Home
        Service Manual - Infiniti Medical
         Contents
1.               Objective Action Result  Change option short press TOOT    to Start  wes i   ight Detector 1     Contrast Cal  Off  CAA Start  No action  required  Please Wait  17 Minutes  No action  required Autozero  Please Wait  E     Prepare for Cal    Connect the a fimin  Check FilterLine to AA  the calibration UU     gas   Connect Gas  Cal  Check    Stop    Tak      Check the Press the gas as Famin  2  measured valve for 15 5 0  values shownin   seconds until U       Vol  in the  EtCO2 digital  display        the readings  stabilize                       Connect G  s    Cal  Check    Stop       Tak op           Calibration is not required if the measured value is the  same as the concentration of the calibration gas             0 83  Vol    e g  Concentration of calibration gas used is    5   CO2   the measured value should be between 4 7    5 3   therefore  calibration is not required           To reset and  return to  measuring  mode   automatically  erasing event  messages        long press    nal       Reset  Please wait               Section 10  Performance Verification    Table 10 1  CO2 Calibration Check  Continued        Objective Action Result       To exit service Turn off the  mode and save monitor   event messages                If the reading is outside these limits  calibration is  required  Refer to Table 10 2 for the Calibration Process  procedure              10 5    Section 10  Performance Verification    Calibration Process    If calibration is required
2.   10 4  13 1    Handheld Capnograph               csesccececeeeeeeeeeeeeeeeeeteeeeeeeees 1 2  LCD Screen and 7 Segment Digital Display                2  1 2  Removing the Battery Pack            ceeecseeceeeeesteeeeeeenteeeeeees 7 2  Removing Mounting Screws             cccccccseeeeeesseeeeeneeeeees 7 3  Separating Rear and Front CoverS             c scceceessseeeeeeees 7 3  Replacing CO   Board and Housing Components               7 4  Replacing the PUM eieiei rirani a RRE 7 5  Fow Syste M aranana aida Gdns eae 7 6  Toh E aE E E E E errr errr recente cee eee ereteeer ar 7 7  EXM ust  LING desia eeetialtetheecdeaeee nde diet 7 7  Lines to Solenoid  Zero line  Input line 1  Input line 2         7 8  LEONE eka a a hen Siete feel 7 9  Replacing LCD and 7 Segment Display             0  c ee 7 10  Replacing EPROM  esiliu irae erea dnia eriu abAus 7 12  Leak Test Jig Connection           eeeececeteeeeeeeeeeeteeeeeteeeeeneees 10 1  Initialization SCrON         cccceceeeeeeeeeceeeeeeeeeeeeeeeeeeeaeeteeeeeees 10 2  Measuring Mode asesan iri renerne ae RaRa ans 10 2  Quick G  de Syra eairt ntaa aA ETATEN i 10 3  Connecting the monitor to Printer PC with   Communication Adapter Kit    10 9    Packing the Handheld Capnograph in Original Packing   11 2    Accessing Service Mode and Changing Parameter    SEUINGSE vias A T T 4 3  Accessing Service Level 2            cccccccceeeseseeeeeeeeseneeeesaees 4 5  Changing Default Alarm Limits             c   ccseeeeceeseeeeeeeeees 4 7  Changing Def
3.   2 way electro pneumatic valve controlling the  directional flow either from the patient or from the  CO   Scrubber  during Autozero      Lithium compound CO  absorber     
4.   SECTION 2  REQUIRED EQUIPMENT  2 1 Required Equipment       2 1 REQUIRED EQUIPMENT    You will need the following equipment to disassemble  replace parts   check  adjust  or calibrate the monitor     Note  For ordering spare parts  contact your local distributor   For Disassembly and Reasssembly    Phillips head screwdriver  medium    IC Extractor for PLCC socket   Threadlocker glue e g  Loctite   222 super screw lock  For Calibration  Checks and Adjustments    Calibration Gas  5  CO2 in air    FilterLine or FilterLine H and a connecting means   Leak Test Jig    Vacuum manometer with a range of 0 500 mBar  e g  MPL 3200 DM  Digital Manometer MicroPneumatic Logic  Inc     Communication Adapter Kit  only necessary to verify print and  PC functions     Seiko DPU 414 printer  only necessary to verify print functions   External power supply  Handheld Capnograph Operator   s Manual    Flow Meter e g  Top Track Mass Flow Meter   model  822 13 OV1 PV1 V1  Sierra Instruments  Inc     Warning  Observe ESD  electrostatic discharge  precautions    when handling  adjusting or performing any procedure with the  monitor internal components        2 1       SECTION 3  CLEANING  3 1 Cleaning       3 1 CLEANING    Always clean the outer surface of the monitor before servicing     Warning  Always wear latex or surgical gloves when cleaning    or servicing the Handheld Capnograph monitor after hospital  use     Warning  Do not use cleaning materials based on  phenols        Warning  Do not all
5.   perform the Calibration Process  after  performing Calibration Check  as shown in Table 10 2    Table 10 2  Calibration Process             Objective Action Result  To start No action oe Ainin        2  Calibration required  T A  Process after Yl io                   calibration check     Connect Gas  Cal  Check    Stop  amp      Cab op        To activate long press  Calibration CO  Gas    Process  e Calibration     Gop   al  gt               To adjust the short press  EA value to A gp  e gas C  P mibration on  concentration you   Stop    Col    gt    are using   To start Press the gas  calibration  valve and Calibration  long press in process  kx           amp     Continue    pressing the    z  1 Calibration  538 Ya ve in process  until the  Disconnect    Gas appears        Monitor performs   Disconnect Gas  an internal from the Calibration  calculation  FilterLine  in process       O       Calibration No action  succeeded  required  Calibration    O K       Return g                       10 6    Section 10  Performance Verification    Table 10 2  Calibration Process  Continued        Objective    Action    Result       Return to  calibration check  mode     long press    Cal  Check    Connect Gas  Kal   Stop E    Cal                      To reset and long press   return to   measuring mode Reset  Please wait    To exit service Turn off the   mode and save monitor     event messages                   Calibration Errors    Table 10 3 describes calibration problems that can 
6.  CO2 board           Beeper does not  beep     One or more of the  Alarm Silence options  has been turned OFF     Turn the respective  alarm ON  Refer to  the Handheld  Capnograph  Operator   s manual           A CO2 component has  failed        Replace with a new  CO2 board        Table 5 2  Advisory Message             Message Probable Cause Corrective Action  Check Unit   A CO2 board component   Check Event  has failed  Messages screen in          Service Mode        5 4          Section 5  Troubleshooting    Table 5 3  Event Messages       Event Message    Probable Cause    Corrective Action       CO2 board      Component Failed      Replace CO2 board          Flow         Loose connections      Kinks or twists in  tubing      Flow system blocked      Pump Malfunction      A CO2 board  component has failed         Inspect all tubing  ends for loose  connections and  ensure tight  connections      Straighten kinks  and undo any twists  in the tubing       Check flow rate and  if necessary change  Flow System  refer  to Section 6  Flow  Calibration Check      Replace pump      Replace with a new  CO2 board          5 5       SECTION 6  FLOW CALIBRATION CHECK    6 1 Introduction  6 2 Flow Rate Check  6 3 Flow Calibration Process       6 1 INTRODUCTION    Perform a Flow Calibration Check after replacing the Flow System and or  Pump  Flow Calibration Check includes Flow Rate Check and Flow  Calibration Process     6 2 FLOW RATE CHECK    To perform the Flow Rate Check  follow t
