low oximeter readings  Oxygenation should be verified  before starting therapies and interventions  especially in preterm  babies and patients with chronic lung diseases     It often happens that patients suffer from low peripheral perfusion  due to hypothermia  hypovolaemia  serious vasoconstriction or  reduced cardiac output etc  These symptoms may lead to a loss in  the oximeter readings     The temperature of the patient and the room must not be too low  since it will affect the measurement outcome     Do not use the oximeter alone while monitoring vital parameters     Page 15    MAGLIFE light    4 3 2 2 Sphygmomanometer cuffs  Non Invasive Blood Pressure     1  Make sure that the NIBP function is on and 
11.  se tesama Waynes eta tonewe A 39  CORSA or Men  s orrasa a E E ana atelanes  41  A A 42  Screen ala DOS ica 42        WSI     42  Technical SITE 43  Tecnnicalsbecificatiohms ai a 43  Ne AAA A O ERE DEDE sstaceuiearanta  43  Technical specitications 0t Moquies sensitivity oe 44  A II O SN DST ES YTRET ERE ES SENDE 44  NIBP Nom Invasive EGR 44  Threshold anges A A A td 45  CIC ANIIAG aa 46  TFOUDIESMOOUIG aid aci  n 47  General CHO al ao aitor 4   A ERE SEERE REESE ERE ESSEN ERE RER ERE ES teenies 48  Maintenance siii ii PEGE NE BES OSS Ea 48  Alaris HALLS OD ane PPPoE     EE A rer cer ere 48  Additional accessories and indicatiONS    ooooccccccccccccnnononancnannnnnnnnnnn cnn rr rn cnn 49  DISDSa iiccsuAmctesthacce SVERRE NS e TEE SR SER ER SBE SN aa Aaa a i 49  Bat ISDN 49  Device disposi is 49  ON 50  Rear   nd POWer SUD aiii is 51    1  Description of MAGLIFE light    MAGLIFE  ight is a monitor designed solely for monitoring vital patient parameters during MRI   Magnetic Resonance Imaging  examinations     Depending on the selected version  MAGLIFE  ight will monitor the following parameters     transcutaneous measurement of arterial oxygen saturation  SpO2  pulse oximeter   pulse rate    blood pressure  NIBP     MAGLIFE  ight may be configured with all combinations of the above parameters    MAGLIFE flight is fitted with batteries as standard and can move with the patient  e g  to and  from the examination room and the adjacent preparation room      MAGLIFE  ight is desig
12. 0459       SCHILLER    MEDICAL S A S     MAGLIFE  ight    User manual    Version 1 2   October 2006       Part No  0 48 0080       The Art of Diagnostics    MAGLIFE light    Revision history    Versi  n   mte   Commens  June 2006 First edition    July 2006 First release  Soft 3  October 2006       Note  Distribution and maintenance information    SCHILLER has an international network of customer service  sales and consulting  agencies  For details about your local representative  please contact the SCHILLER  subsidiary near you  You will find a complete list of all the representatives and  subsidiaries on our website at http   www schiller ch   Sales information is also available from sales schiller ch    Manufacturer   SCHILLER Medical S A S  Tel   33  0  3 88 63 36 00  4  rue Louis Pasteur Fax   33  0  3 88 94 12 82  F  67162 Wissembourg Email  tech support schiller fr    Website  www schiller fr    Registered office    SCHILLER AG Tel   41  0  41 766 42 42  Altgasse 68 Fax   41  0  41 761 08 80  CH 6341 Baar  Switzerland Email  sales schiller ch    Website  www schiller ch    Part number  0 48 0080 Page 2    MAGLIFE light    WARNINGS    This manual shall be deemed to be an integral part of the  described unit     Compliance with its content is a prerequisite for proper device  performance and for patients and operators safety      The manufacturer disclaims all responsibility for the safety   reliability and performance of the device if       assembly  extensions  adjustme
13. 5  20 220   MAP 25 235       Part number  0 48 0080 Page 45    MAGLIFE light    6  Cleaning    MAGLIFE light can be cleaned with common cleaning and disinfecting agents  BURATON   INCIDIN  GG  KORSOLIN or LYSO FORMIN 2000      Follow the manufacturer s instructions for use    Switch the device off and disconnect from the mains before cleaning    Do not remove any covers    If any liquid does penetrate into the device  the device must be cleaned completely and  inspected    Do not expose the device to temperatures above 50   C     The use of an autoclave is not permitted     The device and its accessories need to be decontaminated  risk of pathological contamination   before disposal     The batteries of MAGLIFE light must be disposed of using a special procedure and not merely  scrapped     Part number  0 48 0080 Page 46    MAGLIFE light    7  Troubleshooting    7 1 General errors    Check and correct the  connections   LED  3  must  be on  The screen does not activate Battery discharged Connect the mains   LED  when the device is switched on  4  flashes  e Magnetic field too great e Place the device outside the  e Device faulty field   lt 40mT   e Replace the device    leal E O Acc   device initialise sequence is not  e Device faulty e Replace the device   carried out   The device shuts off e Battery discharged e Connect the mains     automatically e Magnetic field too great o er the device outside the  le    e Non specific printer e Replace the printer with a  printer speci
