1. User s Manual Protector Series Evidence Drying Cabinets Models 3400000 3400001 3400002 3400003 3400004 3400010 To receive important product updates complete your product registration card online at register labconco com Protecting your laboratory environment ax 7 LABCONCO Labconco Corporation 8811 Prospect Avenue Kansas City MO 64132 2696 800 821 5525 816 333 8811 FAX 816 363 0130 E MAIL labconco labconco com HOME PAGE www labconco com Copyright 2010 Labconco Corporation All rights reserved The information contained in this manual and the accompanying products are copyrighted and all rights reserved by Labconco Corporation Labconco Corporation reserves the right to make periodic design changes without obligation to notify any person or entity of such change Warranty Labconco provides a warranty on all parts and factory workmanship The warranty includes areas of defective material and workmanship provided such defect results from normal and proper use of the equipment The warranty for all Labconco products will expire one year from date of installation or two years from date of shipment from Labconco whichever is sooner except the following Purifier Logic Biological Safety Cabinets carry a three year warranty from date of installation or four years from date of shipment from Labconco whichever is sooner SteamScrubber amp FlaskScrubber Glassware Washe
2. Adjustment of the speed control varies the volume of air drawn through the liner while the unit is operating Ideally smoke released into the intake prefilter should traverse the liner and be drawn into the exhaust prefilter in approximately 5 seconds or less Using a small screwdriver adjust the speed control to give the required face velocity See Figure 6 3 to locate the speed control The air velocity is increased by turning the speed control counter clockwise and clockwise to decrease velocity Product Service 1 800 522 7658 27 Chapter 7 Troubleshooting Refer to the following table if the cabinet fails to operate properly If the suggested corrective actions do not solve the problem contact Labconco for additional assistance PROBLEM Cabinet blower and lights won t turn on UV light not working UV light is dim or flickering CAUSE Unit not plugged into outlet Circuit breaker s tripped Door is open UV Security Lock not on Lamp is defective Lamp wiring is disconnected Defective lamp ballast Lamp is defective or is at end of operating lifetime Defective lamp ballast CORRECTIVE ACTION Plug the cabinet into appropriate electrical service Check connection to the top of cabinet Reset circuit breakers Close door UV light will not work with the door open The key switch for the UV Security Lock must be turned on Replace defective lamp Inspect lamp wiring Replac
3. 3400010 only v Open the cabinet door and ensure that the drain strainer is in place Open the lower door and turn the cabinet water supply valve on Use the spray gun in the Shower setting ensuring proper operation Turn the gun and cabinet supply valve off Push the control switch to the Drain Pump position The majority of water should be pumped out of the cabinet and to the floor drain Some water may remain in the bottom of the cabinet due to surface tension or the floor not being level In this case use a small squeegee or sponge to direct the remaining water to the drain Shut off the drain pump when finished Product Service 1 800 522 7658 11 Chapter 4 Performance Features and Safety Precautions The Protector evidence drying cabinet operates using the following principles e Directional airflow e Prefiltration of large particulates by an inlet prefilter in the door and an exhaust prefilter in the top of the liner e Filtration and retention of particulates by a High Efficiency Particulate Air HEPA filter e Adsorption of organic odorous chemicals by a granular activated carbon filter The major components in the cabinet are e The prefilters e The blower to force air through the cabinet e The HEPA filter e The granular activated carbon filter 12 Product Service 1 800 522 7658 Chapter 4 Performance Features and Safety Precautions Directional Airflow Directional airflow plays a crit
4. material around the top of the unit 3 Plug the power cord into the back of the cabinet and then into an appropriate outlet Install the Key into the UV Security lock on the upper panel on those models equipped with a UV light 10 Product Service 1 800 522 7658 Chapter 3 Getting Started Initial Operation Prior to use with evidence you should perform the following operational tests to ensure the cabinet is operating properly e Airflow Smoke Pattern Test v With the cabinet fan ON introduce smoke from a cigarette chemical smoke stick etc into the prefilter at the bottom of the door The smoke should flow directly into the cabinet and then into the upper prefilter located at the top of the cabinet liner The smoke should move from the door to the intake prefilter in approximately 5 seconds or less If the smoke does not flow as described see the Adjusting the Fan Speed section of Chapter 6 e UV Light and Interlock Tests With the door closed turn the Fan UV Light switch to UV and the UV Security Keylock to ON The UV light will then turn on Open the door and the UV light will turn off Close the door and the UV light should relight Turn the UV Security Keylock to the OFF position and the UV light will turn off If the UV light does not work as described call Labconco s Product Service Department at 1 800 522 7658 or 816 333 8811 e Spray Gun and Drain Pump and Tests Model
