1. MEM Indicates the number in the meter display is a stored test value see page 39 ALARM CABLE Indicates the meter is in data transfer mode to PC This function is not available yet Indicates alarm has been set see page 16 CTL This icon indicates a control solution test see page 25 or the stored value is a control solution result see page 39 Displays the date time am pm in 24 hours format see page 12 or type of day averages 7 14 or 30 see page 39 Understanding Your New Blood Glucose System EvenCareG2 BEEPER a Set sound volume see page 18 TEMPERATURE Appears when it is either too hot or too cold to test outside the ranges of 50 F 104 F CLOCK Indicates the meter is ready to set up time and date see page 13 VOICE Indicates the voice feature is on see page 18 ___ UNITS Unit of measurement for your blood glucose is mg dL BLOOD DROP and ARROW Indicate the meter is ready for testing see page 34 EvenCareG2 Understanding Your New Blood Glucose System The EvenCareG2 Blood Glucose Test Strip il Ey EvenCare amp 2 E Contact Points Sampling End Insert this end to the test Apply blood or control strip port on the meter solution here The EvenCare G2 Test Strips are glucose specific biosensor based test strips that require very little blood sample to quickly test glucose in fresh capillary whole blood The test results are plasma calibra
2. Reference 1 American Diabetes Association Standards of Medical Care in Diabetes 2008 Diabetes Care 2008 31 512 S54 38 Viewing Your Meter s Memory EvenCareG2 Viewing Your Meter s Memory Your EvenCare G2 Meter automatically stores up to 300 test and control results with date and time The meter also provides 7 14 and 30 day averaging to help track your blood glucose trend You must set up the date and time on your meter properly to use the memory and day averaging function see Setting Up Your New System on page 12 The meter will not memorize test or control results if the date and time are not set Viewing Stored Readings and 7 14 30 Days Average from Memory You will need y Your EvenCare G2 Meter Step 1 Press and hold Enter to turn on the meter Step 2 Press Enter to enter setup screen Step 3 Press AorY until MEM starts to flash If the voice function is on the meter will say memory recall Press Enter to enter 7s MEM will appear at the upper left corner of the mem display screen together with your most recent 2 Q testing result Press A to see the next result a and the next and so on Holding A will flash LM d Lt through the results ms Laia ro Po 2 EvenCareG2 Viewing Your Meter s Memory END will appear on the display screen after the last recorded test result if less then 300 readings are stored You can press W to go back viewing from the old
3. Causas posibles Qu puede hacer e Soluci n de Control usada e Aseg rese de que est de marca err nea utilizando Soluci n de Control EvenCare G2 e Soluci n de control no est a e Aseg rese de que el entorno de temperatura ambiente prueba est entre 50 F a 104 F e Soluci n de control caducada e Compruebe la fecha de caduci contaminada o da ada tira dad y apertura de la botella la reactiva soluci n de control y en las tiras reactivas Repetir prueba uti lizando una nueva tira reactiva Si el resultado sigue estando fuera de rango use una nueva botella de soluci n de control y vuelva a probar e Mal funcionamiento del Medidor Si el problema persiste por favor llame al Centro de Servicio T c nico Medline al 1 866 274 2131 De lunes a viernes entre las 8 00 am y 5 00 pm Hora Centro Co mun quese con su Proveedor de Antenci n de Salud en cualquier otro momento 79 Probando su Glucemia Table of Contents EvenCareG2 Probando su Glucemia Necesitara Y Su Medidor EvenCare G2 V Una nueva Tira Reactiva de Glucemia EvenCare G2 Y Dispositivo de punci n con una lanceta limpia sin uso o lancetas de seguridad Tapa transparente para Prueba de Sitio Alterno AST en palma y antebrazo Antes de empezar aseg rese de e Configurar su medidor correctamente y realizar una prueba de con trol Ver Configuraci n de su nuevo sistema en la p gina 60 y Prueba de Soluci n de Control en la p gina 7
4. Press AorY until starts to flash If the voice function is on the meter will say Set testing Alarm Press Enter to enter alarm setting Step 4 The display screen will indicate alarm 1 status the default is OFF You can turn it on or off by pressing Aor 16 Setting Up Your New System EvenCareG2 Step 5 Press Enter and the hour will flash Press AorY to select the hour Press Enter to confirm your choice and advance to set minutes Step 6 The current minutes will flash Press AorW to select the correct minutes Press Enter to confirm your choice and advance to set alarm 2 Step 7 Set alarm 2 and 3 following steps 4 through 6 After you have finished setting up alarm 3 press Enter to exit alarm setup NOTE e Alarm will not sound during testing e Alarm will follow the time format 12 hour or 24 hour you choose when setting up the clock e Alarm settings will not be erased when changing batteries Begin testing by inserting a EvenCare G2 Blood Glucose Test Strip see page 30 Testing Your Blood Glucose or press and hold Enter to turn off the meter or see the next section on how to set up the voice feature on your meter 17 EvenCareG2 setting Up Your New System Voice and Beeper Function Your meter has a voice function that can be used as a verbal aid for using the meter and hearing test results The default setting for the voice function is on
5. Slide the battery cover back into the meter firmly NOTE e Meter will prompt you to reconfirm the date and time when turned on See page 13 for Setting the Clock e The date stored results alarm settings volume setting and language choice stay in memory when the batteries are changed 42 Table of Contents EvenCareG2 43 EvenCareG2 Table of Contents 44 EvenCareG2 Table of Contents Caring for the Meter Cleaning the Meter e Ifthe meter gets dirty use a moist NOT WET lint free cloth dampened with a mild detergent e Do not get water inside the meter or hold it under running water e Do not use glass or household cleaners on the meter e Do not try to clean the test strip holder Storage and Precautions e Handle the Meter with care severe shock such as dropping the meter could damage the electronics e The meter and test strips are designed to be used within the temperature ranges between 50 F to 104 F e Avoid leaving the meter in extremely hot or cold place such as near a heat source or in an extremely hot or cold car e Do not store or use the meter or test strips in high humidity levels such as bathroom or kitchen e Always close bottle cap immediately after removing a test strip and make sure it s closed tightly e Do not take the meter apart as doing so will void the warranty If there are technical problems or questions please call Medline s Technical Service Center at 1 866 274 2131 Monda
6. se anular la garant a Si hay problemas t cnicos o preguntas por favor llame al Centro de Servicio T cnico Medline al 1 866 274 2131 de lunes a viernes entre las 8 00 am y 5 00 pm Hora del Centro P ngase en contacto con su Proveedor de atenci n m dica en cualquier otro momento 93 Soluci n de Problemas Soluci n de Problemas Esta secci n detalla los mensajes de la pantalla y c digos de error importantes que se encuentran al utilizar su medidor EvenCare G2 y tiras reactivas EvenCare G2 Lo que ve Z L mre Z a m rr a m L ret Lo que el medidor dice Error 1 Fallo Medidor Error 3 Tira reactiva usada Error 4 muestra suficiente Lo que significa El medidor electr nico o Tira reactiva usada O da ada Used or damaged test strip Tira reactiva usada o da ada No suficiente muestra en la tira reactiva para em pezar Table of Contents EvenCareG2 Qu debe hacer e Retire la bater a y encienda de nuevo el medidor e Retire la tira reactiva e inserte una nueva tira reactiva e Si el problema persiste por favor llame al 1 866 274 2131 De lunes a viernes 8 00 a m 5 00 p m CST Comun quese con su Proveedor de Antenci n de Salud en cualquier otro momento e Repita la prueba con una nueva tira reactiva Recuerde a adir la muestra s lo cuando vea el indicador en la pantalla o el medidor indique verbalmen
7. hueso Presione el dispositivo de punci n firmemente contra el sitio de prueba y presione el bot n de liberaci n no levante el dispositivo de punci n Para sitios de prueba en el antebrazo aplique y libere presi n bombeo por unos pocos segundos mientras mantiene el dispositivo de punci n en constante contacto con la piel La prueba de Palma en la base del pulgar no requiere de bombeo Cuando la sangre es de tama o aproximada mente levante el dispositivo de punci n hacia arriba sin que se corra la sangre NOTA e Frotar el antebrazo hasta que se caliente antes de punci n ayudar a minimizar la diferencia con los resultados del examen de yema del dedos La palma en la base del pulgar no requiere ser frotada antes de la prueba e Silos resultados de las pruebas en sitios alternos no est n de acuerdo con c mo se siente use prueba de yema de dedos 86 EvenCareG2 Table of Contents Paso 4 Pase suavemente la tira reactiva y toque la gota de sangre en un ligero ngulo La tira reactiva act a como una paja para extraer la sangre Mantenga la tira reactiva en la gota de sangre hasta que el medidor emita un pitido para indicar que la la tira reactiva tiene suficiente sangre para la prueba Paso 5 La pantalla empezar la cuenta atr s Despu s 6 segundos el resultado de su prueba de glucosa aparecer en la pantalla del medidor y se anun ciar verbalmente si usted tiene la funci n de
8. la prueba de glucemia no reflejan c mo se siente e Desea practicar el procedimiento de prueba IMPORTANTE e Utilice s lo soluciones de control de glucosa EvenCare G2 de con las tiras reactivas EvenCare G2 Otras marcas de soluci n de control producir n resultados inexactos e Siempre revise la fecha de vencimiento NO use soluciones de control si han caducado e NO CONGELAR Almacene las soluciones de control a temperatura ambiente Necesitar V Soluci n de Control EvenCare G2 baja y alta Y Su Medidor EvenCare G2 Y Una nueva Tira Reactiva de Glucemia EvenCare G2 Paso 1 Saque una Tira Reactiva EvenCare G2 del frasco de tiras reactivas y cierre el frasco inmediatamente Inserte la tira reactiva para encender el medidor 75 Prueba de Soluci n de ControTable of Contents EvenCareG2 IMPORTANTE e Verifique la fecha de caducidad impresa en el frasco de tiras reactivas y su paquete No utilice las tiras reactivas caducadas e Use cada tira reactiva inmediatamente despu s de sacarla de la botella e No utilice tiras reactivas mojadas o da adas e Mant ngase alejado de la luz solar directa y el calor Almacene e frasco de tiras reactivas en un lugar seco y fresco e Registre la fecha de apertura en la etiqueta del frasco Deseche la botella y cualquier tira reactiva restante despu s de 6 meses desde la fecha de apertura e Prueba bajo el rango de temperatura de 50 F a 104 C Su medidor no funcionar f
