Home
LIFE-BASE 1 NG XS - WEINMANN Emergency
Contents
1. Description Order no MEDUCORE Standard Basic plus MEDUMAT Easy CPR WM 9740 MEDUCORE Standard Basic plus MEDUMAT Standard a WM 9745 MEDUMAT Standard WM 9400 MEDUMAT Standard plus MEDUCORE Standard WM 9405 LIFE BASE 1 NG XS with charger interface and MEDUMAT Standard WM 9870 MEDUCORE Standard Basic WM 9882 MEDUCORE Standard Advanced WM 9883 MEDUCORE Standard Pro WM 9880 9 3 Accessories and spare parts You can order accessories and spare parts separately if required A current list of accessories and spare parts can be ordered on the Internet at www weinmann emergency de or via your dealer Scope of supply EN 29 10 Technical data 30 LIFE BASE 1 NG Product class to 93 42 EEC Class Standards applied EN 1789 Dimensions without protective bag WxHxD in mm 415 x 240 x 140 LIFE BASE 1 NG 427 x 240 x 140 LIFE BASE 1 NG XL 220 x 230 x 136 LIFE BASE 1 NG XS Weight Empty LIFE BASE 1 NG approx 2 0 kg LIFE BASE 1 NG XL approx 2 2 kg LIFE BASE 1 NG XS approx 1 0 kg With equipment LIFE BASE 1 NG and LIFE BASE 1 NG XL max 7 0 kg LIFE BASE 1 NG XS max 5 0 kg Maximum load for carrying strap 5 kg Temperature range operation storage 18 C to 60 C 40 C to 70 C Note The temperature range may be limited by any devices fitted Electrical rating for charger interface 12 VDC Imax 8A C
2. El margen de te mperatura puede quedar limitado por aparatos montados Conexi n el ctrica para la interfaz de carga 12 VCC Imax 8A Reservado el derecho a modificaciones constructivas Datos t cnicos ES 61 11 Garant a WEINMANN Emergency otorga al comprador de un produc to original WEINMANN Emergency o una pieza de repuesto montada por WEINMANN Emergency una garant a del fabri cante limitada seg n las condiciones de garant a v lidas para los distintos productos y las duraciones de la garant a a partir de la fecha de compra que se indican a continuaci n Las condiciones de garant a se pueden cargar en Internet en www weinmann emergency de Si lo desea tambi n pode mos enviarle las condiciones de garant a En un caso de garant a dir jase a su distribuidor especializado Duraciones de la Producto garantia Aparatos de emergencia de WEINMANN Emergency incluidos los accesorios excepto 2 a os las m scaras para oxigenoterapia y medicina de urgencia M scaras incluidos accesorios acumulador bater as salvo que se indique otra cosa en la DREE 6 meses documentaci n t cnica sensores sistemas de tubos flexibles Productos para un solo uso Ninguno 62 ES Garant a 12 Declaraci n de conformidad Por la presente WEINMANN Emergency Medical Technology GmbH Co KG declara que el producto cumple las disposiciones pertinentes de la
3. Hang portable system Note If you are using a protective bag ensure that the protective bag is properly located first 1 To hang the portable system in the wall mounting place the portable system in the bottom rail Y Bottom rail Warning Risk of injury from loose portable system Control pin An inadequately secured portable system may fall off or be thrown around the vehicle Ensure that you both hear and feel the catch engage Check that the red control pin is flush with the top Catch Catch edge of the catch housing and does not protrude cross bar housing 2 Swivel the portable system towards the wall until you hear and feel the catch cross bar engage Operation EN 21 Remove portable system 1 For unlatching it is essential to grasp through the han dle so that the portable system does not fall out of the wall mounting and injure someone 2 To unlatch push the catch cross bar upwards 3 Swivel the portable system forwards a little way and lift the portable system up off the bottom rail 5 2 Hang portable system from stretcher or hospital bed LIFE BASE 1 NG LIFE BASE 1 NG XL You can also hang the portable system from a stretcher or a hospital bed as long as the rail diameter does not exceed 35 mm Caution Note that the portable system may not be loaded A N with an additional weight e g be leant on by a person even once it is fixed
4. The strap for the portable system can be attached to this lug 7 Adapter plate The devices can be attached to the portable system with the aid of the adapter plate Overview EN 5 Rear view LIFE BASE 1 NG Protective bag Key 1 Cable duct for charger interface 5 Side access compartment 2 Vibration dampers only with 6 Viewing window MEDUMAT Transport 7 Velcro tab to access the control panel 8 Reflective strip 9 Opening for passing through a patient tube system 10 Adapter plate 3 Hinge for bottom rail 4 Push button to fix protective bag in position 6 EN Overview 1 2 LIFE BASE 1 NG XS Front view 4 Handle 5 Charger interface 3 Lug for tensioning belt short 6 Lug for 2 Adapter plate tensioning belt long 1 Portable system Protective bag for defibrillator 8 Viewing window 9 Reflective strip 10 Side compartment 11 Velcro tape with ventilator only 13 Carrying strap 14 Tensioning belt short Rear view Re 7 Hinge for bottom rail Protective bag for ventilator 12 Loop for Velcro tape with ventilator only Tensioning belts 15 Tensioning belt long Overview EN 7 1 Portable system The portable system is for storing a device and its components 2 Adapter plate You can use the adapter plate to attach the de vice to the portable system 3 Lug for tensioning belt short You can attach a tension
5. 4 4 4 5 4 6 5 1 5 2 5 9 5 4 Vista general 34 LIFE BASE 1 NG LIFE BASE 1NGXL 34 LIFE BASE 1NGXS 37 Soporte mural 39 Se alizaci n especial en el aparato asocia al a 40 Indicaciones de seguridad en estas instrucciones 41 Descripci n del aparato 42 Uso conforme a lo dispuesto 42 Descripci n del funcionamiento 42 Indicaciones de seguridad 44 Disposiciones de seguridad 44 Montaje ooooooomoooo 46 Soporte mural 46 Aparato sin montar para sistemas de transporte con placa adaptadora 46 Bolsa de protecci n LIFE BASE 1 NG LIFE BASE 1 NG XL 47 Bolsa de protecci n LIFE BASE 1 NG XS 47 Bolsa de accesorios LIFE BASE 1 NG LIFE BASE 1 NG XL 49 Correa LIFE BASE 1 NG XS 50 Manejo oooococconoo o 51 Enganchar el sistema de transporte en el soporte mural 51 Enganchar el sistema de transporte a la camilla o la cama de hospital LIFE BASE 1 NG LIFE BASE 1 NG XL 52 Enganchar el sistema de transporte a la camilla o la cama de hospital LIFE BASE 1 NG XS Despu s del USO 53 6 1 6 2 6 3 7 1 7 2 1 3 8 1 9 1 9 2 9 3 10 11 12 Acondicionamiento higi nico Sistema de transporte Bolsa de protecci n y bolsa de accesorios Piezas montables Control
6. E The right to make design modifications is reserved EN Technical data 11 Warranty WEINMANN Emergency gives the customer a limited man ufacturer warranty on new genuine WEINMANN Emergency products and any replacement part fitted by WEINMANN Emergency in accordance with the warranty conditions applicable to the product in question and in ac cordance with the warranty periods from date of purchase as listed below The warranty conditions can be download ed from www weinmann emergency de on the Internet We can also send you the warranty conditions on request In the event of a claim under warranty contact your specialist dealer Product Warranty period WEINMANN Emergency devices including accessories exception masks for oxygen 2 years and emergency medicine Masks including accessories rechargeable battery batteries unless there is information to the contrary in the 6 months technical documentation sensors tube systems Disposable products None Warranty EN 31 12 Declaration of conformity WEINMANN Emergency Medical Technology GmbH Co KG declares herewith that the product complies fully with the respective regulations of the Medical Device Directive 93 42 EEC The unabridged text of the Declaration of Conformity can be found on our website at www weinmann emergency de 32 EN Declaration of conformity Indice 1 2 1 3 1 4 1 5 2 1 2 2 SA 4 1 4 2 4 3