14. LIFE light be installed in the area around the magnet  where the magnetic leakage field is less than or equal to 40 mT  400 Gauss    see typical installation drawings and labels on the device      Part number  0 48 0080 Page 5    MAGLIFE light    Note     Note     Caution     Caution     Caution     Caution     Caution     Caution     Caution     The ambient temperature must be located within the following limits   15  C  lt  Amb  T  lt  35  C  60  F  lt  Amb  T  lt  96  F    The instructions specific to the installation and use of the MAGLIFE light  monitor have been provided in section 4 of this manual  which must be read  carefully before initial use of the monitor     The safety instructions relating to the MRI environment must be followed at all  times  during installation  use  repairs etc    Any work on MAGLIFE light  e g   repairs  must be done outside the area where the risk of magnetic attraction is  non existent     No equipment that hampers blood flow  e g  use of a cuff type  sphygmomanometer  may be used on the limb of a patient undergoing oxygen  saturation monitoring  SpO2   as it could disrupt the correct determination of  the data     Sensors and probes may not come in contact with conductive parts  including  the earth     Do not defibrillate on the accessories  Place the defibrillation electrodes as far  away from the other accessories as possible     The monitor is adapted for the MRI configuration upon power up  Any  subsequent change in the MRI system c
15. LLER representative     MAGLIFE  ight bears the CE 0459 mark in accordance with directive 93 42 EEC relating to medical devices  and fulfils the essential requirements of annex   of that directive     This product complies with the electromagnetic immunity requirements of standard EN 60601 1 2   Electromagnetic Compatibility Medical Electrical Equipment         The radio interference emitted by this device is within the limits specified in the standard EN 55011  class B   When the device is connected to a printer  it is within the limits specified in the standard EN 55011     class A     Part number  0 48 0080 Page 4    MAGLIFE light    Please read and adhere to the following list of warnings  precautions and notes  Some of them have been    WARNINGS  PRECAUTIONS AND NOTES    repeated at appropriate areas throughout this manual     Follow the warnings and precautions stated on the labels affixed on the monitor  Throughout this manual we    have added specific notifications that offer additional information  These notations are designated as     A Note is provided when extra general information is applicable     A Caution is provided when special care is to be exercised by the user and or patient  to avoid injury to the    patient  damage to the device or damage to other property     A Warning is provided when actions may result in a serious outcome  i e  injury  serious adverse effect     death  to the patient or user     Caution     Caution     Warning     Warning     Warnin
16. an alter the performance of the  MAGLIFE  ight monitor  Inform SCHILLER of any such change     The device is designed to operate with electrosurgical devices  Place the  accessories as far away from the electrodes of the electrosurgical device   That device must be installed and used in accordance with the manufacturer s  instructions     The device is designed to operate without any equipotential connection     Part number  0 48 0080 Page 6    MAGLIFE light    Contents   Page   Revision NISTON se 2  WARNINGS osia EO E E EEE EEA RE 3  TOIAN OMG iaa ii 3  WARNINGS  PRECAUTIONS AND NOTES ooonccocccccconnnccnnnnnncconnnnccnnonnnnnnnnnnonennnnnnnnnnnnrnnnnnnrennnnnrrnnannenenzns 5  nn elec aaa a aaa ENa a a E EE 7  1  Description of MAGLIFE GME  icsi eaa aaa E aaa a ed lse le 8  2  Location and description of displays and controls  ne 9  2 1 Frontor MAG ERE Meta dai 9  22 Re     9  Zo Mains  POW Cl TI 9  3  Devicessymbolidenilication ae 10  3 1 MAGLE ERE SVIMDOIS errre que saseumensmedevesnnswa patents vioastaanteemadinecauoumialaneswinacteepenend delicada 10  3 2 J Wai 10  4  USC 0t MAGUIEE MOP ese ssivenosececsveszeaveandoxss o 11  4 1 A eines ual EEE NES arte celal tenet act x Magne ata telat ce AT Bagatelles 11  4 2 DR 13  4 2 1 SEO OP COMO Ly POR ORCATOS OO TE BESAD EGE TEE RENSE TESS S 13  4 2 2 NIBP OPUOR ON aaben ad 13  4 2 3 SEOs and NTE Prop ON 14  4 3 Placing of probes amd Sensors ri iii date dedo ii 14  4 3 1 So 14  4 3 2 Probes and sensors to be used and positioning 
