1. ID7sx Base Application blocks No Content Format 020 Current DeltaTrac Response A B _ Target weight weight value Unit Tolerance value number 2 Write A W 0 2 0 Target weight weight value Unit Tolerance value number 2 021 001 Tare memory Response A B _ j Weight value Unit 2 1 999 Write AjW O x x x x x Weight value _ Unit 021 999 Comment xx xxx 21 001 21 999 021 Tare memory 1 25 Response A B Weight value Unit sos Write A W O x x Weight value _ Unit 045 Comment xx 21 45 The contents of the tare memories 1 25 are identical to the contents of the tare memories 021_001 021_025 046_001 DeltaTrac memory Response A B _ Target value weight value _ Unit _ __ em 1 999 Tolerance value number 2 046 999 Write AW O x x x x x Target value weight value _ Unit Tolerance value number 2 _ _ Comment xx xxx 46 001 46 999 046 DeltaTrac memory Response A B _ Target value weight value Unit MET 1 25 Tolerance value numer 2 070 Write A W 0 x x _ Target value weight value Unit Tolerance value number 2 Comment xx 46 70 The contents of the DeltaTrac memories 1 25 are identical to the contents of
2. 5 1 3 Command description Switch keypad on or off Command R0 Switch on keypad Ril Switch off keypad Response R B Keypad switched on or off Comments e Factory setting Keypad switched on e When the keypad is switched off the terminal cannot be operated manually Set zero Command Z Set gross weight display to zero after weighing platform stabilization effect as when ZERO SET key is pressed Response Z B Weighing platform set to zero Z Command cannot be executed Zero set range dropped below Zit Command cannot be executed Zero set range exceeded Comments e Setting to zero is not possible when the weighing platform stabilizes in the zero set range e With some weighing platform types setting to zero deletes a saved tare weight This is indicated with the message TA see section 5 1 4 Changing over to different weight unit Command U _ Unit Change over weight display to different weight unit U Change over weight display to first weight unit Response U B Weight display changed over to different weight unit Comment Possible units mg g kg Ib ozt oz dwt Operating instructions 22008168A 04 10 53 Interface description ID7sx Base Tare Command T Tare weighing platform After the weighing platform stabilizes the current weight value is sav
3. Command T A C Delete tare weight Response TALCA Weighing platform tared with the specified weight TAa Command not carried out Transmit data record Command SX After the weighing platform comes to a standstill transmit a data record with stabile weight values Effect as when ENTER key is pressed S X I Transmit a data record with stabile or dynamic weight values regardless of whether the weighing platform is at a standstill S X I R Repeatedly transmit a data record with stabile or dynamic weight values regardless of whether the weighing platform is at a standstill Response S X S Application block _ __ Application block A No Data record Data record with stabile weight values transmitted S X D Application block Application block A No Data record Data record with dynamic weight values transmitted S X r Command cannot be executed S X Weighing platform in underload range S X Weighing platform in overload range Comments e Number of application blocks three place with preceding zeros The contents of the corresponding application block is contained in the data record see chapter 6 The standard data record consists of 3 blocks S X S A 0 1 1 Gross weight weight value Unit A 0 1 2 Net weight weight value Unit A 0 1 3 _ Tare weight weight value Unit The continuous transmission of data records started with th
4. data e Fail safe storage of configuration data ID7sx Base Technical data and accessories 8 2 Accessories Applications Order No DataPac ID7 sx Basic functions data communication 22 008 442 DosPac ID7sx Basic functions dispensing filling 22 008 441 FormPac XP ID7sx Basic functions formulation dispensing 22 008 440 based upon database incl PC software FormTool XP Weighing platform connections Order No IDNet ID7sx e Connection for a weighing platform 22 008 443 e Max of 2 additional connections possible Point Ex ID7sx e Connection for an analog weighing platform 22 008 443 e Max of 2 additional connections possible Active CL IDNet ID7sx e Connection for a weighing platform 22 008 443 e Connection of a second ID7sx for use as a second display Serial data interfaces Order No CL20mA ID7sx CL 20 mA interface for installation in ID7sx 22 008 444 RS232 ID7sx RS232 interface for installation in ID7sx 22 008 445 Digital inputs outputs Order No 8 I 0 ID7sx 8 digital inputs 8 digital outputs 22 008 446 Network connection Order No Profibus DP ID7sx Field bus module Connection via PSU ID Ex 22 008 649 Memory module Order No Memory ID7sx Paperless archiving of certification relevant weighing 22 008 447 Operating instructions 22008168A 04 10 103 Technical data and accessories ID7sx Base Barcode reader Order No Barcode kit ID7sx barcode scanner II2G EEx ib IIC T4 II2D T 70
5. PROFIBUS DP Configuring Profibus DP ID7 EXP AB AREA Input of up to three expanded application blocks for constants which can be accessed when writing applications blocks Example Input enables access to 021 application blocks 021 001 to 021 999 046 application blocks 046 001 to 046 999 071 application blocks 071 001 to 071 999 TEST MODE Activation of the information display In line 3 and 4 write and read tables are displayed as follows 3 4 5 6 TEST MODE 0 999 kg Id Val 5432109876543210 I Os 2 00 0000 0000000010000000 00 00 1 00 03E7 0100000000000000 08 00 CANCEL 1 Read table 2 Write table 3 Operating mode internal 4 Value hexadecimal 5 Command response bits 6 Inputs outputs hexadecimal Note Once the Profibus settings are complete the Profibus DP ID7sx module must be de energised Only then do the new settings become active 50 Operating instructions and installation information 22008168A 04 10 ID7sx Base Interface description 5 Interface description To exchange data with a computer the ID7sx Base weighing terminal is equipped with an RS232 or CL20mA interface via the PSU power supply unit Both interfaces can be adjusted individually see section 4 5 To operate the serial interfaces in the dialog mode one of the following METTLER TOLEDO command sets must be selected in the master mode e MMR command set see section 5 1 e METTLER TOLEDO Continuous mode see section
6. CL handshake in receiving direction CL handshake in transmission direction CL handshake in both directions CL handshake With the CL handshake 3 types of interface control are possible Handshake in receiving direction in transmitting direction and in both directions After switch on and after each interruption the ID7sx Base attempts to establish the handshake in both directions This type of CL handshake is suitable for data transmission from the ID7sx Base to the computer 1 The ID7sx Base transmits SYN after switch on 2 The computer transmits the character ACK after switch on or after receiving SYN 3 ID7sx Base then sends the response to a command or to a key actuation after each ACK This type of CL handshake is suitable for data transmission from the computer to the ID7sx Base 1 The ID7sx Base transmits SYN after switch on 2 The computer transmits the character SYN after switch on or after receiving SYN 3 ID7sx Base acknowledges the receipt of SYN again with SYN and signals its readiness to receive with ACK 4 Then the computer can transmit a command after each ACK 1 The ID7sx Base transmits SYN after switch on 2 The computer transmits the character SYN after switch on or after receiving SYN 3 ID7sx Base acknowledges the receipt of SYN again with SYN and signals its readiness to receive with ACK 4 The computer signals its readiness to receive with ACK 5 During operation the ID7s
7. 10 Index A Accessories 103 Acoustic signal 58 73 Additional functions 12 Additive tare function 17 Alphanumeric identification 18 Alphanumeric printout 58 75 Application blocks 57 73 74 87 Applications 4 ASCII characters 51 105 Automatic continuous transmission 40 Automatic taring 9 36 101 Autozero 9 36 B Barcode 21 38 46 58 Basic functions 9 Big Weight Display 5 21 22 32 C Certified weighing platforms 6 Change over weight unit 53 73 101 Check mode 32 Checkweighing 13 Classifying 13 Cleaning 8 Command format 63 Command overview 52 64 Commissioning 6 Communication 39 Configuration printouts 42 D Data formats 51 63 Date 32 101 DeltaTrac 12 31 55 74 101 DeltaTrac memory 14 30 DeltaTrac target values 14 55 74 Dialog mode 39 51 Digital inputs outputs 95 103 Display 5 57 67 99 Display Mode 98 Display update 37 Dynamic weighing 15 33 101 106 E Error messages 60 77 96 F Filling 13 Function keys 5 12 99 G GA46 38 Gross 16 101 H Housing 99 l I O test 47 ID code 6 18 ID5 mode 33 ID7 mode 33 Identifications 18 58 101 Interface CL20mA ID7sx 38 Interface connections 38 99 Interface description 51 Interface type 38 Interface8 l O ID7sx 38 Introduction 3 K Keypad 5 99 L Language 31 M Master mode 26 Memory 20 Memory ID7sx 23 38 METTLER TOLEDO continuous mode 41 MMR command set 40 51 Multiplicative tare fu
8. A No Data record Data record with dynamic weight values transmitted SXI Invalid value S x 1 Weighing platform in underload range S X I Weighing platform in overload range Comments e Number of application block three digit with leading zeros The content of the corresponding application block is contained in data record see chapter 6 Standard data record consists of 3 blocks S X A 0 1 1 Gross weight weight value Unit A 0 1 2 Net weight weight value Unit A 0 1 3 Tare weight weight value Unit The continuous transmission of data records started with the S x I R command can be stopped with the S x or s x 1 command Example Command S X T Response Standard data record E XUL AO T E Ero m OI ep BUM Ay Oe eters pa Ok oO Ol eae JJA uS cactus oe popula ere 56 Operating instructions 22008168A 04 10 ID7sx Base Interface description Read application block Command A R No Read content of application block Response A B __ Information Content of application block transmitted Comments e Transmitted information is dependent on application block see chapter 6 e Number of application block must be entered as 3 digits with preceding zeros Write to application block Comm
9. Weighing functions Tare compensation At the press of a button or automatically up to maximum load subtractive Tare target value e For single range scales over entire weighing range subtractive e For multi range scales depending on national calibration regulations e 999 stored tare memories protected against power failure Tare calculation Tare addition tare multiplication sub tare Tare indicator NET lights up with saved tare weight DeltaTrac e Analog display of dynamic measured values e With optical marks for target value and tolerances e 3 selectable applications e 999 DeltaTrac memories protected against power failure Setting to Zero Automatic or manual Gross changeover Display of weight value can be changed over fo gross weight at press of a button Unit changeover Unit can be changed over to weight units kg g mg Ib oz ozt dwt in dependence on national calibration regulations at press of a button Dynamic weighing e Cycle time adjustable from 1 255 cycles e Automatic printout selectable Stabilization detector 4 step with motion indicator Weighing process 3 step adjustment to weighing sample adapter Vibration adapter 3 step adjustment to ambient conditions Identification data e 4 memories for 20 alphanumeric characters can be recalled with keys A to D e Each memory can be assigned a fixed name which can be written in the marking field next to the corresponding key
10. Comments The possible settings are dependent on the connected weighing platform RESET SCALE Reset weighing platform to factory setting WEIGHING PROC ADAPT universal weighing VIBRATION ADAPTER average conditions STABILITY DETECTOR ASD 2 AUTOZERO on AUTOTARA off RESTART off Operating instructions 22008168A 04 10 37 Settings in the master mode ID7sx Base 38 4 5 Select interface connection Select interface type Possible interface types INTERFACE master mode block gt Select the interface connection in the first block COMI COM5 gt Specify the interface type for the selected interface connection COMI COM5 NOT ASSIGNED CL20mA RS232 GA46 BARCODE MEMORY 8 1 0 PROFIBUS DP IDNET When the selected interface connection is not assigned For COMI COM3 only A CL20mA ID7sx interface must be installed on the interface connection for this purpose For other settings see 4 5 1 for COM2 or COM3 only An RS232 ID7sx interface must be installed on the selected interface connection for this purpose For other settings see 4 5 1 For COMI COM3 only For connection of the GA46 6A46 W printer via the PSU power supply unit in the safe area A CL20mA ID7sx or RS232 ID7sx interface must be installed on the interface connection for this purpose The other setting possibilities are described in the operating and installation instructions GA46 This selection no l
11. ID7sx Base Interface description Specify DeltaTrac target value Command D Y Target weight weight value Unit Tolerance 3 Specify DeltaTrac target value D Y Delete DeltaTrac target value Response D B DeltaTrac target value loaded deleted Comments Observe limit values see section 3 1 1 e Also possible A w 0 2 0 see section 6 2 Example Command b vY 4 5 kis 5 LIS Response D B Transmit content of display Command S Transmit a stabilized weight when weighing platform is stabilized S Transmit a stabilized or dynamic weight independent of weighing platform stabilization Response S Weight value _ Unit Stabilized weight value transmitted S D Weight value _ Unit Dynamic weight value transmitted S I Invalid weight S I Weighing platform in underload range SI Weighing platform in overload range Transmit content of display repeatedly Command S I R Transmit stabilized or dynamic weight values after each measuring cycle independent of weighing platform stabilization S R Transmit the next stabilized weight value after a weight change e g different item and one dynamic and the next stabilized weight value after each deflection 30 d S R Deflection weight weight valu
12. MASTER MODE START POS Select start position for entering the master mode NORMAL Selection of the master mode blocks always begins with the TERMINAL block factory setting LAST POSITION When entering the master mode the last block edited is displayed immediately BIG WEIGHT DISPLAY Switch full display indication of the weight on or off Factory setting BIG WEIGHT DISPLAY ON CONTROL MODE Adjust control mode X10 KEY Activation of control mode with X10 key factory setting CONTROL MODE ON This setting is only possible with non certified scales The weighing terminal always operates with the higher resolution 32 Operating instructions 22008168A 04 10 ID7sx Base Settings in the master mode DYNAMIC WEIGHING Set printing during dynamic weighing NO PRINT Results during dynamic weighing are not automatically printed out factory setting AUTO PRINT Each result during dynamic weighing is automatically printed Dynamic weights are marked with Result on the printout ID5 MODE Deactivating or activating downward compatibility with ID5 If ID5 MODE ON is selected the ID7sx Base is operated with downward compatibility to the ID5 Affected settings Text length of identification data 18 characters Text length for keys CODE A D max 18 characters Date time dd mm yy hh mm ss Barcode print command P
13. TYPE FIELD ENTRY LINE DATA alphanumeric character Entry also possible as ASCII code see below AB DATA Number of application blocks to be output xxx The application block can be further specified with the following keys AB EXT For selecting read only memories Xxx yyy SUB BLK For selecting a sub block XXX Z Of XXX yyy Z RANGE For entering a range XXX XXX OF XXX yyy yyy CHRn DATA up to 240 alphanumeric characters Entry also possible as ASCII code see below TEXT DATA Alphanumeric characters DB FIELD Select database field Entry of To enter data or select database fields the EDIT mode must be active DATA parameter press Fp key repeat if necessary until the assignment of the F5 key changes to EDIT 2 Press the EDIT key an input mask appears 3 Enter data in the format and with the keys offered 4 Complete entry with ENTER Enter ASCII code for LINE and CHRn parameters and TEXT 1 Open the entry mask with the EDIT key Press the key and enter the ASCII code numerically 2 3 Complete the numeric entry with the key 4 Complete entry with ENTER Operating instructions 22008168A 04 10 45 Settings in the master mode ID7sx Base 4 5 2 Set barcode reader BARCODE Set barcode reader TYPE DL900 DL910 Select barcode reader DLL6000 When one of the barcode readers is selected the communication and mode para meters for the selected barcode reader are automaticall
