0340 Electron Capture Detector (ECD)
Contents
1. Figure 340 5 Disconnecting the makeup gas adapter 6 of 24 Detectors Jun 2001 Agilent 6890 Gas Chromatograph Service Manual Electron Capture Detector ECD 340 Replacement procedures Disconnect the ECD vent tube at the rubber sleeve Use a 5 16 inch wrench to disconnect the 1 16 inch Swagelok union on the anode purge line Remove the one Torx T 20 screw securing the top ECD detector cover to the detector pallet and remove the cover Disconnect the anode signal wire from the detector f Anode purge line Detector top cover disconnect fitting Remove ECD vent tube Remove anode signal wire Figure 340 6 Removing the cover anode purge fitting vent tube and anode signal 10 11 wire Fully loosen the four Torx T 20 screws on the detector pallet and lift the pallet and detector from the GC Thread the heater sensor leads out of the detector pallet and remove the detector from the pallet Jun 2001 Detectors 7 of 24 Agilent 6890 Gas Chromatograph Service Manual 340 Electron Capture Detector ECD Replacement procedures WARNING The ECD detector is now properly disassembled for replacement or exchange DO NOT remove the inner thermal cover if you are replacing the entire detector assembly Replacing the heater sensor assembly After removing the ECD detector from the GC you can further disassemble it to replace the heater sensor assembly
2. Agilent 6890 Gas Chromatograph Service Manual 340 Electron Capture Detector ECD Diagnostics If the ECD frequency indicates contamination 2100 check for the following e Contaminated carrier gas trap s and or supply replace carrier gas supply tank and any traps on the carrier supply line e Insufficient column conditioning fully condition the column e Contaminated detector bake out the detector e Column inlet and or septum bleed clean the inlet replace the septum with a conditioned low bleed septum e Leaks perform leak tests on both the inlet and detector systems e Anode current leakage make sure the anode contacts are clean Make sure the anode nut is tight Leak test Note Once you have determined that the flow system components upstream from the detector gas supply tubing inlet column fittings are leak free perform the following ECD detector leak test 1 With the GC on and operating normally set the oven detector and inlet temperatures to ambient 2 Turn off the ECD and then turn off the inlet pressure 3 Turn off the anode and makeup gas flows 4 Cap the ECD exhaust vent with a vent plug part no 5060 9055 5 Set carrier gas pressure at the inlet corresponding to the ECD to 15 psi 103 kPa 6 Wait until the system reaches the setpoint pressure and then turn off the pressure and monitor the actual pressure value for at least 10 minutes 18 of 24 Detectors Jun 2001 Agilent 6890
3. The ECD cell contains radioactive Ni To reduce the risk of exposure wear disposable gloves while handling the ECD cell When you are finished dispose of the gloves and wash your hands with soap and water 1 Remove the detector as described in the Replacing the entire detector detector cell procedure in this section 2 Loosen the locking tab screw on top of the detector slide the locking tab back and pivot it out of the away Figure 340 7 Unlocking the thermal cover 3 Lift the thermal cover up and carefully slide it off of the anode assembly 8 of 24 Detectors Jun 2001 Agilent 6890 Gas Chromatograph Service Manual Electron Capture Detector ECD 340 Replacement procedures Heater sensor assembly Heated blocks Figure 340 8 Removing the heater sensor 4 Slide the heater and sensor out of the heated blocks Jun 2001 Detectors 9 of 24 Agilent 6890 Gas Chromatograph Service Manual 340 Electron Capture Detector ECD Replacement procedures Replacing the makeup gas adapter After removing the ECD detector from the GC you can further disassemble it to replace the makeup gas adapter The makeup gas adapter consists of a line from the detector pneumatics manifold that carries makeup gas to a weldment that screws into the bottom of the ECD detector From there the makeup gas sweeps past the end of the column and carries the column effluent into the ECD cell 1 Remove the
