Home
Troubleshooting Guide
Contents
1. SonoGhec Troubleshooting Guide V PASS Positive result Record result according to your department policy X FAIL Negative result Insufficient energy 4294 OUOS Parameters Possible reasons Corrective actions Low energy Cycle time too short the Test a longer cycle let ultrasonic basket or the load Tost PEE sack EEEE cycle time may absorb too much energy est cycle without load functional test Instrument tray and avoid wire mesh trays Instr load Wire mesh trays absorb ultrasonic energy Solid bottom trays are recommended Water level Reflection of ultrasonic energy Refer to the instruction manual for on the surface may change correct water level energy distribution De gassing Dissolved gasses may absorb Refer to the instruction manual for ultrasonic energy proper de gassing Transducers The efficiency of the transducers Redo the functional test and check may decrease with age or the individual transducers individual transducers may be out of order Note 1 The performance after a corrective action needs to be tested again with a new SonoCheck Note 2 The normal time needed for a colour change is below 1 min but in the case of absorbed energy time may take up to several minutes PEREG GmbH A Porschestra e 12 D 84478 Waldkraiburg Ph 49 8638 84195 MAKE T PEREG Fax 49 8638 84162 Email info pereg de www pereg de VISIBLE