14. E FE 4 5 6 eee TTE Imm He oae Yeo Ft AIG IF EJE 2 3 A 5 6 7 8 9 10 CODE A CODE D keys enter identification data 2 FUNCTION CHANGE key display additional functions when entering weight values switch over unit 3 INFO key recall memory contents and system information Function keys F1 F6 the current assignment is shown in the display above the key SCALE key select scale ZERO SET key set scale to Zero test scale TARA key tare scale TARE SPECIFICATION key enter known tare values numerically CLEAR key clear entries and values ENTER key accept and transfer data 2 ow00 40 wo i Numeric keypad with decimal point and signs Operating instructions 22008168A 04 10 5 Introduction and commissioning ID7sx Base 1 4 1 4 1 1 4 2 ID code Certification Commissioning Control drawings and guides for installers The explosion protected weighing system with the ID7sx Base weighing terminal may only be installed according to the respective guide for installers and the accompanying control drawing depending on the power supply unit Component Guide for installers Control drawing ID7sx ME 22008316 PSU ID7sx PSU ME 22006472 ME 220064 78 PSUX 230V ME 22006386 PSUx 230V ID7sx ME 22006397 PSUx 1 20V ME 22006395 PSUx 1 20V ID7sx ME 22006399 Marking and sealing of certified weighing platforms
15. F ADD T Select Adjust parameter Selectfunction STD Accept parameter of function key ADD settings and F5 STD ADD INS return to INS etc EDIT upper level DEL block PRINT SAVE The printout can be edited with function key F5 STD Select standard setting lights up if factory setting selected ADD Adds a new entry at the end of the printout INS Inserts a new entry in front of the displayed entry EDIT Changes into the EDIT mode for the displayed entry to edit the entry DEL Deletes the displayed entry PRINT Creates a key printout SAVE Confirm changes and return to upper level block Operating instructions 22008168A 04 10 27 Settings in the master mode ID7sx Base 28 4 2 3 Example 4 2 4 4 2 5 Orientation in the master mode If only one parameter appears the display shows the last steps in the path of the current master mode block for improved orientation The upper 4 lines of the display show the following path for selecting the COMMUNICATION interface settings MASTERMODE INTERFACES Pa COMI CL 20mA Current selection COMMUNICATION Entries in the master mode The following basic rules apply to entries made in the master mode e Confirm alpha numeric entries with ENTER e Alphanumeric entries with the ID7sx Base see section 3 11 e To accept the displayed value Press ENTER key Emergency entrance into the master mode If a personal code has been assigned for entering the master mode and
16. METTLER TOLEDO SICS command set Communication syntax and formats Identification _ Weight value _ Unit Framing String of 1 10 l3 Cal characters for characters characters specification of number of command characters ad 4 variable characters Identification _ Status _ Weight value _ Unit Framing String of 1 char 10 chaor 3 char left Cpl characters for right justified justified specification filled in with filled in with of response blank char blank char 1 2 char Tare specification command 7 Tare specification response a A 1 3 2 9 5 k a A A 3j e gq249 5 T E E 3 MES The following symbols are used in the command description 10 numbers with sign and decimal point right justified with preceding blank spaces Unit 3 characters left justified with following blank spaces Text n maximum of n characters left justified Weight value The string framing is mandatory however it is not listed in the following command description Enter commands as upper case letters Text to be entered must always be placed in inverted commas Operating instructions 22008168A 04 10 63 Interface description ID7sx Base 5 3 2 Command overview Command Meani
17. On the ID7sx Base weighing terminal memory contents and system information can be recalled 1 Press INFO key Then the following function key assignment appears DELT TARE TEXT ALIBI DATE VERS Display Display tare Display fixed Recall content Display date Display DeltaTrac weight texts and of alibi and time version values name of keys memory numbers of CODE A This selection installed CODE D only appears software when modules Memory ID7sx is installed 2 Select desired information The information is displayed for approx 5 seconds then the ID7sx Base changes to the weighing mode again Notes e When several values are displayed the ID7sx Base automatically changes to the next value after approx 5 seconds e With the CLEAR key it is possible to switch to the next value or back to the weighing mode Operating instructions 22008168A 04 10 19 Additional functions ID7sx Base 20 3 12 1 3 13 Recall memory Press INFO key 2 Enter number of memory and press DELT TARA or TEXT key depending on desired memory Recall name of CODE A CODE D keys Press INFO key 2 Press one of the keys CODE A CODE D The display shows the current Code Print or transfer data If a printer or computer is connected weighing results can be printed out or transferred to the computer In the master mode you can set the following for this purpose e Data to be print
18. With the ID code it can be checked whether certified weighing platforms have been tampered with since the last certification The ID code can be displayed on the terminal at any time see section 3 10 During certification the currently displayed ID code is recorded and sealed During each change to the configuration the displayed ID code increases It then no longer matches the sealed ID code the certification is not longer valid To mark and certify your weighing system please contact METTLER TOLEDO Service or your local board of weights and measures Check certification 1 Display ID code see section 3 10 press ZERO SET key until IDENTCODE is displayed No value is shown for noncertified weighing platforms but instead IDENTCODE 2 Compare ID code displayed with sealed ID code on ID card The certification of the weighing system is only valid when both values are identical Operating instructions 22008168A 04 10 ID7sx Base Introduction and commissioning Table version Max 15 30 60kg Min 0 1ke e 5 10 20g es VA Max 6kg d 1g IDENTCODE S1 2 iM Lr ww m m ww 1 POSEE PP SEP a SN Panel version C IDENTCODE S1 2 zt TSE SESSA SA SS 3 Press ZERO SET key again The connected weighing platform is checked The display shows CHECK SCALE and after the test is completed SCALE IS OK Then the ID7sx B
19. e g with DeltaTrac can be intentionally written with the format percent by setting the sign to 1 e f a non existent block or an alphanumeric block is selected the ID7 responds with ERROR COMMAND The requested data are supplied in the 16 bit integer mode in the same format as ihe weight value and in the 32 bit floating point mode floating point values are always transmitted The application block number in the write table must be entered as a value Word O in 16 bit integer mode Word 1 in 32 bit floating point mode in the following format for the READ APPLICATION BLOCK and WRITE APPLICATION BLOCK commands Basic application block Sub block no Exp Application block number Bit 15 14 13 12 11 10 9 8 7 6 5 4 32 1 0 Example S S S S E E A A AA AA AAA A AB 10 0000000000001 0 0 AB 20 sub block2 O O 1 0 00 000001 01 0 0 Operating instructions and installation information 22008168A 04 10 ID7sx Base Interface description Expanded application block Condition One or more expanded application blocks are selected in the master mode Example Application block 21 is selected as the 1st expanded application block application block 46 is selected as the 2nd expanded application block Sub block no Exp Index of the expanded AB Bit 15 14 13 1211109 8 7 6 5 4 3 2 0 Example S S S SE E A AA AA AA A A AB 21_007 000 0 j0 1 0 000000 I 1 AB 46_005 SB 1 00017
20. following response A B _ MMR resp A W A ESEEENE GICS Write to application block A W No _ Information MMR command set A W No Information SICS command set The weighing terminal receives the command from the computer to write to the No application block This command is contained in the following description of the application blocks in the MMR version A B MMR command set A wW _ A SICS command set The weighing terminal transmits a confirmation to the computer This response is not contained in the following description of the application blocks Command MMR A W 0 2 1 _ 0 0 1J_ 1 2 0 _ k g _ Command SICS AXWI GDO S b ppp E Bee Ol eee c Write to tare memory 1 Response MMR A B Response SICS A W _ A Notes Only those application blocks can be written to for which the corresponding AW command is listed in the following description An application block can consist of one or more sub blocks and the numbering of the sub blocks begins with 1 The sub blocks of an application block can each contain a maximum of 20 characters The sub blocks are separated with or Hz ASCII character O9 hex 9 deci A W No _ Sub block 1 Sub block 2 Sub block n MMR resp A W No Sub block 1 Sub block
21. the weighing platform is at a standstill S I R Transmit a stabile or a dynamic weight value after each measuring cycle regardless of whether the weighing platform is at a standstill Response S _ S _ Weight value _ Unit Stabile weight value transmitted S _ D _ Weight value _ Unit Dynamic weight value transmitted SJ Invalid value s Weighing platform in underload range S Weighing platform in overload range Comment Stop S 1 R command with s S 1 SiR command or disconnect port 66 Operating instructions 22008168A 04 10 ID7sx Base Interface description Set to zero Command Z Set gross weight display to zero after weighing platform comes to a standstill effect as when ZERO SET key is pressed Response Z _ A Weighing platform set to zero z i Command cannot be executed e g standstill not achieved or another command is currently being executed z l Command cannot be executed Zero set range dropped below Z _ Command cannot be executed Zero set range exceeded Comment Can only be set to zero when the weighing platform comes to a standstill in the zero set range Reset Command Reset weighing terminal to the state maintained after Power On Response I 4 _ A _ text Serial number Comments e Al
22. 0 Stability detector switched off only possible with non certified weighing platforms ASD 1 fast display good reproducibility ASD 2 A W factory setting ASD 3 A v ASD 4 slow display very good reproducibility AUTOZERO Switch automatic zero point correction on or off The automatic zero point correction corrects the weight of minor dirt with the weighing platform unloaded Factory setting AUTOZERO ON Comment On certified weighing platforms the zero point correction is always switched on AUTOTARA Switch automatic taring on or off Factory setting AUTOTARA OFF 36 Operating instructions 22008168A 04 10 ID7sx Base Settings in the master mode RESTART Switch restart function on or off When RESTART ON is set the zero point and tare value remain stored after the power supply is interrupted When the weighing platform is switched on again the terminal shows the current weight Factory setting RESTART OFF SECOND UNIT Select second weight unit Possible units g kg Ib oz ozt dwt Unit Abbreviation Conversion to g Kilogram kg 1000 g Pound Ib 453 59237 g Ounce oz 28 349523125 g Troy Ounce ozt 31 1034768 g Pennyweight dwt 555173843 g Gram g eru Milligram mg 0 001 g Comment On certified weighing platforms only the units permitted by certification appear DISPLAY UPDATE Set display speed of the weight display Select number of updates per second UPS
23. 1 1 to 6 1 3 e A blank space must be entered between AR or AW and the application block number E g A R _ No e The command identification is repeated in the response and a blank space and the character A added A R A Information application block transmitted and A W A application block written e Texts entered or transmitted are always in inverted commas Read application block for CODE A Command A R 0 9 4 Response A R A Article Write application block for CODE A Command A2 w 0 9 4 Article Response A W _ A e Operating instructions 22008168A 04 10 89 Application blocks ID7sx Base 6 2 TERMINAL SCALE application blocks No Content Format 001 Terminal type Response A B _ M e t t 1 r T 0 1 4 0 _ 1 D 7 8 x 002 Program number Response A B _ 1 T 0 7 0 0 x x x _ 004 Serial number Response A B Number 7 006 Transfer key Response A B Kevys 2 4 Write A W 0 0 6 2 4 007 Current gross weight Response A B Weight value _ Unit 2nd weight unit 008 Curre
24. 127 deci see section 9 1 Operating instructions 22008168A 04 10 51 Interface description ID7sx Base 5 1 2 Command overview Command Meaning Page RO RI Switch keypad on off b3 Z Set weight display to zero after weighing platform stabilization 53 U_ Change over terminal to a different weight unit 53 T Tare 54 To Specify tare weight 54 DY_ Specify DeltaTrac target value 55 S Transmit in case of weighing platform stabilization 55 SI Transmit independent of weighing platform stabilization 55 SIR Transmit repeatedly independent of weighing platform stabilization 55 SR Transmit stabilized weight values repeatedly depending on a weight 55 change SR_ Transmit repeatedly depending on weighing platform stabilization with 55 specification of an excursion value SX Transmit data record after weighing platform stabilization 56 SXI Transmit data record independent of weighing platform stabilization 56 SXIR Transmit data record repeatedly independent of weighing platform 56 stabilization ARNO Read information of application block 57 AWNo _ Write to application block 57 D_ Write to display 57 Pa Print alphanumeric characters or barcodes on the GA46 58 58 DS Trigger acoustic signal 58 ID Interrogate terminal identification 58 W_ Actuating digital outputs 59 52 Operating instructions 22008168A 04 10 ID7sx Base Interface description
25. 2 Sub block n SICS Extensive application blocks are displayed so that each sub block begins in a new line To write to individual sub blocks enter the corresponding number of characters If only sub block 1 is written to the characters are eliminated e g sub block 3 written to A w No Sub block 3 MMR resp A W No Sub block 3 SICS Operating instructions 22008168A 04 10 ID7sx Base Application blocks 6 1 3 Example Data formats e n the following description of the application blocks the following data formats are used Weight value 10 digits with sign and decimal point right justified with preceding blank space Unit 3 characters left justified with following blank spaces Number n Number n digits right justified with preceding blank spaces Text n maximum of n characters If the SICS command set is used Text must always be placed in inverted commas Conclude commands and responses with the string frame Cpl ASCII characters Cp OD hex 13 deci Le OA hex 10 deci The string frame is not contained in the following description Read and write application blocks with the SICS command set In the following description the application blocks are shown in the syntax for the MMR command set When used with the SICS command set please observe the following SICS conventions also see sections 6
26. 999 2 Enter text for the selected memory max of 20 characters 3 To load additional fixed texts repeat the first two steps 4 End entry Confirm FIXED TEXT NO without entry with ENTER CLEAR ALL TEXTS Delete all text memories Comment Fixed Text No 20 is displayed during switch on and with a restart DELTA FIX Save target weight tolerance combinations in DeltaTrac memory LOAD DELTA FIX Enter memory number of DELTA FIX No 1 999 Enter target weight DELTA in the displayed unit Enter tolerance TOL in To enter additional Delta Fix repeat the first three steps Or de cy mo End entry Confirm memory number without entry with ENTER CLEAR ALL DELTA Delete all DeltaTrac memories Comment With the FUNCTION CHANGE key you can select the weight unit for entering the DeltaTrac target and tolerance values 30 Operating instructions 22008168A 04 10 ID7sx Base Settings in the master mode DELTATRAC Set DeltaTrac application TYPE Select DeltaTrac application FILLING Weigh in target weight within a tolerance range factory setting CLASSIFYING Evaluate the test samples as good too light or too heavy based on the target weight and tolerance CHECKWEIGHING Determine difference between target and actual weight AUTO PRINT WITHIN Automatic printout when actual weight lies within the specified tolerance TOL PRINT ONLY WITHIN Printout only when actual va
27. A B _ Parameters for Scale 2 Note For service information purposes the internal scale parameters can be read oult printed ihe structure and content are scale dependent 603 Parameters for Scale 3 Response A B Parameters for Scale 3 Note For service information purposes the internal scale parameters can be read oult printed ihe structure and content are scale dependent Operating instructions 22008168A 04 10 93 Application blocks ID7sx Base 6 3 INTERFACE application blocks Application blocks are reserved for the possible interface connections These application blocks can only be read and written to when an interface is actually installed on the interface connection concerned 6 3 1 Serial interfaces No Content Format 101 Description of application Response ID7sx Interfaces 102 Program designation Response IKY7 0 0100 103 Transmit buffer COMI Response Transmit buffer COM Write 0 3 Information 104 Transmit buffer COM2 Response Transmit buffer COM2 Write 0 4 Information 201 Description of application Response ID7sx Interfaces 202 Program designation Response IKY07 0 0100 203 Transmit buffer COM3 Response Transmit buffer COM3 Write 0 3 Information 204 Transmit buffer COMA Response Tran
