Home
        operating manual for the Wescore 5520 Osmometers
         Contents
1.     Instrument                   Broken or       DamagedO  Thermocouple      Is S D  2 0         less    Check for LooseO    TC Screws Yes    ReplaceO  ThermocoupleO    Evaluate SampleD  LoadingO  Technique    Head             Verify That YouLl  are Using 1000  Standards                Is There Visible  Contamination             Yes    No       59    SECTION 5  SOLVING PROBLEMS    5 1 Troubleshooting    The following is a detailed presentation of material covered in the Troubleshooting  Flowchart  This material is indexed to further information found throughout the manual     60    Problem  Display is blank     There is a large  shift in calibration     Solution  Check to see if power indicator is lit     Check to see if power is on   Check for power on mains   Check fuses  see Sections 1 5 and 2 8      Check the display and keyboard connections     To avoid the risk of serious injury  the display and  keyboard connections should only be checked by  qualified service personnel     Reset instrument by turning off power for 3 seconds   Then turn power back on     If none of these steps resolves the problem   contact your dealer or Wescor for assistance     Check the Temperature Drift Scale on the display    If the scale indicates that ambient temperature is  outside acceptable levels  take steps to stabilize the  instrument  See Section 2     SECTION 5    SOLVING PROBLEMS  5 1 Troubleshooting  Problem Solution  There is a large Check the freshness of the calibration standards  
2.     Problem Symptom s     Thermocouple Contamination A shift in calibration  Error message  during calibration or Clean Test     Deformed or Flattened Loss of high range readings or  Thermocouple precision   Broken Thermocouple ERROR message on display or    wildly erratic behavior if the connection  is intermittent     Disconnected TC Head ERROR message on the display   Connector  Loose TC Head Screws Unstable calibration and erratic    readings  ERROR message during  calibration or Clean Test     65    SECTION 5  SOLVING PROBLEMS    5 2 Common TC Head Problems    Problems can often be solved by inspecting and  cleaning the TC head     1 Follow the instructions in Section 4 to remove  the TC head     NOTE    To improve temperature stability inside the osmometer   leave the access cover in place while the TC head is out  of the instrument              Carefully inspect the TC head under a micro   scope  Check for any gross contamination on  the thermocouple or thermocouple mount     Contamination is a natural consequence of normal  use of the osmometer  It may also occur inadvertent   ly during shipping or set up  Contamination changes  the linearity of the instrument response  first  detectable in the lower ranges of osmolality        Contamination does not generally degrade precision   but  depending upon the nature of the contaminating  substance  this can occur  See Section 4 for complete  instructions to detect and remove contamination     NOTE    Contamination can be in
3.    Purified Water   Liquid Dropper   Blow Clean Liquefied Propellant or Equivalent    Pressure limited to 20 psig      NOTE        To remove significant contamination  use the Wescor  cleaning solution followed by numerous successive rins   es with pure water  Wescor   s cleaning solution is approx   imately 8  ammonium hydroxide  Concentrated ammoni   um hydroxide can be used to remove particularly stub   born contamination  as described in Section 4 6  Lint or  dust particles can usually be removed by simply rinsing  with water a number of times     Place a waste container close by on the floor     Use a cotton swab to remove residue from the  surface of the mount surrounding the thermo   couple     CAUTION           Do not contact the thermocouple with the swab     With the dropper  release cleaning solution onto  the thermocouple mount     Immerse the thermocouple and the entire sur   face of the mount in cleaning solution  Let  stand at least 1 minute     SECTION 4  PREVENTIVE MAINTENANCE    43 Cleaning the TC Head          4 Hold the TC head over the waste container        5 Quickly pull the TC head straight down and  away from the droplet of liquid  allowing liquid  to fall into the waste container directly below     6 Immediately apply rinse water before evaporation  can occur  Use purified water with resistivity of 1                        or higher for rinsing  Water of lesser  quality will contaminate the thermocouple     CAUTION    Do not contaminate the rinse wat
4.   plunger     NOTE        Normally  sample droplets will not cling to the outside of  the tip  If they do  they can usually be removed simply by  dragging the tip against the lip of the container as you  remove it  Occasionally a clinging droplet may have to be  removed with a tissue  but be very careful not to wick solu   tion out of the tip     35    SECTION 3    OPERATING THE VAPRO OSMOMETER    3 2 Loading Samples    36    Touch Tip to Sample Disc     Press Disc       lat     With the pipettor tip resting in the notch of the  pipettor guide  position the tip about 5 millimeters  above the center of the sample disc     Smoothly depress the micropipettor plunger to the  stop  The specimen may drop onto the sample  disc  Whether the sample droplet falls onto the  disc or clings to the tip  you must complete Step 6     CAUTION        Never allow the micropipettor tip  sample material  or  the wet disc to touch the outer surface of the sample  holder  If this occurs  abort the measurement and  wipe the sample holder clean before proceeding     CAUTION        Break any air bubbles on the sample disc before  proceeding  A bubble bursting inside the sample  chamber will contaminate the thermocouple     With the plunger still held against the stop  lightly  touch the micropipettor tip to the sample disc  then  lift it away  The tip must briefly contact the sample  disc to press it flat against the holder  The paper  disc should appear fully saturated  with a slight liq   uid menisc
5.  34  45  46  48   54  58  61  62  65  66  85  damage 55  59  62  63  65  66  70  85  deformed 62  63  65  66  68  restoring 69  70  junction  bead  70  head  access cover 12  13  15  46  52    Thermocouple                   cleaning 48 50  55  66  cleaning supplies 18  48                     15  46  51  59  65  66  67  inspecting 66  oxidation 56  removing 45  46  47  66  reinstalling  51  67  screws 59  61  62  64  65  hygrometer 10  76  hygrometry 76  mount 10  65  70  sensor 65  temperature graph 78  79  Tissue Paper 82  Troubleshooting Flow Chart 58  59    U  Unit of Measure 20  101  User s Manual 7  18    Vv   Vapor Equilibration 31   Vapor Pressure  determination 9  equilibration 10  method 9   Viscosity  sample 9  81   Voltage 71   Voltage Selector 16  27  changing 27   Volumetric error 68  88    101    
6.  Aligns and steadies the pipette for precise application of the speci   men onto the sample disc in the sample holder     Sample Holder   Standard sample holder for samples up to 10 microliters in volume   Requires solute free paper discs  provided  for use  Other sample  holders are available for special applications  see Appendix B and D      Sample Slide  Moves the sample holder from the loading position  under the  pipette guide  to the sample chamber     Power Indicator  The green light indicates the instrument is on     INSTRUMENT TOP  TC Head Access Cover    Provides access to thermocouple head for cleaning and mainte   nance  Press down on the right side of the cover to remove     13    Serial Data Port Signal Chime       Model and Serial    Sample Chamber Lever  Numbers     Shown in the Open  Position        Power Entry Module    14    SECTION 1  INTRODUCTION    1 5 Controls and Features    INSTRUMENT RIGHT SIDE    Sample Chamber Lever    Opens and closes the sample chamber  Closing the sample cham   ber locks the sample holder within the chamber  The chamber  should remain closed  except when loading or removing samples   Closing the chamber initiates the measurement cycle or the     Standby    mode  indicated by READY on the display screen  if no  sample is present     INSTRUMENT INTERIOR    The thermocouple  TC  head and thermocouple head connector are  accessed by removing the chamber access cover from the top of the  instrument     INSTRUMENT REAR PANEL    Signa
7.  Clean Test reveals moderate  levels of contamination will save time  Cleaning is much more diffi   cult if you wait until contamination interferes with normal calibration  settings     Cleaning the TC head requires removing it from the instrument   Carefully follow directions to safeguard the thermocouple and  ensure successful completion of the cleaning process     45    SECTION 4  PREVENTIVE MAINTENANCE    4 2 Removing the TC Head       1 Turn the power switch off     Rotate the sample chamber lever to the open    position        Remove the TC head access cover from the  top of the osmometer by pressing down on the  right hand edge  then lifting the raised edge up    and away             Remove the TC head connector by squeezing  the locking tab and lifting        46    SECTION 4  PREVENTIVE MAINTENANCE    4 2 Removing the TC Head    5 Using the 9 64 inch hex driver  completely  loosen  but do not remove from the TC head   the attachment screws     Grasp the top of the TC head  with the attach   ment screws  and lift it straight up and out of  the instrument  Replace the access cover while  the TC head is out of the instrument           7 Remove the attachment screws from the head        CAUTION   To avoid thermocouple damage  invert the TC head  with the thermocouple facing up  before setting it down        47    SECTION 4    PREVENTIVE MAINTENANCE    4 3 Cleaning the TC Head       48       Materials needed to clean the TC head     Wescor Cleaning Solution   Cat    SS 003
8.  SECTION 3    OPERATING THE VAPRO OSMOMETER    3 1 Operation Overview    30      zmolalitu    2932 mmo1 kg      zmolalitu          nnorz ka       Normal Mode    Sample  3  Mean       5 0  1 21  High 291  Low 288       Average Mode    The modes are described below     Normal Mode   For routine running of single samples  Does not dis   play any statistical data  This mode is the default  setting upon power up of the instrument     Average Mode   Will run single samples while maintaining statistical  data on up to 32 samples  These data include the  number of samples run  1 to 32   mean  standard  deviation  highest result and lowest result     NOTE        If a 33rd sample is run  the result will displace the first  sample  a 34th sample will displace the second sample   and so on  The statistical evaluation will always be based  upon the latest 32 samples  if more than 32 samples are  assayed     Average Mode is useful when you need the best  possible accuracy  When the instrument is calibrated  in Average Mode  calibration is adjusted to the mean  value of the assayed samples  Calibration resets the  operation mode to Normal Mode  see Section 3 5      NOTE        We recommend calibrating in Average Mode using 3 or 4  samples of each calibrating solution value     To restart a new set of precision values  bring up the  Mode Menu  the arrow should be pointing at  Average Mode   The next sample assayed will be  sample  1 in the data group  Or  select the Mode  Menu immediately af
9.  Service  MANUALS and INSTRUCTIONAL MATERIALS  M2468 5520 Vapro Osmometer User s Manual  M2469 5520 Vapro Osmometer Service Manual  V 1003 Thermocouple Cleaning Video  VHS Format  specify NTSC  PAL  or    SECAM     74    APPENDIX  C    Theory of Operation    Osmolality is an expression of the total concentration of dissolved  particles in a solution without regard for particle size  density  con   figuration  or electrical charge  Indirect means for the measurement  of osmolality are afforded by the fact that the addition of solute par   ticles to a solvent changes the free energy of the solvent molecules   This results in a modification of the cardinal properties of the sol   vent  i e   vapor pressure  freezing point  and boiling point   Compared with pure solvent  the vapor pressure and freezing point  of a solution are lowered  while its boiling point is elevated  provided  that a single solvent is present in the solution  Solutions containing  more than one solvent generally behave in more complex ways        In single solvent solutions  the relative changes in solution proper   ties are linearly related to the number of particles added to the sol   vent  although not necessarily linearly related to the weight of  solute  since solute molecules may dissociate into two or more ionic  components  Since these properties all change linearly in proportion  to the concentration of solute particles  they are known as  colliga   tive  properties     Osmotic pressure is also a co
