SK-500I Syringe Pump Instruction Manual

image

Contents

1. which completely meets the various requirements of modern clinical treatment on different occasions and is now widely used in hospitals SK 500I syringe pump is an economical product of intellectualization collectivization and high technology which can be used in all clinics or wards thus promoting the previously single pump intravenous infusion control to mass control The patients infusion progress and alarm information e g infusion completion pipe occlusion can all be seen at the Nurse s Station This greatly improves the medical nursing quality reduces the nurse s workload and ensures the patient s nursing safety 1 2 Intended Use It is used in hospitals where patients need intravenous infusion at steady speed or continuous and precise infusion 1 3 Safety Precaution 1 3 1 Unqualified and untrained people are not allowed to operate the pump 1 3 2 Attention for setting up the pump 1 Place the pump in dry place 2 Do not place the pump in places where atmosphere pressure temperature humidity sunshine dust salt and ion air may bring damage to the pump 3 Pay attention to the safety of the pump Do not drop vibrate or bump the pump including during transportation 4 Do not use the pump in the place which keeps chemical medicine or gives out poisonous gas 5 Pay attention to the frequency of voltage and allowable current of voltage 6 Make sure there is no high frequency cacophony maker medical equipment o
2. Electromagnetic Compatibility amp Interference This pump is protected against the effects of external interference including high energy radio frequency emissions magnetic fields and electrostatic discharge for example as generated by electrosurgical and cauterising equipment large motors portable radios cellular telephones etc and is designed to remain safe when unreasonable levels of interference are encountered The RF emissions of this pump are very low and are not likely to cause any interference with the nearby electronic equipment However this pump emits a certain level of electromagnetic radiation which is within the levels specified by IEC EN 60601 1 2 and IEC EN 60601 2 24 If the pump interacts with other equipment measures should be taken to minimize the effects for instance by repositioning of relocation 5 Battery Operation The internal rechargeable battery allows continued operation when the AC power is unavailable for example during patient transfer or AC power failure A fully charged battery will provide over 5 hours at typical infusion rates To Achieve optimum operation ensure that the battery is fully recharged after full discharge before storage and at regular 3 month intervals during storage Charge retention will eventually degrade Where retention is critical the internal battery should be replaced every 2 years It is recommended that only a qualified service engineer replaces the battery SK 500I Syr
3. include the usage or process guarantee of whatever natural condition wherever it is placed and in legal commercial or any other kind of activities It guarantees but by no means confines its guarantee to the product s mechanical function and its adaptation to special purposes Purchasers should clearly agree that the compensation will be paid only under this limited guarantee and that they should respect the announcement under this limited guarantee SK 500I Syringe Pump Instruction Manual 1 Preface Thank you for using SK 500I syringe pump Please read the manual carefully before using it so as to operate the pump correctly and safely Table of Content Brief Introduction Intended Use Safety Precaution Functions and Features Specification System Structure Packing Faceplate Appearance Battery Charge Running Volume Curve Basic Operation Steps BOLUS Function Lock and Unlock NEAR Alarm Change New Brand of Syringe Adjust Occlusion Sensibility Cleaning Storage Transport and Maintenance Common Alarm Signal and Alarm silencing Common Problems and Solutions SK 500I Syringe Pump Instruction Manual 2 1 Brief introduction 1 1 Brief Introduction SK 500I syringe pump is a combination of microelectronic technology and modern nursing an important outcome of microcomputer applied to clinical nursing It can control the infusion rate and total volume precisely and continuously over long hours on large scale
