1.          HEWLETT PACKARD    HP 3478A Digital Multimeter    Service Manual          478A    U           SECTION Ill  OPERATION    3 1  INTRODUCTION    3 2  This section has the information and instructions  for the operation of the 3478A Multimeter  showing  front panel and remote operation  The information is  an abbreviated description of the operation and is writ   ten for a Service Trained Person  rather than an  Operator  For more complete operating instructions   refer to the 3478A   s Operators Manual  To familiarize  yourself with the front and rear panel features or for a  review of instrument operations  refer to Figure 3 1     3 3  The Operation Section is separated into the follow   ing major areas  General Information  Front Panel  Operation  Shifted Operation  Miscellaneous Opera   tion  and Remote Operation  HP IB   It is suggested  that you read the Remote Operation Chapters last  since  you need to know the other operations to understand  the remote operations  The major areas are as follows     a  General Information   paragraph 3 4     Title Paragraph  AC Power Operation 3 5  Turn On 3 7  Display 3 10  Input Terminals 3 12  Input Terminals Cleaning 3 15    b  Front Panel Operation   paragraph 3 17     Title Paragraph  DC Volts Measurements 3 18  AC Volts Measurements 3 22  Resistance Measurements 3 26  DC Current Measurements 3 30  AC Current Measurements 3 34  Ranging 3 38  Triggering 3 43  c  Shifted Operation   paragraph 3 50   Title Paragraph  General 3 
2.     3 22  AC Volts Measurements    3 23  The 3478A uses a True RMS AC to DC Converter  to measure ac voltages from 1uV to 300V in four ranges   300mV  3V  30V  and 300V  The response of the con   verter is from 20Hz to 100KHz on all ranges  to 300KHz  on the 30V Range only   All ranges are protected from  input voltages up to 450V peak  Select the AC Volts  Function for ac volts measurements by pressing the  v  button     3 24  The AC to DC Converter changes the ac input  voltage to dc volts  which is then measured by the  3478A   s A D Converter  All ranging is done in the AC  to DC Converter  applying the same voltage value to the  A D Converter for all full scale inputs     3 25  When the AC Volts Function is selected  the right  side of the display shows       MVAC    or    VAC     depen   dent on the range selected  The reading can be displayed  either as a 54  4   or 3  digit reading  dependent on  the selected Number of Digits Displayed  see paragraph  3 54   A  6 second delay is also applied in the ACV  Function before a reading is taken  also  during a range  change   Refer to Table 1 1 or Table 4 1 for the AC  Volts Function   s accuracy specifications     3 26  Resistance Measurements    3 27  The 3478A is able to make resistance  measurements from  0001 ohms to 30M ohms in seven  ranges  The ranges extend from the 30 ohm to the 30M  ohm range  All ranges are protected from input voltages  up to 350V peak  Resistance measurements can be made  using either the 2 Wir
3.   display can also show alphanumeric messages  sent  remotely   The four characters to the right show the  function  and to a certain extent  the range  and the 8  characters to the left show the reading  e g    12 3657  MVDC   An    OVLD    is displayed if the input is out of  range for the selected range and function     3 12  Input Terminals    3 13  The 3478A has one set of Input Terminals on the  front panel and one set on the rear panel  The front  panel terminals consist of an    INPUT HI        INPUT  LO        Q SENSE HI     Q SENSE LO     and    A      Amps  terminal  Except for the       A       terminal  the rear  panel has the same set of terminals  The two sets of ter   minals are selected by the Front Rear Switch  located  on the front panel   The front terminals are selected  with the switch    out    and the rear with the switch     in        3 14  The INPUT HI and LO Terminals are used for  measuring dc volts  ac volts  and resistance in the  2 Wire Ohms configuration  Refer to Figure 3 2 for a  typical connection  The Q SENSE HI and Q SENSE LO  Terminals  in conjunction with the INPUT Terminals   are used in the 4 Wire Ohms configuration  Refer to  Figure 3 4 for a typical ohms connection  The A  Amps   Terminal with the INPUT LO Terminal is used to          N SENSE INPUT    OO  OO    3A 250Y                               Figure 3 2  Typical Input Measurement Connection          N SENSE INPUT   4 WIRED  2 WIRE              Figure 3 3  Typical Current Measur
4.  2  or 3 1 2 digit reading   dependent on the selected Number Of Digits Displayed   see paragraph 3 54   Refer to Table 1 1 or Table 4 1 for  the AC Current Function   s accuracy specifications     3 38  Ranging    3 39  The 3478A has two range modes  Manual and  Autorange  Manual ranging is selected by pressing the  AUTO MAN button  if the 3478A is in Autorange  or  by pressing either the 4 gt  or X gt  buttons  The    M  RNG    annunciator on the display then turns on  The  following explains the different range modes     3 40  Uprange  The 3478A upranges to the next higher  range each time the   gt  button is pressed  The highest  selectable range depends on the function selected  e g   300V for the AC Volts and DC Volts Functions   If a  function is selected with its highest range lower than the  previous range  the multimeter defaults to the new  highest range     3 41  Downrange  The 3478A downranges to the next  lower range each time the X  gt  button is pressed  The  lowest selectable range depends on the function selected   e g  30mV for the DC Volts Function   If a function is  selected with its lowest range higher than the previous  function  the multimeter defaults to the new lowest  range     3 42  Autorange  The 3478A selects the optimum range  when this mode is selected  The mode is selected when  the 3478A is first turned on or by pressing the  AUTO MAN button  if in Manual Range   If  Autorange is enabled  the 3478A upranges when the  reading is at or above  
5.  3478A Addressing    3 93  HP IB requires that each device on the Bus needs  to be identified as a Listener or a Talker  in order to ex   ecute the Bus Messages and commands  Because of this   each device has its own unique    listen       and       talk       ad   dress  The address of the 3478A is set by the Address  Switch on its rear panel  Setting the 3478A   s Listen Ad   dress also sets its Talk Address     3 94  The address switch is an eight section       DIP        switch with five sections used for addressing  The switch  is shown in Table 3 5  The allowable address settings are  also listed in Table 3 5  The factory address setting of  the 3478A is decimal 23  refer to Section II of this  manual for the switch setting      3 95  Instrument address commands  sent by the con   troller  are usually in this form  universal unlisten   device talk  device listen  The universal unlisten com   mand removes all listeners from the Bus to allow only  the addressed listener s  to receive data  The data is sent  by a talker which is designated by the device talk com   mand     3 96  Talk Only Mode  No Controller     3 97  The 3478A   s Talk Only Mode allows the  multimeter to send measurement data to an external  device  like a printer  without a Bus controller  The  multimeter is placed into the Talk Only Mode by setting  the five address switches  on the rear panel       DIP        switch  to 1  set only the five address switches to the up  position   Measurement data is the
