termômetro infravermelho infrared thermometer mt
Contents
1. i 6 6 Emissivity Table ACCESSORIES MAINTENANCE 8 1 Battery Replacement 8 2 Cleaning WARRANT Y 31 1 INTRODUCTION This instrument is a portable easy to use 3 1 2 digits compact sized digital infrared thermometer with la ser sighing designed for simple one hand operation Meter comes with backlight display sensor cover Auto Hold function and auto power down 10 sec onds approx after releasing trigger to extend bat tery life 2 SAFETY INFORMATION It is recommended that you read the safety and operation instructions before using the infrared ther mometer N WARNING e Pressing the trigger turns the laser beam on and off Take extreme care and do not allow the laser beam to enter your eye or those of any other person or animal Do not look directly into the laser light from the optical system When measuring the temperature of an object which has a mirror finish be careful not to allow the laser light beam to be reflected off the surface into your eyes or those of another person Do not allow de laser light beam to impinge upon any gas which can explode 18 3 SPECIFICATION The symbol f identifies hazardous conditions where the user must be referred to manual 3 1 General Specif
2. protetora contra poeira fun o Hold e desligamento autom tico 10 segundos aprox ap s soltar o gatilho para prolongar a vida til da bateria 2 INFORMA ES DE SEGURAN A recomendado a leitura das instru es de seguran a e opera o antes de usar o term metro infravermelho S N ADVERT NCIA e Pressionando o gatilho faz com que o laser seja ativado Tenha extremo cuidado em n o permitir que o laser seja apontado para seus olhos ou de outra pessoa ou animal Nunca olhe diretamente para uma fonte de laser de um sistema tico Quando medir temperatura de um objeto que seja espelhado tenha cuidado para que o laser n o seja refletido para seus olhos ou de outras pessoas Nunca permita que a luz do laser incida sobre qualquer g s que possa explodir 2 3 ESPECIFICA ES O s mbolo no instrumento indica que o usu rio deve referir se a explica o no manual 3 1 Especifica es Gerais Display Cristal l quido de 3 1 2 d gitos LCD com leitura m xima de 1999 Indica o de Bateria Fraca O mostrado quando a tens o da bateria cair abaixo do n vel de opera o Tampa Protetora para o Sensor Taxa de Amostragem 300ms nominal Ambiente de Opera o 0 C 50 C umidade relativa lt 70 Temperatura de Armazenamento 20 C 60 C umidade relativa lt 80 com a bateria removida do instrumento Desligamento Autom tico Aprox 10s Corrente de Consumo em Repous
3. tica ou perto de uma carga eletrost tica isto pode causar erro e Nunca use o instrumento em ambientes explosivos OU corrosivos o instrumento pode ser danificado ou poder ocorrer uma explos o Nunca deixe ou use o instrumento exposto diretamente a luz solar ou onde ele possa ser exposto a altas temperaturas alta umidade ou condensa o Se isto for feito poder deformar o instrumento sua isola o poder ser danificada ou poder n o operar de acordo com as especifica es e Nunca aponte a lente para o sol ou qualquer outra fonte de luz intensa Se isto for feito o sensor pode ser danificado e Nunca permita que a lente entre em contato com o objeto cuja temperatura est sendo medida Tome cuidado para que a lente n o suje ou seja riscada ou permita que materiais estranhos sejam fixados a ela Isto pode causar erro e Nunca toque ou segure a parte frontal do aparelho A medi o pode ser afetada pela temperatura das m os Nunca coloque o instrumento sobre ou ao redor de objetos quentes 70 C 158 F pois isto pode causar danos ao gabinete do instrumento Se o instrumento exposto mudan as significativas de temperatura ambiente frio para quente ou quente para frio permita que o instrumento estabilize a temperatura por 20 minutos antes de executar a medi o Poder ocorrer condensa o na lente quando houver mudan a de um ambiente frio para um ambiente quente espere 10 minutos para que a co
