Home

Invacare Platinum

image

Contents

1. Remove valve assembly from tubing CAUTION If P E valve is not to be installed IMMEDIATELY plug the P E valve tubing to prevent sieve bed contamination NOTE To ensure proper orientation the Out port barb should face to the right side of the sieve bed when viewing the back of the unit 6 Install new P E Valve assembly IMMEDIATELY by reversing STEPS 3 5 7 Connect sieve bd tubing to each side of P E valve 8 Secure sieve bed tubing with new tie wraps 9 Connect spade connectors to P E valve assembly NOTE After replacing P E valve retiming of the concentrator may be necessary Refer to Timing on page 69 10 Run unit and check for leaks Refer to Leak Test on page 93 11 Reinstall cabinet Refer to Removing Cabinet on page 26 Platinum Oxygen 50 Part No 1118396 SECTION 8 P E VALVE Out Port Barb Sieve Bed Tie Wraps Sieve Tubing Sieve Beds Shown for Reference Sieve Beds Shown for Reference FIGURE 8 1 Replacing P E Valve Part No 1118396 51 Platinum Oxygen SECTION 9 SIEVE BEDS SECTION 9 SIEVE BEDS Replacing Sieve Beds A DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical out let before servicing NOTE This procedure applies to Platinum 5 and Platinum 9 models NOTE For this procedure refer to FIGURE 9 1 on page 53 and FIGURE 9 2 on page 54 NOTE ALWAYS replace sieve
2. Turn power off O and unplug unit Remove cabinet Refer to Removing Cabinet on page 26 Turn unit so the back faces you Loosen and lower the adjustable clamp on the product tank Pull product tank up and back to access the check valve PVC tubing Locate the check valve and PVC tubing to your right Cut the PVC tubing below the check valve in half 9 Bs WN da Check Valve Cut PVC Tubing below Check Valve to install Tap in Kit Adjustable Clamp Product Tank FIGURE 25 1 Disassembly Part No 1118396 101 Platinum Oxygen SECTION 25 SIEVE BED PRESSURE CHECK Installation NOTE For this procedure refer to FIGURE 25 2 1 Insert the T fitting into the ends of the PVC tubing 2 Secure 6 mm 4 inch clamps to each end of the PVC tubing T fitting connection 3 Attach the 6 mm x 5 cm x 2 inch silicone tubing to the new T fitting and secure with a 6 mm inch clamp 4 Insert plug into the open end of the 6mm x 5 cm x 2 inch silicone tubing 5 Secure with a 6 mm inch clamp 6 mm 4 inch x Clamp EE i 0 T Fitting mm x 5 cm a 96 4 2 8 Tubing A x 2 inch dr Tubing D wg Plug 6 mm inch Clamp FIGURE 25 2 Installation Pressure Testing NOTE For this procedure refer to FIGURE 25 3 on page 103 1 Remove plug and connect pressure gauge to the 6 mm x 5 cm x 2 inch silicone tubing and secure
3. Electric current flowing through a metal film resistor heats the disk in excess of 300 C Oxygen molecules contact the electrode of the disk and pick up extra electrons to become oxygen ions These oxygen ions are attracted to the electrode on the bottom of the zirconia sensing disk Because of the crystal structure of the zirconia only oxygen ions can pass through When the oxygen ions reach the bottom electrode the extra electrons are released from the oxygen ions and oxygen molecules return to the air The number of electrons is directly related to the oxygen concentration The electrons travel to the P C board where they are counted and the oxygen concentration reading is calculated A microprocessor on the P C board contains software that interprets the signal being received from the sensor It compares the signal to clinically acceptable limits Signals outside of the clinically acceptable limits generate responses in the form of lights audible indicators and or system shut down Operating Sequence Once the power switch has been turned on I the SensO circuit will wait five minutes for the concentrator to begin producing clinically acceptable oxygen and the oxygen sensor to stabilize The GREEN light will illuminate indicating normal system operation while the oxygen sensor is warming up After 5 minutes if the oxygen purity exceeds 85 2 the GREEN light will continue to illuminate Part No 1118396 15 Platinum Oxygen
4. TRANSFORMER SECTION I 6 TRANSFORMER Replacing Transformer Assembly A DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing NOTE For this procedure refer to FIGURE 16 1 NOTE This procedure applies to Platinum 5 and Platinum 9 models Unplug unit Remove cabinet Refer to Removing Cabinet on page 26 Remove control panel Refer to Removing Control Panel on page 64 Locate the transformer assembly in the upper left corner on the back of control panel Unplug the transformer connector from the P C board Remove the two keps screws that secure transformer to the back of the control panel db gv OT ge en db GES Reverse STEPS 5 6 to install new transformer Torque keps screw to 1 53 N M x 0 11 N M 13 5 1 inch Ibs Reinstall control panel Refer to Control Panel on page 64 Reinstall cabinet Refer to Cabinet on page 26 Control Panel Transformer Assembly Keps Screws P C Board Transformer P C Board Connector xX J N 4 X K No NN FIGURE 16 1 Replacing Transformer Assembly Platinum Oxygen 70 Part No 1118396 SECTION 17 ON OFF SWITCH SECTION I 7 ON OFF SWITCH Replacing On Off Switch A DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing NOTE For this procedure refer to FIGURE 17 1 on page 72 NOTE This procedure ap
5. Remove poppet and O ring from manifold opening Detail A 10 Discard old poppet and O ring 11 Install new O ring into the opening in the plastic manifold housing Detail A Platinum Oxygen 84 Part No 1118396 SECTION 20 4 WAY VALVE CAUTION DO NOT overtighten pilot valve stem when installing it into plastic manifold Damage to the rubber O ring and or plastic manifold may occur 12 Install new slotted poppet into the bottom of the pilot valve stem with tapered end facing to the bottom 13 Install pilot valve stem and new slotted poppet with coil with yoke and locknut intact into manifold opening 14 Use a flatblade screwdriver and tighten pilot valve stem clockwise until snug Torque to 2 8 0 5 25 inch lbs 5 in Ibs 15 Install new label onto pilot valve coil yoke NOTE DO NOT remove spade connectors from pilot valve coils Simply lay coils with spade connectors aside while performing this installation 16 Position pilot valve coil with yoke at an approximate 60 angle to the right as viewed from the front of valve to ensure spade connectors are not pulled or tensioned after reinstallation 17 Use the flat blade screwdriver and hold the pilot valve stem in place 18 Tighten locknut clockwise with the 9 16 inch wrench until snug to secure the stem and coil with yoke DO NOT overtighten Torque to 2 3 0 20 inch lbs 3 in Ibs CAUTION DO NOT overtight
6. SECTION 19 HOUR METER SECTION 19 HOUR METER Replacing Hour Meter A DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing NOTE For this procedure refer to FIGURE 19 1 on page 76 NOTE This procedure applies to Platinum 5 and Platinum 9 models 1 2 N o OND gm Unplug unit Remove cabinet Refer to Removing Cabinet on page 26 Remove control panel Refer to Removing Control Panel on page 64 Remove and code the two hour meter connectors and wires connected to back of hour meter Spread retaining clips on hour meter housing that secure hour meter to control panel Remove hour meter by pushing meter through front of control panel Install new hour meter reversing STEPS 4 6 Reinstall control panel Refer to Removing Control Panel on page 64 Reinstall cabinet Refer to Removing Cabinet on page 26 Part No 1118396 75 Platinum Oxygen SECTION 19 HOUR METER Hour Meter Housing Hour Meter Connectors Hour Meter Wires Retaining Clips Hour Meter Housing Back of Control Panel FIGURE 19 1 Replacing Hour Meter 76 Part No 1118396 Platinum Oxygen SECTION 20 4 WAY VALVE SECTION 20 4 WAY VALVE A DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing Replacing 4 Way Valve NOTE For this procedure refer to FIGURE 20 1 and FIGURE 20 2 on page 78 NOTE
7. SECTION 2 INSTALLATION SEQUENCE OF OPERATION If the oxygen level is not above 85 2 after the first five minutes the system will continue to monitor the O5 and wait for a maximum of 30 minutes from start up to reach 85 2 before activating an alarm Environmental factors such as low voltage high altitude or age of the machine will affect the time required to reach 8576 2 If the oxygen level is not above 85 2 within the first 30 minutes the oxygen concentration alarm sequence will activate and the unit will shut down When oxygen concentration is above 85 2 the sensor measures oxygen purity every 10 minutes If a reading falls below 8576 2 a YELLOW light will illuminate If the oxygen purity falls below 73 376 the RED light Alarm Shut Down mode will activate Platinum Oxygen 16 1118396 SECTION 3 PNEUMATIC DIAGRAM SECTION 3 PNEUMATIC DIAGRAM Plantinum Oxygen Concentrator Ouput Output HEPA Filter Check Flowmeter Valve Patient 9 Output Oxygen Sensor Pressure Regulator SensO2 models Only D gt a c E Pressure Sensor D Product Tank i i_ Check W LA Valves Air Inlet oe Ir Inte poco Y T i i UA EE MUN a 4 Cabinet NI ii a Filter s y In Y A Pa Compressor i Inlet Filter Sieve i Sieve Bed Valve i Bed a i way Compressor Valve i VP i
8. Solenoid Relief uet Pilot Valve Im B NS 1 s _ yh E Part No 1118396 17 Platinum Oxygen SECTION 4 TROUBLESHOOTING SECTION 4 TROUBLESHOOTING Normal Operation Internal Status Indicators RED Off GREEN Off Unit plugged in power switch on Single beep on start up Power Loss Internal Status Indicators RED Off GREEN Off Unit unplugged power switch on alarm off Power Loss Internal Status Indicators RED Off GREEN Off Unit plugged in power switch alarm off battery Free circuit drained Platinum Oxygen No Problems No Problems No power at outlet Power cord Frayed Broken or damaged spade Spade connector from power cord loose or disconnected inside back of unit On Off VO switch a Disconnected wire b Faulty switch Circuit breaker tripped P C board a P C board damaged b Loose or damaged connector System Okay Battery Free Circuit drained Plug in cord and turn power switch ON I to recharge Check electrical outlet with a table lamp or voltmeter set on the appropriate VAC scale If outlet isn t working check protective device in home s electrical panel or consult an electrician Also ensure that unit is properly plugged in DO NOT use extension cords Move to another outlet or circuit a Reattach cord b Replace power cord connectors on plug c Reattach connector Check all electrical connection
9. This procedure applies to Platinum 5 and Platinum 9 models 1 Unplug unit 2 Remove cabinet Refer to Removing Cabinet on page 26 3 Remove control panel Refer to Removing Control Panel on page 64 4 Remove the four mounting screws that secure the 4 way valve to the manifold assembly g Position new 4 way valve onto manifold assembly and loosely install mounting screws through 4 way valve and into manifold assembly NOTE Ensure the manifold gasket is in place on 4 way valve before installing CAUTION Torque sequence and specification MUST be adhered to or possible damage to the 4 way valve could result 6 Tighten mounting screws in the sequence described below 7 Pre torgue the mounting screws to 1 1 0 2 10 in lbs 2 in lbs using the torque sequence 8 The mounting screws can be now be torqued to 2 5 0 2 22 in lbs 2 in lbs in the same sequence CAUTION This is a maintenance free valve Opening the valve will void any and all warranties applicable to the valve Part No 1118396 77 Platinum Oxygen SECTION 20 4 WAY VALVE Manifold Assembly Mounting Screws NOTE The Numbers indicate the mounting screw torque sequence pre torque to 1 1 N M 0 2 N M 10 in lbs 2 in lbs then torque to 2 5 x 0 2 22 in Ibs 2 in Ibs FIGURE 20 1 Replacing 4 Way Valve 5 Liter Models Manufactured Before 8 1 05 Manifold Assembly 4 Way Va
10. for 9 Liter and shift the 4 way valve to the opposite sieve bed If pressure reading is 21 p s i g 0 3 for 5 Liter and 25 p s i g for 9 Liter when both valve shifts are made proceed to STEP 10 If pressure in the concentrator does not fall into the specifications listed above refer to Troubleshooting on page 16 Turn the concentrator off O Disconnect pressure gauge from the HomeFill port and reinstall plug Platinum Oxygen 104 Part No 1118396 SECTION 25 SIEVE BED PRESSURE CHECK Pressure Gauge Back of Ya x 2 inch Concentrator Silicone Tubing with Connector HomeFill Port FIGURE 25 5 Pressure Testing Part No 1118396 105 Platinum Oxygen NOTES NOTES Platinum Oxygen 106 Part No 1118396 LIMITED WARRANTY LIMITED WARRANTY NOTE THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4 1975 This warranty is extended only to the original purchaser who purchases this product when new and unused from Invacare Corporation or a dealer This warranty is not extended to any other person or entity and is not transferable or assignable to any subsequent purchaser or owner Coverage under this warranty will end upon any such subsequent sale or other transfer of title to any other person This warranty gives you specific legal rights and you may also have other legal rights which may vary from state to state Invacare Corporation warrants its P
11. 30 NOTE Perform this procedure as needed depending on the outlet HEPA filter check results 1 Unplug the unit Remove the cabinet Refer to Removing Cabinet on page 26 2 3 Remove the tubing from both sides of the existing HEPA filter and discard old filter 4 Connect tubing to both sides of the new HEPA filter as shown below 5 Install cabinet Refer to Removing Cabinet on page 26 Part No 1118396 29 Platinum Oxygen SECTION 6 PREVENTIVE MAINTENANCE Control Panel Tubing AS JM Tubing HEPA Filter FIGURE 6 3 Replace the Outlet HEPA Filter Replacing the Compressor Inlet Filter NOTE For this procedure refer to FIGURE 6 2 on page 29 and FIGURE 6 4 on page 31 NOTE Perform this procedure as needed depending upon the environment the concentrator is used in 1 Unplug the unit 2 Locate the filter access and humidifier adapter storage plate located on the side of the concentrator 3 Insert a flathead screwdriver in the plate groove on the top edge of the plate and gently pry the humidifier adapter storage plate open to remove the plate Refer to Detail A in FIGURE 6 2 4 Grasp the compressor inlet filter and pull downward 5 Discard the existing compressor inlet filter NOTE Filter 1131249 is fully compatible with all models This filter replaces 1108078 6 Install new compressor inlet filter by inserting the filter opening into the rubber base Detail of F
12. NG Y GG A yd 75 SECTION 20 4 WAY VALVE 00000000000000000000000000000000000000000000000000000000 77 Replacing 4 ValV inii ta oen iai epe NG e 77 Replacing Manifold eroe rota tn creme ico 79 5 Liter Models Manufactured before 7 24 03 eerte 79 5 Liter Models Manufactured after 7 24 03 and 9 Liter 82 Replacing Pilot Valve Poppets and O RINgS ccssessessesesssessessssessessessessessessesessessessessesseeseeseseesseey 84 SECTION 21 PURITY INDICATORS 00000000000000000000000000000000000000000000000 87 SensC Alarm Indicators ie did 87 EF D PDT 87 Indicator Light Explanation SensO Models eee 87 Indicator Light Explanation LX Models GR cusses 88 SECTION 22 TIMING 0000000000000000000000000000000000000000000000000000000000000000000 89 89 GNGWON 91 SECTION 23 LEAK TEST 00000000000000000000000000000000000000000000000000000000000000 93 Leak Test cea o adu EAE 93 SECTION 24 ALARM TEST 0000000000000000000000000000000000000000000000000000000000 96 Alarmiatid Sensor 96 Power Loss Alaf dedic y O Fd AGORA queen De 96 Low Pressur
13. Platinum Oxygen SECTION 20 4 WAY VALVE 5 Liter Models Manufactured after 7 24 03 and 9 Liter Models NOTE For this procedure refer to FIGURE 20 4 on page 83 Tools Required Diagonal Cutters e Flathead Screwdriver e inch ratchet 4 inch socket and inch extension e Pliers e Tie wraps A DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing Unplug unit Remove cabinet Refer to Removing Cabinet on page 26 Remove control panel Refer to Removing Control Panel on page 64 Remove the tie wraps connecting the sieve bed hoses to top of manifold assembly Remove sieve bed hoses from manifold assembly Si gu SP Remove and code the two spade connectors from the valve pilot on top of the manifold assembly N Loosen the adjustable clamp that secures valve manifold input hose to heat exchanger 8 Remove hose from end of heat exchanger Remove the two mounting screws washers 1 per side and bushings 2 per side securing the 4 way valve manifold assembly to top of sound box 10 The 4 way valve manifold assembly can be removed from the concentrator by lifting the assembly straight up through the cutout in the sound box 11 Remove 4 way valve from manifold assembly Refer to Replacing 4 Way Valve on page 77 12 Remove Y tube and throttle muffler assembly from manifold 13 Loosen the adjustable clamp on valve manifold input
