MLS-A Version 2.0
Contents
1. NJ KERN amp Sohn GmbH WAAGEN GEWICHTE BALANCES WEIGHTS Ziegelei 1 Tel 49 0 7433 9933 0 SY D 72336 Balingen Fax 449 0 7433 9933 149 E Mail info kern sohn com Internet www kern sohn com Service manual Electronic Moisture Analyzer KERN MLS A Version 2 0 5 2010 GB MLS A SH e 1020 cc N 10 11 12 13 14 KERN MLS A Version 2 0 5 2010 Service Manual Electronic Moisture Analyzer Table of Contents BASIC NIOMA OM TT T ET 3 BEAU OC UC 0 uu uy n 3 SAG MERE NIMM M 4 Principle of iio i D nen OR va Ex Ens a ee db en n i sss 5 Appliance OVerVIeW niei Eros sein ean rece ans acta ev tee eases ee Pea eU eee 6 SEGUI Ol M Gt 7 2 OVE CW P uuu uq mua ER 8 Alialyzer Menu suu uuu u isa 10 SUD SI visoko A 10 2 MEN u tmm 11 sv ADOCOUNT S MMC a 12 APP AL REM ME 13 6 S Factory NMellluuu u u zu a nus sum uu Umm 14 Changing Healel lu uu uuu ul ute e Di un chusu aao paw ako kasus 16 Changing the Temperature Sensor 20 Temperature Calibration with Temperature
2. 2 JE ago PEFS m w z Eug woe iii 0 145 w a Inm Qmm 57 m B we Fs i 2 s io Dm oe j m gt cers iai 4g Hm I s r anoss a u 1SQH I asn asn 5609 168354 pinbr SISSPUj Josuas JAWA DZT t queg DOT Ld ICE JO IBUUOD 2331 JVAOZI NEM OOF A 10 HEM OOF p n gt ouuasuL 2006 app B OVA s OZT Ovz E DVA0ZI F dwee 2vAOPZ asn4 NO jaueg _ MLS_A SH e 1020 32 y Br PERTH X CONNECTION 2 M ad u y e 74 j j n loud 44 LI 4 OG SP pth I Dd dad hd tad dod bd vdd M EF W W UT Nw PS 19 19 a gt L M C I yw S 7 p s de ers 2 lra LL UNE 2 4 PRINTED CIRCUIT BOARS d al i A t E M 52 A B MLS A SH e 1020 gt 14 Spare Parts Part number Description UNE Top Pan support 34 MLS A SH e 1020
3. 3 Safety HOT SURFACES e During test the cover of the heating chamber can become very hot especially near the top vent Keep all material away from the vent area Do not touch hot surfaces e Samples will be hot after a test Do not use bare fingers to pick up any samples pans or touch the materials The internal surfaces of the weighing chamber may be hot for some time after a test MATERIALS e The MLS A moisture analyzer is not suitable for all materials During the drying process samples will be heated to high temperatures Materials that are flammable or explosive should not be tested e Materials that emit noxious fumes should only be tested in suitable fume hoods or with other ventilation e Materials that will have a film on the surface during drying may cause high internal pressures to be present These materials should not be tested using this method HIGH VOLTAGE The analyzer has high voltage inside and at the heating lamp This high voltage can be lethal When servicing the analyzer use a voltage source protected by a RCD residual current device circuit breaker Test circuit breaker before proceeding with any test Be aware of the dangers and only perform service work you have been trained to do If in doubt contact KERN or your supplier 4 MLS A SH e 1020 4 Principle of Operation Refer to the MLS A user manual for details of operation The basic principle of the moisture analyzer is to weigh a sample in its n
4. Calibration Set 24 Error MESSAGES MEE ETE ETT TET 27 Load C O u CE OR UD RE REGN 28 URN MESS y T RC PN 28 11 2 Beblacing the Load u naa MUS NEU p ha at NA MON UR ans 29 EIGCITODIOS iens dM LED IM Sd IDE uama 31 System DIAGF AI eet et 32 SDS Paris idit terre OLEI Se ene ee 34 MLS_A SH e 1020 1 Basic Information Grundlegende Hinweise The device must be repaired only by trained specialist staff or personnel with professional formation such as a repair specialist accredited by law concerning verification The service manual is obligatory for repair work After repair original conditions of the device have to be restored Only original spare parts should be used Das Gerat darf nur von geschultem oder beruflich ausgebildetem Fachpersonal z B eichrechtlich anerkannter Instandsetzer repariert werden Die Serviceanleitung ist bindend fur Reparaturen Das Gerat muss nach erfolgter Reparatur wieder in den Originalzustand zuruckversetzt werden Es durfen nur Originalersatzteile verwendet werden 2 Introduction This Service Manual will guide you through the normal service and calibration procedures necessary to maintain the moisture analyzer Please read this Manual thoroughly before performing any service This manual should be used in conjunction with the User operation manual MLS_A SH e 1020 3
