Bedienungsanleitung - Buhler Technologies LLC

image

Contents

1. Anschlussleitung 2 5m 91 44 05 00 09 7 7 Beiliegende Unterlagen Datenblatt TGAK 3 DD 45 00 08 Flussplan CS P01 10 4 Konformitatserklarung KX 45 00 06 Dekontaminierungserklarung 18 BXCS3000 02 2012 Art Nr 90 31 034 Bedienungs und Installationsanleitung v Installation and Operation Instruction cr o Tragbares Gasaufbereitungssystem TGAK 3 BU RHHER Portable gas conditioning system TGAK 3 TECHNOLOGIES 1 Introduction The trouble free operation of analysers depends to a very high degree on the individual operation condi tions Sample gas contains quite often moisture and particles which make the operation of the analyser unreliable or even can damage the sensor particularly if the analysis is carried out at different sites The system has to be conditioned for the analyzing process The TGAK 3 provides a portable complete sample conditioning system at reasonable dimensions and weight which can be carried around to the individual sample points The specifications in the attached data sheet must be followed especially those concerning the specific purpose used materials and the range for pressure and temperature This manual describes all models of TGAK3 If there are specific features for a certain model they are de scribed in the manual The specific TGAK model is indicated on the type plate as well as the order number and part number 2 Important Advice Please check before installation of
2. and Operation Instruction 4 o Tragbares Gasaufbereitungssystem TGAK 3 EX UU TLISIES Portable gas conditioning system TGAK 3 TECHNOLOGIES 7 2 Replacing the electrical main fuse gt Disconnect the main power supply The main fuse is below the main power socket The rectangular fuse holder can be removed using a screw driver gt Replace the fuse and insert the fuse holder again The kind of fuse depends on the voltage Assure to have the proper fuse 7 3 Replacement of micro fuse gt Switch off the device and disconnect it from power supply gt Loosen the 4 screws fixing the front panel for this the case should be laid on its back gt Carefully lift the front plate similar to opening a door Caution The display mains socket and hoses are fixed to the front panel If the front panel should be removed completely hosing and connections must be accurately fixed afterwards gt The fuse is placed at the right side on the supply circuit board covered by a plastic cap gt Replace the fuse and refit the cap Regard the supply voltage for choosing the correct fuse see 6 3 gt If the front panel had been removed remount the hoses and electrical connections accurately gt Fix the front panel with the 4 screws gt Reconnect power supply 7 4 Cleaning of the heat exchanger The heat exchanger must be replaced in case of damage or when clogged In the latter case we recom mend to check if the use of a fil
3. order cannot be processed Ger t Serien Nr Device Serial no R cksendegrund Reason for return Ich best tige hiermit dass das oben spezifizierte Ger t ordnungsgem gereinigt und dekontaminiert wurde und keinerlei Gefahren im Umgang mit dem Produkt bestehen herewith declare that the device as specified above has been properly cleaned and decontaminated and that there are no risks present when dealing with the device Ansonsten ist die m gliche Gef hrdung genauer zu beschreiben In other cases please describe the hazards in detail Aggregatzustand bitte ankreuzen Condition of aggregation please check Fl ssig Liquid Fest Solid Pulvrig Powdery Gasf rmig Gaseous Folgende Warnhinweise sind zu beachten bitte ankreuzen The following safety advices must be obeyed please check Explosiv Giftig T dlich Entz ndliche Stoffe Brandf rdernd Explosives Acute toxicity Flammable Oxidizing Komprimierte Gase Gesundheitsgef hrdend Gesundheitssch dlich Umweltgef hrdend Gas under pressure Irritant toxicity Health hazard Environmental hazard Bitte legen Sie ein aktuelles Datenblatt des Gefahrenstoffes bei Please include the current material safety data sheet of the hazardous material Angaben zum Absender Information about the dispatcher Firma Company
4. 3 Aufbauen und AnschlieDBn iie nee 7 3 1 Montag as so oc ee PRSBEDIPEFFPRERCEESETTELE T a SEIEN 7 3 2 Elektrische Anschlisse titt tette aetate tu etit edente a Etro xvn ua nut aav de RER RR Ru 7 20 e 8 4 1 Geratevariante CSP 12201 addet teen ce br e Da een 8 4 2 Ger tevariante CSP 12202 uuunsensnsnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnannnnnnnnnnnnnnnnnnnnnnnnnnnannnennnnnnnnnanann 8 4 3 K ndens tanfall 2 22 ee Le Eee nr ee 9 4 4 Bedienung der Men funktionen 4444444ennnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn mann ann naar nananman 9 4 4 1 bersicht der Men f hrung c cccccsccscccccsessssesesesececssecsesesesesessesesesesececesesseseseseseseseeeneseees 10 4 4 2 Ausf hrliche Erkl rung des Bedienungsprinzips seee nn 11 4 5 Beschreibung der Men funktionen sees eene eene enn nn nennen 11 4 5 1 az iTeinipiH 11 4 5 2 Untermen Peltierk hler Anzeige P5 sssssseeeneeeeene nennen nene 12 5 Instandsetzung Entsorgung e eeeeeeeeeeeeeeeeeennnne nenne nenne nnnm nnmnnn nnmnnn 13 5 1 Fehlerbehebutig eine ee ee ne 13 5 2 iir eas 13 6 Wartung eec ase occ 13 6 1 Warnhinweise m a AAA A AE A 13 6 2 W aU aE E NER 14 EMEN 15 7 1 Fehlersuc
5. Disconnect input and output hoses Take care of the safety instructions Loosen but don t remove the centre knurled head screw Push screw downwards Remove cover Push terminals sideward VV ON ON ON Replace hose and step backwards the above steps for mounting 7 6 Spare parts To order spare parts please indicate type of cooler and serial number see type plate For accessories and enhancement see data sheets and or catalogue The following parts are recommended for stocking Spare part Part no Spare filter elements for types CSP1xxxx CSP4xxxx package with 5 pieces 41 15 00 10 Spare filter elements for type CSP2xxxx package with 8 pieces 41 15 10 50 Condensate vessel type GL4 CS PX 0005 Cock for GL4 in types CSP1x1xx CSP4x1xx 90 08 982 Cock for GL4 in type CSP2x1xx 90 26 503 Quick coupling male DN4 6 for types CSP1xxxx CSP4xxxx 90 08 434 Quick coupling male DN4 6 for type CSP2xxxx 90 08 436 Spare tube 90 terminal for peristaltic pump 0 3 l h only if pump is installed 91 24 030027 Electrical fuse for cooler 230V 5x20 mm 1 25A slow 91 10 00 00 58 115V 5x 20 mm 2 5A siow 91 10 00 00 13 Electrical main fuse 230V 5x20 mm 1 6Aslow 91 10 00 00 10 115V 5x20mm 2 5Aslow 91 10 00 00 12 Connecting cable 2 5m 91 44 05 00 09 7 7 Attached documents Data sheet TGAK 3 DE DA 45 00 08 Gas flow chart CS P01 10 4 Certificate of conformity KX 45
6. a gas cooler with condensate vessel a gas pump filter moisture detector and a flow meter with needle valve A version without the pump is also available The TGAK features a unique fail safe interlock which inhibits the operation of the sample pump until the cooler has reached it s operational temperature of 5 C Additional features include a moisture sensor which also inhibits sample pump operation in the event moisture is detected downstream of the cooler The sample then passes through a fine particulate filter to ensure analyzer cleanliness All wetted parts are available in a variety of materials for corrosive non corrosiveand reactive applications Condensate removal is accomplished with either a condensate vessel a manual drain valve or an integrated peristaltic pump Additional options or design variations are available upon request Please contact one of our application specialists for further information Outlet dew point and alarm threshold adjustable Improved cooling capacity TEN B hler Technologies GmbH D 40880 Ratingen Harkortstr 29 S e DE 45 0008 Tel 49 0 2102 4989 0 Fax 49 0 2102 4989 20 a 08 2007 Internet www buehler technologies com e mail analyse buehler technologies com Technical Data Ambient temperature Outlet dew point Alarm threshold Dew point stability statically over whole range Inlet dew point max Inlet gas temperature Gas flow standard version acid r
7. control continues during setting mode gt You can navigate through the main menu using the keys according to Fig 1 gt To accept the menu item press and the related submenu is activated gt Now the parameters may be set To change the parameters scroll the submenu using the keys and confirm the selected menu item with gt The values can be changes within their limits using the keys t 4 Pressing the enter key stores the set value Afterwards the device returns to the submenu automatically gt Wait for a few seconds without pressing any key to return to the submenu without saving the val ues The same procedure holds for the sub and main menu If you forget to quit the menu the system returns automatically to display mode In this case the preset values are kept instead of being reset Note As soon as the values are saved by pressing the enter key they are accepted for regulation gt Quit the main menu or the submenu by selecting the menu item E Exit 4 5 Description of the menu functions 4 5 1 Main menu Peltier cooler PKE 5xx Peltier cooler This item allows all relevant settings for the Peltier cooler In the corresponding p 6 gt OK submenu nominal temperature and alarm limits may be selected Global settings toP settings Top settings Selection of the global temperature unit either degree Celsius C or degree Fahrenheit F toP Note This men