14. all components Storage 1 Store the repackaged oxygen concentrator in a dry area 2 DO NOT place other objects on top of the repackaged concentrator FIGURE 5 1 Unpacking Part No 1118396 11 Platinum Oxygen SECTION IMPORTANT SAFEGUARDS SECTION I IMPORTANT SAFEGUARDS A WARNING IMPORTANT SAFEGUARDS contains important information for the safe operation and use of this product Accessories Information There are many different types of humidifiers oxygen tubing cannulas and masks that can be used with this device You should contact your local home care provider for recommendations on which of these devices will be best for you They should also give you advice on the proper usage maintenance and cleaning Use of some administration accessories or certain humidifiers not specified for use with oxygen concentrator may impact the performance General Warnings The use of oxygen therapy reguires that special care be taken to reduce the risk of fire Any materials that will burn in air and some that will not are easily ignited and burn rapidly in high concentrations of oxygen For safety concerns keep all sources of ignition out of the room in which this product is located and away from areas where oxygen is being delivered Textiles oil and other combustibles are easily ignited and will burn with great intensity in oxygen enriched air NO SMOKING signs should be prominently displayed A spontaneous
15. canister to ensure the barbed fittings are clear and do one of the following e Replace exhaust canister and muffler if plugged or restricted Install only new exhaust muffler if exhaust canister is in good condition Install new re install existing exhaust canister by reversing STEPS 3 5 Reinstall cabinet Refer to Removing Cabinet on page 26 Part No 1118396 3l Platinum Oxygen SECTION 6 PREVENTIVE MAINTENANCE Manifold Exhaust Tubing Barbed Fittings Exhaust Canister Exhaust Muffler BE m O 4 i ia 7 A p Mounting Screw s ZZ Rear of Sound Box FIGURE 6 5 Replacing Exhaust Canister Exhaust Muffler Platinum 5 Manufactured before 7 24 03 Inspecting Replacing Throttling Muffler Platinum 5 AW Manufactured after 7 24 03 NOTE For this procedure refer to FIGURE 6 7 on page 34 NOTE The following should be performed at 20 000 hour intervals although these procedures can be performed as needed depending upon the environment the concentrator is exposed to 1 Unplug unit Remove cabinet Refer to Removing Cabinet on page 26 Remove the Y tube from the manifold assembly Very carefully remove the blue tie wrap from the Y tube DO NOT sever the Y tube 9 A N Separate the throttle muffler from the Y tube and inspect the muffler for residue build up 6 If excessive buildup remove the throttle muffler and discard Replace with a new
16. correctly The letters VAC are printed on the check valves The check valves MUST be installed with the VAC on the underside of the check valve Low system pressures and eventual bed contamination will result if check valves are not properly installed NOTE To ensure proper orientation the Out port barb should face to the right side of the sieve bed when viewing the back of the unit 5 Install the new check valve s into the 4 inch PVC tubing 6 Make sure VAC is on the underside of the check valve to ensure proper orientation of check valves when installing 7 Secure the check valve s by installing new tie wraps to the 1 4 inch PVC tubing where previously removed NOTE If 14 inch PVC tubing is damaged in any way replace the section of 14 inch PVC tubing before installing check valve s NOTE Ensure check valves are not crossed Left barb product tank check valve goes to left barb on sieve bed Right barb product tank check valve to right barb on sieve bed 8 Run units and check for leaks Refer to Leak Test on page 93 9 Reinstall cabinet Refer to Removing Cabinet on page 26 Part No 1118396 55 Platinum Oxygen SECTION 10 VALVES PVC Tubing m p Check Valves Pd Sa 5 2 Tie wraps Two per Check Valve PVC Tubing oy 5 Product Tank Adjustable Clamp On 5 Liter models the clamp is located near top of product tank On 9 Li
17. models Platinum Oxygen 86 Part No 1118396 SECTION 21 PURITY INDICATORS SECTION 21 PURITY INDICATORS SensO Alarm Indicators A DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical out let before servicing Operation The Platinum with SensO Concentrator is equipped with an oxygen purity indicator This feature monitors the purity level of the oxygen generated by the concentrator If purity falls below alarm thresholds indicator lights on the control panel will illuminate NOTE Concentrator may be used during the initial start warm up time approximately 30 minutes while waiting for the purity to reach maximum When the unit is turned on the GREEN light will come on SYSTEM greater than 85 After five minutes the oxygen sensor will be operating normally and will control the indicator lights depending on oxygen concentration values The explanation of the indicator light functions are as follows Indicator Light Explanation SensO Models NOTE For this procedure refer to FIGURE 21 1 on page 88 NOTE This procedure applies to Platinum 5 and Platinum 9 models NOTE Indicator lights are visible only when lit GREEN light O2 Normal Operation YELLOW light Call supplier IMMEDIATELY You may continue to use the concentrator unless instructed otherwise by your supplier Be certain that backup oxygen is nearby RED light Q Total unit shutdown Swi
18. muffler If no buildup present reattach the throttle muffler 7 Place the new blue tie wrap onto the Y tube 8 Insert the new muffler approximately 0 5 in 1 27 cm into the Y tube opening 9 Adjust the tie wrap so it is resting just above the lip on the Y tube and tighten 10 Reattach Y tube to the manifold assembly 11 Reinstall cabinet Refer to Removing Cabinet on page 26 Platinum Oxygen 32 Part No 1118396 SECTION 6 PREVENTIVE MAINTENANCE 5 Liter AW Models Y Tube Manifold Assembly Tie Wrap Blue Y Tube Lip Throttle Muffler FIGURE 6 6 Inspecting Replacing Throttling Muffler Platinum 5 AW Manufactured after 7 24 03 Inspecting Replacing the Throttle Muffler Platinum 5 and 9 AWQ NOTE For this procedure refer to FIGURE 6 7 on page 34 Unplug unit Remove cabinet Refer to Removing Cabinet on page 26 Remove the throttle muffler from exhaust canister Cut the two tie wraps that secure the exhaust canister together Remove the exhaust canister from the Y tube Separate the two halves of the exhaust canister Detail A Inspect the throttle muffler located inside the exhaust canister Detail A 9 NDT B Perform one of the following If excessive buildup remove the throttle muffler and discard Replace with a new muffler Proceed to STEP 9 e fno buildup present use existing throttle muffler and proceed to STEP 9 9 Install throttle muffler into Y tu
19. on the concentrator or any repairs made OXYGEN CONCENTRATOR Refer To Service Manual DIAGNOSTIC ALARM SYSTEM ADJUSTMENTS BY UNAUTHORIZED PERSONNEL VOIDS WARRANTY LED FLASHES ALARM RED GREEN PROBABLE CAUSE BEEP AT START UP NO PROBLEM SYSTEM OK SHORT BEEP WITH MAIN POWER LOSS LONG PAUSE CONTINUOUS LOW PRESSURE MAJOR LEAK CONTINUOUS HIGH PRESSURE NO SWITCHING CONTINUOUS TIME OUT FAILURE COMPRESSOR CONTINUOUS PILOT VALVE CIRCUIT CONTINUOUS 73 SHUTDOWN LOW O9 PE VALVE COIL CONTINUOUS CONTINUOUS EEPROM FAILURE o o 4 o occom Oo o CONTINUOUS o Er OXYGEN SENSOR FAILURE NOTE CYCLE POWER ON AND OFF FIVE 5 TIMES TO CLEAR INTERNAL FAULT CODE Cabinet Filter s Clean Weekly Replace as Needed PREVENTIVE Heat Exchanger Clean as Needed MAINTENANCE Compressor Inlet Filter Check Annually or During Scheduled PM SCHEDULE Outlet HEPA Filter Check Annually or During Scheduled PM Compressor Rebuild as Needed 1115180 Exhaust Muffler Replace as Needed Platinum Oxygen 100 Part No 1118396 SECTION 25 SIEVE BED PRESSURE CHECK SECTION 25 SIEVE BED PRESSURE CHECK Models without HomeFill Connection On Back Of Unit Installing the Sieve Bed Pressure Tap In Kit and Checking Bed Pressure A DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing Disassembly NOTE For this procedure refer to FIGURE 25 1
20. only Avoid all other components DO NOT leak test 4 way valve body Carefully leak test for pressure leaks around sieve bed hoses at 4 way valve DO NOT allow leak test solution to enter into 4 way valve and system D With leak test solution or equivalent soapy water leak test the following A The two sieve bed hoses at the top of 4 way valve and hose to bottom center port on 4 way valve FIGURE 23 1 NOTE Refer to Preventive Maintenance on page 27 for maintenance of heat exchanger and filters B Hose connections to top and bottom fittings of sieve beds C P E valve hose connections on top of sieve beds FIGURE 23 2 D Hose fittings at product tank cap and check valves FIGURE 23 2 E The front and rear fittings on heat exchanger FIGURE 23 2 N Replace any tubing that appears cracked worn etc 8 Re install cabinet Refer to Removing Cabinet on page 26 Refer to Typical Product Parameters on page 7 for concentration levels If low concentrations are observed refer to Troubleshooting on page 18 Part No 1118396 93 Platinum Oxygen SECTION 23 LEAK TEST 4 d 4 For 5 Liter Models with Canister Muffler manufactured before 7 24 03 4 Way Valve Sieve Bed Hoses to 4 way Valve STEP 6a BEHIND Exhaust Tubing STEP Exhaust Tubing For 5 Liter Models with Throttle Muffler manufactured after 7 24 03 and 9 Liter Models
21. slide the assembly onto the shaft bringing it flush to the motor bearing NOTE Make sure the eccentric set screw is positioned over the flat of the shaft 4 Using the access hole tighten set screw to 4 2 40 inch lbs Detail A of FIGURE 7 6 Hold the sleeve down to the housing with one hand and slowly rotate the eccentric with the other hand to ensure all the components are aligned properly g Fan NOTE For this procedure refer to FIGURE 7 6 on page 44 1 Reinstall the fan onto the shaft in original orientation Valve Plate NOTE For this procedure refer to FIGURE 7 5 on page 43 1 With the sleeve located and firmly seated on the housing replace the valve plate in the same manner as it was 2 Repeat STEP 1 for other valve plate NOTE Make sure the top edge of the sleeve locates in the O ring groove in the bottom of the valve plate CAUTION Make sure that the O ring is not twisted when seated in the groove of the bottom of the valve plate Compressor Heads NOTE For this procedure refer to FIGURE 7 5 on page 43 1 Place the heads on top of the valve plates 2 Tighten the head screws in a criss cross pattern to 4 2 N M 40 inch lbs Part No 1118396 47 Platinum Oxygen SECTION 7 COMPRESSOR Compressor A WARNING To avoid personal injury or property damage rotate the fan by hand prior to connecting the unit to a power source Check for suction at the air inlet by placing
22. timing value is stored in memory for future use Part No 1118396 89 Platinum Oxygen SECTION 22 TIMING NOTE If the unit is turned off before an autotune cycle is completed the P E valve timing will not be stored or updated If the board is new autotuning MUST be successfully completed before the unit can be placed into service 8 After a successful autotune run the unit for 30 minutes and verify that the oxygen concentration output of the unit is within specification 9 If the oxygen output concentration is not within specification rerun the autotune again 10 Reinstall the cover screws P C Board Pressure Sensor Autotune Push Button Switch FIGURE 22 1 Autotuning Platinum Oxygen 90 Part No 1118396 SECTION 22 TIMING Manual Tuning NOTE For this procedure refer to FIGURE 22 2 on page 92 NOTE This procedure applies to Platinum 5 models only 1 Turn power off O and unplug the unit 2 Remove cabinet Refer to Removing Cabinet on page 26 4 Locate the timing switch on the upper right corner of the P C board near the pressure Plug unit in but DO NOT turn on I until instructed to below Sensor Insert a small common or flathead screwdriver into the adjustment screw of the timing switch Set the timing switch initially at the number 7 position Turn unit on I and wait 15 minutes for unit to stabilize 8 If after 15 minutes of run time the concentration is below spec
23. to Platinum 5 and Platinum 9 models Tools Reguired Diagonal Cutters Flathead Screwdriver l4 inch ratchet 4 inch socket and Va inch extension Pliers Tie wraps Unplug unit Remove cabinet Refer to Cabinet on page 26 Remove the two upper mounting screws that secure the control panel p de Loosen the two lower mounting screws that secure the control panel AN WARNING Use caution when removing the pressure sensor tubing from the product tank Injury could result when tubing disconnects from product tank g Cut the tie wraps that secure the 1 8 inch I D pressure sensor tubing to the product tank connection and remove tubing Remove the inch I D tubing from the top and bottom barbed fittings of the flowmeter D N Cut tie wrap and remove 4 inch I D tubing from the patient outlet barbed fitting behind the control panel NOTE On SensO models the oxygen sensor tube attached to the side of the regulator MUST also be released 8 Disconnect the nine or ten pin connector from P C board 9 Disconnect the eight pin connector from P C board 10 Disconnect the transformer harness connector NOTE Before performing STEP 10 code all wires to ensure correct reinstallation Platinum Oxygen 64 Part No 1118396 SECTION 13 CONTROL PANEL 11 Remove spade connectors from circuit breaker hour meter and On Off I O switch Jumper wire harness from On Off switch to circuit breaker doe
24. your finger over the port as you rotate the fan You should feel a slight suction with each rotation of the fan If you DO NOT feel a suction but you feel or hear a thump as you turn the fan DO NOT connect the unit to a power source Review the assembly procedure for possible error NOTE For this procedure refer to FIGURE 7 9 on page 49 1 Reinstall compressor Refer to Replacing Compressor Assembly 5 Liter Models manufactured before July 24 2003 on page 37 2 Run unit and check for leaks around compressor hoses and compressor tubes Refer to Leak Test on page 93 3 Take pressure readings to ensure proper system pressures Refer to Models without HomeFill Connection On Back Of Unit on page 101 FIELD SERVICE PARTS LIST FOR THOMAS 2660 SERIES COMPRESSOR ITEM PART QTY PER QTY PER DESCRIPTION NO NO ASM UNIT See Chart 2 Connecting Rod Eccentric amp Bearing Assembly 2 624277 2 Piston Cup 3 625776 2 Screw Piston Cup Retainer 4 626175 2 Piston Cup Retainer 5 618114 2 Cylinder Sleeve 6 See Chart Head 7 623143 2 O Ring Head Gasket 8 625175 8 Screw Head 9 621591 2 Valve Plate Assembly 638574 Valve Pressure Relief Not Shown 115 V 10 617177 2 Valve Restraint 11 617562 2 4 Valve Keeper Strip 12 621485 2 4 Valve Flapper Intake amp Exhaust 13 621641 2 Valve Plate 14 625094 2 4 Scre
25. 0 GREEN One Flash Compressor Tighten fittings Replace or repair Or a Leaks at fittings or tubing Repair leak or replace RED One Flash b Leaking or defective relief valve DO NOT use extension cords Use GREEN Three Flashes c Insufficient voltage at outlet another outlet d Worn cup seals or gaskets Rebuild top end of compressor Refer Control Panel Indicators to Rebuilding the Thomas Model RED On 2650 and 2660 Compressor on YELLOW Off page 41 GREEN Off Heat exchanger Replace or retighten a Leak at tubing or body chamber Replace or retighten tubing Replace SWitelron continous b Inspect tubing and heat exchanger heat exchanger Refer to Replacing audible alarm Heat Exchanger on page 60 Compressor shut down Regulator cracked or leaking Replace regulator Refer to Replacing Failure to cycle due to Regulator on page 57 low pressure P E valve leaking Replace P E Valve Check voltage at P E valve connector on 24 volt D C scale The P E Valve activates or energizes approximately one second prior to the activation of the 4 way valve with approximately 24 volts If voltage is in excess of 24 volts consistently replace the P C board Refer to Replacing P C Board on page 68 If the P C board voltage acts normally replace the P E valve Refer to Replacing P E Valve on page 50 NOTE Check for leaks starting at the compressor output through all the pneumatic connections Major