4. s 33 Planalto Paulista CEP 04069 000 S o Paulo SP Fax Envie uma c pia do certificado de garantia devidamente pre enchido atrav s do fax 0xx11 2577 4766 e mail Envie os dados de cadastramento do certificado de garantia atrav s do endere o sac minipa com br Site Cadastre o certificado de garantia atrav s do endere o http www minipa com br sac IMPORTANTE Os termos da garantia s ser o v lidos para produtos cujos certificados forem devidamente cadastrados Caso contr rio ser exigido uma c pia da nota fiscal de compra do produto Manual sujeito a altera es sem aviso pr vio Revis o 01 Data Emiss o 15 06 2005 16 TABLE OF CONTENTS 1 2 3 MEASUREMENT CONSIDERATIONS INTRODUCTION 18 SAFETY INFORMATION 18 SPECIFICATION 3 1 General Specification s 3 2 Laser Specification 3 3 Electrical Specification INSTRUMENT DESCRIPTION 21 OPERATION sis sec rone rsen enfer enean Sis 22 5 1 Operation Mode zr 5 2 Highest Temperature Location 5 3 Field of View 6 1 Measurement Theory 6 2 Infrared Radiation 6 3 Structure of Emission Thermometer 6 4 Emissivity eere 6 5 Specials Attentions
5. Minipa s authorized dealer 3 Warranty will be void in case A Ithasbeenmisused altered neglected or damaged by accident orabnormal conditions of operation or handling B Theinstrument showsviolations by anonauthorizedrepaircenter 4 This warranty does not apply to fuses dry cells batteries and accessories as test leads carrying case thermo couple etc 5 Forinstrumentwith software Minipa assumes responsibility that the software will operate in accordance with its functional specifi cations for 90 days Minipa will not guarantee that the software will beerrorfree oroperate withoutinterruption 6 Minipaassumesnoriskfordamageintransitor transportation costs 7 Warranty will be valid only after the registration of this certificate Name Address City State Phone Sales Voucher N Date Serial N Sales Agent Name 31 Warranty Certificate Registration Procedures The registration can be made by the following ways Mail Send a copy of warranty certificate correctly filled to the following address Minipa Ind stria e Com rcio Ltda Att Servi o de Atendimento ao Cliente Alameda dos Tupin s 33 Planalto Paulista CEP 04069 000 S o Paulo SP Fax Send a copy of warranty certificate correctly filled by fax number 0xx11 2577 4766 e mail Scanning this form and attach to your e mail Please send to sac minipa com br Site Register the warranty certificate by http Avww minipa com br sac IM
6. NI CO CO CO S Q NO O Q O O e Manual de Instru es e Bateria 9V instalada e Capa Protetora 8 MANUTEN O 8 1 Troca de Bateria O instrumento alimentado por uma bateria de 9V NEDA 1604 IEC 6F22 O s mbolo aparece no display LCD quando a troca da bateria necess ria Para substituir a bateria siga o procedimento descrito abaixo 1 Retire a tampa do compartimento da bateria cuidadosamente escorregando para baixo a tampa do medidor 2 Desconecte a bateria velha do medidor e substitua por uma nova 3 Enrole o excesso de fio e coloque primeiro a parte de cima da bateria dentro do compartimento Instale a bateria e recoloque a tampa 4 Quando a bateria instalada o instrumento liga automaticamente para verificar a codi o da bateria E automaticamente se desligar ap s aproximadamente 10 segundos sem opera o 8 2 Limpeza Periodicamente limpe a parte externa do instrumento com pano macio umedecido em detergente neutro n o utilize produtos abrasivos ou solventes 9 GARANTIA AN Trinipa O instrumento foi cuidadosamente ajustado e inspecionado Se apresentar problemas durante o uso normal ser reparado de acordo com os termos dagarantia 6 7 GARANTIA S RIE N MODELO MT 350 Este certificado v lido por 24 vinte e quatro meses apartir da data daaquisi o Ser reparado gratuitamente nos seguintes casos A Defeitos de fabrica o ou danos que se ve
