6125665 Serv Galileos EN.book

image

Contents

1. The constancy test phantom is inserted in the bite block holder of the unit for for the constancy and acceptance tests as well as for the check of the mechanical unit adjustment gt 166 Checking the mechanical system adjustment 1 Insert the constancy test phantom in the bite block holder of the unit P3 B2 B2 j e4 NO tog 2 Measure distances B1 B2and B3 between the tube assembly housing and measuring point M on the constancy test phantom positions 1 2 and 3 using the steel tape measure from the service set 3 Then calculate the ideal distance between the tube assembly and measuring point M as follows B1 B2 2 ideal distance 4 Distances B1 B2 and B3 must not deviate more than 2 mm from the calculated ideal distance With deviations exceeding 2mm the unit must be adjusted mechanically via the position of the ring motor gt 193 61 25 665 D3437 166 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 7 Adjusting calibrating the unit Service Manual GALILEOS 7 3 Adjustment and calibration via the Ser
2. GALILEOS Remote control Board Software Board Software DX6 03 05 00 DX42 02 54 00 DX7 02 77 01 DX7 LO 02 29 00 DX7 L1 02 29 00 DX7 L2 02 27 00 DX7 L3 02 27 00 DX7 L4 02 04 00 DX7 L5 02 02 00 DX71 02 54 00 DX11 04 04 01 DX11 FPGA 01 03 00 DX41 02 30 00 DX89 01 55 00 DX89 FPGA 01 54 00 SIDEXIS XG GALILEOS Software RCU GALILEOS Server software Implant V2 5 1 or higher V1 8 V2 0 V2 1 V1 8 Main software V04 07 00 GALILEOS Remote control Board Software Board Software DX6 03 05 00 DX42 02 55 00 DX7 02 79 00 DX7 LO 02 29 00 DX7 L1 02 29 00 DX7 L2 02 27 00 DX7 L3 02 27 00 DX7 L4 02 04 00 DX7 L5 02 02 00 DX71 02 54 00 DX11 04 07 00 DX11 FPGA 01 03 00 DX41 02 30 00 DX89 01 56 00 DX89 FPGA 01 54 00 SIDEXIS XG GALILEOS Software RCU GALILEOS Server software Implant V2 5 1 or higher V1 8 V2 0 V2 1 V1 8 54 61 25 665 D3437 D3437 076 01 15 02 06 2012 Main software V04 07 01 GALILEOS Remote control Board Software Board Software DX6 03 05 00 DxX42 02 55 00 DX7 02 79 00 DX7 LO 02 29 00 DX7 L1 02 29 00 DX7 L2 02 27 00 DX7 L3 02 27 00 DX7 L4 02 04 00 DX7 L5 02 02 00 DX71 02 54 00 DX11 04 07 01 DX11 FPGA 01 03 00 DX41 02 30 00 DX89 01 56 00 DX89 FPGA 01
3. Error code Description Actions required see E1 06 42 Tolerance exceeded e Replace the tube assembly S gt 310 mA nom Error code Description Actions required see E1 06 43 Tolerance exceeded e Replace the tube assembly S gt 310 VH actual value Error code Description Actions required see E1 06 44 Tolerance exceeded kV actual e Replace the tube assembly S gt 310 value Error code Description Actions required see E1 06 45 Tolerance exceeded mA actual e Replace the tube assembly S gt 310 value Error code Description Actions required see E6 06 51 VHmax e Replace the tube assembly S gt 310 Error code Description Actions required see E6 06 52 MAmax e Replace the tube assembly S gt 310 Error code Description Actions required see E6 06 53 KVmax e Replace the tube assembly S gt 310 Error code Description Actions required see E6 06 54 Basic heating pulses not applied e Replace the tube assembly S gt 310 Error code Description Actions required see E6 06 55 Anode voltage too low e Replace the tube assembly S gt 310 Error code Description Actions required see E6 06 56 Error during auto compensation e Please contact the Sirona Customer S 61 Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary e Let the tube assembly cool down for
4. GALILEOS Remote control Board Software Board Software DX6 02 88 00 DX42 02 47 00 DX7 02 60 00 DX7 L0 02 22 00 DX7 L1 02 22 00 DX7 L2 02 22 00 DX7 L3 02 22 00 DX7 L4 01 00 00 DX7 L5 DX71 DX11 02 66 00 DX41 02 30 00 DX89 01 16 00 DX89 FPGA 01 15 00 The main software V 03 05 00 is run compatible as of GALILEOS Software V1 5 51 Main software V03 06 01 Main software V03 06 02 52 GALILEOS Remote control Board Software Board Software DX6 02 90 00 DX42 02 48 00 DX7 02 61 00 DX7 LO 02 22 00 DX7 L1 02 22 00 DX7 L2 02 22 00 DX7 L3 02 22 00 DX7 L4 01 00 00 DX7 L5 DX71 02 40 00 DX11 02 67 01 DX41 02 30 00 DX89 01 17 00 DX89 FPGA 01 15 00 For GALILEOS Comfort The main software V03 06 01 is run compatible from GALILEOS Software V1 5 or above For GALILEOS Compact The main software V03 06 01 is run compatible from GALILEOS Software V1 6 or above GALILEOS Remote control Board Software Board Software DX6 02 90 00 DX42 02 48 00 DX7 02 61 00 DX7 LO 02 22 00 DX7 L1 02 22 00 DX7 L2 02 22 00 DX7 L3 02 22 00 DX7 L4 01 00 00 DX7 L5 DX71 02 40 00 DX11 02 67 03 DX41 02 30 00 DX89 01 17 00 DX89 FPGA 01 15 00 For GALILEOS Comfort The main software V03 06 01 is run compatible from GALILEOS So
5. DX1 X303 m z DX42 X100 Lo SA TT L25 L26 _ T T L108 X103 DX41 L117 X103 Media Konverter L117 L7 L25 y 61 25 665 D3437 D3437 076 01 15 02 06 2012 E Media converter 29 3 Unit description 3 2 Hardware Facescan cabling Sirona Dental Systems GmbH Service Manual GALILEOS as PAA 9 2 L77 L78 FACESCAN L77 30 A Power switch 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 3 Unit description Service Manual GALILEOS 3 2 Hardware Board photos Boards in the slide Boards DX1 DX11 Installed up to unit serial number 8499 for GALILEOS Comfort and 48499 for GALILEOS Compact DX1 X403 X402 X607 J302 J309 X303 X302 X306 X309 X104 X611 J306 X100 X103 REF 59 25 214 DS rua X811 X812 X813 X610 REF 59 24 142 J901 61 25 665 D3437 D3437 076 01 15 02 06 2012 31 3 Unit description Sirona Dental Systems GmbH 3 2 Hardware Service Manual GALILEOS Boards DX1 DX11 Installed from units with a serial number of 8500 and above for GALILEOS Comfort and 48500 and above for GALILEOS Compact IMPORTANT The DX1 DX11V2 board can only be operated with unit
6. 1 Loosen the four screws A and remove cover plate B incl the cable shielding L3 CAUTION Also pull cable L3 off connector X3 and the ground cable off connector X304 on board DX6 Tip The ferrite core and cable shielding can remain on the cover plate 2 Detach cables L5 L6 and L15 from the rubber grommets and pull the cables off of sockets J6 L5 J2 J3 L6 and J5 L15 on board DX6 61 25 665 D3437 312 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 6 X ray tube unit 3 Loosen the two rear screws C on the tube assembly 4 CAUTION The tube assembly is heavy Hold the tube assembly firmly in place loosen the two front screws D 3 4 turns and remove the tube assembly toward the front Preparing the new X ray tube assembly for installation INOTICE eee Oil leakage The screws E from the ring X ray window must not be loosened as otherwise oil may leak out IMPORTANT Note the order for attaching the filters when reattaching the diaphragm unit to the new tube assembly If the filters are installed in the wrong order this will impair the image quality Insert the aluminum filter D first followed by the copper filter M 61 25 665 D3437 D3437 076 01 15 02 06 2012 313 9 Repair 9 6 X ray tube unit Sirona Dental Systems GmbH Service Manual GALILEOS With the Type 1 diaphragm unit
7. 61 25 665 D3437 D3437 076 01 15 02 06 2012 203 Alphabetical list of service routine functions Function Service routine see Acoustic signal for end of exposure activate deactivate 017 15 S gt 268 Automatic adjustment of pulse preheating S005 8 S gt 227 CAN bus logging in the web interface enable S007 5 S gt 235 CAN bus service 012 S gt 249 Configuration service S017 S gt 252 Configure the switching plate for the swivel arm 017 7 S gt 261 Configuring boot mode DYNAMIC DHCP AutolP STATIC 037 3 S gt 281 fixed address Country group code enter 017 3 S gt 254 Current measurement unpulsed 011 9 S gt 248 Default gateway address manual input 037 4 S gt 283 Default gateway address set to defaults 037 2 S gt 279 Displaying the network data 037 1 S gt 277 Dosimetry without ring movement S011 S gt 247 Dosimetry with pulsed radiation 011 12 S gt 247 Error logging memory S007 S gt 229 Error logging memory clear 007 2 S gt 232 Error logging memory display 007 1 S gt 229 Fan test 005 4 S gt 225 Flash file system S009 S gt 241 Flash file system format S009 4 S gt 241 Flash file system test 009 5 S gt 243 Height adjustment sensor system test 018 4 S gt 275 Height adjustment service S018 S gt
8. Error code Description Actions required see E5 11 09 Internal error in program sequence Acknowledge error and repeat procedure of board DX11 If the error occurs repeatedly S gt 61 e Perform a software update bug fix Error code Description Actions required see E7 1111 Wrong unit configuration e Check the unit configuration by running S gt 252 service routine S017 2 and reconfigure if necessary Error code Description Actions required see E7 11 12 Internal error in data management If the error occurs after a module has been S of board DX11 replaced e Query More details with SiXABCon and seek advice from the Sirona Customer Service Center on how to proceed If no module has been replaced e Restart the unit 1 Switch off the unit 2 Wait 1 minute 3 Switch unit on 4 Repeat procedure e Check software versions on the info S gt 238 screen or by running service routine S gt 61 008 2 perform software update if necessary If the error occurs repeatedly S gt 61 e Perform a software update bug fix Error code Description Actions required see E7 11 14 Wrong remote control for this unit Install the correct remote control This error message blocks all unit If necessary obtain a new remote control functions To continue to work with from the manufacturer PIS unit yOu Wu svunpiig Ne Aremote control for another Sirona unit or a remote control and restart the unit
9. S gt 343 e Replace board DX41 Error code Description Actions required see E6 41 03 Invalid commanding or control data e Check the CAN bus S gt 128 This error may also occur in e Please contact the Sirona Customer S gt 61 connection with other causal error Service Center CSC to find out whether messages Please also observe the a bug fix by means of a software update is causal error message It appears possible and perform such an update if only after you acknowledge the first necessary error message Error code Description Actions required see E6 41 04 Data transfer error or dialog error to Check the CAN bus S gt 128 edule Masten SIAE e Please contact the Sirona Customer S gt 61 Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary Error code Description Actions required see E6 41 05 Data transfer error or dialog error to e Repeat the software update S gt 61 Roctieadorobmedule e Check the CAN bus S gt 128 Only occurs in connection with e Replace board DX41 S gt 343 software update 61 25 665 D3437 D3437 076 01 15 02 06 2012 107 Error code Description Actions required see E6 41 06 Module failed in TTP time trigger e Check the CAN bus S gt 128 protocol detected on master side Replace board DX41 S gt 343
10. After implementing repair work protective conductors and device leakage current checks must be carried out see the sections on Checking protective conductor and Checking device leakage current 61 25 665 D3437 D3437 076 01 15 02 06 2012 289 9 Repair Sirona Dental Systems GmbH 9 2 Height adjustment motor M1_4 spindle Service Manual GALILEOS Height adjustment motor M1_4 spindle Preparing for motor replacement Switch the unit on Use the Up Down keys on the control panel to move the slide up Switch the unit off again Remove the covers Intermediate piece Profile covers top and bottom Tip While loosening the screws press the top profile cover down towards the unit and allow it to slide down once the screws are loose Arm cover Slide cover rear center Slide cover rear top Slide cover rear bottom and Slide cover front PON gt Tip If the height adjustment motor is inoperative you can also move the slide manually gt 290 Moving the slide manually Risk of injury due to uncontrolled movement of the slide If the slide can no longer be moved electrically it must be moved mechanically The position of the slide must be secured to ensure that no uncontrolled downward movement occurs during service in cases where the carriage has fewer self locking properties For this purpose Sirona recommends using the free height adjustment service kit REF 62 57 518 Thi
11. Check the software versions of SIDEXIS S gt 61 is available for the current program XG and the unit for compatibility perform parameters software update if necessary Error code Description Actions required see E1 11 88 The unit is in demo mode If user mode is expressly required S gt 81 Occurs when the unit is switched on Switch the demo mode off CAUTION Radiation can be released after the demo mode is switched off Location 12 CAN bus Error code Description Actions required see E6 12 01 CAN controller initialization error on Check the CAN bus S gt 128 DX1 Error code Description Actions required see E6 12 02 CAN malfunction cannot be e Check the CAN bus S gt 128 assigned to module 61 25 665 D3437 D3437 076 01 15 02 06 2012 103 Location 14 Digital extension SIDEXIS XG Error code Description Actions required see E5 14 01 Abort by SIDEXIS XG e Check network connection XG3D plugin installation and software version Error code Description Actions required see E7 1402 Interface version not compatible with SIDEXIS XG e Check the software versions of the unit S008 2 and XG3D plugin and perform software update if necessary S gt 238 Error code Description Actions required see E5 14 04 The network connection was interrupted This error often occurs if SIDEXIS XG is selected before
12. S037 Test step 3 Configuring boot mode 8 13 S037 Network service Selection field Parameters Meaning 1 DYNAMIC Automatic address assignment DHCP AutolP STATIC Fixed address Factory setting 1 Call service routine S037 3 gt 213 STATIC 61 25 665 D3437 D3437 076 01 15 02 06 2012 Once the service routine has been selected the current boot mode of the unit is displayed in selection field 1 281 8 Service Sirona Dental Systems GmbH 8 13 S037 Network service Service Manual GALILEOS 2 Use arrow keys A to select the required boot mode automatic address assignment DYNAMIC or fixed address STATIC in selection field 1 see table Memory key B Easypad or the LED above Memory key B Multipad lights up 3 Save the setting gt 219 4 Exit the service routine gt 220 61 25 665 D3437 282 D3437 076 01 15 02 06 2012 61 25 665 D3437 D3437 076 01 15 02 06 2012 S037 Test step 4 Manual input of static network settings IP address default gateway address and subnet mask This service routine cannot run in DYNAMIC mode T key is blocked Symbol on the control Status panel Function Patient symbol key 1 A The IP address is displayed in selection field 1 or after pressing the T key number pad B1 is selected Patient symbol key 2 B The default gateway is displayed in selection fiel
13. Selection field Code Function 01 00 Welcome screen disabled 01 Welcome screen enabled Factory setting 1 Call service routine 8017 13 gt 213 Once the service routine has been selected the code for the current setting is displayed in selection field 1 2 Use arrow keys A to select the required setting in selection field 1 see table Once the required setting has been selected the Memory key B lights up 3 Save the setting gt 219 4 Exit the service routine gt 220 61 25 665 D3437 D3437 076 01 15 02 06 2012 265 8 Service Sirona Dental Systems GmbH 8 11 S017 Configuration service Service Manual GALILEOS S017 Test step 14 Enable disable certain lines of the welcome screen Selection field Code Meaning Function 1 1 First name Last name Date of birth Patient number Inactive o BR w h Active Factory setting 1 Call service routine 8017 14 gt 213 Once the service routine has been selected the code for the line currently selected is displayed in selection field 1 2 Use the arrow keys A to select the required line in selection field 1 see table The activation status code is displayed in selection field 2 61 25 665 D3437 266 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 11 S017 Configuration service 3 Use t
14. gt 343 TTP Time Trigger Protocol board DX42 replace module if necessary S gt 61 e Replace board DX42 e Please contact the Sirona Customer Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary Error code Description Actions required see E6 42 07 TTP timeout error detected on e Check the CAN bus S gt 128 Save aide e Check power supply 3 3 V of board S gt 343 The module was temporarily not DX11 replace board DX1 or DX11 if addressed by the master necessary npor PONG e PUNE Master Se e Check the signal path from board DX1 to Procedure error in the software board DX42 replace module if necessary Master DX11 receives no return e Replace board DX42 commanding from the module This error may also occur in connection with other causal error messages Please also observe the causal error message It appears only after you acknowledge the first error message TTP Time Trigger Protocol Error code Description Actions required see E6 42 08 General fault detected locally on e Check software versions on the info S gt 238 module slave side CAN controller screen or by running service routine S gt 61 being reinitialized S008 2 perform software update if f Occurs if software of boards is PSS ay Selene incompatible e Check the CAN bus S gt 343 e Replace board DX42 e Please contact the Sirona Customer Service Center CSC
15. 6 Touch the center of the green square in the bottom left corner of the screen The fifth adjustment screen appears on the display 7 Confirm the adjustment settings by clicking the Yes button The touchscreen is now adjusted To repeat the adjustment procedure touch the Vo button To abort the adjustment procedure touch the Abort button 61 25 665 D3437 192 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 7 Adjusting calibrating the unit Service Manual GALILEOS 7 6 Mechanical adjustments Mechanical adjustments Ring center adjustment Perform this adjustment only if the measured values are out of tolerance gt 166 1 Remove the arm cover 2 Move the ring center to the left or right NOTICE Do not undo the screws completely Make sure that spring B does not pop out This spring has a defined prestress Loosen the four screws A slightly 3 Correct the position carefully and then retighten the screws 61 25 665 D3437 D3437 076 01 15 02 06 2012 193 7 Adjusting calibrating the unit Sirona Dental Systems GmbH Service Manual GALILEOS 7 6 Mechanical adjustments X 194 4 Move the ring center to the left or right NOTICE Do not undo the screws completely Loosen the four screws C slightly 5 Correct the position carefully and then retighten the screws If the center
16. IJd wy a X400 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 3 Unit description Service Manual GALILEOS 3 2 Hardware Boards in the stand Board DX32 Up to device serial number 3199 F101 F102 X2 X100 Board DX32 From device serial number 3201 F101 F102 X2 X100 A Line filter 61 25 665 D3437 D3437 076 01 15 02 06 2012 37 3 Unit description Sirona Dental Systems GmbH 3 2 Hardware Service Manual GALILEOS Board DX41 This board is not available as a repair part or spare part Board DX41 is omitted in units with a serial number of 3201 and above DX41 X100 X102 X104 X103 61 25 665 D3437 38 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 3 Unit description Service Manual GALILEOS 3 2 Hardware Board in the remote control Board DX42 This board is not available as a spare part or a repair part DX42 X105 X109 X110 X200 V103 ae X201 X101 V100 X107 61 25 665 D3437 D3437 076 01 15 02 06 2012 39 3 Unit description Sirona Dental Systems GmbH 3 2 Hardware Service Manual GALILEOS Boards in Facescan FACESCAN modular board This modular board is not available as a spare part or a repair part Facescan can only be ordered as a complete unit FACESCAN DASLU L S br 61 25 665 D3437 40 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 3 Unit descr
17. Remove the fabric tube and the pull wire Laying the corrugated tube or cable at the rotation unit Lay the corrugated tubes and cables back in their original position Plug the connectors back in again Reattach the covers ON gt o 61 25 665 D3437 D3437 076 01 15 02 06 2012 365 9 Repair 9 13 Cable 366 Sirona Dental Systems GmbH Service Manual GALILEOS Replacing cable L7 L117 or L108 in cable track 2 N oar w 10 11 12 13 Switch the unit on Move the slide downward to a pleasant working position using the Up Down keys on the control panel Switch the unit off again Remove the arm cover Remove the two cross braces and the cover plate of board DX1 NOTICE Wrap the connector X303 cable L108 with adhesive tape immediately after pulling it off to protect the detent at the connector against breaking off Disconnect the fiber optic cable L7 L117 and cable L108 from board DX1 Switch the unit on Use the Up Down keys on the control panel to move the slide up Tip If the height adjustment motor is inoperative you can also move the slide manually gt 290 Switch the unit off again Remove the covers Intermediate piece and Profile top and bottom Tip While loosening the screws press the top profile cover down towards the unit and allow it to slide down once the screws are loose Remove board DX32 Removing board DX32 gt 294 Detach fiber optic cable L7 L1
18. The adjustment is saved The iris adjustment is now complete 6 Continue the calibration procedure with the shading calibration gt 183 61 25 665 D3437 182 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 7 Adjusting calibrating the unit Service Manual GALILEOS 7 3 Adjustment and calibration via the Service Functions menu 7 3 6 Shading calibration Diaphragm Radiation field Dose Sensor Iris Shading Distortion Geometry c4 gt ie Jiaphragm Shading calibration Radiation field There is no valid shading calibration data on this Nese machine ae Image acquisition Save values v The Service exposure menu is called gt 161 1 Click the Shading tab The corresponding tab card is selected 2 Click on the mage acquisition button SIDEXIS XG makes the unit ready for exposure gt 162 Service routine S010 11 or S010 15 extended shading calibration is displayed on the control panel T amaga Fasten et Dose Seat he mra These Sr ve hcg ela dada on fis 3 Press the R key to move the unit back to the starting position 4 Press the release button Press and hold down the button until the exposure is complete the preview image is displayed in the exposure window and the acoustic signal indicating the end of the exposure double beep sounds if it has been configured 61 25 665 D3437 D3437 076 01 15 02 06 2012 183 Sirona Denta
19. 11 Connect cable L77 WH white to terminal K1 2 orange 12 Fasten cable L76 as low as possible downwards with the cable clamp G 13 Refit the bottom profile cover 61 25 665 D3437 D3437 076 01 15 02 06 2012 337 9 Repair Sirona Dental Systems GmbH 9 10 Head fixation device Service Manual GALILEOS Head fixation device Replacing receptacle element for head fixation for unit with head fixation device Remove defective receptacle element 1 Using the Up Down buttons on the control panel switch the device on and move it to a comfortable working height to remove the acquisition unit 2 Remove the head fixation device see operating instructions 3 fthe bore hole C on the defective acquisition unit is not available Press the locking button A and move the flange B forward to expose the bore hole C 4 Loosen screw D and remove the defective acquisition unit Install the new acquisition unit 1 f the bore hole C on the new acquisition unit is not available Press the locking button A and move the flange B forward to expose the bore hole C 61 25 665 D3437 338 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS 9 Repair 9 10 Head fixation device GALILEOS Adjust the light localizer 61 25 665 D3437 D3437 076 01 15 02
20. 2 To test the flash file system press the Memory key A R key Easypad or LED above R key Multipad lights up followed by the R key B 61 25 665 D3437 D3437 076 01 15 02 06 2012 243 8 Service Sirona Dental Systems GmbH 8 8 S009 Flash file system Service Manual GALILEOS Once the system has passed the test without errors OK appears in selection field 1 Ifthe test fails ERROR is displayed In this case the flash file system must be formatted with service routine S009 4 gt 241 3 Exit the service routine gt 220 61 25 665 D3437 244 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 8 S009 Flash file system S009 Test step 7 Save restore DX89 data 1 Call service routine S009 7 gt 213 After you select this service routine the following can be displayed in selection field 1 DX89 DX11 DX89 gt DX11 D gt Data is transferred from DX89 to DX11 memory key A Easypad or LED above the memory key A Multipad is lit up DX11 gt DX89 Data is transferred from DX11 to DX89 memory key A Easypad or LED above the memory key A Multipad is lit up Data on both boards DX11 and DX89 is valid or the data transfer is not possible all keys Easypad or LEDs Multipad are not lit Only one practical direction of data transfer is offered at any one time If both locations contain valid data is displ
21. 2006 03 07 09 03 26 Message Recording started Value 104 2006 03 07 09 03 45 Message Recording stopped 2006 03 07 09 05 16 Message Recording started Value 104 2006 03 07 09 05 35 Message Recording stopped 2006 03 07 09 07 27 Message Recording started Value 101 2006 03 07 09 07 35 Message Recording cancelled 2006 03 07 09 52 44 Message Recording started Value 9641 2006 03 07 09 52 58 Message Recording stopped Data which might be expected to occur in the error logging memory is explained below to aid you in interpreting it 61 25 665 D3437 D3437 076 01 15 02 06 2012 125 Example of error logging data System time 2006 03 06 20 13 02 System time clock on DX11 Entry type Message General system event Message val General system event with additional value Error Error event DeviceError Data for error event on a module Error Sidexis Network error event Stringname Free status texts Stringsegment Additional data string name RTC Date Time Change Date and time of a SIDEXIS PC PC Date Time Date and time of the DX11 set Compression table Compression table Entry data Message Self test Successful Self test completed successfully Recording started Start of a recording Value 9000 Sequence ID of the recording Recording stopped End of an exposure Recording cancelled Exposure
22. 247 8 Service Sirona Dental Systems GmbH 8 9 S011 Dosimetry without ring movement Service Manual GALILEOS 2 Initiate the radiation Radiation uses 200 pulses and 85kV 42mAs IMPORTANT If you let go of the release button before the maximum radiation time has elapsed radiation is terminated prematurely and the exposure is interrupted The actual radiation time is not displayed If you release radiation during the cool down interval a countdown of the remaining waiting time is displayed in the header of the Easypad or in selection field 2 on the Multipad automatic exposure blocking 3 Exit the service routine gt 220 S011 Test step 9 Current measurement unpulsed 1 Call service routine 8011 9 gt 213 Selection field 1 displays the kVmA level and selection field 2 A XRAY ZN displays the maximum radiation time The kVmA level and the maximum radiation time are preset and 85kV 7mA cannot be changed 4 0s 2 Initiate the radiation The maximum set radiation time has elapsed IMPORTANT If you let go of the release button before the maximum radiation time has elapsed radiation is terminated prematurely and the exposure is interrupted The actual radiation time is notdisplayed If you release radiation during the cool down interval a countdown of the remaining waiting time is displayed in the header of the Easypad or in selection field 2 on the Multipad automatic exposure blocking
23. 310 Error code Description Actions required see E6 06 13 Error when writing to e Acknowledge error and repeat EEPROM procedure Stored data may be lost If the error occurs repeatedly S gt 310 e Replace the tube assembly Error code Description Actions required see E2 06 20 Overtemperature of single tank e Wait until the X ray tube assembly has power pack cooled down e Check fan function by running service S gt 225 routine S005 4 replace fan if S gt 318 necessary e Check temperature sensor in single S gt 226 tank by running service routine S005 5 S gt 310 replace tube assembly if necessary Error code Description Actions required see E6 06 21 Hardware signal of release button Check cable L5 optical fiber replace S gt 141 not detected if necessary S gt 362 e Replace board DX1 S gt 342 e Replace the tube assembly S gt 310 Error code Description Actions required see E6 06 22 Broken temperature sensor e Replace the tube assembly S gt 310 61 25 665 D3437 90 D3437 076 01 15 02 06 2012 Error code Description Actions required see E3 06 23 Hardware signal of release button Check cable L5 applied during powerkon 1 Switch off the unit 2 Pull cable L5 off tube assembly 3 Switch unit on 4 Perform optical check of L5 If light is visible S gt 342 e Replace board DX1
24. 350 Case A V03 03 01 or higher e Inserting a DX77 from another unit Proceed as described in the chapter S gt 352 After changing the DX11 e GALILEOS Comfort board gt 350 Case B e System software version V03 03 01 or higher e Inserting a new DX11 Proceed as described in the chapter S gt 354 After changing the DX11 e GALILEOS Compact board gt 350 Case C e System software version V03 06 01 or higher e Inserting a DX11 from another unit Proceed as described in the chapter S gt 356 After changing the DX11 e GALILEOS Compact board gt 350 Case D e System software version V03 06 01 or higher 61 25 665 D3437 D3437 076 01 15 02 06 2012 347 Board Constellation Actions Page DX6 Replacing a fube assembly including board DX6 Tube e Inserting a new X ray tube assembly Proceed as described in the chapter S assembly Case E e GALILEOS Comfort i e System software version V03 03 01 or higher e Inserting an X ray tube assembly from Proceed as described in the chapter S another unit Case F e GALILEOS Comfort e System software version V03 03 01 or higher e Inserting a new X ray tube assembly Proceed as described in the chapter S e GALILEOS Compact Gase G e System software version V03 06 01 or higher e Inserting an X ray tube assembly from Proceed as described in the chapter S another unit Case H e GALILEOS Compa
25. Facescan Configuration application 2 Click on the Facescanner in browser button A password dialog box opens 3 Inthe field User enter service 4 Inthe field Password enter sirona The Facescan Device Service web dialog opens Select Update dialog gt Inthe menu bar select the menu item UPDATE The Facescan Firmware Update window opens Starting the update 1 Press the Enter Update Menu button A further Facescan Firmware Update window opens 2 Press the Browse button 3 Navigate to the firmware update file and select it 4 Press the Upload mage button 5 Press the Run Update button The update starts The update ends automatically after around 5 minutes with a text message 6 Press the Reboot button 7 Wait around a minute until the green LED on the Facescan unit begins to light up 8 Only if the SIRONA browser is not being used Refresh the browser display e g in Windows Internet Explorer press F5 The update is completed Concluding the update 1 Switch the unit on 2 Only at initial installation Perform a complete unit calibration gt 155 3 Only at initial installation Perform a white balance gt 190 70 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 4 General operating procedures Service Manual GALILEOS 4 3 Configuring the unit Configuring the unit Configuring the unit via the SIDEXIS Manager The X ray compon
26. Move unit to starting position press the R key H3 07 Remove the occlusal bite block H3 21 Close the door H3 23 Close swivel arm H3 24 Gettering in progress please wait H4 03 SIDEXIS XG is not ready for exposure make unit ready for exposure Easypad Multipad Ready for exposure no special display kV level and mAs are displayed Exposure not possible 110 Please wait Progress bar Ready for exposure in XXs XXs X RAY Active LED lights up on control panel If error message E1 11 20 is displayed on the control panel and or the remote control during the calibration process this does not necessarily indicate an equipment error This error message only indicates that the adjustment or calibration data of the unit is incomplete at this point Acknowledge the error message with the R key if applicable and continue the adjustment or calibration procedure For assistance with other help messages or error messages displayed during the adjustment or calibration process please refer to the section of these instructions entitled Messages gt 82 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 7 Adjusting calibrating the unit Service Manual GALILEOS 7 1 General information about unit adjustment and calibration Service Functions menu 8 Message window Tabs index cards submenus 61 25 665 D3437 D3437 076 01 15 02 06 2012 Diaphr
27. NOTICE Do not force on the adjusting knob Make sure the locking function of the adjusting knob works properly when setting the button in place Set the adjusting knob in place with the silicone ring Updating the ID label s rona N Model No 5997221 03437 Senai No XXXXXX a g 85KV 7 MA 0123 25 Al 85 IEC 522 P 100 ag ROHRE TUBE TOSHIBA OF 151R Mode Nc Seriat No XXXKXX Mos ec 336 93 7 MADE IN GERMANY This product complies with DHHS regulations 21 CFR Subchapter J applicable at date of manufacture Date of manufacture XXXXXX 20XX Sirona Dental Systems GmbH FabrikstraBe 31 D 64625 Bensheim Germany Tres comptes with CIOS re ianon 21 CFR subchapter J applicable at of Date of marefacture XXXXXX 29XX Fobrastrate 31 044625 Bensheim GERMANY gt Update the nameplate at the tube assembly cover To do so affix the supplied label as shown in the figure 61 25 665 D3437 D3437 076 01 15 02 06 2012 315 9 Repair Sirona Dental Systems GmbH 9 6 X ray tube unit Service Manual GALILEOS Cables and connectors for replacement of the X ray tube assembly 61 25 665 D3437 316 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 6 X ray tube unit A Cable L5 gt Socket J6 on board DX6 B Cable L6 gt Socket J2 J3 on board DX6 C Grounding cable gt Connector X304 on board DX6 D Cable L3 gt Connector X3 on
28. Risk of electric shock Always switch the unit off before connecting a measuring instrument or carrying out continuity checks NOTICE Risk of damage to boards Please observe the usual precautionary measures for handling printed circuit boards ESD Touch a ground point to discharge static electricity before touching any boards NOTICE Risk of damage to tube assembly Keep to the prescribed cool off periods if several exposures have to be taken to check a measured value IMPORTANT Select the correct current voltage type and adjust the measuring range to match the expected readings Calibrating the unit Unit calibration is described in detail in the chapter entitled Adjusting calibrating the unit gt 155 61 25 665 D3437 D3437 076 01 15 02 06 2012 369 10 Maintenance Sirona Dental Systems GmbH 10 2 Checking the height adjustment Service Manual GALILEOS Checking the height adjustment Check the threaded rod and motor for gt Perform a visual inspection of height adjustment motor A and abrasion spindle B for abrasion If significant abrasion is present Replace the height adjustment motor including spindle gt 290 Check whether the height adjustment gt Use the Up Down keys on the control panel to move the unit up and produces atypical running noises down through its entire adjustment range If the mechanics of the height adjustment is defective a speed
29. This Sor may ae cco e Please contact the Sirona Customer S gt 61 connection with other causal error s Service Center CSC to find out whether messages Please also observe the a bug fix by means of a software update is causal error message It appears f possible and perform such an update if only after you acknowledge the first error messa VOCES RIY ge Error code Description Actions required see E6 41 07 TTP time trigger protocol timeout Check the CAN bus S gt 128 error detected on slave side Check power supply of board DX11 The module was temporarily not measuring point 3 3 V on board DX1 see addressed by the master wiring diagrams Undervoltage on the master side If 3 3 V is present S gt 343 Procedure error in the software e Replace board DX11 Master DX11 receives no return A commanding from the module If 3 3 V is not present S gt 343 This error may also occur in Replace board DXT connection with other causal error messages Please also observe the causal error message It appears only after you acknowledge the first error message Error code Description Actions required see E6 41 08 General fault detected locally on e Check the CAN bus S gt 128 moaie stave Bide PAN controler e Check software versions on the info S gt 238 being reinitialized i screen or by running service routine S gt 61 S008 2 perform software update if i necessary S gt 61
30. arm cover gt Remove the ring motor as described in the chapter Replacing the ring motor gt 301 Removing the covers Removing the motor 61 25 665 D3437 D3437 076 01 15 02 06 2012 303 Sirona Dental Systems GmbH 9 Repair 9 3 Ring motor M1_3 Service Manual GALILEOS Replacing the pinion 1 Loosen the set screws A and pull off the defective pinion B 2 IMPORTANT Ensure that the pinion is seated in the coupling so that the set screws A are sitting on the flattened surface C of the pinion during subsequent tightening to prevent the pinion from turning Insert the new pinion 3 IMPORTANT Apply Loctite 242 to the set screws B before tightening Retighten the set screws A Installing the motor gt Reinsert the motor in the ring route the cable and connect the motor as described in the chapter Replacing the ring motor gt 301 Attaching the covers gt Reattach the covers 61 25 665 D3437 304 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 3 Ring motor M1_3 Laying of cables when replacing the ring motor 1 Lay the cable G parallel to cable L3 and secure it with the clamps 2 Plug connectors X804 I and X813 H into board DX1 What has to be done after replacing the ring motor M1_3 pinion 1 Check the function of the ring motor 2 Perform complete unit adjustment or calibration gt 155
31. booting 4 Repeat procedure and check function If the error occurs repeatedly e Replace remote control see Installation Instructions Error code Description Actions required see E3 42 31 Release button pressed during e see section Error analysis of X RAY S gt 142 power on control signal path gt 142 The hardware signal for radiation release is applied on board DX42 when the unit is switched on 114 61 25 665 D3437 D3437 076 01 15 02 06 2012 Location 71 Multipad board DX71 Error code Description Actions required see E6 71 01 General error during module e Please contact the Sirona Customer S gt 61 initialization Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary If the error occurs repeatedly S gt 343 e Replace board DX71 or Multipad S gt 307 Error code Description Actions required see E6 71 02 Invalid system data or uninitialized Please contact the Sirona Customer S gt 61 module storage data Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary e Acknowledge error and repeat procedure If the error occurs repeatedly S gt 343 e Replace board DX71 or Multipad S gt 307 Error code Description Actions required see E6 71 03 Invalid commandi
32. 0 0 0 0 0 0 0 DX7 0 0 0 0 0 0 0 DX81P 0 0 0 0 0 0 0 DX88 0 0 0 0 0 0 0 Display the CAN Bus trace Attention This page will take about 2 minutes to be displayed completely The CAN bus browser opens 2 Inthe lower area select the CAN bus link 61 25 665 D3437 236 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 6 S007 Error logging memory CAN Bus Trace TimeStamp Type ID Len Data 0183006 776 Rx 0x401 0x08 0x00 0x00 0x00 0x00 0x00 0x00 0x00 0x00 0183010 785 Rx 0x481 0x08 0x00 0x00 0x00 0x00 0x00 0x01 0x00 0x00 0183012778 Rx 0x181 0x08 0x00 0x00 0x00 0x00 0x00 0x02 0x00 0x00 0183020 802 Tx 0x001 0x08 0x18 0x27 0x09 0x19 0x01 Ox6f 0x00 0x00 0183022780 Tx 0x080 0x08 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0183024 803 Tx 0x100 0x08 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0183026 794 Tx 0x200 0x08 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0183028 781 Tx 0x280 0x08 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0183030 801 Tx 0x300 0x08 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0183032806 Tx 0x400 0x08 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0183034 798 Tx 0x500 0x08 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0183036 788 Tx 0x480 0x08 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0183038 802 Tx 0x180 0x08 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0183040 784 Tx 0x700 0x08 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0183042 784 Tx 0x780 0x08 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0183046 790 Rx 0x081 0x08 0
33. 00 V01 10 06 01 13 01 02 61 01 Start SW update Show k logfile 0x42 02 45 06 Dx4 v02 30 00 DX06 v02 88 00 0x07 v02 57 00 Dx07_LO 02 18 00 0x89 V01 10 06 Dx89_F1 01 13 01 v02 60 07 Close 61 25 665 D3437 D3437 076 01 15 02 06 2012 02 45 06 v02 30 00 v02 88 00 V02 57 00 02 18 00 V01 10 06 01 13 01 v02 61 01 Start SW update Show logfile i 4 General operating procedures 4 2 Updating the firmware Update mode You can select two different update modes via the tabs Automatic or Main version Automatic The software of all components is automatically updated to the latest software version The right window displays a list of the modules their installed software version and the latest software version offered by the update Main version The software can be upgraded or downgraded to the desired version This update mode is required for example if a replacement module arrives from the warehouse and features a newer release than the existing main release of the unit In this case a main version update to the overall system status displayed on the info screen must be performed for the corresponding component with the appropriate update file SUI The module is then reprogrammed The colored bars in front of the software releases indicate their validity see chapter entitled Using the update manager to check the program re
34. 01 15 02 06 2012 Sirona Dental Systems GmbH 10 Maintenance Service Manual GALILEOS 10 8 Checking the protective ground wires Checking the protective ground wires Potentially lethal shock hazard It is essential to switch the unit off and to wait at least one more 1 minute before beginning the check 1 Switch the line voltage off at the main switch of the building installation 2 Disconnect the power cable and the second protective ground wire from the building installation 3 Remove the bottom profile top tube assembly and bottom tube assembly covers Measuring setup for protective ground wire test A oo a mod wow A Ammeter B Power source C Measuring point A central protective ground wire D Voltmeter E Measuring points B E 61 25 665 D3437 377 D3437 076 01 15 02 06 2012 10 Maintenance 10 8 Checking the protective ground wires Check whether the ground wire resistance complies with the specifications 378 Sirona Dental Systems GmbH Service Manual GALILEOS A and B GNYE wire 0 10 AandC 2 Protective ground 0 1 Q wire A and D Housing DX32 0 20 AandE Tube assembly 0 20 housing J A power source with a current of at least 0 2 A a no load voltage of 24 V max and 4 V min is required 1 Connect the power source between the measuring points specified in the table for at least 5
35. 02 and SIDEXIS 2 5 1 or above 003 For GALILEOS Comfort Requires unit main software V 03 05 00 and SIDEXIS 2 5 1 or above For GALILEOS Compact Requires unit main software V 03 06 02 and SIDEXIS 2 5 1 or above V1 10 004 For GALILEOS Comfort Requires unit main software V 04 09 00 and SIDEXIS 2 5 3 or higher For GALILEOS Compact Requires unit main software V 04 09 00 and SIDEXIS 2 5 3 or higher 57 4 General operating procedures 4 1 Switching the unit on Sirona Dental Systems GmbH Service Manual GALILEOS General operating procedures Switching the unit on X rays Be sure to observe the radiation protection regulations applicable in your country gt No person may be positioned in the unit when it is switched on NOTICE Damage to the unit Check the room height before you raise the unit gt Ifthe room height is less than 2 27 m 89 3 8 or 2 30 m 90 1 2 for installation with the floor stand you must limit the maximum travel height gt 272 NOTICE Fluctuations in temperature can cause condensation to form in the unit Electrical components are destroyed by short circuits gt Do not switch the unit on until the temperature of the unit has adapted to the ambient temperature and the condensation has evaporated NOTICE The unit must not be switched on off constantly Constant switching on and off reduces the service life of
36. 04 To do this first press the Memory key LED above the R key lights up and then the R key Exit the service routine by pressing the arrow key above selection field 3 Switch off the unit 61 25 665 D3437 D3437 076 01 15 02 06 2012 16 17 18 19 20 21 22 23 24 25 26 27 28 29 61 25 665 D3437 D3437 076 01 15 02 06 2012 Wait for approx 1 minute Then switch the unit back on The error message E6 15 05 undefined system serial number is displayed Acknowledge the error message with the R key Error message E6 15 04 undefined activation data is displayed Acknowledge the error message with the R key Call the service menu gt 210 Call the service routine S008 3 check the serial number and confirm this if necessary gt 240 The unit serial number is located on the nameplate of the unit NOTICE If the serial number is incorrect exit the update process and contact the Sirona Customer Service Center Switch off the unit Wait for approx 1 minute Then switch the unit back on The message No Key should no longer appear Call the service menu gt 210 Call the service routine S017 and perform the unit configuration test step 2 15 gt 252 The board DX41 must be configured using the service routine 017 9 In units with a serial number of 1080 and above the switching plate configuration of the swivel arm must be set or checked using service routine S017