6.  dd       is the address code  of the instrument     3 9  Once the Self Test is completed  the 3478A goes to  its turn on state which is     FUNCTION 2 2052    a aces DCV  RANGE 2 2202  geet a AUTO  TRIGGER                 INTERNAL  AUTOZERO               2  005 ON    NUMBER OF DIGITS DISPLAYED   5  3 10  Display    3 11  The Display is a 12 character alphanumeric Liq   uid Crystal Display  LCD  with 12 annunciators  The  display is normally used to show readings  however  the    3 1       Operation       3478A                   3478A DIGITAL MULTIMETER  HEWLETT    PACKARD    N SENSE INPUT                4 WIRE   2 WIRE              ES                     Eiza be    gt  NGEZN I     Ea  a  I   ag  HAN    AUTO  a    TEST   gs       4 WIRE    S    EA    Gates Lak  Ce  be        Al      ALL TERM   500V     HAX  al                             WARNING  FoR  PROTECTION FROM  ELECTRICAL SHOCK  POWER CORD GROUND    N SENSE   4 WIRE     INPUT   2 WIRE     OMO     IVT    Oe    HEWLETT PACKARD CO   MADE IN USA    MUST NOT BE DEFEATED     BT TRI G       Vv oe    un                      2619A34899      ALL TERM    5 0Vpk  MAX  poe i       1    V 120V 220V 240V  O oO    jei       eige4el  leone  1 5  Hz  6 604z       250mAT         125mAT                Use the measurement keys along the top row to select the  type of measurement you want to make        The 12 character alphanumeric display includes 12 dedicated  annunciators  The display is read directly in engineering units   i e   MV for
7.  milli volts  MOHM for Meg ohm resistance  etc        If the blue key is pressed before another key  the function  shown above that key is executed     These terminals are used for the voltmeter sense leads when  making 4 wire ohms measurements        These terminals are the voltmeter and 2 wire ohms input   4 wire ohms source current terminals    This is the Amps input terminal and is used with the INPUT  LOW terminal   3 Amp fuse protected     Oo  When this switch is out  the front panel input terminals are  selected  If the switch is in  the rear panel terminals are used              3478A 3 1   These keys are used to select special operating features of the  3478A  The blue shift key allows for selecting the       SHIFTED        functions of the bottom row of keys     The Internal Trigger is selected by this key  In this mode  the  3478A triggers itself     Autozero is a function that allows you to enable or disable the in   ternal zero correction circuitry     Pressing the Single Trigger key causes the 3478A to take one  reading and wait for the next trigger impulse  This impulse can  come from either the Single Trigger key or the External Trigger  input  rear panel BNC      The Test Reset key performs an internal self test  then resets the  3478A to its turn on state  Any errors in the seif test are noted  in the display     The front panel SRQ  Service Request  key is an HP IB opera   tion that enables you to manually interrupt the controller  This  key is enabled by s
8. 303099 and downranges at or  below  027000  delete one or two zeroes from the  numbers for the 4 1 2 and 3 1 2 Digit modes  respec   tively   The numeric range points are irrespective of  decimal placement  Refer to Figure 3 5 for the  autorange points  the example is for the DCV Function   other functions are similar      3 43  Triggering    3 44  The 3478A has three local trigger modes  Inter   nal  Single  and External  In addition to these  a Hold    3 5          Operation    a  aM    1 N  1 1  DOWNRANGE POINTS    i 1  i 1  i 1  1 D  1  1       yn  H     L     i  i  i  H  H  i  i       INPUT LEVEL 3478 3 5    Figure 3 5  Autoranging Points    and a Fast trigger mode can be selected  but only over  the HP IB  remote operation   When the multimeter is  triggered  the right most digit on the display blinks   showing that the display is updated   The following  paragraphs explain the trigger modes     3 45  Internal Trigger  In this mode  the measurement  cycle is internally initiated and the 3478A makes the  measurements at the maximum reading rate  The Inter   nal Trigger is selected at instrument turn on or by press   ing the INT TRIG button     3 46  Single Trigger  In this mode  a measurement cycle  is initiated each time the SGL TRIG button is pressed   When the button is initially depressed  the 3478A in   itiates a measurement cycle and then places the  multimeter in the Single Trigger Mode  If the button is  pressed during a measurement cycle  while in the Single  Tr
9. 51  Number Of Digits Displayed 3 54  Autozero 3 57  Self Test Reset Operation 3 60  Calibration 3 63  d  Miscellaneous Operation   paragraph 3 66   Title Paragraph  Voltmeter Complete 3 67    e  Remote Operation   paragraph 3 69     Title Paragraph   General 3 70  3478A Response to Bus   Messages 3 74  3478A Addressing 3 92  Talk Only Mode  No   Controller  3 96  3478A HP IB Programming 3 98  Advanced programming 3 117    3 4  GENERAL INFORMATION    3 5  AC Power Operation    3 6  Before connecting ac power to the 3478A  make  sure the power source matches the power requirement of  the multimeter  as marked on the rear panel   If the in   strument is incompatible with the power source  refer to  Section II of this manual for power requirement  modification     3 7  Turn On    3 8  When the 3478A is first turned on  the multimeter  goes through an Internal Test routine  During the test   the instrument displays    SELF TEST    and  when the test  is completed and it passes  the instrument displays       SELF  TEST OK     If the Self Test fails  refer to Section VII  of this manual for troubleshooting information  For more  information on the Self Test  refer to paragraph 3 60     NOTE    For 3478As with serial numbers 2301A03760  and below or any instruments that have the  Revision B ROM  U502  installed  the instru   ment displays its remote  HP IB  address for  about one second instead of    SELF TEST  OK     The address is displayed as       HPIB  ADRS dd        where      
10. Register as a true condition   They will not cause the SRQ Message to be output  For  example  suppose the Front Panel SRQ condition is the  only one set in the SRQ Mask  If the Data Ready condi   tion is true  but not Front Panel SRQ  no SRQ Message    7  e  s  oa Jus    2   br  oo     Power On u     Service Request Bit    Invalid Calibration  Front Panel SRO    Data Ready   Bit Always O  Syntax Error  Internal Error       Figure 3 6  Status Byte          Operation    Table 3 4  Status Byte Definitions    Octal Decimal Definitions  Code Code    Data Ready   Indicates to the con   troller that measurement data is  ready to be output  The Require Ser   vice Message and bit is cleared when  the controller begins to accept the  data or when the reading is no longer  available  See paragraph 3 114               This bit is always at O  O high               Syntax Error   This shows that an in   valid Program Code s  has been sent  to the 3478A  e g  F9               Internal Error   Shows that a failure  inthe 3478A is detected  This may  be a failure in the Self Test Routine   see paragraph 3 60   the A D Con   verter  or a checksum error in the  Calibration RAM  checked every  time a reading is taken   More infor   mation can be obtained by reading  the 3478A   s error register  See  paragraph 3 120                    Front Panel SRQ   This bit is set  when the 3478A   s Front Panel SRO  button is pressed  See paragraph  3 115               Invalid Calibration   When this bit i
