11752D 3.5 mm Connector Gage Operating and Service Manual
Contents
1. Agilent Technologies 11752D 3 5 mm Connector Gage Operating and Service Manual Manual Part Number 11752 90021 Printed in USA July 2001 Supersedes June 2000 Notice The information contained in this document is subject to change without notice Agilent Technologies makes no warranty of any kind with regard to this material including but not limited to the implied warranties of merchantability and fitness for a particular purpose Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing performance or use of this material Agilent Technologies assumes no responsibility for the use or reliability of its software on equipment that is not furnished by Agilent Technologies This document contains proprietary information which is protected by copyright All rights are reserved No part of this document may be photocopied reproduced or translated to another language without prior written consent of Agilent Technologies RESTRICTED RIGHTS LEGEND Use duplication or disclosure by the U S Government is subject to restrictions as set forth in subparagraph c 1 ii of the Rights in Technical Data and Computer Software clause at DFARS 252 227 7013 for DOD agencies and subparagraphs c 1 and c 2 of the Commercial Computer Software Restricted Rights clause at FAR 52 227 19 for other agencies Agilent Technologies Inc 14
2. is lined up with the zero indicator on the dial CONNECTOR GAGE GAGE MASTER LIMIT POINTERS ZERO SET KNOB HOLD BY BARREL Figure I Agilent 11752D Connector Gage Features 6 Aligent 11752D Operating and Service Manual Measuring Connectors NOTE Push on Measurement Operation Gages should be checked often to make sure that the zero setting has not changed Generally when the gage pointer on a gage that has been recently zeroed does not line up exactly with the zero mark the gage or calibration block needs cleaning Measuring the recession of the center conductor behind the outer conductor mating plane in a connector is done in exactly the same way as zeroing the gage except the zero set knob is not re set when the measurement is made Hold the gage by the barrel only and torque only the connecting nut do not turn the Gage or the device to 8 lb in 00 N cm Recession or protrusion will show as a reading counterclockwise or clockwise respectively from zero see Figure 1 Protrusion of center conductor shoulder male pin and end of female center pin is not allowable on any 3 5 mm connector For maximum accuracy measure the connector several times and take an average of the readings The push on feature of the 11752D female gage should only be used to obtain a quick measurement of the recession of the center conductor It can also be used to verify the presence of a protrusion If an accurate mea
3. 00 Fountaingrove Parkway Santa Rosa CA 95403 1799 U S A Copyright Agilent Technologies Inc 2000 ii Agilent 11752D Operating And Service Manual In This Manual Overview page 1 Instrument Characteristics page 2 Installation page 3 Operation page 4 Replaceable Parts page 9 Agilent 11752D Operating And Service Manual iii Certification Warranty Warranty Custom systems are warranted by contractual agreement between Agilent Technologies and the customer Agilent Technologies Inc certifies that this product met its published specifications at the time of shipment from the factory Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology NIST formerly NBS to the extent allowed by the Institute s calibration facility and to the calibration facilities of other International Standards Organization members This Agilent Technologies system product is warranted against defects in materials and workmanship for a period corresponding to the individual warranty periods of its component products Instruments are warranted for a period of one year During the warranty period Agilent Technologies will at its option either repair or replace products that prove to be defective Warranty service for products installed by Agilent Technologies and certain other products designated by Agilent Tec
4. ING notice until the indicated conditions are fully understood and met CAUTION The CAUTION notice denotes a hazard It calls attention to an operating procedure practice or the like which if not correctly performed or adhered to could result in damage to the product or loss of important data Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met Instrument Markings When you see tis symbol on your Instrument you should reer to the instruments instruction manual for important information Pa AN pls This symbol indicates hazardous voltages This symbol indicates that the instrument requires alternating current ac input Pd The CE mark is a registered trademark of the European Community If it is accompanied by a year it indicates the year the design was proven The C Tick mark is a registered trademark of the Australian Spectrum Management Community The CSA mark is a registered trademark of the Canadian Standards Association ISM1 A This text indicates that the instrument is an Industrial Scientific and Medical Group 1 Class A product CISPER 11 Clause 4 Agilent 11752D Operating And Service Manual vii This symbol indicates that the power Tine switch is ON This symbol indicates that the power line switch is in STANDBY position This symbol indicates that the power Tine switch is OFF Safety Earth This is a Safety Clas