5. the shelf to swing down into a vertical position Drying in the Cabinet Start up Models 34000 00 thru 04 Turn off UV light and turn the UV Security switch to OFF Remove the key and store it in a safe place All Models 20 Ensure the drain strainer is in place Turn the switch to Fan Check the condition of the prefilters and note if the Filter System light is on Wash hands and arms thoroughly with germicidal soap Wear a long sleeved lab coat with knit cuffs and over the cuff rubber gloves Use protective eyewear Wear a protective mask if appropriate Product Service 1 800 522 7658 Chapter 5 Using The Cabinet Wipe Down e Open the door to its full open position e Wipe down the interior surfaces of the cabinet with 70 ethanol or a suitable disinfectant and allow to dry CAUTION Do not use bleach on the stainless steel liner Use disinfectants compatible with stainless steel Loading Evidence e Load the evidence so that air flows around all surfaces of it Do not overload the cabinet Do not place objects directly against the side walls e Do not obstruct the intake prefilter bottom of the door or the exhaust air discharge on top of the cabinet e Large objects should not be placed close together e Slowly close the door and use a padlock not supplied or secure it with a chain of custody seal Unloading Evidence e Wear a long sleeved lab coat with knit cuffs and over the cuff ru
6. the switch to the Drain Pump position starts the drain pump Do not operate the pump dry for more than 5 minutes Filter System Light The Filter System Light will illuminate if the differential pressure across the filters becomes excessive If lit first inspect or replace the two prefilters If lit with clean prefilters replace the HEPA filter NOTE If the Filter System Light illuminates during use finish the drying of the evidence in the cabinet and then correct the problem before using the cabinet again Figure 5 1a F ce co Evidence Drying Cabinet ap Product Service 1 800 522 7658 19 Chapter 5 Using The Cabinet Shelf Operation The three drying shelves on the left side wall can be positioned horizontally to dry small pieces of evidence or swung down against the left side wall of the liner if needed To position the shelves in the up horizontal position 1 Pivot the desired shelf until it is above horizontal the frame of the shelf is higher than the retaining clip on the back wall of the liner Push the shelf back all the way Slowly lower the shelf to a horizontal position ensuring that the frame of the shelf engages the retaining clip on the back wall of the liner The shelf is now ready to use To fold the shelf down vertical position 1 Lift up on the rear frame of the shelf to release it from the rear retaining clip Pull the shelf towards the front of the liner Allow
7. TY PRECAUTIONS Directional Airflow Prefilters HEPA Filters Blower Cabinet Air Intake and Exhaust Ultraviolet UV Lamp Safety Precautions WWwWWwwWWNN dN P moewowmnMspRA 12 13 14 14 15 15 15 16 17 CHAPTER 5 USING THE CABINET Models 34000 00 thru 04 Model 3400010 only Shelf Operation Drying in the cabinet CHAPTER 6 MAINTAINING THE CABINET Routine Maintenance Schedule Service Operations Door Prefilter Replacement Cabinet Prefilter Replacement Granular Activated Carbon and HEPA Filter Replacement Changing the UV Lamp Models 34000 00 thru 04 Resetting a Circuit Breaker Setting the Speed Control CHAPTER 7 TROUBLESHOOTING APPENDIX A COMPONENTS APPENDIX B DIMENSIONS APPENDIX C SPECIFICATIONS APPENDIX D ACCESSORIES DECLARATION OF CONFORMITY 18 18 19 19 20 20 22 23 23 24 24 24 25 26 26 27 28 29 30 31 32 33 34 35 Chapter 1 Introduction Congratulations on the purchase of a Labconco Protector Evidence Drying Cabinet The cabinet is designed to protect you and the room environment from biohazardous aerosols and offensive odors while drying evidence The Protector Series cabinet is the result of years of experience in manufacturing laboratory equipment and users like you suggested many of its features to us The Evidence Drying cabinet models offer many unique features to enhance performance and flexibility To take full advantage of them please acqua
8. ain Requirements If you intend to connect the optional drain kit or if you have Model 3400010 you will need to have the Evidence Drying Cabinet within 6 feet 1 8m of a suitable floor drain Note The floor drain must be located below the level of the drain hose exiting the cabinet for all models even the drain pump equipped model 3400010 This will allow complete drainage of the system during wash down operations The cabinet and floor drain should be located in an area not subject to freezing temperatures Water Supply Requirements 3410010 only If you intend to connect the water supply to wash down the cabinet locate the cabinet within 6 feet 1 8 m of a domestic water line equipped with a water shut off valve ending in a 3 4 inch male garden hose connector Water system pressure should be 10 40 PSIG 0 7 2 8 Bar Product Service 1 800 522 7658 3 Chapter 3 Getting Started Now that the installation site is properly planned you are ready to inspect and install the Protector evidence drying cabinet This chapter covers how to e Unpack and move the cabinet e Install the cabinet e Connect the electrical supply source Tools required for installation the cabinet include two 1 2 wrenches a flat blade screwdriver a 2 Phillips screwdriver scissors or a box knife Note The cabinet weighs approximately 300 Ibs 136 kg The shipping pallet allows for lifting with a mechanical lift truck or floor jack If you mu
