DeVilbiss 5-Liter Series Oxygen Concentrator Service Manual
Contents
1. Phi Me ret P PUES 22 24 Interior VIEWS p etat tt d A Us Mes wu a V catt a i 25 37 Other Figures P mE 38 42 Pneumatic and Wiring Diagrams 1 2 22 43 44 WARRANTY INFORMATION rt n ews ER SEN IRE E S e acea 45 ORDERING INFORMATION AND PARTS LIST Ordering Information ss u l tee Ve e ae er p e ee bd gece ua ad nde at decla dd 46 Parts Return and Ordering 222 4 46 A E A ru pst E RR BR EN RR RI RUM ARR EAR CN RE eed 47 48 SPECIFICATIONS Sovete A A EARS uda Ure RE dead 49 LT 1928 2 GENERAL INFORMATION INTRODUCTION This service manual was designed to provide Sunrise Medical Respiratory Products Division qualified service technicians and homecare providers with the proper maintenance ser vice safety and repair procedures for the DeVilbiss Oxygen Concentrator Read and understand all the information contained in this ser vice manual before attempting to operate or perform any maintenance on the concentrator An oxygen concentrator is a device that delivers highly con centrated oxygen for therapeutic applications Room air is a mixture of 78 nitrogen 21 oxygen 1 argon and other gases Th
2. Precauciones de Seguridad y Advertencias Generales enn DESEMBALAJE Y AJUSTE INSPECCI N Inicial s SaaS 53 Ajuste del Pacie te a cia 53 Instr cciores de F ncionamiento SSSA ES SERENA 54 MANTENIMIENTO Sistema de Alerta dell Paciente RSS SESS Manutencao de Rotna del Pa E Mantenimiento Preventivo Peri dico del Proveedor de Asistencia Domiciliaria Manutencio Entre Pacientes ile ea Resumen de Mantenimiento Preventivo ou ccccsessssssssssesesessscsssscsesescsessssesssesesessssssesesesesessesesesesesessssssesesesessssesesesesesesssseseseseseseeseseseseseseseeseaeseseseseeceneaes SOLUCI N DE PROBLEMAS En to e RR Ere o ee dU ERR e EAE BE NEXU o ERR lr RE RE 57 Secuencia de Funcionamiento Normal iio 57 Soluci n de Problemas Simplificado ett nel iaa Ed Sol ci n de Problemas Guadf o aaa nuan t eer uqa una i au n BRE RR EHE nana liada Soluci n de Problemas Cuadro B S l ci n de Problemas Cuadro sioe ie ei Soluci n de Problemas Cuadro D o Sol ci n de Problemas Cuadro E aae N Sol ci n de Problemas Cuadro EE RD Ge RR a ERRARE PRU
3. 95 Devoluci n de Partes y Politica de rdenes a rro 95 Listado de Partes eo tee ral de Rede c e d be ect a C aa senda 96 97 u u Sm a 98 LT 1928 50 INFORMACION GENERAL INTRODUCCION Este Manual de Servicio esta dise ado para proveer a los t cnicos de servicio calificados y proveedores de Asistencia Domiciliaria pertenecientes a la Division de Productos M dicos Respiratorios de Sunrise Medical con procedimien tos apropiados de mantenimiento servicio seguridad y reparacion para el Concentrador de Oxigeno DeVilbiss Leer y entender toda la informaci n contenida en este man ual de servicios antes de poner en funcionamiento o realizar cualquier maniobra de mantenimiento en el concentrador Cualquier concentrador de ox geno es un equipo que provee ox geno altamente concentrado para aplicaciones terap uticas El aire ambiental es una combinaci n de 78 de Nitr geno 21 de Ox geno 1 de Arg n y otros gases El concen trador absorbe el aire ambiental separa el nitr geno del ox geno y provee ox geno concentrado al paciente a trav s del puerto de ox geno Para un mayor y m s profundo curso de capacitaci n Sunrise Medical provee cursos de servicio de concen tradores de ox geno Para obtener informaci n sobre los cursos de servicio contacte nuestro Departamento de Servicios T cnicos Respiratorios al 1 800 333 4000 814 443 4881 NOTA Sunrise Me
4. LT 1928 FIGURAS DIAGRAMAS Y VISTAS Figura 2 Vista Lateral Externa Cubierta Posterior Cubierta Frontal Tornillo de Gabinete Correa del Cord n El ctrico Cord n El ctrico Rueda Rejilla de Escape LT 1928 72 FIGURAS DIAGRAMAS Y VISTAS Figura 3 Vista Posterior Externa Tornillos de Gabinete Filtro de Aire Puerta de Filtro Puerto de Comunicaci n sin la cubierta Tornillos de Gabinete LT 1928 FIGURAS DIAGRAMAS Y VISTAS Figura 4 Vista Posterior Interior BI5ADS y 515ADZ Tornillos de Babero Filtro de Ingreso de Bacterias QQ Puerto de Comunicaci n Tanque de Acumulaci n Abrazadera Escalonada Manguera de Escape de Presi n Silicona 1 2 pulg D I Compresor Montaje del Motor Conector de Compresor Placa de Montaje del Compresor Cord n Ventilador de Tornillo de Placa de El ctrico y Enfriamiento Montaje Pasacables en unidades de 115 voltios 74 LT 1928 14 FIGURAS DIAGRAMAS Y VISTAS Figura 4 A Vista Inferior Posterior Interior 515 5 y 515AKZ Conector del Conector IEC para los Compresor Modelos 515AKS y 515AKZ LT 1928 FIGURAS DIAGRAMAS Y VISTAS Figura 5 Vista Frontal Interna 515ADS y 515ADZ Placa PC Tamiz Izquierdo V lvula de Aire Direccional Rotativa Arn s de Cableado Tamiz Derecho Ligaduras de Cables
5. Power Failure Blinking red Service Required light and pulsing audible alarm Low Flow Below 0 5 Ipm Continuous red Service Required light and audible alarm Below Normal Oxygen 84 to 75 yellow Low Oxygen light 7576 to 6076 yellow Low Oxygen light and beeping audible alarm Less than 6076 red Service Required light and beeping audible alarm Refer to Specifications for spe cific alarm settings The visible and audible alarms will activate for approximately 15 minutes in a no power situation If the unit is turned On without power or power is removed later no alarm will sound for the first 10 seconds After that time the alarm will produce an audible pulse every few seconds while the visible alarm blinks Power for this alarm is provided by a capacitor on the PC board NOTE If the concentrator has been unused for an extended period the unit must run several minutes before the power fail alarm will activate The PC printed circuit board Figure 5 is responsible for controlling the system and alarms NOTE A high pressure condition is indicated by the audible a popping sound release of pressure from a pressure relief valve located on the compressor head ROUTINE PATIENT MAINTENANCE The oxygen patient should perform the following maintenance Oxygen Humidifier reusable bottles only The patient should clean the humidifier bottle daily The patient should follow the instructions
6. AVM OS I DeVilbiss 5 Liter Series Oxygen For Product Numbers Para productos Numero Concentrator Service Manual pee 515ADZ e e DeVilbiss Concentrador de Oxigeno 515AKS 4 27 515AKZ Serie de 5 Litros Manual de Servicio DANGER NO SMOKING PELIGRO NO FUMAR 2 L o 9 CAUTION Federal U S A law restricts this device to sale by or on the order of a physician SUNRISE MEDICAL PRECAUCION Las leyes Federales EE UU restringen la venta de este equipo por o por orden de un m dico TABLE OF CONTENTS ESPA OL MR 50 GENERAL INFORMATION 2 225 E CO Si aad WEINE WE on eee NER 3 Important Safeguards bi Velde pits e aa E RE Ca da bed NEN A e We beau ea a 4 Safety Precautions and General Warnings 1 4 UNPACKING AND SETUP Initial Inspection hoi ras dade esie dele ee er 5 Patient Setup EROS ISS t e anasu Se SRS XR Dd 5 Operating Instructions 2 eel REDE We RR EE ad SAA ehe 6 MAINTENANCE Patient Alert System veh net tebe see a Cad odes 0 dey dae ie RUPES re ees d ETE 7 Routine Patient Mairiteriance Raw a A veau kuya Ear eR Ve Ok ee re SSA REX E RS 7 Periodic Homecare Provider Prevent
7. Conectar el adaptador recto de humidificador parte 444 506 a la botella haciendo girar la tuerca de mariposa en la unidad del humidificador hasta que est ajustada Asegurar la botella en la correa Conectar un extremo del tubo de ox geno en la salida de ox geno de la unidad y el otro extremo al adaptador pl stico en el humidificador Colocar el humidificador cerca del paciente Cuando est listo para funcionar Conectar la c nula nasal parte CANDO el cat ter o la mas carilla al tubo de ox geno seg n las instrucciones del fabri cante Seguir las instrucciones de funcionamiento que se detallan a continuaci n LT 1928 DESEMBALAJE Y AJUSTE INSTRUCCIONES DE FUNCIONAMIENTO Retirar completamente el cord n el ctrico de la correa Asegurar que el interruptor de encendido est en la posici n OFF Unidades de 115 Voltios Insertar el enchufe en un tomacorriente El Concentrador de Ox geno DeVilbiss utiliza un enchufe polarizado de dos puntas y est doblemente aisla do para protecci n contra choques el ctricos A ADVERTENCIA ADVERTENCIA El enchufe de los concentradores DeVilbiss 515ADZ y 515ADS posee una punta m s ancha que la otra Para reducir el riesgo de choque el ctrico este enchufe s lo se adaptar al tomacorriente en una sola posici n No intentar alter ar esta caracter stica de seguridad A ADVERTENCIA ADVERTENCIA El uso inapropiado del cable el ctrico y los ench
8. 4 Remove the two screws from the back of the compressor mounting plate s Figure 4 5 Remove the two 10 mm hex nuts that secure the mounting plate to the front of the compressor housing Figure 6 These nuts are located on each side of the rotary valve 6 Lift compressor and mounting plate up and out of the com pressor housing area 7 Remove the tubing from the compressor intake port fitting CAUTION If the unit has been running recently the compressor may be hot REPAIR AND REPLACEMENT To inspect and or replace internal components Figures IIA amp 12A Remove the eight screws that hold the compressor heads in place When removing the heads be sure to keep each head and its components with the correct compressor side 2 Check for proper placement of or damage to the gaskets on the bottom of the compressor heads Replace if damaged 3 Remove reed valve plates A reed valve is located on each side of the valve plate 4 The compressor reed valves should be flush with the valve plate If the valve is broken or not flush with the valve plate or foreign matter is detected inside the head clean or replace the compressor reed valves To replace the compressor reed valves Figures 11 amp 12A a Remove the screw holding the compressor reed valves in position on the valve plate and discard the used reed valves b Position the new reed valves so that they are centered and completely cover the holes in t
9. nent part s return s the original purchaser shall be responsi ble to remove the defective component part s For all returns the original purchaser must 1 properly package the unit or parts in a Sunrise approved shipping container 2 properly identify the claim with the Return Authorization Number and 3 send the shipment freight prepaid Service under this warranty must be performed by a qualified Sunrise provider and or an Authorized Service Center Labor costs incurred by a qualified Sunrise provider are excluded and will not be reimbursed Only labor costs incurred by an Authorized Service Center are reimbursed INFORMATION NOTE This warranty does not obligate Sunrise to provide a loaner unit during the time that an oxygen concentrator is undergoing repair NOTE Replacement components are warranted for the unexpired portion of the original Limited Warranty This warranty shall be voided and Sunrise shall be relieved of any obligation or liability if The equipment is operated or maintained outside the parameters indicated in the Sunrise operating and service instructions Unqualified service personnel conduct routine maintenance servicing or warranty repairs Unauthorized parts or components i e regenerated sieve material are used to repair or alter the equipment Unapproved filters are used with the unit THERE IS NO OTHER EXPRESS WARRANTY IMPLIED WAR RANTIES INCLUDING THOSE OF MERCHANTAB
10. y 12 Si los filtros no estan obturados no se registran p rdidas compresor deber ser removido y reparado o reemplazado Para retirar el compresor Asegurar que la unidad est desenchufada del tomacorriente 2 Desconectar los cables del compresor desconectando el conector el ctrico del compresor Figura 4 3 Retirar la abrazadera escalonada y la manguera de la boca de salida en el compresor Figura 4 4 Retirar los dos tornillos en la parte posterior de la s placa s de montaje del compresor Figura 4 5 Retirar las dos tuercas hexagonales de 10 mm que sujetan la placa de montaje al frente del compresor Figura 6 Estas tuercas est n ubicadas a cada lado de la v lvula rotativa 6 Levantar el compresor y la placa de montaje hacia arriba y fuera del alojamiento del compresor 7 Retirar los tubos del accesorio del puerto de entrada del compresor PRECAUCI N Si la unidad ha estado funcionando recientemente el compresor puede estar caliente Para inspeccionar y o reemplazar componentes inter nos Figuras A y 12 A Retirar los ocho tornillos que sujetan los cabezales del compresor en su lugar Al retirar los cabezales asegurarse de mantener cada uno de ellos y sus componentes con el lado correcto del compresor 2 Verificar la ubicaci n apropiada o el da o a las juntas en la parte inferior de los cabezales del compresor Reemplazar si est n da adas 3 Retirar las placas de
11. e 14 Accumulator Pressure Test E NER Ge eS s 15 Capacitor ces RE HH tard E e rr uus ET 15 COITIDFeSSOF x Su Guede o A ea eis etes E HE n ders qe m 15 Cooling Fan seso it estet Ea poteet e sensed EE E 17 Final GhecleValve Po Soa Sac a 17 Flow Meter cine b abal AAA Na SES DEUS 17 Hour Meter ives IE mI EE FERONT item ae to gite 18 Molecular Sieve Beds a ua repara ASA AAA RR EN ER S S ER nh ade dobre 18 Power Cord i a us aa A one A Mes GE DA A A Cae A Se uro qe ifs 18 Power Switch y es Le ee A 18 Pressure Regulator lt gt SE dae weder Find Ste SORESA Ua Esa aa 18 Printed Circuit Board PC board 2 cnt es CESARE DE a DE eo ene AR a ee 19 Rotary Valves Gro A O eiue ce treo ut verte tur e Y 19 Sieve Bed Check Valves ive ROSE Roche RP re e SONS A ea 20 FIGURES DIAGRAMS AND VIEWS Fig re Index 2 555 E E e REQUE adea Uu EE LET Ag alas eal Stand 21 Exterior VIEWS uo u tea
12. la unidad deber dejarse funcionando durante 20 minutos antes de que la alarma de falla de energ a sea habilitada Esta alarma est alimentada por un condensador en la placa PC PRECAUCI N No aplicar fuerza o presi n sobre la placa PC cuando se conecten o desconecten compo nentes electr nicos o neum ticos Podr a da arse el ensamble electr nico Para retirar y reemplazar la Placa PC Figura 7 l Verificar que la unidad est desenchufada del tomacorriente Retirar los gabinetes anterior y posterior y el babero Desconectar todos los cables terminales y conectores Retirar los tubos de 1 8 pulg 3 2 mm conectados al sensor BRAY Retirar los tornillos que ajustan la placa a la unidad y retirar la placa PC 6 Instalar la nueva placa PC y asegurarla utilizando el tornillo 7 Reconectar todos los cables el ctricos terminales conectores y el tubo al sensor 5 8 Reemplazar el babero y los gabinetes anterior y posterior V LVULA ROTATIVA FIGURA 13 La v lvula rotativa temporizada alternativamente distribuye 6 presi n provista por el compresor a los tamices Mientras un tamiz est siendo presurizado el otro est siendo despresurizado a trav s de la v lvula Los gases liberados l pasan por el puerto de escape de la v lvula y salen a trav s de la boca de escape La v lvula contiene dos discos giratorios que reciben su energ a de un motor de velocidad gradual para ciclar la pre si n en
13. n de Sunrise bajo la presente garant a est limitada a 1 reparaci n de producto repuestos y mano de obra en sus instalaciones de fabricaci n o en un Centro de Servicio Autorizado o 2 provisi n de reemplazos de parte s componente s La pre sente garant a no cubre cualquier costo de mano de obra incurrido en el retiro o reemplazo de parte s componente s garantizadas por cualquier otro personal de servicio Los ele mentos de mantenimiento de rutina tales como filtros no est n cubiertos bajo la presente garant a Env o de Reclamos de Garant a El comprador original deber enviar cualquier reclamo de garant a a Sunrise o a un Centro de Servicio Autorizado Luego de la verificaci n del estado de la garant a se emitir n las instrucciones correspondientes Para la devoluci n de parte s componente s el comprador original deber ser responsable de retirar la parte s componente s defectuosa Para todas las devoluciones el comprador original deber 1 embalar apropiadamente la unidad o partes en un contenedor de env o aprobado por Sunrise 2 identificar apropiadamente el reclamo con el N mero de Autorizaci n de Devoluci n y 3 enviar con costo de env o prepago El servicio bajo la presente garant a deber ser realizado por cualquier proveedor califica do de Sunrise y o Centro de Servicio Autorizado Los costos de mano de obra incurridos por un proveedor calificado de Sunrise est n excluidos y no ser n reemb
14. n del acumulador por debajo de los niveles m nimos NOTA Un compresor defectuoso se manifestar por un lento incremento en la presi n La presi n s lo puede alcanzar cierto nivel y luego detenerse Los niveles bajos de concentraci n de ox geno y las presiones del acumulador m s altas de lo normal pueden indicar tamices defectuosos Los tamices severamente contaminados pueden provocar tambi n que la v lvula de alivio de presi n del com presor se abra NOTA Una v lvula rotativa que no funciona adecuadamente tambi n causar alta presi n en el tanque de acumulaci n y la activaci n de la v lvula de alivio de presi n En este caso deber determinarse si el problema se debe a los tamices la REPARACI N Y REEMPLAZO v lvula o ambos CONDENSADOR El condensador permite al compresor comenzar a funcionar proveyendo voltaje a las serpentinas del motor del compresor Un condensador defectuoso resultar en un compresor que funciona m s lentamente o que no enciende PRECAUCI N Los concentradores de las series 515ADS y 515ADZ utilizan un compresor Thomas con un condensador de 15 mfd o un compresor GSE con un condensador de 22 mfd De ser necesario reemplazarlo asegurar la instalaci n de un condensador apropiado PRECAUCI N Los concentradores de las series 515AKS y 515AKZ utilizan un compresor GSE con un condensador de 10 mfd De ser necesario reemplazarlo asegurar la instalaci n de un condensador apropiado
15. Do not use regenerated sieve material LT 1928 UNPACKING AND SETUP INITIAL INSPECTION It is suggested that an initial inspection be performed upon receiving the oxygen concentrator After removing the DeVilbiss Oxygen Concentrator from the carton examine it for any external damage If shipping damage has occurred contact the Sunrise Medical Customer Service Department at 800 333 4000 814 443 4881 for specific instructions Save the carton for possible later return note the position of the unit and placement of the packing material 2 Open the filter door Figure 3 and record the number of hours on the hour meter Check to make sure the air filter is in place 3 Check to be sure the intake bacteria filter Figure 4 is in place 4 Plug the unit into an electrical outlet turn the unit and check the audible and visible alarms 5 Set the flow meter to maximum recommended liter flow and let the unit run for at least 20 minutes 6 Use an oxygen analyzer to check the concentration NOTE If the unit fails to operate properly oxygen concen tration not within specification or if internal damage is found contact the Sunrise Medical Customer Service Department at 1 800 333 4000 814 443 4881 PATIENT SETUP I Position the unit near an electrical outlet the room where the patient spends most of his or her time NOTE Do not connect to an electrical outlet controlled by a wall switch The
16. Exhaust Grill LT 1928 Figure 3 Exterior Rear View Air FIGURES Filter DIAGRAMS Cabinet Screws AND VIEWS Cabinet Screws Filter Door Communication Port with cover removed FIGURES DIAGRAMS AND VIEWS Figure 4 Interior Rear View 515ADS and 515ADZ Bib Screws Intake Bacteria Filter Accumulator Tank Communication Port Ladder Clamp Pressure Exhaust Hose Silicone 1 2 ID Compressor Compressor Connector Motor Mount Compressor Mounting Plate Power Cord amp Cooling Fan Mounting Plate Screw Strain Relief on 115 volt units FIGURES DIAGRAMS AND VIEWS Figure 4A Interior Rear Lower View 515AKS and 515AKZ Compressor IEC Connector for Connector models 51 5AKS and 515AKZ LT 1928 26 FIGURES DIAGRAMS AND VIEWS Figure 5 Interior Front View 515ADS and 515ADZ Wire Harness PC Board Right Sieve Bed Left Sieve Bed Sieve Bed Cable Ties Rotary Directional Air Valve LT 1928 FIGURES DIAGRAMS AND VIEWS Figure 5A Interior Front View 515AKS and 515 7 NOTE Sieve beds shown are used only on models 515AKS and 515AKZ Wire Harness PC Board Right Sieve Bed Left Sieve Bed Sieve Bed Cable Ties Rotary Directional Air Valve LT 1928 28 FIGURES DIAGRAMS AND VIEWS Figure 5B Interior Front View Close Up 515AK
17. Figure 8 Make sure the unit is unplugged from the wall outlet Remove the front and back cabinets From behind the bib remove the 2 hoses from the flow meter dos ob While squeezing tabs on flow meter brackets push the flow meter through the bib 5 Install new flow meter in bib and reconnect hoses COMPONENT TESTING HOUR METER To replace the hour meter Figure 9 Make sure the unit is unplugged from the wall outlet Remove the front and back cabinets and loosen the bib 2 3 Disconnect the hour meter connector from the PC board 4 Remove the meter by carefully inserting a small flat screwdriv er under the outer edge of meter and prying upward Install a new hour meter by applying downward pressure until it snaps into position Connect the hour meter to the PC board CAUTION Do not apply any force or flex the PC Board when connecting or disconnecting electronic or pneumatic components Damage to the electronic assembly is possible MOLECULAR SIEVE BEDS The two molecular sieve beds alternately remove the nitrogen from the air passing through them and provide the patient with a constant supply of oxygen The efficiency of the molecular sieve material will be impaired if it becomes contaminated by moisture Contamination causes the molecular sieve material to lose its nitrogen adsorbing properties resulting in a decrease in oxygen concentration The unit should run for a minimum of 20 minutes