37. 06 2012 2 IMPORTANT Do not tighten the screw yet It should not be possible to rotate the acquisition unit Screw the new receptacle element on to the unit with screw D so that the laser localizer E is facing forward 3 Push the head fixation device F into the acquisition unit see operating instructions 4 Press the locking button A and push the flange B including the head fixation device backwards so that the light localizer E is exposed 1 CAUTION Keep a minimum distance of 100 mm between the eye and the laser Do not look directly into the laser beam Switch the laser light on using the light localizer button on the control panel 339 9 Repair Sirona Dental Systems GmbH 9 10 Head fixation device Service Manual GALILEOS 2 Align the acquisition unit To do this move the rotary knobs on the head fixation device to a vertical position Then align the acquisition unit so that the laser light is shown in the middle of the vertical knobs of the head fixation device and the bite holder 3 Press the locking button A and push the flange B including the head fixation device F back to the front so that the bore hole C is exposed Tighten the screw D firmly IMPORTANT The acquisition unit should not be turned when pushing the head fixation device back and tightening the screw 4 Switch
38. 076 01 15 02 06 2012 310 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 6 X ray tube unit Loosen the two lower screws A approx 2 to 3 turns Push the diaphragm B upward and then toward the front Unscrew the two sleeves C with washers Remove the aluminum filter D Remove the primary diaphragm F OO e ONS Keep diaphragm B safe and store primary diaphragm F and aluminum filter D in a safe place IMPORTANT These components will be reused when attaching the diaphragm unit to the new tube assembly Type 2 diaphragm serial no 2201 and higher Oil leakage The screws E from the ring X ray window must not be loosened as otherwise oil may leak out Loosen the two lower screws A approx 2 to 3 turns Push the diaphragm B upward and then toward the front Unscrew the two sleeves C with washers Remove the aluminum filter D SV PON aS Keep diaphragm B safe and store aluminum filter D in a safe place IMPORTANT These components will be reused when attaching the diaphragm unit to the new tube assembly 61 25 665 D3437 D3437 076 01 15 02 06 2012 311 9 Repair Sirona Dental Systems GmbH 9 6 X ray tube unit Service Manual GALILEOS 1 Loosen the two lower screws A approx 2 to 3 turns 2 Push the diaphragm B upward and then toward the front Removing the defective X ray tube assembly
39. 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 8 X ray detector Installing the X ray detector 1 Hook the new x ray detector into holder K from above using the ring on the unit The dead weight of the X ray detector will cause it to tilt into the correct position 2 Secure it in place using the screw J ooososeodoooo0oo0o000 f ooesosooosesocsse 3 CAUTION Risk of injury The cover plate may have sharp edges Carefully pull the cover plate K upwards to remove it from the X ray detector 4 Plug cable L13 from the ring onto connector X201 on PCB DX89 and use the two screws to secure it 61 25 665 D3437 D3437 076 01 15 02 06 2012 321 9 Repair Sirona Dental Systems GmbH 9 8 X ray detector Service Manual GALILEOS ixmsxt2 D 1x953 M amp 1x953 N 1x95 5 Connect the grounding cable from the ring with screw G as well as with washer L serrated washer M and contact washer N 61 25 665 D3437 322 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 8 X ray detector 6 NOTICE Ensure that cable L13 is correctly laid in the nut O of the X ray detector Route the cable L13 as illustrated in th
40. 132 Fast flashing Fast flashing CAN error no physical communication with CAN bus possible there is probably a short circuit in the CAN cable or on the board of a module e Disconnect CAN cables one after the other set jumpers to inner position until the CAN bus functions again V700 and V701 flash slowly Replace faulty module Off Fast flashing CAN error CAN bus TTP time trigger protocol disturbed by defective constantly transmitting board bus heavy e Disconnect CAN cables one after the other set jumpers to inner position until the CAN bus functions again V700 and V701 flash slowly Replace faulty module Off Off System did not power up DX11 Switch unit off and on again and wait until end of power on time 132 Jumper positions in the CAN bus The jumpers are located on board DX1 at sockets X302 X303 X306 X307 X309 X500 and X503 see also wiring diagrams If a cable is connected to the socket the corresponding jumpers must be set to the outer position If no cable is plugged in the jumpers must be set to the inner position If a jumper is set to the inner position without a cable plugged in the CAN bus is interrupted at this location Modules located behind this location can no longer be connected to the CAN bus and therefore do not function 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS
41. 208 8 3 Basic operating procedures in the Service MENU ccccceceeeeeeeeeeeeeeeeeeees 210 8 3 1 Activating the service Menu c ceecceecceeeeeeeeeeececeeeceeeeseeseeeeeeaseeenes 210 S33 A BaSVOAi2 25 Ae ie Oe a ha ae eee Se Ae see 210 8 3 1 2 Multipad ennea oeta ht ees Aral ECan cate et aad ead 212 8 3 2 Selecting service routines and test StepS ccccccecceeeeeeeeeeeeeeees 213 8 3 2 1 Selecting a service routine cccccceeeeecccieeeeteeeesneceeeeeeeeseseedeeenseenee 213 8 3 2 2 Selecting a test step bas cives cease ieem eh Reese elt eset 214 8 3 2 3 Service routines with Security ACCESS 2 eee ceeeeeeeeeteeeeeeeeeeeeeees 215 8 3 3 Select parameters cccccccceccccecccecceccceccceceeeeeeeceeeeeeeeeeeeeeeeeeeeeeeeess 217 8 34 Saving patametetS aerisire arenira aii daca nant 219 8 3 5 Exiting the test step and service routine 0 0 0 0 eee eeeeeeeeeeeeeeeeeeeeeeeeeees 220 8 4 002 Radiation without rotary movement selectable kV mA level and 221 maximum radiation time 84 1 S002 Test Step Dreis i See eden 221 8 5 S005 General X ray tube assembly service eee eeeeeeeeeeeeeeeeeeenteeeeeeeeeeeeees 223 8 91 S005 TESt Sl6 pi ics cecetessteravvadiedestidoccces causa usveccevel EA KA eeN EERTE 223 8 5 2 S005 Test Step 4 2 ccccccccccccecccecccceccecececeeeeeeeeeeeeeeceeeeeeeeeeeseeeeeeerenss 225 8 5 3 S005 Test Step 5 2 ccccccccccccceccccecceeeceeeeeeeeeeeeceeeeeeeneeeeeeeeee
42. 25 665 D3437 296 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 2 Height adjustment motor M1_4 spindle Installing the new motor Install the height adjustment motor in the reverse order of removal Please observe the following Nuts When fastening the motor make sure that all three screws are tightened uniformly and protrude approx 3 mm out of the nut Acorn nuts If acorn nuts have been installed in the unit turn the acorn nuts to the end stop CAUTION Do not forget to reattach all connectors or cables route them in their original position and reattach all cable ties and cable clamps Make sure that none of the cables are crushed by the cover plates of the DX32 connection box Final work With the height adjustment service kit 1 After reinstalling the spindle screw the first of the two screws E back into the spindle holder 2 Then remove the locking pin and screw in the second of the two screws E 3 Attach the cover D Remove the clamp A 5 Only then should you check the travel function of the slide gt Without the height adjustment service kit Attach the spindle holder F with the two screws E Attach the cover D Reattach the upper limit stop to the previously marked position Only then should you check the travel function of the slide PONS 61 25 665 D3437 D3437 076 01 15 02 06 2012 297 9 Repair Sirona Dental Systems GmbH 9 2 Heigh
43. 3 Exit the service routine gt 220 61 25 665 D3437 248 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS S012 CAN bus service 8 Service 8 10 S012 CAN bus service SR SHZ Function 012 CAN bus service 012 1 No Presence display of modules 61 25 665 D3437 D3437 076 01 15 02 06 2012 SR service routine SHZ security access IMPORTANT The CAN bus service is not yet implemented for the module DX11 S012 Test step 1 Presence display of modules CAN bus events e Presence code behind the counter value P module present L module lost Selection field Parameter Display Range of values 1 Subassembly DX1 DX88 2 e Counter value of gt Call service routine 8012 1 gt 213 249 8 Service Sirona Dental Systems GmbH 8 10 S012 CAN bus service Service Manual GALILEOS Checking the module gt Use the arrow keys A in selection field 1 to select the required module The counter value of the CAN bus events processed so far since the last switch on of the unit of the selected module is displayed in selection field 2 with the presence code of the module L or P see table Once the module has been selected the T key B Easypad or the LED above the T key B Multipad lights up 61 25 665 D3437 250 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service S
44. 3 V not present S gt 343 return commanding from the module e Replace board DX1 This error may also occur in connection with other causal error messages Please also observe the causal error message It appears only after you acknowledge the first error message 61 25 665 D3437 120 D3437 076 01 15 02 06 2012 Error code Description Actions required see E6 89 08 General fault detected locally on e Check the CAN bus S gt 128 module slava side CAN Controller e Check software versions on the info S gt 238 being reinitialized screen or by running service routine S gt 61 S008 2 perform software update if 3 necessary S gt 134 Check board DX89 replace if necessary S gt 343 e Please contact the Sirona Customer Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary Error code Description Actions required see E7 89 10 Module is stuck in bootloader stage Check operating status of board note S gt 134 LED states If the board remains in the bootloader stage S gt 61 e Run software update S gt 134 e Check board DX89 replace if necessary S gt 343 Error code Description Actions required see E7 89 12 Unit is not ready for operation This error is a sequential fault e Restart the unit 1 Switch off the unit 2 Wait 1 minut
45. 6 Troubleshooting Sirona Dental Systems GmbH 6 9 Fault diagnosis of the X ray detector and on board DX89 Service Manual GALILEOS Image memory test Operating voltages 28V 24V 5V 3 3V Operating voltages on DX89 and X ray detector 154 LED on Image memory test is o k LED off Based on the PLL_FPGA LED check whether the FPGA on DX89 has started properly e PLL_FPGA LED off see the action under LED OFF for PLL_FPGA e PLL_FPGALED on Replace board DX89 gt 346 LEDs of operating voltages The unit must be switched off before disconnecting any plugs or cables The four LEDs are powered directly by the four operating voltages and all must light up after the system start If this is not the case check connector X201 for firm seating If the connector is OK and the LEDs still do not light up then replace the X ray detector gt 319 Supply voltage in V Light emitting diode LED 28 V101 24 V109 5 V108 3 3 V107 These two LEDs PLL_FPGA and PLL_CCD must light up following the system start If this is not the case replace the x ray detector gt 319 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS 61 25 665 D3437 D3437 076 01 15 02 06 2012 7 Adjusting calibrating the unit Adjusting calibrating the unit X rays When performing the following tests be sure to observe the radiation protection regulation
46. 665 D3437 D3437 076 01 15 02 06 2012 334 9 Repair 9 9 Facescan Sirona Dental Systems GmbH Service Manual GALILEOS Installing the new PoE module 1 Putthe PoE module P on the thread bolts H of the mains filter plate J the serrated washer must sit behind the fitting plate of the PoE module 2 Screw down the PoE module P to the place provided in the stand with two screws K and a nut L IMPORTANT The locking catch of the RJ45 plug of Ethernet cable L77 is secured with adhesive tape N gt Remove the adhesive tape N from the RJ45 plug of Ethernet cable L77 3 Plug Ethernet cable L77 into the POE socket of the PoE module 4 To fasten slide the cable cover l downwards again as far as possible S NSS lt N EE 335 61 25 665 D3437 D3437 076 01 15 02 06 2012 9 Repair Sirona Dental Systems GmbH 9 9 Facescan Service Manual GALILEOS 5 Pull cable L7 from the SC SC socket 6 Plug cable L7 into the BU2 socket of the PoE module 7 Plug cable L71 into the BU1 socket of the PoE module 8 Plug cable L71 into the SC SC socket IMPORTANT Cable L76 must be routed behind the mains filter plate J 9 Route cable L76 behind the mains filter plate to the terminal K1 61 25 665 D3437 336 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 9 Facescan 10 Connect cable L77 BN brown to terminal K1 1 orange
47. 8 5 S005 General X ray tube assembly service Service Manual GALILEOS If the tube type is invalid 1 Confirm the security access gt 215 Selection field 1 shows the indicator number of the invalid tube type detected by the X ray tube assembly 2 Use the arrow keys in selection field 1 to select the code for the tube type see table The Memory key B Easypad or the LED above the Memory key B Multipad lights up 3 Save the selected parameter gt 219 4 Exit the service routine gt 220 After calling the service routine S005 1 again the security access is no longer requested and the stored tube type is displayed 61 25 665 D3437 224 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 5 S005 General X ray tube assembly service S005 Test step 4 Fan test Selection field Code Function 1 00 Fan off 01 Fan on Factory setting 1 Call service routine S005 4 gt 213 A 2 Use the arrow keys A in selection field 1 to select the code 01 see table 3 Confirm your selection by pressing the R key B The fan starts up 4 Check the fan for running noise 5 Exit the service routine gt 220 Upon exiting the service routine the fan is automatically switched off again 61 25 665 D3437 D3437 076 01 15 02 06 2012 225 8 Service Sirona Dental Systems GmbH 8 5 S005 Gener
48. 82 5 1 Help messages iinis eh ene tes ceed iaoa eoe haaa a At haerai oeta 82 5 2 System MESSAGES 0 eeceeccceccsccceeceaaceaeceaeceeceeececececeeeceeeeceeceeeeeeeeeeeeeeeeeeeteess 84 5 3 Status messages and displayS eee eeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeenaaeeeeeeeeeenaea 84 5 4 Error message Sienen e tonsa tet E redcereesiients 85 5 4 1 Error code Ex yy ZZ eide e nerea aee A Aaa aaka ARARE 85 D42 EX Emor type erpe nities Gan a Si 85 543o YY Cocaine e a a E tees 87 5 4 4 General handling of error MESSAGES cccceeeceeeeeeeeeeeeeeeeeeeeeeteees 87 5 5 _ Listof error MESSAGES aarre aerea eiea AnS AEEA E R EPA ASEENI EEKAN TENAS 88 5 5 1 Location 06 Tube assembly DX6 cccccccceecseceneneeeeteeeeeceeneeeeees 88 5 5 2 Location 07 asypadiDAs scccicichteietieseMadacdacate tisastenticdadeeataeuckesbiainae 94 5 5 3 Location 10 System NardWwales ccsvccsseiat visstin devas dentesondeateele eecueuedes 98 5 5 4 Location 11 Power PC Board DX11 ccccccccssesseeseceeeeeeseeeesaeeees 101 5 5 5 Location 12 CAN DUS ccccceccccecceceeseeseececeeeeseeseseseseceeeeeeseesaeaeanes 103 5 5 6 Location 14 Digital extension SIDEXIS XG 104 5 5 7 Location 15 Configuration Update cccceccecceeeeeeeeeeeeeeeeeeeeetteeees 106 5 5 8 Location 41 Media interface Card ccceccccecceceeseseseceeeeeeseesaeaeenes 107 5 5 9 Location 42 Remote control cccccccseecccccee
49. D3437 308 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 5 Control panel Laying of cables when replacing the user interface Easypad 1 Plug the green cable L10 A into connector X103 on board DX7 2 Plug the gray cable L9 B into connector X102 on board DX7 Multipad 1 Plug the green cable L10 D into connector X103 on board DX71 2 Plug the gray cable L9 C into connector X102 on board DX71 61 25 665 D3437 D3437 076 01 15 02 06 2012 309 Sirona Dental Systems GmbH 9 Repair 9 6 X ray tube unit Service Manual GALILEOS X ray tube unit Potentially lethal shock hazard It is essential to switch the unit off and to wait at least another 4 minutes before starting the repair or taking off a cover panel Replacing the X ray tube assembly Only for Type 3 diaphragm Pull off the adjusting knob with the silicone ring 2 Remove the Front tube assembly and Rear tube assembly covers gt 42 1 Turn the rotating element so that the tube assembly as viewed from the front is located on the right side of the unit i e not above the swivel arm 2 Remove the old diaphragm unit Removing the covers 1 Installing the diaphragm unit Type 1 diaphragm up to serial no 2200 Oil leakage The screws E from the ring X ray window must not be loosened as otherwise oil may leak out 61 25 665 D3437 D3437
50. Dental Systems GmbH 3 2 Hardware Service Manual GALILEOS Remote control Component Designation Function Board DX42 Display board for remote control Facescan FACESCAN Component Designation Function Boards FACESCAN Modular board PoE Power supply board 61 25 665 D3437 24 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 3 Unit description Service Manual GALILEOS 3 2 Hardware Cabling overview IcPz000 59224 Lo zsa P2000 5923 2 P2000 922484 bY O 2000 5922484 DX32 X2 DX6 X3 DX32 X1 DX1 X100 4 8 A Ee ed 3 ct x 3 5 m gt a A Power switch Ww Line filter C Wago terminal 61 25 665 D3437 D3437 076 01 15 02 06 2012 25 3 Unit description Sirona Dental Systems GmbH 3 2 Hardware Service Manual GALILEOS aie ae l 00 DX1 X104 A ol i 19P2009 5922559 TP Ss ra DX7 X102 3 L10 36 DX1 X302 mr T P 5922567 70_P2000 5922567 HLL DX7 X103 L15 DX7 DX71 ae DX1 J309 Fs DX6 J5 A Power switch 6
51. GALILEOS Error code Description Actions required see E3 71 30 Up down keys pressed on power e Restart the unit oe 1 Switch off the unit E3 71 33 Light localizer key pressed during 2 wait 1 minute power on 3 Switch unit ON making sure that the E3 71 34 T key pressed during power on Multipad is not actuated during boot up E3 71 35 R key pressed during power on 4 Repeat procedure and check function Service key actuated during power If the error occurs repeatedly S gt 343 E3 71 36 on e Replace board DX71 or Multipad S gt 307 Memory key actuated during power E3 71 37 on Program selection key actuated E3 71 38 during power on Radiation time key actuated during E3 71 39 power on kV mA key actuated during power E3 71 40 on Patient symbol pressed during E3 71 41 power on 61 25 665 D3437 118 D3437 076 01 15 02 06 2012 Location 89 X ray detector This error may also occur in connection with other causal error messages Please also observe the causal error message It appears only after you acknowledge the first error message Please contact the Sirona Customer Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary Error code Description Actions required see E6 89 01 General error during module e Please contact the Sirona Customer
52. Iris calibration There is no valid iris calibration data on this machine Image acquisition Save values v The Service functions menu is called gt 161 1 Click the ris tab The corresponding tab card is selected pea a en 5 2 Click onthe mage acquisition button SIDEXIS XG makes the unit ready for exposure gt 162 Service routine S010 10 is displayed on the control panel 3 Press the R key to move the unit back to the starting position 4 Press the release button Press and hold down the button until the exposure is complete the preview image is displayed in the exposure window and the acoustic signal indicating the end of the exposure double beep sounds if it has been configured IMPORTANT This process takes approx 2 3 minutes The unit transfers the acquired images to the Reconstruction and Control Unit RCU This process can take 2 3 minutes Once the transfer is complete the evaluation of the sensor calibration is displayed in the message window If the information in the message window indicates that calibration is not OK and or not possible keep repeating the procedure starting with Step b until calibration is OK and or possible If you have repeated the procedure three times and still have not attained a positive result please contact the SIRONA Customer Service Center CSC 5 If the calibration is OK or possible click the Save va ues button
53. Jumper outside XXXX L Ox E __ Jumper inside Ox Ox LJ _ 61 25 665 D3437 D3437 076 01 15 02 06 2012 6 Troubleshooting 6 2 Checking the CAN bus If the jumpers are set to the outer position the module is connected i e the connector is plugged in If the jumpers are set to the inner position the module is not connected i e the connector is not plugged in 133 Checking the boards Check operating state of the board Measure voltages Visual inspection Is the board undamaged Do the LEDs indicate ormal operation Are the voltage levels OK Assign fault to the Assign fault to the board DX1 or board DX1 or other connected other connected board or board or component e g component e g cable replace cable replace component if component if necessary necessary Install replacement board and check unit function Does the unit work Board is OK Continue troubleshooting Replace board or according to list of faults component chapter Messages 61 25 665 D3437 134 D3437 076 01 15 02 06 2012 Important LEDs on the boards see also wiring diagrams Board LEDs Normal operation Malfunction Bootloader DX1 V100 lit not lit V101 lit not lit V108 lit no
54. Manual GALILEOS 8 Service 8 6 S007 Error logging memory S007 Test step 2 Clearing error logging memory 1 Call service routine S007 2 gt 213 The system s readiness to clear the memory is indicated by the display message FFFF in selection field 1 If the error logging memory does not contain any data 0000 is displayed 61 25 665 D3437 232 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 6 S007 Error logging memory 2 To clear the memory press the Memory key A R key B Easypad or LED above the R key B Multipad lights up followed by the R key B 61 25 665 D3437 D3437 076 01 15 02 06 2012 233 8 Service Sirona Dental Systems GmbH 8 6 S007 Error logging memory Service Manual GALILEOS Once the memory has been cleared the message 0000 is displayed in selection field 1 3 Exit the service routine gt 220 61 25 665 D3437 234 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 6 S007 Error logging memory S007 Test step 5 Enabling CAN bus logging in the web interface NOTICE This service routine may only be called up subject to the approval of and with the support of the Sirona Customer Service Center CSC Selection field Parameters Function 1 WS CAN bus off Logging off WS CAN bus on Logging on WS CAN bus ex on Extended logging
55. One 16 28 Ventilation Slots meisneri paraa ieee anar EAA N ESERE i 16 2 9 Qualifications of service personnel cceecsecceeeceeeceeeeeeeeeeeeeeeeeeeeeeeeeeess 16 2 10 Radiation protection cccceccceeeeeceeeeeeeeeceeaeeeaeeeaaeeaaeeaaeeaeeeaeeeaeeeaeeeaeeaeeees 16 BV Ly Safety checks sere eet dae ted inhalants hie a ee a e aa ATT Eaa o Een 16 3 UnitdescriptioNie ia sie e Aa EEE a E EAA RR 17 3 1 Unit classes and versions seiscccescccsscasteicibeaiecaeldknedse sees qlen deaceenndaendiilas taveeiae 17 3 1 1 GALILEOS Comfort and GALILEOS Compact cceeeeeeee 18 Bi HardWare ieie ea eect pcan aaea Sy AE AE EA EEE Sea 19 3 2 1 Information on the unit cs cite shar cccehevedadantcueeesvecadeucerenedabandaceneacentecenas 19 3 2 2 Installation versionS ssssssssseeeresrsrseesecsenesriariororrnrnninenrinnnnnnneeseeeree 20 3 2 3 Modules and components 2 ccccceececceceeeeeeeeeseeecenceeeeeeseseeceeeaeeeeee 21 323 Tr Slide ra aa a aa ee ages a a a aA EE RAE 22 Eo EP ike WO Aaa a a E ORE ere Cee ne ne ee eee ee 23 Be2 oe REMOTE Controler rerien os aaa aar aa AAAA ELETA E TAVARE OEE cease 24 De OA PACES CAN r e a E tes 24 3 2 4 Cabling ON STV WN 2s ta ae oes aaa S ahead Counted ee ENNAN NNa 25 61 25 665 D3437 D3437 076 01 15 02 06 2012 3 3 2 5 Board Pnolosvecni cia uleuiicednanusil aes ae aisha 31 3 2 5 1 Boards In the SNAG sisi ccvvcccvis tienteteneives waves seduvervdaie dose se
56. R key D The selected setting is saved to non volatile memory Multipad You want to run service routine S017 6 to save the selected option Remote control enabled V The LED above Memory key C lights up 1 Press Memory key C The LED above the R key lights up 2 Press R key D The selected setting is saved to non volatile memory 219 D3437 076 01 15 02 06 2012 8 Service Sirona Dental Systems GmbH 8 3 Basic operating procedures in the service menu Service Manual GALILEOS Exiting the test step and service routine Easypad menu for selecting service routines Touch the Service key A or the double arrow key B to go back to the Touch the double arrow key B in the service menu to go back to the main menu 01 17 n gt gt Exception Service routine S017 6 In service menu S017 touch the Service key A to go to the next test step in the service routine Multipad Press the Service key A or the up arrow key above selection field 3 B to go back to the menu for selecting service routines In the service menu touch the up arrow key above selection field 3 B to go back to the main menu Exception Service routine S017 In service menu S017 press the Service key A to go to the next test step in the service routine 61 25 665 D3437 220 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 4 S002 Radiation
57. Remove cable L2 from terminal X100 and pull it downwards from out of the connection box F Remove connector X1 from board DX32 Loosen the two left hand screws G Loosen the two screws on the right I and remove the connection box including board DX32 61 25 665 D3437 D3437 076 01 15 02 06 2012 295 9 Repair Sirona Dental Systems GmbH 9 2 Height adjustment motor M1_4 spindle Service Manual GALILEOS Replacing the height adjustment motor spindle Removing the spindle 1 Loosen the two screws E on the spindle holder F if you have not already done so gt 290 2 Turn spindle holder D with an 18 mm A F socket wrench counterclockwise until the motor comes to rest on the limit stop and spindle L has been turned all the way out of the motor 3 Remove the straight pin K 4 Remove the spindle L Tip First pull spindle L downward along the motor and then diagonally upward and out of the unit Removing the defective motor 1 Pull the pulse generator cable connector X402 off board DX1 2 Detach the motor cable from the cable harness and carefully pull it out of the stand 3 Pull the motor connecting cable off of the filter 4 Loosen the three screws M 5 Remove the motor while carefully pulling the motor cable out of the stand gt Attach the new rubber pads N to the new motor They are included in the scope of supply of a new HA motor 61
58. Remove the energy chain along with the mains filter plate from the stand Installing the new energy chain 1 Lead the energy chain up through the stand from its base 2 Insert new energy chain 1 in the stand IMPORTANT Assembly instructions gt Pay attention to the energy chain s rolling direction 61 25 665 D3437 360 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 13 Cable IMPORTANT Possible assembly errors The one end piece has been removed to make the new energy chain easier to lay The missing end piece is enclosed gt Attach this end piece to the energy chain only when the energy chain has been laid in the stand 3 Screw the new energy chain down in the position marking of the old energy chain Screw down the mains filter plate Screw the power cable A to the terminal K1 and the strain relief 6 Lead the external PE cable B over the mains filter plate through the ferrite core twice C a hose 7 Screw the external PE cable B down on the ground bolts D 61 25 665 D3437 D3437 076 01 15 02 06 2012 361 9 Repair 9 13 Cable Sirona Dental Systems GmbH Service Manual GALILEOS Replacing cables Switch the unit off before you start replacing cables or removing connectors NOTICE Be careful not to twist the cables or kink the fiber optic light guides when installing them Always che
59. S gt 61 initialization Service Center CSC to find out whether S gt 141 a bug fix by means of a software update is possible and perform such an update if S gt 362 necessary S gt 134 Check cable L27 L28 replace if S gt 343 necessary e Check board DX89 replace if necessary Slama e Check x ray detector replace if necessary As of X ray detector serial number 5000 cable L28 can no longer be replaced individually Error code Description Actions required see E6 89 02 Invalid system data or uninitialized Please contact the Sirona Customer S gt 61 module storage data Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary e Acknowledge error and repeat procedure If the error occurs repeatedly S gt 141 e Check cable L27 L28 replace if S gt 362 necessary S gt 134 e Check board DX89 replace if necessary S gt 343 Check x ray detector replace if S gt 319 necessary As of X ray detector serial number 5000 cable L28 can no longer be replaced individually Error code Description Actions required see E6 89 03 Invalid commanding or control data Check the CAN bus S gt 128 S gt 61 61 25 665 D3437 D3437 076 01 15 02 06 2012 119 Error code Description Actions required see E6 89 04 Data transfer error
60. Software version V03 07 02 Version 13 Full revision of the manual including new document structure New version of board DX1 DX11 and new X ray detector integrated Service routine 05 3 added Name change GALILEOS and GALILEOS GAX5 changed to GALILEOS Comfort and GALILEOS Compact Version 14 Software version V04 07 00 and V04 07 01 added Version 15 Software version V04 09 01 added Facescan added 61 25 665 D3437 386 D3437 076 01 15 02 06 2012 We reserve the right to make any alterations which may be required due to technical improvements Sirona Dental Systems GmbH 2012 Sprache englisch Printed in Germany D3437 076 01 15 02 06 2012 A Nr 115711 LESS aaa Sirona Dental Systems GmbH in the USA FabrikstraBe 31 Sirona Dental Systems LLC Order No 61 25 665 D3437 64625 Bensheim 4835 Sirona Drive Suite 100 Germany Charlotte NC 28273 Www sirona com USA
61. Start SW update Show logfile The update is started A message box informs you when the update process is completed NGSwUpdat 11 Confirm the update by clicking the OK button peage A message in the software manager notifies you that a unit restart is required to activate the software update you performed A SW update finished 61 25 665 D3437 D3437 076 01 15 02 06 2012 63 4 General operating procedures 4 2 Updating the firmware 64 Sirona Dental Systems GmbH Service Manual GALILEOS NOTICE Effectiveness of the software update The unit must be restarted after every software update The new DX11 version will not run until the unit has been rebooted see also chapter Measures following replacement of boards gt 347 Any errors with the consecutive numbers 01 03 04 06 or 07 displayed immediately following the software update may be ignored If these messages appear again after the unit is rebooted perform troubleshooting as described in the section entitled Error messages gt 85 If anything conspicuous occurs in connection with unit handling on completion of the software update and restart of the unit please repeat the software update as the first measure 12 Click on the Show logfile button and use the log files to check whether the update was successfully performed If it features entries such as Update of DXxx failed please perform the update again Repeat this procedure a
62. System software version V03 03 01 or higher IMPORTANT Exchange is only possible within the same system class e g the DX11 must come from a GALILEOS Comfort unit if it is to be installed in a GALILEOS Comfort unit After inserting the board you must reconfigure the IP address to match the IP address of the existing X ray component Before you set the unit to anew IP address make sure that the IP address you re assigning has not been assigned to any other unit 1 Switch the unit on Do not acknowledge any error messages at this time 2 Install the current version of the SIDEXIS XG software V2 0 or higher If the current version of SIDEXIS XG is a patch version the previous official main version of SIDEXIS XG must be installed before you can install the current version 3 Perform a software update to version V03 03 01 or higher via automatic update gt 61 If you are using a DX11 that already has the same software status as the overall system a repeated software update to this status must be performed in order that an administrative entry can be made in the memory of the DX11 4 Switch the unit off Wait for approx 1 minute Then switch the unit back on The error message E6 15 05 undefined system serial number is displayed The message No Key is displayed on the Easypad Acknowledge the error message with the R key Call the service menu gt 210 Call service routine S008 3 gt 240 Enter
63. adjustment down An acoustic signal must be audible If no acoustic signal sounds Replace board DX1 gt 342 Checking the fan and temperature sensor Check whether the fan is functioning gt Check the function of the fan using service routine S005 4 gt 225 Ifthe fan is defective Replace the fan gt 318 Check whether the temperature sensor gt Read the temperature in the single tank with service routine is supplying plausible values 005 5 gt 226 Ifthe displayed temperature reading is not plausible Replace the tube assembly Checking the cables for damage Check whether the cables feeding the gt Perform a visual inspection of the power cable protective ground unit are OK wire control cables and data cables If cables exhibit external damage Replace the respective cable gt 362 61 25 665 D3437 D3437 076 01 15 02 06 2012 371 10 Maintenance Sirona Dental Systems GmbH 10 5 Checking the idling rollers Service Manual GALILEOS Checking the idling rollers Check whether the idling rollers A are gt Manually turn the ring B and check it for smooth and easy OK movement If the ring does not move smoothly and easily Remove the housing covers and check the idle rollers A for dirt and foreign particles Clean and remove foreign particles if necessary 61 25 665 D3437 372 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 10 Maintenance Service Manual GALI
64. bug fix by means of a software update is possible and perform such an update if necessary Error code Description Actions required see E6 71 07 TTP time trigger protocol timeout Check the CAN bus S gt 128 error detected on slave side Check power supply of board DX11 The module was temporarily not measuring point 3 3 V on board DX1 see addressed by the master wiring diagrams Undervoltage on the master side If 3 3 V is present S gt 343 Procedure error in the software e Replace board DX11 Master DX11 receives no return r 3 5 commanding from the module If 3 3 V is not present gt 343 This error may also occur in Replace board Di connection with other causal error messages Please also observe the causal error message It appears only after you acknowledge the first error message Error code Description Actions required see E6 71 08 General fault detected locally on e Check the CAN bus S gt 128 moguls islaye SIGE CAN COnU ONCE I Check software versions on the info S gt 238 being reinitialized screen or by running service routine S gt 61 S008 2 perform software update if i necessary S gt 343 e Replace board DX71 or Multipad S gt 307 e Please contact the Sirona Customer Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary Error code Description Actions required see E7 71 10 Module is
65. cancelation Termination state Reason for ending exposure Value 0 Exposure completed Value 1 Exposure cancelation by user Value 2 Exposure cancelation due to internal error Imagetransfer started Start of image data transfer Imagetransfer stopped End of image data transfer Logbook started Corresponds to unit switch on Image state switched to Released Exposure has been delivered to SIDEXIS XG and confirmed by SIDEXIS XG Other entry data which document the occurrence of a rescue event include e Image state switched to Rescue e Rescue request Sidexis Error e Rescue request Sidexis TrackEpilogue e Rescue request Sidexis Timeout These entry data may also occur after Recording stopped or Cancel and indicate an exceptional circumstance You can supply important information for error diagnosis in coordination with the Sirona Customer Service Center Entry data Error E6 07 06 Error code ERR_DX7_TTP_LOST Clear text error display 126 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 6 Troubleshooting Service Manual GALILEOS 6 1 Error logging memory Entry data DeviceError DEV_DX42 Name of module to which the message refers Byte 0 7 Detailed error bytes for an error 0x10 0x00 0x00 0x00 0x00 0x00 0x00 occurrence 0x00 Entry data SidErr ERR_SOCKET_ERROR Detail of network error for Sirona only Error Sidexis SockErr Detail of network error for Sirona only Ent
66. cover on until it snaps in place 61 25 665 D3437 196 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 7 Adjusting calibrating the unit Service Manual GALILEOS 7 6 Mechanical adjustments Diaphragm adjustment Preparations 1 With the Type 3 diaphragm Pull off the adjusting knob with the silicone ring 2 Remove the Front tube assembly and Rear tube assembly covers Adjusting the type 1 diaphragm up to serial no 2200 1 Loosen both screws A slightly approx 1 turn 2 Adjust the diaphragm position by using screws B horizontal adjustment and C vertical adjustment Depending on the adjustment direction of the diaphragm it may be necessary to loosen the corresponding locknuts E slightly before the adjustment CW rotation of screws Moves the diaphragm to the right or upwards CCW rotation of the screws Moves the diaphragm to the left or downwards IMPORTANT To measure the shift refer to the gap between the plastic support and the lead diaphragm D 3 Retighten screws A and locknuts E 4 Repeat the diaphragm exposure gt 167 61 25 665 D3437 D3437 076 01 15 02 06 2012 197 7 Adjusting calibrating the unit Sirona Dental Systems GmbH 7 6 Mechanical adjustments Horizontal and vertical diaphragm adjustment 198 Service Manual GALILEOS Adjusting the Type 2 diaphragm serial no 2201 and higher 1 Loosen both screws A sl
67. current unit height is displayed in selection field 1 in this menu You can then set the unit to the desired height using the UP DOWN keys on the control panel Press the Service key or the double arrow key Easypad or the arrow key above selection field 3 Multipad to exit the height adjustment menu 61 25 665 D3437 156 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 7 Adjusting calibrating the unit Service Manual GALILEOS 7 1 General information about unit adjustment and calibration General information about unit adjustment and calibration Start by checking the mechanical unit adjustment This step is a prerequisite for the subsequent adjustment and calibration of the unit Please adhere to the following order when adjusting and calibrating the system Diaphragm exposure Type 1 Type 2 gt 167 diaphragm or diaphragm exposure Type 3 diaphragm gt 171 Checking the radiation field gt 178 Dosimetry gt 179 Sensor adjustment gt 181 Iris adjustment gt 182 Shading calibration gt 183 Distortion calibration gt 185 Geometry calibration gt 187 Tip It may be helpful to use the SIDEXIS XG coloring function to evaluate the image 61 25 665 D3437 D3437 076 01 15 02 06 2012 157 Help messages Status messages 158 Displays and help messages during adjustment calibration The most frequent help and status messages during calibration are listed below H3 01
68. detector replace if S gt 319 necessary As of X ray detector serial number 5000 cable L28 can no longer be replaced individually Error code Description Actions required see E7 89 25 Image memory error e Restart the unit 1 Switch off the unit 2 Wait 1 minute 3 Switch unit on 4 Repeat procedure and check function e Run software update S gt 61 e Check that the memory modules on board S gt 343 DX89 are firmly fixed replace board DX89 if necessary Error code Description Actions required see E7 89 26 Total exposure time was exceeded e Check cable L13 CAN bus replace if S gt 141 necessary S gt 362 Error code Description Actions required see E7 89 27 At least 10 image segments are e Check cable L13 CAN bus replace if S gt 141 defective necessary S gt 362 If the error occurs repeatedly S gt 134 e Check board DX839 replace if necessary S gt 343 e Replace the X ray detector S gt 319 61 25 665 D3437 122 D3437 076 01 15 02 06 2012 Error code Description Actions required see E7 89 28 FPGA field programmable gate e Restart the unit array on board DX839 is defective or 1 Switch off the unit does not respond 2 Wait 1 minute 3 Switch unit on 4 Repeat procedure and check function If the error occurs repeatedly S gt 61 e Run software update S gt 343 e Replace board DX89 Error code Description Action
69. ee a a KA EEA TERTERA 249 8 1 0 1 S0122 Test Ste pe Tao e a a a ET EaR 249 S017 Configuration service cccccccssccseececceaeeeeeeeeeececceeeceeeeeeeeeeeeeeeeeeeees 252 8 1 S0177 TeSt Step 2 nerro aa a ar 252 8 11 2 S01T TeSt Step 3 irei eei aiee EAEE EEE EE EE E eaa 254 8 11 3 S017 Test step 4 nnnennnnnnnennnennnnnnrnnnennnnnnnrnnnnnnnnnennnennnn reenn nennen 256 8 114 S0177 Test Step Drane a e e a EE 258 8 11 5 S01T Test step Orire nieee eiee EEEE E E aE a 259 8 11 6 S017 Test step T enrera a R ea a aa EEE aE 261 8 117 S017 TestStep Qeon e ab cont a N eare 263 8 11 8 S017 Test step 13 oo ecccscccssccneccaeeeaeeeeeeeeccaecceeeeeeeeeceeeeeeeeeeeeeess 265 8 11 9 S017 Teststep Terreno ereraa na iee ie aaen eaeh 266 8 11105017 Test Step 15 irntraren ea conte a a E r 268 8 11 11S017 Test step 25 oo ceccccscesscceeccaeeeaeeeeeeeeceaeccececeeceeeeeeeeeeeeeeeeess 270 018 Service for height adjustment cccceeeecee cee eeeeeeeeeneeeeeeeeeeeeeseens 272 8 12 1 S018 Test Step 2 oo i a a e a aaa E aE e eE 272 8 12 2 S018 Test step 3 oo ceceesseeceeesneeeaeeaaanaeeeaeecaeeeeeeaeceaeceeeeeeeeeeeeess 274 8 12 3 5018 Teststep 4e ae e aaa rer aee ea EE aE na EN 275 S037 Network service cccceesceesseccceccceesaeeaaeeeceaeccaeceaeceeeeeeeeeeceeeeeeeeseeeeees 277 8 13 1 S037 Test step 1 nnnnnnnnnnnnnnnnnnnrnnrnrnnnnnnnennnnnnnnnnnennnennne ranerne rarene 277 8 13 2 S037 Test St p 2 ine ai e A
70. he E E e E ra a Ea 279 8 13 3 S037 Test Step 3ean eaae aee aeia a aei E aaa 281 8 13 4 S037 Test step 4 oo cesceccceesceecneeaaeeaaeeeceeaeecaeececcaeeeaeeeeeeeeeeeeeeess 283 61 25 665 D3437 D3437 076 01 15 02 06 2012 9 Repair 9 3 9 4 9 5 9 6 9 7 9 8 61 25 665 D3437 D3437 076 01 15 02 06 2012 EEEE EEE ET A E E E E EE acces 288 Safety CHECKS i Sone aeaa aa aos castes a da e aeaaeae 289 Height adjustment motor M1_4 spindle 0 eeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeaaees 290 9 2 1 Preparing for motor replacement ccccceccceceeeeeeeeeeeceeeeeeeteeereees 290 9 2 1 1 Moving the slide manually ssssseeeeeeeeeeeerieeeeeereerrrreeserrrrinrresssreeena 290 9 2 2 Removing board DX32 eeeeeeessesereeseeertirittteetrtnrttteeesrinnnnssssrreena 294 9 2 3 Replacing the height adjustment motor spindle cceeeees 296 9 2 4 Laying of cables when replacing the height adjustment motor 298 9 2 5 What has to be done after replacing the height adjustment motor 300 M1_4 or the spindle FRING motor CITB ss aena an alaonret aA aiae RAPAE AA ERa 301 9 3 1 Replacing the ring Motor cccccecesecsecccceeeeeeeeeseceeeeceeeeseeseceeeeeeeenes 301 9 3 2 Replacing the pinion at the ring MOtOTr seeeeeeeeseeeeeeeeeeeerreerreesesreena 303 9 3 3 Laying of cables when replacing the ring MOtOM ee eeeeeeeeeeees 305 9 3 4 What has to be done after replacing the
71. in relation to This error may also occur in S gt 61 i main software releases perform software connection with other causal error f update if necessary S gt 128 messages Please also observe the causal error message It appears Check the CAN bus S gt 343 A P thefirst ig Check the signal path from board DX1 to ge board DX42 replace module DX42 if necessary e Please contact the Sirona Customer Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary Error code Description Actions required see E6 42 04 Data transfer error or dialog error to Check the CAN bus S gt 128 module master side e Please contact the Sirona Customer S gt 61 Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary 61 25 665 D3437 D3437 076 01 15 02 06 2012 111 Error code Description Actions required see E6 42 05 Data transfer error or dialog error to Repeat the software update S gt 61 Pootbade of mmodule e Check the CAN bus S gt 128 Only occurs in connection with a e Replace board DX42 S gt 343 software update Error code Description Actions required see E6 42 06 Module failed in TTP detected on Check the CAN bus S gt 128 mastersiae Check the signal path from board DX1 to S
72. measured for any particular configuration Has a bus resistance of 60Q been measured for any particular configuration Yes Replace board DX1 Replace defective cable in the respective signal path 61 25 665 D3437 130 D3437 076 01 15 02 06 2012 Check optical CAN bus connection L6 to board DX6 Fault at optical CAN bus connection Can the fault be traced back to this connection Yes Replace optical cable L6 Unstable CAN bus or sporadic interferences Check connection quality of CAN bus with service routine 8012 1 Are all modules detected Faulty connection quality Check individual signal paths and connectors Are signal paths and connectors OK Replace corresponding connectors cables or modules 61 25 665 D3437 D3437 076 01 15 02 06 2012 131 Checking the CAN bus with the diagnostic function of board DX1 Board DX1 features a diagnostic function for diagnosing malfunctions of the CAN bus via LEDs V700 and V701 see wiring diagrams The following table indicates the operating status of the CAN bus and the recommended error correction measures V700 V701 CAN bus operation Error correction Slow flashing Slow flashing CAN bus OK Not required Fast flashing Off CAN error no communication with board DX7 i e no display of error messages e Check cabling Check CAN jumpers Jumper positions in the CAN bus gt
