ANNEXURE-`A` INSTRUMENTS FOR ENT

image

Contents

1. TECHNICAL SPECIFICATIONS 1 Microprocessor based digital display electrosurgical unit with continuous feedback system amp having feather touch controls 2 Pure cut amp 2 blends with varying degree of hemostatis amp under water cutting application 3 Spray coagulation Force amp Soft coagulation mode amp two Bipolar micro amp macro mode 4 Audio alarm on completion of Bipolar Coagulation 5 Capability to auto store last used values on front panel if unit is shut off 6 The unit should have double paddle explosion protected foot switch for monopolar cut and coagulation and dual pad silicon patient plate for continuously contact monitoring system 7 The unit should have facility of dual mono polar coagulation 8 Separate foot switch for Bipolar Coagulation 9 Automatic switch off circuit when there is break in patient connecting wire and patient plate 10 Safety audio visual alarm for patient plate disconnection and continuous activation of unit 11 Pure cut 300 watt 12 Blend 1 Low 250 300 watt 13 Blend 2 200 watt 14 Spray Coag 120 watt 15 Force Coag 120 150 watt 16 Desiccation 100 watt 17 Two Bipolar modes Micro 70 80 watt amp Macro 70 80 watt 18 Standard accessories a Double paddle foot switch with cable b Bipolar foot switch mains cable c Silicon rubber patient plate with cable 3 no d Autoclavable electrosurgical handle 3 nos e Set of 5 electrodes 1 no f
2. House marker pins at 3 4 5 length 16cm 1 Oesophagoscopes 50cm 12x16cm robert jesberg oval tube 1 30cm10x12 1 27cm 8x12 1 broncho oesophagoscope forceps single action jaws sheath diameter 2 5 mm width 3 5 mm working length Biopsy Forceps 55 cm 1 Forceps alligator spoon shaped single action jaws Foreign Body Removal Forceps semi flexible sheath diam 1 mm length 35 cm 1 Suction Long suction L 20 25cm 40cm 2mm diameter 3 Oesophageal Dilators 1 set 1 with light carrier 27 5cm 3 5mm 35 30 7mm Bronchoscopes 45 40cm lumen 9mm negus broyle 3 forceps alligator sheath diam Imm length 40 cm semi Biopsy Forceps flexible 1 Foreign Body Removal Forceps sheath diam 2 5mm width 4mm length 50 cm 1 40 cm 30 cm 25 cm for brochoscopes diameter 2 5mm Suction 2mm 3 Ribbon Right Angled Retractors Lagenback retractor 1 Cricoid Hook Double hook 1 Biopsy Forceps tongue biopsy forceps 1 Foreign Body Removal Forceps laryngeal forign body removal forceps 1 Tracheostomy Set 1 Cricothyrotomy Set With all instruments 1 Ear Cup Forceps round cup forceps 2mm 8 cm 1 fish hartmann ear forceps extra delicate up biting 0 8mm Aural Forceps 8 cm length 1 with ipsi and contra reflex with inbuilt printer SISI Test Impedance Audiometer stranger test with headphones l Page 2 of 28 ENT MICROSCOPE SPECIFICATIONS Binocular tube f 170mm Eye piece 12 5 x Objective lens f 250mm 120 degree coupling floor stand with object
3. mode for applications at low voltage to prevent sparking c Macro mode for applications on tissue with high resistance It should have patient plate monitoring facility and should give audiovisual alarm and deactivate output if contact between patient and patient plate is not proper to eliminate the risk of patient burns The unit should have two handswitching and two Footswitching Monopolar outputs and one handswitching and footswitching bipolar output It should have membrane keyboard for power settings Page 14 of 28 10 11 12 13 14 15 The unit should have individual digital display of power for Bipolar Monopolar cut and Monopolar Coag The unit should have temperature sensing cooling fan which should operate automatically to protect generator from thermal damage It should have RS232 serial port to interface with computer to reduce the time and effort in problem diagnostics The unit should have RF activation port to tell other equipments like ECG or EEG that RF current is being generated The unit should not have RF Leakage current more than 150 mA It should be Compact and light weight weighing less than 10 Kg The unit should be operational between 170V AC to 260V AC 50 Hz should not require stabilizer Lap force FX It should have safety standard of UL CUL IEC 601 2 2 It should be compatible with Tissue Select It should be compatible with Robotics arm Accessories e Monopolar Footswitch 1no Bipola
4. needle tray s s perforated Minor set Total 4 sets required ONE ON eS N N NNNNN RK RK e RK kK eR NnPWNRK TO WON DW MN BW WN Sponge holding forceps 10 Backhaus towel forceps 5 B P Handle no 3 B P Handle no 4 Mosquito artery forceps 5 curved Artery forceps 6 curved Artery forceps 6 straight Allies tissue forceps 6 Kocker artery forceps 6 Kocker artery forceps 8 curved Babcock tissue forceps 6 Dissecting forceps 6 plain Dissecting forceps 6 plain Russian forceps 6 Needle holder 6 Needle holder 6 Right angle clamp 8 Adson dissecting forceps 5 plain Adson dissecting forceps 5 toothed Metzenbaum scissor 6 curved Metzenbaum scissor 6 straight Metzenbaum scissor 8 curved Mayo s scissor 6 curved Suction tips set of 4 Kilner retractor double ended 6 Czerny retractor 1 Nos 1 Nos 1 Nos 2 Nos 2 Nos 1 Nos 1 Nos 1 Nos 1 Nos 1 Nos 1 Nos 1 Nos 1 set 1 Nos 1 Nos 1 Nos 1 Nos 2 nos 6 nos 2 nos 1 nos 6 nos 6 nos 2 nos 2 nos 4 nos 2 nos 2 nos 2 nos 2nos 1 nos 1 nos 1 nos 1 nos 1 nos 1 nos 1 nos 1 nos 1 nos 1 nos 1 set 2 nos 2 nos Page 7 of 28 27 28 Langenbeck retractor 2 Instruments hanger Instruments for Intestinal Surgery 2 set required SO R a Intestinal clamps non crushing straight