17. ast 75 cl and a pressure measuring device   If required  add an inflating bulb with a deflating valve  Activate Calibration   the display on the screen will show the pressure value measured by the  NIBP module  Make several comparisons between the measurements by  MAGLIFE light and those by the external measuring device over the entire  measurement range     Note  Calibration is regulated in some countries  Comply with  applicable laws and regulations     Note  The Calibration submenu is locked by a password   The password is communicated to the people who followed  the technical training     Myx     111   Patient type     A B C D E F G H   I   J    Ax  c m njo   a r s T u   Guy iw x Y Z          Alarms limit jop1f2 3 4 5  6 7 8 9        Calibration       Page 22    MAGLIFE light    4 7 1 2 Oximeter  SpO2  submenu    Windows displayed      with the SpO2 parameter disabled  select SpO2 On Off to enable            with the SpO2 parameter enabled  select SpO2 On Off to disable     On f Off  Average  Patient type    Alarms limit    Exit       e Select the averaging time to be taken into account while calculating  the SpO2 value and the pulse rate  The choice between the two values    8 s and 16s   is made by activating the Average tab     On    Off  Average  Patient type    Alarms limit    Exit       e Patient type selection  The patient type selection only affects the  thresholds     On   Off  Average    Patient type Neonate    Child  Alarms limit Adult    Exit       Part number
18. comes out of the standby  mode  the threshold values are the default values for the type of patient    selected  Adult  Child or Neonate        See section 4 7 1 3 1 Power up and standby mode    Three different sets of default values are available for each type of patient   These values can be set from the Device configuration menu  see    section 4 7 2      Part number  0 48 0080 Page 25    MAGLIFE light    The Operator 1 choice gives access to the following submenu  which is  used to set the thresholds from the values selected by the operator when  the function was last accessed     Alarms limit Adult Operator 1    spO  Pulse A5 75 p min        50 99 di  NIBP SYS 30 155 mmHg    DIA 15 220  to  MAP 20 235 mmHg      Exit   Quick     Default     gt             e The Quick selection is used to set all the values in relation to the  measured physiological values of the patient  The upper values are  set to 20  above and the lower values are set to 20  below those  physiological values    e The Default selection is used to go back to the default values for  the selected patient type    e  gt   provides access to the following submenu     Alarms limit Adult Operator 1    spO  Pulse A5 75 p min    Lo 50 99    NIBP SYS 30 455 mmHg    DIA 15 220  to  MAP 20 235 mmHg    OFF   Operator 2   Exit   a         Is used to go back to the previous submenu  e OFF is used to disable all the audio and visual physiological  alarms    Part number  0 48 0080 Page 26    MAGLIFE light    Caution  Fai
19. device is connected to the mains  whether or not it is operating  While the  battery is charging  indicator  4  flashes  Once the battery is fully charged  it stays on  and steady    MAGLIFE  ight can operate on the battery  new battery  fully charged  for two hours   one hour if intensive use is made of the NIBP function     An alarm message is displayed about 5 to 10 minutes before the battery is completely  discharged and the automatic device shutdown occurs     Operating during a mains failure    MAGLIFE  ight manages its power supply automatically  If the mains supply fails or is  of poor quality  the battery will automatically start supplying power to the device  regardless of the duration of the disruption    Indicators  3  and  4  will go off and the mains present indicator  see section 4 2   and  the battery charge indicator will reflect the new situation in the Status zone after three  seconds     Part number  0 48 0080 Page 17    MAGLIFE light    4 6 Use of the menu    All the functions other than those accessible via the controls on the unit are selected  from the menus displayed on the screen     You can enter the main menu in two ways       by pressing rotary button  6  or      by pressing key  5      Caution  The displayed menus depend on the parameters defined as active in the  configuration  see paragraph 4 7 2       The rotary button is used to        confirm a choice in the menu  when the line is highlighted in blue        make a choice       switch to the 
20. digital entry mode by making the cursor flash       move from one selected field to another without making any change      select a status in a toggle function    Double clicking  two clicks in rapid succession  is used to exit any menu  lt has the  same function as key  5      Button rotation is used to       move in a menu      increase or decrease the value of an entry field    Caution      Menus disappear if no action is taken for more than 10 seconds     4 7 Description of menus    4 7 1 Main menu    The main menu is as follows     NIBP Menu    Alarm limits menu  BO display  Trend Menu    Edition menu    Settings menu       It is used to select submenus   Note  Only the labels corresponding to the installed parameters are displayed     The main menu contains the following submenus     Part number  0 48 0080 Page 18    MAGLIFE light    4 7 1 1 NIBP  Non Invasive Blood Pressure  menu    Windows displayed      with the NIBP parameter disabled  select NIBP On Off to enable     Calibration       e Cycle selection  to select a cycle value from 1 minute to 90 minutes  activate symbol  3    then select the appropriate value by rotating button  6  and confirm by  pressing the button     On   Off    a  Patient type    Alarms limit    Calibration       Part number  0 48 0080 Page 19    MAGLIFE light    e Patient type selection    Depending on the type of patient  the initial inflating pressure is  automatically selected  180mmHg for Adult Child and 120mmHg for  Neonate     Note  Fo