9. bber gloves Use protective eyewear Wear a protective mask if appropriate e Carefully remove objects in the cabinet Evidence that was considered biohazardous when loaded should be handled as biohazardous Models 34000 00 thru 04 Wipe Down e Wear a long sleeved lab coat with knit cuffs and over the cuff rubber gloves Use protective eyewear Wear a protective mask if appropriate e Wipe down the interior surfaces of the cabinet with 70 ethanol or a suitable disinfectant and allow to dry e Inspect and clean the prefilters at the bottom of the door and at the top of the liner e Dispose of rubber gloves appropriately and have lab coat laundered properly e Wash hands and arms thoroughly with germicidal soap Product Service 1 800 522 7658 21 Chapter 5 Using The Cabinet Models 3400010 Wash Down Wear a long sleeved lab coat with knit cuffs and over the cuff rubber gloves Use protective eyewear Wear a protective mask if appropriate Wipe down the interior surfaces of the cabinet with 70 ethanol or a suitable disinfectant and allow adequate contact time Turn the cabinet supply valve on and ensure the spray gun is in the Shower position Turn the switch to Drain Pump Gently wash down the walls and back of the cabinet s interior avoiding splashing whenever possible Do not overfill the bottom of the cabinet liner Allow the pump to operate for 30 seconds to 5 minutes after the cabinet line
10. d filters or prefilters always wear appropriate personnel protective equipment Lift the granular activated carbon and HEPA filters straight up and out of the cabinet top Install the replacement filters with the HEPA filter below the granular activated carbon filter into the cabinet The gaskets on both filters should be on the lower bottom surface Reinstall both bungee cords to secure the filters in place The unit is now ready to resume operation Figure 6 2 Unhook the bungee cord from the front of the unit and then lift the filters straight up to remove Product Service 1 800 522 7658 25 Chapter 6 Maintaining The Cabinet Changing the UV Lamp Models 34000 00 thru 04 Note For optimum performance the UV lamp should be changed on an annual basis The UV lamp and the work area of the cabinet must be thoroughly decontaminated before removing the lamp 1 Thoroughly surface decontaminate the UV lamp and the work area of the cabinet 2 Unplug the cabinet and open the door 3 Remove the UV lamp by rotating it 90 degrees and lifting it straight up and out of its sockets 4 Install new lamp by reversing the removal procedure Resetting a Circuit Breaker To reset any of the circuit breakers located on the top rear of cabinet depress the white button until it sets Figure 6 3 26 Product Service 1 800 522 7658 Chapter 6 Maintaining The Cabinet Setting the Speed Control
11. dispenser 2 lbs 1kg Germicidal Cleaner 1 Gal 3785 ml 3901701 Bottle of liquid disinfectant cleaner for refilling the spray bottle 10 lbs Skg 34 Product Service 1 800 522 7658 Protecting your ae laboratory environment LABCONCO CORPORATION 816 333 8811 800 621 5525 8811 Prospect Avenu AX 816 363 Ps LABCONCO Kansas City MO 64132 2696 www labconco A CE Declaration of Conformity We _Labconco Corporation 8811 Prospect Avenue Kansas City MO 64132 USA Ph 816 333 8811 Fax 816 363 0130 labconco labconco com being the representative party in the EU solely responsible for this certification hereby declare that the following product Equipment Laboratory Equipment Forensic Enclosures Type Model 3400001 02 03 04 Protector Evidence Drying Cabinet 230V 3410020 21 22 23 24 25 26 27 Protector Downdraft Powder Station 230V 3420020 22 24 26 CApture Portable Fuming System 230V Manufacturer Labconco Corporation 8811 Prospect Avenue Kansas City MO 64132 USA Ph 816 333 8811 Fax 816 363 0130 labconco labconco com is in conformity with the standards listed below EN61010 1 EN61326 1 EN55022 EN61000 3 2 3 following the provisions of the following directives 2006 95 EC 89 336 EEC 2004 108 EC when installed and operated in accordance with the manufacturers installation and operating instructions Af ff gt SY Z A KANSAS bh 4 2070 Mark Schm
12. e a stand pipe to prevent backflow into the cabinet 1 Remove the bucket from the lower storage area and save for possible reuse Connect the drain hose assembly to the tail pipe of the drain on the bottom of the drain assembly Push the drain assembly up onto the stand pipe until it is fully seated Hand tighten the compression ring of the drain assembly to the tail pipe until it is secure Route the drain hose through the appropriate hole in the bottom rear panel of the cabinet Route the drain hose to the floor drain and secure the end of the hose to prevent it from moving Pour water into the cabinet drain and inspect the entire drain system for leaks Tighten any leaking fittings Product Service 1 800 522 7658 Chapter 3 Getting Started Drain Options Model 3400010 only NOTE This model is designed to have its drain hose routed to a nearby floor drain To install the drain hose assembly NOTE The drain line should be routed to a floor drain with an air gap only NEVER directly connect the drain line to a sewer or drain system or use a stand pipe to prevent backflow into the cabinet 1 Remove the packing materials securing the hose to the back of the cabinet 2 Route the drain hose to the floor drain and secure the end of the hose to prevent it from moving Water Supply Connection 3410010 only NOTE This model has a cabinet water shut off valve located behind the lower front door The valve