7. Reparaci n Precauci n nicamente debe dejar que realicen las inspecciones y los trabajos de reparaci n el fabricante WEINMANN Emergency o el personal experto Indicaciones de seguridad ES 45 4 Montaje 4 1 Soporte mural El soporte mural para el sistema de transporte est confeccionado para su montaje a la pa red Con un juego de montaje adecuado ver cap tulo 9 3 en la p gina 60 puede montar tambi n el apoyo mural en un riel normalizado de hospital o en una barra 19 mm 38 mm Para el montaje siga las instrucciones de montaje correspondientes que se adjuntan tanto al apoyo mural como al correspondiente juego de montaje 4 2 Aparato sin montar para sistemas de transporte con placa adaptadora Nota En la mayor a de los casos el aparato ya est mon tado en el sistema de transporte S lo tendr que montar el aparato si pide un aparato por separado 1 Destornille la placa adaptadora del sistema de transporte 2 Atornille la placa adaptadora a la parte trasera del aparato 3 Atornille la placa adaptadora con el aparato en el sistema de transporte 46 ES Montaje 4 3 Bolsa de protecci n LIFE BASE 1 NG LIFE BASE 1 NG XL Fijar la bolsa de protecci n Coloque la bolsa de protecci n sobre una superficie plana y firme Ponga dentro el sistema de transporte sobre el fondo de la bolsa de protecci n Fije la bolsa de protecci n al sistema de transporte
8. enganchar el sistema de transporte mediante los ngulos de sujeci n a una camilla una cama de hospital o un riel normalizado en la cl nica En combinaci n con el soporte mural BASE STATION 1 NG o BASE STATION Mini Il el sis tema de transporte se puede fijar con seguridad en el mbito hospitalario o en un medio de salvamento v ase 9 3 Accesorios y piezas de repuesto en la p gina 60 El sistema de transporte tambi n puede fijarse al techo con el soporte de techo BASE STATION 1 NG Los soportes murales BASE STATION 1 NG y BASE STATION Mini ll est n confeccionados para la sujeci n mural El soporte mural se puede ampliar adem s con los siguientes juegos de montaje e Juego de montaje de riel normalizado de hospital BASE STATION 1 NG y BASE STATION Mini Il Fijaci n del soporte mural a un riel normalizado de hospital en veh culos o en el hospital 42 ES Descripci n del aparato e Juego de montaje para montaje en barra BASE STATION 1 NG Fijaci n del soporte mural a una barra 19 mm 38 mm en el hospital La barra puede estar orientada horizontal o vertical e Juego de montaje complementario para riel normalizado de hospital BASE STATION Mini Il Fijaci n del soporte mural a un riel normalizado de hospital en veh culos Puede retirar el sistema de transporte del soporte mural con un asidero Descripci n del aparato ES 43 3 Indicaciones de seguridad 3 1 Disposiciones de seguridad Lea c
9. in position The porta ble system may fall down and injure people or the device may be damaged 1 Undo the knurled screws 2 Extend the hanger the required distance 3 Tighten the knurled screws back up 4 Hang the portable system from a stretcher a hospital bed or a standard rail Note The portable system can only be hung back on a wall mounting once you have pushed the hanger back into its original position 22 EN Operation 5 3 Hang portable system from stretcher or hospital bed LIFE BASE 1 NG XS You can only hang the portable system from a stretcher or a hospital bed if you use a protective bag Caution Note that the portable system may not be loaded with an additional weight e g be leant on by a person even once it is fixed in position The porta ble system may fall down and injure people or the device may be damaged Open the magnetic closures on the straps on the rear of the protective bag sideways 2 Guide the straps round a rail of the stretcher or hospital bed 3 Close the magnetic closures 5 4 After use 1 Open the protective bag if present 2 If present Remove the pressure tube of the oxygen supply or stow it in the side compartment of the protective bag 3 If present Remove the patient tube system 4 Close the protective bag again if present Operation EN 23 6 Hygiene treatment The portable system must be cleaned and disinf
10. inferior 6 Pasador de control 7 Carcasa de enclavamiento 8 Eje de enclavamiento Vista general ES 39 1 4 Se alizaci n especial en el aparato Se alizaci n en el sistema de transporte N Simbolo Significado Letrero del aparato sistema de transporte al A o de fabricaci n N mero de serie del sistema de transporte C Marca CE confirma que el producto cumple las directivas europeas vigentes Sistema de transporte Es No elimine el sistema de transporte tir ndolo a la basura dom stica Se alizaci n en el sopo rte mural N Simbolo Significado Letrero del aparato soporte mural A o de fabricaci n N mero de serie del soporte mural sal SN No elimine el soporte mural tirandolo a la basura dom stica Soporte mural 4 Indica c mo puede liberar el enclavamiento y desprender el sistema de transporte del soporte mural 40 ES Vista general Indicaciones en la correa S mbolo Significado No planchar Lavar a 30 C El No secar en secadora KX vs Tener en cuenta el manual de instrucciones Carga maxima D gt Senalizacion en el embalaje Simbolo Significado a A i Temperatura admisible de almacenamiento 40 C a 70 C Humedad atmosf rica admisible de almacenamiento 95 maximo de 0 de hum rel 0 9
11. instructions for the wall mounting and the attachment set in question e Note that the portable system may not be loaded with an additional weight e g 14 EN Safety instructions be leant on by a person even once it is fixed in position The portable system may fall down and injure people or the device may be damaged Use of the wall mounting Warning e Only the portable systems LIFE BASE 1 NG LIFE BASE 1 NG XL A LIFE BASE 1 NG XS and LIFE BASE Mini Il should be suspended in the BASE STATION 1 NG and BASE STATION Mini Il wall mountings The weight and dimensions of these portable systems are exactly matched to the wall mountings e Hanging third party products or adapters from other manufacturers is not permitted as this does not comply with the use in accordance with purpose of this wall mounting Accessories replacement parts Caution If third party items are used functional failures may occur and fitness for use may be restricted Biocompatibility requirements may also not be met Please note that in such cases any claim under warranty and liability will be voided if neither the acces sories nor genuine replacement parts recommended in the instructions for use are used Maintenance Caution Have servicing and maintenance work carried out only by the manufacturer WEINMANN Emergency or by trained staff Safety instructions EN 15 4 Assembly 4 1 Wall mounting The wall mounting for the portable system is