5. Display in moisture analysis mode Second digital display for indication of the time sequence the current temperature and the moisture percentage in 96 III L LI To switch the display to the various units of the measured result press the key MLS A SH e 1020 7 5 2 Keyboard Overview MLS Button Description Funetion in Weighing mode in Funetion in Weighing mode mode Turn on off Taring TARE e Zeroing Cancel process input e Return to weighing mode en e Switch over display e Increase flashing digit Arrow key Scroll forward in menu Call up drying processes 8 MLS_A SH e 1020 M Store call up drying parameters At numeric input digit selection to the right Arrow key gt In the menu block SETUP one menu step back MENU key Call up configuration menu Arrow key Confirm store settings Fey Fey Invoke user menu drying parameter setting Scroll backwards in menu Arrow key V Decrease flashing digit PRINT PRINT button Calculate weighing data via interface Arrow key At numeric input digit selection to the left START Start Stop e Start Stop drying STOP button Numeric entry Button Description Function O Arrow key Increase flashing digit gr Arrow key V Decrease flashing digit em Arrow key Digit selection to the left amp wu Arrow key gt Digit select
6. cover plate 16 MLS_A SH e 1020 Lift the cover to clear the ring then lift and rotate it Place it to one side The temperature probe will be loose place it so it cannot be damaged during the rest of the procedure m Retaine 4 gt p 3 amp Using needle nose pliers remove the lamp cables from the high temperature switch and the loose wire To remove the lamp remove the one screw on the retainer MLS A SH e 1020 17 The lamp can be now removed To install the new lamp reverse the order of assembly 18 MLS_A SH e 1020 When installing a new lamp do not touch the glass with your bare fingers wear cotton gloves or use a tissue to prevent finger oils or dirt from being transferred to the lamp Ensure the plastic covers over the wires are in the correct location They have 2 pins that align with holes in the cover and must sit firmly over the wires When attaching the lamp retainer to not over tighten the screw Make sure all connections are tight and the wires are positioned so they do not interfere with each other or the action of the door opening The temperature sensor locates on the lamp support position the sensor near to its final location install the cover then make sure the sensor is fitted to the lamp support and is positioned in the centre of the heating area If the plastic protection avers the wires are not fitted correctly the cover will not fit properly MLS A SH e 1020
7. up phase of the first temperature point can be reduced using the MENU button if the temperature constancy is reached before Then enter the temperature value as described before e With the error message TEMP CAL FAIL or inadmissible temperature inputs the appliance returns to weighing mode Calibration will be interrupted repeat calibration procedure al e lf the temperature value for the second point is not entered within 30 minutes the calibration will be interrupted MLS_A SH e 1020 10 Error Messages Symptom Error Message Probable Cause Remedy ERROR TEMP Temp sensor fault Check connector or sensor short circuit ERROR ZERO Pan not fitted or load cell damaged ERROR TEMP HIGH Temperature sensing over 160 deg C Either PT100 fault or solid state relay not switching lamps OFF LID OPEN When lid is closed Micro switch failed TEMP FAIL While carrying out a temp calibration The calibration is cancelled TEMP FAIL will be displayed and neither new figures will be stored Indicates overload at the pan or load cell accompanied by an audible BLEEP damaged MLS A SH e 1020 27 11 Load Cell If the results have been unreliable not repeating or the display is unstable when showing weight the problem may be associated with the load cell First thing to check is that there is nothing to interfere with the movement of the load cell the wires and cables are not touching the pan support
8. 19 8 Changing the Temperature Sensor The temperature sensor used in the moisture analyzer is a PT100 type this means the probe measures 100 Ohms at 0 deg C so at room temperature it should give a resistance reading of about 110 Ohm Only a PT100 supplied by KERN should be used to replace a faulty probe The main reason for changing the probe would be an error message displayed on the MLS A display ERROR TEMP displayed usually means the probe or wiring has a break somewhere in the wires and has lost its continuity check the probe connector is properly attached to the PCB before commencing probe replacement 1 Ensure the moisture analyzer is disconnected from the mains power supply before starting work 2 Remove all four MA case screws 2x underneath at the front and 2x at the rear 3 Lift the upper case clear of the base and disconnect all the connectors from the PCB at the front of the base 4 Remove the M4 nut on the Earth post at the RH rear of the base and remove the earth cable 5 Lift the upper case assembly clear of the base 6 Stand the upper case on the bench the right way up open the lid and remove the glass and the glass retainers 3 small screws secure the retainers 20 MLS_A SH e 1020 7 Remove the 2x screws securing the closing plate and pull the plate forward from the top this gains access to the overheat switch the lamp fixing and the PT100 Heating Bulb Fixin Over heat sw
