ARGUS 707 Infusion Pump Service Manual

image

Contents

1. 9 Go Go G oa A GS green stream ARGUS 707 mi total ml inf h min a ai 9 G 00 0 N G ee sass green Sam Flashing decimal points indicate which display is ready to accept an input by the keys 100 10 amp 1 In the interrogation mode the left hand display shows the address and the right hand dis play shows the according value configured at this address Please refer to chapter 2 6 where the meanings of the addresses are explained To modify any configuration data you have to go into the configuration mode 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 4 38 2 3 Configuration mode The configuration mode alows you to modify the pump configuration manually Please refer to chapter 2 6 where the meaning of the configuration adresses are explained 2 3 1 Step 1 ml total O aa nl inf min xi _ _ Ce 1 o9 1 1 0 1 K S ak ARGUS 707 2 3 2 Step 2 mil total a mi mi inf h min l 9 100 40 4 0 1 m sans TOP i gt ARGUS 707 oo MODE If the PIN code was accepted the pump will show the display of chapter 2 3 1 again but you have now access to all writeable addresses in the list of chapter 2 6 Select therefore any address in the left display see next side 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 5 38 2 3 3 Step 3 a Seal e F 9 Go Go 1 on A Ex MODE hn ARGUS 707
2. 2 5 Changing an existing pin code Enter the configuration mode see chapter 2 3 select address 0 and enter the new pin code 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 7 38 2 6 Address list of the pump configuration Add Address Def Default value of the manufacturer P Protection grade W Write enabled Boe ee E e y jo jo w PIN code active O OO OoOO Oo ono Ies 2 O W Key ON OFF only at STOP valid 3 o w Rate change only at STOP valia 0No Ies 4 o W Key STOP delayed time at 361 Ss RR RR E e o w Static alarm staff alerting system 0 No 1 Yes 7 o wW Display elapsed time in run mode 7820 ______ o Nos 1 Yes je Jo w select remaining tine l ___________ 0 Ne U 10 0 W operation without drop detector Cd O Ny 1 Yes_ Recall ml h rate at next power on 12 0 W Recall ml total end volume at next power on 13 o W Recall ml inf volume inf at next power on 14 O W SBS Step by step function 0 No 1 Yes_ is o W Display VIBI Volume To Be Infused 0 No 1 Yes E 7 a a N N IC CN KVO KOR mode see 60 18 W Drop alarm only if bottle empty O No 1 Yes e fo e 10 0 gt Era as da afaulrt e a eee 20 Jo w Menu Chr Clear ml inf F15 0 a IRI PR PR EI 22 O W Menu trA Transport 10 0 Menu PrL Pressure alarm limit 25 0 W Menu Set Fill 26 es 27 o W Menu dbo Data lock 28 o
3. I yyyy ml 308 0 operating time in min xxxxyyyy xxx min 309 0 operating time n min xxxxyyyy I Yyyy min IN RR RR PR E 311 999 un erime rate in me 999 314 10 W Upstream occlusion sensitivity 0 60 Osas Up 00 de highest sensitivity 316 7 W Downstream pressure limit PrL default 1 10 value Air bubble size microliters 50 1000 Drop rate window center in drops ml 10 65 Set 1 definitions IV IV Set 1 parameters 319 1000 W Correction factor actual nominal 1000 850 1150 er ee refer to chapter 5 5 N DI DI DI II Air summation volume n 50 microliters 2 40 IE II PE II 361 500 W Key ON OFF delay On 45000 per poe e additional key STOP if 4 1 362 8 W Display brightness ea 368 300 W Battery discharge time incl 15min pre 45 300 ie 1 alarm in min NE RR CR PR 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 9 38 add Def e Function Pan I E N 370 0 a Clock seconds o 371 o a Clock minutes O 372 0 a Clock hours oz 374 o a Clock days a 375 0 a Clock months aa 376 0 w Clock years 2000 20999 IE RI O ON N 380 0 Last failure number F XX Refer to chapter 5 7 381 0 Last infusion rate at failure 382 0 383 OI DI IP S YO 384 IC DI IP S y 385 CI DI e G y O 386 Jo 0 387 lo DI e S 388 0 Oldest failure number F XX 389 o____ Oldest infusi
4. external supply must light i hours at a rate of 30 0 ml h 15 Battery check at a rate of 30 ml h The green indicator battery lights up during this Run the battery test until the pump switches test off automatically 16 Printout the pump history refer to chapter 3 4 Check the battery run time by checking the latest history entries gt 4h 30min Check the pump internal time and date 17 Electrical test according to EN60601 1 Measurements attached all measurements made with a power cable 2 5 m Charge the battery after this test ee E The pump has passed the SSC and is safe for use a Date Name Signature 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 38 38