26. 0000000000 17 SECTION 4 TROUBLESHOOTING 00000000000000000000000000000000000000000000000000 18 SECTION 5 CABINET 0000000000000000000000000000000000000000000000000000000000000000000 26 Removing Ea 26 SECTION 6 PREVENTIVE MAINTENANCE 6 6 27 Cleaning the Cabinet Filter a d GYD FA e 27 Replacing the Outlet ne eie etate ost dettes os 28 Check the 28 Replace the Outlet HEPA Filter 28 Replace the Outlet HEPA Filter 2 ean tenau ifydd nd dyd CYH ydd Fn dodd YN FON 29 Replacing the Compressor Inlet Filter ue ee ihe FYD Bodd oM dodid 30 Replacing Exhaust Canister Exhaust Muffler Platinum 5 Manufactured before 7 24 03 31 Inspecting Replacing Throttling Muffler Platinum 5 AW Manufactured after 7 24 03 32 Inspecting Replacing the Throttle Muffler Platinum 5 and 9 33 Cleaning the Heat Exchange 34 Part No 1118396 3 Platinum Oxygen TABLE OF CONTENTS TABLE OF CONTENTS SECTION 7 MM COMPRESSOR 000000000000000000000000000000000000000000000000000000000000 37 Replacing Compressor Assembly 5 Liter Models manufactured before July 24 2003 37 Replacing Compressor Assembly 5 Liter Models manufactured after July 24 2003 38 Replacing Compress
27. 8396 TABLE OF CONTENTS TABLE OF CONTENTS SPECIAL NOTES 090900000000000000000000000000000000000000000000000000000000000000000000000000000 6 TYPICAL PRODUCT PARAMETERS 0000000000000000000000000000000000000000000000000000 7 FEATURES 90900000000000000000000000000000000000000000000000000000000000000000000000000000000000 10 SHIPPING AND HANDLING 000000000000000000000000000000000000000000000000000000000000 1 1 UnpackKing ae E OOF CU OW FNAN E NSPS CULO tm M n RR a a YDD RN ll Y Yd BN Y YNI Af FYFYR efe Y GRUFF 11 SECTION IMPORTANT SAFEGUARDS 66 6 12 Accessories Informatio Ioas t tta oda rias densus baia ebd eite GAD LO 12 General WaEnlhBs Y CY a eiie 12 Maite MAC E a 13 Radio Frequency Interference 13 To Reduce the Risk of Burns Electrocution Fire or Injury to 13 SECTION 2 INSTALLATION SEQUENCE OF OPERATION 14 Installation Verification of Battery Free Power Loss Alarm sse 14 Sequence of 14 SensO Oxygen Sensor Technology Ceramic Zirconia Sensor sss 15 Technical Description GOF 15 Operating SEQUENCE M 15 SECTION 3 PNEUMATIC DIAGRAM 000000000000000000000000000000000000
28. C5LXO2AW 50 dB Average IRCOLXAW OQ For IRCSLXAW IRC5LXO2AW IRC5LXO2AWO Up to 1828 meters 6 000 ft above sea level without degradation of concentration levels NOT RECOMMENDED FOR USE ABOVE 1828 M 6 000 ft For IRC9LXO2AWO Up to 4 000 ft 1230 meters above sea level without degradation of concentration levels For 4 000 ft 1230 meters to 13 129 ft 4000 meters below 9026 efficiency For IRC5LXAW IRC5LXO2AW IRC5LXO2AWO 93 minimum at to 3 L min 91 minimum at 4 L min 87 minimum at 5 L min maximum recommended flow For IRC9LXO2AWO 93 minimum at to 7 L min 91 minimum at 8 L min 8796 minimum at 9 L min 5 Liter is 34 5 3 45 kPa 5 0 5 psi 9 Liter is 62 05 kPa 3 45 kPa 9 0 5 psi For IRC5LXAW IRC5LXO2AW IRC5LXO2AWO 0 5 to 5 L min Flowrates less than L min are not recommend For IRC9LXO2AWO I L min to 9 L min Flowrates less than L min are not recommended For IRCBLXAW IRC5LXO2AW IRC5LXO2AWO 0 L min to 0 5 L min Rapid Audible Alarm Beeping No Accessories Connected IRC9LXO2AWO 0 L min to 1 0 L min Rapid Audible Alarm Beeping No Accessories Connected 320 W IRC5LXO2AWO 395 W IRC5LXO2AW IRC5LXAW 500 W IRC9LXO2AWO 241 kPa 24 1 kPa 35 psi 3 5 psi 0 7 L min Cabinet 2 Outlet HEPA and Compressor Inlet 7 Platinum Oxygen TYPICAL PRODUCT PARAMETERS Safety System Current overload or line surge shutdown High temperature compresso
29. Check pressure at product tank Pressure should rise to 144 79 kPa 21 psi at shift point If not replace P C Board Refer to Replacing P C Board on page 68 Return flow to maximum setting Replace fitting Repair or replace Refer to Replacing Flowmeter on page 73 To accommodate for varying tolerances when replacing components an adjustable timer is used to control the shifting of the Pressure Equalization P E valve Refer to Timing on page 89 a Replace P E valve b Replace P E valve Refer to Replacing P E Valve on page 50 Replace P C board Unit may need retiming after board replacement Refer to Replacing P C Board on page 68 Check pressure at oxygen outlet Adjust regulator If flow is still unstable check for leaks starting at the compressor outlet fitting through all pneumatic connections If no leaks are found and flow is still fluctuating replace the regulator If pressure at test point is within spec 34 5 3 45 kPa 5 0 5 psi for 5 Liter 62 05 kPa 3 45 kPa 9 0 5 psi for 9 Liter replace flowmeter Refer to Replacing Flowmeter on page 73 If low flow conditions persist replace outlet HEPA filter Refer to Replacing the Outlet HEPA Filter on page 28 Part No 1118396 SECTION 4 TROUBLESHOOTING SYMPTOM PROBABLE CAUSE SOLUTION Unit Excessively Loud Pneumatic exhaust a Muffler cracked damaged or missing a Replace Refer to Replacing Exhaust b Muffl
30. E 7 1 Replacing Compressor Assembly 5 Liter Models manufactured before July 24 2003 Replacing Compressor Assembly 5 Liter Models manufactured after July 24 2003 NOTE For this procedure refer to FIGURE 7 2 on page 39 1 Unplug unit 2 Remove cabinet Refer to Removing Cabinet on page 26 3 Loosen clamp and disconnect intake hose and heat exchanger input fitting from compressor assembly 4 Remove heat exchanger cushioned clamp screw from the compressor assembly 5 Remove compressor wires from wire clamps or tie wraps and disconnect motor connector from the main harness connector NOTE Motor connector goes through grommet before reaching the main harness connector 6 Remove protective boot on capacitor 7 Disconnect capacitor wires from the top of the capacitor NOTE Re use the protective capacitor terminal boot if the replacement parts DO NOT provide one Platinum Oxygen 38 Part No 1118396 SECTION 7 COMPRESSOR 8 Remove mounting screws that secure the compressor restraint bracket and remove the bracket 9 Lift heat exchanger tilt compressor assembly toward front of unit and lift out 10 Reverse STEPS 3 7 to install new compressor assembly 11 Run unit and inspect for leaks Refer to Leak Test on page 93 12 Reinstall cabinet Refer to Removing Cabinet on page 26 Intake Hose Grommet Compressors Mounting Screws Protective Boot Capacitor NOTE Models with th
31. FIGURE 23 1 Platinum Oxygen 94 4 Way Valve of 4 way Valve Located B Exhaust Tubing STEP shown for clarity Exhaust Tubing Leak Test Hose fitting on Bottom Center Port of 4 way Valve Located Hose fitting on Bottom Center Port EHIN 6b D NOTE Only the 5 Liter model Part No 1118396 SECTION 23 LEAK TEST Sieve Bed and Product Tank u E Valve Fittings Hose Hoses ST Connections to Top Fittings of Sieve Beds STEP 6b Product Tank Cap Check all hoses valves 2 Check Valve attached STEP 6d Siaya Bede Tubing STEP 6d Hose Connections to Bottom Fittings of Sieve Beds STEP 6b 5 Liter Heat Exchanger Hoses 5 Liter Heat Exchanger Hose 5 Liter Heat Exchanger Hose manufactured before 9 9 05 manufactured after 9 9 05 Front and Rear Fittings 9 Liter Heat Exchanger Hose Front and Rear Fittings FIGURE 23 2 Leak Test Part No 1118396 95 Platinum Oxygen SECTION 24 ALARM TEST SECTION 24 ALARM TEST Alarm and Sensor Tests A DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing Check alarms periodically for proper function Power Loss Alarm NOTE This procedure applies to Platinum 5 and Platinum 9 models With the unit running remove the line cord from power source The power loss alarm mode should sound the audible alarm IMMEDIATELY or within 30 seconds NOTE With t
32. GULATOR Install pressure gauge onto oxygen outlet NOTE Check O pressure at oxygen outlet It should read a steady 34 5 3 45 kPa 5 0 5 psi for 5 Liter 62 05 kPa 3 45 kPa 9 0 5 psi for 9 Liter If pressure is not in specification proceed to STEP 3 If pressure falls within specification no adjustment is needed 3 4 a 6 7 10 11 Unplug unit Remove cabinet Refer to Removing Cabinet on page 26 Move product tank back to access regulator Refer to Replacing Regulator on page 57 Plug unit in and turn power switch on 1 Locate the pressure adjustment screw for 5 Liter models pressure adjustment knob for 9 Liter models in center of pressure regulator top For 5 Liter models only insert a 5 32 inch Allen wrench into the pressure adjustment screw While reading pressure gauge do one of the following e Turn pressure adjustment screw or adjustment knob clockwise to increase output pressure or e Turn pressure adjustment screw or adjustment knob counterclockwise to decrease output pressure Adjust until pressure reads e For 5 Liter models a steady 34 4 kPa 3 45 kPa 5 psi 0 5 psi e For9 Liter models a steady 62 05 kPa 3 45 kPa 9 psi 0 5 psi Once required pressure is achieved reinstall cabinet Refer to Removing Cabinet on page 26 NOTE After ten minutes of run time retest output pressure of concentrator to ensure proper operation and or spec levels 12 Refer
33. IGURE 6 4 7 Push the filter down until the rubber base touches the edge of the filter 8 Reinstall the filter access and humidifier adapter storage plate Platinum Oxygen 30 Part No 1118396 SECTION 6 PREVENTIVE MAINTENANCE DETAIL DETAIL B INSTALLING COMPRESSOR FILTER Filter 1108078 Filter 1131249 Compressor Compressor Label p AR e e E Inlet Filter Gaal P un Compressor a A UN Fiter iL M REMOVE TOP SEAL ONLY REMOVE THIS SEAL ONLY ON CONCENTRATOR MODELS gt gt gt gt WITHOUT lt lt lt lt A FILTER CCESS DOOR PN 1108078 ON NON PLATINUM MODELS X 4 AW W PN 1131249 WES SN ISZ A HERE FIGURE 6 4 Replacing the Compressor Inlet Filter Replacing Exhaust Canister Exhaust Muffler Platinum 5 Manufactured before 7 24 03 NOTE For this procedure refer to FIGURE 6 5 on page 32 NOTE The following should be performed at 20 000 hour intervals although these procedures can be performed as needed depending upon the environment the concentrator is exposed to 1 Unplug unit 2 Remove cabinet Refer to Removing Cabinet on page 26 4 Disconnect and remove the manifold exhaust tubing from the barbed fittings on top Remove the two mounting screws from rear of sound box of the exhaust canister Unscrew the exhaust muffler located on the exhaust canister by turning counterclockwise Inspect exhaust
34. IGURE 7 8 1 Remove the exhaust valve flapper valve restraint and valve keeper strip from the top of the valve plate Discard all 2 Clean the top of the plate with a clean soft cloth Gasket NOTE For this procedure refer to FIGURE 7 8 1 Remove the old gasket from the top of the valve plate and discard 2 Clean the valve plate with a clean soft cloth Gasket Valve Flapper Screw 5 Y Valve Keeper Strip Exhaust Valve Flapper Valve Restraint FIGURE 7 8 Removing Compressor and Compressor Heads Rebuilding the Compressor Exhaust Valve Flapper Valve Restraint and Valve Keeper Strip Part No 1118396 45 Platinum Oxygen SECTION 7 COMPRESSOR Rebuilding the Compressor NOTE Before reassembly wipe any residue from all components with a clean soft cloth Gasket NOTE For this procedure refer to FIGURE 7 8 on page 45 1 Install new gasket seating the gasket firmly in the groove at the top of the valve plate with your finger or blunt object CAUTION Make sure that the gasket is not twisted when seated in the groove of the top of the valve plate Exhaust Valve Flapper Valve Restraint and Valve Keeper Strip NOTE For this procedure refer to FIGURE 7 8 on page 45 1 Install the new exhaust valve flapper valve restraint and valve keeper strip to the top of the valve plate with the valve flapper screw NOTE Torque flapper screws to 1 36 N M 12 inch I
35. IL Pressure DETAIL Sound Box E Regulator Pressure Adjustable TS Adjustment Screw Product Clamp 5 32 inch Allen Tank Lid Regulator Wrench Product Screws Pressure Tank 29 Product Sensor 1 Tank 2 NE En Rese Adjustable NOTE The adjustable clamp hanger is not shown Itis located behind the product tank and is attached NOTE Sieve Beds removed for clarity to the sound box DETAIL 9 LITER MODELS Pressure Pressure Adjustment Regulator Knob Top Mounting Pressure Screws Regulator Pressure Sensor Tubing Product Tank Adjustable Clamp FIGURE 11 1 Replacing Adjusting Regulator 7 Install new regulator onto product tank by reversing STEPS 5 and 6 8 Torque mounting screws to 56 2 11 5 1 in Ibs 9 Re install product tank and secure with the adjustable clamp NOTE On units with SensO gt reinstall 1 8 inch tubing to oxygen sensor housing 10 Adjust regulator pressure if necessary Refer to Adjusting Regulator on page 58 11 Run unit and check for leaks Refer to Leak Test on page 93 12 Reinstall cabinet Refer to Removing Cabinet on page 26 Adjusting Regulator NOTE This procedure applies to Platinum 5 and Platinum 9 models NOTE For this procedure refer to FIGURE 11 1 1 Turn unit on D Platinum Oxygen 58 Part No 1118396 2 SECTION RE
36. Models Manufactured Before 9 9 2005 sse 60 Platinum 5 Models Manufactured After 9 9 2005 cccsssssssessessesssssessesssssssessesstesssessessessssseeses 61 Platinum 9 Concentrato sus eem utt in SG banc YRS 62 SECTION 3 CONTROL PANEL 00000000000000000000000000000000000000000000000000000 64 Removing Control Y EEE pen eise 64 SECTION 14 COOLING FAN 000000000000000000000000000000000000000000000000000000000 66 Replacing Cooling Fan 66 SECTION 15 P C BOARD OD Replacing P G Board 68 SECTION 16 TRANSFORMER TO Replacing Transformer 70 SECTION 17 ON OFF SWITCH 7 Replacing SW iC is acides needed iua 7 SECTION 18 Dez FLOWMETER 000000000000000000000000000000000000000000000000000000000000 73 e UM 73 Part No 1118396 4 Platinum Oxygen TABLE OF CONTENTS TABLE OF CONTENTS SECTION 19 HOUR METER 0000000000000000000000000000000000000000000000000000000000 75 Replacing Hour Mete i i
37. No 1118396 SECTION 2 INSTALLATION SEQUENCE OF OPERATION The electrical activation of the 4 way Valve is accomplished every 8 to 15 seconds by the pressure sensor and P C Board electronics when the pressure reaches a set point of 144 79 kPa 21 psi for 5 Liter and 172 36 kPa 25 psi for 9 Liter The time between cycles is dependent on altitude flow rate and internal environmental factors A valve opens just prior to the shift of the 4 way valve This allows highly concentrated oxygen to enter the just exhausted bed from the top This additional pressure allows the bed to start its cycle at a higher pressure The P E valve will close just after the shift of the 4 way valve If main power is lost the Battery Free Power Loss Alarm will sound a short with a long pause after All units are equipped with a diagnostic alarm system that signals if the pneumatic pressure or electrical systems malfunction The Troubleshooting Guide in this manual explains the alarm system signals and reasons in detail for your convenience SensO Oxygen Sensor Technology Ceramic Zirconia Sensor Technical Description The oxygen being produced by the concentrator flows out of the product tank and into the flowmeter A small flow of oxygen produced by the unit is sent through a precision orifice to the oxygen sensor mounted on the printed circuit board As the oxygen enters the sensor it passes through a screen and contacts the sensing disk
