1. It is used to connect to the computer with the USB Cable main body of SONOST 2000 It is used to connect to the source of electric Power Cable Make sure to connect it with a grounded electric outlet It is used to install the program on the Computer The program is installed on the computer before being delivered Program CD OTO0 2F0423 01 13 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 2 1 5 Articles List P Check the following accessories before installing the system C they are not in good condition contact to OsteoSys or its authorized dealer for this service 2 Stockpiles and Replacement of Articles It connects a foot with the ultrasound probes and helps ultrasound to progress between them Balloon 10 Balloon Housing Housing 3 TA rl Keep the spare balloons and housings in YF acool place It helps ultrasound to progress between a balloon and patient s heel Ultrasound Gel 2 bottles It removes some material which prevents Alcohol ultrasound from progressing Black 1 unit Ee Color 1 unit Spec HP 3820 OT00 2F0423 01 14 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 2 2 Installation of SONOST 2000 Since SONOST 2000 is made up of precise components you shou
2. After taking water into the balloon close it with the stopper Check the surface of the probe through the balloon Tap the balloon to remove bubbles inside of the balloon OTO0 2F0423 01 57 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OT0O1 2R0422 02 MUL 4 2 3 Storage Unplug the power cord Dust on the product can damage the system After use make sure to cover it with a vinyl cover Daily If the hardened gel remains in the device it can effect on BMD storage results Please clean the device everyday 4 2 1 Cleaning Unplug the power cord Do not detach cable of the device connected to plug to prevent dust from coming inside Place cable near the product or at an empty area in its front Storage Put cover on the product to avoid dust for a long time of period Keep the product according to instructions described in 2 2 Installation of SONOST 2000 2 2 Installation of SONOST 2000 OTO0 2F0423 01 58 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 4 3 Safe Use of SONOST 2000 4 3 1 Safety Rules The followings are instructions for safe use of the product Make sure to comply with the instructions when you activate the system Refer to 4 3 2 Caution related to electronic medical equipment 4 3 2 Cautions Related to Use of Electronic Medical Equ
3. OTO0 2F0423 01 38 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Patients List Doctor Name Peter Jackson 15 Kwonseohyun 14 Sungbumsik 13 Leejongcheol 12 Junseungcheol 11 Wi 10 I 3 Cho 8 Jat H Ju 6 Kan 5 Kir 4 Ch 4 amp 2 Ki P The number of patients 15 Gender _ Ethnicity The Year of B 7 Female Male Male Male Korean 1983 2003 01 15 Korean 1973 3 2003 01 15 Korean 1966 8 2003 01 15 Korean 1965 H 2003 01 15 Patient Deletion Do you want to delete the current patient This list display only 200 patients If you want to find another patients Use Patient Search 3 2 2 5 Inquiry amp Deletion of Clinical History of Existing Patients Main Menu 1 If you want to retrieve the clinical history of existing patients select a desired patient and click the Patient Info button on the right side of the Patients List screen Then Patient s Info screen appears that contains the patient s clinical history as shown below On the left amp below side the measurement results list shows If you select a measurement result in the list the relevant data appears in the screen Patient s Info Patient s Name Kwonseohyun gt gt History Measurement Times Left Foot SIS 3 2003 01 24 17456 432 175 6 2 2003 01 15
4. SONOST 2000 OsteoSys WWW SCC OT00 2F0423 01 37 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 3 2 2 4 Revision amp Deletion of Patient s Information 1 If you want to revise the registered patients information select a desired patient using the mouse and click the Info Change button in the Patients List screen 2 Then the Patient s Info Change screen appears in this condition if you enter a new value to make desired changes in the patient s information and click the OK button the patient s information is revised The changeable patient s information refers to name date of birth gender height weight race menopause state foot size and so on 3 If you want to delete the registered patients information select a desired patient using the mouse in the Patients List screen You can select several patients at the same time To do so click those patients while pressing the Ctrl key or if you want to select all patients within a certain click the first and the last line of the area while pressing the Shift key 4 After selecting a desired patient click the Patient Delete button on the bottom of the screen Then the message box Patient Deletion inquiring about whether to delete the information surely appears and if you click the OK button the information is deleted
5. This information is for the report printing Back Finish __ Cance better products for beller life www osfeosys com 3 2 2 2 Registration amp Deletion of Users 1 As soon as you configure the user environment the Doctor Selection window appears as shown below If you are a registered user doctor click the Doctor Select button after selecting your name in the Doctor Selection window using the mouse And then if you enter the password in the edit box Password Enter and click the OK button you can log in the program Eugene Kain 02 1923 1251 RnD Peter Jackson 01x 293 1987 better products for better ife www osfeosys com OT00 2F0423 01 29 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Doctor Selection Password Enter www osteosys com 2 If you are a new user doctor click the New Doctor button in the Doctor Selection window and then the New Doctor window appears If you enter your information name phone number department and password and click the OK button the registration process is completed The log in process is same as the description 1 Name Telephone Department Password www osteosys com OT00 2F0423 01 30 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1
6. amp Bakp las ie ES La a 52K Whole Backup 030212 174324 wdb mw Re Files of type Whole DB vwdb x Cancel EP Exit OsteoSys 4 Clear If you click this button the message box Surely clear doctor inquiring whether to delete the user surely appears If you click the OK button you can clear all users DB I2000 Surely clear DB DB shall be cleared It can be affects other doctor s DB Only be used by device _ Exit OsteoSys OT00 2F0423 01 47 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 5 Recover If you click this button you can recover all users DB All records related the DB are written in this field After selecting the record and clicking the OK button in this screen the DB 1s recovered www ostfeosys com OT00 2F0423 01 48 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 3 2 2 8 Configuration of Environment 1 If you want to configure the user environment again click the Configuration button in the default screen There are four items from Settings to Doctor in this menu 2 Settings G 3 2 2 1 Execution of Program and Configuration of me the Inc