4. To prevent possible hazardous contamination of the area with radioactive material the detector exhaust vent must always be connected to a fume hood or otherwise vented in compliance with the latest revision of Title 10 CFR Part 20 or with state regulations with which the Nuclear Regulatory Commission has entered into an agreement USA only For other countries consult with the appropriate agency for equivalent requirements 1 Record the ECD Output value from the GC display If the number is equal to or greater than 100 continue with this procedure 2 Turn off the anode purge and makeup gas flows 3 Remove the column from the detector 4 Cap the bottom of the makeup gas adapter with a blank column ferrule and column nut 5 Setthe makeup gas flow rate between 50 and 60 mL min Set the detector temperature between 350 and 375 C Jun 2001 Detectors 23 of 24 Agilent 6890 Gas Chromatograph Service Manual 340 Electron Capture Detector ECD Maintaining an ECD detector 6 Set the oven temperature to 250 C 7 Allow thermal cleaning to continue for several hours and then cool the system to normal operating temperatures Performing a radioactivity leak test wipe test ECDs must be tested for radioactive leakage at least every six months Records of tests and results must be maintained for possible inspection by the Nuclear Regulatory Commission and or responsible state agency More frequent tests may be condu
5. detector as described in the Replacing the entire detector detector cell procedure in this section Remove the Torx T 20 screw holding the pneumatics block s to the detector manifold Note There are two pneumatics blocks on the EPC version of the ECD pneumatics manifold The outside block is the anode purge gas line and the inside block is the makeup gas line 3 Slide the makeup gas adapter up and out of the GC 4 When reinstalling the makeup gas adapter ensure the following e Approximately 6 inches of the makeup gas line should reside in the oven after installation e The makeup gas line should be bent into a coil that loops around the bottom of the detector weldment and makeup gas adapter 10 of 24 Detectors Jun 2001 Agilent 6890 Gas Chromatograph Service Manual Electron Capture Detector ECD 340 Replacement procedures Column aie should C protrude about 1 mm Makeup gas Jj adapter U Coil loops around _______ y S f pe bottom of detector and Column nut ER makeup gas adapter Column connection Figure 340 9 Proper configuration of the column and makeup gas tubing e The end of the column should protrude about 1 mm from the top of the makeup gas adapter The total dimension from the back of the column nut to the end of the column will be about 75 to 76 mm Removing the EPC flow manifold The ECD detector uses a Type 1 EPC flow manifold that contains one inlet supply fitting for a purge makeup gas W
6. method If the baseline is acceptable free of peaks and humps and under 1000 Hz throughout a temperature program then the detector and makeup system are clean Any unacceptable baseline problems could indicate contaminated makeup gas EPC module makeup adapter or detector These must be addressed before continuing Jun 2001 Detectors 1 of 24 Agilent 6890 Gas Chromatograph Service Manual 340 Electron Capture Detector ECD Troubleshooting contamination problems If the 6890 GC was manufactured before 6 97 the EPC modules may have O ring contamination See Service Note G1530 14 and follow the procedure given Replace the EPC module Evaluate the carrier side After the detector and the makeup system have been determined to be clean evaluate the carrier inlet and column 1 Remove the makeup adapter and capped off column from the detector Discard the cap Place a new ferrule on the column trim the column end and install it so that 1 to 2 mm extends past the adapter tip Itis best not to use the gigabore liner A better solution is the mixing liner part no G2397 20540 for the u ECD Use the column installation instructions for the u ECD in Volume 1 of the Operating Manual Wipe off the entire makeup adapter with methanol Install it fully check the measurement to be sure Bake out the entire system for another hour at these conditions e Detector 350 C e Inlet Split mode 275 C e Column An appro
7. 340 Electron Capture Detector ECD Theory of operation The ECD is based on the phenomenon that electronegative species can react with thermal electrons present to form negatively charged ions The loss of such electrons is related to the quantity of analyte in the sample In order to produce capturable low energy thermal electrons the carrier gas is ionized by beta particles from a radioactive source in the cell This electron flow produces a current which is collected and measured When the sample molecule is introduced into the cell electrons which would otherwise be captured at the electrode are captured by the sample resulting in decreased current This change is what is recorded and measured for the chromatogram Pneumatic lines Vent line Detector body under cover I Electrometer Electrometer ribbon cable Heater sensor cable Figure 340 1 The Electron Capture Detector ECD Jun 2001 Detectors 1 of 24 Agilent 6890 Gas Chromatograph Service Manual 340 Electron Capture Detector ECD Theory of operation EPC detector The EPC version of the ECD detector has one flow line for the anode purge makeup gas After an initial filter frit it splits into two paths Each line has a proportional valve controlled by a pressure sensor and a non adjustable restrictor frit The makeup gas line sweeps past the end of the column and carries the column effluent into the ECD cell The anode purge flo
