1. Install the new mainboard Re install the modem Re install the paper drive assembly Note Ensure the cables to the power supply board stay securely connected while installing the paper drive assembly 8 Re install the top cover assembly 9 Re install the battery pack 10 Perform the following tests before returning the Atria to service e Functional Checkout on page 2 2 Electrical Leakage Current Tests on page 2 4 e Queue Sensor Calibration on page 2 6 70 00083 01 A Atria 3100 6100 Replace the Atria 6100 Mainboard including Video Adapter Board and Modem The mainboard includes the CR2032 battery but not the video adapter board or the modem If the video adapter board or modem must be replaced disassembly is the same to the mainboard except do not disconnect the video display cables to replace the modem 1 Remove the battery pack see Remove the Battery Pack on page 4 12 2 Remove the top cover assembly see Remove the Atria 6100 Top Cover Assembly on page 4 15 3 Remove the paper drive assembly see Remove the Paper Drive Assembly on page 4 17 4 Remove the video adapter board modem and then the mainboard see Remove Atria 6100 Mainboard Assembly on page 4 20 5 Install the new mainboard 6 Re install the video adapter board 7 Re install the modem 8 Re install the paper drive assembly Note Ensure the cables to the power supply board stay securely connected while installing the paper drive ass
2. The Atria electrocardiograph is a multichannel electrocardiograph with a thermal printer Options may include ECG interpretation and modem or network connectivity System Description The Atria has the following main subsystems Atria 3100 6100 Mainboard contains the host and front end The video adapter board Atria 6100 only and optional modem plug directly into the mainboard Power Supply Board Receives input power from the external power supply or battery pack supplies power to the Atria and recharges the battery pack when connected to external power Paper Drive Assembly Includes the printhead queue sensor to correctly position the paper and the paper drive motor to feed the paper The Top Cover Assembly Contains the keyboard and video display 70 00083 01 A 4 4 Paper Tray Assembly Printhea 100 240 VAC In External Power Supply Battery Pack d Digital Interface VY Equipotential Ground Top Cover Assembly Top Cover Assembly Atria 3100 Video Display Keyboard Character Display Keyboard Display Interface Power amp Digital Control Backlight GND J2 P1 JP23 Video Adapter Board Video Adapter Board Atria 3100 Mainboard J412 JP409 Mainboard JP412 JP409 Modem Optional Modem Optional Figure 1 Atria Subsystems Technical Specifications Specification Dimensions Weight unit only
3. 15 0 x 13 125 x 5 5 381mm x 334mm x 140mm 11 Ibs 5 kg including external power supply Display Atria 3100 2 x 40 character backlit LCD screen Atria 6100 640 x 480 backlit color LCD screen Keyboard Full alphanumeric keypad plus designated quick keys Data Storage Atria 3100 50 records standard Atria 6100 150 records standard optional upgrade to 300 Input Power AC Operation 100 240 VAC 10 50 60 Hz 3 Hz Battery duration 55 minutes continuous printing Printout Printout device 216 mm thermal dot array Paper dimension 8 5 x 11 US letter 210mm x 300mm A4 Paper type Thermal sensitive Burdick Assurance or HeartLine paper recommended Chart speeds 12 5 25 50 mm sec 1 2 70 00083 01 A Atria 3100 6100 Specification Gain 5 10 20 mm mV Chest or Limb may be split Printout formats 3 4 6 or 12 channels additional rhythm formats Acquisition Lead selection II IIl aVR aVL aVF V1 V2 V3 V4 V5 V6 and Alternate Chest Lead chest lead selection V2R through V9R V7 V8 and V9 Modes Frequency response Automatic automatic rhythm or manual rhythm Meets or exceeds ANSI AAMI EC11 1991 and IEC 60601 2 51 standards Input impedance Meets or exceeds ANSI AAMI EC11 1991 and IEC 60601 2 51 standards Electrode offset tolerance 300 mV Sampling 8000 Samples Sec 2 5 LSB Artifact filte
4. Ensure green LED on external power supply is on Ensure the power outlet is working Replace power cord Replace the external power supply These solutions must be performed by qualified service personnel Replace power supply board Atria unexpectedly reboots crashes 3 2 70 00083 01 A The Atria will shut off if both AC and battery power are interrupted at the same time The Atria must have a battery installed during normal operation f operating from the battery ensure the battery is charged and the cable securely connected e f operating from AC power ensure the battery is installed and AC power is available After a crash insert a USB flash drive and select Setup Service Functions Copy crash dump to USB A text file is created that can be sent to Technical Support for analysis Atria 3100 6100 Problem Solution Video Display e Ensure a battery is installed and check the green power LED No display on the back panel to verify AC power is available e Display is too dark or too light Try adjusting the brightness SHIFT NEXT or SHIFT PREVIOUS Press and hold the ON STBY key for 8 seconds to power off and then press ON STBY again to restart the Atria Display is unstable These solutions must be performed by qualified service personnel Ensure keyboard cable is securely connected Atria 6100 only Ensure the display and backlight cables are securely conn