13. e humidity 80 for temperatures up to 88 F 31 C decreasing linearly to 50 relative humidity at 104 F 40 C Main supply voltage fluctuations not to exceed 10 of the nominal voltage Transient overvoltages according to Installation Categories II Overvoltage Categories per IEC 1010 Temporary voltage spikes on the AC input line that may be as high as 1500V for 115V models and 2500V for 230V models are allowed Used in an environment of Pollution degrees 2 i e where normally only non conductive atmospheres are present Occasionally however a temporary conductivity caused by condensation must be expected in accordance with IEC 664 Product Service 1 800 522 7658 Appendix D Accessories Door Prefilters 3872600 Package of 6 polyester prefilters with a MERV 11 rating Precut to fit the door of the cabinet 1 lb 0 5kg Cabinet Prefilters 3872700 Package of 6 polyester prefilters with a MERV 11 rating Precut to fit the inside top of liner 1 1b 0 5kg Starter Kit 3869900 Includes many of the accessories needed to use the cabinet Includes T 6 door prefilters T 6 cabinet prefilters T 10 hanging clips T 10 paper liners die cut paper to fit the bottom of the cabinet to collect material that falls from evidence Paper is waxed on one side T 1 32 oz 946 ml bottle of germicidal cleaner with spray head 15 Ibs 6 8kg Paper Liners 3869800 A tube of 25 paper liners die cut paper t
14. e lamp ballast Replace defective or worn out lamp Replace lamp ballast 28 Product Service 1 800 522 7658 Chapter 7 Troubleshooting Filter System Light Blockage of the Check the inlet prefilters replace if turns on intake prefilters needed Ensure that there is no obstruction of the exhaust on top of the cabinet HEPA filter loading Replace the HEPA filter Product Service 1 800 522 7658 29 Appendix A Components Evidence Drying Cabinet Replacement Parts Item Quantity Part No Description 1 1 3872600 Door Prefilter package of 6 2 1 3872700 Cabinet top Prefilter package of 6 3 1 6938100 HEPA Filter 4 1 6938200 Granular Activated Carbon Filter 5 1 1270100 Lamp UV 30 Product Service 1 800 522 7658 Appendix B Dimensions 7 B 1 28 0 853 mm l 79 0 2408 mm Bye 36 0 1097 mm Interior Dimensions 56 0 1422 mm H x 32 3 820 mm W x 26 5 673 mm D Lower Storage Area Dimensions 12 38 314 mm H x 25 5 648 mm x 26 5 673 mm D Product Service 1 800 522 7658 31 Appendix C Specifications Electrical Data Model Requirements Energy Use 3400000 115 VAC 60 Hz 6 Amps 100W 3400001 2 3 4 230 VAC 50 60 Hz 3 Amps 100W Environmental Conditions 32 Indoor use only Maximum altitude 6562 feet 2000 meters Ambient temperature range 41 to 104 F 5 to 40 C Maximum relativ
15. f the cabinet for service There should be 36 inches 914 mm clearance at the front of the cabinet to allow the door to swing open without hitting any obstructions 2 Product Service 1 800 522 7658 Chapter 2 Prerequisites Location Requirements Note The cabinet is equipped with locking casters to ease installation and maintenance It should be located on a flat floor for proper cabinet alignment and door operation Exhaust Requirements If you intend to connect the cabinet to the optional Exhaust Connection Kit first examine the location to ensure that it accommodates the cabinet s exhaust duct The area directly above the cabinet s exhaust port should be clear of structural elements water and utility lines or other fixed obstructions There should be enough clearance to accommodate a 5 inch diameter duct For further information about the cabinet s exhaust system requirements please refer to Appendix D Accessories Electrical Requirements The evidence drying cabinet models have the following electrical requirements Table 2 1 Model Requirements 34000 00 10 115 VAC 60 Hz 6 Amps 34000 01 20 03 04 230 VAC 50 60 Hz 3 Amps Note An outlet with a circuit breaker rated at 10 amps for 115 volt models 5 amps for 230 volt models should be located as close as possible to the right rear side of the cabinet Power cords supplied with the Evidence Drying cabinet are 3 meters 10 feet long Dr
16. ical role in evidence drying cabinet performance Air is drawn into the inlet prefilter located on the bottom of the door The filtered air then flows upward through the liner passing over the evidence to be dried Before leaving the liner the air then passes through the exhaust prefilter that is designed to protect the HEPA filter from large particulates and may be removed to examine it for the collection of evidence The air is pressurized by the blower and then forced through the HEPA and carbon filters to remove biohazardous particulates and odorous chemicals respectively This is illustrated in Figure 4 1 Figure 4 1 Po Poom Air HEPA filtered Air i Contaminated Air Product Service 1 800 522 7658 13 Chapter 4 Performance Features and Safety Precautions Prefilters Both the door prefilter located on the front of the door and cabinet prefilter located on the inside top of the liner are disposable polyester loft particulate filters Both filters have a Minimum Efficiency Reporting Value MERV of 11 meaning they remove 65 79 of particles 1 0 3 0 microns in size and more than 85 of particles 3 0 10 microns in size See Appendix A for reorder part numbers HEPA Filters HEPA filters are disposable dry type particulate filters The filter material or media is typically made of borosilicate microfibers formed into a thin sheet in a process similar to the production of paper This
