        Associated Research 3500D Series Hypot II Operation & Calibration
         Contents
1.             20    THERMAL FAN  To cool the instrument     INPUT POWER SWITCH  Line voltage selection is set by the position of the  switch  In the down position it is set for 115 volt operation  in the up position it 1s  set for 230 volt operation     REMOTE INPUT OUTPUT  9 pin D subminiature male connector for remote  control of test and reset funtions as well as for monitoring PASS  FAIL  and  PROCESSING output relay signals     CHASSIS GROUND  EARTH  TERMINAL  This safety terminal should be  connected to a good earth ground before operation     INPUT POWER RECEPTACLE  Standard IEC 320 connector for connection to a  standard NEMA style line power  mains  cord     FUSE RECEPTACLE  To change the fuse unplug the power  mains  cord and turn  the fuse cap counter clockwise to remove the fuse     CALIBRATION ENABLE KEY  To enter the calibration mode press this key  while the instrument is being powered ON     INSTALLATION  Introduction    This section contains information for the unpacking  inspection  preparation for use and  storage of your Associated Research  Inc   product     Unpacking and Inspection    Your instrument was shipped in a custom foam insulated container that complies with  ASTM D4169 92a Assurance Level II Distribution Cycle 13 Performance Test Sequence     If the shipping carton is damaged  inspect the contents for visible damage such as dents   scratches or broken meters  If the instrument is damaged  notify the carrier and the  Associated Research customer s
2.      14  To set the Low Resistance Limit on model 3570D only  Please press the SET key until the display shows     I Low   XXXX MQ    Use the Up Down Arrow keys to enter the resistance low limit setting  then press the  EXIT key to exit to the test mode or toggle to another setting using the SET key  The unit  of measure is in megohms with 1   1000 MQ range of setting  A measured value below  the Low Limit setting will result in a failure condition at the end of the Delay time setting     15  To set the Delay Time on model 3570D only  Please press the SET key until the display shows     I Delay   XXX X s    Use the Up Down Arrow keys to enter the delay time setting  then press the EXIT key to  exit to the test mode or toggle to another setting using the SET key     If the delay time is set to    0     the instrument will operate in a continuous ON mode when  the test button is pressed and released  The test will stop only when the manual reset  button is pressed to abort the Test  The minimum delay time setting is 0 5 seconds with  0 1 second resolution     Note  The Delay timer in the IR mode differs from the Dwell timer in the Hipot mode  A  judgment is not made until the Delay time has elapsed  In the Hipot mode a Pass   Fail  judgment is made a soon as the leakage limits are exceeded  Therefore when the Delay  time is set to zero Pass and Fail judgment or indications will not be made     32    C  OPERATING PROCEDURES     1  If the instruments defaults are acceptable then 
3.   FAIL                                   RESE ee   TES o   PROCESSING  X                         O    O       3  PASS 7       p  N FAIL                         REMOTE          1 0    and PROCESSING conditions  These signals are provided by three normally open  internal relays that switch on to indicate the current condition of the tester  These are  normally open free contacts and will not provide any voltage or current  The ratings    of the contacts are 1A   250 VAC  0 5    ADC    Below is a listing that indicates what    conditions activate each pin  When a terminal becomes active the relay closes thereby  allowing the external voltage to operate an external device     Pins 6 and 7 provide the PASS signal   Pins 8 and 9 provide the FAIL signal     Pins 1 and 4 provide the PROCESSING signal     39    The following describes how the relays operate for each test condition     PROCESSING   The relay contact closes the connection between pin  1  and pin  4   while the instrument is performing a test  The connection is opened at the end of the test     PASS   The relay contact closes the connection between pin  6  and pin  7  after  detecting that the item under test passed all tests  The connection is opened when the  next test is initiated or the reset function is activated     FAIL   The relay contact closes the connection between pin  8  and pin  9  after detecting  that the item under test failed any test  The connection is opened when the next test is  initiated or the res
4.  Check to be sure the correct input line voltage has been selected on the   rear panel  Either 115 volts AC or 230 volts AC  Connect the power input  plug into its socket on the rear panel of the instrument  Connect the male end of the plug to  the outlet receptacle     Please be sure that the safety ground on the power line cord is not defeated and that you are  connecting to a grounded power source     d   Turn on the POWER switch located on the lower left hand side of the   front panel  Upon powering the instrument up a POWER ON SELF   TEST  POST  will automatically be performed  This test will check for   the condition of the ram chips  led indicators  pcb s and other critical  components  All of the front panel led s will light temporarily then go out  In addition  you  will see the Associated Research name and Model Number briefly appear on the LCD  readout and then clear itself     ARI HYPOT II    3500D VER E    Current software version  displayed here    SET You should then see the default  l parameters on the LCD meter as  124KVAC follows for models 3500D        SET   Or these defaults for models 3505D  and 3565D   1 24KVAC       Or these defaults for models 3570D   W   SET     The underbar character after the  memory location indicates that an  1 24KVAC Insulation Resistance test will  execute after the Dielectric  Withstand test is completed  successfully        25    These abbreviated parameters stand for the following     SET  This is the parameter settings revie
5.  Fast Acting  5x20mm_     37781   tT Fuse Holder 20mm   37554   1   LCD Display 16 x 2 Characters    137555   1      PowerSwitch2P10A 250V__     376733   1      Test Switch  Green   37497   1   Reset Switch  Red    33189   1     Input Power Cable 10A 7Ft     04040A 08   1       High Voltage Output cable     02100A 13   2     Retumecable                49    SCHEMATIC INDEX    Drawing Number    S03500D Wiring Diagram    S37917 Main Control Board ra  S37961 Ground Continuity Check Board EM  S37903 Ground Continuity Check Board w IR test       50    
6.  MX XXX Xs W Test MX xXXX Xs    X XXKVAC XX   XX mA X XXKVAC XX  XX mA       The instrument will continue to output the desired voltage for the duration of the selected  dwell time for model 3500D or selected Ramp and Dwell times on model 3505D  3565D  and 3570D     34    5  If there is a short circuit in the DUT during the test  the red indicator light will  illuminate on the RESET switch and an alarm will sound  The display will show     models 3500D models 3505D and 3565D    HI FAIL XXX  Xs HI FAIL MX  XXX Xs    X XXKVAC OFL mA X XXKVAC OFL mA       model 3570D    W High MX XXX Xs    X XXKVAC OFL mA       6  If a flash over occurs in the DUT during the test  which results in an OFL condition  due to the current exceeding the maximum metering range  Breakdown will be indicated  as follows  The red indicator light will illuminate on the RESET switch and an alarm will  sound  The display will show     models 3500D models 3505D and 3565D    HI FAIL XXX Xs HI FAIL MX  XXX Xs    X XXKVAC OFL mA X XXKVAC OFL mA       model 3570D    W High MX XXX Xs    X XXKVAC OFL mA       7  If the leakage current exceeds the metering range and neither a short circuit nor flash  over occurred  the red indicator light will illuminate on the RESET switch and an alarm  will sound  The display will show     models 3500D models 3505D and 3565D    HI FAIL XXX Xs HI FAIL MX  XXX Xs    X XXKVAC OFL mA X XXKVAC OFL mA       model 3570D    W High MX XXX Xs    X XXKVAC OFL mA       35    8  If the leakage 
7.  not allow return to the calibration menu     47    a ASSOCIATED    3  Calibration of Insulation Resistance Current model 3570D only  Equipment needed  STANDARD DC MILLIAMMETER with a range of 2000uA and a  fixed Resistor 500K Ohms  50 watts     Please connect a 500KQ resistor in series with the standard DC microammeter and  connect these across the output leads of the instrument  Press the EXIT key  The  instrument will provide around 500VDC on the output and the display will show     Current  __ uA    Please use the Up Down Arrow keys to enter the reading of the standard DC  microammeter into the instrument  Then press the SET key to store the current setting   Press the EXIT or RESET key to return to the calibration menu without changing the  calibration setting     The instrument will output two short    beeps    when attempting to save a missing or bad  parameter and will not allow return to the calibration menu     4  Save Calibration   After the calibration  the instrument has to be turned off to save the calibration  parameters  The power must be turned off and on again to switch from the calibration  mode to the run mode     48    o ASSOCIATED    Replacement Parts List HYPOT IT    Model 3500D  3505D  3565D and 3570D  Rev I  ECO 4671    Designator  37903  37590  37593  37515  37017   1   CW720  Main Control Board SOS  37571   DEA a EE   359   4     Black BananaJacks     37478   2      HighVoltageComector     3759   1        eetkit o     37880   1     Fuse 3 15A  250V 
8.  second resolution     8  To set the Dwell Time  Please press the SET key until the display shows     Models 3500D  3505D  and 3565D Model 3570D    Timer   XXX X s or W Dwell   XXX X s    29    Use the Up Down Arrow keys to enter the dwell time setting  then press the EXIT key to  exit to the test mode or toggle to another setting using the SET key  If the dwell time is  set to    0     the instrument will operate in a continuous ON mode when the test button is  depressed and released  It will stop when the DUT  Device Under Test  goes into failure  or the manual reset button is pressed  The minimum dwell time setting is 0 2 seconds  with 0 1 second resolution     9  To setup the Ground Continuity Check  Please press the SET key until the display shows     Models 3500D  3505D  and 3565D    Continuity   ON or Continuity   OFF    Model 3570D    W Continuity   ON   or W Continuity   OFF    Press the Up Down Arrow key to select the    ON    or    OFF    function of the Ground  Continuity Check mode  the display will show the set status immediately   Press the EXIT key to exit to the test mode or toggle to another setting using the SET key     If the Continuity Function is set to    ON     the Ground Continuity cable has to be  connected between the    Cont  Check    connector on the instrument panel and the chassis  of the DUT  The instrument will test the ground connection of the DUT automatically  If  continuity is present a green indicator will be illuminated on the TEST switc
