Home
Premium V10
Contents
1. Only Error Beep Appears when the batteries cannot provide enough power for a test Replace the batteries immediately 34 MESSAGE WHAT DOES METER SAY WHAT IT MEANS WHAT TO DO E aw Only Error Beep Appears when test strip is removed while counting down or insufficient blood volume Repeat the test with a new test strip Only Error Beep Appears when a used test strip is inserted Repeat with a new test strip Only Error Beep Appears when ambient temperature is below system operation range Only Error Beep Appears when ambient temperature is above system operation range System operation range is 50 F to 104 F 10 C to 40 C Repeat the test after the meter and test strip are in the above temperature range 35 Troubleshooting 1 If the meter does not display a message after inserting a test strip POSSIBLE CAUSE Batteries exhausted WHAT TO DO Replace the batteries Test strip inserted upside down or incompletely Defective meter or test strips Insert the test strip with contact bars end first and facing up Please contact customer service 2 If the test does not start counting down after applying the sample POSSIBLE CAUSE WHAT TO DO Insufficient blood sample Repeat the test using a new test strip with larger volume of blood sample Defective test strip R
2. Micro Kill Plus by Medline EPA Reg No 59894 10 37549 To obtain disinfecting wipes and other information please contact Medline at 1 800 MEDLINE 1 800 633 5463 or visit www medline com Disinfecting Procedures 1 Take out one disinfecting wipe from the package and squeeze out any excess liquid in order to prevent damage to the meter 2 Wipe all exterior surface of the meter including the display and buttons Hold the meter with the test strip slot pointing down and wipe the area around the test slot be careful not to allow excess liquid to get inside Keep the meter moist with disinfection solution contained in the wipe for a minimum of 2 minutes for Micro Kill wipes Follow the instructions on the package label of disinfecting wipe Use two or more wipes if necessary 3 Remove the wipe Allow the meter surface to dry completely 4 Discard the used wipes Improper system cleaning and disinfection may result in meter malfunction If you have a question please contact customer service at 29 1 866 469 2632 for assistance This device has been validated to withstand 5 000 disinfection cycles using the recommended disinfecting wipe towelette The tested number of cycles is estimated by 1 000 tests per year on each meter and around 3 disinfection cycles per meter per day for 5 years the expect life of device Stop using the meter if you see any signs of deterioration such as the LCD display is cracked or clo
3. Al compartir las lancetas se podr a aumentar el riesgo de contraer enfermedades infecciosas Las lancetas no deben ser usadas por m s de una persona Preparando el Lugar de Punci n Para obtener una muestra m s significativa de sangre en el valor de glucosa antes de la extracci n frote el rea de punci n para estimular la perfusi n de sangre La sangre de un lugar que no ha sido estimulado brinda una 54 concentraci n distinta en la medici n que la sangre que ha sido extra da del dedo Cuando el punto de punci n ha sido frotado previamente a la extracci n de la sangre la diferencia se ve significativamente reducida Por favor siga las siguientes sugerencias antes de obtener la gota de sangre L vese y s quese bien las manos antes de empezar e Seleccione el rea de punci n ya sea el dedo u otra parte del cuerpo por favor vea la secci n del Lugar Alternativo de Prueba LAP o c mo escoger un lugar apropiado e Limpie el lugar de punci n con un algod n humedecido con 70 de alcohol y deje secar al aire e Frote el sitio de punci n al menos 20 segundos antes de la penetraci n e Use la tapa transparente incluida en este equipo mientras instala el dispositivo de punci n e Probando en la yema del ded Sujete el dispositivo de punci n firmemente sobre el costado de su dedo Presione al bot n de liberaci n Usted escuchar un clic indicando que la punci n est completa e Sangre d
4. There are eight 8 speaking volume levels to choose from Press until the desired speaking volume appears To confirm your selection press SET Volume 0 indicates that the speaking function is turned off and i will not display during testing Volume 1 to 7 indicates speaking volume from low to high and i will be displayed throughout the testing 5 Choosing a language Press to select L1 L2 The default language for the meter is L1 which is English L2 is the alternate language as indicated in your packaging To confirm your selection press SET Congratulations You have completed all settings NOTE e These parameters can ONLY be changed in the setting mode e f the meter is idles for 3 minutes during the setting mode it will switch off automatically 13 THE TWO MEASURING MODES The meter provides you with two modes for measuring General and QC MODES USE WHEN General any time of day without regard to time since last meal QC testing with the control solution You can switch between the two modes mode by 1 With the meter turned off insert a test strip to turn on the meter The screen will display a flashing blood drop a Y EJ 2 Press to switch between General O z and QC mode Bai Bin icici iG General ac 14 Control Solution Testing Our control solutions contain a known amount of glucose that reacts with test strips and is use
5. Ventana de Confirmaci n Es donde usted confirma si ha puesto suficiente sangre en el orificio absorbente de la tira Barras de Contacto Inserte este extremo de la tira reactiva dentro de la ranura de prueba hasta que no entre mas Los resultados pueden resultar err neos si la barra de contacto no est totalmente insertada en la ranura de prueba La parte frontal de la tira reactiva debe ser colocada hacia arriba cuando es insertada NOTA El medidor ForaCare Premium V10 debe ser usado solamente con tiras reactivas ForaCare Premium V10 Si usa otras tiras reactivas producir n resultados incorrectos 46 Configurando el Medidor Antes de usar el medidor por primera vez o si desea cambiarle las bater as debe verificar y actualizar estas configuraciones Aseg rese de completar los pasos descritos a continuaci n y guardar los nuevos cambios deseados Entrando a la funci n de configuraci n Comience con el medidor apagado sin haber insertado la tira reactiva Mantenga presionando el bot n SET 1 Configurando la fecha Con el a o parpadeando presione Y hasta que el a o correcto aparezca Presione SET Con el mes parpadeando presione hasta que el mes correcto aparezca Presione SET tf Con el dia parpadeando presione Y g ig hasta que el dia correcto aparezca iO Presione SET 2 Configurando la hora 4 Con la hora parpadeando presione hasta que la hora c
6. los resultados del an lisis de la soluci n de control no se almacenar n con los resultados reales del an lisis de glucosa en sangre Debido a ello se producir un an lisis inadecuado de los resultados almacenados VO kai V 3 Aplicando la soluci n de control Agite bien la soluci n de control antes de usarla Disponga una gota en la punta del tap n de la botella y s quela con un tissue o pa o limpio A continuaci n disponga otra gota en la punta del tap n de la botella Para tomar la muestra sostenga el medidor con la tira reactiva insertada y ll velo hasta la parte superior donde est el l quido Toque levemente el canal absorbente de la tira con la gota Una vez que la ventana de 52 confirmaci n est llena el medidor empezar el conteo regresiv Para evitar la contaminaci n de la soluci n de control no aplique directamente soluci n de control sobre una tira ni toque de ning n otro modo la boquilla de la botella 4 Lea y compare los resultados Despu s del conteo regresivo hasta O el resultado de la soluci n de control aparecer en la pantalla Compare este resultado con el rango impreso en el envase de las tiras reactivas el resultado debe estar dentro del rango De no ser as por favor lea las instrucciones una vez m s y repita la prueba de soluci n de control La glucosa en la sangre es n mero por miligramos de decilitro Resultados fuera del rang Si al repetir la prueba el
