Home
        Service Manual
         Contents
1.                   OO  Connecting relationship   Part  Fig  1 2 3 4  I Fig  G  P2 Fig  G  P1 Fig  H  Left Fig  H  Right  Table 3          9 17       Service Manual for RESmart Auto CPAP CPAP V1 1       Components Illumination    1  Device Inside             S N Assembly Figure  1 Insert humidifier locker on the shield bottom   2 Insert humidifier power supply reed on the   tpt Blok    shield bottom as per Fig  E and F        3 Assemble the humidifier locker and button  spring        4 Put fixing on the spring and fix by screw  2X  3mm 8mm            A Power wire to             6 Assemble the shield back on the bottom  fix  by screw  2x3mm 8mm   y  pere    7 Assemble the foam pad  non woven side up         8 Insert the blower unit              Connect the air hole of the pressure sensor  YL on the main board with the bottom hole   9  hole 1  in the front of the blower outlet with a Canal to sensor YL  canal of 12   cm in length and 2 mm in Veen eth    inner diameter   gt     Fi                      10 17    Service Manual for RESmart Auto CPAP CPAP V1 1          10    Clean up the blower wires and wind them with  a tube  Insert the plugs into sockets PM  Put  the lead of thermistor into socket PRT on main  board  Connect the main board to 24V power  supply           11       Draw the shield back backwards and put the  3 pin plug into the corresponding socket on its  back  Lay the main board onto the 4  supporting poles on the shield bottom and set  it into the power board via t
2.                 Appendix B     Spare Parts Configuration          S N Name Quantity Figure                13 17          Service Manual for RESmart Auto CPAP CPAP V1 1                                        1 Shield Cover 1  2 Handle 1  3 Handle Fixin 2  4 Shield Bottom 1  5 Humidifier Clip 1  6 Humidifier Locker 1  7 Humidifier Fixing 1  8 Filter Cover 1  9 Shield Back 1  10 Outlet 1                14 17       Service Manual for RESmart Auto CPAP CPAP V1 1                                        11 PCBA Fixing   12 Infrared Window   13 Panel   14 Humidifier Outer Shield  15 Humidifier Inner Shield  16 Light Window   17 Pole Platelet   18 Humidifier Power Plug  19 Humidifier Plug  left up   20 Humidifier Plug  left down   21 Humidifier Plug  right up           15 17                   Service Manual for RESmart Auto CPAP CPAP V1 1                               22 Humidifier Plug  right down  1   23 Device Pad 6 C  24 Blower Connector 1 B  25 Humidifier Inner Connector 1 r        26 Key Button 1 a    27 Humidifier Scaleboard 1 kan  28 Heater Plane Spring 3 ET    29 Button Spring 1   dk                 Spare Parts List                                     Part Name SIN Qty Unit Classify Use   Main Board PCBA 1220EP0A 2000 1 PC PCBA Main Device   Heater PCBA 1220HW01 1000 1 PC PCBA Humidifier  Power Supply PCBA 123B0048 1 PC PCBA Main Device  Blower Unit  TKD  1220CP0A11130 1 PC Assembly Main Device   Heater Plane 1220HW0112100 1 Set Assembly Humidifier  Power Supply PCBA Wire 1200
3.       Waiting for APAP   s reset    Please disconnect the power supply of the device   wait for over 5 seconds to make sure that the  device is reset  and connect the power cable again     Waiting for the reset of the device          Power on the RESmart   Auto CPAP CPAP device  If the device is turned on at the beginning  please  power off the device and turn on again 5 seconds later     4  After the device is turned on  it will display as below     Upgrading    Click  Upgrade  to start upgrading the device       Click    Upgrade    to start  If succeeded  the right version number will be displayed on display screen  when powered on     Note  If the upgrade failed  please repeat above steps 1 4     If    Error 08    displayed after firmware upgrading  it means unmatched firmware used  for example   applied CPAP firmware to RESmart   Auto CPAP device   In this case  please apply correct firmware  and upgrade again     3 2 Hidden Function    1  Humidifier matching and dis matching  This is automatically for the device  if the  humidifier is not working properly  perform this operation    Available on version 1 19 and later     Assemble humidifier on main device  without water chamber  Enter the maintenance menu  and set  Ramp 40  then press Heated Humidifier Button when the Init P appears  Thus the humidifier is  matched  When    OK    appears on screen  humidifier matching succeeded     If cut the power supply after pressing    Pressure Start Stop Button    when the humidifie
