Home

Super Alcomat - Fisher UK Extranet

image

Contents

1. Power tension 100 240 Vac 15 10 85 264V 50 60Hz Power consumption 10 VA Operating temperature 10 30 C suggested 15 25 C Dimensions 210 x 370 x 380 mm Lx D x H Net weight 8 5 kg Data output RS232 I O Power pack Standard equipment 2 hvdrorepellent floaters interchangeable in mass and volume Container for exhaust liquids Certified hvdroalcoholic solution HYDROSTATIC BALANCE Measurement range for the relative 0 5 2 25 density Sensibility 0 00005 Reading repeatability 0 00005 Autocalibration with internal mass YES Response time 6 sec THERMOMETER Type Thermo probe in platinum Pt100 1 3 DIN Maximum temperature displayed 50 Sensibility 0 05 Reading repeatability 0 05 in the range 10 30 C Readability 0 1 C ALCOHOLIC STRENGTH BY VOLUME ASV Reading range 0 13 99 98 ASV sensibility 0 01 ASV Accuracy and reading repeatability 0 03 ASV between 15 and 25 C TOTAL DRY EXTRACT Range of values 0 0 505 8 g l of sucrose Accuracy and reading repeatability 0 3 g l MUSTS EVALUATION Relative density range 1 04 1 16 ACCESSORIES ON REQUEST Floaters with certificate SIT certified solutions 5 60 ASV FLOATERS INTERCHANGEABLE IN MASS AND VOLUME A particular care is devoted to the construction and calibration of the floaters They are all interchangeable in mass and volume to simplify their use for example it is possible to
2. selezionare dal MENU PRINCIPALE Impostazioni premendo 4 Fig 4 quindi selezionare Imposta ora premendo ancora 4 Evidenziare lora o i minuti od i secondi che si vogliono modificare premendo rispettivamente ORE MINUTI o SEC A questo punto modificare il valore premendo o tenendo conto che i due tasti rispettivamente aumentano e diminuiscono di un unit il valore che si sta aggiornando 6 4 Regolazione del contrasto del display del Super Alcomat Per regolare il contrasto del display del Super Alcomat selezionare dal MENU PRINCIPALE Impostazioni premendo 4 Fig 4 quindi selezionare Contrasto display premendo 5 Il tasto REVERSE Fig 16 permette di invertire il contrasto le scritte diventano bianche su sfondo nero mentre NORMALE riporta al contrasto iniziale scritte nere su sfondo bianco Premendo lo sfondo del display diventa pi scuro mentre lo rende pi chiaro Super Alcomat CONTRASTO DISPLAY PER CORREGGERE NORMALE REVERSE STANDARD Fig 16 Pag 17 di 39 ITALIANO SUPER ALCOMAT 6 5 Lingua Se si vuole cambiare la lingua con la quale sono espressi tutti i messaggi che compaiono sul display selezionare Impostazioni dal MENU PRINCIPALE premendo 4 Fig 4 quindi selezionare Lingua premendo 6 Individuare quindi la lingua desiderata Fig 17 fra italiano inglese francese spagnolo e tedesco premendo il tast
3. Relative density of the distillate at 19 4 C measured by the Super Alcomat Relative density at 20 d20 20 of the distillate Fig 7 ASV and d2020 Of a distillate a short acoustic signal and the ASV value shown with white fonts on black background highlight the achievement of this condition The value remains on the display until the removal of the floater from the suspension the specific gravity shown is then related to 20 in accordance with the tables required by the official method EEC Regulations nr 2676 90 and nr 2870 00 The correction is reliable only for hydroalcoholic mixture and in the liquid temperature range 15 25 C Outside this range the results can undergo variations e if the analysis operations are completed press DRAIN to empty the cylinder 5 1 1 Relative density of a liquid When the specific gravity of the liquid in analysis does not fall within the range of values that correspond to an Alcoholic Strength by Volume the display shows only the relative density of the liquid and the temperature Figure 8 With lt 0 79000 or d20 20 gt 0 99995 the liquid is not a hvdroalcoholic mixture ASV does not appear Super Alcomat 19 4 or 6976 20 Liquid temperature Relative density of the liquid at 19 4 working at 20 the SPECIFIC GRAVITY READ is the 120 20 of the liquid Fig 8 Relative density of a liquid 5 1 2 Correction of the floater or compensati
4. b riempire il cilindro con acqua distillata quantit minima 65 ml ad una temperatura compresa tra 15 e 25 C fare attenzione che sul pescante e nel liquido non ci siano bolle d aria c selezionare Titolo Alcolometrico Volumico dal Men principale premendo 1 Fig 4 premere quindi CORREZ per accedere all opzione Correzione Fig 7 premere H20 Fig 9 d sul display vengono visualizzati il fattore di correzione attuale la densit relativa che dovrebbe avere l acqua alla temperatura misurata dalla termosonda e la densit relativa dell acqua effettivamente misurata dallo strumento Fig 10 e se queste due densit coincidono lo strumento in condizioni di lettura ideale in caso contrario occorre variare il fattore di correzione del pescante Premere i tasti e tenendo presente che il tasto aumenta il valore di densit letto e il tasto lo diminuisce Pag 10 di 39 SUPER ALCOMAT ITALIANO Far coincidere le due densit relative aumenta la densit relativa di 5 unit Densit relativa dell acqua misurata con il Super Alcomat 19 5C Densit relativa dell acqua a 19 5 C da tabella Si ritorna a Compensazione Consente l azzeramento manuale diminuisce la densit videata per TAV pescante in alcool prima della lettura di densit relativa di 5 unit Fig 10 Correzione del pescante con acqua distillata 5 1 3 Stampa del valore di TAV o di de
5. che vengono descritte di seguito 6 1 Modifica del progressivo etichetta Quando si determina il TAV di un distillato o la densit relativa di un liquido possibile identificare il campione in esame tramite un numero da riportare in fase di stampa Questo numero rappresentato dal Progressivo etichetta che come gi detto aumenta di 1 unit ogni volta che si stampa ed il valore di TAV o densit diverso dal precedente Per modificare il Progressivo etichetta selezionare dal MENU PRINCIPALE Impostazioni premendo 4 Fig 4 quindi selezionare Progressivo etichetta premendo 2 Mediante i tasti e si varia di 1 unit tale numero progressivo mentre con 10 100 si selezionano le decine e le centinaia Il valore massimo possibile di etichette pari a 999 Le etichette che contengono il solo valore di densit e di temperatura hanno una numerazione che non pu essere variata dall utente 6 2 Aggiornamento data Per aggiornare la data selezionare dal MENU PRINCIPALE Impostazioni premendo 4 Fig 4 quindi selezionare Imposta data premendo 3 Evidenziare il giorno o il mese o l anno che si vuole modificare premendo rispettivamente GIORNO o MESE ANNO A questo punto modificare il valore premendo o l l tenendo conto che i due tasti rispettivamente aumentano e diminuiscono di un unit il valore che si sta aggiornando 6 3 Aggiornamento ora Per aggiornare
6. 5 correspond on the display the access to the desired corresponding function is granted The single functions are then guided by the indications on the display NOTE For ease of writing in the following text the generic functional key correspondent to a certain text is indicated by the text displayed for example instead of saying press the key correspondent to EXIT it will be said press EXIT Pag 26 di 39 SUPER ALCOMAT ENGLISH Super Alcomat can operate at a temperature also different than 20 as long as in the range 15 25 C and allows through the function 1 ALCOHOL STRENGTH BY VOLUME Alcoholic strength by volume ASV to perform the relative density d20 20 measurements of a distillate to visualise in 6 seconds and completely automatically the Alcoholic Strength by Volume of a distillate from 0 13 to 99 98 ASV in agreement with the official tables EEC Regulations 2676 90 and 2870 00 which refer to the International Alcoholometric Tables published in the International recommendation Nr 22 of the O l M L in 1972 Q trough the function 2 TOTAL DRY EXTRACT to determine the total dry extract of a wine or of a must for specific gravity in the range 0 99 1 16 Q through the function 3 MUSTS VALUATION to evaluate the musts by calculating the values of the Baum Brix and Oechsle degrees for d20 20 in the range 1 04 1 16 5 1 RELATIVE DENSITY AT 20 C
7. Alcomat PLA E EMPLE OF DISTILL amp TE FOR DENSITY MEASURE Density Readout 8 1R 5 G Relative density of the wine must read otherwise 000 is shown Wait for the zero or perform a manual zeroing before the reading of the distillate Fig 13 f press MEASURE the display shows temperature specific gravity read and Alcoholic Strength by Volume of the distillate together with the value of the total dry extract in g l of sucrose Figure 14 Super Alcomat rey 19 51707 2096 g l Resor 0 88470 ji 11 88 i TO ver TO Fig 14 Super Alcomat provides automatically to determine the ASV of the distillate using the correction algorithm valid for hvdroalcoholic mixtures and successively the relative density d20 20 of the distillate itself Super Alcomat applies Tabari s formula to determine the relative density of the alcohol free wine must then refers the correspondent tabulated value of total dry extract Once a stable value is reached the total dry extract value is fixed on the display with white fonts on black background preceded by an acoustic signal g once the total dry extract value is calculated it is possible to print the relative label see next paragraph and or delete from the memory the sample of wine must just analysed by keeping pressed DELETE h to exit the total dry extract function press EXIT until the display shows the MAIN MENU i