9. SH e 1020 8 The display appears 9 Press the key two times The display appears 10 Press the key The display appears DEALER 11 Press the key The display appears SUPERVISOR 12 Press the key two times The display appears FACTORY 13 Press the key to enter the factory menu 14 Use the key to select the required option and confirm with key Factory Menu Menuitem Choice Description MODEL OEM 50g 200g CALIBRATE LOAD 0g LOAD 20g LOAD 50g SERIAL ID Serial number e g OWP0901001 STABILITY 10d 1d 2d AUTO ZERO ON OFF AUTO ZERO 2d 5d 1d 15 Press the VS key to return to normal weighing MLS A SH e 1020 15 7 Changing the Heater To replace the heating lamp over temperature switch or temperature sensor you must use the correct replacement parts supplied by KERN Before proceeding check the voltage of the replacement is correct for the unit A heater that is the wrong voltage will not work properly and can be a hazardous Remove the power cable from the back of the machine There are live circuits associated with the lamp Power must be removed in order to do this repair V lt 2 screws over u gr s Remove 3 screws securing the glass support brackets Remove the cover over the bottom of the ring and the glass Remove the 2 screws securing the cover Remove only the 2 screws in the corners as shown above Do not remove the other 3 screws on the
10. and the temperature is stable The range of the ADC counts will depend upon the load cell used and if the pan support is in place or not Typical values are less than 1 000 000 with no mass on the pan e g 428 295 and about 400 000 greater with 50 grams on the pan support e g 753 482 The temperature ADC counts are normally stable within 10 counts if the temperature sensor is at a stable temperature The value seen is typically 200 000 to 400 000 counts e g 289 852 it will depend upon normal variations in the sensor values Pressing the key escapes to the next item on the dealer menu Pressing VSS key returns to weighing MLS A SH e 1020 6 2 2 CALIBRATE This menu offers two options calibrate the analyzer for correct weight readings CAL MASS or to calibrate the temperature sensor in the test chamber to control heating temperatures CAL TEMP gt not recommended 6 2 2 1 Dealer Balance Calibration CAL MASS e Performs a weight calibration You must have 2 masses available 20 gram and 50 gram for the MLS 65 3A e Press the key to begin e Remove the sample pan from the pan support The calibration is done without a sample pan in place e The display will instruct you to set the no load condition LOAD 0g Check the pan support is empty e When stable stability mark appears press the key The display will ask for the first mass LOAD 20g e Place 20g on the pan support e When stabl
11. ced it is only necessary to plug the new display into the cable and mount the display MAIN PCB The main PCB has an analog section consisting of the analog to digital converter ADC connected directly to the load cell inputs digital microprocessor section input outputs for the keypad and display RS232 interface and a power supply section If a PCB replacement is required contact KERN to order the correct PCB to use It will be sent already programmed into the PCB It will require the user to simply replace the old PCB with the new and then do the Dealer calibration as described in section 6 2 2 1 OTHER ELECTRONIC PARTS The only other parts of note are the Solid State Relay SSR This relay is controlled by the main microprocessor to turn the heater on and off at the correct times to control the heat If a SSR should require replacement it should be replaced with a SSR supplied by KERN TEMPERATURE SENSOR PT100 The temperature sensor inside the weighing chamber is a PT100 type resistance temperature sensor The PT100 sensor has a nominal resistance of 100 ohm at 0 C increasing with temperature To test the sensor simply measure the resistance of the sensor it should be near 110 ohm at room temperature and increase as you warm it MLS_A SH e 1020 31 13 System Diagram NOISIASY weibeig tu ls s 1 dim CEOL SN pelt distin me HE as sassa M a tell