5. and setup the terminal program see chap 3 2 Start to capture text on the PC side before you switch the pump on O p C dd A ce 9 100 10 1 gt ii ape mii nf h Emin l amp START _MODE STOP 0 1 green sifeeam ARGU 707 By saving the transmitted configuration text you may print it using any text program See the following example of a configuration printout ARGUS Medical AG 10 02 04 AS 12 38 3 3 1 Configuration printout example Configuration Sat 17 Jan 2004 19 00 09 Pump type ARGUS707 Inventory number 005 8 407 Software release Wie 031218 75C2 Infused sum 861ml Operating time 15h18min Last service date 00 0 50 0 100 0 150 0 200 0000 250 0000 300 0000 350 0020 01 1 51 0 101 1 151 0 201 1091 251 0000 301 0000 351 0004 02 0 52 0 102 0 152 0 202 1071 252 0000 302 0000 352 0000 03 0 53 0 103 0 153 0 203 1061 253 0000 303 0000 353 0000 04 0 54 0 104 0 154 0 204 1051 254 0000 304 0000 354 0000 05 0 55 0 105 0 155 0 205 1031 255 0000 305 0000 355 0000 06 0 56 0 106 0 156 0 206 1023 256 0000 306 0000 356 0000 07 0 57 0 107 0 157 0 207 1014 257 0000 307 8611 357 0000 08 0 58 0 108 0 158 0 208 1006 258 0000 308 0001 358 0000 09 0 59 0 109 0 159 0 209 0999 259 0000 309 2918 359 0000 10 0 60 0 110 0 160 0 210 0988 260 0000 310 9999 360 0000 11 1 61 0 111 0 161 0 211 0974 261 0000 311 0999 361 0500 12 0 62 0 112 0 162 0 212 0967 262 0000 312 1200
6. 0 188 0 238 0000 288 0000 338 0000 388 0000 39 0 8 9 0 139 0 189 0 239 0000 289 0000 339 0000 389 0000 40 0 90 0 140 0 190 0 240 0000 290 0000 340 0000 390 0000 41 0 91 0 141 0 191 0 241 0000 291 0000 341 0000 391 0000 42 0 92 0 142 0 192 0 242 0000 292 0000 342 0000 392 0000 43 0 93 0 143 0 193 0 243 0000 293 0000 343 0000 393 0000 44 1 94 0 144 0 194 0 244 0000 294 0000 344 0000 394 0000 45 1 95 0 145 0 195 0 245 0000 295 0000 345 0000 395 0000 46 0 96 0 146 0 196 0 246 0000 296 0000 346 0000 396 0000 47 0 97 0 147 0 197 0 247 0000 297 0000 347 0000 397 0000 48 0 98 0 148 0 198 0 248 0000 298 0000 348 0000 398 0000 49 0 99 0 149 0 199 0 249 0000 299 0000 349 0000 399 0707 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 13 38 3 4 History printout 3 4 1 General Connect the ARGUS 707 to your PC see chap 3 1 and setup the terminal program see chap 3 2 Start to capture text on the PC side before you switch the pump on ml total mil inf h min gs c On 100 10 4 0 1 A sans ODE kn green stream ARGUS 707 3 4 2 History header At the top of the history a header will appear showing the device type the software release and if configured the inventory number and the last service date It shows also the pump internal real time at the moment of the printout Please note that there may be one hour summer or winter time deviation 3 4 3 History events Each registered event starts with
7. 2 Present correction factor 1020 Weight of delivered water g 14 9 New correction factor 1013 7 Enter the new correction factor at address 319 e g 1013 in the right hand display and save it by pressing key START then switch the pump off 8 Perform a control measurement according to the steps 2 amp 3 above Repeat the calibration procedure if necessary 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 24 38 5 5 3 Volume calibration with the pump integrated calibration program Needed equipment balance with a resolution of 0 1g at least 1 Insert a new IV set only recommended IV sets may be used see appendix in the user manual in the pump filled with water and connect the infusion line to a measuring cup on a balance 2 Enter the configuration mode as described in chapter in chapter 2 3 Enter the value 123 at address 399 This will force the pump to enter the calibration mode which will be displayed as bAL tArA 3 Press the key START STOP The pump delivers a volume of 5 ml at an infu sion rate of 250 ml h tArA is flashing 4 When tArA stops flashing reset the balance to 0 Press the key START STOP The pump will display tM 215 infusion de livery time is 215 seconds It should now deliver a volume of 15 ml at an infu sion rate of 250 ml h 6 After the delivery time has elapsed the pump stops and shows baL 12 75 in the displays Ent
8. 362 0008 13 0 63 0 113 0 163 0 213 0947 263 0000 313 0010 363 0010 14 0 64 0 114 0 164 0 214 0939 264 0000 314 0010 364 0000 15 0 65 0 115 0 165 0 215 0917 265 0000 315 0000 365 0000 16 0 66 0 116 0 166 0 216 0900 266 0000 316 0007 366 0000 17 1 67 0 117 0 167 0 217 1000 267 0000 317 0250 367 0000 18 0 68 0 118 0 168 0 218 0000 268 0000 318 0020 368 0300 19 1 69 0 119 0 169 0 219 0000 269 0000 319 1000 369 0005 20 0 70 0 120 0 170 0 220 0000 270 0000 320 0000 370 0012 21 0 71 0 121 0 171 0 221 0000 271 0000 321 0000 371 0000 22 0 72 0 122 0 172 0 222 0000 272 0000 322 0000 372 0019 23 1 73 0 123 0 173 0 223 0000 273 0000 323 0000 373 0000 24 1 74 0 124 0 174 0 224 0000 274 0000 324 0000 374 0026 25 0 75 0 125 0 175 0 225 0000 275 0000 325 0000 375 0007 26 0 76 0 126 0 176 0 226 0000 276 0000 326 0000 376 2003 27 0 717 0 127 0 177 0 227 0000 277 0000 327 0000 377 0000 28 0 78 0 128 0 178 0 228 0000 278 0000 328 0000 378 0000 29 0 79 0 129 0 179 0 229 0000 279 0000 329 0000 379 0000 30 0 80 0 130 0 180 0 230 0000 280 0000 330 0000 380 0000 31 0 81 0 131 0 181 0 231 0000 281 0000 331 0000 381 0000 32 0 82 0 132 0 182 0 232 0000 282 0000 332 0000 382 0000 33 0 83 0 133 0 183 0 233 0000 283 0000 333 0000 383 0000 34 0 84 0 134 0 184 0 234 0000 284 0000 334 0000 384 0000 35 0 85 0 135 0 185 0 235 0000 285 0000 335 0000 385 0000 36 0 86 0 136 0 186 0 236 0000 286 0000 336 0000 386 0000 37 0 87 0 137 0 187 0 237 0000 287 0000 337 0000 387 0000 38 0 88 0 138