8. ob sich dies in Zukunft durch den Einsatz eines Demontage und Reinigung des W rmeaustauschers Filters vermeiden l sst gt VON ON Y v VV VV ON ON ON V Beachten Sie die im Kapitel Wartung genannten Warnhinweise Gaszufuhr abstellen Ger t ausschalten und Netzstecker ziehen Die 4 Schrauben der Frontplatte l sen der Koffer sollte dabei liegen Die Frontplatte vorsichtig wie eine T r nach links kippen Achtung Anzeige Netzstecker und Schl uche sind an der Frontplatte befestigt Verschlauchungen Str mungsmesser und Verdrahtungen Netzstecker Anzeige l sen und Frontplatte abnehmen Gasverbindungen und Kondensatablauf des W rmetauschers l sen W rmetauscher nach oben herausziehen K hlnest Loch im K hlblock reinigen W rmetauscher sp len bis alle Verunreinigungen beseitigt sind W rmetauscher an der gek hlten Au enfl che mit Silikonfett einschmieren W rmetauscher mit drehender Bewegung in das K hlnest wieder einschieben Gasverbindung und Kondensatablauf wiederherstellen Verschlauchungen und Verdrahtungen anschlie en und Frontplatte befestigen 7 5 Austausch des Schlauches der peristaltischen Pumpe wenn VON ON ON V Pumpe vorhanden Zu und Abf hrungsschlauch an der Pumpe entfernen Sicherheitshinweise beachten Mittlere R ndelschraube l sen aber nicht ganz abdrehen Schraube nach unten klappen Abdeckkappe abziehen Anschl sse seitlich herausziehen und Schlauch entfernen S
9. 00 06 Declaration of Contamination status 32 BXCS3000 02 2012 Art Nr 90 31 034 sample gas Me gas our CHE quick coupling floumeter i Schnellkupplung Str mungs V messer as il E without pump Re Pumpe IN CSP5xxxxx with bulkhead union pump AN CSPSxxxxx mit Schottverschraubung Pumpe NC heat exchanger Warmetauscher moisture detector Feuchtef hler Schnellkupplung 3 09 1 CU all N HL T quick coupling IN sample gas Me gas yf A panel filter Frontplatten filter adapter DurchfLlu adapter condensate Kondensat AUS condensate h pump De N een condensate P vessel Kondensat sammelgef J CSPxx1xx with condensate peristaltic vessel mit Kondensatsammelgef USFxxZxx with PERF TEE pump mit peristaltische Pumpe ALLE RECHTE a8e ohne Toleranzangabe nach ISO 2768 mK Werkstoff Oberfl chenbear beitungszeichen TGAK3 BUHLER zereme Ney CS POT 1074A S Me Regeltechnik eT SI pmurmes Ratingen res Dan fina Datun Nonel Ere rur WWSETTSSWIETSUNE 7 TT ITE TECHNOLOGIES Tragbare Gasaufbereitung TGAK 3 Die Durchf hrung genauer und sicherer Gasanalysen an wechselnden Einsatzorten erfordert ein kleines kompaktes Gasaufbereitungssystem F r diese Anwendung wurde in einem stabilen Koffer ein vollst ndiges Messgasaufbereitungssystem gesch tzt untergebrac
10. 0V 2 with condensate vessel 1 gives part number CSP 2 2 1 00 Accessories Part no Description Material Package CS PX 00012 Removable case trolley with 50mm wheels Aluminum 1 pcs Spare Parts Part no Description Material Package 41 15 00 10 Filter element for CSP1xx00 and CSP4xx00 Fibre glas 5 pcs 41 15 10 50 Filter element for CSP2xx00 PTFE 8 pcs 91 24 030027 Spare tube 90 connection for peristaltic pump Norprene 1 pc we reserve the right to amend specification EG Konformit tserkl rung gt _ EC declaration of conformity 4277 Lay TECHNOLOGIES Hiermit erkl ren wir dass die nachfolgenden Produkte den wesentlichen Anforderungen der folgenden EG Richtlinie in ihrer aktuellen Fassung entsprechen Herewith we declare that the following products correspond to the essential requirements of the following EC directive in its actual version 2006 95 EG Niederspannungsrichtlinie low voltage directive Folgende weitere Richtlinien wurden ber cksichtigt the following directives were regarded 2004 108 EG EMV EMC Produkte products Tragbares Gasaufbereitungssystem Portable gas conditioning system Typ en type s TGAK3 Zur Beurteilung der Konformit t wurden folgende harmonisierte Normen in aktueller Fassung herangezogen The following harmonized standards in actual revision have been used e EN 61010 1 Sicherheitsbestimmungen f r elektrische Mess Steuer Regel und Laborger te Teil 1 Allgem
11. 89 0 Fax 49 0 2102 4989 20 a 08 2007 Internet www buehler technologies com e mail analyse buehler technologies com Druck rel max au er CSP4xxxx max Technische Daten Umgebungstemperatur 5 bis 40 C Nennk hlleistung bei 25 C einstellbar 2 20 C einstellbar 3 1K und 1 7K um Taupunkt Gasausgangstaupunkt Warnschwellen Netzanschluss oder a E Re A Kaltger testecker Anschlussl nge 1 Leistungsaufnahme Eingangstaupunkt max 60 C Gaseingangstemperatur 110 C 18056 Betriebsbereitschaft ca 10 max 110 I h Abma e ohne Leitung HxBxT ca 10 max 110 I h Durchfluss Standardversion Saurefeste Version Gewicht Standardausf hrung is Messgasanschl sse Schnellkupplung Kondensatablass bei Kondensatsammelgef bei peristaltischer Pumpe 90 kJ h 230 V 50Hz 115 V 60Hz 2 5m max 250 VA nach max 10 min ca 510x355x270 ca 16 kg Ablasshahn Schnellkupplung Unter Ber cksichtigung der zur Verf gung stehenden Gesamtk hlleistung Siehe dazu die Leistungsdaten des PKE 521 Hierzu steht auch unserer Berechnungsprogramm oder die Beratung durch unseren Vertrieb zur Verf gung 180 C Gaseingangstemperatur nur bei der CSP5XX00 Versionen Schottverschraubung im Gaseingang Materialien medienber hrende Teile Glas PPH Viton EPDM Edelstahl rostfrei PA 12 PC Glasfaser Standardversion Filterelement PVDF Glas Teflon Edelstahl rostfrei PTFE Kal
12. Anschrift Address Ansprechpartner Contact person Abteilung Division E Mail Tel Phone Fax Unterschrift Ort Datum Stempel Location date Signature Stamp Buhler Technologies GmbH D 40880 Ratingen Harkortstr 29 Tel 49 0 2102 4989 0 Fax 49 0 2102 4989 20 e mail service buehler technologies com Internet www buehler technologies com
13. Bedienungs und Installationsanleitung ww Installation and Operation Instruction 4277 71 ERX Tragbares Gasaufbereitungssystem TGAK 3 Portable gas conditioning system TGAK 3 Lesen Sie die Bedienungsanleitung vor dem Gebrauch des Ger tes gr ndlich durch insbesondere die Hinweise unter Gliederungspunkt 2 Andernfalls k nnten Gesundheits oder Sachsch den auf treten Die B hler Technologies GmbH haftet nicht bei eigenm chtigen nderungen des Ger tes oder f r unsachgem en Gebrauch Read this instruction carefully prior to installation and or use Pay attention particularly to all advis es and safety instructions to prevent injuries B hler Technologies GmbH can not be held respon sible for misusing the product or unreliable function due to unauthorised modifications BXCS3000 02 2012 Art Nr 90 31 034 B hler Technologies GmbH Harkortstr 29 D 40880 Ratingen ERS Tel 49 0 21 02 49 89 0 Fax 49 0 21 02 49 89 20 Q Internet www buehler technologies com Email analyse buehler technologies com 1 Bedienungs und Installationsanleitung Installation and Operation Instruction ww Tragbares Gasaufbereitungssystem TGAK 3 BUHLER Portable gas conditioning system TGAK 3 TECHNOLOGIES Inhaltsverzeichnis Seite 1 Einlelung ce nu 4 2 Wichtige HiNnWeiS xv sss ccc rece 200000000 nn anna KERN nn ahnen anne 4 2 1 Allgemeine Gefahrenhinweise sse ener nennen sente en renr nnns nnns nennen ns 5
14. Tragbares Gasaufbereitungssystem TGAK 3 BU RHHER Portable gas conditioning system TGAK 3 TECHNOLOGIES 6 Maintenance 6 1 Indication of risk Installation and maintenance of the device shall be performed by trained staff only familiar with the safety requirements and risks The device must be operated within its specifications All repairs must be carried out by B hler authorised personnel only Only perform modifications servicing or mounting described in this manual Only use original spare parts Regard all relevant safety regulations and internal operating instructions during maintenance DANGER Electrical voltage Electrocution hazard Before any manipulation on the device disconnect the electrical equipment from mains power supply Make sure that the electrical equipment cannot be reconnected during repair or maintenance The wiring must be done by trained staff only Regard the correct mains voltage DANGER Toxic corrosive gases Sample gas may be harmful Before maintenance turn off the gas supply and make sure that it cannot be turned on unintentionally Please exhaust sample gas to a safe place Protect yourself against toxic corrosive gas during maintenance Use gloves respirator and face protector under certain circumstances Condensate may be corrosive or toxic Injury of skin and eyes hazard of poisoning Wear gloves respirator and face protection if necessary Regard all accide
15. as conditioning system TGAK 3 TECHNOLOGIES 4 3 Kondensatanfall Kondensat kann tzend oder giftig sein Verletzungen der Haut und Augen Vergiftungsgefahr Tragen Sie ggf Schutzhandschuhe Atem und Gesichtsschutz Beachten Sie die Unfallverh tungsvorschriften und gesetzlichen Regelungen bei der Entsorgung W hrend des Betriebs f llt Kondensat an Dieses wird entweder mittels einer peristaltischen Pumpe abge f hrt oder in einem Kondensatsammelgef gesammelt Bei der peristaltischen Pumpe ist auf ein sicheres Auffangen des Kondensats zu achten Es ist erforderlich den F llstand des Kondensatsammelgef es zu kontrollieren und es von Zeit zu Zeit zu entleeren Dazu den Ablasshahn ffnen und Kondensat entleeren 4 4 Bedienung der Men funktionen Kurzerkl rung des Bedienungsprinzips Benutzen Sie diese Kurzerkl rung nur wenn Sie bereits Erfahrung im Bedienen des im TGAK eingebauten Peltierk hlers besitzen Eine ausf hrliche Erkl rung erhalten Sie unter 4 4 2 und Kapitel 4 5 Tasten Die Bedienung erfolgt mit nur 3 Tasten Sie haben folgende Funktionen Taste Funktionen Wechsel von der Messwertanzeige ins Hauptmen Auswahl des angezeigten Men punktes Annahme eines editierten Wertes oder einer Auswahl Wechsel zum oberen Men punkt Erh hen der Zahl beim ndern eines Wertes oder Wechseln der Auswahl tempor rer Wechsel zur alternativen Messwertanzeige wenn Option vorhanden We