17. in injury For all models Ensure that the cabinet is connected to electrical service in accordance with local and national electrical codes Failure to do so may create a fire or electrical hazard Do not remove or service any electrical components without first disconnecting the cabinet from electrical service NEVER use this cabinet as a chemical storage cabinet When handling evidence or working in the cabinet always wear appropriate personnel protective equipment NEVER attempt to dry evidence containing flammable gases or solvents in the cabinet An open flame should NOT be used in the cabinet HEPA filters only remove particulate matter Operations generating volatile toxic chemicals are prohibited The media of the filter is fragile and should not be touched Avoid puncturing the HEPA filter during installation or normal operation If you suspect that a HEPA filter has been damaged DO NOT use the cabinet contact Labconco at 800 821 5525 or 816 333 8811 for further information 16 Product Service 1 800 522 7658 Chapter 4 Performance Features and Safety Precautions The HEPA filters in the cabinet will gradually accumulate airborne particulate matter from the items placed in the cabinet The rate of accumulation will depend upon the cleanliness of the air operating time and the nature of items being dried in the cabinet Do not operate the cabinet without all of the appropriate filters in place If an odor is detected com
18. ing a 3 Phillips screwdriver tighten both bracket screws until the bracket is secured to the wall Thread the strap through both bracket slots and then through both slots located at the top corners on the rear of the cabinet Push the cabinet back into its final location and then tighten the strap until there is minimal slack in the strap Product Service 1 800 522 7658 Chapter 3 Getting Started Figure 3 1 Secure bracket to wall stud 79 0 2400 cm from the floor to the top of the strap slot Product Service 1 800 522 7658 7 Chapter 3 Getting Started Drain Options Model s 34000 00 thru 04 NOTE These models can either drain to a bucket already located in the lower storage area or the bucket can be removed and an optional Drain Kit 3928200 can be installed The optional drain kit allows the cabinet to drain to a nearby floor drain On units equipped with the drain bucket NOTE The drain bucket had a working capacity of approximately 1 5 gallons Periodically monitor the liquid level in the bucket during decontamination or washdown operations 1 Access the bucket by opening the bottom storage area door on the front of the cabinet Before using ensure the bucket is in its correct position under the drain To install the Drain Kit 3928200 NOTE The drain line should be routed to a floor drain with an air gap only NEVER directly connect the drain line to a sewer or drain system or us
19. ing out of the granular activated carbon it needs to be replaced The weight limit for the fold down shelves is 15 lbs 6 8 kg and the floor of the cabinet and lower storage area is 20 lbs 9 1 kg When surface disinfecting the cabinet Always wear appropriate personnel protective equipment NEVER use high pressure water to clean the liner this practice could generate biohazardous aerosols NEVER spray disinfectants into the blower intake at the top of the liner this may damage the blower Avoid splashing the disinfecting solution on skin or clothing Ensure adequate room ventilation Carefully follow the disinfectant s safety instructions Always dispose of disinfecting solutions in accordance with local and national laws DO NOT allow bleach or disinfectants with high concentrations of free chlorine to contact the stainless steel components of the cabinet for a long period of time Free chlorine will corrode stainless steel after extended contact When handling used filters or prefilters always wear appropriate personnel protective equipment Product Service 1 800 522 7658 17 Chapter 5 Using the Cabinet Models 34000 00 thru 04 Fan UV Light Switch The cabinet Fan UV Light switch is located on the upper front of the cabinet as shown in Figure 5 1 Turning the switch to the Fan position starts the blower drawing air into through the cabinet Turning the switch to the UV On position will allow the UV
20. int yourself with this manual and keep it handy for future reference If you are unfamiliar with how evidence drying cabinets operate please review Chapter 4 Performance Features and Safety Precautions before you begin using the cabinet Even if you are an experienced cabinet user please review Chapter 5 Using the Cabinet it describes the cabinet s features so that you can use it efficiently This manual and other technical information is available in PDF format at our website www labconco com Product Service 1 800 522 7658 1 Chapter 2 Prerequisites Before you install the evidence drying cabinet you need to plan the site for installation Examine the location where you intend to install the cabinet You must be certain that the area is level and of solid construction In addition a source of electrical power must be located near the installation site Carefully read this chapter to learn e Location requirements e Electrical power requirements e Exhaust requirements e Service utility requirements e Space requirements e Drain requirements e Water Supply Requirements Model 3410010 Only Refer to Appendix C Specifications for complete cabinet electrical and environmental conditions specifications and requirements Space Requirements The dimensions for the evidence drying cabinet are shown in Appendix B Dimensions Clearance A minimum clearance of at least 6 inches 150 mm is suggested on the top o
21. itz Place and date of issue Mice President Research and Engineering Labconco Corporation Labconco P N 3696073 Rev C Product Service 1 800 522 7658 35