9. Click OK then a connection to the pump should now be established Restore Defaults Cancel Apply 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 11 38 14 164 A_A707en SM V1 01 doc 3 2 2 Capturing text from the pump transmitted If you intent to print out the configuration or the history events transmitted from the pump you should capture the text transmit ted by the pump into a text file To do this please select Capture Text in the hyper terminal s menu A File box will open type a descriptive name for the file and then click Start The terminal program starts to capture each text received over the serial interface Stop the capturing after you did the re quired pump settings To do this select Stop in the menu Cap ture Text of the terminal program You will then have a text file including the data transmitted by the pump 3 3 Configuration printout O x ARGUS HyperTerminal File Edit wiew Call Transfer Help Dj ala 3 NE Send File Receive File Send Text File Capture ta Printer Creates a File of all incoming text ARGUS HyperTerminal ioj x File Edit View Call Transfer Help oel ala a 3 Fi Send File Receive File Pause N Resume Capture Text Send Text File Capture ta Printer Stop capturing 2 Connect the ARGUS 707 to your PC see chap 3 1
10. EEPROM data invalid E 29 RTC data invalid no RTC etc F 2g Stepper motor power test delayed 5s Fo 5Volt supply out of range ae 24Volt supply out of range delayed 5s Eo Downstream pressure sensor test failed always gt 4 7V delayed 5s Foo Upstream pressure sensor test failed always gt 4 7V delayed 5s E 2 Downstream pressure sensor test failed dynamical test failed F 40 Upstream pressure sensor test failed dynamical test failed F 44 Address invalid for config eeprom F 45 Address invalid for history eeprom F 46 Frequency from uC or RTC out of range F 47 Display print not present F 48 Key s too long active F 49 Sensor print not present E 20 AIL Air in line detector test failed Eo Movement test failed Home Pulse lt Hall Home E 2 Movement test failed Home Pulse gt Hall Home E 53 More than one rotation at STOP without KVO F 54 Infused sum lt gt Calculated sum Rotations F oa Frequency calculation Ea Rotation SW overflow F 38 Volsum control We recommend replacing the main board in case a fault code is not included in this list above 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 26 38 6 REPLACEMENT OF PARTS 6 1 General CAUTION The ARGUS 707 may only be used with accessories and spare parts which have been approved by ARGUS Medical AG for safe technical use CAUTION If a door a housing a pressure sensor or a main board is replaced a full calibration
11. bar 5 Switch the pump off and insert the IV set in the normal direction Now perform a pressure control measurement according to chapter Pressure control measurement Remark s By pressing the key 1 the pump display changes to xxxx Pr S where the left display shows the actual downstream pressure sensor signal in mV Return to the original state by pressing the key 1 again The procedure for the upstream sensor calibration part 2 1s equal to the proce dure for the downstream sensor but with inverse delivery direction 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 22 38 5 4 Pressure control measurement 5 4 1 Downstream control measurement Start an infusion at an infusion rate of 250 ml h according to the user manual Insert a pressure gauge on the downstream patient side which monitors the pres sure in the tube Simulate an downstream occlusion The pump must stop and the alarm must be activated at the default pressure limit of 700 mbar 100 mbar except a different default pressure limit has been configured If the result of this control measurement does not fulfil the stated requirement a pressure calibration according to chapter Pressure calibration has to be done 5 4 2 Upstream control measurement Start an infusion at an infusion rate of 250 ml h with the roller clamp on the up stream side near to the pump Simulate an upstream occlusion by closing the roller clamp To avoid the dr
12. first the calibration procedure of the pressure sensors and afterwards the volume calibration 5 3 Pressure calibration 5 3 1 General The volumetric infusion pump ARGUS 707 contains two pressure sensors One upstream sensor bottle side left input One downstream sensor patient side right output CAUTION A pressure calibration becomes necessary if the pressure control measurement was not accurate enough a new IV set configured or a pressure sensor re placed or the main board is replaced Needed equipment a manometer with a resolution of 0 05 bar a 3 way stop cock the chosen IV set refer to the list including the recommended IV set in the user manual 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 17 38 MAINTENANCE 5 3 2 Enter the pressure calibration mode 1 Enter into the configuration mode see chapter 2 3 t ay Greases green stream ARGUS 7 pu 2 Select address 399 in this display Press then key MODE 4 Press the START STOP key the pump starts to running at a constant rate of 200 ml h The display 3 Enter value 1234 in this Doaa CAL oa display 5 3 3 Calibrating the offset of both sensors Remove any IV set inserted in the pump Press the START STOP key to store both values for the up and downstream sensor this will be acknowledged by a 1 Open the door 3 Press the MODE key to go to the next calibration st
