        0224 Sentech Stage IV Millennium 3 User Guide
         Contents
1.   Always follow the safety  precautions to reduce the risk of fire   electric shock and personal injury     CAUTION   Failure to obey a safety warning  may result in property damage or  personal injury to yourself or others   Always follow the safety precautions  to reduce the risk of fire  electric  shock and personal injury     G  ELECTRICAL HAZARD  Indicates risk of electric shock       l   IMPORTANT INFORMATION    A    CAUTION  Certain medical  conditions and treatments are  contraindicated for use of the STAGE  IV MILLENNIUM 3  Always consult  with the patient s physician before  placing a patient on an Alternating  Pressure system     CAUTION  Bed frames used with the  STAGE IV MILLENNIUM 3 can vary  greatly depending on the specific  health care setting  e g  hospitals   nursing homes  home care  etc   Therefore  it is the responsibility of  the caregiver to take the necessary  precautions to ensure the safety   of the patient  This includes  but is  not limited to  the appropriate use  of side rails to prevent falls and or  patient entrapment     OPERATING INSTRUCTIONS    A WARNING  Do not attempt  to operate this product until you  have read and understood the  safety rules contained in this  manual  Failure to comply can  result in accidents involving fire   electric shock or serious personal  injury  Save owner s manual and  review frequently for continual safe  operation and for instructing  others who may use this product     Electronic Controller    A WARNING  T
2.  5 1 5 Calibration    To ensure optimal performance of your STAGE IV MILLENNIUM 3 system  calibration  should be verified every 12 months  Call SenTech Medical Service Centre on 1300 738 885  for calibration information     5 2 Coverlet  5 2 1 Washing and Disinfecting    If there are visible signs of body fluids and or substances present  coverlets should be  washed between patients  Coverlets can be machine washed using an intermediate level  disinfectant  Disinfectant should be used according to the manufacturer s instructions  To  determine the amount of disinfectant to use  determine the amount of water in the washer  and then follow the manufacturer s instructions  Soak the coverlet in the disinfectant  during the wash cycle  Rinse thoroughly in clean water and dry before use     A CAUTION  DO NOT dry the coverlet using the    heat    cycle  Air dry  or select a  low or    non heat    dry cycle  e g  air fluff     5 2 2 Washing Alternative    If there are no visible signs of body fluids and or substances present  the coverlet can be  sanitised between patients  To sanitise the coverlet     1  Apply an intermediate level disinfectant to the upper surface of the coverlet   Disinfectant may be applied either by spraying or by hand application     2  Ensure surface is completely covered with the disinfectant     3  Allow disinfectant to remain in contact with the surface according to the manufacturer s  instructions     4  Remove disinfectant and rinse     5  Allow to air dr
3.  inadvertent modification of system settings     When LOCK is set  all keys are disabled except for the HELP key  the lock feature off           7  and the key to turn    1  Press and hold the ENTER key       2  When the menu changes while still holding ENTER key Rad  press the blue MENU key  underneath the word    LOCK    to lock and unlock the system setting     3  The lock icon on the display panel indicates the lock status   If open  the system is unlocked  If closed the system is locked     NOTE  The system always powers up with the system unlocked     4 2 3 Help    Provides user assistance and system information     Diagram 8 Help Screen    HELP  ss Enter Key EJ to exit helP      Select One     Setup  gt P System  Patient oPrics Info             1  To access the help feature  press the yellow HELP key ea      2  There are three  3  options on the HELP screen     PATIENT SETUP  HELP TOPICS    Therapy    Basics    Advanced  Other    SYSTEM INFO    Guide for system setup  including setting patient height and weight     General guide for system operation  Use maroon ARROW keys SCS    a Q  to tab through the help screens and green ENTER key Kal  to exit HELP     General description of therapy options     General setup instructions and a description  of special modes features     Explains the Low Air Loss  LAL  and LOCK features     Description of lock feature  system alarms  and    replace filter    notification     Displays hours of system operation since the date of manufa
