Home
lit-045asunoptics surgical ssx300 xenon service manual
Contents
1. ISO 9001and ISO 13485 Certified Reg No 207267 MP21 BEER Sunoptics SURGICAL 55 3005 SolarMaxx 300 Illuminator Service Manual CE 6018 Bowdendale Avenue Jacksonville FL 32216 Customer Service 904 731 5869 FAX 904 733 0012 Toll Free 800 684 6404 LITO045Sunoptics Surgical Rev A English SolarMaxx 300 Illuminator Service Manual TABLE OF CONTENTS Page SER CTE nones rode eda 3 TECHNICAL SUPPORT SERVICES ia a 3 GENERAL THEORY OF 1 2 1 4 5 CIRCUIT BREAKER REPLACEMENT 2 5 5 ree deo e 6 SHUTTER REPLACEMENT ferie IR 6 FILTER ASSEMBLY 7 LAMP BASE REPLACEMENT ela 7 POWER SUPPLY REPLACEMENT psi 8 COOLING ae 8 LAMP REBUILD PROCEDURE scale 9 10 11 REPLACEMENT PARES 225 DRE RM RA aaa 12 epe LEE 13 Figueroa 14 glace 15 LITO045Sunoptics Surgical Rev A English Page 2 of 15 SolarMaxx 300 Illuminator Service Manual INTRODUCTION This manual has been prepared to aid in the repair and maintenance of the SolarMaxx 300 illuminators The procedures and instructions contained in this document are to be used by qualified technical personnel only Some procedures may have live exposed cir
2. Reach in the lamp access door and hold onto the IR filter assembly while removing the 2 mounting screws from the bottom Remove the assembly through the lamp access door Replace the new assembly into the unit through the lamp access door Note that the IR filter is facing toward the back of the unit Secure with the 2 mounting screws from the bottom Place the unit back on its feet Install the lamp being sure it has totally seated and close the access door LAMP BASE REPLACEMENT CAUTION Before performing this procedure be sure the unit has cooled to room temperature Disconnect the illuminator from the power source English 2 gt Remove the 8 cover screws in the sides of the unit 3 Open lamp access door and lift the top straight up and off the unit Set aside 4 Remove the lamp cartridge assembly gently rocking from front to back Set lamp cartridge aside 5 Disconnect the power supply lamp wires from the lamp base igniter module at the and terminals on the igniter circuit board 6 Place the unit on its side 7 From the bottom remove the 4 screws that hold the lamp base to the bottom panel 8 Remove the lamp base from the unit Place the new lamp base igniter assembly into the unit with the banana jacks in the heat shield area Secure with the 4 screws to the bottom of the unit 9 Place the unit back on its feet 10 Reconnect the lamp power supply wires to the igniter module termina
3. fan Air is drawn through the sides of the unit across the power supply and lamp and exhausted through the rear mounted exhaust louver An infrared filter which is located between the lamp lens and the lamp housing provides additional cooling of the light This filter blocks the infra red heat from the output turret while passing the visible light thus lowering the temperature of the instruments and cables Caution should still be used as there are still potentially hazardous temperatures at the turret PARTS REPLACEMENT PROCEDURES The following procedures are meant to aid the technician engineer in replacing defective or damaged components These procedures are meant to be used by qualified personnel only Extreme caution should be used and all necessary safety precautions taken when working on this equipment LIT045Sunoptics Surgical Rev A English Page 5 of 15 Vr PED B ooo SolarMaxx 300 Illuminator Service Manual CIRCUIT BREAKER REPLACEMENT Disconnect the illuminator from the power source Remove the 8 cover screws on the sides of the unit Open lamp access door and lift the top cover straight up and off the unit Set aside Using an ohmmeter verify which circuit breaker is defective Remove the defective circuit breaker and replace with the new one Replace the top cover making sure the lamp access door is over the lamp cartridge Replace the 8 cover screws and secure SHUTTER REPLAC