13. prob lem pressure plate door hinge peristaltic etc Release the pressure again 2 Restart a pressure build up Check for the pressure gauge dis play which must in crease 4 Press the MODE key 3 Press the START STOP key The display shows CAL when a pressure of 0 4 bar has been 1 4b Follow the instruc reached The stored value will be tions on the next side acknowledged by a sound Remark By pressing the key 1 the pump display changes to Pr S xxxx where the right display shows the actual downstream pressure sensor signal in mV Return to the original state by pressing the key 1 again 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 20 38 Remark By pressing the key 1 the pump display changes to Pr S xxxx where the right display shows the actual downstream pressure sensor signal in mV Return to the original state by pressing the key 1 again 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 21 38 MAINTENANCE 5 3 6 Calibrating the upstream sensor part 2 left side 1 Close the roller clamp and install the tube in the reverse direction 2 Open the roller clamp Close the door i 3 Restart a pressure build up 4 Check for the pressure gauge display which must increase Press the START STOP key when a pressure of 0 4 bar has been reached Press key MODE the display shows CAL 1 4b Repeat this point at a pressure of 1 4
14. to be done in accordance to the user and service manuals Visual check for damage cleanness and completeness Housing labels accessories connectors power cable etc Test the function of the stop flow clamp Proper movement of the clamp z 3 Keep MODE pressed while switching on the pump Display pump type and software release Display of 2 4 7 F in numeric display Display of all operation and alarm indicators Connect Disconnect the pump to the mains The indicator external supply lights up Li Test the drop detector by simulating drops Check the green drop indicator fa Test the door switch open and close the door Door open the door indicator lights up g Door close the door indicator turns off Open the door and remove any IV set Indicator air bubble lights up i ie Install a water filled IV set close the door Indicator air bubble turns off Set rate to 333 3 m h press START disconnect the The red indicator drop lights up delayed drop detector The acoustical alarm turns on Press MODE The acoustical alarm mutes a Check the external connector nurse call Relay contact switches see chapter 9 N Calibration of the pressure sensors See chapter 5 3 and 5 4 of the service manual IV set type used Volume calibration See chapter 5 5 of the service manual E Charge the battery while the pump is running 16 The indicator
15. 0 u menu stb Stand by 29 oO fW Menu MEd Medication no eS ee II eee 30 o w Menu tM Timer alarm 32 jo w Menu boL Release Bolus 33 o W Menu boz Bolus rate 32 1 34 10 W Menu tot Bolus total 32 1 38 JO W BOLUS application automatic 32 1 34 1 0 No l Yes _ INI DI E DI NI e ba fai ia Ma a a ee ta 41 lo w Clear ml h after infusion completed 42 jo wW Clear ml total after inf completed 41 1 43 0 wW Air volume accumulated see 350 and 351 a6 0 W Pressure display with indicator 45 1 0No Ives a7 jo W Standby and battery pre alarm low volume 0 No 1 Yes ag 0 W Flashing numeric display at alarm O Oo 1 Yes_ a5 o u Alarm acknowledge only with key MODE 0 No 1 Yes 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 8 38 Ada i Def Funct ron Range SS EE 50 0 W Start with gt lbar downstream pressure allowed O No 1 Yes ow en Ee Set default serial interface to RJ 45 connector O No 1 Yes 60 o wi KVO only after infusion completed O No 1 Yes O for ji set 1 enabled always enabled one 1 es N RR RR PE zra II E 202 1071 w S S 23 1061 w IE 204 Horw o S S 205 1031 pa 206 E EI IE 207 1014 w S S 208 1006 w o S z9 s W a E IN RR RR E E IER COREEI IE arpa wl a2 967 a S S as far w S aa ps w A a IER CARE DORI IE 2i7 poo IEZI E IN RR RR RI E 306 jo infused sum in ml Deer ee ml 307 0 nfused sum n ml xxxxyyyy