4.  of the patient     3  To resume normal operation of the mattress  push the CPR Latch completely down to  the CLOSED position     Diagram 11     Open Position       CPR latch OPEN position         Colour coded for Stage IV and transport       ns me    4 3 3 CPR Trouble Shooting Guide    Problem    CPR Leaks     CPR loose when latch is in  the CLOSED position     Cause    CPR Latch is not  completely down in the  CLOSED position     Material between the CPR  block and the socket is  preventing a proper seal     Belleville washers are not  properly aligned     Screw on brass shaft is  not tight     O ring s  are  missing damaged     Screw on brass shaft is  not tight     CPR components  are missing or  improperly installed         Solution    Make sure the CPR Latch  is completely down in the  CLOSED position     Make sure that nothing  is between the CPR block  and the mattress shell  that is preventing full  closure of the CPR    e g  sheet blanket     Open CPR latch  loosen  screw  align washers onto  brass shaft  and retighten  screw so that head of  screw contacts brass shaft   Close CPR latch     Open CPR latch and  retighten screw   Close CPR latch     Contact a SenTech  Medical Service Centre  on 1300 738 885     Tighten screw     Refer to Section 4 3 4        4 3 4 CPR Removal nstallation    The CPR is permanently attached to the mattress and should not need to be removed   However  if necessary  it can be removed and reinstalled by following the directions below     I
5.  will cause the system to revert to  default pressures based on height and weight     4 2 2 Alternate Menu Options    Diagram 6 Alternate Menu Screen       To access Alternate Menu Options  press and hold the ENTER key Kal  While holding  select one of the following options     ADVANCED Accesses Advanced SETUP screen   DEFLATE Deflates mattress automatically   LOCK Locks and unlocks the control panel     Press ENTER key after selection to return to normal operation         L    4 2 2 1 Advanced Setup    Advance setup allows adjustment of the following system parameters     Parameter Range Values Default Values  MAX FIRM 32 60 mmHg 40 mmHg  Cycle Time 4 9 minutes 5 minutes  Fowler Boost 15 40  20     A CAUTION  Only advanced users with proper training should use the   Advanced Setup feature  Pressing the AUTOSET will reset the system   to the default settings   1  Use the maroon ARROW keys    fo to move the cursor to the parameter to be changed   2  Adjust the setting using the INCREASE and DECREASE keys     Diagram 7 Advanced Setup Screen    Advanced  yE Firm L JIC le Time Fowler Boost  358     mmmH 6 min        INCREASE Increases the highlighted parameter   DECREASE Decreases the highlighted parameter   AUTOSET Loads default settings     4 2 2 2 Deflate Function  Detach the LAL hose from the right side of the controller     Air will be evacuated from the mattress in 15 minutes  After the specified time  the pump  will shut off        4 2 2 3 Lock    The lock feature prevents
6. 0 hPa   20 to  70   c   10 to 100     500 to 1060 hPa   lt  2000m       8 0 Warranty Information    Our goods come with guarantees that cannot be excluded under the Australian Consumer  Law  You are entitled to a replacement or refund for a major failure and for compensation  for any other reasonably forseeable loss or damage  You are also entitled to have the  goods repaired or replaced if the goods fail to be of acceptable quality and the failure does  not amount to a major failure     Limited Warranty    SenTech Medical Systems  Inc      SenTech     warrants each of its products to perform in  accordance with established specifications for the following time periods  starting from  date the product was shipped from Novis Healthcare P L     STAGE IV MILLENNIUM 3  Compressor Pump 3 Years  Electronic Controller 2 Years  Soft Goods 1 Year    During the warranty period  SenTech through it   s distributors will repair or replace at no  charge any products that are not performing in accordance with established specifications   unless the problem failure is due to     1  customer damage  negligence and or misuse  or    2  unauthorized repairs  Items not covered under warranty include  but are not limited  to  stains  punctures  cuts  damages to electrical cords  rips or tears  dents and or lost   missing parts     All products returned for warranty repairs must be assigned a return authorisation  number  prior to return  Returns should include information describing the problem  a