5. Toll Free 1 800 426 0337 Telephone 1 425 402 2000 Email techsupport cardiacscience com Website www cardiacscience com 1 2 70 00083 01 A CE mr 0086 MDSS GmbH Schiffgraben 41 D 30175 Hannover Germany Tel 49 511 62 62 86 30 Fax 49 511 62 62 86 33 Atria 3100 6100 Contents Chapter 1 General Information About the Atria Electrocardiographs esses 1 1 System 1 1 Technical Sp amp cHlicatlQu essais tpa Ee pe 1 2 Chapter 2 Service amp Maintenance When to Perform Service Procedures eee 2 1 Functional Yea EET N 2 2 Required Equipmientu ur ene UI ER DU MI NR EEUU 2 2 Functional Checkout Procedure esses 2 2 Calibration Pulse Check Procedure ess 2 4 Electrical Leakage Current 2 4 Required 2 4 Patient Leakage Current Procedure eese 2 5 Patient Sink Current Procedure uoi estere ro eo eiie 2 5 Queue 2 6 Calibration Procedure acc 2 6 Chapter 3 Troubleshooting Preliminary Ga T rri ends 3 1 Troubleshooting CAE aper
6. anti static mat and wrist grounding strap when working on internal components Caution Burn hazard Components inside the unit may be hot after use Use caution when working on internal components The following section lists the steps to replace each Atria component Replace the Top Cover Assembly The top cover assembly includes the top cover with labels for either Burdick or Quinton keyboard assembly and the video display Note For the Atria 6100 The top cover assembly may be ordered without a video display 1 Remove the battery pack see Remove tbe Battery Pack on page 4 12 2 Remove the top cover assembly see Remove the Atria 3100 Top Cover Assembly on page 4 14 or Remove the Atria 6100 Top Cover Assembly on page 4 15 3 For the Atria 6100 a Remove the paper drive assembly see Remove tbe Paper Drive Assembly on page 4 17 b Disconnect the video display cables see Figure 5 on page 4 20 a and b c Connect the new video display cables d Re install the paper drive assembly 4 Install the new top cover assembly Re install the battery pack Apply the Burdick or Quinton label the new label must be the same as the label on the old top cover 7 Perform the following tests before returning the Atria to service e Functional Checkout on page 2 2 Electrical Leakage Current Tests on page 2 4 e Queue Sensor Calibration on page 2 6 70 00083 01 A Atria 3100 6100 Replace the Paper Drive Assembly
7. A Atria 3100 6100 3 Remove the battery access cover a Slide the plate toward the front of the Atria b Lift up the back end of the plate to remove 4 Remove battery pack a Slide the battery pack out of the battery compartment Caution Equipment damage Do not pull on the wires to disconnect the battery pack connector This may cause the connector to fail Only pull on the connector to disconnect the battery pack b Disconnect the battery pack connector 5 For the Atria 3100 only a Remove the ferrite from the old battery pack by unsnapping fasteners b Place the ferrite around the new battery pack cable approximately in the middle of the cable Atria 3100 6100 70 00083 01 A 4 13 Remove the Atria 3100 Top Cover Assembly Caution Burn hazard Components inside the unit may be hot after use Use caution when working on internal components To remove the top cover assembly 1 Turn the electrocardiograph over and remove the two bottom assembly screws Carefully turn the electrocardiograph back over Remove the paper door and the paper see Remove the Paper Door on page 4 23 4 Remove the two screws and the paper door rails from the paper well 5 Lift the top cover assembly and disconnect the keyboard cable and the video cable 4 14 70 00083 01 A Atria 3100 6100 Remove the Atria 6100 Top Cover Assembly Caution Burn hazard PA Components inside the unit may be hot after
8. Website www cardiacscience com Email techsupport cardiacscience com