4. fill tube relief hold LITO045Sunoptics Surgical Rev A English Page 9 of 15 SolarMaxx 300 Illuminator Service Manual LAMP REBUILD PROCEDURE continued 15 16 20 21 22 23 Secure with 3 lock washers and 6 32 Phillips head screws Apply a thin film of heatsink compound to the inside contact area of the front heat sink Install the new metal spring gasket into the back groove of the front heatsink Place the fingers so they are toward the front Remove the end cap from the front of the lamp Use care not to touch the window of the lamp Apply heatsink compound to the front outside diameter of the lamp Be careful not to get compound on the lamp glass surface If you do clean the lamp with a clean alcohol wipe Press the cooling ring onto the front of the lamp Press the front heatsink assembly onto the lamp cooling ring assembly Align the threaded holds in the bottom of the 2 heatsinks with each other Secure the front heatsink onto the lamp with the lock washer and Allen head screw Align the threaded holes in the bottom of the heatsink with the bottom of the lamp housing and press the entire lamp assembly into the lamp housing so that when the lamp is pointing toward you the housing handle is on the top Install the 2 banana plugs into the bottom of the lamp assembly and tighten PFC LAMP REBUILD PROCEDURE FIG 3 The following procedure will guide you through rebuilding your PFC Xenon lamp car
5. orr swr 1 P1 I3000008 300 watt PwR SUP 1 FI 12000014 12VDC FAN 12000012 12VDC HR METER 1 Li 13000007 XENON 300 i F2 1180006 CORCOM FILTER 1 F3 F4 1180008 INTERNAL FERRITE 2 ss 12000037 HOUR METER RESET swiTcH 1 RETURN LITO045Sunoptics Surgical English Rev A Page 13 of 15 SolarMaxx 300 Illuminator Service Manual FIGURE 2 LITO045Sunoptics Surgical English Rev A Page 14 of 15 SolarMaxx 300 Illuminator Service Manual FIGURE 3 23 o 2 m n 02 3 REQ x z a 5 kK 2 o x a o E lt O COOLING RING BRASS NUT 8 32 SPRING WASHER 2 REQUIRED BLUE HOUSING LIT045Sunoptics Surgical Rev A English Page 15 of 15
6. the front outside diameter of the lamp 15 Press the cooling ring onto the front of the lamp 16 Press the front heatsink assembly onto the lamp cooling ring assembly Align the threaded holes in the bottom of the 2 heatsinks with each other 17 Secure the front heatsink onto the lamp with the copper clips 18 Align the threaded holes in the bottom of the heatsink with the bottom of the lamp housing and press the entire lamp assembly into the lamp housing so that when the lamp is pointing toward you the window of the lamp is pointing to the front panel Install the 2 brass plugs into the bottom of the lamp assembly and tighten LITO045Sunoptics Surgical Rev A English Page 11 of 15 SolarMaxx 300 Illuminator Service Manual REPLACEMENT PARTS DESCRIPTION PART NUMBER Replacement Lamp Cartridge SSR 001 EMI Filter SSR 002 300 Watt Xenon Power Supply SSR 003 Hour meter SSR 004 Circuit Breaker SSR 005 IR Filter Assembly SSR 006 Power Input Module SSR 007 Shutter Assembly SSR 008 Lamp Base Assembly SSR 009 Cooling Fan SSR 010 Reset Switch Assembly SSR 011 Xenon Lamp Rebuild Kit SSR 012 For parts not listed here call our Technical Service Department listed at the front of the manual LITO045Sunoptics Surgical Rev A English Page 12 of 15 SolarMaxx 300 Illuminator Service Manual FIGURE 1 CBI CB2 13000044 7 5 AMP CIRCUIT BREAKER 2 st nz26000 LIMITSWT 102 52 uzessoo on
7. EMENT Disconnect the illuminator from the power source Remove the 8 cover screws in the sides of the unit Open lamp access door and lift the top cover straight up and off the unit Set aside Remove the 3 screws on the bottom of the unit with hold the front panel to the bottom housing Remove the 2 screws on the side of the front panel Remove the screw holding the shutter to the front panel Pull off the shutter assembly Replace the shutter assembly with the new one With the control turned fully clockwise align the shutter in such a way that it does not cover the hole in the front panel Replace the screw and tighten the screw until it holds the shutter in place and still allows for free movement of the shutter Do not over tighten the screw Replace the front panel and secure to the bottom housing with the 5 screws Replace the top cover making sure the lamp access door is over the lamp cartridge Replace the 8 cover screws and secure LITO045Sunoptics Surgical Rev A English Page 6 of 15 SolarMaxx 300 Illuminator Service Manual I R FILTER ASSEMBLY REPLACEMENT CAUTION Before performing this procedure be sure the unit has cooled to room temperature The lamp cartridge and IR filters operate at very high temperatures Disconnect the illuminator from the power source Open the lamp access door and remove the lamp gently rocking from front to back Set lamp aside Place unit on its right side