21. e  Connection  Class    SPO2 accuracy    SPO  display range   Pulse accuracy   Pulse display range   Blocking by defibrillator shock    HF protection    BCI  Fibre optic  CF      2   from 70 to 99      3   from 50 to 69      0     99    5 b min   30     250 b min   10 seconds maximum    protection from electro surgery devices    5 2 2 NIBP Non Invasive Blood Pressure    OEM module  Connection   Class   Measuring principle    Mode    Part number  0 48 0080    CAS   Fast snap  CF  Oscillometric    Manual  Automatic  Continuous    Page 44    MAGLIFE light    Types of patient Neonate  Child  Adult  Sensor accuracy   3mmHg or  2    Pulse accuracy 5 b min  Display range  Adult   Child  systolic  30   255 mmHg  diastolic  15  220 mmHg  mean  20   235 mmHg  Pulse   30   240 P min  Neonate  systolic  30   135 mmHg  diastolic  15 110 mmHg  mean  20   125 mmHg  Pulse   40   240 P min  RF protection protection from electro surgery devices    5 3 Threshold ranges    The upper and lower alarm thresholds are adjustable according to the values in the  table below   The lower limit setting may never be greater than the upper limit setting and vice versa     Parameter Lower limit Upper limit  Neonate patient    Pulse 30 245 35 250  Saturation 50 98 51 99  SYS 30 130 35 135    DIA 15 105 20 110    MAP 20 120 25 125  Child patient    Pulse 30 245 35 250  Saturation 50 98 51 99    SYS 30 250 35 255   DIA 15 115 20 220   MAP 20 230 25 235  Adult patient    Pulse 35 250  Saturation 91 99  SYS 35 25
22. ection class  of the enclosure    Electricity supply    Power consumption    Fuses    Battery  Autonomy    System specifications    SCHILLER Medical S A S   MAGLIFE  ight    2 0 x 216 x 116 mm  10 6    x 8 5    x 4 6     180 x 84 x 68 mm  7    x 3 3    x 2 7       6 Kg  1 3 Kg    IP 21    100  115  230 VAC  50 60Hz  Factory defined voltage    25 VA  2 x 100 mA  T    230 VAC  2 x 200 mA  T    100   115  VAC    12 V 2 Ah Lead  2 hours    Environmental conditions    Operating  Temperature    Relative humidity    Pressure   Magnetic field  Storage   Temperature    Relative humidity    Pressure  Magnetic field    Part number  0 48 0080    15   35  C  60  F   96  F  30     95  non condensing  500   1060 hPa    lt  40mT     10  C   50  C  13   124  F  30     95  non condensing  500   1060 hPa    lt  40mT    Page 43    MAGLIFE light    Display  Connections  Interfaces    Safety standards    EMC    CE marking    Protection class    Colour TFT screen  6 8     98 x 132mm  480 x 640 dots    SPO  and NIBP    RS232  USB 1 1 for connecting the equipment  specified by SCHILLER    IEC 60601 1    IEC 60601 1 2   CISPER 11 Class B    with printer connected  class A   The device may be subject to the following  interference without being affected    Static discharge up to 8 kV   Radio frequency 10 V m  80     2500 MHz  5 Hz  modulating     According to directive 93 42 EEC class Ilb    Class   according to IEC 60601 1    5 2 Technical specifications of modules    5 2 1 Pulse oximeter    OEM modul
23. fied by  SCHILLER  No printing e No power to printer e Connect the printer to the  e Connecting cable loose power source and switch on  e Check and remedy the  connections       Part number  0 48 0080 Page 47    MAGLIFE light    7 2 Errors of modules    Error    No SPO2 measurement    SPO2 signal disturbed    The cuff will not inflate    Doubtful NIBP measurements    Cause  Optical probe faulty  Ambient light artefacts    No perfusion    Module faulty     no light in the SPO2  connector    Incorrect sensor installation  Compression of the  measurement location  Poor perfusion   Pump motor magnetised   because the device has  been subject to an  excessive magnetic field  Faulty cuff or tube    Incorrect cuff installation  Inappropriate cuff    System not airtight    Remedy  Replace the probe  Cover the sensor holder on  the finger  Change the signal taking  location  Replace the device    Move the sensor  Remove the compression    Move the sensor  Replace the device    Replace and check  tightness   Check and adjust the cuff  Replace with an appropriate  cuff   Replace the leaking part    Not locatable       Replace the device    8  Maintenance  e Before each use       check all the device functions    simulate an alarm    check if all the connectors and cable insulators are in good condition    e Once a year       check the leakage current according to paragraph 19 of standard IEC 601 1     8 1 Alarm simulation    To simulate an alarm for verification  proceed as follows    e