4. 01 1 6 2004 Operating mode Continuous operation SK 500I Syringe Pump Instruction f 1 6 System Structure SK 500I1 syringe pump contains the following components 1 The microcomputer system the brain of the whole system giving intellectualized control and management over the whole system and processing the detected signals The two single chip Micyoco SCM systems are used for mutual backup copy and supervision When one SCM goes wrong the other one will give an immediate warning signal and cut the power of the host computer which makes the machine stop working thus ensuring patient s safety 2 The pump device the heart of the whole system and the main driving force of the infusion Driven by step motor the lead screw moves the syringe piston forward 3 The inspection device various kinds of sensors such as displacement sensor detecting liquid infusion rate and infusion volume pressure sensor detecting occlusion etc They can give corresponding signals which will after being magnified be sent to the microcomputer for signal processing After which a control instruction shall be sent for corresponding treatment 4 The alarm device After the signal given by the sensor is processed by the microcomputer an alarm control signal will be received which will be responded by the alarm device to arouse people s attention for right treatment There are mainly two kinds of alarms photoelectric alarm LBD and sound
5. Otherwise it will cause inaccuracy occlusion but not alarm or no occlusion but alarms Step 3 Set the syringe s parameters 1 50ml syringe s parameter setting Under the condition of Power Off with finger pressing CLEAR key first and then another finger pressing eg altogether for 5 seconds and then release both of them together as shown in Diagram 26 The pump will display as shown in Diagram 27 while the 50ml indicator light turns on accordingly Turn the adjusting knob to adjust the value to the one you measured in Step 1 then press adjusting knob to save the value newly set Diagram 20 Diagram 27 2 20ml syringe s parameter setting Under the condition of Power Off with finger pressing BOLUS key first and then another finger pressing 2s altogether for 5 seconds and then release both of them together as shown in Diagram 28 The pump will display as shown in Diagram 29 while the 20ml syringe light turns on accordingly Turn the adjusting knob to adjust the value to the one you measured in Step 1 then press adjusting knob to save the value newly set SK 500I Syringe Pump Instruction Manual 17 Diagram 28 Diagram 29 3 10ml syringe s parameter setting Under the condition of Power Off with finger pressing key first and then another finger pressing 2S altogether for 5 seconds and then release both of them together as shown in Diagram 30 The pump will display as shown in Diagram 31 while the 10ml i
6. Range of infusion rate 10ml syringe 0 1 100ml h 20ml syringe 0 1 200ml h 50ml syringe 0 1 500mlI h Bolus rate 10ml syringe 100ml h 20ml syringe 200ml h 50ml syringe 500ml h Accumulated infusion volume 0 1 999 9ml 3 Power supply AC110 240V 50 60Hz 25W Battery Rechargeable lithium battery 7 4 V 1650mAh in compliance with the requirements of IEC 60601 1 25W running more than 5 hours after being fully charged Battery charge When syringe pump is connected to AC power the power needs to be turned on you may or may not operate the machine the battery will automatically start charging need 8 14 hours Displayed information Flow rate accumulated infusion volume syringe size battery capacity bed No AC power connecting indication Alarm information infusion completion near completion occlusion low battery syringe abnormal control abnormal no AC alarm Operating conditions Environment temperature 5 C 40 C atmosphere pressure 86 106kPa relative humidity 20 90 Storage Environment temperature 15 C 50 C atmosphere pressure 86 106kPa relative humidity20 90 Applicable syringes 10ml 20ml and 50ml syringe The manufacture uses Double Dove syringe to test Standard Comply with the standards of EN I SO14971 2000 A1 2003 IEC60601 1 1995 A1 A2 EN IEC 0601 1 2 2091 IEC60601 2 24 1998 EN 980 2003 EN 04 1998 EN60601 1 4 1996 A 1 1999 EN606