16. ature for effi cient operation Keep the ventilation opening on the right side of the unit clear and free Therefore the space on the right side must be at least 10cm 4 3 1 Mounting Use the quick coupling and an appropriate hose for the connection of the TGAK 3 with the sample point and fit them tight Connect the analyser to the TGAK 3 with an similar hose and keep this hose short and straight The inlets are marked IN for the sample input and OUT for the output The types CSP5xxxx do have a stainless steel fitting instead of a quick coupling It is suitable for gas input temperatures up to 180 C 356 F The quick couplings contain stop valves Therefore the TGAK 3 is only allowed to operate when the quick couplings are both locked in The door of the TGAK 3 must be kept open or be removed when the device is in operating conditions The door can be removed to the top after opening 3 2 Electrical connecting Check the operation voltage of the unit and plug the cable into the socket on the front plate The power supply has to be 230V AC 50 Hz with the CSPx2xxx type and 115V AC 60 Hz with the CSPx1xxx types The minimal external fusing of the power line has to be 4A Wrong mains voltage may damage the device Regard the correct mains voltage as given on the type plate Damage to the device in case of insulation testing Do not proceed insulation tests with high voltage to the device as a whole The device
17. chlauch wechseln und zur Montage obige Schritte in umgekehrter Reihenfolge durchf hren BXCS3000 02 2012 Art Nr 90 31 034 17 TECHNOLOGIES Bedienungs und Installationsanleitung Ww Installation and Operation Instruction Er o Tragbares Gasaufbereitungssystem TGAK 3 BUHEER Portable gas conditioning system TGAK 3 TECHNOLOGIES 7 6 Ersatzteile und Zusatzteile Bei Ersatzteilbestellungen bitten wir Sie Ger tetyp und Seriennummer anzugeben siehe Typenschild Bauteile f r Nachr stung und Erweiterung finden Sie im angeh ngten Datenblatt und in unserem Katalog Die folgenden Ersatzteile sollten vorgehalten werden Ersatzteil Artikel Nr Ersatzfilterelemente f r CSP1xxxx CSP4xxxx 5 St ck pro Packung 41 15 00 10 Ersatzfilterelemente f r CSP2xxxx 8 St ck pro Packung 41 15 10 50 Kondensatsammelgef B GL4 CS PX 0005 Kugelhahn f r GL4 im CSP1x1xx CSP4x1xx 90 08 982 Kugelhahn f r GL4 im CSP2x1xx 90 26 503 Schnellschlusskupplung Vaterteil DN4 6 f r CSP1xxxx CSP4xxxx 90 08 434 Schnellschlusskupplung Vaterteil DN4 6 f r CSP2xxxx 90 08 436 Ersatzschlauch 90 Anschluss f r peristaltische Pumpe 0 3 l h 91 24 03 00 27 nur falls Pumpe eingebaut Feinsicherung Messgask hler 230V 5x20mm 1 25A tr ge 91 10 00 00 58 115V 5x20mm 2 5A tr ge 91 10 00 00 13 Feinsicherung Kaltger testecker 230V 5 x 20mm 1 6A tr ge 91 10 00 00 10 115V 5 x 20mm 2 5A tr ge 91 10 00 00 12
18. chsel zum unteren Men punkt Erniedrigen der Zahl beim ndern eines Wertes oder Wechseln der Auswahl VV VIV V VIV VY tempor rer Wechsel zur alternativen Messwertanzeige wenn Option vorhanden BXCS3000 02 2012 Art Nr 90 31 034 9 Bedienungs und Installationsanleitung Installation and Operation Instruction Tragbares Gasaufbereitungssystem TGAK 3 Portable gas conditioning system TGAK 3 vw u XTTVTLISIET TECHNOLOGIES 4 4 1 bersicht der Men f hrung Anzeige xxxx Hauptmen Anzeige der Temperatur und des Betriebszustandes aktuelle Temperatur Untermen Peltierk hler Anzeige P5__ Einstellbereich 2 C 20 C Anzeige XXxx 1 C 7C Anzeige xxx 1 C 3C Anzeige xxx Solltemperatur Anzeige tEMP 1 afo Alarmhysterese Anzeige A Hi i Alarmhysterese Anzeige A Lo as Exit Untermen Anzeige E Globale Einstellungen Anzeige toP ra of Celsius Fahrenheit Anzeige C F Exit Hauptmen Anzeige E Abb 1 Men bersicht Angezeigt wird die K hlertemperatur in 0 5 C Schritten Durch bet tigen der Enter Taste gelangt man ins Hauptmen Die Temperatur kann wahlweise in Celsius oder in Fahrenheit dargestellt werden Wert einstellen speichert den Wert warte 4s kein speichern Einstellung der K hlerte
19. cht ist Der Statuskontakt ist in der Stellung Alarm Wird der Temperaturbereich erreicht wird die Temperatur dauerhaft angezeigt und der Statuskontakt schaltet um Weiterhin ist bei Inbetriebnahme des TGAK 3 der Feuchtealarm geschaltet rote Warn LED Moisture Um diesen Alarm zur ckzusetzen muss man die RESET Taste dr cken Wenn die Betriebstemperatur des K hlers erreicht ist und kein Feuchtealarm mehr anliegt schaltet sich die eingebaute Messgaspumpe automatisch ein Beim Modell CSP4xxxx ohne Messgaspumpe ist darauf zu achten dass vor Aufnahme der Messung alle Warn LED s aus sind und der TGAK 3 seinen Betriebszustand erreicht hat da es sonst zu Feuchtedurch br chen kommen kann Blinkt w hrend des Betriebes die Temperaturanzeige im Display auf der Frontplatte ist der eingebaute Messgask hler berlastet siehe auch Gliederungspunkt 6 1 Fehlersuche und behebung Die Messgas pumpe wird jetzt automatisch abgeschaltet um den Analysator zu sch tzen Pendelt sich der K hler wieder ein h rt die Temperaturanzeige auf zu blinken und die Messgaspumpe schaltet sich wieder ein Der TGAK 3 ist mit einem Feuchtef hler ausger stet Dieser meldet den Durchbruch von Kondensat falls der K hler defekt oder berlastet ist Unter einem Durchbruch in der Frontplatte sind eine rote LED sowie eine Reset Taste sichtbar Die LED zeigt den Feuchtedurchbruch an Mit der Reset Taste wird das Signal wieder gel scht Sollte es sich u
20. e dew point to high gas temperature to high check limiting parameters install pre cooler Condensate in gas outlet and reduced gas flow no power check power supply quick coupling not locked properly check quick coupling gas path clogged demount and clean heat exchang er and or hoses condensate outlet clogged by ice send cooler for inspection Red LED Moisture is ON Humidity inside the system check performance data install pre cooler dry hoses if necessary drain condensate vessel Display blinks with High Temperature operational temperature not yet reached wait for 20 minutes maximum cooling capacity to low even though cooler is running make sure that air can circulate free and that ventilation louvers are not obstructed the door must be opened gas flow dew point gas tem perature to high check application parameters install pre separator fan broken check fan replace if necessary Low temperature controller defect send cooler for inspection Display of an error E LE i Error 01 short circuit Temperature sensor defect send cooler for repair E r dc Error 02 interruption Temperature sensor defect send cooler for repair 30 BXCS3000 02 2012 Art Nr 90 31 034 Bedienungs und Installationsanleitung ww Installation
21. ectrical connecting uuo tote te etre edet tetto la 22 Pa 0 21 17 11 0 em M 23 4 1 Model CSP 12201 EE 23 4 2 Model ESP ro 23 4 3 Disposal of Condensate ursnssnnnannssnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnannnnnnnnnnnnnnnnnn 24 4 4 Operation of menu functions nennen nennen nrnr en ren tensis sn trnn nnns 24 4 4 1 Overview of the menu items sssssssssesssseeeee eene nnne nnne nennen rennen 25 4 4 2 Detailed description of the operational principle eenn ernennen 26 4 5 Description of the menu functions eesssssssseeeeeeeenenn eene nnn ener nennen nns 26 4 5 1 MIN IGPU eregasten aidaa idaan iae eade iddaa daniati oordaad idana iaaa 26 4 5 2 Submenu Peltier cooler Display P5__ 240r4nn0snsnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn 27 5 Repair angdDispOsal se tren Extr Rx er FE Ee npe el d ip E a E aeu d 28 5 1 arg 28 5 2 Disposal T aeaatate 28 6 Maintenance tme 29 6 1 Indieation ot risk quce ET 29 6 2 Maintenance ERR 29 LEE oec eno none ee eon ere ee eee 30 7 1 Riso 30 7 2 Replacing the electrical main fuse srra iniaa a a aAA RE a 31 7 3 Replacement of micro fuse u2s0urs444s0nnsnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn
22. eine Anforderungen e EN 61326 1 Elektrische Mess Steuer Regel und Laborger te EMV Anforderungen Teil 1 Allgemeine Anforderungen Dokumentationsverantwortlicher f r diese Konformit tserkl rung ist der Unterzeichnende mit Anschrift am Firmensitz The person authorised to compile the technical file is the one that has signed and is located at the company s address Die CE Kennzeichnung wurde angebracht im Jahr The device was CE labelled in 04 Ratingen den 07 10 2010 Stefan Eschweiler Gesch ftsf hrer Managing Director B hler Technologies GmbH Harkortstr 29 D 40880 Ratingen KX 45 0006 Tel 49 0 21 02 49 89 0 Fax 49 0 21 02 49 89 20 Internet www buehler technologies com Dekontaminierungserkl rung IM Declaration of Contamination status rua TECHNOLOGIES Giltig ab valid since 2011 05 01 Revision 0 ersetzt Rev replaces Rev Die gesetzlichen Vorschriften schreiben vor dass Sie uns die Dekontaminierungserklarung ausgef llt und unterschrieben zur ck zu senden haben Die Angaben dienen zum Schutz unserer Mitarbeiter Bringen Sie die Bescheinigung an der Verpackung an Ansonsten ist eine Bearbeitung Ihres Reparaturauftrages nicht m glich Legal regulations prescribe that you have to fill in and sign the Declaration of Contamination status and send it back This information is used to protect our employees Please attach the declaration to the packing Otherwise your repair
