1. 8 Operating transportation and storage conditions Temperature 5 C 45 C Humidity 15 60 CAUTION ENSURE THAT THERE ARE NO PROTRUDING OBJECTS SHARP POINTS OR BED SPRINGS UNDER THE MATTRESS AS THESE COULD PUNCTURE THE AIR CELLS 4 0 System Package Power Unit Package Power unit x 1 Power cord x 1 Operation manual x 1 Mattress Package Double Mattress replacement unit with coverlet x 1 5 0 Features Control Panel Features See Figure 5a on page 5 Power Mute Press power mute button to turn on the power unit and the green LED will light The LED will change to red if abnormal power outage occurs weight kg Simply press weight kg button to adjust the patient weight from 30 kg to 180 kg according to each individual The scale is only an approximation Please adjust the weight setting if the mattress is too soft or firm to suit each patient Caregivers should always perform a hand check by placing their hands underneath patient s pelvis area to check if there is sufficient air support to ensure the patient is not bottoming out Dynamic Static Press Mode to change between dynamic and static mode Dynamic mode will have an alternating cycle time of 10 minutes minutes and static mode to DUAL DOUBLE USER MANUAL 4 pressure monitoring LED Power unit features an integrated pressure senor which can monitor the mattress internal pressure 24 hour to achieve optimal int
2. country the minimum period is 1 year from date of purchase for the power unit and 1 year for the mattress and the coverlet The warranty coverage of any Shelden Healthcare product is contingent up on its purchase from an authorized dealer Warranty for any product in the Shelden Healthcare product line will be honored by the official distributor in your country Warranty coverage will not be extended to any product on which the production lot number has been removed or defaced on which repair has been attempted by any person or agency not authorized by Shelden Healthcare or if in the sole opinion of Shelden Healthcare that the system shows evidence of tampering abnormal or unreasonable abuse negligence accident or operation withoutregard for the restrictions specified in the instructions which accompany the system This warranty does not cover normal maintenance such as cleaning adjustment lubrication and updating of equipment or parts If the damage is result from improper operation a reasonable service fee and part cost will be charged The warranty stated above is the only warranty made and is in lieu of all other warranties whether expressed or implied including any warranty of merchantability or fitness for a particular reason Shelden Healthcare will not be liable for consequential or incidental damages or any kind 15 0 Product Specifications Mattress Dimension MSR207 Individual Model Number Power Cont
3. shelden HEALTHCARE Dual Double User Manual 1 Thank you for purchasing this Shelden Healthcare pressure relieving mattress system PLEASE READ THIS USER MANUAL CAREFULLY BEFORE SETTING UP AND USING THE DEVICE Pay special attention to the warnings and other safety information Use genuine Shelden Healthcare components to maintain the products optimal performance If you do not fully understand all the instructions safety precautions and warnings do not use this device In case you have questions please contact Shelden Healthcare on 01788 833 989 Section 1 0 2 0 3 0 4 0 5 0 6 0 7 0 8 0 Description Page Symbols and Indications Safety Precautions Warnings System Package Features Mattress and Power Unit Installations Program Settings Patient Transfer and Transport 03 03 03 04 04 06 06 06 Section 9 0 10 0 12 0 13 0 14 0 15 0 shelden HEALTHCARE Description Emergency CPR Deflation Cleaning Instruction Routine Maintenance Troubleshooting Returns for Service Warranty Product Specifications Page 07 07 07 08 09 09 09 1 0 Indications Attention Please read enclosed document thoroughly Double insulated Class ll IN C Declaration of Conformity to Medical Device Directive International Electronics Approval IPX0 Do not immerse power unit in liquid or spray liquids directly on power
4. unit PN Type B Equipment Indications This air alternating mattress system is designed for patients who are at high risk of pressure of ulcers or require treatment for existing pressure ulcers This device is intended to treat and prevent pressure ulcers by facilitating blood circulation reactive hyperaemia and decreasing interface pressures Always consult a physician or health professional before using this mattress system This pressure relieving mattress system is suitable for patients assessed to be high risk and up to catagory 3 pressure ulcers Contraindications Certain patient conditions are not suitable for using this type of device Contraindications such as unstable fractures of vertebrae and conditions of the vertebrae Always consult a physician or health protessional before using this device The use of this system does not replace the regular repositioning monitoring and nursing of the patient 2 0 Safety Precautions 1 To ensure proper operation please inspect and verity all parts are setup properly and anchor securely Do not place anything on top of the power unit Make sure power cord is underneath bed frame and free from hazard 2 It is recommended to limit bed linens to a single layer in order to allow the moisture escape efficiently through the coverlet Only breathable incontinent pads are recommended for use with this mattress system 3 Avoid using this device near open