8. ARNING Before proceeding turn off the oven and any heated zones and let them cool down Turn off any detector gases at their supply then turn off the main power switch and unplug the power cord 1 Remove the top plastic covers from the detector and pneumatics areas Also remove the RFI metal shield and the top rear metal cover Jun 2001 Detectors 11 of 24 Agilent 6890 Gas Chromatograph Service Manual 340 Electron Capture Detector ECD Replacement procedures 2 Remove the gas supply fitting from the side of the manifold 3 At the front of the manifold remove the Torx T 20 screw that holds the manifold in its slot Caution Make sure you are properly grounded with an ESD strap before continuing 4 Unlock the detector manifold s ribbon cable from the EPC board and detach the connector The adjacent ribbon cable may have to be removed as well 5 Remove the Torx T 20 screw holding the output pneumatics block s to the manifold and remove the block 6 Slide the manifold from its slot Note There are two pneumatics blocks on the EPC version of the ECD pneumatics manifold The outside block is the anode purge line and the inside block is the makeup gas line 12 of 24 Detectors Jun 2001 Agilent 6890 Gas Chromatograph Service Manual 340 Electron Capture Detector ECD Replacement procedures Figure 340 10 Removing the detector flow manifold 13 of 24 Detectors Jun 2001 Agilent 6890 Gas Chroma
9. Gas Chromatograph Service Manual Electron Capture Detector ECD 340 Diagnostics 7 Check for pressure drop e Ifthe pressure stays stable or drops only 0 5 psi you can consider the ECD leak free e Ifthe pressure drops more than 0 5 psi you have a leak If you are sure none of the upstream flow system components are leaking check for leaks at the column fitting and plugged inlet If you find leaks tighten the fittings and repeat the leak test Note If you can find no other leaks the ECD itself is probably leaking The ECD cannot be disassembled without special license from the Nuclear Regulatory Commission or Agreement State Licensing Agency USA only Jun 2001 Detectors 19 of 24 Agilent 6890 Gas Chromatograph Service Manual 340 Electron Capture Detector ECD Troubleshooting contamination problems Troubleshooting contamination problems Persistent problems with high background or ghost peaks with temperature programming are almost always due to contamination from dirty samples consumables or the carrier makeup gas systems Begin with the procedure in chapter 5 Volume 3 of the Operating Manual If this procedure which can be performed by the user does not solve the problem perform the following steps Ensure clean gas supplies Before continuing verify that the supply gases are of adequate purity 1 Carrier and makeup purity must be gt 99 999 2 After confirming purity verify that the tank re
10. LI ge ae Detector card screw Ribbon cable ECD detector board HW AR SS NES AR Figure 340 13 Detector interface board installed in back detector position 16 of 24 Detectors Jun 2001 Agilent 6890 Gas Chromatograph Service Manual Electron Capture Detector ECD 340 Diagnostics Diagnostics Frequency test Perform this test to make sure that the base frequency for the ECD during a blank run indicates a relatively contaminant free system Note It may take 24 hours for the ECD baseline to completely stabilize especially if you are starting with a cold system and want to assure high sensitivity operation Therefore for the most accurate results run the detector at normal operating conditions for as long as possible at least 2 hours and up to 24 hours before running the frequency test If you will be injecting into an unused inlet you must use low bleed septa Make sure to condition new septa before use in an inlet for several hours with 1 to 5 mL min carrier flow 1 Make sure you are using normal operating conditions and that at least two hours have elapsed since the last run 2 Turn onthe ECD and the corresponding signal 3 Check the displayed Output e lt 100 ECD frequency is acceptable e 2100 Contaminants in system Note Each display count equals a frequency of 5 Hertz e g a display reading of 100 500 Hz Jun 2001 Detectors 17 of 24