5. precision vaporizer for isoflurane and seroflurane anesthesia ventilator Facility for Automated machine check Monitoring system to monitor anesthetic gases inspiratory and expiratory ECG EtCO2 FiO2 Pulse Oximeter airway pressure Nibp IBP as required Rectal skin temperature Environmental Specification Temperature 10degree to 40 S50degree C Humidity 15 95 Relative humidity Page 26 of 28 Altitude 440 to 3565m 500to 800mmHg TECHNICAL SPECIFICATIONS 1 N ON o Compact and modular three gas Anesthesia workstation with integrated ventilator for adult to infants Machine should be suitable for low and minimal flow anesthesia with compliance of breathing circuit fresh gas flow compensation decoupling Machine should have at least 2 3 drawers System should have up to 90mins battery backup Machine should have individual locking front casters Flow Management e Compact and easy to use e Flow meter tubes for Oxygen N20 Air Calibrated in multiple scale range around Oml I5L e Dual air flow tubes standard for higher resolution of low flows Breathing System e Latex free and fully auto clavable Should have change over from spontaneous to bag ventilation with single step Should have an external fresh gas outlet for connective Magill or Bain s circuit Flow sensing capability at inhalation and exhalation ports Fully integrated to minimize misconnections and disconnections Gas Delivery System e S
6. range of 25to 45 cms Wavelength of CFL lamps should be in the range of 420 470 nm and irradiance level should be higher than normal blue tube lights The unit should provide a minimum of irradiance 10Watts m2 The irradiance should be measured and reported to the user institution at the time of installation and thereafter during every subsequent warranty CMC AMC Visit and all breakdown visits Lamp source should be continuous tilt able to 90 degree angle to cover the entire treatment area System should be height adjustable with built in non resettable timer Baby bed should be transparent with up down tiltable facility Should work with input 200 to 240 VAC 50 Hz supply Should have safety certificate from a competent authority CE FDA US STQC CB certificate STQCS certificate or valid detailed electrical and functional safety test report from ERTL Test report from ETDC Copy of the certificate test report shall be produced along with the technical bid The measurement of irradiance should be demonstrated to the technical committee during product demonstration EQUIPMENT SPECIFICATIONS OF ANESTHESIA WORKSTATION Description of function Anesthesia workstation is used for delivering anesthesia agents to the patients The complete unit also monitors the vital signs and ventilate the patients Operational Requirements Anesthesia machine complete and integrated with anesthesia gas delivery system circle absorber system
7. 44 S Retractor 2 Pcs 45 Right Angle Retractor Small Tip 2 Pcs 46 Ovum Fcps Curved Small Head 1 Pc 47 Straight Artery Feps Medium Long 1 Pc 48 Towel Clip Hook Type Medium Large 2 Pes 49 Blunt Trocar 11mm 1 Pc 50 Blunt Trocar 12 5mm 1 Pc 51 Blunt Trocar 16 5mm 1 Pc 52 Trocar 11mm with sleeve with Air valve amp Flap 4 Pcs 53 Trocar 5 5mm with sleeve with Air outflow valve amp 6 Pcs Flap 54 Trocar 12mm with sleeve with Air outflow valve amp 2 Pes Flap 55 Trocar 16 5mm with Air outflow valve amp Flap 2 Pcs 56 Snake Retractor Endoflex 1 Pc 57 Verses Needle 10cm 2 Pcs 58 Verses Needle 15cm 2 Pcs 59 Mochi berci s Needle high quality 2 Pcs 60 Lap Endo knife with Reducer 1 Pc 61 Clip Applicator 10mm with all inserts 200 300 400 2 Pcs 62 Weck Clip Hemo Lock Applicator 10mm 1 Pc 63 Clip Applicator 5mm for LT 200 300 inserts l Pc 64 F O Light Source 24v 250W with 2 outlets 1 Pc 65 F O Cable l Pc Essential 1 All instruments should be Autoclavable 2 Forceps Scissors should have SS316 handles 3 All Insulated Instruments should have Teflon coating on shaft Handles 4 Needle Holders should have Tungstun carbide Jaws amp Handles should be Titanium G 2 5 Total Length of the shaft should be 36cm 6 Bariatric Instruments should be 44cm 7 fibre optic light source should 2 bulbs of 24 volts amp should have 30mm heat filters dual fan in out 8 Warranty for the i
8. ANNEXURE A INSTRUMENTS FOR ENT INSTRUMENT SPECIFICATIONS QTY Nasal Speculum Thudicum 7cm small medium large 6 Laryngeal Mirrors With handle 1 2 3 4 5 8 Nasopharyngeal Mirrors With handle 1 2 3 2 Aural Speculum Hartman 3 4 5 6mm black 4 Ear Suction Microsuction cannula varhoeven different sizes 8 Nasal Suction Suction cannula 2 5mm 3mm cvd 15cm eicken 6 Siegles Pneumatic Speculum 2 Tuning Fork 512 Hz Gardiner brown with base 3 Otoscope 3 5 V halogen bulb with charger 6 Nasal Packing Forceps Tilleys fine 55mm 12 5 cm 5 3 Bulls Lamp halogen light 2 Head Mirror 3 ENT Examination Chair Backward and forward up and down motorized with 2 rotating facility Jobson Horne Probe 2 Bipod Plate Measuring plate for bipod 40 cm 16 1 Tonsil Holding Forceps Denis brown large 20cm 8 1 Tonsil Scissors 18cm 7 1 Freer Elevator Sharp D E sharp blunt edge broad 20 cm 8 1 Rigid Nasal Endoscope 4mm wide angle 0 degree German HD Autoclavable 1 Light Source Cold light source twin bulb twin fan xenon 175mm with FO cable diameter 3 5 mm L 180 cm automatic switch in spare lamps Power supply 100 120 230 240 50 60 hz 400 A mains cord 1 Anterior Commissure medium universal size triangular spatula shaped Laryngoscope lateral outer channels for fo light carrier and suction with FO light carrier for laryngoscope 1 Negus
9. Bipolar forceps of various sizes with cable 3 no g Monopolar cord 3 no with connection cable to electrosurgical unit h Anti rusted mobile trolley i Reusable hand switching pencil 1 no j Properly rated fuses 4 nos 19 Unit should operate on 220 12 5 V AC 50 Hz 20 Should submit certificate of relevant IEC safety standard SPECIFICATION NEONATAL TRANSPORTATION INCUBATOR e Double wall transparent hood section with electronic servo temperature control system Temperature setting range from 29deg C to 38 deg C in 0 1 deg C increments with audible amp visual high temperature alarm e Incubator air temperature display range 20 deg C to 42 deg C in 0 1 deg C increments e Standby heating mode to keep incubator ready for use e Low power voltage alarm e Supplied with 12 V DC rechargeable power pack compatible with 230 V AC 50 60 HZ e Auto cut facility when heater gets over heated e Supplied with incorporated IV pole Page 25 of 28 e Should be 93 W x 41 D x 36 H cm in size and approx 15 7 kg in weight e Fixed on movable trolley with lockable castors amp storage space for battery amp gas cylinder 10 11 12 13 PHOTOTHERAPY UNIT DOUBLE SURFACE Phototherapy based on advanced CFL tube LED Technology Minimum 8 Nos of medical grade Blue CFL lamps on source modules For LED technology the irradiance should cover the entire treatment area Height adjustable mechanism treatment distance to the