18. INSTRUCTIONS Remove the power cord completely from the strap Make sure the power switch is in the Off position 2 115 Volt Units Insert the plug into an electrical outlet The DeVilbiss Oxygen Concentrator uses a two prong polarized plug and is double insulated to protect against electric shock A WARNING WARNING The plug on the DeVilbiss 515ADZ and 515ADS concentrators has one blade wider than the other To reduce the risk of electric shock this plug is intended to fit in a wall outlet only one way Do not attempt to defeat this safety feature A WARNING WARNING Improper use of the power cord and plugs can cause a burn fire or other electric shock hazards Do not use the unit if the power cord is damaged A WARNING S WARNING Oxygen promotes rapid burning Do not smoke when using this unit or when near a per son receiving oxygen therapy Do not operate the oxygen concentrator within a minimum of five feet 1 6m from hot sparking or burning objects or naked flames Do not use in rooms heated by paraffin or portable gas heaters 3 Press the power switch to the On position When the unit is turned on the Service Required light will illuminate and an audible signal will sound the patient alert system momentari ly The Power light also illuminates Only DeVilbiss Oxygen Concentrators with OSD The OSD is an optional device within DeVilbiss concentrators that monitors the oxygen prod
19. INTERRUPTOR DE ENCENDIDO Para reemplazar el interruptor de encendido Figura 8 I Verificar que la unidad est desenchufada del tomacorriente 2 Retirar los gabinetes anterior y posterior y aflojar el babero 3 Tomar nota de la posici n de los cables y el interruptor antes de retirar los cables de los terminales del interruptor 4 Mientras se presionan las pestafias de calce a los costados del interruptor empujar el interruptor fuera de la parte anterior del babero 5 Instalar el nuevo interruptor la orientaci n correcta asegu rando que calce en su posici n 6 Reconectar los cables a los terminales del interruptor REGULADOR DE PRESI N El regulador de presi n estabiliza el flujo de ox geno al paciente y establece la contrapresi n en el sistema Est prea justado a 8 5 psi 58 6 kPa y no deber ser necesario ajustarlo en el rea Para probar el regulador de presi n Encender la unidad ON 2 Ajustar el medidor de flujo a 2 Ipm 3 Utilizar un medidor de presi n vac o parte PVO2D 601 y una boca que calce en la salida de ox geno o en una corta por ci n de tubo conectada a la salida 4 Si el medidor de presi n vac o tiene una lectura diferente a 8 5 0 5 psi 58 6 5 9 kPa con la salida del medidor bloqueada podr a requerirse un ajuste del regulador de presi n En este caso contactar al Departamento de Atenci n al Cliente de Sunrise Medical al 1 800 333 4000 814 443 4881 NO
20. IUE UM Or VILDISS Motor Mount Compressor Mounting Plate Electrical Connector LT 1928 FIGURES DIAGRAMS AND VIEWS Figure 2 GSE Compressor With Head Removed Reed Valve Compressor Rod Compressor Valve Plate Electrical Connector Head Head casket top view Retainer Flapper Screw Sleeve Cup Seal Cup Retainer Pressure Relief Valve Cup Retainer Rod Screw Connecting Tubes Valve Plate Reed Valve Rod Screw bottom view O Ring or Flapper Screw FIGURES DIAGRAMS AND VIEWS Figure 13 Rotary Valve Stepper Motor Valve Wire Harness Left Sieve Right Sieve Bed Fitting Bed Fitting Front View Pressure Exhaust Fitting Pressure Intake Fitting Rear View LT 1928 42 DIAGRAMS AND VIEWS FIGURES Figure 14 Pneumatic Diagram 211 ING YS 01 YHS lt 15 T T BOLON H3dd3ls IOd Nn PM X NN HOION Los i E MOSS 3udNO2 E BOLIOVdVO NOSS3NdNOO wove nny 222209020202 YO ILHA 1 1 0 7 TES EE ine GINA LIOASII 1vViSOWN3Hi Y HIM G3ddI0D3 LON 573000 MO NMOBS gt 11 a JONYNdd J z TSINA LON 02 lt a 71 31 HIM lt omo e magn non NO
21. Tamices contaminados Reemplazar Alarma audible no suena La unidad no ha sido utilizada por un per odo de Reemplazar durante fallas de energ a tiempo prolongado NOTA En este caso la unidad debe funcionar por algunos minutos antes de que la alarma de falla de energ a se active Placa PC defectuosa Reemplazar Interruptor de encendido defectuoso Reemplazar Arn s de cables defectuoso Reemplazar Alarma audible no suena cuan Placa PC defectuosa Reemplazar do la unidad est encendida V lvula de alivio de presi n Placa PC defectuosa Reemplazar Sonido de chasqui v lvula rotativa defectuosa Reemplazar Luz de Servicio Requerido Placa defectuosa Reemplazar no enciende cuando la unidad Luz defectuosa Reemplazar est encendida Conectores de placa PC no est n sujetos Asegurar que las pestafias est n en su lugar LT 1928 SOLUCION DE PROBLEMAS Alarma Visible Parpadeante Otros Sintomas SOLUCION DE PROBLEMAS CUADRO B Alarma Audible Pulsante OFF Causa Posible Compresor Luz de Encendido OFF Soluciones Posibles Ventilador apagado Cable mal instalado defectuoso Insertar enchufe en el receptaculo o reemplazar cable No hay energia en el receptaculo Verificar el interruptor de circuito o fusible Verificar el cableado dom stico por un electricista calificado El circuito puede estar totalmente sobrecargado con otros aparatos Podr a requerir otro recept c
22. Test Point Fitting Accumulator Tank T Fitting E Fixed Orifice em Sieve Bed Test Point Tubing to Flow Meter Sieve Bed Test Point LT 1928 36 FIGURES DIAGRAMS AND VIEWS Figure 10 Interior Rear Lower View Intake Filter Elbow Exhaust Muffler Valve Exhaust Hose 1 2 ID Braided j 7743 5 DON a lt lt lt foe L 1 Pressure Exhaust Hose Silicone 1 2 ID e 7 4 Ladder Clamp Compressor 2 Cooling Fan Guard Cooling Fan Power Cord note correct orienta Connector tion of fan guard Capacitor FIGURES DIAGRAMS AND VIEWS Figure I Compressor Thomas Nexus Intake Fitting Pressure Relief Valve Exhaust Fitting 1 2 1 gt Motor Mount Electrical Connector Compressor Mounting Plate LT 1928 38 FIGURES DIAGRAMS AND VIEWS Figure Thomas Nexus Compressor With Head Removed Reed Valve Head Valve Plate Retainer Screw or Flapper Gasket top view Rod Sleeve Cup Retainer Plate Rod Screw Cup Seal Valve Screw Compressor Reed Valve Compressor Head Pressure Relief Valve Exhaust Fitting Valve Plate O Ring or Flapper bottom view FIGURES DIAGRAMS AND VIEWS Figure 12 Compressor GSE Pressure Relief Valve Intake Fitting Exhaust Fitting HAM ANS i i DT 1 c HUN
23. cm Pressure Relief Valve Operating System 40 psig 5psig 276 kPa 34 5 kPa Time Cycle Pressure Swing For units equipped with an OSD the visible low oxygen indicator will activate at the following level 84 2 The audible alarm will alert at approximately 75 At less than 60 the red service required light will activate Storage Conditions 40 C to 70 C humidity range of 10 to 100 including condensation Equipment Class and Type Class II Equipment Double Insulated Z Applied Part Approval Body and Safety Standard CSA CAN CSA C22 2 No 601 1 90 TUV EN60601 1 1990 A1 A2 A13 EN 1508359 1996 2 Compliance 60601 1 2 Typical Accumulator Pressure at Maximum Recommended Flow Sea Level 750 M 2460 Ft 1500 M 4921 Ft 22 27 psig 150 190 kPa 20 25 psig 138 172 kPa 17 21 psig 120 145 kPa Note The OSD performance at 10 C to 35 C 95 R H through voltage range on the 515ADS verified at 670m Specifications subject to change without notice Note EN 1508359 1996 does not apply to 515AKZ Note Wattage listed is for 50 Hz operation wattage at 60 Hz will be higher TABLA DE CONTENIDOS INFORMACION GENERAL UN Medidas de Seguridad Importantes
24. de Tamices LT 1928 FIGURAS DIAGRAMAS Y VISTAS Figure 5 A Interior Front View 515 5 515AKZ NOTA Los tamices mostrados s lo se usan en los modelos 515AKS y 515AKZ Arn s de Cableado Placa PC Tamiz Derecho Tamiz Izquierdo Ligaduras de Cables de Tamices V lvula de Aire Direccional Rotativa FIGURAS DIAGRAMAS Y VISTAS Figura 5 B Acercamiento Vista Frontal Interna 515AKS y 515AKZ Silenciador de Entrada para los Modelos 515 5 y 515 2 LT 1928 78 FIGURAS DIAGRAMAS Y VISTAS Figure 6 Vista Inferior Frontal Interna BI5ADS y 515ADZ Ranura de j 1 bo Montaje de Babero Ig FUG 77 1 Y Arn s de Cableado de Valvula Rotativa Ligaduras de Cable de Montaje de Valvula Rotativa Valvula de Aire Direccional Rotativa Manguera de Tuerca Hexagonal de Manguera de Tamiz Izquierdo Placa de Montaje de Tamiz Derecho Compresor FIGURAS DIAGRAMAS Y VISTAS Figure 6 A Vista Inferior Frontal Interna 515AKS y 515AKZ NOTA Los tamices mostrados s lo se usan en los modelos 515AKS y 515AKZ Ranura de Montaje de Babero Arn s de Cableado de V lvula Rotativa P Abrazaderas de Manguera de Montaje de V lvula Rotativa V lvula de Aire Direccional Rotativa w gt a Manguera de Tuerca Hexagonal de Manguera de Tamiz Tamiz Izquierdo Placa de Montaje de Derecho Combreso
25. first while the left bed depressurizes to approximately 2 PSI 14 kPa 3 Voltage is again applied to the valve for a short time The valve will stop for approximately 0 7 seconds During this time the sieve bed pressures are equalized 4 short DC voltage signal is again applied to the valve The valve will stop for approximately 7 seconds causing the left bed to pressurize while the right bed depressurizes to approx imately 2 PSI 14 kPa 5 A short DC voltage signal is again applied to the valve The valve will stop for approximately 0 7 seconds During this time the sieve bed pressures are equalized 6 The cycle then repeats with step 2 NOTE In the turn down mode OSD models only the fixed cycle time is decreased to approximately 3 seconds and the bed pressure equalization time to approximately 0 3 seconds NOTE High end sieve bed pressure should not exceed 1 2 PSI 4 kPa above high end accumulator pressures Refer to Specifications for normal pressures obtained during the cycle LT 1928 TROUBLESHOOTING SIMPLIFIED TROUBLESHOOTING The key to simple troubleshooting is to recognize which type of problem exists and select the most effective approach to solving the problem The different types of problems and the approaches for solutions are as follows I The unit runs but a low pressure and flow or high pressure condition exists NOTE Low pressure or flow are indicated by both a v
26. los cables y la manguera Insertar la pesta a del babero en la ranura que se encuentra sobre la v lvula de rotaci n y empujar la unidad hasta que calce en su lugar Asegurar el babero con los dos tornillos PRUEBA DE ACUMULADOR DE PRESI N Para verificar la presi n del acumulador Asegurar que la unidad est apagada OFF Retirar los gabinetes anterior y posterior Utilizar el medidor de presi n vac o parte PVO2D 601 y el conjunto de prueba de presi n parte 303DZ 637 inclui dos en el Kit de Servicio Retirar la tapa de los tubos del accesorio del tanque de acu mulaci n y colocar el tubo de 1 16 pulg 1 6 mm de di metro desde el medidor al accesorio del tanque de acumulaci n recientemente liberado Encender la unidad con la misma velocidad de flujo ajustada al m ximo recomendado Durante cada ciclo la presi n promedio en el acumulador de ox geno aumentar y disminuir NOTA Las presiones normales observadas dependen de la altura y de la velocidad de flujo El incremento de la altura y de la velocidad de flujo disminuir n levemente las presiones del acumulador Las disminuciones en las dos variables incremen tar n levemente las presiones del acumulador La oscilaci n aceptable de presi n del acumulador en varias alturas al flujo m ximo recomendado est n identificadas en la secci n de Especificaciones NOTA Una v lvula de control defectuosa puede causar una r pida ca da en la presi
27. necessary be sure the correct capaci tor is installed CAUTION The 515AKS and 515AKZ concentrators use a GSE compressor with a 10 mfd capacitor If replacement is necessary be sure the correct capaci tor is installed A WARNING WARNING Electric Shock Hazard When replacing the capacitor do not touch the terminals or allow metal objects to come in contact with the terminals on the capacitor The capacitor may hold a charge for several days after the unit is turned off If a defective capacitor is suspected a new one must be installed The capacitor is strapped into a well molded into the bottom of the unit Figure 10 next to the cooling fan To replace the capacitor Make sure the unit is unplugged from the wall outlet Remove the front and back cabinets Remove the compressor Disconnect the two wires from the terminals on the capacitor Cut the nylon cable tie holding the capacitor in place and remove the capacitor Reconnect the wires to the new capacitor Install the new capacitor and secure with a new cable tie Replace the compressor Replace the front and back cabinets and secure with the six screws COMPRESSOR The DeVilbiss Oxygen Concentrator uses a double head oil free compressor The compressor is secured to the compres sor plate with four rubber motor mounts A compressor that is worn or defective may cause pressure to rise slowly cause excessive noise and or vibration cause l
28. ox geno bajo La alarma audible alertar a aproximadamente 75 se activar en este nivel A menos del 60 la luz roja de Servicio Requerido se activar Condiciones de Almacenamiento 40 C a 70 C rango de humedad del 10 al 100 incluyendo condensaci n Tipo y Clase de Equipo Equipo Clase II con Doble Aislamiento Tipo B Parte Aplicada Entidad de Aprobaci n CSA TUV CAN CSA C22 2 601 1 90 EN60601 1 1990 A1 A2 A13 Norma de Seguridad EN 1508359 1996 Cumplimiento de EMC a IEC 60601 1 2 Presi n de Acumulador T pica a M ximo Flujo Recomendado Nivel del Mar 22 27 psig 150 190 kPa 750 M 2 460 pies 20 25 psig 138 172 kPa 1 500 M 4 921 pies 17 21 psig 120 145 kPa Nota El funcionamiento del OSD a 10 C a 35 C 95 H R a trav s de rango de voltage en el 515ADS verificado a 670 m Especificaciones sujetas a cambio sin previo aviso 2Nota EN 1508359 1996 no se aplica al 515AKZ Nota Los vatios enumerados son para el funcionamiento a 50 Hz los vatios ser n mayores a 60 Hz LT 1928 SUNRISE MEDICAL Sunrise Medical Respiratory Products Division 100 DeVilbiss Drive Somerset PA 15501 800 333 4000 In Canada 800 263 3390 814 443 4881 Sunrise Medical 04 07 LT 1928 Rev A
29. pulg Punta 10 Punta Allen 5 32 pulg Punta Allen 5 64 pulg Punta Allen 9 64 pulg Punta Allen 7 64 pulg 2 Medidor de Presi n vac o parte PCO2D 601 Caja de herramientas 2 Accesorios de prueba parte 303DZ 637 Destornillador Torx con puntas Luz de prueba CA CC Llave de matraca de 1 4 pulg Punta hexagonal de 3mm Llave T 15 Torx L Dado de 10 mm con entrada de 1 4 pulg Extensi n con entrada de 1 4 pulg Caja de almacenamiento pl stica Herramienta pl stica de indicador de errores Juntamente con el Equipo de Servicio se necesitar un anal izador de ox geno parte O2ANA para verificar peri dica mente los niveles de concentraci n de ox geno Un volt metro ser necesario para una prueba de voltaje m s precisa NOTA Asegurarse de leer todos los pasos antes de comenzar cualquiera de los procedimientos de este manual NOTA Despu s de reparar o reemplazar un componente hacer funcionar la unidad por 20 minutos verificar la concen traci n de ox geno y verificar si existen p rdidas Comprobar si existen p rdidas utilizando una soluci n de detecci n para tal fin como SWAGELOK MS Snoop o equivalente no debe contener glicol etileno Aplicar la solu ci n de detecci n de p rdidas a todas las juntas y conexiones de mangueras mientras la unidad se encuentra funcionando Si existe una p rdida de aire la soluci n comenzar a burbujear Todas las p rdidas deben ser reparadas antes de
30. refer to low oxygen concentration symptom in Troubleshooting Chart A Both OSD lights are illuminated Defective OSD Replace PC board Yellow Low Oxygen light is illuminated Yellow Low Oxygen light is illuminated and an intermittent audible alarm sounds every five seconds Oxygen level is low Oxygen level is low Check concentration with an oxygen analyzer If the concentration is within specification replace the PC board If the concentration is low refer to low oxygen concentration symptom in Troubleshooting Chart A Check concentration with an oxygen analyzer If the concentration is within specification replace the PC board If the concentration is low refer to low oxygen concentration symptom in Troubleshooting Chart A Refer to Specifications page for oxygen purity levels LT 1928 COMPONENT TESTING REPAIR AND REPLACEMENT PROPER REPAIR PROCEDURES The DeVilbiss Oxygen Concentrator is designed for ease of service To aid service personnel a Service Kit part 444 501 is available which contains the necessary gauges tools and testing instruments to properly service the oxygen concentra tor On parts that are sold separately the part number is indi cated in parenthesis The following parts are included in the Service Kit Slotted bit amp Phillips bit 2 Phillips kit 7 16 Socket 1 4 Drive Crescent wrench 8 Duckbill plie
31. se enciende El nivel de ox geno es bajo Verificar concentraci n con analizador de ox geno Si est dentro de especificaciones reemplazar la Placa Si la concentraci n es baja ver s ntomas de baja concentraci n de ox geno en Soluci n de Problemas Cuadro A Luz amarilla de Bajo Ox geno se enciende y la alarma audi ble intermitente suena cada cinco segundos El nivel de ox geno es bajo Verificar concentraci n con analizador de ox geno Si est dentro de especificaciones reemplazar la Placa Si la concentraci n es baja ver s ntomas de baja concentraci n de ox geno en Soluci n de Problemas Cuadro A Ver la p gina de Especificaciones para niveles de pureza de ox geno LT 1928 PRUEBA DE COMPONENTES REPARACION Y REEMPLAZO PROCEDIMIENTOS DE REPARACION APROPIA DOS El Concentrador de Oxigeno DeVilbiss esta disefiado para facilitar el servicio Para ayudar al personal de servicio se encuentra disponible un Kit de Servicios parte 444 501 el cual contiene los medidores herramientas e instrumentos de prueba necesarios para dar mantenimiento apropiadamente al concentrador de oxigeno Las partes que se venden por separa do muestran el n mero de parte indicado entre par ntesis Las siguientes partes est n incluidas en el Kit de Servicio Punta ranurada Punta Philips 1 Punta Philips 2 Dado de 7 16 pulg con entrada de 1 4 pulg Llave inglesa Pinza Duckbill de 8
32. than nor mal bed pressures a High pressures indicate defective sieve beds b Low pressures indicate occluded filters leaks or defective compressor c Uneven bed pressures indicate valve is not operating cor rectly Check for oxygen leaks at sieve beds flow meter accumulator tank final bacteria filter pressure regulator final check valve outlet port NOTE Check for leaks using a certified leak detection solu tion such as Snoop or equivalent must not contain ethylene glycol Apply leak test solution to all fittings and hose connec tions with unit running If an air leak is present the solution will bubble All leaks should be repaired before putting the unit back in service CAUTION Do not apply leak test solution to any part of the rotary valve or the main PC Board assembly Test accumulator tank pressure If pressure is lower than nor mal then sieve bed check valves are defective NOTE For normal system pressures refer to Specifications TROUBLESHOOTING TROUBLESHOOTING CHART A Visible Alarm Audible Alarm Compressor Power Light OFF OFF ON ON Other Symptoms Possible Cause Possible Remedy Pulsating air noise Excessive noise Intake filter not in place or defective Check filter and replace if necessary Compressor intake hose disconnected Loose or defective motor mounts Reconnect hose Replace motor mounts Defective compr
33. the wall outlet Remove the front and back cabinets and loosen or remove the bib 2 3 Remove the hose from the outlet side of the final bacteria filter 4 Remove the two screws from the back of the oxygen outlet fitting assembly and remove the assembly m Remove the hose from each end of the final check valve 6 Attach the hoses to a new check valve Make sure that the flat side of the check valve is directed toward the oxygen outlet fitting 7 Replace the outlet fitting assembly and connect the hose to the filter 8 Replace the bib and front and back cabinets FLOW METER The pressure compensated flow meter has an accuracy level of 5 at full scale exception 0 5 at 5 Ipm The flow meter on the DeVilbiss Oxygen Concentrator is designed for use at 8 5 psi 58 6 kPa at flow rates up to 5 Ipm To check for leaks in the flow meter tubing Check for leaks using a certified leak detection solution such as Snoop or equivalent must not contain ethylene glycol 2 Apply leak test solution to all fittings and hose connections with the unit running CAUTION Do not apply leak test solution to any part of the rotary valve or the main PC Board assembly 3 If an air leak is present the solution will bubble All leaks should be repaired before putting the concentrator back in service A WARNING WARNING Electric Shock Hazard Use caution when leak testing near electrical connections To replace the flow meter