11. automatically set  except  bit 6  when any corresponding require service condition  is true  For example  you wish to know if the multimeter  has received an incorrect program code  Syntax Error   or has an internal failure  Internal Error   If the condi   tions are true  bits 2 and 3 of the Status Byte are set  but  no Require Service Message is output  The message will  not be output until the SRQ Mask is set to the cor   responding bits of the Status Byte  which are are bits 2  and 3   Since the mask is to be set in octal  the resultant  code for the bits is    14    The mask is set by sending        Mdd        where dd is the octal code for the bits  This is  shown in the following example       U 7 M14    universal unlisten command   set mask  controller       talk       address  where 14 is the  3478A       listen       address octal code for bits  2 and 3    3 104  Only bits 0 to 5 can be set by programming the  SRQ Mask  Bit 6 of the Status Byte is set whenever the  Require Service Message is output and bit 7 is set by the  Power On SRQ switch on the rear panel  see paragraph  3 106   Because of this  only a two digit octal code  00    3 11          Operation    Table 3 6  3478A Program Codes    Program Description  Code    DC Volts   AC Volts   2 Wire Ohms   4 Wire Ohms   DC Current   AC Current   Extended Ohms  see paragraph  3 118                   FUNCTION       most sensitive range on any func   tion   30mV DC  or most sensitive range  on any other function   300
12. case letters generate  characters which do not resemble the letters  Commas   periods  and semicolons can go between characters  The  ASCII characters can be sent in two different modes   The following paragraphs explain the two modes     3 108  In one display mode  the display is continuously  updated  This is enabled by program code       D2        e g      D23478A DMM    where    3478A DMM    is the  message displayed   In this mode  the annunciators con   tinue to be updated  If more than 12 characters are sent  to the display  the extra characters are ignored until a  control character is received  If the control character is  different than HT  Horizontal Tab   VT  Vertical Tab    LF  Line Field   CR  Carriage Return   or FF  Form  Feed   a Syntax Error is generated  A       D2       code locks  the display until a       D1        Selects Normal Display  is  sent  a CLEAR message is sent  an error condition oc   curs  or a front panel button is pressed     3 109  The other display mode is selected by sending  code       D3        This mode is the same as the previous  mode  except the display is not updated and all annun   ciators are turned off  The text on the display remains  on for about 10 minutes and then the display goes  blank  The display can be restored by sending any of the  display codes  D1  D2  or D3  or by pressing the  LOCAL button  if not disabled by the LOCAL  LOCKOUT Message      3 110  Home Commands  The Home Commands   when sent to the 3478A over th
13. e 4 1 2 or 3 1 2 Digit  mode  the 5th and or 6th digits will be output as zeroes   An overload condition  whether plus or minus  will be  output as   9 99999E  9  If a different output is re   quested  Binary  Front Rear Switch position  etc    the  other output supercedes the reading  If a data transfer is  interrupted while being output  the 3478A continues the  output wherever it left off  when addressed again  This  partial output  or any output  can be disabled by a  Group Execute Trigger  GET   Clear Message  DCL or  SDC   sending any valid program code  or pressing the  LOCAL  Shift  and TEST RESET buttons on the front  panel     3 113  Front Rear Switch Position  The 3478A   s  Front Rear Switch position can be remotely determined  by sending program code       S       to the multimeter and  then reading its output  If       0        CR LF  is output the  switch is set to Rear  and if       1        CR LF  is output the  switch is set to Front     3 114  Data Ready Feature  The Data Ready feature of  the 3478A  when enabled  outputs a Require Service  Message  SRQ  after each completed measurement cy   cle  Before the message can be output  bit 0 of the SRQ  Mask must be set  This is done by sending program code           MO1        bit 0 of the Status Byte and Status Register    When the Require Service Message is sent  the front  panel       SRQ          annunciator turns on and bit 0 of the  Status Byte is set  The SRQ condition remains until the  data is read by t
14. e HP IB  are used to set  the 3478A into a predefined operating state  The pro   gram code is       Hn        where n is the number which                     3478A  defines the state  The following lists the home com   mands and the corresponding operating states     a  HO Command  The instrument state for this com   mand is as follows     FUNCTION         c cece eee eee nn DCV  F1   Rangere ens 30mV  Auto  R 2RA   Trigger    Hold  T4   Number of Digits Displayed     4 1 2  N4   AULOZETO    ccc cece cece eee een On  Z1     In addition  the 3478A   s External Trigger is disabled  and the instrument is set into the lowest range  R 2   before Autorange is selected  Also  any data  reading   output to the Bus or the front panel is erased when the  home command is sent  The corresponding program  codes of the instrument state are        FIT4R 2RAZIN4           b  Hl Command  DCV   This command is the same  as the       HO      command  except a trigger occurs and the  resultant output can be read  The corresponding pro   gram codes are        F1R 2RAZ1N4T3           c  H2 Command  ACV   The same as the    H1     command  except the selected function is AC Volts  The  corresponding program codes are      F2R 2RAZINAT3        d  H3 Command  2 Wire Ohms   Same as the       H1        command  except the selected function is 2 Wire Ohms   The corresponding program codes are      F3R 2RAZIN4T3           e  H4 Command  4 Wire Ohms   Same as the    H1     command  except the selected func
15. e or 4 Wire ohms configuration   which are selected by the 2 WIRE Q and 4 WIRE Q but   tons  respectively  Refer to Figure 3 4 for the correct  ohms connections     3 28  Resistance measurements are made by applying a  known current  generated by the 3478A  to the  unknown resistance  The resultant voltage drop is then  measured by the 3478A   s dc circuitry  Input Circuitry  and A D Converter   In the 2 Wire Ohms Function  the  voltage drop is measured across the HI and LO INPUT  Terminals  In the 4 Wire Ohms Function  the voltage is  measured across the HI and LO Q SENSE Terminals   The Ohms Current Source generates the known current  which is applied to the unknown resistance  in both    3 4       3478A          UNKNOWN  RESISTANCE  Rx          3A 250V    i FUSE    2 WIRE OHMS MEASUREMENT                UNKNOWN  RESISTANCE  Rx                      4 WIRE OHMS MEASUREMENT       Figure 3 4  Ohms Connections    ohms functions  from the HI INPUT Terminal  The  current values are as follows     Range Current  30 ohm 1mA  300 ohm ImA  3K ohm ImA  30K ohm 100uA  300K ohm 10uA  3M ohm 1uA  30M ohm  luA    3 29  When an Ohms Function is selected  the right side  of the display shows either    OHM           KOHM           or     MOHM           dependent on the range selected  In addi   tion  the       2 Q    annunciator is on for the 2 Wire Ohms  Function and the    4 Q    annunciator is on for the  4 Wire Ohms Function  The ohms reading can be  displayed either as a 5 1 2  4 1 2  o
