        Ordering Specifications Esco Hospital Pharmacy Isolator – Negative
         Contents
1.   Sufficient clearance of 40 cm  minimum  is recommended between the highest point of the isolator and the    ceiling  If the distance is less than 40 cm  the airflow alarm system may require calibration  Exhaust filter area  especially susceptible to disruptive air currents or air drafts       For purposes of exhaust filter leak scanning  a minimum distance of 50 cm between the highest point of the    isolator and the ceiling is recommended       Adequate space is to be allocated for cleaning behind the isolator and for performing decontamination    procedure       Afar away position from any source of heat sources  heaters  fan converters  etc   for optimum operating    conditions     Global Offices   Kuala Lumpur  Malaysia   Leiden  The Netherlands   Manama  Bahrain  Mumbai  India   Philadelphia  USA   Salisbury  UK   Shanghai  China   Singapore    ESCO    WORLD CLASS  WORLDWIDE     Worldwide Headquarters    Esco Micro Pte Ltd e 21 Changi South Street 1 e Singapore 486777 ROC No  198400165W  Phone  65 6542 0833 e Fax  65 6542 6920 e mail escoglobal com   www escoglobal com    Section X  Site Requirements  Environmental and Electrical Conditions     This isolator shall be designed for operation under the following environment and electrical conditions     ue    a WwW N    Indoor use      Altitudes up to 2000 meters or 6600 ft    20    60  relative humidity      Temperature range from 20 deg C to 30 deg C  68 deg F to 86 deg F   The temperature in the laboratory is to    be 
2.   airflow within work zone and pass thru to enable recovery of chamber  atmosphere to ISO Class 3 conditions within 3 minutes following a worse case contamination event   d  Airlock pass thru to ensure work zone remains sterile during insertion and removal of items   e  Safe glove changing process with zero risk of contamination     4  The isolator shall meet or exceed all applicable safety requirements and also be designed to protect the  product within from airborne contaminants  Work zone and pass thru shall be under negative pressure to the  room in order to maintain operator protection in case of a breach in the barrier isolation system     5  The manufacturer shall individually test every isolator before shipment  A test report shall be provided to the  customer with a duplicate copy permanently attached to the isolator  Ata minimum  the following tests shall be  performed     Filter Leak   Downflow Velocity   Dynamic Pressure   Particle Count or Air Cleanliness   Product Ingress and Egress   Recovery Time   Gauntlet Breach   Operator Comfort  include noise  light and vibration     samoa0o9    6  On request  a representative of the purchaser may witness all tests via a    Factory Acceptance Testing     program  Charge applies     7  All 230VAC 50HZ models shall be CE marked     8  The manufacturer must have a quality management system that is certified to ISO 9001  and an  environmental management system that is certified to ISO 14001     Global Offices   Singapore   Ph
3.  Headquarters    Esco Micro Pte Ltd e 21 Changi South Street 1 e Singapore 486777 ROC No  198400165W  Phone  65 6542 0833 e Fax  65 6542 6920 e mail escoglobal com   www escoglobal com    Section IV  Isolator Control and Alarm System    1     The isolator shall have a microprocessor incorporated to supervise all isolator functions and monitors  downflow air  work zone and pass thru pressures in real time       The isolator shall be equipped with a soft touch keypad and continuously back lit LCD  4 line display  for    controlling the operation of the fan  light  electrical outlet and microprocessor menu  The LCD shall be able  to display time  main chamber  work zone  pressure  pass thru pressure  airflow  status messages and to  provide access to Menu function       Airflow shall be measured and displayed with a thermistor based true velocity airflow sensor mounted on the    isolator       An optional  audible and visual  alarm package shall be available on request for monitoring for any unsafe    conditions  airflow  window position  and hardware errors        The microprocessor shall be incorporated with the following functions     Clock time setting that will not be lost even when the isolator has been switched off   Warm up timer to set time for stabilizing the sensors  fan and control system   Post purge time to set time for purging all residual contaminants in the work zone   Inner door timeout to ensure the inner door has not been left opened for more than a certain p