7. SK 500I Syringe Pump Instruction Manual Product Registration No Yue FDA approved 2005 22100022 Please read the manual before using the product Please keep the manual for reference Limited Guarantee SK 5001 syringe pump is delicately made of top quality components Shenzhen Shenke Medical Instrument Technical Development Co Ltd guarantees that the product won t have any defect in material and technology within 12 months from the purchasing date The company s liability and the agents authorized by Shenzhen Shenke Medical Instrument Technical Development Co Ltd under this limited guarantee should be restrictive under the operation privilege of Shenzhen Shenke Medical Instrument Technical development Co Ltd undertaking the maintenance and the replacement of the syringe pump under the circumstance that it has defect in material and technology through inspection The maintenance and the replacement of any products based on this limited guarantee should not exceed the time limit of the above limited guarantee All the maintenance based on the limited guarantee should be undertaken by qualified and trained people When syringe pump is found defect within the time limit of the limited guarantee the purchaser should notify Shenzhen Shenke Medical Instrument Technical Development Co Ltd or the authorized agents within 30 days to find the products defects The defective syringe pump should be sent directly to Shenzhen Shenk
8. The infusion The alarm can be silenced completion i completes by pressing the adjusting knob The alarm will ring again in 2 minutes Press key to silence the alarm and also stop the infusion at the same time SK 500I Syringe Pump Instruction Manual 21 Infusion pipe occlusion Low battery capacity Syringe disengaged SK 500I Syringe Pump Instruction Manual 22 1 The infusion line 1S blocked 2 The occlusion sensitivity 1s too high 3 There is something wrong with the sensor 1 The battery capacity is too low 2 The battery ages or there is something wrong with syringe pump s charge circuit 1 There is no AC supply 2 There is a problem with power supply circuit 1 The syringe is disengaged or not properly installed 2 The syringe s parameter is wrongly set 3 There is a problem with the syringe pump s potentiometer For cause 1 Press key to silence the alarm and also stop the infusion at the same time Start the infusion again after solving the problem For cause 2 Refer to 2 4 Change new brand of syringe on page For cause 3 Ask the manufacturer to check and repair For cause 1 Connect to AC and charge the battery fully you need to turn the power on For cause 2 Ask the manufacturer to check and repair For cause 1 Check if the power cord is well plugged or not For cause 2 Ask the manufacturer to check and repair For ca
9. alarm buzzer 5 The input and display device The input part is in charge of setting various infusion parameters such as infusion volume and infusion rate etc The display part is in charge of displaying various parameters and showing the current operation progress on the LCD 1 7 Packing 1 Product label pattern pasted on the back shell of the pump Syringe Pump Eno Model SK 5001 LOT sil A IP21 ps4 mee 2007 09 Input AC100 240V 50 60Hz 25VA Shenzhen Shenke Medical Instrument Technical Development Co Ltd 8th Floor Building No 615 2 Bagua Road Futian District Shenzhen 518029 P R China WELLKANG LTD t a WELLAKANG TECH CONSULTING Ec rer Suite B 29 Harley Street i LONDON WIG SOR United Kingdom WY AML ODONI AV 2 Definition and Symbols Lot number Serial No SK 500I Syringe Pump Instruction Manual 5 Attention consult accompanying documents TYPE BF equipment Not for Municipal Waste Date of Manufacturer Manufacturer REP CEE Device complies with the requirements if the EC Directive 93 42 EEC Registered with the CE mark European Representative P2 l Protected against Vertically falling drops of water 3 Inside content of a standard packing Syringe pump x1 AC power cord x 1 Instruction manual x1 Pole clamp x1 Certificate of approval x 1 Maintenance card x1 SK 500I Syringe Pump Instruction Manual 6 Please contac
10. arge automatically The battery capacity indicator light will be on If all 3 lights are on it means the battery is fully charged and it will stop charge automatically When less than 3 lights are on the pump will begin to charge the battery SK 500I Syringe Pump Instruction Manual 8 automatically 2 As shown in Diagram 2 the AC indictor light will be off when the power is turned off This means the syringe pump is using the battery capacity When all the 3 battery capacity indicator lights are off the pump will give an alarm to remind the user to charge the battery Notice It needs 14 hours to charge the battery fully after its power is used up Diagram 1 Diagram 2 2 0 Running Volume Curve model ne K 5001 with domi sings Setup Aow Rate Condition miii two hours H0 mm Hg Cellvery Head E ru Q a Tine imin SK 500I Syringe Pump Instruction Manual 9 modalno K 5001 Setup Aow Rate Condition imiti two hours mm Hg Cellvery Head E E Time min 2 Syringe Pump Operation 2 1 Basic Operation Steps Step 1 Power connection Insert the power cord into the power supply as shown in Diagram 3 Notice The applicable voltage 1s AC100 240V 50 60Hz Diagram 3 Step 2 Turn the power on Keep pressing the Shey for 3 5 seconds and then release it as shown in Diagram 4 The syringe pump begins self diagnosis and all the indicator lights will flash by turns After the self dia