7. IUM 3 is designed  to accommodate patient weights up to  230kg  500Ib      The Alternating Pressure feature provides  pressure relief by deflating and inflating  every other air cell on a timed interval    It is widely recognized that constant  pressure to a bony prominence is a  leading cause of skin breakdown  The  STAGE IV MILLENNIUM 3 provides  continuous movement of air cells that  alleviates constant pressure and enhances  circulation  The deflated air cells provide  pressure relief  while the inflated air cells  support the patient   s weight     A visible and audible alarm function has  a number of indications depending on the  cause of failure     The electronic controllers provide a     real time    display of air pressure for  both the inflated and deflated air cells   The amount of pressure to support a  patient can be set automatically based on  the patient   s height and weight or can be  manually set for custom configurations   All settings are stored in non volatile  memory  If power is interrupted  the  electronic controller automatically  returns to the previous settings   when the power returns     The Controlled Low Air Loss  LAL  feature  on the STAGE IV MILLENNIUM 3 provides  an optimum environment to assist in  patient healing and comfort  LAL therapy  is delivered through a patented SenTech  Coverlet that provides a flow of diffused  air directly to the patients skin through  thousands of microscopic micro vents  In  addition  the mattress has been des
8. Stage IV Millennium 3 Saniech    Operating Instructions Medical Systems  Inc        3 Zone Pressure Relief System with Controlled Low Air Loss       Important Notice    Before operating this medical equipment   it is important to read this manual and to  understand the operating instructions and  safety precautions  Failure to do this could  result in patient injury and or damage to  the products     If you have any questions  please contact  Novis Healthcare on 1300 738 885            ns ae    Table of Contents    1 0 Safety Precautions 4  2 0 Product Overview 6  3 0 Installation 7  4 0 Operation 9    5 0 Maintenance And Cleaning 19    6 0 Troubleshooting Guide 23  7 0 Product Specifications 24  8 0 Warranty Information 25    90 Product Return Procedure 26          1 0 Safety Precautions    The purpose of safety precautions are to  attract your attention to possible dangers   The safety symbols and the explanations  with them  require your careful attention  and understanding     The safety warnings by themselves do not  eliminate any danger  The instructions or  warnings they give are not substitutes for  proper accident prevention measures     The following symbols may appear in  this manual  on the Control Unit  or on  its accessories  Some of the symbols  represent standards and compliances  associated with its use     A SAFETY ALERT SYMBOL  Indicates caution or warning     WARNING   Failure to obey a safety warning can  result in serious injury to yourself or  to others
9. ctrical Controller is disconnected from  mains electricity     2  On the bottom of the unit  there are two  2  filter housings   e White     Air Intake Filter  e Black     Cooling Fan Filter    3  Remove the filter grill covers and remove the filters  DO NOT unscrew the  filter assembly     4  Clean the filters by washing in a mild detergent and allow to air dry   5  Insert the filters into the filter housing and replace the grill covers     6  If the filters cannot be cleaned  or become damaged  contact SenTech Medical Service  Centre for replacement filters       l   IMPORTANT  Good filter maintenance is critical in keeping your STAGE IV  MILLENNIUM 3 controller in optimal operating condition  Failure to keep the  filters clean will result in system downtime and increased repair costs     5 1 3    The exterior of the controller and CPR assembly should be periodically wiped down with a  cloth dampened with disinfectant     A CAUTION  DO NOT spray disinfectant directly on the electrical controller   or immerse the controller in any type of liquid  This could result in a severe  electrical hazard as this equipment has no protection against ingress of water            L    5 1 4 Cleaning and Electricity   O  IMPORTANT  Before plugging in the controller  check the power cord for electrical  hazards  e g  cuts  exposed wires  worn insulation  etc  If hazards are present   take the controller out of operation immediately and call SenTech Medical Service  Centre on 1300 738 885    