4.  accessible from outside the isolator which may lead to environmental  contamination in case seals fail  are not acceptable       The downflow filter shall be angled  matching the slope of the isolator main body   so as to increase airflow    uniformity over the work zone     Global Offices   Kuala Lumpur  Malaysia   Leiden  The Netherlands   Manama  Bahrain  Mumbai  India   Philadelphia  USA   Salisbury  UK   Shanghai  China   Singapore    ESCO    WORLD CLASS  WORLDWIDE        Worldwide Headquarters    Esco Micro Pte Ltd e 21 Changi South Street 1 e Singapore 486777 ROC No  198400165W  Phone  65 6542 0833 e Fax  65 6542 6920 e mail escoglobal com   www escoglobal com    Section Ill  Fan System    1  The isolator shall be equipped with a permanently lubricated  direct drive  motorized impeller fan with  external rotor designs     2  The motor shall be equipped with integrated blades to narrow profile and to eliminate the need for a motor  shaft     3  All rotating parts of the fan shall be unitized and balanced for smooth  quiet  vibration free operations     4  The motor shall be designed for maximum energy efficiency and minimal maintenance  It shall also possess a  compact design and flat profile with a completely integrated assembly to optimized motor cooling     Global Offices   Kuala Lumpur  Malaysia   Leiden  The Netherlands   Manama  Bahrain  Mumbai  India   Philadelphia  USA   Salisbury  UK   Shanghai  China   Singapore    ESCO    WORLD CLASS  WORLDWIDE     Worldwide
5. ESCO    WORLD CLASS  WORLDWIDE     Worldwide Headquarters    Esco Micro Pte Ltd e 21 Changi South Street 1 e Singapore 486777 ROC No  198400165W  Phone  65 6542 0833 e Fax  65 6542 6920 e mail escoglobal com   www escoglobal com    Ordering Specifications  Esco Hospital Pharmacy Isolator     Negative Pressure  Recirculating  HPI 6N_     Section I  General Requirements  1  The isolator shall meet the performance requirements of the following     a  CETACAG 001 2005  USA   b  CETACAG 002 2006  USA  c  USP CHAPTER 797  USA     2  The isolator shall comply with the following international air cleanliness standards     a  ISO Class 3 as per 1SO14644 1  equivalent to Class 1 per US Federal Standard 209E   b  JIS Class 3 as per JIS B9920  Japan     3  The isolator shall provide a sterile negative pressure work environment for the compounding of hazardous  materials  antineoplastic  or cytotoxic compounding applications  It shall also provide a more inherently  superior sterility compared to open front clean air devices such as laminar flow clean benches and biological  safety cabinets  The isolator shall possess the following features for the purpose of maximum patient  protection and sterility     a  Acombination of downflow and exhaust ULPA filters to provide a fully integrated envelope for product  protection in all configurations   b  ULPA filters with a typical efficiency of 99 999  at 0 1 to 0 3 microns provide superior ISO Class 3 air  cleanliness   c  Laminar  unidirectional
6. arters    Esco Micro Pte Ltd e 21 Changi South Street 1 e Singapore 486777 ROC No  198400165W  Phone  65 6542 0833 e Fax  65 6542 6920 e mail escoglobal com   www escoglobal com    Section VI  General Construction   User Features    1  The isolator shall have the following dimensions     Dimensions 6ft   1 8m Isolator  Exterior  without base stand  1950 x 820 x 1320 mm  76 8    x 32 3    x 51 9        Main chamber  work zone  1145 x 610 x 670 mm  45 0    x 24 0    x 26 4        Pass through 355 x 610 x 670 mm  13 9    x 24 0    x 26 4           2  The isolator shall be equipped with 2 electrical outlets  mounted on the right hand rear wall  For 115VAC 60HZ  isolator these outlet s  shall be GFCls  conventional non GFCI outlets are not acceptable for safety reasons      3  The main body of the isolator shall be constructed with 1 2mm  0 05     18 gauge electro galvanized steel with  ovenbaked epoxy antimicrobial powder coated finish    4  Abrasion and corrosion resistant pharmaceutical grade 304 stainless steel shall be used for the back   side  walls  work surface  and drain spillage trough in the isolator  Other grades  which may possess inferior  corrosion resistance properties  shall not be acceptable     5  The internal work zone liner  side and back walls  shall be constructed of a single piece of stainless steel  with  large radius corners for easy cleaning  Three piece construction  left side wall  back wall  right side wall  shall  not be acceptable as joints are diff