16. 01 doc ARGUS Medical AG 10 02 04 AS 32 38 10 124 Main board A707 10 125 Sensor board A707 10 126 Pump unit A707 10 127 Door complete A707 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 33 38 10 128 Casing back plane A707 10 136 Casing back plane A707 230VAC complete 10 129 Casing forepart without door 10 137 Casing forepart complete 12 011 Interface cable 8pol 2m 12 042 Cable staff alert 2m amp 12 012 Interface adapter of 10 119 Battery NiMHd 12V 1500mAh A707 11 221 Sealing A707 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 34 38 D f mi total w mi mi inf a i E fh min Z t O aS aa we Pui i 2 Goo 10 1 Coi 4 Sam Ban aleea ARGUS 707 12 035 Pressure gauge with stopcock ail Ye Pe 11 005 Bottle holder 45 cm 11 043 Bottle holder 60 cm Partno Description 11 237 Identification plate A707 11 238 Label flow direction 11 239 Short instruction German 11 240 Short instruction English 11 241 Short instruction French 11 242 Short instruction Danish 11 243 Short instruction Dutch Part no 11 244 11 245 11 246 11 247 Part no 11 222 Front panel A707 10 135 Edge board complete A707 Description Short instruction Czech Short instruction Portuguese Short instruction Spanish Short instruction Italian 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 35 38 7 SAFETY STANDARD CHECK The safety check should be done
17. 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 10 38 3 CONFIGURATION AND HISTORY PRINTOUT 3 1 Introduction CAUTION The infusion pump has to be disconnected from the patient before and while the serial interface cable is connected to the pump The connection of the ARGUS 707 over the serial interface RS 232 can be done by con necting the interface cable no 10 093 to the serial interface outlet of the serial COM port of a PC A data transfer between ARGUS 707 and a PC can be done without any additional soft ware running on the PC 3 2 Settings of the terminal program 3 2 1 RS232 connection settings of the terminal program ax Open the terminal program on the PC e g ee HyperTerminal which has been included in every eee Microsoft Windows Operating System Enter a name and choose an icon for the connection In the Connection Description menu type for e g Hame ARGLS ARGUS and click OK leon ConnectTo 2 x Please make sure that the correct COM port of the PC has been selected D ARGUS Enter details far the phone number that you want to dial Countryregior Switzerland 41 iu ci E 2x Phone number Port Settings Connect unc ET Bits per second IEEE Data bits E gt Parity None gt Make the appropriate changes on the COM Proper ston tte f 2top DIIS gt ties menu according to the right hand picture Flow control None
18. 2 Requirements for a software update eee eee nana aee aaa 16 ine Bo ulei TA o oa e IRRD RE RIO PRI ORI RR IRAN RIC RI NN MAI RER IRA RR N N RIC RIND E ENE NR IRI 16 MAINTENANCE bss vases doiecaaaas ia A added iata cata DA aci ata a Danaila l uda dau dastea b taie ai aci da daco PEAS ERSP 17 Si le CDC IE sata netromania and i ai d ta et d oa i na i ee 17 PeR eCa DEAT a IERI E ORI RER IE NCN RR PNI RR E E E OR RRC RI EA EE EEE EE ii ia WPS e MO a PEEN PE tarie TERTA TE A TENI T E ATERT E ETE ETT 17 5 4 Pressure control measurement ia saoi omida data sie i seta aa beaboa ba dau edi auda md daa Gt E ta a di 23 Je Volume C IDA een E goala ae ate al ai in ui piele aia i ua ala 24 OPS DEC LLC AOI paiete at a etate und ital di sil ta e il o ti e nat ala e tul 25 Sd AU OSS a II IRT PRR IONIC RE N E AEAEE E E E 26 REPLACEMENT OF PARTIS ponnc a a aa da a a EENE 27 AEE IE E E E EE e ele aci E AT EE ENIE EA O lac in el ta a 6 2 Disassembling of the ARGUS 707 cccccccccccsccsssssssessessessessssesasccececaecccccecscecscceccecceneceeceseeeeessensees 2 Me RI e ei AG ile 1 OE aN E DIR E RI PR RODIE IRI NE RIN N NOR RR DRE E RP IRI RTR SE RANITI 31 0 44 9 DAL Da kc esse i ai em oe m ne os ini aaa steaua adela S 32 SAFED STANDARD C HECE specia eta ue ua 2090 pna Rae dia cine e i ideia stana inta aaa aaa aia mda a 36 TY RING DIAS RAV esti etice iai olita i ad dili ni su culeeccuanssiiatentaracacacetvassaielaacuvenssnccevaatoraasscacacsusani
19. SS Important remark Invalid values entered will be corrected automatically by the pump to the maxima or minima value allowed for the according address 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 6 38 PUMP CONFIGURATIONS 2 4 First activation of a configuration PIN Code The activation of a PIN code allows you to protect the configuration from unauthorized persons To activate the PIN code enter the configuration mode as described in the previous chapter For this use the default PIN code 0 1 To enter into the configuration mode switch the pump on while keeping the keys MODE and START STOP pressed 2 Press key START 3 Press key MODE 5 Enter the default PIN Code 0 To do 4 Press key START again The this press the key 10 while keeping the display will show Cod key 1 pressed Press START key to acknowledge the entered PIN code ee 6 Press the MODE key mi inf The flashing decimal points will change to the right hand display Press key START to acknowledge to new PIN 7 Enter now the new PIN Code max 4 digit number Please remember this code it will never be visible again After you switch the pump OFF and ON again you can enter the configuration mode only if you enter the correct PIN code in display of chapter 2 3 2 Please note The interrogation mode can always be accessed without the PIN