7. SET Felicitaciones Ha terminado de programar su medidor NOTA e Estos par metros pueden ser cambiados SOLO en modo de configuraci n o programaci n e Si el medidor est sin ser utilizado durante el modo de configuraci n por 3 minutos el medidor se apagar autom ticamente 49 LAS DOS MODALIDADES DE MEDICI N El medidor le provee de cuatro modalidades para la medici n General y QC Modalidades Cuando usarlas General Acualquier hora del dia sin tomar en cuenta cuando fue su ltima comida Qc Probando la soluci n de control Usted puede cambiar entre cada funci n i le qi General ac 50 1 Comience con el medidor apagado Inserte la tira reactiva para encender su medidor la pantalla va a mostrar el simbolo de gota de sangre 2 Presione para cambiar entre la funci n General y QC Probando la Soluci n de Control La Soluci n de Control contiene una cantidad conocida de glucose que reacciona con las tiras reactivas y es usada para asegurarle que su medidor y las tiras reactivas trabajan juntas correctamente Es posible que en el kit no se incluyan tiras reactivas soluciones de control o lancetas est riles compruebe el contenido de la caja del producto Se pueden adquirir por separado Aseg rese de que tiene todos los art culos antes de analizarse la glucosa en sangre Haga una prueba para la Soluci n de Control cuando e Al a
8. compacto as como a su f cil operaci n usted podr usar el Sistema de Vigilancia de Glucosa en la Sangre ForaCare Premium V10 para examinar los niveles de la glucosa por usted mismo y en cualquier lugar Si tiene alguna pregunta llame gratuitamente al Centro de servicios de atenci n al cliente de ForaCare al n mero 1 866 469 2632 Le atenderemos 24 horas al d a 7 d as a la semana Tambi n puede visitar www foracare com Version 2 0 2012 02 311 4124100 003 39 IMPORTANTES MEDIDAS DE SEGURIDAD LE SE ANTES DE SER USADO e El medidor y la lanceta son para su uso en un nico paciente No los comparta con nadie ni siquiera con sus familiares No usar en m s de un paciente e Todas las piezas del kit se consideran biopeligrosas y pueden transmitir enfermedades contagiosas incluso despu s de realizar los procedimientos de limpieza y desinfecci n Para obtener m s informaci n visite 1 FDA Public Health Notification Use of Fingerstick Devices on More than One Person Poses Risk for Transmitting Bloodborne Pathogens Initial Communication 2010 http www fda gov MedicalDevices Safety AlertsandNotices ucm224025 htm 2 CDC Clinical Reminder Use of Fingerstick Devices on More than One Person Poses Risk for Transmitting Bloodborne Pathogens 2010 http www cdc gov injectionsafety Fingerstick DevicesBGM html 1 Utilice este sistema SOLAMENTE para el uso previsto 2 NO utilice accesorios que no hay
9. humedad relativa Unidad de medici n mg dL Rango de medici n 20 a 600 mg dL 1 1 a 33 3 mmol L 74 Este aparato ha sido probado para cumplir todos los requerimientos electr nicos y de seguridad de IEC EN 61010 1 IEC EN 61010 2 101 EN 61326 1 IEC EN 61326 2 6 75 76 77 78 79 ForaCare Premium V10 Distributed by Fora Care Fora Care Inc 810 Lawrence Dr Suite 104 Newbury Park CA 91320 USA Products made in Taiwan Toll Free 1 888 307 8188 8 30am 5 00pm PST Mon Fri 1 866 469 2632 24 Hours 7 days week www foracare com usa
10. la tira reactiva El medidor mide la corriente calcula el nivel de glucosa en la sangre y muestra el resultado La fuerza de la corriente producido por la reacci n depende de la cantidad de glucosa en la muestra de sangre 43 Apariencia y Funciones Principales del Medidor 44 Ranura de Prueba Inserte la tira reactiva aqu para encender el medidor y empezar la prueba Expulsi n de la Tira Reactiva Expulse la tira reactiva usada empujando este bot n Pantalla de Visualizaci n Bot n M Muestre la memoria del medidor silencie una alarma de recordatorio cambie un ajuste en modo SET Puerto de Datos Descargue los resultados del an lisis con un cable de conexi n Bot n SET Entrada y confirmaci n de la configuraci n del medidor Pantalla de Visualizaci n El LI 13 GE o mg dL e A LAZIL III SIL a BABE a KO ai 221 aM Me Simbolo de Volumen Insertar Tira Reactiva Simbolo de Tira Reactiva Resultados de la Prueba Simbolo de Gota de Sangre Fecha Hora 10 Simbolo de Memoria Advertencia Cetona Promedio Diario 11 Modo de Control de Soluci n Unidad de Medici n S mbolo de Bater a Agotada 45 Orificio Absorbente Aplique una gota de sangre aqu La sangre ser absorbida autom ticamente Mango de la Tira Reactiva Coja la tira por esta parte para introducirla en la ranura Atenci n PA
11. ForaCare Premium V10 Blood Glucose Monitoring System Owner s Manual Dear ForaCare Premium V10 System Owner Thank you for purchasing the ForaCare Premium V10 Blood Glucose Monitoring System This manual provides important information to help you to use the system properly Before using this product please read and follow the instructions in this Owner s Manual and any package inserts for test strips and control solution Regular monitoring of your blood glucose levels can help you and your doctor gain better control of your diabetes Due to its compact size and easy operation you can use the ForaCare Premium V10 Blood Glucose Monitoring System to easily monitor your blood glucose levels If you have questions just call ForaCare Customer Care Service Center toll free at 1 866 469 2632 We are here for you 24 hours a day 7 days a week You can also visit www foracare com Version 2 0 2012 02 311 4124100 003 IMPORTANT SAFETY PRECAUTIONS READ BEFORE USE e The meter and lancing device are for single patient use Do not share them with anyone including other family members Do not use on multiple patients e All parts of the kit are considered biohazardous and can potentially transmit infectious diseases even after you have performed cleaning and disinfection For more information please visit 1 FDA Public Health Notification Use of Fingerstick Devices on More than One Person Poses Risk for Transmitting Bloodbor
12. L MEDIDOR 25 Revisando los Resultados de la Prueba 25 Revisando los Resultados del Promedio Diario de la Glucosa en 26 la Sangre Descargando los Resultados a la Computadora 27 MANTENIMIENTO 28 Baterias 28 Cuidados de su Medido 29 Cuidados de las Tiras Reactivas 32 Informaci n Importante Sobre la Soluci n de Control 33 LOCALIZACION Y RESOLUCION DE PROBLEMAS EN EL SISTEMA 34 Leyendo Resultados 34 Mensages de Error 35 Localizaci n y Resoluci n de Problemas 36 INFORMACI N DETALLADA 3 ESPECIFICACIONES 38 4 B ANTES DE EMPEZAR Informaci n Importante e La excesiva p rdida de agua y la deshidrataci n severa puede causar lecturas por debajo de los valores reales Si usted cree que est sufriendo de deshidrataci n severa consulte con su medico inmediatamente e Si usted obtiene resultados de la glucosa en la sangre m s altos o m s bajos de lo normal y no presenta s ntomas que lo evidencien repita la prueba Si presenta alg n s ntoma y o contin a obteniendo resultados irregulares siga el tratamiento recomendado por su m dico e Aplique solamente muestra fresca de sangre capilar para realizar la prueba de glucosa La aplicaci n de otras sustancias causar resultados incorrectos e Si usted tiene s ntomas que son inconsistentes con los niveles de la glucosa en la sangre y usted ha seguido todas las instrucciones descritas en este manual contacte con su profesional m dico e No recomendamos usar este pr
13. a target range that works best for you 37 SPECIFICATIONS Model No TD 4124 Dimension amp Weight 88 67 x 55 61 x 23 07 mm 48 8 g Power Source Two 1 5V AAA alkaline batteries Display LCD Memory 450 measurement results with respective date and time External output USB Auto electrode insertion detection Auto sample loading detection Auto reaction time count down Auto switch off after 3 minutes without action Temperature Warning Operating Condition 50 F to 104 F 10 C to 40 C below 85 R H non condensing Storage Transportation Conditions 4 F to 140 F 20 C to 60 C below 95 R H Measurement Units mg dL Measurement Range 20 to 600 mg dL 1 1 to 33 3 mmol L This device has been tested to meet the electrical and safety requirements of IEC EN 61010 1 IEC EN 61010 2 101 EN 61326 1 IEC EN 61326 2 6 38 Estimado usuario del Sistema ForaCare Premium V10 Gracias por adquirir el Sistema de Vigilancia de Glucosa en la Sangre ForaCare Premium V10 Blood Glucose Monitoring System Este manual le dar informaci n importante que le ayudar a utilizar el sistema correctamente Antes de usar este producto lea y siga las instrucciones del presente Manual del usuario y de los posibles prospectos para tiras reactivas y soluciones de control Vigilar regularmente sus niveles de glucosa en la sangre puede ayudarle a usted y a su doctor a controlar m s eficientemente la diabetes Gracias a su tama o