10.  a contaminant  a gross contam   ination or fiber  on the thermocouple that is holding  vapor     If the displayed contamination level is greater than  10  you should perform the thermocouple cleaning  procedures found in Section 4     SECTION 3  OPERATING THE VAPRO OSMOMETER    3 5 Calibrating    For optimum operating accuracy  the instrument must be correctly  calibrated according to the instructions in this section  Calibration  accuracy depends upon three main factors    e Standard solution accuracy    e Thermocouple cleanliness    e Loading technique  repeatability     NOTE   Use Optimol glass encapsulated ampule standards for calibration        Check calibration following the initial equilibration period after you  first set up the instrument  Thereafter  we recommend that you  check calibration before each session of use     NOTE     Calibration values are saved by the instrument when power is interrupted        Instrument Response Characteristics   Most clinical osmolality determinations range from 200 to 1000  mmol kg  The inherent linearity of the vapor pressure method  results in an extremely linear instrument response from 100  mmol kg to the upper limit of the osmometer   s range     41    SECTION 3    OPERATING THE VAPRO OSMOMETER    3 5 Calibrating    42      zmolalitu    2932 nmol ks       SELECT          Osmolality    290 mmol kg    3Calibrate    Clean Test  Delete Result  Meru    Mode       Initial or Routine Calibration    1 Run a 290 mmol kg standard     I
11.  and compensates for changing  ambient temperatures that would otherwise result in  calibration changes  It also maintains a continuous  balance in its thermocouple control circuitry to  ensure convergence of the thermocouple to the pre   cise dew point temperature during the measurement  cycle     These internal functions are necessary to maintain  accurate performance  That is why we recommend  that the osmometer be left under power when not in  use  It is also why long measurement sessions  should be interrupted periodically to allow the instru   ment one full measurement cycle on a dry  empty  chamber     NOTE        Occasionally  after a series of runs  an osmolality reading  appears on the screen after cycling on an empty cham   ber  This may be due to residual moisture on the sample  holder  If this occurs  withdraw the sample slide and thor   oughly clean the sample holder using lint free tissue   Then return the slide to the measurement position and  close the sample chamber     43    SECTION 3  OPERATING THE VAPRO OSMOMETER    3 7 Serial Data Output       The 5520 serial port uses a DB9 connector on the  instrument back panel  This port is for asynchronous  serial communication with a printer or computer  It  uses standard non return to zero  NRZ  format at  RS 232 voltage levels     The instrument senses when the RTS  pin 7  is  active     When a sample is assayed while a device is con   nected to the 5520 s RS 232 port  the display status  line will show     EHTER
12.  can be represented  by a simple model  i e   water as solvent with nonvolatile solutes   will have a linear  uniform relationship among all of the colligative  properties  vapor pressure  freezing point  boiling point  etc    In  addition  most of these same solutions can be uniformly frozen with  few artifacts arising from the freezing process  Thus  one can  expect to obtain very similar results  if not exact duplication   between freezing point and vapor pressure measurements on the  vast majority of clinical specimens     Aside from this broad category of solutions  there is a small but  important class of solutions that may be encountered in clinical work  where the colligative relationships do not necessarily hold  These  are solutions in which non physiological volatile solutes   actually  solvents   are present  In such cases  the interactions among the  various molecules cause the properties of such solutions to be more  complex  They generally do not follow linear relationships  as in  solutions having only a single solvent  It must be remembered that  osmometers for clinical applications  whether based on freezing  point or vapor pressure methodology  determine the osmolality of  solutions by indirect means  When complex solutions are encoun   tered  the results obtained by either of these instruments may not  faithfully represent the osmolality of the solution  Each instrument  will respond to the parameter it is designed to measure  and the  resultant indication
13.  event  you may as well make the attempt   since a badly deformed thermocouple will not func   tion  Because of the delicate nature of the task  you  will need steady hands and a microscope  preferably  stereoscopic  having magnification in the range of  30X to 60X     69    SECTION 5  SOLVING PROBLEMS    5 4 Deformed or Broken Thermocouple    1 Make a tool by cutting a thin sliver or wedge    J from a sheet of ordinary paper        2 Work the pointed end of the paper sliver under  the thermocouple wire     3 Use the paper sliver to lift and reshape the  thermocouple  The paper sliver is sufficiently   flexible to avoid undue stress on the thermo   couple wires  Shape the thermocouple to a  rounded arch that is perpendicular to the sur   face of the TC mount  as illustrated  The junc   tion  bead  should be at the high point of the  arc     4 Thoroughly clean the thermocouple  Section 4   before reinstalling the TC head     Broken Thermocouple   Usually  a broken thermocouple is readily evident   especially under a microscope  On rare occasions   the thermocouple may have an intermittent electrical  connection that will cause highly erratic behavior in  the osmometer  A break at either of the thermocou   ple connection points may require meticulous  inspection to discover  A broken thermocouple  requires replacement of the TC head  Contact your  dealer or Wescor for assistance        70    APPENDIX       A  Instrument Specifications  Sample Volume 10 pL nominal  Larger sample
14.  in depth  information throughout this manual     The suggestions listed here are intended to help you quickly solve  routine problems  For unusual problems which require more  detailed information about the operation of the osmometer  refer to  the Vapro Service Manual     If you have tried all these suggestions and still need help  contact    your dealer or Wescor for assistance  Refer to Customer Service   Section 1 2     57    SECTION 5  SOLVING PROBLEMS    5 1 Troubleshooting          Large Shift inO Poor Linearity inO    No Display Calibration Low Range                  Is PowerO  Indicator Lit            Is Temperature    r    Yes  Stable     No p  Temperature toO  Stabilize    Yes    Check if PowerO  Switch is On    Obtain Fresh O  Standards    Are CalibrationS  StandardsO  Fresh          Yes    Check           Power         Receptacle             MESSAGEL   Error Use OnlyO  With 100     Run  CleanO  Test  Using 1000  mmol kg             Check Fuses       ContaminationL   level  gt  10     Yes           Clean     Thermocouple            Check DisplayO  Connector       RecalibrateO  Instrument         InstrumentO  Repair Required           Performed By Qualified Personnel Only  58    SECTION 5  SOLVING PROBLEMS       5 1 Troubleshooting       MESSAGEO   ERROR CheckO MESSAGEQO AbnormalOl    Erratic Heading  Calibration Error  Display    Thermocouple        Run an Auto 0  Repeat usingO  1000        Check TCO  Connector    RestartO          Verify Calibration  0  Standard
15.  or immedi   ately after running a prior sample but before opening  the sample chamber  After you select a mode  all  samples are processed in that mode until you select  a different mode  Functions are specific actions the  osmometer can perform  The Function Menu is not  available until you have assayed a sample             wexcar            Chamber Oren  Press SELECT to move the selection arrow on the  7 g 1 mmol  kg display  Press ENTER to activate the selected menu  item  If you continue to press SELECT the arrow will  ahorn al Mose return to the top of the menu     Auto Rereat  Process Delay MODE MENU    Mode Menu The Mode Menu appears upon opening the chamber  while the instrument is in the  standby  mode  the  instrument has cycled on an empty chamber   or          Osmolality when you select Mode Menu from the Function  1 Menu       when        press SELECT while the                                                 Calibrate   Clean Test When you activate the Mode Menu  the arrow points  Delete Result to the current mode  When you close the chamber   Mode Menu or press ENTER with the chamber open  the instru     ment executes the sample measurement in the  selected mode  You can change the mode at any  time before the measurement cycle begins  or on  the last sample before the sample chamber is  opened  To activate the Mode Menu after a sample  has been run  either select Mode Menu from the  Function Menu or open the chamber and press  SELECT     Function Menu    29   
16.  shift in calibration  and replace if needed  See Section 4     Run the Clean Test  If the contamination  level is greater than 10  clean the  thermocouple  If the contamination level  is less than 10  recalibrate the instrument     There is poor linearity Check the Temperature Drift Scale on the display to see  in the low range  below if temperature is stable  If necessary  allow the  200 mmol kg   instrument to stabilize  See Section 2 3     Verify that you are using fresh calibration standards   See Section 4     Run the Clean Test  Section 3 4  using the 100  standard  If contamination is greater than 10   clean the thermocouple using the instructions in  Section 4     If you run the Clean Test and the dis   play shows    ERROR Use only with 100  mmol kg standard     this may indicate that  ERROR you incorrectly used a 290 mmol kg or  only with 1000 mmol kg standard for the test  Run    ise  100        ka the Clean Test again using the 100 mmol kg  Standard standard        If while using the100 mmol kg standard you  see    ERROR Use only with 100 mmol kg  standard on the display  check for loose TC  head screws     61    SECTION 5  SOLVING PROBLEMS    5 1 Troubleshooting    Problem  Continued   There is poor linearity  in the low range  below  200 mmol kg      Scrambled or erratic  reading on the display  or poor repeatability     The display shows        62    Solution   If  after taking the above steps  the error  message again appears after running the  Clean Test  che
17.  the  instrument into the outlet  To change the voltage  selector     1 Switch the power off and remove the power  cord from the power entry module     Use a screwdriver to open the fuse door from  the switch  left  side     Pull the voltage selector away from the mount   ing slots     Rotate the indicator until the correct voltage is  facing directly outward  then press it back into  the slots        W N       Change the fuses to match the new voltage  setting  See Section 2 8 for fuse replacement  instructions             6 Close the fuse access door and verify that the  correct voltage now appears in the indicator  window        115Vac   230Vac    27    SECTION 2  GETTING STARTED    2 8 Changing Fuses    To access the main fuses     1 Turn power off and disconnect the power cord        2 Use a small screwdriver to open the fuse door     For continued protection against fire hazard  onlv use  fuses of the correct type and rating   Fuse Specifications   115 V Setting  1 8 ampere time delay type  1 4    x 1  1 4      2 required   230 or 240 V setting  1 16 ampere time   delay type  1 4    x 1 1 4     2 required      Refer to Specifications  Appendix A      28    SECTION 3  OPERATING THE VAPRO OSMOMETER    3 1 Operation Overview    Menu Selections   Two main menus  Mode and Function  are used to  select modes and functions of the osmometer   Modes control how the osmometer processes sam   ples and displays results  Use the Mode Menu to  select the mode before loading a sample 
18.  to Send    To send data to the external device  press ENTER   The display will show     Osmolality    See Appendix F for more information     44    SECTION 4  PREVENTIVE MAINTENANCE    4 1 Preventive Maintenance Overview       Cleaning the thermocouple  TC  head is the only routine mainte   nance required by the Vapro osmometer  This section will guide you  through the necessary steps of removing  cleaning and reinstalling  the TC head  Also included are methods for identifying and resolv   ing more difficult thermocouple contamination situations     During normal use  dust or lint particles gradually accumulate in the  sample chamber  Reasonable care in loading and removing sample  material from the sample holder usually makes it possible to run at  least 100 assays before cleaning becomes necessary     Gross contamination is usually a result of incorrect sample loading  or incompletely removing sample material from the sample holder  following an assay  When correctly operated  sample material will   never contact internal chamber parts  See Section 3 2     Under heavy use  run the Clean Test when the osmometer has  assayed 100 samples  Record the results of this test  When the  Clean Test shows moderate contamination  Clean Test reading of  around 10  try rinsing the TC mount as explained in Section 4 3  If  simple rinsing fails to correct the problem  you will need to perform  the full cleaning procedure as outlined in Section 4 3     Cleaning the TC mount as soon as the