7. as they exhibit poor energy efficiency and high heat dissipation characteristics       The touch control keypad and isolator LCD display shall be mounted in the centre  and sloped downwards     Alternative designs  which may not be clearly visible from a seated position  shall not be acceptable       Unlike conventional circular ports  the gloveports shall be oval shaped to provide maximum comfort to the    operator  The operator shall be able to use the gloves of his her preference       Optional hydraulic stand shall be available on request for enabling operating to adjust the work surface height    to his her height  for both sitting and standing operation     Global Offices   Kuala Lumpur  Malaysia   Leiden  The Netherlands   Manama  Bahrain  Mumbai  India   Philadelphia  USA   Salisbury  UK   Shanghai  China   Singapore    ESCO    WORLD CLASS  WORLDWIDE     Worldwide Headquarters    Esco Micro Pte Ltd e 21 Changi South Street 1 e Singapore 486777 ROC No  198400165W  Phone  65 6542 0833 e Fax  65 6542 6920 e mail escoglobal com   www escoglobal com    Section VIII  Installation and Service    1   2     10     It shall be possible to lift the isolator directly off the pallet from the truck onto the support stand  if ordered    The depth of the isolator shall be under 850mm without any disassembly  to enable it to pass through narrow  spaces      The electrical circuit diagram as well as original factory test report shall be permanently attached to the  isolator  in additio
8. eriod  Measurement unit selection between English and Metric    oao00M      The microprocessor shall enable restrict access to the main menu with password protected administration     Such restriction shall have control over the fan  blower hour meter  and resetting of default values for  parameters such as warm up period  post purge period  PIN for administration and Pin for fan control       It shall be possible for the microprocessor to be set in Maintenance mode for the following purposes     a  Calibration Performance validation  b  Filter scanning  c  Checking of input output diagnostics    Global Offices   Kuala Lumpur  Malaysia   Leiden  The Netherlands   Manama  Bahrain  Mumbai  India   Philadelphia  USA   Salisbury  UK   Shanghai  China   Singapore    ESCO    WORLD CLASS  WORLDWIDE     Worldwide Headquarters    Esco Micro Pte Ltd e 21 Changi South Street 1 e Singapore 486777 ROC No  198400165W  Phone  65 6542 0833 e Fax  65 6542 6920 e mail escoglobal com   www escoglobal com    Section V  Safety Features    1      All panels leading to potentially contaminated and   or dangerous areas shall be colour coded red  Warning    labels alone shall not be acceptable       The blower plenum shall be a permanent design constructed of steel  Cloth or bag based plenums  which are    subject to mechanical failure  shall not be used       A removable  perforated metal face guard shall be installed below the downflow filter to ensure airflow    uniformity inside the work zone as 
9. icult to clean     6  The work surface shall be constructed of a single piece of stainless steel for easy cleaning  with raised edges  on all sides to contain spillages  Flat or perforated work surfaces are not acceptable     7  The drain pan  located below the work surface  shall have wide  open corner radii  The four sides of the drain  pan shall also be angled towards the center of the isolator  The left and right side of the drain pan shall  be directly sealed to the work zone liner to close off the potentially contaminated space behind the side wall   Alternative designs  which exhibit poor cleanability  shall not be acceptable     8  The back return air grille shall be integrated with the back wall  to ensure maximum usable working space   With the exception of the front air grille  there shall be no other air slots   perforations on the work surface     9  All welds shall be ground flush and smooth  There shall be no welded joints to collect contaminants or rust     10  It shall be possible to clean all areas of the isolator  including behind the front window  without any  disassembly     11  The manufacturer shall be able to provide evidence that all surfaces  joints and welds have been type tested  for cleanability     12  The manufacturer shall be able to provide evidence that the isolator can be decontaminated using a gas   phase agent such as formaldehyde and hydrogen peroxide     13  Acomprehensive user and service manual shall be provided with the isolator     G