16. ement Connection    measure ac or dc current  Refer to Figure 3 3 for a  typical current connection     3 15  Input Terminals Cleaning    3 16  The high input impedance of the 3478A requires  that the area surrounding the multimeter   s Input Ter   minals  front or rear  must be free of leakage causing  paths  e g  dirt  fingerprints  etc    The paths can be  removed by using a soft cotton swab dipped in isopropyl  alcohol     3 17  FRONT PANEL OPERATION  3 18  DC Volts Measurements    3 19  The 3478A is able to make dc volts measurements  from  luV to 300V in five ranges  30mV  300mV  3V   30V  and 300V  All ranges are protected from input  voltages up to 450V peak  Select the DCV Function for  dc volts measurements by pressing the  v button     3 20  In the DC Volts Function  ranging is done in the  Input Circuitry of the 3478A  The result is that the input  to the A D Converter  which changes the voltage to  digital information  has the same value in all ranges for    3 3    Operation    all full scale inputs  e g  10V input to the A D for  30mV  300mV  3V  etc  inputs to the 3478A      3 21  When the DC Volts Function is selected  the right  side of the display shows       MVDC    or    VDC     depen   dent on the range selected  The reading can be displayed  either as a 5 1 2  4 1 2  or 3 1 2 digit reading  depen   dent on the selected Number Of Digits Displayed  see  paragraph 3 54   Refer to Table 1 1 or Table 4 1 for the  DC Volts Function   s accuracy specifications 
17. erations  shifted and unshifted  An unshifted  operation  AUTO MAN  INT TRIG  etc   is normally  selected by pressing a single button  A shifted operation  is done by first pressing the blue       SHIFT    button  on  the right end of the upper row of buttons  and then  pressing one button  on the lower row   To select a new  shifted operation  press the blue Shift button again and  the button for the new operation  Table 3 1 lists the  shifted operations and corresponding buttons  The  shifted operations are also shown in blue lettering above  the buttons     3 53  When the blue Shift button is pressed  the     SHIFT    annunciator on the display is on  The annun   ciator remains on until a different button is pressed     Table 3 1  Shifted Operations    Shifted Select Description Of  Operation Button Operation    3 Digit  Disp  AUTO MAN   Selects 3 Digits Displayed   see paragraph 3 54    AN Selects 4 Digits Displayed   see paragraph 3 54    Selects 5 Digits Displayed   see paragraph 3 54    Turns Autozero on or off   see paragraph 3 57    Places the 3478A into its  Internal Test Mode  see  paragraph 3 61    Displays the 3478A   s cur   rent HP IB Address Code   see paragraph 3 92    Places the 3478A into the  calibration mode  see Sec   tion IV of this manual      4 Digit  Disp     5 Digit  Disp  Sr    Az INT TRIG    Test Reset SGL TRIG    Adrs SRQ    Cal LOCAL    3 54  Number Of Digits Displayed    3 55  The 3478A can display readings in either 5 1 2  4  1 2  or 3 1 2 dig
18. etting the SRQ mask        Figure 3 1  3478A Front and Rear Panel Features    3 2             3478A          3478A    The 3478A HP IB Address is displayed when the Address key is  pressed     The Local key returns the 3478A to front panel control from the  REMOTE mode     The 3478A features total electronic calibration  The Cal key is  used as part of that procedure  See Section IV      8  The range keys are used to select the proper range for the  measurement  Press any of the keys to select the manual  range mode  Note the M RNG annunciator in the display  The  Auto Man key will return the meter to autorange      10  The    SHIFTED    functions of the range keys are used to select  alternate numbers of display digits        Operation    The Rear Input Terminals are only used in the DC Volts  AC  Volts  and Ohms Functions  They are selected with the    werner    Front Rear Switch in the    in    position      12  The External Trigger connector is used to externally trigger the    3478A  when in the Single Trigger mode      3  HP IB Connector     The Voltmeter Complete Connector outputs a pulse after each  measurement cycle      1  The Fuse is 250mA for the 100V and 120V operation  or    125mA for the 220V and 240V operation      18  The Option Label shows the instrument s power option      M These switches select the 3478A   s HP IB Address  the Power   On SRQ feature  and correct power line frequency  50Hz 60Hz      Figure 3 1  3478A Front and Rear Panel Features  cont d   
19. h 3 119    Calibrate  see Section IV of this  manual    Read the Error Register  See  paragraph 3 120    Clears Status Register   Set SRQ Mask  Mdd  where dd is  the octal code of the bits   Front Rear Switch Position   O Rear  1  Front     3478A    77  can be sent to program the mask  The mask can be  cleared by sending        M00        Remember  the mask is on   ly set to output a Require Service Message  not the  Status Byte  for a certain SRQ condition     3 105  Clearing Status Register  Status Byte   Bits 2  through 5 and 7 of the register can be cleared by sending  program code       K        and by a device CLEAR message   Bit 0  Data Ready  is not cleared since it represents the  3478A   s current status  the bit is cleared when the  3478A outputs its reading   Bit 6 is the SRQ bit which  can be cleared by a Serial Poll or if the SRQ Mask is set  to 00     3 106  Power On SRQ  This feature is enabled by set   ting the Power On SRQ switch to On  switch  3 on the  rear panel in the up position   When the feature is en   abled  the 3478A Requires Service  generates an SRQ   each time power is cycled or a reset condition is  generated by the instrument  e g  due to an instrument  failure or selecting the Self Test      3 107  Sending Data to the Display  Up to 12 ASCII  characters can be displayed at a time by the 3478A   s  display  The legal characters are decimal 32 through 95  of the 128 ASCII characters  Only upper case letters and  numbers can be displayed  Lower 
20. has failed its internal RAM self  test             U C  ROM FAIL The 3478A has failed its internal ROM self    test indicating an error in the ROM       CAL RAM FAIL                An attempt to write to the Calibration  CMOS RAM was unsuccessful       UNCALIBRATED        The Calibration CMOS RAM has an incor   rect checksum showing that calibration is  needed           A D LINK FAIL The internal CPU  A D Controller  is unable    to communicate with the A D Converter       A D SLOPE ERR        The A D Converter is unable to do a proper  conversion    A D TEST FAIL The A D Converter has failed its self test     3 63  Calibration    3 64  The 3478A does not have any adjustments to  calibrate the instrument  calibration is done elec   tronically  A known good calibration source is applied  to the multimeter and the value of the known source is  entered into the instrument  A reading is then taken and  compared with the value entered  A Calibration Cons   tant is calculated  from the entered value  to correct the  reading to the known value  and then stored into  memory  The correct readings are calculated using the  constants and then displayed     3 65  There are two steps in calibrating the 3478A   First  the front panel Calibration Switch has to be set to    3 7          Operation    CAL ENABLE  Second  the blue Shift button and the  LOCAL  CAL  button have to be pressed for each  calibration step  e g  press the buttons for the DC Volts  Function   s zero calibration and pre