9.  that the test  duration is the same for each product tested     PLC remote inputs and outputs    The standard 9 pin interfaces provide outputs for Pass  Fail  and Test in Process   Inputs include Test and Reset  This gives the user all the basic remotes required to  configure the Hypot II through simple PLC relay control     Front panel LCD displays test parameters and results    A front panel LCD allows the operator to monitor the test  The display holds the  results after a test item failure so that the operator can easily review the test results   Indications of high fail and continuity failure are clearly displayed     Maximum output current 12 milliamps AC and 5 milliamps DC    This makes this instrument a true hipot tester with enough output current to test  capacitive loads in AC mode and allows the instrument to comply with the UL    120 K  ohm    requirements     15    11     12     13     14     15     16     17     18     19     2  ASSOCIATED    Software calibration control    Hypot II is calibrated through the front panel keypad  All calibration information is  stored in non volatile memory  This allows Hypot II to be completely calibrated  without removing any covers and exposing the technician to hazardous voltages     User selectable output voltage frequencies of 50 or 60 hertz    Hypot II was designed for the global market  This feature makes it simple for the  user to select the output frequency in the AC hipot mode so that products can be  tested at the sam
10. E OUTPUT JACK  For the connection of the detachable 5 foot   1 52 m  red high voltage test lead  The silicone rubber insulation is flexible for easy  handling and is rated at 30K VDC  The jack is recessed for safety when this lead is  not being used     HIGH VOLTAGE ARROW  LED INDICATOR   This indicator flashes to warn  the operator that high voltage is present at the high voltage output terminal     RETURN OUTPUT JACK  For the connection of the detachable 5 foot  1 52 m   black return test lead  This lead is always used when performing a test     19                                                                               Y GND F2A  250V  A caution   A WARNING   PLEASE OPERATE THIS INSTRUMENT ACCORDING  RISK OF ELECTRICAL SHOCK  II   ar    O D  1  Do not touch the output terminals the  DO NOT  REMOVE  COVER  product under test while o lest Is behg    REFER SERVICE TO MANU  rind c  gt  c  gt   FACTURER S SERVICE CENTER  2  If there is ony abnormal condition or a 115 de  FOR CONTINUED PROTECTION malfunction tum off the Input power switch      y      gt  HA  RESET  1  AGAINST FIRE REPLACE  ONLY Puello eee eee c  gt      gt  Pa    Peces e input wolloge se  WTH 250V FUSH OF THD sich polcian the re logs the Tabo a  SPECIFIED CURRENT RATING  ment wil be connected to c 5 g 5 5  4  Pleose refer to the instruction manual for O 98  132 192 264VAC  detailed operating instructions before using      yA  gt   2A MAX  50 60HZ  this instrument   C  gt         O                REMOTE 1 0 
11. OPERATION AND SERVICE MANUAL    Hypot    II    MODEL 3500D  AC ONLY HIPOT WITH GROUND CONTINUITY CHECK   MODEL 3505D  ENHANCED AC ONLY HIPOT WITH GROUND CONTINUITY CHECK   MODEL 3565D  ENHANCED AC DC HIPOT WITH GROUND CONTINUITY CHECK   MODEL 3570D  ENHANCED AC DC HIPOT WITH IR AND GROUND CONTINUITY     SERIAL NUMBER    A A E    M O d e   S    Associated Research  Inc   2000  13860 West Laurel Drive   3 5 0 0 D  3  gt  0  gt  D Lake Forest  e 60045 4546   3565D 3570D U S A     Item 37589 Ver 3 18 Printed Nov 6  2000    5 YEAR WARRANTY POLICY    Associated Research  Inc   certifies that the instrument listed in this manual meets  or exceeds published manufacturing specifications  This instrument was calibrated using  standards that are traceable to the National Institute of Standards and Technology  NIST      Your new instrument is warranted to be free from defects in workmanship and  material for a period of  1  year from date of shipment  Options and accessories that are  external to the instrument are warranted for a period of  90  days from date of shipment   You must return the    Owners Registration Card    provided within  15  days from receipt  of your instrument     AR recommends that your instrument be calibrated on a twelve month cycle   Instruments purchased and used in North America only  may have their warranty  extended in one year increments to a maximum of  5  years provided they are returned to  AR at least annually for calibration and inspection  The annual 
12. OUS VOLTAGES MAY BE PRESENT     CHASSIS GROUND SYMBOL     CALLS ATTENTION TO A PROCEDURE  PRACTICE  OR    WARNING CONDITION  THAT COULD POSSIBLY CAUSE BODILY  INJURY OR DEATH   CAUTION CALLS ATTENTION TO A PROCEDURE  PRACTICE  OR    CONDITION  THAT COULD POSSIBLY CAUSE DAMAGE TO  EQUIPMENT OR PERMANENT LOSS OF DATA     io     A Hipot produces voltages and currents which can cause harmful or fatal electric  shock  To prevent accidental injury or death  these safety procedures must be strictly  observed when handling and using the test instrument     SERVICE AND MAINTENANCE   User Service   To prevent electric shock do not remove the instrument cover  There are no user  serviceable parts inside  Routine maintenance or cleaning of internal parts is not  necessary  Any external cleaning should be done with a clean dry or slightly damp cloth   Avoid the use of cleaning agents or chemicals to prevent any foreign liquid from entering  the cabinet through ventilation holes or damaging controls and switches  also some  chemicals may damage plastic parts or lettering  Schematics  when provided  are for  reference only  Any replacement cables and high voltage components should be acquired  directly from Associated Research  Inc  Refer servicing to an Associated Research  Inc   authorized service center     PHONE  1  847  367 4077    ASSOCIATED RESEARCH  INC  1  800  858 TEST  8378   13860 WEST LAUREL DRIVE FAX  1  847  367 4080  LAKE FOREST  IL 60045 4546 U S A  E MAIL   info  asresearc
13. Since a DC hipot does charge the item under test it becomes necessary to discharge  the item after the test     3  DC testing unlike AC testing only charges the insulation in one polarity  This  becomes a concern when testing products that will actually be used at AC voltages   This is a key reason that some safety agencies do not accept DC testing as an  alternative to AC     4  When performing AC hipot tests the product under test is actually tested with peak  voltages that the hipot meter does not display  This is not the case with DC testing  since a sinewave is not generated when testing with direct current  In order to  compensate for this most safety agencies require that the equivalent DC test be  performed at higher voltages than the AC test  The multiplying factor is somewhat  inconsistent between agencies which can cause confusion concerning exactly what  equivalent DC test voltage is appropriate     The Ground Continuity Test    The Ground Continuity test is checking to verify that there is low impedance path  between the chassis and the safety ground wire of the line cord  A low impedance ground  system is critical in ensuring that in the event of a product failure a circuit breaker on the  input line will act quickly to protect the user from any serious electrical shock     International compliance agencies such as CSA  IEC  TUV  VDE  BABT and others   have requirements calling out this test  A low current test merely indicates that there is a  safety ground con
14. T BE DEFEATED BY THE USE OF AN  EXTENSION CORD  POWER CABLE  WITHOUT A PROTECTIVE CONDUCTOR   GROUNDING      This instrument is shipped with a three wire power cable  When this cable is connected  to an appropriate AC power source  this cable connects the chassis to earth ground  The  type of power cable shipped with each instruments depends on the country of destination     Operating Environment    This equipment is intended for indoor use only  The equipment has been evaluated  according to Installation Category Il and Pollution Degree 2 as specified in IEC 664     This instrument may be operated within the following environmental conditions   Temperatures from 32     113   F  0     45   C     Relative humidity of 0 to 95     Altitude up to 15 000 feet  4 600 meters      STORAGE AND SHIPMENT  Environment    This instument may be stored or shipped in environments with the following limits   Temperature     40   to  75  C  Altitude    7 620 meters  25 000 feet   The instrument should also be protected against temperature extremes which may cause  condensation within the instrument     Packaging    Original Packaging  Please retain all original packaging materials if you do not have an  alternate method of repackaging  If you are returning your instrument to us for servicing  please repackage the instrument in its original container or use an alternate packaging  solution  Please do not reuse the original packing material if there appears to be damage  or missing packing mater