10. iladelphia  USA   Leiden  The Netherlands   Salisbury  UK  Kuala Lumpur  Malaysia   Beijing  Shanghai  amp  Guangzhou  China   Mumbai  India   Manama  Bahrain    ESCO    WORLD CLASS  WORLDWIDE     Worldwide Headquarters    Esco Micro Pte Ltd e 21 Changi South Street 1 e Singapore 486777 ROC No  198400165W  Phone  65 6542 0833 e Fax  65 6542 6920 e mail escoglobal com   www escoglobal com    Section Il  Filtration System    iF    The isolator shall be equipped with the following particulate filters     a  One supply filter to provides clean air to the work surface in a gentle vertical laminar flow  b  One intake filter to ensure ensure recirculation of clean air in work chamber  c  One exhaust filter to exhaust air for preventing heat build up      The intake  supply and exhaust filters shall fulfill the following characteristics     ULPA  Ultra Low Penetration Air  type  HEPA filters shall not be used    Mini pleat separator technique  Aluminum separators shall not be used   Lightweight aluminum framed  particle board  wood frame filters shall not be used   Typical efficiency of 99 999  at 0 1 to 0 3 microns   Protected by a face guard   Test for leaks at time of manufacture     gt 9QAo0T        The filters shall be easily accessible for scan testing in the field without accessing any potentially    contaminated areas       Adedicated upstream sampling port shall be provided  accessible from within the isolator  to facilitate filter    scantesting in the field  Sampling ports
11. lobal Offices   Kuala Lumpur  Malaysia   Leiden  The Netherlands   Manama  Bahrain  Mumbai  India   Philadelphia  USA   Salisbury  UK   Shanghai  China   Singapore    ESCO    WORLD CLASS  WORLDWIDE     Worldwide Headquarters    Esco Micro Pte Ltd e 21 Changi South Street 1 e Singapore 486777 ROC No  198400165W  Phone  65 6542 0833 e Fax  65 6542 6920 e mail escoglobal com   www escoglobal com    Section VII  Ergonomics    A     The isolator shall be designed with a sloped front to reduce glare  increase accessibility  and ensure a  comfortable work position for the operator       The isolator shall possess a sliding tray to facilitate material transfer without the operator having to reach into    the passthru interchange area       Each isolator shall be delivered with the following     a  Tool kit containing common hand tools as well as replacement fasteners and miscellaneous hardware  b  Original factory test report   c  1 copy of the User and Service Manual     The factory test report shall be permanently and digitally archived by the manufacturer should it need to be  obtained in the future in the field for isolator re certification       Quick start  flicker free  low glare  electronically ballasted lightning shall be provided in the isolator  The    minimum average light level shall not be lower than 950 Lux to the work surface  The colour of the lighting  shall be warm  5000k fluorescent lamps to be used   Conventional electromagnetic ballasts shall not be  acceptable 
12. m the front and top  enabling    the isolator to be installed next to a wall or other equipment in order to save room space       The front hinged panel of the isolator shall be equipped with gas spring supports  The front hinged panel shall    be secured to prevent accidental access     All service work  excluding installation work  shall be possible with no more than 1 serviceperson  for a 4ft    1 2m isolator   The fan plenum shall be equipped with mechanical assisted lifting provisions to allow for easy  filter replacement     Global Offices   Kuala Lumpur  Malaysia   Leiden  The Netherlands   Manama  Bahrain  Mumbai  India   Philadelphia  USA   Salisbury  UK   Shanghai  China   Singapore    ESCO    WORLD CLASS  WORLDWIDE     Worldwide Headquarters    Esco Micro Pte Ltd e 21 Changi South Street 1 e Singapore 486777 ROC No  198400165W  Phone  65 6542 0833 e Fax  65 6542 6920 e mail escoglobal com   www escoglobal com    Section IX  Site Requirements  Location     1     The isolator shall be installed as far away as the laboratory space allows from   a  Personnel traffic flows   b  Air vents  out    c  Door and Window   d  Any other sources of disruptive air currents or air drafts    As the isolator is fully enclosed  the location and disruptive air currents will not affect performance as much   however it may affect the process of work product movement in and out of the isolator  Placing the isolator  away from disruptive air currents will assure maximum CSP sterility     