9. medidor S lo tiene que utilizar los botones AoY para navegar hasta el icono de la funci n deseada y pulsar el bot n Intro para entrar en la funci n cuando el icono parpadea Navegue hasta el Final en la pantalla para salir de la Pantalla de Configuraci n El Medidor tambi n anunciar cada funci n cuando la funci n de voz est activada mem Q E lt LD ae a ota Pulse Intro para entrar Recuperar Memoria para recuperar los resultados almacenados Su medidor puede memorizar hasta 300 memorias con fecha y hora Ver p gina 87 Bonnet Pulse Intro para entrar en el modo PC Link para la transferencia de datos en un PC Esta funci n no est disponible todav a o Pulse Intro para entrar en Configuraci n de hora y fecha ver p gina 78 dao Pulse Intro para entrar en Configuraci n de Alarma ver p gina 61 Su medidor puede tener hasta tres ajustes de alarma 62 EvenCareG2 Table of Cofitenfigjuraci n de su Nuevo Sistema Q E Pulse Intro para ingresar a la configuraci n de volumen ver p gina 66 Q ET Pulse Intro para entrar a selecci n de idioma ver p gina 67 FIN sets Salir de la pantalla de configuraci n y volver a prueba Ajuste del Reloj Necesitar Y Su Medidor EvenCare G2 NOTA Si la fecha y la hora en el medidor EvenCare G2 no se ha establecido el resultado de prueba o control no se almacenar en la memoria Paso 1 Mantenga pulsada la tecla I
10. mg dL GOTA DE SANGRE y FLECHA Indica que el medidor est listo para la prueba ver p gina 34 59 Entendiento los Resultados ddisbl afdbantentiucd KenCareG2 La tira reactiva de Glucemia EvenCareG2 2 Ey EvenCares2 EE Puntos de Contacto Extremo de Muestreo Inserte este extremo al Aplique sangre o puerto de tira reactiva en soluci n de control aqu el medidor Las tiras reactivas EvenCare G2 son tiras de prueba espec ficas de glu cosa basadas en biosensor que requieren muy peque a muestra de san gre para probar r pidamente la glucosa en sangre entera capilar fresca Los resultados del ensayo son calibrados por plasma para facilitar la com paraci n con los resultados de laboratorio sin embargo los resultados pueden ser de 10 a 15 superiores a los resultados obtenidos a partir de ensayos de glucemia referenciada de sangre entera La tira reactiva est tambi n equipada con detecci n de subllenado para que le avise cuando no hay suficiente sangre para realizar una prueba por lo que puede estar seguro de que cada lectura que se obtiene es un resultado preciso y significativo IMPORTANTE e Cierre inmediatamente con fuerza el tap n del vial de la botella de tiras reactivas despu s de cada uso Mantenga las tiras reactivas no utilizadas en su envase original e NO deje tiras de prueba fuera de la botella mientras no est en uso e NO reutilizar las tiras reactivas Las tiras reactivas son para
11. nuevo para ver el promedio de 14 d as y pulse W de nuevo para ver el promedio de 30 d as El n mero de resultados incluidos en los promedios se muestra al lado del icono Los resultados de soluci n de control no se incluyen en el c lculos de promedio diario 90 EvenCareG2 Table of Contefiemdo la Memoria de su Medidor NOTA e Los resultados de las pruebas de control se marcan por ctl en la esquina inferior izquierda de la pantalla adem s de MEM e Cuando la memoria del medidor est en blanco la pantalla mostrar nil e Cuando no se dispone de datos de promedio diario la pantalla mostrar cuatro guiones e Si la memoria del medidor est llena la adici n de un nuevo resultado de prueba har que el m s antiguo sea borrado e Resultados de pruebas memorizadas y promedios diarios no se bor rar n al cambiar las bater as e Puede empezar a probar en cualquier momento mediante la inserci n de una tira reactiva 91 Viendo la Memoria de su Mediile of Contents EvenCareG2 Cuidado del Medidor El cuidado de su medidor EvenCare G2 es f cil Siga estas simples directrices para mantener su medidor funcionando correctamente Cambio de Bater a Las bater as deben ser cambiadas cuando el icono aparece en la pantalla del medidor Materiales que necesitar Dos bater as alcalinas AAA V Su Medidor EvenCare G2 Paso 1 AverigPresione firmemente sobre la tapa de la bateria y deslicela en la direccion de la
12. original del medidor 97 Especificaciones Especificaciones Tiras Reactivas Rango de Prueba Calibraci n Tiempo de Prueba Tama o de muestra Rango de Hematocrito Tipo de Pantalla Memoria Resultado promedio Dimensi n Peso Bater a Duraci n de la bater a Apagado Autom tico Temperatura de funcionamiento Humedad relativa Condiciones de almacenamiento Table of Contents Tiras reactivas de glucemia EvenCare G2 20 600 mg dL Plasma 6 segundos Mayor que 1uL 30 55 Pantalla LCD grande 300 resultados de an lisis de sangre con fecha y hora 7 14 y 30 d as promedio 93L x 58W x 20 5H mm 79g 2 baterias alcalinas AAA 1000 pruebas de uso continuo o 1 a o Despu s de 2 minutos de inactividad 50 F a 104 F Menos del 85 Medidor a 39 F 122 F Tiras Reactivas at 39 F 86 F Para obtener informacion adicional consulte el inserto Tira Reactiva EvenCare G2 98 EvenCareG2
13. 3 para m s detalles e L vese las manos y el sitio de prueba a fondo con jab n y agua tibia y seque bien e Est haciendo la prueba en un rea comprendida entre los rangos de temperatura de 50 F 104 F Su medidor no har pruebas fuera de este rango y el icono de term metro de la pantalla con indicaci n verbal si la funci n de voz est activada Mueva el medidor a un rea que est entre 50 F 104 F y deje que repose durante 10 a 15 minutos antes probar de nuevo C mo preparar su Dispositivo de Punci n para Prueba Casera Paso 1 Desenrosque la tapa ajustable del dispositivo dwe punci n e inserte la lanceta empujando hacia abajo firmemente hasta que est comple tamente asentada Paso 2 Gire la cubierta protectora de la lanceta Gire la tapa ajustable hacia la derecha de vuelta sobre el dispositivo de punci n EvenCareG2 Table of Contents Probando su Glucemia NOTE e Para la prueba de yema de los dedos use la tapa ajustable opaca e Para pruebas en sitios alternos en palma y antebrazo use la tapa ajustable transparente Paso 3 Ajuste la configuraci n de punci n en la tapa ajustable para el nivel de profundidad de punci n 1 es el m s ligero y 5 es el m s profundo Paso 4 Con una mano sosteniendo la tapa ajustable tire del dispositivo de punci n con la otra mano hasta que escuche un sonido de clic El dispos itivo de punci n ya est listo para tomar mues tras de sangre I
14. AorW to select the correct hour Press Enter to con firm your choice and advance to set minutes 14 Setting Up Your New System EvenCareG2 Step 8 Set Minute The current minutes will flash If the voice function is on the meter will say Set Minutes Press Aor Y to select the correct minutes Press Enter to confirm your choice and finish setting the clock NOTE e Anytime during set up inserting a EvenCare G2 Test Strip will begin testing Any changes made so far are stored e You may need to reset time after changing batteries After date and time has been set the meter will announce the date and time if the voice feature is on Begin testing by inserting an EvenCare G2 Blood Glucose Test Strip see page 30 Testing Your Blood Glucose or press and hold Enter amp to turn off the meter or see the next section on how to set up the alarm feature on your meter 15 EvenCareG2 setting Up Your New System Setting the Alarms You can set up to three alarms on the EvenCare G2 Meter You must set the clock before you can set the alarms When an alarm sounds the meter will turn on and a melody will sound for 30 seconds Pressing Enter or inserting a test strip will silence the alarm You can set up one or all three alarms You will need Y Your EvenCare G2 Meter Step 1 Press and hold Enter to turn on the meter Step 2 Press Enter to enter setup screen Step 3
15. Dear EvenCare G2 Owner EvenCareG2 Dear EvenCare G2 Owner Thank you for choosing the EvenCare G2 Blood Glucose Monitoring System to help you in monitoring your diabetes We designed the EvenCare G2 Blood Glucose Monitoring System to be accurate and easy to use The added convenience of the voice feature can be used as a verbal aid for using the meter and hearing test results by visually impaired users This manual contains all of the information needed to use and to maintain your new blood glucose meter Please read it carefully before use If you need assistance please call Medline s Technical Service Center at 1 866 274 2131 Monday through Friday between 8 00 a m and 5 00 p m Central Time Contact your Healthcare Provider at all other times EvenCareG2 Dear EvenCare G2 Owner Intended Use The EvenCare G2 Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips forearm or palm Testing is done outside the body In Vitro diagnostic use The meter includes voice functionality to assist visually impaired users It is intended for both lay use by people with diabetes and in a clinical setting by healthcare professionals as an aid to monitoring levels in Diabetes Mellitus It is not intended for the diagnosis or screening of diabetes or for neonatal use IMPORTANT e Alternate Site Testing AST with the EvenCare G2 Blood Glucose Mo