38. ONTAINED HEREIN THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGE OR LIMITATION OF HOW LONG AN IMPLIED WARRANTY LASTS SO THE ABOVE EXCLUSION AND LIMITATION MAY NOT APPLY TO YOU THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE PROVINCIAL LAWS AND REQUIREMENTS Part No 1118396 107 Platinum Oxygen Yes you can Invacare Corporation www invacare com USA One Invacare Way Elyria Ohio USA 44036 2125 800 333 6900 Technical Services 800 832 4707 EU Representative Invacare EC Hong A S Factory Ostergade 3 DK 4270 Hong Denmark Tel 011 45 58852722 Fax 01 1 45 58854386 0434 Manufacturer Invacare Medical Equipment Suzhou Co Ltd Unit 2B Export Processing Zone No 200 Suhong Road SIP Suzhou Industrial Park Suzhou Jiangsu PRC 215021 Tel 86 512 62586192 Fax 86 512 62586072 All rights reserved Trademarks are identified by and All trademarks are owned by or licensed to Invacare Corporation unless otherwise noted Teflon is a registered trademark of DuPont Corp Torx is a registered trademark of Textron Inc 9 2007 Invacare Corporation Part No 1118396 Rev D 8 07
39. PRESSOR DESCRIPTION QUANTITY Valve Plate Gaskets 2 Head Mounting Screws 8 Valve Screws 4 Valve Restraints 2 2 Piston Cup Retainer Screws Tools Required Torque wrench that has an inch pound scale for head screws set screws flapper valve screw and pipe plugs e Torx 1 25 drive for torque wrench for head screws e Torx T 27 drive for torque wrench for retainer screws 1 8 inch Allen wrench attachment for torque wrench for eccentric set screw e 5 32 inch Allen wrench attachment for torque wrench e Flat blade screwdriver attachment or 1 4 inch hex socket for torque wrench for flapper valve screw e Phillips screwdriver attachment for torque wrench for retainer screws CAUTION DO NOT lubricate or use oil on any moving parts The compressor eccentric uses a precision sealed bearing Therefore additional lubrication is not necessary Removing Compressor and Compressor Heads Compressor 1 Unplug concentrator and carefully remove compressor from concentrator Refer to Replacing Compressor Assembly 5 Liter Models manufactured before July 24 2003 on page 37 Compressor Head Plate NOTE For this procedure refer to FIGURE 7 5 on page 43 1 Clean loose dirt from the outside of the compressor Use the Torx 1 25 torque wrench and loosen the eight head screws Note head placement before removing compressor heads ps cUm Remove the compressor head plate Platin
40. Platinum Oxygen 8 Part No 1118396 LABEL SYMBOL STATUS TYPICAL PRODUCT PARAMETERS INDICATOR LIGHTS LED SYSTEM OKAY over 85 2 GREEN Indicator Light O Between 73 3 to 85 276 YELLOW Indicator light A YELLOW Solid B YELLOW Flashing Sensor Failure Call a qualified technician LABEL SYMBOL SYSTEM FAILURE O Below 73 3 RED Indicator Light Continuos Audible Sieve Gard Compressor shutdown Call a qualified technician LX INDICATORS STATUS INDICATOR LIGHTS LED SYSTEM OKAY GREEN Indicator Light a Part No 1118396 SYSTEM FAILURE RED Indicator light Continuous Audible Alarm Sieve GARD Compressor Shutdown Call a qualified technician Platinum Oxygen FEATURES FEATURES FRONT VIEW Oxygen Outlet Oxygen Purity and Fault Indicator Lights SensO model Fault and Power Indicator Lights LX model Flowmeter Power Switch Circuit Breaker Hour Meter Cabinet Filter one of two shown FIGURE 4 1 IRC5LXAW IRC5LXO2AW IRC5LXO2AWO IRC9LXO2AWO ACCESSORIES NOT SHOWN Power Cord Homefill ll home oxygen compressor WT IOH200AW Outlet Fitting Contact your local Invacare representative for available accessory information NOTE This outlet fitting is to be used only for filling oxygen cylinders with the HomeFill II home oxygen compressor The outlet fit
41. SECTION 18 FLOWMETER SECTION 18 FLOWMETER Replacing Flowmeter A DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing NOTE For this procedure refer to FIGURE 18 1 on page 74 NOTE This procedure applies to Platinum 5 and Platinum 9 models 1 Unplug unit 2 Remove cabinet Refer to Removing Cabinet on page 26 3 Remove control panel Refer to Removing Control Panel on page 64 NOTE Note the location of the PVC tubing for accurate replacement after new flowmeter is installed 4 Remove palnuts that secure flowmeter to the control panel Remove flowmeter from front of control panel Install new flowmeter reversing STEPS 4 5 Reinstall control panel Refer to Removing Control Panel on page 64 Mu QD UM Run unit and check for leaks where inch PVC tubing engages flowmeter fittings Refer to Leak Test on page 93 Reinstall cabinet Refer to Removing Cabinet on page 26 Part No 1118396 73 Platinum Oxygen SECTION 18 FLOWMETER 9 Liter Models Palnuts Existing 5 psi New 9 psi Flowmeter Flowmeter Flowmeter Control Panel 10 10 9 9 8 8 7 7 y 6 6 5 5 E 4 4 3 3 2 2 l min O l min O Flowmeter Barbed Fittings Back of Flowmeter FIGURE 18 1 Replacing Flowmeter Platinum Oxygen 74 Part No 1118396
42. Service Manual Platinum Oxygen Concentrator LUN Model IRC5LXAW HomeFill II Compatible Model IRC5LXO2AW With SensO2 and HomeFill Compatible Model IRC5LXO2AWO With SensO2 and HomeFill Il Compatible Model IRC9LXO2AWO With SensO2 and HomeFill Il Compatible Dealer This manual MUST be given to the user of the product User BEFORE using this product read this manual and save for future reference For more information regarding Invacare products parts and services please visit www invacare com Yes you can A WARNING DO NOT USE THIS PRODUCT OR ANY AVAILABLE OPTIONAL EQUIPMENT WITHOUT FIRST COMPLETELY READING AND UNDERSTANDING THESE INSTRUCTIONS AND ANY ADDITIONAL INSTRUCTIONAL MATERIAL SUCH AS OWNER S MANUALS SERVICE MANUALS OR INSTRUCTION SHEETS SUPPLIED WITH THIS PRODUCT OR OPTIONAL EQUIPMENT IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS CAUTIONS OR INSTRUCTIONS CONTACT A HEALTHCARE PROFESSIONAL DEALER OR TECHNICAL PERSONNEL BEFORE ATTEMPTING TO USE THIS EQUIPMENT OTHERWISE INJURY OR DAMAGE MAY OCCUR ACCESSORIES WARNING INVACARE PRODUCTS ARE SPECIFICALLY DESIGNED AND MANUFACTURED FOR USE IN CONJUNCTION WITH INVACARE ACCESSORIES ACCESSORIES DESIGNED BY OTHER MANUFACTURERS HAVE NOT BEEN TESTED BY INVACARE AND ARE NOT RECOMMENDED FOR USE WITH INVACARE PRODUCTS NOTE Updated versions of this manual are available on www invacare com Platinum Oxygen 2 Part No 111
43. and violent ignition may occur if oil grease or greasy substances come in contact with oxygen under pressure These substances MUST be kept away from the oxygen concentrator tubing and connections and all other oxygen eguipment DO NOT use any lubricants unless recommended by Invacare For optimum performance Invacare recommends that each concentrator be on and running for a minimum of 30 minutes at a time Shorter periods of operation may be harmful for maximum product life If the concentrator has a damaged cord or plug if it is not working properly if it has been dropped or damaged or dropped into water call Oualified Technician for examination and repair Keep the oxygen tubing cord and unit away from heated or hot surfaces including space heaters blankets stoves and similar electrical appliances DO NOT move or relocate concentrator by pulling on the power cord NEVER drop or insert any object into any opening Platinum Oxygen 12 Part No 1118396 SECTION IMPORTANT SAFEGUARDS NEVER block the air openings of the product or place it on a soft surface such as a bed or couch where the air opening may be blocked Keep the openings free from lint hair and the like Fill humidifier with water to the level shown by the manufacturer DO NOT overfill Invacare recommends that Crush Proof oxygen tubing be used with this product and NOT exceed 15 2 m 50 ft in length Maintenance The Platinum Oxygen Concentrator
44. atinum 9 models 1 2 4 With unit off disconnect the compressor connector from the main wiring harness Plug unit in and turn power on 1 The Time Out Failure alarm should activate within 40 seconds Refer to SensO2 Indicators on page 99 for shutdown mode Replace tubing and tie wrap NOTE If any alarm fails to perform to specification contact Invacare Technical Service Relief Valve Pull Out Compressor Stem to Simulate Leak Compressor Connector Disconnect from Harness FIGURE 24 1 Low Pressure Test Time Out Test High Pressure Test NOTE For this procedure refer to FIGURE 24 2 on page 98 3 e High Pressure Occurs when pressure in product tank rises beyond a preset value typically 158 kPa 7 kPa 23 p s i 1 p s i Main Valve Coil Alarm sequence occurs when the main valve coil connection or P C board circuitry has failed With the unit running and flow set at 5 L min for 5 liter models and 9 L min for 9 liter models remove one orange wire from the main valve coil The main valve coil alarm should activate within 40 seconds Refer to SensO2 Indicators on page 99 for shutdown mode FIGURE 24 3 Replace tubing and tie wrap NOTE If any alarm fails to perform to specification contact Invacare Technical Service Part No 1118396 97 Platinum Oxygen SECTION 24 ALARM TEST 5 Liter models manufactured Before 8 1 05 5 Liter models manufactured After 8 1 05 and 9 Li
45. be Detail A 10 Reconnect both halves of the exhaust canister Detail A 11 Install exhaust canister over throttle muffler that is inserted into the Y tube Detail 12 Inspect remaining throttle muffler 13 Perform one of the following e If excessive buildup remove the throttle muffler and discard Replace with a new muffler Proceed to STEP 14 Part No 1118396 33 Platinum Oxygen SECTION 6 PREVENTIVE MAINTENANCE Ifno buildup present use existing throttle muffler and proceed to STEP 14 14 Install throttle muffler into the exhaust canister approximately 0 5 in 1 27 cm Refer to Detail A 15 Secure both halves of the exhaust canister together with two tie wraps 16 Reinstall cabinet Refer to Removing Cabinet on page 26 Detail Y Tube Throttle Muffler Exhaust Canister Exhaust LE E Canister Nys C Partial AW SY AM Maps Throttle Muffler Throttle Muffler FIGURE 6 7 Inspecting Replacing the Throttle Muffler Platinum 5 and 9 AWO Cleaning the Heat Exchanger N DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing NOTE For this procedure refer to FIGURE 6 8 on page 35 NOTE This procedure applies to Platinum 5 and Platinum 9 models 1 Unplug unit 2 Remove cabinet Refer to Removing Cabinet on page 26 CAUTION Use care not to deform heat exchanger
46. beds in pairs to ensure that both beds are in optimum condition 1 Unplug unit 2 Remove cabinet Refer to Removing Cabinet on page 26 3 Remove tie wraps and disconnect P E valve asm and sieve bed tubing from top of sieve beds 4 For5 Liter models manufactured before 1 27 05 only Remove large tie wraps 5 Loosen adjustable clamps on sieve beds Slide clamps down below the adjustable clamp hangers 6 Lift sieve beds up and rest sieve beds on top rear of base N Remove tie wraps clamps and reinforced tubing from bottom of both sieve beds and remove sieve beds CAUTION DO NOT remove plastic caps not shown from new sieve bed fittings until ready to install new sieve beds Severe sieve contamination can occur if uncapped beds are exposed to air 8 Remove plastic caps from the top and bottom fittings of the new sieve beds 9 Install new sieve beds reversing STEPS 3 7 NOTE After replacing sieve beds retiming is necessary Refer to Timing on page 89 10 Run unit and check for leaks Refer to Leak Test on page 93 11 Reinstall cabinet Refer to Removing Cabinet on page 26 Platinum Oxygen 52 Part No 1118396 SECTION 9 SIEVE BEDS Tie Wraps Sieve Bed Qd y Tubing i E ja Y Tie Wraps Regulator _ Adjustable Adjustable Clamps Clamp Hanger Large Reinforced Tubing attaches Tie Wraps HERE not shown Tie Wrap
47. bs O ring Intake Valve Flapper and Valve Keeper Strip NOTE For this procedure refer to FIGURE 7 7 on page 45 1 Install the new valve keeper strip on top of the new intake valve flapper so that the word up is visible Fasten with the valve flapper screw 2 Install the new O ring seating it firmly into the groove with your finger or blunt object NOTE Torque flapper screws to 1 36 N M 12 inch Ibs Piston Cup NOTE For this procedure refer to FIGURE 7 6 on page 44 1 Place cylinder sleeve over rod top with smooth edge facing up 2 Place piston cup on rod top making sure it is centered 3 Place retainer on top of piston cup and seat retainer in rod top pocket making sure the retainer and cup align properly along with screw holes B Insert new screw into rod top and torque to 3 39 30 inch lbs Connecting Rod Eccentric Assembly NOTE For this procedure refer to FIGURE 7 6 on page 44 1 With gloved hand or clean cloth pre form piston cup if necessary 2 Gently slide cylinder sleeve up and over cup to fully form it Platinum Oxygen 46 Part No 1118396 SECTION 7 COMPRESSOR CAUTION DO NOT lubricate or use oil on any moving parts The compressor eccentric uses a precision sealed bearing DO NOT crimp the piston cup when you replace the sleeve If the cup is crimped it MUST be replaced 3 With bearing side of connecting rod eccentric assembly facing the center motor of the compressor
48. ce transformer assembly Refer GREEN Off to Replacing Transformer Assembly on page 70 Control Panel Indicators RED Off YELLOW Off GREEN Off Power Switch ON Continuous audible alarm Unit Operating Internal repairs required Replace SensO circuit board Refer to Internal Status Indicators Replacing P C Board on page 68 RED Three Flashes GREEN One Flash CONTROL PANEL INDICATORS RED Off YELLOW Flashing GREEN On Low Flow Alarm does not System leak Repair leak in product tank regulator activate on flows less than tubing fittings or flow meter 0 5 L min Defective check valves Replace check valves Refer to Replacing NOTE This feature was Check Valves on page 55 added on units manufactured after November 2002 These units can be identified by a single beep on start up NOTE On models manufactured after 09 03 the error code displayed on the internal status indicators is maintained in memory to aid in repairs The error code must be cleared from the unit s memory after fixing the fault The error code memory will be cleared after turning power on and off five times after the fault was detected Part No 1118396 25 Platinum Oxygen SECTION 5 CABINET SECTION 5 CABINET Removing Cabinet A DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing NOTE For this procedure refer to FIGURE 5 1 NOTE This procedure app
49. ck for purity using a calibrated Oxygen Analyzer at Test Point Oxygen Outlet of the concentrator Compressor inlet filter dirty Compressor a Defective Internal Status Indicators RED Two Flashes GREEN One Flash73 Shutdown Control Panel Indicators SensO ONLY RED On YELLOW Off GREEN Off Faulty capacitor Bad motor windings Worn seals For SensO units the RED indicator will signal extremely low purity and will be accompanied by a continuous audible alarm and a system shutdown Repairs are required Bad bearings Leak at fittings or tubing Leaky or defective relief valve Insufficient voltage outlet Heat exchanger a Leak at tubing or body chamber b Inspect tubing and heat exchanger Clean or replace Refer to Cleaning the Cabinet Filter on page 27 Replacing the Outlet HEPA Filter on page 28 and Replacing the Compressor Inlet Filter on page 30 Replace inlet filter Refer to Replacing the Compressor Inlet Filter on page 30 Replace compressor Refer to Replacing Compressor Assembly 5 Liter Models manufactured before July 24 2003 on page 37 Replace capacitor Refer to Replacing Capacitor on page 40 Replace compressor Refer to Compressor on page 37 Rebuild top end of compressor Refer to Rebuilding the Thomas Model 2650 and 2660 Compressor on page 41 Replace compressor Refer to Compressor on page 37 Replace fit
50. connect concentrator from electrical outlet before servicing NOTE For this procedure refer to FIGURE 12 4 on page 63 1 Unplug unit 2 Remove cabinet Refer to Removing Cabinet on page 26 CAUTION Use care not to deform heat exchanger when installing removing or cleaning 3 Loosen the adjustable clamps on the compressor output hose BLUE and valve manifold input hoses at the front and rear of the heat exchanger Platinum Oxygen 62 Part No 1118396 SECTION 12 EXCHANGER 4 Remove screw securing heat exchanger standoff to the soundbox e Move the heat exchanger out to access the hoses and remove them from the heat exchanger Remove heat exchanger Reverse STEPS 3 6 to install new heat exchanger Run unit and check for leaks Refer to Leak Test on page 93 9 o Reinstall cabinet Refer to Removing Cabinet on page 26 Sound Box Screw Heat Exchanger Standoff Heat Exchanger Adjustable Clamp Compressor Outlet Hose BLUE Valve Manifold Input Hose from compressor lt gt FIGURE 12 4 Replacing Heat Exchanger Part No 1118396 63 Platinum Oxygen SECTION 13 CONTROL PANEL SECTION I 3 CONTROL PANEL Removing Control Panel N DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing NOTE For this procedure refer to FIGURE 13 1 on page 65 NOTE This procedure applies