19. 1  APPENDIX F  Serial Data                                                 93  APPENDIX G  Setup Men                                             95  INDEX                             alba a ra puq acum 97    SECTION 1  INTRODUCTION    1 1 User s Manual Overview    Thank you for purchasing the Vapro    vapor pressure osmometer   You will find it to be a valuable investment and an important partner  in the laboratory     The Vapro User s Manual is your key to efficiently operating this  instrument  We recommend that you become thoroughly familiar  with the operation procedures and troubleshooting techniques  described in this manual     Information is presented in a step by step format to demonstrate the  operation and care of the instrument from a first time user s point of  view  Once you become familiar with the operation of the Vapro  osmometer  the manual will help you maintain the instrument in a  high state of precision and reliability     SPECIFICATION OF SAFE USE     Using this instrument in a manner not specified by Wescor may  impair the safety protection designed into the equipment and may  lead to injury     SAFE USE ENVIRONMENT     This equipment is designed to be safely operated at 5 to 35 C  max   imum relative humidity 8096     FUSE  All fuses in this equipment are time lag  Type T      EXPLANATION OF SYMBOLS FOUND ON EQUIPMENT      7 Alternating Current  AC     Power On    O Power off    JN International Attention Symbol  Calls attention to important  informat
20. Added kg Osmolality Osmolality Osmolality   mmol kg   pL   mg   mmol   mmol kg   mmol kg   mmol kg   289 2500 1953 42 331 340 287  289 5000 3905 85 374 392 285  289 10000 7810 170 459 501 282  289 25000 19525 424 713 798 277  289 50000 39050 849 1138 1400 250    out of cal       1  Instruments gave identical results on serum alone    2  Assuming 100  ethanol  with a relative gravity of 0 78     90    APPENDIX  E    Osmolality Standards    STANDARD INTERNATIONAL  SI  UNITS OF OSMOLALITY    Osmolality  by definition  is an expression of the total number of  solute particles dissolved in one kilogram of solvent without regard  for particle size  density  configuration  or electrical charge     Traditionally  osmolality has been expressed as milliosmols per kilo   gram  with various abbreviations such as mOs kg  mOsm kg  and  mOsmol kg  The letters  Os  signify that osmolality is defined as the  concentration  expressed on a molal basis  of the osmotically active  particles in true solution  Thus  one mole  1000 mmol  of sodium  chloride dissolved in a kilogram of water has an ideal osmolality of  2000 mOsm kg  since a molecule of sodium chloride dissociates in  solution to produce two ions  that is  two osmotically active particles     In fact  a molal solution of sodium chloride has an osmolality value  slightly less than the ideal because the residual mutual attraction of  the hydrated ions reduces their mutual independence due to the  osmotic coefficient  Since this coefficie
21. Clean Test using the instruc   tions in Section 3 4        39    SECTION 3    OPERATING THE VAPRO OSMOMETER    3 4 Running the Clean Test    40      zmolalitu    100                Calibrate  Clear Test   Delete Result   Mode Menu       SELECT    d   nnolzks    Contamination            Z             The Clean Test is a diagnostic feature that compares  two consecutive sample assays and uses the differ   ence to determine the contamination level of the  thermocouple     When To Run the Clean Test   We recommend running the Clean Test before each  session of osmometer use  after calibration  or any   time you notice a significant  10 mmol kg or more   shift in the 100 mmol kg calibration  Check thermo   couple cleanliness before assaying samples that  require good linearity and accuracy in the low range   After cleaning the thermocouple  use the Clean Test  to verify the effectiveness of the cleaning     Instructions  1 Run a 100 mmol kg standard sample in Normal  Mode  Observe the reading  Press CALIBRATE     2 Before opening the sample chamber  press  SELECT to reveal the Function Menu  Press  SELECT again to point the selection arrow at  CLEAN TEST     3 Press ENTER  The instrument performs a sec   ond assay of the loaded sample and displays  the difference between the first and second  assay in approximately 2 to 3 minutes     NOTE    If the thermocouple fails to dry within 4 minutes  the  instrument will report    Check Thermocouple Head     This  indicates the presence of
22. VAPRO      VAPOR PRESSURE OSMOMETER  MODEL 5520    USER   S MANUAL    M2468 4       1995  1998  2000  2002 Wescor  Inc   All rights reserved  Printed in the United  States of America     Wescor  Vapro  Optimol  Osmocoll and  Blow Clean are trademarks of Wescor  Inc     Other trade names used in this manual  may be trademarks of their respective  owners  used here for information only     The information in this manual may  change without prior notice     TABLE      CONTENTS             SECTION   INTRODUCTION   1 1 User s Manual Overview                                7  1 2 Customer                                                    8  1 3 Vapro System                                                   9  1 4 How the Vapro Works                                 10  1 5 Controls and                                               12  SECTION   GETTING STARTED   2 1 Instrument Setup                                             17  2 2            Accessories                                    18  2 3 Setting Up the                                               19  2 4 Micropipettor Information                              22  2 5 Using Optimol   Osmolality Standards                  24  2 6 Using Osmocoll      Serum Control                      26  2 7 Changing Voltage                                             27  2 8 Changing FUSES                                      28  SECTION   OPERATING THE VAPRO  OSMOMETER   3 1 Operation Overview                                  29  3 2 Loa
23. ample  disc or does not spread out over the whole disc naturally  it may be  preferable to eliminate the sample disc and use the pipettor tip to  apply the material as uniformly as possible over the central depres   sion of the sample holder     In other situations  materials can be sampled successfully by  immersing the paper sample disc  which is held in the forceps  into  the specimen to be tested  then carefully transferring the wet disc to  the central depression of the sample holder  Caution must be exer   cised when using this  disc immersion  technique to avoid any con   tact of the wet sample disc with the outer portion of the sample  holder  since this would result in solute material being transferred to  the thermocouple mount and would rapidly contaminate the sample  chamber     In any event  when working with unusual specimens  make certain  the sample occupies the full diameter of the central depression in  the sample holder  as it would if saturated into a paper sample disc   The thickness of the specimen should be as small as possible     88    APPENDIX  D       Special Application Notes    OSMOMETRY WITH MULTI SOLVENT SOLUTIONS    Biological solutions  in general  are aqueous in nature  Most speci   mens submitted to the clinical laboratory for testing  both pathologic  and normal  will exhibit characteristic properties that are essentially  attributable to the cardinal properties of water  as modified by the  dissolved solute particles  Such solutions  which
24. ceive Data  output   3 TXD Transmit Data  input   oe eeo 4 N C No Connection  5 GND Signal Ground  passive   6 7 8 9 6 DSR Data Set Ready  output   7 RTS Request to Send  input   8 CTS Clear to Send  output   9 N C No Connection    DSR is tied true whenever the instrument power is on  DCD and  CTS are internally tied together     The serial port is configured as Data Communications Equipment   DCE   This enables the osmometer to be connected directly to  most computers and printers which are usually configured as Data  Terminal Equipment  DTE   Use a STANDARD PC AT type 9 pin to  25 pin serial cable  Do not use a null modem cable unless your  device is configured as DCE     Data output is in ASCII characters  Upon power up the osmometer  will output the characters  READY  at the serial port  At the comple   tion of a sample assay the instrument looks for RTS to be true  If  this line is high   ENTER to send  is displayed on the top line of the  display  Press ENTER at this time to output the data on the serial  port  The data format is as follows     20 hex  space    Reading    mmol kg    OA hex  line feed    OD hex  carriage return     94    APPENDIX  G    Setup Menu    The Setup Menu allows you to select available lan   guages  English  French  or German and available  measurement units  mmol kg or kilopascals at 25   C      2 5 x mmol kg   It also allows you go run a Self  Test on the instrument to check basic input and out   put functions     Language and Units of Measurm
25. ck the TC mount for gross   visible  contamination  A grossly contaminated  TC mount requires extensive cleaning and may  require replacement  See Sections 5 2  5 3  and  5 4  If these procedures fail to resolve the  problem  contact your dealer or Wescor for  assistance  Section 1 2      Run a 1000 mmol kg standard in Auto Repeat Mode  then  check the standard deviation shown on the display    If less than 2 0  assess your sample loading technique for  possible loading errors  See Section 3 2     If standard deviation is greater than 2 0  check  for loose TC head screws  See Section 5 2  If this is not  the problem  check the TC mount for gross contamination     If these steps fail to resolve the problem  contact  your dealer or Wescor for assistance     Check the TC connector for incomplete contact  Check for  loose TC head screws     Remove the TC head and inspect the TC mount under a  microscope for bent  damaged  or broken thermocouple  If  thermocouple is undamaged  check for gross or visible cont   amination  If contamination is evident  clean the thermocou   ple according to instructions in Section 5 2  If these mea   sures fail to work  contact your dealer or Wescor for assis   tance  Section 1 2         Problem   Calibration deteriorates  after cleaning the TC  mount     Abnormal display  or sample  reading on an empty chamber   or one of following Error  messages            ERROR  IT ERROR  CLE ERROR    SECTION 5  SOLVING PROBLEMS    5 1 Troubleshooting       Solut
26. d  solution from ampules to other containers     e Always follow the instructions in Section 3 5 to calibrate your  Vapro osmometer  and always verify calibration prior to running  any unknowns     See Appendix E for more information     25    SECTION 2  GETTING STARTED    2 6 Using Osmocoll II Serum Control    Osmocoll   II is    processed  stabilized bovine serum control  which  is useful in a quality assurance program for the Vapro osmometer     For optimum results  always adhere to the following guidelines     Upon arrival in your laboratory  refrigerate Osmocoll Il   Under refrigeration  the serum will remain stable until the  labeled expiration date     Once opened  the product has a maximum stable life of 5 days   if kept refrigerated and tightly capped     Instructions    CAUTION     Never calibrate the Vapro osmometer using Osmocoll Il control solution        1 Run a sample of Osmocoll II control     2 If the measured osmolality falls outside of the range listed on  the package labeling  each Osmocoll lot is assayed for osmo   lality   you should suspect the calibration of the instrument   Recalibrate the instrument using freshly opened Optimol  ampule standards  See Section 3 5 for instructions     26    SECTION 2  GETTING STARTED    2 7 Changing Voltage Selector       The voltage selector is set at the factory  If the volt   age shown does not match your power outlet  see  Appendix A for nominal voltage ranges   you must  change the voltage selector before plugging
27. ding Samples                                     34  3 3 Cleaning the Sample                                         39  3 4 Running the Clean                                        40  3 5  Calibrating    eie ease aa RR 41  3 6 Standby or Waiting Periods                            43  3 7 Serial Data                                                44    TABLE      CONTENTS    SECTION    4  PREVENTIVE MAINTENANCE             4 1 Preventive Maintenance Overview                      45  4 2 Removing the TC                                        46  4 3 Cleaning the TC                                         48  4 4 Reinstalling the TC Head                              51  4 5 Equilibration After Cleaning                            53  4 6 Severe or Stubborn                                              54  SECTION   5 1 Troubleshooting            RR x                        57  5 2 Common      Head                                          65  5 3 External Factors Affecting Precision                     68  5 4 Deformed or Broken Thermocouple                     69  APPENDIX A   Instrument Specifications                                  71  APPENDIX B   Accessories  Supplies  and Replacement Parts               73  APPENDIX      Theory of                                                        75  APPENDIX D   Special Application                                            81                                  5    APPENDIX E  Osmolality Standards                                     9