16. MPORTANTE Utilice una lanceta nueva est ril cada vez que haga la prueba para evitar la contaminaci n cruzada Si se utilizan toallitas con alcohol para limpiar los dedos aseg rese de que el rea est seca antes de obtener la muestra de sangre 81 Probando su Glucemia Table of Contents EvenCareG2 Informacion Importante sobre Pruebas en Sitios Alternos AST El Sistema de Monitoreo de glucemia EvenCare G2 puede hacer pruebas de glucosa desde reas distintas a la yema de los dedos como la palma y el antebrazo Pruebas en sitios alter nos o AST La Prueba de sitio al terno puede ser menos dolorosa que la prueba en la yema del dedo pero debido a las diferencias fisiol gicas entre la yema del dedo y palma y an tebrazo1 los resultados de prueba de sitio alterno pueden ser significativamente diferentes a los resultados de las pruebas de la yema del dedo en determinadas condiciones Usted debe consultar con su m dico o profesional de la salud antes de usar pruebas en sitios alternos HACER AST NICAMENTE en los intervalos siguientes e Antes de una comida o en ayunas m s de 2 horas desde la ltima comida e 2 horas o m s despu s de tomar insulina e 2 horas o mas despu s del ejercicio Alternate Site Testing SHOULD NOT be used when e Usted no est consciente de hipoglucemia no puede decir si tiene baja az car en la sangre e Dentro de las 2 horas de una comida ejercicio o medicamentos e Usted va a opera
17. a actual parpadear Si la funci n de voz est activada el indicador va a decir Ajustar Hora presione AoY to select the correct hour para seleccionar la hora correcta Pulse Intro para confirmar su elecci n y avanzar a ajustar los minutos EvenCareG2 EvenCareG2 Table of Cofitenfigjuraci n de su Nuevo Sistema Paso 8 Ajuste de minuto Los minutos actuales parpadear n Si la funci n de voz est activada el medidor va a decir Ajus tar Minutos presione AoY para seleccionar los minutos correctos Pulse Intro para confirmar su elecci n y terminar de ajustar el reloj NOTA e En cualquier momento durante la configuraci n insertar una tira reactiva EvenCare G2 iniciar la prueba Cualquier cambio realizado hasta el momento se almacena e Es posible que deba reiniciar la hora despu s de cambiar las bater as Despu s de configurar fecha y hora el medidor anunciar la fecha y hora si la caracter stica de voz est activada Comenzar la prueba mediante la inserci n de una Tira reactiva de glucemia EvenCare G2 vea p gina 78 Probando su glucemia o mantenga pulsada la tecla Intro para apagar el medidor o vea la siguiente secci n sobre c mo configurar la funci n de alarma en su medidor 65 Configuraci n de su Nuevo Sistesteof Contents EvenCareG2 Configurando las Alarmas Puede configurar hasta tres alarmas en el medidor EvenCare G2 Debe ajustar el reloj antes de po
18. a el diagn stico o detecci n de la diabetes o para uso neonatal IMPORTANTE e Prueba de Sitio Alterno AST con el Sistema de Monitoreo de Glucemia EvenCare G2 s lo puede utilizarse durante condiciones de glucemia en fase estable e Deshidrataci n La deshidrataci n severa puede llevar a resultados inexactos de pruebas de glucemia Si sospecha que est gravemente deshidratado contacte con su profesional de la salud inmediatamente e Rango de Hematocrito Un rango de hematocrito que es superior al 55 o inferior al 30 puede causar resultados inexactos de prueba de glucemia 52 ndice Entendiendo Su Nuevo Sistema de Glucemia EvenCareG2 Table of Contents ndice EvenCareG2 El Medidor EvenCareG2 6 54 Pantalla de Medidor 8 56 Tira reactiva EvenCareG2 10 58 Soluci n de Control de Glucosa EvenCareG2 recomendado 59 Configuraci n de su Nuevo Sistema La Pantalla de Configuraci n 60 Ajuste del Reloj 61 Ajuste de Alarmas 64 Funci n Voz y Beeper 66 Selecci n de Idioma 67 Tabla de Funciones de Voz 68 Prueba de Soluci n de Control Realizaci n de una prueba con soluci n de control 73 Soluci n de problemas de Soluci n de Control 77 Testing Your Blood Glucose C mo preparar su Dispositivo de Punci n 78 Informaci n importante sobre Pruebas en Sitios Alternos AST 80 Realizaci n de una prueba de glucemia 82 Entendiendo los Resultados de su Prueba de Glucemia 86 Viendo
19. ado Propietario EvenCare G2 EvenCareG2 Estimado Propietaried tiven Gare 62 Gracias por elegir el Sistema de Monitoreo de Glucemia EvenCare G2 ara ayudarle en el control de su diabetes Hemos dise ado el Maran ai ren Mec a Ser ENESE f cil de usar La conveniencia de la funci n de voz se puede utilizar como una ayuda verbal para el uso del medidor y escuchar los resulta dos del examen para usuarios con deficiencia visual Este manual con tiene toda la informaci n necesaria para usar y dar mantenimiento a su nuevo medidor de glucemia Por favor lea con cuidado antes de usar Si necesita ayuda llame al Centro de Servicio T cnico de Medline al 1 866 274 2131 de lunes a viernes entre las 8 00 am y las 5 00 pm Hora del Centro Comun quese con su proveedor de atenci n m dica todas los otras veces 51 49 Estimado Propietario EvenCarkBl2 of Contents EvenCareG2 Uso Previsto El Sistema de Monitoreo de Glucemia EvenCare G2 est dise ado para la medici n cuantitativa de glucemia en muestras de sangre en tera capilar fresca tomadas de yema de los dedos el antebrazo o la palma Las pruebas se hacen fuera del cuerpo uso diagn stico in vitro El medidor incluye funcionalidad de vozpara ayudar a los usuar ios con discapacidad visual Est destinado para uso lego por personas con diabetes y en un entorno cl nico de profesionales de salud como una ayuda para controlar los niveles en la Diabetes Mellitus No est dise ado par
20. and can be switched off to conserve battery power When the voice function is off the meter will beep to alert you to important messages You will need Your EvenCare G2 Meter Setting Volume Step 1 Press and hold Enter to turn on the meter Step 2 Press Enter to go into setup screen Press AorY until 4 starts to flash Press Enter to enter volume setting Step3 The display screen will show the current sound volume default is 2 Press AorY to adjust the sound volume from 1 lowest to 5 highest A melody will sound according to the sound volume you have selected Setting to O will turn off the voice function The meter will instead beep to alert you to important messages Once you have determined the suitable volume press Enter again to exit volume setting 18 Setting Up Your New System EvenCareG2 Language Selection Step 1 Press and hold Enter to turn on the meter Step 2 Press Enter to enter setup screen Step 3 Press AorY until E starts to flash The meter will announce Language Selection Step 4 The default language 1 on the display screen is announced Press to change the language Press AorY to select another language Press when you have finished selecting the language and exit language selection English and Spanish are the two options NOTE e Any time during set up you can insert an EvenCare G2 Test Strip and b
21. asco de tiras reactivas en un lugar seco y fresco Registre la fecha de apertura en la etiqueta del frasco Deseche e frasco y cualquier tira reactiva restante despu s de 6 meses desde la fecha de apertura Debe ver un icono de gota de sangre intermitente si la tira reactiva ha sido insertada correctamente Paso 2 Espere a que una gota de sangre intermitente y un icono de flecha aparezcan en la pantalla del medidor El medidor tambi n anunciar Por favor aplique sangre en la tira reactiva si tiene la funci n de voz activada Paso 3 Obtener muestra de sangre 84 EvenCareG2 Table of Contents Probando su Glucemia Para la prueba de yema del dedo Mantenga el dispositivo de punci n use la tapa opaca ajustable o lanceta de seguridad contra el lado de la yema del dedo y presione el bot n de liberaci n para crear una punci n CONSEJO e Frote suavemente su mano y el dedo hacia el lugar de la punci n para formar una gota de sangre No orde e ni apriete alrededor del sitio de punci n e Pinche el lado de la yema del dedo para evitar hinchaz n Para evitar los callos elija un sitio de punci n diferente cada vez 85 Probando su Glucemia Table of Contents EvenCareG2 Para las pruebas de Palma y Antebrazo Mantenga el dispositivo de punci n use la tapa ajustable clara contra el sitio de prueba base de palma o PA antebrazo Evitar un sitio de prueba 4 que tenga venas lunares pelo excesivo o
22. d time month date hour minutes Setting alarm 1 2 or 3 Set Alarm number When sending data to PC Sending data to PC This function is not available yet Finish sending data to PC Data transmission complete This function is not available yet 23 EvenCareG2 voice Function Table Other warning messages see page 44 Solving Problems WHEN the meter will speak WHAT the meter will say Meter date and time Please set date and time has not been set Meter failure Error 1 Meter Failure Used test strip Error 3 used test strip Not enough blood or Error 4 not enough sample control solution 24 Control Solution Testing EvenCareG2 Control Solution Testing The purpose of the control solution testing is to make sure the Even Care G2 Meter and the EvenCare G2 Test Strip are working properly You should perform control solution testing when e Using the meter for the first time e Using a new bottle of EvenCare G2 Blood Glucose Test Strips e You left the test strip bottle cap open for awhile e You dropped the meter e You suspect your meter and test strips are not working properly e The blood glucose test results do not reflect how you feel e You want to practice the testing procedure IMPORTANT e Use only the EvenCare G2 Glucose Control Solutions with the EvenCare G2 Test Strips Other brands of control solution will produce inaccurate results e Always check the
23. der ajustar las alarmas Cuando suena una alarma el medidor se enciende y una melodia sonara durante 30 segundos Pulsar Intro o insertar una tira reactiva silenciara la alarma Necesitar Y Su Medidor EvenCare G2 Paso 1 Mantenga pulsada la tecla Intro para en cender el medidor Paso 2 Pulse Intro para acceder a la pantalla de configuraci n Paso 3 Presione AoW hasta que comience a parpadear Si la funci n de voz est activada el medidor va a decir Configurar Alarma de prueba Pulse Intro para ingresar el ajuste de la alarma Paso 4 La pantalla indicar el estado de alarma 1 El valor predeterminado es OFF Puede activarla o desactivarla presionando AoW EvenCareG2 Table of COntefidsiraci n de su Nuevo SistemaT Paso 5 Pulse Intro y la hora parpadear Presione A OY para seleccionar la hora Pulse Intro para confirmar su elecci n y avanzar a ajustar los minutos Paso 6 Los minutos actuales parpadear n Presione A O Y para seleccionar los minutos correctos Pulse Intro para confirma su elecci n y avanza al ajuste de la alarma 2 Paso 7 Configurar la alarma 2 y 3 siguiendo los pasos 4 a 6 Despu s de haber terminado la configuraci n de la alarma 3 pulse Intro para salir de la configuraci n de alarma NOTA e La alarma no sonar durante la prueba e La alarma seguir el formato de hora 12 horas o 24 horas que usted elija al config
