        Series 2000 Treadmill Operators Manual
         Contents
1.                        1 7  Service Information                                                    1 8  Service Requirements                                                1 8  Equipment Identification      1 9  Equipment Overview                               2 1  Intended  Use  is ji 2 3  Preparation for USe s sinsla nd 2 4  Safe Handling  Guidelines  viral 2 4  Operating Instructions                                                 2 5  Electrical Safety Tests                                                2 5  Operating Contralsz   sente A nee ne io criado 2 5  Controlling the Treadmill                                              2 5  Emergency Stop Switch Check                                        2 6  Maintenance    ml dena trae eawns 3 1  INTTOQUCUON ecient cinerea sm ERIE ees ner dan aa RAAN ERRA 3 3  Recommended Maintenance                                           3 3  Inspection and Cleaning                                                3 4  Visual lns pectin xs a ne NAA bb ki ot NU 3 4  Extenor Cleanings stag tas as dida bids 3 4  Revision B Series 2000 Treadmill i    409110 006       Series 2000 Treadmill  409110 006    Revision B    CE Marking Information       CE Marking Information         Compliance    The Series 2000 Treadmill bears CE mark CE 0459 indicating its  conformitv with the provisions of the Council Directive 93 42 EEC  concerning medical devices and fulfills the essential requirements of  Annexl of this directive  The product is in radio interf
2.     Revision B    Series 2000 Treadmill  409110 006    3 1    For vour notes       3 2 Series 2000 Treadmill Revision B  409110 006    Maintenance  Introduction       Introduction    Recommended Maintenance    See the    Series 20000 Treadmill field service manual     PN 409110 004   for detailed maintenance information     NOTE  Unless vou have an Equipment Maintenance Contract  GE Medical  Systems Information Technologies does not in any manner assume  the responsibility for performing the recommended maintenance  procedures  The sole responsibility rests with the individual or  institution using the equipment  GE Medical Systems Information  Technologies service personnel may  at their discretion  follow the  procedures provided in this manual as a guide during visits to the  equipment site        Revision B    Series 2000 Treadmill  409110 006    3 3    Maintenance  Inspection and Cleaning       Inspection and Cleaning    Visual Inspection    Exterior Cleaning    Perform a visual inspection dailv  Turn off the unit and remove power  before making an inspection or cleaning the unit  If you notice any items  that need repair  contact an authorized service person to makethe  repairs     Inspect the following for excessive wear or damage    m Walking belt   Drive belt   Handrail and hardware   All cords and cables for fraying or other damage    Verify that all cords and connectors are securely seated     Turn the treadmill system off  Clean the exterior surfaces with a dean  
3.  Each page of the document has the document part number and revision  letter at the bottom of the page  The revision letter identifies the  document s update level     The revision historv of this document is summarized in the table below              Revision Date Comment  A 7 July 1998 Initial release of manual   B 6 January 2004 Updated to show the new serial number  configuration and make editorial changes                 Manual Purpose    This manual contains the instructions necessarv to operate the  equipment safelv in accordance with its function and intended use  These  instructions include but are not limited to     m an explanation of the function of controls and indicators   m the sequence of operation     m connection and disconnection of detachable parts and accessories   and      instructions for operator cleaning  preventive inspection and  maintenance     Where necessary the manual identifies additional sources of relevant  information and or technical assistance        Revision B Series 2000 Treadmill 1 3  409110 006       Introduction  Manual Information       Conventions    Safetv Messages    Definitions    These are the conventions used in this manual     DANGER safety messages indicate an imminently hazardous situation  which  if not avoided  WILL result in death or serious injurv     WARNING safety messages indicate a potentially hazardous situation  which  if not avoided  COULD result in death or serious injury     CAUTION safety messages indicate a po