10. cture  and the software revision         ns a    4 2 4 Alarm    Alarm will activate with an audio visual indicator when air mattress does not reach  required pressure within 20 minutes     Diagram 9 Alarm Screen    SYSTEM ALARM    Check 15  Cells 5       4 3 CPR Operation  4 3 1    The Low Profile CPR provides caregivers the ability to rapidly evacuate air from the  support surface in emergency situations  The CPR is located at the foot of the mattress on  the patient s right side and is permanently attached to the mattress  The Low Profile CPR is  shown in Diagram 10 below     Diagram 10     Closed Position       CPR latch CLOSED position          Colour coded for Stage IV and transport    A CAUTION  Loose hoses can be a tripping hazard  Use caution when transporting  the mattress     DO NOT immerse the CPR in any liquids   Avoid exposure to high heat and or solvents   Protect hoses from sharp or pointed objects that could cut  puncture or tear the material     DO NOT attempt to repair the CPR  If damaged or not working properly  call SenTech  Medical Service Centre on 1300 738 885 for service information        4 3 2 Setup  amp  Operation    1  Before operating the system  make sure that the CPR Latch is completely down in the  CLOSED position  See Diagram 10 on previous page      2  To activate the CPR and rapidly evacuate the air from the mattress  pull the CPR Latch  completely up to the OPEN position  See Diagram 11 below   Rate of evacuation is  dependent on the weight
11. e  1  full cycle before  determining if the pressure  increase was sufficient     Re enter pressures needed   DO NOT press AUTO SET     Allow air cells to reach pressure     Turn controller power off for 5  seconds  then on again to reset     Check fuse on power cord socket  by opening the fuse compartment     NOTE  If you need to replace  the fuse  there is a spare 1A  fuse in the fuse compartment     Call Novis Healthcare  on 1300 738 885     See Section 4 3           7 0 Product Specifications    Electronic Controller    Physical Dimensions   Height 30 7cm  12 1 inches   Width 36 8cm  14 5 inches   Depth 15 2cm  6 0 inches   Weight 9 5kg  211b   Electrical Parameters    US and Canada  UL2601 Classification  Power Requirements  Fuse   Maximum Current  Ground    Europe  amp  Australia  Power Requirements  Fuse    Maximum Current    Operating Parameters    Weight Range  Height Range    Environmental Conditions    Operating Ambient Temperature  Relative Humiditys  Atmospheric Pressure  Ambient Temperature  Relative Humidity    Atmospheric Pressure    Storage Shipping    Altitude    Mattress Replacement  Fully Inflated     Physical Dimensions   Height 20 3cm  8 inches   Width 91 4cm  36 inches   Length 203 2cm  80 inches     Weight 10 8 20 5kg  24 45lb         Class   IN  Type B  120VAC  60Hz  3 15A  250V  Fast Acting  UL Approved     220 240V AC 50 60Hz  1A  250V  Fast Acting   lt  0 5A    36 230kg  80 500Ib   127 198cm  50 78 inches      10 to  40   c  30 to 75     700 to 106
12. ely  Contact   Novis Healthcare on 1300 738 885  for repair and service information     Do not return a product for any  reason without first contacting  Customer Service to obtain return  authorisation  See Section 9 0      CAUTION  Do not place any  objects items  such as blankets    on  or over  the electronic controller   Excessive weight on the controller  could result in damage to the  electronic controller     After exposure to extreme high or  low temperatures  allow electronic  controller to equilibrate to room  temperature before operating     WARNING  The controller   circulates room air during operation   Exposure to smoke may cause the  system to fail  Therefore  smoking  by patients or visitors while using  the STAGE IV MILLENNIUM 3 is  strongly discouraged     WARNING  The power cord to the  electronic controller should be  positioned to avoid a tripping hazard  and or damage to the cord  SenTech  Medical Systems recommends  placing the cord under the bed frame  and attaching it to an electrical outlet  by the head of the bed     NOTE  No special skills  training or  knowledge are required to operate  the control unit        2 0 Product Overview    The STAGE IV MILLENNIUM 3 is  microprocessor controlled therapeutic  mattress systems that provide 3 zone  Alternating Pressure and Low Air Loss   LAL  therapies for the treatment   and or prevention of pressure wounds   These therapies can be combined or  operated independently from each other   The STAGE IV MILLENN
