        Dual Low Air Loss User Manual
         Contents
1.  can be opened for maintenance     Air Filter and Filter Cap  3     Shelden Healthcare recommends that the filter to  be clean or replace once a month to ensure optimal    performance of the power unit    6 0 Mattress and Power Unit   Installations    1  Remove existing mattress from bed frame      2  Place Shelden Healthcare mattress replacement  on the bed frame with the logo at foot end  Secure  the mattress at each side using anchor strap  Please  verify all bed functions are working properly without    interference before proceeding to the next step      3  Secure the power unit onto the foot board using    hanging hooks      4  Firmly connect the mattress    air hose couplings  to the CPC quick release couplers on the power    unit s air outlet      5  Plug the power unit into a wall outlet and turn  on the main power rocker switch on the left side  panel  The STANDBY CPR orange LED on control  panel will light to acknowledge the power unit is in    standby mode      6  Press the POWER button on the control panel to  inflate the mattress system  MAX INFLATE may be    used to shorten the initial inflation time if needed   7 0 Program Settings     1  Center the patient on the mattress  Adjust    the mattress    internal pressure according to the  patient weight by using the weight  kg  button on  the control panel of the power unit  If the patient  feels the mattress is too soft or firm  increases   or decreases the mattress    internal pressure one  increment at a time a
2.  into wall outlet  outlet     Check if power surge shuts down A power surge may overload the   power unit  circuitry temporarily  Turn the  unit off and check the fuse for  damage  Turn the unit ON again  with normal procedure     Power unit does not respond to Please contact Shelden  possible solutions  Healthcare on 01788 833 989     Check if there is leakage in air  tubes  or air cells                  Please contact Shelden  Healthcare on 01788 833 989     3  alert LED is lit during  operation     Check if mattress    air hose couplers Secure air hose couplers firmly  are properly connected to power  unit s quick release couplers     into place        Inspect air filter for dust or dirt    Clean or replace air filter        Lift mattress coverlet up to check if Check and adjust air tubes     air tubes are kinked or obstructed  positions        Please contact Shelden  Healthcare on 01788 833 989     5  Alternating or static setting is Possible control failure   not available         Air tube is referring to the mean of directing air flow from the power unit to each individual air cell     DUAL LOW AIR LOSS USER MANUAL  8    13 0 Returns for Service    This device is not self serviceable  Service and  repair must be performed by an authorized  technician or representative  All returned device  must be cleaned and disinfected prior to shipping   Unsanitary or soiled systems will be returned    without servicing     14 0 Warranty    Shelden Healthcare warrants the product to 
3.  mode  UPRIGHT  mode is used to prevent patient from bottoming    out in sit up position  Make sure UPRIGHT mode    DUAL LOW AIR LOSS USER MANUAL   4          wearttaraer       Rear Panel Features Figure 5c    mode is switch off when the patient is laying down  to avoid incorrect pressure setting  Press PANEL  LOCK button to lock out function buttons besides  max inflate and power functions from unauthorized  changes  If you wish to cancel panel lock  simply  press and hold the PANEL LOCK button for 3    seconds     Control Panel  Features    Figure 5a    max inflate  Press the MAX INFLATE button to inflate the mattress  rapidly to maximum pressure  Press the button again  to restore to the original setting or this function will    disable automatically after 20 minutes if unattended     Right Side Panel Features  See Figure 5b    Couplers  1     Quick release CPC couplers are used to secure    mattress air hoses to power unit     Left Side Panel Features    See Figure 5b    Rocker Switch  1     Power unit s main power switch  The power unit  automatically goes into STANDBY CPR mode after the    switch is turned on     Power Receptacle  2    Insert power cord firmly into receptacle     Left Side Panel Features    See Figure 5c    Hanging Hooks  1    Hanging hooks are designed to hang thepower unit  on almost any foot board  Hooks are spring loaded    and will fold against the power unit when not in use     ad LOW AIR LOSS MATTRESS REPLACEMENT SYSTEM    Fuse  2     Fuse holder
4.  state  and or facility requirements    for disposal of the unit     7  The use of this system does not replace the  regular repositioning  monitoring  and nursing of    the patient      8  Operating  transportation  and storage    conditions     Temperature  5  C   45  C  Humidity  15    60     CAUTION  ENSURE THAT THERE ARE NO  PROTRUDING OBJECTS  SHARP POINTS OR  BED SPRINGS UNDER THE MATTRESS AS THESE  COULD PUNCTURE THE AIR CELLS     4 0 System Package    Power Unit Package      Power unit x 1    Power cord x 1      Operation manual x 1    Mattress Package      Mattress replacement unit with coverlet x 1  5 0 Features  Control Panel Features    See Figure 5a on page 5     power  on led  standby cpr led   The system is in STANDBY CPR mode and the    orange LED will light when the main power rocker  switch on the left side panel is turned on  Press the  POWER button to turn on the system and the green  LED will light      Please refer to 9 0 Emergency CPR Deflation for  CPR function     patient weight  Simply press the or button to adjust the patient    weight according to each individual  The scale is  only an approximation  Please adjust the weight  setting if the mattress is too soft or firm to suit each    patient     mode  STANDARD mode is for patient who weights in the  range of 35 kg   140 kg     BARIATRIC mode is for patient who weights in  the range of 160 kg   450 kg  Press and hold the  desired mode button for 10 seconds to switch  between Standard and Bariatric