5. ensure the year 2000 readiness of its information technology and business environment The duration of this warranty extends through January 31 2001 The remedies available under this warranty will be defined in and subject to the terms and limitations of the warranties contained in the contract of sale To the extent permitted by local law this warranty applies only to branded Agilent Technologies products and not to products manufacture by others that may be sold or distributed by Agilent Technologies Nothing in this warranty will be construed to limit any rights or remedies provided elsewhere in the contract of sale with respect to matters other than year 2000 compliance Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products For assistance call your local Agilent Technologies Sales and Service Office refer to Service and Support on page vi Agilent 11752D Operating And Service Manual v Service and Support Any adjustment maintenance or repair of this product must be performed by qualified personnel Contact your customer engineer through your local Agilent Technologies Service Center You can find a list of local service representatives on the Web at http www agilent com find assist If you do not have access to the Internet one of these centers can direct you to your nearest representative United States tel 1 800 452 4844 Latin Ame
6. hnologies will be performed at Buyer s facility at no charge within Agilent Technologies service travel areas Outside Agilent Technologies service travel areas warranty service will be performed at Buyer s facility only upon Agilent Technologies prior agreement and Buyer shall pay Agilent Technologies round trip travel expenses In all other areas products must be returned to a service facility designated by Agilent Technologies For products returned to Agilent Technologies for warranty service Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologies shall pay shipping charges to return the product to Buyer However Buyer shall pay all shipping charges duties and taxes for products returned to Agilent Technologies from another country Agilent Technologies warrants that its software and firmware designated by Agilent Technologies for use with an instrument will execute its programming instructions when properly installed on that instrument Agilent Technologies does not warrant that the operation of the instrument or software or firmware will be uninterrupted or error free LIMITATION OF WARRANTY The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer Buyer supplied software or interfacing unauthorized modification or misuse operation outside of the environmental specifications for the product or improper site preparation or maintenance i
7. lly as well as the connector you are about to measure Foreign material on the gage gage master or connector will make the measurement inaccurate To clean connector surfaces we recommend using compressed air directing the air where it is needed through a plastic not metal nozzle If the compressed air does not remove the foreign material use a cotton swab or if the swab is too large a round wooden toothpick do not use metal wrapped in a single layer of lint free cleaning cloth soaked in isopropyl alcohol See Table 2 Dry the connector and gage surfaces with a brief blast of the compressed air Recommended Cleaning Supplies Part Number Description 8500 0559 Isopropyl alcohol 5080 5400 Cotton swabs 9310 4242 Lint free cleaning cloth Connect the gage to the appropriate gage master and tighten only the connecting nut do not turn gage or master until finger tight Lay the gage and master on a workbench Use two wrenches one standard 5 16 in open end and one 5 16 in torque wrench set to 8 Ib in 90 N cm While holding the fixed nut motionless with the open end wrench use the torque wrench on the connecting nut The connecting nut for the male gage is located on the Aligent 11752D Operating and Service Manual 5 Operation gage master The connecting nut for the female Gage is located on the Gage itself Torque tile connecting nut to 8 lb in 90 N cm Turn the Zero Set Knob Figure 1 until the indicator
8. nd materials identical to those used in factory packaging are available through Agilent offices If the Connector Gage is being returned to Agilent for calibration or service attach a tag indicating the type of service required return address model number and serial number Also mark the container FRAGILE to assure careful handling In any correspondence refer to the instrument by model number and serial number The connector gage should be stored in a clean dry environment The following limitations apply to both storage and shipment Temperature 55 to 75 C Relative humidity less than 95 at 40 C Altitude less than 15 3 km 50 000 ft Aligent 11752D Operating and Service Manual 3 Operating Environment NOTE Operating Precautions CAUTION CAUTION CAUTION Operation Operation For instruction on connecting and care of microwave connectors refer to Application Note 326 Principles 0 Microivaw Connector Care part number 5954 1566 Also available is Microlvave Connector Care part number 08510 90064 Device dimensions change with temperature Therefore the temperature of the gages and all connectors must be stable within 1 C before calibration The operating environment for the connector gage should be within the following limits Temperature 20 to 26 C 1 C 68 to 79 F 1 8 F Relative humidity less than 80 at 26 C dry bulb Altitude less than 4 5 km 15 000 ft Y