7. 240 V C Whether the system is fully grounded D Whether all the connection parts power and optional equipment are connected with the system properly LE je Turning on Power E The product has been factory set for optimum performance Do not attempt to adjust any preset controls or switches except for operation as specified in this manual F If you experience any trouble with the product turn it off immediately and contact OsteoSys or its authorized dealer for assistance G Before electrically or mechanically connection any other manufactures devices to SONOST 2000 contact OsteoSys or its authorized dealer for instructions H Do not operate the product improperly when you use it with other products I Make sure to check whether external parts of the device directly touches the patient J Make sure to check whether the products keep distances at least 10cm from walls CAUTION Equipotential Bonding In hospital doctors and patients are subjected to dangerous uncontrollable compensating currents These currents are due to the potential differences between connected equipment and touchable conducting parts as found in medical rooms The safest solution to the problem is accomplished is consistent equipotential bonding Medical equipment is connected with connecting leads made up with angle sockets to the equipotential bonding network in medical rooms OT00 2F0423 01 61 A4 210 x 297 User s Manual Confi
8. 2R0422 02 MUL 3 If you want to delete registered users select a desired user and click the Doctor Delete button of the Doctor Selection window If there are two or more registered users the message box Patients Shifting inquiring your intention appears as shown below OK All registered users except the user you selected to delete are displayed on the left side of this OK button If you select the user who you want to move the patients data the user you selected to delete has measured until now to and click this button the data moves to the selected user s DB and the user you selected to delete is deleted NO If you click this button the user you want to delete and the patients data the user you selected to delete has measured until now are deleted Cancel If you click this button you can cancel the deletion process Delete all about Doctor Peter Jackson Not shift Not delete Return to Select Doctor www osteosys com 4 If you click the OK or NO button the message box Surely delete doctor inquiring whether to delete the user surely appears If you click the OK button the user information 1s deleted OT00 2F0423 01 31 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Patients Shifting Surely delete doctor Your patient s data sha
9. 3 1 4 Disposing water 3 1 5 Where and how much apply to the Patient s heel 3 1 6 Correct position for Patient s foot amp body 3 1 7 How to apply 1 amp 2 foot positioner according to the Different foot sized patient 3 2 Using S W 3 2 1 Progress Table of Program 3 2 2 Using Program 3 2 2 1 Execution of Program and Configuration of the Incipient Environment 3 2 2 2 Registration amp Deletion of Users 3 2 2 3 BOU Measurement 3 2 2 4 Revision amp Deletion of Patient s Information 3 2 2 5 Inquiry amp Deletion Clinical History of Existing Patients 3 2 2 6 Calibration of SONOST 2000 3 2 2 7 Data Backup 3 2 2 8 Configuration of Environment 3 2 2 9 Inactivation of SONOST 2000 OTO0 2F0423 01 18 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 3 1 Using H W 3 1 1 Applying the Balloon to the Measuring Device Clean the surface of the probe with an alcohol pad Attach the balloon to the probe holder Adjust the balloon housing to the probe holder OT00 2F0423 01 19 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Take water into the balloon using an injector until it is full 65 70cc Close it with the stopper Tap the balloon to remove bubbles inside of the balloon OT00 2F0423 01 20 A4 210 x 297 User s Manual Confidential O
10. and wring water from the cloth and wipe out dust Make sure not to put water or detergent on the computer directly Water or detergent can damage the computer Make sure not to use volatile chemicals such as benzene or thinner because it can damage the surface of main body of the computer Computer OT00 2F0423 01 54 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 Product Example Keyboard Printer Device OTO0 2F0423 01 Soft Cloth DOC No OTO1 2R0422 02 MUL Instructions Dust the keyboard with a soft brush Direct contact with water or detergent can damage the key board Remove paper from the printer and dust the printer with a soft brush carefully If there is strong stain on the printer soak a clean and soft cloth with neutral detergent diluted with water in about a ratio of 5 to 1 and wring water from the cloth and wipe out dust And then clean the stained area with dry cloth Cover the printer with printer cover to avoid water User s instructions for printer After using the device remove the balloons from the device and wash with water The device should be cleaned with a water wringed clean soft cloth everyday because gel remains around the probe or below the device Especially gel stained on the balloons should be moved After this procedure clean the stained area with a dry piece of cloth Make sure not to p
11. desired patients information and click the OK button then the results appears in the screen Patients List 3 8 7 6 5 4 3 Ea 1 Patient Search Search Key x Sen _ By Name B By PatientiD Female C Male By Gender a Example 1976 By Year of Birth This list display only 200 patients If you want to find another patients Use Patient Search OTO0 2F0423 01 34 A4 210 x 297 Main Menu User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OT0O1 2R0422 02 MUL lt Measurement of Patients gt 1 If you want to measure BOL select a desired patient using the mouse in the Patients List screen and click the Measurement button on the right side of the screen Then the Measurement screen appears as shown below Measurement Patient s Name Kwonseohyun gt gt T S c o r e Measured Date Site ar SOS m s BUALdB MH2 BOl Bone Quality Index T Score T Ratio Z2 Scotre Z Ratio Comment Edit Prini Prowse 2 Click the Measurement Start button on the bottom of the Measurement screen Then the Measurement Start window inquiring measuring site and so on If you select inquiring articles and click the Measurement Start button ultrasonic waves are generated from the probe Patien