11. cted when necessary The procedure used is the wipe test A Wipe Test Kit part no 18713 60050 is supplied with each new ECD Refer to the information card supplied in the Wipe Test Kit for instructions on performing the wipe test 24 of 24 Detectors Jun 2001 Agilent 6890 Gas Chromatograph Service Manual
12. detector cell Before proceeding turn off the oven and any heated zones and let them cool down Turn off any detector gases at their supply then turn off the main power switch and unplug the power cord The ECD cell contains radioactive Ni To reduce the risk of exposure wear disposable gloves while handling the ECD cell When you are finished dispose of the gloves and wash your hands with soap and water 1 Remove the detector cover the electronics carrier cover and the right side cover Make sure you are properly grounded with an ESD strap before continuing 2 Disconnect the electrometer ribbon cable from the ECD interface card 3 Disconnect the heater sensor leads from the connector on the right side of the GC 4 of 24 Detectors Jun 2001 Agilent 6890 Gas Chromatograph Service Manual Electron Capture Detector ECD 340 Replacement procedures Heater sensor cable Ribbon cable connector Figure 340 4 Disconnecting the ECD cables 4 Inside the oven remove the insulation cup and disconnect the column from the makeup gas adapter 5 Use a 9 16 inch wrench to loosen the 1 4 inch Swagelok nut on the makeup gas adapter from the bottom of the detector Slide the makeup gas adapter out of the bottom of the detector Jun 2001 Detectors 5 of 24 Agilent 6890 Gas Chromatograph Service Manual 340 Electron Capture Detector ECD Replacement procedures Unscrew Swagelok nut
13. gulators have stainless steel diaphragms equivalent to Agilent part no 8507 0407 3 Install new 1 8 inch copper supply tubing part no 5180 4196 Many times clean tubing from other sources has caused high ECD background At the same time install new traps in both the carrier and makeup supplies Place the moisture trap part no 5060 9084 closest to the tank and the indicating oxygen trap part no 3150 0528 closest to the GC Leak test the entire plumbing setup very carefully Isolate problem to carrier or makeup gas supplies Determine what components of the apparent contamination are from the carrier vs makeup systems in the GC Sharp well resolved peaks that elute during a temperature program with no injection are from the carrier inlet system Broader humps in the baseline are usually from the makeup system Overall high background gt 500 Hz can be contaminated gas from either the carrier or makeup supply or a contaminated detector cell Remove the column from the detector and inspect the installation of the makeup gas adapter It is quite common for the adapter to be installed too low To check this measure from the bottom of the 1 4 inch Swagelok nut to 20 of 24 Detectors Jun 2001 Agilent 6890 Gas Chromatograph Service Manual Electron Capture Detector ECD 340 Troubleshooting contamination problems the bottom of the hex of the makeup gas adapter The measurement should be 19 to 20 mm If it exceeds 22 m
14. m the adapter is installed incorrectly A ridge inside the cell can prevent the adapter from easily seating all the way Wiggle the adapter while installing to allow it to go all the way in Always check the nut to hex measurement to be sure Evaluate the makeup side 1 Remove the makeup adapter from the detector Unscrew the tip and remove the Gigabore liner Inspect the adapter body for carbon graphite deposits Remove all graphite deposits and clean the adapter thoroughly with methanol Soak the removable tip in methanol 2 Reassemble the adapter without the gigabore liner Install a 5 inch length of narrow bore column capped with a new septum so that the other end extends 1 mm past the tip of the adapter Use a Vespel column ferrule rather than graphite Tighten the tip just past finger tight with clean pliers Clean the whole assembly with methanol before installing 3 Install the capped off makeup adapter using a new 1 4 inch Vespel ferrule Be sure it is fully seated check the measurement Retighten after the detector has been heated 4 Set the makeup flow to the original setpoint and bake out the detector at 350 C for 1 hour During this time put the inlet in split mode with 200 to 300 mL min split vent flow gas saver off and bakeout the inlet at 275 C Bakeout the column at its appropriate temperature 5 When bakeout is done do not reconnect the column to the detector Make a series of blank runs with the user s