51. connecting rod and discard Piston Cup NOTE For this procedure refer to FIGURE 7 6 1 Remove the torque screw discard from the piston cup retainer retain for reassembly 2 Remove the piston cup discard and wipe debris from the top of the connecting rod with a clean damp cloth NOTE As the piston travels DETAIL BOTTOM OF up and down it will also rock Cylinder Sleeve COMPRESSOR from side to side This is a pue Eccentric Set Screw Access Hole feature of the WOB L Piston Torque Screw Retainer Piston Cu P Rod Top Eccentric Set Screw hidden FIGURE 7 6 Removing Compressor and Compressor Heads Rebuilding the Compressor Fan Connecting Rod and Eccentric Assembly Piston Cup O ring Intake Valve Keeper and Valve Keeper Strip NOTE For this procedure refer to FIGURE 7 7 on page 45 1 Remove the O ring intake valve flapper and valve keeper strip from the bottom of the valve plate Discard all 2 Clean the bottom of the valve plate with a clean soft cloth Platinum Oxygen 44 Part No 1118396 SECTION 7 COMPRESSOR Valve Flapper Screw Valve Keeper Strip O Ring Intake Valve Flapper FIGURE 7 7 Removing Compressor and Compressor Heads Rebuilding the Compressor O ring Intake Valve Keeper and Valve Keeper Strip Exhaust Valve Flapper Valve Restraint and Valve Keeper Strip NOTE For this procedure refer to F
52. e Moti Edu 96 TimeOut Test 97 High Pressure Y YY CU NOA YA y da 97 UG OO NAE EEE eens 98 P E Valve Coll Testa ee Y WY WY fd Y YF YD yd 99 SECTION 25 SIEVE BED PRESSURE CHECK 000000000000000000000000000000000 101 Models without HomeFill Connection On Back Of Unit see 101 Installing the Sieve Bed Pressure Tap In Kit and Checking Bed Pressure 101 Pressure Testing 102 After Testihg icon eet teris tetur TF YF FO Y GYR cres 103 Models With HomeFill Connection On Back Of Unit serere 104 Pressure quete fitu eni 104 LIMITED WARRANTY 0000000000000000000000000000000000000000000000000000000000000000000 107 Platinum Oxygen 5 Part No 1118396 SPECIAL NOTES SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage Refer to the table below for definitions of the signal words SIGNAL WORD MEANING Danger indicates an imminently hazardous situation which DANGER OD if not avoided will result in death or serious injury Warning indicates a potentially hazardous situation which if WARNING not avoided could result in death or serious
53. eat Exchanger Platinum 5 Models Manufactured Before 9 9 2005 Platinum Oxygen 60 Part No 1118396 SECTION 12 EXCHANGER Platinum 5 Models Manufactured After 9 9 2005 N DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing NOTE For this procedure refer to FIGURE 12 2 on page 62 1 Unplug unit 2 Remove cabinet Refer to Removing Cabinet on page 26 CAUTION Use care not to deform heat exchanger when installing removing or cleaning 3 Remove compressor assembly Refer to Replacing Compressor Assembly 5 Liter Models manufactured before July 24 2003 on page 37 or Replacing Compressor Assembly 5 Liter Models manufactured after July 24 2003 on page 38 4 Loosen hose clamp and separate heat exchanger output fitting from hose elbow 5 Remove heat exchanger from the soundbox assembly 6 Reverse STEPS 3 6 to install new heat exchanger assembly 7 Run unit and check for leaks Refer to Leak Test on page 93 8 Reinstall cabinet Refer to Removing Cabinet on page 26 Part No 1118396 61 Platinum Oxygen SECTION 12 EXCHANGER Heat Exchanger Cushioned Clamp Tube Connection end Exchanger Outlet Compression Fitting Nut FIGURE 12 2 Replacing Heat Exchanger Platinum 5 Models Manufactured After 9 9 2005 FIGURE 12 3 Platinum 9 Concentrators A DANGER To prevent electrical shock ALWAYS dis
54. en locknut when installing it onto pilot valve stem Damage to the plastic manifold may occur 19 Reinstall control panel Refer to Removing Control Panel on page 65 A WARNING Ensure that hoses behind control panel DO NOT contact cooling fan after control panel reinstallation 20 Plug power cord in and turn concentrator ON I to ensure proper operation 21 Reinstall cabinet Refer to Removing Cabinet on page 26 22 If any difficulties are encountered in this installation contact Invacare Technical Service at 800 832 4707 Part No 1118396 85 Platinum Oxygen SECTION 20 4 WAY VALVE Manifold Oben O Ring pening Air Flow Passage DETAIL A Spade Connectors attach here NOTE DO NOT Leave Wire Connectors Intact Locknut disassemble Coil Pilot Valve Slotted Stem Poppet Valve sere zn End Yoke Manifold O Ring New Opening Label FIGURE 20 5 Replacing Pilot Valve Poppets and O Rings 5 Liter Manufactured Before 8 1 05 DETAIL Manifold Opening Air Flow Passage O Ring Spade Connectors attach here Manifold O Ring NOTE DO NOT Leave Wire Connectors Intact Opening disassemble Slotted Locknut Yoke Poppet rr cu Pilot cc T N Coil Stem S Tapered New ds End Pilot Valve Label Stem FIGURE 20 6 Replacing Pilot Valve Poppets and O Rings 5 Liter Manufactured After 8 1 05 and 9 Liter
55. entrator is used in NOTE There are two cabinet filters one located on each side of the cabinet 1 Remove each filter and clean at least once a week depending on environmental conditions NOTE Environmental conditions that may require more frequent cleaning of the filters include but are not limited to high dust smoking air pollutants etc Cabinet Filter located on both sides of 2 Clean the cabinet filters with a vacuum concentrator cleaner or wash in warm soapy water and rinse thoroughly d Dry the filters thoroughly before reinstallation CAUTION DO NOT operate the concentrator without the filters installed FIGURE 6 1 Cleaning the Cabinet Filter Part No 1118396 27 Platinum Oxygen SECTION 6 PREVENTIVE MAINTENANCE Replacing the Outlet HEPA Filter Check the Outlet HEPA Filter NOTE For this procedure refer to FIGURE 6 2 on page 29 NOTE The outlet HEPA filter can be checked by performing the following procedure 1 Turn the concentrator on and adjust the flowmeter to the maximum flow of the unit 2 Observe the flowmeter s flow indicator while connecting a 50 foot 15 2 meters cannula tube to the outlet barb of the concentrator not shown 3 If the flow indicator fluctuates the outlet HEPA filter may need replacement Refer to Troubleshooting on page 18 NOTE To replace the Outlet HEPA filter perform this procedure as needed depending on the Outlet HEPA filter check resu
56. er tubing disconnected or Canister Exhaust Muffler Platinum 5 damaged Manufactured before 7 24 03 on page 3l b Reconnect or replace tubing Refer to Inspecting Replacing Throttling Muffler Platinum 5 AW Manufactured after 7 24 03 on page 32 Compressor inlet filter missing and or a Replace compressor inlet filter Refer orange sticker removed to Replacing the Compressor Inlet Filter on page 30 Compressor removed a Replace Compressor Refer to Replacing Compressor Assembly 5 Liter Models manufactured before July 24 2003 on page 37 Incorrect style of inlet filter aftermarket Replace with factory OEM sound reduced style inlet HEPA filter Refer to Replacing the Outlet HEPA Filter on page 28 Part No 1118396 23 Platinum Oxygen SECTION 4 TROUBLESHOOTING SYMPTOM PROBABLE CAUSE SOLUTION Place unit at least 7 5 cm 3 inches from Unit Overheats Fan a Leads to fan disconnected b Defective fan c Fan installed upside down Heat exchanger a Dirty or plugged b Damaged Compressor a Defective Faulty capacitor Bad motor windings Worn seals Bad bearings Line voltage excessive surge Oxygen Purity P C board defective Good Internal Status Indicators Control Panel Indicators RED Off YELLOW On GREEN On After 30 minutes of run time unit operates normally oxygen purity within normal range GREEN or YELLOW panel indica
57. he unit unplugged and inoperable for a short period of time the battery free circuit will be drained If power failure alarm does not sound with unit unplugged and power switch on the battery free circuit is drained It will recharge when unit is plugged in and switched on Low Pressure Test NOTE For this procedure refer to FIGURE 24 1 on page 97 NOTE This procedure applies to Platinum 5 and Platinum 9 models 1 There are two separate failure modes for Low Pressure Low producttank pressure Pressure in tank drops below a preset value typically 48 kPa 7 p s i e Failure to reach the set point pressure within a prescribed time limit or Timeout Failure Test One 1 With unit running set flow at maximum rating 2 When the main valve switches pull the stem on the compressor relief valve out as far as it will go and hold it 3 The low pressure alarm should activate within 30 seconds Refer to SensO2 Indicators on page 99 for shutdown mode Test Two 1 With unit running set flow at maximum rating 2 Remove the pressure sensor tubing tie wrap from the top of the product tank not shown Platinum Oxygen 96 Part No 1118396 SECTION 24 ALARM TEST 3 The low pressure alarm should activate within 30 seconds Refer to SensO2 Indicators on page 99 for shutdown mode 4 Replace tubing and tie wrap Time Out Test NOTE For this procedure refer to FIGURE 24 1 NOTE This procedure applies to Platinum 5 and Pl
58. hose to bottom center port of manifold assembly 14 Remove valve manifold input hose from bottom center port of manifold assembly 15 Slide 4 way valve manifold assembly complete through the cutout in top of sound box then install the two mounting screws washer and bushings into the manifold to secure the 4 way valve assembly to the top of sound box 16 Reverse STEPS 4 10 17 Reinstall control panel Refer to Removing Control Panel on page 64 18 Plug unit in and turn power On 0 Platinum Oxygen 82 Part No 1118396 SECTION 20 4 WAY VALVE 19 Run unit and check for leaks at hose connections Refer to Leak Test on page 93 20 Check for proper system operation Refer to Typical Product Parameters on page 7 21 Re install cabinet Refer to Removing Cabinet on page 26 Sieve Bed Hoses Tie Wraps Valve Pilot Spade Connectors Manifold Assembly Adjustment Clamp 4 Way Valve Y tube Muffler Assembly Valve Manifold Input Hose DETAIL A MANIFOLD ASSEMBLY Tie Wrap Throttle Muffler Bushing Two 2 per side Sound Box Top FIGURE 20 4 Replacing Manifold Assembly 5 Liter Models Manufactured after 7 24 03 and 9 Liter Models Part No 1118396 83 Platinum Oxygen SECTION 20 4 WAY VALVE Replacing Pilot Valve Poppets and O Rings A DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing NOTE For this pr
59. ification change setting to a number one position lower than the previous setting Let unit run ten minutes If concentration is lower still change the switch to number two positions higher NOTE You will find that the concentrator will perform better at either the higher or lower setting The normal adjustment range is between timing positions 3 and A 10 Continue to change setting one step at a time up or down until the concentration is the highest value achievable 11 Once concentration has reached specification retiming is complete Refer to Typical Product Parameters on page 7 12 Reinstall cabinet Refer to Removing Cabinet on page 26 Part No 1118396 91 Platinum Oxygen SECTION 22 TIMING P C Board Timing Switch Adjustment Pressure Sensor FIGURE 22 2 Manual Tuning Platinum Oxygen 92 Part No 1118396 SECTION 23 LEAK TEST SECTION 23 LEAK TEST A DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing Leak Test NOTE For this procedure refer to and FIGURE 23 1 on page 94 and FIGURE 23 2 on page 95 NOTE This procedure applies to Platinum 5 and Platinum 9 models 1 Unplug unit Remove cabinet Refer to Removing Cabinet on page 26 Plug in unit Turn the concentrator on 1 e oo N Let concentrator run for 20 minutes A DANGER Apply small amount of leak test solution to fittings
60. injury Caution indicates a potentially hazardous situation which if not avoided may result in property damage CAUTION NOTICE The information contained in this document is subject to change without notice A DANGER DO NOT SMOKE while using this device Keep all matches lit cigarettes or other sources of ignition out of the room in which this product is located and away from where oxygen is being delivered NO SMOKING signs should be prominently displayed Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air Failure to observe this warning can result in severe fire property damage and cause physical injury or death Platinum Oxygen 6 Part No 1118396 TYPICAL PRODUCT PARAMETERS TYPICAL PRODUCT PARAMETERS Electrical Requirements Rated Current Input Sound Level Oxygen Output Concentration Levels All 5LXO2AWO 5LXAW 5LXO2AW models Stated concentration levels achieved after initial warm up period approximately 30 minutes Maximum Outlet Pressure Flow Range Low Flow Alarm Average Power Consumption Pressure Relief Mechanism Operational at Change in maximum recommended flow when back pressure of 7k Pa is applied Part No 1118396 230 VAC 10 253 VAC 207 VAC 50 Hz 1 4 A IRCSLXO2AW OQ 2 0 A IRC5LXAW IRC5LXO2AW 2 5 A IRCOLXO2AWO 39 5 dB Average IRC5LXO2AWO 44 dB Average IRC5LXAW IR
61. lacing Capacitor NOTE For this procedure refer to Detail A in FIGURE 7 4 on page 41 NOTE This procedure applies to Platinum 5 and Platinum 9 models 1 Unplug the unit 2 Remove the cabinet Refer to Removing Cabinet on page 26 Platinum Oxygen 40 Part No 1118396 SECTION 7 COMPRESSOR Remove protective boot to reveal capacitor wires Disconnect the spade connectors from the capacitor terminals not shown Remove the screw that secures the capacitor clamp to the soundbox brace Remove the capacitor from the concentrator base assembly Install new capacitor by reversing STEPS 3 6 B Re install the cabinet Refer to Removing Cabinet on page 26 Protective Boot d D Soundbox Brace FIGURE 7 4 Replacing Capacitor Rebuilding the Thomas Model 2650 and 2660 Compressor N DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing This kit is designed for rebuilding the Thomas Model 2660 Series compressor The serial number is located on the front of the compressor If the serial number does not begin in this way this kit will not work The 5 Liter Thomas 2660 Series Compressor Rebuild Kit contains the following DESCRIPTION OUANTITY Teflon Piston Cups 2 Aluminum Piston Sleeves Flapper Valves Intake and Exhaust Valve Keepers Bod A N O Ring Sleeve Part No 1118396 41 Platinum Oxygen SECTION 7 COM
62. latinum Oxygen Concentrators when purchased new and unused to be free from defects in materials and workmanship for a period of five 5 years from date of purchase from Invacare or a dealer with a copy of the seller s invoice required for coverage under this warranty DO NOT OPEN OR ATTEMPT TO SERVICE THE 4 WAY VALVE BODY This will void any and all warranty applicable to the valve body If within such warranty periods any such product shall be proven to Invacare Corporation s satisfaction to be defective such product shall be repaired or replaced at Invacare Corporation s option This warranty only applies to the labor for repairs performed by the Invacare Service Department or Invacare Authorized Service Centers It does not apply to the labor performed by the purchaser or user This warranty does not include normal wear and tear or shipping charges Invacare Corporation s sole obligation and your exclusive remedy under this warranty shall be limited to such repair or replacement Routine maintenance items such as filters are excluded from this warranty For warranty service please contact Invacare Corporation s Service Department at the toll free number on the back cover Upon receiving notice of an alleged defect in a product Invacare Corporation will issue a serialized return authorization It shall cthen be the responsibility of the purchaser to return the entire unit or remove at purchaser s cost the defective component part s identified