24. egin testing Any changes made so far are stored e Your choice of language and volume settings will not be erased when changing batteries Begin testing by inserting an EvenCare G2 Blood Glucose Test Strip see page 30 Testing Your Blood Glucose or press and hold Enter to turn off the meter 19 EvenCareG2 voice Function Table Voice Function Table The voice function can be used as an aid for you to hear the test results and use the meter The following table lists WHEN the meter will speak and WHAT the meter will say Testing blood and control solutions see page 25 Control Solution Test ing and page 30 Testing your blood glucose WHEN the meter will speak Turned on Prompt user to insert test strip Test strip is inserted and the meter is ready to test with flash ing arrow and blood drop Blood Glucose test is completed and the result is shown on screen Blood Glucose test result is BELOW the measurement range of 20 mg dL Blood Glucose test result is ABOVE the measurement range of 600 mg dL 20 WHAT the meter will say Greeting Tone will sound Please insert test strip Blood glucose test Please apply blood to the test strip Your blood glucose reading is number milligram per deciliter Your blood glucose reading is below 20 mg dL Your blood glucose reading is above 600 mg dL Voice Function Table EvenCareG2 WHEN the meter wil
25. est strip icon and L1 will appear on the meter display screen The meter will also announce L1 control solution test please apply control solution onto the test strip If you decide not to do a control solution test Press A or Wuntil ctl disappears from the meter display 26 Control Solution Testing EvenCareG2 Step 4 Squeeze a drop of control solution onto a clean dry non absorbent surface Do not apply control solution to the test strip directly from the bottle Replace the bottle cap on the control solution bottle immediately after use DO NOT touch the Sampling End of the test strip with the bottle of control solution See illustration of Sampling End below a at EvenCareS2 Wil sampling Ena Step 5 Gently touch the drop of the control solution to the tip of the test strip The meter will beep indicating it has received the control solution Step 6 The screen will start to count down After 6 seconds the control test result will appear on the meter s display screen Step 7 Compare the reading on the screen to the Low range printed on the test strip bottle or its package If the reading falls outside the Low range printed on the test strip bottle or its package see Control Solution Trouble Shooting on page 29 NOTE e Control solution testing results will be stored into the meter memory and indicated by ctl icon e Control solution testing results wil
26. est to the most recent readings Step 4 When you reach the most recent result continue to press Y to start viewing day averages The meter will display the 7 day s average press Y again to see 14 day s average and press Y again to see 30 day s average The number of results included in the averages is shown right next tothe icon Control solution results are not included in the day average calculations 40 Viewing Your Meter s Memory EvenCareG2 NOTE e Control testing results are flagged by ctl on the bottom left corner of the display screen in addition to MEM e When the meter memory is blank the display screen will show nil e When there are no day average data available the display screen will show 4 dashes e If the meter memory is full adding a new test result will cause the oldest one to be deleted e Memorized test results and day averages will not be erased when changing batteries e You can begin testing anytime by inserting a test strip 41 EvenCareG2 Caring for the Meter Caring for the Meter Caring for your EvenCare G2 Meter is easy Follow these simple guidelines to keep your meter working properly Changing the Battery Batteries need to be changed when the icon appears on the meter display Materials you will need Two Alkaline AAA Batteries y Your EvenCare G2 Meter Step 1 Press firmly on the battery cover and slide in the direction of the arrow Step 2 Insert new batteries
27. expiration date DO NOT use control solutions if they are expired DO NOT FREEZE Store the control solutions at room temperature You will need y EvenCare G2 Control Solution Low and High Y Your EvenCare G2 Meter Y A new EvenCare G2 Blood Glucose Test Strip Step 1 Take out an EvenCare G2 Test Strip from the test strip bottle and close the bottle immediately Insert the test strip to turn on the meter EvenCareG2 Control Solution Testing IMPORTANT e Check the expiration date printed on the test strip bottle and its package Do not use expired test strips e Use each test strip immediately after removing it from the bottle e Do not use wet or damaged test strips e Keep away from direct sunlight and heat Store the test strip bottle in a dry cool place e Record the date opened on the bottle label Discard the bottle and any remaining test strip after 6 months from date of opening e Test under the temperature range of 50 F to 104 F Your meter will not work outside these ranges e Must insert test strip properly e Must see a flashing blood drop icon if the test strip has been fully inserted Step 2 Wait until the flashing blood drop and arrow icons appear on the meter display screen The meter will also announce Please apply blood onto the test strip if you have the voice feature turned on Press A or Y to enter L1 LOW control solution testing Step 3 ctl icon will appear next to the t
28. flashing blood drop and an arrow icon to appearon the meter display screen The meter will also announce Please apply blood onto the test strip if you have the voice feature turned on Step 3 Obtain blood sample 34 Testing Your Blood Glucose EvenCareG2 For Fingertip Testing Hold the lancing device use the opaque adjustable cap or safety lancet against the side of your fingertip and press the release button to create a puncture TIP e Gently massage your hand and finger toward the puncture site to form a drop of blood Do not milk or squeeze around the puncture site e Lance the side of your fingertip to avoid soreness To avoid calluses choose a different lancing site each time 35 EvenCareG2 Testing Your Blood Glucose For Testing on Palm and Forearm Hold the lancing device use the clear dl E a against the test site base of palm or forearm Avoid a test site that has veins moles excessive hair or bone Press the lancing device firmly against the test site and press the release button do not lift up the lancing device For forearm test sites apply and re lease pressure pumping for a few seconds while keeping the lancing de vice in constant contact with the skin Palm testing at the base of the thumb does not require pumping When the blood is approximately size lift the lancing device straight up without smearing the blood NOTE e Rubbing forearm until wa
29. flecha Paso 2 Inserte baterias nuevas Deslice la bateria Reponga la cubierta en el medidor con firmeza NOTA e El medidor le pedir reconfirmar la fecha y la hora de activaci n Vea pagina 61 para Configuraci n del Reloj e La fecha resultados almacenados ajustes de alarma ajuste de vol umen y elecci n del idioma permanecen en la memoria cuando se cambian las bater as 92 EvenCareG2 Table of Contents Limpieza del Medior Limpieza del Medidor Si el medidor se ensucia utilice un pa o h medo NO MOJADO sin pelusa humedecido con un detergente suave e No introducir agua en el medidor ni ponerlo bajo el chorro de agua e No utilice limpiadores de cristales o caseros en el medidor e No trate de limpiar el soporte de tiras reactivas Almacenamiento y Precauciones Maneje el medidor muicho cuidado contra impactos pues dejar caer el medidor podr a da ar los componentes electr nicos El medidor y las tiras reactivas est n dise ados para ser utilizados en el rango de temperatura de 50 F a 104 C Evite dejar el medidor en lugares extremadamente c lidos o fr os como cerca de una fuente de calor o en un coche extremadamente caliente o fr o No guarde o utilice el medidor o las tiras reactivas en niveles d humedad alta como el ba o o la cocina Siempre tape el frasco inmediatamente despu s de retirar una tira reactiva y aseg rese de que est bien cerrado No despiece el medidor ya que al hacerlo
30. he lancing device with your other hand until you hear a click sound The lancing de vice is now ready for blood sampling IMPORTANT Use a new sterile lancet every time you test to avoid cross contamina tion If alcohol wipes are used to cleanse the fingers make sure the area is dry before the blood sample is obtained 31 EvenCareG2 Testing Your Blood Glucose Important Information on Alternate Site Testing AST The EvenCare G2 Blood Glucose Monitoring System can test for blood glucose from areas other then your fingertip such as palm and forearm alternate site testing or AST Alternate site testing can be less painful then fingertip testing but because of the physiological differ ences between your fingertip and palm and forearm alternate site test results maybe significantly different then results from fingertip testing under certain conditions You should consult with your doctor or healthcare professional before using alternate site testing DO AST ONLY in the following intervals e In a pre meal or fasting state more than 2 hours since the last meal e 2 hours or more after taking insulin e 2 hours or more after exercise Alternate Site Testing SHOULD NOT be used when e You are Hypoglycemic unaware not able to tell if you have low blood sugar e Within 2 hours of a meal exercise or medication e You will be operating machinery or driving a car e During times of stress e You are sick e You t
31. hink your blood glucose is low e Your AST results do not match the way you feel e You are testing for hyperglycemia e Your routine glucose results are often fluctuating 32 Testing Your Blood Glucose EvenCareG2 NOTE e Rubbing forearm until warm before lancing will help minimize the difference with fingertip test results The palm at the base of the thumb does not require rubbing before testing e If results from alternate site testing do not agree with how you feel use fingertip testing instead 1 Jungheim K Koschinsky T Risky delay of hypoglycemia detection by glucose monitoring at the arm Diabetes Care 2001 24 7 1303 04 33 EvenCareG2 Testing Your Blood Glucose Performing a Blood Glucose Test Step 1 Take out an EvenCare G2 Test Strip from the test strip bottle and close the bottle immediately Insert the test strip to turn on the meter IMPORTANT e Check the expiration date printed on the test strip bottle and its package Do not use expired test strips e Use each test strip immediately after removing it from the bottle e Do not use wet or damaged test strips e Keep away from direct sunlight and heat Store the test strip bottle in a dry cool place e Record the date opened on the bottle label Discard the bottle and any remaining test strip after 6 months from date of opening e Must see a flashing blood drop icon if the test strip has been inserted properly Step 2 Wait for a