24. g     Warning     Warning   Warning     Caution     Warning     Warning     MAGLIFE light is a monitor designed solely for use close to Magnetic  Resonance Imagers of 0 2 to 3 T for monitoring of patients undergoing an MRI  examination     The continuous presence of a qualified person is imperative throughout the  examination     Before using the MAGLIFE light monitor  follow the safety instructions below     e MAGLIFE light may only be used by trained healthcare workers who are  familiar with its instructions for use       Make sure that the voltage and frequency of the electricity system match  those stated on the identification plate     Because the device is a class   device  it may only be used in premises with  an electrical system comprising an earth connection     Connecting other devices or sensors to the patient could lead to leakage  currents that may be harmful to the patient  Consult SCHILLER before  interconnecting with other equipment     Take account of the physiological effect of the other devices connected to the  patient     This device may not be installed or operated in explosive environment   This device is not designed for use with inflammable anaesthetic agents     Use only sensors supplied by SCHILLER  They have been designed specially  for use in MRI environments     MAGLIFE light is designed for use with MRI systems with a magnetic field  between 0 2 and 3 Tesla  Contact SCHILLER for use beyond that magnetic  field range     It is imperative the MAG
25. g oxygen saturation measurement as it could disrupt  the correct determination of the data    Several cuff models are available  including     Adult cuff     Child cuff     Neonate cuff    Extension hoses are also available     Part number  0 48 0080 Page 16    MAGLIFE light    4 4    Start up    Connect the mains cord to the mains connector  19  and the low voltage connecting  cord to connectors  20  and  13   Indicator  3  will light up to show that the device is  connected to the mains     Caution  Indicator  4  will illuminate and the battery will be charge automatically  while the device is connected to the mains  even if it is not in operation     Press key  1   the associated indicator  2  will illuminate   After a few seconds  the screen will activate and the system initialised  The initialisation  sequence will last about 10 seconds  After that time  the parameters will be displayed     For battery operation  with the mains cord disconnected   just press key  1   the battery  is integrated in the device   If the battery is charged  associated indicator  2  illuminate  and the device will start up  With a fully charged battery  the device can operate for at  least two hours     Note  The operating of key  1  is disabled during the initialisation phase     4 5 Battery charge    4 5 1    MAGLIFE light may be powered by the internal battery to accompany the patient as the  patient is moved  or if the mains power is absent  The battery charges automatically  whenever the 
26. in       Quality of a  SPO2 signal Plethysmogram  SPO2 area F    60  Pulse rate value  SPO2 value    N       Pulse area  Alarm Thresholds    4 2 2 NIBP option only    Status area     a 29 06 2006 14 27          ADULTE  NIBP area NIBP values  Unit           9 POULS  pimin  Mode N 7 7  Previous measurements           BER 13 min       Piletas  14 24 125  80  94  9 2 14 24    Part number  0 48 0080 Page 13    MAGLIFE light    4 2 3 SPO  and NIBP options       SPO2   9  POULS p min A  100  SPO2 area _J9 8   7 0    a 60 29 06 2006  ER 725114 25  NIBP area  N          Status area  69 39  PNI mmHg          094    en  4 3 Placing of probes and sensors  Warning  The recommendations relate to the type of sensors and probes to be    used  the positioning of the sensors and probes on the patient and the  placing of the cables that connect the sensors and probes on the  patient and the device     The following rules must be followed precisely to avoid the problems described below     4 3 1 Absolute rule    Use only the cables  sensors and cuffs supplied by SCHILLER     4 3 2 Probes and sensors to be used and positioning instructions    4 3 2 1 SpO2 sensor    1  Make sure that the Oximeter function is on and that the oximeter  parameters are set correctly     Caution  Use only the sensors from the list of accessories supplied by  SCHILLER     2  Connect the oximeter patient cable to the oximeter connector and  give the connector a quarter turn to the right hand side to lock it in  place  to
27. instructions  et 14  4 3 2 1 SPOOZ A iaa 14  4 3 2 2 Sphygmomanometer cuffs  Non Invasive Blood Pressure   pe    16  4 4 A O eas ea cate aN ane E ESS R  RES A BER ETS 17  4 5 JU 17  4 5 1 Operans dunne mamis ii 17  4 6 U eor CMa encanta A A A RE EE 18  4 7 DESC POON OF Menis a it S 18  4 7 1 is o FEED EI PE ER FEED E OEE SER  E E SETE V VEDDET 18  4 7 1 1 NIBP  Non Invasive Blood Pressure  menu  pp 19  4 7 1 2 Oximeter Sp OF SED MAC TAU on ns mene Re naa lain arom cen tee valse aes aaruiaeen sl Meiamnee  23  4 7 1 3     ES 25  4 7 1 3 1 Power up and standby Mode     27  4 7 1 4 Niasnete ed 28  4 7 1 5 Trends Ml dolo 28  4 7 1 5 1 Frend erapa Submit 29  4 7 1 5 2 Trend table UD Neco cccrcuecadice spececicetasdceaceeacevene tedden scatterer cd essens lad Salen dadtanersdibauntsteracdeest 30  4 7 1 6 de 31  4 7 1 7 SSL HEB SR 32  4 7 1 7 1 Sound Soc Deia pisito pilar on EEE BESES TERESE SER 32  4 7 1 7 2 ARA A A TES ERNE GS ER ESENER ERE DETS 33  4 7 1 7 3 Screen Driohthess setin Tienie e a EE ce dehctuse Weksaterushes dean ace uuels 33  4 7 2 DENCE CONT SUT ALON A eS 34  4 7 2 1 SO 34  4 7 2 2 UA OR 35  4 7 2 3 Opos SUDINCHU iia  sd adas 37    Part number  0 48 0080 Page 7    MAGLIFE light    4 7 2 4  4 7 2 5  4 7 2 6  4 8  4 8 1  4 8 2  4 9  4 9 1  4 9 2  4 9 3    5 1  3 2  5 2 1  I2  5 3    7 1  1 2    8 1    10   10 1  10 2   11    12     SOtLWare 1pdate Su DML de 37                     38  Reserved SCHILLER SUD e ici 38   SE UCTUNE Ol Menus sica ip ira 39  Eer MEAS pi a EAA