63. le timer is used to control the shift time of the pressure equalization P E valve The SensO board allows for automatic microprocessor controlled adjustments of the P E shift times Non SensO boards allow for manual adjustment Autotuning NOTE For this procedure refer to FIGURE 22 1 on page 90 NOTE This procedure applies to Platinum 5 and Platinum 9 models 1 Turn power off O and unplug the unit 2 Remove cabinet Refer to Removing Cabinet on page 26 3 Plug the unit in and turn the power on I 4 Wait for the main valve shift to indicate one of the following Ifthe board is new the front panel RED indicator light will slowly blink indicating autotune has initiated e Ifthe board is used and has performed autotuning before press and release the small push button switch near the pressure sensor e The front panel RED indicator light will slowly blink indicating the unit is autotuning and automatically adjusting the P E valve timing 5 Setthe output flow to 5 0 L min during autotuning NOTE The GREEN onboard diagnostic light will flash the current value of the P E valve timing number 6 Place the cover back on the unit but DO NOT install the cover screws Let the unit run until autotuning is completed NOTE When autotuning is complete the RED front panel indicator light will change from a slow blinking to a flashing of the final P E valve timing number 7 Turn the unit off and then back on The
64. leaks will cause system pressures to remain below adequate shift exhaust pressures and will cause compressor shutdown Unit plugged in power Defective Parts Defective main valve coil a Replace main valve Refer to Replacing Internal Status Indicators 4 Way Valve on page 77 RED One Flash GREEN b Replace main valve coil Refer to a Flashes Replacing 4 Way Valve on page 77 RED Two Flashes Reattach connector GREEN Three Flashes Defective P E valve coil Resistance 80 Replace P E coil if shorted or open on Control Panel Indicators ohm s 10 resistance check Refer to Replacing P E RED On Valve on page 50 YELLOW Off Defective PC Board Replace PC Board GREEN Off Defective P E valve coil Resistance 80 Replace P E coil if shorted or open on Unit plugged in power ohm s 10 resistance check Refer to Replacing P E switch on continuous Valve on page 50 audible alarm Compressor shut down Connector loose Reattach connector Defective P C board Replace P C board Refer to Replacing P C Board on page 68 Unit Not Operating P C board Failure Replace P C board Refer to Replacing P C Alarm On or Off Board on page 68 Internal Status Indicators RED Two Flashes GREEN Four Flashes Or RED Two Flashes GREEN Five Flashes Platinum Oxygen 20 Part No 1118396 SECTION 4 TROUBLESHOOTING SYMPTOM PROBABLE CAUSE SOLUTION Cabinet filters dirty Low Concentration NOTE Che
65. lectrical shock ALWAYS disconnect concentrator from electrical outlet before servicing CAUTION Only qualified service technicians should make any changes to this equipment Follow these pre cautions to prevent damage to the P C boards Before handling any P C boards you need to be properly grounded to prevent static damage to the components of the board A Static Cuff MUST be worn and properly grounded using an alligator clip Electrical conduit or a water pipe is normally sufficient when a known good ground is not available Care should be taken to ensure that the alligator clip contacts with bare metal surface When removing quick disconnects terminals DO NOT pull on wire itself as damage to the connection may occur Hold down the P C board with one hand and use an upward force with a slight rocking motion to remove the terminals Before installing any P C boards ensure that all insulators are in place NOTE For this procedure refer to FIGURE 15 1 on page 69 NOTE This procedure applies to Platinum 5 and Platinum 9 models 1 Unplug unit Remove cabinet Refer to Removing Cabinet on page 26 Remove control panel Refer to Removing Control Panel on page 64 2 N Disconnect the following A Wiring Harness from nine or ten pin wiring harness connector B Wiring Harness from eight pin wiring harness connector C Transformer harness connector AN WARNING DO NOT remove sensor tubing from P C Board Remove fr
66. lies to both Platinum 5 and Platinum 9 models 1 Unplug unit 2 Remove the four mounting screws that secure cabinet assembly to the base assembly 3 Liftthe cabinet straight up NOTE When required vacuum inside of the cabinet and exposed foam insulation 4 To re install cabinet reverse STEPS 2 3 Cabinet Assembly Mounting Screws ounting Screws x FIGURE 5 1 Removing Cabinet Base Assembly Platinum Oxygen 26 Part No 1118396 SECTION 6 PREVENTIVE MAINTENANCE SECTION 6 PREVENTIVE MAINTENANCE A DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing NOTE The Platinum Concentrators are specifically designed to minimize routine preventive maintenance at intervals of once per year In places with high dust or soot levels maintenance may need to be performed more often The following MUST be performed at a minimum of one year in service to assure years of additional reliability Refer to Troubleshooting on page 18 for plugged filter symptoms Only qualified technicians should perform preventive maintenance on the concentrator Power MUST be disconnected before beginning preventive maintenance on the Invacare concentrator Cleaning the Cabinet Filter NOTE For this procedure refer to FIGURE 6 1 NOTE This procedure applies to Platinum 5 and Platinum 9 models NOTE Perform this procedure as needed depending upon the environment the conc
67. lts Replace the Outlet HEPA Filter NOTE For this procedure refer to FIGURE 6 2 on page 29 NOTE Perform this procedure as needed depending on the outlet HEPA filter check results 1 Unplug the unit 2 Locate the filter access and humidifier adapter storage plate located on the side of the concentrator 3 Insert a flathead screwdriver in the plate groove on the top edge of the plate and gently pry the humidifier adapter storage plate open Detail A 4 Remove plate and set aside Grasp the existing outlet filter and disconnect from the short silicone tube Detail B Disconnect existing outlet HEPA filter from the long silicone tube Detail B Discard existing outlet HEPA filter Attach new outlet HEPA filter to short and long silicone tubes 290 LINE SON Reinstall filter access and humidifier adapter storage plate Platinum Oxygen 28 Part No 1118396 SECTION 6 PREVENTIVE MAINTENANCE DETAIL Plate Groove Filter Access Humidifier Adige Storage Plate DETAIL Long Silicone Tube Long Silicone Tube lt Compressor Ss Inlet Filter Humidifier Bottle Adapter Short Storage Silicone location Tube Outlet HEPA Filter Short Silicone Tube located underneath Outlet HEPA filter FIGURE 6 2 Replace the Outlet HEPA Filter Replace the Outlet HEPA Filter NOTE For this procedure refer to FIGURE 6 3 on page
68. lve 4 Way Valve Face Mounting Screws NOTE The Numbers indicate the mounting screw torque sequence pre torque to 1 1 N M 0 2 10 in Ibs 2 in lbs then torque to 2 5 0 2 22 in lbs 2 in lbs FIGURE 20 2 Replacing 4 Way Valve 5 Liter Models Manufactured After 8 1 05 and 9 Liter Models Platinum Oxygen 78 Part No 1118396 SECTION 20 4 WAY VALVE Replacing Manifold Assembly 5 Liter Models Manufactured before 7 24 03 NOTE For this procedure refer to FIGURE 20 3 on page 81 NOTE This procedure applies to Platinum AW units manufactured before 7 24 03 Units manufactured before 7 24 03 are equipped with an exhaust canister with exhaust hose as shown in FIGURE 20 3 Tools Required Diagonal Cutters e Flathead Screwdriver e inch ratchet 4 inch socket and V4 inch extension e Pliers e Tie wraps Unplug unit Remove cabinet Refer to Removing Cabinet on page 26 Remove control panel Refer to Removing Control Panel on page 64 pet Sot eS Cut and remove the tie wraps connecting the sieve bed hoses to top of manifold assembly Detail A NOTE When removing the sieve bed hose from the manifold assembly use a flat head screwdriver to pry hose up while simultaneously pulling the hose upward 5 Remove sieve bed hoses from manifold assembly Detail A 6 Remove and label the two spade connectors from the valve pilot on top of the manifold assembly Detail A 7 Remo
69. ndicator Light O Between 73 3 to 85 2 YELLOW Indicator light A YELLOW Solid B YELLOW Flashing Sensor Failure Call a qualified technician P E Valve Coil Test SYSTEM FAILURE Below 73 3 8 inch RED light ELLOW light GREEN light Continuous Audible Alarm Sieve Gard Compressor shutdown Call qualified technician D Oxygen Sensor Tubing Clamp Off Flow FIGURE 24 3 Oxygen Sensor NOTE For this procedure refer to FIGURE 24 4 NOTE P E Valve Coil Alarm sequence occurs when the P E Valve Coil connection or P C Board Circuitry has failed 1 With the unit running and flow set at 5 L min for 5 Liter models and 9 L min for 9 Liter models remove one YELLOW wire from the valve coil The valve coil alarm should activate within 10 seconds 2 Reinstall cabinet Refer to Removing Cabinet on page 26 Part No 1118396 99 Platinum Oxygen SECTION 24 ALARM TEST NOTE If any alarm fails to perform to specification contact Invacare Technical Service NOTE If problems arise during test contact Invacare Technical Service for further assistance E Valve Coil YELLOW Wires FIGURE 24 4 Valve Coil Test 3 Refer to Preventive Maintenance Record Invacare Platinum Oxygen Concentrator on page 36 to record date and number of hours when preventive maintenance was performed
70. ng at barbed fitting while pulling tubing away from manifold assembly Reverse STEPS 13 16 to re assemble Slide 4 way valve manifold assembly complete through the cutout in top of sound box then install the two mounting screws washer and four bushings into the manifold to secure the 4 way valve assembly to the top of sound box Detail B Reattach components by reversing STEPS 4 10 Reinstall control panel Refer to Removing Control Panel on page 64 Plug unit in and turn power On 1 Run unit and check for leaks at hose connections Refer to Leak Test on page 93 23 24 Check for proper system operation Refer to Typical Product Parameters on page 7 Re install cabinet Refer to Removing Cabinet on page 26 Platinum Oxygen 80 Part No 1118396 SECTION 20 4 WAY VALVE DETAIL DETAIL Sieve Bed Hoses i Mounting EP 5 Screw Tie Wraps STEP 4 Valve Pilot 4 way Valve Sound Box Top Manifold Assembly Mounting Detail for Manifold Assembly STEP 11 Bottom Center Port Adjustable Clamp STEP 15 Valve Manifold DETAIL Valve Manifold Input Hose Heat Exchanger Input Hose STEPS 9 10 amp 16 To Valve Manifold nput Hose Mounting Screw Placement Exhaust Canister c w Exhaust Hose A STEP 14 Exhaust Muffler FIGURE 20 3 Replacing Manifold Assembly 5 Liter Models Manufactured before 7 24 03 Part No 1118396 81
71. nt as described in the Owner s Manual DO NOT connect the concentrator in parallel or series with other oxygen concentrators or oxygen therapy devices Avoid creation of any spark near medical oxygen equipment This includes sparks from static electricity created by any type of friction Part No 1118396 13 Platinum Oxygen SECTION 2 INSTALLATION SEQUENCE OF OPERATION SECTION 2 INSTALLATION SEQUENCE OF OPERATION Installation Verification of Battery Free Power Loss Alarm When your new Platinum concentrator arrives it should be checked for proper operating conditions 1 If theunit has been in below freezing temperatures allow it to warm up to room temperature before operating 2 The concentrator may need to be turned on for 4 to 5 seconds to charge the Battery Free Power Loss Alarm Connect power cord to outlet and turn the concentrator on Turn flow control knob counterclockwise and flow will begin IMMEDIATELY Set flow rate to five L min Let the unit run for 30 minutes then turn unit off 3 Unplug the power cord and press on off VO switch to the On I position An intermittent audible alarm will sound This confirms proper operation of the Battery Free Power Loss Alarm Turn On Off VO switch Off O 4 Connect power cord to outlet and turn on concentrator Unit will beep on start up 5 Check the oxygen concentration per specifications after 30 40 minutes running time Sequence of Operation Turning on I
72. ocedure refer to FIGURE 20 5 and FIGURE 20 6 on page 86 NOTE This procedure applies to Platinum 5 and Platinum 9 models 1 Unplug unit 2 Remove cabinet Refer to Cabinet on page 26 3 Remove control panel Refer to Removing Control Panel on page 64 4 Locate pilot valve on 4 way valve manifold assembly NOTE Pilot valve can be accessed while the 4 way valve and sieve bed hoses are intact on the concentrator NOTE DO NOT remove spade connectors from pilot valve coils Simply lay coils with spade connectors aside while performing this installation 5 Hold pilot valve stem with the flat blade screwdriver and turn the 9 16 inch locknut counterclockwise one complete turn 6 Unscrew the pilot valve stem assembly from the manifold while leaving the coil with yoke and locknut intact on stem 7 Remove pilot valve poppet from inside the pilot valve stem CAUTION DO NOT remove the coil yoke from the coil The washer between the bottom of the yoke and the bottom of the coil MUST not be removed Otherwise components may be damaged in reassembly 8 Set aside the coil with yoke and pilot valve stem assembly with the wires still intact A WARNING DO NOT use sharp tools to remove O ring from plastic manifold opening Damage to the plastic manifold and or plastic manifold airflow passage may occur Damage to airflow may result in improper oxygen delivery and harm to the user Refer to Detail in FIGURE 20 6 9
73. om tee fitting or regulator ONLY Damage to sensor assembly could occur A faulty sensor may result in improper oxygen delivery and harm to the user D Pressure sensor tubing from product tank not P C Board E If applicable Oxygen Sensor tubing from Regulator Fitting Refer to Replacing Regulator on page 57 Platinum Oxygen 68 Part No 1118396 SECTION 15 BOARD 5 Remove the two mounting screws on top corners of P C board Bend latching tab up to release P C board from control panel Position new P C board 7 Reverse STEPS 4 and 5 Be sure all tie wraps removed from tubing connections are replaced 8 Reinstall control panel Refer to Removing Control Panel on page 64 9 Run concentrator to ensure unit operates to specifications 10 After replacing P C board retiming may be necessary Refer to Timing on page 89 11 Check for leaks at pressure sensor tubing on product tank cap Refer to Leak Test on page 93 12 Reinstall cabinet Refer to Removing Cabinet on page 26 Control Panel Oxygen Sensor Tubing SensO2 models only Pressure Sensor 5 4 Tubing STEP 4D Mounting Screws STEP 5 Transformer Harness Connector STEP 4C Nine or Ten Pin Wiring Harness Connector STEP 4A 8 Pin Wiring Harness Connector STEP 4B FIGURE 15 1 Replacing P C Board Part No 1118396 69 Platinum Oxygen SECTION 16
74. or Assembly 9 Liter Models eene 39 Replacing 40 Rebuilding the Thomas Model 2650 2660 41 Removing Compressor and Compressor Heads sentent 42 Rebuilding the Compressor sis aeri etre tee Cydia 46 SECTION 8 VALVE eee eeeoessoooeesecscccssessscsccosssecccssccosecececcos DO Replacing PE i 50 SECTION 9 SIEVE BEDS 5 2 Replacing Seye a YR dd FO 52 SECTION 10 CHECK VALVES D 5 Replacing Check y 55 SECTION 11 REGULATOR D 7 Replacing Regulate m 57 PRPS ENB SSC ACO Pt meret sede quida atu editt ond deseen ufui 58 SECTION 12 HEAT EXCHANGER OO Replacing Heat Exchanger 245i A a 60 Platinum 5
75. pack the component part s in a manner to avoid shipping damage and to ship the component part s to either Invacare Corporation s plant or service center as specified by Invacare Corporation in advance Defective component part s MUST be returned for warranty inspection using the serial number as identification within thirty 30 days of return authorization date DO NOT return products to our factory without prior consent C O D shipments will be refused please prepay shipping charges LIMITATIONS AND EXCLUSIONS THE FOREGOING WARRANTY SHALL NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE ACCIDENT IMPROPER OPERATION MAINTENANCE OR STORAGE SOOT OR SMOKE FILLED ENVIRONMENTS OR OTHER THAN NORMAL APPLICATION USE OR SERVICE OR TO PRODUCTS MODIFIED WITHOUT INVACARE CORPORATION S EXPRESS WRITTEN CONSENT INCLUDING BUT NOT LIMITED TO MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS OR TO PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE CORPORATION OR TO PRODUCTS DAMAGED BY CIRCUMSTANCES BEYOND INVACARE CORPORATION S CONTROL THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER WARRANTIES WHATSOEVER WHETHER EXPRESS OR IMPLIED INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS C