7. PORTANT The warranty conditions and limitations will be valid only to the certificates correctly registered In case the purchaser did not register asales receipt showing the date of purchase will be required Manual specifications subject to change without notice Revision 01 Date of Issue 15 06 2005 32 Minipa Ind stria e Com rcio Ltda Al dos Tupin s 33 Planalto Paulista S o Paulo CEP 04069 000 CGC 43 743 749 0001 31 Site http Avww minipa com br
8. TERM METRO INFRAVERMELHO INFRARED THERMOMETER MT 350 MANUAL DE INSTRU ES INSTRUCTION MANUAL NDICE 1 INTRODU O 02 2 INFORMA ES DE SEGURAN A 02 3 ESPECIFICA ES 03 3 1 Especifica es Gerais 03 3 2 Especifica es do Laser 04 3 3 Especifica es El tricas 04 4 DESCRI O DO INSTRUMENTO 05 5 OPERA O nn a 06 5 1 Modo de Opera o 06 5 2 Localizar o Ponto de Maior Temperatura 07 5 3 Campo de Vis o sss 08 6 CONSIDERA ES DA MEDIDA 10 6 1 Teoria de Medi o 10 6 2 Radia o Infravermelha 10 6 3 Estrutura do Term metro de Emiss o 11 6 4 Emissividade 6 5 Cuidados Especiais 6 6 Tabela de Emissividade 7 ACESS RIOS essere trente 13 8 MANUTEN O u 8 1 Troca de Bateria ga 8 2 Limpeza 9 GARANTIA retreat tene 15 1 INTRODU O Este instrumento um term metro digital infra vermelho de 3 1 2 d gitos com mira laser port til f cil de utilizar e desenhado para ser operado com apenas uma s m o O medidor possui display LCD iluminado tampa
9. This instrument is supplied with 9V NEDA 1604 IEC 6F22 The symbol appears in LCD display when the replacement is necessary To replace the battery follow the procedure listed below 1 Pull off battery cover by gently sliding it towards the bottom of the meter 2 Disconnect the old battery from the meter and re place with a new unit 3 Wind the excess lead length and put the top of battery beneath the battery chamber Install the battery and put the battery cover 4 When battery is installed the meter turns on auto matically to check the battery conditions Power will turn off automatically after 10 seconds with out operation 8 2 Cleaning Periodically wipe the housing with a damp soft cloth and mild detergent do not use abrasives or sol vents 30 9 WARRANTY AN Thuia ibol This instrument was carefully calibrated and inspected If any failure occurs under normal use this product will be repaired according to warranty conditions and limitations WARRANTY SERIAL N MODEL MT 350 1 Thewarranty periodis24 twenty four months andbegins on the date of purchase 2 Itwillbe repaired free of charge in the following cases A Manufacturing defects or damages occurred under normal use of instrument within the warranty period B The services to correct the failure will be done only in autho rized service center or personal will be allowed to fix this product C If productis purchased through a
10. argem adequada Di metro do Ponto mm n 1000 gt Marca do Laser Dist ncia do Objeto mm O di metro do ponto aumenta com a dist ncia do teste como mostra Di metro do Ponto medido a 90 de Energia 5 2 Localizar o Ponto de Maior Temperatura Aponte o term metro para o objeto e ent o inicie um movimento de varredura ao longo da rea de in teresse movimentando para cima e para baixo at localizar o ponto de maior temperatura 7 5 3 Campo de Medi o Certifique se de que o objeto maior do que a rea de medi o do term metro Quanto menor for o objeto mais pr ximo o term metro dever estar Quando a precis o for cr tica certifique se de que o objeto duas vezes ou mais a rea de medi o A medida que a dist ncia do term metro ao objeto aumenta a rea de medi o aumenta propor cionalmente ERRADO CERT NOTAS e N o recomendado para medir superf cies met licas brilhantes ou polidas a o inoxid vel alum nio etc Verifique a emissividade e O term metro n o mede atrav s de superf cies transparentes tais como vidro Ira medir a 8 temperatura da superf cie do vidro e Vapor p fuma a etc podem prejudicar a precis o das medi es obstruindo o campo de vis o do instrumento N ADVERT NCIA e Nunca use o instrumento perto de qualquer dispositivo que possa gerar radia o eletro magn