28. e each osmometer use session  This  will allow you to monitor the condition of the thermo   couple sensor and the rate at which contamination  builds up in the sample chamber  See Section 3 4  for instructions     SECTION 3  OPERATING THE VAPRO OSMOMETER    3 1 Operation Overview    Delete Result   Deletes the last result from the data group  You can delete multiple  results using this function  This function can only be used while  operating in Average Mode     Returning to Mode Menu  Press ENTER while the arrow points at MODE MENU to exit the  Function Menu and return to the Mode Menu     NOTE    Calibration is a critical element of instrument accuracy  While it is not nec   essary to calibrate the osmometer while familiarizing yourself with it  you  should check calibration before you assay sample material  see instruc   tions in Section 3 5         33    SECTION 3  OPERATING THE VAPRO OSMOMETER    3 2 Loading Samples    When first using the Vapro system  practice the loading procedure using  the micropipettor and the 290 mmol kg standard  Record the value dis   played at the end of the cycle  when the  In Process  display goes out  and the chime sounds  Practice this procedure until you can obtain  sequential results with a spread of less than 6 mmol kg  Consistent tim   ing during loading is important for optimum repeatability  This will come  naturally after a few samples     NOTE    While you practice  do not be concerned if the instrument readings do not  agree with t
29. e of  the Wescor micropipettor     22    SECTION 2  GETTING STARTED    2 4 Micropipettor Information    Sample Volume Considerations    The Vapro osmometer does not demand a high degree of volumet   ric accuracy at the 10 yL sample level  Sample volume variations of    10 percent will not noticeably affect the final result  Gross volu   metric errors  such as might arise from incorrect pipetting technique  or poorly maintained micropipettors  or micropipettors not approved  by Wescor  can cause significant measurement errors     23    SECTION 2                 STARTED    2 5 Using Optimol Osmolality Standards    Wescor s Optimol   ampule osmolality standards are  accurate enough to satisfy the most stringent quality  assurance requirements  Calibration integrity is  assured because ampules provide fresh solution for  each use  Having the accuracy of reference stan   dards  Optimol standards are ideal for routine  osmometer calibration  Optimol standards are manu   factured under strict quality control and have a minimum  storage life of 36 months  Refer to Appendices B and E  for more information     NOTE    Ampule standards are intended for one time use for no  more than a few hours  When you have finished calibra   tion  discard any remaining solution        Instructions   Each ampule contains 0 4 mL of solution  This vol   ume is sufficient to prevent measurable evaporative  concentration for a few hours after the ampule is  opened           1 Flip the stem of the ampule w
30. e per   formed without the necessity for a change in the physical state of  the specimen  It is thus a passive technique of measurement that is  free from measurement artifacts that often occur when the speci   men to be tested must be altered physically  This fundamental dif   ference in methodology gives rise to the many advantages of the  vapor pressure osmometer over the older method     In the Vapro vapor pressure osmometer  a 10 microliter sample of  the solution to be tested is pipetted onto a small  solute free paper  disc which is then inserted into a sample chamber and sealed  A  thermocouple hygrometer is incorporated integrally within the cham   ber  This sensitive temperature sensor operates on the basis of a  unique thermal energy balancing principle to measure the dew point  temperature depression within the chamber  This parameter  in itself  a colligative property of the solution  is an explicit function of solu   tion vapor pressure     PROGRAM STEP 1  EQUILIBRATION AND ZERO SET    The sample is introduced into the chamber and the chamber is  closed  Simultaneously          Process         a countdown by seconds  is displayed   This remains until the end of sequence at Program  Step 4      At this point  there will generally be some difference between the  temperature of the specimen and the temperature of the sample  chamber  Temperature equilibrium occurs within a few seconds  The  vapor pressure may also reach equilibrium during this interval  The  microv
31. e the contami   nant simply by applying a droplet of purified  water to the thermocouple and allowing it to  stand for 30 to 60 minutes     Causes of Unusual Contamination    While there are many potential causes of unusual  contamination  the following are the most common        A severely contaminated thermocouple with visi   ble accumulations of organic matter or salt  deposits is evidence of incorrect or careless  loading procedures        Careless loading of greasy      waxy specimens     Q Failure to clean fingerprints or other deposits from  the sample holder           Oily residue from compressed air lines when an  air jet is used to blow water droplets from the  thermocouple after cleaning        Improper use of Blow Clean  Liquid discharged    from the can onto the thermocouple mount leaves  an oily deposit that is difficult to remove     54    SECTION 4  PREVENTIVE MAINTENANCE    4 6 Severe or Stubborn Contamination    Removing Difficult Contamination   Many contaminants can be detected and removed  under microscopic examination  If cleaning fails to  produce an acceptable clean test  examine the ther   mocouple head under a microscope at 30X to 60X  power     Gross contamination can usually be removed by  repeated cleanings  although mechanical scrubbing   as described below  may expedite the process     NOTE    Concentrated ammonium hydroxide  from local stores   can be used to remove stubborn contaminants  but it  may not be effective on oily  greasy  or waxy ma
32. ea   To Start Cucle surement can be repeated without opening the    chamber and to avoid vapor loss  See Appendix D  for more information        Process Delay Mode    31    SECTION 3    OPERATING THE VAPRO OSMOMETER    3 1 Operation Overview    32        zmolalitu    100                    Calibrate  Clear Test   Delete Result   Mode Menu    Function Menu       jj nnolz ks    Contamination            Z       FUNCTION MENU    To display this menu  press SELECT after a sample  has been assayed  and before opening the sample  chamber  The osmometer performs the function indi   cated by the arrow when you press ENTER     Calibrate Function   Use this function to calibrate the instrument using  the 290  1000  and 100 mmol kg calibration stan   dard  Always begin with the 290 set point  then fol   low with the 1000 and 100 mmol kg  See Section 3 5  for details     Clean Test   The Clean Test consists of two consecutive sample  assays on a 100 mmol kg standard solution  The dif   ference between the first and second assay indi   cates the degree of contamination in the sample  chamber     Run this test if you notice significant changes in the  100 mmol kg calibration level     Always use the Clean Test to check thermocouple  cleanliness before assaying samples which require  good linearity and accuracy in the low range  After  thermocouple cleaning  use the Clean Test to verify  the effectiveness of the cleaning     We recommend that you perform this test on a rou   tine basis befor
33. en  each screw progressively with the 9 64 inch hex  driver  until all four are firmly tightened        4 Reinstall the      head connector        51    SECTION 4  PREVENTIVE MAINTENANCE    4 4 Reinstalling the TC Head       5 Replace the access cover     6 Close the sample chamber     7 Turn on the power  Allow the instrument to complete the initial   ization sequence and reach equilibrium  see Section 4 5      8 Perform the Clean Test  If the test indicates a clean thermo   couple head  you should calibrate the osmometer before pro   ceeding to run samples  If the Clean Test reveals contamination   refer to Section 4 6 and  if necessary  Section 5 1     52    SECTION 4  PREVENTIVE MAINTENANCE    45 Equilibration After Cleaning       Cleaning the thermocouple mount changes the thermal equilibrium  of the instrument and causes a temporary shift in calibration after  the TC head is reinstalled  After reinstalling the thermocouple head   allow the instrument to regain thermal equilibrium     The Temperature Drift indicator will be near center when the  osmometer temperature is stable  See Section 2 3     53    SECTION 4  PREVENTIVE MAINTENANCE    4 6 Severe or Stubborn Contamination       If the Clean Test indicates residual contamination in  spite of a clean appearance     1 Repeat the cleaning procedure and run a sec   ond Clean Test  If there is significant improve   ment  contamination can likely be removed by        repeated cleanings     2 You can often successfully remov
34. ent are initially des   ignated at the factory and stored in non volitile mem   ory  They become default settings on power up of   the osmometer and are displayed briefly at that time     To change these settings     1 Turn off power  Wait aproximetely 10 seconds     2 While pressing both the SELECT and ENTER  keys  turn the on power  Wait several seconds  while the display shows the Wescor logo    default selections  and finally the Setup Menu     Select desired language or unit of measure    3 ment using the SELECT button to move the point   er to your choice  Press ENTER to add selection  to memory     4 Once you have selected your preferences you can  exit the setup menu  The instrument defaults to  these settings until they are changed     95        Accuracy   of standards 24  25    of reported osmolality 25  Ambient Temperature 10  43  60  82  Ammonium Hydroxide 48  55  56  Ampule   organizer 18  24   standards  see Optimol   ASCII Characters 94  Average Mode 30  42  Auto Repeat Mode 31  59  62  68    B  Blow Clean 18  50  74    C  Calibrate Function 32  Calibration  accuracy 25  51  63  adjusted to mean value 30  effects of temperature changes on 21  error 59  64  for maximum accuracy 42  for very large samples 86  for very small samples 84  in Average Mode 30  acceptable limits 42  procedure 41  routine 42  52  shift 38  53  58  60  61  62  standards 24  42  58  61  64  68  verifying 25 32  33  59  71    Carry over Error 88  Chime  signal 14 37 77  warning 20  38  Cla
35. er by touching the tip  of the water dropper  or the water drop  to the water stand   ing on the mount         Dilute any remaining droplets of cleaning solu   tion with pure water     8 Repeat Steps 4  5  and 6        49    SECTION 4  PREVENTIVE MAINTENANCE    4 3 Cleaning the TC Head       Repeat this procedure at least ten times  using  enough water to cover the central depression  and thermocouple        ESS CAUTION     Shaking or tipping the can of Blow Clean will severely  contaminate the thermocouple  The can must remain  flat on the bench     10 Place the Blow Clean upright and level on the   bench  Clear the nozzle with a short burst of  gas  Hold the TC mount about 2 inches from  the nozzle  then aim the nozzle directly at the  thermocouple and release a very short burst   no more than 1 second  to blow away any  remaining droplets        1 1 Inspect the TC mount for any residual conta   mination  If foreign material cannot be  removed using this procedure  refer to  Section 4 6     NOTE    Some contamination is invisible  even under the micro   scope  Inspection can reveal many types of contamina   tion  but cannot replace the Clean Test        50    SECTION 4  PREVENTIVE MAINTENANCE    44 Reinstalling the TC Head       Verify that the sample chamber lever is in the  open position     2                                        CAUTION   The instrument will not hold calibration if the chamber  screws are loose              Start each screw into the threads  then tight
36. f the osmometer reads within   3 mmol kg of  the standard  287 to 293   it is within accept   able calibration limits  In that case  skip to Step  4  If calibration is needed  proceed to Step 2     With the chamber still closed  press SELECT to  reveal the Function Menu  The selector arrow  should be pointing at Calibrate     Press ENTER  The instrument calibrates to the  standard             4 Repeat this sequence using the 1000 mmol kg   and 100 mmol kg standards to establish base    line calibration for these standards  If the read    ing is not within   3 of the standard value  per   form steps 2 and 3     Calibration Method for Maximum Calibration  Accuracy    When you need maximum accuracy  run the following  calibration sequence     1 Select Average from the Mode Menu   Run three consecutive assays using 290  mmol kg standard   3 Select CALIBRATE and press ENTER   The instrument calibrates on the average of the    three samples  This method can also be used for  100 and 1000 mmol kg standards        Chamber Emrtu    READY  gt     E          B           Tenrerature Drift       SECTION 3  OPERATING THE VAPRO OSMOMETER    3 6 Standby or Waiting Periods    When the instrument is not in use  leave the cleaned  sample holder empty and locked in the measure   ment position  If the chamber is left open for longer  than 2 minutes  a warning chime sounds     When in the standby mode  the Vapro osmometer is  not idle  It continuously monitors its internal operat   ing temperature