76. page 40 gt N S Disconnect and remove spade connectors from fan terminals on back side of cooling fan not shown 6 Remove four mounting screws securing the cooling fan to the sound box 24 Remove existing cooling fan CAUTION Fan MUST be positioned properly so air from the fan blows DOWN onto the compressor see air flow arrow on back of fan or damage to the unit will occur 8 Install new cooling fan NOTE Run the self threading mounting screws through the fan housing holes to cut the threads and remove them before mounting the fan for ease of installation 9 Reinstall compressor Refer to Replacing Compressor Assembly 5 Liter Models manufactured before July 24 2003 on page 37 or Replacing Capacitor on page 40 10 Reinstall control panel Refer to Removing Control Panel on page 64 11 Plug unit in and check for leaks Refer to Leak Test on page 93 12 Run unit for at least 30 minutes If unit runs properly replace cover If not call Invacare Technical Service 13 Reinstall cabinet Refer to Cabinet on page 26 Platinum Oxygen 66 Part No 1118396 SECTION 14 COOLING FAN Fan Terminals Hidden from view Spade Connectors not shown Cooling Fan Sound Box Mounting Screws Heat Exchanger FIGURE 14 1 Replacing Cooling Fan Part No 1118396 67 Platinum Oxygen SECTION 15 P C BOARD SECTION 15 P C BOARD Replacing P C Board A DANGER To prevent e
77. plies to Platinum 5 and Platinum 9 models 1 Unplug unit Remove cabinet Refer to Removing Cabinet on page 26 Remove control panel Refer to Removing Control Panel on page 64 Remove and code the four spade connectors from back of existing on off switch Qu gms po je Compress retaining grips on back of existing on off switch and push switch out through front of control panel NOTE Ensure proper orientation of the On Off VO switch and wiring connections before installation Refer to FIGURE 17 1 Detail A and Detail B below CAUTION DO NOT install the On Off I O switch upside down Universal Off O symbol should be at bottom and Universal On symbol should be at the top Possible damage to the concentrator may result if not properly installed Remove switch cover from old part and place it on new switch prior to installation Reverse STEP 4 to secure new On Off VO switch Reinstall control panel Refer to Removing Control Panel on page 64 S ON Reinstall cabinet Refer to Removing Cabinet on page 26 Part No 1118396 71 Platinum Oxygen SECTION 17 ON OFF SWITCH On en 7 KSI DETAIL or 64 s On Off Switch Switch Spade Connector from Brown Wire attached here DETAIL Brown Jumper Wire From Circuit Breaker Spade Connectors Back View from Black Wires attached here FIGURE 17 1 Replacing On Off Switch Platinum Oxygen 72 Part No 1118396
78. r shutdown High Pressure Alarm w compressor shutdown Low Pressure Alarm w compressor shutdown Battery Free Power Loss Alarm SensO Oxygen System SensO Model Low Flow Alarm Weight For IRC5LXAW IRC5LXO2AW IRC5LXO2AWO 23 6 kg kg 52 Ibs 2 Ibs For IRC9LXO2AWO 24 0 kg kg 53 Ibs 2 Ibs Shipping Weight For IRCSLXAW IRC5LXO2AW IRC5LXO2AWO 26 8 kg kg 59 Ibs 2 Ibs For IRC9LXO2AWO 27 2 kg kg 60 Ibs 2 Ibs Operating Ambient Temperature 10 C 35 C 50 F 95 F at 20 60 relative humidity Oxygen Output Less than Ambient 4 F 2 C Cabinet Impact Resistant flame retardant plastic cabinet that conforms to UL 94 VO Regulatory Listing 5LXAW 5LXO2AW 5LXO2AWO ETL certified complying with EN55001 I EN61000 3 2 EN61000 3 3 IEC 60601 1 Al A2 IEC 60601 1 2 5LXO2AW 5LXO2AWO IRC9LXO2AWO Models Only 1508359 MDD CE marked models 9LXO2AWGQ 5LXO2AW 5LXO2AWO Placement No closer than 7 5 cm 3 inches from any wall furniture draperies or similar surfaces Tubing For IRCSLXAW IRC5LXO2AW IRCS5LXO2AWO 7 ft cannula with a maximum 50 ft of Crush Proof Tubing DO NOT pinch For IRC9LXO2AWO Recommended use up to 50 ft high flow tubing with high flow cannula at all flow rates Relative Humidity 20 to 60 Time of Operation Up to 24 hours per day Recommended Storage and Shipping 29 C to 65 C 20 F to 150 F at 15 95 relative humidity Temperature SENSO INDICATORS
79. refer to FIGURE 7 1 on page 38 and FIGURE 7 2 on page 39 1 Unplug unit Remove the cabinet Refer to Removing Cabinet on page 26 Loosen clamp and disconnect intake hose from compressor assembly Disconnect the compression fitting nut that connects the heat exchanger inlet to compressor fitting elbow not shown g Remove compressor wires from wire clamps or tie wraps and disconnect motor connector from the main harness connector not shown NOTE Motor connector goes through grommet before reaching the main harness connector 6 Remove protective boot on capacitor NOTE Reuse the protective boot if the replacement parts do not provide one 7 Disconnect capacitor wires not shown from the top of the capacitor 8 Remove the two mounting screws that secure the compressor restraint bracket and remove the bracket 9 Tilt compressor assembly toward front of unit and lift out 10 Reverse STEPS 3 9 to install new compressor assembly 11 Reinstall the cabinet Refer to Removing Cabinet on page 26 12 Run unit and inspect for leaks Refer to Leak Test on page 93 Part No 1118396 37 Platinum Oxygen SECTION 7 COMPRESSOR Grommet Intake Hose Z Heat Exchanger Mounting Screws LE 7 amp 7 Canister Muffler Wire Clamps or Protective Capacitor Tie wraps Terminal Boot Capacitor Compressor Restraint Bracket FIGUR
80. rottling muffler manufactured after 7 24 03 Restraint Bracket FIGURE 7 2 Replacing Compressor Assembly 5 Liter Models manufactured after July 24 2003 Replacing Compressor Assembly 9 Liter Models NOTE For this procedure refer to FIGURE 7 3 on page 40 1 2 3 Unplug unit Remove the cabinet Refer to Removing Cabinet on page 26 Loosen clamps and disconnect intake hose and compressor output hose blue from compressor assembly Remove compressor wires from wire clamps or tie wraps and disconnect motor connector from the main harness connector Part No 1118396 39 Platinum Oxygen SECTION 7 COMPRESSOR 5 Remove protective boot on capacitor 6 Disconnect capacitor wires from the top of the capacitor NOTE Re use the protective capacitor terminal boot if the replacement parts DO NOT provide one 7 Remove the two mounting screws that secure the compressor restraint bracket and remove the bracket 8 Tilt compressor assembly toward front of unit and lift out 9 Reverse STEPS 3 8 to install new compressor assembly 10 Reinstall the cabinet Refer to Removing Cabinet on page 26 11 Run unit and inspect for leaks Refer to Leak Test on page 93 Motor Main Harness Connection Compressor Output Hose BLUE Intake Protective Boot Capacitor Mi NL Compressor Mountin 5 Restraint Bracket Screws FIGURE 7 3 Replacing Compressor Assembly 9 Liter Models Rep
81. s not shown FIGURE 9 1 Replacing Sieve Beds 5 Liter models manufactured before 1 27 05 Part No 1118396 53 Platinum Oxygen SECTION 9 SIEVE BEDS Tie Wraps P E Valve Assembly Sieve Bed Sieve Bed Adjustable Clamp Hanger Adjustable Clamps Reinforced Tubing attaches Tie Wraps Clamps Here not shown not shown FIGURE 9 2 Replacing Sieve Beds 5 Liter models manufactured after 1 27 05 and 9 Liter models Platinum Oxygen 54 Part No 1118396 SECTION 10 VALVES SECTION I 0 CHECK VALVES Replacing Check Valves N DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical out let before servicing NOTE For this procedure refer to FIGURE 10 1 on page 56 NOTE This procedure applies to Platinum 5 and Platinum 9 models 1 Unplug unit 2 Remove cabinet Refer to Removing Cabinet on page 26 NOTE Replace one or both check valves when performing this procedure The check valves are one way directional and can be checked by passing air through them Air should flow in one direction only gt Loosen adjustable clamp on product tank slide clamp down below the adjustable clamp hanger and pull product tank forward to access check valve s B Remove tie wraps from top and bottom of the check valve s and disconnect check valve s from 4 inch PVC tubing AN WARNING The check valves are one way directional and MUST be installed
82. s not have to be removed 12 Remove control panel and reverse STEPS 3 10 to re install control panel 13 Run unit and check for leaks at pressure sensor tubing on product tank cap Refer to Leak Test on page 93 14 Reinstall cabinet Refer to Removing Cabinet on page 26 Circuit Breaker STEP 10 Upper Mounting Screws Hour Meter STEP 10 1 8 I D Pressure Sensor Tubing STEP 5 ON OFF Switch I O Oxygen Sensor Tube STEP I1 SensO2 Models Only Transformer Harness Patient Outlet Connector Barbed fitting STEP 10 Top and Bottom Barbed fittings Nine or Ten Pn Connector STEP 8 Control Panel 8 Pin Connector STEP 9 Lower Mounting Screws FIGURE 13 1 Removing Control Panel Part No 1118396 65 Platinum Oxygen SECTION 4 COOLING FAN SECTION 14 COOLING FAN Replacing Cooling Fan A DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing NOTE For this procedure refer to FIGURE 14 1 on page 67 NOTE This procedure applies to Platinum 5 and Platinum 9 models 1 Unplug unit Remove cabinet Refer to Cabinet on page 26 Remove control panel Refer to Removing Control Panel on page 64 Remove compressor assembly Refer to Replacing Compressor Assembly 5 Liter Models manufactured before July 24 2003 on page 37 or Replacing Capacitor on
83. s to the On Off VO switch for any disconnected wires If the concentrator does not come on at all and wiring is intact color code and remove wires one at a time Remove and replace with new one Transfer wires from old switch to new switch one at a time to the matching contact Reset breaker NOTE Breaker may trip to safeguard concentrator during a power surge If breaker trips IMMEDIATELY there is a probable short in the unit Check for pinched or charred wires If the breaker does not trip run unit for approximately two hours The circuit breaker should be less than ten ohms If breaker trips again there is an internal problem Contact Invacare Technical Service at 800 832 4707 a Replace P C Board Refer to Replacing P C Board on page 68 b Repair or replace connector Refer to Replacing P C Board on page 68 Part No 1118396 SECTION 4 TROUBLESHOOTING SYMPTOM PROBABLE CAUSE SOLUTION Internal Power Transformer assembly Loss SensO a Faulty Internal Status Indicators RED Off b Connector loose or disconnected GREEN Off c Faulty wiring Alarm may or may not be on Control Panel Indicators RED Off YELLOW Off GREEN Off Fan operates Compressor not operating High Pressure P C board Internal Status Indicators a Malfunction P C board faulty RED One Flash GREEN Two Flashes Unit plugged in power b Disconnected wire switch on continuous alarm Compressor
84. shut down c Shifting valve at pressures greater than 25 p s i 4 way Valve a Not shifting b Bad coil Sieve Beds contaminated Part No 1118396 19 Replace Refer to Replacing Transformer Assembly on page 70 b Reattach connector Replace transformer assembly Refer to Replacing Transformer Assembly on page 70 Replace P C board Unit requires retiming after P C board replacement Refer to Replacing P C Board on page 68 Set flow to max L min for concentrator Check voltage across Pilot Valve 2 on 200 volt scale If meter reads 0 volts when unit is turned on replace P C board Check spade connectors on pilot valve 2 and connectors on P C board Replace P C board Unit requires retiming after P C board replacement Refer to Replacing P C Board on page 68 Check voltage If voltage rises to approximately 24 volts D C when unit shuts down P C board is functioning properly Replace the valve if shorted or open Refer to Replacing 4 Way Valve on page 77 Coil resistance should read 80 ohm 10 Replace 4 way valve Refer to Replacing 4 Way Valve on page 77 Replace Sieve Beds Platinum Oxygen SECTION 4 TROUBLESHOOTING SYMPTOM PROBABLE CAUSE SOLUTION Low Pressure Compressor inlet filter dirty or plugged Replace compressor inlet filter Refer to Internal Status Indicators Replacing the Compressor Inlet Filter on RED One Flash page 3
85. sure Testing After Testing NOTE For this procedure refer to FIGURE 25 4 1 Insert 6 mm V inch plug and 6 mm x 5 cm x 2 inch silicone tubing onto the T fitting and secure with 6 mm V inch clamp 2 Reinstall the cabinet Refer to Removing Cabinet on page 26 6 mm 4 inch Plug 6 mm 4 inch Clamp FIGURE 25 4 After Testing Part No 1118396 103 Platinum Oxygen SECTION 25 SIEVE BED PRESSURE CHECK Models With HomeFill Connection On Back Of Unit Pressure Testing NOTE For this procedure refer to FIGURE 25 5 on page 105 1 Connect pressure gauge to the x 2 inch silicone tubing with connector p n 1131392 2 Remove plug in HomeFill port in back of concentrator 3 Connect x 2 inch silicone tubing with connector to HomeFill port 4 Check each connection to make sure it is tight 5 Turn the concentrator On 1 NOTE Adjust the concentrator to maximum rated Liter flow to 5 L min for 5 Liter and 9 L min for 9 Liter NOTE Wait approximately five minutes to allow system pressures to achieve proper equilibrium before testing 6 The pressure in the concentrator should rise within 20 seconds to 21 p s i g 0 3 and 10 11 for 5 Liter and 25 p s i g for 9 Liter shift the 4 way valve to the opposite sieve bed After the 4 way valve shift the pressure in the concentrator will drop to 12 p s i g 2 before rising within 20 seconds to 21 p s i g 0 3 for 5 Liter and 25 p s i g 0 3 p s i g
86. tch IMMEDIATELY to backup oxygen supply Call supplier IMMEDIATELY GREEN light with YELLOW light flashing Call supplier IMMEDIATELY Oxygen sensor malfunctioning you may continue to use the concentrator Part No 1118396 87 Platinum Oxygen SECTION 21 PURITY INDICATORS OXYGEN PURITY Auto Shut Down e RED Use Backup N Below z Supplier e Normal 4 YELLOW Normal GREEN FIGURE 21 1 Indicator Light Explanation SensO2 Models Indicator Light Explanation LX Models NOTE For this procedure refer to FIGURE 21 2 NOTE This procedure applies to Platinum 5 models only RED light Q Total Unit Shut Down Switch IMMEDIATELY to a back up oxygen supply Call Supplier IMMEDIATELY GREEN light f VO ON OFF System okay Auto Shut Down e a RED Use Backup T Call Supplier Normal e a GREEN FIGURE 21 2 Indicator Light Explanation LX Models Platinum Oxygen 88 Part No 1118396 SECTION 22 TIMING SECTION 22 TIMING A DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing Use extreme care when making adjustments to the timing switch DO NOT allow screwdriver or your hands to contact the P C Board when unit is plugged in and or on Turn unit off and unplug before adjusting timing switch Electrical shock is possi ble To accommodate the varying tolerances when replacing components an adjustab
87. ter models Main Valve Coil pic pu Orange Wires Only one wire needs to be removed FIGURE 24 2 High Pressure Test Oxygen Sensor NOTE For this procedure refer to FIGURE 24 3 on page 99 Alarm sequence occurs when concentration levels fall below a preset value typically 73 3 or 85 2 This failure indicates a catastrophic failure of any component mechanical or electrical Oxygen Sensor Test 1 Turn unit on I Set output flow at 5 L min for 5 Liter models and 9 L min for 9 Liter models 2 Monitor the O level 3 With the O level greater than 8576 after five minutes the GREEN panel indicator light will remain illuminated p Slowly adjust flow beyond maximum rated flow until you can achieve a concentration level greater than 75 but less than 8476 Within 30 minutes the YELLOW panel indicator light will illuminate and the unit will continue to run g Clamp off the 1 8 inch oxygen sensor tubing between the oxygen sensor and the product tank regulator A WARNING Use care not to cut oxygen sensor tubing when clamping the oxygen sensor tubing 6 Within 30 minutes the Oxygen Sensor Alarm should activate and the RED panel light indicator will illuminate The compressor will shut down with the alarm Platinum Oxygen 98 Part No 1118396 SECTION 24 ALARM TEST SENSO INDICATORS LABEL SYMBOL STATUS INDICATOR LIGHTS LED SYSTEM OKAY O over 85 2 GREEN I