21. he 3478A must be in     remote    and       listen       to receive the codes  the RMT  and LSTN annunciators are on when the 3478A is in  remote and listening   An example is as follows     Operation    CMD MODE DATA MODE     U7 F1 RA T3    universal unlisten    command    controller       talk          Single Trigger    address Autorange  3478A    listen        address DCV Function    3 102  The Program Codes are a series of 7 bit  ASCII characters  the parity bit is ignored   All  lower case characters  spaces  commas  and  semicolons are ignored  they may be used with the  codes as separators   All null characters  carriage  return  line feed  form feed  vertical tab  and  horizontal tab characters are also ignored  Other  characters which are not included in Table 3 6   cause a Syntax Error  bit 2 in the Status Register  set   In addition  a Syntax Error is caused if the  characters are sent in a different order than shown  in the table  e g        1F       causes a Syntax Error      Do not send program codes    W    and or     X    to the 3478A  These codes can  under  certain conditions  uncalibrate the instrument   The letters    W    and    X    can  however  be  sent behind program codes    D2    or    D3      which are used to send text to the display   see paragraph 3 107      3 103  Programming the SRQ Mask  The SRQ Mask  must be set for the 3478A to output a Require Service  Message  Setting the SRQ Mask will not set the bits in  the Status Byte  these bits are 
22. he controller  or a Serial Poll is done     3 115  Front Panel SRQ  The Front Panel SRQ feature  of the 3478A outputs a Require Service Message  SRQ   each time the Front Panel SRQ button is pressed   Before the message can be output  bit 4 of the SRQ  Mask must be set  This is done by sending program code     M20     bit 4 of the Status Byte and Status Register    Once this is done  the Require Service Message will be  output and the front panel SRQ annunciator turns on   whenever the SRQ button is pressed  The SRQ condi   tion remains until a Serial Poll is done by the controller     3 13    mm    Operation    3 116  Fast Trigger  This trigger mode can only be  selected over the HP IB  It is the same as the Single  Trigger  except the delays in the AC Volts Function  AC  Current Function  and the high Ohms Ranges are omit   ted  The mode can be selected by sending program codes     T5        3 117  Advanced Programming    3 118  General  The following paragraphs have ad   vanced programming information for the 3478A  It in   cludes Extended Ohms Operation  Reading the Binary  Status Byte  and Reading the Error Register     3 119  Extended Ohms Operation  This operation  or  function  is only available over the HP IB and is used to  measure resistance above 30M ohm  The function is  selected by sending program code    F7    or    H7      Home Command  see paragraph 3 110   The 3478A is  set to the 2 Wire Ohms Function and the 30M ohm  Range  A 10M ohm resistor is connected i
23. igger mode   the 3478A starts a new measurement cy   cle  When the cycle is completed  a new cycle can then  be initiated by pressing the Single Trigger button     3 47  External Trigger  This mode is selected by press   ing the SGL TRIG button and operates the same as the  Single Trigger mode  The difference is that the 3478A  can be triggered from the External Trigger input on the  rear panel  The input is TTL logic compatible and the  trigger pulse should be at least 100nS wide  The 3478A is  triggered on the negative edge of the TTL pulse     3 48  Trigger Hold  In this mode  no triggering is done  by the 3478A  no measurement cycle is initiated   This  mode can only be selected using the 3478A   s remote  operation     3 49  Fast Trigger  This trigger is the same as the Single  Trigger except the delays in the ACV and ACI Func   tions  and high Ohms Ranges are omitted  This trigger  can only be selected using the remote operation     NOTE    When the 3478A is in the Single Trigger  mode and an attempt is made to change    3 6             3478A    range or function  the left portion of the  display goes blank  except for the decimal  point  until another reading is taken  instru   ment triggered      3 50  SHIFTED OPERATION  3 51  General    3 52  The Shifted Operation of the 3478A is used to ex   pand the capabilities of the multimeter using the same  number of front panel pushbuttons  This is done by us   ing the bottom row of front panel buttons for two dif   ferent op
24. isten  if Addressed to Talk    No Extended Listener   No Extended Talker   Service Request Capability   Remote Local Capability with Local Lockout   No Parallel Poll Capability   Device Clear Capability   Device Trigger Capability   No Controller Capability                                     3A    76  Data  The Data Message is used to transfer infor   mation between the 3478A and the controller     f a  The message is used to send data to the multimeter  fand consists primarily of set up information  e g  DC  olts  30V Range  etc    The 3478A is the Listener and  ithe controller is the Talker     b  The message is also used by the controller to  receive data from the 3478A  This includes the  multimeter   s output  readings  and status information   f In this case  the 3478A is the Talker and the controller is  f the Listener       3 77  Trigger  GET  Group Execute Trigger   The  f Trigger message causes the 3478A to initiate a measure   ment cycle  It is an HP IB Trigger and triggers the  multimeter in any trigger mode  since it has priority over  the other trigger modes  If the 3478A is triggered during  a measurement cycle  the cycle is aborted and a new cy   cle is initiated  There may be a delay  up to  5 sec  if a  cycle is in progress when the trigger is received  The  multimeter has to be programmed to       listen       to execute  the trigger     3 78  Clear  DCL or SDC  Device Clear or Selective  Device Clear   A Clear places the 3478A into its turn on   routine  
25. its  The 5 1 2  4 1 2  and 3 1 2 digits  can be selected by first pressing the blue Shift button  and then either the AUTO MAN    gt   or  lt  gt  button   respectively     3 56  The Number Of Digits Displayed affects the  reading rate of the multimeter  This is because the  number of digits determines the integration time of the  A D Converter  In the 4 1 2 digit mode  the integration             3478A    time is 1 60 second  or 1 50 second for the 50Hz op   tion   which is called 1 PLC  Power Line Cycle   In the 3  1 2 digit mode  the time is 1 600 second  or 1 500 sec   ond for the 50Hz option  which is  1 PLC  In the 5 1 2  digit mode  the 3478A takes 10 readings using the 4 1 2  digit mode and averages them together for an extra digit  of resolution  This takes a time of 1 6 second  or 1 5  second in the 50 Hz option  which is 10 PLC  For more  information on run up time and the A D Converter   refer to this manual   s Section VII  Service Group F   A D Converter theory of operation      3 57  Autozero    3 58  The Autozero Function of the 3478A is used to  compensate for offsets that may be present in the  multimeter   s internal circuitry  DC Ohms Input  Amplifier  A D Converter  etc   The method used is to  temporarily connect the input of the amplifier to ground   the INPUT LO Terminal  and make a measurement   the INPUT HI Terminal is open at this time   The off   set reading is then stored into the 3478A   s internal  memory  After that  the short is removed and a 