7.  SPECIFICATIONS    12 5 1 External Power Source  12V DC Medical Grade Adapter    12 5 2 Internal Power Source    Ni MH Rechargeable Battery Pack 7 2V 2 1 A h  intended for continuous  operation        Operating Time   fully charged     Between 3 and 6 hours  depending on power  management These values reflect the performance  of a new battery  age and usage will degrade  capacity        Recharging  Period    Approximately 4 5 hours internal recharging          Charger Type       Internal          12 6 COMPONENTS AND USER INTERFACE    12 6 1  Displays       Graphic LCD  display     128 x 64 dots  with LED backlight dimension  75 mm x 53 mm        Two numeric  fields    3 digits each  using 7 segment LED dimension  22mm x 14mm           Two color alarm  bar       yellow  red          12 6 2 Controls and Indicators          Front Panel       ON OFF switch  Alarm Silence button  Contrast   Value change button  Event Home button  Next   Menu button           12 6 3 Connections       Front Panel    CO2 Input connector       Rear Panel    Clamp connector          Side Panel       Power Supply Communication Adapter Port  Gas  output connector             12 3       SECTION 13  SPARE PARTS    13 1    Spare Parts List       13 1 SPARE PARTS LIST    Spare parts are shown below in Table 13 1  For spare part numbers  contact your local distributor     Table 13 1  Spare Parts List       Item    Description       CO2 board   includes electronic board  sensor  pump  and scrubber  solenoid  L
8.  screen the password  long press  Pump volt   simultaneously Flow Cal   aa e  To change the short press  parameter setting or YON  value C  Pump volt   Flow Cal   To move to the next short press  parameter  and save he  the changed setting  Pump volt   Flow Cal                          4 9       SECTION 5  TROUBLESHOOTING    5 1 Introduction   5 2 Who Should Perform Repairs  5 3 Repair Level Supported   5 4 How to Use This Section   5 5 Obtaining Replacement Parts  5 6 Troubleshooting Guide       5 7    5 8    5 9    5 10    5 11    INTRODUCTION    This section provides information for troubleshooting for the monitor  isolating a failure of the monitor     WHO SHOULD PERFORM REPAIRS    Only qualified service personnel should remove and replace components  of the monitor  If your facility does not have qualified service personnel   contact your local distributor     REPAIR LEVEL SUPPORTED    The monitor has the following replaceable components  CO2 board  Flow  System  Pump Assembly  Housing Components  LCD  7 segment displays  and Software Assembly     The procedures for disassembling and replacing the above mentioned  components are described in Section 7  Disassembly Guide     HOW TO USE THIS SECTION    Failures of the Housing Components  front cover  rear cover  gas outlet   ON OFF button and keypads  are determined by visually inspecting these  components for cracks or deformations  and checking for mechanical  failures     Use the Troubleshooting Guide in Section 5 6 to 
9.  short press      gt           In case you entered a  wrong digit    Password        To restart password  validation process    long press      gt     Enter password      000                To reset and return to  long press   measuring mode    automatically erasing mal     event messages  Reset  Please wait     To exit service mode Turn off the   and save event monitor     messages                Section 4  Service Mode    Default Alarm Limits    The service technician can change the default alarm limits for either  patient mode  Adult or Neonatal      The monitor has the following alarms with adjustable level settings   e EtCOz high and low levels  mmHg   e FiCO2 high level  mmHg     Refer to Table 4 3 for changing the settings of the above mentioned  parameters        Warning  Changing default settings will permanently change the  user   s start up value and should be verified with the responsible  clinical personnel        Table 4 3  Changing Default Alarm Limits             Objective Action Response  To access the Alarm short press  eet Patient Adult   Limits Menu after b EtC0e R60 X     entering last password FicO2 x 8  digit NoResp   30  RR F0 k 3          To change the patient  short press       Patient Neonata    mode SON EtCO2 Fea x       Ficd2 F 8   NoResp   20   RR Rise k 12          To access a new short press    parameter   gt                     FicO2 x 8  NoResp   26  RR F150   12             To change the value short press    VOAN  E        To reset and retur
10. 2    Remove the two CO2 board mounting screws    Remove the CO2 connector screw from the CO2 input connector   Remove the ON OFF button and save for reassembly     Remove the gas outlet from the tubing and save for reassembly     Day SEY R SOPs  ie    Remove the CO2 board and replace with a new CO2 board and  reassemble in reverse order        Warning  The scrubber on the CO2 board is a lithium based  compound  Follow local governing ordinances for disposal              COz Board Mounting Screws    3 7 CO2 Connector Screw     I    CO2 Board          Lower Keypad ON OFF Button    Gas Outlet    Figure 7 4  Replacing CO2 Board and Housing Components       7 4    Section 7  Disassembly Guide    7 4 REPLACING THE PUMP  REFER TO FIGURE 7 5     1   2   3     Note     SO  SN  gt     Note     Open the monitor case described in Section 7 2   Remove the damping cover from the pump     Locate the two tubings  Main line and Exhaust line  connected to  the pump  refer to Figure 7 6     Locate the purple Main line restrictor and the green Exhaust line  restrictor     Disconnect both tubings from their restrictors     Both restrictors must remain with their flow system lines attached  to the CO2 board     Disconnect the pump connector from it   s CO2 board socket   Remove the CO2 board from the front cover  refer to Section 7 3   Unscrew and remove the pump mounting screws    Remove the lower damping and damping holder    Remove the pump and the upper damping     Place the new Pump     Reass
11. CD  7 segment display  flow  system  CO2 input connector  software EPROM and  screws        Software Assembly   includes EPROM chip and last  updated software version        Housing Kit   includes front cover  rear cover   ON OFF button  keypads  gas outlet  screws       7 Segment Digital Display  x2        LCD Assembly   Graphic Display Screen and PC  driver       Calibration Gas  5  COz in air  and FilterLine       Leak Test Jig   vacuum test accessories       Communication Adapter Kit       Battery Pack       Pump Assembly             Flow System       GLOSSARY OF TERMS    BIT   BTPS     Respiration Rate     7 segments LED     Solenoid     Scrubber           Built in Tests     Body Temperature  Pressure and Saturation  Compensation     Where Body Temperature is assumed 37  C with  vapor pressure 47mmHg    The Calculations are made according to   Pcoz   Foo2 x Pb   47   Where     Fcoz is the Fractional concentration of CO  in  Dry gas  Feog     CO   100    Pb   the ambient pressure   Pcog   the partial pressure of COs at BTPS  Erasable Programmable Read Only Memory   Electrostatic Discharge     End Tidal Carbon Dioxide  amount of CO   present  at the end of the exhalation     Fractional Inspired Carbon Dioxide  amount of  CO  present during inhalation     FilterLine Recognition Safeguard   Liquid Crystal Display   Human Machine Interface     Number of respiratory cycles  inhalation and  exhalation  per minute     Graphic display of seven segments of light emitter  diodes   