32. ime on the EvenCare G2 Meter has not been set test or control result will not be stored in memory Step 1 Press and hold Enter to turn on the meter Step 2 Press Enter to go into setup screen then press AorY until start to flash If the voice function is on the meter will say Set date and time Press Enter to enter clock setup Step 3 Set the Year The current year will flash If the voice function is on the meter will say Set Year press AorWto select the correct year Press Enter to con firm your choice and advance to set the month EvenCareG2 setting Up Your New System Step 4 Set the Month The current month will flash If the voice function is on the meter will say Set Month press AorY to select the correct month Press Enter to confirm your choice and advance to set the date Step 5 Set the Date The current date will flash Press AorY to select the correct day If the voice function is on the meter will say Set Day Press Enter to con firm your choice and advance to set time format Step 6 Set Time Format to 12 hour or 24 hour The current time format will flash Press AorW to select for either 12 hour format or 24 hour format Press Enter to confirm your choice and ad vance to set hour Step 7 Set the Hour The current hour will flash If the voice function is on the meter will say Set Hour press
33. l not be used for calculating averages 27 EvenCareG2 Control Solution Testing Step 8 Removed the used test strip The meter will automatically turn off Discard the used test strip IMPORTANT Do not reuse test strips Follow steps 3 through 8 to do High control solution testing and com pare the result with the High range printed on the test strip bottle 28 Control Solution Testing EvenCareG2 Control Solution Trouble Shooting If your control solution testing is out of range too high or too low it may be caused by the following Possible Causes What you can do Wrong brand of Control e Make sure you are using Solution being used EvenCare G2 Control Solution e Control solution not at room e Make sure the testing environ temperature ment is between 50 F to 104 F e Expired or contaminated e Check the expiration and open control solution or damaged bottle date on both the control test strip solution and test strips Repeat testing using a new test strip If the result is still out of range use a new bottle of control solution and retest e Meter malfunction e Ifthe problem persists please call Medline s Technical Service Center at 1 866 274 2131 Monday through Friday between 8 00 a m and 5 00 p m Central Time Contact your Healthcare Provider at all other times 29 EvenCareG2 Testing Your Blood Glucose Testing Your Blood Glucose You will need Y Your EvenCare G2 Meter Y A new EvenCa
34. l speak WHAT the meter will say Temperature is out of range be Temperature is out of range tween 50 F 104 F When the meter is turned off Melody will sound Test Strip is inserted and press A or Y CTL showing next to the test strip icon L1 control solution test Please apply control solution onto the test strip y v A A Blood glucose test A L2 control solution Please apply blood d test Please apply onto the test strip v control solution 21 EvenCareG2 voice Function Table Setting Meter Functions via the Setup Screen see page 12 setting up your new system Pressing Enter to enter setup screen after the meter has been turned on Memory Exit Function Language A 5 D YA Recall y Mode s Selection A PC Link This function is no a i i gt N Set Date A Set Testing Set Sound gt SS and Time v Alarm lt Volume WHEN the meter will speak WHAT the meter will say During date and time setup Set Year setting up the year During date and time setup Set Month setting up the month During date and time setup Set Date setting up the date During date and time setup Set Hour setting up the hour 22 Voice Function Table EvenCareG2 WHEN the meter will speak WHAT the meter will say During date and time setup Set Minutes setting up the Minutes After finishing setting up date The date and time is year an
35. la Memoria de su Medidor 87 Cuidado del Medidor Cambio de la Bater a 90 Limpieza del Medidor 91 53 53 ndice Table of Contents EvenCareG2 Almacenamiento y Precauciones 91 Solucion de Problemas 92 Garantia del Producto 95 Especificaciones 96 54 EvenCareG2 TaBieticidententu Nuevo Sistema de Glucemia Entendiendo Su Nuevo Sistema de Glucemia El Sistema de Monitoreo de Glucemia EvenCare G2 contiene los siguientes elementos e Medidor EvenCareG2 e Gu a del Usuario e Funda de transporte e Libro de registro 55 Entendiento los Resultados dTabl nisdarteriuchwenCareG 2 El Medidor EvenCareG2 Puerto de tira reactiva Inserte la tira reactiva ver p gina 82 para encender el medidor Pantalla del Medidor Muestra los resultados de su prueba de glucemia valores de memoria promedios y otros mensajes Intro y Botones st Mantenga pulsada la tecla Intro por 2 segundos para activar o desactivar el Medidor or Pulse Intro para entrar en en el modo de configuraci n cuando no se prueba Ver p gina 60 Presion AoVY tpara entrar en el modo de control con tira reactiva insertada Ver p gina 74 durante configuraci n del medidor ver p gina 60 o para navegar por los valores de prueba y promedios almacenados ver p gina 87 56 EvenCareG2hnditstitelos Gestet dos de su Prueba de Glucemia Compartimiento de Bater a Contiene dos bater as AAA Puerto de Datos Esta funci n no est ac
36. la hora es a o mes Configurar de fecha y hora fecha hora minutos Ajuste de alarma 1 2 3 Configurar Alarma n mero Al enviar los datos al PC Enviando datos a PC Esta funci n no est disponible todav a Finalizar el env o de datos a PC A Transmisi n de datos completa Esta funci n no est disponible todav a 73 Tabla de Funci n de Voz Table of Contents EvenCareG2 Otros mensajes de advertencia consulte la pagina 92 Resolucion de Problemas CUANDO el medidor hablara La fecha y hora del medidor no se han establecido Fallo de Medidor Tira reactiva usada No hay suficiente sangre o soluci n de control LO QUE el medidor dir Por favor configure fecha y hora Error 1 Fallo de Medidor Error 3 Tiras reactiva usada Error 4 muestra insuficiente 74 EvenCareG2 Table of ContentPrueba de Soluci n de Control Prueba de Soluci n de Control El prop sito de la prueba de soluci n de control es asegurarse que el medidor Evencare G2 y la tira reactiva EvenCare G2 est n funcionando adecuadamente Debe realizar pruebas de soluci n de control cuando e Utilice el medidor por primera vez e Use una nueva botella de tiras reactivas de glucemia EvenCare G2 e Dej la tapa de la botella de tiras reactivas abierta por un tiempo e Se le ha ca do el medidor e Sospecha que su medidor y las tiras reactivas no est n funcionando adecuadamente e Los resultados de
37. ll Medline be liable for indirect or consequential damages or for any loss arising from misuse improper use abuse neglect unauthorized repair or modification This warranty is void and of no force and effect in the event of product misuse improper use abuse neglect unauthorized repair or modifica tion This warranty specifically excludes EvenCare G2 Test Strips and EvenCare G2 Control Solution THIS WARRANTY IS THE SOLE AND EXCLUSIVE WARRANTY TO THE EXCLUSION OF ALL OTHER WARRANTIES EXPRESS OR IMPLIED ALL OF WHICH ARE WAIVED AND DISCLAIMED All warranty claims must be directed to Medline s Technical Service Center at 1 866 274 2131 This warranty is extended only to the original purchaser of the meter 49 Specifications Specifications Test Strips Test Range Calibration Test Time Sample Size Hematocrit Range Display Type Memory Result Averaging Dimension Weight Battery Battery Life Automatic Power off Operating Temperature Relative Humidity Storage Condition Table of Contents EvenCare G2 Blood Glucose Test Strips 20 600 mg dL Plasma 6 Seconds Larger than 1uL 30 55 Large LCD screen 300 blood test results with date and time 7 14 and 30 days averaging 93L x 58W x 20 5H mm 79g 2 Alkaline AAA Battery 1000 tests of continuous use or 1 year After 2 minutes of nonuse 50 F to 104 F Less than 85 Meter at 39 F 122 F 50 EvenCareG2 Estim
38. local government regulation to avoid injury or contamination 37 EvenCareG2 Understanding Your Blood Glucose Test Results Understanding Your Blood Glucose Test Results Blood glucose values will vary depending on food intake medication health stress and exercise The ideal range for adults with and without diabetes should bet e 70 to 100 mg dL before meals and e 70 to 140 mg dL after meals It is important to consult with your physician or healthcare professional to determine an appropriate target range for you What to do if you get a high or low reading If the Meter displays results that are HI or LO or you get a result that is more than 250 mg dL or below 50 mg dL AND you feel ill e Treat your diabetes according to the instruction from your doctor and or consult your healthcare provider If the Meter displays results that are HI or LO or you get a result that is more than 250 mg dL or below 50 mg dL AND you DO NOT feel ill e Test your meter with a control solution refer to page 25 Control Solution Testing e Test again using fingertip If you still get a high or low reading contact your health care professional NOTE e Inaccurate results may occur in severely hypotensive individuals or patients in shock e Inaccurate low results may occur for individuals experiencing a hyperglycemic hyperosmolar state with or without ketosis e Critically ill patients should not be tested with blood glucose meters