88. ter models the clamp is located near middle of product tank FIGURE 10 1 Replacing Check Valves Platinum Oxygen 56 Part No 1118396 SECTION REGULATOR SECTION I REGULATOR Replacing Regulator A DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing NOTE For this procedure refer to FIGURE 11 1 on page 58 NOTE This procedure applies to Platinum 5 and Platinum 9 models 1 Unplug unit Remove cabinet Refer to Removing Cabinet on page 26 gt Loosen adjustable clamp on product tank and slide clamp down below the adjustable clamp hanger Detail A 4 Pull product tank back and up to access regulator NOTE On SensO concentrator models only remove the 1 8 inch tubing to oxygen sensor housing 5 Remove the four mounting screws that secure regulator to product tank cap Detail B or Detail C 6 Remove regulator from product tank lid ensuring that the O rings are removed NOTE Clean mounting surface of product tank cap Ensure that the two O rings for the new regulator are installed onto the ports underside of the new regulator before installation A WARNING Ensure proper installation of regulator Note flow arrow on regulator This MUST face toward the front of unit or flow will be interrupted causing system shutdown Part No 1118396 57 Platinum Oxygen SECTION REGULATOR DETA
89. the power switch applies 230 VAC to the compressor motor hour meter transformer cooling fan and the printed circuit P C board Room air enters the compressor via the cabinet filter and the compressor inlet filter The air is compressed by the wobble pistons in the compressor to a pressure of 144 79 kPa 21 psi for 5 Liter model and 172 36 kpa 25 psi for 9 Liter model As increased pressure creates increased temperature a heat exchanger is used to lower the temperature before the air enters the 4 way valve It is then channeled to a sieve bed containing the adsorption material Restriction downstream of the sieve bed causes pressure to build up inside the sieve bed which is necessary for the adsorption process A small amount of relatively pure oxygen enters the top of the second bed through a restrictive leak in the pressure equalization P E valve with the balance entering a storage tank The nitrogen removed is exhausted back from the bed through the 4 way valve into room air A muffler is located at the exhaust end of the valve to muffle the sound of the exhaust as it exits the concentrator The oxygen not being used to exhaust is channeled into the storage tank The pressurized oxygen is regulated down to 34 5 3 45 kPa 5 0 5 psi for 5 Liter 62 05 kPa 3 45 kPa 9 0 5 psi for 9 Liter enters an accurate flow measuring device flows through Outlet HEPA filter and check valve then out to the patient Platinum Oxygen 14 Part
90. ting does not affect concentrator performance Refer to the HomeFill II owner s manual part number 1116444 for connection and operating instructions When not in use the plug provided with the concentrator should be inserted into the outlet fitting For more information about the HomeFill II contact Invacare TOP VIEW Platinum Oxygen 10 Part No 1118396 SHIPPING AND HANDLING SHIPPING AND HANDLING Before you install and operate the concentrator please refer to the Platinum Owner s Manual 1118389 The concentrator should ALWAYS be kept in the upright position to prevent cabinet damage while being transported The shipping container has been designed to assure maximum protection of the concentrator If the concentrator is to be reshipped by common carrier additional cartons are available from Invacare The air compressor suspension system has been engineered to withstand severe motion and orientation Unpacking 1 Check for any obvious damage to the carton or its contents If damage is evident notify the carrier or your local dealer 2 Remove all loose packing from the carton 3 Carefully remove all the components from the carton NOTE Unless the oxygen concentrator is to be used IMMEDIATELY retain containers and packing materials for storage until use of the concentrator is required Inspection 1 Inspect examine exterior of the oxygen concentrator for nicks dents scratches or other damages Inspect
91. tings or tubing Replace compressor Refer to Compressor on page 37 DO NOT use extension cords Replace or retighten Replace or retighten tubing Replace heat exchanger Refer to Replacing Heat Exchanger on page 60 Regulator cracked or leaking Replace Refer to Replacing Regulator on page 57 Exhaust Throttle muffler dirty or plugged Fan a Not operating Unit overheating b Faulty fan Replace Refer to Replacing Exhaust Canister Exhaust Muffler Platinum 5 Manufactured before 7 24 03 or Inspecting Replacing Throttling Muffler Platinum 5 AW Manufactured after 7 24 03 a Leads to fan disconnected Reconnect b Replace Refer to Replacing Cooling Fan on page 66 Sieve beds defective Replace Refer to Models without HomeFill Connection On Back Of Unit on page 101 Tubing Wine Part No 1118396 21 Platinum Oxygen SECTION 4 TROUBLESHOOTING SYMPTOM PROBABLE CAUSE SOLUTION Low Concentration Continued Fluctuating Flow Platinum Oxygen P C board a Shifts at wrong pressures Flowmeter a Flowmeter opened beyond maximum flow rate b Cracked or broken fitting c Input tubing leaking or loose Timing P E valve a Bad coil b Restrictor blockage Inspect P C board restrictor tubing for kinks or tears Regulator Flowmeter a Incorrectly set regulator b Flowmeter malfunction Outlet HEPA filter a Dirty or plugged 22
92. to Troubleshooting on page 18 if you are unable to adjust or maintain 34 4 kPa 3 5 kPa 5 psi 0 5 psi for 5 Liter model and 62 05 kPa 3 45 kPa 9 psi 0 5 psi for 9 Liter model Part No 1118396 59 Platinum Oxygen SECTION 12 HEAT EXCHANGER SECTION 12 HEAT EXCHANGER Replacing Heat Exchanger Platinum 5 Models Manufactured Before 9 9 2005 A DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing NOTE For this procedure refer to FIGURE 12 1 on page 60 1 Unplug unit 2 Remove cabinet Refer to Removing Cabinet on page 26 CAUTION Use care not to deform heat exchanger when installing removing or cleaning e Remove compressor assembly Refer to Replacing Compressor Assembly 5 Liter Models manufactured before July 24 2003 on page 37 or Replacing Compressor Assembly 5 Liter Models manufactured after July 24 2003 on page 38 Remove screw securing cushioned clamp to the compressor Disconnect the compression fitting nut from it s body Loosen hose clamp and slide out heat exchanger Remove heat exchanger from the compressor assembly Reverse STEPS 3 6 to install new heat exchanger assembly so PND wg Run unit and check for leaks Refer to Leak Test on page 93 10 Reinstall cabinet Refer to Removing Cabinet on page 26 Hose Clamp Heat Exchanger Fitting Nut Cushioned Camp FIGURE 12 1 Replacing H
93. tor should illuminate Platinum Oxygen 24 Base exhaust vent plugged or restricted Cabinet filters dirty or blocked any wall DO NOT place unit on pile or shag carpeting that may restrict air flow Clean or replace Refer to Cleaning the Cabinet Filter on page 27 Reconnect leads Replace fan Refer to Rebuilding the Thomas Model 2650 and 2660 Compressor on page 4l Install fan with air flow arrow pointing down Clean heat exchanger Replace heat exchanger Refer to Replacing Heat Exchanger on page 60 Replace compressor Refer to Replacing Compressor Assembly 5 Liter Models manufactured before July 24 2003 on page 37 Replace capacitor Refer to Replacing Capacitor on page 40 Replace compressor Refer to Compressor on page 37 Replace compressor Refer to Compressor on page 37 Replace compressor Refer to Compressor on page 37 Have line voltage inspected by certified electrician voltage regulator may be required and is obtainable from your local electric company Replace P C board Refer to Replacing P C Board on page 68 Unit may need retiming after P C board replacement Refer to Timing on page 89 Part No 1118396 SECTION 4 TROUBLESHOOTING SYMPTOM PROBABLE CAUSE SOLUTION Unit Not Operating Transformer assembly Internal Status Indicators a Assembly connector disconnected Reattach connector RED Off b Faulty transformer assembly Repla
94. um Oxygen 42 Part No 1118396 SECTION 7 COMPRESSOR Valve Plates NOTE For this procedure refer to FIGURE 7 5 1 Note the position of the plates before removing 2 Lift off the valve plate on one side 3 Repeat for the other side NOTE Note the position of the tab on valve plate for reassembly NOTE To avoid confusion only service one end of the compressor at a time Head Screws Head Compressor Head Plate FIGURE 7 5 Removing Compressor and Compressor Heads Rebuilding the Compressor Compressor Compressor Head Plate Valve Plates Fan NOTE For this procedure refer to FIGURE 7 6 on page 44 1 Remove the fan by pulling it straight off the motor shaft grabbing the center housing DO NOT pull the fan blades NOTE Note position orientation of fan for reassembly Part No 1118396 43 Platinum Oxygen SECTION 7 COMPRESSOR Connecting Rod and Eccentric Assembly NOTE For this procedure refer to FIGURE 7 6 1 Rotate shaft to align the eccentric s set screw with the access hole in the bottom of the compressor housing 2 Insert the 1 8 inch Allen wrench into the access hole in the bottom of the compressor housing Detail A of FIGURE 7 6 3 Loosen the set screw 1 4 turn 4 Rotate connecting rod to top dead center 180 and slide the connecting rod eccentric assembly off the shaft through the opening in the housing 5 Secure the rod assembly in a fixture 6 Remove the sleeve from the
95. ve the mounting screw s that secures exhaust canister assembly to rear of the sound box Refer to Replacing Exhaust Canister Exhaust Muffler Platinum 5 Manufactured before 7 24 03 on page 31 8 Remove exhaust muffler from exhaust canister by turning counter clockwise 9 Loosen the adjustable clamp not shown that secures valve manifold input hose to heat exchanger Details A and C NOTE When removing the valve manifold input hose from the heat exchanger use a flat head screwdriver to pry hose outward while simultaneously pulling the hose outward 10 Remove valve manifold input hose from end of heat exchanger Details A and C 11 Remove the two mounting screws washers 1 per side and bushings 2 per side securing the 4 way valve manifold assembly to top of sound box Detail B 12 The 4 way valve manifold assembly can be removed from the concentrator by lifting the assembly straight up through the cutout in the sound box 13 Remove 4 way valve from manifold assembly Refer to Replacing 4 Way Valve on page 77 Detail A Part No 1118396 79 Platinum Oxygen SECTION 20 4 WAY VALVE 14 15 16 17 18 19 20 21 22 Remove exhaust canister with hoses from manifold Loosen the adjustable clamp on valve manifold input hose to bottom center port of manifold assembly Remove valve manifold input hose from bottom center port of manifold assembly Use screwdriver to pry at connection of tubi
96. w Valve Flapper 15 623137 2 O Ring Valve Plate l6 638281 Fan Gray 17 638282 Fan Black 2660 CHI44 989 COMPRESSOR AW 2660 CHI39 989 COMPRESSOR AWQ ITEM PART NO QTY DESCRIPTION 607731 2 Connecting Rod Eccentric amp Bearing Assembly 6 665167 Head Platinum Oxygen Part No 1118396 17 SECTION 7 COMPRESSOR COMPRESSOR PARTS LEGEND 1 Connecting Rod Eccentric amp Bearing Assembly 2 Piston Cup 3 Screw Piston Cup Retainer 4 Piston Cup Retainer 5 Cylinder Sleeve 6 Head 7 O Ring Head Gasket 8 Screw Head 9 Valve Plate Assembly 10 Valve Restraint 11 Valve Keeper Strip 12 Valve Flapper Intake amp Exhaust 13 Valve Plate 14 Screw Valve Flapper 15 O Ring Valve Plate 16 Fan Grey 17 Fan Black FIGURE 7 9 Compressor Exploded View Part No 1118396 49 Platinum Oxygen SECTION 8 P E VALVE SECTION 8 P E VALVE Replacing P E Valve A DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical out let before servicing NOTE For this procedure refer to FIGURE 8 1 on page 51 NOTE This procedure applies to Platinum 5 and Platinum 9 models 1 Unplug unit Remove cabinet Refer to Removing Cabinet on page 26 Remove spade connectors from P E valve assembly Mm Remove two tie wraps that secure the sieve bed tubing to each side of the P E valve assembly
97. was specifically designed to minimize routine preventive maintenance Only professionals of the healthcare field or persons fully conversant with this process such as authorized or factory trained personnel should perform preventive maintenance or performance adjustments on the oxygen concentrator Radio Frequency Interference Most electronic equipment is influenced by Radio Frequency Interference RFI CAUTION should be exercised with regard to the use of portable communications equipment in the area around such equipment To Reduce the Risk of Burns Electrocution Fire or Injury to Persons DO NOT come in contact with the concentrator while you are wet DO NOT place or store product where it can drop into water or other liquids DO NOT reach for product that has fallen into water Unplug IMMEDIATELY This device is to be used only in accordance with the prescription of a physician and the Platinum 5 Owner s Manual If at any time the patient or attendant concludes that the patient is receiving an insufficient amount of oxygen the supplier and or physician should be contacted IMMEDIATELY No adjustments should be made to the flowrate unless prescribed by a physician therapist The product should NEVER be left unattended when it is plugged in Close supervision is necessary when this product is used by on or near children or physically challenged individuals The Platinum concentrator is intended to be used only as an oxygen suppleme
98. when installing removing or cleaning 3 Remove excess dirt using compressed air or vacuum 4 Reinstall cabinet Refer to Removing Cabinet on page 26 Platinum Oxygen 34 Part No 1118396 SECTION 6 PREVENTIVE MAINTENANCE Heat Exchanger FIGURE 6 8 Cleaning the Heat Exchanger Part No 1118396 35 Platinum Oxygen SECTION 6 PREVENTIVE MAINTENANCE PREVENTIVE MAINTENANCE RECORD INVACARE PLATINUM OXYGEN CONCENTRATOR Model No IRC Serial No On Each Inspection Record Date Of Service Record Elapsed Hours on Hour Meter Clean Cabinet Filters Check Prescribed L min Flow Rate Annually During Preventive Maintenance Schedule Or Between Patien Clean Replace Cabinet Filters Check outlet HEPA Filter Check Compressor Inlet Filter Check Oxygen Concentration SensO models Check Power Loss Alarm As Required Clean Heat Exchanger Replace Exhaust Muffler Rebuild Top End Compressor Every 180 Days LX model Check oxygen Concentration NOTE Refer to Preventive Maintenance section of Service Manual Platinum Oxygen 36 Part No 1118396 SECTION 7 COMPRESSOR SECTION 7 COMPRESSOR A DANGER To prevent electrical shock ALWAYS disconnect concentrator from electrical outlet before servicing Replacing Compressor Assembly 5 Liter Models manufactured before July 24 2003 NOTE For this procedure
99. with a 6 mm clamp 2 Check each connection to make sure it is tight 3 Turn the concentrator On I NOTE Adjust the concentrator to maximum Rated Liter flow 5 L min for 5 Liter and 9 L min for 9 Liter NOTE Wait approximately five minutes to allow system pressures to achieve proper equilibrium before testing 4 The pressure in the concentrator should rise within 20 seconds to 145 kPa 2 kPa 21 p s i g 0 3 p s i g for 5 Liter and 172 36 2 kPa 25 p s i g 0 3 p s i g for 9 Liter and shift the 4 way valve to the opposite sieve bed 5 After the 4 way valve shift the pressure in the concentrator will drop to 96 kPa 14 kPa 14 p s i g 2 p s i g before rising within 20 seconds to 145 kPa 2 kPa 21 p s i g 0 3 p s i g for 5 Liter and 172 36 2 kPa 25 p s i g 0 3 p s i g for 9 Liter and shift the 4 way valve to the opposite sieve bed 6 If pressure reading is 145 kPa 2 kPa 21 p s i g 0 3 for 5 Liter and 172 36 2 kPa 25 p s i g 0 3 p s i g for 9 Liter when both valve shifts are made proceed to STEP 8 Platinum Oxygen 102 Part No 1118396 SECTION 25 SIEVE BED PRESSURE CHECK 7 If pressure in the concentrator does not fall into the specifications listed above refer to Troubleshooting on page 18 8 Turn the concentrator Off O 9 Unclamp the pressure gauge and remove it from the T fitting 6 mm 4 inch Clamp STEP 1 Adjustable Clamp FIGURE 25 3 Pres

Download Pdf Manuals

image

Related Search

Related Contents

OM, Gardena, Programmateur 6040, Art 01275-20, 2004-01  1 of 2  en français - XL-Nord  EA818E−8 ロータリーカッティングヘッド 取扱説明書  Dragon 12  eBOS QRG  

Copyright © All rights reserved.
Failed to retrieve file