26. mV DC  300mV AC  300mA  DC  300mA AC  or most sensitive  range on any other function   3V DC  3V AC  3A DC  3A AC  or  most sensitive range on any other  function   30V DC  30V AC  30 Ohm  or most  sensitive range on any other func   tion   300V DC  300V AC  300 Ohm  or  least sensitive range on any other  function   3K Ohm or least sensitive range on  any other function   30K Ohm or least sensitive range on  any other function   300K Ohm or least sensitive range  on any other function   3M Ohm or least sensitive range on  any other function   30M Ohm or least sensitive range  on any other function    Autorange                                            DIGITS  DISPLAYED    3 1 2 Digits Displayed  4 1 2 Digits Displayed  5 1 2 Digits Displayed            TRIGGER Internal Trigger  External Trigger   Single Trigger   Hold Trigger   Fast Trigger  see paragraph 3 116                     AUTOZERO Autozero Disabled    Autozero Enabled       DISPLAY Normal Display Operation  Display Text and update Annun   ciators   Display Text and turn off Annun     ciators                    HOME  COMMAND    Equivalent to       F1T4R 2RAZ1N4        Equivalent to       F1R 2RAZ1N4T3   Equivalent to       F2R 2RAZ1N4T3     Equivalent to       F3R 2RAZ1N4T3     Equivalent to       F4R 2RAZ1N4T3     Equivalent to       F5R 2RAZ1N4T3        Equivalent to       F6R 2RAZ1N4T3     Equivalent to       F7R 2RAZ1N4T3                        MISC   COMMANDS    Read Status of the 3478A in Binary   See paragrap
27. mple  bits 0 and 5 are true which results in octal        41     the maximum value is       77         Refer to Table 3 7   byte 4  for the definitions of the bits  The Error  Register is also cleared by sending program code       E           3 14       Table 3 7  Binary Status Byte Definition    Bits Function  Range  and Number Of Digits  True Displayed   XXXXXX01   5 1 2 Digits Displayed   XXXXXX10   4 1 2 Digits Displayed   XXXXXX11  3 1 2 Digits Displayed   XXX001XX  30mV DC  300mV AC  30 Ohm   300mA AC or DC  or Extended Ohms  Ranges   XXXO010XX   300mV DC  3V AC  300 Ohm  3A AC  or DC Ranges   XXX011XX   3V DC  30V AC  3K Ohm Ranges   XXX100XX   30V DC  300V AC  30K Ohm Ranges   XXX101XX   300V DC  300K Ohm Ranges   XXX110XX   3M Ohm Range   XXX111XX   30M Ohm Range   001XXXXX   DC Volts Function   O10XXXXX   AC Volts Function   O11XXXXX   2 Wire Ohms Function   100XXXXX   4 Wire Ohms Function   101XXXXX   DC Current Function   110XXXXX   AC Current Function   111XXXXxX   Extended Ohms Function    Bit    1 Instrument Status Bits    o Internal Trigger Enabled  Autorange Enabled  Autozero Enabled  3478A set for 50Hz Operation  Front Rear Switch in Front Position  Calibration RAM Enabled  External Trigger Enabled  Always Zero  not true     SRQ Mask    Data Ready   SRQ for every available  reading to the HP IB   Not used   Syntax Error   SRO if Syntax Error oc   curs   Internal Error   SRO if Hardware Error  occurs   Front Panel SRQ   SRQ if SRO button is  pressed   Calibra
28. n output after each  trigger  Function and range settings are selected from  the front panel        allel  the    pller    ica   om   The   ote    eeds  i ex   is   ad   ress  Ad     IP 39  itch  are   of  this      nly  sent    the  nal   he  ing  P       up  ach          3478A    Table 3 5  3478A Address Codes    INSTRUMENT  ADDRESS    Shown at Factory  Default Address   23   This number  is called the  Primary Address     ASCII Code    Character 5 bit    Decimal Code                 OMNONNO AWN      Q    IT N lt XS lt CHVPIPVOZZTrA  IONMIOHD         2     2 2    gt   0000000000000000     0000    2 0000    0000    0000   00  00  00  00  00  00  00  00  oo gt O0 0 0 0 0 0 0 0 0 0 0 0  0 0 0    2 22 2 2 00000000   2  2  00O000000       3 98  3478A HP IB Programming    3 99  General  The following paragraphs explain how  to program the 3478A over the HP IB  The information  is given using the HP IB format  rather than controller  dependent language  If the information is desired using  controller dependent language  refer to the 3478A   s  Operator   s Manual     3 100  Programming the 3478A is done by DATA  messages  Set up information  Range  Function  etc   is  DATA sent by the controller and is done using program  codes  The DATA received by the controller  from the  3478A  is measurement data  and other data like the  Status Byte and the five bytes from the Output Binary  Status Byte     3 101  Program Codes  All the program codes for the  3478A are listed in Table 3 6  T
29. n parallel  with the input  Measure the resistance first and then  measure the unknown resistance  The unknown  resistance can then be calculated by this formula   Ri   Rt  Rx        Ri   Rt    where Rx is the unknown resistance  Ri is the measured  10M ohm resistor  and Rt is the measured value of the  parallel combination     3 120  Reading the Binary Status Byte  The current  status  or state  of the 3478A can be determined by  reading its Binary Status Byte  The total number of  bytes is five  with each byte 8 bits wide  The bytes can be  read by sending program code       B       to the 3478A and  then reading its output  A small program to read the  output using the  hp  Model 85 Personal Computer is as  follows     10 OUTPUT 723        B        20 ENTER 723 USING       5 1B         B1 B2 B3 B4 B5    The bytes are in variables Bl  B2  B3  B4  and B5  The  bytes and corresponding meanings of the bits  when  true  are in Table 3 7  Program Code       B       clears the Er   ror Register  byte 4  see Table 3 7      3 121  Reading the Error Register  Besides using Binary  Status Byte 4  see previous paragraph   the status of the  Error Register can also be determined by sending pro   gram code       E       to the 3478A and then reading its out   put  The output is a two digit octal number followed by  a carriage return and line feed  The number shows  which bit s  of the register is true  a two digit number  can be used since only 6 bits are used by the register    For exa
30. ndependent of all other HP   IB activity and is sent on a single line called the SRQ  Line  Its state is either true or false  low true and  high   false   The 3478A must be programmed to send  the SRQ Message  This is done by programming the  SRQ Mask  see paragraph 3 103   The front panel SRQ  annunciator is on when the 3478A requires service     3 84  Since more than one device  on the same Bus  can  output the SRQ Message  the devices can be polled by  the controller  by a Serial Poll  to determine if the  3478A  or another device  requires service  The 3478A  then outputs a Status Byte  see paragraph 3 85  which  shows for what reason the multimeter requires service     3 85  Status Byte  The Status Byte is output by the  3478A in response to a Serial Poll  The message has the  same information as the 3478A   s Status Register  see  next paragraph   and sets the corresponding bit true for  any true SRQ condition shown in Table 3 4  whether the  SRQ Mask is set or not   The bit is represented in Figure  3 6     3 86  The 3478A can require service if any condition in  Table 3 4 is true  Since the SRQ Mask must be set to  output the Require Service Message  except for bit 7   which is set by the Power On SRQ switch   the 3478A   s  Status Register is used to monitor the conditions  This  way  only the condition that is set by the mask outputs  the SRQ Message  Other true conditions that can cause  an SRQ Message  but which are not set by the SRQ  Mask  remain in the Status 