5. 601 1 2 1993 Type BF Applied Part    Maximum Weight Capacity    450 kg    Shelden Healthcare reserves the right to change  prices or design parameters and specifications  without incurring any liability in connection with    such modifications     DUAL LOW AIR LOSS USER MANUAL  10    Shelden Healthcare Ltd   Unit 1A   The Wharf   Stretton Under Fosse    Rugby   Warwickshire   CV23 OPR  Telephone  01788 833 989   Fax  0845 127 50 50   Email  sales shelden healthcare co uk    Web  www pressurerelieving co uk    
6. be  free from defects from the date of purchase   Please inspect all accessories when you purchase  our product  If there is any damage or missing  accessories when you receive the product  please  ask for a replacement from your local distributor    within three days of purchase     The warranty periods for Shelden Healthcare  products are according to the regulations in your  country  the minimum period is 1 year from date   of purchase for the power unit and 1 year for the  mattress and the coverlet  The warranty coverage of  any Shelden Healthcare product is contingent up on  its purchase from an authorized dealer  Warranty for  any product in the Shelden Healthcare product line  will be honored by the official distributor in your    country     Warranty coverage will not be extended to any  product on which the production lot number  has been removed or defaced  on which repair  has been attempted by any person or agency  not authorized by Shelden Healthcare or if in  the sole opinion of Shelden Healthcare that the  system shows evidence of tampering  abnormal  or unreasonable abuse  negligence  accident  or operation withoutregard for the restrictions    specified in the instructions which accompany    the system  This warranty does not cover normal  maintenance such as cleaning  adjustment   lubrication  and updating of equipment or parts   If the damage is result from improper operation   a reasonable service fee and part cost will be    charged     The warranty state
7. bed linens to a   single layer in order to allow the moisture escape  efficiently through the coverlet  Only    breathable     incontinent pads are recommended for use with this    mattress system      3  Avoid using this device near open flames   lighters  or cigarettes  Flammability hazard exists   This device draws air from the surrounding   Thus  cigarette smoking may damage internal    component      4  This system should disinfect thoroughly between    patients to avoid cross contamination      5  Verify patient weight does not exceed bed  frame  bed rails  and this mattress system s weight    capacity  3 0 Warnings     1  Use this mattress with the correct side rails to  ensure the gap between the top of the mattress    and the top of the bed rails meets the EN60601 2     3   LOW AIR LOSS MATTRESS REPLACEMENT SYSTEM    52  Failure to do so could result in serious patient    Injury      2  DO NOT disassemble the power unit if you are  not a qualitied technician  Please contact your local    distributor for all service    3  This product is NOT AP APG protected      4  Re position the patient once awhile is still    necessary when using this mattress      5  Incorrect battery replacement may result in risk  of explosion  Replace only with same or equivalent    battery type recommended by the manufacturer      6  If the equipment needs maintenance  contact  Shelden Healthcare  as soon as possible  For  equipment that is no longer functional  make sure  to follow national 
8. d above is the only warranty  made and is in lieu of all other warranties whether  expressed or implied  including any warranty   of merchantability or fitness for a particular  reason  Shelden Healthcare will not be liable for    consequential or incidental damages or any kind     15 0 Product Specifications    Mattress Dimension  MSR208     36  200x90x20cm  39  200x99x20cm  42  200x107x20cm  48  200x122x20cm  54  200x137x20cm  60  200x153x20cm    Number of Air Cells    20 air cells    Power Control Unit Dimension  SR363     29 5 x 22 5 x 12 0 cm    Material  Power Unit    ABS fire retardant UL    9   LOW AIR LOSS MATTRESS REPLACEMENT SYSTEM    Mattress Cell  100  nylon with TPU lamination  Coverlet    100  nylon with PU backing    Power Input    AC100 120V 50 60Hz5A AC220   240V  50 Hz 5A    Power Consumption    Normal Operation  Max 350W    Power Control Unit Weight    4 6 kg    Power Cord    HO5VV F3x1 00mm2    Fuse Rating    T1A  250V    Operating  amp  Transportation  amp  Storage Conditions    Temperature  5  C   45  C  Humidity  15    60     Atmospheric pressure  700   1060 hPA    Electrical Classification    Class Il Type B  double insulated with or without    functional ground wire     IPXO  do not immerse power unit in any liquid or    spray any liquids directly on the power unit     This system is not AP  APG protected  Continuous    operation     Agency Approval   EN60601 1  EN60601 1 2  EN61000 3 2  EN61000   3 3  UL2601 1   Electromagnetic Compatibility    EN60