73. move the clamp A directly underneath the slide The slide is now locked in this position for further repair work Moving the slide without the height adjustment service kit REF 62 57 518 Move the slide 1 Loosen the two screws C and remove the cover D 2 Loosen the 1st of the two screws E on the spindle holder F 3 Attach the socket wrench SW19 to the spindle Hold it firmly in place while you unscrew the 2nd of the two screws E CAUTION If the socket wrench has to be reset secure the spindle holder against turning e g by using a screw 4 Rotate the spindle holder using a socket wrench SW19 to move the slide to the required height Clockwise rotation of spindle slide moves up Counterclockwise rotation of spindle slide moves down 5 After reaching the desired target position secure the position again using the two screws E CAUTION Before replacing the height adjustment motor the slide must be secured in this position 61 25 665 D3437 292 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 2 Height adjustment motor M1_4 spindle Locking the slide 1 Make a mark at the position of the upper limit stop 2 Loosen nut G on the upper profile clamp H and remove the upper limit stop J from the stand 3 Install limit stop J above the lower limit stop so that there is a distance of 31 cm between the
74. of the ring cannot be fully adjusted using the screws C then proceed with adjusting the swivel arm Otherwise the mechanical adjustment is now complete and you may begin calibration 6 Re attach the arm cover 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 7 Adjusting calibrating the unit Service Manual GALILEOS 7 6 Mechanical adjustments Adjusting the swivel arm Perform this adjustment only if the measured values are out of tolerance gt 166 1 Remove the swivel arm cover Move the swivel arm to the entry position loosen the internal grid slightly bend the housing upwards and remove it by pulling towards the pivot joint of the swivel arm 2 Loosen screw A slightly 3 NOTICE Do not forget to tighten screw A again Otherwise the clearance and play of the swivel arm is not ensured Adjust the swivel arm with the eccentric screw B Hold the eccentric screw securely in place and tighten screw A again A IMPORTANT The swivel arm is shown here without the control panel for purposes of clarity 61 25 665 D3437 D3437 076 01 15 02 06 2012 195 7 Adjusting calibrating the unit Sirona Dental Systems GmbH 7 6 Mechanical adjustments Service Manual GALILEOS 4 Re attach the swivel arm cover To do this position the nose C in the groove of the swivel arm and press the
75. on Factory setting 1 Call service routine S007 5 gt 213 2 Use the arrow keys A to select the required setting see table Once the required setting has been selected the T key B Easypad or the LED above the T key B Multipad lights up 3 Touch the T key B to enable the selected setting All CAN bus events occurring from now on during operation of the unit will be logged and can be displayed with a web browser e g Internet Explorer This log will help you when consulting the Sirona Customer Service Center CSC for error diagnosis 4 Exit the service routine gt 220 61 25 665 D3437 D3437 076 01 15 02 06 2012 235 8 Service Sirona Dental Systems GmbH 8 6 S007 Error logging memory Service Manual GALILEOS Displaying the log with a web browser https XXX XXX XXX XXX info CAN htm 1 Enter the following web address on a PC with internet access integrated in a system network A IP address of the unit TENSE UrnaureU Receive Interrupt disabled Clock divider CDR OxcO CAN Mode PeliCAN Mode RX ErrorCounter 0 TX ErrorCounter 0 ErrorCodeCapture 0x00 Bus Timing 0 BTRO 0x03 Bus Timing 1 BTR1 0x18 ErrCnt ErrCnt ErrCntt ErrCnt2 ErrCnt3 0 0 0 0 0 Status Register Device Error comp Overrun Error Receiver warning Receiver Error Transmitter warning Transmitter Erro Bus o DX41 0 0 0 0 0 0 0 DX42 0 0 0 0 0 0 0 Dx6 0 0 0 0 0 0 0 DX61
76. on the Easypad GALILEOS Comfort System code Description Actions required 100 System is starting e Wait no action required 110 Exposure not possible e Restart the unit 1 Switch off the unit 2 Wait 1 minute 3 Switch unit on 4 Repeat procedure 150 Sensor is prepaired XX seconds e Wait no action required The message will be deleted automatically this may take up to 10 minutes Status messages and displays On the control panel Status displays Description Easypad Multipad GALILEOS Comfort GALILEOS Compact Ready for exposure no special display kV level and mAs System is ready for exposure X RAY are displayed X RAY Active LED lights up on Multipad Exposure in progress Please wait Progress bar Unit waiting for operational readiness Ready for exposure in XXs XXs The cooling time countdown is running 84 On the Facescan LED Blue A LED Green B Meaning Off On Standby Blinking Off Ready for exposure or ready to send data On Off Exposure or end of a firmware update via the Facescan USB stick On Blinking Data transfer On On Boot process firmware update or reset to factory settings 61 25 665 D3437 D3437 076 01 15 02 06 2012 Error messages The error messages are displayed as error codes Ex yy zz on the Easypad touchscreen GALILEOS Comfort or on the Multipad display GAL
77. panel to check the function of the height adjustment motor Reset the travel height 61 25 665 D3437 300 D3437 076 01 15 02 06 2012 9 Repair 9 3 Ring motor M1_3 Sirona Dental Systems GmbH Service Manual GALILEOS Ring motor M1_3 Replacing the ring motor Removing the covers gt Remove the arm cover Removing the defective motor 1 Detach the motor cable from the cable harness and pull it off of connector X813 on board DX1 2 Loosen the four screws A on the ring motor and remove the motor including the screws and the serrated washers B 61 25 665 D3437 D3437 076 01 15 02 06 2012 301 9 Repair Sirona Dental Systems GmbH 9 3 Ring motor M1_3 Service Manual GALILEOS Installing the new motor 1 Insert the new motor including coupling and absorber in the ring Tip While inserting the motor turn it back and forth slightly until the pinion engages in the ring gear 2 Use the screws A and serrated washers B to screw the new motor onto the motor support ring 3 Run the ring motor cable along its original path and plug it back into connector X813 on board DX1 IMPORTANT Don t forget to reattach all cable ties and clamps Attaching the covers gt Reattach the covers 61 25 665 D3437 302 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 3 Ring motor M1_3 Replacing the pinion at the ring motor gt Remove the
78. process can take 2 3 minutes Once the transfer is complete the evaluation of the distortion calibration is displayed in the message window If the information in the message window indicates that calibration was not successful check the calibration phantom to make sure that it is not damaged If the phantom checks out OK all balls are present and correctly positioned repeat the procedure starting with point d as often as required until the calibration is OK If you have repeated the procedure three times and still have not attained a positive result please contact the SIRONA Customer Service Center CSC 61 25 665 D3437 D3437 076 01 15 02 06 2012 185 7 Adjusting calibrating the unit Sirona Dental Systems GmbH 7 3 Adjustment and calibration via the Service Functions menu Service Manual GALILEOS Save values 7 Ifthe calibration is OK or possible click the Save values button The calibration is saved The distortion calibration is now complete 8 Remove the distortion phantom again from the X ray detector cover 9 Continue the calibration procedure with the geometry calibration gt 187 61 25 665 D3437 186 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 7 Adjusting calibrating the unit Service Manual GALILEOS 7 3 Adjustment and calibration via the Service Functions menu 3 8 Geometry calibration IMPORTANT For an installed and configured Facescan The face s
79. s 2 Measure the voltage drop with the voltmeter measure the current with the ammeter and calculate the resistance using the formula R U I Ifthe resistance value is greater than indicated in the adjacent table check whether the protective ground wires are fastened according to the specifications Check whether plain washer tooth lock washer and cable lug are installed on the protective ground wire in the correct order and whether the nuts of the protective ground wire connections are firmly tightened If the fastening of the protective ground wires does not meet the specifications fasten the protective ground wires correctly Tip Do not connect the power cable and the second ground wire to the building installation yet Check the device leakage current first gt 382 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 10 Maintenance Service Manual GALILEOS 10 8 Checking the protective ground wires Measuring point A Central ground wire Up to unit serial number GALILEOS Comfort 8 999 GALILEOS Compact 48 999 61 25 665 D3437 D3437 076 01 15 02 06 2012 379 10 Maintenance Sirona Dental Systems GmbH 10 8 Checking the protective ground wires Service Manual GALILEOS From unit serial number GALILEOS Comfort 9 000 GALILEOS Compact 49 000 Measuring points B and C GNYE power connection and 2nd ground
80. selected service routine has several test steps the first selectable test step is displayed in selection field 2 test step 1 in the example 213 8 Service Sirona Dental Systems GmbH 8 3 Basic operating procedures in the service menu Service Manual GALILEOS Selecting a test step wv The service menu must be selected gt 210 J The required service routine must be selected gt 213 gt Select the required test step in selection field 2 with arrow keys C and confirm your selection by pressing Service key B Easypad The selected service routine as well as the test step chosen are displayed in the right hand column S005 4 in the example 61 25 665 D3437 214 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 3 Basic operating procedures in the service menu Multipad The parameters or IDs of the selected service routine are displayed on the Multipad The Multipad does not show which service routine or test step is currently active Service routines with security access A security code is required for accessing service routines involving functions such as radiation release or editing of configuration data or stored values This procedure prevents the inadvertent selection or activation of these service routines To select a service routine or test step with security access proceed as follows 1 Select the service routi
81. service menu Service Manual GALILEOS a a gt Touch the or arrow keys A to select the code 01 B for the Remote control enabled option Once the selected parameter has been changed in this case the p i A code for the activation of the remote control the Memory key C z lights up 01 17 6 Multipad Example You want to run service routine S017 6 to activate the remote control J Once you have selected service routine S017 6 the code 00 is pre selected for the Remote control disabled option gt Press the UP or DOWN arrow key A to select the code 01 B for the Remote control enabled option Once the selected parameter has been changed in this case the code for the activation of the remote control the LED above Memory key C lights up 61 25 665 D3437 218 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS Example lt 01 S17 Example 61 25 665 D3437 8 3 Basic operating procedures in the service menu Saving parameters Once one or a number of parameters have been selected via a service routine the current selection must be saved so that it is applied in the unit Easypad touchscreen You want to run service routine S017 6 to save the selected option Remote control enabled JY The Memory key C lights up 1 Touch the Memory key C The R key D lights up 2 Touch the
82. should no longer appear 18 Call up the Extended Details via SixABCon This generates an XML file with the system parameters which is Filed in the PDATA P2K_Config directory under the network name of the unit e The process is completed 61 25 665 D3437 D3437 076 01 15 02 06 2012 357 9 Repair Sirona Dental Systems GmbH 9 13 Cable Service Manual GALILEOS Cable Replacing energy chain 1 completely Removing the defective energy chain IMPORTANT Remove cable ties For the following steps all necessary cable ties should be removed with wire cutters Danger of fatal electrocution gt Before you remove the energy chain switch off the power supply 1 Disconnect cable L1 from switch 1 2 Disconnect cable L1 from the ground point F 3 Loosen the two screws A 61 25 665 D3437 358 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 13 Cable 4 If not present Mark the position of the screw B on the stand 5 Make a note of the position of the screw B for when you later install the new energy chain 6 Remove the screw B from the energy chain 1 7 Remove the power cable from terminal K1 8 Unscrew the mains filter plate 61 25 665 D3437 D3437 076 01 15 02 06 2012 359 9 Repair Sirona Dental Systems GmbH 9 13 Cable Service Manual GALILEOS 9 Remove the cable covers I and J of the right hand cable duct 10
83. status to be programmed or the software version on the module is newer than the one in the update file this will be indicated by a red triangle with an exclamation mark If the version of the selected update file is lower than the current software version of the unit then there will be no display in the right window The downgrade required in this case is possible only via Main version mode 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 4 General operating procedures Service Manual GALILEOS 4 2 Updating the firmware Updating the Facescan firmware There are two ways of updating the Facescan firmware e Option 1 Update via USB stick gt 67 The program data on the Facescan unit is completely overwritten e Option 2 Update via the network gt 70 The program data is transferred to the Facescan unit Unlike in option 1 the unit configuration data is however not overwritten Option 1 Update using the Facescan USB stick Opening Facescan 1 Unscrew the cover A from the Facescan 2 Pull gray cable L78 4 from slot X2 of the FACESCAN modular board 3 Unscrew the protective plate T from the Facescan unit The FACESCAN modular board is visible 4 Plug cable L78 4 again into slot X2 on the FACESCAN modular board 61 25 665 D3437 D3437 076 01 15 02 06 2012 67 4 General operating procedures Sirona Dental Systems GmbH 4 2 Updating the firmware Service Manual GALILEOS Updating the s
84. stuck in bootloader stage Check board DX71 S gt 134 If the board remains in the bootloader stage S gt 61 e Repeat the software update S gt 343 e Replace board DX71 or Multipad S gt 307 61 25 665 D3437 116 D3437 076 01 15 02 06 2012 Error code Description Actions required see E7 71 12 Unit is not ready for operation e Check the CAN bus S gt 128 This error is a sequential fault e Restart the unit 1 Switch off the unit 2 Wait 1 minute 3 Switch unit on 4 Repeat procedure and observe causal error messages Error code Description Actions required see E6 71 20 Contact to DX11 interrupted during Note error message on remote control S gt 125 operation DX42 and check log memory via S gt 128 extended details e Check the CAN bus FARRU e Check cable L9 replace if necessary Salt 802 Error code Description Actions required see E7 7121 No CAN bus connection DX11 does Start the detail query via SiXABCon S not start If board DX11 responds S gt 362 Occurs in the start screen after e Check the signal path to DX71 repair or S gt 343 pOwerson replace cables connectors if necessary e Replace board DX1 If board DX11 does not respond S gt 343 e Replace board DX11 61 25 665 D3437 D3437 076 01 15 02 06 2012 117 5 Messages 5 5 List of error messages Sirona Dental Systems GmbH Service Manual
85. the Multipad display e On the display of the remote control There are 3 groups of message texts Help messages Hx xx e Help messages are caused by operator errors e The user must take action Error messages Ex yyxx e Error messages indicate unit faults e The user must take action to eliminate the fault s System messages Sxxx e System messages inform the user about the current operating status of the unit e The user does not have to take any action If error messages are displayed on the control panel that are not listed in this section such as message 1311 these messages come from the Windows system In such cases you must check whether the firmware you are using is compatible with the SIDEXIS XG version and run a software update gt 61 if necessary Help messages The help messages are displayed as help codes Hx xx on the Easypad touchscreen GALILEOS Comfort or on the Multipad display GALILEOS Compact as well as on the display of the remote control if present The codes tell you how to operate the system if radiation release is not possible due to a previous operator error The following list provides you with an overview of all help codes their meaning and the action required to eliminate the corresponding problems 61 25 665 D3437 D3437 076 01 15 02 06 2012 IMPORTANT The measures listed only clear help messages that result from operator errors If it is not possible to clear a message by t
86. the adjustment screws to adjust the diaphragm in the X Y or Z direction CW rotation of screws Moves the diaphragm in the X Y or Z direction CCW rotation of the screws Moves the diaphragm in the direction opposite to the X Yor Z direction 4 Retighten the five screws A firmly 5 Repeat the diaphragm exposure gt 167 61 25 665 D3437 200 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 7 Adjusting calibrating the unit Service Manual GALILEOS 7 6 Mechanical adjustments Adjusting the lower lead diaphragm If the exposure taken in the upper jaw diaphragm setting is not OK the lower lead diaphragm must be adjusted 1 Setthe rotary knob on the tube assembly to the maxillary exposure position Loosen both screws B of the lead diaphragm Manually adjust lead diaphragm C Tighten both screws B firmly 5 Repeat the diaphragm exposure gt 167 PON Adjusting the upper lead collimator If the exposure taken in the lower jaw diaphragm setting is not OK the upper lead diaphragm must be adjusted 1 Set the rotary knob on the tube assembly to the lower jaw position Loosen both screws B of the lead diaphragm Manually adjust lead diaphragm C Tighten both screws B firmly ak WON Repeat the diaphragm exposure gt 167 61 25 665 D3437 D3437 076 01 15 02 06 2012 201 Service Overview of service routines List of a
87. the unit off again 61 25 665 D3437 340 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 11 Light barriers Light barriers Replacing the light barriers The following light barriers can be replaced e Light barrier at ring motor starting position of rotation V1_3 e Light barrier at HA motor height adjustment V1_4 61 25 665 D3437 D3437 076 01 15 02 06 2012 341 9 Repair 9 12 Boards Prior to replacing boards Replacing the boards DX6 X ray tube assembly and DX11 or DX89 and DX11 Prior to replacing board DX11 Connector designations on the boards 342 Sirona Dental Systems GmbH Service Manual GALILEOS Boards Important notes about replacing boards Touching the boards can damage them Please observe the usual precautionary measures for handling printed circuit boards ESD Touch a ground point to discharge static electricity before touching any boards You must observe the notes in the chapter This chapter describes all measures required after the replacement of modules or boards provided they were known at the time of publication You will find more up to date information and supplements concerning this subject on the latest GALILEOS XG CD and on the Sirona dealer page on the Internet For this reason you should always check for the latest information on the replacement of modules and performing updates before you star
88. the unit serial number found on the nameplate of the unit gt 240 IMPORTANT Any serial number which is unknown to the unit will not be accepted by the unit The serial number entered must be identical to the number on the nameplate of the unit If an inadmissible serial number is entered the input will not be accepted and the serial number can be entered again 10 Switch the unit off 11 Wait for approx 1 minute Then switch the unit back on In systems that already run with a system software version V03 03 01 or higher please check whether there is a XML file in the PDATAV P2K_Config with the network name of the system This file contains important information about the previous unit configuration The error message E1 11 20 invalid unit calibration is displayed 12 Acknowledge the error message with the R key 13 Call the service menu gt 210 o o oN 61 25 665 D3437 352 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 12 Boards 14 Call the service routine S017 and perform the unit configuration test step 2 15 gt 252 The board DX41 must be configured using the service routine 017 9 In units with a serial number of 1080 and above the switching plate configuration of the swivel arm must be set or checked using service routine S017 7 Inform the customer of the configuration options of the software status for example the welcome screen or acoustic e
89. tisscveescesneRechaadaueitiachevenn date hechedeenonds 344 9 12 2 3Replacing board D ABZ verde vesedescte nts cebteveddyiotrieravtentatas Vere iehehueamteds 345 9 12 2 4Replace board DX89 0 cccccccssccscccecccaccceccecccecceeeeeeeceeceeeeeeeeeeeeeess 346 9 12 3 Measures following replacement of boards eeeeeeeeeeeeeeeeeee 347 9 12 3 1After changing the DX11 board ooo eee e eee eeeeeeeeeeeeeeeeeeenaaee 350 Cablen i ned Ge A a ae Se ee eS 358 9 13 1 Replacing energy chain 1 completely c cceeeeeseeeeteeeeeeeeeeeeeee 358 9 13 2 Replacing cables reanreeee err areien err er er eee Bean ere ne een APSE 362 9 13 2 1Replacing fiber optic cable L5 L6 or L15 cc ccccceceeeceeeeeeeees 363 9 13 2 2Cable replacement L3 L5 L6 and L15 Laying the cable 364 corrugated tube at the rotation unit 9 13 2 3Replacing cable L7 L117 or L108 in cable track 2 ee 366 9 13 2 4Replacing cable L1 or grounding strap in cable track 1 00 368 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS 10 11 12 61 25 665 D3437 D3437 076 01 15 MaAIMOMANC Gs ay ston 8 as ats aaa coed aaaea raai aeaaea aa deeds ea iaaa iresi 369 10 1 Galibrating the Unitas ook cles i naa riete eee haaraa aa ea EA E aaae 369 10 2 Checking the height adjustment sssneeeeeseneesnneesnttrnserrrnnennnrennereserrrnet 370 10 3 Checking the fan and temperature Sensor cc
90. upper edge of the upper screw on board DX41 and the lower edge of the limit stop 61 25 665 D3437 D3437 076 01 15 02 06 2012 293 9 Repair Sirona Dental Systems GmbH 9 2 Height adjustment motor M1_4 spindle Service Manual GALILEOS Removing board DX32 1 Unscrew the bracket C which is located in front of board DX32 2 Move the stand to a height of 1260 Easypad display 3 Disconnect GALILEOS from the power supply 5 Loosen all the left hand screws B of the protective plates E 61 25 665 D3437 294 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS 9 Repair 9 2 Height adjustment motor M1_4 spindle 10 11 12 13 Unscrew the remaining screws from the protective plates E Remove the covering plates E top and bottom from the connection box F of board DX32 Cable L3 can stay on the top covering plate E remove any possible shield terminal Tip The covering plate can be folded away simply at the sides and put away in the stand sideways to protect against scratches suitable padding should be put in between Remove connector X2 from board DX32 and remove the protective conductor
91. with an asterisk on the info screen e g 04 03 01 GALILEOS Remote control Board Software Board Software DX6 02 88 00 DX42 02 45 06 DX7 02 57 00 DX7 LO 02 18 00 DX7 L1 02 18 00 DX7 L2 02 18 00 DX7 L3 02 18 00 DX7 L4 DX7 L5 DX71 DX11 02 61 01 DX41 02 30 00 DX89 01 10 06 DX89 FPGA 01 13 01 The main software V 03 03 02 is run compatible as of GALILEOS Software V1 2 GALILEOS Remote control Board Software Board Software DX6 02 88 00 DX42 02 46 04 DX7 02 58 03 DX7 L0 02 22 00 DX7 L1 02 22 00 DX7 L2 02 22 00 DX7 L3 02 22 00 DX7 L4 DX7 L5 DX71 DX11 02 63 05 DX41 02 30 00 DX89 01 12 07 DX89 FPGA 01 13 01 The main software V 03 04 00 is run compatible as of GALILEOS Software V1 4 61 25 665 D3437 D3437 076 01 15 02 06 2012 Main software V03 04 02 Main software V03 05 00 61 25 665 D3437 D3437 076 01 15 02 06 2012 GALILEOS Remote control Board Software Board Software DX6 02 88 00 DX42 02 46 04 DX7 02 58 03 DX7 LO 02 22 00 DX7 L1 02 22 00 DX7 L2 02 22 00 DX7 L3 02 22 00 DX7 L4 DX7 L5 DX71 DX11 02 64 00 DX41 02 30 00 DX89 01 12 07 DX89 FPGA 01 13 01 The main software V 03 04 02 is run compatible as of GALILEOS Software V 1 4 3
92. without rotary movement selectable kV mA level and maximum radiation time S002 Radiation without rotary movement selectable kV mA level and maximum radiation time SR SHZ Function 002 X ray beam test 002 5 Yes Long term exposure with fixed radiation intervals from any position SR service routine SHZ security access S002 Test step 5 Long term exposure with fixed radiation intervals from any position Unit is radiating X rays Excess exposure to X rays is detrimental to health gt Use the prescribed accessories for radiation protection gt Do not stay in the X ray room during exposure Move as far away from the unit as the coiled cable for the release button allows you to Selection field Parameters Range of values 1 kV mA level 60 kV 8 mA 85 kV 7 mA 2 Radiation time 0 1 s 5 0 s Factory setting 61 25 665 D3437 D3437 076 01 15 02 06 2012 221 Sirona Dental Systems GmbH Service Manual GALILEOS 8 Service 8 4 S002 Radiation without rotary movement selectable kV mA level and maximum radiation time 1 Call service routine S002 5 gt 213 A 2 Use the arrow keys A to select the required kV mA level and the required radiation time see table A X RAY 85kV 7MA 4 0s 3 Initiate the radiation The maximum set radiation time has elapsed IMPORTANT If you let go of the release button before the maximum rad
93. 0 Subnet mask 255 255 255 0 Default gateway 192 168 15 1 Net API version 1 Net API revision 15 Demomode On 61 25 665 D3437 80 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 4 General operating procedures Service Manual GALILEOS 4 5 Using demo mode operation without radiation release Switching off demo mode 1 Switch off the unit Potentially lethal shock hazard It is essential to switch off the unit and to wait at least another 4 minutes before taking off the covers of the X ray tube assembly Remove the cover of the tube assembly Loosen the screws A and remove the cover plate B Set the dip switch S2 DX6 to position 1 Connect cable L5 XRAY to connector J6 DX6 Radiation release is now once again possible ak whr 6 Switch on the unit and check the mode on the info screen Demo mode ON means that Demo mode is switched on radiation release is not possible Demo mode OFF means that Demo mode is switched off radiography X ray radiation are possible 7 Switch the unit off again and reattach the cover plate B and the tube assembly covers by following the dismantling procedure in reverse order 61 25 665 D3437 D3437 076 01 15 02 06 2012 81 5 Messages 5 1 Help messages 82 Sirona Dental Systems GmbH Service Manual GALILEOS Messages The different message texts are displayed e GALILEOS Comfort On the Easypad touchscreen e GALILEOS Compact on
94. 0 0 0 0 ee eeeeeeeeeeeeeeeeeeeeeeeeees 142 6 8 1 Error analysis of X RAY control signal path up to unit serial number 142 3199 with board DX41 6 8 2 Error analysis of X RAY control signal path from unit serial number 147 3201 without board DX41 6 9 Fault diagnosis of the X ray detector and on board DX89 ee 151 6 9 1 LEDs on board DX89 ss saicncteteaseiaivesieed aca yankinieianacaleiiuetewtetiaaeehianlage 152 6 9 2 LED statuses and their significance in case of an error 0 153 6 9 3 LEDs of operating voltages sive weaned reset on sees odes ewleae duce sadeedehadeece 154 Adjusting calibrating the Unit scic chectisecs tees cenecest aeceeneches dedeealcneacedaseeenathoncedceenateee 155 7 1 General information about unit adjustment and calibration 0 ee 157 7 1 1 Displays and help messages during adjustment calibration 158 7 1 2 Service Functions MOM i ceceiuccsiecieceeiderlnseeltins da cleeneeenaeteacdcs 159 7 1 2 1 Calling the Adjustment Calibration MeNU eeeeeeeeteeee teeta 161 7 1 3 Enabling exposure readiness cccccceeeececceeeeeeeeeedeeeeenereeeeeneens 162 7 1 4 Taking an xpOSUIe ciccisiecssdeacaliscncueccenidesiteta ele Gol ea andes 162 Fla SAVE Al E E heteetastigas Oot ene adeiata Mee Fada tateatetpatetatectageee 162 7 1 6 Test phantoms for adjustment and calibration ccceeeeeeeeeeeee 163 7 1 6 1 Distortion phantom wesssssesiondaceeccels ctead
95. 1 2 3 4 Remove the retaining ring L Remove the copper filter M and the aluminum filter D Unscrew the two sleeves C with washers Loosen the two lower screws A approx 2 to 3 turns With the Type 2 diaphragm unit 1 2 Remove the retaining ring L Remove the copper filter M and the aluminum filter D With the Type 3 diaphragm unit gt No preparations are required Installing the new tube assembly 1 314 Hang the new tube assembly on the two front screws of the rotating element and tighten them securely Insert the two rear screws and tighten them firmly Plug cables L3 L5 L6 and L15 as well as the ground cable back onto board DX6 and reattach the cables to the rubber grommets Reattach the cover plate 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 6 X ray tube unit Installing the diaphragm unit gt NOTICE Do not forget to re install the filters or in the case of Type 1 diaphragm unit the primary diaphragm The diaphragm is installed in the reverse order to removal With the Type 3 diaphragm IMPORTANT Note the order for installing the filters The image quality will be impaired if the filters are installed in the wrong sequence Insert the aluminum filter D first followed by the copper filter M Attaching the covers gt Reattach the covers With the Type 3 diaphragm gt
96. 1 25 665 D3437 26 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS 3 Unit description 3 2 Hardware L13 DX1 X404 h it DX1 X803 AY m n L27 T L28 fii L29 a L13 L27 L28 DX7 DX71 61 25 665 D3437 D3437 076 01 15 02 06 2012 Cable L28 cannot be replaced on X ray detectors with a serial number 2 5000 A Power switch 27 3 Unit description 3 2 Hardware Sirona Dental Systems GmbH Service Manual GALILEOS Cabling up to serial number 3199 DX1 3901 U il DX1 X303 DX42 X100 TP 5922633 Lo BA TPO 5922633 DX41 L25 L26 DX41 X100 DX41 X103 Y a Media Konverter L8 L17 E Media converter 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS Cabling from serial number 3201 3 Unit description 3 2 Hardware Bay DX1 J901
97. 11 S017 Configuration service Service Manual GALILEOS S017 Test step 25 Select the diaphragm type Selection field Code Function 1 00 Type 1 diaphragm GALILEOS Compact 01 Type 1 Type 2 diaphragm GALILEOS Comfort 02 Type 3 diaphragm GALILEOS Compact 03 Type 3 diaphragm GALILEOS Comfort Factory setting 1 Call service routine 8017 25 gt 213 Once the service routine has been selected the code for the current setting is displayed in selection field 1 61 25 665 D3437 270 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 11 S017 Configuration service 2 Use arrow keys A to select the required code in selection field 1 see table Once the required setting has been selected Memory key B Easypad or the LED above Memory key B Multipad lights up 3 Save the setting gt 219 4 Exit the service routine gt 220 61 25 665 D3437 D3437 076 01 15 02 06 2012 271 8 Service Sirona Dental Systems GmbH 8 12 S018 Service for height adjustment Service Manual GALILEOS S018 Service for height adjustment SR SA Function S018 Service for height adjustment 018 2 No Set the maximum travel height 018 3 No Undo the maximum travel height setting 018 4 No Check the height adjustment sensor system SR service routine SHZ security access S0
98. 17 and cable L108 from the cable clamps at the rear of the unit and pull the cables through the slit in the slide toward the front into the stand Unscrew the angle brackets on both sides of the cable track 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS 14 15 9 Repair 9 13 Cable Remove the motor side end piece from the cable track With defective cable L7 L117 Unscrew cable L7 L117 from the interface board and remove the shield If cable L7 L117 should be intact and used again this step is not necessary Unless it is not possible to lay down the cable track flat near the stand see next step Remove the cable ties from the cable track and lay the cable track down on a flat surface stretched out CAUTION You must observe the position of connector X303 of cable L108 see image Carefully pull both cables together out of the cable track and the fabric tube With defective cable L108 Wrap the connector X303 of the new cable L108 with adhesive tape to protect the detent against breaking off Lay the cable track down on a flat surface stretched out Fasten the two new cables together with adhesive tape above the flag labels 61 25 665 D3437 D3437 076 01 15 02 06 2012 367 9 Repair Sirona Dental Systems GmbH 9 13 Cable Service Manual GALILEOS 21 CAUTION Push the green cable The white cable is carried along In
99. 18 Test step 2 Set the maximum travel height 1 In user mode move the unit to the required maximum travel height by pressing UP DOWN keys A 2 Call service routine S018 2 gt 213 Once the service routine has been selected the current height position is displayed in selection field 1 Memory key B Easypad or the LED above Memory key B Multipad lights up 61 25 665 D3437 272 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 12 S018 Service for height adjustment 3 To save the maximum travel height press Memory key B R key C lights up followed by R key C 4 Set the mechanical limit stop at the unit Loosen nut D and move mechanical limit stop E for the limit switch until it engages Tighten nut D again The next time the UP key is pressed the unit stops 10 mm below the limit switch 5 Exit the service routine gt 220 it 61 25 665 D3437 D3437 076 01 15 02 06 2012 273 8 Service Sirona Dental Systems GmbH 8 12 S018 Service for height adjustment Service Manual GALILEOS S018 Test step 3 Undo the maximum travel height setting 1 Call service routine S018 3 gt 213 Once the service routine has been selected the current height position is displayed in selection field 1 Memory key B Ea
100. 2 Sirona Dental Systems GmbH Service Manual GALILEOS 5 Messages 5 5 List of error messages Error code Description Actions required see E6 14 10 Clock signals for sensor image e Check cable L13 for crushed spots and S gt 362 transfer not received on board DX1 kinks and check connectors repair or DX11 10 replace if necessary Faulty detection of sensor image e Check board DX89 replace if necessary S gt 134 E6 14 12 transfer data signals on board DX1 S gt 343 DX11 recurring 12 61 25 665 D3437 D3437 076 01 15 02 06 2012 105 Location 15 Configuration update Error code Description Actions required see E7 15 01 Wrong memory modules If a DRAM memory module is plugged into S gt 350 board DX11 e Replace memory module or DX11 If no DRAM memory module is plugged into S gt 350 board DX11 e Replace board DX11 Error code Description Actions required see E7 15 03 Wrong software constellation of e Check the software versions of the unit on S gt 238 modules the info screen or by running service S gt 61 routing S008 2 and run or repeat i software update or downgrade if necessary Error code Description Actions required see E6 15 04 Product activation keys invalid or e Enter release key see Ol not available Occurs after replacement of tube assembly DX6 or board DX11 an
101. 2 96 EC as well as the corresponding national laws This directive requires environmentally sound recycling disposal of the product The product must not be disposed of as domestic refuse Please observe the disposal regulations applicable in your country Disposal procedure We advise that this product is subject to the stipulations in EC guideline 2002 96 governing waste electrical and electronic equipment and must be disposed of in line with these special requirements within the European Union EU Prior to disassembly disposal of the product it must be fully prepared cleaned disinfected sterilized When disposing of equipment permanently please proceed as follows 61 25 665 D3437 384 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS Entsorgung elektrischer und elektronischer Ger te aus Optik uad Medizin 61 25 665 D3437 D3437 076 01 15 02 06 2012 11 Dismantling and disposal 11 2 Disposal In Germany To initiate return of the electrical device please send a disposal order to enretec GmbH 1 You can find a form for placing a disposal order on the company s homepage www enretec de under the menu item Entsorgung elektrischer und elektronischer Gerate Disposal of electric and electronic devices The form can either be downloaded or completed online Fill out the form with the corresponding details and send it either as an online order or fax it to enretec Gmb
102. 25 665 D3437 262 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 11 S017 Configuration service S017 Test step 9 Activate deactivate operation with board DX41 This service routine is used to configure operation of the system with or without module DX41 This configuration is necessary for software updates and module replacement of board DX11 if systems with or without board DX41 should be supported Selection field Code Function 1 00 Board DX41 inactive 01 Board DX41 active Factory setting 1 Call service routine S017 9 gt 213 Once the service routine has been selected the code for the current setting is displayed in selection field 1 61 25 665 D3437 D3437 076 01 15 02 06 2012 263 8 Service Sirona Dental Systems GmbH 8 11 S017 Configuration service Service Manual GALILEOS 2 Use arrow keys A to select the required code in selection field 1 see table Once the required setting has been selected Memory key B Easypad or the LED above Memory key B Multipad lights up 3 Save the setting gt 219 4 Exit the service routine gt 220 61 25 665 D3437 264 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS S17 13 8 11 S017 Configuration service S017 Test step 13 Enable disable the welcome screen
103. 272 IP address enter manually 037 4 S gt 283 IP address set to defaults 037 2 S gt 279 Long term exposure with fixed radiation intervals from any 002 5 S gt 221 position Maximum travel height set 018 2 S gt 272 Maximum travel height undo setting 018 3 S gt 274 Module software version show overview 008 2 S gt 238 Modules presence display 012 1 S gt 249 Network service S037 S gt 277 Radiation without rotary movement selectable kV mA level and S002 S gt 221 maximum radiation time Remote control enable disable display 017 6 S gt 259 Save restore DX89 data 009 7 S gt 245 Select a language 017 4 S gt 256 for GALILEOS Comfort Select a language set 017 5 S gt 258 for GALILEOS Comfort 204 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS 8 Service 8 1 Overview of service routines Function Service routine see Select the diaphragm type 017 25 S gt 270 Subnet mask manual input 037 4 S gt 283 Subnet mask set to defaults 037 2 S gt 279 Temperature sensor test single tank 005 5 S gt 226 Tube assembly service general S005 S gt 223 Unit serial number confirm 008 3 S gt 240 Unit variant configure 017 2 S gt 252 Update service 008 S gt 238 Welcome screen enable disable 017 13 S gt 265 for GALILEOS Com
104. 3 03 01 or higher e System software version V03 03 01 or higher e Up to unit serial number 3199 S Set jumper X109 and X110 to Configuration with DX41 inside jumper e Inserting a new DX42 e Switch the unit on S gt 61 e GALILEOS Compact e Perform a software update of the device to a version V03 06 01 or e System software version higher V03 06 01 or higher gner DX89 Inserting a new DX89 X ray e Inserting a new DX89 e Switch the unit on S gt 61 detector e Perform a software update of the unit to a version V03 03 01 or higher e GALILEOS Compact e System software version V03 06 01 or higher V03 03 01 or higher e Restore the configuration data of S gt 245 board DX89 by using service routine 009 7 e Inserting a new DX89 e Switch the unit on S gt 61 e Perform a software update of the device to a version V03 06 01 or higher e Restore the configuration data of S gt 245 board DX89 by using service routine 009 7 61 25 665 D3437 D3437 076 01 15 02 06 2012 Board DX41 is omitted in unit serial number 3201 and higher Board DX41 is available as a spare part for units up to unit serial number 3199 349 9 Repair 9 12 Boards Case A 350 Sirona Dental Systems GmbH Service Manual GALILEOS After changing the DX11 board New DX11 GALILEOS Comfort For DX11 System software version V03 03 01 or higher For DX1V2 System software version V04 04 00 or hig
105. 437 D3437 076 01 15 02 06 2012 259 8 Service Sirona Dental Systems GmbH 8 11 S017 Configuration service Service Manual GALILEOS 2 Use arrow keys A to select the code for the required setting in selection field 1 see table Once the required setting has been selected Memory key B Easypad or the LED above Memory key B Multipad lights up 3 Save the setting gt 219 4 Exit the service routine gt 220 61 25 665 D3437 260 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 11 S017 Configuration service S017 Test step 7 Configuring the switching plate for the swivel arm Selection field Code Function 1 01 up to unit serial number 1079 02 unit serial number 1080 or higher Factory setting 1 Call service routine S017 7 gt 213 Once the service routine has been selected the code for the current setting is displayed in selection field 1 61 25 665 D3437 D3437 076 01 15 02 06 2012 261 8 Service Sirona Dental Systems GmbH 8 11 S017 Configuration service Service Manual GALILEOS 2 Use arrow keys A to select the required country group code in selection field 1 see table Once the required setting has been selected Memory key B Easypad or the LED above Memory key B Multipad lights up 3 Save the setting gt 219 4 Exit the service routine gt 220 61
106. 54 00 SIDEXIS XG GALILEOS Software RCU GALILEOS Server software Implant V2 5 1 or higher V1 8 V1 9 V2 0 V2 1 V1 8 V1 9 Main software V04 09 01 GALILEOS Remote control Board Software Board Software DX6 03 06 01 DX42 02 56 02 DX6NG 04 07 01 DX7 02 80 05 DX7 LO 02 29 00 DX7 L1 02 29 00 DX7 L2 02 27 00 DX7 L3 02 27 00 DX7 L4 02 04 00 DX7 L5 02 02 00 DX71 02 54 03 DX11 04 09 01 DX11 FPGA 01 03 00 DX41 02 30 00 DX89 01 58 00 DX89 FPGA 01 55 00 SIDEXIS XG GALILEOS Software RCU GALILEOS Server software Implant V2 5 3 or higher V1 10 V2 2 V1 9SP1 61 25 665 D3437 D3437 076 01 15 02 06 2012 Facescan firmware Facescan GALILEOS GALILEOS Software SIDEXIS XG main unit software FS 000001 V04 09 00 or higher V1 10 or higher V2 5 3 or higher or higher 56 GALILEOS Software GALILEOS Software CD index Remarks V1 2 004 Requires unit main software V 03 03 01 and SIDEXIS 2 0 V1 4 005 Requires unit main software V 03 04 00 and SIDEXIS 2 2 V1 4 3 SW 007 Requires unit main software V 03 04 01 and SIDEXIS 2 2 V1 5 009 Requires unit main software V 03 05 00 and SIDEXIS 2 3 V1 6 V1 6 1 011 012 For GALILEOS Comfort Requires unit main software V 03 05 00 and SIDEXIS 2 3 or above For GALILEOS Compact Requires unit main software V 03 06 01 and SIDEXIS 2 3 or above For GALILEOS Comfort Requires unit main sof
107. 6 2012 Sirona Dental Systems GmbH 3 Unit description Service Manual GALILEOS 3 2 Hardware Minimum requirements for Processor DualCore from 1 6 GHz reconstruction PC included in the scope of supply RAM 2 GB RAM Hard disks gt 200 GB Operating system Windows XP Professional Service Pack 2 or newer External drive 1x DVD ROM dual layer Minimum requirements for See SIDEXIS XG Operator s Manual SIDEXIS visualization PC not included in the scope of supply The system requirements are also listed under www sidexis com Network Network 100 MB Ethernet 1 Gbit Ethernet recommended Communication interface RJ45 for LAN cable 61 25 665 D3437 D3437 076 01 15 02 06 2012 47 3 Unit description 3 2 Hardware 48 3 2 7 1 Sirona Dental Systems GmbH Service Manual GALILEOS Diagrams Cooling curve of tube housing Cooling curve of X ray tube Heating curve of tube housing 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 3 Unit description Service Manual GALILEOS 3 2 Hardware Central X ray beam 10 i Anode angle 61 25 665 D3437 D3437 076 01 15 02 06 2012 49 Main software V03 03 02 Main software V03 04 00 50 Software compatibility GALILEOS firmware Any software combinations other than those listed here are not allowed If a module software version does not match the main software version the main software version is identified
108. 61 25 665 D3437 D3437 076 01 15 02 06 2012 255 8 Service Sirona Dental Systems GmbH 8 11 S017 Configuration service Service Manual GALILEOS S017 Test step 4 Select a language Selection field Code Function 1 00 English 01 English 02 French 03 Italian 04 Dutch 05 Spanish 06 Russian 08 Portuguese 10 Chinese PRC 11 Korean 12 Japanese 13 Chinese Taiwan Factory setting varies by order 1 Call service routine S017 4 gt 213 Once the service routine has been selected the code for the current setting is displayed in selection field 1 lt 01 S17 61 25 665 D3437 256 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 11 S017 Configuration service A 2 Use the arrow keys A to select the code for the required language in selection field 1 see table P Once the language has been selected the Memory key B lights up A NU 3 Save the setting gt 219 SI 00 D S17 IMPORTANT If the selected language is not in the installed language 4 set S017 Test step 5 gt 258 English is set by default 4 Exit the service routine gt 220 61 25 665 D3437 D3437 076 01 15 02 06 2012 257 8 Service Sirona Dental Systems GmbH 8 11 S017 Configuration service Service Manual GALILEOS S017 Test step 5 Select a language set IMPORTANT A softwar