4.  Filter aretrademarks of GE Medical Svstems Information  Technologies registered in the United States Patent and Trademark Office     12SL  155L  Access  AccuSpeak  ADVANTAGE  BAM  BODYTRODE  Cardiomatic  CardioSpeak  CD  TELEMETRY  LAN  CENTRALSCOPE  Corolation  EDIC  EK Pro  Event Link Cirrus  Event Link  Cumulus  Event Link Nimbus  HI RES  ICMMS  IMAGE VAULT   IMPACT wf  INTER LEAD  IQA   LIFEWATCH  Managed Use  MARQUETTE PRISM  MARQU ETTE   RESPON DER  MENTOR   MicroSmart  MMS  MRT  MUSE CardioWindow  NST PRO  NAUTILUS  O gt SENSOR  Octanet  OMRS  PHi   Res  Premium  Prism  QUIK CONNECT V  QUICK CONNECT  QT Guard  SMART PAC  SMARTLOOK   Spiral Lok  Sweetheart  UNITV  Universal  Waterfall  and Walkmom are trademarks of GE Medical Svstems  Information Technologies     OGE Medical Systems Information Technologies  1998  2004  All rights reserved        T 2 Series 2000 Treadmill Revision B  409110 006 6 Januarv 2004    Contents       introdu  tioti  iceri isib   rib nr entanti 1 1  Manual Information                                                    1 3  REVISION SON a tn MR mn l 3  Manual Up 0  A une A A re Lidia hts l 3  CONVENIOS reas PET CR a ei ta  tna WANAE eae   in l 4  Safety Information                                                     1 5  Responsibility of the Manufacturer                                       1 5  Generali suis Rats te e l partita de A nl 1 5  Equipment  Sympos sesini eei aa ia i a le En et ayia 1 6  Danger and Warnings                         
5.  soft cloth and a mild dishwashing detergent diluted in water  Wring out  the excess water from the cloth and take care not to drip solutions on the  keyboard or writer assembly   Use anti septic cleaner on the handrails  and walking belt   Avoid contact with open vents  plugs or connectors   Dry the surfaces with a clean cloth or paper towel        3 4    Series 2000 Treadmill Revision B  409110 006    CE    GE Medical Svstems 0459  Information Technologies       gemedical com    World Headquarters  GE Medical Svstems  Information Technologies  Inc   8200 West Tower Avenue  Milwaukee  WI 53223 USA  Tel   1 414 355 5000   1 800 558 5120  US only   Fax   1 414 355 3790    European Representative   GE Medical Systems  Information Technologies GmbH  Munzinger StraBe 3 5   D 79111 Freiburg   Germany   Tel    49 761 4543 0   Fax    49 761 45 43   233    Asia Headquarters   GE Medical Systems   Information Technologies Asia  GE  China  Co   Ltd   24th Floor  Shanghai MAXDO Center    8 Xing Yi Road  Hong Qiao Development Zone  Shanghai 200336  P R  China   Tel    86 21 5257 4650   Fax    86 21 5208 2008    
6. Medical Systems  Information Technologies       Theelectrical installation of the relevant room complies with the  requirements of the appropriate regulations     m The equipment is used in accordance with the instructions for use     This device is intended for use under the direct supervision of a licensed  health care practitioner     Toensure patient safety  use only parts and accessories manufactured or  recommended by GE Medical Systems Information Technologies     Contact GE Medical Systems Information Technologies for information  before connecting any devices to this equipment that are not  recommended in this manual     If the installation of this equipment  in the USA  will use 240 V rather  than 120 V  the source must be a center tapped  240 V  single phase  circuit     Parts and accessories used must meet the requirements of the applicable  IEC 601 series safety standards  and or the system configuration must  meet the requirements of the IEC 60601 1 1 medical electrical systems  standard     The use of ACCESSORY equipment not complying with the equivalent  safety requirements of this equipment may lead to a reduced level of  safety of the resulting system  Consideration relating to the choice shall  include    m use of the accessory in the PATIENT VICINITY  and   m evidence that the safety certification of the ACCESSORY has been    performed in accordance to the appropriate IEC 60601 1 and or IEC  60601 1 1 harmonized national standard        Revision B    Serie