10. Intestinal clamps non crushing curved Payr Intestinal clamps crushing 4 blade Serrations 9 Moynihan cholecystectomy forceps curved Desjardin gall stone forceps Bake bile duct dilators set of 9 malleable stems8 Instrument for Urology 1 set required SOO AON SO Re 0 Randall renal calculus forceps set of 4 Mayo s clinic pedicle clamps Clutton urethral dilators set of 12 export quality Superapublic trocar Pelvis retractor malleable set of 4 Pot scissor Catheter introducer Watson chynae retractor Macdonald tissue dissector Instruments for Neuro Surgery 1 set required Pi Beeler 8 9 10 11 Adson dual forceps fine toothed 5 Dott dual dissecting forces fine toothed 7 Adson dressing forceps bayonet fine serrated jaws 7 Sergent dual scissor straight 5 Jeffson retractor self retaining curved blunt pointed blades 5 1 2 North field rongeur heavy straight compound action 9 Oliverson rongerur angled to side compound action 8 Hudson brace with perforated amp burs Gigli saw handle Gigli saw Cairns Gigli saw guide amp protector Instruments for Vascular Surgery 1 set required 1 Potts scissor acute angled 45 degree 7 1 2 2 Potts satinsky vena cava clamp angled shanks 8 1 2 Satinsky vena cava clamp atraumatic 1X2 serrations 2 nos 2 nos 2 nos 2 nos 2 nos 1 nos 1 nos 1 nos 1 set 1 set Inos 1 set 1 nos 1 set 1 nos 2 nos 1 no
11. Laryngoscope operating laryngoscope for children 15 cm with FO light 1 carrier for distal illumination and suction tube to remove vapour Biopsy Forceps Forceps straight with 2 mm cupped jaws sheath 1 conically reinforced from distal to proximal end working length 21 cm Foreign Body Removal Forceps Foreign body grasping forceps Alligator Grasping 1 Forceps heavy model single action jaws working length 23 cm Laryngeal Suctions 2mm 4mm 6mm 23 cm 3 Kleinsassers Laryngoscope Adult 18cm large or medium with FO light 1 Microlaryngeal Forceps 23 cm serrated str horizontal 1 Microlaryngeal Scissors Straight 23cm 1 Wullstien Ear Needle str 16 5cm 1 Round Knife 16 cm 3 5mm x 2 5mm 1 Microelevator For Ear 16 cm fine straight 1 Graft Knife Sickle knife 6mm extra fine slightly curved 18 cm 1 Page 1 of 28 Graft Press 1 Endaural Retractor lempert with blades 1 40000 rpm high sp surgical straight and angled Electric Drill Motor Handpiece handpeice forward and reverse cutting with inbuilt And Burrs irrigation 1 Set of Burrs Cutting 10 set TC 1 Diamond set of 10 1 House dieter upward matt 8cm extra delicate 0 8mm Malleus Head Nipper width 1 Pick Straight Str matt 16 cm 90 degree 0 5mm 1 Angled Pick long cvd matt 16 cm 45 degree 0 4mm 1 Perforator 0 3 0 6mm 16 cm 1 Prosthesis Crimper 1 Measuring Rod
12. Power consumption 380 480 VA Colour temperature 6000 k 9 Xenon short ARC Lamp 180 watt 1 piece 10 Fiber light cable 2300 mm long with bundle diameter 3 5 mm 1 piece Deram 180w Xenon short ARC Lamp with 5 piece connector used in light sources Longer lead versions also available for Optic units 14V 180W XBOR181 45C Deram 180w Xenon short ARC Lamp with leads 14V 180W XBOR 181 45C For 8 i ALI 180W Xenon light source from at lantch medical denices Ltd ii orion 800 high intensity 180w Xenon light source color temp 6000 A k SPECIFICATIONS FOR LAP TELESCOPES amp INSUFFLATION UNIT amp TROLLY LAPAROSCOPIC TELESCOPES SET HopkinsHRod Lens Telescope with CLS System and Universal Eyepiece and Multi adapoter light connector 0 deg 5mm wide angle 01 HopkinsllRod Lens Telescope with CLS System and Universal Eyepiece and Multi adapoter light connector 30 deg 5mm wide angle 01 INSUFFLATION UNIT Specifications ELECTRONIC PNEUMOFLATOR Special features High degree of patient safety Easy to use Clear adjacent displays for set value and actual value allow easy monitoring of insufflation process Touch keys for precise preselection of set values Optical and acoustic warning signals in the event of patient overpressure Fully automatic electronically controlled gas refill e g in case of gas loss when changing instruments Page 16 of 28 Safety constant monitoring of intra abdominal pressure any overpre
13. abcock tissue forceps 6 Babcock tissue forceps 8 Dissecting forceps 6 plain Dissecting forceps 6 toothed Dissecting forceps 8 plain Dissecting forceps 8 toothed Russian forceps 6 Needle holder 6 Needle holder 8 Mayo s scissor 8 straight Mayo s scissor 8 curved Metzenbaum scissor 8 straight Metzenbaum scissor 8 curved Czerny Retractor Balfour abdominal retractor self retaining Deaver s retractor set of 5 Instruments hanger Right angle clamp 8 BO DOS SOY ON SS a j WWWNNNNNNNNN DN e KR e e KR e Nor COO AN DUN BRWNHRK DO WANA NN FW WN 4 Nos 8 Nos 2 Nos 2 Nos 1 Nos 8 Nos 8 Nos 2 Nos 6 Nos 6 Nos 6 Nos 6 Nos 4 Nos 4 Nos 2 Nos 2 Nos 2 Nos 2 Nos 2 Nos 2 Nos 1 Nos 2 Nos 2 Nos 1 Nos 1 Nos 2 Nos 1 Nos 2 Nos 1 Nos 1 set 3 Nos 1 Nos Page 6 of 28 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 Right angle clamp 9 Adson dissecting forceps plain Adson dissecting forceps 5 toothed Langenbeck retractor 1 Langenbeck retractor 2 Morris retractor 2 blade Morris Retractor 3 blade Doyen s retractor 3 blade Metzenbaum scissor 6 curved Debakey vascular forceps 6 Debakey vascular forceps 8 Suction cannula pools Suction tips set of 4 Bone curette long Malleable probe Suction cannula long Suture