37. he specified concentration of the solution  When you feel comfort   able with the procedure and are able to obtain repeatable results  calibrate the  instrument using the instructions in Section 3 5  then run the Clean Test   Section 3 4         Sample Volume   The optimum sample volume  10 microliters  should fully saturate one  of the SS 033 sample discs  The osmometer accommodates varia   tions in sample volume as great as   10 percent  9 microliters to 11  microliters  without noticeable variation in indicated osmolality     CAUTION     Samples greater than 11 microliters can contaminate the thermocouple        34    SECTION 3  OPERATING THE VAPRO OSMOMETER    3 2 Loading Samples          Sample Loading Procedure   1 Rotate the sample chamber lever upward and pull  the sample slide out from the instrument until it  comes to a stop  bringing the sample holder direct   ly under the pipettor guide     2 Use the forceps supplied with the instrument to  place a single sample disc in the central depres    sion of the sample holder  Make sure you have  picked up only a single disc  If necessary  use the  forceps and a teasing needle to separate discs  If  two discs stick together  the reading will be slightly  elevated  Reject imperfect discs or any that do not  lie flat        3 With a clean tip installed  aspirate a sample into  the micropipettor by depressing the plunger to the    Place Sample Disc in           Sample Holder stop  immersing the tip  and gently releasing the  
38. he time required to achieve temperature stability  depends on the initial instrument temperature   typically  10 to 30 minutes  but may be longer if inital temperature  varies more than 5 degrees from room temperature        NOTE    Under normal circumstances  leave power on to keep the  instrument in a ready state and to maintain stability   See  Section 3 6        21    SECTION 2                 STARTED    2 4 Micropipettor Information    The micropipettor furnished with the Vapro osmometer uses a two   step  aspirate expel  mechanism that dispenses 10 microliters of liq   uid for osmolality assay  This no maintenance micropipettor works  with a wide range of biological solutions and laboratory reagents   Disposable plastic tips eliminate carry over error from sample to  sample  Use the provided micropipettor to assure uniform results  among different operators     We do not recommend three step pipetting  aspirate expel blowout   for loading the osmometer  The blowout step tends to create bub   bles in the specimen that can lead to thermocouple contamination     Positive Displacement Pipettors   A positive displacement pipetting device or alternative loading meth   ods may be more suitable for extremely viscous fluids or complex  specimens     CAUTION    Do not use positive displacement pipetting devices for routine operation   Refer to Appendix D for additional information regarding special applica   tions        The sample loading procedure in this manual presumes the us
39. hen the initialization cycle  is complete  the display screen will appear as shown  on the left     NOTE    Even though the screen at this point indicates the instru   ment is ready  calibration will not be stable until the  instrument reaches thermal equilibration  see below               Temperature Drift Scale   Osmolality determination involves the measurement  of extremely small temperature differentials  The  osmometer is thus sensitive to ambient temperature  changes which induce internal temperature changes     20    SECTION 2  GETTING STARTED    2 3 Setting Up the Osmometer    While the instrument compensates for small chang   es that occur over time  moving the instrument to a  different area or exposing it to too much air circula   tion will shift the osmometer   s reading and calibra   Chamber Empet  tion points  The Temperature Drift scale allows you    R E    D Y s to determine when internal temperature has stabi   lized              m     mmu        Tenrerature Drift  The Temperature Drift scale appears on the screen  whenever the instrument is in the    standby    mode   and it has completed two cycles  The instrument is  considered stable and ready to operate unless the  indicator is against the   or   marks on the scale   indicating a changing internal temperature that can  affect instrument calibration  See Note below        NOTE    It is normal for the osmometer to undergo a significant  temperature drift during the first few minutes of equilibra   tion  T
40. holder  Push the sample hold   er into the chamber and close the chamber     Solid  or some viscous  samples require extended periods to reach  equilibration inside the chamber  On such samples you may want to  make repeated measurements without opening the chamber to  determine the time required to achieve equilibrium  Osmolality val   ues will trend downward until they stabilize  If you know the required  time  simply defer the measurement for that period     86    APPENDIX  D    Special Application Notes    5 Press ENTER to make a measurement  Osmolality is dis   played when the measurement is complete     6 For repeat measurements  leave the chamber closed and  press ENTER  Readings should be lower with each successive  assay until equilibrium is reached     87    APPENDIX  D    Special Application Notes    SAMPLING VISCOUS AND OR  NONHOMOGENEOUS SPECIMENS    The broad range of specimen materials amenable to testing in the  vapor pressure osmometer may require you to adapt your sampling  technique to suit the physical characteristics of unusual samples   Using the micropipettor will assure the application of uniform vol   umes of both test specimen and calibrating solutions  but if the vis   cosity of the sample is extremely high  a positive displacement  micropipettor may be preferable for sampling  These devices are  not recommended for routine use  however  due to their propensity  toward carry over error     If the sample material does not readily saturate the paper s
41. ion  Check for deformed or broken thermocouple     Reset instrument by turning the power off for at least 15  seconds  then turn power back on     Be sure the sample holder is thoroughly clean and dry  before closing the sample chamber  see Section 3 6      If this fails to resolve the problem  contact your dealer or  Wescor for assistance     63    SECTION 5  SOLVING PROBLEMS    5 1 Troubleshooting    64    Problem   The following error  message appears on  the screen     Calibration  ERR  E    Value not  within range       Solution  Be sure that you are using the correct calibration  standard     Check the TC head connector for faulty connection   Check the thermocouple for gross contamination   Check for loose TC head screws  If these steps fail to    solve the problem  contact your dealer or Wescor for  assistance     SECTION 5  SOLVING PROBLEMS    5 2 Common TC Head Problems    Many years of field experience have shown that the majority of  problems encountered with the osmometer are with the thermocou   ple sensor  It is suspended from the thermocouple mount  which  forms the upper half of the sample chamber  The thermocouple  mount is part of the thermocouple head assembly  referred to simply  as the    TC head        Common thermocouple problems affect instrument performance in  distinctive ways  providing significant clues that will be evident in the  behavior of the instrument  These are summarized below in order of  most likely occurrence    COMMON TC HEAD PROBLEMS
42. ion and instructions in the instruction manual     SECTION 1  INTRODUCTION    1 2 Customer Service    Wescor is ready to help resolve any difficulty with the operation or per   formance of your Vapro osmometer  If you cannot solve a problem  using the procedures in this manual  please contact us     Customers within the United States are encouraged to contact us  by telephone  Outside the U S   many of our authorized dealers  offer complete customer service and support  Contact Wescor by  mail  telephone  or fax at the address and numbers listed below     WESCOR  INC  459 South Main Street  Logan  Utah 84321    USA   TELEPHONE   435 752 6011 Extension 0   Operator  171   Orders  172   Service   TOLL FREE  U S  and Canada    800 453 2725 Extension 0   Operator  171   Orders  172   Service   FAX   435 752 4127   E MAIL    service  wescor com   WEB SITE    WWW Wescor com    SECTION 1  INTRODUCTION    1 3 Vapro System Description    The Vapro osmometer is an advanced electronic adaptation of the  hygrometric method of vapor pressure determination  The sensitive  thermocouple and sophisticated electronics provide the means to  measure the dew point temperature depression of a specimen with  resolution to 0 00031   C     Vapor pressure and freezing point are among the colligative proper   ties of a solution  Compared with pure solvent  these properties are  altered in proportion to the number of solute particles dissolved in  each kilogram of solvent  water in the case of biologica
43. is  can cause unwanted calibration shifts  See Section 3 6           9 Remove the specimen from the sample chamber  immediately after a measurement using the  instructions in Section 3 3  If the sample is left in  the chamber for longer than 4 minutes  a warn   ing chime will sound     CAUTION    You can severely contaminate the chamber  or the  thermocouple  in a single loading if you improperly  load the sample or if you fail to thoroughly clean the  sample holder  Severe contamination can make it  impossible to calibrate the osmometer        38    SECTION 3  OPERATING THE VAPRO OSMOMETER    3 8 Cleaning the Sample Holder    To clean the sample holder and prepare for another  sample     Smoothly rotate the chamber lever to the open  position  then withdraw the sample slide     Using a lint free tissue  not facial tissue  or a  cotton swab  carefully remove the wet disc and  any traces of residual liquid from the sample  holder        CAUTION        Remove Sample and  Clean Sample Holder Never use metal forceps to remove wet discs  this can    damage the surface of the sample holder     3 Leave no visible residue on the holder surface   If needed  use a tissue or swab moistened with  deionized water  Always clean the sample hold   er with a fresh tissue or swab to avoid contami   nation  Avoid touching the sample holder with  bare fingers     The sample holder should appear bright  shiny  and  perfectly dry before loading the next specimen     NOTE     Regularly perform the 
44. ith the required equilibration time for a partic   ular type of sample  you can simply leave the chamber closed for  the required time and then press ENTER to begin the osmolality  measurement     The standard sample holder has a diameter of 7 mm and a depth of  1 25 mm  Two optional sample holders are available from Wescor    for measuring samples which are too large for the standard sample  holder     e AC 064 sample holder is 7 mm dia x 2 5 mm deep     e AC 065 sample holder is 9 5 mm dia x 4 5 mm deep     85    APPENDIX  D    Special Application Notes    Instructions    For best precision  use the smallest holder that can accommo   date the sample volume without danger of contaminating the  thermocouple     CAUTION    Never load any sample that extends above the lip of the sample holder   Solid sample material extending above the lip of the sample holder can  severely contaminate or even break the thermocouple        2 Calibrate the instrument  Use the same size sample holder as  will be used for the assayed sample  Match the volume and  shape of the subject sample and the calibration solution as  closely as possible  Several filter paper discs saturated with  standard solution should be used for calibration to dampen the  motion of the solution and to approximate the size and shape  of the sample material     3 Select Process Delay Mode This allows you to delay the mea   surement cycle after closing the chamber until you press  ENTER     4 Place the sample in the sample 
45. ith your finger  or  tap the ampule lightly against a hard surface to  dislodge any solution held by capillary action in   the stem of the ampule     2 Place the ampule in the breaker position of the  Ampule Organizer  Hold the organizer firmly  down against the work surface     3 Slide the provided flexible protection sleeve  over the stem of the ampule     4 Grasp the sleeved stem firmly and snap the  neck of the ampule     24    SECTION 2                 STARTED    2 5 Using Optimol Osmolality Standards    Sample directly from the ampule  using a fresh micropipettor tip  each time to avoid contamination of the solution     Discard any solution that remains after finishing your calibration  procedures     Optimol standards are packaged in 60 ampule cartons designed for  convenient shipment and storage  See Appendix B for ordering  information     Assuring Accurate Measurements    The accuracy of reported osmolality is directly linked to the accuracy  of the calibrating standard solutions  While these solutions have  exact specified osmolality at the time they are opened  osmolality  inevitably increases as water evaporates     Always adhere to the following guidelines when using Optimol  ampule standards     e Since the specified value of osmolality is certain only at the time  the ampule is opened  do not rely on any opened ampule if you  are uncertain of how long its contents have been exposed to  evaporation     e Sample directly from the ampule   do not transfer standar