23. elow the nominal temperature LO Note Default value at delivery is 3 C 5 F unless otherwise agreed Exit submenu Exit E Selecting this item returns to the main menu BXCS3000 02 2012 Art Nr 90 31 034 27 Bedienungs und Installationsanleitung Installation and Operation Instruction b d Tragbares Gasaufbereitungssystem TGAK 3 EXUTUTLISTY Portable gas conditioning system TGAK 3 TECHNOLOGIES 5 Repair and Disposal 5 1 Repair If the device shows irregularities see chapter 6 1 for troubleshooting If you need help or more information call 49 0 2102 498955 or your local agent Please hold the data of the type plate ready If the device doesn t work correctly after elimination of failures and turning power on the device must be checked by the manufacturer Please ship the device with suitable packing to B hler Technologies GmbH Service Harkortstra e 29 40880 Ratingen Germany In Addition attach the filled in and signed Declaration of Decontamination status to the packing Otherwise your repair order cannot be processed The form can be requested by e mail to service buehler technologies com 5 2 Disposal The device contains no coolants Regard the local regulations for disposal of electric and electronic equip ment 28 BXCS3000 02 2012 Art Nr 90 31 034 Bedienungs und Installationsanleitung Ww Installation and Operation Instruction Er o
24. eration of menu functions Overview of the operational principal Use this short description if you have experience with the build in Peltier cooler You will find detailed description in chapter 4 4 2 and chapter 4 5 Keys Operation is carried out by only the keys with the following functions Key Function Switch from measurement display to main menu Selection of the display menu item Accepting the changed value or selection Switch to the upper menu item Increase of the value of switching the selection Temporary display of the alternative measurement display if option is installed Switch to lower menu item Decrease of the value of switching the selection VV VIV V VIV V WV Temporary display of the alternative measurement display if option is installed 24 BXCS3000 02 2012 Art Nr 90 31 034 Bedienungs und Installationsanleitung Installation and Operation Instruction Tragbares Gasaufbereitungssystem TGAK 3 Portable gas conditioning system TGAK 3 vw u XTTVTLISIET TECHNOLOGIES 4 4 1 Overview of the menu items Display of current temperature and operating state Current Temperature Display xxxx Main Menu Submenu Peltier Gas Cooler Display P5__ Setting Range Set Temperature 2 C 20 C Display tEMP Display xxxx a upper alarm threshold 1 C 7C Dis
25. esistant version 5 40 C adjustable 2 20 C adjustable 3 1K and 1 7K with respect to the dew point 0 1K 1 5K 60 C 110 C 180 C appr 10 max 110 In appr 10 max 110 In Cooling capacity at 25 C Power supply Power connection Power consumption Ready for operation after Dimensions without pipe HxWxD Weight of standard version Condensate draining 90 kJ h 230 V 50Hz or 115 V 60Hz 2 5m cable with connector max 250 VA max 10 min appr 510x355x270 appr 16 kg Gas terminals Rel pressure max CSP4xxxx only manual valve quick coupling Condensate vessel Peristaltic pum quick coupling PUB 0 bar 1 bar Under restriction of the available cooling power See the PKE 521 data sheet for specifications You may as well use our gas cooler selection programme or simply call us 180 C inlet gas temperature only with CSP5XX00 version Gas inlet then with bulkhead union Materials in Gas Path Standard filter element Acid resistant materials filter element Please indicate with order Portable gas conditioning system The part number consists of a code for the available options Tick in desired code at x Part no CSP x x 00 O1 059A x 00 Glass PPH Viton EPDM stainless steel PA 12 PC glass fibre PVDF glass teflon stainless steel PTFE Kalrez sintered PTFE marked positions Desc
26. f r das Alarm relais gesetzt werden Eingestellt wird die Alarmgrenze im Bereich von 1 C H L oO 1 8 F bis 3 C 5 F bezogen auf die eingestellte K hlertemperatur Hinweis Der Standardwert bei Auslieferung betragt 3 C 5 F solange nichts anderes vereinbart ist Exit Untermen Exit t Durch Ausw hlen gelangt man zur ck ins Hauptmen 12 BXCS3000 02 2012 Art Nr 90 31 034 Bedienungs und Installationsanleitung v Installation and Operation Instruction cr o Tragbares Gasaufbereitungssystem TGAK 3 UNTER Portable gas conditioning system TGAK 3 TECHNOLOGIES 5 Instandsetzung Entsorgung 5 1 Fehlerbehebung Sollte ein Fehler beim Betrieb auftreten finden Sie unter Gliederungspunkt 6 1 Hinweise f r die Fehlersu che und Beseitigung Sollten Sie Fragen haben wenden Sie sich bitte an unseren Service Tel 49 0 2102 498955 oder Ihre zust ndige Vertretung Ist nach Beseitigung eventueller St rungen und nach Einschalten der Netzspannung die korrekte Funktion nicht gegeben muss das Ger t durch den Hersteller berpr ft werden Bitte senden Sie das Ger t zu die sem Zweck in geeigneter Verpackung an B hler Technologies GmbH Reparatur Service Harkortstra e 29 40880 Ratingen Deutschland Bringen Sie zus tzlich die Dekontaminierungserkl rung ausgef llt und unterschrieben an der Verpackung an Ansonsten ist eine Bearbeitung Ihres Reparaturauftrage
27. he und Beseitigung eese aa ER EENAA A nnne nns 15 7 2 Auswechseln der Hauptsichrung s ssis bsinnn inniinn naana aaa aa essen A ennt nne nnne nnne nnne enne 16 7 3 Auswechseln der Feinsicherung im K hlermodul PKE521 sees 16 7 4 Demontage und Reinigung des W rmeaustauschers sssseeeeen ene 17 7 5 Austausch des Schlauches der peristaltischen Pumpe wenn Pumpe vorhanden 17 7 6 Ersatzteile und Zu satztelle en ee Be C b quee en 18 7 7 Beiliegende Unterlagen sna R E E E 18 4 4 1 Overview of the menu items u 240us2s0onnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnsnnnnnnnnnnnnnnnnnnnnannn 25 4 4 2 Detailed description of the operational principle sseen 26 4 5 1 Main BIST TETTE 26 4 5 2 Submenu Peltier cooler Display Pb cscceseceeeeceeceeeeeeeaeeeeeeeceeeeceaeeeeaaeseeeeeseaeeesaeeseaeeeeeees 27 2 BXCS3000 02 2012 Art Nr 90 31 034 Bedienungs und Installationsanleitung v Installation and Operation Instruction cr o Tragbares Gasaufbereitungssystem TGAK 3 X UTUTLISIES Portable gas conditioning system TGAK 3 TECHNOLOGIES Contents Page 1 IRWOOUCIION iii eee A A T 19 2 Important AGVICG e 19 2 1 General indication of risk sss eene nennen nnne nnn nenne nnns 19 3 Installalon and Connection en ee 22 3 1 Quee eS 22 3 2 El
28. heit ausgew hlt werden Exit Hauptmen Exit Durch Ausw hlen gelangt man zur ck in den Anzeigemodus BXCS3000 02 2012 Art Nr 90 31 034 11 Bedienungs und Installationsanleitung Installation and Operation Instruction ww Tragbares Gasaufbereitungssystem TGAK 3 BUHLER Portable gas conditioning system TGAK 3 TECHNOLOGIES 4 5 2 Untermen Peltierk hler Anzeige P5__ Peltierk hler gt Solltemperatur temperature Temperatur Diese Einstellung setzt den Sollwert f r die K hlertemperatur Der Wert kann in L E n p einem Bereich von 2 C 35 6 F bis 20 C 68 F gesetzt werden Hinweis Der Standardwert bei Auslieferung betr gt 5 C 41 F solange nichts anderes vereinbart Bei ge nderter Temperatur blinkt die Anzeige gegebenenfalls bis der neue Arbeitsbereich erreicht ist Peltierk hler obere Alarmgrenze Alarm high Alarm high Hier kann der obere Schwellwert f r den optischen Alarm sowie f r das Alarmre x lais gesetzt werden Eingestellt wird die Alarmgrenze im Bereich von 1 C gt 1 8 F bis 7 C 13 F bezogen auf die eingestellte K hlertemperatur A Hi Hinweis Der Standardwert bei Auslieferung betr gt 3 C 5 F solange nichts anderes vereinbart ist Peltierk hler gt untere Alarmgrenze Alarm low Alarm Low Hier kann der untere Schwellwert f r den optischen Alarm sowie
29. ht In seiner Grundausf hrung besteht das System aus einem Gask hler mit Kondensatsammelgef einer Gaspumpe einem Filter einem Feuchtef hler und einem Str mungsmesser mit Nadelventil Eine Version ohne Pumpe istebenfalls erh ltlich Unabh ngig von der Umgebungstemperatur k hlt der Messgask hler das Gas auf den voreingestellten Taupunkt Werkseinstellung 5 C ab Damit wird der Taupunkt sicher unterschritten und Feuchtigkeit im Messgas als Kondensat ausgeschieden Eine Sicherheitsschaltung gibt die Messgaspumpe erst frei wenn der K hler seinen Arbeitspunkt erreicht hat Zum Schutz des Analysators ist dem K hler ein Feuchtef hler nachgeschaltet Dieser stoppt die Messgaspumpe wenn ein Wasserdurchbruch oder eine berlastung des K hlers vorliegen sollte Die Materialauswahl kann erfolgen zwischen einer Standardversion mit Materialien f r wenig korrosive Gase oder einer s urefesten Materialzusammen stellung Dies bezieht sich auch auf das Filterelement F r die Kondensatableitung sind zwei Ausf hrungen w hlbar ein Kondensatsammelgef mit Ablasshahn oder eine eingebaute Schlauchpumpe mit Schnell kupplung Weitere Optionen oder Abwandlungen sind gegebenenfalls lieferbar Bitte nehmen Sie unsere Beratung in Anspruch Ausgangstaupunkt und Alarmschwellen einstellbar verbesserte K hlleistung B hler Technologies GmbH D 40880 Ratingen Harkortstr 29 C e DD 45 0008 Tel 49 0 2102 49
30. is equipped with extensive EMC protection If insulation tests are carried out the electronic filter devices will be damaged All necessary tests have been carried out for all concerned groups of components at the factory test voltage 1 kV or 1 5 kV respectively depending on the device If you wish to carry out the insulation test by yourself please test only separate groups of components Disconnect the control board and if installed the peristaltic pump and the ground connector of the moisture detector controller Now you can perform the electrical strength test 22 BXCS3000 02 2012 Art Nr 90 31 034 Bedienungs und Installationsanleitung v Installation and Operation Instruction A Tragbares Gasaufbereitungssystem TGAK 3 YUP AUER Portable gas conditioning system TGAK 3 TECHNOLOGIES 4 Operation li NOTE The door of the TGAK 3 must be kept open or be removed when the device operating The door can be removed to the top after opening Insert the plug of the power cable into the socket on the front plate after correct tubing of the device After turning the device on the actual cooling block temperature is shown in the display Since the set temperature range with respect to the pre set output dew point of the cooler is not yet reached the display blinks and the status contact is set to Alarm During warm up the moisture alarm is set red LED Moisture in the left upper window Press the RESET button to c