8. SECCO TOTALE Intervallo di valori 0 0 505 8 g l di saccarosio Precisione e ripetibilit di lettura 0 3 g l VALUTAZIONE MOSTI Intervallo di densit relativa d20 20 1 04 1 16 ACCESSORI A RICHIESTA Pescanti con certificato Soluzioni certificate SIT 5 60 vol PESCANTI INTERCAMBIABILI IN MASSA E VOLUME Particolare cura viene posta nella costruzione e taratura dei pescanti Essi sono tutti intercambiabili in massa e volume per facilitarne l utilizzo ad esempio per determinare la densit relativa di un mosto di un vino tal quale e del relativo TAV si possono usare tranquillamente due pescanti separati evitando in questo modo inquinamenti Per ottenere quanto sopra esposto sia il filo di sospensione che il trattamento superficiale del vetro sono stati oggetto di studi approfonditi per azzerare i problemi causati dalla tensione superficiale dal volume proprio del filo e dalle cariche elettrostatiche Il volume dei pescanti a richiesta pu essere certificato da Istituti competenti europei come lo L N E di Parigi FI M G C di Torino il PTB in Germania etc realizzato con un materiale amagnetico inossidabile e di diametro minore o uguale a 0 20 mm Pag 3 di 39 ITALIANO SUPER ALCOMAT 2 INSTALLAZIONE SUPER ALCOMAT Super Alcomat stato progettato e costruito per resistere e funzionare anche in condizioni gravose E tuttavia consigliabile installare Super Alcomat su un supporto rigido esent
9. a a p pede iaia dirai 25 MEASUREMENTS WIT SUPERALGOMAT iii bzar ia ea 26 RELATIVE DENSITY 20 A DISTILLATE AND ASV ALCOHOLIC STRENGTH BY VOLUME 27 Relative density Ona a a sca 28 Correction of the floater or compensation i 28 Compensation of the floater with the hydroalcoholic mixture ii 29 Compensation of the floater with distilled Water 29 Printing the value of ASV and of relative density ss seen nanna nnrn tannar nn nn Ent nAnnnnnnntnnzzznnnnnan 30 Zeroing of the balance i e a ai ee a ae A eee 31 TOTAL DRY EXTRAGT iene tin ite e lie Mae ee ea See wai ete sed eed 31 Printing the value of TOTAL DRY EXTRACT AAAAEKHAAENEEAINAMAMNEMAINAMMINAMEINAZEMNEMEINEEEZEEEINEZE 33 EVALUATION OF MUSTS c arti lata 34 Printing the value of the degrees Baum Brix and Oechsle of a MUSt 35 SETTINGS OF THE SUPER ALCOMAT alia E eo 36 Modification of the progressive label i 36 Adjustingrtie date list ieri ia lie line ale lara ia 36 Adjustingithe 5 A ea alia lia ina 36 Setting the contrast of the Super Alcomat display i 36 Languages iaia A a Fa aiar 37 SERVICES SUPPLEMENTARY CALIBRATION 37 SERIAL TRANSMISSION ss oh techie aint tn nad el
10. alcolom VOL 3 95 Densita vino mosto a20 C 1 044 Se la stampante non collegata il display mostra il messaggio lampeggiante STAMPA DELL ETICHETTA ATTENDI ERRORE DI STAMPANTE PREMERE UN TASTO PER CONTINUARE Se si attiva la stampante senza aver raggiunto un valore stabile di estratto secco compare sul display il messaggio DATO INSTABILE OPERAZ IMPEDITA 5 3 VALUTAZIONE MOSTI gradi Baum Brix ed Oechsle sono in relazione con la densit relativa d20 20 di un mosto L utilizzo del Super Alcomat consente di determinare automaticamente la massa volumica ed il valore dei gradi Baum Babo Brix ed Oechsle di un mosto per densit d20 20 comprese tra 1 04 e 1 16 in un range di temperatura compreso tra 15 e 25 Sotto i 15 e sopra i 25 i risultati non sono completamente affidabili Si procede come segue a nel caso in cui il mosto non sia limpido provvedere ad una filtrazione b verificare che il pescante sia in posizione sull apposito supporto c selezionare VALUTAZIONE MOSTI dalla pagina del MENU PRINCIPALE premendo 3 Fig 4 Attendere lo zero della bilancia e premere il tasto MEASURE Il pescante viene immerso automaticamente nella soluzione ha inizio la misura Sul display vengono visualizzati temperatura densit letta densit corretta a 20 C con quattro cifre e massa volumica a 20 C del campione di mosto in esame campione valido per d20 20 Compresa tra
11. at the end of the measurement drain the liquid analysed by pressing DRAIN j at the end of every measurement wash abundantly with distilled water in order to accurately clean the cylinder from any remainder 5 2 1 Printing the value of TOTAL DRY EXTRACT To print the value of TOTAL DRY EXTRACT of the wine must under consideration just press PRINT the option is available on the display at the moment of the determination While printing the display of the Super Alcomat shows the blinking message PRINTING LABEL PLEASES WAIT PRINTING OK 5 For further details see Becchetti Analysis methods of wines and alcoholic mixture 1999 by Gibertini elettronica Pag 33 di 39 ENGLISH SUPER ALCOMAT The label is so composed Date 09 03 2001 Hour 10 24 Sample N 29 The identification number of the sample is chosen during the storing phase see Dry extract 9 1 129 2 previous paragraph l Sample Nr IS NOT updated of 1 unit Alcoholic str ASV 3 95 after every printout and it is the last Spec Gravity wine must at 20 C 1 044 value selected in SAVE If the printer is not connected the display shows the following blinking message PRINTING LABEL PLEASES WAIT PRINTER ERROR PRESS ANY KEY TO CONTINUE If the printer is activated before a stable value of dry extract is reached the display shows the blinking message UNSTABLE DATA OPERAT IMPOSSIBLE 5 3 EVALUATION OF MUSTS T
12. con acqua distillata per pulire accuratamente il cilindro da ogni residuo g premere USCITA per tornare a MENU PRINCIPALE 5 3 1 Stampa del valore dei gradi Baum Babo Brix ed Oechsle di un mosto Per avere una stampa relativa alla valutazione di un mosto basta premere STAMPA l opzione presente sul display al momento della determinazione Durante la stampa sul display del Super Alcomat compare il messaggio lampeggiante STAMPA DELL ETICHETTA ATTENDI STAMPA OK L etichetta cos composta Data 10 03 2001 Ora 10 38 Valutazione mosti Numero aggiornato automaticamente di 1 unit ogni volta che si stampa valore impostato tramite NUMERA Campione Mosto N 16 Temperatura mosto 24 6 C Densit mosto a20 C 1 1060 Grado Baum 13 93 Grado Babo 21 35 Grado Brix 24 9 Grado Oechsle 106 0 Se la stampante non collegata il display mostra il messaggio lampeggiante STAMPA DELL ETICHETTA ATTENDI ERRORE DI STAMPANTE PREMERE UN TASTO PER CONTINUARE Se si attiva la stampante senza aver raggiunto un valore stabile di densit compare sul display il messaggio DATO INSTABILE OPERAZ IMPEDITA 3 Vedere per approfondimenti Becchetti Metodi di analisi dei vini e delle bevande spiritose 1999 ed Gibertini elettronica Pag 16 di 39 SUPER ALCOMAT ITALIANO 6 IMPOSTAZIONI SUPER ALCOMAT Per modificare alcune impostazioni il men opzioni presenta diverse possibilit
13. use two different floaters to determine the relative density of a must or a wine in this way possible sample pollution is avoided In order to achieve what mentioned before the suspension wiret and the superficial treatment of the floater s glass were object of thorough studies to nullify the problems caused by the surface tension by the wire s own volume and by the electrostatic charges The floaters volume can on request be certified by competent organisations such as the L N E Paris the I M G C Turin the PTB Germany etc realised with a non magnetic and stainless material with a diameter less or equal to 0 20 mm Pag 22 di 39 SUPER ALCOMAT ENGLISH 2 INSTALLATION OF SUPER ALCOMAT Super Alcomat was projected and built to resist and operate also in harsh conditions However it is advisable to install the Super Alcomat rigid base without vibrations inan environment with stable temperature and without excessive ventilation using a dedicated electrical line 2 1 INSTALLATION PROCEDURE Level the balance acting on the front supports so that the bubble is the centre of the circle on the level unlock the balance by rotating clockwise the knob on the right side of the instrument until it stops Fig 1 e Fig nr 15 UNLOCKED LOCKED Only to transport Fig 1 hang the floater on the dedicated prop Figure nr 7 connect the circular plug of the transformer to the connec
14. 03 2001 Hour Musts evaluation Must sample N Number automatically increased Temperature of must 24 of 1 unit every time a printout is Spec gravity of must at 20 C 1 performed or value chosen with Degr Baum 13 NUMBER Degr Babo Degr Brix Degr Oechsle If the printer is not connected the display shows the following blinking message PRINTING LABEL PLEASES WAIT PRINTER ERROR PRESS ANY KEY TO CONTINUE If the printer is activated before a stable value of density is reached the display shows the blinking message UNSTABLE DATA OPERAT IMPOSSIBLE 6 For further details see R Becchetti Analysis methods of wines and alcoholic mixture 1999 by Gibertini elettronica Pag 35 di 39 ENGLISH SUPER ALCOMAT 6 SETTINGS OF THE SUPER ALCOMAT To modify some settings the options menu presents different possibilities that are described in the following 6 1 Modification of the progressive label When determining the ASV of a distillate or the relative density of a liquid it is possible to identify the sample under consideration through a number to report while printing This number is represented by the progressive label which as already said increases of 1 unit after every printout and the ASV or density value is different from the previous one To modify the Progressive label number select from the MAIN MENU Settings by pressing 4 Figure 4 then select Progressive lab