15. TPUT JACK  For the connection of the detachable 5 foot   1 52 m  red high voltage test lead  The silicone rubber insulation is flexible for easy  handling and is rated at 30KVDC  The jack is recessed for safety when this lead is  not being used     2  RETURN OUTPUT JACK  For the connection of the detachable 5 foot  1 52 m   black return test lead  This lead is always used when performing a test     3  CONTINUITY CHECK OUTPUT JACK  For the connection of the detachable 5  foot  1 52 m  return test lead  This lead is always used when performing a continuity  test     03 Remote Interlock    Option 03 is a Remote Interlock feature that inverts the present reset logic on the  instrument from normally opened  must close contacts to activate the reset  to normally  closed  must open contacts to activate the reset   Hardware and software have been  reconfigured to provide the interlock connections on pins 2 and 5 of the Remote Interface  Signal Input port  This reset scheme is designed for use with an external safety interlock  device that utilizes a  Fail When Open  configuration on its output interface  The  instrument can still be used without the external reset device as long as the Interlock  Connector 38037  provided with unit  is plugged into the Remote Interface Signal Input  port  If there is nothing connected to the Remote Interface Signal Input port  to provide a  connection to the interlock  reset   then the instrument will not function  Please refer to  page 38 for furthe
16. When the HYPOT II tests the ground conductor of the line cord  if the resistance of the  ground conductor exceeds 10  the hipot will signal a continuity failure  If continuity is  present  high voltage is applied to both conductors of the line cord to insure that all  current carrying conductors in the primary circuit are tested  When testing products with  two prong plugs  do not activate the continuity circuit     41    MODEL 3500D  3505D  3565D and 3570D OPTIONS  Introduction    This section contains a list and descriptions of available factory installed options at the  time of this printing  The list of options contains an option code number which can be  referenced on the model option label on the rear panel of the unit when options are  present     Model Option Label  On the rear panel of the instrument you will find a label that contains the option code     For example  your options code would appear as follows   fitted with option 01 OPT  01    fitted with option 01 and 03 OPT  0103    3500D  3505D  3565D and 3570D Options    Option List    Remote Interlock       Description  01 Rear Outputs  The Rear Outputs option gives the user the capability to access the High Voltage  Return    and Continuity outputs from the rear panel for use in automated systems  The  configuration and descriptions are as follows     42    OPTIONAL REAR PANEL CONTROLS    1 2       kann A CAUTION   HN  END FZA 250V  CAUTION A HIGH YOLTAGE  FN N                            3    1  HIGH VOLTAGE OU
17. a where combustible  materials are present     TEST OPERATOR  Qualifications    This instrument generates voltages and currents which can cause harmful or fatal  electric shock and must only be operated by a skilled worker trained in its use     The operator should understand the electrical fundamentals of voltage  current  and  resistance        Safety Procedures   Operators should be thoroughly trained to follow these and all other applicable safety  rules and procedures before they begin a test  Defeating any safety system should be  treated as a serious offense and should result in severe penalties  such as removal from    the Hipot testing job  Allowing unauthorized personnel in the area during a test should  also be dealt with as a serious offense     Dress  Operators should not wear jewelry which could accidentally complete a circuit     Medical Restrictions  This instrument should not be operated by personnel with heart ailments or devices such  as pacemakers        TEST PROCEDURES    NEVER PERFORM A HIPOT TEST ON ENERGIZED CIRCUITRY OR EQUIPMENT     If the instrument has an external safety ground connection be sure that this is connected   Then Connect the return lead first for any test regardless of whether the item under test is  a sample of insulating material tested with electrodes  a component tested with the high  voltage test lead  or a cord connected device with a two or three prong plug     Plug in the high voltage test lead only when it is being used  Handl
18. afety of equipment repairs     The specific technique varies with each product  but basically  during a dielectric voltage    withstand test  an electrical devise is exposed to a voltage significantly higher than it  normally encounters  The high voltage is continued for a given period of time     If  during the time the component is tested  stray current flow remains within specified  limits  the device is assumed to be safe under normal conditions  The basic product  design and use of the insulating material will protect the user against electrical shock     The equipment used for this test  a dielectric withstand tester  is often called a  hipot    for high potential tester   The  rule of thumb  for testing is to subject the product to  twice its normal operating voltage  plus 1 000 volts     2  ASSOCIATED   gt  gt  INTRODUCTION    However  specific products may be tested at much higher voltages than 2X operating  voltages   1 000 volts  For example  a product designed to operate in the range between  100 to 240 volts  can be tested between 1 000 to 4 000 volts or higher  Most  double  insulated  products are tested at voltages much higher than the  rule of thumb      Testing during development and prototype stages is more stringent than production run  tests because the basic design of the product is being evaluated  Design tests usually are  performed on only a few samples of the product  Production tests are performed on each  and every item as 1t comes off the producti
19. ain reason for this is that most items being hipot tested will operate at AC  voltages and AC hipot testing offers the advantage of stressing the insulation  alternately in both polarities which more closely simulates stresses the product will  see in real use     2  Since AC testing cannot charge a capacitive load the current reading remains  consistent from initial application of the voltage to the end of the test  Therefore   there is no need to gradually bring up the voltage since there is no stabilization    required to monitor the current reading  This means that unless the product is  sensitive to a sudden application of voltage the operator can immediately apply full  voltage and read current without any wait time     Another advantage of AC testing is that since AC voltage cannot charge a load there  is no need to discharge the item under test after the test     AC testing disadvantages       1     A key disadvantage of AC testing surfaces when testing capacitive products  Again   since AC cannot charge the item under test  reactive current is constantly flowing   In many cases the reactive component of the current can be much greater than the  real component due to actual leakage  This can make it very difficult to detect  products that have excessively high leakage current     Another disadvantage of AC testing is that the hipot has to have the capability of  supplying reactive and leakage current continuously  This may require a current  output that is actually mu
20. be sure to connect the appropriate test  leads to the device under test  DUT  or test fixture  Be sure to connect the safety ground   on the rear panel  to a suitable known good ground before energizing this instrument   Then connect the return lead first to the test fixture or the DUT followed by the high  voltage lead     Check your connections to be sure they are making good contact and that the test station  or area is clear of debris or other personnel DO NOT TOUCH THE DEVICE  UNDER TEST ONCE THE TEST HAS BEEN STARTED     WARNING   2  Please follow the setup procedures to set the parameters  The display    will show the Dwell or Delay Timer value  the Test Voltage value  and  the High Limit Setting  On model 3505D  3565D and 3570D the memory location will  also be displayed     models 3500D models 3505D and 3565D    Set XXX  Xs Set MX XXX Xs    X XXKVAC XX XX mA X XXKVAC XX XXmA       On model 3570D the display will show either the Dielectric Withstand test stettings or the  Insulation Resistance test settings depending on what has been selected  as follows     model 3570D    W Set MX XXX Xs or I Set MX XXX Xs    X XXKVAC XX  X X mA X XX KV DC XXX  X MO       On model 3570D the underbar character may be displayed after the memory number  i e   M5_   this indicates that there is another configured test that will execute after the  displayed setup has completed successfully  To view the second test setup  press the  down arrow key  Return to the first test by pressing the 
21. calibration and  inspection must be performed annually each and every year following receipt of  instrument  Any instrument not calibrated and inspected annually will not be eligible for  extended warranty status  This extended warranty is non transferable and is offered only  to the original purchaser     A return material authorization  RMA  must be obtained from AR before  returning this instrument for warranty service  Please contact our Customer Support  Center at 1 800 858 TEST  8378  to obtain an RMA number  It is important that the  instrument is packed in its original container for safe transport  If the original container  in not available please contact our customer support center for proper instructions on  packaging  Damages sustained as a result of improper packaging will not be honored   Transportation costs for the return of the instrument for warranty service must be prepaid  by the customer  AR will assume the return freight costs when returning the instrument  to the customer  The return method will be at the discretion of Associated Research     Except as provided herein  Associated Research makes no warranties to the  purchaser of this instrument and all other warranties  express or implied  including   without limitation  merchantability or fitness for a particular purpose  are hereby  excluded  disclaimed and waived     Any non authorized modifications  tampering or physical damage will void your  watranty  Elimination of any connections in the earth gr