7. Series 2000    Treadmill  Operator   s Manual    409110 006 Revision B          GE Medical Systems  Information Technologies    gemedicalsystems com    NOTE  Due tocontinuing product innovation  specifications in this manual are subject to change without  notice     Listed below are GE Medical Systems Information Technologies trademarks  All other trademarks contained  herein are the property of their respective owners     900 SC  ACCUSKETCH  AccuVision  APEX  AQUA K NOT  ARCHIVIST  Autoseq  BABY MAC  C Qwik  Connect  CardioServ  CardioSmart  CardioSvs  CardioWindow  CASE  CD TELEMETRV  CENTRA  CHART  GUARD  CINE 35  CORO  COROLAN  COROMETRICS  Corometrics Sensor Tip  CRG PLUS  DASH   Digistore  Digital DATAQ  E for M  EAGLE  Event Link  FMS 101B  FMS 111  HELLIGE  IMAGE STORE   INTELLIMOTION IQA  LASER SXP  MAC  MAC LAB  MACTRODE  MANAGED USE  MARQUETTE   MARQUETTE MAC  MARQUETTE MEDICAL SVSTEMS  MARQUETTE UNITV NETWORK  MARS   MAX  MEDITEL  MEI  MEI in the circle logo  MEMOPORT  MEMOPORT C  MINISTORE  MINNOWS   Monarch 8000  MULTI LINK  MULTISCRIPTOR  MUSE  MUSE CV  Neo Trak  NEUROSCRIPT   OnlineABG  OXYMONITOR  Pres R Cuff  PRESSURE SCRIBE  QMI  QS  Quantitative Medicine   Quantitative Sentinel  RAC RAMS  RSVP  SAM  SEER  SILVERTRACE  SOLAR  SOLARVIEW  Spectra  400  Spectra Overview  Spectra Tel  ST GUARD  TRAM  TRAM NET  TRAM RAC  TRAMSCOPE  TRIM  KNOB  Trimline  UNION STATION  UNITY logo  UNITV NETWORK  Vari X  Vari X Cardiomatic   VariCath  VARIDEX  VAS  and Vision Care
8. belt     Keep hands  hair  jewelry  and loose clothing away from  moving parts     Do not place feet under treadmill during elevation  changes  Otherwise  serious injury could result     Routethe AC power cable away from moving parts     Verify proper operation of the stop switch assembly   pn 88380 006  every month     Operate the treadmill with 6 feet of clearance at the rear   end opposite the motor               Revision B    Series 2000 Treadmill  409110 006    1 7    Introduction  Service Information       Service Information    Service Requirements    Refer equipment servicing to GE Medical Svstems Information  Technologies authorized service personnel only  Any unauthorized  attempt to repair equipment under warrantv voids that warrantv     It is the user s responsibility to report the need for serviceto GE Medical  Systems Information Technologies or to one of their authorized agents     Failure on the part of the responsible individual  hospital  or institution  using this equipment to implement a satisfactory maintenance schedule  may cause undue equipment failure and possible health hazards     Regular maintenance  irrespective of usage  is essential to ensure that  the Series 2000 treadmill will always be functional when required    Technical specifications describing the equipment can be found in the     Series 20000 treadmill field service manual     PN 409110 004        Series 2000 Treadmill Revision B  409110 006    Introduction  Service Information       Equ
9. er     The emergency stop switch is a safety device for use in emergency  situations to stop the treadmill  Press the    STOP    push button and the  treadmill promptly stops but the belt will not lock  allowing for removal  of foreign objects  To release the emergency stop switch  turn the push   button 1 4 turn in either direction     Controlling the Treadmill    m Turn the power switch on  I    U se the controlling equipment to start the treadmill  adjust the  treadmill speed and grade  proceed through the exercise phases   terminate the exercise session  and turn off the treadmill        Revision B    Series 2000 Treadmill 2 5  409110 006    Equipment Overview  Operating Instructions             CAUTIONS  Do NOT contact unit or patient during defibrillation     Wait until treadmill belt is moving before placing feet on  belt     Keep hands  hair  jewelrv  and loose clothing away from  moving parts     Do not place feet under treadmill during elevation  changes  Otherwise  serious injury could result           Emergency Stop Switch Check    With the belt moving at a relatively high speed  press the emergency  stop switch  The treadmill belt will stop promptly but the belt will  not lock  allowing for removal of foreign objects  Toreleasethe  switch  turn the button 1 4 inch turn in either direction     Use the controlling equipment to terminate the exercise session and  turn off the treadmill        2 6    Series 2000 Treadmill Revision B  409110 006    3    Maintenance   