46. ity assay by older macro methods   especially since other analytical parameters are very often simultane   ously requested on such specimens     An equally important advantage is that the vapor pressure osmometer  does not physically change the sample  Where biological specimens or  medications are multiphasic  or highly viscous  vapor pressure osmome   try becomes the only reliable method of measurement  For example   feces  sputum  and gastrointestinal aspirate specimens usually contain  variable amounts of mucous material that interferes with or prevents  freezing point depression measurements but does not affect vapor pres   sure osmometry  Neither does the presence of finely suspended insolu   ble material  a feature of radio opaque media  which are often examined  to detect grossly high osmolality values likely to produce rapid dehydra   tion when given to small infants     In general research  the potential applications are too numerous 10 list   However  the vapor pressure osmometer is of value to a wide range of  biologists and microbiologists concerned with fluid and electrolyte bal   ance in all forms of life  especially where specimens are necessarily very  limited in size  and may exhibit unusual viscosity     The instrument is capable of vapor pressure determinations   expressed as osmolality  even on complex specimens such as tis   sue sections  Such specimens should be cut to approximately the  diameter and thickness of the paper sample disc  if possible     F
47. l Application Notes    SPECIAL LOW VOLUME PROCEDURE     Preparing Paper Discs   1 Use a  Mieth or equivalent  1 8 inch diameter punch to create a  supply of paper discs  Punch only one thickness of paper stock  at a time  to prevent paper discs from sticking together  That   along with static electricity will make it difficult to pick up a sin   gle disc with the tweezers     2 After punching  remove paper discs from the retainer of the  punch  Store discs in a clean  static free container     As stated before  very low volume tests require careful and consis   tent technique to achieve reliable results  The following are impor   tant for you to consider when running samples with very low vol   umes     e Use only single sample discs  Because of their small size  you  must be careful not to load more than one     e Discs must be punched cleanly no ragged edges   e The sample holder must be kept very clean     e Do not exceed 4 uL of sample in the special sample holder   Using too much sample fluid can severely contaminate the  thermocouple     e The paper disc must be completely saturated by sample fluid  If  not fully saturated the disc may appear patchy  In this condition   data will be inconsistent and repeatability will be poor     NOTE    Very small samples of less than 2 microliters can be successfully mea   sured using lighter paper for the discs  You should experiment with various  papers  Be cautious that some papers contain electrolytes that make them  unsuitable  S
48. l Chime   Sounds a short chime at the conclusion of each measurement cycle  and a long chime when certain fault conditions exist  Turn the shut   ter wheel to adjust volume     Serial Data Port    For asynchronous serial communication with a printer or computer   The serial port uses a DB9 connector at RS 232 voltage levels   See Appendix F for more information     15    SECTION 1  INTRODUCTION    1 5 Controls and Features    16    Power Entry Module    Fuse Door    Power Switch   1  on  0  off       Voltage Selector    INSTRUMENT REAR PANEL  Continued     POWER ENTRY MODULE  Accepts standard IEC 320 type power cord     Fuse Door  Provides access for fuse replacement  See Section  2 8 for instructions     Power Switch  Switches power on  1  or off  0      Voltage Selector    This selector is set at the factory  If necessary  you can  set the selector to match your local power source  see  Section 2 7   Fuses must match the voltage selection   To change fuses  see Section 2 8 and Appendix A     Using this equipment in a manner not specified by  Wescor may impair the provided safety protection     SECTION 2                 STARTED    2 1 Instrument Setup Sequence    We recommend that you follow this sequence if you are using this  instrument for the first time  Details about these operations are  found in the following sections      i      Inspect accessories and supplies     2  Place the instrument on a suitable work surface in an area free  from drafts     3  Plug in the powe
49. l solutions    Thus  measuring either property is an indirect means of determining  solution concentration or osmolality     The chief advantage of the vapor pressure method is that it does  not require alteration of the physical state of the specimen   Concomitant benefits include     e 10 microliter sample size     e Routine operation on micro samples of any biological solution   including whole blood  serum  plasma  urine  and sweat  as well  as complex specimens such as tissue samples     e None of the measurement artifacts that arise in freezing point  depression measurements due to elevated viscosity  particulate  matter  inhomogeneities  or other physical characteristics of the  sample     e Superior reliability because the measurement involves minimal  mechanical complexity     NOTE   Vapro displays in Standard International  SI  units  mmol kg  See Appendix E        SECTION 1  INTRODUCTION    1 4 How the Vapro Works    A 10 microliter specimen is aspirated into a micropipettor tip  The  specimen is then inoculated into a solute free paper disc in the  sample holder  whereupon the sample holder is pushed into the  instrument and the sample chamber is locked  Locking initiates the  automatic measurement sequence     The sensing element is a fine wire thermocouple hygrometer  This  is suspended in a unique  all metal mount  which when joined with  the sample holder  forms a small chamber enclosing the specimen     As vapor pressure equilibrates in the chamber airspace  
50. lality   AC 062 Sample Holder  7 mm dia  x 1 25 mm deep  supplied with  instrument   AC 063 Sample Holder  4 25 mm dia  x 1 2 mm deep   low sample  volume     SPECIAL PURPOSE SAMPLE HOLDERS  gross samples     AC 064 Sample Holder  7 mm dia  x 2 5 mm deep  AC 065 Sample Holder  9 5 mm dia  x 4 5 mm deep  AC 078 Sample Holder for Kwikdisk     models 5520  5500  5100C    OSMOMETRY STANDARDS CONTROLS  OPTIMOL AMPULE STANDARDS  0 4 mL vial  package of 60     OA 010 Optimol Osmolality Standard Solution  100 mmol kg   OA 029 Optimol Osmolality Standard Solution  290 mmol kg   OA 100 Optimol Osmolality Standard Solution  1000 mmol kg  OSMOLALITY CONTROLS   55 025 Osmocoll I  Standard Control  1 mL vial  package of 6     73    APPENDIX    B  Accessories  Supplies  and Replacement Parts  SUPPLIES  SS 003 Cleaning Solution  for manual cleaning  2 oz dropper bottle   SS 006 Deionized Water  2 oz dropper bottle   SS 026 Blow Clean  SS 028 Kwikdisk     aluminum paper disks  package of 200  requires AC 078   SS 033 Sample Discs  vial of 5000   SS 036 Micropipettor Disposable Tips for AC 037  package of 1000   REPLACEMENT PARTS  CIRCUIT MODULES and ASSEMBLIES  RP 170 Front Panel Keyboard Assembly  310300 Display Assembly  310354 Chamber Switch Assembly  330915 Power Supply Board Assembly  330915X Power Supply Board Assembly  with exchange   340454 Main Board Assembly  340454X Main Board Assembly  with exchange   310346 Head Cable Assembly  FACTORY SERVICE  FS 255 Thermocouple Clean and Check
51. le         AMBIENT TEMPERATURE            AT DEW POINT TEMPERATURELI  DEPRESSION    WET BULB TEMPERATURE    DEW POINT TEMPERATURE    4        PRE CONVERGENCE TEMPERATURE    Typical curve for closed chamber    ie  0  Um  DE   6        a    E            H   n      fa  1   2 i3  4  EQUILIBRATION    COOLING   CONVERGENCE READOUT  SAMPLE HOLD     i   i HJ  gt             n n     TIME  SECONDS     0 n    56 69    CHAMBER CLOSED    APPENDIX        Theory of Operation    THERMOCOUPLE TEMPERATURE VERSUS OSMOLALITY    The graph on the left is a plot of thermocouple temperature versus  time as the instrument cycles through the program  beginning with  chamber closure  time   0   The graph depicts the excursion of  thermocouple temperature that typically occurs during each of the  program steps outlined above  T  is the ambient temperature in the  chamber  Tp is the dew point temperature  and AT is the dew point  temperature depression  The output is proportional to AT     Assuming that the chamber remains closed while the osmometer  displays the final reading at Step 4  the thermocouple temperature  returns to T  after holding at the wet bulb depression temperature  until all of the water has evaporated from the thermocouple  If the  chamber is opened  the water will evaporate almost instantly and  the thermocouple temperature will quickly return to ambient     The relationship between sample osmolality and the reading  obtained by the osmometer is governed by fundamental considera   ti
52. lligative property of a solution  but  unlike the other three  it is not a cardinal property of the solvent   Solution osmotic pressure can be measured directly using a semi   permeable membrane apparatus  but only with respect to those  solute particles that are impermeable  since smaller solute particles  freely transude the membrane and do not directly contribute to  osmotic pressure  Such a measurement is referred to as  colloid  osmotic pressure  or  oncotic pressure   It is expressed in terms of  pressure  in mmHg or kPa  Total osmotic pressure  i e   that which  can be calculated on the basis of total solute concentration  is a the   oretical concept only     The measurement of total solution concentration  or osmolality  can  only be made indirectly by comparing one of the solution colligative  properties with the corresponding cardinal property of the pure sol   vent  The first practical laboratory instruments developed for routine  measurement of osmolality were based upon depression of the  freezing point and  until recent years  all osmometers for large scale  testing were based upon this methodology     75    APPENDIX  C    Theory of Operation    The Vapro osmometer embodies newer technology  It is based upon  a measurement of vapor pressure depression made possible by  thermocouple hygrometry  The vapor pressure method enjoys a sig   nificant intrinsic advantage over the measurement of either freezing  point depression or boiling point elevation in that it can b
53. mometer on a suitable work sur   face     NOTE    Avoid locations where instrument precision will be  altered by thermal gradients or rapid temperature  changes caused by heavy foot traffic  air vents  blow   ers  heaters  or windows        2 Connect the power cord to an electrical outlet  that matches the voltage selected on the rear  panel  Avoid power circuits that are shared by  centrifuges  air conditioners  or other power  equipment  We recommend that you use a  power line surge protector to isolate the  osmometer from spikes and surges              3 Verify that the sample holder is in the measure   ment position  sample slide is pushed com   pletely into the instrument until it stops         Verify that the sample chamber lever is in the  closed position     TURN PAGE TO CONTINUE INSTRUCTIONS  19    SECTION 2  GETTING STARTED       2 3 Setting Up the Osmometer    Turn the osmometer on  I   The POWER indica   tor on the front panel shows green when power  is on     The display screen will briefly show the Wescor  logo  language and unit selection  and the resident  software version  This will be followed by the word   initialization  with a countdown timer     To change the diplayed language or units of mea   sure  see Appendix G  Setup Menu     Chamber Empty NOTE     If you open the chamber lever before the end of the ini     R F A        gt  tialization cycle  the warning chime sounds     The initialization cycle establishes the reference  point for the instrument  W
54. nt varies with the solute con   centration  the relation between osmolality and concentration of  solute is not linear  For this reason  measurements of osmolality  made on laboratory diluted specimens  with subsequent multiplica   tion by the dilution factor to calculate the original solution osmolality   will not give valid results     91    APPENDIX  E    Osmolality Standards    With complex solutions  such as biological fluids  analytical vari   ables are universally expressed as the concentration of specific ions  and of undissociated solute particles  It follows that a molal solution  of NaCl can be analytically expressed as a combination of a molal  solution of sodium ions and a molal solution of chloride ions  The  total concentration of solute particles  the osmolality  is therefore  2000 millimolal  Osmolality can best be expressed simply as 2000  mmol kg without the necessity of introducing the  osmole  concept     The commission on Clinical Chemistry of the International Union of  Pure and Applied Chemistry  IUPAC  and the International  Federation of Clinical Chemistry  IFCC  have recommended that the  unit of osmolality be mmol kg  and this has been adopted by the  American Journal Clinical Chemistry as part of its general accep   tance of Standard International units  Wescor led the industry as the  first osmometer manufacturer to adopt Standard International  SI   units for osmolality     QUALITY ASSURANCE   Wescor calibration solutions are manufactured using 