12. CD and 7 Segment Display                  ee 7 10  7 6 1 Replacing the LCD             ccceeeeeeeeeseeeeteeeeeees 7 10  7 6 2 Replacing the 7 Segment Display                0 7 10       iii    Table of Contents    iv    7 7 Replacing Housing Components                ccccceceeeeeeeteeees 7 11   7 7 1 ON OFF Button and Gas outlet             cee 7 11   7 1 2 Keypads ear E ATER 7 11   7 7 3 FONE COVEN nise aana A 7 11   7 7 4 Rear Covali ria eE TA 7 11   7 8 Updating Software Version          c cccceeeeseessteeeeneeeeees 7 12   Section 8  Electrical Safety T Sts        ccsssscsseesssssessserssssssessserensnsenenes 8 1   8 1 Electrical Safety Tests             ccccceceeseeeesseeeseeeeeeseeeeteneeeeaes 8 1   Section 9  Periodic MainteNanrCe         cccseccccseeesssenccssnensenensesssensenersensees 9 1   9 1 Periodic Maintenance                cceceeeeeeeeeeeeeeeeeeeeeeeteeeeeseeaees 9 1   Section 10  Performance Verification        ccssscccsseerssercsssseesertessseees 10 1   10 1 Flow System Leak Check              ccceecceeeeeeeeeeeeeseteeeeneeeees 10 1   10 1 1   Flow System Leak Check Procedure                10 1   10 2 Performance Verification Procedure              ceceeeeeeeees 10 2   Section 11  Packing For SHIPMEN    1 ssecccsseccsseeensseetessneecsssernsseenens 11 1   11 1 General INStrUCtIONS               cccececcessteeeeeeesneeeeesetaeeeeeseaaes 11 1   11 2 Packing Handheld Capnograph in Original Carton           11 1   11 3 Packing in a Different Carton           cc
13. O2 board number  the number of operating hours  and the  CO2 board   s software version     4 2 2 Event Messages Screen    The monitor automatically performs a series of built in tests  BITs  at  start up and during operation to check sub systems and modules for  errors  The event messages screen lists the last five event messages  the  time each occurred relative to the unit   s operating hours  and a  recommended service action  There are two event messages and each has  a specific response as follows     Event Message  Action   CO2 board Replace COboard  Flow Replace Flow System and or Pump    Refer to Section 5  Troubleshooting       Section 4  Service Mode    Note  To exit the service mode and save event messages  turn off the  monitor  Event messages will be erased when exiting the service    mode by a long press of fi      4 2 3 Interface Test Screen    The interface test screen shows four parameters  listed below  that  provide tools for the user to test  adjust  and calibrate the unit   s  interfacing systems     e HMI Test   e Light Detector   e Contrast Cal   Calibration   e CO2 Calib   Calibration   4 2 3 1 HMI Test    When the HMI Test is turned on  the monitor performs a self test of the  following sub systems        Buzzer  sounds for one second      LCD graphic screen shows a checkerboard pattern  e 7 segment digital display  8s scroll left to right    e Alarm bar  lights red  yellow  off    e Backlight  turns on and off    e Alarm silence indicator  light tur
14. Portable Bedside Capnograph    Service Manual    Cl04185F       TABLE OF CONTENTS    List of Figures       List of Tables  Section 1  INCFOMUCHION   1 1cccccseeensecenncsnencessensesnancessensenensessnnesensesesasanes 1 1  1 1 Manual OVErviEW oerien iana a ia E 1 1  1 2 Warnings  Cautions  and Notes               ccccsssceeesssteeeeeeeees 1 1  1 2 1 Wamigevenenien acid dies 1 1  1 2 2 GAUL i ESEE EA A EESE E A TE TET 1 1  1 2 3 NOtE iesen Oni newer  1 1  1 3 Product DesgriptOnN errire inuri ieee RE 1 2  Section 2  Required EQuiprmenit  u    sccccesecccceeeeesecsesseeeeeeeeeeseeseneeeneneas 2 1  2 1 Required Equipment           ccccccececseceeeeeeneeeeeeenneeeeeeenaeeeeeees 2 1  SOCHON  32 CICANING ai ceieiiesasceceseceeteactiecetceresanczeoneteesdaeeceenerssiertietecesvece 3 1  3 1 GCIGANING 2 st cecies avid Mine ees dees het ieee aaea  3 1  3 1 1 Proceduren ie 3 1  Section 4  Service MOC  a    seccsesncceceeseseseseeeeeeeeeseeeeseneeeneneeeseaeseneeeeneas 4 1  4 1 MOUCHO iii ieii Diana aes Be eee 4 1  4 2 Service Level Wsiiecccedateeacdesentieerii dese R ei eaves 4 1  4 2 1 Information Screen    4 1  4 2 2 Event Messages Screen s    4 1  4 2 3 Interface Test Screen    4 2  4 2 4 Accessing Service Mode and Service Level 1   4 3  4 3 Service Level 2       cccccccceeeseeceeeeeeeneeeeeeeeeeeeeeeeeeeeeseaeeneeeees 4 5  4 3 1 Default Alarm Limits            c cc eeeeeeeeeeeeennneeeeeeeaee 4 7  4 3 2 Default SettingS             ccccceecceeeseteeessteeeeeneeeeees 4 8  4 3 3 Flow Ca
15. Reassemble the new Main line in reverse order        Figure 7 7  Main line    Replacing the Exhaust line  Refer to Figure 7 8   Locate the green restrictor     Carefully disconnect the Exhaust line  at the green restrictor   from the tubing connected to the pump  Leave the green  restrictor in the exhaust line     Reassemble the new Exhaust line in reverse order        Figure 7  8  Exhaust line       Section 7  Disassembly Guide  7 5 2 Replacing the lines to the solenoid  Refer to Figure 7 9     7 5 2 1 Removing lines to the solenoid    1  Carefully disconnect the Zero line from the solenoid        Warning  The scrubber on the Zero line is a lithium based  compound  Follow local governing ordinances for disposal           2  Remove Input line 1 from the solenoid and the other end from  the FilterLine input connector     3  Remove Input line 2 from the solenoid and the other end from  the sensor connector     Zero line       FilterLine Input connector    Figure 7 9  Lines to Solenoid  Zero line  Input line 1  Input line 2     7 8    7 5 2 2    Section 7  Disassembly Guide    Reassembly of the new solenoid lines   Attach the new Input line1 to the upper solenoid connector     Locate the lower solenoid connector which is closest to the  sensor and attach the new Input line 2 to this connector     Bring the Input line 2 tubing over Input line 1  and attach to the  sensor connector     Lead the Input line 1 under the FRS cables  connected to the  FilterLine input connector  not vi
16. anufacturer warranty     Pack the monitor or component s  carefully  If available  use the original  carton and packing materials and follow the instructions in    Packing  monitor in Original Carton     If the original shipping carton and material  are not available  use other suitable shipping materials and container and  follow the instructions in    Packing in a Different Carton        Prior to shipping the monitor or a component  contact your local  representative for a Returned Material Authorization  RMA  number   Mark the shipping carton and any shipping forms with the RMA and the  monitor   s serial number     Caution  Observe ESD  electrostatic discharge  precautions when  packing any monitor components     11 2 PACKING HANDHELD CAPNOGRAPH IN ORIGINAL CARTON    If the original carton and packing material are available  repack the  monitor as follows   Refer to Figure 11 1      1  Place the unit in a plastic bag  Place it into the corresponding  space  in the original foam padding  with the front panel facing up     2  Place the foam padding cover over the top of the unit   3  Seal the carton with packing tape     4  Label the carton with the correct shipping address  return address  and RMA number     11 3 PACKING IN A DIFFERENT CARTON    If the original carton and packing material are not available or if shipping  any monitor component not in its original carton     1  Place the monitor in a plastic bag or component in a plastic anti   static bag   2  Locate a corr