14. aCare Premium V10 Test Strips Using other test strips with this meter can produce inaccurate results 10 Setting the Meter Before using your meter for the first time or if you change the meter battery you should check and update the settings Make sure you complete the steps below and have your desired settings saved Entering the Setting Mode Start with the meter off no test strip inserted Press SET 1 Setting the date I With the year flashing press until the correct year appears Press SET With the month flashing press until the correct month appears Press SET With the day flashing press until the l correct day appears Press SET Buin UP an Leh 2 Setting the time td f With the hour flashing press until the correct hour appears Press SET ry or i AA a Big With the minute flashing press until aL IRE the correct minute appears Press SET 11 hi d I deL deL y ES na q qi q O nn 12 3 Deleting the memory With dEL and a flashing M on the display press If recorded results does not need to be deleted from the memory press SET to skip With dEL and a flashing M on the display press to delete all readings in memory Screen will display and press SET to set the speaking volume press O to turn off the meter 4 Setting the speaking volume
15. an sido suministrados por el fabricante 3 NO utilice este aparato si sospecha que est trabajando err neamente o ha sufrido alg n da o 4 Este aparato NO est dise ado para curar ning n s ntomao enfermedad que los datos de la medici n son SOLO para referencia 5 Antes de utilizar este equipo para realizar una prueba de glucosa en la sangre lea cuidadosamente todas las instrucciones Lleve a cabo todos los procedimientos indicados como se estipula en el manual para verificar el control de calidad 6 Utilizando este instrumento en ambientes secos especialmente si hay materiales sint ticos como ropa sint tica alfombras etc ya que podr a causar descargas est ticas y originar resultados err neos 7 NO use este instrumento cerca de fuentes de radiaci n electromagn ticas ya que podr an interferir en su operaci n CONSERVE ESTAS INSTRUCCIONES EN UN LUGAR SEGURO 40 TABLA DE CONTENIDO ANTES DE EMPEZAR 06 Informaci n Importante 06 Uso Previsto 07 Principios de la Prueba 07 Apariencia y Funciones Principales del Medidor 08 Pantalla de Visualizaci n 09 Tira Reactiva 10 Configurando del Medidor 11 LAS DOS MODALIDADES DE MEDICI N 14 Probando la Solucion de Control 15 Realizando una Prueba con la Soluci n de Control 15 HACIENDO UNA PRUEBA CON MUESTRA DE SANGRE 18 Preparando el Lugar de Punci n 18 Lugar Alternativo de Prueba LAP 20 Realizando la Prueba de Glucosa en la Sangre 22 MEMORIA DE
16. batteries these batteries should be kept away from children If swallowed promptly seek medical assistance e Batteries might leak chemicals if unused for a long time Remove the batteries if you are not going to use the device for an extended period i e 3 months or more e Properly dispose of the batteries according to your local environmental regulations Caring for Your Meter Wash and dry your hands thoroughly before use or handling of the meter and strips What is Cleaning and Disinfection Cleaning and disinfection are different Cleaning is the process of removing dirt e g food debris grease dust disinfection is the process of killing germs e g bacteria and viruses When to clean and disinfect the meter Clean the meter when you see any dirt on it You should disinfect the meter at least once a week to prevent infection How to clean and disinfect the meter The meter must be cleaned prior to the disinfection Use one disinfecting wipe to clean exposed surfaces of the meter thoroughly and remove any visible dirt or blood or any other body fluid with the wipe Use a second wipe to disinfect the meter Do NOT use organic solvents to clean the meter We recommend using the product listed below to clean and disinfect your meter The recommended product contains isopropyl alcohol as its active 28 ingredient It has been shown to be safe when used on the ForaCare Premium V10 blood glucose meter gt Micro Kill
17. commended that you discard the first drop of blood as it might contain tissue fluid which may affect the test result 19 Alternative Site Testing Suggested Test Areas for the Hand Important There are limitations with Alternative Site Testing AST Please consult your health care professional before you perform AST When to use AST Food medication illness stress and exercise can affect blood glucose levels Capillary blood at the fingertip reflects these changes faster than capillary blood at other sites Thus when testing blood glucose during or immediately after a meal physical exercise or any other event take a blood sample from your finger only We strongly recommend that you perform AST ONLY at the following times e Ina pre meal or fasting state more than 2 hours since the last meal e Two hours or more after taking insulin e Two hours or more after exercise Do NOT use AST if e You think your blood glucose is low e You are unaware of hypoglycemia e You are testing for hyperglycemia e Your AST results do not match the way you feel e Your routine glucose results often fluctuate 20 Performing a Blood Glucose Test Ki E CH e qi 1 Insert the test strip to turn on the meter Wait for the meter to display E and flashing 4 Thank you for using this product Please relax during measurement Please apply blood into the strip 2 Select the appropriate meas
18. d to ensure that the meter and test strips are working together correctly Test strips control solutions or sterile lancets may not be included in the kit please check the contents on your product box They can be purchased separately Please make sure you have all items prior to testing your blood glucose Test with a control solution when you first receive the meter you first receive the meter you do your weekly routine quality testing of the meter and test strips you begin using a new vial of test strips you suspect the meter and or test strips are not working correctly your blood glucose test results are not consistent with how you feel or if you think the results are not accurate you are practicing the testing procedure you have dropped or think you may have damaged the meter Performing a Control Solution Test a Q E di 1 Insert the test strip to turn on the r z6 meter L Wait for the meter to display CE and g ig 0 Fi LI 71 1 a ee KI Voice Guide Thank you for using this product Please relax during measurement Please apply blood into the test strip 15 ES a E q b om ie BiG E 47 wo wo WARNING 2 Press until CTL is displayed With CTL displayed the meter will not store your test result in memory If you press again the CTL will disappear and this test is no longer a control solution test When doing the control so
19. do de glucosa en la sangre ser almacenado en la memoria autom ticamente La glucosa en la sangre es n mero por miligramos de decilitro 6 Expulse la tira reactiva y deseche la lanceta Para expulsar la tira reactiva coloque el medidor de forma que la tira est de frente a un basurero y oprima el bot n de expulci n El medidor se apagar autom ticamente despues de haber expulsado la tira Siempre siga las instrucciones del manual del dispositivo de punci n al remover la lanceta ADVERTENCIA La lanceta y las tiras reactivas usadas son consideradas un desecho de riesgo biol gico Por favor deseche de acuerdo a las recomendaciones de su m dico 60 MEMORIA DEL MEDIDOR Su medidor almacena en la memoria los 450 resultados de las pruebas de la glucosa en la sangre m s recientes junto con sus respectivas fechas y horas Para ingresar en la memoria de su medidor comience con el medidor apagado Revisando los Resultados de la Prueba q FI LI i 1 Pulse y suelte M Pulse y suelte O otra vez y la primera lectura que aparecer en pantalla ser el resultado de la ltima medici n de glucosa en la sangre con su fecha y hora as como el tipo de medici n La glucosa en la sangre es n mero por miligramos de decilitro 2 Presione para reviser los resultados guardados en el medidor cada vez que lo presiona Despu s del ltimo resultado de la prueba presi
20. dquirir por primera vez su gluc metro e Al menos una vez a la semana como prueba rutinaria para verificar el funcionamiento del gluc metro y de las tiras reactivas e Al empezar un envase nuevo de tiras reactivas e Si sospecha que el gluc metro o las tiras no est n funcionando correctamente e Silos resultados de la prueba de glucosa en la sangre no son consistentes a c mo se siente o si los resultados no son exactos e Al practicar el proceso de la prueba e Si ha dejado caer el gluc metro o si piensa que ste ha sido da ado Realizando una Prueba de la Soluci n E q ES ui 1 Inserte la tira reactiva para encender r LH S el medidor Espere hasta que el medidor muestre los s mbolos de la tira reactiva CB y la Am gota de sangre 6 Gracias por usar este producto Por 51 favor rel jese mientras se realiza su medici n Por favor aplique sangre en la tira reactiva di amp 2 Pulse hasta que aparezca CTL Cuando se muestra CTL el medidor no n almacenar el resultado del an lisis en la 2 10 g qa memoria Si presiona otra vez CTL LED 7 00 desaparecer y ya no estar m s en la funci n de prueba de soluci n de control ADVERTENCIA Al realizar el an lisis de la soluci n de control deber mostrarse CTL de modo que el resultado del an lisis no se almacene en la memoria Si no tiene el medidor en el modo correcto