12. mediante los botones a presi n Si se aloja un aparato de respiraci n en el sistema de transporte Pase el sistema de mangueras para pacien tes a trav s del orificio situado delante en la bolsa de protecci n Cierre los cierres de velcro en la parte delantera y trasera Quitar la bolsa de protecci n 1 2 3 Coloque el sistema de transporte sobre una superficie plana y firme Abra los cierres de velcro en la parte delantera Abra los botones a presi n situados sobre el sistema de transporte y saque el sistema de transporte hacia atr s de la bolsa de protecci n 4 4 Bolsa de protecci n LIFE BASE 1 NG XS 1 Coloque el sistema de transporte sobre una superficie plana y firme Retire todos los componentes del aparato Para el aparato de respiraci n artificial deslice la parte inferior de la bolsa de protecci n por delante en el hue co entre el sistema de transporte y el aparato de respi raci n artificial y pase la parte superior de la bolsa de protecci n por encima del sistema de transporte hasta Montaje ES 47 el asa de transporte o bien Para el desfibrilador deslice la bolsa de protecci n por delante y pas ndola por encima del sistema de trans porte 4 Cierre la cinta de velcro situada en la parte trasera del sistema de transporte 5 Coloque la bolsa de protecci n alrededor del asa de transporte y cierre las cintas de velc
13. 23 Hygiene treatment 24 Portable system 24 Protective bag and accessories bag 24 Attached parts 25 7 1 7 2 7 3 8 1 9 1 9 2 9 3 10 11 12 Function check 26 Visual inspection 26 Check Caters aci dira 26 Check protective bag if present 26 Servicing oooooommoo 27 Disposal eiii ead ba wee oe 27 Scope of supply 28 General information 28 Standard scope of supply 28 Accessories and spare parts 29 Technical data 30 Warranty 0c eee eee 31 Declaration of conformity 32 Contents EN 3 1 Overview 1 1 LIFE BASE 1 NG LIFE BASE 1 NG XL LIFE BASE 1 NG 4 Push button 5 Handle 6 Lug for strap 3 Knurled screw 2 Hanger 1 Portable system LIFE BASE 1 NG XL 6 Lug for strap 1 Portable system 2 Hanger 7 Adapter plate Key 1 Portable system 2 Hanger The portable system is for storing a combination The hanger is for hanging the portable system of ventilation and therapy devices and the neces from a hospital bed or stretcher sary components y P 3 Knurled screw The knurled screw is for fixing the hanger in position 4 EN Overview 4 Push button The push buttons are used to fix the protective bag to the portable system 5 Handle Use the handle to lift and carry the portable sys tem 6 Lug for strap
14. 5 humedad relativa 1 5 Indicaciones de seguridad en estas instrucciones En este manual de instrucciones se se alizan las indicaciones de seguridad de la siguiente forma Advertencia Advierte de la existencia de peligro de lesiones y posibles da os materiales Precauci n Advierte de da os materiales y posiblemente de resultados de terapia incorrectos Nota Contiene consejos tiles Vista general ES 41 2 Descripci n del aparato 2 1 Uso conforme a lo dispuesto El sistema de transporte sirve para alojar una combinaci n de aparatos de las series MEDUMAT MODUL y MEDUCORE as como los componentes necesarios interfaz de car ga tubos bolsa de accesorios y de protecci n cables el ctricos De este modo el usuario dispone de unidades m viles para el tratamiento de trastornos respiratorios el seguimiento de los par metros cardiovasculares y de asistencia en la reanimaci n Para una fijaci n segura el sistema de transporte se puede enclavar en los soportes murales correspondientes BASE STATION 1 NG y BASE STATION Mini Il Mediante diversos juegos de montaje la BASE STATION 1 NG se puede montar en diferen tes dispositivos de soporte en los medios de salvamento y en el mbito hospitalario 2 2 Descripci n del funcionamiento Todos los aparatos previstos est n montados de forma fija en el sistema de transporte De este modo el transporte y el manejo resultan m s f ciles Puede
15. ION 1 NG can be attached to various retaining devices in emergency vehicles and in hospitals using a variety of attachment sets 2 2 Description of function All the devices provided are securely mounted on the portable system making them easier for you to transport and operate Using the hangers you can hang the portable system from a stretcher a hospital bed or a standard hospital rail The portable system can be secured in a hospital environment or in an emergency vehicle in conjunction with the BASE STATION 1 NG or BASE STATION Mini Il wall mounting see 9 3 Accessories and spare parts on page 29 The portable system can also be secured to the ceiling in conjunction with the BASE STATION 1 NG roof mounting The BASE STATION 1 NG and BASE STATION Mini Il wall mountings are assembled for fixing to the wall You can also expand them by adding the following attachment sets e attachment set for standard hospital rail BASE STATION 1 NG and BASE STATION Mini Il attach the wall mounting to a standard hospital rail in vehicles or in hospital e attachment set for fitting to a rail BASE STATION 1 NG attach the wall mounting to a rail 19 mm 38 mm in hospital The rail can be horizontally or vertically aligned 12 EN Description of device e supplementary attachment set for standard hospital rail BASE STATION Mini II for attaching the wall mounting to a standard hospital rail in vehicles You can remove the portable system
16. You can use the carrying strap to carry the portable system on your shoulder 14 Tensioning belt short The tensioning belt holds the pressure tube of the ventilator if you are not using a protective bag 15 Tensioning belt long The tensioning belt holds the patient tube system of the ventilator if you are not using a protective bag 1 3 Wall mounting BASE STATION 1 NG Catch for wall mounting 6 Control pin 1 Catch 2 Bore 3 Guide pin 4 Charger interface 8 Catch cross bar 7 Catch housing 5 Bottom rail 1 Catch 2 Bore 3 Guide pin 4 Charger interface 5 Bottom rail 1 Catch 2 Bore 3 Guide pin 4 Charger interface 5 Bottom rail 6 Control pin 7 Catch housing 8 Catch cross bar Overview EN 1 4 Special markings on the device Markings on the portable system No Symbol Significance Information plate for porta ble system 1 sal Year of manufacture Serial number of portable system C CE symbol confirms that the product conforms to applicable European directives Portable system 2 X Do not dispose of the portable system in domestic waste a Markings on the wall mounting No Symbol Significance Information plate for wall mounting 3 Year of manufacture Serial number of wall mounting z Do not dispose of the wall mounting with domestic waste Wall mountin