17. ault Settings                 ccecceseeeeeeeeeetteeeeeeeees 4 8  Accessing Flow Calibration          0   c cccccscseseeeeeteeeseseeeees 4 9  Troubleshooting Guide             cccceeeeceeceeeeeeeneeeseeeeeeeneeeesaees 5 2  Advisory Message ieri ririnui knis Tainena AEE 5 4  Event Me SSaQ eS s ecien eaa Erona TAANE 5 5  Flow Calibration aei gara aaa e aeaa ti aaa 6 2  CO2 Calibration CHOCK          cccccssceceeeeeeeeeeeeeeeetetaeeneneees 10 3  Calibration Process atsieti airaa ings 10 6  Calibration Process  Troubleshooting              cccceeeeees 10 7  Accessing Handheld Capnograph Print Functions           10 10  Spare Parts  Listaa ia e e a GR 13 1       SECTION 1  INTRODUCTION    1 1 Manual Overview  1 2 Warnings  Cautions  and Notes  1 3 Product Description       1 1 MANUAL OVERVIEW    This manual contains information for servicing the Handheld  Capnograph  the monitor   Only authorized service personnel should  service this product  The Handheld Capnograph Operator   s Manual is an  integral part of the service procedures  Before servicing the monitor  read  the Operator   s Manual carefully for a thorough understanding of how to  operate the unit     Only use the tools and test equipment as specified in this manual  Only  use original spare parts available from your local distributor     Warning  Incorrect procedures may harm the patient  or damage    the monitor        1 2 WARNINGS  CAUTIONS  AND NOTES  1 2 1 Warning    A warning precedes an action that may result in i
18. cation Adapter Kit  connect the printer to the  monitor  refer to Figure  10 5                       Figure 10 5  Connecting the monitor to Printer PC with    Communication Adapter Kit    Turn the printer ON and select on line mode     10 9    Section 10  Performance Verification    17  Access the monitor print functions as described in Table 10 4 below     Table 10 4  Accessing Handheld Capnograph Print Functions             Objective Action Result  To access the long press  x2   Instrument Setup   gt  aa  menu  first access FicOe OFF  the Alarm Limits Power Momt Normal  menu from any Print Screen    measuring display                    To select Print short press  x4  Coe Scale    C02 Units mmHs  FicOe Ot  Power Mgmt Normal  Print Screen  Be sure the Print short press Tae Seal    COs Scale  option showa VN C02 Units mmHg  Screen  if it doesn t  C  FiC     oft  change to the Screen Power Momt Normal       option Print Screen             To return to long press    Measuring Mode   40  roel SaR  To print long press   simultaneously              printout                    10 10       SECTION 11  PACKING FOR SHIPMENT    11 1 General Instructions  11 2 Packing Handheld Capnograph in Original Carton  11 3 Packing in a Different Carton       11 1 GENERAL INSTRUCTIONS    To ship a monitor or one of its components for any reason  follow the  instructions in this section  Failure to follow the instructions in this  section may result in loss or damage not covered by any applicable  m
19. ccceeeeeeeeeeeeeeeeeeneeee 11 1   Section 12  SPCCIFICATIONS    421 1sseeccceeeceseeeeceeeeeeeseeeseseneeeeeeeeesseneeeeneets 12 1   12 1 PIMYSICAL  aroia E A ERRAT 12 1   T2501      SiZ6tat a ia bee a A a Sa 12 1   12 12 Weigh a aatan a a aa aak iapa 12 1   12 1 3 Noise Emission sdiuir aa a e aana 12 1   12 2 Environmental                 ccccccececeeeeeeeececaecaeeeeeeeeseseeseeneneaees 12 1   12 2 1 Temperature    i a E S 12 1   12 2 2 Relative Humidity   0 0 0    eeeeeeeeeeeeeeeeeeees 12 1   12 2 3 Pressure and Altitude    eeeeeeeeeeeeeeeees 12 1   12 3 Safety Standards  00        ccccccceceeeeeeeeeceeeeeeeeaeeseeeeeeesaeeneneees 12 1   12 4 POITOrMaNnCe wast  aaa tin ee abit aaa eaten 12 2   12 5 Power Specifications           ccccccsceeeeeeeeeeeeeeeeeeeeeeteeeeneeeees 12 3   12 5 1 External Power Source               ecceeeseeeeetteeeeees 12 3   12 5 2 Internal Power SOuICE              ccceeceeeeeeeeeeeeees 12 3   12 6 Components and User interface           0   cccceeeceeeeeeeeeeeeees 12 3   12 6 1 DISplaYyS T EE E 12 3   12 6 2 Controls and Indicators           seseeeseeesee resene ene 12 3   126 3  Connections cecer 12 3   Section 13  Spare Parts  sasien nonae aens anero e ehe daninha denera i taai iania 13 1   13 1 Spare  Parts List Aai na e a 13 1  Glossary of Terms    Table of Contents    LIST OF FIGURES    LIST OF TABLES    1 1  1 2    7 2  7 3  7 4  7 5  7 6  7 7  7 8  7 9  7 10  7 11  7 12  10 1  10 2  10 3  10 4  10 5    4 5    5 2  5 3  6 1  10 1  10 2  10 3
20. defective    Inspect the keypad for    buttons on the  upper or lower  keypad do not  work     cracks  deformations  or other damage   Replace with a new  keypad        A CO2 board component  has failed     Replace with a new  CO2 board        No information  is displayed on  the LCD     The contrast default  value is set too low or  high     Check the contrast  value in the Service  Mode using the  Contrast Cal  option  and correct  accordingly        An LCD component has  failed     Replace with a new  LCD              A CO2 board component   Replace with a new  has failed  CO2 board   LCD Backlight The Power Refer to the Handheld  does not come on   Management option is Capnograph  when the set at LOW  Operator   s manual to  monitor is placed change the Power  in a dark Management option  environment  to NORMAL or  HIGH        An LCD component has  failed     Replace with a new  LCD           A CO2 board component  has failed        Replace with a new  CO2 board              5 3    Section 5  Troubleshooting       Table 5 1  Troubleshooting Guide  Continued           Symptom Probable Cause Corrective Action  7 Segment The Power Refer to the Handheld  Display does not   Management option is Capnograph   light when the set at LOW  Operator   s manual to  unit is change the Power  functioning  Management option    to NORMAL or  HIGH        A 7 Segment display  has failed     Replace with new 7   Segment display        A CO2 board component  has failed     Replace with a new 
21. ed correctly     Note  When operating the monitor for the first time after updating the  software version  the warm up and the self test periods will be  longer than usual              ay  Eo    i    7 oe  A 2 MAREE    LE       EPROM             5 Zp Eai Hi aite   U0    Figure 7 12  Replacing EPROM    7 12       SECTION 8  ELECTRICAL SAFETY TESTS  8 1 Electrical Safety Tests       8 1 ELECTRICAL SAFETY TESTS    Do not return the monitor to the user until the conditions in the following     Warning    statement are met     Warning  After servicing the monitor  always verify product  performance per the procedures in Section 10  Performance    Verification and perform any safety tests required by local  regulatory bodies or mandated by your institution        8 1       SECTION 9  PERIODIC MAINTENANCE    9 1 Periodic Maintenance       9 1 PERIODIC MAINTENANCE    Periodic maintenance is recommended according to operating hours  refer  to section 4 2  Service Level 1       The Pump and Flow System should be replaced every 7 000 operating  hours     For replacing the pump refer to section 7 4  Replacing the Pump      For replacing the flow system refer to section 7 5  Replacing the Flow  System        The monitor should be returned to the manufacturer for periodic  maintenance every 14 000 operating hours     SECTION 10  PERFORMANCE VERIFICATION    10 1 Flow System Leak Check  10 2 Performance Verification Procedure       The performance of the monitor must be verified after all se
22. emble the unit  using the new assembly parts  in reverse  order     Before replacing the pump mounting screws  first dip them into  Loctite   glue or any other threadlocker glue  Carefully twist the  screws into the lower damping  gently pressing against the   CO2 board      Damping cover        Lower damping    Pump connector      Damping holder  Pump mounting screws e  a    ns       Figure 7 5  Replacing the Pump       Section 7  Disassembly Guide  7 5 REPLACING THE FLOW SYSTEM    1  Open the monitor case as described in Section 7 2   The Flow System has 5 lines  refer to Figure 7 6   Two lines go from the pump     Main line    Exhaust line  Three lines go from the solenoid      Zero line    Input Linel      Input Line2               Input line 4     Input line 2  not viewed          green restrictor  Zero line    Main line     a    Purple restrictor    Exhaust line  oO    Figure 7 6  Flow System    7 6    Section 7  Disassembly Guide    7 5 1 Replacing the lines to the pump    7 5 1 1  1   2     Note     7 5 1 2    Replacing the Main line  Refer to Figure 7 7   Locate the purple restrictor     Carefully disconnect the main line from the tubing  connected to the pump  leaving the purple restrictor in the main  line     Remove the tubing from the CO2 sensor     Remove the tubing from the pressure sensor     When replacing the main line  in order to fit the main line on the  board without twists and kinks  coil the tubing around the  pressure sensor  refer to Figure 7 8      