15. 1 04 e 1 16 Fig 15 d dopo qualche secondo al raggiungimento del valore stabile di densit relativa i valori numerici corrispondenti ai gradi Baum Babo Brix ed Oechsle del mosto appaiono sul display preceduti da un breve segnale acustico mono Densita Massa latta volumica 1 a 085 1 68303 RA STAMPA Fig 15 Super Alcomat determina la densit relativa del mosto in esame La massa volumica del mosto ottenuta moltiplicando il valore della densit corretta a 20 C per 0 998203 g ml massa volumica dell acqua a 20 C Pag 15 di 39 ITALIANO SUPER ALCOMAT In corrispondenza della densit corretta a 20 il Super Alcomat valori di densit presenti in tabella riporta il grado Brix etc corrispondente per i valori di densit non presenti esegue una interpolazione POSSIBILE MEMORIZZARE FINO 99 CAMPIONI DI MOSTO Premere NUMERA mediante i tasti 10 10 assegnare ad ognuno un numero identificativo e premere USCITA Il valore numerico impostato viene indicato sul display e sull eventuale stampa Esso viene automaticamente aggiornato ogni volta che si stampa E perci possibile identificare il campione in esame tramite la funzione NUMERA oppure mediante l aggiornamento automatico di 1 unit della numerazione del campione e al termine della misura scaricare il liquido tramite il tasto DRAIN f lavare abbondantemente
16. 12 Piedini regolazione bolla 17 Serbatoio per riempimento cilindro minimo 65 ml 3 1 PANNELLO COMANDI Il pannello comandi del Super Alcomat fig 3 n 11 dispone di un display e otto tasti funzione ai quali associato un significato diverso a seconda del men visualizzato sul display TEMP a SCI 21 6001 raug Tsaa 0 98710 Pag 5 di 39 ITALIANO SUPER ALCOMAT Ogni tasto deve essere mantenuto premuto fino a quando viene emesso un segnale acustico che indica l accettazione del comando da parte dello strumento NOTA Dopo alcuni minuti se il Super Alcomat non viene utilizzato il display si spegne automaticamente Premendo ON il display nuovamente acceso pulsanti permettono l esecuzione dell azione indicata dal display tasti MEASURE e DRAIN attivano sempre delle funzioni dedicate Il tasto IMEASUREJ permette di abbassare automaticamente il pescante per la misurazione della densit e del TAV di alzare e abbassare il pescante premere 1 sec di alzare il pescante il tempo necessario per permettere un azzeramento automatico del sistema premere 3 sec Il tasto DRAIN consente di alzare e mantenere alzato il pescante premere 1 sec di alzare il pescante e contemporaneamente di attivare l apertura di una elettrovalvola che permette lo svuotamento del cilindro in modo automatico premere 3 sec Il tempo di apertura dell elettrovalvola indica
17. 60 50 720 Ed 4 07 2009 ISTRUZIONI PER L USO INSTRUCTION MANUAL ISTRUZIONI PER L USO INSTRUCTION MANUAL ISTRUZIONI PER L USO INSTRUCTION MANUAL ISTRUZIONI PER L USO INSTRUCTION MANUAL ISTRUZIONI PER L USO INSTRUCTION MANUAL ISTRUZIONI PER L USO INSTRUCTION MANUAL ISTRUZIONI PER L USO INSTRUCTION MANUAL ISTRUZIONI PER L USO INSTRUCTION MANUAL ISTRUZIONI PER L USO INSTRUCTION MANUAL ISTRUZIONI PER L USO INSTRUCTION MANUAL ISTRUZIONI PER L USO INSTRUCTION MANUAL ISTRUZIONI PER L USO INSTRUCTION MANUAL ISTRUZIONI PER L USO INSTRUCTION MANUAL ISTRUZIONI PER L USO INSTRUCTION MANUAL ISTRUZIONI PER L USO Super INSTRUCTION MANUAL ISTRUZIONI PER L USO Mhloomat INSTRUCTION MANUAL ISTRUZIONI PER L USO INSTRUCTION MANUAL Tutte le informazioni riportate in questo manuale sono quelle disponibili al momento della stampa Il costruttore si riserva il diritto di apportare modifiche al prodotto in qualsiasi momento senza preavviso Si consiglia di verificare eventuali aggiornamenti All the information given in this manual is what available at the time of printing The manufacturer reserves the right to make changes to the product at any time without notice It is advisable to verify if there are any updates Tutti i diritti riservati E vietata qualsiasi forma di stampa duplicazione riproduzione o pubblicazione di questo manuale o parte di esso senza l autorizzazione scritta da parte della GIBERTINI ELE
18. ALIBRAZIONE FASE 2 ATTENDERE Il display del SUPER ALCOMAT mostra il MENU PRINCIPALE Ora il sistema SUPER ALCOMAT pronto per lavorare seguito da ATTENZIONE Se all accensione Super Alcomat segnala FUORI SCALA lampeggiante premere il tasto per continuare Il messaggio lampeggiante Bilancia in fase di stabilizzazione termica Attendere viene visualizzato sul display Alla fine della stabilizzazione termica Super Alcomat operativo Qualora il messaggio di stabilizzazione termica persista per oltre 15 minuti premere nuovamente il tasto e far controllare Super Alcomat da un tecnico 5 MISURE CON SUPER ALCOMAT Quando il Super Alcomat pronto per lavorare compare il MENU PRINCIPALE vedere paragrafo n 4 Fig 4 Il MENU PRINCIPALE visualizza le diverse funzioni che si possono svolgere con il dispositivo Super Alcomat Premendo uno dei tasti collocati sulla parte inferiore o sul lato destro del display indicati sul display stesso da 1 o 2 o 3 o 4 o 5 si accede alla funzione desiderata corrispondente Le singole funzioni sono poi guidate dalle indicazioni sul display NOTA Per comodit di scrittura nel testo che segue si usa indicare il generico tasto funzione corrispondente ad un determinato testo presente sul display con il testo corrispondente ad esempio invece di dire premere il tasto corrispondente a USCITA si dir premere USCITA Pag 7 di 39 ITAL
19. IANO SUPER ALCOMAT Super Alcomat pu operare ad una temperatura anche diversa da 20 C purch compresa nell intervallo tra 15 e 25 C e consente di tramite funzione 1 TIT ALC VOL Titolo Alcolometrico Volumico TAV effettuare le misure di densit relativa d20 20 di un distillato visualizzare 6 secondi ed modo completamente automatico il Titolo Alcolometrico Volumico di un distillato da 0 13 a 99 98 vol in accordo con le tabelle ufficiali OIV che fanno riferimento alle Tabelle alcolometriche internazionali pubblicate nella Raccomandazione Internazionale N 22 dell O I M L nel 1972 tramite la funzione 2 ESTRATTO SECCO TOTALE determinare il valore dell estratto secco totale di un vino di un mosto per densit comprese tra 0 99 e 1 16 tramite la funzione 3 VALUTAZIONE MOSTI valutare i mosti calcolando il valore dei gradi Baum Brix e Oechsle per comprese tra 1 04 e 1 16 5 1 DENSIT RELATIVA 20 DI UN DISTILLATO E TAV TITOLO ALCOLOMETRICO VOLUMICO Si opera nel seguente modo a verificare che il pescante sia in posizione sull apposito supporto b riempire il cilindro con il liquido quantit minima 65 ml di cui si vuole misurare la densit versandolo nell apposito serbatoio il livello ottimale di riempimento garantito dal sistema di scarico che impedisce la fuoriuscita del liquido stesso fare attenzione che sul pescant
20. JAPAN IA JAPAN PRofCHINA Sy PRof CHINA CHACL CCIBLAC CHINESE 4 A i HONG KONG CHINA HKAS LS INDIA THAILAND SINGAPORE ar MABL TLAS SPRING VIET INDONESIA Buss TAME of KAN BRAZIL SOUTH AFRICA AUSTRALIA NEWZEALAND INMETRO SANAS NATA IANZ MAPS OF WORLDWIDE RECOGNIZED INTERNATIONAL CENTRES EQUIVALENT TO SIT Pag 39 di 39 gt a info gibertini com Tel 39 023541434 ISO 9001 2008 LAT N 094 GIBERTINI ELETTRONICA s r l Via Bellini 37 20026 Novate Mil se Milano Italy Tel 39 02 3541434 Fax 39 02 3541438 www gibertini com sales gibertini com
21. MEASURE and DRAIN keys activate always dedicated functions The MEASURE key allows to automatically lower the floater for the determination of the specific gravity and of the ASV to lift and lower the floater press 1 sec to lift the floater the necessary time to allow an automatic zero setting of the system press 3 sec The DRAIN key allows to lift and hold in position the floater press 1 sec to lift the floater and at the same time to open an electrovalve that allows the automatic emptying of the cylinder press 3 sec A flashing led and an acoustic signal indicate the opening time of the valve After the emptying the valve closes automatically 4 IGNITION OF SUPER ALCOMAT At ignition of the Super Alcomat the software release number is displayed Figure 3 anticipated by an acoustic signal and by the name of the company Super GIBERTINI DENSI ALCO Fig 3 After about two seconds the MAIN MENU appears ALCOHOL STRENGTH BY TOTAL DAY EXTRACT Wats VALLETIOA SETTINGS SERVICES Fig 4 By pressing the function keys 4 or 5 it is possible to access to Settings or Services see paragraph 6 by pressing the function keys 1 or 2 or 3 the display shows the following blinking message Pag 25 di 39 ENGLISH SUPER ALCOMAT Super Alcomat SCALE PERFORMING THERMAL STABILIZATION PLS WAIT Fig 5 Any key pressed in this situat
22. OF A DISTILLATE AND ASV ALCOHOLIC STRENGTH BY VOLUME It is done as follows a verify that the floater is in position on the dedicated prop b fill the cylinder with the liquid minimum amount 65 ml of which the specific gravity must be measured by pouring in the apposite tank the optimal replenishment level is granted by the automatic drainage systems which prevent the leek of the liquid itself pay attention that no bubbles are present on the floater and in the liquid c press 1 to select Alcoholic Strength by Volume The Super Alcomat display shows Super Alcomat Temperature measured by the thermo probe of the Super Alcomat It indicates that the Super Alcomat is in the zero condition it appears at the end of the calibration and or in stand by condition If anticipated by a or a it indicates that the hydrostatic balance is close to the zero Fig 6 Zero condition of the hydrostatic balance d press MEASURE the floater will be automatically lowered read the ASV value on the display the ASV and the relative density at 20 C appear on the display after the measured relative density the numbers change until the data does not reach stability Pag 27 di 39 ENGLISH SUPER ALCOMAT ASV of the distillate at 20 C ASV Temperature of the distillate The temperature indication for values lower than 15 or higher than 25 C is displayed in a box with contrasting colour