28. C 22 008 640 complete with RS232 ID7sx and Viper Ex SW slotcard Other accessories Order No ID7sx installation set for cabinet installation 22 008 439 Wall bracket Completely rust proof 00 504 130 Floor stand Completely rust proof 00 504 132 Stand socket Completely rust proof 00 503 701 Bracket stand Completely rust proof 00 504 128 104 Operating instructions 22008168A 04 10 ID7sx Base Appendix 9 Appendix 9 1 ASCII table hex deci ASCII hex deci ASCII hex deci ASCII hex deci ASCII hex deci ASCII US US US US US 00 0 NUL 34 52 4 68 104 h 9C 156 DO 208 L 01 1 SOH 35 53 5 69 105 i 9D 157 DI 209 02 2 STX 36 54 6 6A 106 j 9E 158 B D2 210 L 03 3 ETX 37 55 7 6B 107 k 9F 159 f D3 211 L 04 4 EOT 38 56 8 6C 108 AO 160 D4 212 E 05 5 ENQ 39 57 9 6D 109 m Al 161 D5 213 F 06 6 ACK 3A 58 6E 110 n A2 162 6 D6 214 T 07 7 BEL 3B 59 3 6F 111 o A3 163 D7 215 08 8 BS 3C 60 lt 70 112 p A4 164 fi D8 216 09 9 HT 3D 6 71 113 q A5 165 D9 217 J OA 10 LF 3E 62 gt 72 114 r A6 166 a DA 218 r OB 11 VT 3F 63 73 115 S A7 167 e DB 219 i Oc 12 FF 40 64 74 116 t A8 168 DC 220 OD 13 CR 41 65 A 75 117 u A9 169 DD 221 I OE 14 SO 42 66 B 76 118 V AA 170 DE 222 I OF 15 SI 43 67 C 77 119 w AB 171 Vo DF 223 m 10 16 DLE 44 68 D 78 120 X AC 172 A EO 224 a 11 17 DC1 45 69 E 79 121 y AD 173 i El 225 B 12 18 DC2 46 70 F 7A 122 Z AE 174 E2 226 T 13 19 DC3 47 7
29. Ex system solution Analog scale In this configuration the following possibilities are available Multi scale operation with up to 3 weighing platforms K x D x or Point Ex system solution e Up to 3 data interfaces and one 1 0 interface for printing for data exchange with a computer for control e g of valves or flaps e Memory module for storing the individual configuration data as well as weighing data 1 2 2 ID7sx Base with PSUx power supply unit Hazardous area Safe area om reo p I z L i Only one weighing platform K x D x with Point Ex A D converter or Point Ex System solution is permissible in this configuration 4 Operating instructions 22008168A 04 10 ID7sx Base Introduction and commissioning 1 3 1D7sx Base weighing terminal 1 3 1 Display 0 005 3 NET 5 DELT DYN UNIT x10 BRUT MODE 6 1 Weight display with sign and decimal point Stability monitor lights up until the weighing platform has levelled out then the weight unit appears here 3 Range display for multi range weighing platforms 4 Number of the weighing platform shows the weighting platform just selected 5 NET symbol for marking net weight values 6 Assignment of the function keys 1 3 2 Keypad
30. Interface description ID7sx Base Transmit stabile weight values repeatedly depending on a weight change Command S R Excursion weight weight value Unit After a weight change greater than the specified excursion weight transmit alternately the next stabile weight value and a dynamic weight value depending on the specified excursion S R If no excursion weight is entered the weight change must be at least 12 5 of the last stabile weight value however at least 30 d Response S S Weight value _ Unit Current stabile weight value transmitted Weight change S D Weight value _ Unit Dynamic weight value transmitted S r Command cannot be executed S _ L Command understood parameters defective s Weighing platform in underload range S Weighing platform in overload range Comment Stop command with command s S 1 S 1 R Q or disconnect the port Example Command S R 1 4 0 k g Responses sS S 2 040 0 0 k g lst item s _ D Sud 54 9 PO ba LS s _ s 4 1 0 5 0 _ k g 2nd item Taring Command T Tare weighing platform After the weighing platform comes to a standstill the current weight value is saved as a tare weight and the weight display set to zero with the weight on the platform Effect a
31. TARE key The tare weight is saved and the weight display set to zero The display shows the NET symbol Notes e When the weighing platform is relieved the saved tare weight is displayed with a negative sign e The weighing platform only saves one tare value Automatic taring Prerequisite AUTOTARA ON must be set in the master mode see section 4 4 gt Place empty container on scale The container weight is automatically saved and the weight display set to zero The display shows the NET symbol Note When the weighing platform is relieved the saved tare weight is cleared Operating instructions 22008168A 04 10 9 Basic functions ID7sx Base 2 2 3 Enter numerically Correct entry Copy tare constant 2 2 4 2 2 5 2 3 Specify tare weight 1 Press TARE SPECIFICATION key 2 Enter tare weight container weight and confirm with ENTER When weighing platform is relieved the entered tare weight is displayed with a negative sign Note With the FUNCTION CHANGE key you can select the weight unit for entering the tare weight gt Clear the entry character by character with the CLEAR key and repeat correctly The ID7sx Base has 999 tare memories for frequently used tare weights programmed in the master mode Enter memory number 1 999 2 Press TARE SPECIFICATION key The display shows the NET symbol and the net weight based on the recalled tare weight Recall currently saved tare weig
32. e 999 memories for frequently used identification data Info function Displays of current weighing data identification data and memories at the press of a button Date and time e For printout or output via the data interface e Quariz controlled 12 or 24 hour display automatic calendar function Europe or US format protected against power failures Operating instructions 22008168A 04 10 101 Technical data and accessories ID7sx Base Profibus DP ID7sx Housing aluminium DIN locking foot Dimensions 42 mm x 106 mm x 115 mm W x H x D Weight 2 400 g Protection type IP20 Supply 24 V DC 20 96 Current consumption 100 mA at 24 V DC CPU core 80C52 32 K RAM 128 K flash Displays 2 status LEDs UART interface RS232 UART buffer max 15 KB Temperature range 0 50 C Max humidity 90 non condensing Profibus DP connection 9 pin Sub D Electrical isolation DP interface and 24 V DC Baud rate up to 12 Mbit s Node address adjustable to between O and 126 in the master mode Factory setting 126 Data width 2 4 IN and 2 4 OUT words consistent for 2 words at operating mode 16 bit integer or for 4 words at operating mode 32 bit floating point Status indication 2 LEDs 102 Operating instructions 22008168A 04 10
33. information 22008168A 04 10 81 Interface description ID7sx Base 82 Read commands Reading and writing application blocks e The read commands Display value Net Gross Tare Key and Application block overwrite the cyclically transmitted display values with the required data The data are transmitted as 16 bit integers or 32 bit floating points As soon as the COMMAND EXECUTED bit is toggled these values must be evaluated immediately by the PLC as in the next cycle the value in the read table is overwritten again with the current weight value e The response to the READ KEY NUMBER command write table bits 1 2 3 0 0 1 is transmitted in the Word O 16 bit integer or in Word 1 82 bit floating point The low byte contains the keyboard code the high byte the function key code The ID7 can store a maximum of 10 keys for being called via the READ KEY NUMBER command If they are not called the oldest key actuations are overwritten After reading out the last stored key the KEY WAS PRESSED bit is reset The key memory is cleared after the device is switched on and after the mastermode is exited e When writing an application block the desired data are simultaneously transferred with Word 3 For this reason writing application blocks is only possible in 16 bit integer 4 word mode Only application blocks with the formats numeric or weight value can be read or written When writing certain tolerance sub blocks
34. is dependent on the application block see chapter 6 e The number of the application block must be entered as a three place number with preceding zeros Operating instructions 22008168A 04 10 73 Interface description ID7sx Base Write application block Command A W No Information Write application block AW No Reset application block A W Ne Delete application block Response AW A Application block written AJW I Application block not present A W L Application block cannot be written Comments e The information to be entered is dependent on the target block see chapter 6 Deleting and resetting have the same effect Specify DeltaTrac target value Command D Y Target weight weight value Unit _ Tolerance Specify DeltaTrac target value D Y Delete DeltaTrac target value Response D Y _ A DeltaTrac target value loaded deleted Comments Observe limit values see section 3 1 1 e Also possible Aw 0 2 0 see section 6 2 Example Command D Y _ 4 5 _ k g _ 5 _ 3 Response D Y _ A 74 Operating instructions 22008168A 04 10 ID7sx Base Interface description Print text or barcode with GA46 printer safe area Command __ Text_48 TETE Print text as per setting Text 48 Print text in small print Text 48 Pr
35. mode These operating modes are suitable for continuous data transmission in real time from the ID7sx Base to METTLER TOLEDO devices e g to a second display The data are even transmitted when the weighing platform is moving or the gross weight O Commands can also be sent to the ID7sx Base weighing terminal permitting remote control of certain keys on the terminal There are 2 different continuous modes e Continuous mode net and tare values are continuously transmitted e Short continuous mode only net values are continuously transmitted 5 2 1 Data output from ID7sx Base Output format Weight values are always transmitted in the following format STX SBI SB2 SB3 DFI DF2 CR CHK STX ASCII characters 02 hex 2 deci character for start of text is required by some printers SB For status bytes see below DFI Data field with 6 digits for the weight value transmitted without a decimal point and unit DF2 Data field with 6 digits for the tare weight is not transmitted in the short continuous mode CR Carriage return ASCII character OD hex 13 deci CHK Checksum 2 part complement of binary sum of 7 lower bits of all previously transmitted characters including STX and CR Status byte SB1 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0 0 Rounding Increment Decimal position Bit4 Bit3 Roundin
36. module With the Memory ID7sx module you can fulfill your recording obligations in certified operation without having to archive paper Memory ID7sx automatically assigns each weighing with consecutive data record number which also appears on the print out saves the net and tare value the date and fime Memory ID7sx operates according to the principle of a ring memory When the capacity limit of approx 700000 data records is reached the oldest data record is deleted and overwritten with data from the latest weighing By entering suitable search criteria you can quickly access the data of a very specific weighing Condition Memory ID7sx installed and COM4 configured as Memory see Section 4 5 Initiate gt Press INFO ALIBI key sequence The function keys change to the following assignment FIND gt gt lt gt Num END Enter search Search for Display data Display Search for Exit Info Alibi criteria next record of directly data record and return to matching weighing following with known normal mode data record carried out weighing data record starting with directly number oldest beforehand Fast search with entry of data record number 1 Press gt Num key 2 Enter number of data record to be searched for and confirm with ENTER Memory ID7sx now searches for the desired data record Notes e The search may take up to 10 seconds e f no data record with the entered number is found the me
37. the DeltaTrac memories 046 001 046 025 071 001 Text memory Response A B Text 20 pes 1 999 Write AjW O x x x x x Text 20 071 999 Comment xx 71 001 71 999 071 Text memory 1 20 Response A B Text 20 Ges Write A W 0O x x Text 20 090 Comment xx 71 90 The contents of the text memories 1 20 are identical to the contents of the text memories 071_001 071_020 Operating instructions 22008168A 04 10 91 Application blocks ID7sx Base No Content Format 091 Barcode EAN 28 Response A B j EAN28 EAN 128 O1 EAN 128 310 EAN 128 EAN 128 330 EAN 28 2 8 Article Check digit Weight Article 4 digit article no from memory Code A Check digit 1 digit calculated by ID7sx Base for the weight Weight 5 digit positive weight value with 3 decimal places between 00 000 kg 99 999 kg EAN 128 01 0 1 Article or 0 1 Article Check digit Or 0 1 0 Article Check digit or 0 1 0 Article Article Article no from memory Code A max 14 digits Check digit 1 digit calculated by ID7sx Base Length total of max 16 digits EAN 128 310 0 1 9 Article Check digit 3 1 0 x Weight or 0 1 9 Article 3 1 0 x Weight Article Article no from memory Code A max 12 or 13 digits Check dig
38. the Profibus DP ID7sx module show the following status LED Status Meaning green switched on e DP data cycles active switched off e DP data cycles inactive DP connection cable not plugged or data cycles not started by the DP master yellow blinking very fast e Bootloader active after every switch on for approx 2 seconds Switched off e UART data exchange to ID7sx active Profibus Short flash impulses every second active switched on e UART data exchange to ID7sx faulty or not Short off impulses every second active Profibus active Switched off e serial data exchange to ID7sx active Profibus short flash impulses every 3 seconds inactive switched on e serial data exchange to ID7sx faulty or not short off impulses every 3 seconds active Profibus inactive 86 Operating instructions and installation information 22008168A 04 10 ID7sx Base Application blocks 6 Application blocks 6 1 6 1 1 Read Response Example Application blocks are internal information memories in which weighing data calculated quantities configuration data or character sequences entered with the keypad are stored The content of the application blocks can be read out or written to with a computer When the GA46 printer is connected via the PSU power supply unit in the safe area the assignment of the application blocks can be printed out see operating instructions for the GA46 printer Syntax and formats
39. to another METTLER TOLEDO weighing terminal for use as a second display Conversely another METTLER TOLEDO weighing terminal can be connected to the ID7sx weighing terminal for use as a second display Both terminals must support the AUTO DIR or TOLEDO CONTINUOUS protocols for this ID7sx used as a second display of another ID7sx e An Active CL IDNet module must be installed in the ID7sx weighing terminal used as the second display e Connection in accordance with terminal diagram 22006478 Sheet 3 see PSU guide for installers 22006472 Index C or greater The AUTO DIR operating mode must be set for the COM port used in the ID7sx serving as the weighing terminal No special settings are required in the second display device e With ID7sx as a second display the weight value fills the entire display BIG WEIGHT display ON Operating instructions 22008168A 04 10 21 Additional functions ID7sx Base 22 3 15 2 3 15 3 3 15 4 ID7sx used as d second display on a weighing terminal in the safe area COMI of the ID7sx weighing terminal must be lead to the safe area via a slotcard interface in the PSU See terminal diagram 22006478 Sheet 2 in PSU guide for installers 22006472 The weighing terminal in the safe area must be connected to the CL output of the PSU power supply via one of its COM ports Ensure that the CL interface configured as active in the weighing terminal is configured as passive in the PSU IDNET
40. used must comply with the specifications contained in the installation instructions Non compliant equipment jeopardises the intrinsic safety of the system cancels the Ex approval and renders any warranty or product liability claims null and void The safety of the weighing system is only guaranteed when the weighing system is operated installed and maintained in accordance with the respective instructions Also comply with the following the instructions for the system modules the regulations and standards in the respective country the statutory requirement for electrical equipment installed in hazardous areas in the respective country all instructions related to safety issued by the owner The explosion protected weighing system must be checked to ensure compliance with the requirements for safety before being put into service for the first time following any service work and every 3 years at least Prevent the build up of static electricity Always wear suitable working clothes when operating or performing service work in a hazardous area Do not use protective coverings for the device Avoid damage to the system components Operating instructions 22008168A 04 10 3 Introduction and commissioning ID7sx Base 1 2 Applications The ID7sx Base weighing terminal can be operated with the PSU or PSUx power supply unit 1 2 ID7sx Base with PSU power supply unit Hazardous area Safe area Point
41. you have forgotten your code you can still enter the master mode gt Enter the character sequence C L E A R as your personal code Operating instructions 22008168A 04 10 ID7sx Base Settings in the master mode 4 3 TERMINAL master mode block 4 3 1 Overview of the TERMINAL master mode block In the TERMINAL master mode block you enter the following system settings EN CH p P DISPLAY DISPLAY ON DISPLAY OFF DISPLAY ERROR INFO DURATION MESSAGES MESSAGES DISPLAY CONTRAST RESET TERMINAL Legend Blocks highlighted in grey are described in detail in the following e Factory settings are printed in bold print Operating instructions 22008168A 04 10 29 Settings in the master mode ID7sx Base 4 3 2 Settings in the TERMINAL master mode block FIX TARE Save tare values protected against power failure to tare memories LOAD FIX TARE Enter memory number of FIX TARE NO 1 999 2 Enter tare weight for the selected memory in the displayed unit 3 To load additional fixed tare values repeat the first two steps 4 End entry Confirm FIX TARE NO without entry with ENTER CLEAR ALL TARES Delete all tare memories FIXED TEXT Save texts protected against power failure to text memories These texts can be assigned for example as identifications or also output during printing LOAD FIXED TEXT Enter memory number of FIXED TEXT NO 1