31. lear the alarm As soon as the operating temperature range is reached and the moisture alarm has been reset the display stops blinking and the status contact switches to Normal For model CSP4xxxx without internal pump make sure that all alarm LEDs are off and the cooler has reached its working range before starting the measurement Otherwise a moisture slip may occur If the display starts blinking during operation the cooler is overloaded see also chapter 6 1 Troubleshoot ing To prevent the analyser from getting damaged the sample gas pump is switched off automatically As soon as the working range of the cooler is reached the display stops blinking and the sample gas pump is switched on again The TGAK 3 is equipped with a moisture detector This detector will signal a water slip in case the cooler is overloaded or has a break down The red LED in the upper left display will turn on The detector controller must be reset by pushing the reset button If the reason for the detection was a short term moisture break through the TGAK 3 will turn to operation again If there s a serious water slip the LED will stay lightened In this case the TGAK 3 should be inspected immediately and the reason for the slip has to be found The sensitivity of the moisture detector can be set with the potentiometer in the left upper window Sensitivi ty is reduced by turning counter clockwise The sensitivity is factory preset and should only be cha
32. llation and Operation Instruction Tragbares Gasaufbereitungssystem TGAK 3 vw u XTTVTLISIET Portable gas conditioning system TGAK 3 TECHNOLOGIES 7 Anhang Beachten Sie die Sicherheitshinweise unter Gliederungspunkt 2 und 6 7 1 Fehlersuche und Beseitigung Problem St rung m gliche Ursache Abhilfe Keine Anzeige Netzspannung unterbrochen Netzanschluss vornehmen Sitz des Netzsteckers berpr fen Sicherung defekt Sicherung berpr fen u ggf wech seln Kondensat im Gasausgang KondensatsammelgefaB voll Kondensatsammelgef entleeren K hlleistung zu gering Unbedingt darauf achten dass die L ftungsschlitze nicht verdeckt werden W rmestau und die T r im Betrieb offen ist Zu gro e Durchflussmenge zu hoher Taupunkt Gastemperatur Grenzparameter einhalten Vorab scheider vorsehen Kondensat im Gasausgang bei vermindertem Gas durchsatz Netzspannung unterbrochen Anschluss pr fen Schnellschlusskupplung nicht richtig eingesteckt Sitz der Kupplung pr fen Gaswege verstopft W rmetauscher oder Schl uche demontieren und reinigen Kondensatausgang vereist K hler einsenden Rote Moisture LED am Ausbruch oben links Feuchtigkeit im System zul ssige Daten einhalten Vorab scheider vorsehen ggf Leitungen trocknen Kondensatsammelgef entleeren Anzeige blinkt bei bertemperatur A
33. m einen kurzzeitigen Feuchtedurchbruch handeln wird die LED nach Bet tigen der Reset Taste erl schen Falls eine dauerhafte oder schwerwiegende St rung vorliegt leuchtet die LED weiter Das Ger t muss dann unbedingt berpr ft werden ber ein Potentiometer an der Frontseite kann die Empfindlichkeit des Feuchtef hlers angepasst werden Durch Drehen des Potentiometers entgegen dem Uhrzeigersinn reduzieren Sie die Empfindlichkeit Die Empfindlichkeit ist ab Werk voreingestellt und sollte nur mit entsprechender Sachkenntnis ver ndert wer den 4 1 Ger tevariante CSP 12201 Die Ger tevariante CSP12201 ist mit einem 2 poligem Wippschalter ausgestattet Dieser dient zur Abschal tung der Messgas und peristaltischen Pumpe w hrend der Messgask hler des Systems weiter arbeitet Bei zeitlich auseinander liegenden Messungen an einem Aufstellungsort kann so die Aufw rmphase des Messgask hlers umgangen werden Alle anderen Funktionen sind baugleich mit dem CSP12200 Messgas koffer wie in dieser Anleitung beschrieben 4 2 Ger tevariante CSP 12202 Die Ger tevariante CSP12202 kann mit geschlossener T r betrieben werden Alle anderen Funktionen sind baugleich mit dem CSP12200 Messgaskoffer Bitte beachten Sie die Bedienungsanleitung des TGAK 3 8 BXCS3000 02 2012 Art Nr 90 31 034 Bedienungs und Installationsanleitung ww Installation and Operation Instruction A Tragbares Gasaufbereitungssystem TGAK 3 URILAR Portable g
34. mperatur im Bereich 2 C 20 C Setzen der oberen Alarmgrenze im Bereich 1 0 7 C Setzen der unteren Alarmgrenze im Bereich 1 0 3 C Einstellen der Anzeigen einheit Wahlweise Celsius oder Fahrenheit VD 00 0001 10 BXCS3000 02 2012 Art Nr 90 31 034 Bedienungs und Installationsanleitung v Installation and Operation Instruction cr o Tragbares Gasaufbereitungssystem TGAK 3 XU AUER Portable gas conditioning system TGAK 3 TECHNOLOGIES 4 4 2 Ausf hrliche Erkl rung des Bedienungsprinzips Die ausf hrliche Erkl rung f hrt Sie Schritt f r Schritt durch das Men des Peltierk hlers gt Schlie en Sie den Peltierk hler an die Stromversorgung an und warten Sie die Einschaltprozedur ab Zu Beginn wird f r kurze Zeit die im Ger t implementierte Software Version angezeigt An schlie end geht das Ger t direkt zur Messwertanzeige ber Durch Dr cken der Taste gelangt man vom Anzeigemodus ins Hauptmen Es ist gew hrleis tet dass die Steuerung auch im Men betrieb weiter l uft Man navigiert mit den Tasten gem Abb 1 durch das Hauptmen Best tigt man einen Hauptmen eintrag wird das zugeh rige Untermen aufgerufen Hier k nnen Betriebsparameter eingestellt werden Zum Einstellen der Parameter durchl uft man das Untermen mit den Tasten und best tigt mit den einzustellenden Men punkt Nun k nnen die Werte innerhalb bestimmter Grenzen durch Bet tigen der Tasten ve
35. n sind diese in der Bedienungsanleitung gesondert beschrieben Welches Modell Sie vor sich haben ersehen Sie aus dem Typenschild Auf diesem finden Sie neben der Auftragsnummer auch die Artikelnummer und Typbezeichnung 2 Wichtige Hinweise Bitte berpr fen Sie vor Einbau des Ger tes ob die genannten technischen Daten den Anwendungs parametern entsprechen berpr fen Sie ebenfalls ob alle zum Lieferumfang geh renden Teile vollst ndig vorhanden sind Der Einsatz der Ger te ist nur zul ssig wenn das Produkt unter den in der Bedienungs und Installationsanleitung beschriebenen Bedingungen dem Einsatz gem Typenschild und f r Anwendungen f r die es vorgesehen ist verwendet wird die im Datenblatt und der Anleitung angegebenen Grenzwerte eingehalten werden berwachungsvorrichtungen Schutzvorrichtung korrekt angeschlossen sind die Service und Reparaturarbeiten die nicht in dieser Anleitung beschrieben sind von B hler Techno logies GmbH durchgef hrt werden Originalersatzteile verwendet werden Diese Bedienungsanleitung ist Teil des Betriebsmittels Der Hersteller beh lt sich das Recht vor die Leis tungs die Spezifikations oder die Auslegungsdaten ohne Vorank ndigung zu ndern Bewahren Sie die Anleitung f r den sp teren Gebrauch auf 4 BXCS3000 02 2012 Art Nr 90 31 034 Bedienungs und Installationsanleitung v Installation and Operation Instruction Er o Tragbares Gasaufbereitungs
36. n instructions 20 BXCS3000 02 2012 Art Nr 90 31 034 Bedienungs und Installationsanleitung ww Installation and Operation Instruction Er o Tragbares Gasaufbereitungssystem TGAK 3 XU AUER Portable gas conditioning system TGAK 3 TECHNOLOGIES DANGER Electrical voltage Electrocution hazard Disconnect the device from power supply Make sure that the equipment cannot be reconnected to mains unintentionally The device must be opened by trained staff only DANGER Toxic and corrosive gases Sample gas can be hazardous Take care that the gas is exhausted in a place where no persons are in danger Before maintenance turn off the gas supply and make sure that it cannot be turned on unintentionally Protect yourself during maintenance against toxic corrosive gases Use gloves respirator and face protector under certain circumstances DANGER Explosion hazard if used in hazardous areas The device is not suitable for operation in hazardous areas with potentially explosive atmospheres Do not expose the device to combustible or explosive gas mixtures BXCS3000 02 2012 Art Nr 90 31 034 21 Bedienungs und Installationsanleitung Ww Installation and Operation Instruction 4 o Tragbares Gasaufbereitungssystem TGAK 3 YAULI Portable gas conditioning system TGAK 3 TECHNOLOGIES 3 Installation and Connection The TGAK 3 has to be set on a levelled and safe surface Do not exceed the ambient temper
37. nditioning system TGAK 3 TECHNOLOGIES GEFAHR Giftige tzende Gase Messgas kann gesundheitsgef hrdend sein Schalten Sie vor Beginn der Wartungsarbeiten die Gaszufuhr ab und sp len Sie die Gasleitungen ggf mit Luft Bitte sorgen Sie ggf f r eine sichere Ableitung des Gases Sch tzen Sie sich bei der Wartung vor giftigen tzenden Gasen Tragen Sie die entsprechende Schutzausr stung Kondensat kann tzend oder giftig sein Verletzungen der Haut und Augen Vergiftungsgefahr Tragen Sie ggf Schutzhandschuhe Atem und Gesichtsschutz Beachten Sie die Unfallverh tungsvorschriften und gesetzlichen Regelungen bei der Entsorgung 6 2 Wartung W hrend des Betriebs f llt Kondensat an Dieses wird entweder mittels einer peristaltischen Pumpe abge f hrt oder in einem Kondensatsammelgef gesammelt Bei der peristaltischen Pumpe ist auf ein sicheres Auffangen des Kondensats zu achten Es ist erforderlich den F llstand des Kondensatsammelgef es zu kontrollieren und es von Zeit zu Zeit zu entleeren Dazu wie folgt vorgehen Ablasshahn ffnen und Kondensat entleeren Der Frontplattenfilter neben dem Str mungsmesser sollte ebenfalls monatlich einmal kontrolliert werden und mindestens alle 6 Monate gewechselt werden Dazu die ger ndelte Kappe abschrauben das Element herausnehmen und ggf durch ein neues Element ersetzen 14 BXCS3000 02 2012 Art Nr 90 31 034 Bedienungs und Installationsanleitung Insta