11. e a ilumina o do display funcionam ao mesmo tempo quando o aparelho ligado Pressione o gatilho para ligar o medidor se ele estiver desligado Se o gatilho solto o valor congelado e o s mbolo HOLD aparece Fun o Desligamento Autom tico Faz com que o instrumento seja desligado em 10s Selecionando a Escala de Temperatura As leitura s o mostradas tanto em graus Celsius C ou Fahrenheit F Quando o term metro ligado ele estar ajustado para a ltima escala de temperatura que foi utilizada quando o term metro foi desligado Para mudar a escala de temperatura puxe a tampa do compartimento da bateria e ent o mude a chave para a escala C ou F 5 1 Modo de Opera o 1 Retire a tampa protetora e ent o pressione o gatilho para ligar o medidor 2 Aponte o instrumento para o objeto cuja tempe ratura ser medida 3 A medi o ser executada durante todo o tempo em que o gatilho estiver pressionado 4 Verifique a figura referente ao di metro com rela o a dist ncia e o ponto do laser no objeto cuja temperatura dever ser medida 5 Recoloque a tampa protetora para prolongar a vida do sensor e prevenir acidentes pelo uso indevido do laser NOTA Mesmo que o campo de vis o ou campo de medi o coincidam o campo real corresponde ao di metro para 90 da resposta tica O objeto cuja temperatura ser medida precisa ser maior que o campo de medi o pelo menos 1 5 a 2 vezes para uma m
12. f view and the spot almost coincide actu ally the field of measurement corresponds to the diameter for 90 optical response The object whose temperature is to be measured needs to be larger than the measurement diameter spot of size by an adequate margin at least 1 5 to 2 times larger Diameter of Spot mm ce 1000 Distance to Object mm Laser Marker Spot size increases with distance from the probetip as shown Spot Diameter measured at 90 Energy 5 2 Highest Temperature Location Aim the thermometer to the object then start a scan movement across the measured area in up and down motion until locate the hottest point 23 5 3 Field of View Make sure that the object to be measured is larger than the unit s spot size As smaller the object closer the meter should be to it When the accuracy is critical make sure the object is at least twice larger than the spot size As the distance of the meter to the object increase the spot size increase proportionally n CORRECT REMARK lt is not recommended for use in measuring shiny or polished metal surfaces stainless steel alumi num etc Check the emissivity e The thermometer cannot measure through trans parent surfaces such as glass It will measure 24 the surface temperature of the glass instead Steam dust smoke etc can impair accurate mea surement by obstructing the unit s field
13. h Emissivity values range from 0 10 for a very reflective object to 1 00 for a black body Factory set emissivity value of 0 95 which cover 90 of typical applications 27 6 5 Specials Attentions e f the surface to the measured is covered by frost or other material clean it to expose the surface e f the surface to be measured is highly reflective apply masking tape or finish black paint to the sur face e f the meter seems to be giving incorrect readings check the front cone There may be condensation or debris obstructing the sensor clean per in struction in the maintenance section 6 6 Emissivity Table We listed below a few emissivities about some sur faces Substance Emissivity Asphalt 0 9 to 0 98 Concrete 0 94 Cement 0 96 Sand 0 90 Earth 0 92 to 0 96 Water 0 92 to 0 96 Ice 0 96 to 0 98 Snow 0 83 Glass 0 90 to 0 95 Ceramic 0 90 to 0 94 Marble 0 94 Plaster 0 80 to 0 90 28 Brick red 0 93 to 0 96 Cloth black 0 98 Human Skin 0 98 Lather 0 75 to 0 80 Charcoal powder 0 96 Lacquer 0 80 to 0 95 Lacquer mat 0 97 Rubber black 0 94 Plastic 0 85 to 0 95 Timber 0 90 Paper 0 70 to 0 94 Chromium Oxides 0 81 Copper Oxides 0 78 Iron Oxides 0 78 to 0 82 Textiles 0 90 7 ACCESSORIES Upon removing your new digital multimeter from its packing you should have the following items e Instruction Manual e Protective Cover e 9V Battery 29 8 MAINTENANCE 8 1 Battery Replacement