14. d front glass Easy to access control buttons on the housing front Dripwater protected dustproof housing Low voltage protection via external 24VDC Mains power supply Picture in Picture display PIP Mirror Imaging Possible Upto 5 different users profiles can be stored Medical grade FDA UL and CE Approval VVVVVVVVVVVV V Specifications HD TFT Flat Screen Monitor with stand size 26 Aspect Ratio 16 10 HD format Brightness 500 cd m2 Max Resolution 1920 1200 Pixels Maximum viewing angle 178 vertical Contrast ratio 800 1 Rated power 115 watts Power Supply 100 240 VAC Screen Dimensions 627 x 427 x 100mm Video Inputs S Video signal to 4 pol Mini DIN socket RGB signal to 5 x BNC socket Composite to BNC socket HD SDI signal to BNC socket DVI signal to DVI D socket SDI signal to BNC socket Accessories External 24VDC Power Supply Mains Cord Pedstal Certified to EN 60601 1 protection class IPX 1 Page 13 of 28 XENON LIGHT SOURCE Xenon 300 1 Xenon Light Sources 300 watts Lamp type Xenon 15V 300 Watt Color Temperatures 6000K Light Outlets 1 Light Intensity Adjustment Continuously adjustable either manually or automatically by cameras video output signal Certified To IEC 601 1 amp UL 544 CE According to MDD protection class 1 CF Fibre Optic Light Cable 1 Thickness 4 8mm Length 300 cms Essential l Supplier should be a company of repute having it s own office in India
15. e humidifier chamber 15 each Sensor cable 15 each Oxygen cells 2 each Flow pressure sensor 20 each SPECIFICATIONS ADULT VENTILATOR The device complies with requirements of Medical Device Directive 93 42 EEC Classification Class I equipment According to IEC EN 60 601 1 Standards IEC EN 60 601 1 Type B IEC EN 60 601 2 12 EN 794 1 IP classification IP 20 Electromagnetic compatibility EMC e Emission According to IEC EN 60601 1 2 Edition 2 2001 e Immunity Extended test to According to IEC EN 60601 1 2 30 v m Patient range Weight Adult all ventilation modes 10 250 kg Infant invasive ventilation 0 5 30 kg Infant NAVA NIV NAVA 0 5 30 kg Infant NIV PS PC 3 30kg Infant Nasal CPAP 0 5 10 kg OPERATING RANGES Operating temperature 10 to 40 C Relative humidity 15 to 95 non condensing Atmospheric pressure 660 to 1060 hPa Lowest pressure in breathing system 400 cmH20 NON OPERATING CONDITIONS Impact Peak acceleration 15 g Pulse duration 6 ms Number of impacts 1000 Page 21 of 28 Storage temperature Storage Relative Humidity Storage Atmospheric Pressure POWER SUPPLY Power supply automatic range Selection Plug in battery module Battery backup Battery capacity Recharge time Battery backup time External 12 V DC Max power consumption Dimensions User Interface Patient Unit Weight Method of triggering Max operating press
16. f Plastic Handle without ratchet Outer Tube insulated Forceps Insert 1 Page 18 of 28 6 Grasping Forceps 2 x 4 teeth rotating with connector pin for unipolar coagulation size 5 mm double action jaws consisting of Plastic Handle with style ratchet Outer Tube insulated Forceps Insert l 7 FUNDUS GRASPER Forceps rotating with connector pin for unipolar coagulation size 5 mm heavy double action jaws consisting of Plastic Handle with MANHES style ratchet Outer Tube insulated Forceps Insert 1 8 METZENBAUM Scissors rotating with connector pin for unipolar coagulation size 5 mm lenght 36 cm blades curved double action jaws length of blades 12 mm consisting ofInsulated Handle without ratchet Outer Tube insulated 1 9 Bipolar Forcep width of the jaws 4 mm serrated size 5 mm consisting of Ring Handle Outer Tube Inner Tube Forceps Insert 1 10 Bipolar High Frequency Cord with 2 x 4mm 1 11 Monopolar High Frequency Cord with 4 mm 1 12 Suction and Irrigation Tube 1 13 Dismantling needle holder 1 14 Coagulating and Dissecting Electrode L shaped with connector pin for unipolar coagulation size 5 mm 1 15 PCOD needle 1 16 Myoma Screw 1 VACUUM EXTRACTOR The equipment should have facility for suction evacuation and vacuum extractor of fetal head Electric requirement as per Indian power supply 240 watts Vacuum pressure suction pressure of the unit should be 0 700 mm Hg minu
17. for at least 15 years 2 Should have supplied to institutions of repute in India 3 Should cover at least 3 years warranty and maintenance 4 Should provided on site installation and demonstration SPECIFICATIONS FOR ELECTROSURGICAL GENERATOR CAUTERY Micro controller based isolated Electrosurgical Generator having both Monopolar and Bipolar outputs designed for all surgical procedures Smart generator should be able to monitor changes in tissue impedance continuously and adjusts power Monopolar outputs should have three cutting modes a Low Cut for delicate tissue or Laparoscopic cases having maximum power of 300 W b Pure cut for clean precise cut in general surgery having maximum power of 300 W c Blend mode for cutting with homeostasis having maximum power of 200 W All cut modes should be able to adjust output power depending on tissue density by less than 15 or 5W whichever is greater It should have three Coag Modes with maximum power of 120 W a Desiccate mode for low voltage contact coagulation suitable for Laparoscopic and delicate tissue work b Fulgurate mode for efficient non contact coagulation in most applications c Spray mode should have randomized spray effect of varying amplitude and frequency for coagulating large tissue areas with minimum depth of necrosis It should have three bipolar modes with maximum power of 70 W a Precise mode to have fine control of desiccation in delicate tissue b Standard
18. hould have pin index yokes for oxygen and N20 besides separate connection for central gas supply for O2 N20 and Air e The machine should have pressure gauges for cylinder and central supply lines mounted on front of Anesthesia machine The gas connections should be non interchangeable e Automatic cut off N2O by oxygen pressure failure e Hypoxic guard to ensure 21 O2 at a minimum of 200ml e Emergency oxygen flush at 25 75 L min bypassing the vapourizer Vapourizer e Tool free installation selectatec manifold interlocks and isolates vapourizer e Machine should be able to mount two vapourizers These should be Temperature Pressure and flow compensated Isoflurane and seroflurane e The mounting of vapourizer should be quick and easy e Should have extended delivery range from 0 8 e Vapourizer must be isolated from gas flow in the off position amp prevent the simultaneous activation of more than one vapourizer 10 Standard Circle Absorber System e Should have adjustable pressure limiting valve breathing circuit pressure measuring device e Should be suitable for low flow e Should have CO2 absorber chamber canister e Should have bag ventilator selecting valve integrated onto the absorber 11 Ventilator e Electronically controlled electronically driven ventilation not requiring any driving gas Should not require changing of bellows for adults and infants Ventilator should have volume control and pressure control funct