34. una liberaci n audible de presi n como un chasquido desde una v lvula de liberaci n de presi n ubicada en el cabezal del com presor MANTENIMIENTO DE RUTINA DEL PACIENTE El paciente que utiliza ox geno deber realizar el siguiente mantenimiento Humidificador de Ox geno s lo botellas reutilizables El paciente deber limpiar la botella del humidificador diaria mente El paciente deber seguir las instrucciones provistas por el fabricante Si no se proveyeron instrucciones de limpieza deberan seguirse estos pasos Lavar la botella del humidificador en una soluci n de agua caliente y detergente lavavajillas Sumergir el humidificador en una soluci n de una parte de vinagre blanco y tres partes de agua caliente por 30 45 minutos La soluci n act a como agente germicida Enjuagar profundamente bajo el agua caliente del grifo y rellenar con agua destilada No sobrecargar C nula Mascarilla y Tubos El paciente deber limpiar y reponer la c nula o mascarilla y los tubos seg n instrucciones del fabricante Filtro de Aire y Conector de Salida de Ox geno El filtro de aire Figura 3 y el conector de salida de ox geno deber n ser limpiados por el paciente por lo menos una vez a la semana Para limpiarlos deber seguir los pasos siguientes Retirar el filtro de aire ubicado en la parte posterior de la unidad Retirar el conector de salida de ox geno si es utilizado del puerto de sa
35. 2 Screws Cabinet 303DZ 628 Sieve Bed 3152 Valve Rotary 515 2 702 Valve Hose Braided Formed 515ADZ 649 Wire Harnesses 51505 08 Tools LT 1928 SPECIFICATIONS Catalog Number DEVILBISS 5 LITER SERIES 515ADS 515ADZ 515AKS 515AKZ Delivery Rate Lower delivery rates 1 to 5 LPM available for low flow applications Maximum Recommended Flow 5 LPM Outlet Pressure 8 5 psig 58 6 kPa Electrical Rating 115V 60 Hz 4 1 Amp 220V 230V 50 Hz 1 7A 230V 60 Hz 2 4A Operating Voltage Range 97 127V 60 Hz 187 264V 50 Hz 200V 264V 60 Hz Oxygen Percentage 1 5 LPM 93 3 Operating Altitude tested at 21 C only 0 1500 M 0 4921 ft 1500 3000 M 4921 9842 ft 3000 4000 M 9842 13123 ft Operating Environment Range 10 C to 35 C humidity range of 30 to 75 Power Consumption Across the voltage range No degradation of performance Tested at nominal voltage only No degradation in performance expected based on tests of similar products Not recommended Not tested No degradation in performance across the operating voltage range tested at sea level 385 Watts Average 515ADS only 320 Watts 2 5 LPM amp below 335 Watts Average 515AKS only 301 Watts 2 5 LPM amp below Weight Sound Level ISO 8359 1996 from front 50 lbs 22 7 Kilograms 48 0 dbA overall average Dimensions 27 75 H x 16 W x 14 D 70 5 x 40 6 x 35 6
36. 2 pulg D I Trenzada i SL REI l Manguera de Escape de Presi n Silicona de 1 2 pulg D I imm e e Abrazadera Escalonada Conector del Protector de Ventilador de Compresor Conector de Ventilador de Enfriamiento Cord n Enfriamiento El ctrico notar la correcta Condensador orientaci n del protector LT 1928 86 FIGURAS DIAGRAMAS Y VISTAS Figura 11 Compresor Thomas Nexus Boca de Entrada Valvula de Alivio de Presi n Boca de Escape 22912142 Montura de Motor Conector El ctrico Placa de Montaje del Compresor FIGURAS DIAGRAMAS Y VISTAS Figura A Compresor Thomas Nexus sin Cabezal V lvula de Boquilla Junta de V lvula Ret n Tornillo Aleta rincipal vista superior Vastago Manga Placa de Ret n de Taza Tornillo de V stago Junta Herm tica Tornillo de V lvula Boca de Escape V lvula de Boquilla Cabezal de V lvula de Alivio de Compresor Placa de Anillo O o Aleta Compresor de Presi n V lvula vista inferior LT 1928 88 FIGURAS DIAGRAMAS Y VISTAS Figura 12 Compresor GSE Boca de Entrada DT Boca de Escape V lvula de Alivio de Presi n TETTE TTD Or VILBISS Montura de Motor Placa de Montaje de Compresor Conector El ctrico LT 1928 FIGURAS DIAGRAMAS Y VISTAS Figura 12 Compresor GSE sin Cabezal Placa de 2 V lvul V
37. 3 Figura 9 Parte superior ADS 2 2 2 2 2 84 Figura 9A Parte s perior AKS y AKZ d lei oes Cv e tg 85 Figura 10 Acercamiento Inferior 2 86 Otras Figuras Figura 11 Compresor Thomas Nexus retirado de la 87 Figura 11 A Compresor Thomas Nexus sin 88 Figura 12 Compresor GSE retirado de la 89 Figura 12 A Compresor GSE sin 90 Figura 13 V lvula RotatiVa cc ri AA A AAA ee UE E DE ea eas 91 Diagramas Figura 14 Diagrama Neum ticos A VER eR ee e 92 Figura 15 LO A ie Veh PIE MN vi epe aie eerie 93 LT 1928 70 FIGURAS Figura Vista Frontal Externa Luces Indicadoras de OSD en unidades con OSD solamente Panel de Luces Puerto de Salida de Ox geno Babero Cubierta Frontal Rueda DIAGRAMAS Y VISTAS Luz de Servicio Requerido Rojo Base Luz de Encendido verde Interruptor de Encendido Perilla de Medidor de Flujo Medidor de Flujo Interruptor de Circuito
38. A ADVERTENCIA ADVERTENCIA Peligro de choque el ctrico Al reem plazar el condensador no tocar las terminales o per mitir que objetos met licos est n en contacto con las terminales del condensador El condensador puede retener carga por varios d as despu s de que la unidad ha sido apagada Si se sospecha un condensador defectuoso debe ser instalado uno nuevo El condensador est sujeto dentro de un molde en la parte inferior de la unidad Figura 10 junto al ventilador de enfriamiento Para reemplazar el condensador Asegurar que la unidad est desenchufada del tomacorriente 2 Retirar los gabinetes anterior y posterior 3 Retirar el compresor 4 Desconectar los dos cables de las terminales del condensador 5 Cortar el cable de nylon que sujeta al condensador en su lugar y retirar el condensador Reconectar los cables al nuevo condensador 7 Instalar el nuevo condensador y asegurarlo con un nuevo cable 8 Recolocar el compresor 9 Recolocar los gabinetes anterior y posterior y asegurar con los seis tornillos COMPRESOR El Concentrador de Ox geno DeVilbiss utiliza un compresor de doble cabezal sin aceite El compresor est asegurado a la placa del compresor con cuatro monturas de motor de goma Un compresor gastado o defectuoso puede e Provocar el incremento lento de la presi n Provocar excesivo ruido y o vibraci n Provocar bajas concentraciones de ox geno Un compresor gastado o defectuo
39. DEX Exterior Views The following figures show the exterior of the DeVilbiss 5 Liter Series Concentrators Figure ii diate A A uu Sy Figure 2 Side pees ies AOL CRX USER UR Machu sae es Figure 3 a Stowe dots oe Ea ele PEEL e ttis Interior Views The following figures show the interior of the DeVilbiss 5 Liter Series Concentrators Figure 4 R ar ADS amp ADZ x nk reef tia seats VPE diate haki bleed a Figure 4A Rear ARSBARZ ii CIA a tad e AR es Figure 5 Front ADS 8 ADZ 2 er erat Me rete ied etre e ri Pa e Peay acl Figure 5A Front AKS AKZ sre o Ver Des em teure Free Cree ur we ER DE ER Figure 5B Frorit close up AKS ae viue eee RR O A Figure 6 Lower front close up ADS amp ADZ Figure 6A Lower front close up AKS amp 2 Figure 7 Upper front close up ADS amp ADZ Figure 7A Upper front close up AKS amp 2 22 Figure 8 Behind bib 2 cen ey e eere rper RU Figure 9 Top ADS RADZI iii rur a ete he oue
40. EBA DE COMPONENTES REPARACI N Y REEMPLAZO Procedimientos de Reparaci n Apropiados nn ua tentent tentent 62 Remoci n del Gabinete sukaman cas Prueba de Acumulador de Presi n Co ser Taro ae ccc UE Ventilador de Enfriamiento ERE ra e E RR eR E Y UR d Ye Ra Medid er de Fluje n ND EE id Tamices Molecular Cord n El CtriCo r u A ese EUR Interruptor de Corriente Reg lador de Presi n pim apa Placa de Circuitos Impresos Placa PC uu tete a s 68 V lvula ROtativa ai anh awaqa n wauu qunas aswaa Qayaw qaa a awasqa ayaka aaa 68 V lvulas de Control de ai aaa 69 FIGURAS DIAGRAMAS Y VISTAS ndice de Figuras Diagramas Y Vistas ii rra 70 Vistas 71 73 Vistas Interiores w Otras RIgUTAS sarria ia coa qu qaku ahua aus a aan ha au auqa aan Diagramas Neum ticos y de Cableado ll u u 92 93 INFORMACI N DE GARANTIA ia 94 INFORMACION DE ORDENESY LISTADO DE PARTES Informacion de rdenes
41. EN LIMITACIONES SOBRE EL PERIODO DE DURACI N DE UNA GARANT A IMPL CITA O LA LIMITACI N O EXCLUSI N DE DANO CONSECUENTE O INCIDENTAL POR LO QUE LA LIMITACI N O EXCLUSI N ARRIBA MENCIONADA PUEDE NO SER APLICABLE A USTED La presente garant a provee derechos legales espec ficos y usted puede adem s poseer otros derechos que var an de estado a estado NOTA Las garantias para modelos internacionales podrian variar LT 1928 INFORMACION DE ORDENES Y LISTADO DE PARTES INFORMACION DE ORDENES Al ordenar componentes guias de instrucciones o manuales de servicio debe proveer lo siguiente N mero de cat logo de la unidad Numero de serie de la unidad N mero de parte Cantidad requerida Gu a de Instrucciones del Concentrador DeVilbiss parte A 515A English Spanish French Gu a de Instrucciones del Concentrador DeVilbiss parte SE 515A English Spanish French German Italian Dutch Guia de Instrucciones del Concentrador DeVilbiss parte P 515A English Portuguese Manual de Servicio Serie DeVilbiss 5 Litros parte LT 1917 English Manual de Servicio Serie DeVilbiss 5 Litros parte LT 1927 English Portuguese Manual de Servicio Serie DeVilbiss 5 Litros parte LT 1928 English Spanish Las rdenes pueden ser realizadas llamando al 800 333 4000 800 333 4000 905 660 2459 814 443 488 Servicio de Atenci n al Cliente e Repuestos Garantizados EE UU Canad Depa
42. ILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE LIMITED TO THE DURATION OF THE EXPRESS LIMITED WARRAN TY AND TO THE EXTENT PERMITTED BY LAW ANY AND ALL IMPLIED WARRANTIES ARE EXCLUDED THIS IS THE EXCLUSIVE REMEDY AND LIABILITY FOR CONSEQUEN TIAL AND INCIDENTAL DAMAGES UNDER ANY AND ALL WARRANTIES ARE EXCLUDED TO THE EXTENT EXCLU SION IS PERMITTED BY LAW SOME STATES DO NOT ALLOW LIMITATIONS ON HOW LONG AN IMPLIED WARRANTY LASTS OR THE LIMITATION OR EXCLUSION OF CONSEQUENTIAL OR INCIDENTAL DAMAGE SO THE ABOVE LIMITATION OR EXCLUSION MAY NOT APPLY TO YOU This warranty gives you specific legal rights and you may also have other rights that vary from state to state NOTE International model warranties may vary LT 1928 ORDERING INFORMATION AND PARTS LIST ORDERING INFORMATION When ordering components instruction guides or service manuals the following must be provided Unit Catalog Number Unit Serial Number Part Number Quantity Required DeVilbiss Concentrator Instruction Guide part A 515A English Spanish French DeVilbiss Concentrator Instruction Guide part SE 515A English Spanish French German ltalian Dutch DeVilbiss Concentrator Instruction Guide part P 515A English Portuguese DeVilbiss 5 Liter Series Service Manual part LT 1917 English DeVilbiss 5 Liter Series Service Manual part LT 1927 English Portuguese DeVilbiss 5 Liter Series Service Manual part LT 1928 En
43. IONOD lt lt E 12 ISOLMOA STT 30 S3dVOINN 01V1S0WH31 NOD SOGVdlnt3 ON SOTIOOW Vaid ee T h Ovid 130 WOVYING 1104 QEZ 30 S30vOIND Wale J lt L BOOWTILNSA 04934 A gt rt f owna L B 0 cx 2 01 ET lt oaran OBS3N NQUAVA OYDAN 30 BOQII3IA wes Tf gt zar YYWINA 30 05 35 ENT er iA x E i nm lt 01I02415 30 1 SOLOA STT 30 S30 OINN vid E oros E 29 YVYWONA 32 H miter gt ror en 0812313 77 5310H1N09 gt gt 9 Sr 30 wor s3om 30 13d tar esr EI 2 gt Lon 5 e n NOIOVOINDINOO 30 O1N3nd oo os 1 INFORMACION DE GARANT A INFORMACI N DE GARANT A Garant a Limitada del Concentrador de Ox geno DeVilbiss 515ADS 515ADZ Susnrise Medical Sunrise garantiza los Concentradores de Ox geno DeVilbiss 515ADS 515ADZ bajo las condiciones limitaciones establecidas a continuaci n Sunrise garantiza que estos equipos est n libres de defectos de manufactura y material por cinco 5 afios desde la fecha de su salida de f brica al comprador original t picamente el proveedor de cuidados de salud Esta garant a est limitada al Comprador de un nuevo equipo adquirido directamente a Sunrise o a uno de sus Proveedores Distribuidores o Agentes La obligaci
44. MAINTENANCE 1 Discard oxygen tubing cannula amp humidifier bottle 2 Wash or replace the cabinet air filter 3 Wash the concentrator cabinet 4 Check oxygen concentration If the unit falls within specifica tion the extended life intake bacteria filter does not need to be replaced between patients PREVENTATIVE MAINTENANCE SUMMARY Patient Daily Clean the humidifier bottle if used Weekly Clean air filter on back of unit Clean exterior of cabinet Other Clean and replace cannula mask and tubing as instructed by manufacturer Homecare Provider Change intake bacteria filter as necessary following requirements in step 3 3 months Check oxygen concentration on non OSD units Check the concentrator environment and set a maintenance interval of less than 3 months if required 2 years Check audible alarm and indicator lights Change final bacteria filter change within 17 520 hours Check system performance Check oxygen concentration on OSD units LT 1928 TROUBLESHOOTING SYSTEM OPERATION The DeVilbiss Oxygen Concentrator uses a pressure swing adsorption system The air is drawn into the unit through air fileers and into a double head compressor A pneumatic diagram of the system is shown in Figure 14 The compressed air passes through a rotary valve Figure 5 which is cycled at a pre determined rate and is directed into one of two sieve beds The sieve beds contain molecular sieve material whi
45. NCIA unidad Utilizar s lo materiales que son compatibles con el ox geno B NS ADVERTENCIA El ox geno promueve incineraci n r pida No fumar mientras utiliza A ADVERTENCIA esta unidad o estando cerca de una persona que recibe terapia de oxigeno No operar el concentrador de J ADVERTENCIA Peligro de Choque El ctrico Al ox geno dentro de un radio m nimo de cinco pies 1 6 reemplazar el condensador no tocar las terminales o m de objetos calientes chispeantes o quem ndose o permitir que objetos met licos est n en contacto con de llamas No utilizar en habitaciones con calefacci n las terminales del condensador El condensador puede a base de parafina o calentadores port tiles de gas contener carga durante varios d as luego de que la MEE unidad ha sido apagada El condensador esta ubicado C No colocar el humidificador con un paciente que requiere M oxigeno menos que est prescrito por un profesional m di la base qe ta unidad a un costado del ventiladorde LE EN enfriamiento co y luego solo debera utilizarse un humidificador de tipo bur buja K Utilizar s lo repuestos y accesorios para concentradores de 4 Sunrise Medical D No conectar el concentrador de oxigeno a un tomacorriente controlado por un interruptor de pared el tomacorriente L No utilizar material para tamices regenerado deber a estar independiente de otros artefactos E utilizar un adaptador el ctrico o cable de extensi n con el concen
46. NE NwOMB l 4 0 E aei 22 gt gt LINDO SUNA ION SH 3904 ouvog 4 SIOHLNOO 5 DINOY19373 9 aNvd 1H911 k 180d NOLLYOINNINNOD Figure 15 Wiring Diagram LT 1928 WARRANTY WARRANTY INFORMATION DeVilbiss 515ADS 515ADZ Oxygen Concentrators Limited Warranty Sunrise Medical Sunrise warrants the DeVilbiss 5 5ADS 515ADZ Oxygen Concentrators under the condi tions and limitations stated below Sunrise warrants this equip ment to be free from defects in workmanship and materials for five 5 years from date of factory shipment to the original purchaser typically the healthcare provider This warranty is limited to the Buyer of new equipment purchased directly from Sunrise or one of its Providers Distributors or Agents Sunrise s obligation under this warranty is limited to 1 prod uct repair parts and labor at its factory or at an Authorized Service Center or 2 supplying a replacement for component part s This warranty does not cover any labor charges incurred in removing or replacing the warranted component part s by any other service personnel Routine maintenance items such as filters are not covered under this warranty Warranty Claims Submissions The original purchaser must submit any warranty claim to Sunrise or to an Authorized Service Center Upon verification of the warranty status instructions will be issued For compo
47. RAMAS Y VISTAS Vistas Exteriores Las siguientes figuras muestran el exterior de los Concentradores DeVilbiss Series 5 Litros Figura gis sss ua ROSSA ERR TEE QUO A ARS AS 71 Figura 2 Lateral aya Sean eh IA beatae we a hid sae d e du de CR e eee ees 72 Figura 3 Posterior citrato ipei dos 73 Vistas Interiores Las siguientes figuras muestran el interior de los Concentradores DeVilbiss Series 5 Litros Figura 4 Posterior ADS y ADZ isis EL RM ees Aly AM ats OT D eR EN 74 Figura 4A Posterior ARS y AKZ eer Y REA de ee eR EL ey us but ua e Rene 75 Figura 5 Frontal ADS y ADZ coc albe OR n AREE er e t ade Y eR T esate ERR UR caet eire d iore 76 Figura 5A Frontal AKS y AKZ I u pee awe LS rb ay vedi gb YN MEN UI Qr Cote Cete Due Ste uode 77 Figura 5 B Acercamiento Frontal AKS 2 78 Figura 6 Acercamiento Inferior Frontal ADS y 2 79 Figura 6A Acercamiento Inferior Frontal AKS y 2 80 Figura 7 Acercamiento Superior Frontal ADS 2 81 Figura 7 Acercamiento Superior Frontal AKS y AKZ 82 Figura 8 ERA a eva Cobh ee abe SS a 8
48. Remove the 1 8 3 2mm tubing attached to the sensor UY Remove the screw that secures the board to the unit and remove the PC board 6 Install the new PC board and secure it using the screw 7 Reconnect all electrical wires terminals connectors and the tubing to the sensor 8 Replace the bib and front and back cabinets REPAIR AND REPLACEMENT ROTARY VALVE Figure 13 The timed rotary valve alternately distributes pressure sup plied by the compressor to the sieve beds While one bed is being pressurized the other bed is being exhausted through the valve Exhaust gases go through the valve exhaust port and exit through the exhaust muffler The valve contains two revolving discs powered by a stepper motor to cycle the pressure between the beds DC voltage is supplied by the PC board to the motor windings causing the internal discs to turn and direct pressure to the proper sieve bed If the rotary valve does not shift properly the same bed may continue to pressurize causing the pressure relief valve to release the excess pressure There are several reasons why the rotary valve could malfunc tion therefore the cause of failure must be determined before corrective action can be taken To test the rotary valve Figures 6 amp 7 Remove the front cabinet 2 Connect pressure gauges to the test points at the top of the sieve beds in order to observe unit cycling and bed pressures Refer to the section
49. S and 515 7 Intake Silencer for models 515 5 and 515AKZ Figure 6 Interior Front Lower View 515ADS and 515ADZ Bib Mounting Slot Rotary Directional Air Valve FIGURES DIAGRAMS AND VIEWS Compressor Mounting Rotary Valve Wire Harness Rotary Valve Mounting Cable Ties FIGURES DIAGRAMS AND VIEWS Figure 6A Interior Front Lower View 515AKS and 515AKZ NOTE Sieve beds shown are used only on models 515AKS and 515AKZ Bib Mounting Slot Rotary Directional Air Valve Left Sieve Bed Hose Compressor Mounting Plate Hex Nut Right Sieve Bed Hose Rotary Valve Wire Harness Rotary Valve Mounting Hose Clamps AND VIEWS FIGURES DIAGRAMS Figure 7 Interior Front Upper View 515ADS and 515ADZ Hour Meter Sieve Bed Oxygen Outlet Test Point Tubing 1 8 ID Oxygen Outlet Sieve Bed Tubing 1 8 ID Test Point Wires OSD PC Board Valve Wire Harness Connector Capacitor for Power Fail Alarm CAUTION Do NOT insert mounting screw in this hole Oxygen Flow Meter Tubing shown disconnected PC Board Screw Fuse for removal of Bib an 32 LT 1928 FIGURES DIAGRAMS AND VIEWS Figure Interior Front Upper View 515 5 and 515AKZ NOTE Sieve beds shown are used only on models 515 5 and 515 7 Hour Meter Wires Sieve Bed Test Point Oxygen Outlet Tubing 1 8 ID PC Board Valv
50. TA Asegurar que no existen p rdidas antes de ajustar el regulador de presi n utilizando una soluci n de detecci n de p rdidas tal como Snoop o equivalente no debe contener gli col etileno PRECAUCI N No aplicar la soluci n de detecci n de p rdidas en ninguna parte de la v lvula rotativa o el ensamble de la placa PC Un mal funcionamiento en el regulador de presi n causar una p rdida o fluctuaci n en el flujo de ox geno que ser visto en el medidor o una disminuci n en la concentraci n de ox geno Para reemplazar el regulador de presi n Figura 9 l Verificar que la unidad est desenchufada del tomacorriente 2 Retirar los gabinetes anterior y posterior 3 Retirar las abrazaderas de los tubos y los tubos del regulador de presi n 4 Desatornillar el regulador del tanque de acumulaci n 5 Instalar un nuevo regulador en el tanque de acumulaci n y colocar los tubos y las abrazaderas 6 Recolocar los gabinetes anterior y posterior LT 1928 PRUEBA DE COMPONENTES REPARACION Y REEMPLAZO PLACA DE CIRCUITO IMPRESA 3 La placa de circuito PC es la responsable de monitorear y controlar el Concentrador de Oxigeno DeVilbiss La placa PC contiene alarmas preajustadas para flujo bajo y 4 fallas de energia alguno de los valores de las alarmas fuera excedido el sistema de alerta del paciente sera activado NOTA Si el concentrador hubiera estado fuera de uso por un per odo prolongado
51. TA Usar un medidor de flujo bajo parte 515LF 607 para las velocidades de flujo menores de Ipm NOTA La unidad puede requerir hasta 20 minutos para que la concentraci n de ox geno y la velocidad de flujo se estabili cen La velocidad de flujo deber ser monitoreada y reajustada si fuese necesario El medidor de flujo tiene un dispositivo de bloqueo Si fuese necesario preajustar y bloquear una velocidad de flujo prescri ta apretar el tornillo de ajuste ubicado en la tuerca hexagonal justo debajo de la perilla de control utilizando una llave Allen de 1 16 pulg No se podr realizar un ajuste sin aflojar el tornillo de ajuste previamente El concentrador de ox geno DeVilbiss est ahora listo para ser utilizado LT 1928 SISTEMA DE ALERTA DEL PACIENTE El Sistema de Alerta del Paciente del Concentrador de Ox geno DeVilbiss detecta fallas en los componentes de la unidad Este sistema est compuesto por alarmas visibles y audibles que aler tan al paciente en el caso de mal funcionamiento de la unidad La alarma visible est localizada en el frente del panel Figura 1 y lee Servicio Requerido El sistema de alarma audible est ener gizado internamente no requiere bater as Cuando las luces indi cadoras se encienden o la alarma audible suena en otro momen to que no sea durante el encendido de la unidad es indicador de que ha ocurrido alg n problema Modelos sin OSD Falla de energ a Luz roja in