20. Service Manual for Volumetric Infusion Pump green stream ARGUS 707 Made in Switzerland C 0120 ARGUS Medical AG CH 3627 Heimberg Switzerland A member of the CODAN group 10 Table of contents INTRODUCTION ses ceai oar a aaa E aaa dia alt d aid taia iaca sati aa aluatul d ai a unul adina aiba dat 3 si GIC TIE acnee taie ei ale ele lt ai St aaa Oi a lt a a ft 3 PUMP CONFIGURA TIONS cipa oaia t atata 8 at aaa d 00 Out Gcal t atasa aa da dia tie aa 3 Dal za Ce Ile e ear ate e dia ada i bei a tati Ma lata a a ad Did ee das e it Bta aa 3 2 2 Mie tO ALI OI IMO 8st a pei Di aia al E O E mb a fa 4 2 5 COMM UAT ON TOG Cvs tza aaa st da iati ne ema i a bin ae ali dis ainu inta o ein Da adia 5 2 4 First activation of a configuration PIN Code eee I Dy eO a ATIC ES I CO Cie E E nies aie tale iata iata 7 2 6 Address list of the pump configuration nene eee eee aaa 8 2 7 SPECIAL COMMO UL AMON OP ONS ge catene ema bea eee inel rusia o ol 8 ceata ti i il arata 10 CONFIGURATION AND HISTORY PRINTOUT wisssssscsssssesssasersdecsszocesascnvedsenssevnssatessedvousagntvensnesosstesstecssavsuseeenes 11 EO C CLOWN ca ara sia E date blat tt ast i ad delta n nl aaa a a 11 3 2 Settings of the terminal Program scoate asa mie Alei a aus a ceia ta i a i aa te nicer 11 Koek OP FIO IN AU a O e a A aaa betia Dzeu 12 A D O E E E AEA R 14 SOEEWARE UPDA FE E E EAEE O 16 AE E n A EAEE E E E A E I E AR IRI EE E A E E AE E T A 16 4
21. a lock off Data lock on Infsum cleared Transport off Transport on Air bubble pump stop Bolus total reached Exit setup mode PC configuration done PC configuration failure KVO Pump start ext changed parameters Upstream occlusion pump stop ARGUS Medical AG 10 02 04 AS 15 38 4 SOFTWARE UPDATES 4 1 General This chapter describes the procedure to perform a software update on the ARGUS 707 in fusion pump To check the installed software release in your ARGUS 707 press the MODE key while switching on the pump Please refer to your local distributor or ARGUS Medical AG to determine the latest soft ware release able to run on your device hardware 4 2 Requirements for a software update To update an ARGUS Medical device the following items are needed A PC with Microsoft Windows RS 232 serial interface cable part no 10 093 connected to the PC Software file AMFlasher and corresponding user instruction pdf file on your PC Data file A707 _xxx txt including the pump software which can be downloaded with the AMFlasher software The file name A707 xxx txt contains the software release xxx version of the pump software release Those items are available from your local distributor or from ARGUS Medical 4 3 Safety aspects Be aware of the following points e For medical device traceability your local distributor or ARGUS Medical AG needs to be informed about every device updates Se
22. a message line Please refer to the complete list men tioned in chapter 3 4 5 which shows the possible messages generated by the user s han dling on the pump or by the pump itself Each event has a time stamp on the right side of the second line 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 14 38 14 164 A_A707en SM V1 01 doc 3 4 4 History printout example KKK History KKK KK Pump type ARGUS7 07 Inventory number 005 8 407 Software release V1 00 031218 75C2 Infused sum gt 388ml Operating time 24h44min Last service date Pump OLE Rate 0040 0ml h Ini SeL 1 Tue 20 Jan 2004 18 29 22 Total 0000 0ml PrLimit 0700mbar Infsum 0000 0ml Status 0x0000 Pump on Rate 0040 0ml h Inf Set l Tue 20 Jan 2003 18 27 18 Total 0000 0ml PrLimit 0700mbar Infsum 0000 0ml Status 0x0000 3 4 5 History messages Tue 20 Jan 2004 19 40 30 Possible messages appearing in the first line of each history event No information available Battery defective Battery low pre alarm Battery low pump stop Bolus start Bolus stop External power off External power on Downstream occlusion pump stop Prlimit change Pump has detected failure Pump off Pump on Pump start Pump stop KVO Rate change Enter setup mode NO drops pulp stop Not enough drops pump stop Too many drops pump stop Door open pump stop Inf Set change Timer alarm pump stop KVO Total volume reached pump stop Dat
23. at least every 10 000 running hours or every 24 months Refer to the form in the appendix of this manual 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 36 38 8 WIRING DIAGRAMM Display Board with Hall Switch Door X01 oooooooo Drip SCl 16 Nurse Call X09 X10 BAO Oo ee 00000000 o Oo o Edge Board X08 me X02 E Stepper Motor o 6 o o X02 X01 a Battery X07 Main Board m xo1 Sensor Board 10 with Hall Switch Motion Trafo N 4 4 AC Main Pressure Sensor Pressure Sensor Air Detector Stop Flow Protection Input Output i 3 9 BLOC SCHEMATIC Buzzer Uh o 2 mL e o 88 O 8 Controller o D Alarm e 2 2 se TN Rey A E Imi total On Off E a KJ h min RS232 Li i Display RS 232 Driver Motor Power Switch Test Switcher Sa EEPROM AC Main EG les RTC L 512 Hz Motor Buzzer Reset Watchdog pe Ext DC Li Batt Z 3 Volt 5V Ni Cd Logic Air Bubble Detector Akku 12V Drop Detector Pressure Sensor in Door Control Pressure Sensor out Motion Control 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 37 38 10 APPENDIX Safety standard check SSC ARGUS 707 ARGUS Medical AG Serial no Inventory no Hospital Department Customer The SSC has to be performed at least every 24 months or after 10 000 hours of operation The check has