9. uet Cio E arp Beck ir emit eee 3 2 Chapter 4 Component Replacement Parts 4 1 Component Replacement ss s civcsecececsnncnaseveseneesiauncssssevessvapiesexecens 4 6 Replace the Top Cover Assembly eee 4 6 Replace the Paper Drive Assembly totiens 4 7 Replace the Atria 3100 Mainboard including Modem 4 8 Replace the Atria 6100 Mainboard including Video Adapter Board and Modem 4 9 Replace the Power Supply Board rect rotate 4 10 Replace the External Power Supply sire Ett eso 4 11 70 00083 01 A Atria 3100 6100 1 4 12 Remove the Battery 4 12 Remove the Atria 3100 Top Cover Assembly 4 14 Remove the Atria 6100 Top Cover Assembly 4 15 Remove the Paper Drive 4 17 Remove Atria 3100 Mainboard Assembly 4 18 Remove Atria 6100 Mainboard Assembly 4 20 Remove Power Supply Board esses 4 22 Remove the Paper Door ssssscccossressssseneccesnonscccssnseeeceses 4 23 Atria 3100 6100 70 00083 01 A 1 General Information What s in this chapter About the Atria Electrocardiographs 1 1 System Description 1 1 Technical Specifications 1 2 About the Atria Electrocardiographs
10. use Use caution when working on internal components To remove the top cover assembly 1 Turn the electrocardiograph over and remove the two bottom assembly screws Carefully turn the electrocardiograph back over Remove the paper door and the paper see Remove tbe Paper Door on page 4 23 4 Remove the two screws and the paper door rails from the paper well Atria 3100 6100 70 00083 01A 4 15 5 Lift the top cover assembly and disconnect the keyboard cable 6 Remove the top cover assembly and set upright behind the Atria Note The two video display cables will still be connected 4 16 70 00083 01 A Atria 3100 6100 Remove the Paper Drive Assembly To remove the paper drive assembly 1 2 Atria 3100 6100 Remove the printer assembly screws 1 and 2 Lift out the paper drive assembly and rotate clockwise 180 to set aside Note When reinstalling the paper drive assembly be sure to push the paper cover fully to the position pins on the print head bracket 3 before tightening the screw Note The paper drive assembly power cables remain connected to the power supply board 70 00083 01A 4 17 3 Disconnect the cable connectors from J3 2 J4 1 J6 3 and J7 4 on the power supply board Remove Atria 3100 Mainboard Assembly The mainboard assembly includes an optional modem board B Note The modem is connected to the mainboard with four multi pin connector
11. Atria 3100 6100 Remove the battery pack see Remove the Battery Pack on page 4 12 Remove the top cover assembly see Remove the Atria 3100 Top Cover Assembly on page 4 14 or Remove the Atria 6100 Top Cover Assembly on page 4 15 Remove the paper drive assembly see Remove the Paper Drive Assembly on page 4 17 Install the new paper drive assembly Note Ensure the cables to the power supply board stay securely connected while installing the paper drive assembly Write the printhead resistance written on the top of the printhead onto the printhead resistance label Attach the printhead resistance label onto the enclosure bottom in the battery compartment Re install the top cover assembly Re install the battery pack Perform the following tests before returning the Atria to service e Functional Checkout on page 2 2 Electrical Leakage Current Tests on page 2 4 e Queue Sensor Calibration on page 2 6 70 00083 01 A 47 4 8 Replace the Atria 3100 Mainboard including Modem If the modem must be replaced disassembly is the same to the mainboard 1 Remove the battery pack see Remove the Battery Pack on page 4 12 2 Remove the top cover assembly see Remove the Atria 3100 Top Cover Assembly on page 4 14 3 Remove the paper drive assembly see Remove tbe Paper Drive Assembly on page 4 17 4 Remove the modem if not replacing and then the mainboard see Remove Atria 3100 Mainboard Assembly on page 4 18
12. SERVICE MANUAL ATRIA 3100 ATRIA 6100 PART NUMBER 70 00083 01 A Copyright 2007 Cardiac Science Corporation All rights reserved This document contains instructions for both the Atria 3100 and Atria 6100 TO RESPONSIBLE SERVICE PERSONNEL The contents of this document are not binding If any significant differences between the product and this document are encountered regarding service work contact Cardiac Science Corporation for more information Cardiac Science Corporation recommends the use of authorized Cardiac Science Corporation personnel for the maintenance and repair of all Burdick equipment Cardiac Science Corporation cannot warrant the operation of the equipment if other than Burdick genuine replacement or exchange parts are used in the service or repair of this equipment and if such service or repair is performed by non authorized personnel This product has been carefully designed to provide a high degree of safety and dependability However we can not guarantee against the deterioration of components due to aging and normal wear CAUTION The Atria electrocardiograph is a restricted device Federal law restricts the sale distribution or use of this device to by or on the lawful order of a health professional DANGER Explosion hazard Do not use this device in the presence of flammable anesthetics P N d CARDIAC SCIENCE iu Cardiac Science Corporation 3303 Monte Villa Parkway Bothell WA 98021 USA