5. ernal pressure and to ensure maximum pressure relief LED indicates the system is monitoring the mattress internal pressure while the compressor is inflating Control Panel Features Figure 5a alert LED This alert notifies medical staff the mattress has insufficient pressure Audible alarm will sound for 15 seconds at each cycle and the indicator will stay on until the problem has been solved Right Side Panel Features See Figure 5b Rocker Switch 1 Power unit s main power switch Couplers 2 Quick release female couplers are used to secure mattress air hoses to power unit Power Receptacle 3 Insert power cord firmly into receptacle Rear Panel Features See Figure 5c Convenient Handle 1 It provides additional gripping surface for user to carry the power unit Hanging Hooks 2 Hanging hooks are designed to hang the power unit on almost any foot board Fuse 3 Fuse holder can be opened for maintenance Air Filter and Filter Cap 4 Carilex recommends that the filter to be clean or replace once a month to ensure optimal pertormance of the power unit Bi PRESSURE RELIEVING ALTERNATING AIR MATTRESS OVERLAY SYSTEM 6 0 Mattress and Power Unit Installations 1 Place Shelden Healthcare mattress overlay on top of a 1 or 2 foam mattress with the logo at foot end Secure the mattress using the elastic straps at four corners Please verify all bed functions are working properly
6. flames lighters or cigarettes Flammability hazard exists This device draws air from the surrounding Thus cigarette smoking may damage internal component 4 This system should disinfect thoroughly between patients to avoid cross contamination 5 Verify patient weight does not exceed bed frame bed rails and this mattress system s weight capacity 3 0 Warnings 1 Use this mattress with proper side rails to ensure 3 PRESSURE RELIEVING ALTERNATING AIR MATTRESS OVERLAY SYSTEM the gap between the side rail and top of the mattress is small enough to prevent patient trom getting his or her head or neck into the gap Fail to do so could result in serious patient injury 2 DO NOT disassemble the power unit if you are not a qualified technician Please contact your local distributor for all service 3 This product is NOT AP APG protected 4 Re position the patient once awhile is still necessary when using this mattress 5 Incorrect battery replacement may result in risk of explosion Replace only with same or equivalent battery type recommended by the manufacturer 6 If the equipment needs maintenance contact Shelden Healthcare as soon as possible For equipment that is no longer functional make sure to follow national state and or facility requirements for disposal of the unit 7 The use of this system does not replace the regular repositioning monitoring and nursing of the patient
7. lorite solution of 1000ppm in line with HSG9518 and BHTA guidelines Please ensure air mattress and cover are completely dry before patient use Machine wash warm water at maximum 71 C Dry mattress cover on a cool tumble setting Do not iron 11 0 Routine Maintenance Remove air filter from the rear panel of the power unit by opening up the filter cap to clean or replace the air filter Inspect the filter for dirt or dust and clean it with mild soap and water Reinsert the dried air filter after cleaning and ensure the cap is secure Replace with genuine air filter once a year is recommended and compressor every two year is strongly suggested to prolong system lifetime Only disinfected and dry systems are to be stored Disconnect the air hoses from the power unit Roll up the mattress starting from head end and working down toward the foot end Use the straps to fix it 7 PRESSURE RELIEVING ALTERNATING AIR MATTRESS OVERLAY SYSTEM 12 0 Troubleshooting 1 Power unit does not function Check if power cord is firmly Secure power cord into wall plugged into wall outlet outlet Check if power surge shuts down A power surge may overload the power unit circuitry temporarily Turn the unit off and check the fuse for damage Turn the unit ON again with normal procedure Power unit does not respond to Please contact Shelden possible solutions Healthcare on 01788 833 989 Check if there is leakage in ai