13. maintained within    2 deg C under all conditions       Main supply voltage fluctuations not to exceed     10  of the nominal voltage  It is recommended that the    voltage fluctuation doesn   t exceed     2  of the nominal voltage at all times       Installation Category  2 0    Installation category  over voltage category  defines the level of transient over voltage which the instrument is  designed to withstand safely  It depends on the nature of the electricity supply and its over voltage protection  means  For example  in CAT Il  which is the category used for instruments in installations supplied from   a supply comparable to public mains such as hospital and research laboratories and most industrial  laboratories  the expected transient over voltage is 2500 V for a 230 V supply and 1500 V for a 120 V supply       Pollution Degree  2 0    Pollution degree describes the amount of conductive pollution present in the operating environment  Pollution  degree 2 assumes that normally only    Global Offices   Kuala Lumpur  Malaysia   Leiden  The Netherlands   Manama  Bahrain  Mumbai  India   Philadelphia  USA   Salisbury  UK   Shanghai  China   Singapore    
14. n to a summary of the service instructions      The distributor and   or manufacturer shall be able  on request  to provide local on site training for the user s    The manufacturer shall have officially trained the distributor      The distributor shall be able  on request  to certify the isolator on site after installation  The distributor may  also engage or refer the customer to a third party service provider  The following tests  at a minimum  shall be  performed after installation    a  Downflow velocity and uniformity   b  Airflow visualization   c  Filter leak scan  both downflow and exhaust filters    Warning  It is the responsibility of the user to ensure the isolator is tested after installation  in order to  ensure all the safety requirements have been met      The manufacturer shall provide a minimum parts warranty of 3 years  supported by the local distributor  for the  isolator with the exception of consumable items  filters  lamps etc       When necessary  the isolator may be installed with    a  6 hook IV bar  b  2 electrical outlets  c  All stainless steel construction  d  Sharps disposal system  e  Cleaning accessories  f  Alarm package  g  Support stand of fixed height with leveling feet or casters  h  Support stand of telescoping height for leaving feet and casters  i    Adjustable hydraulic stand with casters  j  Exhaust activated carbon filter and or external vent system    The isolator shall be designed such that all service work may be carried out fro
15. well as provide secondary protection against damage       The isolator shall possess a robust dual wall construction  All positive pressure work zones shall be    surrounded by negative pressure plenums at the sides and back       All components  with the exception of the fan and filters  shall be located outside of plenum to facilitate service    work without the need to decontaminate the isolator       The isolator shall be permanently marked with safety instructions clearly visible from the front       All electrical components shall be UL listed or recognized or certified to any other accepted international    standard       The isolator shall meet UL 61010 1 IEC 610101 1  EN 61010 1 and CSA C22 2 No 61010 1       The isolator shall be designed to eliminate sharp edges  protrusions in order to minimize the risk of personal    harm to the operator  All metal edges shall be dressed and deburred     10  High quality and durable hypalon sleeves shall be provided together with safe change cuff ring that will    ensure glove change with zero risk of contamination     11  All exterior painted surfaces of the isolator shall be powder coated with an antimicrobial inhibitor to minimize    contamination  which cannot be washed out or diminish by repeated cleaning     Global Offices   Kuala Lumpur  Malaysia   Leiden  The Netherlands   Manama  Bahrain  Mumbai  India   Philadelphia  USA   Salisbury  UK   Shanghai  China   Singapore    ESCO    WORLD CLASS  WORLDWIDE     Worldwide Headqu
    