38. nged with appropriate knowledge of the results 4 1 Model CSP 12201 Model CSP 12201 is equipped with a rocker switch to switch off the sample gas pump and the peristaltic pump whereas the cooler remains working This avoids warm up of the cooler if a certain amount of time lies between two readings All other functions are similar to model CSP12200 as described in this operating manual 4 2 Model CSP 12202 This model can be operated with door closed All other functions are similar to model CSP12200 as de scribed in this manual BXCS3000 02 2012 Art Nr 90 31 034 23 Bedienungs und Installationsanleitung Ww Installation and Operation Instruction Am _ Tragbares Gasaufbereitungssystem TGAK 3 RUILE Portable gas conditioning system TGAK 3 TECHNOLOGIES 4 3 Disposal of Condensate Condensate may be corrosive or toxic Injury of skin and eyes hazard of poisoning Wear gloves respirator and face protection if necessary Regard all accident prevention regulations and leagal regulations for disposal During operation condensate will occur and is collected in a condensate vessel or removed by a peristaltic pump depending on the type of TGAG 3 used With peristaltic pumps take care of the safe collection of the condensate If having a condensate vessel version the level of the condensate has to be checked regularly and the ves sel has to be emptied For emptying open the cock at the bottom of the vessel 4 4 Op
39. nnnnnn 31 7 4 Cleaning of the heat exchanger 24044440snnnannnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnannnnnnnnnnnnnnnn 31 7 5 Replacement of the peristaltic pump s hose just in case pump is fitted 32 7 6 Spare PAINS u a einer ne Vou xe ne Pau pn eder esent edel dea ud 32 7 7 as urztelot 10 Ko 0161 B pn1 1a1L ERPPPEEPSFELFFERBFEPPEBEFRELFFEEBFELPERERFELFFEREFEPELBEFRELTFEEBFELPEERSTERFFEREFEPEEREFRELTFELEFEEPERRFFERFFEREFEPEEN 32 BXCS3000 02 2012 Art Nr 90 31 034 3 Bedienungs und Installationsanleitung Installation and Operation Instruction b d Tragbares Gasaufbereitungssystem TGAK 3 RUA i Portable gas conditioning system TGAK 3 TECHNOLOGIES 1 Einleitung Die einwandfreie Funktion eines Analysenger tes wird in erheblichem Ma e von den jeweiligen Betriebsbe dingungen beeinflusst Da h ufig ein Messgas neben der zu analysierenden Gaskomponente gr ere Mengen Feuchtigkeit und Schmutzpartikel enthalten sind muss das Messgas entsprechend aufbereitet werden Das f hrt besonders an st ndig wechselnden Entnahmestellen sehr h ufig zu Problemen Der TGAK 3 dient der qualifizierten Gasaufbereitung an st ndig wechselnden Einsatzorten Bitte beachten Sie die Angaben der Datenbl tter hinsichtlich spezifischen Verwendungszwecks vorhandener Werkstoff kombination sowie Druck und Temperaturgrenzen Diese Bedienungsanleitung beschreibt mehrere Modelle des TGAK3 Sofern f r ein Modell Besonderheiten gelte
40. ns without pipe HxWxD Weight of standard version Condensate draining 85 BTU h 230 V 50Hz or 115 V 60Hz 8 ft 2 5m cable with connector max 250 VA max 10 min appr 20x14x10 6 in appr 35 Ib manual valve quick coupling Condensate vessel Peristaltic pum quick coupling pump 0 psi 14 5 psi Under restriction of the available cooling power See the PKE 521 data sheet for specifications Call one of our application specialists for more information 356 F inlet gas temperature only with CSP5XX00 version Gas inlet then with bulkhead union Materials in Gas Path Standard filter element Acid resistant materials filter element Please indicate with order Portable gas conditioning system The part number consists of a code for the available options Tick in desired code at x Part no CSP x x 00 QPwWDMD gt X 00 Glass PPH Viton EPDM stainless steel PA 12 PC glass fibre PVDF glass teflon stainless steel PTFE Kalrez sintered PTFE marked position Description TGAK 3 Standard version Acid resistant materials N A Without pump 356 F inlet gas temp acid resistant materials Suppl 115V Supply 230 V Condensate vessel with cock for manual draining Peristaltic condensate pump included Without further options case trolley Example portable gas conditioning system TGAK 3 CSP higher resistive materials 2 supply 23
41. nschalten Das Ger t darf nur von instruiertem fachkundigem Personal ge ffnet werden GEFAHR Giftige tzende Gase Messgas kann gesundheitsgef hrdend sein Sorgen Sie ggf f r eine sichere Ableitung des Gases Stellen Sie vor Beginn der Wartungsarbeiten die Gaszufuhr ab und sichern Sie sie gegen unbeabsichtigtes Aufdrehen Sch tzen Sie sich bei der Wartung vor giftigen tzenden Gasen Tragen Sie die entsprechende Schutzausr stung GEFAHR Explosionsgefahr bei Verwendung in Explosionsgef hrdeten Bereichen Das Betriebsmittel ist nicht f r den Einsatz in explosionsgef hrdeten Bereichen geeignet Durch das Ger t d rfen keine z ndf higen oder explosiven Gasgemische geleitet werden 6 BXCS3000 02 2012 Art Nr 90 31 034 Bedienungs und Installationsanleitung Installation and Operation Instruction b d Tragbares Gasaufbereitungssystem TGAK 3 UNTER j Portable gas conditioning system TGAK 3 TECHNOLOGIES 3 Aufbauen und Anschlie en Es ist darauf zu achten dass der TGAK 3 auf einer ebenen sicheren Unterlage steht Beachten Sie auch dass die zul ssige Umgebungstemperatur eingehalten wird Die Konvektion des K hlers darf nicht behin dert werden An der seitlichen L ftungs ffnung muss ausreichend Platz zum n chsten Hindernis sein min destens 10cm 3 1 Montage Mittels der Schnellschlusskupplung und einer geeigneten Schlauchleitung ist die Verbindung zwischen Ent nahmestelle und TGAK 3 herz
42. nt prevention regulations and legal regulations for disposal 6 2 Maintenance During operation condensate will occur and is collected in a condensate vessel or removed by a peristaltic pump depending on the type of TGAG 3 used With peristaltic pumps take care of the safe collection of the condensate If having a condensate vessel version the level of the condensate has to be checked regularly and the ves sel has to be emptied For emptying open the cock at the bottom of the vessel The filter on the front panel next to the flow meter should be checked monthly and is recommended to be replaced at least every six month Unscrew the clear screw on cover counter clockwise take the filter ele ment out and replace it by a new one BXCS3000 02 2012 Art Nr 90 31 034 29 Bedienungs und Installationsanleitung Installation and Operation Instruction Tragbares Gasaufbereitungssystem TGAK 3 vw u XUTVTLISIET Portable gas conditioning system TGAK 3 TECHNOLOGIES 7 Appendices Regard all safety instructions in chapter 2 and 6 7 1 Trouble shooting Problem Failure Possible cause Solution No display no power check power supply fuse blown Check fuse and change it if neces sary Condensate in gas outlet condensate vessel full drain vessel cooler overloaded make sure that air can circulate free and that ventilation louvers are not obstructed the door must be opened flow to larg
43. play A Hi Display xxx afo lower alarm threshold 1 C 3C Display A Lo Display xxx afe Exit Submenu Display E Displayed is the block temperature with a resolution of 0 5 C By pressing the Enter button brings the display to the Main Menu The unit of temperature is adjustable in the menu Global settings Celsius or Fahrenheit adjust value store value wait 4s no storage Adjust the set point 2 C 20 C Set the upper alarm Threshold above setpoint tEMP 1 C 7 C Set the lower alarm Threshold below setpoint tEMP 1 C 3 C Global Settings Celsius Fahrenheit u or e a ET Display toP Display C F eds E Exit Main Menu Display E VE 00 0001 Fig 1 Overview of the menu BXCS3000 02 2012 Art Nr 90 31 034 25 Bedienungs und Installationsanleitung Ww Installation and Operation Instruction cr o Tragbares Gasaufbereitungssystem TGAK 3 YAULI Portable gas conditioning system TGAK 3 TECHNOLOGIES 4 4 2 Detailed description of the operational principle This detailed description leads you through the menu for the cooler step by step gt Connect the cooler to the power supply and wait until the power up sequence has finished First the version of the implemented software is displayed for a short time Then the device switches to the measurement display gt Pressing the key switches from display mode to main menu It is guaranteed that the
44. r Spannungsfestigkeit mit Hochspannung am Gesamtger t durch Das Ger t ist mit umfangreichen EMV SchutzmaBnahmen ausger stet Bei einer Pr fung der Spannungs festigkeit werden elektronische Filterbauteile besch digt Die notwendigen Pr fungen wurden bei allen zu pr fenden Baugruppen werkseitig durchgef hrt Pr fspannung je nach Bauteil 1 kV bzw 1 5 kV Sofern Sie die Spannungsfestigkeit selbst nochmals pr fen wollen f hren Sie diese nur an den entspre chenden Einzelkomponenten durch Klemmen Sie die Regelplatine und ggf die peristaltische Pumpe sowie die Erdung des Beschaltungsger tes f r den Feuchtef hler ab und f hren Sie dann die Spannungsfestigkeitspr fung gegen Erde durch BXCS3000 02 2012 Art Nr 90 31 034 7 Bedienungs und Installationsanleitung Installation and Operation Instruction b ud Tragbares Gasaufbereitungssystem TGAK 3 eX UTI SII Portable gas conditioning system TGAK 3 TECHNOLOGIES 4 Betrieb W hrend des Betriebes muss die T r des Koffers ge ffnet sein um einen W rmestau zu verhindern Sie kann nach dem Aufschwenken nach oben aus den Scharnieren gehoben werden Stecken Sie den Netzstecker in die Buchse auf der Frontplatte ein wenn das Ger t ordnungsgem ver schlaucht ist Nach dem Einschalten des Ger tes sehen Sie die Blocktemperatur Die Anzeige blinkt so lange der eingestellte Temperaturbereich um den voreingestellten Ausgangstaupunkt des K hlers noch nicht errei