39. m The EvenCare G2 Blood Glucose Monitoring System contains the following items if it arrives as a meter only system e EvenCareG2 Meter e User s Guide e Carrying Case e Log Book The EvenCare G2 Blood Glucose Monitoring System contains the following items if is arrives as a Starter Kit e EvenCareG2 Meter e Lancing Device e 10 Lancets e 10 Test Strips e User s Guide e Carrying Case e Log Book EvenCareG2 Understanding Your New Blood Glucose System The EvenCareG2 Meter Test Strip Port Insert test strip see page 34 to turn on the meter Meter Display Shows your blood glucose test re sults memory values averages and other messages Enter and AY buttons Press and hold Enter for 2 seconds to turn on or off the Meter or press Enter to go into setup mode when not testing see page 12 Press Aor Y to enter control mode with test strip inserted see page 26 during meter setting see page 12 or to navigate the stored test values and averages see page 39 Understanding Your New Blood Glucose System EvenCareG2 Battery Compartment Holds two AAA batteries Data Port This function is not yet available EvenCareG2 Understanding Your New Blood Glucose System Meter Display Screen BATTERY Appears when batteries need to be changed see page 42 END Will appear on the display screen after the last recorded test result if less then 300 readings are stored
40. ma de la tira reactiva El medidor emitir un pitido que in dica que ha recibido lasoluci n de control Paso 6 La pantalla empezar la cuenta atr s Despu s de 6 se gundos el resultado de la prueba de control aparecer en la pantalla del medidor Paso 7 Compare la lectura en la pantalla a la gama baja impresa en el frasco de tiras reactivas o de su paquete Si la lectura cae fuera del rango Bajo impreso en la botella de tiras reactivas o su paquete consulte Soluci n de Problemas de Control en P gina 77 High 135 2203 NOTA e Los resultados de la prueba de soluci n de control se almacenaran en la memoria del medidor y se indican por el s mbolo ctl e Los resultados de la pruebas de soluci n de dontrol no ser n utilizados para el c lculo de promedios 77 Prueba de Soluci n de ControTable of Contents EvenCareG2 Paso 8 Retirada la tira reactiva usada El medidor se apagara automatica mente Deseche la tira reactiva usada IMPORTANTE No vuelva a usar las tiras reactivas Siga los pasos del 3 al 8 para hacer las pruebas de control de soluci n Alta y comparar el resultado con el rango Alto impreso en el envase de tiras reactivas 78 EvenCareG2 Table of ContentPrueba de Soluci n de Control Soluci n de Problemas de Soluci n de Control Si su prueba de soluci n de control est fuera de rango demasiado alta o demasiado baja puede ser causado por lo siguiente
41. mg dL AND you feel ill e Treat your diabetes according to the instruction from your doctor and or consult your healthcare provider e Trate su diabetes de acuerdo a las instrucciones de su m dico y o con sulte con su proveedor de atenci n m dica Si el Medidor muestra resultados que son HI o LO o bien obteine un resultado que es superior a 250 mg dL o inferior a 50 mg dL y NO se siente enfermo e Pruebe su medidor con una soluci n de control consulte la pagina 73 Prueba de Soluci n de Control e Pruebe de nuevo con la yema del dedo Si a n as obtener una lec tura alta o baja contacte con su profesional de la salud NOTA e Resultados inexactos tambi n pueden producirse en individuos severamente hipotensos o pacientes en estado de shock e Los resultados inexactos bajos puede producirse para individuos que experimentan un estado hipergluc mico hiperosmolar con o sin cetosis e Los pacientes cr ticamente enfermos no deben ser probados con medidores de glucemia Referencia 1 American Diabetes Association Standards of Medical Care in Diabetes 2008 Diabetes Care 2008 31 512 S54 88 EvenCareG2 Table of Contefiemdo la Memoria de su Medidor Viendo la Memoria de su Medidor Su medidor EvenCare G2 almacena autom ticamente hasta 300 resultados de pruebas y control con fecha y hora El medidor tambi n dispone de promedios de 7 14 y 30 d as para ayudar a un seguimiento de su tendencia de glucemia Debe configura
42. nitoring System can be used only during steady state blood glucose conditions e Dehydration Severe dehydration may lead to inaccurate blood glucose test results If you suspect you are severely dehydrated contact your healthcare professional immediately e Hematocrit range A hematocrit range that is higher than 55 or lower than 30 can cause inaccurate blood glucose test results Table of Contents EvenCareG2 Table of Contents Understanding Your New Blood Glucose System The EvenCareG2 Meter 6 Meter Display Screen 8 The EvenCareG2 Test Strip 10 EvenCareG2 Glucose Control Solution Recommended 11 Setting Up Your New System The Set Up Screen 12 Setting the Clock 13 Setting the Alarms 16 Voice and Beeper Function 18 Language Selection 19 Voice Function Table 20 Control Solution Testing Performing a Control Solution Test 25 Control Solution Trouble Shooting 29 Testing Your Blood Glucose Preparing Your Lancing Device 30 Important Information on Alternate Site Testing AST 32 Performing Blood Glucose Test 34 Understanding Your Blood Glucose Test Results 38 Viewing Your Meter s Memory 39 Caring for the Meter Changing the Battery 42 Cleaning the Meter 43 EvenCareG2 Table of Contents Storage and Precautions 43 Solving Problems 44 Product Warranty 47 Specifications 48 Understanding Your New Blood Glucose System EvenCareG2 Understanding Your New Blood Glucose Syste
43. ntro para en cender el medidor Paso 2 Pulse Intro para entrar en la pantalla de Configuraci n a continuaci n Presione Ao V until hasta que empiece a parpadear Si la funci n de vozest activada el medidor dir Ajuste la fecha y la hora Pulse Intro para entrar en la configuraci n del reloj Paso 3 Configurar el A o El a o actual parpadeara Si la funci n de voz est activada el medidor dir Ajuste A o presione AoY para seleccionar el a o correcto Pulse Intro para confirmar su elecci n y avance a ajustar el mes 63 Configuraci n de su Nuevo SiStesteof Contents Paso 4 Configurar el mes El mes actual parpadeara Si la funci n de voz est activada el medidor dir Ajuste mes pre sione AoY para seleccionar el mes correcto Pulse Intro para confirmar su elecci n y avance a ajustar la fecha Paso 5 Ajuste de la fecha La fecha actual parpadear Presione AoY para seleccionar el d a correcto Si la funci n de voz est activada el medidor dir Ajuste d a Pulse Intro para confirmar su selecci n y avance para configuraci n de formato de hora Paso 6 Establecer Formato de Hora a 12 horas o 24 horas El actual formato de hora aparecer en pantalla Pulse AoW para Seleccionar ya sea formato de 12 horas o de 24 horas Pulse Intro para confirmar su elecci n y avanzar para configurar la hora Paso 7 Ajuste de hora La hor
44. ontrol de L1 Por favor aplique soluci n de control en la tira reactiva Prueba de A Soluci n de control glucemia Por favor pr L2 Por favor aplique aplique sangre e v soluci n de control la tira reactiva 71 Tabla de Funci n de Voz Table of Content EvenCareG2 Ajuste de las Funciones del Medidor a trav s de la Pantalla de Configuraci n Consulte la p gina 60 configuraci n de su nuevo sistema Pulsar Intro para entrar en la pantalla de configuraci n despu s de que el medidor ha sido encendido A Memoria A Recuperar 4 PC Link This function is not available yet N A A Configurar gt oe A Funci n Salir E Modo Selecci n A Configurar Configurar Sonido E fecha y Hora v Pruebas Alarma v Volumen CUANDO el medidor hablara Durante la configuraci n de fecha y hora configuraci n del a o Durante la configuraci n de fecha y hora configuraci n del mes Durante la configuraci n de fecha y hora configuraci n de la fecha Durante la configuraci n de fecha y hora configuraci n de la hora LO QUE el medidor dir Configurar A o Configurar Mes Configurar Fecha Configurar Hora 72 EvenCareG2 Table of Contents Tabla de Funci n de Voz CUANDO el medidor hablar QU dir el medidor Durante la configuraci n de Configurar Minutos fecha y hora configuraci n de minutos Despu s de terminar la La fecha y
45. ood glucose reading Run the control test to make sure the test strips and the meter are working together properly to practice testing procedure and when using a new lot of EvenCare G2 Blood Glucose Test Strips 11 EvenCareG2 setting Up Your New System Setting Up Your New System The Setup Screen The Setup Screen of the EvenCare G2 Meter lets you take advantage of the advance functions of this meter Simply use AV buttons to navigate to the desired function icon and press the Enter button to enter the function when the icon is flashing Navigate to End at the display screen to exit the Setup Screen The Meter will also announce each function when the voice feature is on mem K E Ln of LL Enter to enter Memory Recall to retrieve stored re sults Your meter can memorize up to 300 memories with date and time see page 39 A Press Enter to enter PC Link Mode for data transfer onto a PC This function is not available yet da Press Enter to enter Time and Date Setup see page 13 D Press Enter to enter Alarm Setup see page 16 Your meter can have as many as 3 alarm settings 12 Setting Up Your New System EvenCareG2 Q PEE Press Enter to enter Volume Setup see page 18 Q ais Press Enter to enter Language Selection see page 19 NG sasz Exit Setup Screen and return to testing Setting the Clock You will need Y Your EvenCare G2 Meter NOTE If date and t
46. r la fecha y la hora en el medidor correctamente para utilizar la memoria y la funci n de promedio de d a vea Configuraci n de su Nuevo Sistema en la p gina 60 El medidor no memorizar resultados de prueba o el control si la fecha y hora no est n ajustadas Visualizaci n de Lecturas Guardadas y Promedio de 7 14 30 d as de la Memoria Necesitar Su Medidor EvenCare G2 Paso 1 Mantenga pulsada la tecla Intro para encender el medidor Paso 2 Pulse Intro para acceder a la pantalla de configuraci n Paso 3 Presione A o Y hasta que MEM comience a parpadear Si la funci n de voz est activada el medidor dir recuperaci n de la memoria Pulse Intro para entrar 7s Sam MEM aparecer en la esquina superior izquierda dA E de la pantalla junto con su m s reciente a resultado de prueba Presione A para ver el resul E nio tado siguiente y el siguiente y as sucesivamente Sostener A destellar los resultados I amp i i i 89 Viendo la Memoria de su Mediile of Contents EvenCareG2 END aparecera en la pantalla despu s del resultado de la ultima prueba registrada si se almacenan menos de 300 lecturas Puede presionar Y para volver viendo desde la m s antigua a las lecturas m s recientes Paso 4 Al llegar al resultado m s reciente contin e presionando Y para iniciar visualizar los promedios diarios El medidor mostrar el promedio de 7 d as pulse W de
