Home

Instruction manual en Manual de instrucciones es testo 460

image

Contents

1. Only operate the measuring instrument properly for its intended purpose and within the parameters specified in the technical data Do not use force Never store the product together with solvents acids or other aggressive substances Only carry out the maintenance and repair work that is described in the documentation Follow the prescribed steps when doing so Use only OEM spare parts from Testo Protecting the environment gt Take faulty rechargeable batteries as well as spent batteries to the collection points provided for them Send the product back to Testo at the end of its useful life We will ensure that it is disposed of in an environmentally friendly manner Specification 13 Specifications Functions and use The testo 460 is a rotational speed measuring instrument It is normally used to measure rotational speed e g of impellers drums and spindles Technical data Measurement data Further instrument data Sensor Protection class IP40 Optical LED Ambient conditions Parameters 0 50 C 32 122 F rpm rps Storage transport conditions Measuring ranges 40 70 C 40 158 F 100 30000 rpm 1 7 500 0 rps Voltage supply Resolutions 2x 1 5 V type AAA 0 1 rpm 100 0 999 9 rpm Battery life 0 1 rps 1 7 17 0 rps 20 h without display light 1 rpm 1000 30000 rpm Dimensions rps 17 0 500 0 rps 119x46x25 mm inc protection cap Accuracies iah
2. de medici n permitido gt Capacidad restante lt 10 min cambiar las pilas Si no hemos podido resolver sus dudas por favor p ngase en contacto con su distribuidor m s cercano o con el Servicio T cnico de Testo Encontrar los datos de contacto en el libro de garant as y en la p gina de internet www testo com
3. range L Residual capacity lt 10 min Change batteries If we could not answer your question please contact your dealer or Testo Customer Service Contact details can be found on the guarantee card or on the Internet at www testo com Instrucciones breves del testo 460 27 Instrucciones breves del testo 460 Cubierta de protecci n posici n de reposo 2 Fotodiodo Visualizador Teclas de funci n Compartimento para pilas parte Esa posterior Efectuar los ajustes b sicos El instrumento est apagado gt presionar durante 2 s gt seleccionar con A confirmar con S Unidad de velocidad de giro rpm rps Funci n Auto off OFF ON Encender el instrumento Pulsar Encender la iluminaci n del visualizador durante 10 s El instrumento est encendido gt pulsar gt Medir gt Mantener presionado a Seleccionar modo de visualizaci n El instrumento est encendido gt seleccionar con D Hold se mantienen los valores de medici n gt Max valores m ximos gt Min valores m nimos NO ajustar el n mero de aspas Apagar el instrumento El instrumento est encendido gt presionar durante 2 s 28 Seguridad y medio ambiente Seguridad y medio ambiente Indicaciones sobre este documento gt Lea atentamente esta documentaci n y familiaricese con el manejo del producto antes de utilizarlo Guarde la presente documentaci n en un lugar accesible de forma q
4. bre el objeto se deber comprobar si en alg n punto reflectante del mismo existen las propiedades reflectantes suficientes para medir el n mero de revoluciones En este caso apuntar verticalmente hacia el objeto y comprobar si se visualiza alg n valor Si el objeto de medici n emite varias reflexiones por vuelta p ej ventilador reflectante de 5 aspas para medir correctamente el n mero de revoluciones se deber introducir en el instrumento el n mero de impulsos por vuelta p ej n mero de aspas del ventilador V ase m s abajo gt Encender el instrumento gt Pulsar Se inicia el modo de medici n gt Encender la iluminaci n del visualizador Y El instrumento est encendido gt Pulsar La iluminaci n del visualizador se apaga autom ticamente 10 segundos despu s de la ltima pulsaci n de cualquier tecla 32 Utilizar el producto gt Cambiar el modo de visualizaci n ajustar el n mero de aspas B Como alternativa a la introducci n de los impulsos por vuelta p ej n mero de aspas del ventilador se puede seleccionar el ajuste NO 1 y calcular manualmente el n mero de revoluciones a partir del valor de medici n valor de medici n dividido entre impulsos n mero de aspas Modos de visualizaci n ajustables ajuste del n mero de aspas Hold se mantienen los valores de medici n Max valores m ximos desde el inicio de la ltima medici n Min valores m nimos desde e
5. he object to be measured has come to a standstill The ideal distance to the measurement object is between 10 and 40 cm During the measurement always aim at the reflective marker from an angle of approx 30 This avoids multiple reflections from the surface If no reflective marker can be attached it must be checked whether the reflective properties of a reflective point on the measurement object are sufficient for rom measurement In this case aim at the measurement object from the vertical and check whether a value is displayed If the measurement object displays several reflections per revolution e g reflective ventilator with 5 vanes the number of the impulses per revolution e g number of ventilator vanes must be entered into the instrument See below in order to obtain correct rom measurements gt Switching the instrument on gt Press O Measuring mode is opened gt Switching the display light on Y The instrument is switched on gt Press The display light goes out automatically if no key is pressed for 10 seconds gt Changing the display view setting the number of blades B As an alternative to entering the impulses per revolution the setting NO 1 can be selected and the rpm calculated manually from the measurement value measurement value divided by impulse number of vanes Adjustable views setting the number of blades Hold Readings are held Max Maximum values since the s