19. ion with PEEP Tidal volume 20m1 1500ml PEEP 0 20mbar Breathing Frequency 4 65 breath min I E Ratio 4 1 to 1 4 Page 27 of 28 Pressure limiting Pmax Upto 80cm H2O Should automatically compensate for compliance of breathing system 12 Display of Ventilator Tidal Volume Inspiratory and expiratory Minute volume Inspiratory Pressure Plateau Pressure Peak and mean PEEP I E Ratio 13 Alarm Limits and Alarm Adjustable high low limits with Audio and visual alarms of minute volume airway pressure inspired oxygen concentration Alarm silence countdown timer 120 to 0 secs apnea alarm Mechanical ventilation on less lt equal to 5ml breath measured in 30secs Mechanical Ventilation off lt equal to 25ml breath measured in 30sec 14 Specifications for Patient Monitor Should be suitable for adult and pediatric population Should have vital parameters as ECG 3 and 5 leads NIBP Systole diastolic mean IBP SPO2 EtCO2 Gas monitoring with M lt AC values inspiratory and expiratory Temperature measurement Automatic Identification and measurement of anesthetic gases EtCO2 02 N20 MAC value FiO2 measurement Should have automatic graphic and tabular trending of all monitored parameters Should have at least 90mins battery backup Facility to store snap shots during critical events Audio visual and graded alarm system Facility for data storage for at least 24hours Registrar Page 28 of 28
20. ive lens 200 mm and 400mm manual magnification 0 4x 2 5x focusing range 40mm INSTRUMENTS FOR PATHOLOGY AUTOMATIC TISSUE PROCESSOR A a ce eo en Sd 14 15 16 17 Complete automatic dehydration Pre programmable provision for dehydration cycle S S Basket rotor slow S S Tissue basket S S Tissue capsule with sliding cover size 28 x 28 x 9 mm 20 x 20 x 10 mm Two litre capacity glass beakers S S Wax bath with thermostat Tissue capsule basket S S with even perforation Divider for tissue capsule Programming of the machine should be simple and user friendly 11 12 13 Should have a main process timer 24 hours delay timer should be provided for weekend and routine delay starting Unit should have provision to prevent double processing and ensure tissue remain embedded in the wax bath until removed Acrylic control panel cover Should offer safety for tissues in case of power failure Suitable to work on 220 V single phase 50 Hz AC supply Equipment should be USA FDA approved TISSUE EMBEDDING STATION l 2 Should be compact and have user friendly operation Safety measures in place to provide safeguard against burn injuries Page 3 of 28 BO GOP SSIS ON a e Bench top unit Easy to maintain Adequate paraffin container with temperature setting range between 50 C to 75 C Should have a paraffin collection tray Cold palte should be substantial with 0 C temperature Programmable sta
21. m 10mm 33cm 2 Pcs 18 Claw Forceps 33cm 5mm 2 Pes 19 Atro Forceps 2 2 small jaw 33cm 5mm Very fine 1 Pe serrations 20 Scissors Curved 33cm 5mm Metzenbaum Saw 2 Pes edged 21 Scissors small micro scissors small tongue 30cm 1 Pe CBD Section 22 Hook Scissors 33cm 5mm 2 Pcs 23 Scissors Matzenbaum Extra Curved Woif Type 33cm 2 Pes 5mm 24 Suction Nozzle Lever Type Blunt Tip 33cm 5mm 1 Pe 25 Suction Nozzle Lever Type open end 23cm 5mm 1 Pc 26 Suction Nozzle Combo 5mm 10mm 33cm Blunt 1 Pe cum open end 27 Normal Intestinal Grasper 33cm 5mm 2 2 2 Pes 28 Hasson Cone 11mm 1 Pe 29 L Hook Dissector 33cm 5mm 2 Pcs 30 Spatula Dissector 33cm 5mm 2 Pcs 31 Aspiration Needle with Coagulation Point 2 Pcs 32 Palapation Probe Magnaetic33cm 5mm 1 Pc 33 Knot Pusher 33cm 5mm 1 Pc 34 Needle Holder RK Sinha type Titanium Straight 1 Pe Toothed Blunt Tungsten Carbide Tip 35 Titanium Needle Holder Curved Handle Type 2 Pes 36 Palanivelu Hydatid Cyst Long Y Trocar 11mm Extra 1 Pe Long 37 10 to 5 Reducer Sleeve 2 Pcs 38 12 to 10 Reducer Sleeve 2 Pcs Page 10 of 28 39 16 5 to 10 Reducer Sleeve 2 Pes 40 5 to 3 Reducer Sleeve 2 Pcs 41 Babcock Grasper 33cm 10mm 1 Pc 42 Babcock Grasper 33cm 5mm 1 Pc 43 Reducer Sleeve Top End Supra Flap 2 Pes
22. nstruments should be minimum 2 years 9 Clip Applicator should have total interchangeable shaft 10 Prompt after sales services and substitute instrument or appliance should be supplied in event of repair 11 Sheath and inserts all instruments should be inter changeable 12 Grip of instruments should be soft and secure and should not leave any mark on tissues Page 11 of 28 SPECIFICATIONS FOR DIGITAL CAMERA The system should be truly Digital HDTV endoscopic video camera The system should have the maximum Resolution of 1920 X 1080 pixels progressive scan and the consistent use of 16 9 format for Input amp Output to guarantee genuine HDTV The system should have Special Features Visibly Improved Imaging CCD sensing chip should optimizes image quality amp Digital Source Sampling thus maximizing hi fidelity image transmission Optimizes to Any Size The system should have Optical Zoom to enhance the quality of Image size amp cross specialty standardization of the camera system regardless of the telescope used Plug and Go The system should automatically optimize all settings The system should be ready to use as soon as it is connected to the camera control unit Any Head Any Time The system should have the facility to use a single camera control unit for all camera heads either single chip or three chip thus minimizing preparation amp maximizes interspeciality standardization The system should have USB port to ca