47. r maquinaria o conducir un coche e Durante los momentos de estr s e Usted est enfermo e Usted cree que su glucemia es baja e Sus resultados de AST no son iguales a como se siente e Usted esta haciendo prueba para hiperglicemia e Sus resultados de glucosa de rutina a menudo son fluctuantes 82 EvenCareG2 Table of Contents Probando su Glucemia NOTA e Frotar el antebrazo hasta que se caliente antes de punci n ayudar a minimizar la diferencia con los resultados del examen de yema del dedos La palma en la base del pulgar no requiere ser frotada antes de la prueba e Si los resultados de las pruebas en sitios alternos no est n de acuerdo con c mo se siente use prueba de yema de dedos 1 Jungheim K Koschinsky T Risky delay of hypoglycemia detection by glucose monitoring at the arm Diabetes Care 2001 24 7 1303 04 83 Paso 1 Saque una Tira Reactiva EvenCare G2 del frasco de tiras reactivas y cierre el frasco inmediatamente Inserte la tira reactiva para encender el medidor Probando su Glucemia Table of Contents EvenCareG2 Realizacion de una prueba de glucemia IMPORTANTE Verifique la fecha de caducidad impresa en el frasco de tiras reactivas y su paquete No utilice las tiras reactivas caducadas Use cada tira reactiva inmediatamente despu s de sacarla de la botella No utilice tiras reactivas mojadas o da adas Mant ngase alejado de la luz solar directa y el calor Almacene e fr
48. ra de alertarle de mensajes importantes Una vez que haya determinado el volumen adecuado pulse Intro de nuevo para salir del ajuste de volumen 68 EvenCareG2 Table of Coftenfiguraci n de su Nuevo Sistema Selecci n de Idioma Paso 1 Mantenga pulsada la tecla Intro para en cender el medidor Paso 2 Pulse Intro para acceder a la pantalla de configuraci n Paso 3 Presione Ao V E hasta que comience a parpadear El medidor anun ciar Selecci n de Idioma Paso 4 El idioma por defecto 1 en la pantalla es anunciado Pulse Intro para cambiar el idioma Presione AoVY para seleccionar otro idioma Presione Intro cuando haya terminado de seleccionar el idioma y salido de la selecci n de idioma Ingl s y espa ol son las dos opciones NOTA e En cualquier momento durante la configuraci n puede insertar una tira reactiva EvenCare G2 y comenzar la prueba Cualquier cambio realizado hasta el momento se almacena e Su elecci n de configuraci n de idioma y volumen no se borrar al cambiar las bater as Comenzar la prueba mediante la inserci n de una tira reactiva de glucemia EvenCare G2 Vea p gina 78 Probando su glucemia o mantenga pulsada la tecla Intro para apagar el medidor 69 Tabla de Funci n de Voz Table of Contents EvenCareG2 Tabla de Funcion de voz La funci n de voz se puede utilizar como una ayuda para que usted pueda escuchar los resultado
49. re G2 Blood Glucose Test Strip Y Lancing device with a clean unused lancet or safety lancets y Clear cap for Alternate Site Testing AST on palm and forearm Before you begin make sure e To set up your meter properly and run a control test See Setting Up Your New System on page 12 and Control Solution Testing on page 25 for details e Wash your hands and the testing site thoroughly with soap and warm water and dry well e You are testing in an area between the temperature ranges of 50 F 104 F Your meter will not test outside of this range and will display thermometer icon with verbal indication if the voice feature is turned on Move the meter into an area that is between 50 F 104 F and let it sit for 10 to 15 minutes before testing again Preparing Your Lancing Device for Home Testing Step 1 Unscrew the adjustable cap of the lancing device and insert the lancet by pushing down firmly until it is fully seated Step 2 Twist the protective cover off the lancet Twist the adjustable cover clockwise back onto the lancing device 30 Testing Your Blood Glucose EvenCareG2 NOTE e For fingertip testing use the opaque adjustable cap e For alternate site testing on palm and forearm use the clear adjustable cap Step 3 Adjust the puncture setting on the adjustable cap for the puncture depth level 1 is the lightest and 5 is the deepest Step 4 With one hand holding the adjustable cap pull back t
50. ries are low What itWhat Should You See MeansYou Do Temperature out of range Test result is higher than 600 mg dL Test result is lower than 20 mg dL EvenCareG2 Table of Contents Solving Problems Change the battery ac cording A page 42 Changingdbe Battery Move the meter into an area that is between 50 F 104 F and allow 10 to 15 minutes for it to reach the new tempera ture Wash and dry your hands and repeat the test on your fingertip with a new test strip If the result is still HI contact your physician or healthcare professional immediately Wash and dry your hands and repeat the test on your fingertip with a new test strip If the result is still LO contact your physician or healthcare professional immediately Solving Problems Not Applicable What What Should You See There are no memorized results in the meter What the meter says Table of Contents EvenCareG2 Check if the date and time on your meter is set up See Setting Up Your New System on page Wat it Start testing your blood glug Testing Your Blood Glucose on page 30 48 EvenCareG2 Table of Contents Product Warranty Product Warranty The EvenCare G2 Blood Glucose Meter is warranted to be free of defects in workmanship and materials for a period of five 5 years from the date of purchase Medline s liability for warranty claims is limited to repair or replacement and in no event sha
51. rm before lancing will help minimize the difference with fingertip test results The palm at the base of the thumb does not require rubbing before testing e f results from alternate site testing do not agree with how you feel use fingertip testing instead 36 Testing Your Blood Glucose EvenCareG2 Step 4 Gently bring the test strip and touch the drop of blood at a slight angle The test strip acts like a straw to pull the blood in Keep the test strip in the blood drop until the meter beeps to indicate the test strip has enough blood to test Step 5 The screen will start to count down After 6 seconds your glucose testing result will appear on the meter display screen and will be announced verbally if you have the voice feature turned on Your test result will be automatically stored into the meter memory CAUTION e If you see HI or LO displayed your blood glucose level may be above 600 mg dL or below 20 mg dL Test again using fingertip testing DO NOT test on forearm If you still receive the same result call your physician or healthcare professional immediately Step 6 Properly discard the used test strip The meter will automatically shut off Step 7 Remove the used lancet from your lancing device according to instructions and discard into proper waste container IMPORTANT e Used lancets and strips are biohazardous materials and can transmit bloodborne diseases Dispose of them according to
52. s de las pruebas y usar del medidor En la tabla siguiente se lista CUANDO el medidor hablar y QU dir el medidor Pruebas de sangre y soluciones de control vea p gina 73 Prueba de de Soluci n de Control y p gina 78 Probando su glucemia CUANDO el medidor hablar LO QUE el medidor dir Encendido Sonar el Tono de Saludo Pide al usuario que inserte la tira reactiva Por favor introduzca la tira reactiva La tira reactiva se inserta y la medidor est listo para probar con flecha parpadeante y gota de sangre Prueba de glucemia Por favor aplique sangre a la tira reactiva El examen de glucemia se completa y el resultado se muestra en la pantalla Su lectura de glucemia es N mero miligramos por decilitro Su lectura de glucemia es El resultado de la prueba por debajo de 20 mg dL de glucemia est por DEBAJO del rango de medici n de 20 mg dL Su lectura de glucemia es Resultado de la prueba de por encima de 600 mg dL glucemia est por ENCIMA del rango de medici n de 600 mg dL 70 EvenCareG2 Table of Contents Tabla de Funci n de Voz CUANDO el medidor hablar LO QUE el medidor dir La temperatura est fuera de La temperatura est fuera de rango entre 50 F 104 F rango Cuando el medidor est Sonar una melod a apagado La tira reactiva se inserta y pulse A o Y CTL se muestra junto al icono de tira reactiva Prueba de soluci n de c
53. te que a ada la muestra e Retire la tira reactiva y repita la prueba con una nueva tira reactiva Ver Probando su Glucemia en la p gina 78 94 Lo que ve os l mg dL 10 18 i teow 95 Lo que el medidor dice Baja Bateria Temper atura est fuera de rango Su lectura de glucemia est por encima de 600 mg dL Su lectura de glucemia est Por de bajo de 20 mg dL Lo que significa Baterias estan bajas Resultado de la prueba es mas alto que 600 mg dL Resultado de la prueba es mas alto que 600 mg dL Resultado de la prueba es mas bajo que 20 mg dL EvenCareG2 Table of Contents Soluci n de Problemas Qu debe hacer Cambiar la bater a de acuerdo a la p gina 90 Cambio de la Bater a Mueva el medidor a un rea que est entre 50 F 104F y permita 10 a 15 minutos para que alcance la nueva temper atura Lave y seque sus manos y repita la prueba en la yema de su dedo con una nueva tira reactiva Si el resultado sigue siendo HI comun quese con su m dico o profesional de salud de inmediato Lave y seque sus manos y repita la prueba en la yema de su dedo con una nueva tira reactiva Si el resultado es todav a LO comun quese con su m dico o profesional de salud Lo que ve Soluci n de Problemas Lo que el medidor dice No es Aplicable Lo que significa No hay resultados memori