22. light to turn on when the door is closed and the UV Security lock is in the ON position UV Security Lock The UV security lock located to the right of the Fan UV Light switch allows the user to lockout the UV preventing its operation Before evidence is loaded into the cabinet it is recommended that the lock be turned to the OFF position and the key removed This will prevent the accidental activation of the UV light while evidence is drying The key should be stored in a convenient secure location known to all of the users to prevent its loss Recommended locations include hanging from the bar at the top of the liner or in the lower storage area Filter System Light The Filter System Light will illuminate if the differential pressure across the filters becomes excessive If lit first inspect or replace the two prefilters If lit with clean prefilters replace the HEPA filter NOTE If the Filter System Light illuminates during use finish the drying of the evidence in the cabinet and then correct the problem before using the cabinet again 18 Product Service 1 800 522 7658 Chapter 5 Using The Cabinet Figure 5 1 W 3 co Evidence Drying Cabinet ay Models 3400010 Fan Drain Pump Switch The cabinet Fan Drain Pump switch is located on the upper front of the cabinet as shown in Figure 5 1a Turning the switch to the Fan position starts the blower drawing air into through the cabinet Turning
23. n fifteen 15 days of delivery Limitation of Liability The disposal and or emission of substances used in connection with this equipment may be governed by various federal state or local regulations All users of this equipment are required to become familiar with any regulations that apply in the user s area concerning the dumping of waste materials in or upon water land or air and to comply with such regulations Labconco Corporation is held harmless with respect to user s compliance with such regulations Contacting Labconco Corporation If you have questions that are not addressed in this manual or if you need technical assistance contact Labconco s Customer Service Department or Labconco s Product Service Department at 1 800 821 5525 or 1 816 333 8811 between the hours of 7 00 a m and 6 00 p m Central Standard Time Part 3898400 Rev B ECO G108 TABLE Or ContTENTS CHAPTER 1 INTRODUCTION CHAPTER 2 PREREQUISITES Space Requirements Clearance Location Requirements Exhaust Requirements Electrical Requirements Drain Requirements Water Supply Requirements Model 3400010 only CHAPTER 3 GETTING STARTED Unpacking the cabinet Moving the cabinet Attaching the cabinet to the Wall Drain Options Models 34000 00 thru 04 Drain Options Model 3400010 only Water Supply Connection Model 3400010 only Preparing the Evidence Drying Cabinet for Operation Initial Operation CHAPTER 4 PERFORMANCE FEATURES AND SAFE
24. n for quiet energy efficient operation Cabinet Air Intake and Exhaust In order to operate properly the air intake at the bottom of the door the exhaust prefilter area in the top of the liner and the top surface of the activated carbon filter must be unobstructed Blocking or restricting any of these areas or allowing the prefilters to become excessively loaded with dust can adversely affect the performance of the cabinet Ultraviolet UV Lamp Models 34000 00 thru 04 The optional UV lamp generates a primary wavelength of light of 254nm A secondary emission is in the visible blue wavelength resulting in the characteristic blue color while operating UV light with a wavelength of 254 nm is biocidal primarily by creating thymine dimers in DNA These dimers prevent the correct transcription of the DNA into RNA resulting in cellular death or viral inactivation In order to be effective the UV light must directly strike the nucleic acid and its effectiveness can be diminished or negated by dissolved proteins or metals or by other UV opaque substances protecting the target nucleic acid Because of its limitations UV light should be used as an adjunct to good surface disinfection practices In order to get optimum performance from the UV light it should be replaced after 6 000 hours of operation or less and the exterior surface of the lamp should be kept clean and free of dust Note UV irradiation is absorbed by the tempered safe
25. o fit the bottom of the cabinet to collect material that falls from evidence Paper is waxed on one side 10 lbs Skg Security Tags 3902400 Package of 100 each tamper evident labeled tags Tags irreversibly indicate if the cabinet door has been opened during the drying cycle 1 1b 0 5kg Hanging Clips 3869500 Package of 10 each hanging clips that allow small articles to be hung from the cabinet s hanger rod or shelf 1 lb 0 5kg Exhaust System Connection Kit 3869600 Allows the cabinet to be connected to an exhaust system via the included 14 356 mm long 5 125 mm diameter flexible duct The exhaust system must be capable of exhausting 150 CFM 4 2 m min at a vacuum of 0 5 of H20 125 Pa 10 Ibs 5kg Product Service 1 800 522 7658 33 Appendix D Accessories Floor Drain Connection Kit 3928200 Allows the cabinet liner drain connector to be routed to a nearby floor drain Includes hose barb adaptor and 10 ft 3 meters of 1 2 13 mm O D polyurethane tubing 5 Ibs 2 3 kg Germicidal Cleaner Individual Wipes 3872400 Package of 100 each individually packaged single use disinfectant wipes 2 Ibs 1kg Germicidal Cleaner Ready to Use Wipes 3872501 Package of 90 ready to use disinfectant wipes in a convenient dispensing tub 2 Ibs 1kg Germicidal Cleaner 32 oz 946 ml 3901801 Bottle of liquid disinfectant cleaner Includes a screw on spray head