13. Verify external cables are undamaged and securely connected Swap out the power cable external transformer and patient cable with known good cables 4 Check for any other visible damage such as cracks in the case broken or cracked cable connectors or rattling when the Atria is moved Atria 3100 6100 70 00083 01 A 3 1 Troubleshooting Chart Problem Solution Artifacts or noise on ECG trace Check the following for general problems refer to the Operating Instructions Troubleshooting section for specific problems and solutions Try to prevent unnecessary patient movement Ensure all leads are clean secure and properly placed Ensure the skin is properly prepared and unbroken Ensure all cables are securely connected Check cables for damage including nicks cuts wear or bulging Position the Atria as far from other electronic equipment as possible Message Waiting for Data is displayed This message is normally displayed for about 15 seconds after beginning an ECG acquisition If it does not go away after about 15 seconds Check the screen for a lead fail indication and then fix that lead Check all the ECG leads Troubleshoot for noise see Artifacts or noise on ECG trace Refer to the Operating Instructions for additional troubleshooting information AC power indicator is off Green LED on back panel Ensure both ends of the power cord are securely connected
14. age and then ensure paper door is properly seated Take paper out and reinstall Follow paper installation instructions in the Operating Instructions Clean queue sensor See illustration below Paper dust may collect on the sensor clean the queue sensor by wiping with a soft cloth Figure 3 Queue sensor These solutions must be performed by qualified service personnel Verify the printhead resistance entered Setup Service Functions Printhead Resistance matches the resistance entered on the printhead e Ensure cables from the paper tray assembly to the power supply board are securely connected Replace cables from the paper tray assembly to the power supply board Replace queue sensor Replace paper tray assembly Replace power supply board Atria 3100 6100 Problem Solution Battery Atria will not operate when disconnected from AC power or only operates for a short time Normal battery life is about 3 hours depending on how the Atria is used if the battery lasts less than half that time or is more than two or three years old it should be replaced To conserve Battery life do not turn on the Atria until the patient is prepared and the leads are connected e Ensure green LED on external power supply is on Check the green power LED on the back panel to verify power is available when charging the battery Ensure the battery is installed and the battery cable
15. attery Pack 14 4 146 0131 00 NiMH 1500mAH Atria 3100 Battery Pack Ferrite 276 0280 00 for the power brick Atria 6100 Battery Pack 14 4V 146 0127 00 NiMH High Capacity ZA WARNING BATTERY MAY EXPLODE Mainboard to Power Supply 175 1607 00 Board Cable E 70 00083 01 A Atria 3100 6100 Part Description and Location Part Atria 3100 Keyboard with Card 60 00040 01 Edge Connector Atria 3100 LCD 14 pin Ribbon 60 00041 01 Cable Atria 6100 Keyboard Cable 60 00039 01 Not Pictured Plastite Screw 4 x 5 16 211 0466 00 Note Do not use non Plastite screws Other types of screws may damage the case Not Pictured Screw M4 X 0 7 X 8 215 0223 00 Not Pictured External Power Supply 010 1684 00 Brick Not Pictured Screw M3 x 6 701950 Not Pictured Screw 6 32x3 8 Pan PH 50 00046 01 Not Pictured Screw 6 32x1 4 Flat 100 PH 50 00045 01 Not Pictured Screw M4x16 Pan PH 017439 065 Not Pictured Screw 4 40x3 19 Flat 100 PH 50 00047 01 Atria 3100 6100 70 00083 01 A Component Replacement AN AN AN 4 6 WARNING Shock hazard Before opening the case ensure the Atria is unplugged and the battery pack is removed If power must be applied while the case is open ensure all personnel and equipment are clear of any energized components Caution Static electricity Several Atria components are extremely sensitive to static electricity Always use an