23. OMAT ENGLISH 5 1 4 Zeroing of the balance The instrument offers the possibility to perform an automatic zeroing during the measuring procedure By keeping pressed MEASURE the floater is lifted for some seconds allowing the balance to perform the zeroing Super Alcomat offers also the possibility to perform a manual zeroing of the balance a if the cylinder is full press DRAIN to drain the sample and lift the floater b press ZERO the display shows zero followed by the initial screen c if the suspension is occupied the display shows ZEROSET IMPOSSIBLE 5 2 DRY EXTRACT The total dry extract includes all matter that is non volatile under specified physical conditions The official European method is densimetric EEC Regulation Nr 2676 90 Attachment 4 Total Dry Extract The total dry extract is calculated indirectly basing upon the value of the specific gravity of the alcohol free must wine i e of the must wine from which the alcohol was eliminated and which was brought back to the initial volume by adding distilled water The total dry extract is expressed by the concentration in g l of a sucrose solution with the same specific gravity of the alcohol free must wine It is done as follows a verify that the floater is in position on the dedicated prop b select TOTAL DRY EXTRACT from the page MAIN MENU by pressing 2 Figure 4 The instrument awaits for the measure of the relative density of
24. POSSIBILE MEMORIZZARE FINO A 99 CAMPIONI DI VINO MOSTO Questa funzione consente di effettuare prima tutte le misure dei campioni di vino mosto per poi richiamarle in seguito al momento della misura della densit dei distillati corrispondenti Per evitare l inquinamento del campione usare un altro pescante previo controllo del fattore di correzione vedere paragrafo 5 1 2 Per memorizzare un campione premere IMEMORIZ Fig 12 mediante i tasti 10 l 10 assegnare ad ognuno un numero identificativo confermare il valore con SALVA operazione consentita solo se il valore di densit relativa a 20 C stabile e passare alla misura di un altro campione di vino mosto svuotando il cilindro scaricare il precedente liquido mediante la pressione di DRAIN lavare abbondantemente con acqua distillata per pulire accuratamente il cilindro da ogni residuo consigliabile un operazione di avvinamento del cilindro per evitare inquinamenti nel caso si disponga di abbastanza liquido versare il distillato alcolico del vino mosto attendere lo 000 della bilancia premere MEASURE per procedere alla nuova misura Per richiamare un campione premere RICHIAM o RICHIAMA mediante i tasti 10 10 selezionare il numero identificativo del campione da richiamare confermare la scelta premendo RICHIAM e passare alla determinazione dell estratto secco premendo CALCOLO e per passare direttamente alla de
25. SELF CALIBRATION ACTIVATED PHASE 2 PLS WAIT During the first and the second calibration phases it is not possible to use the instrument followed by Pag 37 di 39 ENGLISH SUPER ALCOMAT 7 SERIAL TRANSMISSION The serial transmission of tte SUPER ALCOMAT uses the standard EIA RS232 7 1 CONNECTION OF THE SERIAL OUTPUT The connection is the following Sionas gopek arcomat Nog RS 232 Out Pin 2 Pin 3 RS 232 In Pin 3 Pin 2 Ground Pin 5 Pin 5 7 2 RS232 COMMUNICATION PROTOCOL BETWEEN THE SUPER ALCOMAT AND THE PC By means of the RS232 communication protocol it is possible to receive with a computer the data visualised on the Super Alcomat according to the following protocol Configuration of the serial port COMI 9600 8 N 1 COMMANDS TABLE Valid Answer Answer Commands to legal commands to illegal commands T lt CR gt lt Temperature gt lt CR gt lt gt lt CR gt S lt CR gt lt Density read gt lt CR gt lt gt lt CR gt D lt CR gt lt Density corrected at 20 C gt lt CR gt lt gt lt CR gt A lt CR gt lt Alcoholic strength gt lt CR gt lt gt lt CR gt Phase 1 lt Dry extract gt lt gt lt CR gt MMA Phase 2 lt Dry Extract gt lt Alcoholic Strength gt lt CR gt ica M lt CR gt lt Mass gt lt Baume gt lt Babo gt lt Brix gt lt Oe gt lt CR gt lt gt lt CR gt The command is
26. TTRONICA Srl All rights reserved Forbidden any form of print duplication reproduction and publication of this instruction manual or part of it without the written agreement of GIBERTINI ELETTRONICA Srl e W WN N Roihi Rok io 5 3 1 6 1 6 2 6 3 6 4 6 5 6 6 Tal 7 2 es e e 524 or w 5 3 1 6 1 6 2 6 3 6 4 6 5 6 6 7 1 7 2 INDICE INDEX CARATTERISTICHE TECNICHE SUPER ALCOMAT ii 3 INSTALEAZIONE SUPER ALGOMAT i rai AOL Aa Arai isa 4 PROCEDURA D INSTALLAZIONE arie leali 4 DESCRIZIONE SUPER A COMA b A vai adi TA THE 5 PANNELLO COMAND aret ererat AIA RSA AA 5 ACCENSIONE SUPERALGOMAT lle iii erede eal 6 MISURE CON SUPER ALC OMAT p A ia 7 DENSIT RELATIVA 20 C DI UN DISTILLATO E TAV TITOLO ALCOLOMETRICO VOLUMICO 8 Densit relativa di um liglido i 8 i ia A p aria liana alette i ila ansia 9 Correzione del pescante compensazione i 9 Correzione del pescante con la soluzione idroalcolica ii 10 Correzione del pescante con acqua distillata ii 10 Stampa del valore di TAV di densit relativa 00 i 11 Azzeramento bilancia ne JA Ali
27. a SUPER ALCOMAT secondo il seguente protocollo Configurazione porta seriale COMI 9600 8 N 1 TABELLA COMANDI Comandi Risposta Risposta validi a comando legale a comando illegale T lt CR gt lt Temperatura gt lt CR gt lt gt lt CR gt S lt CR gt lt Densit letta gt lt CR gt lt gt lt CR gt D lt CR gt lt Densita corretta a 20 C gt lt CR gt lt gt lt CR gt A lt CR gt lt Grado alcolico gt lt CR gt lt gt lt CR gt Fase 1 lt Estratto secco lt gt lt CR gt E lt CR gt Fase 2 lt Estratto secco gt lt Grado alcolico gt lt CR gt M lt CR gt lt Massa gt lt Baume gt lt Babo gt lt Brix gt lt Oe gt lt CR gt lt gt lt CR gt Il comando definito illegale quando la condizione operativa non ne consente la visualizzazione N B Per lt CR gt si intende il carattere ASCII Carriage Return SUPER ALCOMAT fornisce sempre una risposta a tutti i comandi ricevuti se il comando viene riconosciuto ed il contesto valido la risposta sara il valore richiesto seguito dal carattere di fine riga lt CR gt se il comando viene correttamente riconosciuto ma viene inviato al di fuori di un contesto valido es richiesta estratto secco in fase di misura Alcool la risposta sar una serie di quattro trattini lt gt seguita dal carattere di fine riga lt CR gt se il comando non viene riconosciuto co
28. a nona Dana i e AN 12 ESTRATTO SECCO TOTALE pain ara aa ai in 12 Stampa del valore di ESTRATTO SECCO TOTALE 14 VALUTAZIONEMOSTI LE lea o ee leet are 15 Stampa del valore dei gradi Baum Brix ed Oechsle di UN MOSto 16 IMPOSTAZIONI SUPER ALGOMAT illi g iaia aa 17 Modifica del progressivo etichetta ii 17 Aggiornamento dat s siriana aaa 17 Aggiornamento Ofra tartine Oa pa ALI LO LAI ae 17 Regolazione del contrasto del display del Super Alcomat ii 17 alii ia Li era a 18 SERVIZI CALIBRAZIONE SUPPLEMENTARE 18 TRASMISSIONE SERIALE fai ies rae ieee ra a 19 GOLLEGAMENTO DELEWSGITA SERIALE is fii i a it 19 PROTOCOLLO DI COMUNICAZIONE 5232 FRA SUPER ALCOMAT E PO 19 CONSIGLI D USO E MANUTENZIONE i 20 ini iii io ec iii e N ci Sg 20 TECHNICAL CHARACTERISTICS SUPER ALCOMAT 22 INSTALLATION OF SUPERAEGOMA Ti i alal lalla sea 23 INSTALLATIONiIPROGEDURE mice oein asea d A B dea 23 DESCRIPTION OFSUPERALGOMATI 2 a a 24 COMMAND PANEL ALA VA g 24 IGNITION OF SUPER ALGOMAT i ti a i pa
29. are che il pescante sia in posizione sull apposito supporto b selezionare ESTRATTO SECCO TOTALE dalla pagina del MENU PRINCIPALE premendo 2 Fig 4 Lo strumento rimane in attesa della misura della densit relativa del vino mosto tal quale campione valido con una densit relativa compresa tra 0 99 e 1 16 con il messaggio lampeggiante Super Alcomat POSIZIONARE IL GAMPIONE VINO MOSTO MISIRA DENSITA Fig 11 Densita Latta 19 6 C 000 c attendere lo zero della bilancia oppure premere ZERO per un azzeramento manuale quindi premere il tasto IMEASUREJ d sul display vengono visualizzati temperatura densit letta densit corretta a 20 C e massa volumica a 20 C del campione di mosto vino in esame Fig 12 Pag 12 di 39 SUPER ALCOMAT ITALIANO Super Alcomat Massa volumica del vino mosto tal quale Richiama il campione Memorizza il campione assa direttamente al calcolo di vino mosto di vino mosto dell ESTRATTO SECCO Fig 12 Super Alcomat provvede automaticamente a riportare a 20 C il valore letto di densit AI raggiungimento del valore stabile il valore di densit corretta a 20 C espresso con quattro cifre decimali viene fissato sul display con caratteri bianchi su sfondo nero preceduto da un breve segnale acustico La massa volumica ottenuta moltiplicando il valore della densit corretta a 20 C per 0 998203 g ml massa volumica dell acqua a 20 C E
