1. Please put the microscope on top of a horizontal plane and stable table making a free space at least at 30 cm per side Do not put the microscope near any warm supply bur ners blowlamps nor expose it directly to the sun etc Avoid inflammable or toxic substances in the working area The micrsocope is supplied with a Schuko standard wire Please insert the wire that feeds the AC electric current in the base of current 220V 50Hz 10 provided with earth wire and the other extreme to the micrsocope connector Neither the manufacturer nor the distributor will assume any responsibility for the damages produced to the equipment during its installation or damages to persons suffe red by the improper use of the electric connection The tension should be 220V 50Hz 10 If you are not using the microscope for a long period of time please make sure it is dis connected from the net and protected from dust this way you will avoid accidents and will extend its working life Setting up The observer should be in a comfortable position properly sit down and with the back in a straight position It is advisable to work on a dark table in order to avoid the light that dazzles and reduces the good definition of the images thus avoiding ludicrous fati gue Pay attention to background lighting and install the microscope away from windows Take down the mechanical stage by means of the macrometric coarse controls Screw the obj
2. ture of a microscope objective is a measure of its ability to gather light and resolve fine sample detail at a fixed object distance Thus the bigger it is the brighter and better resol ved the image will be Working distance It is the distance mm between the sampe specimen and the frontal lens of the objective when the microscope is focused Focal distance It is the distance mm between the main plan image of the system and its image focus Resolution It is the reciprocal value of the separating power that is to say the smallest distance between two points on a sample that can still be distinguished as two separate entities It is calculated in lines mm Field s number It expresses the diameter mm of the field diaphragm composed by the eyepiece Field of view Dimensions mm of the real field we are observing 4 INSTALLATION SETTING UP Preliminary inspection Unwrap the microscope take off the involving plastic and take off the poliespan pro tection in which it comes fitted Take off all the protective items and without connecting the microscope to the net make sure that it does not present any damage because of the shipment In case the microscope presents any damage tell it immediately to your trans port agent or dealer so that they can make the claims in the correct time limit Please keep the original wrapping you will always need it for returns enclosed with all the accessories supplied Please
3. Este manual es parte inseparable del microscopio profesional serie 120 Zuzi por lo que debe estar disponible para todos los usuarios del equipo Debe manipularse siempre con cuidado evitando los movimientos bruscos golpes ca das de objetos pesados o punzantes evitar el derrame de l quidos en su interior Nunca desmonte el equipo para repararlo usted mismo adem s de perder la garant a podr a producir un funcionamiento deficiente de todo el equipo as como da os a las per sonas que lo manipulan Para prevenir fuego o descargas el ctricas evite los ambientes secos y polvorientos Si esto ocurre desenchufar inmediatamente el equipo de la toma de corriente Cualquier duda puede ser aclarada por su distribuidor instalaci n puesta en marcha funcionamiento Usted puede tambi n mandarnos sus dudas o sugerencias a la siguien te direcci n de correo del Servicio T cnico Zuzi asistenciaQauxilab com o bien llamando al Tel 807117040 0 30Euros min Este equipo est amparado por la Ley de garant as y bienes de consumo 10 2003 No se consideran en garant a las revisiones del equipo La manipulaci n del equipo por personal no autorizado provocar la p rdida total de la garant a Los fusibles o accesorios as como la p rdida de los mismos no est n cubiertos por dicha garant a Tampoco estar n cubiertos por el periodo de garant a las piezas en su des gaste por uso natural Aseg rese de guardar la factura de
4. del campo real que estamos observando 4 INSTALACI N PUESTA EN MARCHA Inspecci n preliminar Desembale el microscopio retire el pl stico que lo envuelve y quite la protecci n de poliesp n en que viene encajada Retire todas las protecciones y sin conectar el microsco pio a la red el ctrica aseg rese de que no presenta ning n da o debido al transporte De ser asi comuniquelo inmediatamente a su transportista o suministrador Guarde el embalaje ya que siempre se deben realizar las devoluciones en su embalaje original con todos los accesorios suministrados Compruebe los accesorios que usted debe recibir junto al equipo Estativo L mpara hal gena de recambio 6V 20W Cabezal seg n modelo Aceite de inmersi n Objetivos 4X 10X 40X R 100X R I Fusible Oculares WF10X Funda de pl stico Cable Schuko est ndar Manual de instrucciones 10A 250V UNE 20356 EN 60 320 Certificado de garant a Juego de filtros verde y azul Solo aceptamos devoluciones de equipos en los 15 d as posteriores al env o y siempre que vengan completos en su embalaje original 6 CHOOSING OBJECTIVES AND EYEPIECES The observed image loses surface and definition as the magnifications are higher This increase of magnifications must be obtained by means of more powerful objectives and not from higher magnifying eyepieces since the eyepiece only increases the image pro duced by the objective therefore the more
5. n codificadas por unas marcas de la siguiente manera Figura 2 40 Aumento del objetivo 0 65 Apertura num rica 160 Longitud del tubo 0 17 Espesor del portaobjetos Objetivos acrom ticos Aumentos 4X 10X 20X 40X 60X 100X Distancia de trabajo mm 17 50 7 32 1 26 0 63 0 41 0 19 Apertura num rica 0 10 0 25 0 40 0 65 0 85 1 25 Aumento total Es el resultado de multiplicar el aumento del ocular por el aumento del objetivo Objetivos 4X 10X 20X 40X R 60X R 100X R D Oculares Aumento total 40 100 200 400 600 1000 WE10X pi 4 50 1 80 0 90 0 45 0 30 0 18 WI W E Campo ancho R Retr ctil I Objetivo de inmersi n A N Es la capacidad de las lentes de uni n al rayo de luz proyectada Determina las propiedades del objetivo y el condensador La apertura num rica m s grande hace la imagen m s brillante y la resuelve mejor Distancia de trabajo Distancia en mm entre la preparaci n y la lente frontal del obje tivo cuando el microscopio se encuentra enfocado Distancia focal Distancia desde el plano principal imagen del sistema hasta su foco imagen expresada en mm Resoluci n Es el valor reciproco del poder separador el cual representa la m nima dis tancia en la cual dos peque as part culas bajo la lente pueden verse separadas Se calcula en lineas mm N mero de campo Representa el di metro en mm del diafragma de campo que es for mado por el ocular Campo de visi n Tama o en mm