11. ction Manual 15 2 3 Lock and unlock NEAR alarm function 1 Under the condition of Power Off with fingers pressing START CLEAR keys together first and then another finger pressing edl altogether for 5 seconds and then release all of them together as shown in Diagram 21 2 The pump enters into the interface of locking and unlocking of alarm function as shown in Diagram 22 The alarm indicator light will turn on accordingly and the screen will display a value You can adjust this value through the adjusting knob O1 stands for locking the alarm function and 02 stands for unlocking alarm function Press the adjusting knob to save the value newly set Diagram 21 Diagram 22 2 4 Change New Brand of Syringe Step 1 Measure the length of the syringe Put the syringe to the measurer of the Fast Operation Manual as shown in Diagram 23 measure the new brand of syringe by the scale it reads 75mm as shown in Diagram 24 Diagram 23 Diagram 24 Step2 Install the syringe Turn off the syringe pump as shown in Diagram 25 Install the syringe needed on the syringe pump Make sure the syringe s piston is pushed to the end of the syringe Push the compression strut above until it touches the syringe properly SK 500I Syringe Pump Instruction Manual 16 Diagram 25 Notice We recommend Double Dove syringe If you want to change to other brand of syringe please refer to 2 4 Change New Brand of Syringe on page
12. e Medical Instrument Technical Development Co Ltd or the authorized agents for the purpose of inspecting fixing and replacing the syringe pump The postage should be borne by purchaser The products mailed to Shenzhen Shenke Medical Instrument Technical Development Co Ltd or its agents should be appropriate packed in transportation packing box with instruction manual of Shenzhen Shenke Medical Instrument Technical Development Co Ltd Inappropriate packing will cause severe damage to syringe pump which should be borne by purchaser The limited guarantee of Shenzhen Shenke Medical Instrument Technical Development Co Ltd does not apply to products damage or defects resulted from the following causes mechanism damage and incorrect installation caused completely or partially by neglect of duty liquor feeding falling of syringe pump false use and abuse reconstruction of syringe pump by any unprofessional disqualified or untrained people damage due to incorrect packing when mailed to Shenzhen Shenke Medical Instrument Technical Development Co Ltd and its agents Shenzhen Shenke Medical Instrument Technical Development Co Ltd or its agents reserve the rights to issue the products inspection invoice to purchasers provided they are not able to confirm the matter after inspection This limited guarantee is the only complete guarantee related to the products of Shenzhen Shenke Medical Instrument Technical Development Co Ltd It does not
13. egins Diagram 15 Notice The faster the operation indicator lights flashes the higher the infusion rate is Al 14 A 15 Step 8 Infusion completion As shown in Diagram 16 when infusion comes to near completion the indicator light shall be on accompanied with an alarm signal to alert the user Notice The alarm can be silenced by pressing the adjusting knob and the NEAR alarm will ring in 2 minutes As shown in Diagram 17 when infusion is completed the indicator light shall be on accompanied with an alarm signal to alert the user Press key as shown in Diagram 16 and the pump will stop the operation Notice The alarm can be silenced by pressing the adjusting knob and the alarm will ring in 2 minutes Press the adjusting knol the alarm Diagram 16 Diagram 17 SK 500I Syringe Pump Instruction Manual 14 Diagram 18 Step 9 Turn off the power Press 2s key for 3 5 seconds as shown in Diagram 19 to turn off the power Diagram 19 2 2 BOLUS function During the infusion process if you need bolus infusion while the on going rate is lower than the maximum rate of the size of syringe in use you can keep pressing the key as shown in Diagram 20 the pump shall begin infusion at the fastest rate It will go back to the original rate after releasing key Notice the maximum rate for 10ml syringe 100ml h for 20ml syringe 200ml h and for 50ml 500ml h Diagram 20 SK 500I Syringe Pump Instru