9. nd wait for the system to  stabilize before making another change until a    comfortable state is achieved      2  Caregivers should always perform a hand check  by placing their hands underneath patient s pelvis  area to check if there is sufficient air support to    ensure the patient is not bottoming out      3  IMPORTANT  TUCKING THE SHEET IN TIGHTLY  REDUCES THE EFFECTIVENESS OF THE SYSTEM     8 0 Patient Transfer and Transport  Transfer    It is recommended to have the mattress system in  MAX INFLATE mode during transfer  Make sure the    bed is secured before proceeding     Transport    If a transport occasion comes up  detach mattress     air hose couplings from the power unit s CPC quick  release couplers and connect the two air hose  couplings together  Shelden Healthcare Theraflo  system has a safety foam base underneath the air  cells to support the patient for a short period of  time  To resume normal operation  please follow 7 0    Program Setting     DUAL LOW AIR LOSS USER MANUAL l6    9 0 Emergency CPR Deflation    In the case of emergency  press the POWER   button on the control panel to turn the system into  STANDBY CPR mode to deflate the mattress rapidly  in order to perform CPR  Press the POWER button  again and reset the patient weight to resume    therapy     10 0 Cleaning Instruction    Air mattress and power unit must be cleaned  thoroughly between patients to avoid cross  contamination  The following is a suggested  guideline  but local infection con
10. shelden    HEALTHCARE       Dual Low Air Loss  User Manual    Specialist Suppliers to the Healthcare Industry    1      NV AT N DI TACCO NAAN     VV     A J f YK A VAAN  IM WV  _      n    AN  LUVV A   BS _        VIA     TRESS REPLACEMEN    Thank you for purchasing this Shelden Healthcare    pressure relieving mattress system     PLEASE READ THIS USER MANUAL CAREFULLY BEFORE  SETTING UP AND USING THE DEVICE     Pay special attention to the warnings and other   safety information  Use genuine Shelden Healthcare  components to maintain the products optimal  performance  If you do not fully understand all the  instructions  safety precautions  and warnings  do not use  this device  In case you have questions  please contact    Shelden Healthcare on 01788 833 989       SYSTEM    Section  1 0  2 0  3 0  4 0  DU  6 0  7 0    8 0       Description Page    Symbols and Indications   Safety Precautions   Warnings   System Package   Features   Mattress and Power Unit Installations  Program Settings    Patient Transfer and Transport    03    03    03    04    04    06    06    06    Section    9 0    10 0    12 0    13 0    14 0    15 0    Description  Emergency CPR Deflation  Cleaning Instruction  Routine Maintenance  Troubleshooting   Returns for Service  Warranty    Product Specifications    YALL LOW AIR   OSS LISER MANIJAI  JUFAL LVVV AIK ZIJ UOLIA IV AIN AI    shelden    HEALTHCARE       Page  07  07  07  08  09  09    09    1 0 Indications    AY document thoroughly   0  Double ins
11. trol policies should    be followed as well     Mattress cleaning with hand hot water with a PH  neutral detergent  Decontamination by wiping with  a Sodium Hypochlorite solution of 1000ppm in line  with HSG9518 and BHTA guidelines  Please ensure  air mattress and cover are completely dry before    patient use     Machine wash warm water at maximum 71  C  Dry  mattress cover on a cool tumble setting  Do not    iron     WARNING  Always unplug the power unit before  cleaning  Routine cleaning of power unit can be  done by wiping down with damp cloth using  disinfectant and water or mild neutral detergent     Never spray liquids directly on the unit itself     11 0 Routine Maintenance    Remove air filter from the rear panel of the power  unit by opening up the filter cap to clean or replace  the air filter  Inspect the filter for dirt or dust and  clean it with mild soap and water  Reinsert the  dried air filter after cleaning and ensure the cap is  secure  Replace with genuine air filter once a year  is recommended and compressor every two year is    strongly suggested to prolong system lifetime     Only disinfected and dry systems are to be stored   Disconnect the air hoses from the power unit  Roll  up the mattress starting from head end and working    down toward the foot end  Use the straps to fix it     7   LOW AIR LOSS MATTRESS REPLACEMENT SYSTEM    12 0 Troubleshooting    1  Power unit does not function  Check if power cord is firmly Secure power cord into wall  plugged
12. ulated  Class ll     C     Declaration of Conformity to Medical    Attention  Please read enclosed    Device Directive    Afe International Electronics Approval    IPX0    Do not immerse power unit in liquid    or spray liquids directly on power unit    Type B Equipment    Indications    This low air loss mattress system is designed for  patients who are at high risk of pressure of ulcers or  require treatment for existing pressure ulcers  This  device is intended to treat and prevent pressure  ulcers by facilitating blood circulation  reactive  hyperaemia and decreasing interface pressures   Always consult a physician or health professional  before using this mattress system  This pressure  relieving mattress system is suitable for patients  assessed to be high risk and up to catagory 3    pressure ulcers   Contraindications    Certain patient conditions are not suitable for  using this type of device  Contraindications such as    unstable fractures of vertebrae and conditions of    the vertebrae  Always consult a physician or health    professional before using this device     The use of this system does not replace the regular  repositioning  monitoring  and nursing of the    patient     2 0 Safety Precautions     1  To ensure proper operation  please inspect   and verify all parts are setup properly and anchor  securely  Do not place anything on top of the power  unit  Make sure power cord is underneath bed    frame and free from hazard      2  It is recommended to limit 
    