1. John Chubb Instrumentation Ltd Unit 30 Lansdown Industrial Estate Gloucester Road John Chubb Cheltenham GL51 8PL UK Tel 44 0 1242 573347 instru mentation Fax 44 0 1242 251388 email jchubb jci co uk User Manual JCI 140 Static Monitor UM140 issue 21 September 2007 Contents Introduction The JCI 140 Static Monitor Use of instrument CE Conformance RoHS and WEEE Directives Specification features NMnNBWN RK Registered in England No 5798844 VAT GB 392 007 172 Registered Office Unit 30 Lansdown Industrial Estate Gloucester Road Cheltenham GL51 8PL UK USER MANUAL JCI 140 STATIC MONITOR A compact instrument for easy sensitive and reliable measurements and for monitoring of electrostatic conditions 1 INTRODUCTION Static electricity causes problems in many areas of industry e ignition of flammable gases e shocks to personnel e attraction of airborne dust and debris e cling of thin films and light fabrics e damage to semiconductor devices e upset of operation of microelectronic system operation and on the other and there are many constructive applications for static electricity Static electricity arises when materials in contact are separated and the speed of charge movement on either surface and away to earth is long compared to the time for the surfaces to move apart Many operations in industry and normal life involve the contact and separation of materials that are poo
2. Introductory Workshop in Electrostatics on the JCI Website www jci co uk Electrostatics Wshop 00 html 3 USE OF INSTRUMENT 3 1 Getting started By the time you read this User Manual you will already have opened the JCI 140 carrying case Now take out the instrument Hold the instrument in the hand with the liquid crystal display upwards and the gold plated sensing aperture away from the body Take the earth bonding lead yellow out of the case and connect it to the earth bonding point on the right hand side of the back of the instrument Connect the other end of the lead to an earth point using either the Durable Dot connector or the crocodile clip Now switch on the instrument The ON OFF switch with the black slider is towards the top right hand side of the back plate The switch has 3 positions left is off middle is ON for a sensitivity of 2000V full scale and fully to the right is for a sensitivity of 20 000V full scale When the instrument is switched on you should hear the motor inside start and see the LCD start to show numbers A piece of plastic polythene bag biro shaft rubbed and held in front of the sensing aperture will make the displayed numbers change and give larger readings when closer 3 2 Aspects of operational use Earth bonding The instrument needs to be bonded to earth via the bonding point on the back if the readings are to be meaningful Handheld operation The instrument can be held by the side g
3. major build up of static on people on industrial plant or on materials This requires effective linking to earth With metal plant and equipment this means good and reliable earth bonding links you can see and check With many materials it is not so easy Most plastics are bad conductors of electricity 1 e charge does not move easily over or through the material To enable such materials to dissipate static fairly quickly they need to have additives included in their manufacture or have surface treatment Static charge needs to be able to leak away to earth over and through materials within a fraction of a second Role of measurements How do you know if you have problems with static how do you find the source and cause of problems how do you know if remedial actions are effective and how do you choose appropriate material The answer is measurements JCI manufactures a number of instruments to measure static conditions and to assess the suitability of materials for particular applications and to avoid risks and problems JCI 155v5 Charge Decay Test Unit 2 THE JCI 140 STATIC MONITOR The JCI 140 is a compact sensitive field mill instrument set to measure the voltage of surfaces at a defined distance The voltage sensitivity can be switched between 2000 and 20 000 volts full scale for 100mm 4 separation Measurements are shown on a 3 digit liquid crystal display with decimal point and low battery indication An audio alarm gives warnin
4. ve brown 8 Signal outputs via 8 pin miniature DIN socket signal output 2 0V for full scale of LCD sensitivity range 0 2V with alarm state indication by superimposed pulse signal remote range setting control from low sensitivity range by applying 5V to sensitivity indication line Earth bonding combination 10mm Durable Dot and 4mm bayonet pin earth bonding point Supplied with earth bonding cord Dimensions 34 x 66 x 149 5mm overall Weight 320g Calibration JCI 140 instruments are set up in manufacture with measurements whose accuracy is traceable to NPL standards Option for formal Calibration to BS 7506 Part 2 1996 JCI 140 Static Monitor