109. 61 25 665 D3437 D3437 076 01 15 02 06 2012 305 9 Repair Sirona Dental Systems GmbH 9 4 Rotary knob on the swivel arm Service Manual GALILEOS Rotary knob on the swivel arm Replacing the rotary knob Removing the rotary knob 1 Slide the plastic ring A toward the rear 2 Turn the rotary knob and find the opening B If no opening B appears you can now simply pull off the rotary knob or gt if an opening appears Loosen setscrew C with an Allen key 2mm 3 Pull the rotary knob off Attaching the rotary knob Install the rotary knob by performing the steps above for dismantling in reverse order 61 25 665 D3437 306 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 5 Control panel Control panel Replacing the Easypad user interface GALILEOS Comfort or Multipad GALILEOS Compact oe GALILEOS Compact 1 Press into slit A of the housing cover with a screwdriver do not pry and remove the defective user interface from the control panel 2 Pull cables L9 and L10 off of connectors X102 L9 and X103 L10 on board DX7 Easypad GALILEOS Comfort or DX71 Multipad GALILEOS Compact of the defective user interface 3 Plug the cables into connectors X102 L9 and X103 L10 of board DX7 Easypad GALILEOS Comfort or DX71 Multipad GALILEOS Compact on the new user interface 4 Clip the new user interface onto t
110. 7 D3437 076 01 15 02 06 2012 Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary Error code Description Actions required see E3 07 30 Up down keys pressed on power e Restart the unit S gt 307 lke 1 Switch off the unit E3 07 33 Light localizer key pressed during 2 Wait 1 minute power on 7 3 Switch unit on E3 07 34 T key pressed during power on f 4 Repeat procedure and observe causal E3 07 35 R key pressed during power on error messages E3 07 36 A APAR pressed during power If the error occurs repeatedly e Replace user interface with electronics DX7 Error code Description Actions required see E7 07 40 No valid language set found Check selected language set by running S gt 258 service routine S017 5 correct if necessary Check whether selected language setis S gt 61 already installed perform software update if necessary Please contact the Sirona Customer S gt 61 97 Location 10 System hardware programmable logic component version for current operating mode compatibility replace board DX1 if necessary Error code Description Actions required see E7 10 01 EEPROM cannot be written e Acknowledge error and repeat procedure If the error occurs repeatedly S gt 350 e Replace board DX11 Error code Descriptio
111. 7 Inform the customer of the configuration options of the software status for example the acoustic exposure signal Activate these functions if they are required If the travel height of the unit has to be limited Set the travel height with service routine S018 2 gt 272 Perform a software update to the current software version gt 61 This updates all modules in accordance with the configuration The error message E1 11 20 invalid unit calibration is displayed Acknowledge the error message with the R key Perform a complete unit calibration gt 155 After a successful unit calibration has been performed the error message should no longer appear Call up the Extended Details via SixABCon This generates an XML file with the system parameters which is Filed in the PDATA P2K_Config directory under the network name of the unit The process is completed 355 9 Repair Sirona Dental Systems GmbH 9 12 Boards Service Manual GALILEOS Case D e DX11 from another unit e GALILEOS Compact e System software version V03 03 01 or higher IMPORTANT Exchange is only possible within the same system class e g the DX11 must come from a GALILEOS Compact unit if it is to be installed in a GALILEOS Compact unit After inserting the board you must reconfigure the IP address to match the IP address of the existing X ray component Before you set the unit to anew IP address make sure that the IP address you re as
112. Does the unit work Replace the cable 61 25 665 D3437 D3437 076 01 15 02 06 2012 141 Error analysis of X RAY control signal path Error analysis of X RAY control signal path up to unit serial number 3199 with board DX41 Error and help messages with remote control installed E3 42 31 E3 13 40 E3 41 20 occur in combination after the unit is switched on with the door contact closed Switch the unit off and then on again Make sure that the release button is not actuated during Was the release button actuated when the unit was switched on switch on Do the error messages recur Short circuit in the coiled cable of the release button or on the membrane keyboard of the remote control Replace release button with Yes No coiled cable Replace remote control Problem solved 61 25 665 D3437 142 D3437 076 01 15 02 06 2012 E3 42 31 occurs once after the unit is switched on E3 41 20 occurs once after the unit is switched on E3 41 24 occurs once during operation of the unit 61 25 665 D3437 D3437 076 01 15 02 06 2012 Hardware error on display board DX42 Replace remote control Is the door contact open Yes Short circuit in the coiled cable of the release button or on the membrane keyboard of the remote control Replace release button with coiled cable Replace remote control e Close door contact Ye Switching the unit ON an
113. EOS 10 9 Checking the device leakage current 1 Switch the line voltage off at the main switch of the building installation 2 Disconnect the power cable and the second protective ground wire from the building installation 3 Remove the profile cover 4 Check whether the unit power switch is turned on 5 Connect a high resistance measuring voltage source between the short circuited power cable B and the protective ground wire A 6 Measure the voltage drop via MD The measured value must not exceed 5 mA Ifthe leakage current is not correct Troubleshoot the problem following the instructions in Device leakage current too high gt 140 7 Reconnect the unit to the building installation see the installation instructions for the unit 61 25 665 D3437 D3437 076 01 15 02 06 2012 383 11 Dismantling and disposal Sirona Dental Systems GmbH 11 1 Dismantling and reinstallation Service Manual GALILEOS Dismantling and disposal Dismantling and reinstallation When dismantling and reinstalling the system proceed according to the installation instructions for new installation in order to guarantee its proper functioning and stability The X ray unit must be recalibrated whenever structural alterations in the area surrounding the X ray room or new installations have been performed Disposal Your product is marked with the adjacent symbol Within the European Economic Area this product is subject to Directive 200
114. ESCAN gt 24 optional D O gt 61 25 665 D3437 D3437 076 01 15 02 06 2012 21 3 Unit description Sirona Dental Systems GmbH 3 2 Hardware Service Manual GALILEOS Slide Comfort GALILEOS Compact Component Designation Function Boards DX1 Open loop closed loop control in general DX11 Controller board DX6 Open loop closed loop tube assembly DX7 Easypad touchscreen GALILEOS Comfort DX71 LED display on Multipad GALILEOS Compact DX89 Image memory of the X ray detector Motor MU Rotary movement of rotating element Light barriers LS Position control of the ring cycle LS Position control of the swivel arm not available as individual repair part see spare parts list 61 25 665 D3437 22 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS 3 Unit description 3 2 Hardware Stand MHV e DX32 DX41 Component Designation Function Boards DX32 Power supply board DX41 Interface board Motor MHV Linear movement of height adjustment From unit serial number 3101 new units are supplied with a new version of the board DX32 see chapter Board photos gt 31 Starting with unit serial number 3201 new units will be delivered without board DX41 61 25 665 D3437 D3437 076 01 15 02 06 2012 23 3 Unit description Sirona
115. H 8 5 S005 General X ray tube assembly service Service Manual GALILEOS A progress indicator is displayed during the service function 3 After the adjustment has been performed exit the service routine gt 220 2mV VH 61 25 665 D3437 228 D3437 076 01 15 02 06 2012 007 Error logging memory SR SHZ Function S007 Error logging memory 007 1 No Display error logging memory 007 2 Yes Clearing error logging memory 007 5 No Enabling CAN bus logging in the web interface SR service routine SHZ security access S007 Test step 1 Display error logging memory In addition to service routine S007 1 you can also use the extended detail query in SiXABCon to check the error logging memory Easypad touchscreen Symbol on the control Status Function panel Patient symbol key 1 B is selected Step width for scrolling between error events 1 Patient symbol key 2 C is selected Step width for scrolling between error events 10 Patient symbol key 3 D is selected Step width for scrolling between error events 100 Selection field Selection display 1 Error event 2 Error code for the selected event 3 Date and time of the selected error event Factory setting 1 Call service routine S007 1 gt 213 2 Use the patient symbol keys B C D to select the step width for scrolling between the error events se
116. H at 49 0 3304 3919 590 You can also get in touch with the following contacts for disposal orders and any questions relating to this you may have Phone 49 0 3304 3919 500 By e mail pickup eomRECYCLING com Mailing address enretec GmbH Gesch ftsbereich eomRECYCLING Kanalstrasse 17 16727 Velten Your equipment that is not permanently installed will be collected in the practice while your permanently installed equipment will be collected curbside at your address at the agreed time and date All disassembly transport and packaging costs are to be borne by the owner operator of the equipment The disposal itself is free of charge Worldwide outside Germany Please contact your local dental equipment specialist for country specific information on disposal The X ray tube assembly for this product contains an X ray tube with a potential implosion hazard a small amount of beryllium a lead lining and mineral oil The unit contains counterbalancing weights made of lead 385 Service Manual History Version 1 Software version V03 03 01 Version 2 General revision supplements and corrections in chapter 4 Calibrating the unit Version 3 Software version V03 04 00 01 supplements and corrections in chapters 1 General information software versions and cable L13 and 4 Calibrating the unit mechanical adjustment and dosimetry menu supplements and corrections in chapter 6 Repair laying cables supple
117. ILEOS Compact as well as on the remote control display if there is one The codes provide you with error type error location and troubleshooting information Error code Ex yy zz The error messages are encoded according to the following pattern Ex Error type Troubleshooting classification for the user yy Location Module subsystem or logical function unit ZZ Consecutive Identification of error number Ex Error type Identifier x is intended to help you reach a decision quickly on how to proceed with the corresponding error x Description Error group Actions required 1 System warning System message This error group includes all errors that indicate still acceptable tolerance variations or messages about states which do not directly affect system operation e Acknowledge the error message If the error occurs again Restart the unit 1 Switch off the unit 2 Wait 1 minute 3 Switch unit on 4 Repeat procedure If the error occurs again e Run an error diagnosis gt 88 2 Errors caused by system overload This error group includes states that indicate temporary overtemperatures or similar for example The cause of the error disappears automatically after a certain waiting time e Acknowledge the error message e Repeat the procedure step after a certain waiting time If the error occurs again e Extend the waiting
118. If no light is visible S gt 310 e Replace the tube assembly Error code Description Actions required see E5 06 30 Total radiation time exceeded If a CAN bus error had been reported before e Check the CAN bus S gt 128 e Please contact the Sirona Customer S gt 61 Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary Error code Description Actions required see E5 06 31 Partial radiation time exceeded If a CAN bus error had been reported before e Check the CAN bus S gt 128 e Please contact the Sirona Customer S gt 61 Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary Error code Description Actions required see E5 06 32 Minimum preheating time not If a CAN bus error had been reported observed before e Check the CAN bus S gt 128 e Please contact the Sirona Customer S gt 61 Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary Error code Description Actions required see E1 06 40 Tolerance exceeded e Replace the tube assembly S gt 310 VH nom Error code Description Actions required see E1 06 41 Tolerance exceeded kV nom e Replace the tube assembly S gt 310 61 25 665 D3437 91 D3437 076 01 15 02 06 2012
119. LEOS 10 6 Checking the grounding straps Checking the grounding straps Grounding strap in the stand 61 25 665 D3437 D3437 076 01 15 02 06 2012 373 10 Maintenance Sirona Dental Systems GmbH 10 6 Checking the grounding straps Service Manual GALILEOS Grounding strap on the image detector Check whether the grounding straps 1 Perform a visual and hands on inspection of the grounding straps to have complete and firm contact ensure that they have complete and firm contact at the positions marked If the grounding straps do not have correct contact Fasten the grounding straps correctly 2 Perform a visual inspection of the grounding straps for damage If the grounding straps are damaged Replace the grounding straps 61 25 665 D3437 374 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 10 Maintenance Service Manual GALILEOS 10 7 Checking the cable shields Checking the cable shields Shield on the tube assembly 61 25 665 D3437 D3437 076 01 15 02 06 2012 375 10 Maintenance 10 7 Checking the cable shields Sirona Dental Systems GmbH Service Manual GALILEOS Shield on the housing Check whether the cable shielding is OK gt Perform a visual and hands on inspection of the cable shields to ensure that they have complete and firm contact at the positions marked If the cable shields do not have correct contact Fasten the cable shields correctly 61 25 665 D3437 376 D3437 076
120. New as of 06 2012 GALILEOS Service Manual English Sirona Dental Systems GmbH Service Manual GALILEOS Table of contents 1 About this Service MANU cl ecucts casts casicaid ces orate Suse al dietcmeuluGues duae ou hed dea bay iaset eiate 12 1 1 SCOPE a oe RAAE EE AREARE EAEE AE A E e R E A 12 1 2 Other documentation required c cccesseceeecccenceceeeeeeeceeeeeeceeaeceeeeeeeeeeeeees 12 1 3 Tools and auxiliary Materials iz ceccicsty sees dees eeiioridaes yas deetede ola etal uaelnadees 13 1 4 Structure of the document ticcicccewstsSecsctokcccsdd ems ancecteksteay eteceeeeieelcocacnas 14 1 4 1 Identification of danger leVelS eee eeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeteeaaees 14 1 4 2 Formats and symbols used seeseeeeeeeerieetttterinrrertesssinnnrnssssrrnnn 14 2 Safety INSMUCUONSG iaioo aeiaai iaa a e a aiaa 15 2 1 Modifications to the unit ccccceeeeeeeseeseceee cee eeeeeneneeeneeeeeeeseeneneeeeneereeeeaes 15 22 HIKED connectiON kisss iar iaae tnd a EAE EA E N R t 15 2 3 Electromagnetic compatibility 2 20 0c cceceeccccceeeeeeeeeeeceeeeeeeeeseeneeeenereeeeeeane 15 24 El trostatic CISCNAIGG sorrie agesaes nana uted vetdnies aga eaa 15 27 Switching the CIE SOM isa ccs a ote hse a a nal Gad Coad elo Oe elcid 15 2 6 SGONOSNS AVON s mosses siesta Se Bo eaae dis eh AA ah See caucasian 16 2 7 Laser light localizer s no oie ie SR Ad Ae eek oe
121. OS Error code Description Actions required see E1 10 23 Board DX89 does not have valid e Perform service routine S009 7 copy data S gt 245 data via the X ray detector from DX11 to board DX89 S gt 319 The X ray detector has been e Replace the X ray detector E1 10 24 replaced Board DX89 does not Please report this event to the Customer have valid data via the X ray Service Center to help us improve the detector product This error message should not occur in the application Board DX89 was replaced and must ef 1028 be registered in the system 9 y e Perform service routine S009 7 copy data The X ray detector has not been from DX11 to board DX89 E1 1026 initialized Board DX89 does not Replace the X ray detector have valid data via the X ray detector This error message should not occur in the application Please report this event to the Customer Service Center to help us improve the product 100 61 25 665 D3437 D3437 076 01 15 02 06 2012 Location 11 Power PC Board DX11 Error code Description Actions required see E6 11 01 Program sequence error e Please contact the Sirona Customer S gt 61 Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary e Acknowledge error and repeat procedure If the error occurs again e R
122. TANT For Facescan units The FaceScan must be removed gt 325 from units with FacescanFacescan fitted before the X ray detector can be replaced Removing the covers gt Remove the x ray detector cover Remove cable 1 CAUTION Risk of injury The cover plate may have sharp edges Carefully pull cover plate A upwards to remove it from the X ray detector 2 Loosen the screws B and C as well as clamps D and E 3 Remove cable L13 from connector X201 F on board DX89 61 25 665 D3437 D3437 076 01 15 02 06 2012 319 9 Repair Sirona Dental Systems GmbH 9 8 X ray detector Service Manual GALILEOS 4 NOTICE Make sure that the grounding cable does not slip into the ring Secure it with a cable tie or piece of adhesive tape if necessary Loosen screw G and disconnect the grounding cable H Removing the X ray detector gt CAUTION The x ray detector is heavy Loosen the screw J swing the X ray detector slightly upwards and lift it out of the holder on the ring IMPORTANT Depending on the unit hardware version involved there may be a second screw located on the side opposite screw J If so this screw must be loosened in order to remove the X ray detector This second screw does not have to be used during Q reassembly ooososoodeoooooooo oo e oo ooooo0o00000000 0000000 61 25 665 D3437 320 D3437
123. a service routine 009 4 gt 241 e Run software update gt 61 e If this step does not lead to the desired result board DX89 must be replaced gt 346 If all LEDs light up after the power up phase this leads to conclusions concerning a defect on board DX89 See the procedure outlined above for troubleshooting NOTICE The unit must be switched off before disconnecting any plugs or cables LED on There is a connection to the CCD sensor in the camera head LED off There is no connection to the CCD sensor in the camera head Action e Check the plug connections and connection cables between board DX89 and the x ray detector if necessary replace cable L28 or the x ray detector NOTE In x ray detectors with a serial number 2 5000 cable L28 cannot be replaced If all LEDs light up after the power up phase this indicates that there is a defective FPGA on board DX89 For troubleshooting see the action under LED OFF for PLL_FPGA LED on Gettering is o k LED flashing after a waiting Gettering is not o k period of 12 minutes Action e Replace the X ray detector gt 319 The free ions are pumped out of the vacuum of the X ray detector by the getters hence the name getter pump The getter current is measured during operation If this does not drop below a certain value within 12 minutes the gettering is not o k In this case the error is probably caused by a defective x ray detector 153
124. agm Radiation field Dose Sensor lis Shading Distortion Geometry gt Eikii Diaphragm A Diaphragmexposue SS tadiation field Shading Dist Message window Tabs Preview image Status column m ol olo gt Tools pictograph The message window displays text messages about the calibration process The Service exposure menu guides you through the calibration process and contains the following tabs via which you can open and change back and forth between eight index cards submenus e Diaphragm e Radiation field e Dose e Sensor e Iris e Shading e Distortion e Geometry e Service 159 7 Adjusting calibrating the unit Sirona Dental Systems GmbH 7 1 General information about unit adjustment and calibration Service Manual GALILEOS Preview image The Diaphragm and Shading submenus each contain a preview image that symbolizes the exposure to be taken during the calibration step Due to the varying geometry of the GALILEOS Comfort and the GALILEOS Compact diaphragms the preview images displayed in these submenus differ slightly GALILEOS Comfort GALILEOS Compact We use only the display of the GALILEOS Comfort in these instructions unless explicit reference is made to the GALILEOS Compact Status column To the right of the menu you can see the status column This column provides information about the system s curre
125. aking the measures listed another type of error is the cause In such cases you should run an error diagnosis gt 88 Help code Description Actions required H3 01 R button move into starting position e Press the R key e Panoramic unit moves to starting position H3 20 R button confirm exposure data e Press the R key e Exposure data are confirmed H3 21 Close the door e Close the door or check door contact H3 23 Swivel pendant into end position e Move the swivel arm to its end position completely open or completely closed H3 24 The X ray detector preparation is in progress Wait until the X ray detector is ready This can take up to 10 minutes H4 03 Switch SIDEXIS to ready for exposure state Make SIDEXIS XG ready for exposure H4 07 SIDEXIS 3D Vorauswahl korrigieren Correct SIDEXIS XG 3D preselection H4 08 SIDEXIS 3D Aufnahme w hlen Correct SIDEXIS XG 3D preselection H4 20 Get existing exposure IMPORTANT Do not switch the system off until the help message has disappeared e Get exposure with Sirona Control Admin see SIDEXIS XG Operator s Manual REF 59 62 134 61 25 665 D3437 D3437 076 01 15 02 06 2012 83 5 Messages 5 2 System messages Sirona Dental Systems GmbH Service Manual GALILEOS System messages System codes are only displayed on the Multipad GALILEOS Compact and the remote control System messages are displayed in plain text
126. al X ray tube assembly service Service Manual GALILEOS S005 Test step 5 Temperature sensor test single tank 1 Call service routine S005 5 gt 213 After the service routine has been selected selection field 1 displays the single tank temperature in C The display is updated once per second 2 Exit the service routine gt 220 61 25 665 D3437 226 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 5 S005 General X ray tube assembly service S005 Test step 8 Automatic adjustment of pulse preheating 1 Call service routine S005 8 gt 213 An inactive progress indicator in selection field 1 and the message FFFF in selection field 2 signal that the system is ready for compensation 2 Start the automatic adjustment by pressing and holding the release button IMPORTANT Keep pressing the release button until adjustment is completed and the new offset value for preheating is displayed If you interrupt the adjustment procedure prematurely by letting go of the release button the message EEEE appears in selection field 2 This message must be acknowledged by pressing the R key When pressing the release button radiation is released for 2 s to warm up the tube assembly to operating temperature This is followed by the automatic tuning routine 61 25 665 D3437 D3437 076 01 15 02 06 2012 227 8 Service Sirona Dental Systems Gmb
127. al present e Check cable L17 replace if necessary S gt 141 software signal not present S gt 362 e Check board DX42 replace if necessary S gt 134 S gt 343 e Check board DX41 replace if necessary S gt 134 S gt 343 110 61 25 665 D3437 D3437 076 01 15 02 06 2012 Location 42 Remote control Error code Description Actions required see E6 42 01 General module initialization error Check unit configuration with or without S gt 263 Error generated during module self DAS ey super SOGE routne S0173 S gt 61 iest correct the configuration if necessary e Please contact the Sirona Customer 9 1343 Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary e Replace board DX42 Error code Description Actions required see E6 42 02 Invalid system data or uninitialized Please contact the Sirona Customer S gt 61 module storage data Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary e Acknowledge error and repeat procedure If the error occurs repeatedly S gt 343 e Replace board DX42 Error code Description Actions required see E6 42 03 Invalid commanding or control data e Run service routine S008 2 to check S gt 238 software version of DX42
128. aluate the image 168 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS 7 Adjusting calibrating the unit 7 3 Adjustment and calibration via the Service Functions menu A B Adjustment OK C Adjustment not OK diaphragm not centered D Adjustment not OK diaphragm too small E Permissible tolerance 65 pixels 5 pixels 61 25 665 D3437 D3437 076 01 15 02 06 2012 169 7 Adjusting calibrating the unit Sirona Dental Systems GmbH 7 3 Adjustment and calibration via the Service Functions menu Service Manual GALILEOS The brightness distribution along the surrounding border must be uniform A A shadow in the image indicates horizontal or vertical displacement of the diaphragm The distance between the bottom edge and the lowest point in the image should be 65 5 pixels measure with SIDEXIS scale No surrounding gray shadow should be visible D A surrounding gray shadow in the image indicates that the diaphragm is too small If the distance between the bottom edge and the lowest point in the image is out of tolerance E or the brightness distribution along the surrounding border is not uniform C or a surrounding gray shadow is visible in the image diaphragm opening too small D the diaphragm must be adjusted mechanically gt 197 5 If the exposure is OK A B confirm this by clicking the check box underneath the message wi
129. alue greater than 1500 61 25 665 D3437 D3437 076 01 15 02 06 2012 275 8 Service Sirona Dental Systems GmbH 8 12 S018 Service for height adjustment Service Manual GALILEOS 2 Use UP DOWN keys A on the control panel to move the unit up and a down and use patient symbol keys F to check the switching states 3 Exit the service routine gt 220 i sirona 61 25 665 D3437 276 D3437 076 01 15 02 06 2012 S037 Network service SR SA Function S037 Network service 037 1 No Displaying the network data 037 2 Yes Delete network addresses or set them to factory defaults 037 3 Yes Set boot mode DYNAMIC DHCP AutolP STATIC fixed address 037 4 Yes Manual input of static network settings IP address default gateway address and subnet mask 61 25 665 D3437 D3437 076 01 15 02 06 2012 SR service routine SHZ security access S037 Test step 1 Displaying the network data If all network data is set to default the system is in UDP boot mode Symbol on the control Status Meaning panel Patient symbol key 1 A The IP address is displayed in selection field 1 Patient symbol key 2 B The default gateway is displayed in selection field 1 Patient symbol key 3 C The subnet mask is displayed in selection field 1 Factory setting Selection f
130. ance 120Q or 0Q Resistance 609 Resistance lt 509 Fault in the electrical CAN bus connection Yes For units starting from serial number 3201 Check cable L117 or manual trigger A2 without remote control For units up to serial number 3199 Check correct position of jumpers on board DX42 Is cable L117 or manual trigger A2 correctly connected Are the jumpers set correctly Yes Yes No Check correct position of jumpers Plug in jumpers on board DX1 correctly Are the jumpers set No Plug in jumpers correctly correctly Fault in cabling at plug connectors or in the modules es Check CAN connector from board No Establish plug DX1 to the connected modules connection Yes 61 25 665 D3437 D3437 076 01 15 02 06 2012 129 Locate the problematic component or defective cable using the elimination process Cyclical test with bridging the components Replacement cable available Cyclical test with wandering replacement cable Make the connection with the replacement cable for each module in succession and measure the ohmic resistance on board DX1 at connector X310 between measuring pins CAN H and CAN L in each case Jumper the individual bus outputs to the modules in succession and re measure the ohmic resistance on board DX1 at connector X310 between measuring pins CAN H and CAN L in each case Has a bus resistance of 600 been
131. any changes to settings or processes that are not described in these operating instructions This may lead to a dangerous exposure to radiation Ventilation slots Never cover the ventilation slots on the unit under any circumstances since this may obstruct air circulation This can cause the unit to overheat Qualifications of service personnel Installation and startup may be carried out only by personnel specifically authorized by Sirona Radiation protection The valid radiation protection regulations and measures must be observed The statutory radiation protection equipment must be used During an exposure the service engineer should move as far away from the X ray tube assembly as the coiled cable of the manual release permits With the exception of the service engineer no other persons are allowed to stay in the room during an exposure In case of malfunctions cancel the exposure immediately by letting go of the exposure release button Safety checks After implementing repair work protective conductors and device leakage current checks must be carried out see the sections on Checking protective conductor and Checking device leakage current 61 25 665 D3437 D3437 076 01 15 02 06 2012 3 Unit description 3 1 Unit classes and versions Sirona Dental Systems GmbH Service Manual GALILEOS Unit description Unit classes and versions A GALILEOS unit B Control panel C Facescan
132. approx 30 mins and repeat this procedure If the error occurs repeatedly S gt 310 e Replace the tube assembly 61 25 665 D3437 92 D3437 076 01 15 02 06 2012 Error code Description Actions required see E6 06 60 TDI signal from board DX11 to e Replace cable L15 S gt 362 board DX6 is disturbed e Replace board DX1 S gt 342 BE Signal to start synchronized e Replace the tube assembly S gt 310 readout sequence and to prepare the next exposure Error code Description Actions required see E6 06 65 Tube current or tube voltage is too Replace the tube assembly S gt 310 high in standby mode Error code Description Actions required see E6 06 66 Impermissible tube type e Check tube type of tube assembly S using extended detail query or service S gt 223 routine S005 1 replace tube assembly i if necessary S gt 310 Error code Description Actions required see E6 06 67 Light guide input TDI is active during Check TDI signal Swen On 1 Switch off the unit TDI Signal to start synchronized 2 Disconnect cable L15 at board DX11 readout sequence and to prepare 3 Switch unit on the next exposure 4 Perform visual check at socket J5 e f light is visible Replace board DX11 S gt 342 e fno light is visible Replace the tube S gt 310 assembly Error code Description Actions required see E6 06 68 Tube assembly output aft
133. ard DX89 The LEDs on the board can provide information about the possible cause of the error LEDs on board DX89 gt 152 61 25 665 D3437 D3437 076 01 15 02 06 2012 151 6 Troubleshooting Sirona Dental Systems GmbH 6 9 Fault diagnosis of the X ray detector and on board DX89 Service Manual GALILEOS LEDs on board DX89 The diodes PLL_FPGA PLL_CCD as well as A image memory test and D gettering in particular should be observed LED statuses and their significance in case of an error gt 153 Teeeede 00 28V 24V 5V 3 3V Image memory test Operating voltage on board DX89 Operating voltage on image amplifier oloo gt Gettering 61 25 665 D3437 152 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS 6 Troubleshooting 6 9 Fault diagnosis of the X ray detector and on board DX89 PLL_FPGA PLL_CCD Gettering 61 25 665 D3437 D3437 076 01 15 02 06 2012 LED statuses and their significance in case of an error For X ray detector errors it is usually necessary to send the extended details of the unit to the Sirona Customer Service Center CSC The results of the LED inquiry described below also must be added to the extended details The LED statuses specified here apply to the booted system LED on FPGA on DX89 has started properly LED off FPGA on DX89 has not started properly Action e Format flash file system vi
134. atic e Check fuses F100 and F103 fuses trip only up to unit serial number 3199 and replace if necessary e Switch on the unit and check its function Press on the automatic fuses and check unit function Replace board DX32 Does the unit work Fault has been removed 61 25 665 D3437 D3437 076 01 15 02 06 2012 137 Checking the motors Perform visual inspection of motor Is the motor visually damaged Are the plug connections defective or non existent Temporarily install replacement motor and check unit function Yes Replace the motor or fix the plug connections Does the unit work Replace the End motor No Motor is OK Continue troubleshooting according to list of faults chapter Messages 61 25 665 D3437 138 D3437 076 01 15 02 06 2012 Checking the light barriers Is the cable connection between light barrier and board OK Yes Isa replacement light barrier available Yes Temporarily install replacement light barrier and check unit function Does the unit work Light barrier is OK Continue troubleshooting according to list of faults chapter Messages Replace light barrier 61 25 665 D3437 D3437 076 01 15 02 06 2012 Clean the plug If defective correct the plug or replace defective component Actuate light barrier manually and check signal change on pin 3 of light barrier connec
135. avenets 31 3 2 5 2 Boards in the sland xcc i2 Onan Ra eee Ole ais 37 3 2 5 3 Board in the remote control ices est acter Gene rcenversee eda eeedalice 39 3 2 0 4 Boards in FACESCAN deena Gander sein a A 40 D2 O COVENS ieee eA aee TTE ale ic tiwvodtecutecv A EE Ate Oheenoed 42 32 6 Technical dataivcs catiisictea selene G ae eas 45 32T DIAQrAIM S aiee aE E E A E E E E E E 48 3 3 Software compatibility cn cccenn2e2is sac cx rede capbeaasudecene nctegaenaaa ates vetuaees aon tar saeecenetaseee 50 3 3 1 GALILEOS firmware ccccccec cece cece cece cece eee E aree e ra eiaeiiai 50 3 3 2 FACES CAN firmware oie ccieeadees hee ndei ak aicacendeaaaatincceeeneedealneedeet enn 56 3 3 9 GALILEOS SORWANGs ez nic siocd n a aia r a Tei siaSi 56 General operating PrOCECUIES ccccccccccecccceecceceeceececeeeeeceeeeeeeeeeeeeeeeeeeeeeeneeeeeees 58 4 1 Switching the unit on pic csecccaccscnaveece Sebnccnghanhavtieiedacueandbaaeesiasennaetadsecbelasteebiawede 58 4 1 1 Switching the GALILEOS Comfort On sseesesseseeeeeeereessreeeerresesrena 59 4 1 2 Switching the GALILEOS Compact On s s sseseseeeeeeeeeessreeeerresesreena 60 4 1 3 Factory setting after switch on ee ccc eee eeeeeneeeeeeeeeeeeeeeeeeaaaees 60 4 2 Updating the firmware os c5 2cecceatseacecGteseietasie ts wotdek suatnant padetiad ds aduedeetepeazen 61 4 2 1 Updating the unit firmware c cccecccececceceeeeeeeeeeeneeeeneeeeeeeeeeeeneene 61 4 2 1 1 Update m
136. ayed 61 25 665 D3437 D3437 076 01 15 02 06 2012 245 8 Service 8 8 S009 Flash file system Sirona Dental Systems GmbH Service Manual GALILEOS 2 To trigger the memory process press the Memory key A R key B Easypad or LED above the R key B Multipad lights up followed by the R key B The data are transferred During the data transfer a progress indicator is displayed in selection field 1 3 Exit the service routine gt 220 61 25 665 D3437 246 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 9 S011 Dosimetry without ring movement S011 Dosimetry without ring movement Unit is radiating X rays Excess exposure to X rays is detrimental to health gt Use the prescribed accessories for radiation protection gt Do not stay in the X ray room during exposure Move as far away from the unit as the coiled cable for the release button allows you to SR SHZ Function S011 Dosimetry without ring movement 011 9 Yes 4s Continuous radiation with 85kV 7mA for current measurement 011 12 Yes Dosimetry with pulsed radiation A X RAY 85kV 42mAs 61 25 665 D3437 D3437 076 01 15 02 06 2012 SR service routine SHZ security access 011 Test step 12 Dosimetry with pulsed radiation 1 Call service routine 8011 12 gt 213 Selection field 1 displays 85kV 42mAs
137. bH Service Manual GALILEOS 9 9 Facescan 5 Read the serial number on the identification plate S and record this on the device certificate 6 Screw the right end cap J with two screws K down on the scan unit 7 f network configuration is to be completed later using the Facescan USB stick Do not screw the cover cap A tight or gt If network configuration is to be completed later using a network cable Screw the cover cap A down with two screws What has to be done after replacing the scanner unit 1 Perform a white balance gt 190 2 Perform a complete unit adjustment or calibration gt 155 3 Perform some test exposures 61 25 665 D3437 332 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 9 Facescan Replacing the PoE module Removing the faulty POE module 1 Remove the bottom profile cover gt 42 2 Remove cable L76 from the cable clamp G 3 Detach cable L77 WH white and L77 BN brown from terminals K1 2 and K1 1 orange 4 Pull cables L71 and L7 from sockets SC SC and BU1 61 25 665 D3437 D3437 076 01 15 02 06 2012 333 Sirona Dental Systems GmbH Service Manual GALILEOS 9 Repair 9 9 Facescan 5 Push the cable cover I upwards 6 Unplug the Ethernet cable L77 from the POE socket of the PoE module RAK NRL z OANA 7 Unscrew the two screws K and the nuts L and remove the PoE module 61 25
138. bH 8 Service Service Manual GALILEOS 8 13 S037 Network service 4 Now use the patient symbol keys to select the required number pad 1 4 A D see also table A Number pad B1 B Number pad B2 C Number pad B3 D Number pad B4 The patient symbol key selected in each case Easypad or the LED above the patient symbol key selected in each case Multipad lights up Digit No 12 3 The digit currently selected for changing is displayed in selection 192 168 015 178 field 1 Digit No 3 in the example T T j T Important The number of the digit always refers to the currently selected number pad The current value of the corresponding digit is displayed in selection field 2 2 in the example Digit No 3 61 25 665 D3437 D3437 076 01 15 02 06 2012 285 8 Service Sirona Dental Systems GmbH 8 13 S037 Network service Service Manual GALILEOS Digit No 12 5 Use arrow keys E to select the digit to be changed in selection field 192 168 015 178 1 Digit No 12 in the example The corresponding patient symbol key D or the LED above the corresponding patient symbol key D lights up B4 Selection field 2 displays the value of the currently selected digit 61 25 665 D3437 286 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 13 S037 Network service Digit No 12 6 To change the value for the digit use arrow keys F in selec
139. board DX1 replace if necessary S gt 134 S gt 343 Error code Description Actions required see E1 89 34 X ray detector voltages inaccurate Check cable L27 DX89 power supply S gt 141 replace cable if necessary S gt 362 e Check x ray detector replace if S gt 319 necessary Error code Description Actions required see E2 89 35 Error in iris diaphragm positioning Restart the unit 1 Switch off the unit 2 Wait 1 minute 3 Switch unit on 4 Repeat procedure and check function If the error occurs repeatedly S gt 319 e Check x ray detector replace if necessary Error code Description Actions required see E7 89 37 Video amplification outside e Check board DX839 replace if necessary S gt 134 tolerance S gt 343 Error code Description Actions required see E2 89 38 Error in image signal during e Check cable L13 replace if necessary S gt 141 exposure S gt 362 e Check board DX89 replace if necessary S gt 134 S gt 343 e Check board DX1 replace if necessary S gt 134 S gt 343 Error code Description Actions required see E1 89 39 Error during X ray detector e Repeat procedure preparation If the error occurs repeatedly S gt 319 e Check x ray detector replace if necessary 61 25 665 D3437 124 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 6 Troubleshooting Service Manual GALILEOS 6 1 Error logging memory Troubleshootin
140. board DX6 E F G Laying cables correctly on the cover plate Cable L15 gt Socket J5 on board DX6 Cable routed on left side of tube assembly 2x L21 and L20 H Cable L12 routed on right side of tube assembly What has to be done after replacing the X ray tube assembly IMPORTANT Since board DX6 is also replaced with the tube assembly you MUST also follow the instructions in the chapter entitled Measures following replacement of boards gt 347 1 Perform a complete unit adjustment or calibration gt 155 2 Perform an acceptance test for Germany only without calling in an expert 61 25 665 D3437 D3437 076 01 15 02 06 2012 317 9 Repair Sirona Dental Systems GmbH 9 7 Fan X ray tube assembly Service Manual GALILEOS Fan X ray tube assembly Replacing the fan 1 Remove the Front tube assembly cover 2 NOTICE Cable Loosen the three screws A and carefully remove the cover plate including the fan 3 Pull the fan cable off of connector X2 on board DX6 4 Install the new fan in the reverse order of removal What has to be done after replacing the fan gt Check the function of the fan using service routine S005 4 gt 225 61 25 665 D3437 318 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 8 X ray detector X ray detector Replace X ray detector IMPOR
141. canner is automatically calibrated during the geometry calibration of GALILEOS In this case make sure that the normal room lighting is switched on during the calibration process The room does not have to be darkened during calibration ice Diaphragm Radiation field Dose Sensor Iris Shading Distortion Geometry je gt oe p Jiapnragm Geometric calibration There is no valid geometry calibration data on this machine Seriennr Kalibrierphantom Image acquisition Save values v The Service functions menu is called gt 161 1 Click the Geometry tab The corresponding tab card is selected Seriennr Kalibrierphantom 2 Read off the serial number of the geometric phantom from the ID Pas 2345 label of the phantom and enter it in the Seriennr Kalbrierphantom text box of the Geometry tab card 3 Insert the geometric phantom in the bite block holder of the unit gt 164 4 For Facescan Check to make sure that the room lighting is switched on Switch it on if necessary 5 Click on the mage acquisition button SIDEXIS XG makes the unit ready for exposure gt 162 Service routine 010 13 is displayed on the control panel 6 Press the R key to move the unit back to the starting position 61 25 665 D3437 D3437 076 01 15 02 06 2012 187 7 Adjusting calibrating the unit Sirona Dental Systems GmbH 7 3 Adjustment and calibration via the Service Functions m
142. ccceeeeeeeeeeeeeeeeeeeeeteeeenaeees 371 10 4 Checking the cables for damage ccccccccesscceencceeeeeseeseseeeeeeseeeeeseeneeeee 371 10 5 Checking the idling rollers ceceeecceeeeeecceeeeeeeeeeeeeseeeeeeeeaeeeeeesnaeeeeneaees 372 10 6 Checking the grounding straps ccccceeeeecesseseeeneceeteeeeeeeseneeceeeeeeeeneenenene 373 10 7 Checking the cable Shields ac cncccn cin tiat a ise tees aoe pees cert 375 10 8 Checking the protective ground wires ccceeeeeeeeeeeeeeeeeeeeeeneeeeeteenaeeeteeaees 377 10 9 Checking the device leakage current 0 eee eeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeneeenaaees 382 Dismantling and CSOSA se ccs acer vede srrenns tone See adeta cadens tvteeti dace sesieen eeeemnaghiee met enenrtegls 384 11 1 Dismantling and reinstallation 2 0 2 0 ccc cess eeeee eee eeeeeeeeeeeeeeeeeeeeeeeeeeeaaees 384 W122 DISPOSAL easiest hae atlas aac EE Gun uth ea EE 384 Service Manual PISLONY v2 cactccacevesdetinactihe doeeedpracii senusdrasenatond eva eetusl AKE E Eao oaa 386 02 06 2012 11 1 About this Service Manual 1 1 Scope 12 Sirona Dental Systems GmbH Service Manual GALILEOS About this Service Manual Scope This Service Manual describes the servicing of the GALILEOS Comfort and GALILEOS Compact digital volume tomographs It is intended for use exclusively by trained and authorized distributors and service technicians Other documentation required In addition to this manual
143. cceseeeeceeeseceeeeeeseeeaeaenees 111 5 5 10 Location 71 Multipad board DX71 cccccccccseeeeeesseceeeeeeseesaeaeeees 115 5 5 11 Location 89 X ray detector 2 cecececseccceeeeeeesesseceeeeteeeeseseeeeneseeeeees 119 6 TPOUBISSIOOEING ia iere eaa EEA AEE PAAR EEAS SOEREN RENERT EE 125 6 1 Error logging MEMOLSY eccccesceescecseecseeeceeceeeaaeeaeecaeecaeceeceaeceeeeeeeeeeeeeaeeess 125 6 1 1 Example of error logging data ccccccccecececeeeeeeeeeeeeeeeeeeeeeeteeeeeeees 126 6 2 Checking the CAN DUS airidas diee E aa AAEE e Ea 128 6 2 1 Checking the CAN bus with the diagnostic function of board DX1 132 6 2 2 Jumper positions in the CAN DUS ccccccecececeeeeeeeeeeeeeeeeeeeeeeenees 132 6 3 GheCKING the DoardS siter eie iaie inaa tle ia skh aeecasewe ethcneteaeectoes 134 6 3 1 Checking board DX32 eseeeesseeieeseteeerrrttetttrrrirtttesnsrinrnnnnsssreenn 136 6 4 Checking the Motors cccccecceececeeeeeeeeee eee e ence kenne nnne n nrnna nenene nnna rennene 138 6 5 Checking the light barriers ccccccccccccccceccececceceeeececeeececeeeeeeeeeeeeeeeeteeenees 139 6 6 Device leakage current too high sessssssssssiriesteeerrritttteeerririrrnsssreennnnssene 140 6 7 Checking the cableS nnnnsnnsnnnsnrrrrrrrrrrrrrrnrrnnrrnnrnnnnnnnnnnnn nannt nanenane nnne 141 61 25 665 D3437 D3437 076 01 15 02 06 2012 6 8 Error analysis of X RAY control signal path