30. defined as illegal when the operating condition does not allow its visualisation N B lt CR gt means the ASCII character Carriage Return SUPER ALCOMAT provides always a correct answer to all the commands received if the command is recognised and the context is valid the answer will be the value requested followed by the end line character lt CR gt if the command is correctly recognised but it is sent out of a valid context for example request of the total dry extract while measuring alcohol the answer will be a series of 4 dashes lt gt followed by the end line character lt CR gt if the command is not recognised different command than those indicated in the table the answer will be a lt gt followed by the end line character lt CR gt Pag 38 di 39 SUPER ALCOMAT ENGLISH 8 USE AND MAINTENANCE SUGGESTIONS It is advisable to group the hydroalcoholic solutions to examine in homogeneous groups for example from 9 to 12 ASV from 15 to 20 ASV from 40 to 45 ASV This is made to prevent pollution between a determination and the following In other words it is NOT advisable to pass from 10 to 40 and vice versa unless after doing an abundant rinsing with wine if enough liquid is available When passing from 10 to 40 wash profusely with H20 to prevent contamination Wash with H20 or alcohol at the end of the work A dedicated microprocessor provides besides controlling t
31. e da vibrazioni in un ambiente a temperatura stabile e senza eccessive ventilazioni utilizzando una linea elettrica privilegiata 21 PROCEDURA DI INSTALLAZIONE Mettere in bolla la bilancia agendo sui piedini anteriori in modo tale che la bolla d aria sia posta all interno del cerchio presente sulla bolla sbloccare la bilancia ruotando in senso orario e fino al suo fine corsa la manopola che si trova sul lato destro dello strumento Fig 1 e Fig 3 n 15 SBLOCCO BLOCCO Solo per il trasporto Fig 1 posizionare il pescante sull apposito supporto figura n 7 collegare lo spinotto circolare dell alimentatore al connettore presente sul Super Alcomat indicato in Fig 2 col n 4 Inserire quindi la spina alimentatore in una presa di rete 1 Bolla 2 Connettore stampante 3 Connettore RS232 4 Connettore alimentatore 5 Matricola 6 Collegamento seriale di servizio Fig 2 Vista posteriore SUPER ALCOMAT Pag 4 di 39 SUPER ALCOMAT ITALIANO 3 DESCRIZIONE SUPER ALCOMAT Lo strumento costituito da una bilancia idrostatica elettronica con sensibilit quinta cifra decimale e da un modulo elaboratore integrato nella bilancia stessa Fig 3 7 Supporto pescante 13 Pomoli del centra cilindro 8 Tubo di scarico 14 Centra cilindro 9 Cilindro e pescante 15 Manopola di blocco della bilancia 10 Elettrovalvola per il trasporto 11 Pannello comandi Display 16 Termosonda
32. e e nel liquido non ci siano bolle d aria c Premere 1 per selezionare Titolo Alcolometrico Volumico Il display del Super Alcomat apparir come nell immagine seguente Super Alcomat TOL ALCOLOM VOL SVOL 20 C Temperatura letta dalla termosonda del Super Alcomat Indica che il Super Alcomat si trova nella condizione di zero compare al termine della calibrazione e o in condizioni di riposo Preceduto da un o un indica che la bilancia idrostatica si trova nei pressi dello zero Fig 6 Condizione di zero della bilancia idrostatica d premere MEASURE all avvio della procedura di misurazione il pescante verr automaticamente abbassato per effettuare la misura leggere il valore di TAV sul display il TAV e la densit relativa a 20 C compaiono sul display successivamente alla densit relativa letta le cifre cambiano fino a quando il dato non raggiunge la stabilit Pag 8 di 39 SUPER ALCOMAT ITALIANO TAV del distillato 20 vol Temperatura del distillato L indicazione della temperatura per valori inferioria 15 C o superiori a 25 viene visualizzata in un riquadro con colore contrastante Super Alcomat algm LO VOL VOL a 20 C Densit relativa de distillato a 19 4 4 misurata dal Super Alcomat Densit relativa a 20 d20 20 del distillato Fig 7 TAVe d20 20 di un distillato breve segna
33. el by pressing 2 By pressing and that progressive number is varied of 1 unit while with 10 100 tens and hundreds are selected The maximum possible value for the labels is 999 The labels that include only the value of the specific gravity and the temperature have a numbering that cannot be varied by the user 6 2 Adjusting the date To adjust the date select Settings from the MAIN MENU by pressing 4 Figure 4 then select Set the date by pressing 3 Highlight the day or the month or the year to modify by pressing respectively DAY or MONTH or YEAR At this point modify the value by pressing or Keeping in mind that the two keys increase and decrease respectively of 1 unit the value under adjustment 6 3 Adjusting the time To adjust the time select Settings from the MAIN MENU by pressing 4 Figure 4 then select Set the time by pressing 4 Highlight the hour or the minutes or the seconds to modify by pressing respectively HOUR or MIN or SEC At this point modify the value by pressing or keeping in mind that the two keys increase and decrease respectively of 1 unit the value under adjustment 6 4 Setting the contrast of the Super Alcomat display To adjust the contrast of the Super Alcomat display select Settings from the MAIN MENU by pressing 4 Figure 4 then select Display Contrast by pressing 5 The REVERSE key Figure 16 allows inverti
34. es To prevent sample pollution use another floater given that the compensation factor is correctly set see paragraph 5 1 2 To store a sample press SAVE Figure 12 through the keys 10 10 assign to every sample an identification number confirm the value with SAVE operation allowed only if the value of relative density at 20 C is stable and pass to the measurement of another sample of wine must by emptying the cylinder drain the previous liquid by pressing DRAIN wash abundantly with distilled water in order to accurately clean the cylinder from any remainder Rinsing with wine is suggested to prevent contamination in case a sufficient amount of liquid is available pour the alcoholic distillate of wine must wait for the 000 of the balance press MEASURE to proceed to the next measurement To recall a sample press RECALL through the keys l 10 10 select the identification number of the sample to recall confirm the selection by pressing RECALL and pass to the determination of the dry extract by pressing CALCULAT e to pass directly to the determination of the dry extract press CALCULAT The instruments awaits to measure the relative density of the distillate obtained by the wine must under consideration valid samples have a relative density d20 20 between 0 79000 and 0 99995 with the blinking message Fig 13 Pag 32 di 39 SUPER ALCOMAT ENGLISH Super
35. gge possono essere usati per tutte le procedure di certificazione omologazioni ed accreditamento di prodotti servizi e sistemi di qualita aziendale La Gibertini quindi pu emettere certificati di taratura che garantiscono il mantenimento della riferibilit degli apparecchi usati dal Centro a campioni nazionali delle unit del Sistema Internazionale delle Unit S 1 la correttezza metrologica delle procedure di misura adottate dal Centro A RICHIESTA soluzioni idroalcoliche certificate SIT dal 5 vol al 60 vol NETHERLANDS BELGIUM NORWAY SWEDEN DENMARK FINLAND _RVA__ BELTEST NA SWEDAC DANAK FINAS IRLAND UNITED KINGDOM AUSTRIA GERMANY CZECH REPUBLIC SLOVAKIA NAB UKAS BMWA DASMIN DATech CAI SNAS PORTUGAL SPAIN FRANCE SWITZERLAND ITALY ISRA L IPQ ENAC COFRAC SAS SINAL SIT ISRAC aut 1 Ae CANADA lt SCC sa 21 ICB ES REP of KOREA KOLAS JAPAN JAB JAPAN PROTCHINA PRof CHINA CCIBLAC CHINESETAIPEI CHLA HONG KONG CHINA HKAS gt INDIA THAILAND SINGAPORE a NABL TLAS SPRING VIET INDONESIA Buss TAME BRAZIL SOUTH AFRICA AUSTRALIA NEWZEALAND INME TRO SANAS NATA IANZ MAPPA DEI CENTRI INTERNAZIONALI CORRISPONDENTI AL SIT Pag 20 di 39 ENGLISH SUPER ALCOMAT 1 TECHNICAL CHARACTERISTICS SUPER ALCOMAT GENERAL CHARACTERISTICS
36. gravity correct at 20 C for 0 998203 g ml density of water at 20 C Pag 34 di 39 SUPER ALCOMAT ENGLISH In correspondence of the density correct at 20 C the Super Alcomat reports the degree Brix etc correspondent if the values of density are present in the table u performs an interpolation for values of density not present in the table IT IS POSSIBLE TO ENUMERATE UP TO 99 MUST SAMPLES press NUMBER by pressing 10 10 assign to every sample an identification number and press EXIT the chosen numerical value is shown on the display and on the possible printout It is automatically updated every time a printout is made It is possible therefore to identify the sample under consideration through the function NUMBER or through the automatic updating of 1 unit of the sample enumeration e at the end of the measurements drain the liquid by pressing DRAIN f wash abundantly with distilled water in order to accurately clean the cylinder from any remainder g press EXIT to go back to the MAIN MENU 5 3 1 Printing the value of the degrees Baum Babo Brix and Oechsle of a must To have a printout relative to the evaluation of a must just press PRINT the option is available on the display at the moment of the determination While printing the display of the Super Alcomat shows the blinking message PRINTING LABEL PLEASES WAIT PRINTING OK The label is so composed Date 10