9. our fingers are a heat source Thus handle the devices as little as possible during calibration To protect the sensitive microcircuits that may be connected to the connectors you are calibrating always wear an electrostatic discharge grounding strap connected to a conductive bench mat when working near sensitive equipment Do not use hydrocarbons such as acetone trichlorethylene carbon tetrachloride or benzene to clean connector or gage surfaces See Inspecting and Cleaning the Gages on the next page Do not spray any liquid solvent directly onto connector or gage surfaces 4 Aligent 11752D Operating and Service Manual Using Connector Gages Inspecting and Cleaning the Gages Table 2 Zeroing the Gage Operation Connector wear eventually degrades performance To prolong the life of your gages 1 Clean all surfaces that comes in contact during calibration 2 Do not measure a damaged connector doing so could damage the gage Replace damaged connectors 3 Keep the protector caps on the gages and masters when not in use to prevent contamination 4 Turn only the connector nut not the gage itself when making connection to avoid damage to the connector and gage Torque should not exceed 81b in 90 N cm During normal use the connector gages should be inspected cleaned and zeroed prior to taking measurements Using an illuminated magnifying glass inspect the connector gage and the gage master carefu
10. repairs are needed contact your nearest Agilent sales and service office listed inside the cover of this manual We recommend the gages be calibrated annually at an Agilent Technologies service office A listing of Agilent sales and service offices is located inside the cover of this manual 10 Aligent 11752D Operating and Service Manual
11. rica Canada tel 305 269 7500 fax 305 269 7599 tel 1 877 894 4414 fax 905 282 6495 Europe tel 431 20 547 2323 fax 31 20 547 2390 New Zealand tel 0 800 738 378 fax 64 4 495 8950 J apan tel 81 426 56 7832 fax 81 426 56 7840 Australia tel 1 800 629 485 fax 61 3 9210 5947 Asia Call Center Numbers Country Phone Number Fax Number Singapore 1 800 375 8100 65 836 0252 Malaysia 1 800 828 848 1 800 801664 Philippines 632 8426802 632 8426809 1 800 16510170 PLDT Subscriber Only 1 800 16510288 PLDT Subscriber Only Thailand 088 226 008 outside Bangkok 66 1 661 3714 662 661 3999 within Bangkok Hong Kong 800 930 871 852 2506 9233 Taiwan 0800 047 866 886 2 25456723 People s Republic of China 800 810 0189 preferred 10800 650 0021 10800 650 0121 India 1 600 11 2929 000 800 650 1101 vi Agilent 11752D Operating And Service Manual Safety and Regulatory Information Review this product and related documentation to familiarize yourself with safety markings and instructions before you operate the instrument This product has been designed and tested in accordance with international standards WARNING The WARNING notice denotes a hazard It calls attention to a procedure practice or the like that if not correctly performed or adhered to could result in personal injury Do not proceed beyond a WARN
12. s I product provided with a protective earthing Ground 2 terminal An uninterruptible safety earth ground must be provided from the main power source to the product input wiring terminals power cord or supplied power cord set Whenever it is likely that the protection has been impaired the product must be made inoperative and secured against any unintended operation Before Applying Power Verify that the product is configured to match the available main power source as described in the input power configuration instructions in this manual If this product is to be powered by autotransformer make sure the common terminal is connected to the neutral grounded side of the ac power supply viii Agilent 11752D Operating And Service Manual Description Instruments Covered Manual Changes Supplement Overview This operating manual contains information about initial inspection performance tests adjustments operation and troubleshooting of the Agilent 11752D connector gage The 11752D connector gage is used to measure connector pin depth of 3 5 mm connectors The connector gage contains 3 5 mm connector gage male 3 5 mm connector gage female 3 5 mm gage master male 3 5 mm gage master female Protective caps Storage box This instrument has a two part serial number The first four digits and the letter comprise the serial number prefix The last five digits form a sequential suffix which is unique to each in