12. e stability mark appears press the key The display will ask for the second mass LOAD 50g e Place 50g on the pan support e When stable stability mark appears press the key The calibration will be completed and the display will return to the CAL MASS parameter e Remove the weight MLS A SH e 1020 13 6 2 2 2 Dealer Temperature Calibration CAL TEMP gt not recommended To calibrate the balance using temperatures measured in the chamber use the temperature calibration as described in section 9 This was just a factory setting which should only be used by the factory for specific PT100 probes Please just use the temperature calibration with temperature calibration set 6 3 Factory Menu Access to factory menu is controlled by password 14 The display appears Pressing the key while in normal weighing gives access to the supervisor menu Use the arrow keys to select PASSCODES and confirm with key Use the arrow keys V to select OPERATOR and confirm with key The display appears Use the arrow keys to enter the currently set password and confirm with E key only at first input acknowledge L If the password has been forgotten enter 0000 15 as emergency password NEW CODE 1 I Use the arrow keys to enter the dealer passcode which is ie 18 DLR as D 4 1 12 R 18 letter of the alphabet Press the key MLS_A
13. e PC Begin a timed test The analyzer will output the weight to the PC observe the weight during the time of the test The weight should be constant within about 50mg during the time the test is running If the drift is much worse than this it is an indication the load cell has been damaged If the drift is excessive but less than about 100mg it may be the sample pan or coin has some moisture on it that was driven off It would be wise to repeat the test before making a decision to replace the load cell 11 2 Replacing the Load Cell The load cell used in the MLS A has been especially selected for low drift performance It should only be replaced with another load cell supplied by KERN CAUTION The load cell is very fragile due to the fine resolution requires and therefore it can be damaged very easily Do not use excessive force to remove or replace screws Do not press down on the screws to remove or install them To remove the load cell it is necessary to remove the pan support 2 screws then the cover over the load cell 2 screws Unsolder the wires noting the colour at each connection Hemove the load cell support from the base Handle the support by the casting only not the load cell The load cell is most vulnerable when it is being handled in this way Remove the load cell from the casting Before installing the new load cell loosen the overload stop screws and lower the stops a few turns This prevents them from
14. hese parameters is controlled by password 1 Pressing the key while in normal weighing gives access to the supervisor menu 2 Use the arrow keys V to select PASSCODES and confirm with key 3 Use the arrow keys V to select OPERATOR and confirm with key 4 he display appears 5 Use the arrow keys to enter the currently set password and confirm with key only at first input acknowledge L If the password has been forgotten enter 0000 16 as emergency password NEW CODE 6 The display appears 7 Use the arrow keys V to enter the dealer passcode which is G4 Ie 18 DLR as D 4 1 12 R 18 letter of the alphabet Press the 8 The display appears 9 Press the key two times The display appears MLS_A SH e 1020 11 10 11 12 13 14 Press the key The display appears DEALER Press the key The display appears SUPERVISOR Press the key The display appears DEALER Press the key to enter the dealer menu Use the key to select the required option and confirm with key Available options on the dealer menu are 6 2 1 ADC COUNTS The analyzer displays the weight ADC counts or the heater temperature sensor ADC counts 12 Pressing the key switches the display between the weight LOADCELL and temperature HEATER ADC counts The weight ADC counts are normally stable within a few counts if the load cell is functioning correctly the environment is stable