13. eplace with a STAGE IV  MILLENNIUM 3 Mattress assembly     3  Position mattress and hoses so  that the   e CPR hose assembly is to the left of  the footboard  e Low Air Loss  LAL  hose is to the  right of the footboard    4  Ensure CPR latch is completely down in  the CLOSED position  See 4 3 for Setup  and operation     5  There are two  2  sets of straps with  D rings on each side of the mattress  and one  1  at the head of mattress   Use these straps to secure the mattress  replacement to the bed frame     O  IMPORTANT  Make sure that the  attachment of the mattress does not  interfere with the bed movement  or operation     3 0 Installation  continued     3 2 2 Electronic Controller    1     Hang the controller on the outside edge  of the footboard on the bed frame     A CAUTION  Placing the   electronic controller in the patients  bed  in contact with the patient  or  under and coverings  eg  sheets  could  result in serious injury and or affect  the systems performance       Attach the six  6  colour coded hoses    to the corresponding colour coded  connectors on the left side of the  controller i e  red to red  green to  green  etc   See Diagram 1 on right        Attach the LAL hose to the single    connector on the right side of the  controller   See Diagram 2 on right        Plug into a grounded 240 Vac 50 Hz    electrical outlet        Diagram 1       Diagram 2    O    LOW AIR LOSS              as ae    4 0 Operation    4 1 Key Functions and Settings    Control Pane
14. flowing into  the Low Air Loss Coverlet     Display readings  appear scrambled     Controller is inoperable     CPR hoses become disconnected  from exterior block     Cause    The alarm is activated if the air  cells do not reach programmed  pressure  Check the CPR  connections  See Section 4 3     The pressures may be set too low  for the patient s weight     The AUTOSET key may have been  pressed by mistake which would       overwrite    all customized settings     The internal pump gives priority  to the air cells in the mattress   Once the air cells are inflated to  the selected pressure  air will  then be directed to the coverlet     Power surges can  cause the controller to  temporarily malfunction     May be caused by a power surge  substantial enough to overload  the internal circuitry     May be caused by  internal damage     Excessive force applied to  hose connections     Solution    Be sure all hoses are properly  connected to the controller     Check all hoses along the inside of  the mattress  Each hose should be  tightly connected     Check each air cell to ensure  there are no leaks   It will be  easier to detect a possible leak  if you place the system in the  MAX FIRM mode      Once the leak has been resolved   the alarm light will automatically  turn off after three cycles    To reset the system more quickly   turn the power off and then on  again to reset     Increase bed pressure  An  increase of 3 5 mmHg is usually  sufficient  However  wait at  least on
15. his device is not  suitable for use in the presence of  a flammable anaesthetic mixture  with air  or in the presence of a  flammable anaesthetic mixture with  oxygen or nitrous oxide     Only plug into a grounded power  receptacle and use the power  cord supplied with the STAGE IV  MILLENNIUM 3     Exposure of the electronic  controller to any liquid while it is  plugged in could result in a severe  electrical hazard     CAUTION  Only use fuses that   have the same specified rating  See  Section 7 0 Product Specifications    Using fuses with higher ratings  could result in damage and or injury     WARNING  The 240 VAC  50   Hz models of the STAGE IV  MILLENNIUM 3 incorporates fusing  only in the ungrounded phase  conductor  They must be used in  health facilities on grounded