28. is exposed to a force of attraction  due to the magnetic field  that  rises rapidly as the distance from the magnet decreases  This area must  be indicated by markings on the floor     Caution     Important note     Part number  0 48 0080    The monitor may be installed on a non magnetic trolley supplied as  an option by SCHILLER  The trolley has wheels for turning and  moving the monitor  If the minimum distance is not kept  the  performance of the device may be disrupted  The trolley bearing  the monitor may not be placed in or moved to an area where the  field is greater than 40 mT  400G   regardless of whether the  monitor is on or off  The force of attraction of the magnet could pull  the monitor against the walls of the tunnel    The wheels of the trolley can be locked when the trolley is installed  in the permitted area     The magnetic field is always present  even when the imager is not  being used for an examination     Page 11    MAGLIFE light    Typical plan of an MRI system with a MAGLIFE  ight monitor    1  Faraday cage   2  MAGLIFE fight monitor   3  Mains connector   4  Marking on the floor  40 mT line     40 mT 400G  line    Forbidden area        Authorised area           Figure 1    Part number  0 48 0080 Page 12    MAGLIFE light    4 2 Display screens    4 2 1 SPO  option only    Patient type  Time    Miscellaneous  Date    j    Batterv status       S Sound alarm  status   29 06 2006 14 26 DENG  O   a ADULTE   Status area        SPO2   POULS    99       p m
29. lure to monitor the condition of the patient may lead to  death     e Selection Operator 2 provides access to the second set of values  defined by the operator when the function was last accessed  The  values can then be edited    e  Exitis used to close this submenu     4 7 1 3 1 Power up and standby mode    When the device is powered up  the threshold values are the default values  for the type of patient selected  and therefore not necessarily the values  used when the device was last used for monitoring   The default thresholds  are factory set and can be adapted from the configuration menu   Patient  type selection is offered for 12 seconds after device power up and also  when the device comes out of standby mode  If the operator does not  select the patient type  the default values of the neonatal type are selected  automatically  The possibility to select the values used last is available with  the Previous button     NEONATE       Patient      in a Child    Note  Selecting the Previous option makes it possible to obtain  trend continuity  as all the other selections result in the  insertion of a separating mark between patients in the trends     Part number  0 48 0080 Page 27    MAGLIFE light    4 7 1 4 Magnetic field display    The BO display button is used to display the magnetic field measurement  on the screen  Values Bx  By and Bz show the value of the magnetic field  in mT along the three orthogonal directions in space  The B value shows  the corresponding vecto
30. ned to be placed on a non magnetic mobile stand     Part number  0 48 0080 Page 8    MAGLIFE light    2  Location and description of displays and controls    2 1 Front of MAGLIFE light     see photo in section 11      1     2     10     11     12     Main device on off key   Device    ON    indicator      MAINS    connection indicator        BATTERY    Indicator  showing that the battery is being charged  flashing   or is fully charged   steady      Main    MENU    key  Access to the main menu and for exiting a menu from any other place   Navigation button for menu selection   Display for curves  parameters  menus and messages        NIBP    key  for starting up or stopping a measurement or a series of measurements of the  blood pressure via a cuff        Alarm silence    key  enables or disables the two minute or permanent audio alarm function   physiological and technical alarms      Device    STANDBY    key   SPO optical oximeter sensor connector     NIBP cuff connector     2 2 Rear of MAGLIFE light     see photo in section 12      13    14     15     16     17     18     Connector for the low voltage cord for connecting to the Mains power unit   Loudspeaker   USB connector   RS 232 connector   Auxiliary connector compartment cover     Identification label     2 3 Mains power unit   see photo in section 12      19     20     Mains plug     Low voltage plug for connecting the MAGLIFE  ight     Part number  0 48 0080 Page 9    MAGLIFE light    3  Device symbol identification