24. ep The display shows roll CLAM Remark By pressing the key 1 the actual value of the offset voltage in mV 1s displayed for the upstream sensor in the left hand display for the downstream sensor in the right hand display Return to the original state by pressing the key 1 again 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 18 38 MAINTENANCE 5 3 4 Calibrating the upstream sensor part 1 left side 1 Insert a new IV set filled with water place the roller 2 Close the clamp on the upstream side near the pump and close roller the door Open the roller clamp and let the pump clamp and running for 5 seconds at least wait 10 seconds 3 Press the START STOP key to store the value for an upstream oc clusion This will be ac knowledged by a sound 4 Press key MODE to enter the next calibration procedure The display shows CAL 0 4b 5 Open the roller clamp for the next step Remark By pressing the key 1 the pump display changes to xxxx Pr S where the left display shows the actual upstream pressure sensor signal in mV Return to the original state by pressing the key 1 again 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 19 38 MAINTENANCE 5 3 5 Calibrating the downstream sensor right side 1 Simulate a downstream occlusion by the 3 way stop cock The pump must generate a pressure of at least 1 4 bar Otherwise there might be a mechanical
25. er now the value of the balance e g 14 90 g This value must be within the range of 12 75 17 25 Otherwise switch the pump off and restart the volume calibration 7 Press the key START STOP to acknowledge the entered value 8 The pump displays the new correction factor e g 1013 9 Press the key START STOP to store the new correction factor acknowledged by a buzzer sound The pump display changes back to bAL tArA again 10 Switch the pump off Do not remove the infusion set 11 Perform a control measurement with an infusion rate of 250 ml h and an infu sion total of 15 ml Repeat the calibration procedure if necessary 5 6 Pump specifications Please refer to the user manual for the specifications chapter 8 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 25 38 5 7 Fault codes A technical failure will be signalled by the pump with a continuous alarm display and a continuous sound During this state the fault code which causes the pump to fail can be displayed by pressing the key MODE IF the pump was switched OFF after a detected failure the fault code will be stored in the configuration of the pump please refer to chapter 2 6 Addresses 380 389 The possible fault codes registered in the configuration are listed in the table below ale ma Failure E Ad ROM test E 2a ROM check Runtime 23 RAM test check F 24 XRAM test check F 23 CPU test Foo Invalid function menu Bee
26. es 37 BPLECOC SCHEMATIC sc onceeae ianon en eiefna adie coc eat alia E E E De daaaaa cateii iaca taactileea 37 APPENDIX scsiaeiac toaca iaz c 6080 300 aaa adi andann D ga nea aud 3 d da aasian a da alta datata dada dia b taia calea 38 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 17 02 04 AS 2 38 1 INTRODUCTION 1 1 General IMPORTANT This service manual is intended for the exclusive use of authorized persons who have been trained by ARGUS Medical AG in the maintenance and repair of the ARGUS 707 infusion pump The service manual is meant to be used together with the user manual IMPORTANT ARGUS Medical AG shall not assume any responsibility for any manipulations which have been carried out on the ARGUS 707 infusion pump by a non authorized person This manual contains the latest data available It is subject to further modifications in ac cordance with technical improvements 2 PUMP CONFIGURATIONS 2 1 General CAUTION The configuration possibilities constitute a modification of the pump and may only be carried out by authorized persons CAUTION After changing the configuration a function check and a control measurement has to be performed 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 3 38 2 2 Interrogation mode With the interrogation mode you can read the present configuration of the pump without the possibility to modify any configurations mi total O mi ni inf min A eh oe