17. a tensora 14 Correa tensora corta 15 Correa tensora larga Vista general ES 37 1 Sistema de transporte El sistema de transporte sirve para alojar un aparato y sus componentes 2 Placa adaptadora Puede fijar el aparato al sistema de transporte con la ayuda de la placa adaptadora 3 Ojal para la correa tensora corta Puede fijar a este ojal la correa tensora corta 4 Asa de transporte Con el asa de transporte puede levantar y llevar el sistema de transporte Puede fijar la correa al asa de transporte 5 Interfaz de carga Con la interfaz de carga puede cargar el aparato en el sistema de transporte 6 Ojal para la correa tensora larga Puede fijar a este ojal la correa tensora larga 7 Eje para el riel de sujeci n inferior Este eje sirve para colgar el sistema de transporte en el soporte mural 8 Mirilla Le permite leer la pantalla del dispositivo con la bolsa de protecci n cerrada 9 Tiras reflectantes La bolsa de protecci n tiene tiras reflectantes visibles en la oscuridad 38 ES Vista general 10 Compartimento lateral Puede almacenar los componentes y accesorios del aparato en los compartimentos laterales Para el aparato de respiraci n artificial el comparti mento de la izquierda permite acceder a la co nexi n de gas a presi n y ofrece espacio para el ubo de presi n y el compartimento de la dere cha permite acceder a la bater a y al filtro de en rada del aparato Pa
18. ar 4 4 Protective bag LIFE BASE 1 NG XS 1 Place the portable system on a firm level surface 2 Remove all the device components 3 On the ventilator push the underside of the protective bag into the space between the portable system and the ventilator from the front and pull the top of the pro tective bag over the portable system as far as the handle Alternatively on the defibrillator push the protective bag over the portable system from the front Assembly EN 17 Remove protective bag 18 EN Assembly 4 Close the Velcro tab on the rear of the portable system 5 Place the protective bag around the handle and close the Velcro tabs under and next to the handle 6 Connect all components to the device 7 On the defibrillator stow the electrodes under the defibrillator 8 Close the viewing window of the protective bag 9 On the ventilator clamp the patient tube system in the space between the portable system and the ventilator and fix it in position with the Velcro tabs on the protec tive bag 10 Stow the accessories and small parts in the side compartments of the protective bag 1 Place the portable system on a firm level surface 2 On the ventilator open the Velcro tapes on the patient tube system and pull it out of the space between the portable system and the ventilator 3 On the ventilator open the viewing window of the pro
19. del funcionamiento Control visual Comprobar el enclavamiento Comprobar la bolsa de protecci n si dispone de ella Mantenimiento EMI ACI cti See Volumen de suministro Indicaciones generales Volumen de suministro de serie Accesorios y piezas de repuesto Datos t cnicos Garant a Declaraci n de conformidad Indice ES 54 54 55 55 56 56 56 57 58 58 59 59 59 60 61 62 63 33 1 Vista general 1 1 LIFE BASE 1 NG LIFE BASE 1 NG XL LIFE BASE 1 NG 4 Bot n a presi n 5 Asa de transporte 6 Ojal para correa 3 Tornillo moleteado 2 ngulo de O 1 Sistema de sujeci n transporte 5 Asa de transporte 4 Bot n a presi n 3 Tornillo moleteado 6 Ojal para correa 1 Sistema de transporte 2 ngulo de sujeci n 7 Placa adaptadora Leyenda 1 Sistema de transporte 2 ngulo de sujeci n El sistema de transporte sirve para el alojamiento El ngulo de sujeci n sirve para enganchar el sis de una combinaci n de aparatos de respiraci n tema de transporte en una cama de hospital o una artificial y de terapia as como los componentes camilla discal 3 Tornillo moleteado El tornillo moleteado sirve para fijar el ngulo de sujeci n 34 ES Vista general 4 Bot n a presi n Con los botones a presi n puede fijar la bolsa de protecci n al sistema de transporte 5 Asa de transporte Con el asa de tra
20. directiva 93 42 CEE para productos sanitarios El texto completo de la declaraci n de conformidad se encuentra en www weinmann emergency de Declaraci n de conformidad ES 63 WEINMANN Emergency Medical Technology GmbH Co KG Frohboesestrasse 12 22525 Hamburg GERMANY E customerservice weinmann emt de www weinmann emergency de T 49 40 88 18 96 120 F 49 40 88 18 96 481 Center for Production Logistics Service WEINMANN Emergency Medical Technology GmbH Co KG Siebenstuecken 14 24558 Henstedt Ulzburg GERMANY n uu uu Ln oO Q mM i me Ke is Ko partner for life WEINM NN medical Mtechnology
21. e instructions for cleaning disinfecting and if appropriate sterilizing in the associated instructions for use Hygiene treatment EN 25 7 Function check You must subject the portable system to a function check before it is used each time and every time it has been dismantled but in any case at least every 6 months If you discover faults during the function check you may not use the portable system again until the faults have been rectified To guarantee that the portable system works perfectly observe the fol lowing points 7 1 Visual inspection Look at the portable system closely The portable system may not be damaged in any way which could result in parts coming loose which may restrict its function or which may cause injuries If you find any damage relating to one of these points the portable system must be re paired or replaced 7 2 Check catch 1 Hang the portable system in the wall mounting and allow it to engage The portable system must be located firmly in the wall mounting The red control pin of the catch cross bar must be flush with the top edge of the catch housing 2 Check unlatching next To do this reach through the handle and pull the catch cross bar upwards The catch cross bar must be easy to move The portable system may not catch on removal 7 3 Check protective bag if present Check push buttons Open and close the push buttons The push butto
22. ected after every use Always perform a function check afterwards see 7 Function check on page 26 6 1 Portable system The portable system is cleaned without its devices by the scouring and wiping method You can take all the devices off the portable system for this purpose e See the instructions for use for the disinfectant used e For disinfecting by wiping we recommend terralin protect e We recommend wearing suitable gloves e g household or disposable gloves for disinfecting Note For further information on hygiene treatment and a list of all cleaning agents and disinfectants which can be used please see our Internet brochure at www weinmann emergency de 6 2 Protective bag and accessories bag Subject the bags to hygiene treatment in accordance with the table below Bag Cleaning Disinfecting Sterilizing PVC protective bag In hot water with a mild Es s PVC accessories bag detergent Disinfectant wipe Not permitted CORDURA In hot water with a mild 30 C cycle in a washing rotective bag detergent machine no spin using a P e Alternatively suitable disinfecting additive Not permitted CORDURA E fe 30 C cycle in a washing disinfecting detergent or accessories bag A machine no spin impregnation agent Note To prevent unattractive washing powder deposits we recommend using a liquid detergent 24 EN Hygiene treatment 6 3 Attached parts Follow th