21. e otros lugares que no son las yema del dedo Reemplace el dispositivo de punci n con la tapa transparente Deslice el bot n de expulsi n hacia atr s hasta que haga clic Cuando obtiene sangre en la palma de la mano el antebrazo el brazo la pantorrilla o el muslo tenga cuidado de la venas evidente debido a la hemorragia excesiva 55 NOTA Escoja una rea diferente cada vez que hace el examen Repetir la punci n en el mismo lugar puede causarle dolor o crearle callosidades e Antes de realizar la prueba LAP por favor primero consulte con su profesional m dico e Es recomendable descartar la primera gota de sangre ya que por lo general contiene otras sustancias que podr an afectar el resultado Lugar Alternativo de Prueba LAP reas de an lisis sugeridas para la Pantorrilla Importante Existen limitaciones para llevar a cabo el LAP Lugar Alternativo de Prueba Por favor consulte con su profesional medico antes de realizar el LAP Cu ndo utilizar el LAP La comida medicamentos enfermedad tensi n y los ejercicios pueden afectar los niveles de glucosa en la sangre La sangre capilar en la yema de los dedos refleja estos cambios m s r pido que en otras partes del cuerpo Cuando haga la prueba de glucosa en la sangre durante o inmediatamente despu s de comer o de haber hecho ejercicios u otro 56 evento de este tipo siempre tome la muestra de sangre nicamente de su dedo Le recomendamos
22. eemplazando la bateri Aseg rese de que el medidor est apagado antes de sacar las bater as 1 Retire la tapa de las pilas 2 Retire las pilas antiguas y coloque dos pilas alcalinas de tama o AAA y 1 5 V 3 Cierre la tapa de las pilas prestando especial atenci n a la polaridad de las pilas Si las pilas se han insertado correctamente oir un pitido NOTA e Reemplazando las bater as no afecta los resultados guardados en la memoria 64 e Como con todas las bater as peque as stas deben estar fuera del alcance de ni os peque os En caso de que ser ingeridas por favor busque ayuda m dica lo m s pronto posible e Las bater as podr an derramar qu micos si el aparato no se usa por mucho tiempo Remueva las bater as si usted no va a usar el aparato por un largo per odo de tiempo por Ej 3 meses o m s Deseche las bater as de acuerdo con las regulaciones ambientales de su localidad Cuidados de su Medidor L vese y s quese las manos a fondo antes de usar o manipular el medidor y las tiras Qu es la limpieza y qu la desinfecci n La limpieza y la desinfecci n son cosas distintas La limpieza es el proceso de quitar suciedad como restos de comida grasa polvo mientras que la desinfecci n es el proceso de matar los g rmenes como bacterias y virus Cu ndo limpiar y desinfectar el medidor Limpie el medidor cuando vea suciedad en l Debe desinfectar el medidor al menos una vez a la se
23. ente hasta obtener otra gota de sangre Por favor tenga cuidado de no esparcir la muestra de la sangre El volumen de la muestra debe ser de al menos 0 5 microlitos e tama o real 58 4 Aplicando la muestra Sujete el medidor con la tira hacia abajo de modo que el orificio absorbente de la tira reactiva toque la gota de sangre La sangre llegar a la tira Una vez que la ventana de confirmaci n est totalmente llena el medidor comienza la cuenta atr s No quite la tira de la gota de sangre hasta que el medidor emita un pitido NOTA e No oprima la zona de punci n contra la tira reactiva ni trate de aplicar una muestra de sangre que se haya esparcido El medidor se apagar autom ticamente si no aplica la muestra de sangre en 3 minutos Si esto ocurriera deber de remover y reinsertar la tira reactiva y comenzar el procedimiento de nuevo e La sangre debe llenar la ventana de confirmaci n completamente antes de la cuenta regresiva del medidor Si encuentra que la ventana no est llena NUNCA trate de a adir m s sangre a la tira reactiva Deseche la tira reactiva y haga la prueba otra vez con una tira nueva e Si tiene problemas llenando la ventana de confirmaci n contacte con su m dico o con el servicio al cliente de su localidad para recibir m s ayuda 59 5 Leyendo sus resultados El resultado de la glucosa en la sangre aparecer en su medidor despu s del conteo regresivo llegue a O Este resulta
24. epeat the test with a new test strip Sample applied after automatic switch off 3 minutes after last user action Defective meter Repeat the test with a new test strip Apply sample only when flashing 6 appears on the display Please contact customer service 3 If the control solution testing result is out of range POSSIBLE CAUSE WHAT TO DO Error in performing the test Read instructions thoroughly and repeat the test again Control solution vial was poorly shaken Shake the control solution vigorously and repeat the test again Expired or contaminated control solution Check the expiry date of the control solution Control solution that is too warm or too cold Control solution meter and test strips should be at room temperature 20 C to 25 C 68 F to 77 F before testing Defective test strip Repeat the test with a new test strip Meter malfunction Please contact customer service 36 DETAILED INFORMATION The meter provides you with plasma equivalent results Time of day Normal plasma glucose range for people without diabetes mg dL Fasting and before meal lt 100 mg dL 5 6 mmol L 2 hours after meals lt 140 mg dL 7 8 mmol L Source American Diabetes Association 2010 Clinical Practice Recommendations Diabetes Care 33 Supplement 1 S1 S100 Please consult your doctor to determine
25. er resultados exactos Las condiciones de almacenamiento deben estar entre temperaturas de 35 6 F a 86 F 2 C a30 C NO la congele 69 LOCALIZACION Y RESOLUCION DE PROBLEMAS EN EL SISTEMA Si usted sigue las acciones recomendadas pero el problema persiste o hay mensajes de error distintos a los que aparecen a continuaci n por favor llame a su agente de servicio al cliente local No trate de reparar por s mismo el aparato y bajo ninguna circunstancia trate de desensamblar el medidor Lectura de resultados MENSAJE Cu ndo Glucosa Aparece cuando su resultado est por debajo del l mite de la medida que es menor de 20 mg dL 1 1 mmol L Aparece cuando el resultado es igual o mayor que 240 mg dL 13 3 mmol L Si usted tiene diabetes tipo 1 esto indica la posibilidad de que haya una acumulaci n de cetona Busque asistencia m dica de inmediato Aparece si el resultado del an lisis de glucosa en sangre es superior al l mite superior de medida que es 600 mg dL 33 3 mmol L 70 Mensajes de Error MENSAJE QU DICE EL QUE SIGNIFICA ACCION MEDIDOR E E p Solo pitido de error P Repase las E 3 Solo pitido 3 a a instrucciones y repita de error gl el an lisis Si el Problema con el medidor problema cont nua por favor contacte con E A Solo pitido are a el servicio al cliente de error para la asistencia Code EJ 8 Solo pitido de e
26. formaci n acerca del Health Care System Software o para obtener un cable USB visite www foracare com o llame al n mero del servicio de atenci n al cliente para solicitar ayuda 1 Adquiriendo el cable requerido e instalando el software Para descargar el software Sistema de Cuidados de la Salud Health Care Software System en su computadora visite la p gina web de ForaCare Inc www foracare com 2 Conectando a una computadora personal Conecte el cable de transmisi n a un puerto USB de su computadora Con el medidor apagado conecte el otro extremo del cable USB al puerto de datos del medidor Aparecer en pantalla USb indicando que el medidor est en funci n de comunicaci n 3 Transmisi n de datos Siga las instrucciones suministradas por el software para transmitir los datos Los resultados ser n transmitidos con la hora y la fecha Una vez transmitidos los resultados quite el cable y el medidor se apagar autom ticamente ADVERTENCIA No se podr realizar ninguna prueba de glucosa en la sangre mientras el medidor est conectado a su PC 63 MANTENIMIENTO Su medidor viene con dos bater as alcalinas AA A 1 5V Se al de bater a agotada o baja El medidor mostrar el mensaje siguiente para avisarle cuando se est n agotando las pilas pae 4 El s mbolo aparecer con E b y low E 5 bateria no tiene suficiente bateria para ejecutar la prueba Tiene que cambiarla inmediatamente R