16. ected The graphic below shows the Video Display cable improperly connected the cable is not fully inserted into the jack and the right tab is not fully engaged to lock in the cable Figure 2 Incorrect Video Cable Installation Replace the top cover assembly Replace the power supply board Replace the mainboard Keyboard Select Setup Service Functions Keyboard Test to verify No response from one or more keyboard operation keys f ON STBY key does not work ensure the battery is installed Key sticks and check the green power LED on the back panel to verify AC power is available Restart the Atria Clean the outside surface keyboard with isopropyl alcohol These solutions must be performed by qualified service personnel e Verify the keyboard cable is securely connected Clean the underside of the keypad and the keypad contacts with isopropyl alcohol Replace the keyboard cable Replace the top cover assembly Replace the mainboard Atria 3100 6100 70 00083 01 A 3 3 Problem Solution Printing Paper does not feed Paper feeds but does not print Printing too light or too dark Out of Paper message in error 3 4 70 00083 01 A Select Setup Service Functions Printer Test Sequence to verify printer operation e Ensure Burdick paper is used Assurance or Heartline Verify paper type in System Settings matches the paper installed Check paper door gear for dam
17. embly 9 Re install the top cover assembly 10 Re install the battery pack 11 Perform the following tests before returning the Atria to service Atria 3100 6100 e Functional Checkout on page 2 2 Electrical Leakage Current Tests on page 2 4 Queue Sensor Calibration on page 2 6 70 00083 01A 4 9 4 10 Replace the Power Supply Board 1 Remove the battery pack see Remove the Battery Pack on page 4 12 2 Remove the top cover assembly see Remove the Atria 3100 Top Cover Assembly on page 4 14 or Remove the Atria 6100 Top Cover Assembly on page 4 15 3 Remove the paper drive assembly see Remove the Paper Drive Assembly on page 4 17 4 Remove the power supply board see Remove Power Supply Board on page 4 22 5 Replace the power supply board Note A rubber spacer must be installed under the screw hole near P1 see the board detail on Figure 7 on page 4 22 Ensure the other rubber spacer to the right of P1 is left intact The spacer may adhere to the power supply board or bottom assembly when the board is removed and is easily lost When reinstalling the power supply board ensure the spacer is in place to prevent the board from warping when the screw is tightened 6 Re install the paper drive assembly Note Ensure the cables to the power supply board stay securely connected while installing the paper drive assembly 7 Re install the top cover assembly Re install the battery pack Perform the followi
18. ia and disconnect the AC power cord 1 Plug the AC power cord into a wall outlet Wait until PRESS ON STBY to power on the unit displays on the Battery Charging screen Press the ON STBY key to power on the Atria Verify the Atria boots up to the Home screen without errors Remove any paper if installed and note the printhead resistance in the battery compartment Select Setup Service Functions Printhead Resistance and verify the resistance matches the printhead resistance label in the battery compartment Note If the mainboard is replaced the printhead resistance must be re entered If the paper tray assembly is replaced a new printhead resistance must be entered and the label in the battery compartment must be changed both should match the printhead resistance written on the printhead Reinstall the paper and verify the paper type matches the type selected in Setup System Paper Type Select Setup Service Functions Printer Test Sequence The Atria prints 3 pages of test characters Verify all characters print evenly across the page Note The test characters may appear lighter or darker depending on the type of paper used but the printing should still be clear and even across the page Press the HOME key Turn on the patient simulator 2 2 70 00083 01 A Atria 3100 6100 10 11 12 19 14 15 16 Atria 3100 6100 Connect the patient simulator to the Atria On the Home screen verif
19. is securely connected Ensure the battery has been charged for at least four hours e Ifthe Atria has not been used for an extended period several months the battery may be fully discharged and will only accept a minimal charge Try disconnecting AC power and then charge the battery again These solutions must be performed by qualified service personnel Replace the battery Replace the power supply board Additional patients cannot be added to the patient directory Normally when the directory reaches the maximum number of patients the oldest records will be deleted as new records are added However this only happens if the records have been printed or sent If none of the records have been printed or sent additional patients cannot be added to the directory Print or send old records Date and time inconsistent or cannot be set These solutions must be performed by qualified service personnel e Replace 3V button cell on mainboard Replace mainboard assembly Communications Errors See the Communications Troubleshooting Help pages ALT H for a list of specific errors General checks include Ensure the serial phone network or USB cables are securely connected to the correct jack Check the communications settings For network problems check with the network administrator to ensure the destination is available to display the IP Address of the Atria select Setup Service Func