144. ck the cables before replacing them gt 141 The cables are labeled with small flags They specify the designation and part number of the cable The plugs and sockets on the cables are designated both on the boards and cables Check the designation when you pull off the cables Some cables feature markings of green adhesive tape Mark the corresponding positions on the unit before removing an old cable Lay the new cable so that the cable markings again come to rest at the corresponding positions marked on the unit while removing the old cable An overview of all cables can be found in the chapter 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 13 Cable Replacing fiber optic cable L5 L6 or L15 IMPORTANT If a radius limiter is not yet installed When replacing one cable all existing fiber optic cables L5 L6 or L15 should be retrofitted with the radius limiters included with delivery The radius limiters improve torsional and bending force tolerance 1 Remove the defective fiber optic cable os PON 2 NOTICE Do not kink or twist fiber optic cables the bending radius may not be less than 20mm otherwise it is at risk of breaking Attach the radius limiter A close to the connector which is plugged onto board DX1 onto the cable 3 Plug the connector of the new fiber optic cable to the same color assignment on
145. cover 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 3 Unit description Service Manual GALILEOS Chassis X ray tube assembly Detector Facescanner optional Geometry Scanning operation 61 25 665 D3437 D3437 076 01 15 02 06 2012 Technical data Model designation Nominal voltage Permissible fluctuation Permissible drop under load Rated current Nominal power output Current time product Nominal frequency Internal line impedance Main building fuse Power consumption Focal spot size acc to IEC 60336 measured in the central X ray beam kV mA Pulsed mode Total filtration of X ray tube assembly Cone beam angle High voltage generation frequency Type Image intensifier I 1 Thales or Siemens Active input window size Camera Model designation Maximum current Power consumption Weight Source I I converter coating distance central X ray beam Source isocenter distance central X ray beam Source skin distance minimum distance Orbital angle 3 2 Hardware GALILEOS 200 V 240 V 10 10 6A 0 6 kW at 85 kV 7 mA 42 mAs 50 Hz 60 Hz max 0 8 ohms 25 A slow blow 16 A for single line 0 9 kVA 0 5 85 kV 5 mA 7 mA 10 ms 30 ms gt 2 5 Al 90 IEC 60522 collimated to approx 24 80 kHz 100 kHz 215 mm 8 1 2 diameter Pixels 10002 FPS 15 30 Dynamics 12 bits 4096 brightness values 60
146. cription Actions required see E7 07 10 Module is stuck in bootloader stage Check user interface with electronics DX7 note LED states If the board remains in the bootloader S gt 61 stage e Repeat the software update e Replace user interface with electronics S gt 307 DX7 Error code Description Actions required see E7 07 12 Unit is not ready for operation e Check the CAN bus S gt 128 Therefore the error can only be This error is a sequential fault displayed on the remote control e Restart the unit DX42 1 Switch off the unit 2 Wait 1 minute 3 Switch unit on 4 Repeat procedure and observe causal error messages Error code Description Actions required see E6 07 20 Contact to DX11 interrupted during Note error message on remote control S operation DX42 and check log memory via extended details e Check the CAN bus S gt 128 e Check cable L9 replace if necessary S gt 141 S gt 362 Error code Description Actions required see E7 07 21 No CAN bus connection DX11 does Start the detail query via SiXABCon not start lf DX11 responds S gt 362 Occurs in the start screen after e Check the signal path to DX7 repair or S gt 342 power on replace cables connectors if necessary e Replace board DX1 If DX11 does not respond S gt 342 e Replace board DX11 61 25 665 D3437 96 D3437 076 01 15 02 06 2012 61 25 665 D343
147. ct e System software version V03 06 01 or higher DX7 e Inserting a new Easypad incl DX7 e Switch the unit on S gt 61 Easypad e System software version e Perform a software update for the V03 03 01 or higher unit to version V03 03 01 or higher After replacement of the Easypad the language set on the board DX7 is set to the factory default setting 00 German English French Italian If the configured unit language set which can be queried by running service routine S017 5 or via the Extended Details in SiXABCon has a configuration other than 00 this configuration will be copied to board DX7 by the update function DX71 Inserting a new Multipad incl DX71 e Switch the unit on S gt 61 a e System software version e Perform a software update of the VO3 06 07 or higher device to a version V03 06 01 or higher DX32 e Inserting a new DX32 No further action is required Stand e System software version V03 03 01 or higher DX41 e Inserting a new DX41 e Switch the unit on S gt 61 sang e Only for GALILEOS up to serial e Perform a software update of the unit number 3200 to a version V03 03 01 or higher 61 25 665 D3437 348 D3437 076 01 15 02 06 2012 e GALILEOS Comfort e System software version Board Constellation Actions Page DX42 Inserting a new DX42 Silat e Inserting a new DX42 e Switch the unit on S 61 contro e GALILEOS Comfort e Perform a software update of the unit to a version V0
148. d all calibration data in the list are set to invalid red LEDs State Diaphragm Re field Diaphragm Radiation field Dose Sensor Iris Shading Distortion Geometry KI Distortion calibration There is no valid distortion calibration data on this machine 7 Geometry Distortion calibration Seriennr Kalibrierphantom Image acquisition Save values v The Service functions menu is called gt 161 1 Click the Distortion tab The corresponding tab card is selected Seriennr Kalibrierphantom 2 Read off the serial number of the distortion phantom from the ID label 12365 2345 of the phantom and enter it in the Seriennr Kalbrierphantom text box of the Distortion tab card 3 Clip the distortion phantom onto the X ray detector cover gt 163 4 Click on the mage acquisition button SIDEXIS XG makes the unit ready for exposure gt 162 Service routine S010 12 is displayed on the control panel 5 Press the R key to move the unit back to the starting position 6 Press the release button Press and hold down the button until the exposure is complete the preview image is displayed in the exposure window and the acoustic signal indicating the end of the exposure double beep sounds if it has been configured IMPORTANT This process takes approx 2 3 minutes The unit transfers the acquired images to the Reconstruction and rac Control Unit RCU This
149. d possibly after software updates See also the section titled Measures following replacement of boards gt 347 Ol Operating instructions Error code Description Actions required see E6 15 05 Unit serial number invalid or not e Run service routine S008 3 and confirm or S gt 238 available enter the unit serial number on the unit Occurs during first power on after replacement of board DX6 or DX11 See also the section titled Measures following replacement of boards gt 347 Error code Description Actions required see E6 15 10 Update file for module is corrupt e Obtain latest update file from the Sirona S gt 61 Customer Service Center CSC or the Sirona home page and perform software update 61 25 665 D3437 106 D3437 076 01 15 02 06 2012 Location 41 Media interface card Error code Description Actions required see E6 41 01 General module initialization error If the error is a software error known to the S gt 61 Sirona Customer Service Center CSC a S gt 343 software update bug fix must be p performed e Replace board DX41 Error code Description Actions required see E6 41 02 Invalid system data or uninitialized Ifthe error is a software error known to the S gt 61 module storage data Sirona Customer Service Center CSC a software update bug fix must be performed e Acknowledge error and repeat procedure If the error occurs again
150. d 1 or after pressing the T key number pad B2 is selected Patient symbol key 3 C The subnet mask is displayed in selection field 1 or after pressing the T key number pad B3 is selected Patient symbol key 4 D or after pressing the T key number pad B4 is selected Factory setting Selection field Parameter Display Meaning 1 IP address default unit gateway or subnet mask of the or after pressing the T key selected digit 2 default Fixed address static Fixed address modified setting dynamic Automatic address assignment Factory setting 283 8 Service 8 13 S037 Network service 192 168 15 240 DEFAULT 284 Sirona Dental Systems GmbH Service Manual GALILEOS Call service routine 8017 4 gt 213 Once the service routine has been selected the IP address of the unit is displayed in selection field 1 Easypad DEFAULT STATIC or DYNAMIC is displayed in selection field 2 see table You can display various items of network data in selection field 1 by pressing the patient symbol keys A B or C see table The patient symbol key selected in each case Easypad or the LED above the patient symbol key selected in each case Multipad lights up To change the selected parameter first press the T key E 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems Gm
151. d OFF again Is the door contact open Short circuit in signal path between boards DX41 and DX42 when switching the unit on Yes Replacing cable L17 Replace remote control Replace board DX41 Does the error message recur Yes The fault is corrected Short circuit in signal path between boards DX41 and DX42 when switching the unit on Replacing cable L17 Replace remote control Replace board DX41 143 6 Troubleshooting Sirona Dental Systems GmbH 6 8 Error analysis of X RAY control signal path Service Manual GALILEOS E6 13 43 occurs once during operation of the unit Was the door contact opened during the exposure e Acknowledge error message with R key Close door contact Repeat the exposure e Check door contact repair if necessary Replacing cable L17 Replace remote control Replace board DX41 H321 is triggered at start of exposure e Close door contact Start exposure again Is the door contact open Is the error message displayed again e Check door contact The fault is repair if necessary corrected Replacing cable L17 61 25 665 D3437 144 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS 6 Troubleshooting 6 8 Error analysis of X RAY control signal path E3 13 40 E3 41 20 occur in combination after the unit is switched on 61 25 665 D3437 D3437 076 01 15 02 06 2012 Error messages without i
152. d hold down the Service key until the patient symbol keys light up approx 2 s Then press the patient symbol keys in the sequence b d a within the next 4 seconds After you have entered the key combination correctly the service routine S017 test step 1 select confirm system class is started automatically The Memory key lights up Acknowledge any additional error messages with the R key Confirm the GALILEOS Comfort system class 03 To do this first press the Memory key R key lights up and then the R key Exit the service routine with the double arrow key Switch off the unit 61 25 665 D3437 D3437 076 01 15 02 06 2012 16 17 18 19 20 21 22 23 24 25 26 27 28 29 61 25 665 D3437 D3437 076 01 15 02 06 2012 Wait for approx 1 minute Then switch the unit back on The error message E6 15 05 undefined system serial number is displayed Acknowledge the error message with the R key Error message E6 15 04 undefined activation data is displayed Acknowledge the error message with the R key Call the service menu gt 210 Call the service routine S008 3 check the serial number and confirm this if necessary gt 240 The unit serial number is located on the nameplate of the unit NOTICE If the serial number is incorrect exit the update process and contact the Sirona Customer Service Center Switch off the unit Wait for approx 1 minute Then switch the u
153. dB Facescan 6 25 A 0 945 kVA 5 7 kg 510 mm 20 1 16 333 mm 13 1 8 approx 220 mm 8 5 8 204 45 3 Unit description 3 2 Hardware Reconstruction X ray tube 46 Scan time Number of single exposures Marking of focal spot Automatic exposure blocking Class device Degree of protection against electric shock Degree of protection against ingress of water Year of manufacture Mode of operation Long term power output Anode material Exposure parameters for determining leakage radiation Continuing current for leakage radiation measurements Transport and storage temperature Basic unit Detector Air humidity Admissible operating temperature Operating altitude Toshiba DF 151R or Siemens SR 120 15 60 Sirona Dental Systems GmbH Service Manual GALILEOS approx 14 s 200 The duration of automatic exposure blocking cooling period depends on the set kV mA level and the actual exposure time Depending on the tube load interval times of 8 s to 300 s are automatically set by the system 1 Type B device Ordinary equipment without protection against ingress of water m l 20XX on the rating plate Continuous operation 100 W Tungsten 7mA 85 kV 0 14 mA 40 C 70 C 40 F 158 F 30 C 55 C 22 F 131 F 10 95 without condensation from 10 C to 35 C 50 F 95 F lt 3000 m 61 25 665 D3437 D3437 076 01 15 02 0
154. dependent hammering noise may occur which points to bearing damage at the height adjustment motor If a hammering noise occurs Replace the height adjustment motor including spindle gt 290 Check whether precise jolt free height If the unit is not used for a longer period of time a slight jolt may occur the adjustment is possible first time it starts moving However the next time it starts moving it must execute a jolt free soft start gt Use the Up Down keys on the control panel to move the unit and observe the movement of the slide The slide must start in gentle starting and then change over to a faster movement If the height adjustment cannot be correctly positioned in detail using the gentle start Lubricate the spindle with a light coat of Chesterton 622 61 25 665 D3437 370 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 10 Maintenance Service Manual GALILEOS 10 3 Checking the fan and temperature sensor Check whether the height adjustment limit switches are functioning properly gt Manually press the actuators C of both limit switches D one after the other while the height adjustment motor is running The motor must stop If the motor does not stop Check the corresponding microswitch and replace if necessary Check cable L19 replace if necessary Check whether an audible signal canbe gt Use the Up Down keys on the control panel to move the unit up and heard during height
155. depending on service routine Memory key Save selection Service key Different functions depending on service routine Most however confirm a selection or the activation of the next test step T est rotation key Start a test R eturn key Move the unit to the starting position or confirm a save operation 61 25 665 D3437 D3437 076 01 15 02 06 2012 209 8 Service Sirona Dental Systems GmbH 8 3 Basic operating procedures in the service menu Service Manual GALILEOS Basic operating procedures in the service menu Activating the service menu Easypad The structure of the touchscreen user interface on the control panel is subdivided into 4 levels e Level 1 Main menu e Level 2 Program Settings e Level 3 Basic Settings e Level 4 Service menu When the unit starts up the main menu appears on the touchscreen 1 To select level 2 Se ect program settings touch the blue arrow in the top right corner of the touchscreen A 2 To select level 3 Se ect basic settings touch the left hand blue arrow in the top right corner of the touchscreen B 61 25 665 D3437 210 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 3 Basic operating procedures in the service menu 3 To select level 4 service menu access touch the wrench symbol C 4 Switch to the service menu Press and hold down the Service key D until
156. dge the error message self test which may occur spontaneously e Run an error diagnosis gt 88 and without any related operator action They may be caused by Further operation of the unit is possible system self tests 7 Unrecoverable system This error group includes all errors Run an error diagnosis gt 88 error which may occur spontaneously and without any related operator action They may be caused by system self tests In this case it is absolutely certain that continued system operation is not possible 61 25 665 D3437 86 D3437 076 01 15 02 06 2012 yy Location Identifier yy defines the location or logical function unit where the error has occurred yy Location Function unit Board 06 X ray tube assembly DX6 07 Easypad user interface GALILEOS Comfort DX7 71 Multipad user interface GALILEOS Compact DX71 10 System hardware DX11 DX1 11 System software DX11 DX1 12 CAN bus DX11 DX1 13 Stand peripherals DX11 DX1 14 Digital extension DX11 DX1 15 Configuration update wrong software wrong module constellation etc DX11 DX1 41 Media interface card DX41 42 Remote control DX42 89 X ray detector DX89 The location may be a DX module number standing for an entire HW function unit or a logical SW function unit on board DX11 central control General handling of error messages Error messages must always be acknowledged with the R key If tr
157. door contact e Start exposure again Fault has been removed e Acknowledge error message with R key e Close door contact e Repeat the exposure Is the help message displayed again 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 6 Troubleshooting Service Manual GALILEOS 6 8 Error analysis of X RAY control signal path Error messages without installed remote control E3 13 40 occurs after the unit is switched on Switch the unit off and then on again Ensure that release button A2 is not activated while Was the release button A2 activated while switching on the unit switching on Does the error message recur Short circuit in coiled cable of release button A2 e Replace release button Yes A2 No e Replace cable L108 Fault is removed 61 25 665 D3437 D3437 076 01 15 02 06 2012 149 6 Troubleshooting Sirona Dental Systems GmbH 6 8 Error analysis of X RAY control signal path Service Manual GALILEOS Error messages with and without installed remote control Error code Description Actions required see E3 13 40 Short circuit in signal path between Replace cable L117 or L108 S gt 362 board DX11 and release button A2 Replace board DX1 S gt 343 during power on e Replace board DX11 E6 13 41 Release signal missing on board e Replace cable L117 or L108 S gt 362 DX ral ster Oh exp oeure e R
158. e 3 Switch unit on 4 Repeat procedure and check function If the error occurs repeatedly e Repeat procedure and observe causal error messages Error code Description Actions required see E5 89 13 Error when writing to EEPROM e Acknowledge error and repeat procedure S gt 61 IMPORTANT Stored data may be e Run software update lost If the error occurs again S e Check log memory via extended details S gt 319 Check x ray detector replace if necessary Error code Description Actions required see E6 89 20 Faulty voltage supply of DX89 e Check cable L13 replace if necessary S gt 141 S gt 362 61 25 665 D3437 D3437 076 01 15 02 06 2012 121 Error code Description Actions required see E6 89 21 File system error e Check board DX839 replace if necessary S gt 134 S gt 343 Error code Description Actions required see E5 89 22 The power supply of the X ray e Check board DX839 replace if necessary S gt 134 a respond or is the S gt 343 Check x ray detector replace if S gt 319 necessary Error code Description Actions required see E5 89 23 Camera head in the X ray detector e Check cable L27 L28 replace if S gt 141 does not respond or wrong version necessary S gt 362 e Check board DX89 replace if necessary S gt 134 S gt 343 Check x ray
159. e Replace board DX41 e Please contact the Sirona Customer Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary Error code Description Actions required see E6 41 09 Board DX41 is installed in the unit e Perform device configuration via service S gt 343 but not configured routine S017 9 61 25 665 D3437 108 D3437 076 01 15 02 06 2012 Error code Description Actions required see E7 41 10 Module is stuck in bootloader stage Check operating status of board note S gt 134 LED states If the board remains in the bootloader stage S gt 61 e Repeat the software update S gt 343 e Replace board DX41 Error code Description Actions required see E7 41 12 Unit is not ready for operation This error is a sequential fault S gt 343 e Restart the unit Switch off the unit Wait 1 minute Switch unit on PONS Repeat procedure and check function Check unit configuration with or without DX41 by running service routine S017 9 correct the configuration if necessary If the error occurs repeatedly e Repeat procedure and observe causal error messages Error code Description Actions required see E3 41 20 Release signal applied during e Restart the unit power on 1 Switch off the unit 2 Wait 1 minute 3 Switch on the unit making sure that the release button is not pressed du
160. e Socket wrench 13 mm A F 17 mm A F 18 mm A F e Side cutting pliers e Spirit level e Digital Multimeter Accuracy Class 1 e Mult O Meter 512L e Soldering tool for repairing cables e Cable ties e Teflon tape e Loctite 61 25 665 D3437 D3437 076 01 15 02 06 2012 13 1 About this Service Manual Sirona Dental Systems GmbH 1 4 Structure of the document Service Manual GALILEOS Structure of the document Identification of danger levels To prevent personal injury and material damage please observe the warning and safety information provided in this document Such information is highlighted as follows DANGER An imminent danger that could result in serious bodily injury or death A possibly dangerous situation that could result in serious bodily injury or death A possibly dangerous situation that could result in slight bodily injury NOTICE A possibly harmful situation which could lead to damage of the product or an object in its environment IMPORTANT Application instructions and other important information Tip Information on making work easier Formats and symbols used The formats and symbols used in this document have the following meaning J Prerequisite Requests you to do something 1 First action step 2 Second action step or gt Alternative action Result See Formats and symbols Identifies a reference to another used gt 14 text passa
161. e ad lead cle once e in a E 325 9 9 1 Replacing the scan UMIby 2 sees ieecces scent eet eieeeet soe anacceere 325 9 9 1 1 Removing the defective scan unit eee eeeeeeeeeeeeeeeeeeeettteeeeeeeeeeenees 325 9 9 1 2 Attaching new SCAM unit sot cnadeviceiss ncaectes se edetnures i ad esetuearotgunethenoanes 327 9 9 1 3 What has to be done after replacing the scanner unit 332 9 9 2 Replacing the PoE MOdule ccccccceeeeeeeeeeeereeeeeeeeeeeennaeeeeeeeeeeenaee 333 9 9 2 1 Removing the faulty POE module eee cceeeeeeeeeeeeeeeeneeeeeeeeeeeeeeee 333 9 9 2 2 Installing the new PoE module uu eeeeeeeeeeeee eee eeeeeneeeeeeeeeeeeees 335 Head fixation device 22 2 ccccceceeececcceeeeeeeeenenceceeeeeeeeesseeeeeneceesensececaeenseeetees 338 9 10 1 Replacing receptacle element for head fixation for unit with head 338 fixation device Light barriers euietad ie icecetdacccids vauatedbaenevabeatal en tide bneeadaueadl texeelenatela de i ENEE Eseia 341 9 11 1 Replacing the light barriers eee ceeeeeeeeeeeeeeeeeeeeeeeeeeenaeeeeeeeeeeeeea 341 BO AROS sr doc cares doaet tet sashes a ir aa etree teh Seidl e a p pie Hes chic hia 342 9 12 1 Important notes about replacing bOardS cceeeeeeneeeeeeeeeeeeneee 342 9 12 2 Replacing boardS cia Oe eek i acne GA aioe 343 9 12 2 1TReplacing PC board DX1 arose cic dedet snctctionde oes braetdicoeteceey deine 344 9 12 2 2Replacing board DX11 sesssccces see
162. e diagram and attach the cover shielding on the X ray detector housing using the 2 clamps D and screws B Depending on the unit hardware version the brackets on your unit may differ slightly from those shown 7 Secure the cable with clamp E and screw C 8 Re attach the cover plate A 61 25 665 D3437 D3437 076 01 15 02 06 2012 323 9 Repair 9 8 X ray detector Attaching the covers Updating the ID label Sirona Dental Systems GmbH Service Manual GALILEOS Remove the plastic cap P in front of the input window of the new X ray detector Depending on the hardware version of your unit the plastic cap may differ slightly from the one shown in this diagram Use the two screws Q to attach the lower cover part to the X ray detector NOTICE The tab on the upper cover part must be pushed underneath the ring cover Then place the upper cover part on the lower one and screw it tight using the four screws R as well as a fifth screw S sirona TL Morne Ta ae gt Update the nameplate on the detector cover To do this affix the supplied label as shown in the figure IMPORTANT For Facescan units The FaceScan must be refitted gt 327 on units with Facescan fitted after the X ray detector has been replaced What has to be done after replacing the X ray detector 1 2 324 Perfo
163. e previous official main version of SIDEXIS XG must be installed before you can install the current version Perform a software update to version V03 06 01 or higher using automatic update gt 61 If multiple units are installed in a single network Set the IP address via SiXABCon Switch off the unit Wait for approx 1 minute Then switch the unit back on The error message E1 10 03 format flash file system is displayed The message Wo Key is displayed on the Multipad Press the R key to acknowledge the error message The formatting of the flash file system is started automatically Error message E1 10 04 is displayed during the entire process approx 5 6 min When the formatting is finished the error message is automatically acknowledged by the system and error message E6 11 07 undefined system class is displayed Press the R key to acknowledge the error message The access level for the service menu level 4 is automatically started Press and hold down the Service key until the LEDs above the patient symbol keys light up approx 2 s Then press the patient symbol keys in the sequence b d a within the next 4 seconds After you have entered the key combination correctly the service routine S017 test step 1 select confirm system class is started automatically The LED above the Memory key lights up Acknowledge any additional error messages with the R key Confirm the GALILEOS Compact system class
164. e table The selected patient symbol key lights up 61 25 665 D3437 D3437 076 01 15 02 06 2012 229 8 Service Sirona Dental Systems GmbH 8 6 S007 Error logging memory Service Manual GALILEOS A A 3 Use the arrow keys A in selection field 1 to select the required error event 66 in the example nae The corresponding error message is displayed in selection field wv A 2 gt 82 Selection field 3 displays the date and time of the error event lt 66 gt S7 4 Exit the service routine gt 220 E6 81 07 1 2004 12 14 12 31 09 Multipad 1 Call service routine 8007 1 gt 213 2 Use the arrow keys A in selection field 1 to select the required error event 66 in the example Use the patient symbol keys to set the increment for scrolling between the error numbers Patient symbol key 1 left increment 1 factory setting Patient symbol key 2 increment 10 Patient symbol key 3 increment 100 The LED above the selected patient symbol key is lit up 61 25 665 D3437 230 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 6 S007 Error logging memory 3 Use the arrow keys B to scroll and display the corresponding error code gt 82 the time and the date of the error event in selection field 4 Exit the service routine gt 220 61 25 665 D3437 D3437 076 01 15 02 06 2012 231 Sirona Dental Systems GmbH Service
165. e update must be performed gt 61 every time the language set changes in order to install the corresponding languages in the system Selection field Code Function 1 00 German English French Italian 01 German English French Dutch 02 German English Spanish Russian 03 German English Korean Japanese 04 German English Spanish Portuguese 05 German English Chinese PRC Chinese Taiwan Factory setting varies by order 1 Call service routine 8017 5 gt 213 Once the service routine has been selected the code for the current setting is displayed in selection field 1 2 Use the arrow keys A to select the code for the required language set in selection field 1 see table Once the language set has been selected the Memory key B lights up 3 Save the setting gt 219 4 Exit the service routine gt 220 61 25 665 D3437 258 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 11 S017 Configuration service S017 Test step 6 Enable disable the remote control Selection field Code Function 1 00 Remote control disabled 01 Remote control enabled Factory setting 1 Call service routine S017 6 gt 213 Once the service routine has been selected the code for the current setting is displayed in selection field 1 61 25 665 D3
166. ed on an electronic data carrier These are delivered together with the unit Connector pins or sockets bearing ESD warning labels must not be touched or interconnected without ESD protective measures See also Electrostatic discharge and Electromagnetic compatibility gt 15 Single use devices are identified with the symbol shown on the left They must be disposed of immediately after use Do not use single use devices more than once 19 3 Unit description Sirona Dental Systems GmbH 3 2 Hardware Service Manual GALILEOS Installation versions i RT ct The unit can be equipped with e a1 3mcoiled cable with release button inside the treatment room A Or e aremote control with or without coiled cable B C located outside the X ray room see also installation instructions 61 25 665 D3437 20 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 3 Unit description Service Manual GALILEOS 3 2 Hardware Modules and components Slide Stand Remote control gt 24 FAC
167. eeeeeeees 226 8 5 4 5005 Test Step 8 vcs eccvisestasvadseecbiid ea E e conten N A REEERE 227 61 25 665 D3437 D3437 076 01 15 02 06 2012 8 6 8 7 8 8 8 9 8 10 8 11 8 12 8 13 S007 Error logging Memory eich cstcuecetcesecue se acted setandeatinl vad gadedeed dase ateetee 229 8 0 1 S007 TEStStep Tanie a a a a a r aE 229 8 6 2 S007 Test step 2 oo cccsccsscccsenseceeaeeaaeeaeeeaeceaeeeeceeeceeeceeeeeeeeeeeeess 232 8 6 3 S007 Test Step Dirie e vs Ea Eae panata 235 8 6 3 1 Displaying the log with a web browser ccccccccccecceeeeeeeeeeeeees 236 S008 Update Service 0 00 0 ececeeccecseccececceecsaeeaaeeeceeeececeececeeeeeeeeeeseeeeeeeeeeeseess 238 8 7 1 SOO8 Test Step 2 ccccccsscecsecsaneeeaaeeaeeeaeceaeeeeceeeeeeeceeeeeeeeeeeeess 238 8 7 2 S008 Test step Savin sccedccccstances cobte eiee eE EE E EEEE EA 240 S009 Flash file system 0 0 0 0 ccc ceeeceeseeeeeeeneeeaeeeaeeaeeeaeecaeeceesaeceaeeeeeeeeeeeeeeess 241 8 8 1 9009 Test Ste iF kerrie neeaae E NaN E a EGEE OEE saueectenss 241 8 8 2 SOQ9 Test Step D rror ea a e a EE E R ATAT 243 8 8 3 S009 Test step 7 0 cee ccecseccsseccceeseeceaeeaaeeaeeeaeecaeceeceaeceeeceeeeeeeeeeeeess 245 S011 Dosimetry without ring movement sessssesseseeeeesesserrrrressrrrrrrrreseeeee 247 8 9 1 S011 Teststep 12 rerin e a e a a a 247 8 9 2 5011 Test step 9 ccccccscccsensanecaeeaaeeaeeeaeeeaeceeceeeceeeceeeeeeeeeeeeees 248 SOTZ CAN b s Service r ean
168. eesdeoetadsdeveecdade ale seek ineeeee 163 7 1 6 2 Geometry DIMAUON ais e2 red aactes eaten cas sateen eri ed oncaahaets asda event rere 164 7 1 6 3 Constancy test PhaMmtoMm se sccccse0 ac ocseiceeecatoucse Santen Gavleneegees 166 7 2 Checking the mechanical system adjustment e ee eeeeeeeeeeeeeeeeeneeeeeeeeees 166 7 3 Adjustment and calibration via the Service Functions meni 167 7 3 1 Diaphragm image sagivce haan vc scspucetevastae nage vad Aaecoigdsseeiddeageenteed 167 7 3 1 1 Type 1 Type 2 diaphragim cccccccccecseceeeccceeeeeeeneeeneeeeneeeeeeesnees 167 7 39 12 TYP GS diaphragm Spieaestiieo ie ieee aa aA eais 171 7 3 2 Checking the radiation field ccccceececcececneeeeeeeeneesnaceeeeeeeeteseees 178 TO DIOSIMGUY eck EET E E coed Meena aadudic TE 179 7 3 4 Sensor AC ASUNIG NC ecatecs tots steelers ati led eereb seal toh cuateh ies Ceci teu ck siaeN 181 7 39 hris ACUSUMOM sack iccasiesinigescdessereanadtulsealdnetetigecat tar aucannleeuiatseetawianeetoeds 182 7 3 6 Shading Calibration wai A oie icee eRe seh ote Rae eee 183 7 3 7 Distortion Call bration sieve scsx gece nc tte ieena iee eheti 185 7 3 8 Geometry calibration cccsccscdeictiecctesscneewuntasiiiaursiviawendenelvceetueneeniaetecs 187 TSD ONICO ii seai ina dadin eaan a SATE beatin ot dened A AN N Ean aN 189 7 4 Performing a white balance for FaceScan eeeeeeeeeeeeeeerrieeseerrrrirresssrreeees 190 7 5 Checking and adj
169. en in advance to determine whether the current software constellation is permissible If this is not the case the version number of the main software is labeled with an asterisk e g V03 03 01 In the event of software incompatibilities perform a software update or downgrade gt 61 Always perform the measures described below in the given sequence and do not carry out any other actions between the steps The following table provides an overview of various possible replacement situations and cross references to detailed descriptions of the actions required for the corresponding situations following board replacement Board Constellation Actions Page DX1 Inserting a new DX7 GALILEOS Comfort System software version V03 03 01 or higher e Switch the unit on S gt 155 e Perform a complete unit adjustment or calibration GALILEOS Compact System software version V03 06 01 or higher DX11 Replacing a DX11 Board DX11 with software version V02 62 01 or higher requires a SIDEXIS software version of V2 0 or higher in order to perform an update This requires an overall system update to software version V03 03 01 or higher or SIDEXIS V2 0 or higher The SW on the CD included with the DX11 must be installed on the entire system Inserting a new DX11 GALILEOS Comfort System software version Proceed as described in the chapter S gt 350 After changing the DX11 board gt
170. ent OK Adjustment not OK Permissible tolerance 30 pixels 5 pixels The brightness distribution along the border surrounding the image on all sides must be uniform A The distance between the bottom edge and the lowest point in the image should be 30 5 pixels measure with SIDEXIS scale B If the distance between the bottom edge and the lowest point in the image is out of tolerance E or the brightness distribution along the surrounding border is not uniform C the diaphragm must be adjusted mechanically gt 197 5 For GALILEOS Compact Evaluate the image 172 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS 7 Adjusting calibrating the unit 7 3 Adjustment and calibration via the Service Functions menu A B Adjustment OK C Adjustment not OK diaphragm not centered D Adjustment not OK diaphragm too small E Permissible tolerance 65 pixels 5 pixels 61 25 665 D3437 D3437 076 01 15 02 06 2012 173 7 Adjusting calibrating the unit Sirona Dental Systems GmbH 7 3 Adjustment and calibration via the Service Functions menu Service Manual GALILEOS 6 Collimator open Gk 7 Adjustment of the Upper jaw s diaphragm setting JV v 1 RE osc Sharo Citaten Geomar e l gt 2 3 4 The brightness distribution along the surrounding border must be uniform A A shadow in the image
171. ent must be set up and enabled using the SIDEXIS Manager For more information refer to SIDEXIS XG Installation Instructions Configuring Facescan Facescan is generally configured via the integrated web dialog on the FaceScan unit gt 75 There are however two basic options for configuring the Facescan unit e Type 1 Configuration using the Facescan USB stick gt 71 e Type 2 Configuration over a network cable peer to peer gt 75 Type 1 Configuration using the Facescan USB stick Connecting the USB stick 1 Pull gray cable L78 4 from slot X2 of the FACESCAN modular board 2 Unscrew the protective plate T from the Facescan unit The FACESCAN modular board is visible 3 Plug cable L78 4 again into slot X2 on the FACESCAN modular board 61 25 665 D3437 D3437 076 01 15 02 06 2012 71 4 General operating procedures Sirona Dental Systems GmbH 4 3 Configuring the unit Service Manual GALILEOS Starting the configuration 1 Plug the Facescan USB stick into a PC 2 Important Check the entry USB_STICK_MODE The value must be at Config 3 Using a text editor program edit the configuration file facescan_settings cfg on the Facescan USB stick and save this see section Syntax of the configuration file facescan_settings cfg gt 74 4 NOTICE Improper removal of the USB stick can lead to loss of data on the USB stick Remove the USB stick using the safely remove function operating system of
172. ent symbol key 2 B lit The default gateway is displayed in selection field 1 Patient symbol key 3 C lit The subnet mask is displayed in selection field 1 1 Call service routine S037 2 gt 213 Once the service routine has been selected the network data is displayed as in test step 1 Easypad The Memory key and the R key also become visible 192 168 15 240 The Memory key Easypad or the LED above the Memory key DEFAULT Multipad lights up mAs Y 61 25 665 D3437 D3437 076 01 15 02 06 2012 279 8 Service Sirona Dental Systems GmbH 8 13 S037 Network service Service Manual GALILEOS 2 Before restoring the factory settings check the network data that is still in the system A Show IP address B Show default gateway C Show subnet mask The patient symbol key selected in each case Easypad or the LED above the patient symbol key selected in each case Multipad lights up 3 To reset the network data press Memory key E R key Easypad or LED above R key Multipad lights up followed by R key F The default network data factory default setting is displayed To 192 168 15 240 switch between the display of the different network data proceed as in test step 1 4 Restart the unit 5 Exit the service routine gt 220 DEFAULT 61 25 665 D3437 280 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS
173. enter CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary Error code Description Actions required see E6 06 03 Invalid commanding of control data e Check the CAN bus S gt 128 CAN bus error e Please contact the Sirona Customer S gt 61 This error may also occur in Service Center CSC to find out connection with other causal error whether a bug fix by means of a messages Please also observe the software update is possible and causal error message It appears perform such an update if necessary only after you acknowledge the first error message Error code Description Actions required see E6 06 04 Data transfer error or dialog error to e Check the CAN bus S gt 128 module master side e Please contact the Sirona Customer S gt 61 Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary Error code Description Actions required see E6 06 05 Data transfer error or dialog error to Repeat the software update S gt 61 bootloader of module e Check the CAN bus S gt 128 Only occurs in connection with If the error occurs repeatedly or the module S 310 software update is no longer addressable e Replace the tube assembly 88 61 25 665 D3437 D3437 076 01 15 02 06 2012 Err
174. enu Service Manual GALILEOS 7 Press the release button Press and hold down the button until the exposure is complete the preview image is displayed in the exposure window and the acoustic signal indicating the end of the exposure double beep sounds if it has been configured IMPORTANT This process takes approx 2 3 minutes The unit transfers the acquired images to the Reconstruction and Control Unit RCU This process can take 2 3 minutes Once the transfer is complete the evaluation of the calibration is displayed in the message window If the information in the message window indicates that calibration was not successful check the calibration phantom to make sure that it is not damaged If you have repeated the procedure three times and still have not attained a positive result check the mechanical geometry of the unit gt 166 Adjust the unit if necessary and then repeat the calibration If this still does not lead to a positive result please contact the SIRONA Customer Service Center CSC Save values 8 If the calibration is OK or possible click the Save values button The calibration is saved Calibration of the geometry is now complete 9 Remove the geometric phantom from the bite block holder of the unit Calibration of the unit is now complete IA test image can be generated gt 189 via the Service tab card if necessary 61 25 665 D3437 188 D3437 076 01 15 02 06 2012 Sirona Dental S
175. eometry c4 gt State Dose measurement I Dose is within range of 1 2 2 3 mGy Image acquisition v The Service functions menu is called gt 161 Y The distortion phantom is clipped onto the cover of the X ray detector for protection against scratching 1 Click the Dose tab The corresponding tab card is selected 2 Attach the Mult O Meter sensor approximately in the middle of the distortion phantom mounted on the X ray detector 3 Click on the mage acquisition button SIDEXIS XG makes the unit ready for exposure gt 162 Service routine S011 8 is displayed on the control panel 4 Press the R key to move the unit back to the starting position 5 Press the release button Press and hold down the button until the exposure is complete the preview image is displayed in the exposure window and the acoustic signal indicating the end of the exposure double beep sounds if it has been configured 61 25 665 D3437 D3437 076 01 15 02 06 2012 179 7 Adjusting calibrating the unit Sirona Dental Systems GmbH 7 3 Adjustment and calibration via the Service Functions menu Service Manual GALILEOS 6 Then read off the dose from the Mult O Meter The value must be between 1 2 and 2 3 mGray If the value is outside the permissible range 1 2 to 2 3 mGray check the X ray tube assembly 7 To confirm that the dose is within the permissib
176. eplace board DX1 S gt 343 e Replace board DX11 E3 13 42 Short circuit in signal path between Replace cable L117 or L108 S gt 362 board DX11 and release button A2 e Replace board DX1 S gt 343 during operation of the unit e Replace board DX11 61 25 665 D3437 150 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 6 Troubleshooting Service Manual GALILEOS 6 9 Fault diagnosis of the X ray detector and on board DX89 Fault diagnosis of the X ray detector and on board DX89 Do not damage the image tube The image tube of the X ray detector is sensitive to mechanical stress and therefore must be handled with extreme care Avoid bumps and jolts Please consider this point during transport and installation For error messages in connection with board DX89 it is important to determine whether the fault concerned is attributable to a defect on board DX89 or to a defect in the X ray detector To do this proceed as follows 1 Move the unit down using the Up Down keys 2 Switch off the unit 3 Remove the x ray detector cover 4 CAUTION Risk of injury The cover plate has sharp edges Carefully pull the cover plate upwards to remove it from the X ray detector see also Replace X ray detector gt 319 5 WARNING Potentially lethal shock hazard Do not touch any live parts while observing board DX89 Remove the cover plate of board DX89 Switch the unit on again and observe bo
177. eplace board DX11 146 61 25 665 D3437 D3437 076 01 15 02 06 2012 Error analysis of X RAY control signal path from unit serial number 3201 without board DX41 Error and help messages with remote control installed E3 42 31 E3 13 40 occur in combination after the unit is switched on with the door contact closed Switch the unit off and then on again Ensure that release button A2 is not activated while Was the release button A2 activated while switching on the unit switching on Do the error messages recur Short circuit in coiled cable of release button A2 or on membrane keyboard of remote control Yes e Replace coiled cable e Replace display board DX42 of remote control No Fault is removed E3 42 31 occurs once after the unit is switched on Hardware fault on display board DX42 or short circuit in coiled cable of release button A2 or on membrane keyboard of remote control e Replace release button A2 e Replace remote control 61 25 665 D3437 D3437 076 01 15 02 06 2012 147 E6 13 43 occurs once during operation of the unit Was the door contact opened during the exposure Yes e Check door contact repair if necessary e Replace cable L117 e Replace remote control H321 is triggered at start of exposure Is the door contact open e Check door contact repair if necessary e Replace cable L117 148 e Close
178. er e Replace the tube assembly S gt 310 exposure does not match the expected value 61 25 665 D3437 93 D3437 076 01 15 02 06 2012 Location 07 Easypad DX7 Error code Description Actions required see E6 07 01 General error during module e Please contact the Sirona Customer S gt 61 initialization Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary If the error occurs repeatedly S gt 307 e Replace user interface with electronics DX7 Error code Description Actions required see E6 07 02 Invalid system data or uninitialized Please contact the Sirona Customer S gt 61 module storage data Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary e Acknowledge error and repeat procedure If the error occurs repeatedly S gt 307 e Replace user interface with electronics DX7 Error code Description Actions required see E6 07 03 Invalid commanding or control data Check the CAN bus S gt 128 This error may also occur in e Please contact the Sirona Customer S 61 connection with other causal error Service Center CSC to find out messages Please also observe the whether a bug fix by means of a causal error message It appears software update is possible and only afte