23. ess Service Mode   During self test  Self test occurs and the  press and hold Information Screen  simultaneously appears     and i a    gt   lt  RAE    ven    Self Test       Service Mode  12345  Operating Hours 5 783  Firmware    Z X XX       To access Event long press    Messages    gt                 To access Interface long press  Test Screen  D Light Detector  Contrast Cal   CO  Calib   To change the short press  parameter setting or HMI Test  value YOA Light Detector  Co     Contrast Cal   CO  Calib                          4 3    Section 4  Service Mode       4 4    Table 4 1  Accessing Service Mode and    Changing Parameter Settings  Continued        To move to the next  parameter  moving to  the next parameter  accepts and saves  changed values     short press      gt     HMI Test On  1  Contrast Cal  Off  Co  Calib  Off       To change Contrast  Cal  setting to set    short press  Wm    C    HMI Test  Light Detector     Contrast Cal     CO  Calib           To start Contrast Cal     short press      gt        Contrast Cal     flashes        To Adjust Contrast  Cal  middle point    press  wm     4    until desired  setting is reached    Pressing on the left  lightens the screen   and pressing on the  right darkens the  screen      Screen lightens or  darkens       To accept new setting    short press      gt     HMI Test  Light Detector  Contrast Cal                    ICO  Calib  _  To reset and return to long press  measuring mode   automatically erasing mal   the e
24. ewed in Figure 7 9  and  connect to the FilterLine input tubing connector     Connect the Zeroline to its solenoid connector and place the  scrubber close to the pump under the coiled main line tubing   refer to Figure  7 10     Zero line          Scrubber    NX  N    Figure 7 10  Zero line    7 9    Section ieee en eee  7 6 REPLACING LCD AND 7 SEGMENT DISPLAY  REFER TO FIGURE 7 11   7 6 1 Replacing the LCD   1    Remove the CO2 board as described in 7 3    2  Unscrew and remove the four LCD mounting screws     3  Carefully remove the LCD and replace with a new LCD   7 6 2 Replacing the 7 Segment Display  1  Remove the CO2 board as described in Section 7 3     2  Carefully remove the faulty 7 segment display from its socket and  replace with a new 7 segment display  Note the correct  orientation of the 7 segment display when replacing it     LCD Mounting Screws    LCD               7 Segment Display       Figure 7 11  Replacing LCD and 7 Segment Display    7 7    7 7 1    7 7 2    7 7 3    7 7 4    Section 7  Disassembly Guide  REPLACING HOUSING COMPONENTS    The Housing kit includes the following replaceable components   Front cover  Rear cover  ON OFF switch    Upper Keypad  Alarm silence button and Contrast Value change  button     Lower Keypad  Print Home button and Next Menu button   Gas outlet  Screws    Washers    Note  When reassembling the monitor after replacing any part  be sure to  follow the disassembly steps in reverse order     ON OFF Button and Gas Outlet  Fig
25. he steps below   1  Connect the FilterLine to the Handheld Capnograph   2  Turn on the monitor     3  Connect the other end of the FilterLine to the Flow meter gas  outlet     4  Access the Flow Calibration screen of the Service Mode  refer to  Section 4 3 3  Flow Calibration     5  Check that the Flow Meter reading is 50  5 ml min   at sea level   If the Flow Meter reading is 50  5 ml min     5a Perform Flow Calibration Process  refer to Section 6 3  Flow  Calibration Process     If the Flow Meter reading is not 50  5 ml min     5b Change the pump voltage value until the flow rate displayed  in the flow meter is 50  5 ml min  refer to Section 4  Service  Mode  Table 4 5  Accessing Flow Calibration     5c Perform Flow Rate Check  5d Perform Flow Calibration Process     If you cannot set the flow rate to 50  5 ml min after adjusting the  pump voltage value     5e Change the Flow System  refer to Section 7 5  Replacing the  Flow System     5f Perform Flow Rate Check  5g Perform Flow Calibration Process     If after changing the Flow System you still cannot set the flow rate  display in the flow meter to 50  5 ml min     5h Change the Pump  refer to Section 7 4  Replacing the Pump   51 Perform Flow Rate Check    5j Perform Flow Calibration Process        Section 6  Flow Calibration Check    6 3 FLOW CALIBRATION PROCESS    To perform the Flow Calibration Process  refer to Table 6 1     Table 6 1  Flow Calibration                Objective Action Response  To access Flow Simultaneo
26. isolate failures of the CO2  board  Flow System  Pump Assembly  LCD and 7 segment displays  Once  a failure has been isolated  refer to Section 7  Disassembly Guide for  instructions for removing and replacing a component of the monitor     OBTAINING REPLACEMENT PARTS    Your local distributor provides technical assistance information and  replacement parts  Refer to parts by the part name listed in Section 13   Spare Parts     5 1    Section 5  Troubleshooting       5 12       5 2    TROUBLESHOOTING GUIDE    If you encounter a problem that cannot be resolved through a visual  inspection  refer to Table 5 1 which provides a list of symptoms  probable  causes  and recommended actions to correct the problem  It is  recommended that corrective actions be performed in the order presented   For a symptom that is not listed in Table 5 1  contact your local  distributor     If an Advisory message is displayed on the LCD  refer to Table 5 2 for the  indicated failure and corrective action  Caution messages appear on the  display screen followed by    and Advisory messages are followed by       Check in the Service Mode for any Event Messages and refer to Table 5 3  for the indicated failure and corrective action     Table 5 1  Troubleshooting Guide                Symptom Probable Cause Corrective Action  The unit does not   The battery pack is Check if there is a  turn on when the   missing or is installed battery pack in the  ON OFF button incorrectly  monitor  if not  place  is swi
27. libration         cccececceecceeeeeeeeeeeeeeeeeeneeeeeaees 4 9  Section 5  TTOUDICSHOOLINAG   1 ssssncceccecececeeeceeeeeeceeseseeeeeneneeeseeeeeneeeneneas 5 1  5 1 MOO UCTION EEE E ETEEN A T E E 5 1  5 2 Who Should Perform Repairs             ccccecceeseeeeeteeeeeseeeeseees 5 1  5 3 Repair Level Supported              c ccceceeeeeeeeeeeeeeesecaeeeeneeeees 5 1  5 4 How to Use This Section            cccccececeeeeeeeeeeeeeeeeeteeeeeeeeees 5 1  5 5 Obtaining Replacement Parts             cccccseeeeeeeeesteeeeeeeees 5 1  5 6 Troubleshooting Guide             cceceeecececeeeeeeeeeeeeeeeeeeeeetaees 5 2  Section 6  Flow Calibration Check     1cssscccseeersssecccseeensnersesseennsnersessees 6 1  6 1 INtrOCUCTION   nec  a eee EEEE ANEA AA 6 1  6 2 Flow Rate Check          cccccecceceeeeeseeeceeeeeeseaaeseeeeeeseaeeeeeneeesaas 6 1  6 3 Flow Calibration Process            cccccceccceesceeeeeeeeeeseeeeeneeeteas 6 2  Section 7  Disassembly Guide      11  csecccceecceceeceeeseseseeeeeeeeeeessseneneeeeneas 7 1  7 1 Introduction   eanan ded dude eo ane vate 7 1  7 2  Opening the Handheld Capnograph Case              c eee 7 2  7 3 Replacing the CO2 Board           c  cccceceeeeeeeeeeeeeesseeeteneeetens 7 4  7 4 Replacing the PUMP   asssssssssssrressssrnesnnnnensrnnnensrnnnnnennnnnnnnnn 7 5  7 5 Replacing the Flow System        eseeeesseesseereeerreerreerrenn 7 6  7 5 1 Replacing the lines to the pump    7 7  7 5 2 Replacing the lines to the solenoid                      7 8  7 6 Replacing L