6. The diaphragm must not be used to reduce the Picture 7 Diaphragm s aperture lighting intensity Security The microscope must be used by previously qualified staff that knows how to operate it thanks to the user manual You should put the microscope in a horizontal plane stable table having a safety area of at least 30 cm per side Do not place the microscope near any warm supply burners blowlamps etc nor expose it directly to the sun Avoid vibrations dust and dry environments WI IMPORTANTE Empiece siempre la observaci n con el objetivo de menos aumentos lo que facilita el enfoque imposibilitando as que se estropeen las preparaciones o se ensu cien los objetivos Adem s es indispensable para regular la iluminaci n previa a la obser vaci n de mayor aumento y se obtiene una imagen de conjunto de la topograf a de la estructura para centrarse posteriormente en los puntos de mayor inter s para su obser vaci n a mayores aumentos Accionando el rev lver se pasar a otro objetivo de mayor aumento Al ser stos para focales bastar con retocar el enfoque con el mando microm trico Regule la altura del condensador para conseguir iluminar uniformemente el campo de visi n menor campo cuanto mayor sea el aumento as la regulaci n del condensador ser m s precisa cuanto mayor sea el aumento En 40 aumentos objetivo 4x y ocular 10x el condensador se bajar para cubrir un campo m s amplio y evitar un
7. a photographic camera or a video Condenser Abbe double lens Numerical aperture 1 25 Moveable filter holder It includes blue and green filters 32 mm diameter Vertically variable by pinion and rack Y INDEX OF CONTENTS 1 USES OF THE INSTRUMENT 18 6 CHOOSING OBJECTIVES AND EYEPIECES 7 CAUSES OF A DEFECTIVE IMAGE oc sssssssssssssssssecsesesssssissnnnnnssssnesssseeseseeeeet 27 8 TROUBLESHOOTING Le ecsesssssssescsestecsnssecnsusecessseccnssecsnececsnsesensscssnssesensersenseseousenenesenenseas 28 9 USEFUL ADVICE caia 28 10 ACCESSORIES 1 ee inn 90 ANNEX I CE CERTIFICATE 1 USES OF THE INSTRUMENT Zuzi professional microscopes series 120 are very hard equipments designed to provi de the highest satisfaction in a wide range of demands They are equipped with high quality optics and a wide range of accessories to complete the microscope Zuzi series 120 are suitable for education as well as for veterinary clinical fields etc 2 DESCRIPTION Picture 1 1 1 Head 2 1 2 Eyepieces 1 3 Dioptric compensation ring 1 4 Screw for holding the head 1 1 5 Base 1 6 Nosepiece 1 7 Objectives 1 8 Stage 5 1 9 Stage clips HA 1 10 Coaxial control EA 5 1 11 Macrometric knob coarse adjustment TES gt P A 10 1 12 Micrometric knob fine adjustment i 1 13 On off switch AED RR 12 1 14 Light intensity control 14 v 1 15 Condenser 18 13 1 16 Iris diaphragm 17 1 17 Fil
8. compra para tener derecho de reclamaci n o pres taci n de la garant a En caso de enviar el equipo al Servicio T cnico adjuntar factura o copia de la misma como documento de garant a Rellene y env e la garant a antes de los 15 d as despu s de la compra El fabricante se reserva los derechos a posibles modificaciones y mejoras sobre este manual y equipo ATENCI N A NO SE ADMITIRA NINGUN APARATO PARA REPARAR QUE NO ESTE DEBIDAMENTE LIMPIO Y DESINFECTADO INDICE DE IDIOMAS Castellano 2 16 ARNOR 1509001 2000 ia AUXILAB S L E Beyotreda CE DECLARATION OF CONFORMITY ZUZI MICROSCOPES SERIES 120 of Auxilab S L for the Directive of Machines 89 392 CEE modified and the regulations adopted for their transposition NAME OF THE MANUFACTURER IMPORTER AUXILAB S L ADDRESS Poligono Morea Norte 8 31191 Beri in Navarra WE STATE THAT ZUZI MICROSCOPES SERIES 120 CODES 50120006 50120007 and 50120008 Are designed and manufactured according to Directive 89 392 CEE including the modifiCations and the nacional regulations that transpose them Directive 73 23 CEE modified over the electric security Directive 89 336 CEE modified over the electromagnetic compability And that the following harmonized rules have been applied or part of them UNE 292 1 UNE 292 2 UNE 292 2 A1 UNE 614 1 UNE 1050 UNE 294 UNE 894 1 UNE 894 2 UNE 60204 UNE 61010 1 BERIAIN 11th May 2006 Signed by ALFONS
9. hubiese cualquier suciedad en las superficies externas de las lentes l mpielas con un pa o suave que no desprenda pelusa humedecido con un poco de xilol o tolueno Para quitar el polvo que se haya posado sobre las lentes sople con una pera o l mpielo con un cepillo o pincel suave de pelo natural o mediante alguna gasa especial para lentes En las partes mec nicas utilice lubricantes no corrosivos teniendo especial cuidado de no tocar las partes pticas ESTE DEBIDAMENTE LIMPIO Y DESINFECTADO ATENCI N A NO SE ADMITIRA NINGUN APARATO PARA REPARAR QUE NO 6 ELECCI N DE OBJETIVOS Y OCULARES La imagen observada pierde superficie y nitidez a medida que los aumentos son supe riores Este incremento de aumentos debe obtenerse mediante objetivos cada vez m s potentes y no a partir de oculares de m s aumento ya que el ocular solo aumenta la ima gen dada por el objetivo por lo que contra m s aumentos tenga el ocular mayor es la p r dida de nitidez claridad y superficie que presenta la imagen Para las observaciones rutinarias utilice el ocular de menos aumentos con objetivos m s potentes El ocular de gran aumento se reservar para casos particulares teniendo pre sente el hecho de que disminuye la definici n y que no incrementa la resoluci n 7 CAUSAS DE UNA MALA IMAGEN En caso de una imagen deficiente comprobar Que la iluminaci n est bien realizada que la intensidad lum nica no sea excesiva ni demasi
10. magnifications It is the result of multiplying the eyepiece magnification and the objective magnifications Objectives 4X 10X 20X 40X R 60X R 100X R 1 Eyepieces Total magnific 40 100 200 400 600 1000 Wrox sa Coe 450 180 0 90 045 030 0 18 Wo 8 LOCALIZACI N DE PROBLEMAS Consulte los siguientes puntos antes de remitir el microscopio al servicio t cnico Si la l mpara no enciende Compruebe que exista una buena conexi n a una toma de corriente apropiada Compruebe que el fusible y la l mpara est n en buen estado Si el campo de visi n aparece recortado Compruebe que el rev lver se encuentra perfectamente encajado en su posici n Para ello gire el rev lver ligeramente a ambos lados hasta que este encaje correctamente Compruebe el correcto centrado del condensador Esta operaci n se realiza mediante tres tornillos situados alrededor de la montura del condensador Si existe polvo o suciedad visible en el campo de visi n Compruebe la existencia de polvo en la lente del precondensador lente superior del condensador y ocular as como la limpieza de la muestra Una vez localizada la zona cau sante del problema proceder a su limpieza seg n lo citado anteriormente Si la preparaci n se desenfoca mientras realizamos la observaci n Comprobar la tensi n del anillo de fricci n ya que probablemente sea demasiado d bil y provoque una ca da involuntaria de la platina Si un sector del campo de visi