52. able Marr n o Cables Dobles Negros Cable Rojo Cables Dobles Negros o Cable Marr n Cualquier cable rojo puede ser conec tado a cualquier terminal izquierda del interruptor Cinta de Conexi n de Panel de Luces Pesta a de Babero LT 1928 FIGURAS DIAGRAMAS Y VISTAS Figura 9 Vista Superior Interna 515ADS y 515ADZ Boca de Punto de Orificio Fijo Medidor de Reductor de Regulador de Prueba de Tanque de Tiempo Tubos de 1 4 Presi n Acumulaci n V lvula de Control x 1 8 pulg V lvula de de Tamiz Control de Tamiz Punto de Prueba de Tamiz Arn s de Cableado de Puerto Tubo al Medidor de Flujo Punto de Prueba de de Comunicaci n Tamiz LT 1928 FIGURAS DIAGRAMAS Y VISTAS Figura 9 A Vista Superior Interna 515 5 515AKZ NOTA Los tamices mostrados s lo se usan en los modelos 515AKS y 515AKZ V lvula de Reductor de Tubos Control de Tamiz de 1 4 x 1 8 pulg Medidor de Tiempo Arn s de Cableado de Puerto de Comunicaci n Regulador de Presi n V lvula de Control de Tamiz Boca de Punto de Prueba de Tanque de EE Acumulaci n E j Boca en T de Tanque de Acumulaci n Orificio Fijo Punto de Prueba Punto de Prueba de Tamiz de Tamiz Tubo al Medidor de Flujo FIGURAS DIAGRAMAS Y VISTAS Figura 10 Vista Inferior Posterior Interna Codo de Filtro de Entrada Amortiguador de Escape Manguera de Escape de V lvula de 1
53. acitor Replace capacitor Defective compressor Replace compressor Unit warm to the touch and cannot be restarted for several minutes Compressor overheated due to Occluded filters Restricted input or output air passage Low or high line voltage Defective cooling fan Clean or replace filters Remove obstruction Check line voltage use alternate circuit independent of other appliances Replace cooling fan Defective compressor TROUBLESHOOTING CHART E Visible Alarm Audible Alarm Compressor Replace compressor Power Light ON Other Symptoms ON Possible Cause ON ON Possible Remedy Fluctuating or no flow System pressure below 9 psi 62 1 kPa due to Leak in system Defective compressor Check for leaks in all hoses and fittings Replace compressor LT 1928 TROUBLESHOOTING CHART F OSD UNITS ONLY Compressor TROUBLESHOOTING Power Light ON ON Other Symptoms Possible Cause Possible Remedy No OSD lights are illuminated Defective OSD Check concentration with an oxygen analyzer If the concentration is within specification replace the PC board No OSD lights are illuminated but red Service Required light is illuminated accompanied by a beeping audible alarm Oxygen level is low Check concentration with an oxygen analyzer If the concentration is within specification replace the PC board If the concentration is low
54. al number Hour meter reading for each concentrator Account number Company name and address Description of problem LT 1928 ORDERING INFORMATION AND PARTS LIST Accessories Fung Straight Hum Adapter Components 5I5AKS 515 5I5ADZ 5ISAKZ Cabinet Parts osas SISADS EOS SISADZ 603 SISADZ 603 SISADZ 6167 50102 60 5540233 51540263 55402 65 CAUTION The 515A series concentrators use either a CAUTION 515ADS and 515ADZ series concentra Thomas compressor or a GSE compressor Be sure to tors use either a Thomas compressor with a 5 mfd order the correct part number when replacing rebuild capacitor or a GSE compressor with a 22 mfd capacitor ing the compressor Capacitor is included with purchase The 515AKS and 515AKZ series concentrators use of compressor GSE compressor with 10 mfd capacitor If replacement is necessary be sure the correct capacitor is installed LT 1928 ORDERING INFORMATION AND PARTS LIST Components continued 5I5AKS 515ADS 5IS5ADZ 5I5AKZ Fittings Accumulator nk Te Flow Meters AE z Hoses 18 ID Blue 14 ID Blue G Lg 12 ID Slicone 2 Lg 444 56 Hour Meter PVSLD 617 Light Panel 50502 615 Motor Mounts 50512609 Nut Compressor Mounting 303DZ 630 Oxygen Outer Port 30302 60 PC Board SISAKZ 622 Power Cord PVSLD 618 Power Cord Strap MC29D 657 Power Cord Strain Relief 505DZ 645 Power Switch 5052506 Pressure Regulator MC29D 612 MC29D 61
55. apagar y la luz roja de Requiere Servicio se encender NOTA Ver Especificaciones para ajustes de alarmas espec ficos Cuando la unidad con OSD se encuentra encendida ON las cuatro luces indicadoras Encendido Requiere Servicio Ox geno Bajo y Ox geno Normal en el panel frontal se encender n por un momento Luego de algunos segundos s lo las luces Encendido y Ox geno Normal permanecer n encendidas NOTA Luego de encender la unidad ON el OSD lleva a cabo un diagn stico evaluativo continuo para verificar fallas electr nicas Si el OSD detecta esta condici n durante el fun cionamiento del concentrador la luz verde del OSD Ox geno Normal se apagar se activar una se al audible y parpadear la luz roja de Requiere Servicio De no ocurrir esto y durante los primeros quince minutos de funcionamien to la luz verde de Ox geno Normal permanecer encendida durante el proceso de estabilizaci n de ox geno Luego de este tiempo el OSD comenzar a monitorear la pureza del ox geno cada segundo Mover lentamente la perilla de medidor de flujo hasta que la bola del medidor est centrada en la l nea junto a la velocidad de flujo apropiada NOTA Cuando la perilla del medidor de flujo es girada en la direcci n de las agujas del reloj el flujo decrece y eventual mente cerrar el flujo de ox geno Cuando la perilla es girada en el sentido contrario el flujo aumenta NO
56. ar el condensador para determinar la necesidad de reemplazarlo el condensador est incluido con la compra del compresor Si el condensador tiene el valor incorrecto para el compresor o desea reemplazarlo ver la secci n Condensador 3 Reconectar los tubos a la boca de entrada del compresor 4 Colocar el compresor en la base de la unidad para que los pernos en la placa de montaje est n alineados con las muescas en el frente de la base de la unidad 5 Asegurar la placa de montaje con dos tornillos en la parte posterior e instalar las tuercas en la parte anterior de la placa 6 Reconectar la manguera a la boca de salida del compresor REPARACI N Y REEMPLAZO 7 Reconectar el conector el ctrico del compresor Para retirar el compresor de la placa de montaje Voltear el compresor hacia abajo para que descanse sobre los cabezales Retirar las cuatro tuercas hexagonales de montaje del compresor y la placa de montaje Para retirar las monturas del motor Desatornillar manualmente las monturas del motor de los pies del compresor VENTILADOR DE ENFRIAMIENTO El ventilador de enfriamiento provee un flujo de aire constante para enfriar el compresor El ventilador est ubicado en la parte inferior de la unidad debajo del compresor Figura 10 Un ventilador defectuoso puede causar que el dispositivo ter moprotector interno del compresor se active y apague el compresor Si esto ocurriera el compresor requerir varios m
57. as Instalar una nueva v lvula rotativa revirtiendo el procedimiento mencionado NOTA Utilizar la menor fuerza posible al ajustar las ligaduras de los cables alrededor de la v lvula rotativa El exceso de ajuste de las mismas puede transmitir ruido al gabinete incre mentando el nivel de ruido LT 1928 PRUEBA DE COMPONENTES REPARACION Y REEMPLAZO VALVULAS DE CONTROL DE TAMICES Una valvula de control Figura 9 ubicada entre la salida de cada tamiz y el tanque de acumulaci n permite que el ox geno pase a trav s de los tamices al tanque de acumulaci n cuando la presi n del tamiz es mayor que la presi n del tanque de acumulaci n Estas v lvulas tambi n previenen un flujo rever tido de ox geno desde el acumulador a los tamices Realizar la prueba de presi n del acumulador descrita en la secci n Prueba Reparaci n y Reemplazo de Componentes La misma explica los detalles sobre la colocaci n del medidor La oscilaci n de presi n aceptable a diversas alturas est incluida en la secci n de Especificaciones Una v lvula de control defectuosa provocar la disminuci n de la concentraci n de ox geno y menor presi n del acumulador Para reemplazar una v lvula de control defectuosa retirar el tubo a cada lado de la misma e instalar una nueva v lvula ase gurando que la punta de salida est en direcci n al tanque de acumulaci n LT 1928 FIGURAS DIAGRAMAS Y VISTAS iNDICE DE FIGURAS DIAG
58. as disponibles para aplicaciones de flujo bajo 515AKS 515AKZ Flujo M ximo Recomendado 5 LPM 8 5 psig 58 6 kPa Presi n de Salida Valuaci n El ctrica 115V 60 Hz 4 1 Amp 220 V 230 V 50 Hz 1 7 A 230 V 60 Hz 2 4 A 97 127V 60 Hz 187 264 V 50 Hz 200 264 V 60 Hz Rango de Voltaje Operativo Porcentaje de Ox geno 1 5 LPM 93 3 Altura Operativa probado a 21 C solamente A trav s de rango de voltaje 0 1 500 M 0 4 921 pies Sin degradaci n del rendimiento Probado a voltaje nominal solamente 1 500 3 000 M 4 921 9 842 pies Sin degradaci n del rendimiento esperado basado en pruebas en productos similares 3 000 4 000 M 9 842 13 123 pies No recomendado No probado Rango de Ambiente Operativo Sin degradaci n de rendimiento a trav s 10 C a 35 C rango de humedad del rango de voltaje operativo de 30 a 75 probado al nivel del mar Consumo de Energ a 385 vatios Promedio 335 vatios Promedio 515ADS solamente 320 vatios 515AKS solamente 301 vatios e 2 5 LPM y menos 2 5 LPM y menos Peso 50 lbs 22 7 kg Nivel de Sonido ISO 8359 1996 desde el frente 48 0 dbA promedio general Dimensiones 27 75 Alto x 16 Ancho x 14 Prof 70 5 x 40 6 x 35 6 cm V lvula de Alivio de Presi n 40 psig 5psig 276 kPa 34 5 kPa Sistema Operativo Ciclo de Tiempo Oscilaci n de Presi n Para unidades equipadas con OSD 84 2 el indicador visible de
59. aters C Do not place a humidifier with an oxygen patient unless pre scribed by a physician and then only a bubble type humidifier should be used D Do not connect the oxygen concentrator to an electrical out let controlled by a wall switch the outlet should be indepen dent of other appliances E Do not use an electrical adapter or extension cord with the oxygen concentrator Only operate the oxygen concentrator with all filters in place do not operate if the air filter is wet INFORMATION A WARNING WARNING Electric shock hazard Do not remove cabi net The cabinet should only be removed by a qualified Sunrise Medical homecare provider A WARNING WARNING Disconnect the power cord from the wall outlet before attempting repairs on the unit Extra care should be taken if it is necessary to operate the unit with the cabinet removed A WARNING WARNING Do not use oils greases or any petrole um based solvents cleaners on or near the unit Use only materials that are compatible with oxygen A WARNING WARNING Electric Shock Hazard When replacing the capacitor do not touch the terminals or allow metal objects to come in contact with the terminals on the capacitor The capacitor may hold a charge for several days after the unit is turned off The capacitor is located in the base of the unit next to the cooling fan Use only Sunrise Medical concentrator replacement parts and accessories
60. ative Maintenance 1 8 Between Patient Maintenance AA SANS SSE has dis aed ie p PK a urere 8 Preventative Maintenance Summary 1 1 4 2 2 8 TROUBLESHOOTING System Operation ete Var er eg Mara Sua aa ea d Shed ta p Sua Wl 9 Normal Operating Sequence 1 4 2 2 2 9 Simplified Troubleshooting vua T T Ts tae ee eas Gene ae eb a hee Abend QU VIA a 10 Troubleshooting Chart Avui ion bed RR Dodo ee RASS II Troubleshooting Chart B u a aaa ayu am aaa eene Hautes RI ARE o IA ed er 12 Troubleshooting Chart C vicio aa RE HR gia E e Aa Mana r 12 Troubleshooting Chart u all LAUREA MES Qu bui ER PURI US HG haley Alger uqa a RE Lara waa ee a 12 Troubleshooting Chart D ey dated a bred e She deed bee eee Pes tae 12 Troubleshooting Chart cui Das Ver Da Glas cheat she dae d uate ube ane 13 COMPONENT TESTING REPAIR AND REPLACEMENT Proper Repair Procedures u a u a uy a E Y MEME Looe tae ee hui Se ee US 14 Cabinet Removal enn dead ee IE abe ee
61. before turning Off to prevent problems associated with moisture contami nation in the system To test the sieve beds Remove the 1 16 plugged piece of tubing from the top of each sieve bed 2 Connect the pressure vacuum gauges to the sieve bed test points Figure 9 in order to observe unit cycling and bed pressures Refer to the section on Normal Operating Sequence 3 If it is determined that the valve did not shift refer to the sec tion on Rotary Valve testing However if the unit is cycling prop erly allow unit to run while observing the sieve bed pressures 4 After 20 minutes of operation check the oxygen concentra tion levels Low oxygen concentration and high pressures indi cate contaminated sieve beds NOTE If the molecular sieve material is found to be no longer effective first locate the source of the malfunction or cause such as leaks for contamination and take corrective action To replace the molecular sieve beds Figures 5 6 amp 7 Make sure any contamination problem has been corrected before replacing 2 Make sure the unit is unplugged from the wall outlet 3 Cut the plastic cable ties that secure the sieve beds to the internal structure of the unit 4 Remove the tubing from the fittings at the top of each sieve bed 5 Remove the hose clamps and hose from the bottom of the sieve beds REPAIR AND REPLACEMENT Install new sieve beds in reverse order using new plastic cable ti
62. ch is a synthetically produced inorganic silicate It is very porous and has the unique ability to selectively adsorb nitrogen from the air as it passes through the sieve bed As one bed is being pressurized the other bed is quickly depressurized This allows the nitrogen that was adsorbed during its pressurization cycle to be exhausted from the sieve material The nitrogen is released through exhaust ports located on the rotary valve assembly The ports are connected to a single piece of hose running from the valve to the exhaust muffler Also during each bed pressurization a small amount of oxygen flows through an orifice Figure 9 from the pressurized bed into the depressurizing bed The orifice is clamped inside a long piece of blue tubing connecting the outlets of the two sieve beds This helps purge the nitrogen from the depressurizing bed The beds will continue to be alternately pressurized and depressurized as the unit operates Oxygen leaving the sieve beds is directed through a check valve to the accumulator tank A pressure regulator Figure 9 on the tank controls the oxygen pressure as it leaves the accu mulator and enters the flow meter The flow meter allows the oxygen flow to be controlled and adjusted to the level pre scribed by the patient s physician From the flow meter the oxygen passes through the final bacteria filter Figure 8 a check valve and finally the oxygen outlet port to the patient The DeVilbiss O
63. cteria filter and cause a decrease in the unit performance Exterior Cabinet The patient should clean the concentrator exterior cabinet by using a damp cloth or sponge with a mild household cleaner and wiping it dry A WARNING WARNING Do not apply liquids directly to the cabinet or utilize any petroleum based solvents or cleaning agents LT 1928 PERIODIC HOMECARE PROVIDER PREVENTATIVE MAINTENANCE Every DeVilbiss Oxygen Concentrator is tested at the factory To assure continued trouble free performance the following preventative maintenance should be performed by the home care provider during periodic oxygen patient visits Failure to properly maintain the unit will void the warranty 1 Check the oxygen concentration with an oxygen analyzer part 2 3 months on non OSD units or every two years on OSD units a Calibrate the oxygen analyzer prior to checking the oxy gen concentration The analyzer should be properly cali brated using the manufacturer s recommended procedure NOTE Changes in temperature altitude or humidity may affect the analyzer s oxygen concentration reading The analyz er should be calibrated in similar conditions to the location of the concentrator b The concentrator must operate for a minimum of 20 min utes before checking the oxygen concentration c Connect the analyzer to the unit s oxygen outlet port Figure 1 and wait until the display stabil
64. cuit break er Mark these wires accordingly Tilt the top of the bib forward to release it from the slot in the body of the concentrator Remove the hose connected to the bottom of the flow meter To reassemble bib Reconnect the wires and hose WARNING 2 Insert the tab into the slot above the rotary valve and push until it snaps into place WARNING When servicing the DeVilbiss Oxygen 3 Secure bib with two screws Concentrator be absolutely certain that the correct tools are used and that the parts are free of oil and grease or any material not compatible with oxygen Teflon tape is recommended and must be applied to Teflon is a registered trademark of DuPont the male threads omitting the first thread to eliminate LOX 8 is a trademark of Fluoramics Inc Snoop is a registered trademark of SWAGELOK LT 1928 14 COMPONENT TESTING REPAIR AND REPLACEMENT ACCUMULATOR PRESSURE TEST To check accumulator pressures I Make sure the unit is Off 2 Remove front and back cabinets 3 Use the pressure vacuum gauge part PVO2D 601 and pressure test assembly part 303DZ 637 included in the Service Kit 4 Remove the tubing cap from the accumulator tank fitting and attach the 1 16 1 6 mm diameter tubing from the gauge to the accumulator tank fitting just vacated above 5 Turn the unit On with the flow rate set to maximum recom mended flow During each timed cycle the average press
65. dical se reserva el derecho de alterar o cam biar el dise o de las series de Concentradores de Ox geno DeVilbiss Por lo tanto pueden existir diferencias menores en la construcci n o en los componentes entre la unidad que usted posee y la descrita en este manual LT 1928 INFORMACION GENERAL MEDIDAS DE SEGURIDAD IMPORTANTES A ADVERTENCIA Leer todas las instrucciones antes de hacer funcionar el con centrador de ox geno La informaci n importante se encuentra G ADVERTENCIA Peligro de choque el ctrico No reti resaltada por estos t rminos rar el gabinete El gabinete deber ser removido por un proveedor calificado de Asistencia Domiciliaria de A ADVERTENCIA Sunrise Medical ADVERTENCIA Informaci n de seguridad sobre peli A ADVERTENCIA gros que pudieran causar lesiones serias o muerte PRECAUCI N Informaci n para prevenir da os al H ADVERTENCIA Desconectar el cord n el ctrico del producto tomacorriente antes de realizar reparaciones en la NOTA Informaci n a la que usted deber a prestar especial unidad Se deber n tomar medidas de precauci n adi atenci n cionales si fuese necesario hacer funcionar la unidad una vez removido el gabinete PRECAUCIONES DE SEGURIDAD Y ADVERTENCIAS GENERALES A ADVERTENCIA A La Ley Federal EE UU restringe la venta de este equipo por o por orden de un m dico ADVERTENCIA No utilizar aceites grasas o solventes limpiadores a base de petr leo sobre o cerca de la A ADVERTE
66. dificador de burbuja entre el concentrador y el paciente nica Conexi n Detubos de Ox geno Figura 1 2 Insertar la c nula parte CN100 en el puerto de salida de ox geno Colocar el tubo de ox geno de 5 32 pulg 4 mm D I parte HOSTO7 OST15 OST25 OST50 Conexion de tubo de Oxigeno con Humidificacion Si el profesional m dico ha prescrito un humidificador de oxigeno como parte de la terapia del paciente seguir estos pasos Llenar la botella del humidificador parte HUM16 con agua destilada No sobrecargar Si se utiliza una botella precargada ver el Paso 3 Colocar la tuerca de mariposa ubicada en la parte superior de la botella del humidificador al puerto de salida de ox geno para que quede suspendida Comprobar que est bien ajustada Colocar el tubo de ox geno de 5 32 pulg 4 mm D I parte OST07 OSTI5 OST25 6 5 50 no exceder los 50 pies 15 metros directamente a la boca de la botella del humidificador NOTA Para un funcionamiento ptimo el Concentrador de Ox geno DeVilbiss posee una presi n de salida nominal preestablecida de 8 5 psi 58 6 kPa Utilizar solamente humidi ficadores de tipo burbuja No utilizar humidificadores de tipo jet NOTA La condensaci n desde el humidificador puede ocur rir en tubos m s largos o si el tubo yace sobre un piso fr o Esto puede ser reducido utilizando un pie de humidificador removible parte HMC44DM 509 Para utilizar el pie 2 3
67. dor de flujo El medidor de flujo permite controlar y ajustar el flujo de ox geno a los niveles prescritos por el m dico del paciente Desde el medidor de flujo el ox geno pasa a trav s de un filtro final de bacterias Figura 8 una v lvula de control y finalmente del puerto de salida de ox geno al paciente El Concentrador de Ox geno DeVilbiss opera en un ciclo cronometrado El ciclo es controlado por la Placa PC La Placa PC env a el voltaje a la v lvula ocasionando que cambie y presurice los tamices de manera alterna La Placa PC tambi n activa el sistema de alarma electr nico Una condici n de alta presi n se indicar con un sonido de chasquido producido por la liberaci n de presi n por una v lvula de alivio de presi n en el cabezal del compresor El flujo bajo y las fallas de energ a son indicadas por alarmas audi bles y visibles El sistema operativo de los Modelos 515 OSD incorpora tec nolog a decreciente La Placa PC constantemente monitorea la velocidad de flujo y disminuir el ciclo cuando la misma sea inferior a 2 5 Ipm Por lo tanto decrece el ciclo basado en la menor demanda de ox geno Como resultado la unidad fun ciona m s fr a con menor consumo de energ a SECUENCIA DE FUNCIONAMIENTO NORMAL Cuando el Concentrador es encendido la siguiente secuencia c clica puede ser observada colocando medidores de presi n en los puntos de prueba de los tamices La v lvula rotativa es c