31. on the range selected  The readings can be  displayed either as a 5   4   or 3  digit reading   depending on the selected Number of Digits Displayed   see paragraph 3 54   Refer to Table 1 1 or Table 4 1 for  the DC Current Function   s accuracy specifications     NOTE    Current inputs of greater than about 1 amp  may cause the current shunt   s   1 ohm  Resistor  value to change slightly due to self   heating  This may cause inaccuracies in the  measurement  Sufficient time should be allow   ed for the circuitry to settle after the measure   ment is complete and before other critical  current measurements are made     3 34  AC Current Measurements    3 35  The 3478A can make ac current measurements  from 1uA to 3A in two ranges  300mA and 3A  The fre     Operation    quency response is from 20Hz to 20KHz  The ranges are  protected from excessive currents and voltage by a 3A  fuse  Select the AC Current Function to measure ac cur   rents by pressing the   a button     3 36  The AC Current Function is similar to the DC  Current Function  see paragraph 3 30   a voltage drop  across a resistor is measured  The difference is that the  resultant ac voltage is changed from ac to dc using the  AC to DC Converter  Similar to the AC Volts Function   all ranging is done in the converter     3 37  When the AC Current Function is selected  the  right side of the display shows    MAAC    or    AAC      dependent on the range selected  The readings can be  displayed either as a 5 1 2  4 1
32. owing lists the different 3478A remote  operations     a  3478A Response to Bus Messages  paragraph  3 74      Title Paragraph  Data 3 76  Trigger  GET  3 77  Clear  DCL or SDC  3 78  Remote 3 79  Local 3 80  Local Lockout 3 81       Clear Lockout and Set Local 3 82    Require Service  SRQ  3 83  Status Byte 3 85  Status Bit 3 89  Pass Control 3 90  Abort 3 91    b  3478A Addressing  paragraph 3 92  and Talk   Only Mode  paragraph 3 96      c  3478A HP IB Programming  paragraph 3 98      Title Paragraph   General 3 99  Program Codes 3 101  Programming the SRQ   Mask 3 103  Clearing Status  Register  Status Byte  3 105  Power On SRQ 3 106  Sending Data to the Display 3 107  Home Commands 3 110  Reading Data from the   3478A 3 111  Front Rear Switch Position 3 113  Data Ready Feature 3 114  Front Panel SRQ 3 115  Fast Trigger 3 116    d  Advanced Programming  paragraph 3 117      Title Paragraph    3 118  3 119    General   Extended Ohms Operation   Reading the Binary Status  Byte 3 120   Reading the Error Register 3 121    3 74  3478A Response to Bus Messages    3 75  The following paragraphs explain the 3478A   s  response to Bus Messages  The multimeter   s Bus  capabilities are in Table 3 3     Table 3 3  3478A   s Bus Capabilities      Mnomonie   Interface Function Name    Source Handshake Capability   Acceptor Handshake Capability   Talker  Basic Talker  Serial poll  Talk Only  Mode  Unaddressed to Talk if Addressed to  Listen    Listener  Basic Listener  Unaddressed to L
33. r 3 1 2 digit reading           478A    he       3478A    dependent on the selected Number Of Digits Displayed   see paragraph 3 54   A  03 second or a  3 second delay  is also applied before each reading in the 3M ohm and  30M ohm Ranges  respectively  Refer to Table 1 1 or  Table 4 1 for the Ohms Function   s accuracy specifica   tions     NOTE    To insure 4 Wire Ohms accuracy when using  long leads  it is suggested to keep the lead  resistance as follows     1  Ohms Sense Lead   lt  10K ohms on all  ranges   2  Ohms LO Source Lead   lt  1 30 ohms of  full scale on all ranges    3  Ohms HI Source Lead   lt  1 30 ohms of  full scale on the 3K through 30M ohm  ranges and  lt  3K ohms on the 30 and 300  ohms ranges     3 30  DC Current Measurements    3 31  The 3478A can make dc current measurements  from luA to 3A in two ranges  300mA and 3A  The  ranges are protected from excessive currents and  voltages by a 3A 250V fuse  Select the DC Current  Function to measure dc currents by pressing the  A  button     3 32  In the DC Current Function  the current is ap   plied between the INPUT LO and A Terminals  Since a  known value resistor   1 ohm  is connected between the  terminals  a voltage proportional to the unknown cur   rent and the resistor is generated  This voltage is  measured by the 3478A   s dc circuitry  Input Circuitry  and A D Converter      3 33  When the DC Current Function is selected  the  right side of the display shows       MADC    or    ADC      dependent 
34. regular  input measurement is made  The offset reading is then  subtracted from the input reading and the compensated  reading is displayed     3 59  The Autozero Function is enabled when the  3478A is turned on and after doing a Self Test  see  paragraph 3 60   The feature can be disabled by press   ing the blue Shift button and then the INT TRIG but   ton  The    AZ OFF    annunciator on the display will  light  showing that the function is off  After the func   tion is turned off  the multimeter immediately takes an  offset reading and stores it into memory  This last  reading is then subtracted from the input measurements  that follow  Since no more offset readings are taken  the  reading rate of the 3478A is faster  up to twice as fast    If a range  function  or digit change is made  or an at   tempt is made to calibrate the 3478A  a new offset  reading is taken  With Autozero off  the 3478A   s input  circuitry remains in a static state  This is useful when  making measurements in extremely high impedance cir   cuits where the internal switching transients of the  3478A may affect the reading accuracy     NOTE    The 3478A   s long term stability may be af   fected if the Autozero feature is disabled     3 60  Self Test Reset Operation    3 61  The 3478A uses an Internal Self Test to check its  display and internal circuitry  The multimeter goes  through the test at turn on and also when the Self Test  is selected  The test can be selected by pressing the blue  Shift b
35. s  set  an attempt to calibrate the  3478A has failed            This is the SRQ bit  This bit is true  only if a Require Service Message is  output              Power On SRQ   Shows that a  power on reset has occurred  See  paragraph 3 106         Note  More than one bit in the Status Byte can be true  see  paragraph 3 87      is output  the Status Register   s Data Ready bit is true    The only way the SRQ Message is output is if the Front  Panel SRQ condition is true     3 87  More than one bit in the Status Byte Message can  be true  For example  bit 0  2  and 4 are true  remember   bit 6 is true for any SRQ condition  if the SRQ Mask is  set for the condition  making the resultant Status Byte  look like the following     b7 bO  oT TOP forty oft      NOTE    A    1    in this example shows a true condi   tion     3 10    3478A    3 88  The byte is output as shown in the previous exam   ple with the corresponding octal number of the example  at 125 shown as follows     01 010 101    mm    3 89  Status Bit  The 3478A does not respond to a  Parallel Poll  The Status Bit is used only for Parallel  Poll and should not be confused with the bits in the  Status Byte Message     3 90  Pass Control  The 3478A does not have controller  capabilities     3 91  Abort  Interface Clear   All HP IB communica   tion is terminated  including the 3478A   s Bus com   munication   Control is returned to the controller  The  Abort Message does not remove the 3478A from remote  control     3 92 