11. magnifications the eyepiece has the more the loss of definition brightness and surface of the image it provokes For routine observations you must use the lower magnifying eyepiece with the most powerful objectives The higher magnifying eyepieces will be reserved for especial obser vations taking into account that they reduce the definition and do not increase the reso lution 7 CAUSES OF A DEFECTIVE IMAGE In case of a defective image you must check that The lighting has been properly done the lighting intensity cannot be excessive nor too weak Never adjust it with the condenser s diaphragm Both the condenser and the lamp must be well centered Check that there is no diffusor filter between the field s diaphragm and the aperture one Eyepieces fit well and objectives are properly tightened The cleanliness of the whole optical system The process should be as follows a Make the eyepieces rotate and if the specks move clean them b Make the head rotate You must never disassemble the head However you can clean carefully those accesible surfaces of the prisms blowing them with a rubber pear c Rotate the objective If the parasit images also rotate clean it with a small dry brush trying to remove the dust d Observe the frontal surface with a magnifying glass or an inverted eyepiece There should be enough immersion oil and it must not contain bubbles or impurities The thickness of microscope slides micros
12. n est fuera de foco Comprobar si el objetivo est correctamente situado en la trayectoria del haz luminoso Si no es as girarlo hasta que se inserte correctamente en su posici n Comprobar si la muestra se encuentra situada correctamente en la platina y asegurarse de que queda firmemente sujeta por la pinza Si la lente frontal del objetivo toca la preparaci n durante la operaci n de enfoque o al pasar de un objetivo de menor aumento a otro objetivo superior Comprobar si la preparaci n est colocada al rev s el portaobjetos encima del cubre objetos y proceder a su correcta colocaci n si fuese necesario Comprobar si el cubreob jetos tiene el espesor apropiado 0 17 mm 9 RECOMENDACIONES PR CTICAS Como el cabezal es giratorio en todos los modelos de la serie 120 es preferible la obser vaci n por la parte frontal de la platina ya que facilita el acceso a los mandos mec nicos Cuando utilice el objetivo de inmersi n debe tener en cuenta las siguientes considera ciones a Para utilizar toda la apertura num rica del objetivo de inmersi n tanto el objetivo WI de inmersi n tanto el objetivo como la preparaci n estar n sumergidos en aceite de inmersi n Para ello proceda del siguiente modo a 1 Enfoque la preparaci n con un objetivo de pocos aumentos 4x a 2 Ponga una gota de aceite de inmersi n en el cubre a 3 Gire el rev lver y sit e el objetivo de inmersi n retoque el enfo
13. objective touches the preparation while focusing or while changing from a low magnifying objective to a higher one Check if the sample specimen is upside down the microscope cover glass on top of the microscope slide and if no proceed to place it correctly Check also if the cover glass has the suitable thickness 0 17mm 9 USEFUL ADVICE As all professional microscopes series 120 have a rotary head it is advisable to proceed to observation by the frontal part of the stage as it enables accesing to the mechanical controls When you use the immersion objective you have to take into account the following a To use all the numerical aperture of the immersion objective both the objective and the sample specimen should be immersed in oil To do so W Base Control coaxial mandos verticales Dimensiones 130 x 140 mm Area de recorrido 75 X x 50 Y mm Escalas graduadas a 0 1 mm Diafragma Iris Situado bajo el condensador y la base mec nica Regula la cantidad de luz que pasa al condensador Sistema de iluminaci n L mpara hal gena de baja tensi n de lampara 6V 20W toma de corriente incorporada 220 230V 50 60Hz 10 con potenciador del control de la intensidad Especificaciones pticas Objetivos acrom ticos Objetivo que corrige las aberraciones pticas Dicho sistema ptico puede transmitir la luz blanca sin descomponer en sus colores constituyentes Las caracter sticas de cada objetivo est
14. y a las reglamentaciones adoptadas para su transposici n NOMBRE DEL FABRICANTE IMPORTADOR AUXILAB S L DIRECCI N Pol gono Morea Norte 8 31191 Beri in Navarra DECLARAMOS QUE MICROSCOPIO ZUZI SERIE 120 REFERENCIAS 50120006 50120007 y 50120008 Est n dise ados y fabricados de acuerdo a Directiva 89 392 CEE incluidas las modificaciones de la misma y las regla mentaciones nacionales que la transponen Directiva 73 23 CEE modificada sobre seguridad el ctrica Directiva 89 336 CEE modificada sobre compatibilidad electromagn tica Y que se han aplicado las siguientes normas armonizadas o parte de ellas UNE 292 1 UNE 292 2 UNE 292 2 A1 UNE 614 1 UNE 1050 UNE 294 UNE 894 1 UNE 894 2 UNE 60204 UNE 61010 1 BERIAIN a 11 de Mayo de 2006 Fdo ALFONSO AINCIBURU SANZ DIRECTOR GERENTE Pol gono Morea Norte 8 31191 Beriain Navarra Spain Tel 948 310 513 Fax 948 312 071 Internet www auxilab es Email correo auxilab es WI Thank you for choosing this equipment We sincerely wish that you enjoy your profes sional microscope Zuzi series 120 We highly recommend looking after this equipment according to what is stated in this manual Zuzi develops its products according to the CE marking regulations as well as empha sizing the ergonomics and security for its user The correct using of the equipment and its good quality will permit you to enjoy this equipment for years The improper use of th
15. 5 con diafragma y portafiltros Platina Platina fija mec nica con mov 130x140 mm Recorrido 75 x 50 mm Pinza soporte para muestras Iluminaci n Transmitida l mpara hal gena 6V 20W Dimensiones 180 mm x 380 mm x 230 mm Estativo Met lico muy estable con mandos coaxiales a ambos lados para enfoque macro y micro El mando derecho esta provisto de un anillo de control de fricci n y el izquierdo de palanca de bloqueo de platina bloqueando la altura m xima de la platina para no haya ninguna posibilidad de que el objetivo haga contacto por accidente con el portaobjetos y da e la muestra o el propio objetivo Cabezal La serie 120 Zuzi dispone de diversos cabezales dependiendo del modelo Monocular inclinado 45 y giratorio 360 Puede fijarse en cualquier posici n Binocular inclinado 45 y giratorio 360 Graduaci n sim trica interpupilar con escala entre 55 y 75mm y anillo de compensaci n di ptrica 5 dioptr as Triocular dispone de un cabezal binocular inclinado 45 y giratorio 360 Graduaci n sim trica interpupilar con escala entre 55 y 75mm y anillo de compensaci n di ptrica 5 dioptr as Y un tubo vertical para adaptarlo a una c mara fotogr fica o a un video Condensador Abbe de doble lente Apertura num rica 1 25 Porta filtros m vil Incluye filtros verde y azul 32 mm de di metro Variable verticalmente por pi n y cremallera Wy a 1 Focus the preparatio