27. ice for assistance 5 Read your result The result of your blood glucose test will appear after the meter counts down to O This blood glucose result will automatically be stored in the memory The blood glucose is number milligrams per deciliter 6 Eject the used test strip and remove the lancet To eject the test strip point the strip into a disposal container for contaminated items Slide the Test Strip Ejector forward The meter will automatically turn itself off after the test strip is ejected When removing the lancet follow the instructions from the lancing device insert The used lancet and test strip may be biohazardous Please follow the proper procedure for discarding biohazardous waste in your area 23 METER MEMORY The meter stores up to 450 most recent blood glucose test results with date and time stamp in its memory To enter the meter memory function start with the meter turned off Reviewing Test Results i mi LI 1 gt coco as DAL KO mg dL am M 24 1 Press and release M will appear on the display Press again and the first reading you see is the last blood glucose result along with date time and the measurement mode The blood glucose is number milligrams per deciliter 2 Press again to recall the next test results stored in the meter each time you press After the last test result press O again and the meter will tu
28. iente 68 F a 77 F 20 C a 25 C antes de realizar la prueba Tira reactiva defectuosa Repita la prueba con una nueva tira reactiva Malfuncionamiento del medidor Por favor contacte con servicio al cliente INFORMACION DETALLADA El medidor proporciona resultados equivalentes en plasma 73 Hora del d a Rango normal del plasma de la glucosa para personas sin diabetes mg dL En ayunas o antes de comer lt 100 mg dL 5 6 mmol L 2 horas despu s de comer lt 140 mg dL 7 8 mmol L Source American Diabetes Association 2010 Clinical Practice Recommendations Diabetes Care 33 Supplement 1 S1 S100 Por favor consulte con su m dico para determinar cual es el rango que mejor se adapta a sus necesidades ESPECIFICACIONES Modelo No TD 4124 Dimensiones y peso 88 67 L x 55 61 W x 23 07 H mm 48 8 g Fuente de energ a dos bater as alcalinas AA A 1 5V Pantalla LCD Memoria 450 resultados de mediciones con su fecha y hora respectiva Salida externa USB Auto detecci n al insertar electrodo Auto detecci n en la muestra de carga Auto reacci n de tiempo en cuenta regresiva Se desconecta autom ticamente despu s de 3 minutos sin acci n Advertencia de Temperatura Condiciones de operaci n 50 F a 104 F 10 C a 40 C bajo un 85 de humedad relativa no condensando Storage Transportation Conditions 4 F a 140 F 20 C a 60 C bajo un 95 de
29. liza an lisis e Si el medidor es usado por otra persona ser necesario descontaminar tanto el medidor como la lanceta antes de su uso por la segunda persona NO permita que la soluci n de limpieza y desinfecci n acceda a la ranura para tiras reactivas al compartimiento de las pilas o al bot n de expulsi n de tiras e Si moja la ranura para tiras reactivas limpiela con el pico de un tissue Seque a fondo el medidor antes de su uso No roc e el medidor directamente con soluciones de limpieza en especial las que contengan agua agua jabonosa pues de este modo la soluci n podr a acceder a la carcasa interior y da ar los componentes o circuitos electr nicos 67 Almacenamiento del medidor e Condiciones de almacenamiento 4 F a 140 F 20 C a 60 C bajo un 95 de humedad relativa e Siempre guarde o transporte el medidor en su estuche original e Evite que el medidor se caiga o que sufra impactos fuertes e Evite la luz solar directa y la alta humedad Cuidados de las Tiras Reactivas e Condiciones de almacenamient 39 2 F a 104 F 4 C a 40 C bajo un 85 de humedad relativa NO las congele e Almacene las tiras reactivas solamente en su envase original No las transfiera a otro contenedor e Guarde el envase de las tiras reactivas en un lugar fresco y seco Mant ngalo alejado del calor y de la luz solar directa e Cierre bien el envase inmediatamente despu s de remover la tira reactiva Siempre mani
30. lution test CTL must be displayed so that the test results will not be stored in the memory Failure to have the meter in the correct mode will result in having the control solution test results stored with your actual blood glucose test results This will cause an inaccurate analysis of your stored results 3 Apply control solution Shake the control solution bottle thoroughly before use Squeeze out a drop onto the tip of the bottle cap and wipe it off using a clean tissue or cloth Then squeeze out another drop onto the tip of the bottle cap Hold the meter so that the strip is pointing downward and move the absorbent hole of the test strip to touch the drop Once the confirmation window fills completely the meter will begin counting down To avoid contaminating the control solution do not directly apply control solution onto a strip or otherwise touch the nozzle of the bottle 16 4 Read and compare the result After counting down to O the control solution test result will appear on the display Compare this result with the range printed on the test strip vial It should fall within the range If not please read the instructions again and repeat the control solution test The blood glucose is number milligrams per deciliter o gt Normal 100 165 Out of range results If you continue to have test results fall outside the range printed on the test strip vial the meter and strips may not be worki
31. mana para impedir infecciones C mo limpiar y desinfectar el medidor Es necesario limpiar el medidor antes de su desinfecci n Utilice un pa o desinfectante para limpiar las superficies expuestas del medidor a fondo y retire la suciedad sangre u otro l quido corporal visible con el pa o Utilice un segundo pa o para desinfectar el medidor siguiendo el procedimiento de desinfecci n que se indica a continuaci n NO utilice solventes org nicos para limpiar el medidor 65 Recomendamos el uso del producto indicado a continuaci n para limpiar y desinfectar el medidor El producto recomendado contiene alcohol isoprop lico como ingrediente activo Se ha demostrado que se puede usar con seguridad en el medidor de glucosa en sangre ForaCare Premium V10 gt Micro Kill Micro Kill Plus de Medline EPA Reg No 59894 10 37549 Para obtener pa os desinfectantes y otra informaci n p ngase en contacto con Medline en el n mero 1 800 MEDLINE 1 800 633 5463 o visite Procedimientos para la desinfecci n 1 Saque un pa o desinfectante del paquete y esc rralo para quitarle el posible exceso de l quido con el fin de evitar da ar el medidor 2 Limpie todas las superficies exteriores del medidor incluida la pantalla y los botones Sujete el medidor con la ranura para tiras reactivas hacia abajo y limpie la zona que la rodea cuidando no mojar en exceso el interior de la ranura Mantenga el medidor h medo con la soluci
32. n desinfectante del pa o durante al menos 2 minutos o bien el tiempo de contacto m nimo que especifique el fabricante del pa o desinfectante Siga las instrucciones que aparecen en la etiqueta del envase del pa o desinfectante Utilice dos o m s pa os en caso necesario 66 3 Retire el pa o Deje que se seque totalmente la superficie del medidor 4 Deseche los pa os usados La limpieza y desinfecci n inadecuadas del sistema puede provocar aver as en el medidor Si desea realizar alguna pregunta p ngase en contacto con el servicio al cliente en el tel fono 1 866 469 2632 Se ha validado que el dispositivo puede soportar 5 000 ciclos de desinfecci n usando el pa o de desinfecci n recomendado El n mero de ciclos comprobado se estima en 1 000 an lisis al a o en cada medidor y unos 3 ciclos de desinfecci n de cada medidor al d a durante 5 a os la vida til esperada del dispositivo Deje de usar el medidor si ve alg n signo de deterioro como que la pantalla LCD parezca rasgada o con niebla que los botones no funcionen o que la carcasa exterior est rajada P ngase en contacto con el servicio local al cliente en el tel fono 1 866 469 2632 para solicitar un medidor de repuesto si observa alg n signo de deterioro El medidor debe sustituirse despu s del n mero validado de ciclos de desinfecci n o del per odo de garant a lo primero que suceda NOTA e NO limpie ni desinfecte el medidor mientras rea
33. ne Pathogens Initial Communication 2010 http www fda gov MedicalDevices Safety AlertsandNotices ucm224025 htm 2 CDC Clinical Reminder Use of Fingerstick Devices on More than One Person Poses Risk for Transmitting Bloodborne Pathogens 2010 http www cdc gov injectionsafety Fingerstick DevicesBGM html Use this device ONLY for the intended use described Do NOT use accessories which are not specified by the manufacturer Do NOT use the device if it is not working properly or if it is damaged BWNHER This device does NOT serve as a cure for any symptoms or diseases The data measured is for reference only 5 Before using this device to test blood glucose read all instructions thoroughly and practice the test Carry out all the quality control checks as directed 6 Use of this instrument in a dry environment especially if synthetic materials are present synthetic clothing carpets etc may cause damaging static discharges that may cause erroneous results 7 Do NOT use this instrument in close proximity to sources of strong electromagnetic radiation as these may interfere with the accurate operation KEEP THESE INSTRUCTIONS IN A SAFE PLACE TABLE OF CONTENTS BEFORE YOU BEGIN 06 Important Information 06 Intended Use 07 Test Principle 07 Meter Overview 08 Display Screen 09 Test Strip 10 Setting the Meter 11 THE TWO MEASURING MODES 14 Control Solution Testing 15 Performing a Co