8. r tubes or air cells Please contact Shelden Healthcare on 01788 833 989 3 alert LED is lit during operation Check if mattress air hose couplers Secure air hose couplers firmly are properly connected to power unit s quick release couplers into place Inspect air filter for dust or dirt Clean or replace air filter Lift mattress coverlet up to check if Check and adjust air tubes air tubes are kinked or obstructed positions Please contact Shelden Healthcare on 01788 833 989 5 Alternating or static setting is Possible control failure not available Air tube is referring to the mean of directing air flow from the power unit to each individual air cell 9 9 P DUAL DOUBLE USER MANUAL 18 13 0 Returns for Service This device is not self serviceable Service and repair must be performed by an authorized technician or representative All returned device must be cleaned and disinfected prior to shipping Unsanitary or soiled systems will be returned without servicing 14 0 Warranty Shelden Healthcare warrants the product to be free from defects from the date of purchase Please inspect all accessories when you purchase our product If there is any damage or missing accessories when you receive the product please ask for a replacement from your local distributor within three days of purchase The warranty periods for Shelden Healthcare products are according to the regulations in your
9. rol Unit SR363 Mattress SR208 Coverlet SR611 Electrical Inout AC100 120V 50 amp 60Hz or AC220 240V 50Hz Agency Approval EN60601 1 EN60601 1 2 EN61000 3 2 EN61000 3 3 UL2601 1 Dimensions Power Control Unit 29 5 x 22 5 x 12cm Mattress Set Standard inflated 200 x 140 x 18cm King Size inflated 200 x 15 2 x 18cm Weight Power Unit approx 2 5kg Mattress Set 10kg 9 PRESSURE RELIEVING ALTERNATING AIR MATTRESS OVERLAY SYSTEM Power Consumption Max 20W Material Power Unit ABS fire resistant UL Mattress Cell 100 nylon with TPU lamination Coverlet 100 nylon with PU backing Operating Transportation amp Storage Temperature Humidity 50C 450C 15 60 Maximum Weight Capacity 150kg The power unit is not AP APG protected Shelden Healthcare reserves the right to change prices or design parameters and specifications without incurring any liability in connection with such modifications DUAL DOUBLE USER MANUAL I 10 Shelden Healthcare Ltd Unit 1A The Wharf Stretton Under Fosse Rugby Warwickshire CV23 OPR Telephone 01788 833 989 Fax 0845 127 50 50 Email sales shelden healthcare co uk Web www pressurerelieving co uk
10. sport Transfer lt is recommended to have the mattress system at maximum weight setting during transfer Make sure the bed is secured before proceeding Transport If a transport occasion comes up two options are available 1 Detach the mattress air hose couplings from the power unit s quick release couplers and connect the two air hose couplings together to retain air in the mattress and the mattress will stay inflated for approx 2 hours 2 Unplug the power unit s power cord from the wall outlet and the mattress should stay inflated for approx 40 50 minutes To resume normal operation please follow 7 0 Program Setting DUAL DOUBLE USER MANUAL 16 9 0 Emergency CPR Deflation In the case of emergency pull hard to remove the mattress CPR latch for emergency deflation and turn off the power unit by pressing the power mute button on the control panel at the same time The air will discharge from the mattress with patient s own weight To resume normal operation simply re insert CPR latch securely press the power mute button again and reset the patient weight 10 0 Cleaning Instruction Air mattress and power unit must be cleaned thoroughly between patients to avoid cross contamination The following is a suggested guideline but local infection control policies should be followed as well Mattress cleaning with hand hot water with a PH neutral detergent Decontamination by wiping with a Sodium Hypoch
11. without interference before proceeding to the next step 2 Secure the power unit onto the foot board using hanging hooks 3 Check if the mattress CPR latch is firmly inserted and the CPR tag is in place 4 Firmly connect the mattress air hose couplings to the quick release couplers on the power unit s air outlet 5 Plug the power unit into a wall outlet and turn on the main power rocker switch on the right side panel 6 Press power mute button to inflate the mattress system Please wait for approx 30 seconds until the power unit is ready before making any program adjustment 7 Wait for approx 30 minutes for the mattress to inflate fully 7 0 Program Settings 1 Center the patient on the mattress Adjust the mattress internal pressure according to the patient weight by using the weight kg button on the control panel of the power unit Ifthe patient feels the mattress is too soft or firm increases or decreases the mattress internal pressure one increment at a time and wait for the system to stabilize before making another change until a comfortable state is achieved 2 Caregivers should always perform a hand check by placing their hands underneath patient s pelvis area to check if there is sufficient air support to ensure the patient is not bottoming out 3 IMPORTANT TUCKING THE SHEET IN TIGHTLY REDUCES THE EFFECTIVENESS OF THE SYSTEM 8 0 Patient Transfer and Tran