68. e Wire Harness Connector LLL mc mai pee a Em Capacitor for Power Fail Alarm Fuse PC Board Screw Sieve Bed Test Point Tubing to Flow Meter Oxygen Outlet Tubing 1 8 ID OSD Oxygen Accumulator Tank LT 1928 FIGURES DIAGRAMS AND VIEWS Figure 8 Interior Behind Bib View Final Bacteria Filter Oxygen Outlet Handle Port Screws Power Switch Final Check Valve Light Panel Ribbon Connector Flow Meter Circuit Breaker Bib Tab Switch Detail Red Wire Brown Wire or Double Black Wires Red Wire Double Black Wires or Brown Wire Either red wire can be connected to either left hand terminal of the switch LT 1928 FIGURES DIAGRAMS AND VIEWS Figure 9 Interior Top View 515 5 and 515ADZ Fixed Orifice Sieve Bed Hour Meter 1 4 x 1 8 Sieve Bed Pressure Regulator Accumulator Tank Check Tubing Reducer Check Valve Test Point Fitting Valve Sieve Bed Communication Port Test Point Wire Harness Tubing to Flow Meter Sieve Bed Test Point LT 1928 FIGURES DIAGRAMS AND VIEWS Figure 9A Interior Top View 515 5 and 515 7 NOTE Sieve beds shown are used only on models 515AKS and 515AKZ Hour Meter Sieve Bed 1 4 x 1 8 Check Valve Tubing Reducer Communication Port Wire Harness Pressure Regulator Sieve Bed Check Valve Accumulator Tank
69. e atr s del babero Inclinar la parte superior del gabinete anterior hacia delante hasta que pueda ser empujado fuera de la base de la unidad La mayor a de todos los servicios y reparaciones pueden realizarse sin retirar el babero anterior Sin embargo para tener acceso a los componentes detr s del babero ste puede ser aflojado o removi do Para aflojar el babero Figura 4 Retirar los dos tornillos ubicados directamente sobre el medidor de tiempo que sostienen el babero a la estructura interna de la unidad Esto permitir el acceso a los compo nentes detr s del babero Para retirar el babero completamente Figura 8 Retirar los dos tornillos seg n se explica m s arriba centrador nuevamente en servicio 2 Desconectar el conector de cinta de la Placa PC PRECAUCI N No aplicar la soluci n de detecci n de p rdidas a ninguna parte de la v lvula rotativa o el Tefl n es una marca registrada de DuPont ensamblaje de la Placa PC LOX 8 es una marca comercial de Fluoramics Inc Snoop es una marca registrada de SWAGELOK LT 1928 62 PRUEBA DE COMPONENTES Desconectar las lineas del interruptor de encendido y del interruptor de circuito Marcar estos cables apropiadamente Inclinar la parte superior del babero hacia delante para liberar lo de la ranura en el cuerpo del compresor Retirar la manguera conectada a la parte inferior del medidor de flujo Para ensamblar el babero 3 Reconectar
70. e cab inet The cabinet should only be removed by a qualified Sunrise Medical homecare provider A WARNING WARNING Disconnect the power cord from the wall outlet before attempting repairs on the unit Extra care should be taken if it is necessary to operate the unit with the cabinet removed CABINET REMOVAL To remove front and back cabinets Figures 2 amp 3 2 Ensure the unit is unplugged from the wall outlet Using a screwdriver remove the six screws that hold the back cabinet to the internal structure and the bib NOTE All six screws are the same size Remove the back cabinet by sliding it toward the rear until clear Remove the front cabinet by pushing the top shoulders toward the back of the unit then outward away from behind the bib Tilt the top of the front cabinet forward until it can be pulled out of the base of the unit The majority of all the servicing and repairs can be done with out removing the front bib However to gain access to the components behind the bib it may be loosened or removed To loosen the bib Figure 4 Remove the two screws located directly above the hour meter that hold the bib to the unit s internal structure This will allow access to the components behind the bib To remove the bib completely Figure 8 l 2 3 4 5 Remove the two screws as above Disconnect the ribbon connector from the PC board Disconnect the lines from the power switch and cir
71. e concentrator draws in room air separates the nitrogen from the oxygen and delivers concentrated oxy gen to the patient through an oxygen port For more in depth classroom type training Sunrise Medical holds oxygen concentrator service schools For service school information contact the Respiratory Technical Service Department at 1 800 333 4000 814 443 4881 NOTE Sunrise Medical reserves the right to alter or change the design of the DeVilbiss Oxygen Concentrator series Hence slight differences in construction or components may exist between the unit in hand and what is described in this manual GENERAL IMPORTANT SAFEGUARDS Read all instructions before operating the oxygen concentra tor Important information is highlighted by these terms A WARNING WARNING Safety information for hazards that might cause serious injury or death CAUTION Information for preventing damage to the product NOTE Information to which you should pay special attention SAFETY PRECAUTIONS AND GENERAL WARNINGS A Federal U S A law restricts this device to sale by or on the order of a physician A WARNING B S WARNING Oxygen promotes rapid burning Do not smoke when using this unit or when near a person receiving oxygen therapy Do not operate the oxygen concentrator within a minimum of five feet 1 6m from hot sparking or burning objects or naked flames Do not use in rooms heated by paraffin or portable gas he
72. e main PC Board assembly A malfunction in the pressure regulator will cause either a loss or fluctuation in the oxygen flow which will be seen on the flow meter or a decrease in oxygen concentration To replace the pressure regulator Figure 9 Make sure the unit is unplugged from the wall outlet 2 Remove the front and back cabinets 3 Remove the tubing clamp and tubing from the pressure regulator 4 Unscrew the regulator from the accumulator tank 5 Install a new regulator on the accumulator tank and attach the tubing and tubing clamp 6 Replace the front and back cabinets PRINTED CIRCUIT BOARD The printed circuit PC board is responsible for monitoring and controlling the DeVilbiss Oxygen Concentrator The PC board has preset alarms for low flow and power fail ure Should any of the alarm values be exceeded the patient alert system will activate NOTE If the concentrator has been unused for an extended period the unit must run 20 minutes before the power fail alarm will be enabled This alarm is powered by a capacitor on the PC board CAUTION Do not apply any force or flex to the PC Board when connecting or disconnecting electronic or pneumatic components Damage to the electronic assembly is possible To remove and replace the PC board Figure 7 Make sure the unit is unplugged from the wall outlet Remove the front and back cabinets and the bib Disconnect all wires terminals and connectors
73. eba de voltaje y los voltajes que deber an estar presentes en cada uno Polo Polo Lectura de Negativo Positivo Voltaje Naranja Amarillo o 2 Naranja Gris 5 VCC Naranja Marr n 5 VCC Intermitente Negro o verde Amarillo 2 3 VCC rojo azul blanco NOTA La lectura de voltaje 2 3 se encendera y apagara mien tras la valvula est en ciclo Las Series 515A poseen un tiempo de ciclado fijo de aproximadamente 7 segundos Sin embargo cuando el modelo ADS con OSD pasa al modo descanso el tiempo de ciclo es acortado a aproximadamente 3 segundos Si no se obtiene alguna de estas lecturas verificar por cables sueltos o dafiados en el arn s Si el arn s de cableado esta correcto reemplazar la placa PC principal Si se obtienen los voltajes apropiados reemplazar la valvula Para reemplazar la valvula rotativa Verificar que la unidad esta desenchufada del tomacorriente Retirar el gabinete anterior y aflojar el babero Desenchufar el arn s de cableado de la valvula Cortar las ligaduras plasticas en las mangueras de los tamices Aflojar las abrazaderas r pidas y retirar las mangueras de los tamices en ambos lados de la v lvula Aflojar las abrazaderas escalonadas y retirar las mangueras de presi n entrante y saliente en la parte posterior de la v lvula Reemplazar los amortiguadores de espuma de ser necesario con los que se incluyen en el nuevo paquete de v lvul
74. eber exceder 1 2 PSI 4 kPa por sobre las presiones finales del acumulador Ver Especificaciones para las presiones normales obtenidas durante el ciclo LT 1928 SOLUCION DE PROBLEMAS SOLUCION DE PROBLEMAS SIMPLIFICADA La clave para la soluci n simple de problemas es reconocer qu tipo de problema existe y seleccionar la soluci n m s efectiva Los diferentes tipos de problemas e intentos de solu cionarlos son los siguientes Tipo La unidad funciona baja presi n y flujo existe una condici n de alta presi n NOTA La presi n o el flujo bajos son indicados por alarmas visibles y audibles La presi n alta es indicada por un sonido de chasquido causado por la v lvula de alivio de presi n Conectar los medidores a los puntos de prueba de los tamices Figura 7 2 Ver Secuencia de Funcionamiento Normal para asegurar el ciclo correcto de la unidad 3 Si la presi n de los tamices aumenta lentamente verificar fil tros obturados y p rdidas severas Si los filtros est n limpios y no existen p rdidas entonces el compresor es defectuoso 4 Si la v lvula de alivio de presi n est liberando presi n obser var si la unidad est ciclando o no 5 Si la unidad no est ciclando o las presiones de los tamices son desiguales esto indica que la v lvula rotativa no est funcio nando correctamente Ver Prueba de V lvula Rotativa descrito en el cap tulo Prueba de Componentes Reparaci n y R
75. eemplazo 6 Si la unidad est ciclando juntamente con presiones de tamices muy altas esto indica fallas en los tamices Tipo II El compresor no arranca cuando la unidad es encendida Verificar que el ventilador de enfriamiento est funcionando En caso contrario determinar por d nde se est perdiendo energ a 2 Verificar el voltaje del compresor en el conector del compresor 3 Si existe voltaje entonces el condensador o el compresor est n da ados 4 Si no existe voltaje el arn s de cables est defectuoso A ADVERTENCIA ADVERTENCIA Peligro Mec nico Mantenga alejados los dedos o ropas sueltas etc cuando trabaja en el compresor Tipo III El concentrador funciona y contin a ciclando pero tiene baja concentraci n de ox geno Conectar los medidores de prueba y verificar las presiones m s altas o bajas que lo normal a La presi n alta indica tamices defectuosos b La presi n baja indica filtros obturados p rdidas o com presor defectuoso c Las presiones de tamices desiguales indican que la v lvula no est funcionando correctamente Verificar por p rdidas de ox geno en tamices medidor de flujo tanque de acumulaci n regulador de presi n puerto de salida filtro final de bacterias v lvula final de control NOTA Verificar la existencia de p rdidas utilizando una solu ci n de detecci n de p rdidas tal como Snoop o equivalente no debe contene
76. entration Leak in system Check for leaks in all hoses and fittings Defective sieve bed check valve Replace check valve Defective compressor reed valve Replace compressor reed valve Defective compressor Replace compressor Rotary valve not operating correctly Replace valve Occluded filters Clean or replace filters Contaminated sieve beds Replace sieve beds Audible alarm does not sound during power failure Unit has not been used for an extended period of time NOTE If the concentrator has been unused for an extended period the unit must run several minutes before the power fail alarm will activate Allow unit to run for 20 minutes and retry Defective PC board Replace PC board Defective power switch Replace power switch Defective wire harness Replace wire harness Audible alarm does not sound when unit is turned Defective PC board Replace PC board Pressure relief valve activated popping sound Defective PC board Replace PC board Defective rotary valve Replace valve Service Required light does not illuminate when unit is turned On Defective PC board Replace PC board Defective light Replace light PC board connectors not properly latched Be sure tabs are pushed completely into place LT 1928 TROUBLESHOOTING TROUBLESHOOTING CHART B Visible Alarm Audible Alar
77. es Position the new beds so that the bed serial number label is at the top of the unit NOTE Make sure that the sealing caps remain on the new sieve beds until just prior to connecting hoses and tubing Leak test all connections with a certified leak detection solution such as Snoop or equivalent must not contain ethylene gly col Apply leak test solution to all fittings and hose connec tions with unit running If an air leak is present the solution will bubble All leaks should be repaired before putting the unit back in service CAUTION Do not apply leak test solution to any part of the rotary valve or the main PC Board assembly POWER CORD To replace the power cord 115 volt units only Figure 4 Make sure the unit is unplugged from the wall outlet 2 Remove the back cabinet 3 Disconnect the power cord connector NOTE Loosening or removing the compressor mounting plate may make it easier to service the power cord 4 Note wire colors and socket locations before removing wires 5 Using a pair of duckbill pliers squeeze the power cord strain relief and pull it out of the base of the unit 6 Insert a new power cord through the hole in the base of the unit and secure with strain relief 7 Insert sockets into connector housing and then reconnect the power cord connector 8 Replace back cabinet and secure with the six screws POWER SWITCH To replace the power switch Figure 8 Make sure the unit i
78. essor Replace compressor Defective cooling fan Replace cooling fan Fluctuating oxygen flow Occluded humidifier Clean or replace humidifier Use of improper humidifier Use only a bubble type humidifier Occluded filters Clean or replace filters Occluded or defective cannula and tubing Detach cannula from oxygen delivery tubing If proper flow is not attained check tubing for kinks or other obstructions Clean or straighten as required or replace tubing if necessary Use of excess oxygen tubing The unit is designed to deliver 5 Ipm with a cannula on 50 feet 15 meters of approximately 5 32 4 mm inside diameter tubing Smaller diameter tubing or the addition of any other flow restriction may prevent obtaining the desired flow rate Defective flow meter Leak in system Replace flow meter Check for leaks in all hoses and fittings Defective compressor Replace compressor Defective compressor reed valve Replace compressor reed valve Defective check valve Replace check valve Pressure regulator not adjusted properly or defective Adjust or replace pressure regulator Little or no oxygen flow Flow meter not adjusted properly Adjust flow meter Hose disconnected to flow meter Oxygen delivery tubing is kinked or blocked Reconnect hose Straighten tubing or remove obstruction Occluded humidifier Clean or replace humidifier Low oxygen conc
79. ficaci n el filtro de ingreso de bacterias de vida prolongada no necesitar ser reemplazado entre pacientes RESUMEN DE MANTENIMIENTO PREVENTIVO Paciente Diariamente Limpiar la botella del humidificador si se uti liza Semanalmente Limpiar el filtro de aire en la parte posterior de la unidad Limpiar el exterior del gabinete Otros Limpiar y reemplazar la c nula mascarilla y tubos seg n instrucciones del fabricante Proveedor de asistencia domiciliaria Cambiar el filtro de ingreso de bacterias cuando sea necesario siguiendo los requerimientos en el paso 3 3 meses Verificar la concentraci n de ox geno en unidades sin OSD Verificar el ambiente del concentrador y establecer intervalos de mantenimiento de menos de 3 meses si se requiere 2a os Verificar las alarmas audibles y las luces indicadoras Cambiar el filtro final de bacterias cambio dentro de las 17 520 horas Verificar el rendimiento del sistema Verificar la concentraci n de ox geno en unidades con OSD LT 1928 SOLUCION DE PROBLEMAS FUNCIONAMIENTO DEL SISTEMA El Concentrador de Oxigeno DeVilbiss utiliza un sistema de absorci n oscilante de presi n El aire es llevado dentro de la unidad a trav s de los filtros de aire y dentro de un compre sor de doble cabezal En la Figura 14 se muestra un diagrama neum tico del sistema El aire comprimido pasa a trav s de una v lvula rotativa Figura 5 el cual circula a un rango determinad
80. geno La falta de mantenimiento preventivo de la unidad anular los t rminos de la garant a Verificar la concentraci n de ox geno con un analizador de ox geno parte O2ANA cada 3 meses en unidades sin OSD o cada dos a os en unidades con OSD a Calibrar el analizador de ox geno previo a la verificaci n de la concentraci n de ox geno El analizador deber ser cali brado apropiadamente utilizando los procedimientos recomendados por el fabricante NOTA Los cambios en temperatura altitud o humedad pueden afectar la lectura del analizador de concentraci n de ox geno El analizador deber ser calibrado en condiciones similares a las de la ubicaci n del concentrador b El concentrador deber estar encendido por un m nimo de 20 minutos antes de verificar la concentraci n de ox geno c Conectar el analizador al puerto de salida de ox geno de la unidad Figura y esperar hasta que la pantalla se estabil ice d Tomar nota de la lectura Verificar las luces indicadoras y la alarma audible cada dos a os Cuando el interruptor de encendido est en ON escuchar la alarma audible y verificar si las luces del panel indi cador frontal est n funcionando 3 Cambiar el filtro de ingreso como se indica a continuaci n a Filtro de ingreso de bacterias de vida prolongada parte 515DZ 605 Inspeccionar una vez al a o Cambiar si fuera necesario sin exceder las 8 760 horas Abrir la puerta del fi
81. glish Spanish Orders may be placed by calling Customer Service 800 333 4000 Warranty parts U S A 800 333 4000 Canada 905 660 2459 International Department 814 443 4881 PARTS RETURN AND ORDERING POLICY ALL DEFECTIVE COMPONENTS THAT ARE STILL UNDER WARRANTY MUST BE RETURNED TO THE FACTORY IN SOMERSET PA WITHIN 30 DAYS AFTER SHIPMENT OF THE NEW COMPONENTS IF THE COMPONENTS ARE NOT RECEIVED WITHIN THIS PERIOD AN INVOICE WILL BE ISSUED TO YOUR ACCOUNT REBUILT EXCHANGE PARTS PRICING IS AVAILABLE ONLY WITH THE RETURN OF DEFECTIVE PARTS WITHIN 30 DAYS COMPONENTS WILL THEN BE BILLED AT THE REBUILT COST THERE WILL BE A CHARGE FOR SHIPPING IF THE DEFECTIVE COMPONENT IS NOT RECEIVED WITHIN 30 DAYS THEN A NEW COMPONENT INVOICE WILL BE ISSUED TO YOUR ACCOUNT COMPONENTS THAT ARE OUT OF WARRANTY AND NOT ONA REBUILT EXCHANGE PROGRAM DO NOT HAVE TO BE RETURNED TO THE FACTORY Before returning parts or units to the factory call the Sunrise Medical Customer Service Department 800 333 4000 814 443 4881 to obtain a return authorization number Include in the package a note indicating the return authoriza tion number along with your company name address phone number and account number The return authorization number should also be written on the outside of the package To expedite your order for warranty or non warranty parts the following information should be given to the representative Catalog number Seri
82. he valve plate c Place the metal retainer on the reed valves and secure with the reed valve screw 5 Check for proper placement of or damage to the rubber o ring on the bottom of the valve plate Replace if damaged 6 Remove piston sleeves by pulling upward and inspect cup seal on pistons Replace if badly worn or damaged To replace cup seal Figures 11 12A Remove rod screw from top of piston Remove the cup retainer plate Discard defective cup seal Place new cup seal into position Replace cup retainer plate mo ange Secure with screws 7 Reposition sleeve on piston NOTE In some cases it may be easier to position sleeve on piston before installing a new cup seal and retainer plate 8 Place valve plates on the compressor so that heads of reed valve screws are aligned with the indentation in top of pistons 9 Install the compressor heads so that the holes in the heads are aligned with the holes in the compressor housing 10 Secure compressor heads with the screws To replace the compressor NOTE For mounting plate and motor mount removal refer to sections below Also refer to steps used in removing the compressor CAUTION The 515ADS and 515ADZ concentrators use either a Thomas compressor with 15 mfd tor or a GSE compressor with a 22 mfd capacitor The 515AKS and 515AKZ concentrators use a GSE com pressor with a 10 mfd capacitor If replacement is nec essary be sure the co
83. icar por p rdidas o cables da ados entre el conector del compresor y el interruptor o el arn s de cables Si se registra voltaje en el conector del compresor entonces el condensador o el mismo compresor pueden estar defectuosos Para probar el compresor para una salida apropiada NOTA Si el compresor no est proveyendo una salida lo sufi cientemente alta el sistema de alerta del paciente puede ser activado Retirar los gabinetes anterior y posterior Conectar los medidores de presi n vac o a los puntos de prueba en los tamices Ver Prueba de Presi n de Tamices en la secci n Prueba de Componentes Reparaci n y Reemplazo para detalles al colocar el medidor Un compresor defectuoso ser indicado por un lento incremento en la presi n La presi n s lo puede alcanzar cierto nivel y luego detenerse Si se observan estas condiciones entonces El los filtro s de la unidad puede n estar obturado s verificar el filtro de aire y el filtro de entrada por oclu siones Puede haber una severa p rdida el sistema verificar por p rdidas de aire utilizando una soluci n de detecci n de p rdidas tal como Shoop o equivalente no debe con tener glicol etileno PRECAUCI N No aplicar soluci n de detecci n de p rdidas a ninguna parte de la v lvula rotativa o el ensamblaje de la Placa PC Las v lvulas de boquilla del compresor junta herm tica o el mismo compresor pueden estar defectuosos Figura