4.     Outlet  left side     ol                                                       Hole  1  2 and 3 in Fig  B  connect to sensor YL via silicon rubber canal  Specified in Table 1        Fig  B    Fig  A    Specification of canal       1          Hole  1 on sensor YL d 2mm    4mm  Li2cm             Table 1    6 17    Service Manual for RESmart Auto CPAP CPAP V1 1       Fig  C   Power PCBA  front side  Sockets figuration                            rel   ees   a  00000 Z                      Fig  D     Wire connecting on power PCBA  1 2    SS       h Blue hum               HZ                fl Brown  3 pin plug  AC Socket    Fig  E   Power supply connecting  1          Orange              Orange                   2 pin plug Reed    6 pin plug  Fig  F     Power supply to humidifier    TAT    Service Manual for RESmart Auto CPAP CPAP V1 1                               Connecting relationship                                         2  Humidifier  Fig  G   Humidifier PCBA  back side           m          P1             ro  ET                    8 17    Fa  Pan 1 2 3 4  D Back of device Fig  C  CN1  E Fig  C  CN2 Fig  A  P24 Fig  F  Left Fig  F  Right  Blower 8 pin plug Fig  A  PM  Table 2       Service Manual for RESmart Auto CPAP CPAP V1 1       Fig  H     Power supply to humidifier    Power Supply to Humidifier                                                          Fig       Heater PCBA       4 pin Plug           3 pin Plug       ana    Power Reed to Humidifier                   
5. 02 1 Set Wire Main Device  Main Board PCBA Wire 120003 1 Set Wire Main Device  Silicon Rubber Canal 290001 1 PC Assembly Main Device  Handle 210030 1 PC Plastic Main Device                      16 17          Service Manual for RESmart Auto CPAP CPAP V1 1                                                                                                       Handle Fixing 210011 2 PC Plastic Main Device  Shield Bottom 210031 1 PC Plastic Main Device  Humidifier Clip 210004 1 PC Plastic Main Device  Humidifier Locker 210012 1 PC Plastic Main Device  Humidifier Fixing 210013 1 PC Plastic Main Device  Filter Cover 210005 1 PC Plastic Main Device  Shield Back 210006    PC Plastic Main Device  Outlet 210007    Set Plastic Main Device  PCBA Fixing 210014 2 PC Plastic Main Device  Infrared Window 210010 2 PC Plastic Main Device  Panel 210036 1 PC Plastic Main Device  Power Supply PCBA Fixing 210015 1 PC Plastic Main Device  Humidifier Outer Shield 210038 1 PC Plastic Humidifier  Humidifier Inner Shield 210039 1 PC Plastic Humidifier  Pole Platelet 210025 1 PC Plastic Humidifier  Light Window 210023 1 PC Plastic Humidifier  Button Spring 230001 1 PC Metal Main Device  Humidifier Scale Board 230004 1 PC Metal Humidifier  Heater Plane Spring 230005 3 PC Metal Humidifier  Humidifier Power Plug 220004 1 PC Rubber Main Device  Humidifier Plug  left up  220005 1 PC Rubber Main Device  Humidifier Plug  left down  220006 1 PC Rubber Main Device  Humidifier Plug  right up  220007 1 PC Rubber Mai
6. BI     RESmart  Auto CPAP CPAP System    Service Manual    Service Manual for RESmart Auto CPAP CPAP V1 1       Table of Contents    1  Introduction  1 1 Intended Use  1 2 Implements        2  Structure  2 1 System Feature                          2 2 Internal Structure  2 3 Spare Parts         3  Software                 3 1 Upgrading Device Firmware  3 2 Hidden Function      Filon  4  Malfunction and Countermeasure    4 1 Error Code Table and Information   4 2 Malfunction Determination                     Appendices      Appendix A   Device Configuration  1  Main device       2  Humidifier     Components Illumination  1  Device Inside       Appendix B             Spare Parts Configuration  Spare Parts List       Service Manual for RESmart Auto CPAP CPAP V1 1       IMPORTANT    All data in this manual have been verified correctly  It is sufficient for servicing and repairing  this device  If applying this manual on other purpose  the validation should be confirmed by  BMC Medical Co   Ltd  Otherwise  BMC Medical Co   Ltd  has no responsibility for the result   All information in this manual is protected by law and regulation  All copyrights belong to  BMC Medical Co   Ltd     IMPORTANT  CAUTIONS and WARNINGS in this manual are to emphasize dangers to service  people     WARNING  If operate improperly  it may cause damage to people and environment   CAUTION  If operate improperly  it may cause damage to instrument     IMPORTANT  Important information for servicing and repairi