12. incorporated herein Incorrect operation or failure of the user to maintain the equipment relieves the manufacturer or his agent of the system s noncompliance with specifications or of responsibility for any damage or injury The following conventions are used throughout the manual to denote information of special emphasis WARNING Warning is used to indicate the presence of hazard that can cause severe personal injury death or substantial property damage if the warning is ignored CAUTION Caution is used to indicate the presence of hazard that will or can minor personal injury and property damage 1f the caution is ignored Note is used to notify the user of installation operation or maintenance information that is important but not hazard related The symbols which are shown in this manual or SONOST 2000 OT00 2F0423 01 3 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL OTO0 2F0423 01 The noticed information which should be concerned with explanation in this manual The noticed information when the device is operated The reference page or section Applied Part Type B I and O on power switch represent ON and OFF respectively The Attention symbol that marks warning and important information in the user s manual The conductor provides a connection between equipment and the potential equalization bus b
13. 15378 497 926 1 2005 01 15 15373 485 920 Right Foot No Measur 50 No Measur SOS BUA _BQI T Score MP TM BUA BOQI T Score MP TM 37 SE CH 0 6 K Wes 0 6 xo X S Gab 15541 526 1000 IU PSPL E 2 2003 01 15 1 2003 01 15 155 MP Menopause or not TM Take Medicine or not 0 2 Ka 0 3 x X Result Taking Medicine or not Comment Measured Date Site S05 m s BUACdB MHz Bol Bone Quality Index T Score Ratio Z2 Score Ratio Menopause or not OTO0 2F0423 01 39 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL There are two Print Preview buttons in the Patient s Info screen comparing to the Measurement screen and each button has the different function If you click the Print Preview on the right side the result you selected with a mouse appears in the Patient s Info screen Print SITE REPORT Input Hospital s Name Input Hospital s Address Prhed Dat 20301 24 Phone hpt Hospitals Phone Namber vong AdtT Rat 953 Age MatiedZ Rati 0 9 T hg Mediche or rot Doctor Name Petr Jackson SONOST 2000 OsteoSys MN er Et If you click the Print Preview on the left side all results of the selected patient appears in the Pati
14. 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Fuse Replacement Close 400 120V 200 240V 3 15A Open the fuse drawer on the upper side of the appliance inlet there will be the two small fuse holder Push the fuse holder toward the arrow direction and Pull the fuse holder toward the upper side of the appliance inlet Remove the old fuse by pulling up Install the new fuse by pushing to the fuse holder Insert the fuse holder to the appliance inlet At this time the arrow direction in the upper side of the fuse holder should be in accordance with that on the fuse draw Also the same method is used to exchange the other fuse folder Close the fuse the drawer CAUTION For continuous protection against a risk of fire hazard replace only with same type and same rating of fuse Checking points after use of product A Make sure to clean the product so that use Deactivate the product and turn the power off in order B Do not pull the power cord forcibly OTO0 2F0423 01 63 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL C Comply with followings to keep the product properly Avoid places that can wet the product Avoid places where air pressure temperature and humidity drastically change and where there are direct ray of sun dirt salt and sulfur
15. Deletion of USsers ccsccssccsccssceeeees 29 OTO0 2F0423 01 5 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 3 2 2 3 BQI INICASUPCIMON E 33 3 2 2 4 Revision amp Deletion of Patient s Information 38 3 2 2 5 Inquiry amp Deletion of Clinical History of Existing Patients 39 3 2 2 6 Calibration of SONOST 2000 cccecccscesccescesceeeees 44 Se Data BACKUD E 46 3 2 2 8 Configuration of Environment sccsccscccscccsccsscescces 49 3 2 2 9 Inactivation of SONOST 2000 cceccsccsccsscsccsccseees 51 4 Maintenance and Repair of SCONOST 2000 52 4 1 Resolution to Problems kk 53 4 2 Maintenance and Repalll cccssccssccssccesccesscesccesccessceeses 54 A2 E EE EE 54 4 2 2 Stockpile and Replacement of Articles ccceccssceesceescees 56 BZD NORA E 58 4 3 Safe Use of SCONOST 2000 see 59 AS E E e AT TEE 59 4 3 2 Cautions Related to Use of Electronic Medical Equipment 60 5 Specifications amp Software Uporades cc ceeeeeeees 65 SA SDC CHI CATIONS searen a N 66 5 2 Software U ET aesetacssaipeaxwaaetereieaatansanks 68 6 Product MEIER EE 69 OTO0 2F0423 01 6 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Chapter 1 Introduction 1 1 General Produc
16. Do not lean shake or rock the product D Clean accessories after use E Make sure to clean the product so that users can use it immediately If the product is out of order do not try to repair it for yourself Do not modify use the product If any modification is desired ask OsteoSys or its authorized dealer for this service Check maintenance and repair Cautions in times of transportation of cart If you have to move the cart on incline floor you should detach the product from the cart so that the product does not slip to the ground Environmental protection If the products are at the end of their useful lives do not dispose them at your option Contact OsteoSys or its authorized dealer for this service Others Read the user s instructions carefully and use the product properly OTO0 2F0423 01 64 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Chapter 5 Specifications amp Software Updates 5 1 Specifications 5 2 Software Updates OTO0 2F0423 01 65 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 5 1 Specifications General Classification Class Type B Applied part Device Dimensions 497mm x 279mm x 210mm Weight 7 9 Kg AC 100V 240V Input Voltage Free Voltage Single Phase Frequency 50 60Hz Power 150W max Trans
17. Measurement Start Site Left Foot C Right Foot Menopause ornot e C Yes Taking Medicine or not re Yes Comment 1 Make feet parellel 2 Make foot position fixed with foot guide 3 Measuring operated with the calf fully laid on the calf supporter OTO0 2F0423 01 35 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 3 When ultrasonic waves are generated the word on the Measurement Start button is changed into Measurement Stop In case of emergency click this button and you can stop measuring In the normal condition the message box Measurement Completion inquiring about whether to restart to measure BO or not appears with an alarm sound after measurement being finished Measurement Patient s Name Kwonseohyun gt gt Jre E re re Bry re oe er we a m Bea Mia STE o Saat a Baba ba AE ve E ct ibd 94 a Bde Bambam Bo af a Bales ee ek ct EN El rr 0 o oO o HI zs Measurement Completion daf BQI 174 2 T Score 3 8 e za am eu Em H ee eee mr mi T Score 3 8 T Ratio 167 5 Z2 Score 4 7 Z Ratio 171 9 4 If you click the OK button the results appears in the screen as shown below Measurement Patient s Name Kwonseohyun gt gt s O o HI Beau Measured Date 2003 01 24 15 10 Site Left Foot SO