36. see paragraph 3 7   In addition  the  multimeter   s address switch  see paragraph 3 92  is read  and the SRQ Mask is set to zero or octal 200  if the  Power On SRQ switch is on  see paragraph 3 106   If  during the turn on routine an error is detected  the hard   ware error bit in the serial poll register is set     3 79  Remote  The Remote Message allows the 3478A to   be controlled over the HP IB  In remote  the front panel  buttons  except the LOCAL and Front Panel SRQ but   tons  are disabled  The Local and Front Panel SRQ but   tons are only disabled when the 3478A is in remote and  local lockout  see paragraph 3 81   The instrument state  in remote is determined by the local state before being  placed in remote  The RMT annunciator on the display  will also be on with the 3478A in remote     3 80  Local  This message clears the remote operation  of the 3478A and enables its front panel operation   Pressing the front panel LOCAL button also places the  multimeter in the local state  if the button has not been    Operation    disabled by the Local Lockout Message  see next  paragraph      3 81  Local Lockout  All the front panel buttons are  disabled with this message  if the 3478A is in remote   The message is in effect until cleared over the HP IB or  power is cycled     3 82  Clear Lockout and Set Local  This message plac   ed the 3478A into local and the Local Lockout Message  is cleared     3 83  Require Service  SRQ   The Require Service  Message  SRQ Message  is i
37. ss the buttons again  for the function   s full scale calibration   Refer to Sec   tion IV of this manual for the calibration procedures     NOTE    The CAL ENABLE Switch on the front  panel should not be in the CAL ENABLE  position under normal use  It should only be  in that position to calibrate the instrument     3 66  MISCELLANEOUS OPERATIONS  3 67  Voltmeter Complete    3 68  The Voltmeter Complete connector is aBNC con   nector which outputs a signal at the end of the A D cy   cle  The signal is a negative going TTL compatible pulse  with a time period of approximately 14S  The Voltmeter  Complete connector is located on the rear panel     3 69  REMOTE OPERATION  3 70  General    3 71  The following paragraphs gives device dependent  information necessary to remotely operate the 3478A  over the Hewlett Packard Interface Bus  HP IB    Directions for mechanical interface connections to the  HP IB are given in Section II of this manual   Familiarize yourself with the front panel operation   local  before attempting to use the multimeter in  remote  HP IB      NOTE    HP IB is Hewlett Packard   s implementation  of IEEE Std  488 1978        Standard Digital  Interface for Programmable Instrumenta   tion    and ANSI MC 1 1     3 72  A general description of the HP IB is in this  manual   s Appendix A  Refer to the appendix for any  non 3478A related HP IB information  It is assumed  in  the following paragraphs  that you are knowledgeable  about the HP IB     3 73  The foll
38. tion Error   SRQ if CAL pro   cedure failed   Always Zero  not true    Power On SRQ   PON SRQ switch on  last time power was turned on or DCL  message was received    Internal Error Information    Set if any of the Calibration RAM loca   tions have incorrect checksums or if a  range with an incorrect checksum is  selected   The Main CPU RAM Self Test has failed  The Control ROM Self Test has failed  An A D Slope Error was detected  The A D has failed its Internal Self Test  A failure in the A D link  between U403  and U462    Always Zero   Always Zero    A D DAC Value    A decimal value between O to 63  represents the setting of the internal  Digital to Analog Converter  DAC     Refer to this manual   s Section VII for       
39. tion is 4 Wire Ohms   The corresponding program codes are      F4R 2RAZINAT3        f  HS Command  DCI   Same as the    H1    com   mand  except the selected function is DC Current  The  corresponding program codes are      F5SR 2RAZINAT3        g  H6 Command  ACI   Same as the    H1    com   mand  except the selected function is AC Current  The  corresponding program    codes are      F6R 2RAZINAT3        h  H7 Command  Ext  Ohms   Same as the       H1      com   mand  except the selected function is Extended Ohms  see  paragraph 3 119   The corresponding program codes are      F7R 2RAZIN4T3           3 111  Reading Data from the 3478A  Data in the form  of readings can be output by the 3478A over the HP IB   other data that can be output is discussed in later  paragraphs   To output a reading  if available   the  3478A has to be addressed to       talk        The readings are  output using 13 ASCII characters and are in the follow   ing form      Operation      amp D DDDDDE D CR LF EON    Polarity of reading  a           is sent for ac readings     Reading expressed in scientific  notation  D is a decimal digit     Carriage Return    Line Feed  with End or  Identify Line set        3 112  Each character in the output statement  except  EOD is one byte  which adds up to 13  The exponent  will be in engineering notation  E 3  E 0  E 3  or  E  6  and the mantissa will always have 1  2  or 3 digits  before the decimal point  D DDDDD  DD DDDD  or  DDD DDD   If the 3478A is in th
40. utton and then the SGL TRIG button  When the  test is selected from the front panel  all the segments of  the display are on except are on as long as the test but   ton is pressed  After that     SELF TEST    will be    Operation    displayed during the time of the test  When the test is  completed        SELF TEST OK    will be displayed  if the  test passes   The 3478A then resets to its turn on state   see paragraph 3 9   not the previous state before the test  was selected  For 3478As with serial numbers 2301 A03760  and below or any instruments that have the Revision B  ROM  U502  installed  the instrument turns on all the  segments in the display for about five seconds  The  display then may show a certain message for about 1 4  second  When the Self Test is completed  the multimeter   s  displays its remote  HP IB  address for about one second   The address is displayed as       HPIB ADRS dd        where     dd    is the address code of the instrument     3 62  If the Self Test fails  an error message will be  displayed indicating the type of failure  If there are  multiple failures  only one failure will be displayed  The  3478A then attempts to operate normally  even if a test  fails   If another failure is noted while trying to operate  normally  the new failure will be displayed  This con   tinues until the test s  passes or the instrument is taken  out of the Self Test mode     Table 3 2  3478A Error Messages    Error Message Definition    U C  RAM FAIL The 3478A 
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Hardware Overview  AKD-C Central Power Supply Installation Guide  Contrôleur et enregistreur de données de CO2  Operation Service Manual  User Manual - JG ANSCHÜTZ GmbH & Co. KG  APM AAL - 9684 User's Manual  LAVE-VAISSELLE Plate-forme DIVA-2  組立説明図をPDFで確認  Limit FPAG02 Amplifier    Copyright © All rights reserved. 