55. oltmeter reads the amplifier voltage to establish the reference  for the measurement     76    APPENDIX        Theory of Operation    PROGRAM STEP 2  COOLING    An electrical current is passed through the thermocouple  cooling it  by means of the Peltier Effect to a temperature below the dew point   Water condenses from the air in the chamber to form microscopic  droplets upon the surface of the thermocouple     PROGRAM STEP 3  DEW POINT CONVERGENCE    Electronic circuitry    pumps    thermal energy from the thermocouple  via Peltier cooling in such a way as to cancel out heat influx to the  thermocouple by conduction  convection  and radiation  Given this   the temperature of the thermocouple is controlled exclusively by the  water condensing upon its surface  Thermocouple temperature   depressed below the dew point in Step 2  rises asymptotically  toward the dew point as water continues to condense  When the  temperature of the thermocouple reaches the dew point  condensa   tion ceases  causing the thermocouple temperature to stabilize     PROGRAM STEP 4  END OF SEQUENCE AND  READOUT    The reading on the display is proportional to the vapor pressure of  the solution  When this final reading is reached  a chime sounds  and the    In Process    changes to    Osmolality        The result is displayed in SI units of osmolality     mmol kg     77    APPENDIX    C    Theory of Operation    78    THERMOCOUPLE TEMPERATURE    THERMOCOUPLE TEMPERATURE VS  TIME  with 290 mmol kg samp
56. ons  Vapor pressure depression  a linear function of osmolality  has  been identified as one of the colligative properties of a solution  The  relationship between vapor pressure depression and dew point tem   perature depression is given by    AT   Ae S  where AT is the dew point temperature depression in degrees  Celsius  Ae is the difference between saturation and chamber vapor  pressure and S is the slope of the vapor pressure temperature func   tion at ambient temperature  The Claussius Clapeyron equation  gives S as a function of temperature  T   saturation vapor pressure          and latent heat of vaporization  A    S  Eo  RT   where R is the universal gas constant     The dew point temperature depression  AT  is measured as a volt   age signal from the thermocouple  This voltage is equal to AT mul   tiplied by the thermocouple responsivity which is approximately 62  microvolts per degree Celsius  After voltage amplification by a pre   amplifier  the signal is processed by the microprocessor to provide  calibrate and compensate functions and display the reading     79    APPENDIX  D       Special Application Notes    Clinical and General Research    The Vapro osmometer has unique advantages in many aspects of clini   cal chemistry due to its very small sample requirement  This is particu   larly true in pediatric practice  For example  the amount of sample col   lected for sweat  fecal  sputum  duodenal  and gastric analysis is fre   quently too small to allow osmolal
57. or experimental purposes  large volume sample holders are available   These sample holders will accommodate gross specimens that are not  amenable to testing with the standard shallow sample holder    Contact Wescor for more information     81    APPENDIX  D    Special Application Notes    PROCEDURE FOR VERY SMALL SAMPLES    You can measure samples with very low volumes  under 4 uL   using the following procedures     Sample discs must be hand made from high grade filter paper   Whatman  1 or equivalent  using a high precision 1 8 inch diame   ter paper punch to produce discs with a very clean edge     Required Equipment    e Wescor low volume sample holder  AC 063    e High quality round hole paper punch  1 8  diameter  Mieth or  equivalent    e High quality 2 pL pipette  which will deliver precisely 2 microliters  or less   e Pipette tips  short    e Tweezers   e Teasing needle   e Whatman  1 Filter Paper or equivalent   e Lint free tissue paper   e Cotton tipped applicators    NOTE    Maintain a stable ambient temperature  Heat  cold  air currents and temper   ature fluctuations which vary more than approximately 0 3   C within a 10 to  15 minute time frame  generally will result in poor quality data  You should  monitor the Temp Drift Scale for ambient temperature fluctuations which  will interfere with instrument accuracy        Technique  including timing  is vitally important to obtaining good  data while conducting very low volume tests     82    APPENDIX  D    Specia
58. possible sources of  thermal disturbance  as outlined in Section 2 3     Use AUTO REPEAT to evaluate the precision of the osmometer     Run the instrument with 1000 mmol kg standard in AUTO REPEAT  to determine if instrument can repeat well  If it does  consider the  possibility of loading errors causing poor repeatability     SECTION 5  SOLVING PROBLEMS    5 4 Deformed or Broken Thermocouple       The thermocouple is well protected while the TC  head is in the instrument  Cleaning procedures  detailed in this manual should not harm the thermo   couple  but it can be deformed or broken if contact   ed by any object while it is out of the instrument     e  f the thermocouple is only slightly deformed   the instrument will automatically adjust for the  deformed thermocouple and function normally        e Abadly deformed thermocouple will still function  but will show a noticeable loss of measurement  precision     e Adeformed or flattened thermocouple with its  bead lying close to  or touching  the surface of the  mount will not cool to normal temperature  depression during the measurement cycle   Appendix C   As a result  the instrument may  display a meaningless value        Restoring a Deformed Thermocouple   You can usually salvage even a severely deformed  thermocouple by carefully lifting it into normal posi   tion  Although the thermocouple wires are only 0 025  mm in diameter  they are quite malleable and are  generally amenable to straightening and reshaping     In any
59. r cord and switch power on       Allow for temperature equilibration  observe Temperature Drift  Scale      5  Practice loading samples   6  Perform a Clean Test and clean the thermocouple if necessary   7  Check instrument calibration and recalibrate if necessary     8  Assay samples     50050 oo LIL I      SECTION 2  GETTING STARTED    2 2 Vapro Accessories    The following accessories and supplies are furnished with the Vapro  osmometer     Vapro User s Manual  Micropipettor  Micropipettor Disposable Tips  Forceps  Paper Sample Discs  Optimol   Osmolality Ampule Standards  Osmocoll      Osmolality Serum Control  Ampule Organizer  9 64 inch Hex Driver  Thermocouple Head Cleaning Supplies  consisting of   Cleaning Solution  Deionized Water  Blow Clean      U S  48 states only     In addition to the above  you need a supply of lint free tissue paper  or cotton swabs for cleaning the sample holder between specimens     CAUTION    Never use facial or other soft tissue to clean the sample holder  Such  tissues produce excessive lint residue that will contaminate the thermo   couple sensor          Compressed or liquefied pure gas suitable for blowing dust from delicate  surfaces or precision mechanisms  Available under various trade names     18    SECTION 2                 STARTED       2 3 Setting Up      Osmometer    Carefully unpack the instrument and compare the  contents with the packing list to be certain that  everything needed for operation is at hand     1 Place the os
60. reference data  on the concentrative properties of sodium chloride in water from the  Handbook of Physics and Chemistry  CRC Press  For quality assur   ance  each lot is compared by replicate osmolality measurements to  reference solutions prepared from dried  high purity sodium chloride  obtained from the National Institute of Standards and Technology   NIST      Wescor guarantees the accuracy of its calibration solutions within  the combined overall accuracy of the reference solution formula   tions and the control measurements  100   2 mmol kg  290    3  mmol kg  1 000    5 mmol kg     92    APPENDIX  F    Serial Data Output       The 5520 serial port uses a DB9 connector on the instru   ment back panel  This port is for asynchronous serial com   munication with a printer or computer  It uses standard  non return to zero  NRZ  format at RS 232 voltage levels     The 5520 checks to see if the RTS  pin 7  is active  When    a sample is assayed while a device is connected to the  5520 s RS 232 port  the display status line will show     ENTER to Send    To send data to the external device  press ENTER  The  display will show     Osmolalitu    SERIAL OUTPUT TECHNICAL DATA    Output voltage level     Nominal   9 volts  Maximum    15 volts  Minimum   5 volts    Data protocol     1200 bps  1 Start bit  8 Data bits  No parity  1 Stop bit    93    APPENDIX    F  Serial Data Output  Pin Diagram   Pin   Mnemonic Description  f 1 DCD Data Carrier Detect  output   Pint 2 3 4 5 2 RXD Re
61. rom Wescor if needed  Refer to Section 1 2     67    SECTION 5  SOLVING PROBLEMS       5 3 External Factors Affecting Precision    68    Problems with instrument precision have a number of possible  sources  Often  poor reproducibility is caused by external factors  that are entirely independent of the instrument itself  The following  are some of these factors     e Incorrect Use of Calibration Standards   Instrument accuracy  and linearity  depend upon the correct use  of osmolality calibration standards  Refer to Section 2 5  2 6 and  Appendix E for further information     Sampling Error   Sampling error tends to be amplified when dealing with  specimens of 10 microliters or less  You can prevent errors by  using consistent technique and appropriate methods of transfer   ring samples  See Section 3 2 for details     Micropipettor Caused Errors   Unlike the maintenance free micropipettor supplied by Wescor   many micropipetting devices require routine maintenance   Without proper maintenance  micropipetting devices can  exhibit significant volumetric error  in excess of 50     and cause corresponding variations in indicated osmolality   Positive displacement micropipettors are not recommended   as an alternative to the Wescor micropipettor  except when  dealing with samples of very high viscosity     e Poor Precision  Determine whether the problem is with the instrument or is  caused by external factors  such as the micropipettor     Check the location of the osmometer for 
62. s 9  cycle 10  15  20  31  43  errors 21  position 19  range 71  reference point 10  repeatability 71  resolution 71  sequence 10  time 71  86  units 20  101  Menu  Function 29  31  Mode 29  30  42  selection 29    Micropipetting  technique 23  35  36  errors 68  Micropipettor 18  21  positive displacement 21  88  three step 21  two step 21  tip 10  18  21  35  36    Mode s   Auto Repeat 31  Average 30  Menu 29  changing 29  selecting 29  Normal 30  Process Delay 31   Standby  15  21  29  38  43  Multi solvent Solutions  measuring 89    N    Nonhomogeneous Samples 88    O  Optimol Osmolality Standards 18  24  58  59  evaporation 25  opening 24  Osmocoll 1 Osmolality Control Reference  18  26  control value 26  refrigerating 26  Osmolality 75  display 37  58  59  93  on empty chamber 43  range 26  readout 71  standards 18  24  25 58  59  64  68  92  Osmometry  boiling point 76 77  89  freezing point 75  77  89  90  vapor pressure 75  77  89  90  Osmotic pressure 75    p  Packing List 19    P  Paper punch 82  Paper Sample Disc s  10  18  35  36  39  75  for special applications 82  83  84  86  measuring without 88  Pipette  guide 12  13  36  tip 10  35  special purpose 82  Particles  solute 75  Power  connecting 19  60  consumption 71  cord 16  19  indicator 12  13  19  58  60  Entry Module 14  16  switch 16  46  52  58  Surge Protector 19  Preventive Maintenance 45  Process Delay Mode 31  85  Properties  cardinal 75  Protection Sleeve 24    R  Radio opaque media 81  Replacemen
63. s must be interpreted accordingly     89    APPENDIX  D    Special Application Notes    You must be aware of this phenomenon if you are to correctly inter   pret results  By way of illustration  the table below depicts the  results of solution osmolality measurements made by both vapor  pressure and freezing point osmometers for varying amounts of  ethanol in human blood serum  Note that in the vapor pressure  instrument  concentrations of ethanol anywhere within the clinically  significant range do not appreciably affect the indication of osmolali   ty  This is because the vapor pressure of a water ethanol solution  does not change measurably with small concentrations of ethanol   On the other hand  the freezing point osmometer tends to overesti   mate the actual number of ethanol particles in the solution  as the  freezing point falls disproportionately with increasing amounts of  ethanol  Thus  neither instrument faithfully reports osmolality in the  case of water ethanol mixtures  In clinical practice  the unique  response of the vapor pressure osmometer is usually an advantage  inasmuch as it allows the clinician or attending physician to monitor  the patient s serum metabolites  other than alcohol  independently  of the patient s blood alcohol level     TABLE   ETHANOL IN HUMAN BLOOD SERUM  VAPOR PRESSURE VERSUS FREEZING  POINT OSMOLALITY DETERMINATION     1   2   2   2  Calculated Measured Measured  Serum Ethanol Ethanol Ethanol Total       V P   Osmolality Added   kg 