42. 0 1200 2400 4800 9600 19200 baud MODE Set operating mode STANDARD Set operating mode to factory setting SETTING MMR dialog mode no handshake no auto transmission no continuous transmission transfer string Standard string framing CpLe DIALOG MODE For dialog between ID 7sx Base weighing terminal and computer For other settings see next section PRINT MODE To print weighing data e g on a form printer For other settings see page 42 Operating instructions 22008168A 04 10 39 Settings in the master mode ID7sx Base Set dialog mode STRING FRAMING DIALOG MODE Set dialog between ID7sx Base weighing terminal and computer MMR For information on dialog mode with the MMR command set see section 5 1 HANDSHAKE Possible settings e NO HANDSHAKE e CL HANDSHAKE for additional information on the CL handshake see page 41 e XON XOFF PROTOCOL AUTOMATIC Possible settings CONTINUOUS e NO AUTO TRANSMISSION TRANSMISSION M sed e AUTO SIR after each measuring cycle a stabilized or dynamic weight is transmitted AUTO SR after each weight change which is greater than the set value a motionless weight value and then a dynamic weight value are sent AUTO DIR weight values are transmitted as with AUTO SIR and additionally the Special characters in the display are transmitted for a second display Fixed communications parameters 9600 baud 7 data bits 2 stop bits parity even AUTO XIR ex
43. 1 0 0 000000 I 1 Input of tolerances in If the sign bit 15 in Word 3 is set to 1 tolerance specifications can be written accurately down to one decimal place in 96 This rule applies in the same way for Word O 16 bit integer and Word 1 32 bit floating point when reading Example Decimal Binary 1514 131211109 8 7 6 5 4 3 2 0 100 0 1000 100000 1 1 7 1 10 1 0 0 1 10 1001 10 00000 0 1 0 0 0 1 96 10000000000000 Write commands The write command PRESS KEY requires the low byte keyboard code and the high byte function key code as parameters e The function key code is based on the active function keys and must be correctly specified for each PRESS KEY command A function key change can also auto matically be forced by changing the function key code e g from REF 10 3301 hex to X10 0004 hex e The setpoints loaded via the WRITE SETPOINT X commands e g Setpoint 1 write table bits 4 5 6 O O 1 are deleted after switch on and each time the mastermode is run The Tolerance parameter in the setpoint modes Checking and Filling must be specified in the 16 bit integer mode with 2 decimal places e g 1025 for 10 25 Operating instructions and installation information 22008168A 04 10 83 Interface description ID7sx Base Keyboard codes Key Code Dec Code Hex Key Code De
44. 1 6 7B 123 AF 175 E3 227 T 14 20 DC4 48 72 H 7C 124 BO 176 E4 228 X 15 21 NAK 49 73 7D 125 Bl 177 E5 229 o 16 22 SYN 4A 74 J 7E 126 B2 178 i E6 230 u 17 23 ETB 4B 75 K 7F 127 4 B3 179 E7 231 T 19 25 EM 4D 77 M 81 129 B5 181 4 EQ 233 8 1A 26 SUB 4E 78 N 82 130 B6 182 EA 234 Q 1B 27 ESC 4F 79 O0 83 131 B7 183 1 EB 235 5 1C 28 FS 50 80 P 84 132 a B8 184 J EC 236 co 1D 29 GS 51 81 Q 85 133 B9 185 J ED 237 1E 30 RS 52 82 R 86 134 BA 186 EE 238 E 1F 31 US 53 83 S 87 135 C BB 187 3 EF 239 N 20 32 SP 54 84 T 88 136 BC 188 1 FO 240 21 33 55 85 U 89 137 BD 189 4 Fl 241 22 34 56 86 V 8A 138 BE 190 d F2 242 gt 23 35 57 87 W 8B 139 i BF 191 F3 243 lt 24 36 S 58 88 X 8C 140 i CO 192 L FA 244 f 25 37 59 89 Y 8D 141 i C1 193 L F5 245 26 38 amp 5A 90 Z 8E 142 C2 194 T F6 246 i 27 39 i 5B 91 8F 143 A C3 195 H F7 247 x 28 40 5C 92 90 144 C4 196 F8 248 29 4 5D 93 91 145 ce C5 197 F9 249 2A 42 5E 94 92 146 C6 198 F FA 250 2B 43 5F 95 03 147 C7 199 I FB 251 Y 2C 44 60 96 94 148 C8 200 L FC 252 n 2D 45 61 97 a 95 149 c9 201 F FD 253 2E 46 62 98 b 96 150 CA 202 JL FE 254 2F 47 63 99 C 97 151 CB 203 T FF 255 30 48 0 64 100 d 98 152 y CC 204 LE 31 49 65 101 e 99 153 0 CD 205 32 50 2 66 102 f 9A 154 CE 206 dL 33 51 3 67 103 g 9B 155 e CF 207 2 Operating instructions 22008168A 04 10 105 Index ID7sx Base
45. 1 Enter identification An identification may contain a maximum of 30 characters Enter 1 Press one of the keys CODE A CODE D numerical identification 2 Enter identification data Code A Code D via the numeric keypad and confirm with ENTER Enter 1 Press one of the keys CODE A CODE D alphanumeric The function keys are given the following assignment identification ABCDE FGHIJ KLMNO PQRST UVWXY Z Selection of Selection of Selection of Selection of Selection of Selection of letters A to E letters F to J letters K to O letters P to T letters U to Y letter Z and special characters 18 Operating instructions 22008168A 04 10 ID7sx Base Additional functions Recall fixed text memory 3 11 2 3 12 2 Select desired group of letters e g press KLMNO key 3 Select desired letter The display changes again to the above selection 4 Repeat entry in steps 2 and 3 for additional characters Note Letters and numbers can be combined as desired The ID7sx Base weighing terminal is equipped with 999 memories for fixed texts which can be programmed in the master mode and used as identifications 1 Enter memory number 1 999 2 Press a key CODE A CODE D The saved fixed text is now assigned to the selected identification Code A Code D Clear identifications gt Press desired key CODE A CODE D and clear memory content with CLEAR key Recall information
46. 5 2 e METTLER TOLEDO SICS command set see section 5 3 5 1 MMR command set 5 1 Syntax and formats of communication Command format when transmitting Identification Weight value _ Unit Framing weight formats Char sequence for 1 8 digits 1 3 chor Definable in specification of number of digits number of master mode command variable characters factory setting 1 4 char variable Cal Response format when transmitting Identification Weight value _ Unit Framing weight formats Char sequence for 10 digits right 3 chor definable in specification of justified left justified master mode response filled out with filled out with factory setting 2 3 char blank spaces blank spaces Cale Example Command Tare specification TE E SERES 2 SIKES Response Tare specification T B HI_ _ _ _ _ 1 3 2 9 5 _ k g _ Data formats The following symbols are used in the following command description Weight value 10 characters with sign and decimal point right justified with preceding blank spaces Unit 3 characters left justified with following blank spaces Text_n maximum of n characters left justified e The string framing is mandatory however it is not contained in the following command description e Enter commands as ASCII characters The following ASCII characters are available 20 hex 32 deci 7F hex
47. 68A 04 10 79 Interface description ID7sx Base 5 4 Commands and responses All commands available to the PLC and the corresponding responses are shown in ihe following two tables Data direction PLC gt ID7sx Write table Data direction ID7sx gt PLC Read table Write table 16 Bit Integer Word 0 Word 1 2 Words 16 Bit Integer Word 0 Word 1 Word 2 Word 3 4 Words 32 Bit Floating Word O Word1 Word 2 Word 3 Point Bit Value Value Command 16 Digital 1 0 AB data 16 Bit 32 Bit Floating Point IEEE 754 0 Sign Command valid Toggle bit for all commands Bits 1 2 3 Selection of read table value read write AB 0 0 0 Display 1 0 0 Net 2 0 0 1 Key No 1 0 1 Read AB 0 1 0 Gross 1 1 0 Tare 3 0 1 1 Write AB 1 1 1 Not in use 4 Bits 4 5 6 Selection of write table val ae el its Selection of write table value ID7sx outputs writing on Exponent 0 0 0 Empty 1 0 0 Tare specification application 5 0 0 1 Setpoint 1 1 0 1 Setpoint 2 block 0 1 0 Key No 1 1 0 Fixed Text No 6 0 1 1 Setpoint 3 1 1 1 Setpoint 4 or 7 Tarin Tolerance Mantissa g specifica 3 i tions are 8 Delete tare Displaying or rondes 9 Set fo zero evaluating eens inputs of M set 10 ENTER key external I O 11 Input mode module 12 Mantissa Switch keyboard on off 13 Reserved 14 Bits 14 15 Selection of weighing
48. G POINT 4W Al POINT 32 BIT FLOATING POINT 4W AO SETPOINT MODE Set type and use of setpoint UNIVERSAL Each setpoint can be set and read independently of others CHECKWEIGHING As soon as setpoints 1 and 2 are set DeltaTrac CHECKWEIGHING will be activated with SP1 setpoint and SP2 tolerance in in 16 bit integer mode with 2 decimal places In read table current state BELOW SP1 GOOD SP2 or ABOVE SP3 can be read off FILLING As soon as setpoints 1 and 2 are set DeltaTrac CHECKWEIGHING will be activated with SP1 setpoint and SP2 tolerance in in 16 bit integer mode with 2 decimal places In addition SP3 and SP4 can also be loaded as any desired setpoints In read table current state GOOD SP 1 ABOVE SP2 SP3 REACHED SP3 or SP4 REACHED SP4 can be read off INPUT MODE Set request for identification data in Input mode After setting the user data command INPUT MODE in the write table the selected request for input is automatically carried out and the entries are saved in the application blocks 094 to 097 The user data response INPUT MODE RUNNING remains set while the input mode is active A Code A is requested A B Code B and Code A are always requested A B C Code C Code B and Code A are always requested A B C D Code D Code C Code B and Code A are always requested Operating instructions and installation information 22008168A 04 10 49 Settings in the master mode ID7sx Base
49. ICSO commands x3 Version or implemented SICS1 commands x4 Version or implemented SICS2 commands xb Version or implemented SICS3 commands des Command understood cannot be executed at this time Comments e On the SICS level only fully implemented levels are executed e With the SICS version all levels are specified Transmit scale data Command I Transmit data from weighing terminal and weighing platform s Response I 2 A text Example I 2 A ID7sx Base IZ 18 32 000 kg Operating instructions 22008168A 04 10 65 Interface description ID7sx Base Transmit scale software version Command 1 3 Transmit software version from weighing terminal and weighing platform s Response 3 _ A _ text Example 3 A IPYA 0 0100 1Z19 0 0103 Transmit serial number Command 4 Transmit serial number of weighing terminal Response 4 _ A _ text Example 4 _ A _ 1234567 Comment The response to 14 appears automatically following switch on and after the Reset command Transmit display contents Command S Transmit a stabile weight value when the weighing platform is at a standstill 8 1 Transmit a stabile or a dynamic weight value regardless of whether
50. METTLER TOLEDO MultiRange METTL ID7sx Base weighing terminal These operating instructions 22008168A describe the following components ID7sx Base IDNet ID7sx 1 x standard CL20mA ID7sx 1 x standard RS232 ID7sx 8 l O ID7sx Memory ID7sx Profibus DP ID7sx ID7sx Base Contents Contents Page 1 Introduction and commissioning eese 3 1 1 Safety precautions o petiere ac Pete donat etra 2g 3 1 2 ADDIICGTIOTIS s iri ensi ibeo tede ete eost caet eode nada deor daten teet ri ctt 4 1 3 ID7sx Base weighing terminal ssssssssem ee eeeee b 1 4 COMMISSIONING sss nasodp tied fep nisse nis Ld pop HORE loei toti Atem voies 6 1 5 op TP 8 2 Basic FUNCTIONS ceci co oo ri t erret re Erie Fu yin ERR ER uS 9 2 1 Setting NO Zero se chs tats ince a eto cd e ved tb dto 9 2 2 TAN i e a O E a ias OAA A a 9 2 3 S 6 aT o o POPRATITI 10 24 Switch over welghirig pIGTO TUL ssa ter er t eet dec 11 3 Additional functions eeeeeeeeeeeeeeeeee eene 12 3 1 Weighing with the DeltaTrac 12 3 2 Dynamic weighing osi tenths idle Sale e beret robe led se tete enses ie ots 15 3 3 Change weight unit sat iuueni uota o Ria or teda eb E 15 3 4 Working in a higher resolution zu na apo er Dee dis 16 3 5 Display gross WIDE oiii sedet eere duode kdo etaed e rode erretra 16 3 6 Specifying dynamic set points sss rrearen 16 3 7 Multipl
51. RNAL The PLC controls the outputs of the ID7sx via the write table The ID7sx reads internal inputs and displays these in the read table 5 4 5 Messages in display The following messages may appear briefly in the display Message Meaning PROFIBUS NOT ACTIVE DP e nitialisation processes are still running on Profibus DP e The ID7sx is not yet connected to the Profibus PROFIBUS ACTIVE Readiness restored e g after switch on exiting mastermode or following a bus interruption PROFIBUS ERROR BCC RX PROFIBUS ERROR BCC TX ID7sx or field bus module have detected a BCC error PROFIBUS ERROR DATA RX PROFIBUS ERROR DATA TX Communication error ID7sx lt gt Field bus module e g not ETX Uart error etc PROFIBUS TIMEOUT ID7 Communication error ID7sx lt gt Field bus module The ID7sx does not respond within the defined time PROFIBUS ERROR CONF The field bus module has not received the configuration data properly 5 4 6 GSD file The GSD file required for communication with the Profibus DP ID7sx is available from METTLER TOLEDO Service or can be downloaded from the Profibus GSD Library at http www profibus com Operating instructions and installation information 22008168A 04 10 Interface description ID7sx Base 5 4 7 Status LEDs on the Profibus DP ID7sx module The 4 diagnosis LEDs on
52. S 1EAN13 PS 2Code 39 PS 3EAN13 Factory setting IDb MODE OFF ID7 MODE Deactivating or activating compatibility with ID7 The ID7sx Base is as compatible with the ID7 as possible If ID7 MODE ON is selected all known differences are made irrelevant Factory setting ID7 MODE OFF DISPLAY DURATION Set display duration for messages ERROR MESSAGES Set display duration for error messages factory setting 1 second INFO MESSAGES Set display duration for informational messages factory setting 2 seconds DISPLAY CONTRAST Set contrast of the LCD display Change the contrast in increments with the and function keys Operating instructions 22008168A 04 10 33 Settings in the master mode ID7sx Base RESET TERMINAL Reset all terminal functions to the factory setting DELTATRAC Filling MASTER MODE START POS Normal BIG WEIGHT DISPLAY On DYNAMIC WEIGHING No printout CONTROL MODE X 10 key ID5 MODE Off ID7 MODE Off Comment The memories are not affected by this 34 Operating instructions 22008168A 04 10 ID7sx Base Settings in the master mode 4 4 4 4 1 WEIGHING PROC DAPT 1 VIBRATION ADAPTER STABILITY DETECTOR AUTOZERO AUTOTARA RESTART SECOND UNIT DISPLAY UPDATE RESET SCALE Legend SCALE master mode block In the first block the weighing platform is selected SCALE 1 SCALE 3 The
53. The syntax and formats are dependent on the command set selected in the dialog mode see page 40 Read application block A R No MMR command set A RI_ No SICS command set The weighing terminal receives the command from the computer to read out the content of the No application block Possible formats for No are XXX Entire application block XXX ZZ Sub block of an application block XXX yyy Read only memory XXX yyy zz Sub block of a read only memory This read command is not contained in the following description of the application blocks A B lnformation MMR command set A R A Information SICS command set As a response the weighing terminal transmits the content of the No application block to the computer This response is contained in the following description of ihe application blocks in the MMR version Command MMR A R 0 2 1 0 0 1 Command SICS A R 0 2 1 0 0 1 Read out tare memory 1 Response MMR A B 10e 5d ese Response SICS A R _ A 1 03 5 Ror Operating instructions 22008168A 04 10 87 Application blocks ID7sx Base 88 Write Response Example N If n ote an application block is not in use the weighing terminal transmits the corresponding umber of blank spaces in place of the data For example when Tare Memory 1 is not in use the weighing terminal transmits the
54. ains and reconnect again If the message appears again contact METTLER TOLEDO Customer Service INTERF C X BREAK Wrong cabling Perform cabling according to control drawing contains too many blocks e Interface module C2 C3 not gt Insert or change interface module connected in the ID7sx Base or defective e Interface card and or RS CL20mA Insert or change RS CI20mA module module in the PSU power supply unit not connected or defective TRANSMIT BUFFER FULL No transmission gt Check handshake Too many key messages and baud gt Increase baud rate rate too low KEY BUFFER FULL e Data string currently being edited gt Remove blocks from data string ERROR BARCODE The specified application block contains no data Wrong sub block selected e g sub block O Select application block which contains data Select permissible sub block NO BLOCK Entered application block does not exist Enter different application block Operating instructions 22008168A 04 10 97 What to do if ID7sx Base Error Display Possible causes Remedy BUFFER IS FULL e Data string of transfer key contains more than 10 application blocks gt Change configuration of transfer key DISPLAY MODE e Weighing cell defective e 2 weighing platforms with same scale number connected Contact METTLER TOLEDO Customer Service Contact METTLER TOLEDO Custom