14. ication Display 3 1 2 digit liquid crystal display LCD with maximum reading of 1999 Low Battery Indication The indicator is displayed when the battery voltage drops below the operating level Sensor Cover Measurement Rate 300ms nominal Operating Environment 0 C 50 C RH lt 70 Storage Environment 20 C 60 C RH lt 80 with battery removed from meter Auto Power Off 10 seconds Standby Consuming Current lt 5uA Battery Standard 9V NEDA 1604 IEC 6F22006P Battery Life 9 hours continuous typical with la ser and backlight Dimensions 148 A x 105 L x 42 P mm Weight Approx 157g including battery 3 2 Laser Specification Laser Safety Classification Class II Wavelength Red 630 670nm Operating Distance 2 to 50 feet 0 6 15m Output Power lt 1mW Class Il 3 3 Electrical Specification Temperature Range 30 C 550 C 22 F 1022 F Display Resolution 0 5 1 C automatic 1 F Accuracy t 2 C 4 F to 30 C 100 C 22 F 212 F 2 reading to 101 C 550 C 213 F 1022 F Temperature Coefficient 0 2 of reading or 0 2 C 0 36 F whichever is greater change in accuracy per C or F change in ambient operating temperature above 28 C 82 4 F or below 18 C 64 4 F Response Time 300ms Spectral Response 6 to 14um Fixed Emissivity 0 95 Detection Element Thermopile Optical Lens Fresnel Lens Sighting 1 Bea
15. idade baseada no material que o objeto feito e o acabamento de sua superf cie O valor da emissividade varia de 0 10 para um mate rial muito reflexivo at 1 00 para um corpo negro O valor de emissividade ajustado de f brica de 0 95 para aplica es t picas 6 5 Cuidados Especiais e Se a superf cie a ser medida estiver coberta por gelo ou outro material limpe o para expor a superf cie e Seasuperf cie a ser medida altamente reflexiva aplique uma fita ou tinta preta na superf cie e Se o medidor parecer obter leituras incorretas verifique o cone frontal Pode ter ocorrido condensa o ou fragmentos est o obstruindo o sensor limpe seguindo as instru es na se o de manuten o 6 6 Tabela de Emissividade Est listado abaixo a emissividade de algumas superf cies Subst ncia Emissividade Asfalto 0 9 a 0 98 Concreto 0 94 Cimento 0 96 Areia 0 90 Terra 0 92 a 0 96 Agua 0 92 a 0 96 Gelo 0 96 a 0 98 Neve 0 83 Vidro 0 90 a 0 95 Cer mica 0 90 a 0 94 M rmore 0 94 Reboco 0 80 a 0 90 Tijolo vermelho a 0 96 Pano preto Pele Humana Espuma Carv o Vegetal p Verniz Verniz fosco Borracha preta Pl stico Madeira Papel Oxido de Cromo Oxido de Cobre Oxido de Ferro Tecidos 7 ACESS RIOS Ao receber seu instrumento por favor verifique a exist ncia dos seguintes acess rios a 0 80 a 0 95 a 0 95 a 0 94 a 0 82 S200000000000000 tO INI Oo NI CO C0 to CO O
16. m Laser Marker lt 1mW Class II Field of View 100mm at 1000mm of distance 10 1 20 4 INSTRUMENT DESCRIPTION Fig 1 1 Liquid Crystal Display 2 Trigger 3 Infrared Sensor and Laser Sighting 4 Battery Compartment 5 C F Selection Switch inside battery compartment 21 5 OPERATION Trigger The laser and backlight function work at the same time wen the power on Pull the trigger to turn on the meter when power off Ifthe trigger release the value will be held and HOLD displayed Auto Power Off Function It will auto power off for about 10 seconds Selecting the Temperature Scale Readings are in either degrees Celsius C or de grees Fahrenheit F When the thermometer is turned on it is set to the temperature scale that was in use when the thermometer was turned