64. s or samples as small as 2 pL    can be measured reliably with special procedures                          Measurement Range Typically 0 to 3200 mmol kg    25   C ambient  Measurement Time 80 seconds   Resolution 1 mmol kg   Repeatability Standard deviation  lt  2 mmol kg   Linearity 2  of reading from 100 to 2000 mmol kg   Readout 10 X 6 8 cm LCD   Operating Temperature 15   to 37   C ambient  instrument should be at stable    temperature before calibrating              Calibration Automatic using Optimol    osmolality standards  Serial Output RS 232  ASCII format   Electrical  Line Voltage 100 to 120 V or 220 to 240 V nominal   set at  factory  user selectable with fuse change  50 to 60 Hz  Power Less than 5 watts  Fuses 1 8 ampere  1 4  x 1 1 4    time delay type for 100 to 120    volts  2 required  1 16 ampere  1 4    x 1 1 4    time delay  type for 220 to 240 volts  2 required           Size  Height 17 cm  6 6    Width 29 cm  11 5      Depth 34 cm  13 5      Weight 3 6 kg  8 Ibs         mmol kg is the Standard International  SI  unit of osmolality  Refer to Appendix E     71    APPENDIX  B    Accessories  Supplies  and Replacement Parts    ACCESSORIES    AC 037 Micropipettor  10 microliter  AC 061 Ampule Organizer  AC 066 Thermocouple Head Assembly  model 5520 0 to 3200 mmol kg  AC 067 Thermocouple Head Assembly  model 5520 above 3200 mmol kg  OM 275 Hex Driver  9 64  screwdriver handle   OM 300 Forceps  5 inch  stainless steel  STANDARD SAMPLE HOLDERS  for solution osmo
65. t Parts 67  73  Reproducibility 68  RS 232 15  93    S  Sample s   Chamber 10  52  76  lever 14  35  37  39  51    99    5  Sample s   complex 31  81  89  holder 10  12  35  36  45  cleaning 39  45  63  84  special purpose 81  82  leaf disc  85  loading 34  35  59  82  84  mutiphasic 81  nonhomogeneous 88  physical characteristics 9  81  89  slide 12  13  19  solid s  85  tissue 81  85  very small 82  very large 85  viscosity 9  81  volume 23  71  large 85 87  very small 83 4  variations in 34  Sampling Error 68  Screen  see display screen   SELECT Keypad 12  13  32  40  42  Sensing Element 10  Serial Data  output 44  93  94  pin diagram 94  Port 14  15  44  93  94  voltage data 93  Signal Chime 14  37  77  shutter wheel 15  Software  resident version 20  Solute Particles 9  Solution  colligative properties of 9  76  concentration 9  75   see also  osmolality      100    Special Applications 81 90  Clinical and General Research 81  Specimen  feces 81  gastrointestinal aspirate 81  physical state 9  removing 38  sputum 81  sweat 81  Standard Deviation 30  59  62  Standard International Units 9  38  71  77   91  Swab Stick 55    T  Teasing needle 82  Temperature  ambient 10  20  21  43  71  considerations during setup 19  dew point 10  Drift Scale 20  21  53  60  61  82  stability 58  67  82  Thermal Equilibration 20  21  52  53  60   67  76  85  Theory of Operation 75 79  Thermocouple  TC   cleaning 40  45 50  55  56  cooling 77  control circuitry 43  contamination 18  21  31 
66. ter running a sample  before  opening the chamber  Select Average Mode and  press ENTER  The last sample assayed will be  1 in  the data group     SECTION 3                                         5                  3 1 Operation Overview    Auto Repeat Mode  Checks the repeatability of the osmometer on the    Sensor Dpuaind same sample  The instrument will automatically run  999 Lk 10 consecutive measurements on the sample   usu   nma 3 ally a 1000 mmol kg Optimol sample  and display    the statistical data  Since the chamber is not opened  between measurements as in other modes  a short  delay occurs between each measurement while  water evaporates from the thermocouple  During this  time the display screen shows  Sensor Drying      Low osmolality samples  below 200 mmol kg  may                       show a difference between the first        any subse   quent readings if the chamber is contaminated  see  998 mmaol  ks Section 3 4    Sample  3  Mean The Auto Repeat sequence can be interrupted at  pun any point simply by opening the sample chamber   Auto Repeat Mode Process Delay Mode    Complex samples  such as leaf and other samples  from which water may not readily evaporate  require  long periods to reach vapor equilibrium  Process    Process Delay Delay Mode delays the measurement cycle after you       1    mmol ka close the chamber until you press ENTER  In  research applications  this lets you delay measure   Press ENTER ment until vapor equilibrium is achieved  The m
67. terials   For these more difficult situations  try cleaning agents  such as acetone or a laboratory detergent such as  Alconox        To remove deposits   1 Apply cleaning agents using the methods    described in Section 4 3     2 Cut a wooden swab stick on a sharp angle to  form a fine point        3 Scrub the surface of the mount with the swab  stick and rinse     Performed under the microscope  this procedure is  unlikely to damage to the thermocouple itself  With  patience  and repeated use of cleaning agents  even  the most severely contaminated thermocouple can  be cleaned     55    SECTION 4  PREVENTIVE MAINTENANCE    4 6 Severe or Stubborn Contamination    To clean dark or corroded copper connection points   1 Apply a droplet of concentrated ammonium hydroxide  NH4OH     28 to 30   to the TC mount  Soaking with this solution for a    few minutes will reduce oxidation and restore the bright copper  color     2 Rinse the thermocouple with pure water     56    SECTION 5  SOLVING PROBLEMS    5 1 Troubleshooting    This section describes problems you might encounter in using the  Vapro osmometer  with suggested solutions  The first part is a flow  chart to help you identify problems by symptom  Beginning with the  apparent symptom  trace through the flow chart to identify possible  causes and solutions for the difficulty  Each set of symptoms and  solutions is repeated and discussed in greater detail in the following  pages of Section 5 1  The solutions are indexed to more
68. the thermo   couple senses the ambient temperature of the air  thus establishing  the reference point for the measurement  Under electronic control   the thermocouple then seeks the dew point temperature within the  enclosed space  giving an output proportional to the differential tem   perature     The difference between the ambient temperature and the dew point  temperature is the dew point temperature depression   an explicit  function of solution vapor pressure     Dew point temperature depression is measured with a resolution of  0 00031   C  The microprocessor controlled measurement cycle  requires 80 seconds     Appendix C contains the theory of operation of the vapor pressure  osmometer     10            Head        Thermocouple and TC Mount  Connector         Thermocouple Head and Sample  Chamber Assembly    11    Sample Slide   Sample Holder      Head Access Cover Press          to Remove  Cover    Pipette Guide    Display Screen                              Gazer                                   Power Indicator SELECT Keypad ENTER Keypad    12    SECTION 1  INTRODUCTION    1 5 Controls and Features    INSTRUMENT FRONT PANEL    Display Screen   10 x 7 cm LCD  Displays menu selections  osmolality readings   countdown of measurement time in seconds  operating status  fault  conditions  and other information     SELECT Keypad  Press to call up menus and to select operation mode     ENTER Keypad  Press to engage a selected menu item or operation mode     Pipette Guide 
69. uccessful results have been achieved using standard labora   tory lint free tissue        83    APPENDIX  D    Special Application Notes    Instructions    1 Calibrate the instrument using 2 pL of standard     2 Load a single paper disc into the center of the special sample  holder  You may need to use the teasing needle and the  tweezers to separate discs that        stuck together     3 Place sample into the center of paper disc  Be sure to touch  the pipette onto the disc as in regular procedure  Be sure the disc  is completely saturated     4 Close the sample chamber to begin the measurement cycle     5 When the measurement is complete  open the sample cham   ber and retract the sample slide     6 Thoroughly clean the sample holder of all sample material  using lint free tissue and a cotton tipped applicator     84    APPENDIX  D    Special Application Notes    MEASURING LARGE SAMPLES       Measuring large samples requires consideration of the nature and  size of the sample  You should experiment with these procedures to  find the best approach for your particular application     Samples such as leaf discs  tissues  and other solids often require  considerable time to reach equilibrium  The Process Delay Mode  allows you to delay the measurement indefinitely or to take succes   sive readings without opening the chamber     The time required to achieve equilibration can be determined by  taking measurements until the readings no longer decrease  Once  you become familiar w
70. us on its surface     SECTION 3  OPERATING THE VAPRO OSMOMETER    3 2 Loading Samples    Gently push the sample slide into the instru   ment until it stops   Never close the chamber  unless the sample holder is in this position                      8 Grasping the sample chamber lever between  thumb and forefinger  rotate it smoothly to the  closed position     NOTE    Since the sample may concentrate slightly before the  chamber is sealed  steps 5 through 8 should be performed  with consistent timing  A warning chime sounds If the  chamber is left open longer than 2 minutes              Closing the lever starts the measurement cycle  The  display screen shows    In Process    and counts down  the remaining time          in Process         seconds    When the measurement is completed  the chime  sounds  The display screen shows the osmolality of  the specimen       zmolalitu    227 nnolzka       37    SECTION 3  OPERATING THE VAPRO OSMOMETER    3 2 Loading Samples    The display shows this final reading until the cham   ber is opened and closed once again     NOTE     Vapro reports osmolality measurements in Standard  International  SI  units  mmol kg        CAUTION    During long  uninterrupted measurement periods   occasionally allow the instrument to return to the  standby mode by initiating an operating cycle on an  empty chamber  This is necessary for the instrument  to readjust itself to any temperature change that may  have occurred during the interval  Failure to do th
71. ussius Clapeyron equation 79  Clean Test 32  34  39  40  45  50  52  54  58   61  Colligative properties 9  75  89  Contamination  see Thermocouple   Customer Service 8    D  Delete Result Function 33  Dew Point Convergence 77  Dew Point Temperature 10  Dew Point Temperature Depression 10  79  Display  abnormal 63  blank 60  language 20  101  screen 9  12  20  34  36 37  40  42  58  61  64  77  93  94  units 20  101    E   Electronic Malfunction 67  70   ENTER Keypad 12  13  40  42  44  85  86   93  94   Equilibration  see thermal equilibration    Erratic Readings 59  70   Error Message s  58  59  61  63  64   Ethanol in Human Blood Serum 90    F   Facial Tissue 18  Feces 81   Filter Paper 82  85  Forceps 18    97    Function  Calibrate 32  33  41  42  Clean Test 32  Delete Result 33  Menu 29  32  42  Fuse s  16  changing 28  door 16  28  specifications  28  71    H   Hex Driver 18  47  Human Blood Serum 90  Hygrometric Method 9    I  Initialization cycle 20  Inhomogeneities 9  In Process Display  37  Instrument  dimensions 71  front panel 13  interior 15  moving 21  precision 65  68  rear panel 16  response characteristics 41  right side 14  15  setup 17  specifications 71  weight 71  International Attention Symbol 9  93    K   Keypad s   ENTER 12  13  29  31 32  42  44  86  87   93  94  SELECT 12  13  29  32  42    98    L  Language  changing 20  95  display 20  95  Linearity  in low range 61  62  66  Line Voltage 71    M  Malfunction  see Electronic   Measurement  artifact
72. visible to the eye  even if the  thermocouple appears to be clean  it may not give an  acceptable clean test  Section 3 4   In this case  follow  the instructions in Section 4 6           3 Check if the thermocouple has been deformed  or broken  Refer to Section 5 4 for information  on identifying thermocouple deformation and  how to restore it to normal shape     4 Inspect the TC head connector and the mating  pins for distortion or misalignment     66    SECTION 5    SOLVING PROBLEMS  5 2 Common TC Head Problems    NOTE     Always switch power off before connecting or disconnecting the TC head        If the connector is damaged  electrical connection may be compro   mised or fail altogether  A failed connection produces an ERROR  message on the display  as with a broken thermocouple  A poor  connection can cause erratic performance     If the source of difficulty is still unknown  at least the most frequently  occurring problems will have been eliminated     Testing Osmometer Performance    1 Reinstall the TC head to continue troubleshooting   2 Set the instrument up using the procedure outlined in Section 3   3 Allow 30 minutes for thermal equilibrium     4 If you have a problem performing any of the steps of the setup  procedure  a malfunctioning electronic module is likely  Contact  Wescor or your dealer for assistance  Replacement parts are  available for user installation  or the entire instrument can be  returned to Wescor for repair  Loan instruments are available  f
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Manual do utilizador externo  Teclado sem fios  FAX-/DATENMODEM BENUTZERHANDBUCH  MANUAL DE INSTRUÇÕES  here - Athletics Direct  PB-80C24 Manual  ESPECIFICACIÓN DE PRODUCTO VEGEStart Complet  FND4011SX取扱説明書【PDF460KB】  Télécharger au format PDF (6.23 Mo) - Arles kiosque  AP - Science de la communication et des médias    Copyright © All rights reserved. 
   Failed to retrieve file