15. priate temperature Reload the user s method and make a series of blank no injection runs to see if the problem has been cured Note that a single well resolved peak could be due to the O ring contamination problem mentioned earlier Address per service note If the contamination persists perform a complete inlet maintenance including thorough cleaning of the shell weldment Replace the gold seal and liner Install a known good 30 m 320 um HP5 checkout column to rule out column contamination Peaks from the inlet side are usually due to contaminated carrier gas supply EPC module insert weldment inlet or liner or column 22 of 24 Detectors Jun 2001 Agilent 6890 Gas Chromatograph Service Manual Electron Capture Detector ECD 340 Maintaining an ECD detector Maintaining an ECD detector Caution WARNING ECD bakeout If your ECD baseline is noisy or the display frequency is too high i e 2100 you should perform a thermal cleaning also called a bakeout of the detector Before performing a bakeout verify that the carrier supply gas and flow system are leak and contaminant free Detector disassembly and or cleaning procedures other than thermal should be performed only by personnel trained and licensed appropriately to handle radioactive materials Trace amounts of radioactive Ni may be removed during these other procedures causing possible hazardous exposure to B and x radiation bremsstrahlung
16. tograph Service Manual 340 Electron Capture Detector ECD Replacement procedures Removing the signal board 1 Remove both the electronics top cover and the right side cover 2 Remove the Torx T 20 screw securing the top cover and remove the cover 3 Disconnect the signal wire from the signal board interconnect Remove detector top cover Remove anode signal wire Figure 340 11 Removing the top cover and anode signal wire 14 of 24 Detectors Jun 2001 Agilent 6890 Gas Chromatograph Service Manual Electron Capture Detector ECD 340 Replacement procedures 4 Remove the screw and clamp on the electrical interconnect 5 Remove one Torx T 20 screw from each end of the signal board Do not remove the screw on the top of the cover Figure 340 12 Removing the ECD signal board 6 Unlock and detach the ECD signal board s ribbon cable from the detector s interface board and lift the signal board from the detector pallet 7 Reassembly is the reverse of removal Jun 2001 Detectors 15 of 24 Agilent 6890 Gas Chromatograph Service Manual 340 Electron Capture Detector ECD Replacement procedures Replacing the detector interface card The ECD detector interface board plugs into the main board as shown below 1 Unlock and unclip the ribbon cable from the interface board 2 Remove the screw at the top of the card where it mounts to the main board and pull the card out Va
17. w sweeps the upper part of the ECD and cleans sample deposition from the detector Filter Proportional Pressure ARNORG frit valves sensors Restrictors purge and makeupgas P Ea PS AN s fl p in f Anode purge Vent 63Ni plating Makeup PS NN Ei Capillary adapter Pressure control loops Figure 340 2 EPC ECD flow diagram Manually controlled detector The manually controlled ECD has a pneumatic supply line for the anode purge gas and for the makeup gas Both lines have an on off solenoid valve and a nonadjustable restrictor frit A pressure regulator allows for adjusting the 2 of 24 Detectors Jun 2001 Agilent 6890 Gas Chromatograph Service Manual Electron Capture Detector ECD 340 Theory of operation makeup gas and anode purge flows The makeup gas line sweeps past the end of the column and carries the column effluent into the ECD cell Pressure regulator On off solenoid Anode 1 valve Restrictors purge and Be gt lt makeup E gas in Vent 63Ni plating On off solenoid valve Restrictors Capillary adapter Figure 340 3 Manual ECD flow diagram Jun 2001 Detectors 3 of 24 Agilent 6890 Gas Chromatograph Service Manual 340 Electron Capture Detector ECD Replacement procedures Replacement procedures WARNING WARNING Caution Replacing the entire detector
Download Pdf Manuals
Related Search
Related Contents
Snapper 4219 User's Manual Scheda tecnica - Chiudi e vai sul sito ien TeSys motor starters User Manual - Austin Hughes À vos pinceaux! Vinyle en rouleau : mode d`emploi Un entrepreneur SoftBank 105SH 取扱説明書 TEFAL FV4880D0 Instruction Manual NEC MATWORX NWA-008862-001 User's Manual Copyright © All rights reserved.