5. 80 100 120 140 160 Separation distance mm It also needs to be noted that lower readings will be seen with small targets and where the target is against an earthed backing surface as shown below Response of JCI 140 to targets of various diameters with open and earthed backing at 0 20 40 60 80 100 120 140 160 Diameter of target mm JCI 140 fequency response 0 20 40 60 80 100 120 140 160 180 200 Frequency Hz Frequency response The response of standard JCI 140 instruments is 3dB at 120Hz and for JCI 140F instruments is 3dB at 400Hz Signal noise Noise of the analogue output signals is typically around 1mV r m s for standard JCI 140 instruments and around 10mV for JCI 140F Calibration The JCI 140 can be formally calibrated at JCI using measurements whose accuracies are traceable to National Standards Calibration procedure is described in the British Standard Methods for measurements in electrostatics BS 7506 Part 2 1996 See also JCI Website Calibration remains valid for 12 months so long as there has been no unauthorised access inside of the instrument Alarm setting The threshold for triggering the audio alarm may be set between zero and full scale of the LCD using a 2 mm diameter screwdriver through the hole to the left of the zero setting hole in the back plate The threshold may be set using a piece of charged plastic to create the desired reading The threshold level of t
6. d the cover slid closed Batteries should not be left in the instrument for extended periods as any fluid leakage from an old battery could cause corrosion and damage Well over 4 hours continuous operation can be expected from a new PP3 battery External power supply JCI 140 and JCI 140F instruments may be used with an external power supply for long term continuous monitoring observations The power supply required is 12V isolated from earth Power may be supplied to the instrument either via the 8w mini DIN connector or via the 2 1mm d c power connector in the back cover A 12V 250mA regulated supply e g from an isolated supply wall cube with center positive will be suitable Attachments JCI 140 instruments can be used in conjunction with proprietary JCI attachments for measurement of charge JCI 147 characteristics of air ionisation JCI 145 and for zero current drain measurement of electrostatic voltages JCI 148 4 CE CONFORMANCE John Chubb Instrumentation Ltd of Unit 30 Lansdown Industrial Estate Gloucester Road Cheltenham GL51 8PL declares as designer and manufacturer of the JCI 140 range of Static Monitor instruments that the design and construction of these instruments JCI 140 and JCI 140F conform to the requirements of the EC Directive on Electromagnetic Compatibility EMC 89 336 EEC to Standards EN 50081 1 1992 and EN 50082 1 1992 These instruments also conform to the requirements of the Electrical Equipment Sa
7. fety Regulations 1994 S I 1994 3260 5 RoHS and WEEE Directives JCI electrostatic measuring instruments are not required to conform to the RoHS Directive because they come within Category 9 exemption To comply with the requirements of the EC WEEE Waste Electrical amp Electronic Equipment Directive all JCI instruments at the end of their useful life should be returned to JCI so they can be disposed of or recycled in an environmentally appropriate manner 6 SPECIFICATION FEATURES Sensitivity 2000V amp 20 000V full scale 1V and 10V resolution at 100mm Response 3dB at 120Hz for JCI 140 3dB at 400Hz for JCI 140F Noise within 3mV pk pk on analogue output equivalent to 3V pk pk at 100mm 2000V range on standard JCI 140 Zero stability within 10 volts on 2000V range Accuracy within 2 FSD Linearity within 1 FSD Display 3 digit liquid crystal display of surface voltage in kilovolts at 100mm with polarity and LO BATT indication Audio alarm pulsing audio output when reading above user set level or over range Controls on off slide switch OFF ON range 1 ON range 2 screwdriver zero adjustment hole Z in back cover screwdriver alarm level setting hole adjacent to Z hole remote sensitivity selection via red analogue output lead Power supply PP3 replaceable battery External power supply unit 12V 250mA floating from earth via 2 1mm d c connector or via 8w mini DIN ve blue 6
8. g of any voltages above a user set threshold level The electric field sensitivity at the sensing aperture is about 23V m fora reading of 001 on the 2kV range The frequency response of instruments enables measurements to be made of voltages and electric fields created by mains frequency sources As afield mill instrument there is no need to switch on in a static free environment and no need for measurements to be made within a limited time as applies with simple induction probe type instruments The novel mode of operation of the JCI 140 with no earthing of the rotor provides quiet stable operation with fast response and long operational life Interpretation of the readings obtained with the JCI 140 needs to be approached with care For instance a charged plastic film lying on an earthed surface may show only a low surface voltage but this and the associated electrostatic energy can increase dramatically when it is lifted up The significance of readings depends very much on the practical situation and the type of risk or problem involved The user needs to be aware of the need to assess the significance of readings appropriately If in doubt it will be helpful to consult some of the Codes of Practice or Standards available for example The control of undesirable static electricity BS 5958 Part 1 1991 and or Methods for measurements in electrostatics BS 7506 Part 1 1995 and Part 2 1996 There are also the Notes from the