16. LACI N PUESTA EN MARCHA 5 MANTENIMIENTO Y LIMPIEZA on 6 ELECCI N DE OBJETIVOS Y OCULARES ccoo 12 7 CAUSAS DE UNA MALA IMAGEN ooo icsseecssssecssesessueesssseessutessasessseeesssesnseees 12 8 LOCALIZACI N DE PROBLEMAS caciones 13 9 RECOMENDACIONES PR CTICAS anne 13 1 APLICACIONES DEL INSTRUMENTO Los microscopios profesionales serie 120 Zuzi son unos equipos muy firmes espec fica mente dise ados para brindar la mayor satisfacci n en un gran campo de necesidades Est n equipados con una ptica de gran calidad y una gran cantidad de accesorios para completarlo Los microscopios profesionales serie 120 Zuzi son adecuados para educaci n veterina ria hospitales 2 DESCRIPCI N Figura 1 1 1 Cabezal 1 2 Oculares 1 3 Anillo compensaci n di ptrica 1 4 Tornillo sujeci n del cabezal 1 5 Estativo 1 6 Revolver 1 7 Objetivos 1 8 Platina 1 9 Pinza 1 10 Mando movimiento coaxial axial 1 11 Mando macrom trico 1 12 Mando microm trico 1 13 Interruptor 1 14 Regulador intensidad de luz 1 15 Condensador 1 16 Diafragma iris 1 17 Portafiltros 1 18 Precondensador ai 3 ESPECIFICACIONES T CNICAS Especificaciones t cnicas 120 6 120 7 120 8 Cabezal Monocular Binocular Triocular Oculares A y y y Rev lver Cu druple D D 4X A N 0 10 10X A N 0 25 40X R AN 0 65 100X R I A N 1 25 Objetivos Acrom ticos antif ngico s s s Condensador Abbe N A 1 2
17. MICROSCOPIOS PROFESIONALES PROFESSIONAL MICROSCOPES gt A Serie Series 120 Este manual es parte inseparable del aparato por lo que debe estar disponible a todos los usuarios del equipo Le recomendamos leer atentamente el presente manual y seguir rigurosamente los proce dimientos de uso para obtener las m ximas prestaciones y una mayor duraci n del mismo This manual should be available for all users of these equipments To get the best results and a higher duration of this equipment it is advisable to read carefully this manual and follow the processes of use Gracias por haber adquirido este equipo Deseamos sinceramente que disfrute del microscopio profesional serie 120 Zuzi Le recomendamos que cuide el equipo conforme a lo expuesto en este manual Zuzi desarrolla sus productos seg n las directrices del marcado CE y haciendo hinca pi en la ergonom a y seguridad del usuario La calidad de los materiales empleados en la fabricaci n y el correcto proceder le permitir n disfrutar del equipo por muchos a os El uso incorrecto o indebido del equipo puede dar lugar a accidentes descargas el ctri cas cortocircuitos fuegos lesiones etc Lea el punto de Mantenimiento donde se recogen aspectos de seguridad LEA DETALLADAMENTE ESTE MANUAL DE INSTRUCCIONES ANTES DE OPERAR CON ESTE EQUIPO CON EL FIN DE OBTENER LAS MAXIMAS PRES TACIONES Y UNA MAYOR DURACI N DEL MISMO Tenga especialmente presente lo siguiente
18. NICAL SPECIFICATIONS Technical specifications 120 6 120 7 120 8 Head Monocular Binocular Trinocular Eyepieces WF 10x Nosepiece Quadruple 4X N A 0 10 10X N A 0 25 40X R N A 0 65 100X R I N A 1 25 Objectives Achromatic anti fungus Condenser Abbe N A 1 25 diaphragm amp filter holder Stage Fixed mechanical with mov 130x140 mm Movement range 75 x 50 mm Clips for samples Lighting Transmitted halogen lamp 6V 20W Dimensions 180 mm x 380 mm x 230 mm Base It is metallic very stable provided with coaxial controls at both sides to enable macro metric and micrometric focusing The right control has a friction control ring and the left one has a stage blocking handle The mentioned handle stops the stage at the higher point in order to avoid that the objective may crash the slide and so damage either the sample specimen or the objective itself Head We can supply you with different heads depending on the model Monocular inclined 45 and rotary 360 It can be fixed in any position Binocular inclined 45 and rotary 360 Symmetrical interpupilary graduation with scale between 55 and 75mm and dioptric compensation ring 5 dioptres Trinocular it is provided with a binocular head inclined 45 and rotary 360 Symmetrical interpupilary graduation with scale between 55 and 70mm dioptric compensation ring 5 dioptres and a vertical tube to be adapted to
19. O AINCIBURU SANZ DIRECTOR MANAGER Poligono Morea Norte 8 31191 Beriain Navarra Spain Tel 948 310 513 Fax 948 312 071 Internet www auxilab es Email correo auxilab es Y 10 ACCESSORIES This professional series can be completed with several Zuzi accessories that increase the range of use Phase contrast We have available two models of phase contrast kits Simple phase contrast code 90100022 Turret code 90137022 If we change the condenser we can choose the magnifications we need using the most suitable objectives Thermostatic stage code 90100597 It allows to keep the sample in a suitable temperature itis very useful for the observa tion of living samples e g cells microorganisms semen Fluorescent attachment code 90137035 Eyepieces We have available a wide range of eyepieces of different magnifications wide field with pointer with micrometer etc Ask your distributor dealer or enter the web www auxi lab es Objectives There is also a wide range of objectives of different magnifications Achromatic semi plan plan etc Ask your distributor dealer or enter the web www auxilab es Eyepiece videocamera For models monocular binocular trinocular Video pal system code 59140050 USB codes 59140060 and 59150060 WI INDICE DE CONTENIDOS 1 APLICACIONES DEL INSTRUMENTO cacon 3 2 DESCRIPCI N coccccccccccconicnicnninnns 3 ESPECIFICACIONES T CNICAS 4 INSTA
20. a iluminaci n demasiado intensa Deber elevarlo progresivamente con los objetivos mayores para con centrar el haz de luz y ganar en iluminaci n y contraste Con el objetivo de inmersi n deber elevar el condensador a su m xima altura El diafragma iris permite disminuir la apertura hasta un valor similar al del objetivo y de este modo evitar una iluminaci n marginal que disminuya el contraste Por lo tanto estar m s cerrado cuando se utilice un objetivo de poco aumento y de escasa apertura num rica En realidad solamente se usar su apertura total con el objetivo de inmersi n Para realizar el ajuste se debe partir de la m xima apertura A medida que aumente el contraste por estar el diafragma m s cerrado aparecen fen menos de difracci n que limi tan la resoluci n El ajuste correcto se obtendr cuando el contraste sea m ximo sin que se altere la resoluci n Una forma pr ctica de realizar el ajuste es la siguiente a Se retira el ocular y se observa la pupila de sali da del objetivo a trav s del tubo vac o ajustando la detomeme apertura del diafragma Generalmente es preferible dejar la apertura del diafragma sobre el 70 80 de 50 20 la apertura num rica del objetivo fig 7 b El cierre del diafragma reduciendo la apertu ra num rica del sistema ptico aumenta la pro fundidad de enfoque o poder de penetraci n No eros is se debe sobrepasar el l mite de difracci n bajo el pretexto de aumen