systems  where conditions of maintenance  and supervision ensure that only  qualified persons will service the  electrical distribution system     The electronic controller is a  precision electronic product   Use care when handling or  transporting  Dropping  or other  sudden impacts  may result in  damage to the controller     WARNING  Risk of electric shock  DO  NOT OPEN  Do not attempt to repair  or service the electronic controller     L    1 0 Safety Precautions  continued         Repairs and service should only   be done by a SenTech Medical  Systems nominated Service Centre   If the controller is not functioning  properly  or has been damaged   unplug the unit and take it out of  service immediat
16. igned  to provide an anti shear anti friction  surface for patients     The system includes a rapid release  CPR handle for emergency deflation        NOTE  It is recommended that all shipping    and packing material be saved in the event    that the product has to be sent back to a  SenTech Medical Service Centre     3 1 Unpacking and Inspection    Carefully remove the controller  mattress  and all accessories from the shipping  cartons  Inspect all items for any damage  that may have occurred during shipping   Any damage  or missing components  should be reported to a SenTech Medical  Service Centre as soon as possible     Mattress Replacement  The box  contains a completely assembled  mattress replacement system     This system consists of    e 5cm 2 inch Foam Mattress   e Air Cell Assembly with CPR pull handle  e Top Coverlet with Low Air Loss Hose   e CPR Hose Assembly    Electronic Controller  The electronic  controller is in a separate box containing     e Electronic controller   e Power cord   e Operating Instructions  e Quick Setup Guide    The mattress is treated as an applied part     3 2 Installation    The STAGE IV MILLENNIUM 3 is  designed to operate in a controlled  environment  which is free from  extreme temperatures  high humidity  and or excessive amounts of airborne  particulates  such as dust and smoke        L    3 0 Installation    3 2 1 Mattress Replacement  1  Unroll the mattress replacement     2  Remove the current mattress from the  bed frame and r
17. just the percentage of Fowler Boost  see Advanced SETUP  Section 4 2 2 1     4 2 1 3 Low Air Loss  LAL     Provides a gentle flow of air through the coverlet   1  Pressing the blue MENU key underneath    LAL    turns the LAL on and off     2  When LAL is on  the LAL on the screen is highlighted and the animated LAL icon is  displayed     NOTE  Air for LAL is delivered when the air cells are not being inflated     4 2 1 4 Max Firm    Provides a flat stable surface for patient transfer and medical treatment   1  Select MAX FIRM by pressing the blue MENU key underneath the words    MAX FIRM        2  When the transfer treatment is complete  return to desired mode of therapy by pressing    appropriate MENU key  If system remains in MAX FIRM for more than 20 minutes it  will revert to normal operation        4 2 1 5 Standard Setup  Diagram 5 Standard Setup Screen    Patient HeiSht  Fagin  Weight  185 lb        1  Press the green ENTER key to access SETUP screen   Setup instructions can be accessed by pressing the HELP key                Set patient height using the INCREASE and DECREASE keys     Use the maroon ARROW keys    fo to move cursor to weight     Enter the patient   s weight using the INCREASE and DECREASE keys     YS 2 B    Press AUTOSET key     4 2 1 6 Custom Setup    1  Use the maroon ARROW keys    fo to move the cursor to the desired zone  head  trunk  or foot      2  Adjust pressure using the INCREASE and DECREASE keys     NOTE  Do not press AUTOSET  Pressing AUTOSET
18. l  The STAGE IV MILLENNIUM 3 control panel is shown in Diagram 3 below     Diagram 3      3 Zone m Alternating Pressure    Controlled Low Air Loss      Millennium 3                   Graphic Blue Menu Enter Arrow Help  Display Keys Key Keys Key  Keys   ENTER KEY  Green  Takes you into and out of the SET UP screen   HELP KEY  Yellow      Accesses HELP screens   Q    ARROW KEYS  Maroon     MENU KEYS  Blue     POWER SWITCH    S Tabs through the SET UP menu     Selects MENU options     This switch is located next to the electrical cord on the  bottom right side of the unit  It is used to control the AC  power to the unit        4 2 Basic Features    