20. librated ECG simulator with a 1 mV CAL pulse with rise time of not greater than SmS Verify each lead by connecting the lead to be tested to the input of the appropriate cal pulse connector while all other leads are connected to their appropriate connection points Using AAMI EC11 Fig 3 test circuits with a calibrated source a Create 1 mV Cal pulse by setting the signal source in Fig 3 to 1 volt 196 with a rise time no greater than 5mS and ensure S5 is open while S1 2 and S2 are closed Also ensure S3 is set to position a b Verify each lead by connecting the lead to be tested to the P1 input of the test circuit while all other leads are connected to P2 Electrical Leakage Current Tests Use these tests to verify Atria electrical safety All leakage tests must be performed after any internal repair i e the case is opened or to check for possible damage e g the Atria is dropped Required Equipment Dynatech Nevada 235A Safety Analyzer Note The Dynatech Safety Analyzer is recommended for this procedure if a different safety analyzer is used consult the user manual for proper setup and testing procedures 2 4 70 00083 01 A Atria 3100 6100 Patient Leakage Current Procedure Note Ensure that the Kelvin cable is not connected 1 2 Plug the safety analyzer into the 264 VAC line voltage Plug the Atria power cable into the test receptacle on top of the safety analyzer Connect a patie
21. ng tests before returning the Atria to service e Functional Checkout on page 2 2 Electrical Leakage Current Tests on page 2 4 Queue Sensor Calibration on page 2 6 70 00083 01 A Atria 3100 6100 Replace the External Power Supply 1 Power down the Atria and disconnect external power 2 Remove the ferrite from the old external power supply cable by unsnapping fasteners 3 Place the ferrite around the new power supply cable 2 inches 1 2 inch from the 5 pin din connector 4 Re install the power supply cable Atria 3100 6100 70 00083 01 A 4 11 Disassembly gt 0 5 4 12 The procedures in this section detail removing Atria components Remove the Battery Pack WARNING Fire explosion or contamination hazard Properly dispose of batteries in accordance with local regulations Burning heating or improper disposal may cause explosion fire or contamination Caution Loss of data The battery pack must be installed at all times for proper operation Operating without a battery pack may result in lost or unsaved waveform data Caution Patient injury Do not remove the battery pack within the patient vicinity 1 Disconnect external power WARNING Shock hazard Verify the external power cord is disconnected before removing or replacing battery pack A blank display is not a reliable indication of disconnected external power 2 Pull out the paper door and remove the paper 70 00083 01
22. nt Type Degree of protection from electric shock Class Ila Council Directive 93 42 EEC MDD IEC 60601 1 Class Type CF Defibrillator Proof Mode of operation Continuous Schematics Available on request Warranty 3 years with return of warranty card 1 4 70 00083 01 A Atria 3100 6100 2 Service amp Maintenance What s in this chapter When to Perform Service Procedures 2 1 Functional Checkout 2 2 Electrical Leakage Current Tests 2 4 Queue Sensor Calibration 2 6 When to Perform Service Procedures Perform the following procedures as indicated Atria 3100 6100 Functional Checkout on page 2 2 Used to verify normal operation Must be performed before returning the Atria to the user Electrical Leakage Current Tests on page 2 4 Used to verify Atria electrical safety Must be performed after any internal repair i e the case is opened or to check for possible damage e g the Atria is dropped Queue Sensor Calibration on page 2 6 Used to ensure paper feeds correctly Must be performed if The paper tray assembly is removed or replaced The queue sensor board is replaced The power supply board is replaced The mainboard is replaced 70 00083 01A 24 Functional Checkout Use this procedure to verify normal operation before returning the Atria to the user Required Equipment Patient Simulator Functional Checkout Procedure If necessary turn off the Atr