18. ONOST 2000 Do not install it in following places A Places where the product is exposed to water vapor B Places where the product is exposed to spray or splashing water C Places where the product is exposed to high density oil vapor D Places where the product is subject to excessive shocks or vibrations E Places where the angle of inclination of the mounting surface exceeds 10 degrees F Places where air pressure temperature or humidity drastically changes G Places exposed to direct sunlight or places where air is polluted due to dirt salt and sulfur H Places where chemicals are kept or toxic gas is leaked I Places where the AC power line voltage fluctuates heavily J Places where the AC power line voltage fluctuates heavily while this product is operating K Places where there is a ventilation problem L Do not lean shake or rock the product during transportation M Adjust frequency voltage or current to the product N The product is portable but make sure to put a brake on it when you stop the product O Ground the product by using ground wire P Do not share the power outlet OTO0 2F0423 01 60 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Check followings before using the product A Whether power is off before putting a plug in the socket B Whether the power supply line is suitable with that of the system AC 100
19. S m s 1746 4 BUA dB MH2 43 5 BOl Bone Quality Index 174 2 T Score 3 8 T Ratio 167 5 Z Score 4 7 Z Ratio 171 9 SES e OTO0 2F0423 01 36 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 5 If you need to write or modify comment click the Comment Edit button on the bottom of the screen Then the Comment Edit window you can write or modify comment a nx mn gt vi u i a VI Ze dt eter peed KSE TA of H 55 saline ell gll EI e o K Eis See eben ener MN lth yf nm NAUK Be ART Haiti E to D A 1 i Biber fe CES GC H aa EI i DA a a f H e p GP 6 If you click the Print Preview button on the bottom of the screen that shows the results after measuring BQI you can see the results that will be printed out in advance as shown below If you click the Print button on the top of this screen the results is printed out The Print Preview function can be set in the Configuration menu REPORT Input Hospital s Name Input Hospital s Address Prhed Dat 20301 24 Phone spatHospitats Phone Namber R gl Dap 01 15 CT Osteopenia EI Osteoporosis Left Foot SOS eee eee 1746 4 BUAREMHI as vong AditT Rath Age MatiedZ Rat 17139 T hg Mediche or wot Ho Doctor Name Petr Jackson
20. SONOST 2000 User s Manual Model SONOS T 2000 Council Directive 93 42 EEC Concerning Medical Device OsteoSsys http www osteosys com OsteoSys Co Ltd User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OT0O1 2R0422 02 MUL SONOST 2000 User s Manual CAUTION 1 You must be well acquainted with this manual before using it 2 This manual should be placed where the user could read it whenever necessary Software Version 2 01 01 Documentation Version 012 2003 03 17 Part Number OTO1 2R0422 02 MUL Manufacturer and EC Authorized Representative Information e Manufacturer OsteoSys Co Ltd 551 1 Davit Tower 2F Seongnae dong Kangdong gu 134 030 Seoul Korea Tel 82 2 2041 5600 Fax 82 2 2041 5601 e EC Authorized Representative Ecornet Medical GmbH High t Part Mainstr 6c d D 45768 Marl Germany Tel 49 2365 92 437 0 Fax 49 2365 92 437 55 OTO0 2F0423 01 2 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Thank you for purchasing SONOST 2000 Ultrasound Bone Densitometer To ensure safe operation and long term performance stability it 1s essential that you fully understand the functions and operating maintenance instructions by reading this manual before operating the equipment Particular attention must be paid to all warnings cautions and notes
21. Sys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 3 2 Using S W 3 2 1 Progress Table of Program ct SONOS 2000 e Print Result 4 e Print History 3 OTO0 2F0423 01 21 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 3 2 2 Using Program 3 2 2 1 Execution of Program and Configuration of the Incipient Environment CAUTION If you hear any strange noise from the screen switch it off and contact to OsteoSys or its Authorized dealer 1 Turn on the power switch 2 10 seconds after calibration that checks functions of the system and memory device the program is automatically activated and you should configure the incipient environment 3 Settings Patient ID Mask Set the type of patients ID Default Settings Set the default measurement site Patients gender menopause state ethnicity the function Print Preview etc Language Select your own language to use the program www osteosys com OT00 2F0423 01 28 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 4 Hospital Set the name address and phone number of your hospital Hospital Hospital Information Hospital Name AMAA Hospital Address input Hospital s Address Hospital Telephone input Hospital s Phone Number
22. ar of the electrical installation The date of manufacture Table of Contents 4 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Di He OCOD EE 7 1 1 General Product TInformatton ue 7 2 Configuration of SONOST 2000 9 2 1 Composition of SONOST 2000 cccsssccssscceescceescceesees 10 2 1 1 Shape and Components of SONOST 2000 ccecsscseeeees 10 2 1 2 Shape and Components of Measurement Equipment 11 ZA D Keyboar ME 12 21A Accessories LEIS EE 13 ZLS Articles List eorne axtewlesbesdsewhednataeteataeusedendeens 14 2 2 Installation of SONOST 2000 15 2 3 Before Turning on ENEE 16 3 Using SONOST 2000 18 3E USME Ee 19 3 1 1 Applying the Balloon to the Measuring Device sseeees 19 3L Usine the OC Etgen E 21 3 1 3 Patient Measurement Procedure c sccsccsscsccscccsccssccscescees 22 JEA DISPOSING DEER 24 3 1 5 Where and how much apply to the Patient s heel 24 3 1 6 Correct position for Patient s foot A body NEEN 25 3 1 7 How to apply 1 amp 2 foot positioner according to the Different foot sized AVION E 26 3 2 Using ND VV EEN 27 3 2 1 Progress Table of Trau 27 3 22 Usino vr er EE 28 3 2 2 1 Execution of Program and Configuration of the Incipient Environment ccccccsscecscccsscecsccessceess 28 3 2 2 2 Registration amp
23. as shown below When you click the Daily Test Start button in this screen ultrasonic waves are generated from the probe Daily Test BUA 42 0 db MHz Bandwidth 0 52 MHz Center Frequency 0 32 MHz Nol BUA Bandw CenterF SOS Gain 42 2 0 52 0 32 W 0o u m ewe E j fa g SOS 1753 4 mis Gain 124 i Clean the alcohol pad 3 When ultrasonic waves are generated the word on the Daily Test Start button is changed into Daily Test Stop In case of emergency click this button and you can stop daily testing In the normal condition the message box Daily Test Success appears as shown below with an alarm sound after progressing calibration of the system and if you click the OK button the progress of calibration is completed OTO0 2F0423 01 44 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL TE e aT Wr gt a E H BUA 45 0 db MHz __ Bandwidth 0 55 MHz Center Frequency 0 32 MHz t a ny GES i e WC a Leena el GIE WI Es NEE e 4 af H E z FA em ak e ki sesi f V Ae E 2 e at VI eK He d H tape spe sie amp fy x tins ft D E as Raf au Ba DR S A m Sq se nisip EE Y we d fi u vw Ki 3 dh i In case of the screen shows the calibration has failed do not u