8. ENSITY CONTROL A rotating stainless steel disc that is placed in front of the lamp controls intensity The disc contains holes in varying sizes and patterns Manual lamp intensity control is made via a front panel mounted knob LAMP POWER AND IGNITION SYSTEM The lamp used in the SolarMaxx 300 illuminators is a 300 watt ceramic arc lamp For ignition of the lamp to occur it takes a high voltage pulse of approximately 20 KV The power supply generates a pulse of approximately 400 600 volts which is then stepped up to the 20 KV pulse by the igniter module circuitry Connection to the lamp is made via two jumbo banana jacks LIT045Sunoptics Surgical Rev A English Page 4 of 15 SolarMaxx 300 Illuminator Service Manual The lamp power and ignition system consists of the lamp power supply lamp base assembly and the lamp cartridge assembly In the event you have a power supply failure it is recommended that the power supply be returned to SUNOPTIC TECHNOLOGIES for servicing When power supply is switched on the DC supplies come up to voltage immediately The lamp power supply has a built in delay of 1 to 2 seconds before it will attempt to start the lamp If the lamp is unsuccessful at igniting the power supply will try 6 to 10 times in rapid succession before ceasing After successful ignition the supply switches to a 15 VDC output at approximately 20 amps of current COOLING SYSTEM Cooling is provided by a 100 CFM 12 VDC
9. cuitry and wiring which could be hazardous if contacted with Use extreme caution when working on equipment that has power applied to it TECHNICAL SUPPORT SERVICES In the event that you experience difficulty or need technical assistance please contact our technical support staff at 800 684 6404 or by fax at 904 733 0012 Please have the following information ready when you call e MODEL NUMBER SERIAL NUMBER e DETAILED DESCRIPTION OF THE PROBLEM LITO045Sunoptics Surgical Rev A English Page 3 of 15 SolarMaxx 300 Illuminator Service Manual GENERAL THEORY OF OPERATION The SolarMaxx 300 Illuminator general operation is as follows Please consult Fig 1 for wiring information A C POWER DISTRIBUTION The SolarMaxx 300 llluminators are based around a universal input power supply Input line voltages of 100 120V 220 240V 7 at 50 60 Hz are applied at the Power Input Model via a Hospital grade power cord The input power is then filtered by a low leakage current EMI Filter Over current protection is provided by two 7 5 amp circuit breakers double pole interlock switch provides operator safety which is located on the back left side of the lamp heat shield DC POWER DISTRIBUTION DC power for the other system components is generated by the lamp power supply In addition to the 15 volts DC lamp power it generates 12 VDC The 12 VDC is used to power cooling fan and the elapsed lamp hour meter INT
10. e hour meter wiring harness connector from the power supply Remove the 3 screws holding the back panel to the bottom housing and lay down the panel Remove the fan mounting screws and nuts This will free the fan louver also Replace the fan assembly with the airflow blowing out the back of the unit The fan leads should be coming out of the fan at the tip and against the back panel Secure the fan mounting screws with the nuts Slide the back panel into the mounting slots Reconnect the fan lead connector and the hour meter wiring harness connector to the power supply Replace the top cover making sure the lamp access door is over the lamp cartridge Replace the 8 cover screws and secure LITO045Sunoptics Surgical Rev A English Page 8 of 15 SolarMaxx 300 Illuminator Service Manual LAMP REBUILD PROCEDURE See Fig 2 and Fig 3 PFC units for reference When rebuilding the Xenon lamp cartridge it is recommended to replace several key components of the cartridge assembly These components are included when you purchase a replacement rebuild kit from Sunoptic Technologies CAUTION Before performing this procedure be sure the unit has cooled to room temperature WARNING Lamps are under high pressure and must be handled with A face shield should be worn when handling this lamp The following procedure will guide you through rebuilding your Xenon lamp cartridge Tool required 2 Philips scr