15. ion to the right gt E MLS A SH e 1020 B 6 Analyzer Menus The MLS A has a number of submenus for setting the operation of the analyzer and the communications In addition to the Supervisors menu for setting the analyzer to the needs of the user there is a Dealer MENU that allows a service person to calibrate the analyzer observe the ADC counts from the load cell The main topics of the Supervisors Menu are PASSCODES to control unauthorised access to the menu system SETUP Setting of date and time Date format YMD MDY DMY User ID Key Buzzer On Off Test Buzzer On Off Backlight On Off Auto Filter and Auto zero settings Language Selection SERIAL Baud rate Parity Continuous output Time for continuous output CALIBRATION Weight Calibrate Temperature Calibration Cal Report On Off 6 1 Supervisor Menu Pressing the key while in normal weighing gives access to the menu Details of this menu are given in the user manual 10 MLS_A SH e 1020 6 2 Dealer Menu The dealer has access to a menu to allow calibration of weight or temperature sensor without the normal limitations put on the calibration during the user methods of setting them In addition it is possible to view the values delivered from the analog to digital converter for both the chamber temperature and the weight This section describes the parameters available to the dealers for setting up the analyzer Access to t
16. itch Plate forward To Remove att NOTE The overheat switch is designed to cut the power to the lamp if the control circuit fails and the software can no longer control the heater it is designed to operate at chamber temperatures of around 240 deg C If the switch was to fail OPEN then the lamp would not light Other causes of lamp failure could be the lamp itself being faulty or either of the micro switches either failing or being incorrectly adjusted Any of these faults would need the rework method described 8 To remove the PT100 slide out the cable securing clamp from the PT100 side if the hinge the PT100 should now pull out from the wire location 9 Close the lid carefully and turn the lid assembly upside down MLS A SH e 1020 21 10 The 4x large M5 screws on the underside of the upper case assembly hold the heating assembly to the upper case remove these screws and the P clips holding the cables in place Earth Connection oo eee Z Stars a SSSR q Cb gd 11 Lift the upper case away from the heating assembly carefully pulling the cables through the case 12 With the lid assembly on the bench remove the M4 screws securing the large heat sink plate also disconnect the EARTH connection Lid Damper Spring 22 MLS_A SH e 1020 13 The removal of the plate reveals the micro switches lid damping spring and hinge assembly and allows for the PT100 to be remo
17. ormal condition then apply heat to drive off any volatile components This usually refers to moisture but in some products it may be alcohol oils or other volatile materials It is the operator s responsibility to understand the materials to be used in the moisture analyzer and the effects of heat on the materials The material may need to be prepared before a successful test can be done It may be the material will need to be ground to a powder spread between sheets of glass fibre paper or otherwise prepared If the material is not uniform a number of tests may need to be done to find an average moisture content See user manual for more suggestions of how the samples might be prepared for testing MLS_A SH e 1020 5 5 Appliance Overview ltem Oo oO O N Description Foldable hood with heating element Temperature sensor Sample dish Adjustable foot Display Keyboard T 8 101112 Item Description 7 opirit level 8 HS 232 interface 9 Fuse 10 Not documented 11 Mains cable connection 12 Master switch MLS A SH e 1020 5 1 Overview of Display gt 0 lt Date y n A A i i A GICAL Custom BPTG 1 VJ V LAU LA 01 Q In Pos MS z P oy kare 4 5 b Display Description gt 0 lt Zeroing display w Stability display Net Net weight z Busy Display in weighing mode Status display in German and English current temperature WEIGHT 23 c rere LILILILI
18. putting a load on the load cell accidentally during the reassembly Heassemble in the opposite order being very careful not to stress the load cell while tightening screws or handling the support Before the load cell and pan support are secured place the cover over the base to make sure the pan support aligns with the hole in the weighing chamber This hole must align or the pan may rub on the sides of the cover causing incorrect weight to be shown MLS A SH e 1020 29 Secure the load cell solder the wires and retest by observing the ADC counts as described above After the load cell is secure the overload stops can be set lt is not necessary to have the analyzer operating during this process Place the pan support in place and put a sample pan on it Add a mass equal to the capacity of the analyzer Carefully raise the overload stops until they just touch the plastic pan support then back off turn Lock in place with the lock nut Repeat for the opposite side Reassemble the cover to allow the display and keypad to work Apply power to the analyzer Verify the stops are not touching by moving the mass around the pan front back right and left while watching the weight display There should be no signs the pan support is rubbing against the stops or hitting the stops prematurely 30 MLS_A SH e 1020 12 Electronics DISPLAY PCB The display is a custom LCD designed for the purpose lf the display should be repla