7. asuring  Pressure meter    2  Structure  2 1 System Features                          Handle    j  a Hi Display Screen   Sem  SS                                User Buttons                Humidifier Controller  Air Outlet    Humidifier Power    Medical Product Note  bottom        AC Inlet  Communication Connector             1 17    Service Manual for RESmart Auto CPAP CPAP V1 1            Indicator  Outlet Port Light    Fill Line  Water Chamber    2 2 Internal Structure   See Appendix A   2 3 Spare Parts   See Appendix B   3  Software   3 1 Upgrading Device Firmware    Note  Once the RESmart   Auto CPAP CPAP device is powered on  the LED display screen will show  the software version of the device as the following picture  Please remember the software version in  the red frame     1 88 AL RSS       1364DF58    1  Power off the RESmart   Auto CPAP CPAP device  and then connect the device to PC through  the USB data cable     2  Run the software    BMC RESmart Auto CPAP CPAP Tools for Upgrad    from PC and it will display  as below       BEC Resmart Auto CPAP Tools for Upgrade             Upgrade File   1 98 AL upd    Browse    Serial Port   COM1 xl             ok   Cancel         a  Select the upgrade file which has same suffix letter as the machine version   For example V  ersion 1 88 AL should choose file name 1 98 AL     b  Select the right serial port  default is COM1   3  Click    OK     it will display as below     2 17    Service Manual for RESmart Auto CPAP CPAP V1 1 
8. function and Countermeasure  4 1 Error Code Table and Information                                                 Error Problems Probable Cause   Index   Error01   Over temperature inside or temperature sensor fault Malfunction on main board   Error02   Motor stops working eee On Motor  or  Main   Error03   Low speed of motor ea elan on  motor     or man   Error04   Low temperature inside Malfunction on main board   Error07   Pressure sensor fault Malfunction on main board   Error08   Wrong firmware version Firmware   Error09   Host parameter error Malfunction on main board   Error12   Over temperature inside or flowrate sensor fault Malfunction on main board   Error13   Temperature below 0  C or flowrate sensor fault Malfunction on main board   Error14 The motor is over temperature or motor temperature   Malfunction on motor or on motor  sensor fault temperature sensor   Error15   Motor temperature sensor fault Ye net  n on  motor  temperature       417       Service Manual for RESmart Auto CPAP CPAP V1 1       4 2 Malfunction Determination                           Can Not Strat Up             Kjddng zomogq  apo Jomog    Self Check                   nod  QV PUD    Replace Main boar          Vad Jayipruiny     Replac             Refer error  code table    Inner Airway Connecte          Rewrite correct softwar    5 17    Service Manual for RESmart Auto CPAP CPAP V1 1       Appendices  Appendix A   Device Configuration    1  Main device    Fig  A  Main board  backside  Fig  B 
9. he fixing pole on  PCB and get the main board fixed    Put blower wires on the side into the gap  between blower unit and shield bottom        11 17             Service Manual for RESmart Auto CPAP CPAP V1 1       2  Device Outside                                                 SIN Assembly  1    Put the handle in the socket of the shield cover   2 Assemble handle fixing to handle and fix by screw   4 3 8 mm    3   Assemble the infrared window  ee  4   Assemble and fix the panel   5   Put on the key buttons   Assemble shield cover on the bottom  adjust the  6   position of outlet and make sure it is clamped in  the right socket   Screw