21. ado d bil No regularla nunca con el diafragma del condensador El condensador y la l mpara deben estar bien centrados Comprobar que entre el diafragma de campo y el de apertura no hay ning n filtro difusor Que los oculares est n bien encajados y los objetivos bien enroscados La limpieza de todo el sistema ptico El procedimiento ser de la siguiente manera a Hacer girar los oculares contemplando si las motitas se mueven si es as limpiarlos b Hacer girar el cabezal en su conjunto Nunca debe desmontarse el cabezal pero si pueden limpiarse delicadamente soplando con una pera aquellas superficies accesibles de los prismas c Girar el objetivo Si las im genes par sitas giran limpiarlo con la ayuda de un pincel seco intentando eliminar el polvo d Observar la superficie frontal con una lupa o con un ocular invertido Que el aceite de inmersi n sea suficiente y que no contenga burbujas ni impurezas Que el grosor de portaobjetos cubreobjetos y medio de montaje no sea demasiado grue so e impida el enfoque a medios y grandes aumentos Existen unas dimensiones est ndar tanto para el portaobjetos como para el cubreobjetos Los portaobjetos y cubreobjetos deben estar limpios Comprobar que el primero est bien colocado y que no haya dos cubreobjetos superpuestos WE Wide Field R Retractile I Immersion objective N A Numerical aperture it determines the objective s properties The numerical aper
22. check that all the accessories are enclosed with the equipment Base Spare halogen lamp of 6V 20W Head depending on the model supplied Immersion oil Objectives 4X 10X 40X R 100X R D Fuse Eyepieces WF10X Plastic cover Standard Shuko cable User manual 10A 250V UNE 20356 EN 60 320 Set of filters green and blue Warranty certificate We will only accept the equipment s return within 15 days after delivery provided it comes in its original wrapping Stage Coaxial control vertical controls Dimensions 130 x 140 mm Movement range 75 X x 50 Y mm Graduated scales at 0 1 mm Iris diaphragm It is placed under the condenser and the mechanical stage It controls the light going through the condenser Lighting system Low tension halogen lamp 6V 20W built in power input 220 230V 50 60Hz 10 with intensity control potentiometer Optical specifications Achromatic objectives They correct optical aberrations This optical system is able to transmit white light without separating constituent colours The characteristics of each objective are codified with some marks as shown above Picture 2 40 Objective magnification 0 65 Numerical aperture 160 Tube s length 0 17 Microscope slide s thickness Achromatic objectives Magnifications 4X 10X 20X 40X 60X 100X Working distance mm 17 50 7 32 1 26 0 63 0 41 0 19 Numerical aperture 0 10 0 25 0 40 0 65 0 85 1 25 Total
23. cope cover glasses and assembly medium so as to stop the focusing at medium or high magnifications There are standard dimensions for both microscope slides and cover glasses Both microscope slides and cover glasses should be clean You should check that the slides are properly placed and that there are no superimposed cover glasses Picture 8 Changing the lamp If it is necessary to change the fuse loosen the protective lid that is on the base of the microscope Fuse pic 9 change the fuse 0 5A and place it in the same position putting again the protective lid Please use the plastic cover provided to keep the microscope away from dust lying on the optical parts when it is not used for a long period of time Please keep the original packaging to transport the equipment either when it is not being used for a long time orin case you send it for an overhaul Picture 9 Changing the fuse Cleaning Never use scourers or substances that can grate for cleaning metallic parts such as stain less steel aluminium coatings etc as they damage the microscope and produce an early ageing of the equipment Use a fluff free cloth dampened with soaped water that does not contain abrasives Lenses must not be disassembled by the user Were there any dust or dirt to be cleaned you should clean it with a natural horse hair brush or a smooth piece of cloth fluff free dampened with a bit of xilol or toluene To remove the dust sett
24. de pl stico del microscopio siempre que ste no est en uso para evitar que el polvo se pose sobre las partes pticas cuando vaya a estar fuera de uso por un per odo prolongado fin de semana vacaciones etc En aquellos modelos provistos con sistemas de observaci n binocular ante todo deben regularse correctamente los dos tubos oculares Esto depen der de a La distancia interpupilar fig 5 Regule la dis tancia interpupilar separando juntando los ocu lares hasta conseguir una total fusi n de las dos imagenes Figura 4 Inserci n de los oculares b La compensaci n di ptrica de los tubos Cerrando alternativamente un ojo y despu s el otro se apreciar una diferencia m s o menos acu sada del enfoque Para corregir este defecto ajuste el anillo de compensaci n di ptrica a cero Mirando por el ocular derecho enfoque la pre paraci n accionando los mandos macrom trico y microm trico y una vez enfocada mire por el ocular izquierdo y ajuste con el anillo de compensaci n di ptrica hasta visualizar una imagen n tida Si el observador tiene astigmatismo debe conservar sus gafas puestas verificando previamente que exista una distancia de observaci n suficientemente alejada de la lente ocular para evitar que sta roce con el cristal de las gafas Figura 5 Distancia interpupilar Una vez hallados estos valores ser muy til memorizarlos sobre todo si el microsco pio es compartido por m s d
25. de under the European regulations for electrical security electromagnetic compati bility and security on machines 5 MAINTENANCE AND CLEANING To get the best results and a higher duration of this equipment it is essential to follow the processes of use Note All the processes of use mentioned below will not have any value unless you keep a continued and careful maintenance Please follow the processes of use of this manual This manual should be available for all users of this equipment Always use original components and supplies Other devices can be similar but they can damage the equipment The microscope is supplied with a Schuko standard wire and it should be connected to a current wire provided with an earth wire it should be handy to be disconnected in case of emergency Never try to repair the microscope by yourself since you will lose the warranty and may provoke damages to the general operating system or the electrical installation as well as injuries to the people that usually handle the microscope burns hurts or dama ges in nearby equipments In the event of breakdown please contact your distributor to overhaul through Zuzi Technical Assistance Department Make sure you do not use lamps with higher power as you may provoke an overheating or any other malfunction If the lamp burns out change it with an Zuzi halogen lamp 6V 20W being careful not to touch the bulb with your hands IMPORTANT Be