14. gnosis finishes the LCD shall display as shown in Diagram 5 Notice When the orange light is on it means the syringe pump is using AC power and that the battery is getting charged automatically SK 500I Syringe Pump Instruction Manual 10 Press cs key just on touch for night vision in the dark Diagram 4 Diagram 5 Step 3 Install the syringe 1 Pull the compression strut of the syringe up to its top and then turn it leftward around 90 then let it go as shown in Diagram 6 2 Press clutch button to pull its piston to the suitable place while putting the syringe filled with liquid into the mounting groove properly 3 Turn the compression strut rightward and compress it onto the syringe The indicator light corresponding with the type of syringe will be on as shown in Diagram 7 Install the syringe into the Syringe s handspike must touch mounting groove N the pisto i correct Diagram 6 SK 500I Syringe Pump Instruction Manual 11 ii N E J a Syringe s compres must touch syringe p Diagram 7 Step 4 Set the infusion rate As shown in Diagram 8 when the syringe pump is in STOP status the LCD will turn to display rate while the rate indicator light is on You can set the rate by turning the adjusting knob Notice Turn the adjusting knob clockwise to increase the infusion rate or anti clockwise to decrease the infusion rate Turn the adjusting Diagram 8 Step 5 Elimina
15. inge Pump Instruction Manual 20 6 Disposal 6 1 Information on Disposal for Users of Waste Electrical amp Electronic Equipment This sm Symbol on the product and or accompanying documents means that used electrical and electronic products should not be mixed with household waste If you wish to discard electrical and electronic equipment please contact your Cardinal Health affiliate office or distributor for further information Disposing of this product correctly will help to save valuable resources and prevent any potential negative effects on human health and the environment which could otherwise arise from inappropriate waste handling 6 2 Information on Disposal in Countries outside the European Union This symbol is only valid in the European Union The product should be disposed of taking environmental factors into consideration To ensure no risk or hazard remove the internal rechargeable battery and the Nickel Metal Hydride battery from the control board and dispose of as outlined by the local country regulations All other components can be safely disposed of as per local regulation 7 Eliminate the alarm and solve problem 7 1 Common Alarm Signals and Alarm Elimination Display Near The infusion is The alarm can be silenced completion i near by pressing the adjusting i completion knob The alarm will ring again in 2 minutes Press key to silence the alarm and also stop the infusion at the same time Infusion
16. liquid off 4 Do not use something like xylene acetone or something analogous to clean the syringe pump These chemicals will cause damage to the outer shell 3 2 Storage and transport The pump should be stored in the environment with temperature between 15 50 C atmosphere pressure between 86 106kPa and relative humidity between 20 90 3 3 Preventive inspection Interval Routine Maintenance Procedure As per Hospital Policy Thoroughly clean external surfaces of the pump before and after prolonged period of storage At least once per year 1 Inspect AC power supply plug and cable for damage Refer to TSM for identification of parts 2 Perform functional tests as outlined in the SK 500I Syringe Pump Instruction Manual 19 Technical Service Manual 3 Operate the pump on battery power until the low battery alarm then charge the battery to confirm operation and charging 3 4 Maintenance The warranty is year from the purchasing day We can offer free repair service within the warranty time on condition that the product is operated properly The following situation is not within the range of free maintenance and repair 1 Malfunction caused by false use repair or reconstruction by any unprofessional unqualified or untrained people 2 Malfunction or damage caused during transportation 3 Malfunction or damage caused by fire salt poisonous gas earthquake hurricane flood abnormal voltage and other nature factors 4