22. ch higher than is really required to monitor leakage current  and in most cases is usually much higher than would be needed with DC testing   This can present increased safety risks as operators are exposed to higher currents     DC testing characteristics       During DC hipot testing the item under test is charged  The same test item capacitance  that causes reactive current in AC testing results in initial charging current which  exponentially drops to zero in DC testing     DC testing advantages       1     Once the item under test is fully charged the only current flowing is true leakage  current  This allows a DC hipot tester to clearly display only the true leakage of the  product under test     The other advantage to DC testing is that since the charging current only needs to be  applied momentarily the output power requirements of the DC hipot tester can  typically be much less than what would be required in an AC tester to test the same  product     DC testing disadvantages       l     Unless the item being tested has virtually no capacitance it is necessary to raise the  voltage gradually from zero to the full test voltage  The more capacitive the item the  more slowly the voltage must be raised  This is important since most DC hipots  have failure shut off circuitry which will indicate failure almost immediately if the  total current reaches the leakage threshold during the initial charging of the product  under test     2  ASSOCIATED   gt  gt  INTRODUCTION    2  
23. current exceeds the high limit setting  but does not exceed the metering  range  the red indicator light will illuminate on the RESET switch and an alarm will  sound  The display will show     models 3500D models 3505D and 3565D    HI FAIL XXX  XS HI FAIL MX XXX Xs    X XXKVAC XX XX mA X XXKVAC XX XX mA       model 3570D    W High MX XXX Xs  X XXKVAC XX XX mA       9  If the leakage current does not exceed the low limit setting on model 3505D  3565D  and 3570D  the red indicator light will illuminate on the RESET switch and an alarm will  sound  The display will show     models 3505D and 3565D model 3570D    LO LMT MX XXX Xs W  Low MX XXX Xs    X XXKVAC XX  XX mA X XXKVAC XX   XX mA       10  To stop the alarm  please press the RESET switch once  The alarm will stop and  the display will retain the failure information  If the RESET switch is pressed again  the  data on the display screen will be cleared and the display will indicate the setting data  from the last test  The instrument is now ready for the next test     11  If the DUT passed the test  the instrument will output a short audible beep tone to  indicate the DUT has passed the test  The display will show     models 3500D models 3505D and 3565D    Pass XXX Xs Pass MX XXX Xs    X XXKVAC XX  XX mA X XXKVAC XX  XX mA       model 3570D    W  Pass MX XXX Xs    X XXKVAC XX   XX mA       The instrument is now ready to perform another test     36    12  If the operator elects to abort a test in process this can be accomplis
24. default  settings on the instrument  The default settings  shown will remain in memory unless you choose to override them with your own test  program  The default settings will appear in memories 1 through 5 for models 3505D   3565D and 3570D  The instrument default settings are as follows     DEFAULTS    Input Voltage   Lock     Test     Dielectric Withstand Test Mode    Voltage Output   High Current Trip   Low Current Trip   Ramp Timer   Dwell Timer   Voltage Type     AC Output Frequency     Continuity     115 or 230 volts AC country specific     rear panel switch selectable     OFF    W   1  3570D only     1 240 volts AC   HI LMT  10 mA   LO LMT 0mA_  3505D and 3565D only   0 second  3505D 3565D 3570D only   1 second   AC    60Hz    OFF    Insulation Resistance Mode  model 3570D only     24    Voltage Output   High Resistance Trip     Low Resistance Trip     Delay Timer     1 000 volts DC  I   High  0 MQ  I  Low  1 MQ    2 seconds    Qne     a   Unpack this instrument from its special shipping container  Be sure to save all  packaging materials in case you need to return it to the factory for service     WARNING        Locate a suitable testing area and be sure you have read all safety    instructions for the operation of the instrument and suggestions on the test  area set up in the SAFETY section of this manual  Locate a three prong grounded outlet   Be sure the outlet has been tested for proper wiring before connecting the instrument to it     QUICK START    CAUTION c  
25. del 3505D  3565D and 3570D  SETTING 0 and 0 2   999 9 seconds  0 1 second   step    0 ramp setting   0 1 seconds fixed ramp    Model 3500D and 3505D  Ramp   0 1 seconds fixed    FAILURE SETTINGS   AC mode  High limit   0 10   12 00 mA  0 01 mA   step  Low limit   0 00  0 10   12 00 mA  0 01 mA   step  0  OFF    3505D  3565D and 3570D only   Accuracy     2   of setting   0 02 mA   DC mode  High limit   0 02   5 00 mA  0 01 mA   step  Low limit   0 00  0 02   5 00 mA  0 01 mA   step  0  OFF    3505D  3565D and 3570D only   Accuracy      2   of setting   0 02 mA        2  ASSOCIATED    Specifications    METERING Voltmeter  4 digits   Range   AC 0 00   5 00 KV    DC 0 00   6 00 KV  Resolution    01 KV  Accuracy     2  of reading   10 V   Ammeter  4 digits   Range   AC 0 10   12 00 mA    DC 0 02   5 00 mA  Resolution   0 01 mA  Accuracy     2   of reading   0 02mA     TIMER DISPLAY Range   0 0   999 9 seconds  Resolution   0 1 second  Accuracy     0 1  of reading   0 05 seconds     DISCHARGE TIME  lt  300 ms  The maximum capacitive load vs output voltage    0 20 uF      lt  IKV 0 050 uF      lt  4KV  0 10 uF      lt  2KV 0 040 uF      lt  SKV  0 06 uF      lt 3KV 0 015 uF      lt  6KV    Current   DC 0 1 A  0 01A  fixed  CONTINUITY CHECK   Max  ground resistance   1 Q  0 10  fixed  Output Voltage Range  100   1000 Volts DC  Resolution  10 volt step  Accuracy     2  of reading   5 volts   Voltage Display Range  0   1000 V  Resolution  10 volt step  Accuracy     2  of reading   2 co
26. e frequency they will be used at     Output voltage fine adjustment    To make Hypot IT usable in all types of applications  the operator can manually  bring the voltage up or down in 10 volt increments by simply pressing the up and  down arrow keys  This makes it very easy to adjust the output voltage even while  Hypot II is in the dwell mode so you can analyze test results at different voltages     Built in Continuity test mode   Hypot II meets ground continuity test requirements called out by UL and other safety  agencies     Flashing high voltage indicator   A flashing LED located directly over the high voltage terminal clearly indicates when  high voltage is active to provide maximum operator safety     Comes complete with a test box for products terminated in a line cord    The standard US style receptacle box allows easy testing of hipot and continuity on  line cord terminated products  As an additional benefit  termination boxes are  available for testing products with line cords configured for other countries     User selectable input voltage   Hypot II can be switched for either 115 or 230 volt input operation through an easy  access rear panel mounted switch to allow it to be used in any country     Low current sense     Monitors the minimum level of current flow  thus ensuring that the DUT is properly  connected and that the hipot test is being performed     Electronic ramping     Provides a gradual and timed method to increase output voltage to the DUT   minimizi
27. e its clip only by the  insulator   never touch the clip directly  Be certain that the operator has control over  any remote test switches connected to the Hipot  Double check the return and high  voltage connections to be certain that they are proper and secure     NEVER TOUCH THE ITEM UNDER TEST OR ANYTHING CONNECTED TO IT WHILE HIGH  VOLTAGE IS PRESENT DURING THE HIPOT TEST     When testing with DC  always discharge the capacitance of the item  WARNING   under test and anything the high voltage may have contacted  such as test  fixtures  before handling it or disconnecting the test leads     HOT STICK probes can be used to discharge any capacitance in the item under test as a  further safety precaution  A hot stick is a nonconducting rod about two feet long with a  metal probe at the end which is connected to a wire  To discharge the device under test   two hot sticks are required  First connect both probe wires to a good earth ground  Then  touch one probe tip to the same place the return lead was connected  While holding the  first probe in place  touch the second probe tip to the same place where the high voltage  lead was connected     KEY SAFETY POINTS TO REMEMBER     Keep unqualified and unauthorized personnel away from the test area   Arrange the test station in a safe and orderly manner    Never touch the product or connections during a test    In case of any problem  turn off the high voltage first    Properly discharge any item tested with DC before touching conn
28. eakage Current Limit  Please press the SET key until the display shows     Models 3500D  3505D  and 3565D Model 3570D    HI LMT  XXXX mA or W  High   XX XX mA    Use the Up Down Arrow keys to enter the leakage current high limit setting  then press  the EXIT key to exit to the test mode or toggle to another setting using the SET key  The  unit of measure is in milliamperes with 0 10 AC and 0 02 DC milliamperes as the  minimum setting     6  To set the Low Leakage Current Limit on models 3505D  3565D and 3570D only  Please press the SET key until the display shows     Models 3505D  3565D Model 3570D    LO LMT  XX XX mA or W Low   XX XX mA    Use the Up Down Arrow keys to enter the leakage current low limit setting  then press the  EXIT key to exit to the test mode or toggle to another setting using the SET key  The unit  of measure is in milliamperes with 0 00 milliamperes AC and DC as the minimum  setting     7  To set the Ramp Time on models 3505D  3565D and 3570D only  Please press the SET key until the display shows     Models 3505D  3565D Model 3570D    Ramp T   XXX Xs or W Ramp   XXX Xs    Use the Up Down Arrow keys to enter the ramp time setting  then press the EXIT key to  exit to the test mode or toggle to another setting using the SET key  If the ramp time is  set to    0     the instrument will raise the voltage up to the test voltage in a fixed ramp time  of 0 1 seconds AC and DC and then begin the dwell cycle  The minimum ramp time  setting is 0 2 seconds with 0 1