26. r has drained Then turn off the pump Turn the cabinet water shut off valve off Remove and clean out the drain strainer Inspect and clean the prefilters at the bottom of the door and at the top of the liner Dispose of rubber gloves appropriately and have lab coat laundered properly Wash hands and arms thoroughly with germicidal soap Models 34000 00 thru 04 Shutdown Close the door If UV disinfection is desired insert the UV Security Lock Key and turn it to the ON position Turn the Fan UV Light switch to UV The UV light will light Turn the UV Light off after an appropriate disinfection time typically 1 8 hours Models 3400010 Shutdown Close the door and push the switch to the center Off position 22 Product Service 1 800 522 7658 Chapter 6 Maintaining the Cabinet The common service operations necessary to maintain the Evidence Drying cabinet for peak performance are listed below Routine Maintenance Schedule Weekly e Using 70 ethanol or a suitable disinfectant surface disinfect the inside of the cabinet e Using an appropriate glass cleaner or LabSolutions Glass amp Surface Wipes Labconco part 1570000 clean the door and the surface of the UV lamp e Operate the cabinet blower noting if the Filter System Light turns on e Ensure the cabinet water shut off valve is in the off position Monthly or more often as required e Using a damp cloth or LabSolutions Glas
27. rs carry a two year warranty from date of installation or three years from date of shipment from Labconco whichever is sooner Blood Drawing Chairs carry a ten year warranty Carts carry a lifetime warranty Glassware is not warranted from breakage when dropped or mishandled This limited warranty covers parts and labor but not transportation and insurance charges In the event of a warranty claim contact Labconco Corporation or the dealer who sold you the product If the cause is determined to be a manufacturing fault the dealer or Labconco Corporation will repair or replace all defective parts to restore the unit to operation Under no circumstances shall Labconco Corporation be liable for indirect consequential or special damages of any kind This statement may be altered by a specific published amendment No individual has authorization to alter the provisions of this warranty policy or its amendments Lamps and filters are not covered by this warranty Damage due to corrosion or accidental breakage is not covered Returned or Damaged Goods Do not return goods without the prior authorization from Labconco Unauthorized returns will not be accepted If your shipment was damaged in transit you must file a claim directly with the freight carrier Labconco Corporation and its dealers are not responsible for shipping damages The United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier withi
28. s amp Surface Wipes Labconco part 1570000 clean the exterior surfaces of the cabinet particularly the front and top of the cabinet to remove any accumulated dust e All weekly activities e Check the prefilters replace both if needed Annually e All monthly activities e Replace HEPA and granular activated carbon filters Product Service 1 800 522 7658 23 Chapter 6 Maintaining The Cabinet Service Operations Door Prefilter Replacement 1 Pull the prefilter frame straight away from the door as shown in Figure 6 1 Working from the inside surface of the prefilter frame pull out the old prefilter and install its replacement such that the orange side of the prefilter faces you away from the front of the prefilter frame Press the prefilter frame back into the door engaging the fastener strips Figure 6 1 C 1 24 abinet Prefilter Replacement Open the door and locate the cabinet prefilter frame on the upper right side of the liner Grasping the front edge of the old prefilter pull the prefilter out if its frame Install its replacement such that the orange side of the prefilter faces up Product Service 1 800 522 7658 Chapter 6 Maintaining The Cabinet Granular Activated Carbon and HEPA Filter Replacement 1 Use an appropriate stepladder release the two bungee cords securing the granular activated and carbon and HEPA filters as shown in Figure 6 2 When handling use
29. sheet is folded or pleated to increase its surface area The pleats are held in place by aluminum diffusers or by beads of glue that add rigidity to the media pack The pack is then set into a frame and sealed as shown in Figure 4 2 The HEPA filter manufacturer establishes the efficiency of the filter by challenging it with an aerosol of known particle size The number of particles that penetrate the filter are quantified and this establishes the efficiency of the filter The filters used in the evidence drying cabinets are at least 99 99 efficient in removing particles 0 3 micron Note The HEPA filter media is very fragile DO NOT touch the media If you think the media of a HEPA filter is damaged DO NOT USE THE CABINET Call Labconco s Product Service Department at 1 800 522 7658 or 816 333 8811 Note HEPA Filters are only effective against particulate material Gases will pass through the filter Figure 4 2 Filter frame Polyurethane seal DE y e Between media pack and filter frame A ke Filter media pack A continuous sheet of filter media Li oi 14 Product Service 1 800 522 7658 Chapter 4 Performance Features and Safety Precautions Blower The blower assembly pulls room air into the intake prefilter on the front of the door through the liner and exhaust prefilter and then forces it through the HEPA and granular activated carbon filter The evidence drying cabinet uses a motorized backward curved fa