17. ndicator light turns on accordingly Turn the adjusting knob to adjust the value to the one you measured in Step 1 then press adjusting knob to save the value newly set Diagram 30 Diagram 3l 2 5 Adjust occlusion sensibility 1 Under the condition of Power Off with finger pressing START key first and then another finger pressing egl altogether for 5 seconds and then release both of them together as shown in Diagram 32 2 The pump enters into the interface of setting pressure alarm value as shown in Diagram 33 The occlusion alarm indicator light will turn on accordingly and the interface displays the pressure alarm value Adjust the pressure alarm value by the adjusting knob The smaller the value is the more sensitive it shall be Press the knob to save the value newly set SK 500I Syringe Pump Instruction Manual 18 Diagram 32 Diagram 33 Notice The recommended occlusion value is 12 20 when the machine operates at medium speed It enters into KVO mode when occlusion alarm sounds KVO mode shall end when the occlusion alarm 1s silenced 3 Cleaning Storage Transport and Maintenance 3 1 Cleaning 1 Use apiece of cloth to wipe the syringe pump clean when any liquid drops on the pump 2 Special attention must be paid when cleaning the syringe pump Use a tampon moistened with 75 of alcohol to wipe the outer shell 3 Be sure to turn the syringe pump off and disconnect it from AC power supply before wiping the
18. r cell phone etc near the pump 1 3 3 Attention before using the pump 1 Confirm the power cord is safely and correctly connected 2 Confirm the ON OFF function and operation of the pump are ok 3 Confirm there is no high frequency cacophony maker medical equipment or cell phone etc near the pump 4 Confirm the patient s vein passage has been established 5 Confirm the pump is mounted within 1 0m above or below the patient s heart 1 3 4 Attention during using the pump 1 Do not exceed the time the doctor diagnoses for the treatment 2 Keep supervising the working condition of the pump and the situation of the patient continuously 3 When finding anything abnormal with the pump or patient consider the patient s safety first then stop the infusion and take proper action SK 500I Syringe Pump Instruction Manual 3 1 3 5 Attention after finishing using pump 1 Turn off the power and pull out the power cord as required 2 Clean the pump before storing it for convenient use next time 1 4 Functions and Features 1 Precise control of the infusion rate 2 Precise control of the infusion volume 3 Smooth flow no pulsant 4 Audible and visible alarm when there is infusion completion near completion occlusion low battery syringe abnormal control abnormal and no AC power 5 Intellectualized control of infusion 1 5 Specification Type SK 5S00I 500ml h Different syringes have different maximum rate
19. ss the key for about 3 5 seconds 012 START key Start infusion after installing the syringe properly 13 STOPkey Press the key to stop infusion Keep pressing this key during infusion the pump will infuse 14 BOLUS key at the maximum rate After releasing this key the pump shall go back to its original rate Press the key to select infusion rate accumulated volume eee and bed number oe The knob is used to adjust infusion rate bed number etc D oD cocou um Press the knob to confirm the value newly set Keep pressing the button to move the piston freely When 17 Clutch button letting go of the fingers the clutch will gear and can only be moved by the engine SK 500I Syringe Pump Instruction Manual 7 l 12 Use the piston to push the syringe Check the type of the syringe 19 syringe pressure lever This light shall be on when the 50ml syringe is placed 50 ml syringe indictor light properly This light shall be on when the 20ml syringe is placed 20 ml syringe indictor light properly This light shall be on when the 10ml syringe is placed 10 ml syringe indictor light properly User friendly operation Operation guide for nurse s reference instruction 23 22 1921 20 15 1418 17 i 2 3 45 6 7 8 910 1112 13 16 1 9 Battery Charge 1 As shown in Diagram 1 after the pump is connected to AC power supply and turned on the AC indicator light is on and the battery begins to ch
20. syringe Reinstall the syringe key to stop the syringe pump Press SELECT key until RATE indicator light is on Turn the adjusting knob to get the needed value Press the adjusting knob and the value can be saved Connect the syringe pump to AC power be sure the power is turned on to charge the battery Refer to 2 4 Change new brand of Syringe on page Refer to 2 4 Change new brand of Syringe on page and set the parameter again Manaufacturer Shenzhen Shenke Medical Instrument Technical Development Co Ltd Address 8th Floor Building No 615 2 Bagua Road Futian District Shenzhen 518029 P R China Tel No 86 755 82402696 82438546 82438005 Fax No 86 755 82438567 Website www sk medical cn www skmedica com SK 500I Syringe Pump Instruction Manual 23 For Technical Documents please contact our European Representative Company Wellkang Ltd t a Wellkang Tech Consulting Website www wellkang com Address Suite B 29 Harlev Street London W1IG9QR United Kingdom SK 500I Syringe Pump Instruction Manual 24