23. nt cable between patient input connector on the Atria and the appropriate patient terminals on the top of the safety analyzer Press the ENABLE button for the ECG LEADS group on the safety analyzer Turn the SELECT knob to select ALL Verify the leakage current is less than 10 pA for these conditions e Polarity Normal Atria power on e Polarity Normal Open Ground Atria power on e Polarity Normal Open Neutral Atria power on e Polarity Reversed Atria power on e Polarity Reversed Open Ground Atria power on e Polarity Reversed Open Neutral Atria power on Patient Sink Current Procedure WARNING Shock hazard This test applies AC Line Voltage to the patient cable Note Ensure that the Kelvin cable is not connected Atria 3100 6100 Turn the SELECT knob to select ISOLATION in the ECG LEADS group and wait for the numbers to display Press the ISO TEST button and verify that the highest measured value is less than 50 pA 70 00083 01 A 2 5 Queue Sensor Calibration 2 6 Used to measure and record queue sensor voltages based on queue sensor characteristics and internal physical alignment Required for the following conditions The paper tray assembly is removed or replaced The queue sensor board is replaced The power supply board is replaced The mainboard is replaced Note This procedure is not required if the paper door is replaced or after minor adjustments If this calibration fail
24. oard connectors Note When reinstalling the video cable ensure the connector is fully inserted and the side clips are completely engaged The figure below shows an incorrectly installed cable Figure 6 Incorrect Video Cable Installation 4 Remove the 7 screws labeled 1 7 with black arrows on Figure 5 Atria 3100 6100 70 00083 01A 4 21 Remove Power Supply Board Figure 7 Power supply board removal 1 Disconnect the ground connector to P1 2 2 Disconnect the cable to the mainboard 1 Remove the 4 screws one in each corner Note A rubber spacer must be installed under the screw hole near P1 see the board detail on Figure 8 Ensure the other rubber spacer to the right of P1 is left intact The spacer may adhere to the power supply board or bottom assembly when the board is removed and is easily lost When reinstalling the power supply board ensure the spacer is in place to prevent the board from warping when the screw is tightened Figure 8 Power supply board detail 4 22 70 00083 01 A Atria 3100 6100 Remove the Paper Door To remove the paper door Pull the door out and lift the door through the side rail at the back of the _ unit Atria 3100 6100 70 00083 01 A 428 E 70 00083 01 A Atria 3100 6100 Cardiac Science Corporation 3303 Monte Villa Parkway Bothell WA 98021 USA CARDIAC SCIENCE W 70 00083 01 A Toll Free 1 800 426 0337 Telephone 1 425 402 2000
25. r response 40 Hz 3db Storage 500 samples second 2 5 LSB Pacemaker display capability Meets or exceeds ANSI AAMI EC11 1991 standard and IEC 60601 2 51 Interpretation optional Diagnosis measurements reasons statements based on five demographic criteria Environmental Operating temperature 50 F to 104 F 10 C to 40 C Operating relative humidity 20 to 75 non condensing Operating atmospheric pressure 1060 hPa to 700 hPa 500 ft to 10 000 ft reference to sea level Storage temperature 4 F to 113 F 20 C to 45 C Storage relative humidity 10 to 90 non condensing Storage atmospheric pressure 1060 hPa to 190 hPa 500 ft to 40 000 ft reference to sea level Harmful Ingress of Water Normal Input Output Units with communications Ethernet RJ45 USB type A Atria 3100 6100 Standard RS 232 9 pin D 70 00083 01 A 1 3 Specification Analog output 3 5 mm phono plug Optional Upgrades 802 11 Bluetooth Telephone line interface RJ11 Conforms to Standards CAN CSA C22 2 No 0 M91 CAN CSA C22 2 No 601 1 M90 CAN CSA C22 2 No 601 1S1 CSA C22 2 No 601 2 25 UL 60601 EN 60601 1 EN 60601 2 25 amp Amend 1 EN 60601 1 2 AZ NZS 2064 1 2 ANSI AAMI EC11 1991 EN 60601 2 51 Safety Leakage current patient 10 pA chassis 100 pA Defibrillation protection to 5000V 360J Equipme
26. s one in each corner and a nylon standoff When removing the modem the standoff may need to be clipped for easier removal a new standoff is included with a new modem or new mainboard When installing the modem check for bent pins and seat the modem carefully ensure all the pins line up 4 18 70 00083 01 A Atria 3100 6100 Figure 4 Mainboard Removal 1 Disconnect the keyboard cable connector from JP409 labeled d in Figure 4 2 Disconnect the power supply board cable from J412 c 3 Disconnect the LCD Display cable from JP23 located just above the keyboard connector JP409 Atria 3100 6100 70 00083 01A 4 19 4 20 Remove Atria 6100 Mainboard Assembly The mainboard assembly includes the video adapter board A in the figure below and an optional modem board B Note The modem is connected to the mainboard with four multi pin connectors one in each corner and a nylon standoff When removing the modem the standoff may need to be clipped for easier removal a new standoff is included with a new modem or new mainboard When installing the modem check for bent pins and seat the modem carefully ensure all the pins line up Figure 5 Mainboard Removal 70 00083 01 A Atria 3100 6100 Disconnect the keyboard cable connector from JP409 labeled d in Figure 5 Disconnect the power supply board cable from J412 c Disconnect the video display cable and ground a and backlight cable b from the video adapter b