9. he instrument as supplied is set at an LCD reading around full scale The setting may be reduced by turning the 20 turn potentiometer clockwise until the alarm sounds at the desired LCD reading level The reading may be varied around the threshold by varying the distance from the piece of charged plastic Flammable atmospheres The JCI 140 and JCI 140F instruments are not Certified for use in areas where flammable gases are or may be present Only BASEEFA or similarly Certified instruments should be used in such atmospheres Care of instrument JCI 140 instruments are designed to be easy to use in normal work areas They will withstand normal handling but may be damaged if knocked or dropped Care should be taken in use and storage to avoid dust dirt or fluff entering the sensing apertures This will impair operation as will be shown by unstable readings and or an appreciable shift of zero setting It may be possible to remove dust or dirt with a fine brush or by blowing or using an aerosol spray cleaner If the problem persists the instrument should be returned to JCI for servicing Low Battery LO BATT indication on the display will start to flash when the battery voltage has fallen to about 7 3V Performance will remain satisfactory for a time after this but the battery should be renewed for reliable operation Effect of ionized air Operation is generally immune to the presence of ionized air This is because ionization currents are in quadra
10. r or bad conductors of electricity notably plastics Common examples are the occurrence of sparks when combing the hair or taking off an acrylic jumper shocks when getting out of a car and shocks after walking across a nylon carpet and touching earthed conductors The separation of such surfaces is often associated with retention of static charge Voltages of several thousand volts and up to 20kV can be generated on our bodies as we go about normal daily activities When we touch a metal object and feel a shock or for example when we hear a crackle as we take off an acrylic jumper we create an electrical discharge This may give us a shock damage a microelectronic circuit or couple into the computer and upset computer operation Manufacturing operations may involve the handling of thin sheet materials for instance in the printing and packaging industries Static can cause sheets to stick together or be attracted to nearby surfaces This attraction is a particular problem with lingerie that may tend to cling to the body It will cause attraction of dust and debris from the air so packages become dirty when left on the Supermarket shelf With powders for instance in the pharmaceutical industry material flow may be restricted In flammable atmospheres static build up leading to the occurrence of sparks can cause explosions and fires with the risk of loss of life and damage to plant and people Controlling static The secret is to prevent any
11. rooves with the display uppermost facing towards the operator The hand should be towards the rear of the instrument to minimise the influence on sensitivity Switch on The instrument is switched on by the slide switch at the right side of the back 2 surface The switch positions are marked 0 for off 1 for the 2 kV range of sensitivity and 2 for the 20 kV range Static conditions at switch on No special precautions need be taken about electric fields around the instrument at switch on Zeroing It is best to use a clean metal cup over the sensing aperture touching the front of the instrument for checking the zero reading and during zero adjustment A hand over the sensing aperture is convenient but not as good The zero setting is adjusted through hole Z in the back plate Allow a few seconds for stabilising after switch on or adjustment Sensitivity The analogue output signal may be used for remote display and for recording on a paper chart recorder or using a digital storage oscilloscope e g a Picoscope While the sensitivity is manually set for low sensitivity the operating sensitivity can be controlled remotely by holding the range output line high 5V to select the 2kV range The sensitivity varies with separation distance from surfaces as shown VARIATION OF SENSITIVITY OF JCI 140 WITH DISTANCE N a N Reading for 1kV on large plate an o a o 0 20 40 60
12. ture to electric field signals and so are removed by phase sensitive detection Signal outputs The 8w miniature DIN connector in the back cover provides the following signal outputs It also provides opportunity for power supply input for long term continuous monitoring operation 8 pin miniature DIN l black earth 2 white cable sheath range alarm 3 red level and pulse signal 4 yellow signal 5 violet 2 0V signal FSD of LCD ve supply 6 blue 7 green ve supply 8 brown The signal cable supplied with the instrument should be used to ensure conformance to EMC requirements The analogue output signal may be used for remote display and for recording on a paper chart recorder or using a digital storage oscilloscope A Picoscope www picotech com provides a convenient and low cost way to turn a PC into a digital storage oscilloscope with opportunity to display observations and to transfer these into a Spreadsheet for numerical analysis Manual range sensitivity selection can be over ridden remotely by holding the range output line low 0V for 2kV range or high 5V for 20kV range Batteries JCI 140 and JCI 140F instruments are operated by a 9V PP3 battery Access to the battery compartment is achieved by turning the instrument over and sliding the battery compartment cover forward Take the battery out and carefully remove the connector The snap on connector should be securely connected to the new battery before the battery is replaced an
Download Pdf Manuals
Related Search
Related Contents
los proyectos hacen los objetos eternos, las modas los FR - Implants dentaires les transparents Samsung 320W 2.1Ch Soundbar HW-J551 User Manual HP SW350 Product Information For Dummies Microsoft Silverlight 4 DIVISION 08 – APERTURAS Fortress USB User Manual MyScreen Mix antiBacterial Copyright © All rights reserved.