29. ections     2  ASSOCIATED   gt  gt  INTRODUCTION    GLOSSARY OF TERMS     as used in this manual     Alternating Current  AC  Current which reverses direction on a regular basis  commonly in the  U S A  60 times per second  in other countries 50 times per second     Breakdown  The failure of insulation to effectively prevent the flow of current  sometimes evident by  arcing  If voltage is gradually raised  breakdown will begin suddenly at a certain voltage level  Current  flow is not directly proportional to voltage  Once breakdown current has flown  especially for a period of  time  a repeated application of voltage will often show breakdown beginning at a lower voltage than  initially     Conductive  Having a volume resistivity of no more than 103 ohm cm or a surface resistivity of no  more than 105 ohms per square     Conductor  A solid or liquid material which has the ability to let current pass through it  and which has  a volume resistivity of no more than 103 ohm cm     Current  The movement of electrons through a conductor  Current is measured in amperes   milliamperes  microamperes  nanoamperes  or picoamperes  Symbol   I    Dielectric  An insulating material which is positioned between two conductive materials in such a way  that a charge or voltage may appear across the two conductive materials     Direct Current  DC  Current which flows in one direction only  The source of direct current is said  to be polarized and has one terminal which is always at a higher pot
30. ential than the other     Hipot Tester  Common term for dielectric withstand test equipment   Hypot    Registered trademark of Associated Research  Inc   for its dielectric withstand test equipment     Insulation  Gas  liquid or solid material which has a volume resistivity of at least 1012 ohm cm and is  used for the purpose of resisting current flow between conductors     Insulation Resistance Tester  An instrument or a function of an instrument capable of measuring  resistance s in excess of 200 megohms  Usually employs a higher voltage power supply than used in  ohmmeters measuring up to 200 megohms     Leakage  AC or DC current flow through insulation and over its surfaces  and AC current flow through a  capacitance  Current flow is directly proportional to voltage  The insulation and or capacitance is thought  of as a constant impedance  unless breakdown occurs     Resistance  That property of a substance which impedes current and results in the dissipation of power  in the form of heat  The practical unit of resistance is the ohm  Symbol   R    Trip Point  The minimum current flow required to cause an indication of unacceptable performance  during a dielectric voltage withstand test     Voltage  Electrical pressure  the force which causes current through an electrical conductor   Symbol   V    INTRODUCTION    The importance of testing    User safety   In an era of soaring liability costs  original manufacturers of electrical and electronic  products must make sure ever
31. et function activated     2  SIGNAL INPUTS OF REMOTE I O  The Hypot II remote connector enables remote operation of the TEST and  RESET functions  A normally open momentary switch can then be wired  across pins 3 and 5 to allow remote operation of the TEST function  A  normally open momentary switch can also be wired across pins 2 and 5  which allows remote operation of the RESET function  For safety the front  panel RESET switch remains active even when a remote reset switch is  connected so that high voltage can be shut down from either location        DO NOT CONNECT VOLTAGE OR CURRENT TO THE SIGNAL INPUTS  THIS  COULD RESULT IN DAMAGE TO THE CONTROL CIRCUITRY     40    E  ADAPTER BOX CONNECTION     The following diagram shows how to connect the adapter box to the HYPOT II and to the  device under test     WARNING Never connect the adapter box to the hipot while the high voltage    outputs are energized     3500D  3505D  3565D and 3570D 36544  38202                               Test Connections    The adapter box allows you an easy way to connect an item which is terminated in a two  or three prong line cord  If you are testing a product which is terminated in a three prong  plug  you are also required to perform a continuity or ground bond test on the ground  conductor of the DUT to the chassis or dead metal of the product  With the HYPOT IL   you can perform both the hipot and continuity tests at the same time by activating the  continuity function on the instrument     
32. g 5000V AC     Please connect a standard ac voltmeter with 5000V minimum full scale range to the high  voltage and return connectors    Then press the SET key on the front panel  The instrument will provide around  5000V AC on the output connectors and the display will show     Voltage  _ V    Please use the Up Down Arrow keys to enter the reading off the AC Voltmeter into the  instrument  Then press the SET key to store the voltage setting  Press the EXIT or  RESET key to return to the calibration menu without changing the calibration setting     The instrument will output two short    beeps    when attempting to save a missing or bad  parameter and will not allow return to the calibration menu     2  Calibration of Dielectric Withstand Current  Equipment needed  STANDARD AC MILLIAMMETER with a range of 25mA and a  fixed Resistor 500K Ohms  50 watts     Please connect a 500KQ resistor in series with the standard ac milliammeter and connect  these across the output leads of the instrument  Press the Down Arrow key  The  instrument will provide around 5000V AC on the output and the display will show     Please use the Up Down Arrow keys to enter the reading of the standard AC  milliammeter into the instrument  Then press the SET key to store the current setting   Press the EXIT or RESET key to return to the calibration menu without changing the  calibration setting     The instrument will output two short    beeps    when attempting to save a missing or bad  parameter and will
33. h and the  operator may proceed with the hipot test  The Hypot II will continue to test the ground  connection of the DUT during other tests  If no continuity is present while the operator  tries to initiate a test  the instrument will signal a continuity failure     10  To set the Frequency 50 60 Hz  Please press the SET key until the display shows     Models 3500D  3505D  and 3565D    Freq   50 HZ or Freq   60 HZ  Model 3570D  W  Freq   50 HZ   or   W Freq   60 HZ       Press the Up or Down Arrow keys to toggle the frequency selection  50 or 60 hertz  then  press the EXIT key to exit to the test mode or toggle to another setting using the SET key     11  To Set Key Lock or Memory Lock   The Hypot Il is equipped with two different lockout selections  The first selection is Key  Lock  Within this selection  you can lockout all front panel control functions except  TEST and RESET  The second selection is Memory Lock and is used in conjunction with  the Key Lock  Within this selection  you can lockout all front panel control functions  except TEST  RESET and memory select     You can toggle this selection only from a power OFF state  Please turn the POWER  switch to the off position then press and hold the SET key  then turn the power ON     Momentarily you will see one of the following screens to indicate if the keys are locked  or unlocked     Key Lock   ON or Key Lock   OFF    To toggle the Key Lock function on and off use the Up or Down Arrow keys  then press  EXIT to selec
34. h com    www asresearch com    Service Interval   The instrument and its power cord  test leads  and accessories must be returned at least  once a year to an Associated Research authorized service center for calibration and  inspection of safety related components  Associated Research will not be held liable for  injuries suffered if the instrument is not returned for its annual safety check and  maintained properly        User Modifications   Unauthorized user modifications will void your warranty  Associated Research will not  be responsible for any injuries sustained due to unauthorized equipment modifications or  use of parts not specified by Associated Research  Instruments returned to Associated  Research with unsafe modifications will be returned to their original operating condition  at your expense     TEST STATION   Location   Select an area away from the main stream of activity which employees do not walk  through in performing their normal duties  If this is not practical because of production  line flow  then the area should be roped off and marked for HIGH VOLTAGE  TESTING  No employees other than the test operators should be allowed inside     If benches are placed back to back  be especially careful about the use of the bench  opposite the test station  Signs should be posted   DANGER   HIGH VOLTAGE  TEST IN PROGRESS   UNAUTHORIZED PERSONNEL KEEP AWAY      Power   Dielectric Voltage Withstand Test Equipment must be connected to a good ground  Be  certain that 
35. hed by pressing  the RESET switch at anytime  The instrument will stop the test process immediately and    the display will show     models 3500D models 3505D and 3565D    Abort MX XXX  Xs  X XXKVAC XX  XX mA    Abort XXX Xs    X XXKVAC XX  XX mA       model 3570D    W ABRT MX  XXX Xs    X XXKVAC XX   XX mA       Please press the TEST switch to initiate another test     Insulation Resistance Test Mode model 3570D only    13  To initiate a test press the TEST switch on the front panel  The red High Voltage  Arrow indicator will flash and the display will immediately show     I Test MX XXX Xs    X XXKVDC XXX   X MO       14  If the DUT passed the test  the instrument will output a short audible beep tone to  indicate the DUT has passed the test  The display will show     I Pass MX XXX Xs  X XXKVDC XXX  X MQ       15  If the operator elects to abort a test in process this can be accomplished by pressing  the RESET switch at anytime  The instrument will stop the test process immediately and    the display will show     I ABRT MX XXX Xs    X XXKVDC XXX   X MO       37    16  If the insulation resistance exceeds the high limit setting  but does not exceed the  metering range  the red indicator light will illuminate on the RESET switch and an alarm    will sound  The display will show     I High MX XXX Xs    X XXKVDC XXX   X MQ       17  If the insulation resistance exceeds the high limit setting  and exceeds the metering  range  the red indicator light will illuminate on the RESET 