23. pture still images in Full HD resolution as well as Video Sequences in SD to a USB mass storage device Camera system should be compatible with Communication Bus system for remote controlled operation of the various features of the camera along with other equipment Technical Specifications Image sensor 3X1 3 CCD Chip Pixels 1920 x 1080 AGC Microprocessor controlled Lens Integrated Zoom Lens Video output Composite signal to BNC socket Y C signal to S VHS socket 2 x RGB signal to D sub socket HDTV signal to DVI D socket Input Keyboard input for character generator 5 pole DIN socket Control Output Input Page 12 of 28 3 5mm Stereo Jack Plug ACC1 ACC2 Serial Port at RJ 11 USB Port with ICM 2x Mains Cord BNC BNC Video Cable Length 180 cm S Video Y C Connecting Cable Length 180 cm Special RGB Connecting Cable Length 180 cm Connecting Cable for controlling peripheral units length 180 cm 2 DVI D Connecting Cable Length 300 cm Keyboard with US English Character Power Supply 100 240 VAC 50 60 Hz Certified to IEC 601 1 601 2 18 CSA 22 2 No 601 UL 2601 and CE according to MDD protection class1 CF DIGITAL MONITOR FOR HD CAMERA Special Features HDTV Display in original 16 10 HDTV format 1080p 50 and 1080p 80 display possible On Screen menu for monitor setting possible in several languages Liquid crystal display Max Resolution 1920 1200 Pixles Anti reflection coate
24. r Moveable condenser is provided with iris diaphragm having 1 2 N A 3 PART DIFFERENTIAL HEMATOLOGY ANALYSER 1 Automated hematology analyser capable of analyzing the following Parameters WBC RBC Hb Hct MCV MCH MCHC RDW MPV RBC and Plt Histograms Differential counts 18 parameters Throughput minimum of 50 samples per hour Closed or open vial sampling with one step sample procuring with single button start switch Auto probe wiping facility Built in printer with large screen for display Patient ID auto numbering Data storage for 200 to 300 patient results and QC storage Levey Jennings graphs Should be USA FDA approved SOOO et SON ON Se ee tS Page 5 of 28 EQUIPMENTS INSTRUMENTS FOR SURGERY OPEN OT INSTRUMENTS SPECIFICATIONS 1 S S of Grade AISI 410 amp AISI 420 using for the manufacturing of Surgical Instruments 2 All the Instruments will be Mat Finish 3 Warranty 6 Months for any manufacturing Defect 4 During warranty and thereafter for repair substitute provided Laprotomy Set total 6 Sets required Sponge Holding Forceps 10 Backhaus Towel Forceps 5 B P Handle no 3 B P Handle no 4 B P Handle no 7 Mosquito Artery Forceps 5 Curved Artery Forceps 6 curved Artery Forceps 6 straight Artery forceps 8 curved Artery forceps 8 straight Allies Tissue forceps 6 Allies tissue forceps 8 Kocker artery forceps 8 straight Kocker artery forceps 8 curved B
25. r footswitch 1no Reusable Handswitching Pencil 2nos Disposable Patient Plate Monitoring System 50 nos Bipolar Forceps 1no Forceps Cord 2nos Universal Adapter 1no a ale aa OUTPUT CHARACTERISTICS Mode Maximum Power Crest Factor Watts typical Bipolar with Precise 70 1 5 Coagulation and Standard 70 1 5 cutting facilities Macor 70 1 5 Monopolar Cut Low 300 15 Pure 300 1 5 Blend 200 2 5 Monopolar Desiccate 120 5 Coag with Flug High Crest Factor 120 7 0 contact and non Flug Low Crest Factor 120 5 5 contact facilities Spray 120 8 Essential 1 Suppler should be a company of repute having its own office in India 2 Instrument supplied should USFDA approved 3 Should cover comprehensive 3 years warranty SPECS FOR BASIC SET FOR CYSTOSCOPE AND ACCESSORIES Fae aca OA Telescope 0 degree 4 mm wide angled autoclavable 1 piece Telescope 30 degree 4 mm wide angled autoclavable 1 piece Cystoscope sheath with obturator 20 FR 1 piece Cystoscope sheath with obturator 23 FR 1 Piece Page 15 of 28 5 Adaptor for cystoscope with one instrument port 1 Piece 6 Flexible grasping forceps 7 FR for cystoscopy with working length 365 mm 1 piece 7 Flexible biopsy forceps 7 FR working length 370 mm 1 piece 8 Light source 180w Xenon short ARC lamp and power cable mains power supply 100 240 V A C 50 60 HZ 1 piece Technical data Classification CF
26. rting time Each functional areas should have individual adjustable temperature 10 220 V 50 Hz 11 Equipment should be USA FDA approved ROTARY MICROTOMES 1 Should have stability which is vital 2 Safety should be insured by providing safeguards to the blade edge 3 Use comfort 4 Section thickness 0 5 mm to 60mm 5 Setting values of variable increments of 0 5 mm 1 mm 2 mm 5 mm 6 Trim setting 1 mm 60 mm 7 Max specimen size 50 x 60 x 40 mm 8 Specimen orientation 9 Should have smooth running hand wheel 10 Universal knife holder lateral adjustment 11 Sections waste tray 12 Cassette black holder Knife holder base Standard knife holder 13 Equipment should be USA FDA approved STUDENTS MEDICAL MICROSCOPE ee a ee eo Stand Stable amp Robust diecast traditional Horse Shoe base Body Inclined through 90 deg standard tube length of 160 mm Focusing Separate coarse and fine focusing knobs with graduation on one side Mechanical Stage For X Y movements detachable mechanical stage Nosepiece Triple revolving Nosepiece is provided with positive click stop for proper centering Stage Fixed square stage 125 x 125 mm with stage clips to hold the slide Illumination Illumination through a 50 mm diameter plano concave mirror mounted on gimbal mount Objectives Antifungal coated Achromatic Objectives of 10x amp 40x SL 100x Oil SL Page 4 of 28 9 Eyepieces Two Eyepiece of 6x 10x or 15x 10 Condense