37. he degrees Baum Brix and Oechsle are in relation with the relative density d20 20 of a must The use of the Super Alcomat allows to determine automatically the density and the value of the degrees Baum Babo Brix and Oechsle of a must for density between 1 04 and 1 16 in the range of temperature 15 25 C Below 15 and above 25 C the results are not completely reliable It is done as follows a in case the must is not clear perform a filtration b verify that the floater is in position on the dedicated prop c select MUSTS EVALUATION in the MAIN MENU screen by pressing 3 Figure 4 Wait for the zero of the balance and press MEASURE The floater is automatically submerged in the liquid and the measurement begins The display shows temperature specific gravity read specific gravity correct at 20 C with four decimal numeral and density at 20 C of the must sample under consideration the sample is valid for d20 20 between 1 04 and 1 16 Figure 15 d after some seconds after a stable relative density value is reached the numerical values correspondent to the degrees Baum Babo Brix and Oechsle of the must are shown on the display preceded by a short acoustic signal Super Alcomat Danoity readout 20 6 Fig 15 Super Alcomat determines the relative density d20 20 of the must under consideration The density of the must is obtained by multiplying the value of the specific
38. he most important functions of the Super Alcomat to diagnose internal problems of the balance which mav have effects on the precision suggesting a programmed service after 10000 hours of functioning with qualified personnel The displaved message is The instrument requires maintenance press a kev to continue which however does not compromise the use of the balance 9 SIT CALIBRATION Gibertini Elettronica is SIT CENTER n 94 law 273 91 accredited for the characterization of electronic balances with capacity till 30 kg The certifications issued have an official value they are documents that according to the law can be used for all the procedures of certification homologations and accreditation of products services and systems of business quality Gibertini therefore can emit the calibration certificate that guarantees the maintenance of the referral of apparatus used by the Centre with national samples of unity of the International System of Units 5 1 the metrological correctness of the measuring procedure adopted by the Centre UPON REQUEST SIT Certified hydroalcoholic solutions from 5 ASV to 60 ASV NETHERLANDS BELGIUM NORWAY SWEDEN DENMARK FINLAND RvA BELTEST BKO OBE SWEDAC DANAK _ FINAS IRLAND UNITED KINGDOM AUSTRIA GERMANY CZECH REPUBLIC SLOVAKIA NAB UKAS BMWA DASMIN DATech S AS PORTUGAL SPAIN FRANCE SWITZERLAND ITALY ISRAEL IPQ ENAC COFRAC SAS SINAL SIT ISRAC REP of KOREA KOLAS
39. i modifica invece il cosiddetto fattore di correzione del pescante in modo tale che tutti i pescanti diano la stessa risposta nelle medesime condizioni standard Tali condizioni sono la determinazione della densit relativa dell acqua distillata e la determinazione del TAV o grado alcolico di una soluzione idroalcolica a titolo noto certificata SIT fornibile a richiesta dalla nostra azienda Le due possibilit sono selezionabili dalla tastiera premendo H20 o ALCOOL rispettivamente Pag 9 di 39 ITALIANO SUPER ALCOMAT Sul display appare il valore attuale del fattore di correzione che pu essere aumentato o diminuito di cinque unit alla volta premendo i tasti e il fattore di correzione rimane memorizzato anche se lo strumento viene spento E consigliabile controllare con regolarit tale fattore 5 1 2 1 Correzione del pescante con la soluzione idroalcolica Gibertini certificata SIT Si procede come segue a verificare che il pescante sia in posizione sull apposito supporto b riempire il cilindro soluzione idroalcolica a titolo noto quantit minima 65 ml ad una temperatura compresa tra 15 e 25 C fare attenzione che sul pescante e nel liquido non ci siano bolle d aria selezionare Titolo Alcolometrico Volumico dal Menu principale premendo 1 Fig 4 accedere all opzione Correzione con il tasto corrispondente Fig 7 sul display vengono visualizzati il fattore di correzione attuale
40. ion Fig 5 brings back to the main menu i e the Super Alcomat during thermal stabilization cannot be used to perform measurements but the keyboard remains active The blinking message if the keyboard is not touched remains on the display until the Super Alcomat is in the thermal stabilization phase about 5 minutes At the end of the thermal stabilization phase the following blinking message appears on the Super Alcomat display SELF CALIBRATION ACTIVATED PHASE 1 PLS WAIT SELF CALIBRATION ACTIVATED PHASE 2 PLS WAIT The display of the Super Alcomat shows the MAIN MENU The system is now ready to work followed by ATTENTION If at ignition the Super Alcomat displays the blinking message OUT OF RANGE press the key that indicates on the Super Alcomat to continue The blinking message SCALE PERFORMING THERMAL STABILIZATION PLS WAIT is displayed At the end of the thermal stabilization the Super Alcomat is operating In case the thermal stabilization message persists for more than 15 minutes press again and have the Super Alcomat checked by a technician 5 MEASUREMENTS WITH SUPER ALCOMAT When the Super Alcomat is ready to work the display shows the MAIN MENU see paragraph 4 Figure 4 The MAIN MENU presents now the different functions that can be performed by the Super Alcomat By pressing one of the keys located on the inferior part or on the right side of the display to which 1 o 2 o 3 o 4 o
41. ion with the corresponding key Figure 7 the display shows the current compensation factor the relative density of the hydroalcoholic mixture the specific gravity of the same mixture corrected at 20 C and the alcoholic strength Figure 9 if the value of the alcoholic strength shown by the display is equal to that declared on the certificate of the mixture used as reference then the instrument is in the ideal reading conditions on the contrary it is necessary to vary the compensation factor of the floater Press or keeping in mind that increases the alcoholic strength read and decreases it It is necessary to point out that the variation of the compensation factor and the variation of the alcoholic strength don t correspond exactly Moreover these variations are very slow proceed therefore waiting some seconds before reading the value on the display and stop on the value that is closest to the reference value d e The ASV measured with the Super Alcomat and that on Decrease the ASV the mixture certificate must agree Increase the ASV ASV or alcoholic strength calculated starting from the relative specific gravity reaa Relative density of the hydro alcoholic mixture read with Super Alcomat Compensation of the It allows a manual zeroing before Previous screen gt floater in water the reading Fig 9 Compensation of the floater with the hydroalcoholic solution 5 1 2 2 Compensation
42. la densit relativa della soluzione idroalcolica la densit della stessa corretta a 20 C ed il grado alcolico Fig 9 se il valore di grado alcolico letto sul display coincide con quello dichiarato sul certificato della soluzione idroalcolica usata come riferimento allora lo strumento in condizioni di lettura ideale in caso contrario occorre variare il fattore di correzione del pescante Premere i tasti e tenendo presente che il tasto aumenta il grado alcolico letto e il tasto lo diminuisce Si fa notare che tra le variazioni del fattore di correzione e quelle del grado alcolico non esiste una corrispondenza esatta e che tali variazioni avvengono molto lentamente procedere quindi attendendo qualche secondo prima di leggere il valore sul display e fermarsi sul valore pi vicino a quello di riferimento c d Devono coincidere il TAV misurato con Super Alcomat e il valore sul certificato della soluzione idroalcolica TAV BUCA TAV TAV o grado alcolico calcolato a partire dalla densit relativa letta Densit relativa della miscela idroalcolica letta con Super Alcomat Compensazione Consente azzeramento manuale Videata precedente pescante in acqua prima della lettura Fig 9 Correzione del pescante con la soluzione idroalcolica 5 1 2 2 Correzione del pescante con acqua distillata Si procede come segue a verificare che il pescante sia in posizione sull apposito supporto
43. la eee Gi ei heed Sed 38 CONNECTION OF THE SERIAL OUTPUT i ia a a a f a a i 38 RS232 COMMUNICATION PROTOCOL BETWEEN THE SUPER ALCOMAT AND THE 38 USE AND MAINTENANCE SUGGESTIONS iii 39 SIP CALIBRATION A on a p alari 39 SUPER ALCOMAT ITALIANO 1 CARATTERISTICHE TECNICHE SUPER ALCOMAT CARATTERISTICHE GENERALI Tensione d alimentazione 100 240 Vac 15 10 85 264V 50 60Hz Assorbimento di corrente 10 VA Temperatura d utilizzo 10 30 consigliata 15 25 C Dimensioni 210 x 370 x 380 mm Lx Px A Peso netto 8 5 kg Uscita dati RS232 I O Alimentatore Dotazione standard 2 pescanti idrorepellenti intercambiabili in massa e volume Contenitore per scarico dei liquidi Soluzione idroalcolica certificata BILANCIA IDROSTATICA Intervallo di misura densit relativa 0 5 2 25 Sensibilit 0 00005 Ripetibilit di lettura 0 00005 Calibrazione interna SI Tempo di risposta 6 sec TERMOMETRO Tipo Termosonda al platino Pt100 1 3 DIN Temperatura massima visualizzata 50 C Sensibilit 0 05 Ripetibilit di lettura 0 05 C nel range 10 30 C Leggibilit 0 1 C TITOLO ALCOLOMETRICO VOLUMICO TAV Intervallo di lettura 0 13 99 98 vol Sensibilit 0 01 vol Precisione e ripetibilit di lettura 0 03 vol tra 15 e 25 C ESTRATTO