84. iclada varias veces para liberar presi n acumulada y prevenir una condici n est tica en el compresor Este ciclo r pido s lo ocurre en el inicio y es claramente audi ble pues la presi n es r pidamente liberada varias veces a trav s de la boca de escape conectada a la v lvula La Placa PC aplica una breve se al de voltaje DC a la v lvula La v lvula se detiene por aproximadamente 7 segundos cau sando que el tamiz derecho se presurice primero mientras que el izquierdo se despresuriza a aproximadamente 2 PSI 14 kPa El voltaje es aplicado nuevamente a la v lvula por un corto tiempo La v lvula se detiene por aproximadamente 0 7 segun dos Durante este tiempo las presiones de ambos tamices son iguales Una breve se al de voltaje DC es nuevamente aplicada a la v lvula La v lvula se detiene por aproximadamente 7 segundos causando la presurizaci n del tamiz izquierdo mientras que el derecho se despresuriza a aproximadamente 2 PSI 14 kPa Una breve se al de voltaje DC es nuevamente aplicada a la v lvula La v lvula se detiene por aproximadamente 0 7 segun dos Durante este tiempo las presiones de ambos tamices son iguales El ciclo se repite desde el paso 2 NOTA En el modo decreciente s lo en modelos OSD el tiempo de ciclo fijo decrece a aproximadamente 3 segundos y el tiempo de igualaci n de la presi n de los tamices es de aproximadamente 0 3 segundos NOTA La presi n final de los tamices no d
85. icol etileno Aplicar la soluci n de detecci n de p rdidas a todas las bocas y conexiones de manguera mientras la unidad se encuentra funcionando PRECAUCI N No aplicar la soluci n de detecci n de p rdidas a ninguna parte de la v lvula rotativa o el ensamble de la Placa PC Si existiera una p rdida de aire la soluci n comenzar a burbujear Todas las p rdidas deben ser reparadas antes de colocar el con centrador nuevamente en servicio A ADVERTENCIA ADVERTENCIA Peligro de choque el ctrico Tener m xima precauci n al probar p rdidas cerca de conexiones el ctricas Para reemplazar el medidor de flujo Figura 8 l 2 Verificar que la unidad esta desenchufada del tomacorriente Retirar los gabinetes anterior y posterior Desde atras del babero retirar las 2 mangueras del medidor de flujo Mientras se empujan las pesta as de los soportes del medidor de flujo empujar el medidor de flujo a trav s del babero Instalar un nuevo medidor de flujo en el babero y reconectar las mangueras MEDIDOR DE TIEMPO Para reemplazar el medidor de tiempo Figura 9 2 Verificar que la unidad esta desenchufada del tomacorriente Retirar los gabinetes anterior y posterior y aflojar el babero Desconectar el conector del medidor de tiempo de la placa PC Retirar el medidor insertando cuidadosamente un peque o destornillador plano debajo del borde exterior del medidor y haciendo palanca hacia arriba Ins
86. informaci n deber ser provista al representante N mero de cat logo Numero de serie Lectura del medidor de tiempo por cada concentrador Numero de cuenta Nombre y direcci n de la compa a Descripci n del problema LT 1928 INFORMACION DE ORDENES Y LISTADO DE PARTES Accesorios Componentes 515AKS 515ADS 515 2 SISAKZ 510220 50162 60 50102631 PRECAUCI N Los concentradores de series 515A PRECAUCI N Los concentradores de series 515ADS utilizan tanto un compresor Thomas como un com 5I5ADZ utilizan tanto un compresor Thomas con un presor GSE Asegurarse de ordenar el n mero de condensador de 15 mdf o un compresor GSE con un parte correcto al reemplazar o reconstruir el compre condensador de 22 mdf Los concentradores de series sor El condensador est incluido con la compra del 515AKS y 515AKZ utilizan un compresor GSE con un compresor condensador de 10 mdf Si fuera necesario un reem plazo asegurar la instalaci n del condensador correc to El condensador est incluido con la compra del compresor LT 1928 96 INFORMACION DE RDENES Y LISTADO DE PARTES Componentes continuaci n 515 515ADS 5I5ADZ 5I5AKZ Medidores de Flujo i8 DI Azul ETa Herramientas 20302637 LT 1928 ESPECIFICACIONES SERIES DEVILBISS 5 LITROS 515ADS 515ADZ 1a5 LPM N mero de Cat logo Velocidad de Administraci n Velocidades de administraci n m s baj
87. inutos para que el dispositivo termoprotector se reinicie Si el ventilador de enfriamiento est defectuoso 99 debe ser reemplazado Verificar que la unidad esta desenchufada del tomacorriente Retirar los gabinetes anterior y posterior Retirar el compresor Desconectar los terminales del ventilador Notar la posici n del ventilador y del protector del ventilador antes de retirar los cuatro tornillos de sujeci n que aseguran el ventilador a la base de la unidad Retirar el ventilador defectuoso y asegurar el ventilador de reemplazo en posici n con los cuatro tornillos de sujeci n NOTA Al instalar el ventilador asegurar que la flecha direc cional del flujo de aire al costado del ventilador est dirigida lejos del compresor y que el protector del ventilador es rein stalado apropiadamente Reconectar el conector el ctrico Reinstalar el compresor V LVULA DE CONTROL FINAL Esta v lvula est ubicada entre el filtro final de bacterias y la boca de salida de ox geno Esta v lvula de control permite al ox geno fluir s lo hacia el exterior de la unidad Cuando la unidad es apagada y el flujo de ox geno se detiene la v lvula de control se cierra para prevenir que el agua ingrese a la unidad Una v lvula de control defectuosa puede ocasionar que el agua proveniente de la botella del humidificador ingrese a la unidad cuando la misma es apagada Esto puede obturar el filtro final de bacterias y o e
88. isible and audible alarm High pressure is indicated by a popping sound caused by the pressure relief valve Connect test gauges to sieve bed tests points Figure 7 Refer to the Normal Operating Sequence to make sure the unit is cycling properly If bed pressure is rising slowly check for occluded filters and severe leaks If filters are clean and there are no leaks then the compressor is defective If the pressure relief valve is releasing pressure observe whether the unit is cycling or not If the unit is not cycling or has uneven bed pressures this indi cates that the rotary valve is not operating correctly Refer to the Rotary Valve Testing described in the chapter Component Testing Repair and Replacement If the unit is cycling in conjunction with very high bed pres sures this indicates defective sieve beds Type II The compressor will not start when the unit is turned on Verify that the cooling fan is running if it is not determine where you are losing power Check for compressor voltage at the compressor connector If voltage is present then the capacitor or compressor is defective If voltage is not present the wire harness is defective A WARNING WARNING Mechanical Hazard Keep fingers loose clothing etc away when working on compressor Type Ili The concentrator runs and continues to cycle but has low oxygen concentrations M Connect test gauges and check for higher or lower
89. izes d Record the reading 2 Check the audible alarm and indicator lights every two years When the power switch is turned On listen for the audible alarm and check to see if the front panel indicator lights are operating 3 Change intake filter as follows a Extended life intake bacteria filter part 515DZ 605 Inspect once a year Change as necessary not to exceed 8760 hours b Open the filter door and replace filter as required 4 Change the final bacteria filter part PV5LD 651 every two years or 17 520 hours a Unplug the unit remove the cabinet and loosen the bib b Remove the hose from each end of the filter Figure 8 and discard the filter c Install the new final bacteria filter with the IN fitting toward the flow meter d Tighten the bib and replace the cabinet 5 Check the system performance every two years of operation by measuring the accumulator pressure swing Use the Accumulator Pressure Test described in the chapter Component Testing Repair and Replacement NOTE This PM Schedule reflects 5000 hour usage equal to one year anormal clean operating environment The homecare provider is responsible for determining the condition of the concentrator operating environment determining a preventative maintenance interval frequency not to exceed the schedule stated above which takes into consideration the specific operating environment BETWEEN PATIENT
90. je usar circuitos alternativos independientes de otros aparatos Ventilador de enfriamiento defectuoso Reemplazar Compresor defectuoso Reemplazar SOLUCI N DE PROBLEMAS CUADRO E Alarma Visible Alarma Audible Compresor Luz de Encendido ON ON ON ON Otros S ntomas Causas Posibles Soluciones Posibles Flujo fluctuante o negativo Presi n del sistema por debajo de 9 psi 62 1 kPa debido a P rdida en el sistema Compresor defectuoso Verificar p rdidas en todas las mangueras y accesorios Reemplazar compresor LT 1928 SOLUCION DE PROBLEMAS SOLUCION DE PROBLEMAS CUADRO F SOLAMENTE UNIDADES OSD Compresor Luz de Encendido ON ON Otros Sintomas Causa Posible Soluciones Posibles Luces OSD no encienden OSD defectuoso Verificar concentraci n con analizador de ox geno Si est dentro de especificaciones reemplazar la Placa PC Luces OSD no se encienden pero la luz roja de Servicio Requerido se enciende acom pa ada por una alarma audible intermitente El nivel de ox geno es bajo Verificar concentraci n con analizador de ox geno Si est dentro de especificaciones reemplazar la Placa Si la concentraci n es baja ver s ntomas de baja concentraci n de ox geno en Soluci n de Problemas Cuadro A Ambas luces de OSD se encienden OSD defectuoso Reemplazar la Placa PC Luz amarilla de Bajo Ox geno
91. l Departamento de Atenci n al Cliente de Sunrise Medical al 800 333 4000 814 443 4881 AJUSTE DEL PACIENTE Colocar la unidad cerca de un tomacorriente en la habitaci n donde el paciente pasa la mayor parte de su tiempo NOTA No conectar a un tomacorriente controlado por un interruptor de pared La toma deber a ser independiente de otros artefactos Colocar la unidad a por lo menos 6 pulgadas 16 cm de las paredes cortinas u otros objetos que pudieran prevenir el flujo de aire apropiado fuera del concentrador de ox geno Ubicar la unidad a un m nimo de 5 pies 1 6 m de chimeneas radiadores estufas y rejillas de aire caliente A ADVERTENCIA S ADVERTENCIA El ox geno promueve incin eraci n r pida No fumar mientras utiliza esta unidad o estando cerca de una persona que recibe ter apia de ox geno No operar el concentrador de ox geno dentro de un radio m nimo de cinco pies 1 6 m de objetos calientes chispeantes o quem ndose o de llamas No utilizar en habitaciones con calefacci n a base de parafina o calentadores port tiles de gas ADVERTENCIA Peligro de choque el ctrico S lo los proveedores calificados de Asistencia Domiciliaria de Sunrise Medical pueden retirar el gabinete Conectar los accesorios de ox geno apropiados tubos de ox geno o humidificador al puerto de salida de ox geno NOTA Se permiten un m ximo de 50 pies 15 metros de tubo m s 7 pies 2 1 metros de c nula m s un humi
92. l medidor de flujo causando una restricci n en el flujo y haciendo dif cil ajustar la velocidad de flujo Para reemplazar la v lvula de control final Figura 8 l 2 Verificar que la unidad est desenchufada del tomacorriente Retirar los gabinetes anterior y posterior y aflojar o retirar el LT 1928 PRUEBA DE COMPONENTES REPARACION Y REEMPLAZO 8 2 babero Retirar la manguera del costado de salida del filtro final de bacterias Retirar los dos tornillos de la parte posterior del ensamble de la boca de salida de oxigeno y retirar el ensamble Retirar la manguera a cada lado de la valvula de control final Instalar las mangueras a una nueva valvula de control Asegurar que el lado plano de la v lvula de control esta en direcci n hacia la boca de salida de oxigeno Reemplazar el ensamble de boca de salida y conectar la manguera al filtro Volver a colocar el babero y los gabinetes anterior y posterior MEDIDOR DE FLUJO El medidor de flujo de presi n compensada tiene un nivel de precisi n del 5 a escala completa excepci n 0 5 a 5 Ipm El medidor de flujo en el Concentrador de Ox geno DeVilbiss est dise ado para utilizar a 8 5 psi 58 6 kPa a velocidades de flujo de hasta 5 Ipm Para verificar p rdidas en los tubos del medidor de flujo Verificar p rdidas utilizando una soluci n de detecci n de p r didas certificada tal como Snoop o equivalente no debe contener gl
93. lida de ox geno Figura 1 Lavar en una soluci n de agua tibia y detergente lavavajillas Enjuagar profundamente bajo el agua tibia del grifo y secar con toalla El filtro deber estar completamente seco antes de ser reinstalado A ADVERTENCIA ADVERTENCIA No operar la unidad sin el filtro de aire o mientras el mismo se encuentra todav a h me do NOTA El filtro de aire deber ser monitoreado m s exhausti vamente en ambientes con grandes cantidades de polvo y pelusa PRECAUCI N El funcionamiento del Concentrador de Ox geno DeVilbiss en ambientes extremos o sin el filtro de aire causar n la oclusi n prematura del filtro de ingreso de bacterias y un decaimiento en el rendimiento de la unidad Gabinete Exterior El paciente deber limpiar el gabinete exterior del concen trador utilizando un pa o o esponja h medos con un limpiador dom stico suave Secar con un pa o A ADVERTENCIA ADVERTENCIA No aplicar l quidos directamente sobre el gabinete o utilizar cualquier solvente o agentes de limpieza a base de petr leo LT 1928 MANTENIMIENTO PREVENTIVO PERIODICO DEL PROVEEDOR DE ASISTENCIA DOMICILIARIA Todos los Concentradores de Oxigeno DeVilbiss son proba dos en la fabrica Para asegurar un funcionamiento continuo libre de problemas el siguiente mantenimiento preventivo deber ser realizado el proveedor de asistencia domiciliaria durante las visitas peri dicas a los pacientes que usan Ox
94. ltro y reemplazar el mismo seg n sea necesario Cambiar el filtro final de bacterias parte PV5LD 651 cada dos a os 17 520 horas a Desenchufar la unidad retirar el gabinete y aflojar el babero Retirar la manguera a cada lado del filtro Figura 8 y descartar el filtro c Instalar el nuevo filtro final de bacterias con la indicaci n IN hacia el medidor de flujo d Ajustar el babero y reinstalar el gabinete 5 Verificar el rendimiento del sistema despu s de funcionar durante dos a os midiendo la oscilaci n de presi n del acumu lador Utilizar la Prueba de Presi n del Acumulador descrita en el cap tulo Prueba de Componentes Reparaci n y Reemplazo NOTA Este Programa de Mantenimiento Preventivo refleja 5 000 horas de uso equivalentes a un a o ambiente de funcionamiento limpio y normal El proveedor de asistencia domiciliaria es responsable de determinar la condici n del ambiente de funcionamiento del concentrador determinar una frecuencia de mantenimiento preventivo que no exceda el programa establecido m s arriba el cual toma en consideraci n el ambiente de fun cionamiento espec fico MANTENIMIENTO ENTRE PACIENTES Descartar los tubos de ox geno la c nula y la botella del humidificador Lavar o reemplazar el filtro de aire del gabinete 3 Lavar el gabinete del concentrador Verificar la concentraci n de ox geno Si la unidad se encuentra dentro de la especi
95. lvula de Conector El ctrico V stago Cabezal de Junta Principal vista Ret n Boquilla o de Compresor Compresor superior Aleta Tornillo Manga Junta Herm tica Junta Ret n de Herm tica Taza V lvula de Alivio de Presi n Ret n de Taza Tornillo de Tubos de V stago Conexi n cada aay V lvula de Tornillo de V stago V lvula vista Anillo O Boquilla o Tornillo inferior Aleta LT 1928 90 FIGURAS DIAGRAMAS Y VISTAS Figura 13 V lvula Rotativa Motor Escalonado Arn s Cableado de V lvula Boca de Tamiz Izquierdo Baca de Tamiz Derecho Vista Frontal Boca de Escape de Presi n Boca de Entrada de Presi n Vista Posterior lt F gt gt lt z lt 9 lt FIGURAS Figure 14 Diagrama Neum tico 31N3lONd Tv 4m INL 4 ON3 MOSN3S 20 VIMHVINIHd d SIYWINITION XO e uv TWNOVYO QVOI2013 LT 1928 DIAGRAMAS Y VISTAS FIGURAS Figura 15 Diagrama de Cableado NAH HOLON oro S3B4W023 OWS3N wz 102 SOLNOA OEZ 30 SIOWOINN OLVLSOWHIL NOD SOQVdInO3 5073004 Vtivd l Te OW 30 y 0704 n Y WNOIONNS 30 E JOSTHAANDO YOOWSN
96. m Compressor Power Light Blinking Pulsing OFF OFF Other Symptoms Possible Cause Possible Remedy Fan off Line cord not properly installed or defective Insert plug in receptacle or replace line cord No power at receptacle Check building circuit breaker or fuse or have house wiring checked by qualified electrician Circuit may be fully loaded with other appliances and another receptacle may be required Oxygen concentrator circuit breaker activated Press the circuit breaker reset button If unit circuit breaker opens again check internal wiring Line cord quick connect terminal inside unit is disconnected Reconnect quick connect terminal Defective power switch Replace power switch Defective circuit breaker TROUBLESHOOTING CHART C Replace circuit breaker Visible Alarm Audible Alarm Compressor Power Light Blinking Pulsing ON ON OFF Other Symptoms Possible Cause Possible Remedy Fan and compressor operating Blown fuse on PC board Replace fuse Pressure relief valve activated Defective PC board Replace PC board popping sound TROUBLESHOOTING CHART D Visible Alarm Audible Alarm Compressor Power Light ON ON OFF ON Other Symptoms Possible Cause Possible Remedy Fan operating Main wiring harness disconnected defective Reconnect replace wiring harness Loose compressor wire Tighten or attach wire Defective cap
97. m ev e essem a eoe PE MERE Figure 9A Top CU EE ERE v Va etri oua TS Figure 10 Lower rear close up ke RARE Shady s hen ronde ele Gnu se des Other Figures Figure 11 Compressor Thomas Nexus removed from base Figure Compressor Thomas Nexus with head Figure 12 Compressor GSE removed from base Figure 12A Compressor GSE with head removed Figure Rotary valY A x Y hee uU ev rede edes EB ON Hed ES Diagrams Figure 14 Pneumatic Diagram isi cor see cele ee e e ex MERE eq eie Biers eh tee ae un eR OR ea Figure 15 Wiring Diagraimn sess a SD rene Ve Gad ge ep AM D ES LT 1928 FIGURES DIAGRAMS AND VIEWS Figure Exterior Front View OSD Indicator Lights Service Required OSD units only Light Red Light Panel Oxygen Outlet Port Bib Front Cover Caster Base Power Light green Power Switch Flow Meter Knob Flow Meter Circuit Breaker LT 1928 FIGURES DIAGRAMS AND VIEWS Figure 2 Exterior Side View Front cover Rear cover Cabinet screw Power Cord Strap Power Cord Caster
98. o y es dirigido hacia uno de los dos tamices Los tamices contienen material de colado molecular que es un silicato inorg nico producido sin t ticamente Es muy poroso y tiene la capacidad nica de absorber nitr geno selectivamente del aire cuando pasa a trav s de los tamices Mientras uno de los tamices est siendo presurizado el otro es r pidamente despresurizado Esto permite al nitr geno absorbido durante la presurizaci n circular hasta ser liberado del material de colado El nitr geno es liberado a trav s de los puertos de escape ubi cados en el ensamble de las v lvulas rotativas Los puertos est n conectados a una nica pieza de manguera que va desde la v lvula hasta la boca de escape Adem s durante la presurizaci n de cada tamiz una peque a cantidad de ox geno fluye a trav s del orificio Figura 9 desde el tamiz presurizado hacia el tamiz despresurizado El orificio est sujeto dentro de una larga pieza de tubo conectado a las salidas de los dos tamices Esto facilita la purga del nitr geno desde el tamiz despresurizado Los tamices continuar n siendo alternativamente presurizados y despresurizados mientras la unidad se encuentre en fun cionamiento El ox geno proveniente de los tamices es dirigido a trav s de una v lvula de control a un tanque de acumulaci n Un regu lador de presi n Figura 9 en el tanque controla la presi n del ox geno mientras ste sale del acumulador e ingresa al medi
99. observan las presiones de los tamices Luego de 20 minutos de operaci n verificar los niveles de concentraci n de ox geno La baja concentraci n de ox geno y presiones altas indican la contaminaci n de los tamices NOTA Si el material de los tamices moleculares ya no fuera efectivo primero ubicar la fuente del mal funcionamiento o causa p rdidas por ejemplo por contaminaci n y tomar las medidas correctivas necesarias LT 1928 PRUEBA DE COMPONENTES Para reemplazar los tamices moleculares Figuras 5 6y 7 l Verificar que cualquier problema de contaminaci n ha sido corregido antes del reemplazo 2 Verificar que la unidad est desenchufada del tomacorriente 3 Cortar las ligaduras del cable pl stico que sujetan los tamices a la estructura interna de la unidad 4 Retirar los tubos de las bocas en la parte superior de los tamices 5 Retirar las abrazaderas de las mangueras y la manguera en la parte inferior de los tamices 6 Instalar los nuevos tamices siguiendo el orden inverso utilizan do nuevas ligaduras de cable pl stico Posicionar los nuevos tamices para que la etiqueta del n mero de serie quede en la parte superior de la unidad NOTA Verificar que las tapas de cierre permanezcan en los nuevos tamices moleculares hasta justo antes de conectar las mangueras y los tubos 7 Revisar por p rdidas todas las conexiones con una soluci n certificada de detecci n de p rdidas tal como Snoo