45. rbeitspunkt noch nicht erreicht Warten max 20 min K hlleistung zu gering obwohl Unbedingt darauf achten dass der K hler arbeitet L ftungsschlitze nicht verdeckt werden W rmestau und die T r ge ffnet ist Zu gro e Durchflussmenge zu Grenzparameter einhalten Vorab hoher Taupunkt Gastemperatur scheider vorsehen Stillstand des eingebauten Venti berpr fen ggf austauschen lators Untertemperatur Regelung defekt K hler einsenden Fehlermeldungen im Display E Ul Error 01 Kurzschluss Temperaturf hler defekt K hler einsenden E 6 c Error 02 Unterbrechung Temperaturf hler defekt K hler einsenden BXCS3000 02 2012 Art Nr 90 31 034 15 Bedienungs und Installationsanleitung Ww Installation and Operation Instruction 45 F n o Tragbares Gasaufbereitungssystem TGAK 3 URILAR Portable gas conditioning system TGAK 3 TECHNOLOGIES 7 2 Auswechseln der Hauptsicherung gt gt Den TGAK 3 durch L sen der Steckverbindung von der Netzspannung trennen Die Hauptsicherung befindet sich auf der Frontplatte in der Kaltger testeckdose Die rechteckige Sicherungsaufnahme unterhalb der Kontakte kann durch Aufhebeln mit einem geeignetem Werk zeug entfernt werden Sicherung austauschen und Sicherungsaufnahme wieder durch eindr cken einsetzen Beachten Sie die Netzspannung f r die Auswahl der richtigen Sicherung Spannungsversorgung d
46. rez PTFE gesintert S urefeste Version Filterelement Bestellhinweise Tragbare Messgasaufbereitungen Die Artikelnummer wird aus den einzelnen Kenndaten zusammen gesetzt wobei die mit x bezeichneten Stellen f r eine Ziffer stehen die die Ausr stung angibt Art Nr CSP Bezeichnung TGAK 3 Standardversion Materialien s urefeste Version nicht belegt Ohne Messgaspumpe 180 C Gaseingangstemp s urefeste Version 1 Spannung 115 V 2 Spannung 230 V 1 Kondensatsammelbeh lter mit manueller Entleerung 2 Peristaltische Kondensatpumpe eingebaut 00 ohne weitere Optionen x x 00 oakWND 4 X Koffer mit Trolli Beispiel Tragbare Gasaufbereitung TGAK CSP Materialien mit erh hter Anforderung 2 Spannung 230V 2 mit Ausr stung Kondensatsammelgef 1 Art Nr CSP 2 2 1 00 Zubeh r Art Nr Bezeichnung Werkstoff VE CS PX 00012 Abnehmbarer Koffer Trolli mit 50mm Rollen Aluminium 1 St Ersatzteile Art Nr Bezeichnung Werkstoff VE 41 15 00 10 Filterelement f r CSP1xx00 und CSP4xx00 Glasfaser 5 St 41 15 10 50 Filterelement f r CSP2xx00 PTFE 8 St 91 24 030027 Ersatzschlauch 90 Anschlu f r peristaltische Pumpe Norprene 1 St Technische nderungen vorbehalten TT ITE TECHNOLOGIES Portable Gas Conditioning System TGAK 3 The Buhler Model TGAK 3 is a complete portable gas conditioning system in a rugged metal case The basic model consists of
47. ription TGAK 3 Standard version Acid resistant materials N A Without pump 180 C inlet gas temp acid resistant materials Supply 115 V Supply 230 V Condensate vessel with cock for manual draining Peristaltic condensate pump included Without further options case trolley Example portable gas conditioning system TGAK 3 CSP higher resistive materials 2 supply 230V 2 with condensate vessel 71 gives part number CSP 2 2 1 00 Accessories Part no Description Material Package CS PX 00012 Removable case trolley with 50mm wheels Aluminium 1 pcs Spare Parts Part no Description Material Package 41 15 00 10 Filter element for CSP1xx00 and CSP4xx00 Fibre glas 5 pcs 41 15 10 50 Filter element for CSP2xx00 PTFE 8 pcs 91 24 030027 Spare tube 90 connection for peristaltic pump Norprene 1 pc we reserve the right to amend specification BURN TECHNOLOGIES Portable Gas Conditioning System TGAK 3 The Buhler Model TGAK 3 is a complete portable gas conditioning system in a rugged metal case The basic model consists of a gas cooler with condensate vessel a gas pump filter moisture detector and a flow meter with needle valve A version without the pump is also available The TGAK features a unique fail safe interlock which inhibits the operation of the sample pump until the cooler has reached it s operational temperature of 41 F Additional features include a moisture sensor which also inhibit
48. rstellt werden Best tigt man die Einstellung mit wird der eingestellte Wert vom System gespeichert Im Anschluss gelangt man automatisch zur ck ins Untermen Da ein manuelles R ckspringen aus den einstellbaren Bereichen nicht vorgesehen ist kann einfach einige Sekunden gewartet werden Das System wechselt dann zur ck ins Untermen hnlich verh lt es sich mit dem Unter oder auch mit dem Hauptmen Falls vergessen wird das Men regul r zu verlassen wechselt das System selbstst ndig zur ck in den Anzeigemodus ohne die Werte zu speichern Dabei werden hier allerdings die zuvor gespeicherten Parameter beibehal ten und nicht wieder zur ckgesetzt Hinweis Sobald Werte mit der Enter Taste gespeichert wer den werden diese f r die Reglung bernommen Verlassen des Haupt bzw Untermen s erfolgt durch Auswahl des Men punktes E Exit 4 5 Beschreibung der Men funktionen 4 5 1 Hauptmen Peltierk hler PKE 5xx Peltierk hler Von hier aus gelangt man zu allen relevanten Einstellm glichkeiten des Peltier p G k hlers Im zugeh rigen Untermen k nnen Solltemperatur und die Alarm schwellen ausgew hlt werden Globale Einstellung ToP Settings Top Settings Auswahl der globalen Temperatureinheit Wahlweise Grad Celsius C oder Grad Fahrenheit F toP Hinweis Zu diesem Hauptmen punkt gibt es keinen Untermen punkt Es kann von hier aus direkt die Temperaturein
49. rungen vor und vermeiden Sie dadurch Personen und Sachsch den Der f r die Anlage Verantwortliche muss sicherstellen dass Sicherheitshinweise und Betriebsanleitungen verf gbar sind und eingehalten werden Unfallverh tungsvorschriften der Berufsgenossenschaften beachtet werden in Deutschland Allgemeine Vorschriften VBG 1 und Elektrische Anlagen und Betriebsmittel VBG 4 die zul ssigen Daten und Einsatzbedingungen eingehalten werden Schutzeinrichtungen verwendet werden und vorgeschriebene Wartungsarbeiten durchgef hrt werden beider Entsorgung die gesetzlichen Regelungen beachtet werden Wartung Reparatur Reparaturen an den Betriebsmitteln d rfen nur von B hler autorisiertem Personal ausgef hrt werden Nur Umbau Wartungs oder Montagearbeiten ausf hren die in dieser Bedienungs und Installationsan leitung beschrieben sind Nur Original Ersatzteile verwenden Bei Durchf hrung von Wartungsarbeiten jeglicher Art m ssen die relevanten Sicherheits und Betriebsbe stimmungen beachtet werden BXCS3000 02 2012 Art Nr 90 31 034 5 Bedienungs und Installationsanleitung ww Installation and Operation Instruction Er o Tragbares Gasaufbereitungssystem TGAK 3 UAU Portable gas conditioning system TGAK 3 TECHNOLOGIES GEFAHR Elektrische Spannung Gefahr eines elektrischen Schlages Trennen Sie das Ger t bei allen Arbeiten vom Netz Sichern Sie das Ger t gegen unbeabsichtigtes Wiederei
50. s nicht m glich Das Formular kann per E Mail angefordert werden service buehler technologies com 5 2 Entsorgen Das Ger t enth lt kein K hlmittel Bei der Entsorgung sind die gesetzlichen Vorschriften insbesondere f r die Entsorgung von elektronischen Bauteilen zu beachten 6 Wartung 6 1 Warnhinweise Das Ger t darf nur von Fachpersonal installiert und gewartet werden das mit den Sicherheitsanforderun gen und den Risiken vertraut ist Das Ger t darf nicht au erhalb seiner Spezifikationen betrieben werden Reparaturen an den Betriebsmitteln d rfen nur von B hler autorisiertem Personal ausgef hrt werden F hren Sie nur Umbau Wartungs oder Montagearbeiten aus die in dieser Bedienungs und Installati onsanleitung beschrieben sind Verwenden Sie nur Original Ersatzteile Beachten Sie bei der Durchf hrung von Wartungsarbeiten jeglicher Art die relevanten Sicherheits und Betriebsbestimmungen GEFAHR Elektrische Spannung Gefahr eines elektrischen Schlages Trennen Sie das Ger t bei allen Wartungsarbeiten vom Netz Sichern Sie das Ger t gegen unbeabsichtigtes Wiedereinschalten Der Anschluss darf nur von geschultem Fachpersonal vorgenommen werden Achten Sie auf die korrekte Spannungsversorgung BXCS3000 02 2012 Art Nr 90 31 034 13 Bedienungs und Installationsanleitung ww Installation and Operation Instruction 4 o Tragbares Gasaufbereitungssystem TGAK 3 XU AUER Portable gas co
51. s sample pump operation in the event moisture is detected downstream of the cooler The sample then passes through a fine particulate filter to ensure analyzer cleanliness TEN Buhler Technologies LLC DA 45 0008 08 2007 1030 West Hamlin Road Rochester Hills MI 48309 P AN Phone 248 652 1546 Fax 248 652 1598 e 5 All wetted parts are available in a variety of materials for corrosive non corrosiveand reactive applications Condensate removal is accomplished with either a condensate vessel a manual drain valve or an integrated peristaltic pump Additional options or design variations are available upon request Please contact one of our application specialists for further information Outletdew point and alarm threshold adjustable Improved cooling capacity QU Internet www buhlertech com e mail sales buhlertech com Technical Data Ambient temperature Outlet dew point Alarm threshold Dew point stability static over whole range Inlet dew point max Inlet gas temperature Gas flow standard version acid resistant version Gas terminals Rel pressure max CSP4xxxx only 1 41 104 F adjustable 35 6 68 F adjustable 3 1K and 1 7K with respect to the dew point 0 1K 1 5K 140 F 230 F 356 F appr 10 max 110 In appr 10 max 110 In Cooling capacity at 77 F Power supply Power connection Power consumption Ready for operation after Dimensio