34. ng properly Do NOT test your blood using this meter Go to www foracare com or call your customer service for help NOTE e The control solution range printed on the test strip vial is for control solution use only It is not a recommended range for your blood glucose level e See the Meter Maintenance section for important information about your control solutions 17 TESTING WITH BLOOD SAMPLE Preparing the Lancing Device for Blood Testing Please follow the instructions in the lancing device insert for collecting a blood sample WARNING To reduce the chance of infection e Never share a lancet or the lancing device e Always use a new sterile lancet Lancets are for single use only e Avoid getting hand lotion oils dirt or debris in or on the lancets or the lancing device e Wash and dry your hands thoroughly before and after handling the meter lancing device and test strips to prevent infection For more information please refer to the Cleaning and Disinfection section e If the meter is being operated by a person who is providing testing assistance to the user the meter and lancing device should be decontaminated prior to use by that person Sharing the lancing device and lancets may increase the risk of spreading infectious diseases Lancing device must not be used on more than one person Preparing the Puncture Site Rubbing the puncture site before blood extraction has a significant influence on the acc
35. ngs lower than actual values may occur for individuals experiencing a hyperglycemic hyperosmolar state with or without ketosis Please consult a healthcare professional before use e Critically ill patients should not be tested with a blood glucose monitoring system Intended Use The ForaCare Premium V10 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites palm forearm upper arm calf and thigh It is intended for use by people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program It is not intended for the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates Professionals may test with capillary and venous blood sample home use is limited to capillary whole blood testing The alternative site testing in the ForaCare Premium V10 Blood Glucose Monitoring System can be used only during steady state blood glucose conditions This meter has some speaking functions but is not intended for use by the visually impaired Test Principle Your system measures the amount of sugar glucose in whole blood The glucose testing is based on the measurement of electrical current generated by the reaction of glucose with the reagent of the strip The meter measures the current calculates the blood glucose level and displays the result The s
36. nsfer to another container e Store test strip packages in a cool dry place Keep away from direct sunlight and heat e After removing a test strip from the vial immediately close the vial cap tightly e Touch the test strip with clean and dry hands e Use each test strip immediately after removing it from the vial e Write the opening date on the vial label when you first opened it Discard remaining test strips after 3 months of first opening e Do not use test strips beyond the expiry date This may cause inaccurate results e Do not bend cut or alter a test strip in any way e Keep the strip vial away from children since the cap and the test strip may be a choking hazard If swallowed promptly see a doctor for help For further information please refer to the test strip package insert 31 Important Control Solution Information 32 Use only our control solutions with your meter Do not use the control solution beyond the expiry date or 3 months after first opening Write the opening date on the control solution bottle and discard the remaining solution after 3 months of first opening It is recommended that the control solution test be done at room temperature 68 F to 77 F 20 C to 25 C Make sure your control solution meter and test strips are at this specified temperature range before testing Shake the bottle before use discard the first drop of control solution and wipe off the dispenser tip to ensure a pu
37. ntrol Solution Test 15 TESTING WITH BLOOD SAMPLE 18 Preparing the Puncture Site 18 Alternative Site Testing 20 Performing a Blood Glucose Test 21 METER MEMORY 24 Reviewing Test Results 24 Reviewing Blood Glucose Day Average Results 25 Downloading Results onto a Computer 26 MAINTENANCE 27 Battery 27 Caring for Your Meter 28 Caring for Your Test Strips 31 Important Control Solution Information 32 SYSTEM TROUBLESHOOTING 33 Result Readings 33 Error Messages 34 Troubleshooting 36 DETAILED INFORMATION N SPECIFICATIONS 38 5 BEFORE YOU BEGIN e Severe dehydration and excessive water loss may cause readings which are lower than actual values If you believe you are suffering from severe dehydration consult a healthcare professional immediately e If your blood glucose results are lower or higher than normal and you do not have any symptoms of illness repeat the test If you have symptoms or continue to get results which are higher or lower than normal consult with your healthcare professional e Use only fresh whole blood samples to test your blood glucose Using other substances will lead to incorrect results e If you are experiencing symptoms that are inconsistent with your blood glucose test results and you have followed all the instructions given in this owner s manual contact your healthcare professional e We do not recommend using this product on severely hypotensive individuals or patients in shock Readi
38. oducto en individuos seriamente hipotensos o pacientes en estado de shock Pueden ocurrir resultados m s bajos de lo normal si el individuo experimenta un estado hiperglic mico hiperosmolar con o sin cetosis Por favor consulte con su profesional m dico antes de utilizar este aparato e No se deben realizar an lisis a pacientes gravemente enfermos con un sistema de control de glucosa en sangre 42 El Sistema de control de glucosa en sangre ForaCare Premium V10 se ha dise ado para su uso en la medici n cuantitativa de glucosa en sangre entera y fresca de capilar del dedo y de los siguientes sitios alternativos palma de la mano antebrazo brazo pantorrilla y muslo Est indicado para su uso para personas con diabetes mellitus en casa como ayuda en el control de la eficacia del programa de control de diabetes No est indicado para el diagn stico o screening de diabetes mellitus como tampoco para su uso en neonatos Los an lisis en sitios alternativos para el Sistema de control de glucosa en sangre ForaCare Premium V10 solo se pueden realizar en condiciones de glucosa en sangre estable Este medidor tiene funci n de voz habla pero no est dise ado para su uso por los discapacitados visuales Principios de la Prueba Su sistema mide la cantidad de azucar glucosa en la sangre La prueba de glucosa en la sangre se basa en la medici n del fluido el ctrico generado por la reacci n de la glucosa con el agente reactivo de
39. one O otra vez y el medidor se apagar La glucosa en la sangre es n mero por miligramos de decilitro 61 Revisando los Resultados del Promedio Diario de la Glucosa dy 1 Pulse y suelte Cuando M aparezca en pantalla mantenga el bot n pay Libere el bot n y el resultado in ess del promedio de 7 d as en function O 0 00 general aparecer en la pantalla 2 Pulse el bot n para revisar los resultados medios de 14 21 28 60 y 90 d as almacenados en memoria 3 Salir de la funci n de memoria Mantenga presionado el bot n y el medidor se apagar mostrando el ltimo resultado NOTA e En cualquier momento que desee salir de la funci n de memoria mantenga Presionado O por 5 segundos o simplemente no realice ninguna acci n por 3 minutos y el medidor se apagar autom ticamente e Los resultados de la soluci n de control NO est n incluidos en el promedio diario e Si est usando el medidor por primera vez era mostrado cuando solicita los resultados de prueba o revisa el resultado promedio Esto significa que no hay resultados de prueba en la memoria 62 Descargando los Resultados a la Computadora Transmisi n de datos por cable Usted puede usar su medidor con un ForaCare 4 pin USB cable y el software del Sistema de Cuidados de la Salud Health Care Software System para mirar los resultados de las pruebas en su computadora personal Para obtener m s in