off To change the temperature scale pull off the battery compartment and then slide the switch to select C or F 5 1 Operation Mode 1 Take the protective cap and them pull the trigger to turn on the meter 2 Point the lens at the object whose temperature is to be measured 3 Measurement is performed as long as trigger is kept 22 4 Referring to the spot size figure aim the laser beam at the object whose temperature is to be measured 5 Pull the cap on to extend life of the sensor and to avoid danger caused by wrong way to use laser REMARK Although the field of measurement or field o
17. ndensa o se dissipe antes de executar a medi o Este instrumento n o a prova de gua ou poeira portanto n o o utilize em ambientes muito contaminados ou midos 6 CONSIDERA ES DA MEDIDA 6 1 Teoria de Medi o Todo o objeto emite energia infravermelha de acordo com a sua temperatura Medindo se a quantidade dessa energia radiante poss vel determinar a temperatura do objeto emissor 6 2 Radia o Infravermelha Radia o infravermelha uma forma de luz radia o eletromagn tica e tem a propriedade de passar 10 facilmente atrav s do ar enquanto facilmente absorvida por mat rias s lidas Com um term metro de emiss o que opera detectando radia o infravermelho poss vel uma medi o precisa Independente da temperatura do ar ou da dist ncia de medi o 6 3 Estrutura do Term metro de Emiss o A radia o que foi emitida pelo objeto focalizada em um sensor de radia o infravermelho via um sistema tico Isto inclui uma lente que transparente para a radia o infravermelho e um filtro de corte de 5 3um A sa da do sensor infravermelho injetado em um circuito eletr nico com sa da de sinal padr o de sen sor de temperatura Termopilha 6 4 Emissividade Todos os objetos emitem energia infravermelho invis vel A quantidade de energia emitida proporcional a temperatura do objeto e sua capacidade em emitir energia IV Esta capacidade chamada de emissiv
18. nit is not constructed to be water proof or dust proof so do not use it in a very dusty envi ronment or in one where it will get wet 6 MEASUREMENT CONSIDERATIONS 6 1 Measurement Theory Every object emits infrared energy in accordance with its temperature By measuring the amount of this radiant energy it is possible to determine the temperature of the emitting object 6 2 Infrared Radiation Infrared radiation is a form of light electromagnetic radiation and has the property that it passes easily through air while it is easily absorbed by solid mat ter 26 With an emission thermometer which operates by detecting infrared radiation accurate measurement is possible irrespective of the air temperature or the measurement distance 6 3 Structure of Emission Thermometer Infrared radiation which has been emitted from the object is focused upon an infrared radiation sensor via an optical system This includes a lens which is transparent to infrared radiation And 5 3um cut off filter The output signal from infrared radiation sen sor is input to an electronic circuit along with the output signal from a standard temperature sensor Thermopile 6 4 Emissivity All objects emit invisible infrared energy The amount of energy emitted is proportional to the object s tem perature and its ability to emit IR energy This ability called emissivity is based upon the material that the object is made of and its surface finis