44. le acustico ed il valore del TAV visualizzato con caratteri bianchi su fondo nero evidenziano il raggiungimento di questa condizione Il dato rimane fissato sul display fino alla rimozione del pescante dalla sospensione la densit letta viene corretta 20 C in accordo con le tabelle previste dal metodo ufficiale OIV e n 2870 00 La correzione affidabile solo per le miscele idroalcoliche e nell intervallo di temperatura del liquido in esame compreso tra 15 e 25 Al di fuori di quest intervallo i risultati possono subire delle variazioni e se le operazioni di analisi sul liquido sono completate premere DRAIN per svuotare il cilindro 5 1 1 Densit relativa di un liquido Quando il valore di densit del liquido in esame non rientra nel campo dei valori che corrispondono ad un Titolo Alcolometrico Volumico il display visualizza soltanto la densit relativa del liquido e la temperatura Fig 8 Per lt 0 79000 gt 0 99995 non si ha una miscela idroalcolica TAV non compare Super Alcomat TITOL 19 4 0 5 aroc 1 00185 SENS Temperatura del liquido Densit relativa liquido 19 4 operando a 20 C la DENS LETTA la d20 20 del liquido Fig 8 Densit relativa di un liquido 5 1 2 Correzione del pescante o compensazione Mediante la procedura di Correzione del pescante non si modifica n la massa n il volume del pescante che si sta usando S
45. mando diverso da quelli elencati in tabella la risposta sara un lt gt seguito da un carattere di fine riga lt CR gt Pag 19 di 39 ITALIANO SUPER ALCOMAT 8 CONSIGLI D USO E MANUTENZIONE Si consiglia di raggruppare le soluzioni idroalcoliche da esaminare in gruppi omogenei es da 9 a 12 TAV da 15 a 20 TAV da 40 a 45 TAV Quanto sopra per evitare inquinamenti tra una determinazione e l altra In altre parole NON consigliabile passare da 10 a 40 e viceversa a meno di non effettuare un abbondante avvinamento se si dispone di liquido a sufficienza Dovendo passare da 10 a 40 lavare abbondantemente con H20 per evitare inquinamenti Al termine del lavoro lavare con oppure alcool Un microprocessore dedicato provvede oltre al pilotaggio delle pi importanti funzioni del Super Alcomat a diagnosticare problemi interni alla bilancia che potrebbero avere riflessi sulla precisione consigliando dopo 10000 ore di funzionamento un intervento programmato effettuato da personale qualificato Il messaggio visualizzato Lo strumento richiede manutenzione premere un tasto per continuare che tuttavia non compromette l utilizzo della bilancia 9 TARATURA SIT La Gibertini Elettronica CENTRO SIT 94 legge 273 91 accreditato per la caratterizzazione delle bilance elettroniche con portata fino a 30 kg Le certificazioni emesse hanno valore ufficiale sono documenti che a norma di le
46. n breve segnale acustico g una volta calcolato il valore di estratto secco totale possibile stampare l etichetta relativa vedere paragrafo successivo e o cancellare dalla memoria il campione di vino mosto appena esaminato premendo CANCEL h per uscire dalla funzione estratto secco totale premere USCITA fino ad avere sul display il MENU PRINCIPALE i al termine della misura scaricare il liquido analizzato tramite il tasto DRAIN j al termine di ogni misura lavare abbondantemente con acqua distillata per pulire accuratamente il cilindro da ogni residuo 5 2 1 Stampa del valore di ESTRATTO SECCO TOTALE Per avere una stampa del valore di ESTRATTO SECCO TOTALE del vino mosto in esame basta premere STAMPA l opzione presente sul display al momento della determinazione Durante la stampa sul display del Super Alcomat compare il messaggio lampeggiante STAMPA DELL ETICHETTA ATTENDI STAMPA OK 2 Vedere per approfondimenti Becchetti Metodi di analisi dei vini e delle bevande spiritose 1999 ed Gibertini elettronica Pag 14 di 39 SUPER ALCOMAT ITALIANO L etichetta cos composta Data 09 03 2001 Ora 10 24 Campione N 29 Il numero identificativo del campione viene scelto in fase di memorizzazione vedere Estratto secco g l 129 2 paragralo precedente Campione N NON viene aggiornato di 1 unit ogni volta che si stampa ed l ultimo valore selezionato in IMEMORIZ Titolo
47. ng the contrast writings become white on black background while NORMAL brings back to the initial contrast black writings on white background By pressing the display background becomes darker while makes it lighter Super Alcomat DISPLAY CONTRAST HW II E L eae TO CORREGT DEFAULT Fig 16 Pag 36 di 39 SUPER ALCOMAT ENGLISH 6 5 Language To change the language used to express all the messages that appear on the display select Settings from the MAIN MENU by pressing 4 Figure 4 then select Language by pressing 6 Identify the desired language Figure 17 among Italian English French Spanish and German by pressing the correspondent key Press 1 to exit Super Alcomat Fig 17 All the messages that appear on the display will be shows in the new language chosen The selection will be maintained also if the instrument is turned off 6 6 SERVICES SUPPLEMENTARY CALIBRATION It must be remembered that the instruments provides automatically after about 10 minutes at rest to start the calibration procedure If the operator senses it necessary he can activate a supplementary calibration Select Services from the MAIN MENU by pressing 5 Figure 4 then select Calibration by pressing 2 The display of the Super Alcomat shows CALIBRATION PROCEDURE ACTIVATED PLS WAIT then the blinking message SELF CALIBRATION ACTIVATED PHASE 1 PLS WAIT
48. nsit relativa Per avere una stampa del valore di TAV del distillato o di densit relativa del liquido in esame basta premere STAMPA l opzione presente sul display al momento della determinazione Se la stampante non collegata il display mostra il messaggio lampeggiante STAMPA DELL ETICHETTA ATTENDI ERRORE DI STAMPANTE PREMERE UN TASTO PER CONTINUARE Se si attiva la stampante senza aver raggiunto un valore stabile di estratto secco compare sul display il messaggio DATO INSTABILE OPERAZ IMPEDITA Durante la stampa sul display del Super Alcomat compare il messaggio lampeggiante STAMPA DELL ETICHETTA ATTENDI STAMPA OK La stampa cos composta nei due casi possibili Data 06 03 2001 Ora 16 07 Pr 012 Data 06 03 2001 Ora 15 20 Pr 001 Titolo alcolom VOL 13 24 Temp distillato 23 7 Temp liquido 24 5 C Densita a 20 C 0 98270 Densita distillato 0 98170 Densita liquido 1 10510 NOTA Sull etichetta stampata compare un numero identificativo indicato con Pr ossia progressivo etichetta Quando lo strumento viene acceso la numerazione parte da 1 Ogni volta che si effettua una stampa la numerazione viene aggiornata La numerazione pu essere modificata vedere paragrafo 6 1 Pag 11 di 39 ITALIANO SUPER ALCOMAT 5 1 4 Azzeramento bilancia Lo strumento offre la possibilit di effettuare un azzeramento automatico nel corso della misura Tenend
49. o corrispondente Premere 1 per uscire Super Alcomat Fig 17 Tutti i messaggi che compaiono sul display saranno visualizzati nella nuova lingua impostata La selezione sar mantenuta anche se lo strumento viene spento 6 6 SERVIZI CALIBRAZIONE SUPPLEMENTARE Va ricordato che lo strumento provvede automaticamente dopo circa 10 minuti di riposo ad avviare la procedura di calibrazione Se l operatore lo ritiene necessario pu attivare una calibrazione supplementare Selezionare Servizi dal MENU PRINCIPALE premendo 5 Fig 4 quindi selezionare Calibrazione premendo 2 Sul display del Super Alcomat compare ATTIVATA PROCEDURA CALIBRAZIONE ATTENDERE quindi il messaggio lampeggiante STRUMENTO IN AUTOCALIBRAZIONE FASE 1 ATTENDERE STRUMENTO IN AUTOCALIBRAZIONE FASE 2 ATTENDERE Durante la prima e la seconda fase di calibrazione non possibile utilizzare lo strumento seguito da Pag 18 di 39 SUPER ALCOMAT ITALIANO 7 TRASMISSIONE SERIALE La trasmissione seriale del SUPER ALCOMAT utilizza lo standard EIA RS232 7 1 COLLEGAMENTO DELL USCITA SERIALE Il collegamento il seguente Segnali H RS 232 Out Pin 2 Pin 3 RS 232 In Pin3 Pin 2 Ground Pin 5 Pin 5 7 2 PROTOCOLLO DI COMUNICAZIONE RS232 FRA SUPER ALCOMAT E PC Mediante il protocollo di comunicazione RS232 possibile ricevere con un personal computer i dati visualizzati d
50. o premuto MEASURE il pescante viene sollevato per qualche secondo permettendo alla bilancia l operazione di azzeramento Super Alcomat offre anche la possibilit di azzerare manualmente la bilancia a se il cilindro pieno premere DRAIN per scaricare il campione e sollevare il pescante b premere il tasto zero sul display appare zero seguito poi dalla videata iniziale c se la sospensione occupata sul display appare ESECUZ AZZERAMENTO IMPOSSIBILE 5 2 ESTRATTO SECCO TOTALE Per estratto secco totale s intende l insieme di tutte le sostanze che in condizioni fisiche determinate non volatilizzano Il metodo ufficiale europeo densimetrico Regolamento CEE N 2676 90 Allegato 4 Estratto secco totale L estratto secco totale viene calcolato indirettamente in base al valore della densit del mosto vino dealcolizzato cio del mosto vino dal quale stato eliminato l alcool e che stato riportato al volume iniziale con acqua distillata L estratto secco viene espresso dalla concentrazione in g l di una soluzione di saccarosio avente la stessa densit del mosto vino dealcolizzato L utilizzo di Super Alcomat consente di determinare il valore dell estratto secco totale di un vino di un mosto per densit comprese tra 0 990 e 1 160 in un range di temperatura compreso tra 15 C e 25 C Sotto i 15 C e sopra i 25 C i risultati non sono completamente affidabili Si procede come segue a verific