24. covered 7 OsteoSys Co Ltd will not be responsible for any loss damage or injury resulting from delay in rendering service under this warranty 8 OsteoSys Co Ltd will not be responsible to you for incidental or consequential damages of any kind arising from or connected with the use of its equipment OTO0 2F0423 01 69 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL BG What Should be Noted in times of Request of repair 1 If the product is out of order stop using it and read the manual carefully 2 Before you request for repair switch off the product and contact a designated service agency after checking the name of model serial number and date of purchase OsteoSys Co Ltd will replace the defective product at no cost to you OsteoSys Co Ltd will pay the shipping and insurance costs of product sent to you 3 Defective equipment shipped from you to OsteoSys Co Ltd must be packed in the replacement cartons Shipping and insurance costs for the return of the defective product must be pre paid by you 4 If you request the product information to repair the defective product yourself we provide you with the proper information This limited is in lieu of all other warranties expressed or implied including warranties of merchant ability or fitness for any particular use No representative or other person is authorized
25. dential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Equipotential Connection Lead Terminal Socket E Main Body Equipotential Connector Earth in Medical Room NOTE Try to place the system far from power generators X ray machines broadcasting Stations and transmission lines to avoids electrical noise during scanning Otherwise abnormal images may result An independent circuit and a safely grounded outlets are strongly recommended for SONOST 2000 Poor or abnormal images may occur if the system shares a power source with other electrical or electronic equipment NOTE The classification according to standard IEC EN 60601 I Type of protection against electric shock Class I Degree of protection against electric shock Type B Degree of protection against harmful ingress of water IPX 0 Methods of sterilization or disinfection See Chapter 4 2 1 Not suitable for use in the pressure of flammable anesthetic mixtures Check followings while the device is in operation A Whether the product and the patient are normal B Whether the products keep distances with one another C If you find any abnormality related to the product or patient stop its operation and check if it can cause any danger to the patient D If the fuses are gone exchange the same type 250V T315AL Otherwise a fire may occur or you may receive electric shock OTO0 2F0423 01 62 A4 210 x
26. ducer amp Balloon Center Frequency 0 5MHz 10 AuW cm Ultrasound Output Bandwidth gt 30 Diameter 31 75mm Focus Type Flat Connector Type UHF OTO0 2F0423 01 66 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Measuring Circuit jaan Aaaloeue eni Ultrasound Transducer 2 channel P SE Bandwidth 0 3 1MHz Output Analogue signal Amplitude 200Vp p Pulse width 4 usec A D Converting 8 bit 32MHz Measurement of Electric signal The number of Data 1024 Hnvironmental 30 75 relative humidity 0 90 relative humidity non condensing Not to exceed 2G in six millisecond Install system in a clean well ventilated area Excessive dust and other airborne debris 10 40 Dust Fumes Airborne debris contaminants can impair sensitive parts We recommend instituting a No Smoking policy in system area OTO0 2F0423 01 67 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 5 2 Software Updates 1 Updates to SONOST 2000 software are available at our homepage http www osteosys com 2 Updates to SONOST 2000 software are free regardless of the product warranty 3 Updates to SONOST 2000 software are irregular we will inform you through your email whenever SONOST 2000 software needs to be updated 4 For details call t
27. ean Kimjaehong 3 Korean Choijungja Korean njungja Korean Kimminsoo Male Korean Leejun Korean S 9 8 7 6 5 4 3 e 1 This Patients List screen displays only 200 patients for improving the system speed If you need to find other patients use the Patient Search menu 2 If you want to register new patients click the New Patient button on the bottom of the Patient List screen Then the New Patient window you can enter the information of new patients appears Enter the information and click the OK button Then the patient is registered and the information is displayed OTO0 2F0423 01 33 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL tte New Patient PatientID 15 PatientID Mask Consecutive Numbers Daten 16 a Name e Birth Date i i VYYY MM DD g 8 7 6 5 4 3 2 1 This list display only 200 pati Gender Female C Male ie Ethnicity Koreaan Height Cm Weight Kg Menopause or not No Yes Taking Medicine or not No C Yes Foot Size mm Foot Supporter fi Di 2 Comment 3 If you want to retrieve registered patients click the Patient Search button and then the Patient Search window contains the retrieval items name ID gender and date of measurement appears If you enter the
28. een Then the message box Result Deletion that inquiring about whether to delete the result or not app Patient s Info Patient s Name Kwonseohyun gt gt History Result Result Deletion Measurement T Do you want to delete the current result Menopause or not Ze No Taking Medicine or not e No i Yes Comment 4 2005 01 24 1746 4 43 5 3 2003 01 24 17456 43 2 2 2003 01 15 1537 8 49 7 1 2005 01 15 15373 48 5 E T 1 9 g Right Foot No Measur SOS 2 2003 01 15 1554 1 1 2003 01 15 1553 0 BUA 52 6 51 2 BOL T Score MP T 100 0 0 2 x x 99 1 03 XIIA MP Menopause or not TM Take Medicine or not OT00 2F0423 01 43 A4 210 x 297 ves Main Menu User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 3 2 2 6 Calibration of SONOST 2000 1 Make sure to calibrate the system to reduce a margin of error and to get accurate values in times of examining bone density It 1s to raise credibility of examination results through checking of system reliability on a daily basis Whenever you calibrate the system you should use the QC Phantom included in the accessories list 3 1 2 Using the QC Phantom 2 To calibrate the system click the Daily Test button in the default screen Then the Daily Test screen appears