179. ervice Manual GALILEOS 8 10 S012 CAN bus service Clearing the counter for the module 1 To delete the counter press the T key B The counter is then reset to 0 2 Exit the service routine gt 220 61 25 665 D3437 D3437 076 01 15 02 06 2012 251 017 Configuration service SR SA Function S017 Unit configuration 017 2 Yes Confirming the unit version 017 3 Yes Enter the country group code 017 4 Yes Select a language for GALILEOS Comfort 017 5 Yes Select a language set for GALILEOS Comfort 017 6 Yes Enable disable the remote control 017 7 Yes Configure the switching plate for the swivel arm 017 9 Yes Enable disable operation with board DX41 017 13 Yes Enable disable the welcome screen only for GALILEOS Comfort 017 14 Yes Enable disable certain lines of the welcome screen for GALILEOS Comfort 017 15 Yes Activate deactivate the acoustic signal for end of exposure 017 25 Yes Select the diaphragm type 252 SR service routine SHZ security access S017 Test step 2 Configuring the hardware version Selection field Code Function 1 0100 CB 0500 CB incl Facescan Factory setting 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 11 S017 Configuration service 1 Call service routine S017 2 gt 220 Once the serv
180. eset the entire calibration of the unit and S gt 155 readjust the unit e Replace board DX11 S gt 350 Error code Description Actions required see E6 11 02 Watchdog error e Acknowledge error and repeat procedure If the error occurs repeatedly S gt 350 e Replace board DX11 Error code Description Actions required see E6 11 03 Operating system resource error e Acknowledge error and repeat procedure If the error occurs repeatedly S gt 350 e Replace board DX11 Error code Description Actions required see E7 11 04 Implausible data in EEPROM e Check the unit configuration via service S gt 252 routines S017 and S018 and reconfigure if S gt 272 necessary e Check calibration with diaphragm test S exposures If the calibration is not OK S gt 155 e Recalibrate the unit If the calibration is OK e Make the individual unit settings again e g programming of the patient symbol keys see Operating Instructions Error code Description Actions required see E6 11 05 RAM allocation failed e Replace board DX11 S gt 350 Error code Description Actions required see E7 11 07 Unknown or invalid definition of unit e Take the action required after replacing a S gt 347 class Occurs during first power on after replacement of board DX6 or DX11 board 61 25 665 D3437 D3437 076 01 15 02 06 2012 101
181. far as the contact and do not apply force NOTICE Risk of damage to boards Please observe the usual precautionary measures for handling printed circuit boards ESD Touch a ground point to discharge static electricity before touching any boards IMPORTANT After replacing boards or modules containing boards check to make sure that the software version of the module corresponds to the current software status of the system The software versions of the modules can be queried by running service routine S008 2 or using the extended detail query in SiXABCon You can also check the info screen in advance to determine whether the current software constellation is permissible If this is not the case the version number of the entire software is labeled with an asterisk e g V03 03 01 61 25 665 D3437 288 D3437 076 01 15 02 06 2012 9 Repair Sirona Dental Systems GmbH 9 1 Safety checks Service Manual GALILEOS IMPORTANT When replacing modules be sure to note which ones contain boards and follow the instructions in the section titled Measures following replacement of boards Also check whether the current software CD or the SIRONA dealer page contains any additional more up to date information about module replacement Be sure to follow the instructions about how to proceed following module replacement You will find this information at the end of each set of repair instructions Safety checks
182. fort Welcome screen enable disable lines 017 14 S gt 266 for GALILEOS Comfort 61 25 665 D3437 D3437 076 01 15 02 06 2012 205 8 Service Sirona Dental Systems GmbH 8 2 Service menu and service routines Service Manual GALILEOS Service menu and service routines You can use the service routines to check the function of certain unit components and modules as well as to set important unit parameters This chapter describes all service routines that can be selected and started via the service menu on the control panel Service routines S010 and S030 cannot be selected manually and therefore are not described here They are used solely for unit adjustment and calibration Displays and symbols in the service menu Easypad There are many different control symbols and display fields on the touchscreen these are activated on a context sensitive basis depending on the procedure step ATW lt l gt 27 H O A O H gt 8333 e D000 206 115 02 062012 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 61 25 665 D3437 D3437 076 01 15 02 06 2012 8 2 Service menu and service routines A X RAY Radiation can be released X RAY Active Caution Radiation is being released B Selection field 1 Display fields for service routines test C Selection field 2 steps values
183. ftware V1 5 or above For GALILEOS Compact The main software V03 06 01 is run compatible from GALILEOS Software V1 6 or above 61 25 665 D3437 D3437 076 01 15 02 06 2012 Main software V03 07 00 GALILEOS Remote control Board Software Board Software DX6 02 93 00 DX42 02 48 00 DX7 02 63 00 DX7 LO 02 26 00 DX7 L1 02 26 00 DX7 L2 02 26 00 DX7 L3 02 26 00 DX7 L4 02 03 00 DX7 L5 02 01 00 DX71 02 40 00 DX11 02 71 00 DX41 02 30 00 DX89 01 18 00 DX89 FPGA 01 15 00 SIDEXIS XG GALILEOS Software RCU GALILEOS Software GALILEOS Server software Compatibility Update Implant V2 5 1 or higher V1 7 x V1 7 x V1 7 4 or higher V1 7 x V1 8 V2 0 V2 1 V1 8 Masowa v 09 0702 GALILEOS Remote control Board Software Board Software DX6 02 94 00 DX42 02 48 00 DX7 02 63 00 DX7 L0O 02 26 00 DX7 L1 02 26 00 DX7 L2 02 26 00 DX7 L3 02 26 00 DX7 L4 02 03 00 DX7 L5 02 01 00 DX71 02 40 00 DX11 02 73 00 DX41 02 30 00 DX89 01 19 00 DX89 FPGA 01 15 00 SIDEXIS XG GALILEOS Software RCU GALILEOS Software GALILEOS Server software Compatibility Update Implant V2 5 1 or higher V1 7 x V1 7 x V1 7 4 or higher V1 7 x V1 8 V2 0 V2 1 V1 8 61 25 665 D3437 D3437 076 01 15 02 06 2012 53 Main software V04 04 00
184. g DANGER Potentially lethal shock hazard It is essential to switch off the unit and wait at least 1 minute before removing a cover Switch OFF the X ray unit before connecting a measuring instrument Perform continuity tests only on units which are switched off NOTICE Risk of damage to unit Select the correct current voltage type and adjust the measuring range to match the expected readings Keep to the prescribed cool off periods if several exposures have to be taken to check a measured value Please observe the usual precautionary measures for handling printed circuit boards ESD Touch a ground point to discharge static electricity before touching any boards CAN bus cable When unplugging CAN bus cables it is essential to unplug the power supply as well Error logging memory The error logging memory is part of the extended details Timestamp Categorie Message 2006 03 06 19 57 40 Message Logbook started 2006 03 06 20 13 02 Message Recording started Value 9000 2006 03 06 20 13 22 Message Recording stopped 2006 03 06 20 48 34 Message Recording started Value 9000 2006 03 06 20 48 54 Message Recording stopped 2006 03 07 15 45 38 Error Sidexis E5 14 04 ERR_SOCKET SidErr ERR_SOCKET_ERROR SockErr EPIPE 2006 03 07 08 57 05 Message Logbook started 2006 03 07 08 58 30 Message Recording started Value 104 2006 03 07 08 58 49 Message Recording stopped
185. ge and specifies its page number e List Identifies a list Command menu item Identifies commands menu items or quotations 61 25 665 D3437 14 D3437 076 01 15 02 06 2012 61 25 665 D3437 D3437 076 01 15 02 06 2012 Safety instructions Modifications to the unit Modifications to this unit which might affect the safety of the system owner patients or other persons are prohibited by law For reasons of product safety this product may be operated only with original Sirona accessories or third party accessories expressly approved by Sirona The user is responsible for any damage resulting from the use of non approved accessories Fixed connection Potentially lethal shock hazard Fixed connection Installing a mains plug instead of the specified fixed connection infringes international medical regulatory actions and is prohibited In case of error this puts patients users and other parties seriously at risk Electromagnetic compatibility The unit complies with the requirements of standard IEC 60601 1 2 Medical electrical equipment is subject to special EMC preventive measures It must be installed and operated as specified in the document Installation Requirements If high voltage systems radio link systems or MRI systems are located within 5 m of the unit please observe the specifications stated in the installation requirements Portable and mobile RF communications equipment may interfere wi
186. guration with the Configure Network button The Facescan restarts with the modified settings 75 4 General operating procedures 4 3 Configuring the unit sirona Facescan Service Functions pines Switch Leds On Switch Leds Of Get Device State ids Weis R Show Service Log Show Appiication Log Download Application Context Sirona Dental Systems GmbH Service Manual GALILEOS Resetting the Facescan configuration to factory default settings Opening the web dialog 1 In SIDEXIS Manager start the Facescan Configuration application 2 Click on the Facescanner in browser button A password dialog box opens 3 Inthe field User enter service 4 Inthe field Password enter Sirona The Facescan Device Service web dialog opens Selecting the service dialog gt Inthe menu bar select the menu item SERVICE The Facescan Service Functions window opens Reset 1 Press the Settings Reset button A reset dialog opens Complete loss of user data When the unit is reset to factory settings all user data is overwritten including white balance and calibration data The network settings are however retained 2 Press the Settings Reset button The configuration is reset to the factory settings 61 25 665 D3437 76 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 4 General operating procedures Service Manual GALILEOS 4 4 Reading unit data AA Readi
187. he Facescan restarts The white balance is complete 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 7 Adjusting calibrating the unit Service Manual GALILEOS 7 5 Checking and adjusting the touchscreen Checking and adjusting the touchscreen 1 Switch the unit on Once the unit software has started up the start screen is displayed on the touchscreen 2 As soon as the start screen appears press the 3 keys light localizer T and R simultaneously see A IMPORTANT These keys must be pressed while the unit is booting and the start screen is still displayed The first adjustment screen appears on the display You can abort the adjustment procedure at any time by pressing the Abort button 3 Touch the top left corner of the screen The second adjustment screen appears on the display 4 Touch the bottom right corner of the screen The third adjustment screen appears on the display 5 Touch the center of the green square in the top right corner of the screen IMPORTANT Touch the green squares as close to the center as possible so that the black dots that then appear will be positioned exactly in the center of the squares The fourth adjustment screen appears on the display 61 25 665 D3437 D3437 076 01 15 02 06 2012 191 7 Adjusting calibrating the unit Sirona Dental Systems GmbH 7 5 Checking and adjusting the touchscreen Service Manual GALILEOS
188. he arrow keys B to select the code for the required state of the line selected in selection field 1 in selection field 2 see table Once the required setting has been selected the Memory key C lights up 4 Save the setting gt 219 5 Exit the service routine gt 220 61 25 665 D3437 D3437 076 01 15 02 06 2012 267 8 Service Sirona Dental Systems GmbH 8 11 S017 Configuration service Service Manual GALILEOS S017 Test step 15 Selection field Code Function 1 00 Acoustic signal indicating the end of the exposure is disabled 01 Acoustic signal indicating the end of the exposure is enabled Factory setting Activate deactivate the acoustic signal for end of exposure 1 Call service routine 8017 15 gt 213 Once the service routine has been selected the code for the current setting is displayed in selection field 1 61 25 665 D3437 268 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 11 S017 Configuration service 2 Use arrow keys A to select the required setting in selection field 1 see table Once the required setting has been selected Memory key B Easypad or the LED above Memory key B Multipad lights up 3 Save the setting gt 219 4 Exit the service routine gt 220 61 25 665 D3437 D3437 076 01 15 02 06 2012 269 8 Service Sirona Dental Systems GmbH 8
189. he control panel sirona T MTT Serial No 12345 i Model No 600726 aris al No 12348 61 25 665 D3437 D3437 076 01 15 02 06 2012 307 9 Repair Sirona Dental Systems GmbH 9 5 Control panel Service Manual GALILEOS 5 Update the nameplate at the control panel cover To do so affix the supplied label as shown in the figure 6 For GALILEOS Comfort Easypad only Cable L10 green cable must be equipped with ferrite core B unless this has already been done What has to be done after replacing the user interface IMPORTANT So that the board is also replaced with the user interface you MUST also follow the instructions in the chapter titled Measures following replacement of boards gt 347 1 Check that the user interface and the display elements are functioning correctly When the unit is switched on all of the display elements must light up briefly 2 Perform a software update to the latest version gt 61 Easypad only Following replacement of the user interface the language set on board DX7 is set to the factory setting by default 00 German English French Italian If the configured unit language set which can be queried by running service routine S017 5 or using the extended detail query in SiXABCon has a configuration other than 00 this configuration will be copied to board DX7 by the update function 61 25 665
190. her After a new DX11 is inserted the IP address is initially reset to the factory setting Before you set the unit to a new IP address make sure that the IP address you re assigning has not been assigned to any other unit 1 2 10 11 12 13 14 Switch the unit on Do not acknowledge any error messages at this time Install the current version of the SIDEXIS software V2 0 or higher If the current version of SIDEXIS is a patch version the previous official main version of SIDEXIS XG must be installed before you can install the current version Perform a software update to version V03 03 01 or higher using automatic update gt 61 If multiple units are installed in a single network Set the IP address via SiXABCon Switch off the unit Wait for approx 1 minute Then switch the unit back on The error message E1 10 03 format flash file system is displayed The message Wo Key is displayed on the Easypad Press the R key to acknowledge the error message The formatting of the flash file system is started automatically Error message E1 10 04 is displayed during the entire process approx 5 6 min When the formatting is finished the error message is automatically acknowledged by the system and error message E6 11 07 undefined system class is displayed Press the R key to acknowledge the error message The access level for the service menu level 4 is automatically started Press an
191. i sensor after the unit is switched on Sensor being prepared message goes During this period the message out Sensor being prepared orato S If this error is displayed without attainment of displayed During this time the unit l exposure readiness is not ready for operation If exposure readiness is reached e Check cable L13 between board DX11 during this time error message E1 and board DX839 replace if necessary 10 07 appears e Check cable L28 between the camera head and board DX89 in the X ray detector replace if necessary e Check cable L27 in the X ray detector replace if necessary e Replace board DX89 e Replace board DX1 e Replace the X ray detector e Replace board DX11 As of X ray detector serial number 5000 cable L28 can no longer be replaced individually Error code Description Actions required see E1 10 20 Board DX11 does not have valid e Perform service routine S009 7 copy data S gt 245 data about the X ray detector from DX89 to board DX11 S gt 155 e Perform service routine S009 7 copy data E110 21 Board DX11 does not have valid from DX89 to board DX11 Ral odours ano e Perform service routine S009 7 copy data from DX89 to board DX11 E1 10 22 X ray detector was replaced and e Calibrate the unit must be registered in the system 61 25 665 D3437 99 D3437 076 01 15 02 06 2012 5 Messages 5 5 List of error messages Sirona Dental Systems GmbH Service Manual GALILE
192. ialog 1 In SIDEXIS Manager start the Facescan Configuration application 2 Click on the Facescanner in browser button A password dialog box opens 3 Inthe field User enter service 4 Inthe field Password enter sirona The Facescan Device Service web dialog opens sirona Facescan Service Functions Enter Auto White Bau Show Service Log Show Application Log segs Rene Reboot Device j Selecting the service dialog gt In the menu bar select the menu item SERVICE The Facescan Service Functions window opens Reading unit settings 1 78 Press the Get Device State button A dialog box to read the unit settings opens Press the Press Here to Download button Save the archived unit settings to the hard disk 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 4 General operating procedures Service Manual GALILEOS 61 25 665 D3437 D3437 076 01 15 02 06 2012 4 5 Using demo mode operation without radiation release Using demo mode operation without radiation release For demo use the X ray detector dummy for GALILEOS Order No 61 19 007 should be used instead of the actual X ray detector For further information please refer to the instructions included with the dummy If the volume tomograph is to be presented as a demo unit at trade fairs or exhibitions it must be ensured that
193. iation time has elapsed radiation is terminated prematurely and the exposure is interrupted The actual radiation time is notdisplayed When you release radiation during the cool down interval a countdown of the remaining waiting time is displayed in the header of the control panel or in selection field 2 on the Multipad automatic exposure blocking 4 Exit the service routine gt 220 61 25 665 D3437 222 D3437 076 01 15 02 06 2012 S005 General X ray tube assembly service SR SA Function S005 General X ray tube assembly service 005 1 No If tube type is valid Read out tube type Yes f tube type is not valid Select tube type 005 4 No Fan test 005 5 No Temperature sensor test single tank 005 8 Yes Automatic adjustment of pulse preheating 61 25 665 D3437 D3437 076 01 15 02 06 2012 SR service routine SHZ security access S005 Test step 1 Read or select the X ray tube assembly type IMPORTANT Only CB tube D151 R indicator number 03 is permissible for use with the GALILEOS volume tomography unit gt Call service routine S005 1 gt 213 Selection field 1 shows the indicator number of the installed tube type IMPORTANT The X ray tube assembly automatically queries the tube type information If no defined value is saved the security access will be shown instead of the tube type 223 8 Service Sirona Dental Systems GmbH
194. ible with the main software version of the unit IMPORTANT The cover plates of the DX1 REF 59 24 142 and REF 62 82 052 board versions are not compatible with each other In the case of boards with REF 59 24 142 DX1 and REF 59 25 214 DX11 The cover plate of board DX11 must be removed gt 344 1 CAUTION Touch a ground point to discharge static electricity before touching any boards Pull the defective DX11 board to remove it from the DX1 board 2 Insert the new DX11 board on the DX1 and reattach the cover plate 61 25 665 D3437 344 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 12 Boards In the case of boards with REF 62 82 052 DX1 and REF 63 17 056 DX11 The cover plate of board DX11 must be removed gt 344 1 CAUTION Touch a ground point to discharge static electricity before touching any boards Pull the defective DX11 board using the removal tool D to remove it from the DX1 board IMPORTANT The removal tool D is included in the delivery scope of the DX11 board 2 NOTICE Ensure that the connector strips of boards DX1 and DX11 are aligned precisely above one another and are not offset before pressing the boards together firmly Attach the new DX11 board on to the DX1 3 Check that the DX11 board is correctly attached to the DX1 board There must be no visible gap between the connector strips 4 Reattach the cover
195. ice mode the control options that are available on the control panel are determined by SIDEXIS XG and the service routine currently selected General control of the unit by means of the control panel as in the user mode is not possible in this mode Easypad Service mode is displayed on the Easypad via the PC service image SERVICE is displayed on the Multipad to indicate that the service mode is active 61 25 665 D3437 D3437 076 01 15 02 06 2012 161 7 Adjusting calibrating the unit Sirona Dental Systems GmbH 7 1 General information about unit adjustment and calibration Service Manual GALILEOS Enabling exposure readiness To take an exposure in SIDEXIS XG the system must first be made ready for exposure J Call the Service functions menu V Select the corresponding tab Image acquisition gt Click on the mage acquisition button The exposure window opens in SIDEXIS XG It indicates the current status of exposure readiness The service routine used for the corresponding exposure is displayed on the control panel along with the specific exposure parameters Taking an exposure J Callthe Service functions menu V Select the corresponding tab SIDEXIS XG must be ready for exposure 1 Press the R key to move the unit into the starting position The procedure is complete as soon as the Ready for exposure status message is displayed on the touchscreen Easypad only and the
196. ice routine has been selected the code for the current setting is displayed in selection field 1 Memory key A Easypad or the LED above Memory key A Multipad lights up 2 Use the arrow keys A to select the code for the required hardware version in selection field 1 see table Once the hardware version has been selected Memory key B Easypad or the LED above Memory key B Multipad lights up 3 Save the setting gt 219 4 Exit the service routine gt 220 61 25 665 D3437 D3437 076 01 15 02 06 2012 253 8 Service Sirona Dental Systems GmbH 8 11 S017 Configuration service Service Manual GALILEOS S017 Test step 3 Enter the country group code Selection field Code Function 1 00 Worldwide 01 Asia 02 US Factory setting 1 Call service routine 8017 3 gt 213 Once the service routine has been selected the code for the current setting is displayed in selection field 1 61 25 665 D3437 254 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 11 S017 Configuration service 2 Use arrow keys A to select the required country group code in selection field 1 see table Once the country group code has been selected Memory key B Easypad or the LED above Memory key B Multipad lights up 3 Save the setting gt 219 4 Exit the service routine gt 220
197. ield Parameter Display Meaning 1 IP address default gateway or subnet mask of the unit 2 default Fixed address static Fixed address modified setting dynamic Automatic address assignment Factory setting 2117 8 Service Sirona Dental Systems GmbH 8 13 S037 Network service Service Manual GALILEOS 1 Call service routine 8037 1 gt 213 Once the service routine has been selected the IP address of the unit is displayed in selection field 1 default static or dynamic is displayed in selection field 2 see table 192 168 15 240 DEFAULT 2 You can display various items of network data in selection field 1 by pressing the patient symbol keys A B or C see table The patient symbol key selected in each case Easypad or the LED above the patient symbol key selected in each case Multipad lights up 3 Exit the service routine gt 220 61 25 665 D3437 278 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 13 S037 Network service S037 Test step 2 Setting the default IP address default gateway address and default subnet mask IMPORTANT The network address can only be restored to the factory setting default value in fixed address boot mode STATIC or no DHCP Symbol on the control Status Meaning panel Patient symbol key 1 A llit The IP address is displayed in selection field 1 Pati
198. ightly approx 1 turn 2 Adjust the diaphragm position using screws B horizontal adjustment and C vertical adjustment CW rotation of screws Moves the diaphragm to the right or upwards CCW rotation of screws Moves the diaphragm to the left or downwards 3 Retighten screws A firmly 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 7 Adjusting calibrating the unit Service Manual GALILEOS 7 6 Mechanical adjustments Adjusting the diaphragm size In order to adjust the size of the diaphragm opening adjust the diaphragm distance 1 Loosen screw D slightly approx 1 turn 2 Adjust the diaphragm distance with knurled nut E Turn the knurled nut toward the rear The image on the X ray detector becomes arger Turn the knurled nut toward the front The image on the X ray detector becomes smaller 3 Retighten screw D firmly 4 Repeat the diaphragm exposure gt 167 61 25 665 D3437 D3437 076 01 15 02 06 2012 199 7 Adjusting calibrating the unit 7 6 Mechanical adjustments Sirona Dental Systems GmbH Service Manual GALILEOS Adjusting the complete diaphragm unit Adjusting the Type 3 diaphragm If the exposure taken in the diaphragm open diaphragm setting is not OK the complete diaphragm unit must be adjusted 1 Set the rotary knob on the tube assembly to the open diaphragm position 3 If necessary use
199. iguration 61 25 665 D3437 D3437 076 01 15 02 06 2012 4 General operating procedures 4 3 Configuring the unit Type 2 Configuration over a network cable peer to peer Connecting the Facescan unit to a PC J APC with an installed web browser must be available v The factory setting of the IP address of the Facescan is 192 168 16 240 J ThePC employed must be in the 192 168 16 xx network otherwise no network connection can be achieved gt Using a network cable connect the PC directly to the GALILEOS media converter with installed Facescan Opening the web dialog 1 Switch GALILEOS on 2 Open up a web browser on the PC 3 Enter the Facescan IP address into the web browser http lt IP Adresse gt A password dialog box opens 4 Inthe field User enter service 5 Inthe field Password enter sirona The Facescan Device Service web dialog opens Selecting the configuration dialog gt Inthe menu bar select the menu item CONFIGURATION The Facescan Configuration window opens Starting the configuration 1 When selecting with DHCP Set the DHCP State field to On or gt When selecting without DHCP Set the DHCP Siate field to Off 2 Inthe PAdresse field enter the desired IP address factory setting 192 168 16 240 3 Inthe Netmask field enter the desired subnet mask factory setting 255 255 255 0 Completing the configuration gt Confirm the confi
200. indicates horizontal or vertical displacement of the diaphragm The distance between the bottom edge and the lowest point in the image should be 65 5 pixels measure with SIDEXIS scale No surrounding gray shadow should be visible D A surrounding gray shadow in the image indicates that the diaphragm is too small If the distance between the bottom edge and the lowest point in the image is out of tolerance E or the brightness distribution along the surrounding border is not uniform C or a surrounding gray shadow is visible in the image diaphragm opening too small D the diaphragm must be adjusted mechanically gt 197 If the exposure is OK A B confirm this by clicking the check box located to the right of the Co limator open button The box will appear checked The adjustment for the open diaphragm diaphragm setting is now complete The Upper jaw button is selectable Continue the calibration procedure with the adjustment of the upper jaw diaphragm setting The Service functions menu is called gt 161 The Diaphragm tab is selected The Upper jaw is selectable in the mage acquisition menu area Set the rotary knob on the tube assembly to the maxillary exposure position Click on the Upper jaw button SIDEXIS XG makes the unit ready for exposure gt 162 Service routine S030 5 is displayed on the control panel Take an exposure 85kV 0 21 mAs gt 162 Eva
201. individual unit components and results in increased power consumption gt After switching the unit off wait for approx 60 seconds before switching it on again 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 4 General operating procedures Service Manual GALILEOS 4 1 Switching the unit on Switching the GALILEOS Comfort on NOTICE The surface of the touchscreen is sensitive The touchscreen can be damaged or its surface scratched gt Never use pointed objects such as ballpoint pens pencils etc to operate the touchscreen gt Only use your fingertips to operate the touchscreen IMPORTANT After the unit is switched on the touchscreen has only limited readability for several minutes until the background lighting has completed its warm up phase After the unit is switched off with the main switch the touchscreen remains lit for approx another 3 to 5 seconds 1 Turn the main switch A to position I 2 Wait for approx 1 minute The X ray radiation indicator B lights up for approx 1 second as a functional check After approx 2 seconds the green LED C in the upper part of the control panel lights up This LED remains lit as long as the unit is on The start screen is displayed on the touchscreen for several seconds The program selection is then displayed on the touchscreen 3 Check whether the patient symbols on the touch
202. iption Service Manual GALILEOS 3 2 Hardware PoE power supply board PoE 61 25 665 D3437 D3437 076 01 15 02 06 2012 41 3 Unit description Sirona Dental Systems GmbH 3 2 Hardware Service Manual GALILEOS Covers When removing covers always remember that direct sunlight or bright room lighting can cause system malfunctions due to activated light barriers Therefore avoid direct sunlight and bright room lighting above the unit Reattach all covers When attaching the covers be sure to screw the sheet metal cover back on IMPORTANT For reasons of electromagnetic compatibility be sure to fasten all screws 61 25 665 D3437 42 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 3 Unit description Service Manual GALILEOS 3 2 Hardware GALILEOS 61 25 665 D3437 D3437 076 01 15 02 06 2012 43 3 Unit description 3 2 Hardware 44 Sirona Dental Systems GmbH Service Manual GALILEOS Profile covers top and bottom Intermediate piece Tube assembly cover front Tube assembly cover rear Cover for ring center in units without head fixation device nim olol gt Acquisition unit for head fixation device in units with head fixation device 0 Ring cover I Support cover Swivel arm cover Arm cover Slide cover front Slide cover top rear Slide cover center rear Slide cover bottom rear Ozgan Alo X ray detector
203. l GALILEOS Replacing boards Risk of damage to boards Please observe the usual precautionary measures for handling printed circuit boards ESD Touch a ground point to discharge static electricity before touching any boards Board DX41 is omitted in units with serial numbers of 3201 and higher 61 25 665 D3437 D3437 076 01 15 02 06 2012 343 Replacing PC board DX1 IMPORTANT The software version of the DX1 DX11 board must be compatible with the main software version of the unit 1 Remove the arm cover 2 Disassemble both cross braces A 3 CAUTION Touch a ground point to discharge static electricity before touching any boards Remove the cover plate B of the board DX11 4 Pullall cables off of board DX1 5 Disassemble and remove the defective board DX1 6 NOTICE You must observe the notes in the chapter titled Replacing board DX11 gt 344 Install the DX11 board from the defective DX1 on the new DX1 7 Reinstall the DX1 board in the unit and reattach the connectors 8 Reassemble both cross braces A 9 Reattach the covers NOTICE Once you have removed the cross braces A the unit must be completely readjusted or recalibrated After replacing the board DX1 you must observe the notes provided in the chapter Replacing board DX11 IMPORTANT The software version of the DX1 DX11 board must be compat
204. l Systems GmbH Service Manual GALILEOS 7 Adjusting calibrating the unit 7 3 Adjustment and calibration via the Service Functions menu em parera The expos mas uen actor 0 922 ep ma aceta The shading exposure and the evaluation of the shading calibration is displayed If the information in the message window indicates that calibration is not OK and or not possible keep repeating the procedure starting with Step b until calibration is OK and or possible If you have repeated the procedure three times and still have not attained a positive result please contact the SIRONA Customer Service Center CSC IMPORTANT No foreign bodies may be visible on the shading exposure If this is the case check the beam path for foreign bodies remove them if necessary and repeat the calibration 5 Ifthe calibration is OK or possible click the Save values button The calibration is saved Diacteage Radaton fk Doo Sansor iie Shodno Datosion Geometny 4 f The shading calibration is now complete 6 Continue the calibration procedure with the distortion calibration gt 185 61 25 665 D3437 184 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 7 Adjusting calibrating the unit Service Manual GALILEOS 7 3 Adjustment and calibration via the Service Functions menu Distortion calibration IMPORTANT When a new distortion calibration is save
205. le range between 1 2 and 2 3 mGray click the check box underneath the message window The box will appear checked Dove ncaa eka pan feaefac ecaet mune sacel cis Standing Ditoton Goomety sel The dosimetry is now complete 8 Remove the sensor from the distortion phantom and take the phantom off the X ray detector 9 Continue the calibration procedure with the sensor adjustment gt 181 61 25 665 D3437 180 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 7 Adjusting calibrating the unit Service Manual GALILEOS 7 3 Adjustment and calibration via the Service Functions menu 7 3 4 Sensor adjustment ice Diaphragm Radiation field Dose Sensor ris Shading Distortion Geometry 4 gt or p Jiaphragm Sensor calibration n field There already is valid sensor calibration data on this machine Image acquisition Save values v The Service functions menu is called gt 161 1 Click the Sensor tab The corresponding tab card is selected 2 Click on the mage acquisition button SIDEXIS XG makes the unit ready for exposure gt 162 T manaon fesaontaa Dove T seewm ia a mans socnd nee Thee seat wad commer calteahon data on bn mache Service routine S010 8 is displayed on the control panel 3 Press the R key to move the unit back to the starting position 4 Press the release button Press and hold down the butto
206. leases 65 4 General operating procedures 4 2 Updating the firmware DX42 wa DX41 DX61 DX07 DX07_L5 wa DX71 wa 0X91 DX81P DX81P_F1 wa DX81C wa DX81C_ F1 DX88 DX88 F1 wa DX89 wa DX89 Fi DX06 DXi1 66 V02 52 04 V00 00 00 V03 05 00 V02 74 07 V02 02 00 V00 00 00 V00 00 00 V02 33 00 V03 08 00 V00 00 00 V00 00 00 V03 00 03 V01 23 00 00 00 00 V00 00 00 V03 04 00 V04 03 00 V02 52 04 V02 23 00 V03 05 00 V02 74 07 V02 02 00 V02 52 01 V02 44 00 V02 33 00 V03 08 00 V02 33 00 V03 08 00 V03 00 03 V01 23 00 V01 54 01 V01 53 00 V03 04 00 04 03 00 Sirona Dental Systems GmbH Service Manual GALILEOS Check program releases You can use the right hand window of the software manager to check which modules are connected to the unit and what their latest program release is NOTE From unit serial number 8500 for GALILEOS Comfort or 48500 for GALILEOS Compact and above a DX11_FPGA version is also displayed in addition to this list Modules which are connected and whose program release corresponds to the latest main software version see chapter entitled are identified by a continuous green bar Modules which the system does not recognize are identified by a broken red bar If the actual status of the module cannot be polled for the update the actual SW version will be displayed as V00 00 If a module has a hardware incompatibility to the program
207. ll service routines available for selection Service routine Function see 002 Radiation without rotary movement selectable kV mA level and S gt 221 maximum radiation time 002 5 Long term exposure with fixed radiation intervals from any S gt 221 position 005 General X ray tube assembly service S gt 223 005 1 Read select tube type S gt 223 005 4 Fan test S gt 225 005 5 Temperature sensor test single tank S gt 226 005 8 Automatic adjustment of pulse preheating S gt 227 S007 Error logging memory S gt 229 S007 1 Display error logging memory S gt 229 S007 2 Clearing error logging memory S gt 232 S007 5 Enable the CAN bus logging in the Miniweb S gt 235 S008 Update service S gt 238 008 2 Overview of the module software versions S gt 238 008 3 Confirming the unit serial number S gt 240 S009 Flash file system S gt 241 009 4 Formatting flash file system S gt 241 009 5 Test flash file system S gt 243 009 7 Save restore DX89 data S gt 245 011 Dosimetry without ring movement S gt 247 011 9 Current measurement unpulsed S gt 248 011 12 Dosimetry with pulsed radiation S gt 247 012 CAN bus service S gt 249 012 1 Presence display of modules S gt 249 S017 Configuration service S gt 252 017 2 Configuring the hardware ver
208. lt detected locally on e Check the CAN bus S gt 128 module slave side CAN controller e Check software versions on the info S gt 238 being reinitialized screen or by running service routine gt 61 S008 2 perform software update if i necessary e Please contact the Sirona Customer S gt 61 Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary e Replace the tube assembly S gt 310 61 25 665 D3437 D3437 076 01 15 02 06 2012 89 Error code Description Actions required see E7 06 10 Module is stuck in bootloader stage Check board DX6 note LED states S 134 If the board remains in the bootloader S gt 61 stage e Repeat the software update e Replace the tube assembly S gt 310 Error code Description Actions required see E7 06 12 Unit is not ready for operation e Check the CAN bus S gt 128 This error may also occur in If this error occurs in combination with connection with other causal error Other errors messages Please also observe the e Restart the unit causal error message It appears only after you acknowledge the first 1 Switch off the unit error message 2 Wait 1 minute 3 Switch unit on 4 Repeat procedure and observe causal error messages e Replace the tube assembly S gt
209. luate the image 61 25 665 D3437 D3437 076 01 15 02 06 2012 Upper jaw vV 61 25 665 D3437 D3437 076 01 15 02 06 2012 Adjustment OK Adjustment not OK The upper edge of the lower lead diaphragm must be within tolerance i e lie inside of the auxiliary lies A If the edge is out of tolerance B the diaphragm must be adjusted mechanically gt 197 If the exposure is OK A confirm this by clicking the check box located to the right of the Upper jaw button The box will appear checked The adjustment for the Upper jaw diaphragm setting is now complete The Lower jaw button is selectable Continue the calibration procedure with the adjustment of the Lower jaw diaphragm setting 175 7 Adjusting calibrating the unit Sirona Dental Systems GmbH 7 3 Adjustment and calibration via the Service Functions menu Service Manual GALILEOS Adjustment of the Lower jaw diaphragm setting The Service functions menu is called gt 161 The Diaphragm tab is selected The Lower jaw is selectable in the mage acquisition menu area Set the rotary knob on the tube assembly to the lower jaw position EOS SS 2 Click on the Lower jaw button SIDEXIS XG makes the unit ready for exposure gt 162 Dischuage Rasaton fekt Doca Saroa tte Shading Cutan Geomemy 4 Service routine S030 5 is displayed on the control panel 3 Take an expos
210. ments to sections 7 6 Checking the grounding straps and 7 7 Checking the cable shields in chapter 7 Maintenance Version 4 Software version V03 05 00 supplements and corrections in chapters 1 General information software versions and 4 Calibrating the unit input of calibration phantom serial numbers supplements and corrections in chapter 6 Repair rotary knob smaller corrections and supplements in complete manual Version 5 Supplements and corrections in chapter 1 General information dimensions changed due to shorter wall holder and cable due to discontinuation of board DX41 and chapters 2 and 3 due to discontinuation of board DX41 and new board DX32 Furthermore changes in chapter 6 replacement of cables in the cable tracks and corrections due to design changes Version 6 GALILEOS GAX5 added software version V03 06 01 Version 7 Head fixation device updated Version 8 Software version V03 06 02 Version 9 Diaphragm adjustment Version 10 Software version V03 07 00 supplements and corrections to chapters 1 General information software versions and 4 Calibrating the unit calibration of diaphragm Type 3 supplement to chapter 5 Service routines S017 25 diaphragm configuration and supplements to chapter 6 Repair replacing the tube assembly Version 11 Supplements to chapter 2 list of error messages Ex 89 xx supplements to chapter 6 Repair Version 12
211. n Actions required see E7 10 02 FPGA of DX1 is not addressable e Replace board DX1 S gt 343 FPGA Field Programmable Gate Array Error code Description Actions required see E1 10 03 The flash file system must be e Acknowledge error and repeat procedure formatted The flash file system is formatted and error Occurs after replacement of board code E110 04 is displayed DX11 Error code Description Actions required see E1 10 04 Flash file system formatting in e Wait until the error code automatically progress disappears approx 2 3 mins Error code Description Actions required see E1 1005 Flash file system is not ready for e Execute service routine S009 4 and S gt 241 operation format flash file system The contents of the error memory are thus lost If the error occurs repeatedly S gt 350 e Replace board DX11 Error code Description Actions required see E7 10 06 Incompatible DX1 FPGA e Check the hardware version of DX1 for S gt 343 98 61 25 665 D3437 D3437 076 01 15 02 06 2012 Error code Description Actions required see E1 10 07 The unit is not ready for operation If this error is displayed after a longer period of S gt 141 Following longer periods of disuse disuse and he attainment of exposure S gt 362 A readiness gt 200 h a preparation time of up to S gt 319 ten minutes is required for the e Acknowledge the error and wait until the
212. n until the exposure is complete the preview image is displayed in the exposure window and the acoustic signal indicating the end of the exposure double beep sounds if it has been configured IMPORTANT This process takes approx 2 3 minutes ees enum ar Serr uac eat nel soe cs The unit transfers the acquired images to the Reconstruction and Control Unit RCU This process can take 2 3 minutes Once the transfer is complete the evaluation of the sensor calibration is displayed in the message window If the information in the message window indicates that calibration is not OK and or not possible keep repeating the procedure starting with Step b until calibration is OK and or possible If you have repeated the procedure three times and still have not attained a positive result please contact the SIRONA Customer Service Center CSC Save values 5 If the adjustment is OK or possible click the Save values button The adjustment is saved The sensor adjustment is now complete 6 Continue the calibration procedure with the iris adjustment gt 182 61 25 665 D3437 D3437 076 01 15 02 06 2012 181 7 Adjusting calibrating the unit Sirona Dental Systems GmbH 7 3 Adjustment and calibration via the Service Functions menu Service Manual GALILEOS 7 3 5 Iris adjustment State Diaphragm Radiation field Dose Sensor lis Shading Distortion Geometry 4 gt
213. ndow The box will appear checked Calibration of the diaphragm is now complete 6 Continue the calibration procedure with the radiation field check gt 178 61 25 665 D3437 170 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 7 Adjusting calibrating the unit Service Manual GALILEOS 7 3 Adjustment and calibration via the Service Functions menu 7 3 1 2 Type 3 diaphragm Service functions Diaphragm Radiation field Dose Sensor lis Shading Distortion Geomety lt 4 gt Diaphragm exposure Please tum collimator switch to the center position Adjusting the diaphragm open J diaphragm setting S Jv The Service functions menu is called gt 161 The Diaphragm tab is selected The Collimator open is selectable in the mage acquisition menu area Set the rotary knob on the tube assembly to the open diaphragm position 2 Click on the Collimator open button Pa levramretnd eeel Shading Ditoten Geom sel SIDEXIS XG makes the unit ready for exposure gt 162 Service routine S030 5 is displayed on the control panel 3 Take an exposure 85kV 21 mAs gt 162 4 For GALILEOS Comfort Evaluate the image 61 25 665 D3437 D3437 076 01 15 02 06 2012 171 7 Adjusting calibrating the unit Sirona Dental Systems GmbH 7 3 Adjustment and calibration via the Service Functions menu Service Manual GALILEOS Adjustm