51. of the floater with distilled water It is done as follows a verify that the floater is in position on the dedicated prop b fill the cvlinder with distilled water minimum amount 65 ml at a temperature between 15 and 25 pay attention that no bubbles are present on the floater and in the liquid c select Alcoholic Strength by Volume on the main menu by pressing 1 Figure 4 then press COMP to access the Compensation option Figure 7 and press H20 Figure 9 d the display shows the current compensation factor the relative density that water should have at the measured temperature and the relative density of the water actually measured by the instrument Figure 10 Pag 29 di 39 ENGLISH SUPER ALCOMAT e if the two densities agree the instrument is in the ideal reading conditions on the contrary it is necessary to vary the compensation factor of the floater Press or keeping in mind that increases the alcoholic strength read and decreases it Make the two densities agree increases the relative density of 5 units Relative density of the water measured with the Super Alcomat at 19 5 T LIQUID DEMNSITY B 19 5 0 99995 NSPLAY ON Dengity H 1 000065 Relative density of water 26 at 19 5 from tables Go back to the Compensation of the It allows a manual zeroing before decreases the relative ASV screen floater in alcohol the reading density of 5 unit
52. on Through the procedure of Floater compensation both the mass and the volume of the floater in use are not modified The so called compensation factor of the floater is modified instead so that all the floaters used will give the same response in the same standard conditions These conditions are the determination of the relative density of distilled water and the determination of the ASV or alcoholic strength of a hydroalcoholic mixture of known ASV and SIT certified provided on request by our company It is possible to select the two possibilities with the keyboard by pressing or ALCOHOL respectively Pag 28 di 39 SUPER ALCOMAT ENGLISH The display shows the current value of the compensation factor that can be increased or decreased by five units at a time by pressing and the compensation factor remains stored even if the instrument is turned off It is advisable to check regularly this factor 5 1 2 1 Compensation of the floater with the Gibertini SIT certified hydroalcoholic mixture It is done as follows a verify that the floater is in position on the dedicated prop b fill the cylinder with a hydroalcoholic mixture of known strength minimum amount 65 ml at a temperature between 15 and 25 C pay attention that no bubbles are present on the floater and in the liquid c select Alcoholic Strength by Volume from the main menu by pressing 1 Figure 4 access to the option Compensat
53. s Fig 10 Compensation of the floater with distilled water 5 1 3 Printing the value of ASV and of relative density To have a printout of the value of the ASV or of the relative density of the liquid under consideration just press PRINT the option is available on the display at the moment of the determination If the printer is not connected the display shows the following blinking message PRINTING LABEL PLEASES WAIT PRINTER ERROR PRESS ANY KEY TO CONTINUE If the printer is activated before a stable value of dry extract is reached the display shows the blinking message UNSTABLE DATA OPERAT IMPOSSIBLE While printing the display of the Super Alcomat shows the blinking message PRINTING LABEL PLEASES WAIT PRINTING OK The printout is so composed in the two possible cases Date 06 03 2001 Hour 16 07 Pr 012 Date 06 03 2001 Hour 15 20 Pr 001 Alcoholic strength 5 13 24 Temp of distillate 23 7 C Temp of liquid 24 5 C Spec gravity at 20 0 98270 Spec gravity of distillate 0 98170 Spec gravity of liquid 1 10510 NOTE On the label printed an identification number appears indicated with Pr i e progressive label When the instrument is turned on the numbering begins with 1 Every time a printout is made the number is updated It is possible to modify the numbering For details see paragraph 6 1 Modification of label number Pag 30 di 39 SUPER ALC
54. terminazione dell estratto secco premere CALCOLO strumento rimane in attesa di poter misurare la densit del distillato ottenuto dal vino mosto in esame campione valido con densit relativa d2020 tra 0 79000 e 0 99995 con il messaggio lampeggiante Fig 13 Pag 13 di 39 ITALIANO SUPER ALCOMAT Super Alcomat POSIZIONARE IL CAMPIONE DISTILLATO PER WISURA DENBITA Densita Letta a 18 5 C 0 9949 Densit relativa del vino mosto appena letta altrimenti compare 000 Attendere lo zero o fare azzeramento manuale prima della lettura del distillato Fig 13 f premere MEASURE sul display vengono visualizzati temperatura densit letta e Titolo Alcolometrico Volumico del distillato insieme al valore di estratto secco totale in g l di saccarosio Fig 14 Super Alcomat provvede automaticamente a determinare il TAV del distillato utilizzando l algoritmo di correzione valido per le miscele idroalcoliche e successivamente la densit relativa d2020 dello stesso Super Alcomat ESTRATTO SECCO oj 19 20 g l Fig 14 0 98470 ki 2 Super Alcomat applica la formula di Tabari per determinare la densit relativa d20 20 del vino mosto dealcolizzato quindi riporta il valore tabellare corrispondente di estratto secco totale AI raggiungimento del valore stabile il valore di estratto secco totale viene fissato sul display con caratteri bianchi su sfondo nero preceduto da u
55. the wine must as it is valid samples have specific gravity between 0 99 and 1 16 with the blinking message Super Alcomat BAMPLE OF O WIME MUST FOR DENSITY HEASURE DISPLAY ON Density Asadout g 5 DOD O O fn Fig 11 c wait for the zero of the balance or press ZERO for manual zeroing and press MEASURE d the display shows temperature specific gravity read specific gravity correct at 20 C and density at 20 C of the wine must sample under consideration Figure 12 Pag 31 di 39 ENGLISH SUPER ALCOMAT Super Alcomat Density of the wine must as it is To recall the sample of wine must To go directly to the calculation of the To store the sample of wine must DRY EXTRACT Fig 12 Super Alcomat relates automatically the specific gravity value read to 20 C Once a stable value is reached the specific gravity value correct at 20 C expressed with four decimal numerals remains fixed on the display with white fonts on black background preceded by an acoustic signal The density is obtained by multiplying the specific gravity value correct at 20 C for 0 998203 g ml density of water at 20 C IT IS NOW POSSIBLE TO STORE UP TO 99 SAMPLES OF WINE MUST This function allows performing all the measurements on the samples of wine must in advance to recall them afterwards at the moment of measuring the specific gravity of the correspondent distillat
56. to da un led lampeggiante e da un segnale acustico Terminato lo svuotamento l elettrovalvola si chiude in modo automatico 4 ACCENSIONE SUPER ALCOMAT Alf accensione del Super Alcomat sul display Fig 3 appare il numero di release del software preceduto da un segnale acustico e dal nome della societ Super GIBERTINI DENSI AL I Fig 3 Dopo circa due secondi appare il MENU PRINCIPALE TIT ALG VOL GR ALCOLICO ESTRATTO SECCO TOTALE VALUTAZIONE MOSTI IMPOSTAZIONI SERVIZI Fig 4 Premendo i tasti funzione 4 o 5 possibile accedere ad Impostazioni o Servizi vedere il paragrafo n 6 premendo invece i tasti funzione 1 o 2 o 3 il display mostra il seguente messaggio lampeggiante Pag 6 di 39 SUPER ALCOMAT ITALIANO Super Alcomat BILANGIA IN FASE DI STABILIZZAZ TERMICA ATTENDERE Fig 5 Qualsiasi tasto venga premuto in questa situazione Fig 5 riporta al MENU PRINCIPALE ossia il SUPER ALCOMAT in fase di stabilizzazione termica inutilizzabile per compiere misure mentre la tastiera rimane attiva Il messaggio lampeggiante se non si tocca la tastiera rimane sul display fino a quando il Super Alcomat in fase di stabilizzazione termica circa 5 minuti AI termine della fase di stabilizzazione termica compare sul display del Super Alcomat il messaggio lampeggiante STRUMENTO IN AUTOCALIBRAZIONE FASE 1 ATTENDERE STRUMENTO IN AUTOC
57. tor the Super Alcomat indicated Fig 2 with nr 4 Insert then the plug in the network socket 1 Level bubble 2 Printer connector 3 RS232 connector 4 Power socket 5 Serial number 6 Service connector Fig 2 Rear view Pag 23 di 39 ENGLISH SUPER ALCOMAT 3 DESCRIPTION OF SUPER ALCOMAT The instrument is constituted by a hydrostatic balance with sensibility to the fifth decimal numeral and an elaborating module integrated in the balance itself Fig 3 7 Floater prop 13 Knob of the cylinder centering device 8 Drainage pipe 14 Cylinder centering device 9 Cylinder and floater 15 Knob for blocking balance 10 Electrovalve to transport 11 Command panel Display 16 Thermo probe 12 Supports for level regulation 17 Tank for cylinder filling minimum 65 ml 3 1 COMMAND PANEL The keyboard of the Super Alcomat Fig 3 nr 11 is equipped with a display and eight functional keys to which a different meaning depending on the displayed menu is associated TITO ALCOLOM OL EEQMOLJAZBJO TE 0 98710 Pag 24 di 39 SUPER ALCOMAT ENGLISH Every key must be pressed until an acoustic signal is uttered This indicates that the instrument accepted the command NOTE After some minutes when the Super Alcomat is not used the display shuts off automatically By pressing ON the display turns on again The keys allow the execution of the displayed action The

Download Pdf Manuals

image

Related Search

Related Contents

Cisco MX200/MX300 User Guide  Targus APTA19  Notes for Users Bemærkninger til brugere Note per gli utenti  Référentiel des métiers - Fédération des Conservatoires d`espaces  Mpman FRA102  Samsung SGH-E100 User Manual  厚 生 労 働 省 告 示 第 二 百 九 十 四 号 ( ) ( 薬 事 法 昭 和  StarTech.com 2 Port Quad Monitor Dual-Link DVI USB KVM Switch with Audio & Hub  Caisse Nationale de l`Assurance Maladie  5201 IP Benutzerhandbuch  