26. e aseg rese de desconectar el microscopio de la red el ctrica Para efectuar el cambio de l mpara disponen de un dispositivo en la parte de abajo fig 8 Suelte el tornillo y qu telo afloje los tornillos del portal mparas y c mbiela agarr n dola firmemente Centre el portal mparas de nuevo si es necesario WI Assemble the head by adjusting it with the built in screw pic 4 Insert the eyepieces WF10x The binocular tube is normally placed in the same direction as the microscope front but if necessary it can placed in any other direction Always use the microscope dust cover when the microscope is not being used in order to avoid dust settling on the optical parts especially for long periods of time weekend holidays Models provided with binocular observation system must have the eyepiece tubes properly adjusted This will depend on a Interpupillary distance pic 5 Adjust the interpupillary distance by separating or joining the eyepieces until there is a total fusion of the two images b Dioptric compensation of the tubes You will appreciate a difference in the focusing more or less marked closing alternatively one eye then the other To correct this defect you should adjust the dioptric ring to zero Looking through the right eyepiece you have to focus the sample specimen Picture 5 Interpupillary distance using the macro and micrometric controls Once you have focused the sample look through t
27. e equipment can cause accidents and electric discharges circuit breakers fires damages etc Please read the point of Maintenance where we expose the security notes TO GET THE BEST RESULTS AND A HIGHER DURATION OF THE EQUIP MENT IT IS ADVISABLE TO READ THOROUGHLY THIS MANUAL BEFORE OPERATING WITH THE EQUIPMENT Please bear in mind the following This manual is inseparable from the professional microscope Zuzi series 120 so it should be available for all the users of this equipment You should carefully handle the microscope avoiding sudden movements knocks free fall of heavy sharp objects on it Avoid spilling liquids inside the equipment Never dismantle the different pieces of the microscope to repair it yourself since it could produce a defective use of the whole equipment and a loss of the product warranty as well as injuries on people that handle the microscope To prevent fire or electric discharges avoid dry or dusty environments In case it may happen unplug the equipment immediately If you have any doubt about setting up installation or functioning do not hesitate in contacting your wholesaler You can also tell us any doubts or suggestions you have by contacting Zuzi Technical Assistance Department by email to asistencia auxilab com or by telephone 34 807 117 040 0 30 Euros min This equipment is protected under the Warranties and consumer goods regulation 10 2003 Overhaul is not covered by the mic
28. e un usuario para evitar tener que repetir la localizaci n de los valores id neos cada vez que el microscopio es manipulado Antes de colocar la preparaci n sobre la platina sta debe bajarse a una distancia superior a la dis tancia de trabajo del objetivo de menos aumentos Hecho esto se coloca la preparaci n en el carro mec nico encaj ndola en la escuadra y sujet ndola mediante la pinza m vil Mirando por fuera se cen tra la muestra respecto a la apertura de la platina Mirando con el ocular y el objetivo de menos aumentos suba la platina con el mando macrom trico hasta que aparezca la imagen En este momen to se empieza a accionar el mando microm trico hasta conseguir un correcto enfoque fig 6 Figura 6 Enfoque During its functioning dangerous materials such as flammable or pathological substan ces must be out of the safety area When you are not using the microscope for a long period of time please make sure it is unplugged from the net in order to avoid possible accidents It is essential to have the equipment switched off and unplugged from the net before cleaning checking components or replacing any piece e g replacement of a fuse Never try to repair the microscope by yourself since you will lose the warranty and may provoke damages to the general operating system or the electrical installation as well as injuries to the people that usually handle the equipment burns hurts Ma
29. ectives into the nosepiece in an ascendant order 4x 10x 40x 100x in clockwise direction pic 3 Picture 3 Placing the objectives VERY IMPORTANT You must never hold the microscope by the mechanical stage or the tube thus all weight would be supported by the micrometric screw whose notches would be worn out slowly You should always hold it by the base Y Figura 8 Cambio de l mpara Si fuese necesario remplazar el fusible desenros que la tapa protectora que se encuentra en la parte posterior del microscopio Fuse fig 9 reemplace el fusible de 0 5 A y vuelva a colocarlo en su misma posici n enroscando de nuevo la tapa protectora Utilice la funda de pl stico siempre que el micros copio no est en uso para evitar que el polvo se pose sobre las partes pticas Guarde el embalaje original para transportarlo as como cuando no vaya a utilizarse durante mucho tiempo o cuando haya que enviarlo a revisar Figura 9 Cambio de fusible Limpieza Para la limpieza de las partes met licas acero inoxidable aluminio pinturas etc nunca utilice estropajos o productos que puedan rayar pues limitan la vida til del equipo Para la limpieza del equipo recomendamos se utilice un trapo libre de pelusa humedeci do con agua jabonosa que no contenga productos abrasivos Para limpiar los componentes pticos del microscopio lentes oculares stos no deben ser desmontados por el usuario Si
30. fore changing the lamp bulb or the fuse make sure that the micros cope is disconnected from the net In order to enable the changing of the lamps these microscopes are provided with a device at the bottom of its base pic 8 Loosen the screw and take it off then loosen the screws of the lamp holder and replace it holding it firmly Center the lamp holder again if necessary UW IMPORTANT You should always start operating by using the objective with lesser mag nification as it facilitates focusing and makes impossible to ruin the slides or get the objec tives dirty Furthermore it is essential to adjust the previous lighting to using the higher magnification and it helps you to see the whole topographic structure of the sample so as you can concentrate on the more appealing parts to observe them at higher magnifications By turning the nosepiece you will pass to an objective with higher magnification As they are paracentric you will only have to adjust focusing with the micrometric control You should adjust the condenser s height to obtain a homogeneous lighting of the field of view the higher magnifying objective is used the smaller will be the field of view Thus the condenser s precision will be higher as we use high magnification values At 40x objec tive 4x and eyepiece 10x you should descend the condenser to obtain a wider field of view and avoid having excessive lighting The condenser will have to be raised progressivel
31. he left eyepiece and then adjust the dioptric adjusting ring until you can obtain a clear sharp image In case the observer wears glas ses he must do the adjustment with them on making sure he she maintains a certain dis tance from the eyepiece lens so as to avoid grazing the glasses Picture 4 Placing the eyepieces Once the values are found it will be very useful to learn them by heart mostly if the microscope is shared with other users This is highly advisable to avoid repeating the adjustment each time the microscope is used Before placing the sample specimen on the stage this one should be descended to a certain distance higher than the working distance of the objective with the lower magnifying values Then you should place the specimen sample on the mechani cal stage fitting it into the set square with the help of the moveable clips Finally you should centre the specimen sample regarding the stage aperture by looking from the outside Looking by the eyepiece and with the lower mag nifying objective raise the stage with the macrome tric control until the image appears Then move the micrometric control until obtaining a correct focusing pic 6 Picture 6 Focusing WI Installation Before using this instrument it is convenient for you to familiarize with its components and basic essentials PLEASE READ THOROUGHLY THE INSTRUCTIONS BEFORE CONNECTING AND OPERATING WITH THIS EQUIPMENT