55. ake up to 10 seconds e f no data record with the entered values is found the message NO MATCHING DATE RECORD appears e f no search criterion has been entered the oldest data record is displayed 24 Operating instructions 22008168A 04 10 ID7sx Base Additional functions 3 16 4 Example Scroll Display data records The data records found are shown in the display DATE 02 04 98 TIME 09 25 51 NUM 000987 NET 25 000 KG TARE 100 346 KG PT With the keys gt gt lt and gt you can scroll within the data records found Notes e f during scrolling with the gt gt key all entries of the Memory ID7sx have been searched the message END OF FILE appears e f during scrolling with the keys lt and gt the file end is reached the display begins again at the first or last data record Operating instructions 22008168A 04 10 25 Settings in the master mode ID7sx Base 4 Settings in the master mode 4 1 Overview of the master mode In the master mode you adapt the ID7sx Base weighing terminal to meet your needs Depending on the configuration the master mode is divided into 4 or 5 master mode blocks which are in turn divided into further blocks zz Pul eee gt gt gt gt DE aapuumsssuua OK OK O K O K i j O K od EIS O K oLa EIS TERMINAL For system settings such as entering the date and time or loading permanent texts see section 4 3 2 PAC To set a
56. and A W No Information Write to application block A W No Reset application block A W No Delete application block Response A B Written to application block Comments e Information to be entered is dependent on target block see chapter 6 Deleting and resetting have same effect Write to display Command D Text 20 Write to display D Switch display to dark D Set display to normal status Response DB Written to display Comments e Character stock ASCII characters 20 hex 32 deci 7F hex 127 deci see section 9 1 e Watch capitalization Operating instructions 22008168A 04 10 57 Interface description ID7sx Base Alphanumeric printout on GA46 printer safe area Command P Text 48 Print text as per setting P _ 1 Text_48 Print text in small type P _ 2 Text_48 Print text in normal type P _ 3 Text_48 Print text in large type P _ S A Text_48 Print text in small type and bold print P S B Text 48 Print text in normal type and bold print P S 1 C Text 48 Print text in large type and bold print P Print blank line Response P B Alphanumeric characters printed Comments e Character stock ASCII characters 20 hex 32 deci 7F hex 127
57. ase automatically returns to normal operation Operating instructions 22008168A 04 10 7 Introduction and commissioning ID7sx Base 1 5 Cleaning DANGER OF SHOCK gt Do not open ID7sx Base weighing terminal to clean CAUTION gt Make sure that unused connection sockets are covered with the corresponding protective caps Ensure the correct position of the seals when doing so gt Do not use high pressure cleaners Cleaning Wipe off ID7sx Base weighing terminal with a commercially available glass or plastic cleaner gt When used in category 2 zone 21 hazardous areas the weighing terminal must be cleaned regularly See also standard EN 50281 1 1 50281 1 2 Operating instructions 22008168A 04 10 ID7sx Base Basic functions 2 Basic functions 2 1 2 2 2 2 1 2 2 2 Setting to zero Setting to zero corrects the influence of minor dirt on the load plate In the case of excessive dirt which cannot be compensated by setting to zero the display shows OUT OF RANGE Manual zero set 1 Relieve weighing platform 2 Press ZERO SET key The display shows 0 000 kg Automatic zero set On certified weighing platforms the zero point of the weighing platform is automati cally corrected when the weighing platform is relieved The automatic zero set can be switched off in the master mode on noncertified weighing platforms Taring Manual taring 1 Place empty container on scale 2 Press
58. atus 1 _ Time 1 _ Status 2 _ Time 2 Status 4 Time 4 Status 5 Trigger time sequence of status changes of digital outputs wi W Reset all outputs to logical O Status Each output is assigned a value The total of the values of those outputs which are to be closed is indicated as the Status Digital output 1 Digital output 2 2 Digital output 3 4 Digital output 4 8 Digital output 5 16 Digital output 6 32 Digital output 7 64 Digital output 8 128 All outputs open 0 All outputs closed 255 Time 1 99999 ms Response W B Digital outputs set Comments e Max 5 statuses Status and 4 intervals Time are possible After sequence has been run digital outputs freeze in last status Status A break in the port has no effect on the outputs e f terminal receives a new W command before time sequence has been run ongoing sequence will be aborted immediately Examples Command w 5 Digital outputs 1 and 3 are closed all others opened Command w 1 1 0 0 0 3 2 5 0 0 0 3 3 5 0 0 triggers following sequence 1s 0 5s Output 1 ss E L Output 6 L Operating instructions 22008168A 04 10 59 Interface description ID7sx Base 60 5 1 5 Terminal messages only with CL20mA and RS232 In the dialog mode the ID7sx Base weighing terminal transmits an acknowledgement to the computer each time a key is pressed When this pressing of a key is replaced with an interface command th
59. c Code Hex Function key F1 01 Set to Zero 14 OE Function key F2 2 02 Taring 15 OF Function key F3 3 03 Tare specification 16 10 Function key FA 4 04 Enter 17 11 Function key F5 b 05 Clear 18 12 Function key F6 6 06 ON OFF 20 14 CODE A 7 07 3 1F CODE B 8 08 Decimal 46 2E CODE C 9 09 Number key O 48 30 CODE D 10 OA Number key 1 49 31 Function change 11 OB Info 12 OC Number key 9 57 39 Scale 13 OD Function key codes Function key Code Dec Code Hex Standard keys of ID7sx Base 00 00 Extended tare keys of ID7sx Base 02 02 Standard keys of Pac 51 33 Extended keys of Pac 52 34 etc Only when the Pac is equipped with more than one function key page i e more than 6 function keys 84 Operating instructions and installation information 22008168A 04 10 ID7sx Base Interface description Digital I Os The operating mode of a module 8 l O ID7sx installed on the ID7sx is dependent on the parameters CONTROL INPUTS CONTROL OUTPUTS Outputs Inputs No I Os on ID7sx The ID7sx controls external outputs via the read table The ID7sx reads external inputs from the write table and executes predefined actions I Os on ID7sx inputs and outputs configured to CONTROL INTERNAL The ID7sx controls internal outputs and displays these in the read table The ID7sx reads internal inputs and executes predefined actions the PLC has no access I Os on ID7sx inputs and outputs configured to CONTROL EXTE
60. ck numbers etc e Commands or the corresponding responses with a total of 16 bits e Control of 8 digital I Os Operating instructions and installation information 22008168A 04 10 ID7sx Base Interface description 5 4 3 Handshake As certain commands can not always be executed immediately by the scale e g taring with a restless weighing platform 3 handshake bits of the PLC allow clear monitoring of the success of its commands 1 The PLC starts a command by setting the corresponding command bit and also toggles COMMAND VALID in the write table All other command bits are O 2 The weighing terminal responds with the current data of the read table If it was possible to completely process the command the COMMAND EXECUTED bit is toggled Otherwise COMMAND EXECUTED remains unchanged 3 The PLC recognises whether it can transmit the next command or must repeat the last one from COMMAND EXECUTED and transmits the write table to the weighing terminal 4 The weighing terminal recognises from the status change of the COMMAND VALID bit that it should carry out the next command In addition the weighing terminal also detects whether the last command has been executed or is still running If the PLC attempts to start new commands before the previous one has been confirmed by the weighing terminal with a status change of COMMAND VALID the weighing terminal ignores this new command Operating instructions and installation information 220081
61. d for METTLER TOLEDO weighing platforms of the D x K x series Point Ex system solution and analog weighing platforms with AWU 3 6 or Point Ex e In the configuration with the PSU power supply unit 2 additional IDNet connections IDNet ID7sx are possible Serial interface connection e CL20mA connection standard e n the configuration with the PSU power supply unit max 2 additional serial interface connections CL20mA ID7sx or RS232 ID7sx possible see guide for installers and control drawing ME 22006478 Operating instructions 22008168A 04 10 99 Technical data and accessories ID7sx Base Terminal Digital inputs outputs 8 l O ID7sx module as an option e 8 digital inputs 8 digital outputs galvanically separated passive e Signal level input logically O not supplied with current 0 3 mA logically 1 supplied with current 4m e Intrinsically safe connection values see PSU guide for installers and control drawing ME 22006478 Storage of certification relevant weighing data Memory ID7sx module as an option e Paperless archiving of certification relevant weighing data Capacity approx 700000 data records with net weight tare weight date and time e Convenient search function Ring data buffer which overwrites the oldest data record when the memory capacity is exceeded 100 Operating instructions 22008168A 04 10 ID7sx Base Technical data and accessories
62. d or gt Eliminate contact weighing sample and surroundings e Weighing platform tilted gt Level weighing platform e Wrong weighing platform selected gt Select right weighing platform PLUG IN Weighing platform cable not gt Connect weighing platform cable connected correctly according to control drawing e Scale module not installed gt Mount scale module correctly WRONG CODE e Wrong personal code gt Enter right personal code SCALE NO ERROR e Error in weighing cell gt Repeat test gt If the message appears again contact METTLER TOLEDO Customer Service 96 Operating instructions 22008168A 04 10 ID7sx Base What to do if Error Display Possible causes Remedy OUT OF RANGE e Zero set range exceeded gt Relieve weighing platform Gross weight negative Taring range exceeded Entered value outside permissible range Relieve weighing platform and set to zero Relieve weighing platform and set to Zero Enter permissible value NOT ALLOWED e Wrong cycle time for dynamic gt Enter cycle time between 1 and 255 weighing cycles Weighing platform does not exist gt Connect weighing platform e Print with negative weight value gt Relieve weighing platform set to zero and repeat weighing NOT EXISTENT e Recalled memory not assigned gt Recall other memory NO DATA TRANSFER Weighing platform does not transmit data to the terminal Disconnect power supply unit from the m
63. deci see section 9 1 e Text is printed in last selected type size e Watch capitalization Barcode printout on GA46 printer safe area Command P _ s 1 Text 20 barcode specific Print Code 39 PI S 2 Text 8 barcode specific Print EAN 8 PI I 3 Text 18 barcode specific Print EAN 13 P _ s 4 Text_20 barcode specific Print EAN 128 P _ s 5 Text_20 barcode specific Print Code 2 of 5 P _ s 6 Text_20 barcode specific Print Code 2 of 5 interleaved PI J 7 Text 20 barcode specific Print Code 128 P _ s 8 Text_20 barcode specific Print EAN 128 PI Print blank line Response P B Barcode printed Comments e Character stock ASCII characters 20 hex 32 deci 7F hex 127 deci see section 9 1 e With Code 39 3 barcodes can be printed next to each other Separating characters S or Hy ASCII character 09 hex 9 deci Arrangement of barcodes Barcode 2 Barcode 1 Barcode 3 Acoustic signal Command D S Generate short acoustic signal beep tone in terminal Response D B Acoustic signal generated in terminal Identification Command I D Interrogate identification of terminal Response I D 7 Program number of Pac 58 Operating instructions 22008168A 04 10 ID7sx Base Interface description Actuating digital outputs Command W __ Status Switch individual digital outputs on or off W _ St
64. dynamic set point ascending DJ dynamic set point descending Fixed set point Set point value is specified in the master mode and cannot be changed in the weighing mode Dynamic set point Set point value is specified in the weighing mode see Section 3 6 Ascending Digital output is set when the value of the application block concerned is greater than or equal to the set point value Decending Digital output is set when the value of the application block concerned is less than or equal to the set point value b Application block Weight value to which the set point refers All application blocks with a valid weight unit are possible Factory setting Application block 012 net weight c Scale WI W3 or ALL for all scales d Set point value With dynamic set points the weight value is entered in the normal mode see Section 3 6 Operating instructions 22008168A 04 10 ID7sx Base Settings in the master mode 4 5 4 Configuring Profibus DP ID7 PROFIBUS DP Configuring Profibus DP ID7 NODE ADDRESS Select desired node address in range 001 to 126 Factory setting 126 OPERATING MODE Set type and word length of user data parameter VALUE Consistent over valid module pair in GSD file 16 BIT INTEGER 2 words 16 BIT INTEGER 2 2 W Al 2 WORDS 16 BIT INTEGER 2 2 W AO 16 BIT INTEGER 2 words 16 BIT INTEGER 2 2 W Al use 2x 4 WORDS 16 BIT INTEGER 2 2 W AO use 2x 32 BIT FLOATING 4 words 32 BIT FLOATIN
65. e _ Unit Transmit the next stabilized weight value and depending on the specified deflection a dynamic weight value after a weight change greater than the specified deflection value Response S Weight value _ Unit Transmit stabilized weight value repeatedly S D Weight value _ Unit Transmit dynamic weight value repeatedly Comment Stop command with s S 1 command or by interrupting the interface Example Command S R 1 4 0 k g Responses 2 00 0 0 _ k g Ist item So eer eb Oy E SS Spe hee ee pod ee poe a eo 2nd item Operating instructions 22008168A 04 10 Interface description ID7sx Base Transmit data record Command SX Transmit a data record with stabilized weight values after weighing platform stabilization Effect as if ENTER key is pressed SIX Transmit a data record with stabilized or dynamic weight values independent of weighing platform stabilization S X I R Transmit data records with stabilized or dynamic weight values repeatedly independent of weighing platform stabilization Response S X Application block Application block A No Data record Data record with stabilized weight values transmitted S X D Application block Application block
66. e S x I R command can be stopped with the commands s x or s x 1 Example Command S X I Response Default data record S X D A 0 1 1 l li CE REO GS RI NIE eS pe a SO et cp ese Rees pe Oe ea eS 72 Operating instructions 22008168A 04 10 ID7sx Base Interface description Switch keyboard on or off Command R 0 Switch on keyboard E Switch off keyboard Response R 0 A Keyboard switched on R 1 A Keyboard switched off Comments e Factory setting Keyboard switched on e When the keyboard is switched off the terminal cannot be manually operated Changing over to different weight unit Command U Unit Change over weight display to different weight unit U Change over weight display to the first weight unit Response U A Weight display switched over to another weight unit U r Impermissible weight unit Comment Possible units mg g kg Ib ozt oz dwt Acoustic signal Command D S Generate short acoustic signal beep in the terminal Response D S A Acoustic signal generated in the terminal Read application block Command A R No Read contents of the application block Response A R A Iformation Contents of the application block transmitted Comments The transmitted information
67. e acknowledgement only differs in the second character in the response format which is part of the command Function Key Acknowledgement Set zero ZA Tare T A see command T Specify tare weight T A H see command T Change over unit U A Unit Transmit data record in S T see command SX case of weighing platform stabilization Switch over weighing S A n platform n weighing platform 1 3 Dynamic weighing A A 0 1 6 Weight value __ Unit Identification A D A D Kx Identification X B C D 20 characters right justified Function keys F1 F6 K F _ x X J K L M N Fault messages Fault messages always consist of 2 characters and a string frame The string frame can be defined in the master mode section 4 5 1 ET Transmission error The terminal transmits a transmission error for errors in the received bit sequence e g parity errors missing stop bit ES Syntax error The terminal transmits a syntax error when the received characters cannot be processed e g command does not exist E L Logic error The terminal transmits a logic error when a command cannot be executed e g when an attempt is made to write to a write protected application block Operating instructions 22008168A 04 10 ID7sx Base Interface description 5 2 METTLER TOLEDO continuous
68. ed as the tare weight and the weight display is set to zero with the weight placed on the platform Effect as when TARE key is pressed T Tare weight weight value _ Unit Specify tare weight The content of the tare memory is overwritten with the specified tare weight and the net weight is displayed Effect as when TARE ENTRY O 9 ENTER key sequence is pressed po Delete tare weight Response T B _ _ Tare weight weight value _ Unit Weighing platform is tared T B H Tare weight weight value Unit Weighing platform is tared with specified weight Ta Command cannot be executed Tare range dropped below T Command cannot be executed Tare range exceeded Comments e Taring is only possible when the weighing platform stabilizes within the tare range e The tare weight is always transmitted in the first weight unit e Each taring command overwrites the content of the tare memory with the new tare weight e Taring with an unloaded weighing platform deletes the tare memory On some weighing platform types a zero set is carried out in the unloaded state This is displayed with the message ZA see section 5 1 4 On not certified weighing systems the tare weight is automatically rounded to the current increment On certified weighing systems Tare range for MultiRange only in first increment range Example Command T Response TB 1112 464 5 0 k 9 54 Operating instructions 22008168A 04 10