10. erence protection  class A in accordance with EN 55011     The country of manufacture can be found on the equipment labeling     The product complies with the requirements of standard EN 60601 1 2   Electromagnetic Compatibility   Medical Electrical Equipment        The safety and effectiveness of this device has been verified against  previously distributed devices  Although all standards applicable to  presentiv marketed devices may not be appropriate for prior devices  i e   electromagnetic compatibility standards   this device will not impair the  safe and effective use of those previously distributed devices  See user s  information     Exceptions    The Series 2000 Treadmill EMC  Immunitv Performance    Users should be aware of known RF sources  such as radio or TV  stations and hand held or mobile two way radios  and consider them  when installing a medical device or system     Be aware that adding accessories or components  or modifying the  medical device or system may degradethe EMI performance  Consult  with qualified personnel regarding changes to the system  configuration        Revision B Series 2000 Treadmill CE 1  409110 006    CE Marking Information       For vour notes       CE 2 Series 2000 Treadmill Revision B  409110 006    l    Introduction       Revision B    Series 2000 Treadmill  409110 006    1 1    For vour notes       1 2 Series 2000 Treadmill Revision B  409110 006    Introduction  Manual Information       Manual Information    Revision Historv   
11. he end of the bed assembly to the floor        MD1128 022B       2 4 Series 2000 Treadmill Revision B  409110 006    Equipment Overview  Operating Instructions       Operating Instructions    Electrical Safety Tests    Operating Controls    Power Switch    Emergency Stop Switch    The electrical safety of this installation is the responsibility of the  customer  not GE Medical Systems Information Technologies  In  hospitals  contact your in house biomedical technician  electrician  or  technicallv qualified personnel  Outside of hospitals  contact your  hospital affiliation for these services  Otherwise  contact GE Medical  Svstems Information Technologies and open a customer billable service  call     Before using the treadmill  perform the tests listed below   m AC line voltage test to verify the power outlet is properly wired       Ground continuity test to verify all exposed metal is properly  grounded     m Leakage tests to verify the equipment passes all applicable leakage  tests     Your in house biomedical technician  electrician  or technically qualified  personnel can find instructions for performing these tests in Chapter 3      Maintenance        The treadmill has two operating controls  the power switch and the  emergency stop switch  The power switch is located on the rear panel   and the emergency stop switch is on the front handrail     The power switch controls the AC power to the treadmill  The on  position  1  applies power  The off position  0  removes pow
12. ipment Identification  The Equipment Identification tag that contains the Serial Number is  located where shown below        Equipment  Identification                     yA    UU           A  AN  AN    Serial Number  Every GE Medical Systems Information Technologies device has a  unique serial number for identification  An explanation of the Serial  Number code is shown below     HHH HH HH HHHH HE      RA    E F    A Product Code  ABM     Series 2000 Treadmill     V ear Manufactured  00 99   00   2000   01  2001   02   2002    and so on     Fiscal Week Manufactured  Production Sequence Number  Manufacturing Site    Mm MO O    Miscellaneous Characteristic       Revision B Series 2000 Treadmill 1 9  409110 006    Introduction  Service Information       For vour notes       1 10 Series 2000 Treadmill Revision B  409110 006    2    Equipment Overview       evision    eries 2000 Treadmill  409110 006    For vour notes       2 2 Series 2000 Treadmill Revision B  409110 006    Equipment Overview  Intended Use       Intended Use    The Series 2000 Treadmill is intended for use with anv oneof the several  GE Medical Systems Information Technologies exercise testing systems  or the MTC 1  manual treadmill controller  for administering a  controlled exercise load during a diagnostic stress test  Standard  features include the emergency stop switch  a full handrail set  and a  long  60 inch walking surface     The emergency stop switch is intended for emergency situations where  immedia