36. ial  Contact our customer support department  1 800 858 8378   for an RGA  return goods authorization  number  Please enclose the instrument with all  options  accessories and test leads  Indicate the nature of the problem or type of service  needed  Also  please mark the container  FRAGILE  to insure proper handling  Upon  receipt your instrument will be issued an AR service number  Please refer to this number  in all correspondence     22    o ASSOCIATED    Other Packaging  If you do not have the original packaging materials please follow these  guidelines     1      2      3      4    5    6      7      Wrap the instrument in a bubble pack or similar foam  Enclose the same information  as above     Use a strong double wall container that is made for shipping instrumentation  350 lb   test material is adequate     Use a layer of shock absorbing material 70 to 100 mm  3 to 4 inch  thick around all  sides of the instrument  Protect the control panel with cardboard     Seal the container securely   Mark the container  FRAGILE  to insure proper handling   Please ship models 3500D  3505D  3565D and 3570D via Federal Express or UPS air     Please refer in all correspondence to your AR service number     Field Installation Of Options    There are no field installable options on the model 3500D  3505D  3565D or 3570D     23    Qne     QUICK START    QUICK START    This quick start guide assumes the operator has some familiarity with automated hipot  testing and desires to use the  
37. instrument executes only the test selected  When    W I    is selected  first the Dielectric  Withstand test will execute and then the Insulation Resistance test will execute  selecting     LW    executes the two tests in the reverse order  To change the Test type on model  3570D only  please press the SET key until the display shows     Press the Up or Down Arrow keys to toggle the test selection  then press the EXIT key to  exit to the test mode or toggle to another setting using the SET key     or       Note  The Test mode type will effect the number of items displayed in the menu  The  menu will display only the items that are relative to the Test mode that has been selected     3  To set the AC   DC mode for models 3565D and 3570D only  Press the SET key until the display shows     Models 3565D    Mode   AC or Mode   DC    Model 3570D    W Mode   AC or W Mode   DC    Press the Up or Down Arrow keys to toggle the output function  AC or DC  then press the  EXIT key to exit to the test mode or toggle to another setting using the SET key     4  To set the Output Test Voltage  Please press the SET key until the display shows     Models 3500D  3505D  3565D Model 3570D    Voltage   XXX KV or W Volt   XXX KV    Please use the Up Down Arrow keys to enter the desired test voltage  then press the EXIT  key to exit to the test mode or toggle to another setting using the SET key  The maximum  voltage which may be entered is 5000 volts AC or 6000 volts DC    28    5  To set the High L
38. is manual a copy of the  Certificate of Calibration   It is recommended that you have  this instrument recalibrated and a safety check done at least once per year  AR  recommends you use  Calibration Standards  that are NIST traceable  or traceable to  agencies recognized by NIST to keep this instrument within published specifications  End  user metrology standards or practices may vary  These metrology standards determine the  measurement uncertainty ratio of the calibration standards being used     Calibration Equipment Required   The following equipment will be needed to properly calibrate your instrument     1   Digital Multimeter with the minimum ranges   AC Voltage  5 000 volts  AC Current  25mA range   DC Current  2000uA range    2   500KQ 50Watt resistor   CALIBRATION PROCEDURE    Model 3500D  3505D and 3565D   To enter the calibration mode the instrument must be in the OFF position  Using a pen or  small screw driver  press and hold the recessed calibration key on the rear panel  then    turn on the Input Power Switch     For model 3500D  3505D and 3565D the display will show        Press the SET key for kilovolts calibration  Down Arrow key  v  for milliamps     For model 3570D the display will show     CAL Mode EXIT IR    SET V V A       Press the SET key for kilovolts calibration  Down Arrow key  v  for milliamps  and the  EXIT key for insulation resistance calibration     46    1  Calibration of Voltage  Equipment needed  STANDARD AC VOLTMETER capable of measurin
39. nection  it does not completely test the integrity of that connection that  would be accomplished when using much higher currents in the range from 10 to 30amps   The high current test is know as a Ground Bond test     The Ground Continuity test verifies that the ground circuit between the Continuity and  Return panel jacks on the Hypot H is below lohm while test at a current of 100mA  This  function can be enabled or disabled through software  when a ground circuit is not  provided on the device under test or test requirement are not needed     The Insulation Resistance Test       Some  dielectric analyzers today come with a built in insulation resistance tester   Typically the IR function provides test voltages from 500 to 1 000 volts DC and  resistance ranges from kilohms to gigaohms  This function allows manufacturers to  comply with special compliance regulations  BABT  TUV and VDE are agencies that  may under certain conditions require an IR test on the product before a Hipot test is  performed  This typically is not a production line test but a performance design test     The insulation resistance test is very similar to the hipot test  Instead of the go no go  indication that you get with a hipot test the IR test gives you an insulation value usually in  Megohms  Typically the higher the insulation resistance value the better the condition of    10    Qne     the insulation  The connections to perform the IR test are the same as the hipot test  The  measured value rep
40. ng any damage from quickly over applying high voltage to sensitive DUTs     20 Test setup memories       The operator can store and recall test setups  thus eliminating the need to re enter  setup information     Qne     21  Insulation Resistance Test Mode     This allows the operator to perform an Insulation Resistance test independently or in  sequence with the Dielectric Withstand  Hipot  test  thus eliminating the need for a  separate piece of test equipment when IR testing is required     SPECIFICATIONS    22 Insulation Resistance Adjustable Output     The Insulation Resistance test mode provides an adjustable output range of 100 1000  volts DC  This gives the user the ability to meet most IR test requirements       Only available on model 3505D  3565D and 3570D      Only available on model 3570D    17       FRONT PANEL CONTROLS    1 2 3 4       CAUTION  HIGH VOLTAGE                   5 OKVAC MAX   6 0KVDC MAX                                      RETURN                            QRR           5 6 T7 8 9 10 11 12    RESET SWITCH  This is a momentary contact switch  If an out of range reading  is detected during a hipot test or an IR test or if continuity failure occurs  the red  failure lamp within the switch will light  To reset the system for the next test  press  and release this switch  This switch may also be used to abort a test in progress     TEST SWITCH  This is a momentary contact switch  Press the green switch to  turn on the high voltage output when in test m
41. ode  The indicator lamp within the  switch will light when continuity is    good     if continuity mode is enabled     LCD DISPLAY  The Liquid Crystal Display is the main readout for the operator  and programmer of the test settings and test results     CONTINUITY CHECK OUTPUT JACK  For the connection of the detachable 5  foot  1 52 m  black return test lead or three prong receptacle adapter box  This jack  is always used when performing a continuity test  Please refer to page 41 for details  on connecting the adapter box between Hypot II and the device under test     POWER SWITCH  Rocker style switch with international ON       and OFF  0   markings     SET KEY  Use this key to advance forward through the setup menus     DOWN ARROW  v   Use this key to decrement numeric values in the setup mode   This key also used to toggle ON OFF functions  Also may be used to decrease output  voltage during a test in 10 volt increments        10     11     12     FRONT PANEL CONTROLS    1 2 3 4            CAUTION  RESET HIGH VOLTAGE             S OKVAC MAX   6 0KVDC MAX                                      RETURN                      AC DC WITHSTAND VOLTAGE TESTER       5 6 T 8 9 10 11 12    UP ARROW  A   Use this key to increment numeric values in the setup mode  This  key also used to toggle ON OFF functions  Also may be used to increase output  voltage during a test in 10 volt increments     EXIT KEY  Use this key when you desire to enter the Run Mode to initiate a test     HIGH VOLTAG
42. omated test sequence using the defaults  If a failure  occurs you will HEAR an audible alarm go off  To stop the alarm you must depress the  RED button marked RESET  This will silence the alarm and reset the instrument to begin  another test  This RESET button may also be used as a safety button to quickly ABORT a  test and cut off the HIGH VOLTAGE     When HIGH VOLTAGE is present a RED flashing indicator located above   the HV receptacle will remain flashing until the HIGH VOLTAGE is OFF    If the device under test PASSED the test then no audible alarm will sound    You will hear a brief BEEP to let you know the item was successfully   tested and it PASSED  In the case of a FAIL condition the instrument will  provide a memory of the test condition results on the LCD display that will remain until the  next test is initiated  Depressing the reset button will reset the instrument alarm while  keeping the last test results on the display  Depressing the reset button a second time will  clear the display     26    OPERATION INSTRUCTIONS FOR Hypot Il  Model 3500D  3505D  3565D and 3570D    A  POWER UP     Check to be sure the correct input line voltage has been selected on the  rear panel  Either 115 volts AC or 230 volts AC  Connect the power  input plug into its socket on the rear panel of the instrument  Connect the male end of the  plug to the outlet receptacle     Please be sure that the safety ground on the power line cord is not defeated and that you  are connecting to a g