28. n to   long press    measuring mode  Eal Reset  Please wait                To exit service mode Turn off the  and save event monitor   messages                   4 7       Section 4  Service Mode    4 3 2 Default Settings       The service technician can change two default settings  refer to Table 4 4      e 3 Min Alert       BTPS  body temperature  pressure  saturation     When the 3 min Alert is ON  the unit will beep every three minutes to    remind the user that an alarm  or all alarms  are set to OFF     When BTPS is ON  the CO2 value is corrected for this factor   Table 4 4  Changing Default Settings       Objective    Action    Response       To access the Default  Settings Menu  after  accessing the Alarm  Limits Menu     long press      gt     3 Min Alert oii  BTPS On       To change the setting    short press    wm  C    BTPS On       To access BTPS  parameter  and save    short press      gt                 messages             the changed setting  3 Min Alert  BTPS  To reset and return to long press  measuring mode  mal  Reset  Please wait     To exit service mode Turn off the  and save event monitor           Section 4  Service Mode    4 3 3 Flow Calibration    The Flow Calibration screen allows the service technician to perform flow  calibration after flow adjustment and or pump or flow system  replacement  refer to Table 4 5     Table 4 5  Accessing Flow Calibration                   Objective Action Response  To access the Flow At any screen after  Calibration
29. njury or death to the  patient or user  Warnings are boxed and highlighted in boldface type     1 2 2 Caution    A caution precedes an action that may result in damage to  or malfunction  of  the monitor  Cautions are highlighted in boldface type     1 2 3 Note    A note gives information that requires special attention     1 1    Section 1  Introduction       1 3 PRODUCT DESCRIPTION    The device is a Handheld Capnograph that continuously monitors end  tidal carbon dioxide  EtC Oz  and respiratory rate  The monitor is for  attended monitoring only and must be used in the continuous presence of  a qualified healthcare provider  It is intended for use in any environment  where continuous  noninvasive monitoring of these parameters is desired   including hospital and mobile use  when protected from excessive  moisture such as direct rainfall      The monitor is intended for use on adult  pediatric  and infant neonatal  patients                       Figure 1 1  Handheld Capnograph    The monitor is operated using a four key keypad  Measurements are  displayed on the front panel on the 7 segment LED digital displays  LED   and on the Liquid Crystal graphic display  LCD  as shown in Figure 1 2   Refer to the Handheld Capnograph Operator   s Manual for complete  operating instructions     7 segment LED digital displays                         EtCO2   amp   min  m l  4g id  LCD graphic display Bose  40 mmHg    i    Figure 1 2  LCD Screen and 7 Segment Digital Display       1 2     
30. ns on and off     When the HMI test is completed the status field returns automatically to     OFF        4 2 3 2 Light Detector    The Light Detector parameter allows the service technician to set the  threshold value of the phototransistor  The default value is 1 and the  range is from 1 to 5 with 1 as the most sensitive  backlight on at all times   and 5 as the least sensitive  backlight off   The threshold value should be  set according to the lighting conditions of the working enviroment  A  higher threshold value saves on battery power consumption     Note  When the monitor   s power management is set to Low  the backlight  is off        4 2    Section 4  Service Mode    4 2 3 3 Contrast Cal   calibration     The Contrast Cal  gives the service technician the option to change the  middle point default value of the contrast intensity of the LCD     Table 4 1 describes how to change the contrast cal  settings    4 2 3 4 CO2Calib   calibration    The CO2 Calib  allows the service technician to perform a calibration  check and calibration process to ensure the accuracy of the monitor     Section 10  Performance Verification describes the calibration check and  calibration process procedure     4 2 4 Accessing Service Mode and Service Level 1   Table 4 1 describes how to access the service mode  change parameter  settings and how to change Contrast Cal  settings     Table 4 1  Accessing Service Mode and  Changing Parameter Settings          Objective Action Response   To acc
31. occur  the message    displayed and the corrective action   Table 10 3  Calibration Process   Troubleshooting                  Stop       Flestarter              Calibration process  exceeds 3 minutes       Message Possible Causes Action  Gas concentration Check setup   warning  does not match the correct  ee  concentration value   probable    Stop      Restarim    selected in the first causes  and  calibration screen  recalibrate  Problems with gas unit as  sampling above   Flow system  problems  purging   blockage or no  FilterLine   CO2 Sensor problem   Contact  Calibration your local  Failed distributor      Restart  If  the  problem  persists   contact  your local  distributor           12  Disconnect the monitor from the AC mains supply  Verify the unit  continues to work from its battery pack and the battery icon appears       10 7    Section 10  Performance Verification    13     on the lower right corner of the display     Reconnect the monitor to mains power with the AC adapter  Verify  that the unit functions  and the plug icon appears     Note  If the battery pack has a low charge level  the battery icon will    appear instead of the plug  The battery icon will fill indicating the  battery charging process     10 8    Section 10  Performance Verification    To verify printing functions     14     15     16        Adjust the printer to the following communication settings   Input Serial    Data Length 8 bits  Parity Settings None  Baud 9600 bps    Using the Communi
32. omponents     Use the following procedures to disassemble the monitor and replace  parts as needed  Reassemble the monitor in reverse order  The  manufacturer recommends that you follow the disassembly procedure in  the order presented     Note  After replacing the Flow System and or Pump  perform a flow  calibration check as described in Section 6  Flow Calibration Check     Note  After reassembling any part of the monitor  be sure to complete the  performance verification as described in Section 10  Performance  Verification     7 1    Section 7  Disassembly Guide  7 2 OPENING THE HANDHELD CAPNOGRAPH CASE    1  Be sure the monitor is disconnected from mains power and remove  the battery pack from the monitor  push the release button at the    bottom of the monitor and pull out the battery pack as shown in  Figure 7 1              1  Press  S      V    Figure 7 1  Removing the Battery Pack    7 2    Section 7  Disassembly Guide    2  Place the monitor on a clean  dry nonabrasive surface with the back  of the monitor facing up     3  Remove the two bottom and two back mounting screws from the  rear panel of the monitor as shown in Figure 7 2     Back Mounting Screws       Bottom Mounting Screws    Q    Figure 7 2  Removing Mounting Screws    4  Carefully lift the rear case cover up as shown in Figure 7 3           Figure 7 3  Separating Rear and Front Covers    7 3    Section 7  Disassembly Guide  7 3 REPLACING THE CO2 BOARD    Open the monitor case as described in Section 7 