13. s 2000 Treadmill 1 5  409110 006    Introduction  Safety Information       Equipment Svmbols     gt   gt            gt     The following symbols appear on the equipment     This symbol means that you must pay attention to the documents delivered with this  equipment  It calls attention to the things to which you must pay special attention during  operation and when the equipment is operated in conjunction with other equipment     In Europe  this symbol means dangerous or high voltage  In the United States  this symbol  represents the caution notice below           CAUTION  To reduce the risk of electric shock  do NOT remove  cover  or back   Refer servicing to qualified personnel           Type B equipment  Type B equipment is suitable for intentional external and internal  application to the patient  excluding direct conductive connection to the patient s heart     Alternating current  AC     Equipotential  This is the ground lug      Protective earth  ground     M13495  M13504  M13864  M13571  M13574  M13573       Series 2000 Treadmill Revision B  409110 006    Introduction  Safety Information       Danger and Warnings          DANGER  DoNOT use in the presence of flammable anesthetics                 WARNINGS  Replace only with the same type and rating of fuse     Thisis Class   equipment  The mains plug must be  connected to an appropriate power supply     Do NOT contact unit or patient during defibrillation     Wait until treadmill belt is moving before placing feet on  
14. tely stopping the treadmill is required to deliver appropriate  emergency care to the patient or health care provider  as implied by the  American Heart Association    Guidelines for Clinical Exercise Testing  Laboratories     1995   It is not intended for routinely stopping the  treadmill        Revision B    Series 2000 Treadmill 2 3  409110 006    Equipment Overview  Preparation for Use       Preparation for Use    Safe Handling Guidelines    The Series 2000 Treadmill ships preset with an approximate 2  grade   This slight elevation provides for free wheel movement and prevents the  shroud from scraping the floor     NOTE Assembiv and Technical specifications describing the equipment  can be found in the    Series 20000 treadmill field service manual      PN 409110 004     If you are moving the treadmill after it has been in operation  use the  controlling equipment to set the grade to approximately 7   Then  remove power and disconnect all cables to the treadmill before moving  the unit     Follow the steps below to move the treadmill  We recommend that two  people work together since the treadmill may be too heavy for some  individuals to lift and lower safely     1  Lift the end of the bed assembly to about knee height  keeping knees  bent and backs straight as you lift     2  Rotate the treadmill in the direction you want to go  the treadmill  will pivot on its wheels  and push forward     3  When you have maneuvered the treadmill into its new location   gently lower t
15. tentially hazardous situation  which  if not avoided may result in minor or moderate injury     NOTE messages provide additional user information     m  Items shown in Bold text are keys on the keyboard  text to be  entered  or hardware items such as buttons or switches on the  equipment     m Items shown in Italicized text are software terms which identify  menu items  buttons  or options in various windows       Toperform an operation which appears with a plus  sign between  the names of two keys  you press and hold the first key while  pressing the second key once  This is called a keystroke combination     For example     Press Ctrl Esc    means to press and hold down the  Ctrl key while pressing the Esc key     m When instructions are given for typing a precise text string with one  or more spaces  the point where the spacebar must be pressed is  indicated as   lt Space gt   The purpose of the  lt  gt  brackets is to ensure  you press the spacebar when required       Enter means to press the    Enter    or    Return    key on the keyboard   Do not type    enter           Series 2000 Treadmill Revision B  409110 006    Introduction  Safety Information       Safetv Information    Responsibilitv of the Manufacturer    General    GE Medical Svstems Information Technologies is responsible for the  effects of safety  reliabilitv  and performance only if       Assembly operations  extensions  readjustments  modifications  or  repairs are carried out by persons authorized by GE 
    