43. on line     The hipot tester must also maintain an output voltage between 100  and 120  of  specification  The output voltage of the hipot must have a sinusoidal waveform with a  frequency between 40 to 70 Hz and has a peak waveform value that is not less than 1 3  and not more than 1 5 times the root mean square value     Advantages and Disadvantages of AC Testing and DC Testing       Please check with the Compliance Agency you are working with to see which of the two  type of voltages you are authorized to use  In some cases a Compliance Agency will  allow either AC or DC testing to be done  However in other cases the Compliance  Agency only allows for an AC test  If you are unsure which specification you must  comply with please contact our CUSTOMER SUPPORT GROUP at 1 800 858 TEST   8378      Many safety agency specifications allow either AC or DC voltages to be used during the  hipot test  When this is the case the manufacturer must make the decision on which type  of voltage to utilize  In order to do this it is important to understand the advantages and   the disadvantages of both AC and DC testing     AC testing characteristics       Most items that are hipot tested have some amount of distributed capacitance  An AC  voltage cannot charge this capacitance so it continually reads the reactive current that  flows when AC is applied to a capacitive load     AC testing advantages       1  AC testing is generally much more accepted by safety agencies than DC testing   The m
44. ounding system or by passing  any safety systems will void this warranty  This warranty does not cover batteries or  accessories not of Associated Research manufacture  Parts used must be parts that are  recommended by AR as an acceptable specified part  Use of non authorized parts in the  repair of this instrument will void the warranty     TABLE OF CONTENTS    SECTION I OPERATORS MANUAL          cccssssssssscsscssssssssssessssncssessessessesssssssssssessoses 1  SAFETY enop a a 2  INTRODUCTION 2 e T a a a 6  SPECIFICA TION Sicario ias e de 12  CONTRO ES iaa 18  INSTALLATION orie a na oats a Geass Ses A A a aa a 21  QUICK STAR Domene eat saul pidas 24  OPERATION  0000 A E A EA A AA AAA Ed 21  OPTIONS egudnin mnnn ana Ra A A R An 42   SECTION II SERVICE MANUAL cosita 45  CALIBRATION caida idad 46  PARTS LISTS 49  SCHEMATICS a 50    lil    SECTION 1  OPERATORS MANUAL    SAFETY PRECAUTIONS REQUIRED FOR HIGH  VOLTAGE TESTING     GENERAL   This product and its related documentation must be reviewed for familiarization with  safety markings and instructions before operation     This product is a Safety Class I instrument  provided with a protective earth terminal    Before applying power verify that the instrument is set to the correct line voltage  115 or    230  and the correct fuse is installed     INSTRUCTION MANUAL SYMBOL  PLEASE REFER TO THE  INSTRUCTION MANUAL FOR SPECIFIC WARNING OR  CAUTION INFORMATION TO AVOID PERSONAL INJURY  OR DAMAGE TO THE PRODUCT     INDICATES HAZARD
45. r information about the remote interface    43    04 Arc Detection    This option provides the user with the capability of detecting low level arcing  The arc  detection circuit has nine preset sensitivity levels     The specifications for the Arc Detection option are as follows   FAILURE SETTINGS   Arc for AC and DC mode  Option 04 only   Sensitivity range   1 9    Arc detection   minimum duration  lt  10us       SETUP PROCEDURE  1  To activate Arc Detection and set the sensitivity level   Please press the SET key until the display shows     Use the Up Down Arrow keys to select the arc sensitivity level  The arc sensitivity level  may be set between the values of 1 and 9  1 being the least sensitive  9 being the most  sensitive  To deactivate the Arc Detection set the arc sensitivity to 0  Press the EXIT key  to exit to the test mode or toggle to another setting using the SET key     OPERATION PROCEDURE  1  If there is an arcing condition over the Arc sensitivity limit  the red indicator light will  illuminate on the reset switch and an alarm sound  The display will show     models 3500D models 3505D and 3565D    Arc Fail XXX Xs Arc MX XXX Xs    X XXKVAC XX XX mA X XXKVAC XX XX mA       model 3570D    Arc MX XXX Xs    X XXKVAC XX XX mA       SECTION 2  SERVICE MANUAL    CALIBRATION PROCEDURES    This instrument has been fully calibrated at the factory in accordance to our published  specifications  It has been calibrated with standards traceable to NIST  You will find in  th
46. resents the equivalent resistance of all the insulation which exists  between the two points and any component resistance which might also be connected  between the two points     INTRODUCTION    Although the IR test can be a predictor of insulation condition it does not replace the need  to perform a Dielectric Withstand test     TYPES OF FAILURES DETECTABLE ONLY WITH A HIPOT TEST    Weak Insulating Materials  Pinholes in Insulation   Inadequate Spacing of Components  Pinched Insulation    IF YOU SHOULD HAVE ANY QUESTIONS RELATING TO THE OPERATION       OF YOUR INSTRUMENT CALL 1 800 858 TEST 8378  IN THE U S A     11    2  ASSOCIATED    Model 3500D  3505D  3565D and 3570D  Functional Specifications    Unless otherwise stated  accuracy s are relative to a laboratory standard measurement     DIELECTRIC WITHSTAND TEST MODE    Specifications    INPUT VOLTAGE 115   230V selectable    15   variation  47   63 Hz    FUSE 115 VAC  230VAC    3 15A fast acting 250V AC    OUTPUT Rating AC 0 5000V  2V   step  12mA  DC 0  6000V  2V   step  5mA   DC mode on 3565D and 3570D only   Regulation     1   of output   5V     VOLTAGE SETTING   OV   Max output rating  10 volts step  Accuracy     2  of Setting   5V   relative to displayed  output   Can be adjusted during operation via UP and DOWN arrow  keys     OUTPUT 3500D  3505D   50 60 Hz selectable  3565D and 3570D   DC and 50   60 Hz selectable    0 and 0 2   999 9 seconds  0 1 second   step  SETTING    0    for continuous running  RAMP TIME Mo
47. rounded power source  Also connect the rear panel chassis ground  for additional safety     Turn on the POWER switch located on the lower left hand side of the front panel  Upon  powering the instrument up a POWER ON SELF TEST  POST  will be automatically  performed  This test will check for the condition of the ram chips  pcb   s and other critical  components  In addition the display will show the following message  with the actual  model number displayed in place of model   below     ARI HYPOT II    model   VER  X X       The instrument will recall the last setup which was active and the display will show the  parameters which were programmed into the unit  The instrument is now ready for  operation     B  SETUP PROCEDURE     1  To select the Memory location on models 3505D  3565D and 3570D only  Please press the SET key until the display shows    Use the Up Down arrow keys to select memory locations 1 through 5  After selecting the  memory press the SET key to view the settings that have been recalled from memory or  to make any changes to these settings     To store any changes to the settings press the Exit key  The parameters are automatically  updated into the memory location that was selected  when exiting to the test mode     27    2  To set the test type and sequence for model 3570D only   On model 3570D only the Insulation Resistance  I  and Dielectric Withstand  W  test can  be run independently or as a sequence of tests  When    W    or    I    is selected the  
48. s are  made through front panel keypad in a restricted access  calibration mode  Calibration information stored in non   volatile memory        2  ASSOCIATED    KEY FEATURES  amp  BENEFITS OF Hypot I       10     No load setup of trip current and output voltage   This provides the operator with an easy and safe way to set trip currents and output  voltages since parameters are set without the high voltage activated     Automatic storage of test program   Hypot II powers up with the parameters that were used during the last test to avoid  operator set up errors     All parameters for the setups can be adjusted through a simple menu driven program   The easy to follow setup screens ensure that the operator correctly sets up all test  parameters     Tamper proof front panel controls   This makes it possible to limit user access to the setup screens so that only authorized  personnel with a security code can change test parameters     Line and load regulation    This system maintains the output voltage to within 1  from no load to full load and  over the line voltage range to ensure that test results remain consistent and within  safety agency requirements      The LCD display allows monitoring of current down to 100 microamps AC and 20  microamps DC    This allows the Hypot II to be used even when test requirements only allow a very low  level of acceptable leakage current     Electronic dwell settings   The electronic dwell control helps keep test results consistent by ensuring
49. switch and an alarm will    sound  The display will show     I High MX XXX Xs  X XXKVDC  gt 1000MQ       18  If the leakage current does not exceed the low limit setting and is within the metering  range the red indicator light will illuminate on the RESET switch and an alarm will    sound  The display will show     I Low MX XXX  XS  X XXKVDC XXX  X MO       19  If the leakage current does not exceed the low limit setting and is not within the  metering range the red indicator light will illuminate on the RESET switch and an alarm    will sound  The display will show     I Low MX    X XXKVDC       38    Qne     D  CONNECTION OF REMOTE I O     A 9 pin    D    type connector is mounted on  INPUT OUTPUT control and information     OPERATION    the rear panel which provides REMOTE   This connector mates with standard 9 pin    D       subminiature connector provided by the user  The Remote I O mates to a female   receptacle  connector  For best performance a shielded cable should be used  To avoid  ground loops the shield should not be grounded at both ends of the cable  Suggested AMP  part numbers for interconnecting to the Remote I O are shown below     205203 3 RECEPTACLE SHELL    745253 7 CRIMP SNAP IN SOCKET CONTACT  for receptacle   745171 1 SHIELDED CABLE CLAMP  for either plug or receptacle     741184 3 JACKSCREW SET  2   REMOTE INTERFACE REAR PANEL     1  SIGNAL OUTPUTS ON REMOTE I O       The rear panel connector provides outp          ut signals to remotely monitor PASS
