1. Operating Instructions    863 Valley View Road  Eighty Four PA 15330 USA  Tel  724 941 9701 Fax  724 941 1369 www skcinc com       Flowmeter  Catalog No  303    SKC primary standard flowmeters are designed  manufactured  and tested for accuracy  no calibration  is needed  Read the operating instructions thoroughly before operating the flowmeter     Principle of Operation   The air sampling pump to be calibrated is connected to and pulls air through the flowmeter   s  volumetric glass tube where a flat soap bubble  film  is interposed into the flow path  As the  airflow causes the film to move up the volume marks  travel time is measured using a stop   watch  Flow rate can be calculated using the travel time and known tube volume     Assembly  SKC Model 303 is shipped ready to use  Film solution is supplied as well as a length of rubber  tubing to connect the flowmeter to the pump     Preparing for Operation  Remove the flowmeter from the case  and place it upright on a level surface       Remove the large rubber stopper from the top of the flowmeter    3  Tip the flowmeter at an angle  and pour a small amount of film solution into the top of  the glass tube  The liquid should run down into the rubber bulb at the bottom    4  Continue adding film solution until it completely fills the bulb and rises to just below  the side arm inlet  Squeeze the rubber bulb occasionally to release trapped air    5  Replace the stopper on top of the flowmeter     Wetting the Walls of the Tube  
2. The interior walls of the flowmeter must be wet for the film to travel without rupture  There  are two methods     Method 1   1  Tip the flowmeter to a horizontal position  Allow the film solution to run out of the rub   ber bulb and into the tube    2  Maintaining the horizontal position  rotate the flowmeter so that film solution coats the  entire inside surface of the glass  Caution  Do not allow the solution to enter the metal  tube in the rubber stopper at the top of the flowmeter    3  Return the flowmeter to an upright position and allow the liquid to flow back down  into the rubber bulb  Squeeze the bulb occasionally to release trapped air     eae 2  Connect a pump to the flowmeter  see Connecting a Pump     Turn on the pump  Each successive film that rises up the tube will wet the glass  The  walls are sufficiently wet when the film bubbles successfully reach the highest volume  mark on the flowmeter     Connecting the Pump   Note  If the pump to be calibrated draws air  connect it to the rubber tubing on the flowmeter   s  top rubber stopper  If the pump blows air  use rubber tubing to connect it to the lower side  arm near the base of the flowmeter     1   2     3     Remove the rubber cap from the side arm inlet near the base of the flowmeter   Using rubber tubing  connect one end to the appropriate port on the flowmeter and the  other end to the inlet of the sampling medium to be used  if applicable    Using flexible tubing  connect the outlet of the sampling med
3. ium to the inlet of the  pump   Sample  Media  Outlet              Calibration  Chamber    Outlet   upper  port     Inlet    Tygon  Tubing            Sample    Pump Flowmeter           Inlet  lower    Pum  p side arm port     Inlet       Measuring the Flow    A    3     Turn on the pump    Squeeze the rubber bulb on the flowmeter to force the liquid above the inlet  Bubbles   film  will form    Using a stopwatch  start timing the film as it passes the zero line and stop timing as the  bubble reaches the 100 ml line  For best results  form several bubbles  five to six seconds  apart  and time the last bubble     Caution  Do not squeeze the bulb continuously  This causes froth to form on the walls mak   ing timing difficult     The flow rate can be determined from the Flow Chart on the back page  or calculated using  the following equation     Flow  ml min    60 x  volume traveled        Time  in seconds     Storage or Transport   Disconnect the pump and replace the rubber cap on the lower inlet of the flowmeter  The  instrument can be placed in any position without solution loss  SKC recommends that it be  transported in its case     Volumetric Calibration of Hand Pumps  SKC Model 303 Flowmeter can be used to determine the volume of manual hand pumps such  as the bellows or piston pumps used with detector tubes  Follow these procedures     Bellows Pumps  i e   Drager      1  Connect a detector tube to the pump according to the manufacturer   s instructions    2  Connect the in
4. let of the tube to the rubber hose on top of the flowmeter    3  Compress the bellows pump and release slowly  Squeeze the rubber bulb on the flow   meter  Bubbles  film  will begin to form    4  When the film reaches the zero line on the flowmeter  compress the bellows pump com   pletely and release  The film will move up the flowmeter and stop when the bellows  pump is filled to capacity    5  Record the point at which the film stops  Most bellows pumps are designed to draw  100 ml  Lines on the flowmeter are graduated at 90  95  100  105  and 110 ml  Visually  extrapolate the volume reading if the film stops between lines    6  Calculate and record the volume     Note  Pumps should be repaired if they fall outside of the 90 to 110 ml range     Piston Pumps  i e   Gastec    MSA  Kitagawa    1  Connect a standard detector tube to the piston pump according to the manufacturer   s  instructions    2  Connect the inlet of the tube to the rubber hose on top of the film flowmeter    3  While pulling the piston slowly back  squeeze the rubber bulb on the flowmeter until  bubbles  film  form    4  Watch the film movement closely  When the film is exactly at the zero line on the flow   meter  push the pump piston all the way in  Then  draw the piston out to full capacity  and lock it in place  The film will move up the flowmeter and stop when the piston  pump is filled to capacity    5  Record the point at which the film stops  Most piston pumps are designed to draw  100 ml  Lines o
5. n the flowmeter are graduated at 90  95  100  105  and 110 ml  Visually  extrapolate the volume reading if the film stops between lines    6  Calculate and record the volume     Note  Pumps should be repaired if they fall outside the 90 to 110 ml range     Notice  This operating instruction may not address all safety concerns  if any  associated with  this product and its use  The user is responsible for determining and following the appropriate  safety and health practices and regulatory limitations  if any  before using the product  The  information contained in this document should not be construed as legal advice  opinion  or  as a final authority on legal or regulatory procedures     Flow Chart for 303 Calibrator    100 ml line 100 ml line   100 ml line 100 ml line  Time Flow Time Flow Time Flow Time Flow   sec   ml min   sec   ml min   sec   ml min     Using 100 ml line   12 500   13 462   14 429   15 400   16 375   17 353   18 333   19 316   20 300   21 286   22 273   23 261   24 250   25 240  231  222  214 Time  amp  Flow Rate  207 using 50 ml line  200 50  193 49  187 49  182 47  176  171  167  162  158  154  150  146  143       Accessories and Spare Parts    Description Cat  No   Film Solution  1 pint   473 ml  302 4011  Precision Digital Stopwatch 303 01 1  Replacement Connecting Hose  2 pieces  P3032101  Replacement Squeeze Bulb P303231 1  Replacement Flowmeter Sidearm Cap P3032312  www skcinc com    Form  3701     Rev 0803    
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
LG TS-C122YDA0 User's Manual  Manuale di istruzioni  Le régime disciplinaire  OmniMount NC200C flat panel wall mount  Smart Investing Owners Manual - Smart Investing with Brian Golly  pv reunion du groupe de travail sensibilisation et  Philips EcoHalo Halogen capsule lamp 872790025292725    Copyright © All rights reserved. 