Hole  7   Fix the shield by 4 pieces screws  M3 10 mm       Device Pad  8   Stick the 4 pieces of device pads   9   Putin the filter and assemble the cover  i ZL           12 17       Service Manual for RESmart Auto CPAP CPAP V1 1                         Plug  up   10   Put on all plugs if necessary  without humidifier   l  Power hi  Plug down   3  Humidifier  S N Assembly       As per Fig  H  assemble the heater plane and  wire  Fix them by screw  3 8 mm            Put springs on each of the three poles on  heater plane        3 Assemble the scaleboard on the heater plane  and fix by screws  3 8 mm            4 As per Table 3  assemble heater PCBA and  wires              5 Put the humidifier inner connector on the  shield  LS       6 Put on light window        7 Assemble the inner and outer shield              8 Fix on the screws  M3 10mm   a
10. n Device  Humidifier Plug  right down  220008 1 PC Rubber Main Device  Device Pad 220001 6 PC Rubber All  Key Button 220002 1 PC Rubber Main Device  Blower Connector 220003 1 PC Rubber Main Device  Humidifier Inner Connector 220009 i PC Rubber Humidifier  Screw 1 8 PC Screw All  Screw 2 9 PC Screw All  Sunk Screw 4 PC Screw Main Device                      17 17       
11. ng     Safety Notice    Electric Shock  There is high voltage over 100VAC inside this device  please be sure to repair the device after power  off     Chemical Safety   There may be risk of harmful bacteria and viruses after using by patient  please clean the device or  wear protective glove before servicing and repairing    Please wash your hands with disinfectant after operating the device    Please deal with the waste according to your local regulation     Service Manual for RESmart Auto CPAP CPAP V1 1       1  Introduction    1 1 Intended Use    The RESmart  Auto CPAP CPAP system is a CPAP  Continuous Positive Airway Pressure  device  designed for the treatment of adult Obstructive Sleep Apnea  OSA  only    The RESmart  Auto CPAP CPAP system is to be used only on the instruction of a licensed health  care professional  Your home care provider will configure suitable pressure settings according to  your health care professional s prescription     This service manual is used to help service engineers to maintain RESmart  Auto CPAP CPAP  system more efficiently  Instructions in this manual may help the device to work in the best condition   Service engineers can find instructions they need from this manual quickly     IMPORTANT   Read and understand the instructions in this manual before operating the device     IMPORTANT   This device is not intended for life support     1 2 Implements    1  Implements for repairing  Type         screwdriver  Pincers    2  Implements for me
12. r is matching   the humidifier will be dis matched     2  Reset  Available on version 1 28 and later     Press and hold the Ramp Button until the user menu appears  about 3 seconds  when the device is  on standby  press     User Buttons to access the Date Setting  set    YYYY MM DD     2097 07 01    then press    Pressure Start Stop Button    when    Hour    setting appears  thus all device settings and  therapy records  doesn t include Use Days and Use Time  will be reset     If set    Minute     16  during the above procedure  then not only device settings and therapy records   but also patient information will be reset     If set    Minute     26  during the above procedure  then not only device settings  therapy records and  patient information  but also Use Days and Use Time will be reset     3 17    Service Manual for RESmart Auto CPAP CPAP V1 1       3  Pressure calibration  Available on version 1 58 and later     Connect pressure meter to the air outlet of the device  in the Date and Time Setting menu  set   YYYY MM DD   2098 08 18   press    Pressure Start Stop Button    when Hour Setting appears to start  the Pressure Calibration  The device output pressure is based on 20 hPa  cmH20   and a number will  appear on the screen  when the output pressure is stable  if the value on the pressure meter is not 20  hPa  cmH20   press       User Buttons to adjust the output pressure   finally press    Pressure  Start Stop Button    to finish the Pressure Calibration     4  Mal
    