31. ntinuous mode     Caution  To ensure the correct vascularization of the member on whom    the blood pressure is measured  the repetition of the  continuous mode is not advised     Is used to select the frequency of  Cycle after CONT mode   automatic measurements after the end    of the continuous mode     Caution  These configuration settings must be in accordance with the    laws applicable in the country of use     Page 36    MAGLIFE light    4 7 2 3 Options submenu    The Options menu is used to enable or disable one or both parameters of  MAGLIFE  ight      5p02 SM    Non Invasive Blood Pressure das       4 7 2 4 Software update submenu    The person performing the update must have the requisite skills to make  the relevant functional and safety checks and is entirely responsible for  such checking    The software is updated by means of a USB key  standard 1 1 or above   Make sure that the software on the USB key in the root directory  only that  software  is compatible with the device you want to update     Adresse  ES G  maglight update          Dossiers x Nom   Taille Type Date de modification     S Disque amovible  G  A El analog  bin   20Ko Fichier BIN 20 12 2005 10 47  2 E  madlight El  boot  script 1 Ko Fichier 21 12 2005 18 22  a update ES  dear  boot  script 1 Ko Fichier 21 12 2005 18 22  E  hardware compatibility 1 Ko Fichier 21 12 2005 18 22   E  kernel  bin 1401 Ko Fichier BIN 21 12 2005 18 05  COScHILLER  tar 1670Ko Winzip File 21 12 2005 18 23      soft 1 Ko
32. nts  modifications and  repairs are not performed by the manufacturer or by  manufacturer authorised persons       the electrical installation of the premises does not comply  with locally applicable requirements       the device is not used in compliance with this instructions  for use     This manual describes the device at the printing time     Upon request  the supplier will provide circuit diagrams  lists of  components  descriptions  calibration instructions or any other  information required by the user   s qualified technical personnel  to repair those parts of the device that have been stated as     repairable    by the device manufacturer  The supply of such  information shall not in any event constitute permission or  approval to modify or repair a device     All rights reserved for the devices  circuits  processes and  names appearing in this manual     The device is not designed for any use that is not specifically  provided in this manual  which may be hazardous     Foreword  Part number  0 48 0080 Page 3    MAGLIFE light    This MAGLIFE light manual provides the information required for proper device performance     Knowledge of monitoring and the understanding of the characteristics and functions of MAGLIFE light are  required for proper use of the device     Do not use the monitor before you read these instructions     Device maintenance information is provided in the service manual of the MAGLIFE light  For more  information  please contact your nearest SCHI
33. ou cannot define a customised colour  combination    e The allowed patient selection button is used to  restrict or extend the patient selection list           Part number  0 48 0080 Page 35    MAGLIFE light    Button  2 is used for special alarm configuration  It opens the    following submenu      4 4 SpO2 alarm    Physiological alarms    2 alarm rejection at start    Fy       4   SpO2 alarm Enabling or disabling the alarm function if the SPO  drops    more than 4      Physiological alarms A physiological alarm may be latched  i e  confirmation by    2 alarm rejection at start    E    Part number  0 48 0080    pressing key  9  is required to stop it  even when the value of  the parameter has exceeded the threshold and is back in the  non alarm range  If the alarm is not latched  it stops as soon  as the value is back within the non alarm range     This option is used to disable audio alarms for two minutes  after the device is switched on  allowing for time required to  apply the sensors for the measured values to stabilise     This button is used to permit or prevent the ability to mute  audible alarms permanently with key  9   lf the alarms are  permanently muted  a reminding audio signal  double beep  is  emitted after every two minutes     Button  NIBP is used to define the parameters of the NIBP    measurement in the continuous mode       CONT mode length    Cycle after CONT mode         lect the duration of th  CONT mode length S used to select the duration of the  co
34. r modulus                4 7 1 5 Trends menu    The Trends menu contains two main submenus   Graphs and Tables   and    two options  lt enables the user to erase the trends and select the period  for trend tables     Sampling rate       Part number  0 48 0080 Page 28    MAGLIFE light    4 7 1 5 1 Trend graph submenu    Part number  0 48 0080    pa a pa e    The Trend graph menu is used to select the trend curves to be displayed  and contains the following tabs        Note  Only two waveforms can be shown at the same time     Pulse f SpO2    pimin th    60       50    10 46 00 10 53 25    BLU GR     0      10 08 30 10 16 00 10 23 30 10 31 00 10 38 30       Navigation buttons    Back one page    Forward one page    Reduce time line from 24 hours to 1 hour    Increase time line from 1 hour to 24 hours    Adapt graph scale    Exit Exit the trend graphs display    Page 29    MAGLIFE light    4 7 1 5 2 Trend table submenu    Part number  0 48 0080    Trend table show the numerical values of the physiological parameters  according to the frequency that has been defined earlier from the trend  type selection menu        Navigation buttons    Forward to the end of the table  most recent value    Forward to next line   Back to previous line   Back to the start of the table  oldest value    Exit the trend tables display   If a non invasive pressure measurement does not strictly    follow the table frequency rules  it is inserted at the  appropriate location     Page 30    MAGLIFE light 