30. should be in the off position whenever the unit is not being washed down 1 Connect the water supply hose the stainless steel braided hose to a domestic water line equipped with a shut off valve and ending in a 3 4 inch male garden hose connector 2 Ensure the water supply valve is in the off position 3 Slowly open the water line shut off valve 4 Check the cabinet s water supply hose for leaks 5 Slowly open the cabinet shut off valve and check again for leaks 6 Set the spray gun nozzle to shower and slowly pull the trigger 7 When water flows steadily from the gun the cabinet shut off valve can then be turned to the off position Product Service 1 800 522 7658 9 Chapter 3 Getting Started Preparing the Evidence Drying Cabinet for Operation After the evidence drying cabinet has been removed from its pallet and positioned in its final location you must 1 Remove the tie and or tape securing the door the lower storage area door and the filters on top of the unit Note The following are located in a box in the lower storage space User s Manual UV Security Lock Key s model s 34000 00 04 only Power Cord Product Registration Card If you did not receive one or more of the components listed for the cabinet or if any of the components are damaged contact Labconco Corporation immediately for further instructions 2 Using a step ladder remove the tape and or packaging
31. st lift the product manually use at least six 6 persons and follow safe lifting guidelines Unpacking the Cabinet Carefully remove the outer carton and inspect the cabinet for damage that may have occurred in transit If the cabinet is damaged notify the delivery carrier immediately and retain the entire shipment intact for inspection by the carrier Note United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier within fifteen 15 days of delivery Do not return goods without the prior authorization of Labconco Unauthorized returns will not be accepted If the cabinet was damaged in transit you must file a claim directly with the freight carrier Labconco Corporation and its dealers are not responsible for shipping damages 4 Product Service 1 800 522 7658 Chapter 3 Getting Started Do not discard the carton or packing material for the cabinet until all of the components have been checked installed and tested The cabinet is secured to the pallet using four transport brackets two on each side To remove the brackets do the following 1 Using the 12 inch wrenches loosen and remove the bolts and locknuts that secure the brackets to the pallet 2 Using a 12 inch wrench loosen and remove the nuts and bolts securing the transport brackets to the bottom of the cabinet 3 Discard the brackets shipping nuts and bolts The cabinet is now ready to be removed from the palle
32. t NOTE A ramp has been provided attached to the wood pallet Follow the instructions on the face of the cabinet for safe removal Moving the Cabinet When moving the cabinet on its casters DO NOT push on the front or back of the unit Move the cabinet by pushing or pulling it sideways This will minimize the possibility of the unit overturning when in motion Product Service 1 800 522 7658 5 Chapter 3 Getting Started Attaching the Cabinet to the Wall WARNING The evidence drying cabinet must be secured to a wall stud or other suitable wall structure using the included strap and bracket This will minimize the possibility of the unit overturning when in use 1 Remove the wall bracket secured to the strap located on the back of the unit on top Locate a stud inside the wall directly behind the cabinet s installation location NOTE The screws provided with the wall bracket are intended to be used when attaching the bracket to a wooden wall stud If your wall is made of different materials cinder block concrete etc you will need to attach the wall bracket to the wall using an anchor or fastener approved for that type of construction 3 Position the bracket centered on the stud with the top of the strap slot 79 inches above the floor Mark the top of each screw hole as shown in Figure 3 1 Using a 1 8 3 mm drill drill pilot holes at both marked locations to a depth of at least 1 inches 38 mm Us
33. ty glass of the door Independent research has shown that the level of UV irradiation on the outside of the cabinet s door is equal to background radiation levels Note The UV sensitivity of a target organism varies depending on the UV output of the lamp the genus and species of the organism the medium the agent is suspended in etc Contact the Health and Safety Officer at your facility for UV light use and recommendations Product Service 1 800 522 7658 15 Chapter 4 Performance Features and Safety Precautions Safety Precautions For models 3400010 only DO NOT directly connect the drain line to a sewer drain system or place in a vertical standpipe DO NOT use the wash down gun to clean up biohazardous material Disinfect any surface or material BEFORE using the gun ALWAYS wear appropriate personal protective equipment including eye and face protection when washing down the cabinet Ensure that the drain line is not kinked or obstructed and flows to a floor drain with an air gap The Drain Pump should only be operated to remove standing liquid in the bottom of the cabinet Do not operate the pump dry for more than 5 minutes For models 34000 00 thru 04 only NEVER expose ANY EVIDENCE plastic or coated materials to ultraviolet UV radiation Never bypass the UV safety interlock that only allows the UV light to work when the door is closed UV exposure to the skin is hazardous and can result in serious eye or sk