69. ed or transferred e Manual or automatic data transfer Key which triggers printing or data transfer Factory setting e Manual triggering with the ENTER key e The content of the display is transferred or printed Operating instructions 22008168A 04 10 ID7sx Base Additional functions 3 14 3 14 1 Example 3 14 2 Example 3 15 3 15 1 Enter values with barcode reader If you have connected an explosion protected barcode reader to the ID7sx Base weighing terminal you can make all required entries such as identifications or target Specifications easily with the barcode reader Read in any desired entries with the barcode reader Read in identification Code A 1 Press CODE A key the ID7sx Base expects the entry of Code A 2 Enter identification Code A with the barcode reader 3 Confirm barcode entry with ENTER Read in a frequently used entry directly with the barcode reader If your working procedure repeatedly requires the same entry you can configure the barcode reader in the master mode see section 4 5 2 so that no additional keys need to be pressed on the ID7sx Base terminal for barcode entry Barcodes are automatically read in as Code A If the working procedure requires the entry of Code A gt Enter identification Code A with barcode reader The read in information is automatically processed by ID7sx Base as Code A Working with a second display The ID7sx weighing terminal can be connected
70. ely the function code for the start of the function K _ B _ x Or K A x for the end of the function is transmitted Response K A Command understood or function successfully executed K r Command understood but currently cannot be executed e g no keyboard present K L Command understood parameters defective Key codes K R x Key x was pressed briefly and released again immediately x Key x was pressed for approx 2 sec 68 Operating instructions 22008168A 04 10 ID7sx Base Interface description Response Function codes x The function codes are dependent on the command transmitted x K3 x K4 1 Set fo zero Tare 2 X10 2 Setto zero 3 Switch on off tare 3 Transfer key 4 Enter master mode 4 Enter master mode 5 Transfer key 5 Exit master mode 6 FI 7 Test 7 F2 10 Unit switchover 8 FS unit switchover 11 X10 9 F4 but not X10 12 RESETALL 10 F5 13 FM 11 F6 but not MODE 14 F2 21 CODE A 15 F3 22 CODE B 16 F4 23 CODEC 17 F5 24 CODED 18 F6 25 Function change key 21 CODEA 26 INFO 22 CODEB 27 SCALE 23 CODEC 28 24 CODED 29 Decimal point 25 Function change key 30 0 26 INFO TM 27 SCALE 39 9 28 40 CLEAR 29 Decimal point 30 0 39 9 40 CLEAR Comments e The factory setting is active after switch on after the Reset command and after exiting the master mode e Only one K command is ever active at one time Operating instructions 22008168A 04 10 69
71. er Service 98 Operating instructions 22008168A 04 10 ID7sx Base Technical data and accessories 8 Technical data and accessories 8 1 Technical data Terminal Display e Backlit LC display with graphics capabilities 64 240 pixels display field 39 x 132 mm e Cover of scratch resistant hardened antireflection glass Keypad e Tactile touch membrane keypad with acoustic acknowledgement e Scratch resistant marking 3 color e 4 keys A to D for identification data 6 function keys with function change and info key 4 scale function keys numerical keypad e Alphanumeric input possible with function keys Housing e All nickel chromium steel DIN X5 CrNi 1810 e Weight net 3 2 kg without cables Type of protection e 2G EEx ib IIC T4 10 C 40 C e 2D IP65 T55 C Protection type IEC 529 e Dust and water tight as per IP65 68 e Resistant to high pressure and steam jet cleaning as per IPX9K Pollution degree Installation category Power supply connectioni e via PSU ID Ex power supply unit in the safe area e via PSUX in the hazardous area Ambient temperature e operation 10 C 40 C for weighing platforms of certification class III 0 40 C for weighing platforms of certification class II e storage 25 60 C Relative humididy 20 80 non condensing Weighing platform connection e IDNet connection standar
72. for the current gross weight on scale 1 at 1 200 kg 725 Set point 2 Response equal to 724 Write equal to 724 x 2 5 726 Set point 3 Response equal to 724 Write equal to 724 x 2 6 727 Set point 4 Response equal to 724 Write equal to 724 x 2 7 8 place binary value Bit8 Bit7 Bit Bit8 output input 8 Bit output input 1 Operating instructions 22008168A 04 10 95 What to do if ID7sx Base 7 What to do if Error Display Possible causes Remedy Display is dark No mains voltage gt Check power supply unit e Wrong cabling gt Cabling according to control drawing Underload e Load plate not in place gt Apply load plate e Preload not applied gt Apply preload e Weighing range dropped below gt Sef zero Overload e Weighing range exceeded gt Relieve weighing platform e Weighing platform locked gt Release lock Weight display unstable e Agitated set up location gt Adjust vibration adapter e Draft gt Avoid drafts e Agitated weighing sample gt Weigh dynamically e Contact between load plate and or gt Eliminate contact weighing sample and surroundings Power malfunction gt Check mains Wrong weight display e Wrong setting to zero of weighing gt Relieve weighing platform set to zero platform and repeat weighing e Wrong tare weight gt Delete tare or enter right tare value e Contact between load plate an
73. g Bit2 Bit 1 BitO Decimal Increment position 0 0 0 0 XXXX00 0 0 XXXXXO 5 0 0 XXXXXX 0 XXXXX X 0 0 XXXX XX 0 XXX XXX 0 XX XXXX X XXXXX Operating instructions 22008168A 04 10 61 Interface description ID7sx Base 62 5 2 2 Status byte SB2 Bit 6 Bit5 Bit4 Bit 3 Bit 2 Bit 1 Bit 0 0 O Ib O Stabiliza O Normal O Positive O Gross tion status sign value 1 kg 1 Movement 1 Underload 1 Negative 1 Net overload sign value Status byte SB3 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0 0 0 O Basic state Weight value Print request Bit 2 Bit 1 Bit 0 Weight value 0 0 0 kg Ib SB2 Bit 4 0 0 g 0 0 t 0 OZ 0 0 ozt 0 dwt 0 fon free unit Commands to ID7sx Base Individual command characters can be transmitted to the ID7sx Base in the text format One function each on the terminal is assigned to these command characters After a command character is received the following functions are executed Command Function C Clear tare P Print or transmit transfer string T Tare Z Set zero Operating instructions 22008168A 04 10 ID7sx Base Interface description 5 3 5 3 1 Command format when transmitting weight values Response format when transmitting weight values Example Data formats
74. gger time sequence of status changes of digital outputs wi W Reset all outputs to logical O Status Each output is assigned a value The total of the values of those outputs which are to be closed is indicated as the Status Digital output 1 Digital output 2 2 Digital output 3 4 Digital output 4 8 Digital output 5 16 Digital output 6 32 Digital output 7 64 Digital output 8 128 All outputs open 0 All outputs closed 255 Time 1 99999 ms Response W A Digital outputs set Comments e Max 5 statuses Status and 4 intervals Time are possible After sequence has been run digital outputs freeze in last status Status e A break in the port has no effect on the outputs e f terminal receives a new W command before time sequence has been run ongoing sequence will be aborted immediately Examples Command w 5 Digital outputs 1 and 3 are closed all others opened Command w 1 1 0 0 0 3 2 5 0 0 0 3 3 5 0 0 O triggers following sequence 1s 0 5s Output 6 __ L 76 Operating instructions 22008168A 04 10 ID7sx Base Interface description 5 3 4 Error messages Error messages always consist of 2 characters and a string limit The string limit can be defined in the master mode section 4 5 1 E T Ei un Operating instructions 22008168A 04 10 Transmission error The terminal transmits a transmission error for errors in the received bit sequence e g parity error missing
75. ht The saved tare weight can be recalled at any time gt Enter INFO TARE SPECIFICATION key sequence The saved tare weight is displayed Clear tare weight gt Relieve weighing platform and tare _or gt Specify tare weight O _or gt Enter TARE SPECIFICATION CLEAR key sequence Weighing Weighing without taring gt Lay weighing sample on weighing platform Gross weight total weight is displayed Weighing with taring 1 Place the empty container on the weighing platform and tare 2 Pour in weighing sample The display shows the net weight and the NET symbol Operating instructions 22008168A 04 10 ID7sx Base Basic functions Weighing with tare specification 1 Place filled container on weighing platform The display shows the gross weight total weight 2 Specify tare weight or recall tare memory The display shows the net weight container content and the NET symbol Note If a multi range weighing platform is chosen a display for the currently active range appears above the scale symbol 2 4 Switch over weighing platform Up to 3 weighing platforms can be connected to the ID7sx Base The weighing platform currently selected is shown on the terminal gt Press SCALE key The next weighing platform is selected gt Enter number of weighing platform and press SCALE key The desired weighing platform is selected Operating instructions 22008168A 04 10 11 Additional function
76. icative tare function sss mm 17 3 8 Additive tare fUNCHON ct cd sett exeat de tede bge fette Medos cede eios dede Ptca 17 3 9 Sandwich ACIS 1 arate sich TEC 17 3 10 Display ID code and test weighing platform ssessssss 18 3 11 ds millPaii vo meer 18 3 12 Recall TOP TOT osa it dta dat ac Don ta fa ue ood 19 3 13 Print or transfer d ta sexincaciwnits meueus BERE IE trip bete 20 3 14 Enter values with barcode reader eee 2 3 15 Working with a second display dt dea d e RR Rn bacs 21 3 16 Recall data from memory module sssen 23 4 Settings in the master mode eeeeeeeeeeeeeeeeeeee 26 4 Overview of the master mode ssssem erene 26 4 2 Operating the master MOS sce oui eeu drea tia i iil 27 4 3 TERMINAL master mode block 29 4 4 SCALE master mode block 2 3 1s ilea ais pitt c acted alist aan Sarina 35 4 5 INTERFACE master mode block 3 5 ae so ae eid 38 5 Interface description eese rrr 51 5 1 MM comirdrid S6l 6 nte EE rh LE HER D bo aeaa sarea 5 5 2 METTLER TOLEDO continuous mode a 61 5 3 METTLER TOLEDO SICS command set 63 5 4 Profibus DP communication with a PLC acc 78 Operating instructions 22008168A 04 10 1 Contents ID7sx Base 6 Application blocks eeeee serrer eene rrr nnn 87 6 1 SVRIGX and TOHITIGHS S oso toteoiet ipo i
77. ich format the designation and content of the application block are printed adjustment possibilities TYPE STYLE AB DB Designation and content in grouped style Designation and content in two lines grouped style Designation and content separated with extra blank spaces Content alone left justified Content alone centred Content alone right justified TEXT Left justified Centred dU b Right justified CRLF parameter Force line feed the CRLF parameter is only available for e _ Text left justified e Content alone left justified Ee Designation and content separated with extra blank spaces e Type CHRn FILL parameter Show content with leading blank spaces up to maximum available length the FILL parameter is only available for e Designation and content separated with extra blank spaces e Content alone left justified e Content alone centred PAD parameter Show designation and content separated with x blank spaces Possible settings O 63 extra blank spaces The PAD parameter is only available for e Designation and content separated with extra blank spaces e Content alone left justified 44 Operating instructions 22008168A 04 10 ID7sx Base Settings in the master mode DATA FIELD parameter Depending on the TYPE selected DATA or FIELD is available DATA
78. ight of these boxes can be subtracted with this function Press SANDWICH T key 2 Place sandwich tare e g box on scale and confirm with ENTER The net weight is retained Operating instructions 22008168A 04 10 17 Additional functions ID7sx Base 3 10 Display ID code and test weighing platform Each time the weighing platform configuration is changed the ID code counter is increased by 1 On certified weighing platforms the displayed ID code must match the ID code on the ID code sticker otherwise the calibration is no longer valid Display ID code gt Press ZERO SET key and hold until IDENTCODE appears in the display and press again Test weighing platform gt Press ZERO SET key again The connected weighing platform is checked The display shows CHECK SCALE and then SCALE IS OK after completing the test Note If weighing platform is defective display shows SCALE ERROR 3 11 Identifications The ID7sx Base weighing terminal is equipped with 4 identification data memories for storing identification data Code A Code D The memories have a name e g Article No and a content which identifies the current weighing e g 1234567 The memories are named in the master mode and the names can be noted on the keyboard When the CODE keys are pressed the name appears in the display Identification data Code A Code D can be entered or recalled for each weighing and are printed immediately 3 11
79. int text in normal print Text 48 Print text in large print Text_48 Print text in small type and bold print Text_48 Print text in normal type and bold print Text_48 Print text in large type and bold print Text 20 barcode specific Print code 39 Text 8 barcode specific Print EAN 8 Text 13 barcode specific Print EAN 13 Text 20 barcode specific Print code 128 Text 20 barcode specific Print code 2 of 5 Text 20 barcode specific Print code 2 of 5 interleaved Text 20 barcode specific Print code 128 00 3 o lal i OY DO el bo e Jd Text_20 barcode specific Print EAN 128 Print blank line Response P A P 5 Alphanumeric characters printed no GA46 present Comments Character stock ASCII character 20 hex 32 dec 7F hex 127 dec see section 9 1 Printing is carried out in the font size last selected Watch uppercase and lowercase letters Operating instructions 22008168A 04 10 75 Interface description ID7sx Base Actuating digital outputs safe area Command W __ Status Switch individual digital outputs on or off W _ Status 1 Time 1 Status 2 _ Time 2 Status 4 Time 4 Status 5 Tri
80. it 1 digit calculated by ID7sx Base X O 6 decimal places of weight value Weight 6 digit net weight value EAN 128 330 3 3 0 x Weight X O 6 decimal places of weight value Weight 6 digit gross weight value 092 Barcode EAN 29 Response A B _ 2 9 Article Check digit Weight Comment Article 4 digit article no from memory Code A Check digit 1 digit no calculated from ID7sx Base for the weight Weight 5 digit positive weight value with 3 places to right of point between 00 000 kg 99 999 kg 093 Barcode EAN 29 A Response A B 2 9 Article Weight Comment Article 5 digit article no from memory Code A Weight 5 digit positive weight value with 3 places to right of point between 00 000 kg 99 999 kg 094 Identification data Response A B _ Name text 20 Identification text_30 097 Code A Code D Write A W 0 x x _ Name text 20 5 Identification text_30 Comment xx 94 97 098 Number of last Alibi Response A B Number 6 entry Note The data record number is output with leading zeros 92 Operating instructions 22008168A 04 10 ID7sx Base Application blocks No Content Format 601 Parameters for Scale 1 Response A B j Parameters for Scale 1 Note For service information purposes the internal scale parameters can be read oult printed the structure and content are scale dependent 602 Parameters for Scale 2 Response
81. ive animals To do this specify the number of weighing cycles for which the mean weight value is fo be taken 1 Set container on the weighing platform 2 Tare weighing platform 3 Place weighing sample in container 4 Press DYN key and enter number of weighing cycles Possible values 1 255 m Start dynamic weighing with ENTER key 6 Mfter cycle time has expired center line of display shows RESULT x xxxx kg This display is retained until the next weighing is started or until it is cleared Delete result gt Press CLEAR key Notes e Dynamic weighing results are automatically printed when AUTO PRINT is set in the master mode see section 4 3 2 e Dynamic weighing can also be started with the interface command AWO16 see section 6 2 3 3 Change weight unit If an additional second weight unit is configured in the master mode it is possible to switch back and forth between the two weight units gt Press UNIT key The weight value is shown in the second unit Note Possible second weight units are mg g kg Ib oz ozt dwt Operating instructions 22008168A 04 10 15 Additional functions ID7sx Base 3 4 3 5 3 6 Use Working in a higher resolution Depending on the setting of the master mode block CONTROL MODE see page 32 the weight value can be displayed in a higher resolution continuously or when called Weight values in a higher resolution are marked with a Displa