33. ow liquids to drip inside the housing     Warning  Do not clean internal components     halogen producing compounds  strong organic acids  or oxygen  producing compounds        3 1 1 Procedure  1  Unplug the unit from AC mains supply     2  With a damp cloth  gently wipe the outer case of the monitor with  a disinfectant based on aldheyde  alcohol or quaternary  ammonium producing compounds     3 1       SECTION 4  SERVICE MODE    4 1 Introduction  4 2 Service Level 1  4 3 Service Level 2       4 1 INTRODUCTION    The monitor supports a service mode built into the instrument   s firmware   The architecture of the monitor   s service mode divides the service mode  into two levels as described in the sections below       Service Level 1    Service Level 2    To access the service mode  levels 1 and 2   refer to Tables 4 1 and 4 2     4 2 SERVICE LEVEL 1    The Service Level 1 provides information and tools on three service  screens to test and calibrate specific functions without disassembling the  monitor       Information Screen    Event Messages Screen    Interface Test Screen    From these display screens the service technician can check the operating  hours counter and event messages  test the Human Interface functions   HMI   set the light detection threshold  adjust the LCD contrast middle  point and do a calibration check and a calibration process  if necessary   Refer to Table 4 1     4 2 1 Information Screen    The information screen displays the monitor   s serial  C
34. owing procedure     1     2  3   4    Note     Connect the monitor to mains supply using the AC adapter   Connect the FilterLine to the monitor   Turn ON the monitor     The Initialization Screen appears and the       ea  pam    Self Test bar fills  refer to Figure 10 2  Microstream   VHHA  Verify that an audio tone sounds  This  verifies proper operation of the Alarm Figure 10 2   interfacing  Check that the LEDs on the Initialization Screen    alarm bar turn red  yellow and then off  The  7 segment display shows 8s moving from  left to right     After initialization  the monitor  automatically goes to Measuring Mode  The  message CO2 Warmup appears on the Silent  Advisory Message area  refer to   Figure 10 3        Figure 10 3  Measuring  Mode    After CO2 Warmup  verify that the message  changes to Ready     If BTPS is on  then the message will read BTPS ON Ready     Disconnect the FilterLine and verify that the message FilterLine  is displayed on the Silent Advisory Message Area and verify that  the pump stops     Reconnect the FilterLine and verify that the message FilterLine  disappears and the pump is operating     Section 10  Performance Verification    10  Verify that all buttons on the unit work properly  Refer to the  Quick Guide for their applications  see Figure 10 4     Quick Guide          ovv      Power On Short Press Long Press    es Changes Displays   Accesses  Selects Parameters Menus                                                      Home    EaI RANNAR E
35. rase Trend  Changes    YA   values   Contrast Quick Scroll  KA Sound ON   OFF Accesses  Alarm Menu  F   gt  Data Transfer  f   ON   OFF                                     Figure 10 4  Quick Guide    11  Verify the CO2 Calibration Check process functions  Follow the  procedure as described below  When completed  return to the  Measuring Mode screen     CO2 Calibration Check    Caution  Do not check CO  values from the measuring mode if the  BTPS setting is ON  factory default   This mode corrects the CO2  value for BTPS  The calibration check mode disables this  correction     In order to ensure accuracy  a CO2 calibration check needs to be done once  a year  Calibration gas and a FilterLine are needed for this procedure   Start the process from the Interface Test Screen  refer to Table 4 2  as  follows in Table 10 1     Note  Connect the FilterLine to the monitor before starting CO2 Calibration Check   Use the calibration kit  5  gas in air and connecting means  supplied and  approved by Oridion Medical 1987 Ltd     Note  If the CO2 Calibration Check is performed on a battery powered  device  ensure that the batter is fuly chaged prior to calibration    Table 10 1  CO2 Calibration Check          Objective Action Result  Access CO2 short press  x3   Calib HMI Test  mie Light Detector    Contrast Cal     ICO  Calib                          10 3    Section 10  Performance Verification    10 4    Table 10 1  CO2 Calibration Check  Continued                                       
36. rvice procedures  Perform all  procedures in this section to verify the correct functioning of the monitor     10 1 FLOW SYSTEM LEAK CHECK    Perform the Flow System Leak Check after all service procedures  described in this manual     10 1 1 Flow System Leak Check Procedure                1  Connect the vacuum manometer to the Leak Test Jig as shown in  Figure 10 1   Yacuum  To Input Manometer  Connector  Clamp  To Gas  Outlet 4    HI            Syringe    Figure 10 1  Leak Test Jig Connection  Open the Leak Test Jig clamp and empty the syringe   s contents     Connect the open ended tubing  as shown in Figure 10 1  of the  Leak Test Jig to the Gas Outlet of the monitor     4  Connect the other open ended tubing of the Leak Test Jig to the  CO   input connector     5  Pull the syringe valve out until the pressure drops  as displayed on  the vacuum manometer  to approximately  300 mBar  or  equivalent  and close the clamp     Wait 30 seconds   7  The change in reading should be less than 20 mBar  or equivalent      Section 10  Performance Verification    10 2    8     If the reading drops more than 20 mBar  there is a leak in the flow  system       Open the unit       Check for loose fittings and if any are found  reconnect  them    Repeat the above process  steps 1 7       If the reading continues to show a drop of more than 20    mBar  replace the Flow System  refer to Section 7 5     PERFORMANCE VERIFICATION PROCEDURE    Verify the performance of the monitor using the foll
37. t  to 15 000 ft           12 3 SAFETY STANDARDS    The monitor was designed to comply with EN60601 1 1990  A1 1998     A2 1995  EN60601 1 1 1992  A1 1996  UL 2601 1 and CSA 22 2 No  601 1   M90  EN864 1997     12 1    Section 12  Specifications       12 4    12 2    PERFORMANCE       Sampling Rate    50  7 5 ml min          CO2 Range 0 99 mmHg  0 13 2 kPa and 0 13 0 Vol   at sea  level   Accuracy Accuracy   EtCO2 readings   From power up until steady state is reached  the CO2    Respiration Rate    reading accuracy is   0   38 mmHg     4 mmHg   39 99 mmHg     12  of reading     The CO2 reading reaches its steady state accuracy 20  minutes after power up    0 38 mmHg    2 mmHg   39 99 mmHg    5  of reading   0 08  for every 1  mmHg above 40mmHg     Equivalent values for kPa and Vol Frequency Response     0 150 breaths min        Warm up Time    30 seconds  typical                 Frequency EtCO2 accuracy is maintained up to 80 breaths    Response min   For maintaining accuracy for respiration rate  over 60 bpm  use the neonatal mode   From 81 to  150 bpm accuracy is  12   if the EtCO 2 is higher  than 18 8 mmHg in neonatal mode    System 2 45 seconds  typical   2 9 seconds maximum   Response Time    delay time    Ambient Compensated internally   automatic   Pressure   Alarms EtCO2 high   EtCO2 low   FiCO2 high   No Breath          message         The accuracy of these alarms is the same as the accuracy of the EtCO2    readings        Section 12  Specifications       12 5 POWER