ALTERNATING PRESSURE Provides pressure relief therapy by inflating and deflating  alternate air cells     FLOAT Provides a low pressure static surface    LOW AIR LOSS  L A L   Provides a gentle flow of air directly to patient   s skin    MAX FIRM Provides a firm surface for patient transfer and medical  treatment    LOCK Prevents inadvertent changes to the control panel    FOWLER BOOST Increases pressure in the Trunk zone for patients in an    inclined position     DEFLATE Automatically deflates the mattress     4 2 1 Main Run Screen    The Main Run Screen has four  4  menu options  Press the blue MENU keys ES  select  the desired option  Selected options are highlighted when selected   Diagram 4 Main Screen    Alternating Pressure 4444 fn    oo i   i     Hh lS    He   fa for  c9 16 SetuF  A A A       Max Firm    During normal operati
19. nd or requested repair and be sent to a SenTech service centre by prepaid transportation   Novis Healthcare will return the repaired replaced product at no charge  Warranty repairs  do not extend the length of the warranty period     Neither SenTech  its distributors  officers  directors  employees or agents shall be liable for  consequential or other damages  including but not limited to personal injury  loss  or any  other expense  directly or indirectly arising from the use of its products  The sole remedy  for breach of the limited warranty granted herein shall be repair or replacement of the  SenTech products     All product specifications are subject to change without notice           9 0 Product Return Procedure    The STAGE IV MILLENNIUM 3 SYSTEM has been designed to provide you with years of  trouble free service  However  in the event that the product needs to be returned for any  reason  such as calibration or repair  the following return procedure must be followed   Failure to follow this procedure may result in unnecessary delays     Return Procedure  Before returning a product to a SenTech Medical Service Centre     1  Contact the medical equipment distributor from where the product was purchased   Alternatively  contact Novis Healthcare to obtain a    Return Authorisation Number      RAN  and    Return Goods Form     RGF      2  Package the product in its    approved    packaging     3  Reference RAN on the packaging and delivery documents     4  Ship product t
20. nstallation    1  Place Spring on the brass shaft of  the CPR     2  Slide brass shaft through the hole in  the CPR socket     3  With latch in open position  align two  holes in CPR block with two screws   with spacers  on CPR socket     4  Squeeze CPR block and socket together  and place the flat washer onto the  brass shaft     5  Place the two  2  Belleville washers  onto the brass shaft  as noted    ZL  IMPORTANT  Make sure the two  2   Belleville washers are oriented so  that when they are together they  form a    pocket        6  Thread the screw into the end of the  brass shaft  Tighten ensuring that the  washers are centred on the brass shaft  and head of screw contacts brass shaft     Removal    1  Remove screw in the end of the  brass shaft     Remove two  2  Belleville washers   Remove flat washer     Slide CPR assembly out of the CPR socket     a   2 fF    Remove the spring        Diagram 12     Installation    Black CPR  Socket in  Mattress Shell    Brass Shaft Belleville Washers    Flat Washer        L    5 0 Maintenance and Cleaning    e   l IMPORTANT  All disinfection should be done with a    hospital grade     disinfectant in accordance with the manufacturer   s specified instructions   Check manufacturer s instructions before use     5 1 Electrical Controller    The electronic controller is easy to maintain   5 1 1 Fuse Replacement     A CAUTION  Only use fuses that have the same rating as specified  See Section 7 0    Using fuses with higher ratings could 
21. o the attention of Novis Healthcare  address details on back cover            Notes          A       Tile    Medical Systems  Inc     TM       Serviced and distributed by    uw    NOVIS    HEALTHCARE    M Novis Healthcare Pty Ltd   Unit 12 12 Mars Road Lane Cove NSW 2066 Australia ABN 45 102 735 491  T 1300 738 885 F 1300 738 886 E novis novis com au  WWw novis com au    0224 10 12    