27. s twice contact Technical Support Calibration Procedure Caution Equipment damage Use only Assurance or Heartline standard paper Using any other paper may damage the Atria and void the warranty 1 Insert paper into the paper drive so that the print head is approximately half way between the perforations of the paper 2 Power on the Atria and verify the ECG screen is displayed press the HOME key 3 Select Setup Service Functions Queue Sensor Calibration and press the ENTER key Press HOME Press the FORM FEED key and verify the paper queues correctly Select Setup Waveform Preferences Speed 12 5mm s Press the MANUAL RHYTHM key and print 3 pages of ECG data Verify additional headers are not printed on the pages 8 Press STOP Verify the paper queues correctly 9 Select AUTO RHYTHM Verify a one page report prints and the paper correctly queues Soe Ol P 10 Repeat for waveform speeds of 25 mm s and 50 mm s 70 00083 01 A Atria 3100 6100 3 Troubleshooting What s in this chapter Preliminary Checks 3 1 Troubleshooting Chart 3 2 Preliminary Checks Before using the troubleshooting chart below check the following 1 Ensure users are correctly following instructions listed in the Operating Instructions Verify proper skin preparation lead placement lead integrity etc 2 Check the Atria settings Setup Print Setup Report to ensure the Atria is configured as expected 3
28. tions Get IP Address Atria 3100 6100 70 00083 01 A 3 5 ESSET 70 00083 01 A Atria 3100 6100 4 Component Replacement What s in this chapter Parts List 4 1 Component Replacement 4 6 Disassembly 4 12 This section contains instructions for removing ordering and replacing the Atria subassemblies Parts List To locate components refer to Disassembly on page 4 12 and to this parts list Part Description and Location Part Mainboard 671 1299 00 1 Video Adapter Board 671 1301 00 Atria 6100 only 2 Modem w nylon standoff 671 0001 00 3 Battery CR2032 146 0052 00 Note Mainboard includes CR2032 Battery Atria 6100 The video adapter board and modem must be ordered separately Atria 3100 6100 70 00083 01A 44 Part Description and Location Part Power Supply Board 671 1300 01 Atria 3100 Top Cover Assembly 651 1503 01 E 70 00083 01 A Atria 3100 6100 Part Description and Location Part Atria 6100 Top Cover Assembly 651 1429 00 with Video Display Atria 6100 Top Cover Assembly 551 0002 00 without Video Display Paper Tray Assembly 10 00040 01S Queue Sensor 30 00005 01S Paper Tray Assembly Queue Sensor Atria 3100 6100 70 00083 01 A Part Description and Location Part Paper Door 10 00041 01S Atria 3100 Battery Access Cover 671 0003 01 Atria 6100 Battery Access Cover 671 0002 01 Atria 3100 B
29. y The patient name and ID number or No Name No Id display Note The patient name and ID number display when patient demographic information has already been entered For the Atria 3100 FAIL DRIFT or NOISE messages do not display For the Atria 6100 SENSORS OK message displays e AC displays Disconnect the AC power cord for 1 minute and verify the Atria continues to operate normally AC will change to BAT Note If the Atria shuts down immediately after disconnecting the AC power ensure the battery is installed and securely connected Reconnect the power cord and verify BAT changes back to AC Press the ECG key to acquire and print out an Auto ECG The printout will consist of a 12 lead formatted ECG Verify the following Correct printing of the lead waveforms Stat ID number entered in the ID field Correct date and time when the ECG was acquired Interpretation indicates Normal ECG Press the HOME key Press the DIRECTORY key to verify the ECG is saved in the patient directory Press the HOME key Disconnect the patient simulator 70 00083 01 A 2 3 Calibration Pulse Check Procedure To perform the calibration pulse check if necessary refer to the Operating Instructions apply a cal pulse individually to Leads I H and V1 V6 C1 C6 and verify they reproduce a 1 mV signal referred to input within a tolerance of 596 This can be accomplished by one of the following methods Using a Ca