23. emergency LIFE BASE 1 NG LIFE BASE 1 NG XL LIFE BASE 1 NG XS Portable system Sistema de transporte Device description and instructions for use Descripci n del aparato e instrucciones de manejo WEINM NN medical Mtechnology Ica English Espa ol 34 Contents 4 3 4 4 4 5 4 6 5 1 5 2 5 3 5 4 6 1 6 2 6 3 Overview 0 cc cece eens 4 LIFE BASE 1 NG LIFE BASE 1 NG XL 4 LIFE BASE 1 NG XS 7 Wall mounting 9 Special markings on the device 10 Safety information in these INSTEUCHONS Ki ias cone Boh 1 Description of device 12 Intended use 2 Description of function 12 Safety instructions 14 Safety information 4 Assembly 000eeeeee 16 Wall mounting 6 Device not mounted in the case of portable systems with an adapter plate seen aie id 16 Protective bag LIFE BASE 1 NG LIFE BASE 1 NG XL 16 Protective bag LIFE BASE 1 NG XS 17 Accessories bag LIFE BASE 1 NG LIFE BASE 1 NG XL 19 LIFE BASE 1 NG XS carrying strap 20 Operation oooooo o 21 Hang portable system in wall Mounting 0 eee eee 21 Hang portable system from stretcher or hospital bed LIFE BASE 1 NG LIFE BASE 1 NG XL 22 Hang portable system from stretcher or hospital bed LIFE BASE 1 NG XS 23 AFET A ides s
24. from the wall mounting in just one maneuver Description of device EN 13 3 Safety instructions 3 1 Safety information Read these instructions for use carefully It is a component part of the following portable systems LIFE BASE 1 NG LIFE BASE 1 NG XL LIFE BASE 1 NG XS The instructions for use must be available at all times Use the portable system exclusively for the intended purpose described see 2 1 Intended use on page 12 For your own safety and the safety of your patients and in accordance with the requirements of Directive 93 42 EEC please note the following Use of the portable system Warning Use the portable system only if you are medically trained and have received instruc tion in ventilation aspiration Severe physical damage may be caused by incorrect use Caution e Before working with the portable system you must have understood how to use it e To avoid infection or bacterial contamination follow the section entitled 6 Hygiene treatment on page 24 e Do not use the portable system in the vicinity of a magnetic resonance tomography MRT device e Use the portable system only for the intended use described see 2 1 Intended use on page 12 e During the journey flight the portable system must be hung in the wall mounting intended for it as otherwise it might fall over or be thrown around the vehicle The wall mounting must be securely attached Follow the fitting
25. g 4 Shows how the catch unlatches and the portable system is removed from the wall mounting 10 EN Overview Markings on the carrying strap Symbol Significance Do not iron Wash at 30 C ex Do not dry in a tumble dryer KX vs Follow instructions for use Maximum load D gt 2 Markings on the packaging Symbol Significance A Permitted temperature for storage 40 C to 70 C rh 0 95 Permitted humidity for storage max 95 relative humidity 1 5 Safety information in these instructions Safety information in these instructions for use is marked as follows Warning war of risk of injury and potential material damage Caution Warns of material damage and potentially false therapy results Note Contains useful tips Overview EN 11 2 Description of device 2 1 Intended use The portable system is for storing a combination of devices from the MEDUMAT MODUL and MEDUCORE series together with the required components charger interface tubes accessories bag protective bag and electrical cables This provides a user with mobile units for treating respiratory disorders for monitoring cardiovascular parameters and for resus citation support The portable system can be latched in the associated BASE STATION 1 NG and BASE STATION Mini Il wall mountings to fix it securely to the wall The BASE STAT
26. iento inferior 5 2 Enganchar el sistema de transporte a la camilla o la cama de hospital LIFE BASE 1 NG LIFE BASE 1 NG XL 52 ES Manejo Tambi n puede enganchar el sistema de transporte a una camilla o una cama de hospital El di metro del tubo no debe ser mayor de 35 mm BR w N Precauci n Observe que el sistema de transporte incluso cuando est fijado no se puede cargar con un peso adicional p ej apoyando el cuerpo sobre l El sis tema de transporte podr a caer y lesionar a perso nas o se podr a da ar el aparato Afloje los tornillos moleteados Extraiga el ngulo de sujeci n a la distancia necesaria Apriete de nuevo los tornillos moleteados Enganche el sistema de transporte a una camilla una cama de hospital o un riel normalizado Nota S lo podr volver a enganchar el sistema de trans porte en un soporte mural cuando que haya de vuelto a su posici n original el ngulo de sujeci n 5 3 Enganchar el sistema de transporte a la camilla o la cama de hospital LIFE BASE 1 NG XS Si utiliza una bolsa de protecci n s lo puede enganchar el sistema de transporte a una camilla o cama de hospital Precauci n Observe que el sistema de transporte incluso cuando est fijado no se puede cargar con un peso adicional p ej apoyando el cuerpo sobre l El sis tema de transporte podr a caer y lesionar a perso nas O se podr a da ar el aparato 1 Los cie
27. ing belt short to this lug 4 Handle The portable system can be lifted and carried by the handle You can attach the carrying strap to the handle 5 Charger interface You can charge the device on the portable system using the charger interface 6 Lug for tensioning belt long You can attach a tensioning belt long to this lug 7 Hinge for bottom rail This hinge is used to hang the portable system in the wall mounting 8 Viewing window This is where to read off the device display when the protective bag is closed 9 Reflective strip The reflective strip means the protective bag is also visible in the dark D 8 EN Overview 10 Side compartment You can store the components and accessories of the device in the side compartments With the ventilator the left hand side compartment pro vides access to the compressed gas connection and space for the pressure tube and the right hand side compartment provides access to the device inlet filter With the defibrillator the left hand side compartment provides access to the re chargeable battery and the right hand side com partment provides space for components such as cables for example 11 Velcro tape with ventilator only The Velcro tapes are used to attach the patient tube system to the protective bag 12 Loop for Velcro tape with ventilator only The Velcro tapes can be attached to these loops which hold the patient tube system 13 Carrying strap
28. irecci n a su delegado de medio ambiente o en su ayuntamiento El embalaje del aparato el cart n y las piezas interiores puede eliminarlo en el reci claje de papel 58 ES Mantenimiento 9 Volumen de suministro 9 1 Indicaciones generales Para el sistema de transporte es aplicable e Elsistema de transporte se suministra completamente montado seg n la variante de equipamiento e Cada sistema de transporte se suministra con estas instrucciones de manejo Los soportes murales BASE STATION 1 NG y BASE STATION Mini ll no se incluyen en el vo lumen de suministro y en caso de necesitarse deben solicitarse por separado v ase 9 3 Accesorios y piezas de repuesto en la p gina 60 9 2 Volumen de suministro de serie Descripci n to ds pedido LIFE BASE 1 NG sin interfaz de carga y MEDUMAT Transporte sin medici n de CO WM 9630 MEDUMAT Transporte con medici n de CO WM 9635 LIFE BASE 1 NG con interfaz de carga y MEDUMAT Transporte sin medici n de CO WM 9620 MEDUMAT Transporte con medici n de CO WM 9625 LIFE BASE 1 NG XL con interfaz de carga y MEDUCORE Standard Advanced WM 9720 MEDUCORE Standard Advanced y MEDUMAT Easy CPR WM 9725 MEDUCORE Standard Advanced y MEDUMAT Standard a WM 9730 MEDUCORE Standard Pro WM 9755 MEDUCORE Standard Pro y MEDUMAT Easy CPR WM 9760 MEDUCORE Standard Pro y MEDUMAT Standard a WM 9765 Volumen de suminis