11. ewdriver 7 64 Allen wrench Needle nose pliers and nut driver Be sure lamp has cooled to room temperature Remove the 2 jumbo banana plugs from the bottom of the unit Press the lamp assemblies from the back out of the lamp housing Set the lamp face down on its front Heatsink Remove the 3 Phillips screws and lock washers from the back heatsink Gently pull the lamp out of the back heatsink Remove the Allen head screw in the side of the front heatsink Gently pull the lamp out of the front heatsink With the needle nose pliers remove the cooling ring and from the inside of each lamp heatsink the metal spring gaskets Clean off all the parts with a rag or paper towel alcohol may help in heatsink compound removal 10 Remove the new lamp from its package Record its serial number and date installed in the illuminator manual 11 Setthe lamp face down with its protective cap on its lens If the lamp has a rubber cap on the fill tube on the rear of the lamp remove it 12 Apply a thin film of heatsink compound to the back of surface of the lamp and the inside contact area of the front and back heat sinks Mi M Oo 5 CAUTION Only apply thin layer of heatsink compound to areas 13 Install the new metal spring gasket into the front groove of the back heatsink Place the fingers so they are toward the back 14 Press the lamp into the back heatsink making sure the fill tube is aligned with the
12. ls 11 Install the lamp cartridge being sure it is totally seated 12 Replace the top cover making sure the lamp access door is over the lamp cartridge 13 Replace the 8 cover screws and secure LITO045Sunoptics Surgical Rev A Page 7 of 15 On SolarMaxx 300 Illuminator Service Manual POWER SUPPLY REPLACEMENT Disconnect the illuminator from the power source Remove the 8 cover screws in the sides of the unit Open lamp access door and lift the top cover straight up and off the unit Set aside Disconnect the fan lead connector the AC power input connectors and the lamp output power wires from the power supply Set the unit on its side While holding the power supply with one hand remove the 4 mounting screws from the bottom of the unit Place the new power supply in the unit so all 4 holes line up with the power supply standoffs Secure with the 4 mounting screws from the bottom Place the unit back on its feet Reconnect the fan lead connector the AC power input connectors and the lamp output power wires to the power supply Replace the top cover making sure the lamp access door is over the lamp cartridge Replace the 8 cover screws and secure COOLING FAN REPLACEMENT Disconnect the illuminator from the power source Remove the 8 cover screws in the sides of the unit Open lamp access door and lift the top cover straight up and off the unit Set aside Disconnect the fan lead connector and th
13. tridge Tool required Screw driver and adjustable wrench 1 sure lamp has cooled to room temperature 2 Remove the Brass plugs from the bottom of the unit Press the lamp assemblies from the back out of the lamp housing 3 Setthe lamp face down on its front Heatsink 4 Remove the copper clips from both heatsink 5 Gently pull the lamp out of the heatsinks 6 Remove the cooling ring from the lamp 7 Clean off all the parts with a rag or paper towel alcohol may help in heatsink compound removal 8 Remove the new lamp from its package Record its serial number and date installed in the illuminator manual 9 Set the lamp face down with its protective cap on its lens If the lamp has a rubber cap on the fill tube on the rear of the lamp remove it 10 Apply a thin film of heatsink compound to the metal surface of the lamp and the inside contact area of the front and back heat sinks Be careful LITO045Sunoptics Surgical Rev A English Page 10 of 15 SolarMaxx 300 Illuminator Service Manual not to get compound on the lamp glass surface If you do clean the lamp with a clean alcohol wipe CAUTION Only apply a thin layer of heatsink compound to areas 11 Press the lamp into the back heatsink making sure the fill tube is aligned at six OCLOCK 12 Install copper clips 13 Remove the end cap from the front of the lamp Use care not to touch the window of the lamp 14 Apply heatsink compound to