82. l running applications and functions are cancelled e The tare memory is reset to Zero Write display Command D Text 20 Write display p Darken display Response D A Display written the complete text appears left justified in the display marked with a symbol e g with D A Display written the end of the text appears left justified in the display with the beginning cut off marked with a symbol e g with D i Command cannot be executed D L Command understood parameters defective Comment A symbol in the display e g indicates that an invalid weight value is displayed Weight display Command D W Switch over main display into the weight mode Response D W A The main display shows the current weight value D W I Command understood but cannot be executed Operating instructions 22008168A 04 10 67 Interface description ID7sx Base Keyboard monitoring Command Imt When a key is pressed execute the function but do not transmit anything factory setting Kc When a key is pressed do not execute the function and do not transmit anything K 4 3 When a key is pressed do not execute the function but transmit the key code K _ C _ x or when the key is pressed longer transmit Ki Rx and KC x K 44 When a key is pressed execute the function and transmit the function code K A x If the function cannot be executed immediat
83. lue lies within the specified tolerance TOL MIN DELTA Specify minimum target weight 40 d LANGUAGE Select dialog language Possible settings German English French Dutch Italian Spanish KEYSABCD Name identification keys CODE A CODED A Identification data CODE A Set name and number of characters B Identification data CODE B Set name and number of characters C Identification data CODE C Set name and number of characters D Identification data CODE D Set name and number of characters Note Max 30 characters possible factory setting 20 characters Operating instructions 22008168A 04 10 31 Settings in the master mode ID7sx Base DATE TIME Enter date and time TYPE EUROPE Select European notation Day Month Year 24 Hours Minutes Seconds US Select American notation Month Day Year 12 Hours Minutes Seconds AM PM DATE Enter date according to the type selected TIME Enter time according to the type selected Comments e Enter single place numbers with a preceding zero e Change over between AM and PM Press FUNCTION CHANGE key e Date and time can be printed out e The clock continues to run after the terminal is switched off PERSONAL CODE Load or delete code for entering the master mode CODE Enter code with a maximum of 8 alphanumeric characters Comment If no code is entered access to the master mode is unrestricted
84. must be set as the operating mode of COMI at the ID7sx in mastermode The proper baud rate is selected automatically here The operating mode AUTO DIR or TOLEDO CONTINUOUS with checksum must be set in the weighing terminal in the safe area for the COM port used With TOLEDO CONTINUOUS with checksum the following communication parameters must be set explicitly 9600 baud 7 data bits even parity 1 stop bit With AUTO DIR the correct communication parameters are set automatically With ID7 as a second display the weight value fills the entire display BIG WEIGHT display ON Weighing terminal in the safe area as a second display of an ID7sx COMI COM2 or COMS of the ID7sx weighing terminal must be lead to the safe area via a slotcard interface in the PSU See terminal diagram 22006478 Sheet 2 in guide for installers 22006472 The weighing terminal in the safe area must be connected to the CL output of the PSU power supply via its ID Net interface Ensure that the CL interface is configured as passive in the PSU here AUTO DIR must be selected as the operating mode of the selected port at the ID7sx No special settings are required at the second display in the safe area Operation possibilities on second display The following functions are also possible on the second display Setting to zero Taring Operating instructions 22008168A 04 10 ID7sx Base Additional functions 3 16 3 16 1 3 16 2 Recall data from memory
85. nction 17 P Personal code 32 Print 20 Print mode 39 42 Printout 58 Profibus DP 49 R Recall information 19 Representable characters 105 Reset scale 37 Reset terminal 34 Response format 63 Restat 37 RS 38 S Safety precautions 3 Sandwich tare 17 Second display 21 Second unit 37 Serial interfaces 51 Set zero 9 53 67 101 SICS command set 40 63 Stability detector 36 Status byte 61 62 String framing 40 51 63 Switch keypad on off 53 73 Switch over weighing platform 11 T Tare 9 54 101 Tare memory 10 30 Taring 70 Technical data 99 Terminal 99 Test weighing platform 18 Text memory 19 30 Time 32 101 Transfer data 20 Transfer string 40 Transmit buffer 94 Transmit data record 56 72 Transmit weight value 55 66 U Unit changeover 101 V Vibration adapter 36 101 Ww Weighing 10 Weighing functions 101 Weighing platform connections 99 103 Weighing process adapter 36 101 What to do if 96 Operating instructions 22008168A 04 10 Ne rQv Texvodoyies ABEE l paka 113 T 0 67934 l tO l 15344 l pakaq Tm 210 6654544 Z yton Fax 210 6654545 2hyavon marketing nefton gr ZULUSPQWON www nefton gr 22008168A Subject to technical changes Mettler Toledo Albstadt GmbH 04 10 Printed in Germany 22008168A Mettler Toledo Albstadt GmbH D 72458 Albstadt Tel 49 7431 14 0 Fax 4 49 7431 14 232 Internet http Avww mt com
86. ng Page Level 0 lO Transmit list of all available SICS commands 65 11 Transmit SICS level and SICS versions 65 12 Transmit scale data terminal platform 65 I3 Transmit scale software version program number 66 14 Transmit serial number 66 S SI SIR Transmit display contents 66 Z Set to zero 67 Reset 67 Level 1 D Write display 67 DW Weight display 67 K Keyboard monitoring 68 SR Transmit stabile weight values repeatedly depending on a weight change 70 T Taring 70 TI Tare immediately 71 TA Specify tare weight 71 TAC Delete tare weight 72 Level 2 SX SXI SXIR Transmit data record 72 RO R1 Switch keyboard on or off 73 U Change over to different weight unit 73 DS Acoustic signal 73 Level 3 AR Read application block 73 AW Write application block 74 DY Specify DeltaTrack target value 74 P Print text or barcode 75 W Actuating digital outputs 76 64 Operating instructions 22008168A 04 10 ID7sx Base Interface description 5 3 3 Command description Transmit SICS commands Command I 0 Transmit SICS commands Response I 0 B T 0 _ 0 _ 10 O _JOJ_J H o 1 D 0 Nu 2 u SX 0 3 AR 0 A Transmit SICS levels and SICS versions Command To Transmit SICS levels and SICS versions Response II A er I x2 o e3 I x4 x5 xl 0123 Scale with SICS levels O 1 2 and 3 x2 Version or implemented S
87. nt net weight Response A B j Weight value _ Unit 2nd weight unit 009 Current tare weight Response A B j Weight value _ Unit 2nd weight unit Write A W 0 0 9 Weight value __ Unit 010 Current weighing Response A B Number 2 platform Write A W 0 1 0 Number 2 Switch over weighing platform 011 Current gross weight Response A B Weight value _ Unit 1st weight unit 012 Current net weight Response A B j Weight value _ Unit 1st weight unit 013 Current tare weight Response A B j Weight value _ Unit 1st weight unit Write A W 0 1 3 Weight value Unit 014 Content of display Response A B _ Display Display Text 20 or weight value 015 Date Response A B _ Date Write A W 0 1 5 _ Date Date DD MM YY or DD MM YY 016 Dynamic weighing Response A B Weight value _ Unit Write A W 0 1 6 _ No of cycles Start weighing cycle Comment No of cycles 1 255 018 Difference target Response A B Weight value _ Unit actual weight 019 Date and time Response A B D D MjM Y Y _ 11 1_ 5h h m m s s Europe A B MMM D D Y Y A P M h h m m s s USA Write A jW 0 1 9 D D M M 7 Y Y S S h h n S s Europe A W 0 1 9 _ M M D D Y Y S A P Mlh h m m s s USA Date instead of also Time instead of also or 90 Operating instructions 22008168A 04 10
88. nts Reference sample Limits Clear DeltaTrac target value Preset DeltaTrac target values 1 Press DELT key 2 Enter target weight and confirm with ENTER 3 Enter tolerance in of target weight and confirm with ENTER Note With the FUNCTION CHANGE key you can select the weight unit for entering the DeltaTrac target values With the CLEAR key the entry is corrected character by character The ID7sx Base weighing terminal has 999 DeltaTrac memories for frequently used target values and tolerances which are programmed in the master mode Enter number of DeltaTrac memory 1 999 2 Press DELT key Press DELT key Lay sample on weighing platform and confirm with SCALE key Only for FILLING and CLASSIFYING Enter tolerance and confirm with ENTER Remove sample from weighing platform Qe Ne s Minimum target value 40 Digit Maximum target value configured maximum load Minimum tolerance Digit Maximum tolerance 10 96 for the applications FILLING CHECKWEIGHING 50 for the application CLASSIFYING Note If the limits are not observed a message appears in the display e g MIN DEL for too small a target value gt Press DELT CLEAR key sequence DELTA CLEARED appears briefly in the display then the weight is shown Operating instructions 22008168A 04 10 ID7sx Base Additional functions 3 2 Dynamic weighing With the dynamic weighing function you can weigh restless weighing samples e g l
89. onger appears when a GA46 printer is already configured For COM2 or COMS only For connection of an explosion protected barcode reader An RS232 ID7sx interface must be installed on the selected interface connection for this purpose For other settings see 4 5 2 For COM4 only A Memory ID7sx module must be installed for this purpose No further settings are required in the master mode This selection no longer appears when a Memory ID7sx module is already configured For COM5 only An 8 l O ID7sx interface must be installed on the selected interface connection for this purpose For other settings see 4 5 3 For COMI only A Profibus DP ID7sx module must be installed on COMI For other settings see 4 5 4 For COMI only In this setting COMI has the same functionality as an IDNet interface This allows the ID7sx Base to be connected to a weighing terminal in the safe area as a second display For other settings see 3 15 Operating instructions 22008168A 04 10 ID7sx Base Settings in the master mode 4 5 1 Settings in the master mode blocks CL20mA and RS232 CL20mA RS232 COMMUNICATION Set communication parameters factory settings are shown in bold print All parameters are shown on a display page and can be set there BITS PER Possible settings 7 bits 8 bits CHARACTER STOPBITS Possible settings 1 stop bit 2 stop bits PARITY Possible settings Even Odd No BAUDRATE Possible settings 150 300 60
90. or CHECKWEIGHING Note With the DeltaTrac signals you can also control lamps flaps or valves see section 4 5 3 Operating instructions 22008168A 04 10 ID7sx Base Additional functions Application FILLING Application CLASSIFYING Application CHECKWEIGHING For weighing in to a target weight with tolerance monitoring Example Target weight 1 000 kg tolerance 1 96 FF 0 560 kg x Ld Target weight not reached yet 0 991 kg ww Fw KA __ Weight within the tolerance 1 000 kg ww Fr I KA1 ee Target weight reached exactly To evaluate test samples as OKAY TOO LIGHT or TOO HEAVY based on a target weight and specified tolerances Example Target weight 1 000 kg tolerance 1 0 900 kg Y Y sai TOO LIGHT 0 0 58 i I 1 Weight below the tolerance wy 1 002 kg jp amp a1 OKAY I I l 1 weight within the tolerance 1 100 kg T Y a1 TOO HEAVY I IEEE Weight above the tolerance For determining the difference between the target and actual weight Example Target weight 1 000 kg tolerance 1 0 900 kg vyv Weight below the tolerance DELTA 0 100 kg A1 Difference 0 100 kg 1 002 kg DELTA 0 002 kg ga1 Weight within the tolerance Difference 0 002 kg 1 100 kg Y Weight above the tolerance Difference 0 100 kg DELTA 0 100 kg Operating instructions 22008168A 04 10 13 Additional functions ID7sx Base 3 1 1 Enter numerically Correct entry Copy consta
91. other setting possibilities are the same for all connected weighing platforms Overview of the SCALE master mode block In the SCALE master mode block the following settings for the weight can be carried out CONDITIONS CONDITIONS CONDITIONS Des es qe es qe RI I 1I Ie Ie Is e Blocks highlighted in grey are described in detail in the following e Factory settings are printed in bold print Operating instructions 22008168A 04 10 35 Settings in the master mode ID7sx Base 4 4 2 Settings in the SCALE master mode block WEIGHING PROC Adapt weighing platform to weighing sample ADAPT UNIVERSAL WEIGHING For solid bodies coarse filling or checkweighing factory setting STATIC WEIGHING For solid bodies and weighing under extreme conditions e g strong vibrations or weighing animals FINE FILLING For liquid or powdered weighing samples VIBRATION ADAPTER Adapt weighing platform to the vibration influences of the environment AVERAGE CONDITIONS Factory setting EXTREME CONDITIONS The weighing platform operates more slowly however is less sensitive e g suitable with building vibrations and vibrations at the weighing location IDEAL CONDITIONS The weighing platform operates very quickly however is very sensitive e g suitable with very calm and stabile weighing location STABILITY DETECTOR Adapt automatic stability detector Possible settings ASD
92. panded AUTO DIR operating mode for adaptation to an ID7sx Sys in the safe area TRANSFER STRING Possible settings e STANDARD gross net tare e USER DEFINED enter numbers of the application blocks which are to be transmitted or printed out Possible settings e lt CR gt lt LF gt Factory setting e lt STX gt lt ETX gt e BLOCK CHECK CHAR e lt CR gt SICS Dialog mode with Standard Interface Command Set SICS see section 5 3 STANDARD Standard setting no handshake no auto transmission HANDSHAKE Possible settings as MMR see previous page AUTOREPEAT Possible settings as MMR see previous page AUTO DIR not possible with SICS 40 Operating instructions 22008168A 04 10 ID7sx Base Settings in the master mode DIALOG MODE Set dialog between ID7sx Base weighing terminal and computer TOLEDO CONTINUOUS CHECKSUM ON For the continuous transmission of net and tare values to METTLER TOLEDO devices e g to a second display For a description see section 5 2 Checksum byte active factory setting CHECKSUM OFF Checksum byte inactive the transfer format is shortened by 1 character TOLEDO SHORT For the continuous transmission of net values to METTLER TOLEDO devices e g to CONTINUOUS a second display For a description see section 5 2 CHECKSUM ON Checksum byte active factory setting CHECKSUM OFF Checksum byte inactive the transfer format is shortened by 1 character
93. platform 0 0 None 1 0 Scale 1 15 Sign 0 1 Scale 2 1 1 Scale 3 Sign 80 Operating instructions and installation information 22008168A 04 10 ID7sx Base Interface description Read table 16 Bit Integer Word 0 Word 1 2 words 16 Bit Integer Word 0 Word 1 Word 2 Word 3 4 words 32 Bit Floating Word O Word 1 Word 2 Word 3 Point Bit Value X Value Command 16 Digital 1 0 Not in 16 Bit 32 Bit Floating Point Use IEEE 754 0 m Command executed 3 Toggle bit for all commands Error command 2 Movement 3 Net 4 E load underload Sa at oe rror scale overload underload reading of 5 Key s was were pressed ID7sx inputs 6 Input mode active 7 Mantisa Setpoint 1 reached or 8 Setpoint 2 reached 9 Setpoint 3 reached Displaying or setting outputs 10 Setpoint 4 reached of extemal 1 0 11 1 keyboard blocked O keyboard unblocked module 12 Mantissa Reserved 13 Reserved 14 Bits 14 15 Current weighing platform 0 0 None 1 0 Scale 1 15 Sign 0 1 Scale 2 1 1 Scale 3 Notes on commands If the command requires parameters they will be transmitted either as an integer value or as a floating point value depending on the operating mode set Exception The commands READ WRITE APPLICATION BLOCK and PRESS KEY always expect integer values as parameters Operating instructions and installation
94. pplication specific parameters This block does not appear with ID7sx Base SCALE To select one of the connected weighing platforms For each selected weighing platform the parameters are then set which concern the weight value e g stability detector unit etc see section 4 4 INTERFACES To select an interface The communication parameters are then set for each interface see section 4 5 SERVICE For configuring the weighing platform s On IDNet weighing platforms only for METTLER TOLEDO service technicians 26 Operating instructions 22008168A 04 10 ID7sx Base Settings in the master mode 4 2 Operating the master mode 4 2 1 Enter the master mode Press MODE key If the current function key assignment does not contain MODE change to the assignment with MODE by repeatedly pressing the FUNCTION CHANGE key 2 Enter personal code if configured The display shows the first master mode block TERMINAL 4 2 2 Assignment of function keys in the master mode Dependinig on the number of parameters offered in the display in the master mode the function keys are assigned as follows parameter c ji END OK Change to Change to next Exit level and Exit the Recall lower previous block within a return to master mode level block or block within a level higher level and return to confirm level block normal mode selection Several parameters lt gt lt gt