6. l inicio de la ltima medici n NO ajustar el n mero de aspas Modificar el valor con a gt Pulsar S varias veces hasta que aparezca el modo de visualizaci n deseado gt Medir NA Advertencia Peligro de lesiones debido a las piezas en rotaci n gt Respetar las distancias de seguridad y las medidas de protecci n indicadas por el fabricante gt Mantener presionado a rpm o bien rps parpadea en el visualizador en cuanto haya se ales de medici n aceptables El punto de luz marca el centro de la superficie de medici n Tras finalizar la medici n soltar a el instrumento cambia al modo de visualizaci n Hold el valor de medici n se mantiene gt Apagar el instrumento gt Presionar G hasta que se apague el visualizador Mantenimiento del producto 33 Mantenimiento del producto gt Cambiar las pilas 1 Abrir el compartimento para pilas deslizar hacia abajo la Sai tapa 2 Retirar las pilas agotadas e insertar pilas nuevas 2 de 1 5 V tipo AAA Respetar la polaridad 3 Cerrar el compartimento para pilas deslizar hacia arriba la tapa Limpiar la carcasa gt Limpiar la suciedad de la carcasa con un pa o h medo agua jabonosa No utilizar productos de limpieza o disolventes agresivos Consejos y ayuda Problemas y soluciones Problema Posibles causas soluciones Hio Lo Valores de medici n fuera del rango de medici n demasiado altos o demasiado bajos respetar el rango
7. nstrumentos 29 Descripci n de los instrumentos Funciones y aplicaci n El testo 460 es un instrumento de medici n del n mero de Sai revoluciones por minuto Se utiliza generalmente para medir el numero de revoluciones de molinetes cilindros cintas Datos t cnicos Datos t cnicos Otros datos del instrumento Sensor Clase de protecci n IP40 LED ptico Temperatura ambiental Par metros de medici n 0 50 C 32 122 F rpm rps Temperatura de almacenaje transporte Rangos de medici n 40 70 C 40 158 F 100 30000 rpm 1 7 500 0 rps Alimentaci n Resoluci n 2 pilas de 1 5 V tipo AAA 0 1 rpm 100 0 999 9 rpm Vida til de las pilas a a nA En o n 20 h sin iluminaci n del visualizador rpm rpm iiag 1 rps 17 0 500 0 rps ero 119x46x25 mm incl pilas y cubierta Exactitud a temperatura nominal de de protecci n 22 C 1 d gito Peso 85 g i i i 85 g incl pilas y cubierta de 0 02 del v m protecci n Frecuencia de medici n 0 5s Directivas normas y controles Directiva CE 89 336 CEE Garant a Duraci n 2 a os Condiciones de garant a v ase el libro de garant as 30 Descripci n del producto Descripcion del producto Vista general Cubierta de protecci n posici n de reposo 2 Diodo LED Visualizador Teclas de funci n Compartimento para pilas parte posterior Primeros pasos gt Col
8. ocar las pilas 1 Abrir el compartimento para pilas deslizar hacia abajo la tapa 2 Colocar las pilas 2 de 1 5 V tipo AAA Respetar la polaridad 3 Cerrar el compartimento para pilas deslizar hacia arriba la tapa gt Efectuar los ajustes b sicos modo de configuraci n Funciones ajustables Unidad de velocidad de giro rpm rps Funci n Auto off OFF desconectado ON conectado el instrumento se desconecta autom ticamente 10 minutos despu s de la ltima pulsaci n de cualquier tecla 1 Al encender el instrumento presionar hasta que aparezca A y lt en el visualizador modo de configuraci n Se mostrar la funci n que se puede ajustar El ajuste actual parpadea 2 Pulsar a A varias veces hasta que parpadee el ajuste deseado 3 Pulsar S 4 para confirmar la entrada 4 Repetir los pasos 2 y 3 para todas las funciones El instrumento cambia al modo de medici n Utilizar el producto 31 Utilizar el producto Para conseguir mediciones fiables se deben utilizar cintas reflectantes Colocar la cinta reflectante sobre el objeto que se va a medir s lo cuando se encuentre inm vil La distancia ptima hasta el objeto de medici n est Ea comprendida entre 10 y 40 cm Durante la medici n apunte siempre sobre la cinta reflectante manteniendo un ngulo de 30 De esta forma se evitan las reflexiones m ltiples de la superficie Si no se puede colocar ninguna cinta reflectante so