100. olsados S lo se reembolsar n los costos de mano de obra incurridos por un Centro de Servicio Autorizado NOTA La presente garant a no obliga a Sunrise a proveer una unidad en pr stamo durante el tiempo que un concen trador de ox geno est siendo reparado NOTA El reemplazo de componentes est garantizado por la porci n no vencida de la Garant a Limitada original La presente garant a ser anulada y Sunrise ser liberado de cualquier obligaci n o responsabilidad si El equipo es operado o mantenido fuera de los par metros indicados en las instrucciones de funcionamiento y servicio indicadas por Sunrise Personal de servicio no calificado conduce mantenimientos de rutina servicios o reparaciones de garant a Se utilizan partes o componentes no autorizados ej material de tamiz regenerado para reparar o alterar el equipo Se utilizan filtros no autorizados con la unidad NO EXISTE NINGUNA OTRA GARANT A EXPRESA LAS GARANT AS IMPL CITAS INCLUYENDO AQUELLAS DE COMERCIABILIDAD Y CAPACIDAD PARA UN PROP SITO EN PARTICULAR EST N LIMITADAS A LA DURACI N DE LA GARANT A LIMITADA EXPRESA Y DE ACUERDO A LO PERMITIDO POR LA LEY CUALQUIERA Y TODAS LAS GARANT AS IMPL CITAS EST N EXCLUIDAS LA PRE SENTE ES LA SOLUCI N EXCLUSIVA Y RESPONSABILI DAD PARA DANOS CONSECUENTES E INCIDENTALES BAJO CUALQUIER Y TODA GARANT A Y QUE EST N EXCLUIDOS AL L MITE QUE SEA PERMITIDA POR LA LEY ALGUNOS ESTADOS NO PERMIT
101. on Normal Operating Sequence 3 If it is determined that the valve did not shift properly or the bed pressures are uneven continue testing with step 4 4 The stepper motor on the rotary valve has multiple windings so there are several voltage readings that need to be checked in order to determine if the problem is being caused by the PC board or the valve itself Testing for proper voltage is done at the wire harness connectors on the valve or on the PC Board using a voltmeter Below are the valve voltage test posi tions and voltages that should be present at each one Negative Lead Voltage Reading Yellow or White 12VDC svo Orange Brown BC VDC intermittent Black or Green or Yellow or White 2 3 VDC Red or Blue NOTE The 2 3 Volt reading will go on and off as the valve cycles 515A Series have a fixed cycle time of approximately 7 seconds However when the ADS model with OSD goes into its turn down mode the cycle time is shortened to approxi mately 3 seconds 5 If any of these readings are not obtained check for loose or broken wires in the harness If wire harness is okay replace the main PC board 6 If proper voltages are obtained replace valve To replace the rotary valve Make sure the unit is unplugged from the wall outlet 2 Remove the front cabinet and loosen the bib 3 Unplug valve wire harness from the valve LT 1928 COMPONENT TESTING REPAIR AND REPLACEMENT 4 C
102. outlet should be independent of other appliances 2 Position the unit at least 6 inches 16 cm from walls draperies or any other objects that might prevent the proper flow of air in and out of the oxygen concentrator 3 Locate the unit a minimum of 5 feet 1 6 meters from fire places radiators heaters and hot air registers A WARNING S WARNING Oxygen promotes rapid burning Do not smoke when using this unit or when near a person receiving oxygen therapy Do not operate the oxygen concentrator within a minimum of 5 feet 1 6 meters from hot sparking or burning objects or naked flames Do not use in rooms heated by paraffin or portable gas heaters WARNING Electric Shock Hazard Only qualified Sunrise Medical homecare providers may remove the cabinet 4 Attach the appropriate oxygen accessories oxygen tubing or humidifier to the oxygen outlet port NOTE A maximum of 50 feet 15 meters of tubing plus 7 feet 2 1 meters of cannula plus a bubble humidifier is allowed between the concentrator and the patient Oxygen Tubing Only Connection Figure 1 Thread the cannula fitting part CN100 onto the oxygen outlet port 2 Attach the 5 32 4 mm I D oxygen tubing part 4 OSTO7 5 OST25 or OST50 Oxygen Tubing with Humidification Connection If the physician has prescribed an oxygen humidifier as part of the patient s therapy follow these steps l Fill the humidifier bottle part HUM 16 wi
103. ower oxygen concentrations A worn or defective compressor can be caused by a defective internal component such as reed valve o ring gasket cup seal These components are included in the Compressor Service Kit part 515DZ 643 Thomas or 515ADZ 643 GSE CAUTION The 515A series concentrators use either a Thomas compressor or a GSE compressor Be sure to order the correct part number when rebuilding the compressor NOTE A built in thermal cutoff switch will shut the compressor off if it becomes overheated This protects the compressor from damage caused by heat build up Some models have an auxiliary thermostat mounted within the compressor compartment NOTE A pressure relief PR valve is located on the pressure head to prevent high pressure build up in the system should a component malfunction occur COMPONENT TESTING To test the compressor operating voltage Figure 4 The compressor requires line voltage to operate If the com pressor does not start when the unit is turned on the voltage input must be tested This voltage can be checked at the compressor connector using a voltmeter or test light connected to the brown and blue wires The voltmeter is the best way to test 2 If no voltage is detected disconnect power and check for loose or broken wires between the compressor connector and switch or wire harness 3 If there is voltage at the compressor connector then either the capacito
104. p o equivalente no debe contener glicol etileno Aplicar la solu ci n de detecci n de p rdidas a todas las bocas y conexiones de manguera mientras la unidad se encuentra funcionando Si existiera alguna p rdida de aire la soluci n comenzar a bur bujear Todas las p rdidas deben ser reparadas antes de poner la unidad nuevamente en funcionamiento PRECAUCI N No aplicar la soluci n de detecci n de p rdidas a ninguna parte de las v lvulas rotativas o el ensamble de la placa PC CORD N EL CTRICO Para reemplazar el cord n el ctrico solamente unidades de 115 voltios Figura 4 l Verificar que la unidad est desenchufada del tomacorriente 2 Retirar el gabinete posterior 3 Desconectar el conector del cord n el ctrico NOTA Aflojar o retirar la placa de montaje del compresor podr a facilitar el servicio al cord n el ctrico 4 Tomar nota de los colores de los cables y la ubicaci n de los enchufes antes de retirar los cables 5 Utilizando un par de pinzas Duckbill apretar el pasacables del cord n el ctrico y tirar de l hasta retirarlo de la base de la unidad 6 Insertar un nuevo cord n el ctrico a trav s del orificio en la base de la unidad y asegurarlo con el pasacables 7 Insertar los enchufes en el recept culo del conector y luego reconectar el conector del cord n el ctrico 8 Recolocar el gabinete posterior y asegurar con los seis tornillos REPARACI N Y REEMPLAZO
105. poner el con A ADVERTENCIA ADVERTENCIA AI reparar el Concentrador de Ox geno DeVilbiss asegurarse de utilizar las herramien tas correctas y que todas las partes est n libres de aceites y grasa o cualquier material no compatible con el ox geno Se recomienda la cinta Teflon y debe ser aplicada al roscado macho omitiendo la primera rosca para eliminar la posibilidad de que las part culas de cinta ingresen al sistema de ox geno El sellador LOX 8 debe ser utilizado en lugar de cinta Teflon A ADVERTENCIA ADVERTENCIA Peligro de choque el ctrico No reti rar el gabinete El gabinete s lo debe ser removido por un proveedor de Asistencia Domiciliaria calificado de Sunrise Medical A ADVERTENCIA ADVERTENCIA Desconectar el cord n el ctrico del tomacorriente antes de intentar reparar la unidad Se deber n tener cuidados adicionales si fuese necesario operar la unidad una vez removido el gabinete REMOCI N DEL GABINETE Para retirar los gabinetes anterior y posterior Figuras 2 y 3 Asegurar que la unidad est desenchufada del tomacorriente Utilizando un destornillador retirar los seis tornillos que suje tan el gabinete posterior a la estructura interna y al babero NOTA Los seis tornillos tienen la misma medida Retirar el gabinete posterior desliz ndolo hacia atr s hasta el final Retirar el gabinete anterior empujando los bordes superiores hacia la parte posterior de la unidad luego hacia fuera desd
106. r LT 1928 80 FIGURAS DIAGRAMAS Y VISTAS Figura 7 Vista Superior Frontal Interna 515ADS y 515ADZ Tubo de Salida Tubo de Salida de Ox geno de Punto de Prueba Cables de Medidor Punto de Prueba de Ox geno de 1 8 pulg D I de Tamiz de Tiempo de Tamiz 1 8 pulg D I OSD Placa PC Conector de Arn s de Cableado de V lvula Condensador para Alarma de Falla de Energ a PRECAUCI N NO insertar el tornillo de montaje en este orificio Tubo de Medidor de Flujo desconectado para retiro de Tanque de Tornillo de Placa PC Babero Acumulaci n de Oxigeno Fusible FIGURAS DIAGRAMAS Y VISTAS Figura 7 A Vista Superior Frontal Interna 515AKS y 515AKZ NOTA Los tamices mostrados s lo se usan en los modelos 515AKS y 515AKZ Cables d Tubo de Salida Me dor de Prueba de Oxigeno de Punto de Prueba Tiempo 118 pulg D I de Toni Tubo de Salida de Ox geno de 1 8 pulg D I OSD Placa PC Tanque de nog de Acumulaci n Cableado de de Ox geno V lvula Tornillo de Placa PC Condensador Tubo al Medidor de para Alarma de Flujo Falla de Energ a Fusible LT 1928 82 FIGURAS DIAGRAMAS Y VISTAS Figura 8 Vista de Babero Trasero Interno Filtro Final de Bacterias Tornillos de Puerto de Manija Salida de Ox geno Interruptor de Encendido V lvula de Control Final Medidor de Flujo Interruptor de Circuito Detalle del Interruptor Cable Rojo C
107. r glicol etileno Aplicar la soluci n de detec ci n de p rdidas a todas las juntas y conexiones de tubos mientras la unidad est funcionando Si se presenta p rdida de aire la soluci n comenzar a burbujear Todas las p rdidas deber n ser reparadas antes de hacer funcionar la unidad nue vamente PRECAUCI N No aplicar la soluci n de detecci n de p rdidas en ning n punto de la v lvula rotativa o del ensamblado de la Placa PC Verificar la presi n del tanque de acumulaci n Si la presi n es inferior a lo normal las v lvulas de control de los tamices est n defectuosas NOTA Para presiones normales del sistema ver Especificaciones LT 1928 SOLUCION DE PROBLEMAS SOLUCION DE PROBLEMAS CUADRO A Alarma Visible Alarma Audible Compresor Luz de Encendido OFF Otros Sintomas OFF ON Causas Posibles ON Soluciones Posibles Sonido de aire pulsante Filtro de entrada fuera de lugar o defectuoso Verificar filtro Reemplazar Manguera de entrada de compresor desconectada Reconectar Ruido excesivo Monturas de motor sueltas o defectuosas Reemplazar Compresor defectuoso Reemplazar Ventilador de enfriamiento defectuoso Reemplazar Flujo de oxigeno fluctuante Humidificador ocluido Limpiar o reemplazar Uso de humidificador inapropiado Usar solo el de tipo burbuja Filtros ocluidos Limpiar o reemplazar C nula o tubos ocluidos defectuoso
108. r or the compressor itself is defective To test the compressor for proper output NOTE If the compressor is not providing a high enough output the patient alert system may be activated Remove the front and back cabinets 2 Connect pressure vacuum gauges to the sieve bed test points See the Sieve Bed Pressure Test in the Component Testing Repair and Replacement section for details on attaching the gauge defective compressor will be indicated by slowly rising pres sure Pressure may only reach a certain level and then stop If these conditions are observed then The unit filter s may be occluded check the air filter and intake filter for occlusions There may be a severe leak in the system check for air leaks using a leak detection solution such as Snoop or equivalent must not contain ethylene glycol CAUTION Do not apply leak test solution to any part of the rotary valve or the main PC Board assembly The compressor reed valves cup seal or the compressor itself may be defective Figures 11 amp 12 If the filters are not occluded and no leaks are found the com pressor must then be removed and repaired or replaced To remove the compressor Make sure the unit is unplugged from the wall outlet 2 Disconnect the compressor wires by disconnecting the com pressor electrical connector Figure 4 3 Remove the ladder clamp and hose from the outlet fitting on the compressor Figure 4
109. rrect capacitor is installed LT 1928 COMPONENT TESTING Inspect the motor mounts Replace if damaged Secure the mounting plate s to the bottom of the new compressor using the four compressor mounting hex nuts N Inspect the capacitor to determine if replacement is necessary capacitor is included w compressor purchase If capacitor is wrong value for compressor or replacement is desired refer to Capacitor section 3 Reconnect tubing to the compressor intake fitting 4 Position compressor on the base of the unit so that the studs on the mounting plates are aligned with notches on the front of the unit base 5 Secure mounting plate with two screws on the back and install nuts on the front side of the plate 6 Reconnect hose to the fitting at compressor outlet 7 Reconnect the compressor electrical connector To remove compressor from the mounting plate l Turn compressor upside down so that it is resting on the heads 2 Remove the four compressor mounting hex nuts and mount ing plate To remove motor mounts Unscrew studded motor mounts from compressor feet hand COOLING FAN The cooling fan provides a constant air flow to cool the compressor The cooling fan is located in the bottom of the unit below the compressor Figure 10 A defective cooling fan may cause the compressor s internal thermo protective device to activate and shut the compressor off Should this condition occur
110. rs TIO Bit 5 32 Allen bit 5 64 Allen bit 9 64 Allen bit 7 64 Allen bit 2 Presure Vacuum gauge part PVO2D 601 Tool box 2 Test Fittings part 303DZ 637 Torx screwdriver w bits AC DC test light 1 4 Ratchet wrench 3mm Hexbit T 15 Torx L wrench 10mm Socket 1 4 Drive 1 4 Drive extension Plastic storage case Plastic error indicator tool In addition to the Service Kit an oxygen analyzer part O2ANA is needed to periodically check oxygen concentration levels A volt meter will be needed for more accurate voltage testing NOTE Be sure to read all of the steps involved before begin ning any of the procedures in this manual NOTE After repairing or replacing a component run the unit for 20 minutes check the oxygen concentration and test for leaks Test for leaks using a certified leak detection solution such as SWAGELOK MS Snoop or equivalent must not contain ethylene glycol Apply leak test solution to all fittings and hose connections with the unit running If an air leak is present the solution will bubble All leaks should be repaired before putting the concentrator back in service CAUTION Do not apply leak test solution to any part of the rotary valve or the main PC Board assembly the possibility of tape particles entering the oxygen sys tem LOX 8 sealant may be used in place of Teflon t A WARNING WARNING Electric shock hazard Do not remov
111. rtamento Internacional DEVOLUCI N DE PARTES Y POL TICA DE RDENES TODOS LOS COMPONENTES DEFECTUOSOS QUE SE ENCUENTREN BAJO GARANT A DEBER N SER DEVUEL TOS A LA F BRICA EN SOMERSET PA DENTRO DE LOS 30 D AS LUEGO DEL ENV O DE LOS NUEVOS COMPO NENTES SI LOS COMPONENTES NO SON RECIBIDOS DENTRO DE ESTE PERIODO UNA FACTURA SER EMITI DA A SU CUENTA LOS PRECIOS DE PARTES DE RECAMBIO RECONSTRU IDAS ESTAN DISPONIBLES SOLO CON LA DEVOLUCI N DE LAS PARTES DEFECTUOSAS DENTRO DE LOS 30 D AS LOS COMPONENTES SER N ENTONCES FACTURADOS A COSTO DE RECONSTRUCCI N EXISTIR UN CARGO POR EMBARQUE SI EL COMPONENTE DEFECTUOSO NO ES RECIBIDO DENTRO DE LOS 30 D AS ENTONCES UNA NUEVA FACTURA POR COMPONENTES SER EMITIDA A SU CUENTA LOS COMPONENTES QUE SE ENCUENTRAN FUERA DE GARANT A Y NO EN UN PROGRAMA DE RECONSTRUCCI N RECAMBIO NO TIENEN OBLIGACI N DE SER DEVUELTOS A LA F BRICA Antes de devolver repuestos o unidades a la f brica contactar al Departamento de Atenci n al Cliente de Sunrise Medical 800 333 4000 814 443 4881 para obtener un numero de autorizaci n de devoluci n Incluir en el embalaje una nota indicando el n mero de autorizaci n de devoluci n junta mente con el nombre de su compa a direcci n tel fono y n mero de cuenta El n mero de autorizaci n de devoluci n tambi n deber constar en la parte exterior del embalaje Para acelerar su orden por repuestos en garant a o no la siguiente
112. s Retirar la c nula del tubo de ox geno De no obtener flujo apropiado verificar tubos por torceduras u obstrucciones Limpiar o enderezar o reemplazar tubos Uso de excesivos tubos de ox geno La unidad est dise ada para proveer 5 Ipm con una c nula de 50 pies 15 metros de aproximadamente 5 32 pulg 4 mm de di metro interno de tubo Un di metro inferior o la adici n de otras restricciones de flujo pueden impedir obtener el flujo deseado Medidor de flujo defectuoso Reemplazar P rdida en el sistema Verificar p rdidas en mangueras y accesorios Compresor defectuoso Reemplazar V lvula de boquilla de compresor defectuoso Reemplazar V lvula de control defectuosa Reemplazar Regulador de presi n flojo o defectuoso Ajustar o reemplazar Flujo bajo o negativo Medidor de flujo no ajustado correctamente Ajustar Manguera desconectada del medidor de flujo Tubo de ox geno retorcido o bloqueado Reconectar Enderezar tubo o retirar la obstrucci n Humidificador ocluido Limpiar o reemplazar Baja concentraci n de ox geno P rdida en el sistema Verificar p rdidas en mangueras y tubos V lvula de control de tamiz defectuosa Reemplazar V lvula de boquilla de compresor defectuosa Reemplazar Compresor defectuoso Reemplazar V lvula rotativa no funciona correctamente Reemplazar Filtros ocluidos Limpiar o reemplazar
113. s unplugged from the wall outlet 2 Remove the front and back cabinets and loosen the bib 3 Note the position of the wires and switch before removing the wires from the switch terminals 4 While squeezing the locking tabs on the sides of the switch push the switch out of the front of the bib 5 Install the new switch in the correct orientation making sure that it locks into position 6 Reconnect the wires to the switch terminals PRESSURE REGULATOR The pressure regulator stabilizes the flow of oxygen to the patient and establishes back pressure on the system It is pre set at 8 5 psi 58 6 kPa and should not have to be adjusted in the field To test the pressure regulator Turn the unit On 2 Set the flow meter at 2 Ipm LT 1928 COMPONENT TESTING 3 Use a pressure vacuum gauge part PVO2D 601 and a fitting suitable to fit on the oxygen outlet or on a short piece of tub ing connected to the outlet 4 f the pressure vacuum gauge reads anything other than 8 5 5 psi 58 6 5 9 kPa with gauge outlet blocked adjustment to the pressure regulator may be required If so call Sunrise Service Department at 1 800 333 4000 814 443 4881 NOTE Make sure no leaks exist before adjusting the pressure regulator by using a certified leak detection solution such as Snoop or equivalent must not contain ethylene gly col CAUTION Do not apply leak test solution to any part of the rotary valve or th
114. so puede ser provocado por un componente interno defectuoso tal como LT 1928 PRUEBA DE COMPONENTES REPARACION Y REEMPLAZO V lvula de boquilla Anillo O Junta Junta herm tica Estos componentes est n incluidos en el Kit de Servicio del Compresor parte 515DZ 643 Thomas 515ADZ 643 GSE PRECAUCION Los concentradores de series 515A utilizan un compresor Thomas o GSE Asegurarse de ordenar el numero de parte correcto al reconstruir el compresor NOTA Un interruptor incorporado de cierre t rmico apa gara el compresor si el mismo se recalienta Esto protege al compresor de da os por recalentamiento Algunos modelos poseen un termostato auxiliar montado dentro del compar timiento del compresor NOTA Una v lvula de alivio de presi n PR se encuentra ubicada en el cabezal de presi n para prevenir que la alta pre si n se acumule en el sistema si ocurriera el mal fun cionamiento de alg n componente Para probar el voltaje de funcionamiento del compre sor Figura 4 El compresor requiere de una l nea de voltaje para poder fun cionar Si el compresor no arranca cuando la unidad es encen dida debe probarse el voltaje de entrada Este voltaje puede ser verificado en el conector del compresor utilizando un voltimetro o luz de prueba conectado a los cables marr n y azul El voltimetro es el mejor m todo para probarlos Si no se detecta voltaje desconectar la energ a y verif
115. supplied by the manufacturer If no cleaning instructions were supplied these steps should be followed Wash the humidifier bottle in a solution of hot water and dishwashing detergent Soak the humidifier in a solution of one part white vinegar to three parts hot water for 30 45 minutes This solution acts as a germicidal agent Rinse thoroughly with hot tap water and refill with distilled water for use Do not overfill Cannula Mask and Tubing The patient should clean and replace the cannula or mask and tubing as instructed by the manufacturer Air Filter and Oxygen Outlet Connector The air filter Figure 3 and oxygen outlet connector should be cleaned at least once a week by the patient To clean these steps should be followed Remove the air filter located in the door on the back of the unit Remove the oxygen outlet connector if used from oxygen out let port Figure 1 2 Wash in solution of warm water and dishwashing detergent 3 Rinse thoroughly with warm tap water and towel dry The fil ter should be completely dry before reinstalling A WARNING WARNING Do not attempt to operate the unit with out the air filter or while the filter is still damp NOTE The air filter should be monitored more closely in environments with abnormal amounts of dust and lint CAUTION Operation of the DeVilbiss Oxygen Concentrator in extreme environments or without the air filter will prematurely occlude the intake ba