29. ent Info screen REPORT Input Hospital s Name Input Hospital s Address Prhtd Dat 20301 24 Phone hpit Hospitals Phone Namber Docpr Name Petr Jackson SONOST 2000 OsteoSys MM SCC OT00 2F0423 01 40 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL REPORT OsteoSys 2F DAYIT Tower 551 1 Sungnae Dong Kangdong Ku Seoul Korea Printed Date 2005 02 14 Phone 82 2 2044 5600 Patient s Info Menopause or not Foot Supporter Regi Date Osteopenia Osteoporosis SOS ims BUA AA MAz Young AduitT Fatioy 6 0 s Age Matchedt Z Rotor G22 Taking Medicine or not te ER e A E Comment Doctor Mame 2005 01 15 Test SONOST 2000 Osteaays Print Sample OT00 2F0423 01 41 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OT0O1 2R0422 02 MUL REPORT OsteosSys ZE DAVIT Tower 551 1 Sungnae Dong Kangdong Ku Seoul Korea Printed Date 2003 02 14 Phone 62 2 2041 S600 Patient s info kimminso0 1976 10 31 ender eit Bick CELL Eee 0 0 Cm Ethnicity Korean Menopause or not Foot Size Foot Supporter Measured Times Regi Date Normal Osteopenia 2 EI Osteoporosis Measurement Times Result Measured Date T 5core selected 2003 01 15 0 9 Recent 2003 01 15 0 9 2003 01 15 1 0 2003 01 15 0 9 SS e
30. gen Comment Doctor Mame 2003 01 15 Test SONOST 2000 Osteosys WML ohne ss OTTI Print Sample OT00 2F0423 01 42 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 2 After revising patients information menopause state taking drug etc if you click the Result Update button in the result screen the message box Result Updating inquiring about whether to update the information surely appears If you click the OK button the relevant history is updated Patient s Info Patient s Name Kwonseohyun gt gt History Result Result Updating Measurement Do you want to update the current result Left Foot No Measur SOS BUA 4 2005 01 24 17464 435 3 2003 01 24 17456 432 1 2 2003 01 15 15376 497 1 2003 01 15 153735 465 Right Foot No Measur SOS BUA BOI T Score MP TM 2 2003 01 15 15541 526 1000 0 2 x x L 200L0t1b 15530 Sie OI 0 3 x x Menopause or noti Ze No C Yes Taking Medicine or not Ze No C Yes Comment 7 MP Menopause or not TM Take Medicine or not me Ble Print Preview Result Update Result Delete Print Preview 3 If you want to delete the clinical result of patients select a desired result and click the button Result Delete in the Patient s Info scr
31. hape and Components of Measurement Equipment FOOT GUIDE 1 6 BALLOON 1 4 CALF SUPPORT 1 5 N STATUS LAMP 1 7 kgn TOP COVER 1 1 SWITCH PANEL 1 3 BOTTOM PLATE 1 2 OT00 2F0423 01 11 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 2 1 33 Keyboard Keyboard Activate SONOST 2000 by pressing a key on the keyboard according to the instructions shown on the monitor The manual shows keys required to activate SONOST 2000 as shown below e Enter key OT00 2F0423 01 12 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OT0O1 2R0422 02 MUL 2 1 4 Accessories List d Check the following accessories before installing the system KL D e o D D they are not in good condition contact to OsteoSys or its authorized dealer for this service User s Manual 1 Make sure to Keep it at a designated place so that users can read it anytime they need to Use them in accordance with patients foot size Foot Positioner It is used to calibrate the system C Phantom Q oe Keep it in the specified place and protect gt it against deformation by heat or pressure Injector It helps to take water into the balloon easily Guide It helps to take water into the balloon easily It prevents dust from gathering the inside of the Device Cover device
32. he OsteoSys Co Ltd Customer Service Department or your OsteoSys Co Ltd distributor OTO0 2F0423 01 68 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Chapter 6 Product Warranty Scope and Period of Warranty 1 OsteoSys Co Ltd guarantees all new product against faults in materials or workmanship for a period of one year from date of shipment to purchaser OsteoSys Co Ltd will as sole and exclusive remedy at no charge replace any such defective unit returned to its service designated warranty period 2 This warranty applies only to failures from operating the product under conditions for which it was designed Warranted product is to be used only for the intended and labeled indications presented in the literature accompanying the product 3 OsteoSys Co Ltd will charge service fees in following cases If the product is out of order after the warranty period expires If the product is out of order due to natural disasters such as fire flood windstorm hail lightning and earthquake If the product is out of order due to improper transportation after installation and careless use If the product is out of order due to improper repair or renovation by someone other than service personnel of the company 6 Cosmetic defects or deterioration will not be refinished or replaced The costs of replacement of fuses are not
33. ipient Environment New Standard i IDMask and L Default Value Hospital Infor Doctor s Infor www osteosys com 3 Hospital 1 5 13 2 2 1 Execution of Program and Configuration of Incipient Environment OTO0 2F0423 01 49 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Configuration Settings Hospital Doctor Hospital Information Hospital Name iput Hospital s Name Hospital Address input Hospitals Address i Hospital Telephone input Hospital s Phone Number l This information is for the report printing better products for better ite www osteosys com 4 Doctor You can set your name phone number and department If you click the PW Change button you can change your password Configuration _ Settings Hospital Doctor Basic Information Name Emna Telephone 02 1345 6756 Department FA New Password New Password Confirmation If you want to change the password click the Change PW b After editing the password click Change PW button again for GE Password Change Current Password IT The only one doctor use this program better products for better life www osteosys com OT00 2F0423 01 50 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 3 2 2 9 Inactiva
34. ipment Do not open it because there are some high tension components inside It might cause serious danger Check if the cable coating is not damaged before turning on the system to prevent electric shock or leakage of electricity If it 1s damaged or broken immediately contact an agency CAUTION Do not share the power outlet with other products If you share the outlet it will affect the product Read user s instructions carefully before using the product Place the user s instructions at a designated place so that users can read it anytime they need it CAUTION If you hear a strange noise or if you find any abnormality on the screen after turning on the product contact the headquarters or an agency Make sure to turn off the power switch before cleaning the product OT00 2F0423 01 59 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 4 3 2 Cautions Related to Use Electronic Medical Equipment Safety rules described below are based on Paragraphs 495 of laws and ordinances promulgated by Ministry of Welfare of Japan The safety rules pertain to use of electric medical equipment Read them carefully for safe use of the product and make sure to refer to 4 3 1 Safety rules 4 3 1 Safety Rules Nobody can handle SONOST 2000 except for doctors and authorized personnel Comply with followings in times of installation of S