19. the pan is lined up with the opening in the case the stainless steel shield under the pan is installed correctly and not rubbing on the bottom of the pan If this is all OK then go to the Dealer Menu Section 6 2 and select ADC counts The ADC counts should be stable within 3 5 counts regardless of weight on the pan If the counts are less than about 100 000 divisions with no load on the pan or greater than about 1 million with 50grams load on the pan may mean the load cell has been damaged and the offset voltage is out of specification The ADC counts do vary as it depends upon the individual load cell and ADC converter If the load cell has been damaged it will not be stable for weighing or it may exhibit excessive drifting during a heating test 11 1 Test Method A method to test the performance of the load cell while heating the chamber Set up a test method with the following settings 1 Moisture Result oingle heat Single 160 Interval 05 seconds Time Stop Stop Time 30 minutes os See Je Manual Start Print Test On eo 9 Format Comp oet the RS232 parameters to do a continuous printout at 5 seconds Interface the analyzer with a terminal program HyperTerminal for example to view the results of the test 28 MLS A SH e 1020 Place a sample pan on the pan support and tare the weight Place a metallic object of approximately 5grams coin for example on the pan support observe the weight on th
20. ved in full pull the PT100 cable through the hinge opening 14 Replacement is a straight reversal of all disassembly operations care must be taken to ensure no cables are trapped and that the cables are correctly routed under the heat sink plate e gt wo Een 2 E E Route cables sides lik 15 Once reassemble the earth continuity should be confirmed either by using a tester or by checking the continuity with a multi meter Only qualified testers should use the PAT tester If the multi meter is used check continuity between the outer reflector and the earth pin on the mains chord MLS_A SH e 1020 23 9 Temperature Calibration with Temperature Calibration Set We recommend sometimes to check the temperature value of the device using the optional temperature calibrating set MLS A04 Before you do this allow the device to cool down for at least 3 hours after the last heating phase Push the probe as closely as possible to the thermal sensor of the MLS The temperature is measured at two points and it is possible to correct it at both temperature points Preparation Push fixing clip under the cover of the weighing chamber according to illustration Insert the probe into the clip Push the probe as closely as possible to the thermal sensor of the MLS Fix the position using the side screw Sample dishes must be in place Use the VS button to cancel the calibration process at an
21. y time 24 MLS A SH e 1020 n weighing mode press the button the first menu item SERIAL is displayed SERIAL Press arrow keys V repeatedly until CALIBRATE is displayed CALIBRATE gt Confirm by pressing the button Use the arrow keys to select CAL TEMP and confirm using button CAL TEMP Calibration starts Every minute sounds the acoustic signal in order to display the permanent calibration CAL TEMP gt Temperature calibration of first point will take 10 min after which you will hear an acoustic signal The display for entering the first temperature point will appear MLS A SH e 1020 25 26 Read temperature on the MLSA 04 Enter the temperature value using arrow keys the active position flashes FIRST 24 Confirm by pressing the button Calibration of the second point will start Heating up to approx 150 C SECOND 27 LiLI LI I Temperature calibration of the second will finish approx 20 min then you will hear an acoustic signal Press the temperature point will appear 29 button the display for entering the second SECOND 150 BENIN I Read temperature on the MLS A04 Enter the temperature value using arrow keys V and confirm with button The appliance returns to weighing mode TARE Otherwise press the VSS key WEIGHT BEE LI LI C 23 ILI E g e The 10 minutes heating