38. tched ON one in the monitor  Be  when the unit is sure the battery pack  operated with is installed correctly   batteries   The battery pack   s Install a fully charged  charge is low  battery pack  charge  the empty battery  pack   The battery pack   s Install a new battery  contacts are defective  pack and check the    monitor functions        The contacts for the Connect the monitor  battery pack on the CO2   to mains line power  board are defective or a   and if the monitor       CO2 board component functions  replace the  has failed  CO2 board   The unit does not   The ON OFF button has   Inspect the button for  turn on when the   a mechanical defect  cracks  deformations  ON OFF button or other damage   is switched ON Replace with a new  when the button   monitor is    connected to  mains power        A CO2 board component   Replace with a new  has failed  CO2 board                    Section 5  Troubleshooting    Table 5 1  Troubleshooting Guide  Continued        Symptom    Probable Cause    Corrective Action       The unit does not  turn on when the  ON OFF button  is switched ON  when the  monitor is  connected to  mains power     The wall socket is not  receiving power or is  defective     Check the wall socket  for power supply or  mechanical defects  If  necessary  use a  different wall socket        The AC adapter is  defective     Replace with new AC  adapter           A CO2 board component   Replace with a new  has failed  CO2 board   One or more The keypad is 
39. ugated cardboard shipping carton  approximately the    size of the monitor or component  with at least 200 pounds per  square inch  psi  bursting strength     3  Fill the bottom of the carton with at least 2 inches of packing  material        Section 11  Packing For Shipment    4  Place the bagged monitor or component on the layer of packing  material and fill the box completely with packing material such  that there is at least 2 inches of packing material around all sides  of the item     Seal the carton with packing tape     6  Label carton with shipping address  return address and RMA  number     Top padding    Handheld Capnograph    Foam padding    Shipping Carton    Figure 11 1  Packing the Handheld Capnograph in  Original Packing       SECTION 12  SPECIFICATIONS    12 1 Physical    12 2 Environmental   12 3 Safety Standards   12 4 Performance   12 5 Power Specifications   12 6 Components and User Interface       12 1 PHYSICAL  12 1 1 Size    206 mm H x 88 mm W x 53 mm D  8 11 H x 3 46    W x 2 06   D     12 1 2 Weight  including battery pack     750 grams  1 66 lb      12 1 3 Noise Emission    maximum 45 dB a     12 2 ENVIRONMENTAL    12 2 1 Temperature       Operating    0  C to 45  C  32  F to 113  F           Storage        35  C to 70  C   31  F to 158  F           12 2 2 Relative Humidity    10 to 95   noncondensing     12 2 3 Pressure and Altitude  for operating and storage         Pressure    430 mmHg to 795 mmhg       Altitude           380m to 4 570m   1 250 f
40. ure 7 4     1  Open the monitor as described in Section 7 2   2  Remove the gas outlet from the tubing and replace with a new gas  outlet     3  Remove the ON OFF button from the CO2 board as described in  Section 7 3 and replace with a new button     Keypads  Figure 7 4   1  Open the monitor as described in Section 7 2   2  Remove the CO2 board as described in Section 7 3     Remove the old keypad  When replacing the new keypad  be sure  the icons are facing the correct direction  refer to  Figure 1 1     Front Cover  Figure 7 4     1  Open the monitor as described in Section 7 2   2  Remove the CO2 board as described in Section 7 3     3  Remove the keypad and ON OFF button  Place them in the new  front cover  When replacing the keypad  be sure the icons are fac   ing the correct direction  refer to Figure 1 1     4  Place the CO2 board in the new cover     Rear Cover  Figure 7 3     1  To replace the rear cover  open the monitor as described in  Section 7 2  and replace the old rear cover with a new rear cover     Section 7  Disassembly Guide  7 8 UPDATING SOFTWARE VERSION    Caution  Observe ESD  electrostatic discharge  precautions when  disassembling and reassembling the monitor and when handling  any of the components of the monitor     1  Open the monitor as described in Section 7 2     2  Remove the EPROM from its socket  on the CO2 board  using an IC  extractor  refer to Figure 7 12     3  Insert the new EPROM   Note  When replacing the EPROM  be sure it is position
41. usly  Calibration screen long press  KA Flow Cal   awit        To select Flow Cal  short press      Pump volt        To start Calibration short press    VA Pump volt             EE    short press      gt  Pump volt   Flow Cal  In progr                Wait until either a Fail or Pass message is displayed       If FAIL is displayed Replace Flow  System and or       Pump Pump volt    refer to Section 5   Troubleshooting    and repeat above   procedure       If PASS is displayed   Exit Service Mode    Pump volt                          Section 6  Flow Calibration Check    Table 6 1  Flow Calibration Continued                    Objective Action Response  To reset and return to long press  measuring mode  mal  Reset  Please wait     To exit service mode Turn off the  and save event monitor    messages                6 3       SECTION 7  DISASSEMBLY GUIDE    7 1 Introduction   7 2 Opening the Handheld Capnograph Case  7 3 Replacing the CO2 Board   7 4 Replacing the Pump   7 5 Replacing the Flow System   7 6 Replacing the LCD and 7 Segment Display  7 7 Replacing the Housing Components   7 8 Updating the Software Version       7 1 INTRODUCTION    The monitor can be disassembled and the following components can be  replaced     CO2 board   Pump Assembly   Flow System   Housing components  EPROM   Liquid Crystal Display  LCD   7 segment displays    Caution  Observe ESD  electrostatic discharge  precautions when  disassembling and reassembling the monitor and when handling  any of its c
42. vent messages    memory  Reset  Please wait     To exit service mode Turn off the  from any service monitor    screen without erasing  the event messages  memory                Section 4  Service Mode    SERVICE LEVEL 2    The Service Level 2 gives the service technician tools on three service    screens       Default Alarm Limits      Default Settings    e Flow Calibration    To access this service level  the service technician must enter a valid  password  1627   refer to Table 4 2     Note  The password should not be disclosed to avoid unauthorized setting    of parameters     Table 4 2  Accessing Service Level 2       Objective    Action    Response       To access Service  Level 2    After entering the  service mode    long press x3      gt     Password Screen  appears    Enter password      W200       Enter Password  1627       Enter the 1st digit    short press  right   xl    wm  CF    Pressing on the  right arrow scrolls  the numbers up   the left arrow  scrolls the numbers  down     Enter password            Move to the next digit    short press         Enter password      100 0       Enter the 2nd digit       short press  right   x6  wm  Co        Enter password     1900             Repeat the same procedure as above for the next two digits  2  7               4 5    Section 4  Service Mode       4 6    Table 4 2  Accessing Service Level 2  Continued        Objective    Action    Response       To access the Alarm  Limits Menu after  entering last password  digit   
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
MONITOR DAP  TomTom Urban Rider TomTom Rider Pro  VE 570  User Manual  BARBECUE BARBECUE  Quick User Guide OrangeHRM 2.4  PL-702 - PLANET Technology Corporation.  MBP26, MBP26/2, MBP26/3, MBP26/4  入 札 公 告  That`s Rights Import Users Guide    Copyright © All rights reserved. 
   Failed to retrieve file