214. ne or the test step and confirm your selection with the Service key gt 213 After you have confirmed your selection a 0 appears in selection field 2 61 25 665 D3437 D3437 076 01 15 02 06 2012 215 8 Service Sirona Dental Systems GmbH 8 3 Basic operating procedures in the service menu Service Manual GALILEOS 2 Confirm security access by once again selecting the number of the main routine 2 in the example with the arrow keys in selection field 2 C and press the Service key B to confirm your selection 61 25 665 D3437 216 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS A X RAY 60kV 8mA Example lt 00 gt S17 61 25 665 D3437 D3437 076 01 15 02 06 2012 8 Service 8 3 Basic operating procedures in the service menu Following this double selection and confirmation via the Service key the service routine is activated Select parameters Easypad touchscreen If arrow keys are displayed in the selection fields once the required service routine has been selected you can use these arrow keys to choose between different parameters You want to run service routine S017 6 to activate the remote control J Once you have selected service routine S017 6 the code 00 is pre selected for the Remote control disabled option 217 8 Service Sirona Dental Systems GmbH 8 3 Basic operating procedures in the
215. ng or control data Check the CAN bus S gt 128 This error may also occur in e Please contact the Sirona Customer S gt 61 connection with other causal error Service Center CSC to find out whether messages Please also observe the a bug fix by means of a software update is causal error message It appears possible and perform such an update if only after you acknowledge the first necessary error message Error code Description Actions required see E6 71 04 Data transfer error or dialog error to e Check the CAN bus S gt 128 modieninaster ste e Please contact the Sirona Customer S gt 61 Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary Error code Description Actions required see E6 71 05 Data transfer error or dialog error to Repeat the software update S gt 61 pootloaderormedule e Check the CAN bus S gt 128 Only Occ rs in connection with e Replace board DX71 or Multipad S gt 343 software update S gt 307 61 25 665 D3437 D3437 076 01 15 02 06 2012 115 Error code Description Actions required see E6 71 06 Module failed in TTP time trigger e Check the CAN bus S gt 128 protocol detected on master side Replace board DX71 S gt 343 e Please contact the Sirona Customer S gt 61 Service Center CSC to find out whether a
216. ng unit data Reading the unit data of the GALILEOS via Extended Details 1 Start the S DEX S Manager under Start Programs SIDEXIS SIDEXIS XG 2 Click on Configuration of the X ray components The Configuration of the X ray components menu opens 3 Select the Affributes tab 4 Click on the More details button The current parameters are read from the unit and filed in an XML file under the network name of the unit in the PDATAV P adders paasa Delete component SurdnsGommy E Coe P2K_Config folder The process can take up to 30 seconds After Siena Conte Server ENTWICKLUNG the parameters are read an editor displaying the data is opened automatically Approved ta Pa comensert g ENTWICKLUNG serialnumber System Software version 10 Orthophos 00 10 19 00 25 1 192 165 Default Gateway Net API version Name of flat module oxd1 Code Number 925214 K 109256042 2 2004 Sep 17 5 PCB5STZPB004 30 59 2000 3u1 07 30 3 Oxff Oxff Oxff Oxff Oxff Oxff Oxff Oxff Extended Configuration ox 11 un 000000201 OxdFpgaversi 3 OxiFpgarevis 3 OxdHwerston 1 OxLimMemorystze 335544320 Dxdicpuctock 80020000 61 25 665 D3437 D3437 076 01 15 02 06 2012 77 4 General operating procedures 4 4 Reading unit data Sirona Dental Systems GmbH Service Manual GALILEOS Reading Facescan unit data over the network Opening the web d
217. nit back on The message No Key should no longer appear Call the service menu gt 210 Call the service routine S017 and perform the unit configuration test step 2 15 gt 252 The board DX41 must be configured using the service routine 017 9 In units with a serial number of 1080 and above the switching plate configuration of the swivel arm must be set or checked using service routine S017 7 Inform the customer of the configuration options of the software status for example the welcome screen or acoustic exposure signal Activate these functions if they are required If the travel height of the unit has to be limited Set the travel height with service routine S018 2 gt 272 Perform a software update to the current software version gt 61 This updates all modules in accordance with the configuration The error message E1 11 20 invalid unit calibration is displayed Acknowledge the error message with the R key Perform a complete unit calibration gt 155 After a successful unit calibration has been performed the error message should no longer appear Call up the Extended Details via SixABCon This generates an XML file with the system parameters which is Filed in the PDATA P2K_Config directory under the network name of the unit The process is completed 351 9 Repair Sirona Dental Systems GmbH 9 12 Boards Service Manual GALILEOS Case B e DX11 from another unit e GALILEOS Comfort e
218. nstalled remote control Switch the unit off and then on again Make sure that the release button is not actuated during switch on Was the release button actuated when the unit was switched on Do the error messages recur Short circuit in coiled cable of release button Replace release button with coiled cable Replace board DX41 145 Error messages with and without installed remote control Error code Description Actions required see E3 41 20 Faulty detection of release signal by Replace board DX41 S gt 343 DX41 processor when the unit is switched on E6 41 23 Faulty detection of release signal by Replace board DX41 S gt 343 DX41 processor during operation of the unit E6 41 25 The DX41 detects no release signal Replace board DX41 S gt 343 when the exposure is started E3 13 40 Short circuit in signal path between Replace cable L7 S gt 362 boards DAI vane Dae duning e Replace board DX41 S gt 343 switch on e Replace board DX1 e Replace board DX11 E6 13 41 Release signal missing on board e Replace cable L7 S gt 362 NUE at Stan Ch Oxp as ire e Replace board DX41 S gt 343 e Replace board DX1 e Replace board DX11 E3 13 42 Short circuit in signal path between Replace cable L7 S gt 362 boards DATI and DAAT during e Replace board DX41 S gt 343 operation of the unit e Replace board DX1 e R
219. nt calibration state v Green and Valid data record checked calibration is in progress No calibration required O Green Calibration data record present calibration has not been performed calibration may be insufficient Red Invalid data record or no record present Calibration required Tools pictograph The tools pictograph shows which if any test phantom must be used for this particular calibration step Click Cancel to quit the Service functions menu 61 25 665 D3437 160 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 7 Adjusting calibrating the unit Service Manual GALILEOS 7 1 General information about unit adjustment and calibration Calling the Adjustment Calibration menu You can call the Service functions menu via SIDEXIS XG Utilities Constancy test 1 3D l Select X ray device l Service exposure Password prompt Select X ray component Service functions The Select X ray device and Select X ray component prompts are only displayed if more than one unit has been set up in SIDEXIS XG Password protection The Service functions menu is password protected Enter the first four digits of the current system date PC in reverse order as the password For example On 05 30 2010 the service password is 5003 Service mode When you open the Service functions menu the unit switches from user mode to the PC service mode logged by the PC In PC serv
220. o unit serial number 8 699 GALILEOS Comfort and 48 699 GALILEOS Compact 164 Geometry phantom IMPORTANT Make sure that the phantom is securely fastened and in an upright position in the bite block holder of the unit For the geometry calibration gt 187 you must insert the geometry phantom in the block holder of the unit This phantom must be removed again for all other calibration steps IMPORTANT This geometry phantom is not suitable for calibration with Facescan T 4 I 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS 7 Adjusting calibrating the unit 7 1 General information about unit adjustment and calibration From unit serial number 8 700 GALILEOS Comfort and 48 700 GALILEOS Compact T IMPORTANT Once it has been inserted into the bite block holder the geometry phantom must be aligned vertically and horizontally with the spirit level so that calibration can be performed correctly gt Insert the geometry phantom A into the pan bite block holder B on the unit and secure it with the screw C 61 25 665 D3437 D3437 076 01 15 02 06 2012 165 7 Adjusting calibrating the unit Sirona Dental Systems GmbH 7 2 Checking the mechanical system adjustment Service Manual GALILEOS Constancy test phantom
221. ode ssir ie a e e ae aaae Ee ae a ae a ea 65 4 2 1 2 Check program releases 2c accesso cases teed wks Gecncdveseh boot pleved eran 66 4 2 2 Updating the FaceScan firmware cceceececeeeeeeeeeeeeeeeeeeeeeeeeeenaes 67 4 2 2 1 Option 1 Update using the FaceScan USB stick ee 67 4 2 2 2 Option 2 Firmware update via the network cccccceeeeeeeeeeeeeees 70 4 3 Configuring US CII od sete ace orev Saute aa ee tee ange wands denice A Eiet e seat 71 4 3 1 Configuring the unit via the SIDEXIS Manager 71 4 3 2 COnmMQuriniG Face SCA ic festeveseheeteenstvedtvioct olan asl eti entail uneetan 71 4 3 2 1 Type 1 Configuration using the Facescan USB stick 06 71 4 3 2 2 Type 2 Configuration over a network cable peer to peer 75 4 3 2 3 Resetting the FaceScan configuration to factory default settings 76 4 4 Reading Unit datas iiciscccssvessevivecssenvaccawes quate e a eaea aaa a a e aa aaa 77 4 4 1 Reading the unit data of the GALILEOS via Extended Details 77 4 4 2 Reading FACESCAN unit data over the network 78 4 5 Using demo mode operation without radiation release cccceee 79 4 5 1 Switching on demo MOdGC ccccceeeeeseseceeeeeeeeneeeeenseeeesessececeneeeeeee 79 4 5 2 Switching off demo Mode 2 cscacicniauniceinteeacenceeenaey 81 61 25 665 D3437 D3437 076 01 15 02 06 2012 5 MESSAGES seers Sed fe ae accede tendbackd stash iaa i a A a a ae iE iE E
222. of an error The USB_STICK_MODE entry in the facescan_settings cfg file is reset to CONFIG 3 Wait until the green status display LED goes out The blue LED should then light up The update is completed 4 Switch GALILEOS off 5 Remove the Facescan USB stick safely from the USB port 61 25 665 D3437 68 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 4 General operating procedures Service Manual GALILEOS 4 2 Updating the firmware Checking the update 1 Plug the Facescan USB stick into a PC 2 Open the facescan_settings log log file in a text editor 3 Check the entries in the log file Ifthe update was successful the log file should state Facescan device software updated with version successful 4 Remove the USB stick from the PC Closing Facescan 1 Pull gray cable L78 4 from slot X2 of the FACESCAN modular board 2 Screw the protective plate T onto the Facescan unit 3 Plug cable L78 4 again into slot X2 on the FACESCAN modular board 4 Screw down the cover A onto the Facescan Concluding the update 1 Switch the unit on 2 Perform a complete unit calibration gt 155 3 Perform a white balance gt 190 61 25 665 D3437 D3437 076 01 15 02 06 2012 69 4 General operating procedures 4 2 Updating the firmware Sirona Dental Systems GmbH Service Manual GALILEOS Option 2 Firmware update via the network Opening the web dialog 1 In SIDEXIS Manager start the
223. of the sensor surface H IMPORTANT Possible faulty alignment Displacement of the installation aid C can be caused through further fastening of the scan unit with the safety screws E gt When screwing in the safety screws E please ensure that the installation aid C always remains aligned 61 25 665 D3437 328 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 9 Facescan 8 Move the safety screw E on both sides from the parked position L delivered condition to the function position M 9 Screw the scan unit down from below with the safety screws E In so doing the screws should pierce through the housing of the X ray detector The distance K between the plate of the scan unit and the cover of the X ray detector should be 6 mm 10 Screw down the 6 screws D 11 Remove the installation aid C stays with the customer 12 Unscrew the transport lock B Keep the transport lock B and the wing screws A in case the Facescan unit has to be returned for repairs these stay with the customer 61 25 665 D3437 D3437 076 01 15 02 06 2012 329 9 Repair 9 9 Facescan Connecting scan unit electrically 1 2 Sirona Dental Systems GmbH Service Manual GALILEOS Loosen the two screws S Remove the protective plate T The FACESCAN modular board is open i Read the MAC address on the FACESCAN modular boa
224. oftware 1 Plug the Facescan USB stick into the PC 2 Open the facescan_settings cfg configuration file in a text editor 3 Enter the value UPDATE for the USB_STICK_MODE entry for example USB_STICK_MODE UPDATE 4 For selection by DHCP Enter the value ON for the DCHP_STATE entry for example DHCP_STATE ON or gt For selection without DHCP Enter the value OFF for the DCHP_STATE entry for example DHCP_STATE OFF 5 Enter the value OFF for the DCHP_STATE entry for example DHCP_STATE OFF 6 Specify the IP address and the subnet mask in the IP and Netmask entries 7 Save your changes 8 NOTICE Incorrect removal of the USB stick can lead to loss of data on the USB stick Remove the USB stick using the Safely Remove function operating system of the PC Starting the update J The unit must be switched off Always switch the device off before inserting the USB stick Before the Facescan USB stick can be inserted into the USB socket of the FACESCAN modular board GALILEOS must be switched off Otherwise the update will not be completed Instead the configuration data stored on the USB stick will be uploaded to GALILEOS 1 Insert the USB stick into the USB port U of the FACESCAN modular board 2 Switch the unit on again The Facescan will be updated Both LEDs in the status display light up during the update The process takes around 5 minutes around 1 minute in the event
225. on Actions required see E6 07 07 TTP timeout error detected on e Check the CAN bus S gt 128 slave side e Check power supply of board DX11 The module was temporarily not measuring point 3 3 V on board Dx1 addressed by the master see wiring diagrams Undervoltage on the master If 3 3 V is present S gt 342 side e Replace board DX11 e Procedure error in the software If 3 3 V is not present S gt 342 e Master DX11 receives no e Replace board DX1 return commanding from the module This error may also occur in connection with other causal error messages Please also observe the causal error message It appears only after you acknowledge the first error message TTP Time Trigger Protocol Error code Description Actions required see E6 07 08 General fault detected locally on e Check the CAN bus S gt 128 module slave side CAN controller e Check software versions on the info S gt 238 being reinitialized screen or by running service routine g gt 61 008 2 perform software update if i necessary e Replace user interface with electronics S gt 307 DX7 e Please contact the Sirona Customer S gt 61 Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary 61 25 665 D3437 D3437 076 01 15 02 06 2012 95 Error code Des
226. optional The unit is prepared for optional operation with Facescan from the following unit serial numbers Unit class from serial number GALILEOS Comfort 9 000 GALILEOS Compact 49 000 Unit software version V04 09 00 or higher must be installed to operate the GALILEOS with Facescan 61 25 665 D3437 D3437 076 01 15 02 06 2012 17 3 Unit description Sirona Dental Systems GmbH Service Manual GALILEOS 3 1 Unit classes and versions GALILEOS Comfort mmm 3 MAAA dDY Zz a 20 go00 S kW mA 18 GALILEOS Comfort and GALILEOS Compact The Comfort and Compact unit classes differ by the equipment of the control panel B While the GALILEOS Comfort has a control panel with a color touchscreen Easypad the Compact has a simpler control panel with a single line display Multipad There are slight variations in how the two unit classes operate because of the difference in the control panels 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS Accompanying documents 9 gt Electrostatic discharge ESD gt Identification of single use devices e 61 25 665 D3437 D3437 076 01 15 02 06 2012 3 Unit description 3 2 Hardware Hardware Information on the unit The following symbols are applied to the unit Observe accompanying documents The Operating Instructions are provid
227. or code Description Actions required see E6 06 06 Module failed in TTP detected on Check the CAN bus S gt 128 master side e Please contact the Sirona Customer S gt 61 This error may also occur in Service Center CSC to find out connection with other causal error whether a bug fix by means of a messages Please also observe the software update is possible and causal error message It appears perform such an update if necessary only after you acknowledge the first error message TTP Time Trigger Protocol Error code Description Actions required see E6 06 07 TTP timeout error detected on e Check the CAN bus S gt 128 slave side e Check power supply of board DX11 The module was temporarily not measuring point 3 3 V on board Dx1 addressed by the master see wiring diagrams e Undervoltage on the master If 3 3 V present S gt 342 side e Replace board DX11 e Procedure error in the software If 3 3 V is not present S gt 342 e Master DX11 receives no e Replace board DX1 return commanding from the e Check cable L6 replace if necessary S gt 141 module S gt 362 Enis Or may alse Occur Ip e Check tube assembly DX6 replace if S gt 310 connection with other causal error necessary messages Please also observe the causal error message It appears only after you acknowledge the first error message TTP Time Trigger Protocol Error code Description Actions required see E6 06 08 General fau
228. or dialog error to Check the CAN bus S gt 128 module m ster side e Please contact the Sirona Customer S gt 61 Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary Error code Description Actions required see E6 89 05 Data transfer error or dialog error to Repeat the software update S gt 61 Bootbad er Onmacule e Check the CAN bus S gt 128 DUOC EHTS Canine Chon with e Check x ray detector replace if S gt 319 software update necessary Error code Description Actions required see E6 89 06 Module failed in TTP time trigger e Check the CAN bus S gt 128 Piotecol JECTION MASTER side e Check x ray detector replace if S gt 319 necessary S gt 61 e Please contact the Sirona Customer Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary Error code Description Actions required see E6 89 07 TTP time trigger protocol timeout e Check the CAN bus S gt 128 error detected on slave side Check cable L13 replace if necessary S gt 141 ULE PEA ial ine not e Check power supply of board DX11 S gt 362 y measuring point 3 3 V on board DX1 see e Undervoltage on the master wiring diagrams side If 3 3 V present S gt 343 e Procedure error in the software Replace board DX11 e Master DX11 receives no If 3
229. ort locks gt Screw the transport lock B to the Facescan using the wing screws A for return shipping 1 Unscrew the securing screws E 2 Unscrew the six screws D and detach the scan unit from the X ray detector brackets to the rear Attaching new scan unit Risk of damage Mechanical stress can cause damage to the scanning unit gt For transporting and aligning the scan unit hold on to the transport lock B only gt Loosen one wing nut A of the transport lock B NOTICE gt Do not remove the wing nuts Attaching the scan unit IMPORTANT Risk of damage When hanging the scan unit the housing of the X ray detector can get scratched gt Hang the scan unit carefully on the X ray detector 61 25 665 D3437 D3437 076 01 15 02 06 2012 327 9 Repair Sirona Dental Systems GmbH 9 9 Facescan Service Manual GALILEOS 1 Hang the scan unit from behind in the inlets of the X ray detector 2 Attach the scan unit loosely with 6 screws D do not screw tightly 3 Mount the installation aid C from below on both sides onto the first cooling vent H of the face scan unit 4 Insert the fastening element F into the installation aid C 5 Turn the holding plates G over the chassis plate of the face scan unit 6 Push the installation aid C with the scan unit on to the sensor surface H of the X ray detector 7 Align the installation aid C to the center
230. ouble free operation is possible after the error is acknowledged then no further action is necessary If error messages occur again or frequently or if fault free operation is not possible run an error diagnosis see chapter entitled Troubleshooting gt 125 In some cases it can be advisable to obtain more information about the history or frequency of the errors from the error logging memory S007 and from SiXABCon l Properties More details see chapter entitled Opening Extended Details see also chapter entitled Error logging memory gt 125 61 25 665 D3437 D3437 076 01 15 02 06 2012 87 List of error messages In the following table the error codes are sorted by the location or function unit where the error has occurred For enhanced clarity the corresponding ID in the error code is printed in bold type Location 06 Tube assembly DX6 Error code Description Actions required see E6 06 01 General error during module e Please contact the Sirona Customer S 61 initialization Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary If the error occurs repeatedly S gt 310 e Replace the tube assembly Error code Description Actions required see E6 06 02 Invalid system data or uninitialized Please contact the Sirona Customer S gt 61 module storage data Service C
231. outine gt 220 DX11 V02 61 01 DX41 V02 30 00 Dx42 V02 45 06 DX6 V02 88 00 DX89 V01 10 06 DX89_F1 V01 13 01 DX7 V02 57 00 DX lang_0 V02 18 00 61 25 665 D3437 238 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 7 S008 Update service Multipad 1 Call service routine S008 2 gt 213 SYSTEMSOFTWARE is displayed on the display line of the Multipad 2 Select the required module in selection field 1 with the arrow keys A and confirm your selection by pressing the Memory key B The software version of the selected module is displayed in selection field 1 3 Exit the service routine gt 220 61 25 665 D3437 D3437 076 01 15 02 06 2012 239 8 Service Sirona Dental Systems GmbH 8 7 S008 Update service Service Manual GALILEOS S008 Test step 3 Confirming the unit serial number IMPORTANT If the backup copy of the old unit serial number does not match the new one after replacing a module the entry of the serial number is activated If an incorrect serial number is entered the message FFFF appears on the display In this case the service routine can be run again 1 Call service routine S008 3 gt 213 2 Confirm the serial number displayed by pressing the R key A 3 Exit the service routine gt 220 61 25 665 D3437 D3437 076 01 15 02 06 2012 240 Sirona Dental Sys
232. plate After replacing the board DX11 you must observe the notes provided in the chapter Replacing board DX32 The removal of the board DX32 is described in the chapter Removing board DX32 gt 294 Install the board by following the same procedure in reverse order After replacing the board DX32 you must observe the notes provided in the chapter 61 25 665 D3437 D3437 076 01 15 02 06 2012 345 9 Repair Sirona Dental Systems GmbH 9 12 Boards Service Manual GALILEOS Replace board DX89 1 Remove the x ray detector cover 2 Carefully pull cover plate A upwards to remove it from the X ray detector 3 Remove the four screws B and remove the defective DX89 board from the X ray detector 4 Pull the connectors of cables L13 X201 L27 X203 and L28 X400 off of the defective board DX89 5 Install the new board DX89 by following the steps for removal in reverse order IMPORTANT After replacing the board DX89 always observe the notes provided in the chapter 61 25 665 D3437 346 D3437 076 01 15 02 06 2012 Measures following replacement of boards After replacing boards or modules containing boards check to make sure that the software version of the module corresponds to the current software status of the system The software versions of the modules can be queried by running service routine S008 2 or using the extended detail query in SiXABCon You can also check the info scre
233. r you acknowledge the first perform such an update if necessary error message Error code Description Actions required see E6 07 04 Data transfer error or dialog error to e Check the CAN bus S gt 128 module master side e Please contact the Sirona Customer S gt 61 Service Center CSC to find out whether a bug fix by means of a software update is possible and perform such an update if necessary Error code Description Actions required see E6 07 05 Data transfer error or dialog error to e Repeat the software update S gt 61 bootloader of module e Check the CAN bus S gt 128 Only occurs in connection with e Replace user interface with electronics S gt 307 software update DX7 94 61 25 665 D3437 D3437 076 01 15 02 06 2012 Error code Description Actions required see E6 07 06 Module failed in TTP detected on Check the CAN bus S gt 128 master side e Please contact the Sirona Customer S gt 61 This error may also occur in Service Center CSC to find out connection with other causal error whether a bug fix by means of a messages Please also observe the software update is possible and causal error message It appears perform such an update if necessary only after you acknowledge the first e Replace user interface with electronics S gt 307 error message DX7 TTP Time Trigger Protocol Error code Descripti
234. radiation release is blocked Switching on demo mode When operated in demo mode the unit must not release any radiation For this reason you must take the following safety measures 1 Switch off the unit Potentially lethal shock hazard It is essential to switch off the unit and to wait at least another 4 minutes before taking off the covers of the X ray tube assembly 2 Remove the cover of the tube assembly 3 Loosen the screws A and remove the cover plate B 79 4 General operating procedures Sirona Dental Systems GmbH 4 5 Using demo mode operation without radiation release Service Manual GALILEOS 4 Set dip switch S2 DX6 to position 2 IMPORTANT If switch S2 is not set to position 2 in demo mode before switching off the unit various error messages will display when the unit is turned back on 5 Pull cable L5 XRAY off connector J6 DX6 Radiation release is now no longer possible 6 Switch on the unit and check the mode on the info screen Demo mode ON means that Demo mode is switched on radiation release is not possible Demo mode OFF means that Demo mode is switched off radiography X ray radiation are possible 7 Switch the unit off again and reattach the cover plate B and the tube assembly covers by following the dismantling procedure in reverse order Bootmode Default Type 100 Network Name GALILEOS Net MAC address 00023181247 IP address 192 168 15 24
235. rd at position X and note this down Plug gray cable L78 3 in slot X1 on the FACESCAN modular board Attach gray cable L78 3 to the chassis at the side 8 9 Screw down the protective plate T Lay cable L78 4 including the ferrite core on the protective plate as shown Plug gray cable L78 4 into slot X2 modular board FACESCAN Fasten cable L78 4 with the clip Z 10 Fasten cable L78 3 with three cable ties to cable L78 4 330 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS Attaching the cover 61 25 665 D3437 D3437 076 01 15 02 06 2012 9 Repair 9 9 Facescan 1 Remove the protective caps E from the four cameras IMPORTANT Permitted cleaning agents A dry lint free cloth A cleaning agent approved by Sirona An up to date list of approved agents can be downloaded from the Internet at the address www sirona com Select the SERVICE Care and cleaning menu items in the navigation system and then open the Care and cleaning agents document If you have no access to the Internet please contact your dental depot to request the list REF 59 70 905 2 Clean the surface of the mirror and the vision panel inside the Facescan cover 3 Screw down the covering bonnet F from below with 6 screws G 4 Connect the cable H with the display board I to the covering bonnet F 331 9 Repair Sirona Dental Systems Gm
236. rent software versions Inthe left hand window of the software manager you can now select the update modes Aufomatic and Main version using the tabs SW Update SW update mode Automatic Main version Module Actual vers EH DXx42 02 45 06 Y02 45 06 EA ODX41 Y02 30 00 02 30 00 EA Dxo6 v02 88 00 02 98 00 E 0x07 yo2 57 00 02 57 00 E DX07_LO 02 18 00 02 18 00 s u Y01 10 06 Y01 10 06 p Y01301 01 13 01 Y0260 07 02 61 01 Start SW update Show logfile 9 Select the desired mode for the software update see chapter entitled Update mode 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 4 General operating procedures Service Manual GALILEOS 4 2 Updating the firmware NOTICE Unit inoperability Before starting the software update make sure that no unit movements are active Otherwise the system may become inoperable in rare cases The X ray detector must be installed as part of the update Exposure readiness must be deselected in SIDEXIS XG and the unit must not already be in service mode 10 Click on the Start SW update button SW Update SW update mode Automatic Main version Module Actual vers _ Target vers BA DXx42 02 45 06 02 45 06 BR ODX41 02 30 00 02 30 00 EA DX06 02 98 00 02 88 00 EA Dx07 yo2 57 00 02 57 00 EA DX07_LO 0218 00 0218 00 shee s 01 10 06 01 10 06 e 01 1301 01 13 01 Y0260 07 02 61 01
237. ring booting 4 Repeat procedure and check function If the error occurs again S gt 142 e Check signal path for short circuit according to wiring diagrams replace component if necessary Error code Description Actions required see E6 41 21 CAN bus connection to board DX41 e Check cable L17 replace if necessary S gt 141 interrupted S gt 362 Board DX41 cannot address board Check board DX42 replace if necessary S gt 134 DX42 via the separate CAN bus connection S gt 343 e Check board DX41 replace if necessary S gt 134 S gt 343 61 25 665 D3437 D3437 076 01 15 02 06 2012 109 Error code Description Actions required see E6 41 23 Hardware fault at controller input on see section Error analysis of X RAY S gt 142 board DX41 control signal path up to unit serial Board DX41 detects a wrong signal number 3199 with board DX41 gt 142 level of the hardware signal for radiation release Error code Description Actions required see E6 41 24 Short circuit in radiation release e see section Error analysis of X RAY S gt 142 signal path between board DX42 control signal path up to unit serial and board DX41 cable L17 number 3199 with board DX41 gt 142 The release signal was detected on boards DX11 and DX41 but not on board DX42 Error code Description Actions required see E6 41 25 X RAY hardware sign
238. ring motor M1_3 pinion 305 Rotary knob on the swivel AFM cceesecceeecceeeeceeeeaeeeeceeaeeeaeeeeeeeeeeeeeeees 306 9 4 1 Replacing the rotary knob ccccceccccccecceeceeeceeeeeeeeeeeeeeeeeeeeeeeneess 306 Control Pane viscecrenduensscavceserquewasetwnceluacacavedeehideantaeatds E ae EE E Raik 307 9 5 1 Replacing the Easypad user interface GALILEOS Comfort or 307 Multipad GALILEOS Compact 9 5 1 1 What has to be done after replacing the user interface 308 9 5 2 Laying of cables when replacing the user interface cceeeeeee 309 Xray tube Units ieee ara i a a eaae i a a aS 310 9 6 1 Replacing the X ray tube assembly ccccccccceeeeeeeeeeeeeeeeeeeeeeeeees 310 9 6 2 Cables and connectors for replacement of the X ray tube assembly 316 9 6 3 What has to be done after replacing the X ray tube assembly 317 Fan X ray tube assembly icc sacssnaciteeeea te ceticewieiawscriieseaetenedavienety 318 9 7 1 Replacing the fan ise 25 ie Src ela he fable Sa he Oke cae 318 9 7 2 What has to be done after replacing the fan ccccceeceeeeteteeees 318 X ray detetom iranerne E AESSR ENEAN EEEE e AE Ranea 319 9 8 1 Replace X ray detector ccceseesesccceeeeceeeseccenenteneesetsdeedeneeneeenteees 319 9 8 2 What has to be done after replacing the X ray detector 0 324 10 9 9 9 10 9 11 9 12 9 13 FACS SCAN is Fac crane ces hs e ea t
239. rm a software update to the current main software version V03 03 01 or higher gt 61 Save the configuration data from board DX89 to board DX11 via service routine S009 7 gt 245 Perform a complete unit adjustment or calibration gt 155 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 9 Facescan Facescan Replacing the scan unit Removing the defective scan unit Removing the covers 1 Undo the four screws K and remove the two covers J 2 Disconnect the cable H from the display board I on the inside of the cover F 3 Undo the six screws G and remove the cover F 61 25 665 D3437 D3437 076 01 15 02 06 2012 325 9 Repair 9 9 Facescan Sirona Dental Systems GmbH Service Manual GALILEOS Disconnecting electrical connections Disconnect cable L78 3 from cable L78 4 Remove the cable L78 4 from the clamp Z Pull cable L78 4 from socket X2 of the FACESCAN modular board Disconnect the gray cable L78 3 from the chassis to the side Slacken the two screws S and remove the panel T EY The FACESCAN modular board is open AUN S siu IK 6 Pull gray cable L78 3 from socket X1 off the FACESCAN modular board 7 Refit the panel T with the two screws S 61 25 665 D3437 326 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 9 Facescan Attaching the transp
240. rom those during the subsequent Facescan process this will lead to inaccurate colors gt Ensure that the lighting conditions during white balancing correspond with those of the later Facescan operation Without white balance the pictures will be tinged A white balance can be performed during e The first installation e Making changes to environmental lighting e Following a software update Opening the web dialog 1 In SIDEXIS Manager start the Facescan Configuration application 2 Click on the Facescanner in browser button A password dialog box opens 3 Inthe field User enter service 4 Inthe field Password enter sirona The Facescan Device Service web dialog opens White balance type 1 J 1 white sheet of paper such as DIN A3 1 Attach the sheet of paper A to where the patient s head goes so that the cameras are pointed at the white surface area 2 Select the menu item SERVICE 3 Press the Enter Auto White Balance button The Auto White Balance dialog box opens 4 Wait until the status LEDs light up 5 Press the Auto White Balance button The white balance starts During this time the LEDs of the Facescan are on for the duration of the white balance The white balance takes a minimum of 2 minutes up to 20 minutes in the event of errors In the event of errors If the white balance does not work then the white balance procedure must be repeated T
241. rt the Facescan USB stick into a PC 2 With a text editor program open the log file facescan_settings log 3 Check the entries in the log file Ifthe configuration has been successful the log file should state Network configuration successful 4 Remove the USB stick from the PC Closing Facescan _ Pull gray cable L78 4 from slot X2 of the FACESCAN modular board 2 Screw the protective plate T onto the Facescan unit w Plug cable L78 4 again into slot X2 on the FACESCAN modular board 4 Screw down the cover A onto the Facescan 61 25 665 D3437 D3437 076 01 15 02 06 2012 73 Sirona Dental Systems GmbH Service Manual GALILEOS 4 General operating procedures 4 3 Configuring the unit Syntax of the configuration file facescan_settings cfg IMPORTANT Pay attention to the syntax Text entries should never have a space before and after Examples Correct USB_STICK_MODE CONFIG Incorrect USB_STICK_MODE CONFIG Configuration without DHCP Text entry factory setting USB_STICK_MODE CONFIG DHCP_STATE OFF IP 192 168 16 240 Netmask 255 255 255 0 IMPORTANT Changes to network addresses gt Adjustthe entries P IP address and Netmask subnet mask as required Configuration with DHCP Text entry DHCP_STATE ON 61 25 665 D3437 74 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS sirona Facescan Conf
242. ry data Stringname Key Act Activation transaction Key Ok Activation transaction Entry data Stringsegment 7YFWDUFV E4MMRJBW e g activation or confirmation code for activation transaction 061 00133 e g counter ID counter reading Entry data Tried to change to e g Tried to change to RTC Date Time Change YYYY MM DD HH MM SS 2006 Nov 30 11 32 13 Entry data YYYY MM DD HH MM SS 2006 Nov 30 11 32 13 PC Date Time 61 25 665 D3437 D3437 076 01 15 02 06 2012 127 6 Troubleshooting Sirona Dental Systems GmbH 6 2 Checking the CAN bus Service Manual GALILEOS Checking the CAN bus NOTICE Risk of damage to unit The power supply MUST be plugged in and switched on when cables are attached and plugged in For example if no power cable is connected to the DX71 the module has no ground connection to the unit and there is no potential equalization If the CAN cable is plugged in the CAN transceiver IC on the DX71 can be destroyed by the voltage difference In other words when the unit is switched on CAN cables may only be plugged in on modules that are connected to the power source and ground For troubleshooting you can disconnect the CAN bus cable and or plug it back in and observe the unit s behavior 61 25 665 D3437 128 D3437 076 01 15 02 06 2012 Re measure the ohmic resistance between the CAN H and CAN L measuring pins of connector X310 Resist
243. s applicable in your country see Operating Instructions X rays Radiation is signaled by the message X RAY active a beep and an X RAY LED You will need the following accessories to perform unit calibration e Geometry phantom gt 164 e Distortion phantom gt 163 e Steel tape measure 300mm IMPORTANT If you encounter problems with unit calibration check whether the required EMC conditions have been met No heavy duty electric equipment e g air conditioners fan motors etc should be present in the vicinity of the unit Tip Move the unit to a typical working height bite block height A approx 1 520 mm 60 with the Up Down keys on the control panel before commencing calibration 155 7 Adjusting calibrating the unit Sirona Dental Systems GmbH Service Manual GALILEOS GALI Eos y As of unit software version V 03 04 00 you can even adjust the height of the unit during the calibration procedure If the unit is ready for an exposure after the mage acquisition button has been pressed in the SIDEXIS XG service menu the corresponding service routine S002 6 010 10 14 S011 8 and S030 5 is displayed on the control panel All of these service routines allow the height adjustment menu to be opened by pressing the Test key The
244. s often as necessary until the failed messages no longer appear P p2ksw_ 20061120 1504 log Editor RT 5 Reading image data from SIOEXIS jaximun available memory for Softwar update 10485760 Automatic version update to version vO3 03 01 B nothing to do for 0x42 S Nothing to do for Dsl B Nothing to do for 0X06 8 nothing to do for 0x07 8 Nothing to do for DxO7_LO amp Nothing to do for 089 8 nothing to do for ox69_Fl jpdate of D1 V226 60 CV02 60 07 to V34 61 V02 61 01 Image CompImg v34 61 D11 PROC_PPC V 00 2 8 15 2000 13 34 56 Update of 0x11 done 08 15 2000 13 35 03 version descriptor successfully written 13 Restart the unit now 14 Use the software manager or the service routine S008 2 see chapter entitled Unit software versions and compatibility to check whether all modules have been updated to the latest release of the program see chapter entitled Using the update manager to check the program releases 15 Call up More details via SiXABCon This generates an XML file with the system parameters which is Filed in the PDATA P2K_Config directory under the network name of the unit 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS SW Update 02 45 06 02 30 00 02 88 00 V02 57 00 v02 18 00 01 10 06 01 13 01 V02 60 07 J SW update mode Automatic Main version V02 45 06 02 30 00 V02 88 00 v02 57 00 02 18
245. s required see E7 89 29 Memory test error during system e Restart the unit bootup 1 Switch off the unit 2 Wait 1 minute 3 Switch unit on 4 Repeat procedure and check function If the error occurs repeatedly S gt 343 e Check that the memory modules on board DX89 are firmly fixed replace board DX89 if necessary Error code Description Actions required see E7 89 30 Flash memory component does not e Restart the unit respond 1 Switch off the unit 2 Wait 1 minute 3 Switch unit on 4 Repeat procedure and check function If the error occurs repeatedly S gt 343 e Replace board DX89 Error code Description Actions required see E6 89 32 TDI Signal to start synchronized e Check cable L13 replace if necessary S gt 141 readout sequence and to prepare the next exposure pulses not aes detected during the exposure e Check board DX89 replace if necessary S gt 134 S gt 343 e Check board DX1 replace if necessary S gt 134 S gt 343 61 25 665 D3437 D3437 076 01 15 02 06 2012 123 Error code Description Actions required see E6 89 33 Board DX89 has detected an image Check cable L13 replace if necessary S gt 141 signal at the wrong point of time S gt 362 e Check board DX89 replace if necessary S gt 134 S gt 343 e Check
246. s service kit is used to prevent automatic downward movement of the slide during service by fixing the slide and the spindle holder The clamp A should be clamped under the slide The locking pin B is used to secure the spindle holder against twisting It must be ensured that no one is located underneath the ring arm during the repair 61 25 665 D3437 290 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 2 Height adjustment motor M1_4 spindle Moving the slide with the height adjustment service kit REF 62 57 518 Move the slide 1 Insert the clamp through the opening A in the stand and rotate it 90 Tighten the nut securely 2 Loosen the two screws C and remove the cover D 3 Remove the 1st screw E on the spindle holder F 4 Insert the locking pin B into this opening 5 Remove the 2nd screw E on the spindle holder 6 Attach the socket wrench SW19 to the spindle Remove the locking pin and then turn the slide up to the desired height using the socket re wrench gt Clockwise rotation of spindle slide moves up Counterclockwise rotation of spindle slide moves down 7 Reinsert the locking pin 61 25 665 D3437 D3437 076 01 15 02 06 2012 291 9 Repair Sirona Dental Systems GmbH 9 2 Height adjustment motor M1_4 spindle Service Manual GALILEOS Locking the slide gt Now
247. screen can be selected in exactly the right position If problems occur during selection adjust the touchscreen gt 191 4 Press the R key The unit moves to its starting position 5 Switch on the PC 6 Start SIDEXIS XG As long as no connection has been made to SIDEXIS XG the message Switch SIDEXIS to ready for exposure state is displayed in the comment line of the control panel 61 25 665 D3437 D3437 076 01 15 02 06 2012 59 4 General operating procedures Sirona Dental Systems GmbH 4 1 Switching the unit on Service Manual GALILEOS Switching the GALILEOS Compact on 1 Turn the main switch A to position I 2 Wait for approx 1 minute The X ray radiation indicator B lights up for approx 1 second as a functional check After approx 2 seconds the green LED C in the upper part of the control panel lights up This LED remains lit as long as the unit is on a 3 Press the R key The unit moves to its starting position 4 Switch on the PC 5 Start SIDEXIS XG Help message H401 remains displayed on the Multipad as long as there is no connection with SIDEXIS XG Factory setting after switch on The unit has the following factory configuration on delivery e Start settings Starting position from the front right VO1 for GALILEOS Comfort V04 for GALILEOS Compact Patient symbol 2 85 kV 21mAs e The acoustic signal for end of expos
248. service routine exposure parameters are visible 2 Press the release button Press and hold down the button until the exposure is complete the preview image is displayed in the exposure window and the acoustic signal indicating the end of the exposure double beep sounds if it has been configured Save values Adjustment or calibration must be OK i e the values in the menu s text boxes must be equal to zero gt To save the adjustment or calibration values click the Save values button The adjustment or calibration is saved The saved adjustment or calibration is identified in the structure tree by a check mark 2D adjustment or a green traffic light symbol with a check mark 3D adjustment 61 25 665 D3437 162 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 7 Adjusting calibrating the unit Service Manual GALILEOS 7 1 General information about unit adjustment and calibration Test phantoms for adjustment and calibration Distortion phantom You must clip the distortion phantom onto the X ray detector cover for the radiation field check gt 178 and for the distortion calibration gt 185 This phantom must be removed again for all other calibration steps 61 25 665 D3437 D3437 076 01 15 02 06 2012 163 7 Adjusting calibrating the unit Sirona Dental Systems GmbH 7 1 General information about unit adjustment and calibration Service Manual GALILEOS Up t