19. o lt 5uA Bateria Padr o 9V NEDA 1604 IEC 6F22006P Duragao da Bateria 9 horas continuos tipico com laser e ilumina o Dimens es 148 A x 105 L x 42 P mm Peso Aprox 157g incluindo a bateria 3 2 Especifica es do Laser Classifica o de Seguran a do Laser Classe Il Comprimento de Onda Vermelho 630 670nm Dist ncia de Opera o 2 a 50 p s 0 6 15m Pot ncia de Sa da lt 1mW Classe Il 3 3 Especifica es El tricas Faixa de Temperatura 30 C 550 C 22 F 1022 F Resolu o do Display 0 5 1 C autom tico 1 F Precisao 2 C 4 F de 30 C 100 C 22 F 212 F 2 leitura de 101 C 550 C 213 F 1022 F Coeficiente de Temperatura 0 2 leitura ou 0 2 C 0 36 F o que for maior devera ser acrescido na precis o por C ou F quando a temperatura ambiente estiver acima de 28 C 82 4 F ou abaixo de 18 C 64 4 F Tempo de Resposta 300ms Resposta Espectral 6 a 14um Emissividade Fixa 0 95 Elemento Detetor Termo Pilha Lente Otica Lente Fresnel Mira 1 Marcador Laser lt 1mW Classe Il Campo de Vis o 100mm 1000 mm de dist ncia 10 1 4 DESCRI O DO INSTRUMENTO Fig 1 1 Display de Cristal L quido 2 Gatilho 3 Sensor Infravermelho e Mira Laser 4 Compartimento de Bateria 5 Chave de Sele o C F dentro do compartimento de bateria 5 OPERA O Gatilho O laser
20. of view Ay WARNING Do not use the unit near any device which gener ates strong electromagnetic radiation or near a static electrical charge as these may cause er rors Do not use the unit where it may be exposed to corrosive or explosive gases The unit may be damaged or explosion may occur Do not keep or use the unit in an environment were it will be directly illuminated by sunshine or were it will be exposed to high temperatures high humidity or condensation If you do it may be de formed its insulation may be damage or it may no longer function according to specification Do not point the lens at the sun or at any other source of strong light If you do the sensor may be damaged Do not contact the lens against the object whose the temperature is to be measured or get it dirty allow it to be scratched or allow any foreign ma terial to adhere to it Doing so may cause errors Do not touch or hold by the front case Tempera ture reading can be affected by heat from hand 25 Do not place the meter on or around hot objects 70 C 158 F It may cause damage to the case e f the meter is exposed to significant changes in ambient temperature hot to cold or cold to hot Allow 20 minutes for temperature stabilization be fore taking measurement e Condensation may form on the lens when going from a cold to hot environment wait 10 minutes for condensation to dissipate before taking mea surements e This u
21. rificar porusocorreto do aparelho no prazo acima estipulado B Os servi os de repara o ser o efetuados somente no departamento de assist nciat cnica por n s autorizado C Aquisi o for feita em um posto de venda credenciado da Minipa Agarantia perde a validade nos seguintes casos A Mauuso alterado negligenciado ou danificado por acidente ou condi es anormais de opera o ou manuseio B Oaparelho foi violado por t cnico n o autorizado Esta garantia n o abrange fusiveis pilhas baterias e acess rios tais como pontas de prova bolsa paratransporte etc Casoo instrumento contenha software a Minipa garante que o soft ware funcionar realmente de acordo com suas especifica es funcionais por 90 dias A Minipa n o garante que o software n o contenha algum erro ou de que venha a funcionar seminterrup o A Minipa n o assume despesas de frete e riscos de transporte A garantia s ser v lida mediante o cadastramento deste certificado devidamente preenchido e semrasuras Nome Endere o Cidade Estado Fone Nota Fiscal N Data N S rie Nome do Revendedor 15 Cadastramento do Certificado de Garantia O cadastramento pode ser feito atrav s de um dos meios a seguir Correio Envie uma c pia do certificado de garantia devidamente pre enchido pelo correio para o endere o Minipa Ind stria e Com rcio Ltda At Servi o de Atendimento ao Cliente Alameda dos Tupin
Download Pdf Manuals
Related Search
Related Contents
BR-NX10(S) - オンキヨー株式会社 Janome 744D Manual Imation Defender F200 Biometric Flash Drive, 16 GB Corberó - Instructions Manuals conecta-parte1 Bibliographie - Créer son blog Bases de datos - RUA - Universidad de Alicante (la radio logicielle) YPC MOBILE DIV. 2264ALB User Manual Copyright © All rights reserved.