32. icture 11 Focusing limit control Y 8 TROUBLESHOOTING Before sending the microscope to our Technical Assistance Department you should check the following points If the lamp does not light Check that the plug is in good conditions and that it is on a suitable earth connection Check the fuse and the lamp are in good conditions If the field of view appears to be cut out Check that the nosepiece is correctly fitted in its position To check it rotate the nose piece slightly clockwise and counterclockwise until it fits properly Check the condenser is well centered You should do this by adjusting three small screws placed around the condenser mount If there is visible dust or dirtiness on the field of view Check if there is dust in the precondenser lens upper lens of the condenser and eye piece as well as the cleanliness of the sample specimen Once you have found the area causing the problem clean it as previously instructed If the preparation takes out of focus while we carry out the observation Check the tension adjustment ring since it may be too weak and it could provoke an involuntary fall of the stage If_a section of the field of view is taken out of focus Check that the objective is properly placed in the light beam path If not rotate it until itis properly fitted Check if the sample is properly placed on the stage and make sure that it is firmly held by the clip If the frontal lens of the
33. led on the lens blow with a rubber pear or clean it with a small soft brush You should use non corrosive lubricants in metallic parts being careful of not touching optical parts ATTENTION IF EQUIPMENTS ARE NOT PROPERLY CLEAN AND DISINFECTED THEY WOULD NOT BE ALLOWED TO REPAIR BY OUR TECHNICAL SERVICE Instalaci n Antes de comenzar a utilizar el instrumento es conveniente familiarizarse con sus com ponentes y fundamentos b sicos as como con las funciones de sus controles LEA DETALLADAMENTE ESTE MANUAL DE INSTRUCCIONES ANTES DE OPERAR CON ESTE EQUIPO CON EL FIN DE OBTENER LAS MAXIMAS PRES TACIONES Y UNA MAYOR DURACI N DEL MISMO Coloque el microscopio sobre una mesa horizontal plana y estable creando un espa cio libre al menos de 30 cm por cada lado No coloque el equipo en zonas pr ximas a fuen tes de calor mecheros sopletes ni lo exponga directamente a la luz del sol etc Evite en el lugar de trabajo productos inflamables o t xicos El microscopio se suministra con un cable Schuko est ndar Inserte el cable de alimentaci n de corriente alterna CA a la base de corriente 220V 50Hz 10 provista de toma de tierra y por el otro extremo al conector del microscopio Ni el fabricante ni el distribuidor asumir n responsabilidad alguna por los da os oca sionados al equipo instalaciones o lesiones sufridas a personas debido a la inobservan cia del correcto procedimiento de conexi n el c
34. n tica y seguridad en m quinas 5 MANTENIMIENTO Y LIMPIEZA Para un adecuado funcionamiento del microscopio es necesario seguir algunas reco mendaciones Nota Todas las normas de utilizaci n citadas anteriormente carecer n de valor sino se realiza una continua labor de mantenimiento Siga las instrucciones y advertencias relativas a este manual Tenga este manual siempre a mano para que cualquier persona pueda consultarlo Utilice siempre componentes y repuestos originales Puede ser que otros dispositivos sean parecidos pero su empleo puede da ar el equipo El microscopio dispone de un cable de red Schuko este debe conectarse a una toma de corriente que est conectada a tierra debiendo quedar a mano para poder desconectarlo en caso de emergencia No intente repararlo usted mismo adem s de perder la garant a puede causar da os en el funcionamiento general del microscopio as como lesiones a la persona quemaduras heridas y da os a la instalaci n el ctrica o equipos el ctricos cercanos o pr ximos En caso de aver a dir jase a su proveedor para su reparaci n trav s del Servicio T cnico de Zuzi Si se funden las l mparas reempl celas por otras de 6V 20W originales Zuzi teniendo cuidado de no tocar la ampolla con las manos desnudas No utilice l mparas de mayor potencia ya que podr a producir un sobre calentamiento u otra mala funci n IMPORTANTE Antes de reemplazar la bombilla o el fusibl
35. n with a low magnifying objective 4x a 2 Put a drop of immersion oil in the cover glass a 3 Rotate the nosepiece and place the immersion objective then adjust focusing with the micrometric control b You should avoid the appearance of bubbles in the oil film since it spoils the resul ting image c It is advisable to dry the frontal lens after each observation Never use a yellow dus ter It is better to dry it with a special paper for the cleaning of glasses or with a piece of linen cloth The objective must never be submerged in any kind of liquid since it can pro voke the detachment of the lens Depending on the observant the tension of the macrometic focal controls can be modified since they are provided with a tension adjustment ring that is placed next to the right macrometric control of the microscope pic 10 However you must avoid to loosen it excessively since this could pro duce that the stage falls and therefore focusing accuracy problems There is also a device for the control of the focu sing limit All models of this series are provided with a ring that avoids that the stage raises beyond a certain point that can be determined once the sample has been focused and this fact can avoid possible accidental collisions while observing pic 11 This mechanism also allows the possibility of getting a quick focusing accele rating the observation process when you are wor king with different samples P
36. que con el mando microm trico b Debe evitar la formaci n de burbujas en la pel cula de aceite ya que se deteriora en gran medida la imagen resultante c Se recomienda secar la lente frontal despu s de cada observaci n Nunca debe usar se gamuza Debe hacerse con un papel especial para la limpieza de gafas o con un trozo de tela de hilo No debe sumergirse nunca el extremo del objetivo en ning n tipo de l qui do dado que puede provocar el desprendimiento de la lente Dependiendo de las preferencias del observador se puede modificar la tensi n de los mandos de enfoque macrom trico Para ello se dispone de un anillo de control de fricci n junto al mando macro m trico derecho del microscopio fig 10 Sin embargo se debe evitar aflojarlo excesivamente ya que esto podr a causar una ca da involuntaria de la platina con los consiguientes problemas de preci si n en el enfoque Tambi n existe un dispositivo de control del l mite de enfoque Todos los modelos de esta serie disponen de un anillo que evita que la platina se eleve m s all de un punto que se puede determi nar una vez enfocada la muestra y de este modo evitar posibles colisiones accidentales durante la observaci n fig 11 Este mecanismo tambi n per mite la posibilidad de realizar un r pido enfoque acelerando el proceso de observaci n cuando se trabaja con diferentes muestras Figura 11 Control de l mite de enfoque WI 3 TECH