22. on  the following words appear above the four Menu Keys   Pressing a key results in the corresponding action     THERAPY Pressing this key toggles the system between Alternating  Pressure and Flotation therapy modes    FOWLER Toggles the Fowler Boost on and off    L A L  Toggles Low Air Loss  LAL  on or off     4 2 1 1 Therapy    1  Press this blue MENU key to select either Alternating Pressure  AP   or FLOAT modes     2  This key remains    highlighted            L    3  The table below shows the function of each therapy     Float Mode AP Mode    All air cells in each zone  Head  Alternate air cells in each zone are inflated to the  Trunk  amp  Foot  are inflated to the selected pressure   selected pressure     All air cells remain at the On a timed cycle  alternate air cells sequentially  selected pressure  inflate and deflate    Pressure of all air cells in each Air pressure in the deflated and inflated air cells in  zone is displayed on the screen  each zone are displayed on the screen     Time remaining in the cycle is also displayed on  the screen     4 2 1 2 Fowler Boost    This feature allows the user to increase the pressure in the trunk zone for patients who  are in an inclined  or Fowler position  Fowler Boost is designed to prevent patients from     bottoming out    when the weight is concentrated in the trunk area while in the   inclined position     1  Pressing the blue MENU key underneath    FOWLER BOOST    increases the pressure  in the trunk zone     2  To ad
23. result in damage and or injury     One  1  replacement fuse is provided with your controller and is located in the  compartment on the electrical cord socket  To replace a fuse     1  Ensure the power switch is in the OFF position     A WARNING  Ensure the Electrical Controller is disconnected from  mains electricity     2  Remove the power cord from the electrical socket on the side of the controller     3  Using a small sized flat head screwdriver  pry the fuse holder away from the socket and  slide it out of the socket     4  Remove the    blown    fuse from the fuse holder clip and discard     5  Remove the spare fuse from the storage compartment and install it into the fuse holder  clip  See Diagram Below      Diagram 13     Fuse Holder  Top View     Active Fuse    Spare Fuse    Fuse Holder Clip  c       6  Push the fuse holder completely back into the electrical socket until it    snaps     into place     NOTE  The fuse holder must be properly oriented to slide in correctly  Do not force     7  Replace power cord and turn on the controller        5 1 2 Filter Maintenance    The microprocessor in the controller is programmed to notify you to clean the  filters after   800 hours of use  See the Filter Maintenance screen below      Diagram 8 Filter Maintenance Screen    REPLACE FILITER l    T  Remind Filter  he  later       It is very important that you clean the filter when notified  To clean the filter     1  Unplug the electronic controller     A WARNING  Ensure the Ele
24. y before use   e  l IMPORTANT  It is recommended the system is cleaned between patients and  approximately every two weeks if in constant use        5 3 Outside Shell    Wipe down shell using an approved disinfectant  ensuring that all surfaces come in  contact with the disinfectant  Rinse off with clean damp cloth and allow to air dry     5 4 Air Cell Assembly    A CAUTION  DO NOT machine wash or dry the air cells  The air cell assembly does  not routinely need to be cleaned or disinfected between patients       l   IMPORTANT  DO NOT disassemble the STAGE IV MILLENNIUM 3 mattress   unless cleaning disinfection is required     If cleaning disinfection is required  disassemble the mattress using the  colour coded zippers       l   IMPORTANT  DO NOT disconnect the hoses from the air cells     5 5 Foam Mattress    The foam mattress is fully enclosed in a nylon urethane cover and should not require  cleaning  However  if it does become    visibly    soiled  it may be wiped down with  disinfectant  ensuring that all surfaces come in contact with the disinfectant  Wipe off with  a clean damp cloth and allow to air dry     5 6 CPR Assembly    The exterior of the CPR Assembly can be periodically wiped using a cloth dampened  with disinfectant            ns a    6 0 Trouble Shooting Guide    Problem    Alarm is on     Patient is sinking or    bottoming  out    while lying flat     The pressure setting was  increased  but the pressure does  not appear changed     Air is not constantly 
    