13. strument The contents of this manual apply directly to instruments having the serial number prefix 2923A A connector gage manufactured after the printing of this manual may have a serial number prefix that is not listed If your connector gage has a prefix number different from 2923A look for a yellow manual changes supplement that documents the differences In addition to change information the supplement may contain information for correcting errors in the manual To keep this manual as current and accurate as possible Agilent recommends that you periodically request the latest manual changes supplement The supplement is keyed to the manual print date code E0489 April 1989 and part number both of which appear on the front cover Copies of the supplement are available on request from your nearest Agilent Technologies office For information concerning a serial number prefix not listed here or in the Manual Changes supplement contact your nearest Agilent Technologies Office Aligent 11752D Operating and Service Manual 1 NOTE Instrument Characteristics Instrument Characteristics The characteristics listed in Table 1 represent the typical characteristics or conditions that describe gage performance Gage resolution depends on your ability to discern the position of the gage needle between divisions The specification for this characteristic is based on tests of many people making the same measurement and independentl
14. surement is required the connecting nut must be threaded and torqued to the connector being measured Aligent 11752D Operating and Service Manual 7 Operation To use the push on feature simply retract the connecting nut as shown in Figure 2 While holding the gage by the barrel see Figure 1 push the connector gage gently firmly and evenly onto the center conductor of the connector being measured P1752 60002 3 5mm FEMALE CONNECTOR Gar Figure 2 Female gage with connecting Nut Retracted Helpful hints Do not twist or rock the gage or apply pressure to the gage head Uneven pressure on the center conductor or gage head will result in an uneven measurement Practice Use the female master to get a feel for the amount of pressure needed to obtain a zero reading on the gage 8 Aligent 11752D Operating and Service Manual Replaceable Paris Replaceable Parts All replaceable parts are shown in Figure 3 There are no internal replaceable parts within the gages themselves STORAGE CASE 11752 80024 FQAM SET 11752 80006 MALE GAGE FEMALE GAGE 11752 60020 11752 60021 PROTECTIVE CAP 1401 0200 MALE GAGE FEMALE GAGE MASTER MASTER 11752 60022 11752 60023 Figure 3 Agilent 11752D Connector Gage and Accessories Aligent 11752D Operating and Service Manual 9 Service Calibration Replaceable Parts Do not attempt to repair the gage or gage master Any attempt to do so will void warranty If
15. v Agilent 11752D Operating And Service Manual Assistance NO OTHER WARRANTY IS EXPRESSED OR IMPLIED AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE EXCLUSIVE REMEDIES THE REMEDIES PROVIDED HEREIN ARE BUYER S SOLE AND EXCLUSIVE REMEDIES AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT INDIRECT SPECIAL INCIDENTAL OR CONSEQUENTIAL DAMAGES WHETHER BASED ON CONTRACT TORT OR ANY OTHER LEGAL THEORY YEAR 2000 Agilent Technologies warrants that each Agilent Technologies hardware software and firmware product on Agilent Technologies Corporate Price List dated July 1 1998 or later delivered under the product s contract of sale will be able to accurately process date data including but not limited to calculating comparing and sequencing from into and between the twentieth and twenty first centuries and the years 1999 and 2000 including leap year calculations when used in accordance with the product documentation provided that all other products that is hardware software firmware used in combination with such Agilent Technologies product s properly exchange date data with it If the agreement requires that specific Agilent Technologies products must perform as a system in accordance with the foregoing warranty then that warranty will apply to those Agilent Technologies products as a system and Customer retains sole responsibility to
16. y indicating their readings Table 1 Characteristics Characteristics and Conditions Typical Limits Comments Measurement range 0 005 in 0 127 mm to In 0 0001 in 0 00254 mm increments 0 005 in 0 127 mm Gage resolution 0 00002 in 0 00051 mm Fifth of an increment Gage calibration accuracy 0 00004 in 0 00102 mm Gage repeatability 0 000025 in 0 00064 mm Quarter of an increment Master accuracy 0 00004 in 0 00102 mm Total uncertainty Worst case RSS 0 000125 in 0 00318 mm Sum of resolution repeatability Gage and 0 000062 in 0 00157 mm master accuracy limits Root sum of the squares Operating Environment Temperature range Humidity at 26 C maximum dry bulb Maximum altitude 20 to 26 C No temperature fluctuations greater than 1 C 0 to 80 4 5 km 15 000 ft Net weight 1 kg 2 lb 3 oz 2 Aligent 11752D Operating and Service Manual Initial Inspection Original Packaging Storage and Shipping Environment Installation Installation Inspect the shipping container for damage If the shipping container or packaging material is damaged it should be kept until the contents of the shipment have been checked If there is mechanical damage or if the Gages do not pass the performance tests notify the nearest Agilent office Keep the damaged shipping materials if any for inspection by the carrier and an Agilent representative Containers a