40. orrecta aparezca g el g Mal Presione SET LI UM WAN ay FU ALIZLILI ILILILF 47 yer ici OO OO HES na O i qi reo LOL t CT O po 48 Con el minuto parpadeando presione hasta que el minuto correcto aparezca Presione SET 3 Borrando la Memoria Con los s mbolos dEL y M parpadeando en pantalla presione y seleccione no para conservar los resultados en memoria luego presione SET para saltarlos Cuando parpadeen los s mbolos dEL y M en la pantalla pulse para eliminar todas las lecturas de la memoria La pantalla mostrar los s mbolos y pulse el bot n SET para ajustar el volumen del altavoz y para apagar el medidor 4 Seleccionando el Volumen de la Voz Hay ocho 8 opciones que usted puede escoger para el volumen de voz Pulse hasta que aparezca el volumen de altavoz que desee Para confirmar su selecci n presione SET El volumen en 0 indica que la funci n de voz est apagada y di no aparecer durante la prueba El volumen de 1 a 7 indica el sonido de la voz de 1 volumen bajo a 7 alto y di aparecer en pantalla mientras se realiza la prueba 5 Seleccionando el lenguaje Presione para seleccionar el L1 L2 El lenguaje predeterminado es L1 que es ingl s L2 es el idioma alternativo indicado en el envase Para confirmar la selecci n presione
41. pule la tira reactiva con las manos limpias y secas e Use cada tira reactiva inmediatamente despu s de sacarla del envase e Escriba la fecha de expiraci n en la etiqueta cuando abra el frasco por primera vez Despu s de 3 meses deseche las tiras reactivas restantes e Nunca use tiras reactivas que est n expiradas ya que puede causar resultados inexactos e No doble corte o altere la tira reactiva de ninguna manera e Mantenga el envase de tiras fuera del alcance de ni os ya que de ingerirse podr a causar asfixia De ser ingerido busque atenci n m dica inmediatamente Para mayor informaci n por favor refierase al inserto en el paquete de tiras reactivas 68 Informaci n Importante sobre la Soluci n de Use solamente la soluci n de control con su medidor Nunca use soluciones de control que fueron abiertas por m s de 3 meses o que est n expiradas Escriba la fecha en la que abri el envase de la soluci n de control en la etiqueta del mismo y des chela despu s de 3 meses Es recomendable que la prueba para soluci n de control se lleve a cabo a temperatura ambiente de 68 F a 77 F 20 C a 25 C Aseg rese de que su soluci n de control su medidor y sus tiras reactivas est n dentro del rango especificado antes de realizar la prueba Agite bien el envase antes de usarlo descarte la primera gota de soluci n de control y limpie la punta del dispensador para asegurarse de que la muestra est pura y pueda obten
42. r as Tira reactiva colocada al rev s o de modo incompleto Inserte la tira reactiva hacia arriba con las barras de contacto primero Medidor defectuoso Por favor contacte con servicio al cliente 72 2 Si la prueba no comienza despu s de colocar la muestra CAUSAS POSIBLES ACCI N Muestra de sangre insuficiente Repita la prueba con una nueva tira reactiva aumentando la cantidad de la muestra de sangre Tira reactiva defectuosa Repita la prueba con una nueva tira reactiva Muestra aplicada despu s del apagado autom tico 3 minutos despu s de la ltima acci n del usuario Repita la prueba con una nueva tira reactiva Apliqu la muestra solamente cuando aparece en la pantalla Medidor defectuoso Por favor contacte con servicio al cliente 3 Si el resultado de la soluci n de control est fuera de rango CAUSAS POSIBLES ACCI N Error en la realizaci n de la prueba Lea las instrucciones cuidadosamente y repita nuevamente la prueba No agit lo suficiente la soluci n de control Agite vigorosamente la soluci n de control y repita la prueba nuevamente Soluci n de control expirada o contaminada Verifique la fecha de expiraci n de la soluci n de control La soluci n de control est muy caliente o fr a La soluci n de control el medidor y las tiras reactivas debe estar a temperatura amb
43. re sample and an accurate result Store the control solution tightly closed at temperatures between 35 6 F and 86 F 2 C and 30 C Do NOT freeze SYSTEM TROUBLESHOOTING If you follow the recommended action but the problem persists or error messages other than the ones below appear please call your local customer service Do not attempt to repair the meter yourself and never try to disassemble the meter under any circumstances Result Readings MESSAGE WHAT IT MEANS Appears when your blood glucose test result is below the lower measurement limit which is less than 20 mg dL Appears when your result is equal to or higher than 240 mg dL This indicates the possibility of ketone accumulation for type 1 diabetes Please seek medical assistance immediately Appears when your blood glucose test result is higher than the upper limit of measurement which is 600 mg dL 33 Error Messages LI a Only Error Beep MESSAGE WHAT DOES WHAT IT MEANS WHAT TO DO METER SAY cg E LI Only Error Beep E J Only Error Beep Review the instructions and repeat the test If Problem with the the meter still does not E A meter work contact the local Only Error Beep distributor for assistance m Li Tr Only Error Beep Problem with the meter Repeat the test with a new test strip If the meter still does not work please contact the customer service for assistance a im 0 OF
44. resultado sigue estando fuera de rango el sistema podr a estar trabajando incorrectamente NO utilice el sistema para hacer ex menes de sangre Contacte a su distribuidor local o bien al servicio al cliente NOTA e El rango impreso en el envase de las tiras reactivas deben ser utilizados nicamente para verificar la soluci n de control Este rango NO es recomendado para los niveles de su glucosa en la sangre e Vea la secci n de Mantenimiento para obtener mayor informaci n sobre la soluci n de control 53 HACIENDO UNA PRUEBA CON MUESTRA DE SANGRE Preparando el Dispositivo de Punci n para Hacer la Prueba Por favor siga las instrucciones en el manual del dispositivo de punci n para obtener una muestra de sangre ADVERTENCIA Para reducir el riesgo de infecci n Nunca comparta la lanceta o el dispositivo de punci n Siempre utilice una lanceta nueva y esterilizada Las lancetas deben ser utilizadas SOLO una vez Al utilizar las lancetas o el dispositivo de punci n evite el uso de crema de manos aceite polvo o cualquier tipo de sustancia contaminante Lave y s quese las manos a fondo despu s de manipular el medidor la lanceta y las tiras de muestra para evitar infecciones Para obtener m s informaci n consulte el apartado Limpieza y desinfecci n Si el medidor es usado por otra persona que ayude al usuario sera necesario descontaminar tanto el medidor como la lanceta antes de su uso por la segunda persona
45. rn off La glucosa en la sangre es n mero por miligramos de decilitro Reviewing Blood Glucose Day Average Results EJ 4 1 Press and release When M appears on the display keep pressing M for 3 seconds until the flashing Bax Big E Du appears Release and then your 7 day ee 10 00 average result measured in general mode will appear on the display 2 Press to review 14 21 28 60 and 90 day average results stored in memory 3 Exit the meter memory Press after the 90 day average is displayed and the meter will turn off NOTE e Any time you wish to exit the memory hold press O down for 5 seconds or leave the meter idle for 3 minutes The meter will turn off automatically e Control solution results are NOT included in the day average If there are no records in memory displays when you recall the test results or review the average result 25 Downloading Results onto a Computer Data Transmission Via Cable You can use the meter with the ForaCare 4 pin USB cable and the Health Care System Software to view your test results on your personal computer To learn more about the Health Care System Software or to obtain an USB cable please visit www foracare com or call our customer care service number for further assistance 1 Obtaining the required cable and installing the software To download the Health Care System Software please visit the ForaCare Inc website a
46. rror Repita la prueba con a una nueva tira reactiva E E Solo pitido Problema con el medidor Si el problema cont de error nua por favor contacte con el servicio al cliente para la asistencia E ki n Parece que la bater a no Remplace la ken Solo pitido tiene suficiente energ a bater a de error para hacer otra prueba inmediatamente 71 MENSAJE QU DICE EL MEDIDOR QU SIGNIFICA ACCI N aw Aparece cuando la tira reactiva se retira durante Haga la prueba operacional del sistema E F Solo pitido la cuenta atr s o cuando eon una hueva tira de error el volumen de sangre no reactiva es suficiente gt a Aparece cuando se inserta Haga la prueba E LJ Solo pitido una tira reactiva que esta con una nueva tira de error usada reactiva a Aparece cuando la El rango operacional E E Solo pitido temperature ambiental es Del sistema es 50 F tou d s tiok m s baja que la temperature a 104 F 10 C a operacional del sistema 40 C Repita la E Aparece cuando la prueba despu s de E E Solo pitido temperature ambiental que el medidor y la de error est por encima del rango __ tira reactiva est n a temperatura ambiente Localizaci n y Resoluci n de Problemas 1 Si el medidor no muestra ning n mensaje despu s de introducer una tira reactiva CAUSAS POSIBLES ACCI N Bater a agotada Reemplace las bate