27. s Inos 1 nos 1 nos 1 nos 1 nos 2nos 1 nos 1 nos 1 nos 1 pair 10 nos 1 nos 2 nos Page 8 of 28 double angled jaws 3 Medium 40 mm 9 mm deep 10 debakey peripheral vascular clamp light weight artraumatic 1X2 4 Serrations curved 6 5 Debakey bull dog clamps straight cross action atraumatic 1X2 6 Serrations 20 mm jaw 3 7 Serrations 30 mm jaw 3 1 2 debakey bull dog clamps curved cross action atraumatic 1X2 serrations 8 20mm jaw 3 9 Debakey tissue forceps straight atraumatic 1X2 serrations 6 10 Debakey tissue forceps straight atraumatic 1X2 serrations 8 11 Needle holder fine for vascular surgery 7 spring action Paediatric Set 1 Set required 1 Mosquito artery forceps 3 curved Mosquito artery forceps 3 2 straight Metzenbaum scissors 4 curved Needle holders 5 fine Allis tissue forceps 4 Intestinal clamp 5 straight non crushing Bulldog clamps Malleable retractor size 4 x8 x8 1 x8 set of 3 Sinus probe malleable 10 Finochetto chest retractor infant CH NAA RWHN Tracheostomy set 4 set required 1 Mosquito artery forceps 5 Artery forceps 6 B P Handle no 4 B P Handle no 7 Dissecting forceps 6 plain Dissecting forceps 6 toothed Needle holder 6 Allis tissue forceps 6 Towel clip cross action 3 1 2 Single hook retractor Double hook retractor Langenbeck retractor Metzenbaum sci
28. sible factor PEEP regulation Rise time expiratory flow measurement Expiratory flow range ALARMS Airway pressure upper Adult invasive ventilation Adult Non Invasive Ventilation Infant Invasive ventilation Infant Non Invasive Ventilation Can be attached to the mobile cart a table rail or pole 15 30 mm diameter TFT LCD module 31 cm 12 1 diagonal 246 0 x 184 5 mm Max 6 cmH20 at a flow of 1 1 s Max 0 1 ml emH20 Microprocessor controlled valve 0 to 3 3 I s 0 to 0 55 I s Max 3 cmH20 at a flow of 1 I s Max 0 1 ml emH20 Microprocessor controlled valve lt 12 ms for 10 90 lt 12 ms for 10 90 response at flow of 0 05 3 2 I s 0 to 3 2 I s 16 120 cmH20 16 40 cmH20 16 90 cmH20 16 40 cmH20 Expired minute volume Upper alarm limit Adult Infant 0 5 60 l min 0 01 30 l min Expired minute volume Loweralarm limit Adult Infant Respiratory frequency High end expiratory pressure Low end expiratory pressure High continuous pressure O2 concentration Gas supply 0 5 40 I min 0 01 20 I min 1 160 breaths min 0 55 cmH20 0 47 cmH20 Set PEEP level 15 cmH20 exceeded for more than 15 sec Set value 5 vol or 18 vol Below 200 kPa 2 0 bar 29 PSI and above 650 kPa 6 5 bar 94 PSI Page 23 of 28 VENTILATION MODES INVASIVE VENTILATION Controlled ventilation PC VC PRVC Optional with Universal Ba
29. sic Edition Supported ventilation PS CPAP VS Optional with Universal Basic Edition Combined ventilation SIMV VC PS SIMV PC PS SIMV PRVC PS Bi Vent Automode Control mode VC lt gt Support mode VS Control mode PC lt gt Support mode PS Control mode PRVC lt gt Support mode VS VENTILATION MODES INVASIVE NIV PC NIV PS Nasal CPAP Real time waveforms up to 4 waveforms displayed simultaneously Pressure Flow Volume CO2 Edi SPECIFICATIONS CAUTERY MACHINE High end diathermy with vessel sealing system Electrosurgical unit with monopolar and bipolar functions Foot switch with switching facilities to change between monopolar and bipolar functions HF power limitation 1 to 300W for monopolar cutting adjustable in 1W steps Auto start for both bipolar outputs possible Facility to change between moderate coagulation modes Adjustable delay time for bipolar auto start function Minimum programs including TUR P TUR Vap TUR BT EndoLAP Thorax Micro Neuro CF according to EN standard Accessories Reusable monopolar pencil blade and needle electrodes reusable connecting cables ground plates etc All accessories needful for cardiac and thoracic surgery neuro surgery Uro amp Gastro surgery Page 24 of 28 Additional Facility e Complete reusable instrument for laproscopic vessel sealing e Special Bipolar TUR programs for resection under saline for urology TUR
30. ssor 6 curved Trachestomy dilator SOOO Sy OO San NS pk A U N Ready Made Instrument Sets Neurosurgery set Spine set 2 nos 2 nos 2 nos 2 nos 2 nos 2 nos 2nos 2 nos 2 nos 2nos Snos 4 nos 2 nos Inos 2 nos 2 nos 2 nos 1 set 1 nos Inos 4 nos 1 nos 1 nos 1 nos 1 nos 1 nos 1 nos 2 nos 4 nos 2 nos 2 nos 2 nos 2 nos 2 nos 1 nos 1 nos Page 9 of 28 SPECIFICATIONS FOR LAPAROSCOPIC AND INSTRUMENTS S No Description Quantity 1 Maryland Kelly s 5mm 33cm 2 Pes 2 Original Maryland University Type 5mm 33cm 1 Pc 3 Wave Intestinal Grasper 5mm 33cm 2 Pes 4 Universal Grasper 33cm 5mm 2 Pes 5 Da Vinci Medium Grasper 33cm 5mm 2 Pcs 6 Endoclinich 33cm 5mm 2 Pes 7 Duckbill Grasper 33cm 5mm 2 Pcs 8 Right Angle 33cm 5mm 1 Pc 9 Right Angle 33cm 10mm 2 Pcs 10 Hunter Grasping Forceps 5mm 33cm Double Action 2 Pcs 2 2 Normal Head 11 Hunter Grasping Forceps 5mm 33cm Double Action 2 Pcs type 2 2 Small Head 12 2 2 Medium size Grasper 33cm 5mm 2 Pes 13 2 Intestinal Grasper 33cm 5mm Medium size 2 Pcs 14 Ovum Fenestrated Forceps 33cm 5mm 2 Pes 15 Maxigrip with Ratchet 33cm 5mm 2 Pes 16 Claw Forceps 10mm 33cm Gall Bladder Extractor 2 Pes 17 Stone Extractor Mediu
31. ssure is reduced immediately Technical specifications Gas flow min gt 0 to 30 lit Pressure mmHg gt 0 to 30 mmhg Intra abdominal pressure guage gt 0 to 50 mmhg Power supply gt 100to 240VAC 50 60Hz Certified to gt IEC 601 1 CE Label according to MDD Should be compatible to the Communicating Computer System so as to function as an integral part of the digitally controlled Operating Room under the command of the operating Surgeon ENDOSCOPIC TROLLEY IN OT INDIAN MAKE Special features Rides on four antistatic casters 2 Equipped with locking brakes Shelves fixed 3 and inclinable 1 7 Equi potential plugs 2 Handles 1 Cable channel mounted on the left or right side of the column 4 sets of non sliding stands for units Electrically supply terminal strip with 6 plugs A camera mount A drawer unit Technical specifications Videocart 680 x 1730 x 640 mm w x hx d Shelf 630 x 420 mm w x d Caster diameter 150 mm Maximum loading capacity per shelf 40 kg Maximum total weight 150 kg Page 17 of 28 INSTRUMENTS FOR GYANACOLOGY NEW SPECIFICATIONS DIGITAL VIDEO COLPOSCOPE DVC 10000 Resolution should be more than 850 lines Single Light LED circle should be there Minimum 10 to 11 Mega Pixels Digital Matrix Processor Recommended Lamp life 20000 hours Optical Zoom from 1 36x Magnification 1 40 Corner to Corner Uniform Brightness Screen should display the magnification indicator Ace