249. signing has not been assigned to any other unit 1 Switch the unit on Do not acknowledge any error messages at this time 2 Install the current SIDEXIS XG software version V2 3 or higher If the current version of SIDEXIS XG is a patch version the previous official main version of SIDEXIS XG must be installed before you can install the current version 3 Perform a software update to version V03 06 01 or higher via automatic update gt 61 If you are using a DX11 that already has the same software status as the overall system a repeated software update to this status must be performed in order that an administrative entry can be made in the memory of the DX11 4 Switch the unit off Wait for approx 1 minute Then switch the unit back on The error message E6 15 05 undefined system serial number is displayed The message No Key is displayed on the Easypad Acknowledge the error message with the R key a Call the service menu gt 210 Call service routine S008 3 gt 240 Enter the unit serial number found on the nameplate of the unit gt 240 IMPORTANT Any serial number which is unknown to the unit will not be accepted by the unit The serial number entered must be identical to the number on the nameplate of the unit If an inadmissible serial number is entered the input will not be accepted and the serial number can be entered again 10 Switch the unit off 11 Wait for approx 1 minute Then
250. sion S gt 252 017 3 Enter the country group code S gt 254 017 4 Select a language S gt 256 for GALILEOS Comfort 017 5 Select a language set S gt 258 for GALILEOS Comfort 017 6 Activate the remote control display S gt 259 017 7 Configure the switching plate for the swivel arm S gt 261 017 9 Activate deactivate operation with board DX41 S gt 263 017 13 Enable disable the welcome screen S gt 265 for GALILEOS Comfort 202 61 25 665 D3437 D3437 076 01 15 02 06 2012 Service routine Function see 017 14 Enable disable certain lines of the welcome screen S gt 266 for GALILEOS Comfort 017 15 Activate deactivate the acoustic signal for end of exposure S gt 268 017 25 Select the diaphragm type S gt 270 S018 Service for height adjustment S gt 272 018 2 Set the maximum travel height S gt 272 018 3 Undo the maximum travel height setting S gt 274 018 4 Check the height adjustment sensor system S gt 275 037 Network service S gt 277 037 1 Displaying the network data S gt 277 037 2 Setting the default IP address default gateway address and S gt 279 default subnet mask 037 3 Configuring boot mode DYNAMIC DHCP AutolP STATIC S gt 281 fixed address 037 4 Manual input of static network settings IP address default S gt 283 gateway address and subnet mask
251. software version V04 04 00 or higher X402 DX1 X403 100 X104 X607 X611 X804 X802 X803 i a TS FET X811 X812 X813 X610 TLL REF 62 82 052 MMU LOE REF 63 17 056 61 25 665 D3437 32 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS 3 Unit description 3 2 Hardware Board DX6 This board is not available as a spare part or a repair part X ray tube assemblies can only be ordered as complete units DX6 X305 X306 X2 X404 61 25 665 D3437 D3437 076 01 15 02 06 2012 33 3 Unit description Sirona Dental Systems GmbH 3 2 Hardware Service Manual GALILEOS Board DX7 This board is only used in the GALILEOS Comfort not in GALILEOS Compact The board is not available as a spare part or a repair part The Easypad can only be ordered as a complete unit DX7 5 61 25 665 D3437 34 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 3 Unit description Service Manual GALILEOS 3 2 Hardware Board DX71 This board is only used in the GALILEOS Compact not in GALILEOS Comfort DX71 X104 X106 X203 X202 X103 X102 61 25 665 D3437 D3437 076 01 15 02 06 2012 35 36 3 Unit description Sirona Dental Systems GmbH 3 2 Hardware Service Manual GALILEOS Board DX89 DX89 X203 X201 EVI ZHWEES 904S GWIS VI S13COVOS99EW ws eet NEEL 8190
252. switch the unit back on In systems that already run with a system software version V03 03 01 or higher please check whether there is an XML file in the PDATA P2K_Config with the network name of the system This file contains important information about the previous unit configuration The error message E1 11 20 invalid unit calibration is displayed 12 Acknowledge the error message with the R key 13 Call the service menu gt 210 o ON 61 25 665 D3437 356 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 12 Boards 14 Call the service routine S017 and perform the unit configuration test step 2 15 gt 252 The board DX41 must be configured using the service routine 017 9 In units with a serial number of 1080 and above the switching plate configuration of the swivel arm must be set or checked using service routine S017 7 Inform the customer of the configuration options of the software status for example the welcome screen or acoustic exposure signal Activate these functions if they are required 15 If the travel height of the unit has to be limited Set the travel height with service routine S018 2 gt 272 16 Perform a software update to the current software version gt 61 This updates all modules in accordance with the configuration 17 Perform a complete unit calibration gt 155 After a successful unit calibration has been performed the error message
253. sypad or the LED above Memory key B Multipad lights up 2 Toundo the maximum travel height setting press Memory key B R key C Easypad or LED above R key C lights up followed by R key C 3 Exit the service routine gt 220 61 25 665 D3437 274 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 12 S018 Service for height adjustment S018 Test step 4 Check the height adjustment sensor system This service routine is used to move the unit up or down as far as the limit switches using the Up Down keys on the control panel The soft limit positions set by the software are ignored in this case Display on the control Status Meaning panel Patient symbol key 1 lit Correction switch activated not lit Correction switch not activated Patient symbol key 2 lit Lower limit switch activated not lit Lower limit switch not activated Patient symbol key 3 lit Upper limit switch activated not lit Upper limit switch not activated 1 Call service routine S018 4 gt 213 Once the service routine has been selected the current height position is displayed in selection field 1 Patient symbol keys 1 to 3 F show the switching state of the limit switches see table If the patient symbol key Easypad or the LED above the patient symbol key Multipad is lit the corresponding switch is activated i e the unit is at a position v
254. t adjustment motor M1_4 spindle Service Manual GALILEOS Laying of cables when replacing the height adjustment motor 1 Plug connector X2 O into board DX32 2 Connect the protective ground wire P and lay it as shown in the photo 61 25 665 D3437 298 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 9 Repair Service Manual GALILEOS 9 2 Height adjustment motor M1_4 spindle 3 Attach cable L2 first to the lower strain relief photo on the left and then to the upper strain relief photo on the right Q of board DX32 Connect cable L2 to board DX32 R and attach the protective ground wire S 5 Run cable L3 T and the motor cable U around the height adjustment motor 61 25 665 D3437 D3437 076 01 15 02 06 2012 299 9 Repair 9 2 Height adjustment motor M1_4 spindle Sirona Dental Systems GmbH Service Manual GALILEOS 8 Lay the motor cable in the cable harness V on the rear of the unit and secure in position with the cable clamps Route the cable into the arm IMPORTANT The green mark must lie in the recess X Plug connector X402 W into board DX1 What has to be done after replacing the height adjustment motor M1_4 or the spindle 1 2 After inserting the new spindle above and below the height adjustment motor grease it thoroughly with Chesterton 622 Use the Up Down keys on the control
255. t lit V110 lit not lit V200 flashing at 1 Hz lit or not lit V610 lit not lit flashing at 2 Hz DX6 V1 flashing at 1 Hz not lit flashing at 2 Hz V203 lit not lit DX7 V100 lit not lit V101 lit not lit V102 lit not lit DX71 V101 lit not lit V103 lit not lit V107 flashing at 1 Hz not lit flashing at 2 Hz DX32 V132 lit not lit V133 lit not lit DX41 V103 flashing at 1 Hz not lit flashing at 2 Hz V202 lit not lit V204 lit not lit DX42 V101 lit not lit V103 lit not lit DX89 V201 lit not lit V202 lit not lit V203 lit not lit V204 flashing at 1 Hz not lit flashing at 2 Hz V205 lit not lit V207 lit not lit Starting with unit serial number 3201 new units will be delivered without board DX41 61 25 665 D3437 D3437 076 01 15 02 06 2012 135 Checking board DX32 Perform visual inspection of board DX32 Is the board visually damaged Are the LEDs V111 and V112 illuminated on board DX32 Yes Measure voltages on DX1 AA107 AA108 28 V 10 AA109 AA108 40 V 10 Yes Are the voltage levels OK and the LEDs V100 and V101 illuminated on board DX1 Yes Board DX32 is OK 61 25 665 D3437 136 D3437 076 01 15 02 06 2012 Check cable L4 Is the cable OK No Replace cable L4 End Check automatic fuses at F101 and F102 on board DX32 Check fuses on board DX32 e Switch off the unit and wait 7 minutes because of discharge Did the autom
256. t replacing any modules or boards Never replace these boards at the same time After replacing one of these boards you must first perform the measures specified in the chapter and then restart the unit Only then may you begin replacement of the other module If the old DX11 is still working Call the Extended Details via SixABCon and check the switching plate configuration for the swivel arm If it deviates from 01 this must be configured again after inserting a new DX11 using service routine 017 7 gt 261 For GALILEOS Comfort If the old DX11 is still working Call up the Extended Details via SiXABCon search for the Language Set ID under Extended Configuration DX7 and note the configuration of the language set If it deviates from 00 the language set must be configured again after inserting a new DX11 using service routine 017 5 gt 258 Following replacement of board DX11 the user preferences patient symbols initial position default contrast mode etc are lost Instruct the user accordingly or set these values after replacing the board provided that they were properly noted down before the board was replaced The connectors on the boards are labeled on delivery of the system Tip Check the designations on the connectors when pulling off the cables and label them correctly if necessary 61 25 665 D3437 D3437 076 01 15 02 06 2012 9 Repair 9 12 Boards Sirona Dental Systems GmbH Service Manua
257. tact to DX11 interrupted during Check the signal path from board DX1 to S gt 343 operation board DX42 replace module if necessary S gt 128 e Check connection of remote control see S gt 141 Installation Instructions e Check the CAN bus Sil ee2h e Check cable L17 replace if necessary ole 134 S gt 343 61 25 665 D3437 D3437 076 01 15 02 06 2012 113 Error code Description Actions required see E7 42 21 No CAN bus connection DX11 does e Check configuration with or without S gt 343 not start DX41 by running service routine S017 9 er S gt 343 A correct the configuration if necessary Occurs in the start screen after S gt 128 power on e Check the signal path from board DX1 to i board DX42 replace module if necessary S gt 259 e Check the CAN bus S Check remote control by running service routine S017 6 configure if necessary e Start the detail query via SiIXABCon If board DX11 responds S gt 362 e Check the signal path to DX42 repair or S gt 343 replace cables connectors if necessary e Replace DX1 If DX11 does not respond S gt 343 e lf error persists Replace board DX11 Error code Description Actions required see E3 42 30 R key pressed during power on e Restart the unit 1 Switch off the unit 2 Wait 1 minute 3 Switch on the unit making sure that the remote control is not pressed during
258. tems GmbH 8 Service Service Manual GALILEOS 8 8 S009 Flash file system S009 Flash file system IMPORTANT The unit has to be completely recalibrated after formatting the flash file system gt 155 When the flash file system is formatted the content of the error logging memory is lost SR SHz Function S009 Flash file system 009 4 Yes Initializing the flash file system 009 5 No Test flash file system 009 7 Yes Save restore DX89 data SR service routine SHZ security access S009 Test step 4 Formatting flash file system 1 Call service routine S009 4 gt 213 2 To initialize the flash file system press the Memory key A R key Easypad or LED above R key Multipad lights up followed by the R key B 61 25 665 D3437 D3437 076 01 15 02 06 2012 241 8 Service Sirona Dental Systems GmbH 8 8 S009 Flash file system Service Manual GALILEOS Flash file system formatting in progress This process takes approx 5 6 mins and is visualized by a progress indicator The end of this process is indicated by the message 0000 in selection field 2 3 Exit the service routine gt 220 Prog 61 25 665 D3437 242 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 8 Service Service Manual GALILEOS 8 8 S009 Flash file system S009 Test step 5 Test flash file system 1 Call service routine S009 5 gt 213
259. th medical electrical equipment Therefore the use of mobile wireless phones in medical office or hospital environments must be prohibited Electrostatic discharge Electrostatic discharge abbreviated ESD ElectroStatic Discharge Electrostatic discharge from people can damage electronic components when the components are touched Touch a ground point to discharge static electricity before touching any boards Switching the unit on Due to the risk of injury caused by malfunction no person may be positioned in the unit when it is switched on 15 2 Safety instructions 2 6 Condensation LASER KLASSE 1 CLASS 1 LASER PRODUCT APPAREIL A LASER DE CLASSE 1 Date of manufacture August 2007 _ _ _ _ S EE Complies with 21 CFR 1040 10 and 1040 11 except for deviations pursant to Laser Notice No 50 dated June 24 2007 Sirona Dental Systems GmbH Fabrikstr 31 D 64625 Bensheim Germany Manufactured in Bensheim 16 Sirona Dental Systems GmbH Service Manual GALILEOS Condensation Extreme fluctuations of temperature may cause condensation inside the unit Do not switch the unit on before it has reached normal room temperature See also Technical Data Laser light localizer The system incorporates Class 1 laser products A minimum distance of 10 cm 4 is required between the eye and the laser Do not stare into the beam Do not use the system with any other lasers and do not make
260. the Software update button The dialog box for entering the service password opens Please enter the service password tad 5 Enter the service password Enter the first 4 digits of the current system date in reverse order as the service password e g on 05 24 1995 5042 must be entered as the service password If an incorrect service password or no password at all is entered the limited update menu for users will be started This only supports an automatic update option The dialog box for selecting the installation source opens Select installation source Rad 6 Click on the button with the 3 dots PR EE imagefile is the default installation source for the software update Image file CACB_UPDATE_IMAGE_V03_03_01 SUI BI corea 61 25 665 D3437 D3437 076 01 15 02 06 2012 61 4 General operating procedures 4 2 Updating the firmware 62 Sirona Dental Systems GmbH Service Manual GALILEOS 7 Select the path or the desired update file from the list and click Open to confirm the selection The update file is located on the unit software CD It is delivered with each DX11 replacement board and also included in the country set The contents of the CD can be downloaded from the Dealer domain of the SIRONA Internet home page under Product Info X ray Systems www sirona com 8 Click on the OK button The software manager opens The left hand window of the software manager displays the modules and their cur
261. the PC 5 Switch GALILEOS on 6 Wait until the green LED of the Facescan status display lights up The Facescan is now ready for operation Always switch the device on before inserting the USB stick Before the Facescan USB stick can be inserted into the USB socket of the FACESCAN modular board GALILEOS must be switched on Otherwise the Facescan configuration data will be reset to the factory settings 7 Insert the Facescan USB stick into the USB port of the FACESCAN board The Facescan will be configured Both LEDs of the status display light up 8 Wait until the light on both the LEDs of the status display goes out process lasts some 10 secs NOTICE Faults during configuration If there has been an error during configuration only the blue LED of the status display goes out Starting the device again is not necessary here gt Inthe event of an error check the log file facescan_settings log on the USB stick gt Make sure to read section Syntax of the configuration file facescan_settings cfg gt 74 gt Repeat the configuration process 9 IMPORTANT Remove the Facescan USB stick from the USB socket 10 Perform a restart of the device The Facescan configuration is complete 61 25 665 D3437 72 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 4 General operating procedures Service Manual GALILEOS 4 3 Configuring the unit Checking the configuration 1 Inse
262. the board DX1 4 Lay the fiber optic cables up to point B and clip the radius limiter A at point B approx 900 mm from the connectors on DX1 onto the cable 5 Guide the fiber optic cable to board DX6 and plug the connector of the new fiber optic cable to the same color assignment on board DX6 61 25 665 D3437 D3437 076 01 15 02 06 2012 363 9 Repair Sirona Dental Systems GmbH 9 13 Cable Service Manual GALILEOS Cable replacement L3 L5 L6 and L15 Laying the cable corrugated tube at the rotation unit The connectors and cables must be protected by inserting them in the fabric tube A supplied with the cables Prepare the cable exchange 1 Remove the covers 2 Pull the connectors off board DX6 Replacing cables 1 For L3 Remove the connector of the cable with tool W1 2 Remove the corrugated tube and the spiral spring B from the cable loom 3 Remove the defective cable and run the new cable up to the rotary ring in the original position 4 Bunch the cables together again to form a loom 5 Fasten the defective cable to the loom and use it as a pull wire to pull the loom through the fabric tube A 6 Pull the fabric tube over the connector and as far over the cable loom as possible 7 Use the pull wire to pull the fabric tube into the spiral spring B 61 25 665 D3437 364 D3437 076 01 15 02 06 2012 TA 6 TULL Slide the corrugated tube over the spiral spring
263. the patient symbol keys light up E H approx 2 s Then press the patient symbol keys in the sequence F H E within the next 4 s After you have entered the key combination correctly the service menu is displayed You can return to the next higher level with the double arrow key I at any time 61 25 665 D3437 D3437 076 01 15 02 06 2012 211 8 Service Sirona Dental Systems GmbH 8 3 Basic operating procedures in the service menu Service Manual GALILEOS Multipad 1 Press and hold down Service key D until the patient symbol keys light up E H approx 2 s 96 2 Then press the patient symbol keys in the sequence F H E within the next 4 s After you have entered the key combination correctly the service menu is displayed You can return to the next higher level with the arrow key above selection field 3 I at any time 61 25 665 D3437 212 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS x No 005 Z S5 61 25 665 D3437 D3437 076 01 15 02 06 2012 8 Service 8 3 Basic operating procedures in the service menu Selecting service routines and test steps Selecting a service routine V The service menu must be selected gt 210 gt Select the desired service routine via the arrow keys in the selection field 1 A and confirm the selection via the service key B If the
264. the unit is ready for selection e Acknowledge error and quit service domain on unit and in SIDEXIS XG Restart the unit Switch off the unit Wait 1 minute Switch unit on Repeat procedure and check function PON gt the error occurs repeatedly e Perform network diagnosis with the support of the Sirona Customer Service Center CSC and check the setting of the network card if necessary again seeking assistance from the Sirona Customer Service Center Checksum offload for patient names with 15 characters with several network cards preferably for onboard systems e Check and if necessary replace network components PC network card Cat5 cable hub switch router media converter L25 26 e Check the software versions of the unit on the info screen or by running service routine S008 2 and XG3D_plugin and perform software update if necessary S gt 238 S gt 61 Error code Description Actions required see E6 14 05 Service of DHCP server is not available e Have network configuration of dental practice checked by the administrator in charge e Ensure proper functioning of the DHCP server Error code Description Actions required see E6 14 06 The bootline of board DX11 had to be preassigned with default values e Reconfiguration of network data via SiXABCon required 104 61 25 665 D3437 D3437 076 01 15 02 06 201
265. third party manufacturer unit may have been connected Error code Description Actions required see E7 1115 A diaphragm not suitable for the e Run service routine S017 25 to modify the S gt 270 diaphragm configuration was diaphragm configuration as appropriate detected by the unit for the installed diaphragm 61 25 665 D3437 102 D3437 076 01 15 02 06 2012 Error code Description Actions required see E111 19 No image data available e Check TDI signal synchronized readout S gt 362 sequence cable L13 replace cable L13 if necessary e Replace board DX89 S gt 343 e Replace board DX1 S gt 343 Error code Description Actions required see E1 11 20 The calibration data on the unit is e Calibrate the unit S gt 155 invalid or does not match the serial if the error occurs again and no modules were S gt 343 numbers of the modules replaced e Replace board DX11 If the error occurs again and modules were replaced e This error is a sequential fault watch for additional causal error messages and take the respective action Error code Description Actions required see E2 11 22 The default iris table is write e Check the software versions of SIDEXIS S gt 61 protected XG and the unit for compatibility perform software update if necessary Error code Description Actions required see E111 23 No matching iris diaphragm setting
266. this way you can prevent the sensitive fiber optic cable from being damaged Push both cables together into the cable track up to the cable markings New cables do not have cable markings Orientate yourself according to the marking on the second old cable and make sure that both cables protrude equally far out of the cable track once they have been drawn in Then make a mark on the new cable 22 NOTICE The cable ties should only fix the position of the cables They must not be tightened too much otherwise fiber optic cable L7 could be damaged Before installing the cable track in the stand fix the cables at both ends of the cable track with a cable tie 23 Reinstall the cable track in the stand Installation of the cable track is performed in reverse order of the removal Replacing cable L1 or grounding strap in cable track 1 The procedure for replacing cable L1 and the grounding strap is basically analogous to the procedure described in chapter Replacing cable L7 L117 or L108 in cable track 2 gt 366 61 25 665 D3437 368 D3437 076 01 15 02 06 2012 10 Maintenance 10 1 Calibrating the unit Sirona Dental Systems GmbH Service Manual GALILEOS Maintenance DANGER Potentially lethal shock hazard It is essential to switch the unit off and to wait at least 1 minute or 4 minutes if disconnecting the tube assembly cable L3 before starting the maintenance or taking off a cover panel
267. time If the error occurs again e Run an error diagnosis gt 88 61 25 665 D3437 D3437 076 01 15 02 06 2012 85 x Description Error group Actions required 3 The system detects that This error group includes all errors Restart the unit a key was pressed that indicate invalid signal states 4 Switch off the unit during power on of keys and safety signals during power on 2 Wait 1 minute 3 Switch unit on 4 Repeat procedure If the error occurs again e Run an error diagnosis gt 88 4 Malfunction or This error group includes all errors Acknowledge the error message and make mechanical obstruction that indicate problems with the sure that the movements of the unit are not of unit movements motor controlled movements on obstructed he Cus cr Qe unit e Repeat the last procedure step or exposure If the error occurs again e Run an error diagnosis gt 88 5 Malfunction during the This error group includes all errors Acknowledge the error message exposure or during resulting from a certain system e Repeat the last procedure step or exposure exposure preparation action triggered by the user which could not be performed because a If the error occurs again required internal partial function e Run an error diagnosis gt 88 software or hardware is not ready or fails 6 Error during system This error group includes all errors Acknowle
268. tion field 192 168 015 179 T Memory key G Easypad or the LED above Memory key G B4 Multipad lights up 7 Save the setting gt 219 8 Exit the service routine gt 220 9 Restart the unit 61 25 665 D3437 D3437 076 01 15 02 06 2012 287 9 Repair Sirona Dental Systems GmbH Service Manual GALILEOS Repair DANGER Potentially lethal shock hazard It is essential to switch the unit off and to wait at least 1 minute or 4 minutes if disconnecting the tube assembly cable L3 before starting the repair or taking off a cover panel When replacing parts in the vicinity of the power connection power switch board DX32 or X ray tube assembly the unit must be disconnected from the junction box of the main building Make sure to reattach all ground cables to ensure correct grounding of all modules Product safety Modifications to this unit which might affect the safety of the system owner patients or other persons are prohibited by law For reasons of product safety this product may be operated only with original Sirona accessories or third party accessories expressly approved by Sirona The user is responsible for any damage resulting from the use of non approved accessories NOTICE Do not damage the cables Be careful not to kink the cables when removing or installing them Take particular care with fiber optic cables L5 L6 L7 and L15 Tighten cable ties only as
269. to find out whether a bug fix by means of a software update is possible and perform such an update if necessary 61 25 665 D3437 112 D3437 076 01 15 02 06 2012 e Check board DX42 replace if necessary e Check board DX41 replace if necessary Tip If the error cannot be eliminated immediately the unit can be temporarily reconfigured and operated with a release button located directly on it see Installation Instructions Error code Description Actions required see E7 42 10 Module is stuck in bootloader stage Check board DX42 note LED states S gt 134 If the board remains in the bootloader stage S gt 61 e Repeat the software update e Replace remote control see Installation Instructions Error code Description Actions required see E7 42 12 Unit is not ready for operation This error is a sequential fault S gt 263 e Restart the unit 1 Switch off the unit 2 Wait 1 minute 3 Switch on the unit making sure that the release button is not pressed during booting 4 Repeat procedure and check function e Check unit configuration with or without DX41 by running service routine S017 9 correct the configuration if necessary If the error occurs repeatedly S gt 343 e Repeat procedure and observe causal error messages e Check the signal path from board DX1 to board DX42 replace module if necessary Error code Description Actions required see E6 42 20 Con
270. tor Pin Signal 1 GND 2 3 3Vor5V 3 Signal 4 GND IMPORTANT When inspecting the light barrier check to see whether ambient light can influence its function Are the signals OK Replace light barrier 139 Device leakage current too high Disconnect tube assembly cable L3 from connector X3 on board DX6 and measure the leakage current Yes Is the leakage current OK Check cable shields and cables for visual damage Are the cable shields and cables OK Replace the tube assembly Correct the cable shields or replace the defective cable Replace board DX32 61 25 665 D3437 140 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 6 Troubleshooting Service Manual GALILEOS 6 7 Checking the cables Checking the cables You can use a standard Cat5 cable as a test cable for L8 up to unit serial number 3201 L10 L12 L40 and L37 This cable must not be permanently installed IMPORTANT Most cables have the same plug at both ends and are connected 1 1 Clean the plug If defective correct the plug replace cable if necessary Is the cable connection of the cable OK Yes For shielded cables Temporarily connect the replacement cable if available and check unit function Correct the shield replace cable if necessary Is the shield support of the cable OK Assign fault to module and replace component
271. tware V 03 05 00 and SIDEXIS 2 3 or above For GALILEOS Compact Requires unit main software V 03 06 02 and SIDEXIS 2 3 or above 013 For GALILEOS Comfort Requires unit main software V 03 05 00 and SIDEXIS 2 3 or above For GALILEOS Compact Requires unit main software V 03 06 02 and SIDEXIS 2 3 or above V 1 7 1 014 For GALILEOS Comfort Requires unit main software V 03 05 00 and SIDEXIS 2 3 or above For GALILEOS Compact Requires unit main software V 03 06 02 and SIDEXIS 2 3 or above V 1 7 1 1 015 For GALILEOS Comfort Requires unit main software V 03 05 00 and SIDEXIS 2 3 or above For GALILEOS Compact Requires unit main software V 03 06 02 and SIDEXIS 2 3 or above 61 25 665 D3437 D3437 076 01 15 02 06 2012 61 25 665 D3437 D3437 076 01 15 02 06 2012 GALILEOS Software CD index Remarks V1 7 2 016 For GALILEOS Comfort Requires unit main software V 03 05 00 and SIDEXIS 2 3 or above For GALILEOS Compact Requires unit main software V 03 06 02 and SIDEXIS 2 3 or above V 1 7 4 002 For GALILEOS Comfort Requires unit main software V 03 05 00 and SIDEXIS 2 3 or above For GALILEOS Compact Requires unit main software V 03 06 02 and SIDEXIS 2 3 or above 001 For GALILEOS Comfort Requires unit main software V 03 05 00 and SIDEXIS 2 5 1 or above For GALILEOS Compact Requires unit main software V 03 06
272. unit parameters etc D Selection field 3 E Arrow keys Touch the and arrow keys to select unit parameters in the selection fields gt 217 F S1 S37 Selected service routine G 1 n Selected test step H Patient symbol keys Different functions depending on service routine l Memory key Save selection J Service key Different functions depending on service routine Most however confirm a selection or the activation of the next test step T est rotation key Start a test L R eturn key Move the unit to the starting position or confirm a save operation M Double arrow key Return to the main menu N Wrench symbol Displayed if level 4 Service menu is activated 207 8 Service Sirona Dental Systems GmbH 8 2 Service menu and service routines Service Manual GALILEOS 8 2 1 2 Multipad LOOO OO OOO 61 25 665 D3437 208 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH Service Manual GALILEOS 8 Service 8 2 Service menu and service routines X ray lamp Caution Radiation is being released Selection field 1 8 digits Selection field 2 4 digits Selection field 3 4 digits Display fields for service routines test steps values unit parameters etc Arrow keys Touch the and arrow keys to select unit parameters in the selection fields Select parameters gt 217 Patient symbol keys Different functions
273. ure 85kV 0 21 mAs gt 162 4 Evaluate the image 61 25 665 D3437 176 D3437 076 01 15 02 06 2012 A Adjustment OK B Adjustment not OK The lower edge of the lower lead diaphragm must be within tolerance i e lie inside of the auxiliary lies A If the edge is out of tolerance B the diaphragm must be adjusted mechanically gt 197 asec tons 7 5 If the exposure is OK A confirm this by clicking the check box located to the right of the Lower jaw button The box will appear checked Diaphragm adjustment is now complete 6 Continue the calibration procedure with the radiation field check gt 178 61 25 665 D3437 D3437 076 01 15 02 06 2012 177 7 Adjusting calibrating the unit Sirona Dental Systems GmbH 7 3 Adjustment and calibration via the Service Functions menu Service Manual GALILEOS Checking the radiation field IMPORTANT The illumination must be checked once the collimator has been adjusted Diaphragm Radiation field Dose Sensor Iris Shading Distortion Geometry 4 r State I Badiation field position corect Image acquisition v The Service functions menu is called gt 161 1 Clip the distortion phantom onto the X ray detector cover gt 163 2 Click the Radiation field tab The corresponding tab card is selected 3 Click on the mage acquisition button SIDEXIS XG makes the unit read
274. ure is activated For GALILEOS Comfort only e The unit language is preconfigured as ordered e The welcome screen is switched on e The first name last name and date of birth lines are displayed on the welcome screen If the customer requires a different configuration this can be implemented via service routine S017 gt 252 61 25 665 D3437 60 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 4 General operating procedures Service Manual GALILEOS 4 2 Updating the firmware Updating the firmware Updating the unit firmware Read the information provided on the software CD supplied with the unit and on the SIRONA dealer page on the Internet very carefully These sources always contain the latest information on software updates 1 Start the S DEX S Manager under Start Programs SIDEXIS SIDEXIS XG 2 Click on Configuration of the X ray components The Configuration of the X ray components menu opens fF Configuration of the X ray components Local approvals Sensors Attributes Print amp bout Name xna a i Add component Ethernet address oovo1s002s1a Change Name IP address 192 168 15 20 ie Delete component Subnet mask 255 255 2550 Details Functional test Standard Gateway fis2t68151 More details Sirona Control Server ENTWICKLUNG X Approved for the computers ep ENTWICKLUNG OK Abbrechen bemehmen 3 Select the Attributes tab 4 Click on
275. usting the touchscreen eee eeeneeeeeeeeeeeeeteneeeeeeeeees 191 61 25 665 D3437 D3437 076 01 15 02 06 2012 7 6 Mechanical agjuStments cs iccrsireteradilecetenndvemeacvedevaara alent suadasadedae beep otepiddes 193 7 6 1 Ring center adjustment 0 s ccpccctecce vt ate canectev nsdn ieteectea eens 193 7 6 2 Adjusting the Swivel arm c cceeesesecccceeeeeeeenescenenceeeesesseeeeneeeeeee 195 7 6 3 DIADACAGIY ACIUSUMONI sed citecturs sherk niente nc tiies erties seul donk deeds das beemataiialdes 197 7 6 3 1 Adjusting the type 1 diaphragm up to serial no 2200 197 7 6 3 2 Adjusting the Type 2 diaphragm serial no 2201 and higher 198 7 6 3 3 Adjusting the Type 3 diaphragm ccccceccseeecnceeeeseeeeeeeeeceenees 200 8 SEIVICG co cc tea enone aie eaeh ys career eevee 202 8 1 Overview of service routines 22 ccccceeeeseeeee cee eeeeeeeseeeeeeeneeeseeneneeeeneeeeeentaes 202 8 1 1 List of all service routines available for Selection eeeeeeeeeees 202 8 1 2 Alphabetical list of service routine functions sssseeesrerrrrerrrereen 204 8 2 Service Menu and service routines cece eee eee ee eeeeee eee e tere teeeeeneaeeeeeeeeeeeaea 206 8 2 1 Displays and symbols in the service MENU cccceeeeeeeeeeeereeeeeeees 206 8 2 1 1 Easypad cupete cca cuslea ti ceeacte desde ia eae EAEE NAE 206 82 12 MUNI PAG a rng a e a aa a A a oia iana tastes
276. vice Functions menu N Adjustment and calibration via the Service Functions menu 7 3 1 Diaphragm image 7 3 1 1 Type 1 Type 2 diaphragm Diaphragm Radiation field Dose Sensor Iris Shading Distortion Geometry 4 gt Diaphragm exposure F Even continuous white edge visible Image acquisition v The Service functions menu is selected 1 Click on the mage acquisition button SIDEXIS XG makes the unit ready for exposure gt 162 Service routine S030 5 is displayed on the control panel 2 Take an exposure 85kV 21 mAs gt 162 3 For GALILEOS Comfort Evaluate the image 61 25 665 D3437 D3437 076 01 15 02 06 2012 167 7 Adjusting calibrating the unit Sirona Dental Systems GmbH 7 3 Adjustment and calibration via the Service Functions menu Service Manual GALILEOS Adjustment OK Adjustment not OK Permissible tolerance 30 pixels 5 pixels The brightness distribution along the border surrounding the image on all sides must be uniform A The distance between the bottom edge and the lowest point in the image should be 30 5 pixels measure with SIDEXIS scale B If the distance between the bottom edge and the lowest point in the image is out of tolerance E or the brightness distribution along the surrounding border is not uniform C the diaphragm must be adjusted mechanically gt 197 4 For GALILEOS Compact Ev
277. wire A Power cable to the unit W Second protective ground wire C Unit 61 25 665 D3437 380 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 10 Maintenance Service Manual GALILEOS 10 8 Checking the protective ground wires Measuring points D and E Board cage DX32 and tube assembly housing D Board cage DX32 E Tube assembly housing 61 25 665 D3437 D3437 076 01 15 02 06 2012 381 10 Maintenance Sirona Dental Systems GmbH 10 9 Checking the device leakage current Service Manual GALILEOS Checking the device leakage current Potentially lethal shock hazard Itis essential to switch the unit off and to wait at least one more 1 minute before beginning the check A high resistance measuring voltage source at line frequency and a measuring circuit compliant with the requirements of IEC 60601 are required Complete test units e g the Bender tester fulfill these requirements Measuring setup for testing the device leakage current a PE 7 1kOhm 0 01 5HF 10kOhm i A Power cable to the unit Voltage source 200 240V 50Hz 60Hz D Voltmeter Ww 61 25 665 D3437 382 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 10 Maintenance Service Manual GALIL
278. x00 0x00 0x00 0x00 0x00 0x00 0x04 0x00 0183048 790 Rx 0x101 0x08 0x00 0x00 0x00 0x00 0x00 0x00 0x00 0x00 0183050 809 Rx 0x201 0x08 0x00 0x00 0x00 0x00 0x00 0x04 0x00 0x00 0183054 791 Rx 0x301 0x08 0x00 0x00 0x00 0x00 0x00 0x03 0x00 0x00 0183056 800 Rx 0x401 0x08 0x00 0x00 0x00 0x00 0x00 0x00 0x00 0x00 0183060 793 Rx 0x481 0x08 0x00 0x00 0x00 0x00 0x00 0x01 0x00 0x00 0183062808 Rx 0x181 0x08 0x00 0x00 0x00 0x00 0x00 0x02 0x00 0x00 0183070 812 Tx 0x001 0x08 0x18 0x27 0x09 0x19 0x01 Ox6f 0x00 0x00 0183072 794 Tx 0x080 0x08 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0183074 806 Tx 0x100 0x08 0x55 0x55 0x55 0x55 0x55 0x55 0x55 0x55 The CAN bus protocol is displayed in the browser and can be saved as an HTML page printed out or sent to the Sirona Customer Service Center CSC 61 25 665 D3437 D3437 076 01 15 02 06 2012 237 8 Service Sirona Dental Systems GmbH 8 7 S008 Update service Service Manual GALILEOS S008 Update service SR SHZ Function S008 Checking the software versions 008 2 No Overview of the module software versions 008 3 No Input confirm query unit serial number SR service routine SHZ security access S008 Test step 2 Overview of module software versions Easypad touchscreen 1 Call service routine S008 2 gt 213 The software versions currently installed on the modules are displayed on an info screen on the touchscreen display Systemsoftware V03 03 01 2 Exit the service r
279. xposure signal Activate these functions if they are required 15 If the travel height of the unit has to be limited Set the travel height with service routine S018 2 gt 272 16 Perform a software update to the current software version gt 61 This updates all modules in accordance with the configuration 17 Perform a complete unit calibration gt 155 After a successful unit calibration has been performed the error message should no longer appear 18 Call up the Extended Details via SixABCon This generates an XML file with the system parameters which is Filed in the PDATA P2K_Config directory under the network name of the unit e The process is completed 61 25 665 D3437 D3437 076 01 15 02 06 2012 353 9 Repair 9 12 Boards Case C 354 Sirona Dental Systems GmbH Service Manual GALILEOS New DX11 GALILEOS Compact For DX11 System software version V03 03 01 or higher For DX1V2 System software version V04 04 00 or higher After a new DX11 is inserted the IP address is initially reset to the factory setting Before you set the unit to a new IP address make sure that the IP address you re assigning has not been assigned to any other unit 1 2 10 11 12 13 14 Switch the unit on Do not acknowledge any error messages at this time Install the current version of the SIDEXIS XG software V2 3 or higher If the current version of SIDEXIS XG is a patch version th
280. y for exposure gt 162 Service routine S002 6 is displayed on the control panel 4 Press the R key to move the unit back to the starting position 5 Press the release button Hold down the release button and observe the distortion phantom The lighting strips on the distortion phantom A must not light up If the strips on the phantom light up at all the system is overexposed and you cannot continue the adjustment In this case repeat the diaphragm adjustment procedure and then check the radiation beam field again If the lighting strips still light up during the re check of the beam field contact the SIRONA Customer Service Center CSC to solve the problem 6 To confirm that the lighting strips on the distortion phantom are notlit click the check box underneath the message window The box will appear checked The radiation field check is now complete 7 Continue the calibration procedure with the dosimetry gt 179 Tip Leave the distortion phantom on the unit for the next calibration step 61 25 665 D3437 178 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 7 Adjusting calibrating the unit Service Manual GALILEOS 7 3 Adjustment and calibration via the Service Functions menu 7 3 3 Dosimetry A dosimeter for pulsed radiation e g Mult O Meter 512L is required for dosimetry ice Diaphragm Radiation field Dose Sensor Iris Shading Distortion G
281. you need the following documents Spare parts list e GALILEOS List of Spare Parts Order no 61 25 699 Wiring diagrams e GALILEOS Wiring References Order no 61 25 640 Installation Instructions e GALILEOS Comfort Order no 61 25 574 e GALILEOS ORTHOPHOS XG 3D Software installation Order no 61 42 389 e GALAXIS Operator s Manual Order no 61 23 488 e SIDEXIS XG Digital Radiography Installation Instructions Order no 59 67 356 You can order the technical documentation in paper form free of change from our Customer Service Center by specifying the above mentioned order numbers REF A large portion of the technical documentation is also available our the product DVD To call up the instructions insert the DVD in the DVD drive of the PC The DVD then starts automatically and a start screen opens The latest documentation can always be downloaded from the Sirona homepage www sirona de HOME Service gt Technical Documentation 61 25 665 D3437 D3437 076 01 15 02 06 2012 Sirona Dental Systems GmbH 1 About this Service Manual Service Manual GALILEOS 1 3 Tools and auxiliary materials Tools and auxiliary materials e GALILEOS service set Order No 6146562 e Screwdriver set slot and Phillips e Torx offset screwdrivers TX10 TX20 TX25 included in the scope of supply e Hexagon socket head screwdriver hexagon socket head screw size 6mm included in the scope of supply Open end wrench 13 mm A F
282. ystems GmbH 7 Adjusting calibrating the unit Service Manual GALILEOS 7 3 Adjustment and calibration via the Service Functions menu Service You can create a test image with the Service tab card It is not necessary to execute this menu for the calibration of the unit Radiation field Dose Sensor Iris Shading Distortion Geometry Service 4 gt ee i Jiaphragm field Image acquisition v The Service functions menu is called gt 161 1 Click the Service tab The corresponding tab card is selected 2 Click on the mage acquisition button SIDEXIS XG makes the unit ready for exposure gt 162 Service routine S032 41 is displayed on the control panel 3 Press the R key to move the unit back to the starting position 4 Press the release button Press and hold down the button until the exposure is complete the preview image is displayed in the exposure window and the acoustic signal indicating the end of the exposure double beep sounds if it has been configured The CCD test image is displayed 61 25 665 D3437 D3437 076 01 15 02 06 2012 189 7 Adjusting calibrating the unit 7 4 Performing a white balance for Facescan Sirona Dental Systems GmbH Service Manual GALILEOS 190 Performing a white balance for Facescan IMPORTANT Constant lighting conditions If the lighting conditions during the white balance process are different f

Download Pdf Manuals

image

Related Search

Related Contents

Prüf- und Wartungsanleitung für die Messgeräte - ABC  Istruzioni- per l`uso  Ma carte de séjour ? - Association des Revues Plurielles  radio remote controls manual K Series    Glacier Bay 67092-6002 Installation Guide  グラビティスタンド 2 ! !  Samsung Galaxy S6 edge+ Užívateľská príručka  Q-Logic MKII-BASE16 User's Manual  