21. t your syringe pump sales agent if there is any component above mentioned missing 1 8 Product Faceplate Appearance The 3 indicator lights mean the level of battery capacity Battery capacity indicator from low to high as marked from below to top If all 3 lights light on it means the battery is full charged When this light is on it means the pump has already been AC Power indicator light connected to AC power and the battery starts charging automatically After the infusion begins the three lights flash alternately Operation indicator light from right to left When stopping the infusion all the three lights are off When this light is on it means the value displayed on the LCD screen is the accumulated volume When this light is on it means the value displayed on the LCD screen is bed number When this light is on it means the value displayed on the LCD screen is infusion rate This is an alarm light indicating that infusion will soon complete This is an alarm light indicating occlusion in infusion pipe and meanwhile the syringe pump stops automatically This is an alarm light indicating infusion completion Accumulated volume indicator light Bed number indicator light Infusion rate indicator light NEAR alarm indicator light indicator light indicator light CLEAR key PTOL condition press the key to clear the accumulated Turn on Press the key for about 3 5 seconds 0 E key Turn off Pre
22. te the accumulated volume 1 Press key as shown in Diagram 9 The screen will display the accumulated volume when the accumulated volume indicator light is on as shown in Diagram 8 2 Press key as shown in Diagram 10 the accumulated volume will become O and start to calculate a new Notice The accumulated volume can only be viewed during the infusion process If you need to eliminate it you have to press first SK 500I Syringe Pump Instruction Manual 12 Diagram 9 Diagram 10 Diagram 11 Step 6 Set bed number necessary when used together with infusion supervision system Press key as shown in Diagram 12 the screen will display the bed number and the bed number indicator light shall be on as shown in Diagram 13 Then you can turn the knob to the bed number and press the knob to save it In about 5 seconds the screen will go back to the display of RATE as shown in Diagram 10 and the bed number adjusting is finished Notice Turn the adjusting knob clockwise to increase the number or anti clockwise to decrease the number Diagram 12 Diagram 13 Step 7 Start infusion SK 500I Syringe Pump Instruction Manual 13 After all values needed are well set and that the syringe is properly installed press key as shown in Diagram 14 to start the infusion the operation indicator lights will flash alternately from right to left The piston shall drive the syringe s handspike from right to left slowly as the infusion b
23. use 1 install the Syringe properly again For cause 2 Refer to 2 4 Change new brand of Syringe on page and set the parameter again For cause 3 Ask the manufacturer to check and repair Control abnormal Il the 3 alarm indictc on at the same time 7 2 Common problems and solutions Press START key but the LED light is not on Turn the adjusting knob but the infusion rate does not change The syringe pump cannot turn with the battery on The occlusion alarm signal sounds soon after the syringe pump starts operation The indicator light cannot detect the syringe s size Manaufacturer The syringe pump does not work The infusion rate does not change The pump turn on syringe cannot The occlusion light is on and the syringe pump stops working Indicator light corresponding to the syringe size is not on information The syringe s parameter is wrongly set The syringe pump s motor cannot run properly The syringe is not properly installed The rate cannot be adjusted in operating status Battery used up The occlusion sensitivity is too high The syringe pump cannot identify the syringe s size For cause 1 Refer to 2 4 Change new brand of Syringe on page and set the parameter again For cause 2 Ask the manufacturer to check and repair Check if the indicator light is on or not after installing the