54. ted for easy comparison to lab results however results may be 10 to 15 higher than results obtained from whole blood referenced blood glucose assays The test strip is also equipped with under fill detection to alert you when there is not enough blood to perform a test so you can be assured that every reading you get is an accurate and meaningful result IMPORTANT e Immediately close the vial cap of the test strip bottle tightly after each use Keep the unused test strips in its original bottle e DO NOT leave any test strips outside the bottle while not in use e DO NOT reuse test strips Test strips are for single use only e Carefully discard used test strips and lancets in proper waste containers 10 Understanding Your New Blood Glucose System EvenCareG2 EvenCare G2 Glucose Control Solution Recommended don S00 USA Made in Call your local supplier or pharmacy to order control solution If they do not stock it please contact us at 1 866 274 2131 between 8 00 am and 5 00 pm central time Monday through Friday EvenCare G2 Glucose Control Solution is the standard glucose concen tration solution required to perform a Glucose Control Solution Test See Control Solution Testing It is available at 2 levels EvenCare G2 LOW Glucose Control Solution helps you to validate the performance of your system at a low range blood glucose reading while HIGH solution helps you to validate your system performance at a high range bl
55. tualmente disponible 57 Entendiento los Resultados ddisbl afdbantentiucd KenCareG2 Pantalla de Medidor CABLE BATER A Indica que el medidor est en modo Aparece cuando las bater as de transferencia de datos a PC necesitan ser cambiadas Esta funci n no est disponible todav a ver p gina 90 FIN Aparecer en la pantalla despu s del ltimo resultado de prueba grabado si hay menos de 300 lecturas almacenadas MEM Indica que el n mero en la pantalla del medidor es un valor de prueba almacenado ver p gina 87 ALARMA Indica que la alarma se ha fi jado ver p gina 64 CTL Este icono indica una prueba de soluci n de control ver p gina 73 o que el valor almacenado es un resultado desolaci n de contral Muestra la fecha hora am pm ver p gina 87 En formato 24 horas vea la pagina 12 o el tipo de promedios d a 7 14 o 30 vea la p gina 39 III E DAY CICI TITIA ore Ll i ave cect cece PM 58 EvenCareGehnditxtitelos Gesteat sdos de su Prueba de Glucemia ZUMBADOR Configuraci n del volumen de sonido ver p gina 66 TEMPERATURA Aparece cuando est demasiado caliente o demasiado fr o para prueba fuera de los rangos de 50 F 104 F RELOJ Indica que el medidor est listo para configurar la hora y fecha ver p gina 61 VOZ Indica que la funci n de voz est activada ver p gina 66 UNIDADES Unidad de medida de glucemia es
56. uera de estos rangos e Debe insertar la tira reactiva correctamente e Debe ver un icono de gota de sangre intermitente si la tira reactiva ha sido insertada completamente Paso 2 Espere hasta que los iconos de gota de sangre y flecha parpadeantes aparezcan en la pantalla del medidor El medidor tambi n anunciar Por favor aplique sangre en la tira reactiva si usted tiene la caracter stica de voz activada Presione A o Y para introducir prueba de soluci n de control L1 BAJA Paso 3 El icono ctl aparecer junto al icono de la tira reactiva y L1 aparecer en la pantalla del medidor El medidor tambi n anunciar prueba de soluci n de control L1 por favor aplique la soluci n de control en la tira reactiva Si usted decide no hacer una prueba con soluci n de control pulse A o Y hasta que ctl desaparezca de la pantalla del medidor 76 EvenCareG2 Table of ContentPrueba de Soluci n de Control Paso 4 Exprima una gota de soluci n de control en una superficie limpia seco absorbente No aplique soluci n de control a la tira reactiva directamente de la botella Vuelva a colocar la tapa de la botella en la botella de soluci n de control in mediatamente despu s de su uso NO toque el Extremo de Muestreo de la tira reactiva con la botella de soluci n de control Vea la ilustraci n del Extremo de Muestreo a continuaci n Paso 5 Toque suavemente la gota de la soluci n de control con la ye
57. un solo uso e Elimine con cuidado las tiras reactivas y lancetas usadas en contenedores de residuos apropiados 60 EvenCareGehnditstitelos esidtdsdos de su Prueba de Glucemia Soluci n de Control de Glucosa EvenCare G2 recomendado Llame a su proveedor o farmacia local para pedir la soluci n de control Si no la tienen por favor contacte con nosotros al 1 866 274 2131 entre las 8 00 am y 5 00 pm hora del centro de lunes a viernes La Soluci n de Control de Glucosa EvenCare G2 es la soluci n de con centraci n de glucosa est ndar requerida para realizar una Prueba de Soluci n de Control de Glucosa V ase Ensayo de Soluci n de Con trol Est disponible en dos niveles EvenCare G2 Soluci n de Control de Glucosa BAJA le ayuda a validar el rendimiento de su sistema en un rango de lectura de glucemia baja mientras que la soluci n ALTA le ayuda a validar el rendimiento del sistema en una lectura de glucosa en sangre de rango alto Ejecute la prueba de control para asegurarse de que las tiras reactivas y el medidor est n trabajando juntos de man era adecuada para practicar los procedimientos de prueba y cuando utilice un nuevo lote de Tiras de Prueba de Glucemia EvenCare G2 61 Configuraci n de su Nuevo Sistesteof Contents EvenCareG2 Configuraci n de su Nuevo Sistema La Pantalla de Configuraci n La Pantalla de Configuraci n del medidor EvenCare G2 le permite aprovechar funciones avanzadas de este
58. urar el reloj e Los ajustes de alarma no se borran al cambiar las bater as Comenzar la prueba mediante la inserci n de una tira reactiva de glucemia EvenCare G2 ver p gina 78 Probando su Glucemia o mantenga pulsada la tecla Intro para apagar el medidor o cosulte la siguiente secci n sobre c mo configurar la funci n de voz de su medidor 67 Configuraci n de su Nuevo Sistesteof Contents EvenCareG2 Funcion de Beeper y Voz Su medidor tiene una funci n de voz que se puede utilizar como una ayuda verbal para usar el medidor y escuchar los resultados de la prueba La configuraci n predeterminada para la funci n de voz es acti vada y puede ser apagada para ahorrar bater a Cuando la funci n de voz est apagada el medidor emitir un pitido para avisarle de mensajes importantes Necesitar Y Su Medidor EvenCare G2 Ajuste de volumen Paso 1 Mantenga pulsada la tecla Intro para encender el medidor Paso 2 Pulse Intro para entrar en la pantalla de con figuraci n Presione Ao Y until hasta que comience a parpadear Pulse Intro para acceder a la configuraci n de volumen Paso 3 La pantalla mostrar el volumen de sonido actual por defecto es 2 Presione AoY para ajustar el volumen del sonido del 1 menor al 5 mayor Una melod a sonar de acuerdo con el volumen del sonido que ha seleccionado Ajustar a O desactivar la funci n de voz El medidor dar un pitido pa
59. voz activada El resultado de la prueba se almacena autom ticamente en la memoria del medidor PRECAUCI N Probando su Glucemia e Si aparece HI o LO su nivel de glucemia puede estar por encima de 600 mg dL o por debajo de 20 mg dL Pruebe de nuevo con la prueba de yema del dedo NO pruebe en el ante brazo Si sigue recibiendo el mismo resultado llame a su m dico o profesional de salud inmediatamente Paso 6 Deseche la tira reactiva usada El medidor se apagar autom ticamente Paso 7 Retire la lanceta del dispositivo de punci n de acuerdo con las instrucciones y deseche en un contenedor de residuos adecuado IMPORTANTE e Lancetas y tiras usadas son materiales biopeligrosos y pueden transmitir enfermedades de transmisi n sangu nea Des chelos seg n las normativas locales 87 ca 1108 14 10 Entendiendo los Resultados de su Prueba de Glucemia los valores de glucemia pueden variar dependiendo de la ingesta de ali mentos medicamentos salud estr s y ejercicio El rango ideal para adultos con y sin diabetes debe ser de 1 e 70 a 100 mg dL antes de las comidas y e 70 a 140 mg dL despu s de las comidas Es importante consultar con su m dico o profesional de la salud para determinar un intervalo de referencia apropiado para usted Qu hacer si obtiene una lectura alta o baja If the Meter displays results that are HI or LO or you get a result that is more than 250 mg dL or below 50
60. y through Friday between 8 00 a m and 5 00 p m Central Time Contact your Healthcare Provider at all other times 45 Solving Problems Table of Contents EvenCareG2 Solving Problems This section details the significant display screen messages and error codes you will encounter when using your EvenCare G2 meter and EvenCare G2 test strips What What the What it What Should Bfoorgee Tinetertays Removdikabaey Bod Meter electronic or turn on the meter again r q Failure test strip is Remove the test strip and grr it damaged insert a new test strip again e Ifthe problem persists please call 1 866 274 2131 Monday through Friday 8 00 a m 5 00 p m CST Contact your Healthcare Provider at all other times Error 3 Used or e Repeat the test with a new used test damaged test strip Remember to a strip test strip add sample only when err you see the prompt on the display screen or the meter verbally prompts you to add sample Error 4 Not enough Remove the Test Strip not sample on and repeat the test with E h enough the test strip a new test strip See LI 1 sample to start Testing Your Blood Glu cose on page 30 46 II l mg dL 10 18 i teow 47 Low Bata Tempera ture is out of range Your blood glucose reading is above 600 mg dL Your blood glucose reading is below 20 mg dL Batte
61. za dos en el medidor Table of Contents EvenCareG2 Qu debe hacer e Compruebe si la fecha y hora en su medidor est n configuradas Ver Confi uraci n de su Nuevo Si tema en la P gina 60 e Comience su prueba glucemia vea Probando su Glucemia en la p gina 78 96 EvenCareG2 Table of Contents Garant a del Producto Garant a del Producto El medidor de glucemia EvenCare G2 est garantizado libre de defectos de fabricaci n y materiales por un per odo de cinco 5 a os a partir de la fecha de compra La responsabilidad de Medline para reclamaciones de garant a est limitada a la reparaci n o reemplazo y en ning n caso Medline ser responsable de da os indirectos o conse cuentes o por cualquier p rdida que surja por el mal uso uso inapropi ado abuso descuido reparaciones no autorizadas o modificaci n Esta garant a ser nula y sin vigencia y efecto en el caso mal uso uso inapropiado abuso descuido reparaciones no autorizadas o modificaci n del producto Esta garant a excluye espec ficamente las tiras reactivas EvenCare G2 y la Soluci n de Control EvenCare G2 ESTA GARANT A ES LA NICA Y EXCLUSIVA GARANT A CON LA EXCLUSI N DE CUALQUIER OTRA GARANT A EXPRESA O IMPL CITA TODAS LAS CUALES SON ANUADAS Y RENUNCIADAS Todas las reclamaciones de garant a deben ser dirigidas al Centro de Servicio T cnico de Medline al 1 866 274 2131 Esta garant a se ex tiende nicamente al comprador