50. t  or press the SET key again to reach the Memory Lock screens     The memory lock screens appear as follows     Memory Lock   ON or Memory Lock   OFF    To toggle the Memory Lock function on and off use the Up or Down Arrow keys  then  press EXIT to select  or press the SET key again to reach the Key Lock screens     Use the following chart to set the Key Lock and Memory Lock for your application                             Key Lock   Memory Lock   Accessible keys   Off Off All   Off On All   On Off TEST  RESET  and Up Down Arrows for memory selection  On On TEST and RESET          12 To set the IR output Test Voltage on model 3570D only  Please press the SET key until the display shows     I Volt   XXX KV    Please use the Up Down Arrow keys to enter the desired test voltage  then press the EXIT  key to exit to the test mode or toggle to another setting using the SET key  The maximum  voltage which may be entered is 1 00 kilovolts DC     31    13  To set the High Resistance Limit on model 3570D only  Please press the SET key until the display shows     I High   XXXX MO    Use the Up Down Arrow keys to enter the resistance high limit setting  then press the  EXIT key to exit to the test mode or toggle to another setting using the SET key  The unit  of measure is in megohms with 0   1000 MQ range of setting  A setting of O MQ disables  the high limit setting  A measured value above the High Limit setting will result in a  failure condition at the end of the Delay time setting
51. the power wiring to the test bench is properly polarized and that the proper  low resistance bonding to ground is in place    Power to the test station should be arranged so that it can be shut off by one prominently  marked switch located at the entrance to the test area  In the event of an emergency   anyone can cut off the power before entering the test area to offer assistance     Work Area  Perform the tests on a nonconducting table or workbench  if possible     There should not be any metal in the work area between the operator and the location  where products being tested will be positioned     Position the tester so the operator does not have to reach over the product under test to  activate or adjust the tester  If the product or component being tested is small  it may be  possible to construct guards or an enclosure  made of a non conducting material such as  clear acrylic  such that the item being tested is within the guards or enclosure during the  test  and fit them with switches so that the tester will not operate unless the guards are in  place or the enclosure closed     Keep the area clean and uncluttered  All test equipment and test leads not absolutely  necessary for the test should be removed from the test bench and put away  It should be  clear to both the operator and to any observers which product is being tested  and which  ones are waiting to be tested or have already been tested     Do not perform Hipot tests in a combustible atmosphere or in any are
52. unts   Resistance Display Range  1   1000 MQ  4 Digit  Auto Ranging   Resolution  500VDC 1000VDC  MQ MQ MQ  0 01 1 00   40 00 1 00   80 00  0 1 35 0   999 9 75 0   999 9  Accuracy     3  of reading   2 counts   at test voltage  gt  500V     7  of reading   2 counts     at test voltage  lt  500V       2  ASSOCIATED    High Resistance Limit Range  0  1000 MQ  0  Off     Delay Timer Range  0  0 5   999 9 sec  0   Constant   Resolution  0 1 sec step  Accuracy     0 1    0 05 sec     GENERAL    REMOTE CONTROL The following input and output signals are provided through  AND SIGNAL the 9 pin D type connector   OUTPUT 1  Remote control   test and reset   2  Outputs   pass  fail and test in process    PROGRAM MEMORY  5 Sets  3505D  3565D and 3570D only     SECURITY Key Lock capability to avoid unauthorized access to all test  parameters  Memory Lock capability to avoid unauthorized  access to memory locations   prong grounding plug    TERMINATIONS 5ft  1 52m  high voltage and return leads  2  with clips and a  standard U S  style  NEMA 5 15  remote receptacle box for  testing items terminated with a line cord  International  receptacles also available    MECHANICAL Tilt up front feet    Dimensions   W x H x D  11 x 3 5 x 14 56 in   280 x 89 x 370 mm   Weight  20 lbs  9 Kgs    Operating Temperature   32     113  F  0     45  C   Relative Humidity   0 to 95     CALIBRATION Traceable to National Institute of Standards and Technology   NIST   Calibration controlled by software  Adjustment
53. up arrow key  If a second test is  not configured an audible beep will be heard when the arrow keys are pressed     Dielectric Withstand Test Mode    3  If you are performing a Ground Continuity Check the Ground Continuity must be set  to the    ON  Enabled     position  The ground check cable has to be connected properly to  the chassis of the DUT and the green indicator on the TEST switch should be illuminated  before activating the Test switch  Otherwise  the instrument will signal a continuity  failure when the test switch is activated  Please refer to page 41 for the proper test  connections using the adapter box  to test items terminated in a three prong plug     33    Qane     The display will show     OPERATION    models 3500D models 3505D and 3565D    CONT Fail         KVAC       model 3570D       If you are not performing a Ground Continuity Check the Ground Continuity must be set  to the    OFF     Disabled  position  Itis not necessary to connect the    Cont Check    cable  since this test will not be performed if the circuit is disabled  The instrument will proceed  with the Hipot Test when the    Test    switch is activated     4  To initiate a test  press the TEST switch on the front panel  The red High Voltage  Arrow indicator will flash and the display will immediately show     models 3500D    Testing XXX  Xs    X XXKVAC XX   XX mA       models 3505D and 3565D    Ramp MX  XXX Xs Dwell MX XXX Xs  X XXKVAC  XX XXmA X XXKVAC XX  XX mA       model 3570D    W Ramp
54. upport department immediately  Please save the shipping  carton and packing material for the carriers inspection  Our customer support department  will assist you in the repair or replacement of your instrument  Please do not return your  product without first notifying us and receiving an RMA  return materials authorization   number     Preparation for Use  Power Requirements and Line Voltage Selection  This instrument requires a power source of either 115 volts AC    CAUTION 15   47 63 Hz single phase or 230 volts AC  15   47 63 Hz single  phase  Please check the rear panel to be sure the proper switch setting  is selected for your line voltage requirements before turning your instrument on  In  addition  please be sure the correct fuse is selected and installed while the instrument is in  the off position   Do not switch the line voltage selector switch located on the rear panel while the  instrument is on or operating  This may cause internal damage and represents a safety  risk to the operator     NOTE    For operation at 115 Volts AC and 230 Volts AC use a 3 15A fast acting fuse     21    o ASSOCIATED    Power Cable    WARNING   BEFORE CONNECTING POWER TO THIS INSTRUMENT  THE    PROTECTIVE GROUND  EARTH  TERMINALES OF THIS INSTRUMENT  MUST BE CONNECTED TO THE PROTECTIVE CONDUCTOR OF THE LINE  MAINS   POWER CORD  THE MAIN PLUG SHALL ONLY BE INSERTED IN A SOCKET OUTLET   RECEPTACLE  PROVIDED WITH A PROTECTIVE GROUND  EARTH  CONTACT  THIS  PROTECTIVE GROUND  EARTH  MUST NO
55. w screen    W   SET  This is the parameter settings review screen for Dielectric Withstand test  mode    M1  Indicates the memory location of the current settings    Mi   Indicates that an Insulation Resistance test will execute after the Dielectric  Withstand test is completed successfully   3570D only    1 0 S  The dwell timer is set to a test duration of 1 second     1 24KVAC  The test voltage is set to 1 240 volts AC   10 00 mA  The high leakage current trip point is set to 10 milliamps     If you wish to not use any one of these parameters you must change your parameters  Please  refer to the Operation Instructions section in this manual for further details     e   If the instrument defaults are acceptable then be sure to connect the appropriate test  leads to the device under test  DUT  or test fixture  Be sure to connect this safety ground to  a suitable known good ground before energizing this instrument  Then connect the return  lead first  black  to the test fixture or item followed by the high voltage output lead  red      WARNING        Please check your connections to be sure they are making good contact    and that the test station or area is clear of debris and other personnel  DO  NOT TOUCH THE DEVICE UNDER TEST ONCE THE TEST HAS BEEN STARTED   To initiate the test press the GREEN test button on the front panel  This is a momentary  button and does not need to be held in the pressed position during the test  The instrument  will then cycle ON and begin the aut
56. y item is as safe as possible  All products must be designed  and built to prevent electric shock  even when users abuse the equipment or by pass built  in safety features     To meet recognized safety standards  one common test is the  dielectric voltage   withstand test   Safety agencies which require compliance safety testing at both the  initial product design stage and for routine production line testing include  Underwriters  Laboratories  Inc   UL   the Canadian Standards Association  CSA   the International  Electrotechnical Commission  IEC   the British Standards Institution  BSD  the  Association of German Electrical Engineers  VDE   the Japanese Standards Association   JSI   These same agencies may also require that an insulation resistance test and high  current ground bond test be performed     The Dielectric Withstand  Hipot  Test       The principle behind a dielectric voltage   withstand test is simple  If a product will  function when exposed to extremely adverse conditions  it can be assumed that the  product will function in normal operating circumstances     The most common applications of the dielectric withstand test are     e Design  performance  Testing     determining design adequacy to meet service  conditions     e Production Line Testing     detecting defects in material or workmanship during  processing     e Acceptance Testing     proving minimum insulation requirements of purchased parts   e Repair Service Testing     determine reliability and s
    