95. rem potete et reti d Podio emo tote p do s 87 6 2 TERMINAL SCALE application blocks e 90 6 3 INTERFACE application blocks ssssssee 94 7 What 10 do If 2 iet eoe re cedes 96 8 Technical data and accessories eeeeeeeeneeeenenennnnnn 99 8 1 Technical data uote ci ecd rni e ed EE teu 99 8 2 ACCOSSOTLOS s e bod Moa senda ucla utut usb tet deus I 103 9 ADDONGIX Seem E 105 9 1 PS CMMODIG ta I ser noesthan aa contr cibo enses ne scaled cae dora ihe tret 105 10 Ds T 106 Operating instructions 22008168A 04 10 ID7sx Base Introduction and commissioning 1 Introduction and commissioning 1 1 Safety precautions The ID7sx Base weighing terminal is approved for operation in zone 1 and 21 hazardous areas It may only be used in areas in which the causes of static electricity build up which lead to propagating brush discharges have been eliminated If the ID7sx Base weighing terminal is used in hazardous areas special care must be taken The code of practice is oriented to the Safe Distribution concept drawn up by METTLER TOLEDO Competence A Ex approval A Operation A The weighing system may only be installed maintained and repaired by authorised METTLER TOLEDO service personnel No modifications may be made to the terminal and no repair work may be per formed on the modules Any weighing platform or system modules that are
96. s ID7sx Base 3 Additional functions 3 1 The assignment of the 6 function keys of the ID7sx Base weighing terminal differs depending on the weighing task The current assignment is shown above the function keys With the FUNCTION CHANGE key it is possible to switch over to other function key assignments Independent of the application software the ID7sx Base has the following additional functions DELT DYN UNIT X10 GROSS MODE Weighing Dynamic Change Increase res Display Activate with the weighing weight unit olution see gross weight master mode DeltaTrac see 3 2 see 3 3 3 4 This key see 3 5 see see 3 1 is not as Chapter 4 signed when the control mode is con tinually switched on MULT TARE ADD TARE SANDWICH T Multiplicative tare function Additive tare function see 3 8 Sandwich tare see 3 9 see 3 7 If at least one dynamic switching point is configured in the master mode see page 48 the second row of function keys is given the following assignment SETP MUL T ADD T SW T Set dynamic Multiplicative Additive tare Sandwich set points fare function function fare see 3 9 see 3 6 see 3 7 see 3 8 Weighing with the DeltaTrac The DeltaTrac is an analog display which makes it easier to read the weighing results In the master mode you can select how the DeltaTrac is displayed for the various weighing tasks FILLING CLASSIFYING
97. s when TARE key is pressed Response T _ S _ Tare weight weight value Unit Weighing platform tared stabile tare value T I Taring not carried out T Command cannot be executed Tare range dropped below T _ Command cannot be executed Tare range exceeded Comments e Each taring command overwrites the contents of the tare memory with the new tare weight e Taring with unloaded weighing platform clears the tare memory On some weighing platform models setting to zero is carried out in the unloaded state e On non certified weighing systems the tare weight is automatically rounded off to the current increment e On certified weighing systems Tare range with MultiRange only in first increment range 70 Operating instructions 22008168A 04 10 ID7sx Base Interface description Tare immediately Command m 7 Tare weighing platform immediately Response __ Tare weight weight value __ Unit Weighing platform tared stabile tare value Tare weight weight value Unit Tol L Weighing platform tared dynamic tare value Taring not carried out Command cannot be executed Command cannot be executed Tare range dropped below Command cannot be executed Tare range exceeded Comments e Each taring command overwrites the contents of the tare memory with the new tare weight Following a dynamic tare value a s
98. sfer key on off When AUTO PRINTOUT ON is selected a printout for the transfer key is automatically created for each weight change gt x digits Possible settings 1 255 digits factory setting 30 digits 42 Operating instructions 22008168A 04 10 ID7sx Base Settings in the master mode Change configuration Display page The setting of the parameters of an entry appears in a clear layout on a display page example TRANSFER KEY EDIT 2 7 TYPE AB STYLE CRLF YES FILL NO PAD 01 DATA 011 013 First display line Information for orientation in an entry e Key name e Mode EDIT INS or ADD e Number of the display entry and total number of entries for the current printout TYPE parameter Selection possibilities AB TEXT CHRn LINE DB Operating instructions 22008168A 04 10 Output content of an application block with or without designation Print out any desired text Insert n of any desired ASCII characters in the line e g for tables selection of character via DATA parameter Blank line or separator line with any desired alphanumeric characters Accesses a database field When a field is printed out all entries of the field are listed The option DB is only available when the software application supports access to a database The offered database fields are application specific 43 Settings in the master mode ID7sx Base STYLE parameter STYLE determines in wh
99. smit buffer COM4 Write 0 4 Information 701 Description of application Response ID7sx Interfaces 702 Program designation Response IKY07 0 0100 Comments on the transmit buffers e The entered information is transmitted directly via the selected interface e A transmit buffer contains a maximum of 246 characters 94 Operating instructions 22008168A 04 10 ID7sx Base Application blocks 6 3 2 Digital inputs outputs The following application blocks are only available when interface 8 I O ID7sx is installed When the weighing terminal checks the outputs the blocks concerned cannot be written to and the E L error message appears No Content Format 706 Digital outputs 1 Response A B 8 place binary value Write A W 7 0 6 8 place binary value 707 Digital inputs 1 Response A B 8 place binary value 724 Set point 1 Response A B _ Set point Text 2 zc AXX X YYY Z Z Scale Text_3 mee Set point value weight value beg Write AW 7 2 x Set point type Text 2 AXX X Y Y Y Z4 Z S Scale Text 3 18 Set point value weight value Note x 4 Set point type Ft F4 Dt D4 Scale W1 W2 W3 ALL T Dec 24 Hex 14 l Dec 25 Hex 15 Axxx yyy zz Application block Example A w 7 2 5 F T A 0 1 1 W 1 1 2 0 0 Kg fixed ascending switching point
100. ssage NO MATCHING DATA RECORD appears Operating instructions 22008168A 04 10 23 Additional functions ID7sx Base 3 16 3 Search with other search criteria gt Press FIND key The function keys are given the following assignment DATE TIME NET TARE START Enter Enter Enter net Enter Start search date as time as value as tare value as with entered search search search search search criterion criterion criterion criterion criteria All offered search criteria can be combined with each other The entered search criteria are shown in the display in clear text This enables you to search for a find a specific weighing Enter date gt Press DATE key and enter complete date in DD MM YY form Enter time gt Press TIME key and enter desired time in one of following formats Format HH all weighings between HH 00 00 and HH 59 59 are found Format HH MM all weighings between HH MM OO and HH MM 59 are found Format HH MM SS only the weighing at the time HH MM SS is found Enter net tare value Press NET or TARE key 2 Enfer weight value and confirm with ENTER The function key assignment changes back again for selection of the search criteria Note With the FUNCTION CHANGE key you can select the weight unit for entering the weight values Start search gt Press START key Memory ID7sx searches for the oldest data record which meets the entered search criteria Notes e The search may t
101. stop bit Syntax error The terminal transmits a syntax error when it cannot process the received characters e g command not present Logic error The terminal transmits a logic error when a command cannot be executed e g when an attempt is made to write an non writeable application block 77 Interface description ID7sx Base 9 4 5 4 1 5 4 2 Write table Read table 78 Profibus DP communication with a PLC Overview The Profibus DP ID7sx is designed for operation as a slave on the Profibus DP This provides the following possibilities with a master PLC also connected to the Profibus DP e Access to the weight values of the weighing platform connected to the weighing terminal Operation of the weighing platforms connected to the weighing terminal zero set taring setting specified tare values etc e Triggering key presses transmitting data strings or display of texts Profibus DP Profibus SPS DP ID7sx with Profibus interface Data formats All user data are transmitted in a compressed up to 4 word long format Format for transmitting user data from the PLC to the Profibus DP ID7sx Format for the transmission of user data from Profibus DP ID7sx to the PLC Structure of the write and read table The write and read table are similarly structured and contain the following sections Value 16 bit integer or 32 bit floating point for the transmission of weight values application blo
102. tabile weight value can be specified However this value is not exact Specify tare weight Command T A Tare weight weight value __ Unit Specify tare weight The contents of the tare memory are overwritten with the specified tare weight and the net weight is displayed Effect as when the key sequence TARE ENTRY O 9 ENTER is pressed Response Tare weight weight value __ Unit r Command not carried out p im Weighing platform tared with the specified value Command understood parameters defective Command cannot be executed Tare range dropped below Command cannot be executed Tare range exceeded Comments e The contents of the tare memory are overwritten with the specified tare value e On non certified weighing systems the tare weight is automatically rounded off to the current increment e On certified weighing systems Tare range with MultiRange only in first increment range Example Command Response d gt N oO Ul o P D EN N O ESCORT Operating instructions 22008168A 04 10 71 Interface description ID7sx Base Delete tare weight
103. ting instructions 22008168A 04 10 ID7sx Base Additional functions 3 7 Multiplicative tare function The multiplicative tare function is particularly suitable when pallets with identical containers are filled If the number of containers and tare of the individual container are known the ID7sx Base weighing terminal calculates the total tare 1 Press MULT TARE key 2 Enter known tare weight of individual container and confirm with ENTER 3 Enter number of containers and confirm with ENTER When the weighing platform is relieved the total tare value is shown in the display with a negative sign Note With the FUNCTION CHANGE key you can select the weight unit for entering the tare weight Additive tare function With the additive tare function you can subtract the tare of additional containers with a know tare weight for related weighings e g if containers with different weights are filled on one pallet 1 Place container on scale and press ADD TARE key 2 Enter known tare weight and confirm with ENTER The total net weight appears in the weight display Note With the FUNCTION CHANGE key you can select the weight unit for entering the tare weight Sandwich tare With the sandwich tare function you can detect additional tare weights for related weighings without loosing the total gross and total net Example In production or shipping boxes are laid between individual layers in the transport container The we
104. tting Output 1 Delta low Output 2 Delta ok Output 3 Delta high Output 4 Stable Output 5 Setpoint 1 Output 6 Setpoint 2 Output 7 Setpoint 3 Output 8 Setpoint 4 SETPOINT MODE With SETPOINT MODE ON 4 configurable fixed or dynamic set points are available see page 48 Outputs are independent of the weighing functions Set the outputs via the AW706 command see section 6 3 2 I O TEST Testing of the function and state of the inputs and outputs If an input or output is set high the display indicates its number If an input or output is not set low the display indicates Set outputs Switch over the outputs with the keys 1 to 8 of the numerical keypad Set inputs Set inputs e g by connecting a supply voltage 24 V Exit 1 0 TEST Exit the I O test and the master mode with the O key of the numerical keypad Comments e During the I O tests only the keys ZERO SET TARE and ENTER are active e Serial interfaces can be used during the 1 0 test Operating instructions 22008168A 04 10 47 Settings in the master mode ID7sx Base SETPOINT MODE ON defining set points 48 After SETPOINT MODE ON is selected the following input mask appears Example SPI FT A012 WI 1 2345 KG SP2 FL A013 w2 0 5678 KG SP3 DT A0OI2 ALL SP4 Di AOI ALL 4 parameters can be set for each set point a Type of set point FT fixed set point ascending Fl fixed set point descending DT
105. x Base receives data and transmits ACK when it is ready to receive data again The computer receives data and transmits ACK when it is ready to receive data again Operating instructions 22008168A 04 10 41 Settings in the master mode ID7sx Base Set print mode TRANSFER KEY CODE A KEY CODE D KEY DYNAMIC KEY Pac keys PRINT MODE Configure printout on an external printer HANDSHAKE Possible settings e NO HANDSHAKE e XON XOFF PROTOCOL LINE LENGTH Enter number of characters per line Possible settings 1 80 characters Factory setting 40 characters LINE FRAMING Enter ASCII character for line framing Possible settings ASCII O 255 Factory setting ASCII 013 010 Cele REPORT TYPE Assignment of one of two possible printout formats to the configured printer Possible settings e REPORT TYPE A e g for barcode printer e REPORT TYPE A e g for A4 printer CONFIGURATION Configuration of the printouts assigned to the individual keys PRINTOUTS For each offered key the current configuration can be printed out with the key sequence CHANGE CONFIGURATION F possibly several times and PRINT Configuration options e DELETE ALL All blocks of the data string are deleted e DEFAULT SETTING Key specific if existent e CHANGE CONFIGURATION See next section e PAPER FEED Adjustment range O 9 lines e REPORT ON OFF Switch key printout on off AUTOMATIC PRINTOUT Switch automatic printout for tran
106. y set OTHER For other explosion protected barcode readers Settings in the sub blocks COMMUNICATION and MODE as for the blocks RS232 RS422 RS485 CL20mA see section 4 5 1 The PRINT MODE setting is not possible when using barcode readers DESTINATION BLOCK 000 00 Enter the number of the application block and of the subsequent block with which the barcode entry is to be described When a target block is selected barcode information can be read directly into this block without having to press a key beforehand see section 3 14 2 4 5 3 X Configure inputs outputs Condition An 8 l O ID7sx interface is installed and configured on COM5 8 1 0 Configure inputs outputs CONTROL INPUTS Operate inputs internally or externally INTERNAL Factory setting Additional settings CONFIGURE INPUTS Select the desired setting for every input Factory setting Input 1 not in use Input 2 zero setting Input 3 taring Input 4 entry ENTER key Input 5 as not in use Input 8 EXTERNAL Inputs are independent of the weighing functions Read status of the inputs with the AR707 command see section 6 3 2 46 Operating instructions 22008168A 04 10 ID7sx Base Settings in the master mode 8 1 0 Configure inputs outputs CONTROL OUTPUTS INTERNAL EXTERNAL Operate outputs internally or externally Factory setting Additional settings CONFIGURE OUTPUTS Select the desired setting for every output Factory se
107. ying weight values in higher resolution Press X 10 key The weight value is displayed in at least a 10x higher resolution The higher resolution is displayed until the X 10 key is pressed again Note With certified weighing platforms the weight value only appears in a higher resolution as long as the X 10 key is pressed Display gross weight The gross weight can only be displayed when a tare weight has been saved gt Press GROSS key and hold down The gross weight is displayed Specifying dynamic set points Conditions e 8 l O ID7sx interface installed and connected e SETPOINT MODE ON and at least one dynamic set point is configured in the master mode If the specified set point values are exceeded or dropped below digital outputs are set e g for controlling lamps flaps valves etc Dynamic set points can be set for each weighing procedure individually The set points are retained until they are overwritten with a new value or deleted Specifying set points Press the SETP key the entry prompt for the first dynamic set point appears 2 Enter the desired weight value and confirm with ENTER 3 If additional dynamic set points are configured the entry prompt appears for the next dynamic set point 4 Enter the desired weight value and confirm with ENTER b Repeat the procedure until all set points have been entered Deleting set points gt Press the SETP key and delete the value with the CLEAR key Opera

Download Pdf Manuals

image

Related Search

Related Contents

Descargar manual de usuario - Domestic 50 Kw  Was_ist_neu_in_SIMIT_SCE_Version_5_SP1 - kleissler  Le journal d`information du Foyer Schaerbeekois  INSTALLATION INSTRUCTIONS  Que du bonheur ?  Pro-Motion Development Software User's Guide    - Oracle Documentation  USER'S MANUAL TABLE OF CONTENTS FIRE SAFETY TIPS  Mode d`emploi du DELPOU-VAC  