37. roscope warranty Operations made by non qualified staff will automatically produce a loss of the micros cope warranty Neither fuses nor accessories including their loss are covered by the product s warranty The warranty neither covers piece s deterioration due to the course of time Please make sure you keep the invoice either for having the right to claim or asking for warranty coverage In case you have to send the equipment to Zuzi Technical Assistance Department you should enclose the original invoice or a copy as guarantee Please do not forget filling the warranty certificate and send it before 15 days after the date of purchase Manufacturer reserves the right to modify or improve the manual or equipment ATTENTION N IF EQUIPMENTS ARE NOT PROPERLY CLEAN AND DISINFECTED THEY WOULD NOT BE ALLOWED TO REPAIR BY OUR TECHNICAL SERVICE INDEX OF LANGUAGES
38. tar la profundidad de enfoque c El diafragma no debe usarse para reducir la Figura 7 Apertura del diafragma intensidad de la iluminaci n Seguridad Los microscopios deben ser utilizados por personal cualificado previamente conocien do el equipo y su manejo mediante el manual de uso Coloque el microscopio sobre una mesa horizontal plana y estable creando un espacio libre al menos de 30 cm por cada lado de ella No coloque el microscopio en zonas pr ximas a fuentes de calor mecheros sopletes no exponga el equipo directamente a la luz del sol y evite las vibraciones el polvo y ambientes muy secos Durante el funcionamiento el material peligroso como l quidos inflamables o material patol gico deben estar fuera de esta rea Cuando no vaya a hacer uso del equipo por largos periodos de tiempo aseg rese de que est desconectado de la red para evitar posibles accidentes Para cualquier manipulaci n de limpieza verificaci n de los componentes o sustituci n de cualquier componente Ej sustituci n de fusible es imprescindible apagar el equipo y desconectarlo de la toma de corriente No intente repararlo usted mismo Adem s de perder la garant a puede causar da os en el funcionamiento general del equipo as como lesiones a la persona quemaduras heridas y da os a la instalaci n el ctrica Fabricado seg n las directivas europeas de seguridad el ctrica compatibilidad electro mag
39. ter holder 1 18 Precondenser 10 ACCESORIOS La serie profesional puede ser completada con multitud de accesorios Zuzi accesorios que ampl an su mbito de aplicaci n Contraste de fases Hay disponibles dos modelos de contraste de fases con los correspondientes objetivos Contraste de fases simple referencia 90100022 Torreta referencia 90137022 si se cambia el condensador se pueden elegir los aumen tos que desee usando los objetivos m s adecuados Platina termost tica referencia 90100597 Permite mantener la muestra a la temperatura adecuada Es muy til para la observa ci n de muestras vivas c lulas microorganismos semen Equipo de epifluorescencia referencia 90137035 Oculares Disponemos de una amplia gama de objetivos de diferentes aumentos campo amplio con puntero con micr metro etc Consulte a su distribuidor o en la p gina web www auxilab es Objetivos Disponemos tambi n de una amplia gama de objetivos de diferentes aumentos Acrom ticos semiplanos planos etc Consulte a su distribuidor o en la p gina web www auxilab es C mara ocular Para modelos monocular binocular y triocular Video pal system referencia 59140050 USB referencias 59140060 y 59150060 WI ANEXO I CERTIFICADO CE ARNOR 1509001 2000 ba AUXILAB S L E Best DECLARACI N CE DE CONFORMIDAD MICROSCOPIO ZUZI SERIE 120 de Auxilab S L a la Directiva de M quinas 89 392 CEE modificada
40. trica La tensi n debe ser de 220V 50Hz 110 Cuando no vaya a hacer uso del microscopio durante largos per odos de tiempo ase g rese de que est desconectado de la red y prot jalo del polvo evitando as posibles accidentes y prolongando la vida til del equipo Puesta en funcionamiento El observador debe adoptar una postura c moda bien sentado y con la espalda recta Es conveniente trabajar sobre una mesa oscura a fin de eliminar toda luz par sita que des lumbra y disminuye la buena definici n de las im genes evitando as una fatiga absurda Preste atenci n a la luz ambiental y coloque el microscopio alejado de las ventanas Haga bajar la platina mediante los mandos de enfoque macro Enrosque los objetivos en el rev l ver porta objetivos siguiendo un orden ascendente 4x 10x 40x 100x en el sentido de las agujas de un reloj fig 3 Figura 3 Colocaci n de objetivos MUY IMPORTANTE Nunca coja el microscopio por la platina ni por el cabezal ya que de esta manera todo el peso del aparato descansar a sobre el tornillo microm trico cuyas muescas ser an lentamente erosionadas C jalo por el estativo o la base WI Monte el cabezal ajust ndolo con el tornillo que lleva incorporado fig 4 Inserte el los ocular es WF10x El tubo binocular se coloca normalmente en direcci n del frente del microscopio pero si fuese necesario puede colocarse en cualquier otra direcci n Utilice la funda
41. y using higher objectives to concentrate the light beam and get a better illumination and con trast With the immersion objective the condenser will have to be raised to its maximum height Iris diaphragm enables decreasing the numerical aperture to a value similar to objective s Thus you avoid marginal illumination that decreases contrast and therefore it will be more closed when you use a lesser magnification objective and with a small numerical aperture You will only use its total aperture when you use the oil immersion objective In order to properly carry out the adjustment you should start with the maximum aper ture As contrast increases because the diaphragm is closing diffraction phenomenons appear thus limiting resolution The correct adjustment will be obtained when we have the maximum contrast without altering resolution A useful way of making the adjustment is the following one a Take off the eyepiece and observe the objective Objective s s exit pupil through the empty tube adjusting the output pupil diaphragm aperture It is normally convenient to keep the diaphragm aperture about aprox 70 80 70 80 30 20 XX of the objective s numerical aperture pic 7 b The closing of the diaphragm reducing the numerical aperture of the optical system increases the focusing depth or penetration power The dif is diaphragm fraction limit must never be surpassed under pre text of increasing the focusing depth c