35. ld install the product according to instructions below Do not install and keep SONOST 2000 at excessively high or low temperatures Proper ye temperature e LI Ay Sem I oe te eee 18 27 e d Do not install the system in the place with polluted air and high humidity and do not expose the machine to direct ray sunlight It is advisable to use SONOST 2000 in the place equipped with air conditioning or Jf heating humidity IS Sa dat ack 3 E F je ack t Do not share the power outlet with other products through an extension cord oe Proper 20 80 Do not cover ventilation of the system or place it near the wall High inside temperature of the machine might cause a fire OTO0 2F0423 01 15 Do not install or keep SONOST 2000 in the place where the machine can be rocked or shaken Make sure to place the system S x parallel to the de A JE J Ty e KR ef Wiel Install SONOST 2000 in the place where it is free from water or chemicals Do not use the system near equipment that generates strong magnetic fields KS V Les 100 240V is usable When SONOST 2000 is taken out of a warehouse insulating transformers should be set to 100 240V Since sudden power outage can remove all data saved in the product o you should install hich eo SONOST 2000 in IE the place where a power supply is stabilized A4 210 x 297 User s Manual Confidential OsteoSys Co L
36. ll be shifted Not shift Not delete Return to Select Doctor www osteosys com 5 If there is only one registered user when you click the Doctor Delete button in the Doctor Selection windows the message box The Last Doctor inquiring to advise data backup and to delete the user surely appears If you click the OK button the deletion process starts If you are only one user check the checkbox The only one doctor use this program on the bottom of the Doctor Selection window Then you can use the system without entering password 3 Doctor Selection The Last Doctor Backup is needed Anyway delete www osteosys com OT00 2F0423 01 32 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 3 2 2 3 BOU Measurement lt Registration amp Retrieval of Patients gt 1 If you click the BQI Measurement button in the default screen you can move to the Patient List screen contains all information of registered patients E Regi Date jiz 2003 01 15 2003 01 15 2003 01 15 2003 01 15 2003 01 15 2003 01 15 2003 01 15 2003 01 15 2003 01 15 2003 01 15 2003 01 15 2003 01 15 2003 01 15 2003 01 15 2003 01 15 Kwonseohyun Korean Sungbumsik Korean Leejongcheol Korean Korean Korean Korean Chonamkyung Korean Jangsungjin Korean Jungwoojin Korean Kangdongkun I Kor
37. njury The system should be placed at least 20cm from the wall However the USB cable length should not be longer than 3 0m Never remove any system covers If the fuses are gone exchange the same type 250V T315AL Otherwise a fire may occur or you may receive electric shock In case of changing the PC or the printer for other products check them according to IEC EN60601 1 1 Unplug the power cord before you move or clean the device and after using SONOST 2000 For installation and operation of this equipment national regulation shall be considered OTO0 2F0423 01 8 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Chapter 2 Configuration of SONOST 2000 2 1 Composition of SONOST 2000 2 1 1 2 1 2 2 1 3 2 1 4 2 1 5 Shape and Components of SONOST 2000 Shape and Components of Measurement Equipment Keyboard Accessories List Articles List 2 2 Installation of SONOST 2000 2 3 Before Turning on Power OTO0 2F0423 01 9 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 2 1 Composition of SONOST 2000 2 1 1 Shape and Components of SONOST 2000 w Oe ge oa lt Front View gt OT00 2F0423 01 10 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 2 1 2 S
38. se the system and execute Daily Test once again If calibration fails continually contact to OsteoSys or its authorized dealer OT00 2F0423 01 45 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 3 2 2 7 Data Backup 1 If you need to save the data click the Backup button in the default screen If you click the button the default screen of Backup menu appears as shown below SC STEE Backup Operations Backup ki Retrieve DB will be saved for backup DB will be retrieved DB will be cleared bel DB will be recoverd It can be affects other doctor s DB Only be used by device 12000 EP Exit OsteoSys 2 Backup If you click this button you can save all users DB in the PC OTO0 2F0423 01 Backup Operations Save as type Whole DB wdb x Cancel It can be affects other doctor s DB Only be used by device A 46 A4 210 x 297 1 ZV000 EP Exit OsteoSys User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 3 Retrieve If you click this button you can exchange all users DB into others in the PC Q Backup Operations 1 ZV000 ZA D lt A DB will be saved for backup Open 2 x Lookin
39. steoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 3 1 2 Using the QC phantom 3 2 2 6 Calibration of SONOST 2000 Insert the standard block positioner Apply gel on the both side of the QC Phantom Put the QC Phantom between the balloons OTO0 2F0423 01 21 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 3 1 3 Patient Measurement Procedure Put the foot positioner in accordance with patient s foot size Ex Using the 1 foot positioner Ex Using the 2 foot positioner If you use the 2 foot positioner necessarily put it on the 1 foot positioner Complete feature Clean the both side of patient s heel with an alcohol pad OTO0 2F0423 01 22 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Apply plenty of gel on the both side of patient s heel Insert patient s foot Let a patient move his foot up and down for the balloons to be covered with ultrasound gel sufficiently Complete feature Completer feature Start to measure after adhering patient s calf to the calf supporter OT00 2F0423 01 23 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 3 1 4 Disposing water Dispose water inside
40. t Information Osteoporosis is one of the serious diseases This device is a bone densitometer which estimates a bone mineral density of the calcaneus by ultrasound It takes about 15 seconds to measure the density and to display the shape of ultrasonic wave by computing simulation on the monitor To ensure safe operation and long term performance stability it is essential that you fully understand the functions operating and maintenance instructions reading this manual before operating the equipment OT00 2F0423 01 7 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Features and Cautions This device is an ultrasound bone densitometer which has 0 2 T score precision error The equipment must be operated only by or under supervision of a qualified person When you give the appropriate commands in SONOST 2000 software ultrasonic waves are generated The generated ultrasonic waves pass through the Patient s calcaneus and the electric signal is treated by SONOST 2000 algorithm SONOST 2000 measures speed of sound SOS and the frequency dependent attenuation of the sound waves BUA and combines them to form a clinical measure called Bone Quality Index BQI All operators must understand the potential hazards in the use of medical electronic devices They must be able to recognize hazards and protect themselves and others from i