52. saufbereitungssystem TGAK 3 YUP AUER Portable gas conditioning system TGAK 3 TECHNOLOGIES disconnect from mains Warning against hazard ous situation Warning against possible explosive atmospheres Warning against electri AN Warning against hot sur wear respirator cal voltage face P Warning against respira tion of toxic gases wear face protection Warning against acid and corrosive substanc es wear gloves ppp 9098 Installation of the device shall be performed by trained staff only familiar with the safety requirements and risks Check all relevant safety regulations and technical indications for the specific installation place Prevent failures and protect persons against injuries and the device against damage The person responsible for the system must secure that safety and operation instructions are accessible and followed local safety regulations and standards are obeyed performance data and installation specifications are regarded safety devices are installed and recommended maintenance is performed national regulations for disposal of electrical equipment are obeyed Maintenance and repair Repairs on the device must be carried out by B hler authorized persons only Only perform modifications maintenance or mounting described in this manual Only use original spare parts During maintenance regard all safety regulations and internal operatio
53. system TGAK 3 EX UT TLISIES Portable gas conditioning system TGAK 3 TECHNOLOGIES 2 1 Allgemeine Gefahrenhinweise Begriffsbestimmungen f r Warnhinweise Signalwort f r wichtige Information zum Produkt auf die im besonderen Ma e aufmerksam gemacht werden soll Signalwort zur Kennzeichnung einer Gef hrdung mit geringem Risiko die zu einem Sachschaden oder leichten bis mittelschweren K rperverletzungen f hren kann wenn sie nicht vermieden wird Signalwort zur Kennzeichnung einer Gef hrdung mit mittlerem Risiko die m glicherweise Tod oder schwere K rperverletzungen zur Folge hat wenn sie nicht vermieden wird Signalwort zur Kennzeichnung einer Gef hrdung mit hohem Risiko die unmittelbar Tod oder schwere K rperverletzung zur Folge hat wenn sie nicht vermieden wird Warnung vor einer all gemeinen Gefahr Warnung vor explosions gef hrdeten Bereichen Netzstecker ziehen ov Spannung Oberfl che Atemschutz tragen Warnung vor elektrischer AN Warnung vor hei er Warnung vor dem Ein atmen giftiger Gase Gesichtsschutz tragen Warnung vor tzenden Fl ssigkeiten Handschuhe tragen Qoo Das Ger t darf nur von Fachpersonal installiert werden das mit den Sicherheitsanforderungen und den Risiken vertraut ist Beachten Sie unbedingt die f r den Einbauort relevanten Sicherheitsvorschriften und allgemein g ltigen Regeln der Technik Beugen Sie St
54. ter will avoid repeating of such clogging gt Regard warnings mentioned in the chapter Operation and maintenance Turn off gas supply Switch off the device and pull mains connector Loosen the 4 screws fixing the front panel for this the case should be laid down VV ON Y Carefully lift the front plate similar to opening a door Caution The display mains socket and hos es are fixed to the front panel v Disconnect the hoses flow meter and electrical connections display mains socket and remove the front panel Loosen gas fittings and condensate drain fitting of the heat exchanger Pull out the heat exchanger upwards out of the cooler Clean the cooling nest hole in cooling block Rinse the heat exchanger until all impurities are disposed Grease the heat exchanger on the cooling surfaces with silicone grease Introduce the heat exchanger with turning movement into cooling nest Re establish fittings for gas tubes and condensate drain VON VV ON ON ON Y Reconnect the hoses to the flow meter and electrical connections display mains socket and fix the front panel v Reconnect power supply BXCS3000 02 2012 Art Nr 90 31 034 31 Bedienungs und Installationsanleitung Ww Installation and Operation Instruction Er o Tragbares Gasaufbereitungssystem TGAK 3 BUHLER Portable gas conditioning system TGAK 3 TECHNOLOGIES 7 5 Replacement of the peristaltic pump s hose just in case pump is fitted
55. the device that the technical data matches the application parameters Also check that the delivery is complete Operation of the device is only valid if the product is used under the conditions described in the installation and operation instruction the in tended application according to the type plate and the intended use the performance limits given in the datasheets and in the installation and operation instruction are obeyed monitoring devices and safety devices are installed properly service and repair is carried out by B hler Technologies GmbH unless described in this manual only original spare parts are used This manual is part of the equipment The manufacturer keeps the right to modify specifications without advanced notice Keep this manual for later use 2 1 General indication of risk Definitions for warnings NOTE Signal word for important information to the product Signal word for a hazardous situation with low risk resulting in damaged to the device or the property or minor or medium injuries if not avoided Signal word for a hazardous situation with medium risk possibly resulting in severe injuries or death if not avoided Signal word for an imminent danger with high risk resulting in severe injuries or death if not avoided BXCS3000 02 2012 Art Nr 90 31 034 19 Bedienungs und Installationsanleitung v Installation and Operation Instruction cr o Tragbares Ga
56. u item has no sub item The temperature unit is directly selected Exit main menu Exit Selecting this item returns to the display mode E 26 BXCS3000 02 2012 Art Nr 90 31 034 Bedienungs und Installationsanleitung v Installation and Operation Instruction Ex o Tragbares Gasaufbereitungssystem TGAK 3 YAULI Portable gas conditioning system TGAK 3 TECHNOLOGIES 4 5 2 Submenu Peltier cooler Display P5__ Peltier cooler gt Nominal temperature Temperature This item allows setting of the nominal temperature for the cooler The value can L E n p be set within a range from 2 C 35 6 F to 20 C 68 F Note Default value at delivery is 5 C 41 F unless otherwise agreed Peltier cooler gt upper alarm threshold Alarm high Alarm high This item allows setting of the upper alarm threshold for the optical alarm as well as for the alarm relay The upper alarm threshold may be set in the range from A H 1 C 1 8 F to 7 C 13 F above the nominal temperature Note Default value at delivery is 3 C 5 F unless otherwise agreed Peltier cooler gt lower alarm threshold Alarm low Alarm low This item allows setting of the upper alarm threshold for the optical alarm as well r as for the alarm relay The upper alarm threshold may be set in the range from A 1 C 1 8 F to 3 C 5 F b
57. urch Aufstecken der Steckverbindung wieder herstellen 7 3 Auswechseln der Feinsicherung im K hlermodul PKE521 gt Ger t ausschalten und durch L sen der Steckverbindung von der Netzspannung trennen Die 4 Schrauben der Frontplatte l sen der Koffer sollte dabei lie gen Die Frontplatte vorsichtig wie eine T r nach links kippen Achtung Anzeige Netzstecker und Schl uche sind an der Frontplatte befes tigt Falls die Frontplatte ganz abgenommen werden soll m ssen al le Verschlauchungen und Verdrahtungen anschlie end sorgf ltig wieder verbunden werden Die Sicherung befindet sich auf der rechten Seite auf der Netzplati ne unter einer Plastikkappe Sicherung austauschen und Kappe wieder Aufdr cken Beachten Sie die Netzspannung f r die Auswahl der richtigen Sicherung s 6 3 Falls die Frontplatte ganz abgenommen wurde alle Verschlauchun gen und Verdrahtungen sorgf ltig wieder anschlie en Frontplatte mit den 4 Schrauben befestigen Spannungsversorgung durch Aufstecken der Steckverbindung wieder herstellen 16 BXCS3000 02 2012 Art Nr 90 31 034 Bedienungs und Installationsanleitung Installation and Operation Instruction Tragbares Gasaufbereitungssystem TGAK 3 Portable gas conditioning system TGAK 3 vw u XTVTLISIET 7 4 W rmetauscher m ssen nur ausgetauscht oder gewartet werden wenn sie verstopft oder besch digt sind Sollten sie sich zugesetzt haben empfehlen wir zu pr fen
58. ustellen Auf dichte Anschl sse ist zu achten Der Gasausgang wird mit dem Analysator ber Schnellschlusskupplung und einer geeigneten Schlauchleitung verbunden Die Schlauchlei tungen sollten m glichst kurz und knickfrei sein Die Gasanschl sse sind mit Messgas IN f r Gaseingang und Messgas OUT f r Gasausgang gekenn zeichnet Bei der CSP5xxxx Ausf hrung ist der Gaseingang mit einer Edelstahl Schottverschraubung ausgestattet Dieses erm glicht eine Gaseinganstemperatur von max 180 C Die Schnellschlusskupplungen sind mit Ventilen ausgestattet Beim Entkoppeln schlie en sich die Ventile automatisch Der Messgaskoffer darf nur bei eingekuppelten Anschl ssen in Betrieb genommen werden Der Messgaskoffer darf nur mit ge ffneter T r betrieben werden Die T r des Koffers kann nach dem Aufschwenken nach oben aus den Scharnieren gehoben werden 3 2 Elektrische Anschl sse Der TGAK 3 verf gt ber ein Netzkabel welches am Ger t in einen Kaltger testecker eingesteckt werden muss Die Spannungsversorgung betr gt 230V AC 50Hz bei den CSPx2xxx Ausf hrungen und 115V AC 60 Hz bei den CSPx1xxx Ausf hrungen Das Ger t kann an jede Schukosteckdose mit mindes tens 4 A Absicherung angeschlossen werden Falsche Netzspannung kann das Ger t zerst ren Bei Anschluss auf die richtige Netzspannung gem Typenschild achten Besch digung des Ger tes bei Durchf hrung der Isolationspr fung F hren Sie keine Pr fung de