29. los cierres magn ticos Desbloquee y bloquee los cierres magn ticos Los cierres magn ticos no se deben abrir de nuevo Control del funcionamiento ES 57 8 Mantenimiento El sistema de transporte y el soporte mural no necesitan mantenimiento alguno Sin em bargo observe los plazos del control peri dico del funcionamiento para las piezas acopla das En caso necesario consulte las instrucciones de uso correspondientes Le aconsejamos que encargue las tareas de mantenimiento tales como inspecciones y tra bajos de reparaci n al fabricante WEINMANN Emergency o a personal experto 8 1 Eliminaci n LIFE BASE 1 NG sin interfaz de carga BASE STATION 1 NG sin interfaz de carga BASE STATION Mini II sin interfaz de carga No elimine el aparato tir ndolo a la basura dom stica Para la eliminaci n correcta del apa rato dir jase a una empresa de eliminaci n de residuos autorizada y certificada Pregunte la direcci n a su delegado de medio ambiente o en su ayuntamiento El embalaje del apa rato el cart n y las piezas interiores puede eliminarlo en el reciclaje de papel LIFE BASE 1 NG LIFE BASE 1 NG XL LIFE BASE 1 NG XS con interfaz de carga BASE STATION 1 NG con interfaz de carga BASE STATION Mini II con interfaz de carga No elimine el aparato tir ndolo a la basura dom stica Para la eliminaci n correcta X del aparato dir jase a una empresa de reciclaje de electr nica autorizada y certifica da Pregunte la d
30. ncionamiento del sistema de transporte observe los puntos siguientes 7 1 Control visual Examine detalladamente el sistema de transporte El sistema de transporte no debe presentar ning n da o como consecuencia del cual se pudieran desprender elementos que puedan limitar el funcionamiento o que pudieran causar lesiones En caso de que detecte defectos en los cuales fuera aplicable alguno de los puntos citados es necesario que repare o cambie el sistema de transporte 7 2 Comprobar el enclavamiento 1 Enganche el sistema de transporte en el soporte mural y enc jelo El sistema de transporte debe estar sujeto firmemente en el soporte mural El pasador de control rojo del eje de enclavamiento tiene que estar enrasado con la parte superior de la carcasa de enclavamiento 2 Compruebe a continuaci n el desenclavamiento Para ello pase la mano por el asidero y tire del eje de enclavamiento hacia arriba El eje de enclavamiento debe poder accionarse con facilidad El sistema de transporte no debe quedar bloqueada al retirarlo 56 ES Control del funcionamiento 7 3 Comprobar la bolsa de protecci n si dispone de ella Comprobar los botones a presi n Abra y cierre los botones a presi n Los botones a presi n no se deben abrir de nuevo Comprobar los cierres de velcro Abra los cierres de velcro La cinta afelpada y la cinta de ganchitos deben estar libres de suciedad Control de
31. ns may not pop open again 26 EN Function check Check Velcro tabs Open the Velcro tabs Both types of tape should be free of contamination Check magnetic closures Open and close the magnetic closures The magnetic closures may not pop open again 8 Servicing The portable system and the wall mounting require no servicing but observe the intervals for regular function checks on the attached parts Look these up in the associated instruc tions for use if appropriate We recommend having maintenance measures such as servicing and repair work carried out by the manufacturer WEINMANN Emergency or by specialist staff 8 1 Disposal LIFE BASE 1 NG without charger interface BASE STATION 1 NG without charger interface BASE STATION Mini II without charger interface Do not dispose of the device in domestic waste To dispose of the device properly contact a licensed certified disposal company You can find an address from your en vironment officer or from your municipal authorities The device packaging card board and inserts can be disposed of in paper recycling facilities LIFE BASE 1 NG LIFE BASE 1 NG XL LIFE BASE 1 NG XS with charger interface BASE STATION 1 NG with charger interface BASE STATION Mini II with charger interface Do not dispose of the device in domestic waste To dispose of the device properly X contact a licensed certified electronic scrap disposal merchant You can find an ad dress from you
32. nsporte puede levantar y llevar el sistema de transporte 6 Ojal para correa En este ojal puede fijar la correa para el sistema de transporte 7 Placa adaptadora Los aparatos pueden fijarse al sistema de transporte con la ayuda de la placa adaptadora Vista general ES 35 Vista posterior LIFE BASE 1 NG Bolsa de protecci n Leyenda 1 Paso de cable para la interfaz de carga 2 Amortiguadores exclusivamente en MEDUMAT Transport 3 Eje para la gu a de asiento inferior 4 Bot n a presi n para fijar la bolsa de protecci n 36 ES Vista general 5 Compartimento de acceso lateral 6 Mirilla 7 Cierre de velcro para abrir el panel de mando frontal 8 Tiras reflectantes 9 Orificio para el paso de un sistema de mangueras para pacientes 10 Placa adaptadora 1 2 LIFE BASE 1 NG XS Vista frontal Vista posterior 4 Asa de transporte 3 Ojal para la correa tensora corta 5 Interfaz de carga 6 Ojal para la 2 Placa adaptadora correa tenso ra larga 1 Sistema de transporte 7 Eje para el riel de sujeci n inferior Bolsa de protecci n Bolsa de protecci n para para el desfibrilador el aparato de respiraci n artificial 8 Mirilla 9 Tiras reflectantes 10 Compartimento lateral 11 Cinta de velcro s lo en el aparato de respiraci n artificial 12 Lazo para la cinta de velcro s lo en el aparato de respiraci Correa n artificial 13 Correa Corre
33. on atenci n estas instrucciones de uso Forman parte de los siguientes sistemas de transporte LIFE BASE 1 NG LIFE BASE 1 NG XL LIFE BASE 1 NG XS Las instrucciones de uso deben estar disponibles en todo momento Utilice el sistema de transporte exclusivamente para el fin descrito v ase 2 1 Uso conforme a lo dispuesto en la p gina 42 Por su propia seguridad personal as como por la de sus pacientes y para cumplir con los requerimientos estipulados en la Directiva 93 42 CEE observe lo siguiente Uso del sistema de transporte Advertencia uN Utilice el sistema de transporte s lo si cuenta con formaci n m dica y ha recibido ins trucci n en la t cnica de respiraci n artificial y de aspiraci n Su utilizaci n incorrecta puede provocar lesiones personales graves Precauci n e Antes de trabajar con el sistema de transporte debe haber comprendido su manejo e Para evitar una infecci n o una contaminaci n bacteriana tenga en cuenta el apartado 6 Acondicionamiento higi nico en la p gina 54 e No utilice el sistema de transporte en la proximidad de un equipo de tomograf a por resonancia magn tica MRT e Utilice el sistema de transporte nicamente para la finalidad de uso descrita v ase 2 1 Uso conforme a lo dispuesto en la p gina 42 e Durante el viaje vuelo el sistema de transporte se tiene que enganchar en el soporte mural previsto al efecto de lo contrario podr a