41. t of measurement turns 3 2 2 3 BQI Measurement out to be abnormal 3 2 2 6 Calibration of SONOST 2000 Is the desired part of the body properly placed i 9 Despite measurement Are the balloons adjusted properly according to the instructions Are bubbles in the balloon its result turns out to be abnormal Is the hardened gel ticked on the surface of the balloons OTO0 2F0423 01 53 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 4 2 Maintenance and Repair 4 2 1 Cleaning CAUTION Turn off the power before cleaning the system Dust the monitor with a soft brush every day Clean the screen of the 771 monitor with a soft cloth If there is a strong stain on the cover soak a clean and soft cloth with neutral detergent diluted with water in about a ratio of 5 Al to 1 and dry the cloth and remove the a stain with it And then wipe out the stained area with soft and dry cloth again Brush _ Monitor Make sure not to put water or detergent on the monitor directly Water or detergent can damage the monitor If you use volatile chemicals such as benzene or thinner surface of the monitor can be damaged Dust the computer with a soft brush everyday If there is too much dirt on the computer soak a clean and soft fe d cloth with neutral detergent diluted d with water in about a ratio of 5 to 1
42. td Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 2 3 Before Turning on Power 1 Connection of Power amp USB Cable with the device Plug in the cable Plug into the connector When you detach the cable from the main body of SONOST 2000 do not hold the cable and instead hold the connector Otherwise the wire could be down 2 Connection of Power Cable with the outlet Avoid sharing the outlet OTO0 2F0423 01 16 Make sure that the covering of cable is not damaged to prevent electric shock or short circuit If you find any damage or any sign of it immediately contact an agency Put the Power cord into only one plug Make sure not to share the power outlet to prevent SONOST 2000 from being affected A4 210 x 297 User s Manual Confidential DOC No OTO1 2R0422 02 MUL OsteoSys Co Ltd Model Designation SONOST 2000 3 Adjust the balloons in the device 4 2 2 Stockpile and Replacement of E A Articles 4 Make sure that paper is properly provided to the printer Supply of paper A EN 4 2 2 Stockpile and Replacement of Articles A4 210 x 297 OTO0 2F0423 01 17 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Chapter 3 Using SONOST 2000 3 1 Using H W 3 1 1 Applying the Balloon to the Measuring Device 3 1 2 Using the QC Phantom 3 1 3 Patient Measurement Procedure
43. the measurement device Replacement period 1 3 week If there is air in the balloons fill up water promptly Otherwise it causes an error when you measure patients BMD 3 1 5 Where and how much apply to the Patient s heel Before measuring using alcohol to rub patient s heel protection for removing air bubbles surface of patient s skin and other infections Apply plenty of ultrasound gel to both side of patient s heel ENOUGH Ankle Jelly Good NOT ENOUGH OT00 2F0423 01 24 No Good A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 3 1 6 Correct position for Patient s foot amp body Adhere a heel to the measuring device closely as shown below Make sure measuring device and body should be one line as shown below Device GOOD NO GOOD OT00 2F0423 01 25 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 3 1 7 How to apply 1 amp 2 foot positioner according to the Different foot sized patient Standard Block Positioner 1 Foot Positioner 2 Foot Positioner 1 In case patient s toes placed the 1 foot positioner using the 1 foot positioner Standard Block Positioner 1 Foot Positioner 2 Foot Positioner 26 A4 210 x 297 OTO0 2F0423 01 User s Manual Confidential Osteo
44. tion of SONOST 2000 If you click the Exit button in the default screen the message box Program Exit inquiring about whether to exit the program appears and if you click the Ok button the program is inactivated e www osteosys com OT00 2F0423 01 51 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Chapter 4 Maintenance and Repair of SONOST 2000 4 1 Resolution to Problems 4 2 Maintenance and Repair 4 2 1 Cleaning 4 2 2 Stockpile and Replacement of Articles 4 2 3 Storage 4 3 Safe Use of SONOST 2000 4 3 1 Safety Rules 4 3 2 Cautions Related to Use of Electronic Medical Equipment OTO0 2F0423 01 52 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL 4 1 Resolution to Problems P If SONOST 2000 shows following symptoms check the system as follows C Is the device properly plugged PT A ere Isn t the power cord damaged with the power on Aren t the fuses gone Computer is not activated with Is the device properly plugged the power on Isn t the power cord damaged gt Monitor is not activated with Is the monitor switched on the power of the computer on Screen of the monitor is not bright enough Result of calibration of the system turns out to bel 3 2 2 6 Calibration of SONOST 2000 abnormal Resul
45. to represent or assume for OsteoSys Co Ltd any warranty liability beyond that set forth herein OTO0 2F0423 01 70 A4 210 x 297
46. ut water or detergent on the device directly because it can cause fatal damage in the electric circuit contained inside A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Product Example Instructions After using the device remove the Balloon balloons from the device and wash kt na CNR them with water 4 2 2 Stockpile and Replacement of Articles P Make sure to keep following articles of consumption all the time C 2 1 5 Articles List Specifications Use white A4 paper Printing Paper D l P Refer to user s instructions for details Replace ink cartridge with a new cartridge if the printed text or picture is not vivid enough Ink Cartridge User s Instructions for Printer OTO0 2F0423 01 56 A4 210 x 297 User s Manual Confidential OsteoSys Co Ltd Model Designation SONOST 2000 DOC No OTO1 2R0422 02 MUL Specifications Gel is a necessary article when measuring BMD Please purchase new one before it being out of stock Gel Purchase ones at our authorized dealer when it needs to be used lt Replacement procedure of the balloon gt Remove the hardened gel from the probe Probe surface Clean the surface of the probe with an alcohol pad After attaching the balloon to the probe holder Adjust the Balloon balloon housing to the probe holder balloon of ST

Download Pdf Manuals

image

Related Search

Related Contents

大切なものを遺したい〜市政活動報告 第一号〜 平成23年9  Manual do Utilizador do Nokia 6110 Navigator  ダウンロード  GA-EP35-DS3P  Radio Shack 61-2425 User's Manual  Philips SHAVER 9000 SensoTouch 3D Electric shaver HQ9020  Guia do Usuário do Cliente da Web Avigilon™ Control Center  Greensmaster 3150  Media Spinner 50_AT.indd  DISTANCIOMETRO SÓNICO CLAVE ZE8009  