9. t i Weight 85 g inc batteries and Nominal temperatue 22 C 1 Digit dar A n 0 02 of reading Measuring rate Directives standards and tests 0 5s EC Directive 89 336 EEC Warranty Duration 2 years Warranty conditions see guarantee card 14 Product description Product description At a glance Protection cap Park position Photo transistor Display Control keys Battery compartment on rear o o First steps gt Inserting batteries 1 To open the battery compartment push the battery cover down 2 Insert batteries 2x 1 5 V type AAA Observe the polarity 3 To close the battery compartment push the battery cover back on gt Basic settings configuration mode Adjustable functions Unit of rotational speed rpm rps Auto off function OFF ON instrument switches off automatically if no key is pressed for 10 minutes 1 When switching the instrument on press and hold until A and appear on the display configuration mode The adjustable function is displayed The current setting flashes 2 Press A several times until the desired setting flashes 3 Press D lt 4 to confirm the input 4 Repeat steps 2 and 3 for all functions The instrument changes to measuring mode Using the product 15 Using the product For reliable measurements it is necessary to use reflective a markers To apply the reflective marker wait until t
10. tart of the last measurement Min Minimum values since the start of the last measurement NO Set the number of blades Change the value with a gt Press S several times until the desired view appears 16 Using the product gt Measuring Warning Danger of injury from rotating parts gt Keep to the safe distances and comply with the protective measures stipulated by the manufacturer gt Press and hold fpm or rps flashes in the display as soon as useful measuring signals are available The light soot marks the centre of the measuring surface When measurement has been completed release a the instrument changes to the Hold display view and the reading is held gt Switching the instrument off gt Press G and hold until the display goes out Maintaining the product 17 Maintaining the product gt Changing batteries E 1 To open the battery compartment push the battery cover down 2 Remove used batteries and insert new batteries 2x 1 5 V type AAA Observe the polarity 3 To close the battery compartment push the battery cover back on gt Cleaning the housing gt Clean the housing with a moist cloth soap suds if it is dirty Do not use aggressive cleaning agents or solvents Tips and assistance Questions and answers Question Possible causes solutions Hi or Lo Readings outside the measuring range too high too low Keep to the permitted measuring
11. testo 460 Instruction manual en Manual de instrucciones es Short manual testo 460 11 Short manual testo 460 Protection cap Park position Photo transistor ae O Display Control keys Battery compartment on rear Basic settings Instrument off gt press and hold 2s gt select with A confirm with 63 Unit of rotational speed rpm rps Auto off function OFF ON Switching the instrument on Press Switching the display light on for 10s Instrument on gt press Measuring gt Press and hold Select display mode Instrument on gt select with Z Hold Readings are held gt Max Maximum values gt Min Minimum values gt NO Set the number of blades Switching the instrument off Instrument on gt press and hold O 2s 12 Safety and the enviroment Safety and the environment About this document gt Please read this documentation through carefully and familiarise yourself with the product before putting it to use Keep this document to hand so that you can refer to it when necessary Hand this documentation on to any subsequent users of the product Pay particular attention to information emphasised by the following symbols With the signal word Warning Warns against hazards which could result in serious physical injury if the precautionary measures indicated are not taken E Important Avoid personal injury damage to equipment gt
12. ue se pueda consultar cuando sea necesario Entregue la documentaci n a posteriores usuarios de este producto gt Preste especial atenci n a la informaci n resaltada mediante los siguientes s mbolos Con la palabra Advertencia Esta informaci n advierte de peligros que pueden suponer graves lesiones corporales en caso de que no se respeten las medidas de precauci n indicadas Bl indicaci n importante Evitar da os personales materiales gt Utilizar el producto s lo de acuerdo con el uso previsto y observando los par metros predeterminados que figuran en los datos t cnicos No forzar el instrumento gt No almacenar nunca el producto junto con disolventes cidos u otras substancias agresivas gt Realizar nicamente los trabajos de mantenimiento que vienen descritos en la documentaci n respetando siempre los pasos indicados Utilizar solamente piezas de repuesto originales de Testo Protecci n del medio ambiente gt Llevar las bater as averiadas y las pilas agotadas a los puntos de recogida previstos al efecto gt Enviar el producto a Testo al t rmino de su vida til Nosotros nos ocuparemos de eliminarlo ecol gicamente E Atenci n su producto est marcado con este simbolo Significa que los productos el ctricos y electr nicos ma Usados no deber an mezclarse con los residuos dom sticos generales Existe un sistema de recogida independiente para estos productos Descripci n de los i

Download Pdf Manuals

image

Related Search

Related Contents

SOYO Motherboard SY-6IEB User's Manual  Samsung SGH-U608 用户手册  V7 Sleeve with additional pockets for all iPads - Green  User`s manual and tutorial for the image rotating RISTRA OPO  パーソナルpHメータ PH82  BRCF1760W Frost Free Fridge Freezer  User Guide PDF  Amplificadores H1 360.4  取扱説明書 フロントウインカー リロケーションキット  LIVELLO LASER ROTANTE  