47. seriamente que usted realice el LAP SOLAMENTE en los siguientes intervalos e Antes de una comida o en ayunas espere m s de 2 horas desde la ltima comida e Dos horas o m s despu s de aplicar insulina e Dos horas o m s despu s de haber hecho ejercicios NO utilice el LAP si e Usted cree que la glucosa en la sangre est baja e A menudo no se dan cuenta cuando su nivel de glucosa en sangre esta bajo e Usted cree que su glucosa en la sangre esta alta e Los resultados del LAP no corresponden a como se siente e Los resultados rutinarios de la glucosa fluct an con frecuencia 57 Realizando la Prueba de Glucosa en la Sangre ET di ES Wi 1 Inserte la tira reactiva para encender PL S el medidor L Espere hasta que el medidor muestre los Bein s mbolos de la tira reactiva CB y la O gota de sangre 6 Gracias por usar este producto Por favor rel jese mientras se realiza su medici n Por favor aplique sangre en la tira reactiva 2 cSeleccione el modo de medici n adecuado General o QC pulsando Y Para seleccionar la funci n de medici n por favor refi rase a la secci n LAS DOS MODALIDADES DE MEDICI N 3 Obteniendo la muestra de sangre VS Utilice el dispositivo de punci n anteriormente preparado y haga la punci n en el lugar deseado Despu s de la penetraci n descarte la primera x gota de sangre con un algod n limpio Presione el lugar de punci n gentilm
48. t www foracare com 2 Connecting to a personal computer Connect the cable to a USB port on your computer With the meter turned off connect the other end of the USB cable to the meter data port USb will appear on the meter display indicating that the meter is in communication mode 3 Data transmission To transmit data follow the instructions provided with the software Test results will be transmitted with date and time stamp After the results have been transmitted remove the cable meter will automatically turn off WARNING While the meter is connected your computer it will not be able to perform a blood glucose test 26 MAINTENANCE Your meter comes with two 1 5V AAA size alkaline batteries Low Battery Signal The meter will display the message below to alert you when the meter power is getting low 4 The appears with E b and low 5 The power is not enough to perform testing Please change the batteries immediately E Replacing the Battery To replace the batteries make sure that the meter is turned off 1 Remove the battery cover 2 Remove the old batteries and replace with two 1 5V AAA size alkaline batteries 3 Close the battery cover paying close attention to the polarity of the batteries If the batteries are inserted correctly you will hear a beep NOTE e Replacing the batteries does not affect the test results stored in the memory 27 e As with all small
49. trength of the current produced by the reaction depends on the amount of glucose in the blood sample Meter Overview 1 Test Strip Slot Insert test strip here to turn the meter on for testing 2 Test Strip Ejector Eject the used strip by sliding up this button 3 Display Screen 4 M Button Display the meter memory silence a reminder alarm change a setting in SET mode 5 Data Port Download test results with a cable connection 6 SET Button Enter SET mode and confirm the meter settings Display Screen menn UN la jij HILL ce mg dL u aaa M 1 Volume Symbol 2 Insert Test strip 3 Test Strip Symbol 4 Test Result 5 Blood Drop Symbol 6 Date 7 Time Memory Symbol Ketone Warning Day Average Control Solution Mode Measurement Unit Low Battery Symbol Test Strip Absorbent Hole Touch to a drop of blood here blood will de Confirmation Window be automatically Confirm if enough blood drawn in has been applied to the absorbent hole of the strip Test Strip Handle ii Hold this part to insert the test strip Contact Bars into the slot Insert this end of the test strip into the test strip slot ATTENTION The front side of test strip should face up when inserting test strip Test results might be wrong if the contact bar is not fully inserted into the test slot NOTE The ForaCare Premium V10 monitor should only be used with For
50. udy the buttons no longer function or the outer casing is cracked Please contact the customer service at 1 866 469 2632 for a replacement meter if any of the signs of deterioration are noticed The meter should be replaced after the validated number of disinfection cycles or the warranty period which ever comes first NOTE Do NOT clean or disinfect the meter while performing a test f the meter is being operated by a second person the meter and lancing device should be decontaminated prior to handling by an assistant Do NOT allow cleaning or disinfecting solution to get in the test slot battery compartment or strip ejection button f you do get moisture in the test strip slot wipe it away with the corner of a tissue Always dry the meter thoroughly before using it Do not spray the meter directly with cleaning solutions especially those containing water i e soapy water as this could cause the solution to enter the case and damage the electronic components or circuitry 30 Meter Storage e Storage conditions 4 F to 140 F 20 C to 60 C below 95 relative humidity e Always store or transport the meter in its original storage case e Avoid dropping and heavy impact e Avoid direct sunlight and high humidity Caring for Your Test Strips e Storage conditions 39 2 F to 104 F 4 C to 40 C below 85 relative humidity Do NOT freeze e Store your test strips in their original vial only Do not tra
51. uracy of your test results Blood from an alternate a site that has not been rubbed exhibits a measurably different glucose concentration than blood from the finger When the puncture site was rubbed prior to blood extraction the difference was significantly reduced 18 Please follow the suggestions below before obtaining a drop of blood Wash and dry your hands before starting e Select the puncture site either at fingertips or another body part please see section Alternative Site Testing AST on how to select the appropriate sites e Clean the puncture site using cotton moistened with 70 alcohol and let it air dry e Rub the puncture site for about 20 seconds before penetration e Use a clear cap while setting up the lancing device e Fingertip testing Press the tip of the lancing device firmly against the lower side of your fingertip Press the release button to puncture your finger a click indicates that the puncture is complete e Blood from sites other than the fingertip Replace the lancing device cap with the clear cap for AST Pull the cocking control back until it clicks When lancing the forearm upper arm hana thigh or calf avoid lancing the areas with obvious veins to avert excessive bleeding NOTE e Choose a different spot each time you test Repeated punctures at the same spot may cause soreness and calluses e Please consult your health care professional before you begin AST e It is re
52. uring mode General or QC by pressing For selecting the measurement mode please refer to the TWO MEASURING MODES SYY 3 Obtaining a blood sample Use the pre set lancing device to puncture the chosen site After lancing wipe away the first drop of blood with a clean cotton swab Gently squeeze the punctured area to obtain another drop of blood Be careful NOT to smear the blood sample The volume of blood sample must be at least 0 5 microliter UL of volume o actual size 21 4 Apply the sample Hold the meter with the strip pointing downward and the absorbent hole of the test strip touching the blood drop Blood will be drawn in to the strip Once the confirmation window is completely filled meter begins counting down Do not remove the strip from the blood drop until the meter beeps NOTE 22 Do not press the punctured site against the test strip or try to smear the blood e If you do not apply a blood sample to the test strip within 3 minutes the meter will automatically turn off You must remove and reinsert the test strip to start a new test e The confirmation window should be filled with blood before the meter begins to count down NEVER try to add more blood to the test strip after the drop of blood has moved away Discard the used test strip and retest with a new one If you have trouble filling the confirmation window please contact your health care professional or our customer care serv