32. tely adjustable Suction system should have piston cylinder self lubricating There should be 2 collection bottles small of 1 5 2 L and large of 3 4 L It should be supplied with both metal and silicon suction cups at least 3 metal cupsi e 3 small 25 30 mm medium 40 45 and large 50 60 mm internal diameter which can be sterilized and should be rust proof Atleast two Silastic cups medium and large 50 and 60 mm also to be supplied There should be separate suction tube for the both suction and vacuum extractor There should be float valve for each bottle The power control should be foot switch also The unit should be mounted on a mobile trolley which is rust proof with a tray to keep the instruments which is made up of stainless steel and should be rust proof Trolley should have locking device in wheels Page 19 of 28 e The unit should have ISO 9001 2 certificate or equivalent which should be submitted with tender e Prior demonstration in the department is necessary INSTRUMENTS FOR PEDIATRICS SPECIFICATIONS FOR NEONATAL VENTILATOR FDI Approved 1 Bssenny components Ventilator Air Compressor Reusable Circuit with online bacterial filter Humidifier Stand for circuit Operator manual Service manual 2 Type of ventilator Continuous flow time cycled pressure control 3 Modes available CPAP IMV SIMV Assist Control PSV VG BiPAP 4 Range of set parameters Peak inspiratory pressure 0 50 cms Posi
33. tic acid Test Timer should display on screen In built Three Step Electronic Green Filter with no light loss Facility to measure the picture in millimeters should be there Picture can be saved for single colposcopy in software Statistics mode with automatic flow chart facility should be provided Printing of multi format diagnosis report like single double four photo report Foot switch should be provided to save the pictures Imaging system for high resolution images Automatic Gamma correction is in built High Speed Fast auto focusing all the time during colposcopy Multi point Glare Reduction Three modes of Electronic digital green filter technology no light diminished and can help in observing fine vessels Convenient vertical floor stand and control handgrip Provide examination and surgical operation with a convenient operation environment FDA approved and prior demonstration SPECIFICATIONS FOR LAP AND GAYNE S No Product Description 1 VERESS Pneumoperitoneum Needle i 2 Trocar size 11 mm Trocar only with pyramidal tip Cannula without valve with insufflation stop cock Multifunctional Valve 3 3 Trocar size 6 mm Trocar only with pyramidal tip Cannula without valve with insufflation stop cock Multifunctional Valve 3 4 Reducer 11 5 mm 1 5 Maryland Dissecting and Grasping Forceps rotating with connector pin for unipolar coagulation size 5 mm double action jaws consisting o
34. tive end expiratory pressure 0 20 cms Fraction of inspired oxygen 21 100 Inspiratory time 0 1 3 secs Rate 0 150 bpm Gas flow 5 15 Lpm 5 Display Both digital and analog All set parameters as mentioned above Measured parameters PIP PEEP Mean airway pressure FiO2 Ventilator rate Denived parameters Te I E ratio Leak percentage Tidal volume Minute ventilation Pressure amp flow waveforms and loops Alarm message Calibration Silenced alarm 6 Alarms Both audio amp visual Low amp high pressure Compressor failure Failure of sensor s Tube obstructed Power failure 7 Humidifier Flow resistance upto 1 cm H2O L Sec Temperature range 31 40 degrees C Temperature control 2 C Digital display of temperature range of display 5 80 C Water level indicator Warm up time less than 15 minutes Alarms Heater wire on Page 20 of 28 Airway temp Tracking 2 C from set temp Chamber temp If chamber temp varies 4 C from set temp for 20 minutes or alarm immediately if set chamber is exceeded by 10 C Heater wire off Airway temp Fixed at 41 C high amp 29 5 C Low Chamber temp limited to 60 C max Should be compatible with both reusable amp Disposable chambers and reusable amp disposable Circuits 8 Power 220 240 V 50 60 Hz Uninterrupted power back up for at least one hour 9 Spares with each ventilator Reusable circuit with Y piece heater wire temp probe amp adapter 15 each Bacterial filter 15 each Reusabl
35. ure Bias flow Adult Infant GAS SUPPLY Inlet gas pressure Air O2 29 94 PSI Connection standards available Unavailable gas loss of gas pressure 25 to 60 C 13 to 140 F lt 95 470 to 1060 hPa 100 120 V AC 10 50 60 Hz or 220 240 V AC 10 50 60 Hz Two battery modules are delivered with the ventilator Up to six battery modules can be included Rechargeable 12 V 3 5 Ah each Approximately 3 h battery At least 3 h when using six batteries 12 0 V 15 0 V DC 10 A At 100 120 V 2 A 190 VA 140 W At 220 240 V 1 A 190 VA 140 W W 355 x D 53 x H 295 mm W 300 x D 205 x H 415 mm Approximately 20 kg Patient Unit 15 kg User Interface 5 kg Flow and pressure Approximately 115 cmH20O 2 l min 0 5 l min 200 650 kPa 2 0 6 5 bar AGA DISS NIST or French standard The flow from an unavailable gas air or O2 is automatically compensated for so that the patient gets the preset volume and pressure PATIENT SYSTEM GAS CONNECTORS Conical fittings Gas exhaust port USER INTERFACE Weight Male 22 mm female 15 mm In accordance with ISO 53561 Male 30 mm cone Approximately 5 kg Page 22 of 28 Attachment SCREEN Type Size Viewing area INSPIRATORY CHANNEL Pressure drop Internal compressible factor Gas delivery system Inspiratory flow range Adult Infant EXPIRATORY CHANNEL Pressure drop Internal compres

Download Pdf Manuals

image

Related Search

Related Contents

Notice d`utilisation  Indice  Fluidmaster 7510 Installation Guide  User manual  l.1"LI~~~r~~r:J~ !E~ llrf.` ~e3r~! 1=~LIJ`~Wli~~I~~~~ :J~`JLI rf.`LlJlj  アブソデックス高応答タイプ  ROHO® Hybrid Elite® Cushion Operation Manual  Hotte aspirante Cooker Hood Campana extractora Exaustor  第一種フロン類回収業者登録申請について 業務用冷凍空調  Ecoreco Owners manual  