27. on rate at failure II DI DI DI 390 0 W Last service gate i yyww year and week 39 0 l2 last service date yyww year and week 392 0 3 last service date yyww year and week 393 W Service interval in months 0 24 394 W Service interval in hours of 0 9999 operation 0 disabled 395 W Own address for SCI 0 127 A a o XXXX yyyy 397 18yyy W Serial number of the pump Lovee A AA A Data xxxx gt clears protection key 2 7 Special configuration options 1 Configuration of a reminder alarm for the safety standard check If a safety standard check reminder alarm is configured the pump will display Ctrl after power up when the time has come to perform a standard safety check To configure a reminder alarm please follow next steps First the service interval has to be entered either in months or in hours of operation or both address 393 394 Next the last service date has to be entered on address 390 Any value higher than 0 entered at the address 393 and or 394 will release the reminder alarm after the ser vice interval has elapsed check also the correct settings of the internal clock 2 PC configuration tool AConfig With this additional software the pump may be configured from a PC over the serial port This software may be available from your local distributor or ARGUS Medi cal AG Caution AConfig may only be used with software versions greater or equal to 1 01 14
28. op alarm release simulate falling drops on the drop detector The pump must stop after several seconds and an upstream alarm right hand LED in the pressure bar graph must be activated If the result of this control measurement does not fulfil the stated requirement a pressure calibration according to chapter Pressure calibration has to be done 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 23 38 5 5 Volume calibration 5 5 1 General There are two ways to calibrate the volume delivered by the ARGUS707 volumet ric pump select one e By entering the correction factor e With the internal calibration program of the pump Needed equipment balance with a resolution of 0 1g at least 5 5 2 Volume calibration by entering the correction factor 1 Insert a new IV set only recommended IV sets may be used see appendix in the user manual in the pump and perform a warm up infusion of 10 ml at an infusion rate of 999 9 m h 2 Infuse now a volume of 15 ml of water in a measuring cup on a zeroed balance at an infusion rate of 250 ml h Determine the weight of the delivered water Enter the configuration mode as described in chapter 2 3 Go to address 319 and read the present correction factor oe a A Calculate the new correction factor as follows l Present factor weight of delivered water New correction factor Present factor weight of delivered water 15 or Pre set volume of point
29. pressure sensors and volume calibration is required 6 2 Disassembling of the ARGUS 707 CAUTION Disconnect the mains cable from the power outlet before opening the housing Observe the antistatic protection rules when disassembling the ARGUS 707 the use of an antistatic table mat and a grounded clip are recommended 6 2 1 Disassembly of the case 5 ihoni adi 9 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 27 38 6 2 2 Removing the pump door 6 2 3 Disassembling the pump unit 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 28 38 6 2 4 Assembling the pump 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 29 38 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 30 38 Mind the torque for the screws listed below Combination clamp 1 0 Nm Transformer 1 0 Nm All other screws not mentioned in this chapter 0 6 Nm 6 3 Replacements of parts CAUTION The ARGUS 707 may only be used with accessories and spare parts which have been approved by ARGUS Medical AG for safe technical use For the part numbers of replacement parts consult the following chapter 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 10 02 04 AS 31 38 6 4 Spare parts 10 087 Combination clamp 10 089 External drop detector 10 093 Interface cable docking 10 120 Display board A707 pumps 10 121 Pressure sensor 10 122 Air detector A707 A707 14 164 A_A707en SM V1
30. rial number you performed e Do not make any software updates when the device is used and or connected to a patient CAUTION A standard safety check see appendix has to be performed after every software up date 14 164 A_A707en SM V1 01 doc ARGUS Medical AG 17 02 04 AS 16 38 5 MAINTENANCE 5 1 General 3 2 CAUTION Only authorized persons who have been trained by ARGUS Medical AG or by the local distributor are allowed to service the ARGUS 707 infusion pump In case of repair request send the unit to the local distributor with a report outlining the exact nature of the failure More information is available from ARGUS Medical AG CH 3627 Heimberg Switzerland E mail info argusmedical com CAUTION The safety standard check has to be performed at least every 24 month or after 10 000 hours of operation The check has to be done in accordance to the chapter 7 No special maintenance of the ARGUS 707 infusion pump is necessary There are no wear and tear parts Recalibration 5 2 1 General Unless otherwise specified by the customer the ARGUS 707 has been calibrated by the manufacturer with the CODAN L86 infusion set If a different infusion set 1s used see recommended list in the appendix of the user manual a recalibration is required CAUTION A recalibration with a different V set always requires a recalibration of both pres sure sensors and a volumetric calibration CAUTION It is mandatory to execute