34. porte mural Enganchar el sistema de transporte Nota Si utiliza una bolsa de protecci n antes de la utili zaci n preste atenci n a que la bolsa de protecci n est correctamente sujeta 1 Para enganchar el sistema de transporte en el soporte mural coloque el sistema de transporte en la gu a de asiento inferior Advertencia Peligro de lesiones por sistema de transporte Gu a de asiento E g p P i suelto Un sistema de transporte que no est suficientemente fijado puede caer o ser proyectado por el veh culo inferior Pasador de control Aseg rese de que el enclavamiento quede encajado de forma audible y perceptible Compruebe que el pasador de control rojo est Eje de Carcasa de enrasado con la parte superior de la carcasa de enclava enclavamiento enclavamiento y no sobresale miento 2 Gire el sistema de transporte en direcci n a la pared hasta que el eje de enclavamiento quede encajado de forma audible y visible Manejo ES 51 Sacar el sistema de transporte 1 Al desenclavar el sistema es absolutamente necesario pasar la mano a trav s del asidero con el fin de evitar que el sistema de transporte pueda caerse del soporte mural y causar lesiones a alguien Para desbloquear presione el eje de enclavamiento hacia arriba Gire el sistema de transporte ligeramente hacia delante y s quelo hacia arriba de la gu a de as
35. r environment officer or from your municipal authorities The device packaging cardboard and inserts can be disposed of in paper recycling facilities Servicing EN 27 9 Scope of supply 9 1 General information The following applies to the portable system e The portable system is supplied fully assembled depending on the equip ment variant e Each portable system is supplied with these instructions for use The BASE STATION 1 NG and BASE STATION Mini Il wall mountings are not included in the scope of supply and should be ordered separately if required see 9 3 Accessories and spare parts on page 29 9 2 Standard scope of supply Description Order no LIFE BASE 1 NG without charger interface and MEDUMAT Transport without CO measurement WM 9630 MEDUMAT Transport with CO measurement WM 9635 LIFE BASE 1 NG with charger interface and MEDUMAT Transport without CO measurement WM 9620 MEDUMAT Transport with CO measurement WM 9625 LIFE BASE 1 NG XL with charger interface and MEDUCORE Standard Advanced WM 9720 MEDUCORE Standard Advanced plus MEDUMAT Easy CPR WM 9725 MEDUCORE Standard Advanced plus MEDUMAT Standard a WM 9730 MEDUCORE Standard Pro WM 9755 MEDUCORE Standard Pro plus MEDUMAT Easy CPR WM 9760 MEDUCORE Standard Pro plus MEDUMAT Standard a WM 9765 MEDUCORE Standard Basic WM 9735 28 EN Scope of supply
36. ra el desfibrilador el compartimento de la izquierda permite acceder a a bater a y el compartimento de la derecha ofrece espacio para los componentes como p ej cables 11 Cinta de velcro s lo en el aparato de respiraci n artificial Las cintas de velcro fijan el sistema de tubos flexibles para el paciente a la bolsa de protecci n 12 Lazo para la cinta de velcro s lo en el aparato de respiraci n artificial Las cintas de velcro que sujetan el sistema de tubos flexibles para el paciente pueden fijarse a estos lazos 13 Correa Con la correa puede llevar el sistema de ransporte al hombro 14 Correa tensora corta La correa tensora sujeta el tubo de presi n del aparato de respiraci n artificial en caso de que no utilice una bolsa de protecci n 15 Correa tensora larga La correa tensora sujeta el sistema de tubos flexi bles para el paciente del aparato de respiraci n artificial en caso de que no utilice una bolsa de protecci n 1 3 Soporte mural BASE STATION 1 NG 1 Bloqueo 2 Agujero Bloqueo soporte mural 6 Pasador de control 3 Pasador de gu a 4 Interfaz de carga 8 Eje de 7 Carcasa de enclavamiento enclavamiento 5 Gu a de asiento inferior 1 Bloqueo 2 Agujero 3 Pasador de gu a 4 Interfaz de carga 5 Gu a de asiento inferior 1 Bloqueo 2 Agujero 3 Pasador de gu a 4 Interfaz de carga 5 Gu a de asiento
37. ready for fitting to a wall Using a suitable attachment set see Section 9 3 on page 29 you can also fit the wall mounting to a stan dard hospital rail or an ordinary rail 19 mm 38 mm When fitting follow the relevant fitting instructions enclosed with the wall mounting and the attachment sets in question 4 2 Device not mounted in the case of portable systems with an adapter plate Note In most cases the device is already mounted on the portable system You only need to mount the device if you order a device separately 1 Unscrew the adapter plate from the portable system 2 Screw the adapter plate onto the rear of the device 3 Screw the adapter plate and the device to the portable system 4 3 Protective bag LIFE BASE 1 NG LIFE BASE 1 NG XL Attach protective bag 1 Place the protective bag on a firm level surface 2 Place the portable system inside on the base of the protective bag 3 Attach the protective bag to the portable system using the push buttons 16 EN Assembly If there is a ventilator on the portable system Route the patient tube system through the front opening in the protective bag Close the velcro tabs on the front and rear Place the portable system on a firm level surface Open the Velcro tabs on the front Open the push buttons on the portable system and re move the portable system from the protective bag from the re
38. rios LIFE BASE 1 NG LIFE BASE 1 NG XL Hay cuatro tipos de bolsas de accesorios bolsa de accesorios para aparatos de respiraci n artificial montada lateralmente bolsa de accesorios para desfibrilaci n respiraci n montada lateralmente bolsa de accesorios para desfibrilaci n montada en el sistema de transporte Montaje ES 49 Montar bolsa de accesorios para aparatos de respiraci n artificial 1 Abra los cierres de velcro en la parte trasera de la bolsa de accesorios 2 Pase la cinta de velcro a trav s del lazo en la bolsa de protecci n 3 Cierre los cierres de velcro en la parte trasera de la bolsa de accesorios Montar bolsa de accesorios para desfibrilaci n respiraci n 1 Enganche la cremallera de la bolsa de accesorios en la cremallera de la bolsa de protecci n 2 Cierre la cremallera Nota la bolsa de accesorios para la desfibrilaci n se suje ta a la izquierda de la bolsa de protecci n la bolsa de accesorios para la respiraci n a la derecha de la bolsa de protecci n Montar bolsa de accesorios para la desfibrilaci n 1 Presione la bolsa de accesorios contra la pared posterior del sistema de transportes 4 6 Correa LIFE BASE 1 NG XS 1 Coloque la cinta de velcro de la correa alrededor del asa de transporte y ci rrela 50 ES Montaje 5 Manejo 5 1 Enganchar el sistema de transporte en el so
39. ro de debajo y al lado del asa 6 Conecte todos los componentes al aparato 7 Para el desfibrilador guarde los electrodos en la parte inferior del desfibrilador 8 Cierre la mirilla de la bolsa de protecci n 9 Para el aparato de respiraci n artificial fije el sistema de tubos flexibles para el paciente en el hueco entre el sis tema de transporte y el aparato de respiraci n artificial y suj telo a la bolsa de protecci n con las cintas de velcro 10 Guarde los accesorios y las piezas peque as en los compartimentos laterales de la bolsa de protecci n Retirar la bolsa de protecci n 1 Coloque el sistema de transporte sobre una superficie plana y firme 2 Para el aparato de respiraci n artificial abra las cintas de velcro del sistema de tubos flexibles para el paciente y s quelo del hueco entre el sistema de transporte y el aparato de respiraci n artificial 3 Para el aparato de respiraci n artificial abra la mirilla de la bolsa de protecci n 48 ES Montaje 4 Para el aparato de respiraci n artificial retire el sistema de tubos flexibles para el paciente 5 Abra la cinta de velcro situada en la parte trasera del sistema de transporte 6 Abra los cierres de velcro de debajo y al lado del asa de transporte y retire la bolsa de protecci n del asa 7 Tire del sistema de transporte hacia atr s para retirarlo de la bolsa de protecci n 4 5 Bolsa de acceso