116. talar un nuevo medidor de tiempo aplicando presi n hacia abajo hasta que calce en posici n Conectar el medidor de tiempo a la placa PC PRECAUCI N No aplicar fuerza o doblar la placa PC al conectar o desconectar los componentes electr ni cos o neum ticos Es posible ocasionar da os al ensamble electr nico TAMICES MOLECULARES Los dos tamices moleculares remueven alternativamente el nitr geno del aire que pasa a trav s de los mismos y proveen al paciente una carga constante de ox geno La eficiencia del material de los tamices moleculares ser menor si se encuentra contaminada por humedad La contami naci n hace que el material de los tamices moleculares pierda sus propiedades de adsorci n de nitr geno provocando una disminuci n en la concentraci n de ox geno La unidad deber funcionar por lo menos un m nimo de 20 minutos antes de encenderse para prevenir problemas asociados con la contam inaci n del sistema por humedad Para probar tamices Retirar la pieza de tubo de 1 16 pulg conectada a la parte superior de cada tamiz Conectar los medidores de presi n vac o a los puntos de prueba de los tamices Figura 9 para observar la unidad ciclando y las presiones de los tamices Ver la secci n Secuencia de Funcionamiento Normal Si se determina que la v lvula no cambi ver la secci n Prueba de V lvula Rotativa Sin embargo si la unidad est ciclando apropiadamente permitir que funcione mientras se
117. te NOTE When the flow meter knob is turned clockwise the flow decreases and eventually will shut off the oxygen flow When the knob is turned counter clockwise the flow increases NOTE Use low output flow meter part 515LF 607 for flow rates under Ipm NOTE The unit may require up to 20 minutes for the oxygen concentration and flow rate to stabilize The flow rate should be monitored and readjusted if necessary The flow meter has a locking device If it is necessary to preset and lock in the prescribed flow rate tighten the set screw located on the hex nut just below the control knob using a 1 16 Allen bit No adjustment can be made without loosening the set screw The DeVilbiss oxygen concentrator is now ready for use LT 1928 PATIENT ALERT SYSTEM The DeVilbiss Oxygen Concentrator patient alert system will detect unit component failure This system is comprised of both visible and audible alarms which signal the patient if a malfunction should occur The visible alarm located on the front panel Figure reads Service Required The audible alarm system is internally powered no batteries are required When the indicator lights illuminate or the audible alarm sounds other than during unit start up a problem has occurred Non OSD models Power Failure Blinking red Service Required light and pulsing audible alarm Low Flow Continuous red Service Required light and audible alarm OSD models
118. termitente Servicio Requerido y alarma audible pulsante Flujo Bajo Luz roja continua Servicio Requerido y alarma audible Modelos con OSD Falla de energ a Luz roja intermitente Servicio Requerido y alarma audible pulsante Flujo Bajo por debajo de los 0 5 Ipm Luz roja continua Servicio Requerido y alarma audible 2 Ox geno por debajo de lo normal 84 a 75 luz amarilla Ox geno Bajo 75 a 60 luz amarilla Ox geno Bajo y alarma audible intermitente menos de 6076 luz roja Servicio Requerido y alarma audible intermitente Ver Especificaciones para ajustes de alarma espec ficos Las alarmas visible y audible se activar n aproximadamente 15 minutos en una situaci n de falta de energ a Si la unidad es encendida ON sin electricidad o la misma se interrumpiera posteriormente la alarma no sonar durante los primeros 10 segundos Luego de este tiempo la alarma producir un pulso audible con intervalos de algunos segundos mientras que la alarma visible parpadea La energ a para esta alarma es provista por un condensador en la Placa PC NOTA Si el concentrador no ha sido utilizado durante un per odo prolongado la unidad deber mantenerse funcionando durante algunos minutos antes de que la alarma de falla de energ a se active La placa PC circuito impreso Figura 5 es responsable de controlar el sistema y las alarmas NOTA Una condici n de alta presi n est indicada por
119. th distilled water Do not overfill If using a prefill go to Step 3 2 Thread the wing nut located on the top of the humidifier bot tle to the oxygen outlet port so that it is suspended Make sure it is securely tightened 3 Attach the 5 32 4 mm I D oxygen tubing part OSTO7 5 OST25 or OST50 not to exceed 50 feet 15 meters directly to the humidifier bottle outlet fitting NOTE For optimum performance the DeVilbiss Oxygen Concentrator has a preset nominal output pressure of 8 5 psi 58 6 kPa Use only bubble type humidifiers Do not use jet type humidifiers NOTE Condensation from the humidifier may occur in longer lengths of tubing or if the tubing is laying on a cold floor This can be reduced by using a removable humidifier stand part 440 509 To use the stand Attach a straight humidifier adapter fitting part 444 506 to the bottle by turning the wing nut on the humidifier until it is tight on the fitting 2 Secure the bottle in the strap 3 Attach one end of the oxygen tubing to the oxygen outlet on the unit and the other end to the plastic adapter fitting on the humidifier Locate the humidifier near the patient When ready for operation Attach the nasal cannula part CANO0 catheter or face mask to the oxygen tubing per the manufacturer s directions 2 Follow the Operating Instructions on the next page LT 1928 UNPACKING AND SETUP OPERATING
120. the compressor will require several minutes for the thermo protective device to reset If the cooling fan is defective it must be replaced Make sure the unit is unplugged from the wall outlet Remove the front and back cabinets Remove the compressor Disconnect the cooling fan terminals Ga 9 RO Note the position of the fan and fan guard before removing the four retaining screws that secure the fan to the base of the unit 6 Remove the defective fan and secure the replacement fan in position with the four retaining screws NOTE When installing the fan be sure the air flow directional arrow on the side of the fan is directed away from the compressor and fan guard is reinstalled properly Reconnect the electrical connector 8 Reinstall the compressor FINAL CHECK VALVE This check valve is located between the final bacteria filter and the oxygen outlet fitting This check valve allows oxygen to REPAIR AND REPLACEMENT flow only out of the unit When the unit is turned off and oxy gen flow stops the check valve closes to prevent water from being drawn into the unit A defective final check valve may allow water to be drawn in from the humidifier bottle when the unit is turned off This may occlude the final bacteria filter and or the flow meter causing a restriction of flow and making it difficult to adjust the flow rate To replace the final check valve Figure 8 Make sure the unit is unplugged from
121. tica en posici n Reemplazar la placa que sujeta la taza Asegurar con tornillos 7 Recolocar la manga en el pist n NOTA En algunos casos puede ser mas sencillo colocar la manga el pist n antes de instalar una nueva junta herm tica y placa de sujeci n 8 Colocar placas de v lvula en el compresor para que las cabezas de los tornillos de v lvula de boquilla est n alineados con la indentaci n en la parte superior de los pistones 9 Instalar los cabezales del compresor para que los orificios en los mismos est n alineados con los orificios en el alojamiento del compresor 10 Asegurar los cabezales del compresor con los tornillos Para reemplazar el compresor NOTA Para retirar la placa de montaje y la montura del motor referirse a las secciones a continuaci n Tambi n ver los pasos utilizados en la remoci n del compresor PRECAUCI N Las series de concentradores 515ADS y 5I5ADZ utilizan un compresor Thomas con concen trador de 15 mdf o un compresor GSE con un conden sador de 22 mdf Las series de concentradores 515AKS y 515AKZ utilizan un compresor GSE con un conden sador de 10 mdf Si es necesario su reemplazo asegu rar la instalaci n del condensador correcto l Inspeccionar la montura del motor Reemplazar si est da ada Asegurar la s placa s de montaje a la parte inferior del nuevo compresor utilizando las cuatro tuercas hexagonales de mon tura del compresor 2 Inspeccion
122. trador de ox geno S lo debe hacer funcionar el concentrador de ox geno con todos los filtros en su lugar no hacer funcionar si el filtro de aire est mojado LT 1928 52 DESEMBALAJE Y AJUSTE INSPECCION INICIAL Se sugiere realizar una inspecci n general al momento de recibir el concentrador de ox geno Luego de sacar el Concentrador de Ox geno DeVilbiss de la caja de cart n examinarlo por da os exteriores Si hubiera ocurrido da o durante su traslado contactar al Departamento de Atenci n al Cliente de Sunrise Medical al 1 800 333 4000 814 443 4881 para obtener instrucciones espec ficas Guardar la caja de cart n en caso de posible devoluci n notar la posici n de la unidad y del material de embalaje Abrir la puerta del filtro Figura 3 y anotar la cantidad de horas en el contador de horas Verificar que el filtro de aire se encuentra en su lugar Verificar para asegurar que el filtro de ingreso de bacterias Figura 4 est en su lugar Enchufar la unidad al tomacorriente encender la unidad ON y verificar las alarmas visibles y audibles Ajustar el medidor de flujo al m ximo flujo por litro recomen dado y dejar la unidad funcionando por lo menos durante 20 minutos Utilizar un analizador de ox geno para verificar la concen traci n NOTA Si la unidad no funciona apropiadamente concen trador de ox geno fuera de las especificaciones o si se detecta alg n da o interno contactar a
123. tre los tamices El voltaje CC es provisto por la placa PC a los devanados del motor causando la rotaci n de los dis cos y dirigiendo la presi n al tamiz apropiado 6 Si la v lvula rotativa no alterna apropiadamente el mismo tamiz podr a continuar presuriz ndose haciendo que la v lvu 7 la de alivio de presi n descargue el exceso de presi n Existen varias razones para el mal funcionamiento de la v lvula 8 rotativa por lo tanto la causa de la falla debe ser determinada antes de tomar cualquier acci n correctiva Para probar la v lvula rotativa Figuras 6 y 7 Retirar el gabinete anterior 2 Conectar los medidores de presi n a los puntos de prueba en la parte superior de los tamices para observar el ciclo de la unidad y la presi n de los tamices Ver la secci n Secuencia de Funcionamiento Normal Si se determina que la v lvula no est alternando apropiada mente o que las presiones de los tamices son desiguales con tinuar realizando pruebas a partir del punto 4 El motor de velocidad gradual de la v lvula rotativa contiene varios devanados por lo que habr diversas lecturas de voltaje que necesitan ser verificadas para determinar si el problema est siendo causado por la Placa PC o la v lvula misma La prueba del voltaje apropiado es realizada en los conectores del arn s de cableado en la v lvula o en la Placa PC utilizando un voltimetro A continuaci n se detallan las posiciones de la pru
124. uced by the unit The OSD operates as follows Normal Oxygen green light oxygen purity normal Low Oxygen yellow light oxygen purity low requires ser vicing NOTE If the oxygen purity continues to fall an audible signal will sound intermittently If the oxygen purity continues to fall to a low enough level the yellow Low Oxygen light will turn off and the red Service Required light will turn on NOTE Refer to Specifications for specific alarm settings When the unit with the OSD is turned On all four indicator lights Power Service Required Low Oxygen and Normal Oxygen on the front panel will briefly illuminate After a few seconds only the Power and Normal Oxygen lights will remain on NOTE After Power On the OSD conducts a continuous diagnostic evaluation to check for a fault in the piezo electron ics If this condition is detected by the OSD electronics at any time during concentrator operation the green Normal Oxygen OSD light will turn off and the beeping audible and blinking red Service Required light alarms activate Otherwise for the first fifteen minutes of operation the green Normal Oxygen light will remain illuminated during the oxygen stabilization process After that time the OSD will begin monitoring the oxygen purity every second Slowly turn the flow meter knob until the flow meter ball is centered on the line next to the appropriate flow ra
125. ufes puede causar quemaduras fuego u otros peligros de choque el ctrico No utilizar la unidad si el cable el ctrico est da ado A ADVERTENCIA S ADVERTENCIA El ox geno promueve incineraci n r pida No fume mientras utiliza esta unidad o estando cerca de una persona que recibe terapia de ox geno No opere el concentrador de ox geno dentro de un radio m nimo de cinco pies 1 6 m de objetos calientes chispeantes o quem ndose o de llamas No uti lizar en habitaciones con calefacci n a base de parafina o calentadores port tiles de gas Presionar el interruptor de encendido a la posici n ON Cuando la unidad est encendida la luz de Service Required se Requiere Servicio se encender y una se al audible sonar moment neamente el sistema de alerta del paciente La luz Power Encendido tambi n se encender S lo el Concentrador de Ox geno DeVilbiss con OSD El OSD es un dispositivo opcional dentro de los concen tradores DeVilbiss que monitorea el oxigeno producido por la unidad El OSD funciona asi Ox geno Normal luz verde pureza de ox geno normal Ox geno Bajo luz amarilla pureza de ox geno baja requiere servicio NOTA Si la pureza del ox geno contin a decreciendo se escuchar una se al audible intermitente Si la pureza del ox geno a n contin a decreciendo a un nivel lo suficiente mente bajo la luz amarilla de Low Oxygen Ox geno Bajo se
126. ulo Interruptor de circuito del concentrador de ox geno activado Presionar el bot n de reinicio del interruptor de cir cuito Si ste vuelve a activarse verificar el cableado interno Terminal de conexi n r pida de cables en el interior Reconectar de la unidad desconectado Interruptor de encendido defectuoso Reemplazar Interruptor de circuito defectuoso Reemplazar SOLUCI N DE PROBLEMAS CUADRO C Alarma Visible Alarma Audible Compresor Luz de Encendido Parpadeante Pulsante ON ON OFF Otros S ntomas Causas Posibles Soluciones Posibles Ventilador y compresor fun Fusible quemado en la Placa PC Reemplazar cionando V lvula de alivio de presi n activada sonido Placa PC defectuosa Reemplaza de chasquido SOLUCI N DE PROBLEMAS CUADRO D Alarma Visible Alarma Audible Compresor Luz de Encendido ON ON OFF ON Otros S ntomas Causas Posibles Soluciones Posibles Ventilador funcionando Arn s de cableado principal desconectado defectuoso Reconectar reemplazar Cable de compresor suelto Ajustar o conectar cable Condensador defectuoso Reemplazar Compresor defectuoso Reemplazar Unidad tibia al tacto y no puede ser reiniciada por varios minutos Compresor recalentado debido a Filtros ocluidos Pasaje de entrada salida de aire restringido Baja o alta l nea de voltaje Limpiar o reemplazar Retirar oclusi n Verificar l nea de volta
127. ure in the oxygen accumulator will rise and fall NOTE Normal pressures observed depend on altitude and flow rate Increases in altitude and flow rate will slightly decrease accumulator pressures Decreases in the two variables will slightly increase accumulator pressures Acceptable accumu lator pressure swing ranges at various altitudes at the maxi mum recommended flow are identified in the Specifications NOTE A defective check valve may cause a rapid drop in accumulator pressure below the minimum value NOTE A defective compressor will be indicated by slowly rising pressure Pressure may only reach a certain level and then stop Low oxygen concentration levels and accumulator pressures higher than normal may indicate defective sieve beds Severely contaminated beds may also cause the pressure relief valve on the compressor to open NOTE A malfunctioning rotary valve will also cause high accumulator tank pressure and activation of the pressure relief valve In this case it should be determined whether the prob lem is with the sieve beds valve or both CAPACITOR The capacitor enables the compressor to start and run by sup plying voltage to the windings of the compressor motor A defective capacitor will result in the compressor running slower or not starting CAUTION The 515ADS and 515ADZ concentrators use either a Thomas compressor with a 15 mfd capaci tor or a GSE compressor with a 22 mfd capacitor If replacement is
128. ut plastic cable ties on sieve bed hoses 5 Loosen speedy clamps and remove bed hoses from each side of valve 6 Loosen ladder clamps and remove pressure intake and exhaust hoses from backside of valve 7 Replace foam dampers if necessary with ones included in new valve package 8 Install the new rotary valve by reversing the above procedure NOTE Use minimal force when tightening the cable ties around the rotary valve Overtightening the cable ties can transmit sound to the cabinet increasing noise level SIEVE BED CHECK VALVES A check valve Figure 9 located between the outlet of each sieve bed and the accumulator tank allows oxygen to pass through the sieve beds to the accumulator tank when the bed pressure is greater than the accumulator tank pressure These valves also prevent reverse flow of oxygen from the accumula tor to the sieve beds Run the accumulator pressure test described in the Component Testing Repair and Replacement section It covers details of attaching the gauge Acceptable pressure swing ranges at vari ous altitudes are included in the Specifications A defective check valve will result in a decrease in oxygen con centration and lower accumulator pressures To replace a defective check valve remove the tubing on either side of it and install a new valve making sure the outlet end is toward the accumulator tank LT 1928 FIGURES DIAGRAMS AND VIEWS FIGURES DIAGRAMS AND VIEWS IN
129. v lvulas de boquilla Una v lvula de boquilla est ubicada a cada lado de la placa de v lvulas 4 Las v lvulas de boquilla del compresor deber n estar al mismo nivel de la placa de la v lvula Si la v lvula est rota o no est al mismo nivel de la placa de v lvulas o una materia extra a es detectada en el interior del cabezal limpiar o reemplazar la v lvula de boquilla Para reemplazar las v lvulas de boquilla del com presor Figuras A y 12 a Retirar el tornillo que sujeta las v lvulas de boquilla del com presor en su lugar en la placa de v lvulas y desechar las v lvu las de boquillas usadas b Colocar las nuevas v lvulas de boquilla para que est n cen tradas y cubran completamente los orificios en la placa de v lvulas c Colocar el sujetador met lico sobre las v lvulas y asegurar con el tornillo de v lvula de boquilla LT 1928 PRUEBA DE COMPONENTES 5 Verificar la ubicaci n apropiada o el da o en el Anillo O de goma en la parte inferior de la placa de v lvulas Reemplazar si est da ado 6 Retirar las mangas de pist n tirando hacia arriba e inspec cionar la junta herm tica de los pistones Reemplazar si est n seriamente gastados o da ados Para reemplazar la junta herm tica Figuras 11 A y 12A a Retirar el tornillo de la parte superior del pist n Retirar la placa que sujeta la taza Desechar la junta herm tica defectuosa Colocar una nueva junta herm
130. xygen Concentrator operates on a timed cycle The cycling is controlled by the PC board The PC board will send voltage to the valve causing it to shift and alternately pressurize the sieve beds The PC board also activates the electronic alarm system A high pressure condition will be indicated with a popping type sound produced by release of pressure from a pressure relief valve on the compressor head Low flow and power failure are indicated by audible and visible alarms The 515 OSD Models operating system incorporates turn down technology The PC board constantly monitors the flow rate and will decrease the cycle time whenever the flow rate is less than 2 5 LPM Therefore it turns down the cycle based on lower oxygen demand As a result the unit runs cooler with less power consumption NORMAL OPERATING SEQUENCE When the concentrator is turned On the following cycling sequence can be observed by attaching pressure gauges to the sieve bed test points The rotary valve is quickly cycled several times to relieve resid ual bed pressure preventing a static condition in the compres sor This rapid cycling only happens on start up and is clearly heard as pressure is being quickly exhausted several times through the exhaust muffler that is connected to the valve 2 The PC board applies a short DC voltage signal to the valve The valve will stop for approximately 7 seconds causing the right bed to pressurize