35. r subsequent measurements  the inflating pressure is automatically  selected  about 50mmHg above the measured systolic pressure for  Adult Child and 30mmHg for Neonate     Patient type Neonate  Child  Unit   Adult    Alarms limit    Calibration       e Measuring unit selection  The Unit button is used to switch between the two units   kPa and mmHg     On   Off  a  a  Patient type  Unit    Alarms limit    Calibration       Part number  0 48 0080 Page 20    MAGLIFE light    e Use the Alarm limits button to manage thresholds  For more details  refer  to section 4 7 1 3 Alarm limits and 4 9 Alarms    On f Off    D    Patient type    Unit  Alarms limit    Calibration       e NIBP alarm limit setting  Select a limit by rotating the navigation knob  6   press to confirm    High SYS    Low SYS    High DIA  Low DIA  High MAP    Low MAP       Select the limit value by rotating the navigation knob  6  again  press to  confirm the selected value     High SYS  Low SYS  High DIA    Low DIA    High MAP    Low MAP       Note  At the switch on time  the alarm levels corresponds to the standard values  of the selected patient  The operator selection is done in general alarm limit  management window refer to section 4 7 1 3 Alarm limits    Part number  0 48 0080 Page 21    MAGLIFE light    Part number  0 48 0080    The Calibration submenu is used to check if the NIBP pressure module is  measuring the pressure correctly  Connect the external verification system  with an expansion vessel of at le
36. that the NIBP  parameters are set correctly     2  Connect the NIBP tube connector to the NIBP measurement  connector  12   Use only the cuffs listed in section 8     Warning  If the cuff is placed too tightly on the limb  the results will be  excessively high readings  Among other considerations  the use of  a cuff suited to the patient has a direct bearing on the accuracy of  the NIBP measurements obtained  Select the cuff according to the  circumference of the patient s limb     Warning  Cuffs can soften over time and get permanent creases that may  mark the limb temporarily  Replace any cuffs which have that  characteristic     Caution  Make sure that the pressure connectors are not flattened or  blocked   Caution  Position the cuff slightly above the elbow  the Velcro    turned up     Since the cuff does not contain a microphone  specific positioning  is not necessary     Warning  The cuff must be tightened around the arm  but must not apply  pressure on the blood vessels before the measurement  Roll the  cuff around the arm and close the Velcro        Caution  Check to ensure the hose is not tangled  check any incorrect  placement of the arms  legs or body of the patient  any clamps  etc    as device operation will be disrupted by a flattened or  restricted hose     Warning  Check that the blood is flowing correctly in the relevant limb  arm   leg   except during the measurement      Caution  A cuff type sohygmomanometer must not be used on the patient s  limb undergoin
37. the device is switched  on and keeping it pressed until the configuration menu is displayed     Alarm limits    e    Releases  Reserved SCHILLER    Device OFF       Note  To exit the Device configuration mode  the device must be  switched off     4 7 2 1 Alarm limits submenu    The default thresholds are set for the three type of patients   Adult  Child  and Neonate  Select one type of patient  press button  6  and make the  setting in the following tab    Neonat  Child    Adult    Magnetic Field       Select the parameter by navigation button  6  counter clockwise and  confirm by pressing button  6   The first value will flash  Turn as required to  modify the value  Save the value by pressing button  6   The next value is  automatically selected  When the last parameter value has been entered   the selection cursor goes back to the parameter selection column     Part number  0 48 0080 Page 34    MAGLIFE light    Alarms limit Neonate Default    spO  Pulse 90 200 p min      85 100      Ne SYS 49 140  o  DIA 30 100    o  MAP 39 120  o       4 7 2 2 Monitor submenu    Note  The Language  Colour  Alarms  NIBP and Patients  selection list buttons are used for settings  whilst the Serial  number and Hardware number buttons are provided for  information only and cannot be modified by the user     e A certain number of languages are available  The  language selection does not affect the operating of  MAGLIFE  ight    e The screen colour combination can be selected with  button      Y
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Permabond 217  Samsung MCM-A202A Felhasználói kézikönyv (XP)  BUTLER E400/TWIN/TRIPLE/QUATTRO  liens authentiques  Burnham PVG Boiler User Manual  2015 Sistema Infotainment Cadillac CUE    Attagraph User Manual  PDF形式 - 柏市入札情報  Hilfe zum Hauptmenü    Copyright © All rights reserved. 
   Failed to retrieve file