40. rres magn ticos de las correas situados en la parte posterior de la bolsa de protecci n se abren hacia los lados 2 Colocar la correa alrededor de una barra de la camilla o cama de hospital 3 Bloquear los cierres magn ticos 5 4 Despu s del uso 1 Abra la bolsa de protecci n si dispone de ella 2 Si dispone de l Retire el tubo de presi n del suministro de ox geno o gu rdelo en el compartimento lateral de la bolsa de protecci n Si dispone de l Retire el sistema de mangueras para pacientes 4 Cierre la bolsa de protecci n si dispone de ella Manejo ES 53 6 Acondicionamiento higi nico Despu s de cada uso es preciso limpiar y desinfectar el sistema de transporte A continuaci n debe llevar a cabo un control del funcionamiento v ase 7 Control del funcionamiento en la p gina 56 6 1 Sistema de transporte El sistema de transporte sin aparatos se limpia seg n el procedimiento de fregado y frotamiento Para este fin puede quitar todos los aparatos del sistema de transporte e Tenga en cuenta el manual de instrucciones del agente desinfectante utilizado e Para el frotado de desinfecci n recomendamos terralin protect e Recomendamos llevar guantes adecuados p ej guantes para fregar o guantes desechables para la desinfecci n Nota Encontrar m s informaci n sobre el tratamiento higi nico y una lista de todos los productos limpiadores y agentes de
41. sinfectantes utilizables en un folleto en nuestra p gina web en www weinmann emergency de 54 ES Acondicionamiento higi nico 6 2 Bolsa de protecci n y bolsa de accesorios Prepare las bolas seg n la siguiente tabla CORDURA Bolsa de accesorios ciclo de aclarado a 30 C sin centrifugar en la lavadora Bolsa Limpieza Desinfecci n Esterilizaci n Bolsa de protecci n PVE aa calientecom un Desinfecci n por frotamiento No permitido Bolsa de accesorios detergente suave PVC 5 Eau caliente cont Ciclo de aclarado a 30 C sin Bolsa de protecci n detergente suave centrifugar en la lavadora e CORDURA o bien a ada un desinfectante No permitido adecuado detergente desinfectante o impregnante Nota Para evitar dep sitos antiest ticos de detergente en polvo recomendamos utilizar detergente l quido 6 3 Piezas montables Observe las indicaciones para la limpieza la desinfecci n y en caso necesario la esterilizaci n en las instrucciones de uso correspondientes Acondicionamiento higi nico ES 55 7 Control del funcionamiento El sistema de transporte se tiene que someter a un control de funcionamiento antes de cada uso y despu s de cada desmontaje pero como m nimo cada 6 meses Si se descubrie ran anomal as en el control de funcionamiento no debe utilizar el sistema de transporte hasta que hayan sido eliminadas las anomal as Para garantizar el correcto fu
42. tective bag 4 On the ventilator detach the patient tube system 5 Open the Velcro tab on the rear of the portable system 6 Take the portable system out of the protective bag backwards 4 5 Accessories bag LIFE BASE 1 NG LIFE BASE 1 NG XL There are four types of accessories bag accessories bag for ventilators fitted on the side accessories bag for defibrillation ventilation fitted on the side accessories bag for defibrillation fitted on the portable system Fit accessories bag for ventilators 1 Open the Velcro tabs on the rear of the accessories bag 2 Pass the Velcro tabs through the loops on the protective bag 3 Close the Velcro tabs on the rear of the accessories bag Assembly EN 19 Fit accessories bag for defibrillation ventilation 1 Hook the zip on the accessories bag into the zip on the protective bag 2 Close the zip Note The accessories bag for defibrillation is attached to the left of the protective bag the accessories bag for ventilation to the right of the protective bag Fit accessories bag for defibrillation 1 Press the accessories bag against the rear wall of the portable system 4 6 LIFE BASE 1 NG XS carrying strap 1 Place the Velcro tab of the carrying strap around the handle and close it 20 EN Assembly 5 Operation 5 1 Hang portable system in wall mounting
43. tro ES 59 N mero de Descripci n pedido MEDUCORE Standard Basic WM 9735 MEDUCORE Standard Basic y MEDUMAT Easy CPR WM 9740 MEDUCORE Standard Basic y MEDUMAT Standard a WM 9745 MEDUMAT Standard WM 9400 MEDUMAT Standard y MEDUCORE Standard WM 9405 LIFE BASE 1 NG XS con interfaz de carga y MEDUMAT Standard WM 9870 MEDUCORE Standard Basic WM 9882 MEDUCORE Standard Advanced WM 9883 MEDUCORE Standard Pro WM 9880 9 3 En caso necesario puede pedir accesorios y piezas de repuesto por separado Puede obtener Accesorios y piezas de repuesto una lista actualizada de los accesorios y las piezas de repuesto en Internet en www weinmann emergency de o a trav s de su distribuidor especializado 60 ES Volumen de suministro 10 Datos t cnicos LIFE BASE 1 NG Clase de aparato seg n 93 42 CEE Clase Normas aplicadas EN 1789 Dimensiones sin bolsa de protecci n LXAxA en mm 415 x 240 x 140 LIFE BASE 1 NG 427 x 240 x 140 LIFE BASE 1 NG XL 220 x 230 x 136 LIFE BASE 1 NG XS Peso Vac o LIFE BASE 1 NG aprox 2 0 kg LIFE BASE 1 NG XL aprox 2 2 kg LIFE BASE 1 NG XS aprox 1 0 kg Con equipamiento LIFE BASE 1 NG y LIFE BASE 1 NG XL m x 7 0 kg LIFE BASE 1 NG XS m x 5 0 kg Correa de carga m xima 5 kg Margen de temperatura Funcionamiento Almacenamiento 18 C a 60 C 40 C a 70 C Nota
44. volcar o ser proyectado por el veh culo El soporte mural debe estar fijado con seguridad Siga las instrucciones de montaje del soporte mural y del correspondiente juego de montaje 44 ES Indicaciones de seguridad e Observe que el sistema de transporte incluso cuando est fijado no se puede cargar con un peso adicional p ej apoyando el cuerpo sobre l El sistema de transporte podr a caer y lesionar a personas o se podr a da ar el aparato Uso del soporte mural Advertencia e Los sistemas de transporte LIFE BASE 1 NG LIFE BASE 1 NG XL A N LIFE BASE 1 NG XS y LIFE BASE Mini Il solo pueden engancharse en los soportes murales BASE STATION 1 NG y BASE STATION Mini Il El peso y las dimensiones de estos sistemas de transporte est n adaptados exactamente al soporte mural Por lo tanto no se permite utilizar otros sistemas de transporte e Nose permite enganchar productos o adaptadores de otros fabricantes dado que esto no corresponde al uso conforme a lo dispuesto de este soporte mural Accesorios piezas de repuesto Precauci n Si se utilizan art culos de otros fabricantes pueden darse fallos en el funcionamiento y limitarse la aptitud para el uso Adem s pueden no cumplirse los requerimientos de biocompatibilidad Tenga en cuenta que en estos casos perder cualquier derecho de garant a o de indemnizaci n si no utiliza los accesorios recomendados en el manual de instrucciones ni las piezas de repuesto originales