Home

OWNERS MANUAL HF – Micro 1000 Laboratory Turbidimeter 0

image

Contents

1. Accessory or Replacement Part Micro 1000 PRMEIME Calibration Set includes 0 02 NTU 10 0NTU 100 NTU amp 1750 NTU Calibration Standards High Range PRIMEIMECalibration Standard for White Light 10000 NTU Micro 1000 IRPRIMEIME Calibration Set includes 0 02 NTU 10 0NTU 100 NTU amp 1750 NTU Calibration Standards High Range PRIMENME Calibration Standard for IR light 10000 NTU Formazin Stock Solution Kit Formazin 4000 NTU Stock Solution 500 ml Lamp Module Tungsten Filament Lamp Module IR Turbidity Free Water Pour Through Assembly Flow Through Assembly Cuvette Stand holds 11 cuvettes Sample Cuvettes 3 pack Sample Cuvettes 10 pack Micro 1000 Manual 22588 RS232 Cable for Serial Printer 19798 Rev 1 9 25 11 Glossary Formazin Turbidity Units FTU see Nephelometric Turbidity Units Indexing a Cuvette The United States Environmental Protection Agency US EPA recommends that cuvettes used for turbidimeter calibration or sample measurement be indexed To index a cuvette with a sample in it slowly rotate the cuvette throughout one complete revolution 360 While rotating the sample cuvette observe the display and locate the position that the cuvette is in which provides the lowest turbidity reading This position is the indexed position of the cuvette Nephelometric Turbidity Units NTU Unit of measure used when comparing the light scattered by a liquid media to the light scattered by a known concentration
2. If you wish to exit into the return to the normal operating mode press the EJ key at this time If you wish to reperform the calibration use either the q or D keys to maneuver back to the standard that you wish to work with Rev 1 9 12 Notes 1 You must use the 0 02 NTU standard at the beginning and end of the calibration cycle to ensure accurate calibration of the instrument 2 During calibration the instrument performs some system self diagnostics Several warning or error messages may be displayed If there is an error in calibration the associated warning or error message s will be displayed in the lower row of the LCD screen Refer to the section on troubleshooting to determine the cause of the warning or error 3 Atany point in time during calibration you can cycle through the required calibration points 0 02 NTU 10 NTU 100 0 NTU 1750 0 NTU and 10000 0 N TU by pressing either the q or D key to individually calibrate with a particular calibration standard If you wish to exit the calibration mode you may do so at any time by simply pressing the or D key until the word END is shown on the lower row of the display and then pressing the EJ key However exiting the calibration process without completing all of the steps for calibration may cause the accuracy of the instrument to be diminished Instrument Configuration SETUP Mode The instrument allows you to configure it according to your needs at any time during
3. months if they are to be used for EPA reporting Rev 1 9 9 1 Slowly rotate the calibration standard one complete revolution 360 While rotating the standard observe the measured turbidity and locate the position of the cuvette having the lowest turbidity reading 2 With the calibration standard positioned at the location having the lowest turbidity reading install the Indexing Ring over the black light shield on the standard so that the pointer of the Ring faces forward toward the operator When using the standards in the future always insert the standard so that the pointer of the index ring faces forward Slowly rotate the standard back and forth about 5 to find the lowest point The standard is now indexed and ready to use 5 3 Calibrating the Instrument Even though it is possible to calibrate the instrument using any sequence of the prescribed calibration standards to achieve the stated accuracy you must use the procedure below to calibrate the instrument 1 Select the calibration function of the instrument by pressing either the up or down arrow key until the Cal block is lit Press the EJ key The cal block will now be lit and the turbidity displayed in the lower row of the display should read 0 02 NTU This is the first standard that must be used in calibration 2 Insert the 0 02 NTU calibration standard into the sample well and index the standard if you have not already done so see section 7 1 At this
4. point press the EJ key to initiate calibration on the 0 02 NTU standard The instrument will calibrate on the 0 02 NTU level the Cal block and the Store block will flash on and off for a period of time and you will see a countdown from 90 Note The 0 02 NTU standard must be evaluated in calibration at both the start and end of calibration to ensure accurate measurements in the ratio mode and when recording FAU measurements Rev 1 9 10 3 When the instrument has completed the calibration the Cal and the Store blocks will quit flashing and the lower row of the display should now indicate that the 1750 NTU standard is the next standard to be used for calibration 1 75 K Press the J key to indicate to the instrument that you plan to calibrate on this point 4 Insert the 1750 NTU calibration standard into the sample well and calibrate as before When the instrument has completed the calibration the Cal and the Store blocks will quit flashing and the lower row of the display should now indicate that the 10000 NTU standard is the next standard to be used for calibration 10 0 K 5 If you plan to measure turbidity levels above 1750 NTU you must calibrate using a 10000 NTU calibration standard not included in the calibration Kit If you do not plan to measure turbidity levels above 10000 NTU then use the arrow buttons until the lower row of the display shows 100 0 and then proceed to step
5. user see section 8 3 The access code is required as an input whenever the user wishes to gain access to either the calibration routine or the configuration setup routine when the feature is enabled Use either the C or V7 keys to select the mode that you wish to enter and then press the J key If the security code option is turned on you will see the following display This security code has four numbers that are selectable one at a time The number that is flashing is the first number for you to select Use either the C or V7 keys to select this number in the code then press the EJ key to move to the next number Follow this process through the four numbers in the access code If you have selected a valid access code the instrument will be directed to either the calibration or the configuration user setup mode depending on your original intention If the wrong access code is selected the instrument will return to the normal automatic mode Calibration Procedures The instrument was calibrated and tested prior to leaving the factory using calibration standards that are traceable to the primary turbidity calibration standard Formazin Therefore it is possible to use the instrument directly out of the box However re calibration of the instrument is recommended to help you become familiar with the operation of the instrument and the calibration procedures Rev 1 9 8 Under normal conditions re calibration is recommended at least on
6. wrong standards or the lamp failed during calibration Rev 1 9 22 7 2 System Error Message s The instrument generates error messages when problems are detected during normal operation When these messages are observed and if you do not understand the instructions shown below contact the HF scientific inc Technical Services department to determine a resolution to the problem The instrument indicates an error message when an E XX is displayed on the lower row of the display The instrument has several error codes each assessing a different component or system of components in the instrument The following table lists the error messages and their associated meanings ERROR ASSOCIATED MEANING TYPICAL CAUSE The A D has failed to report One or several components in the A D circuit correctly has failed call technical service immediately SFE General Electronics Failure One or many electrical components have failed in the instrument If an error is noted please contact the HF scientific inc Technical Services Department immediately to rectify the error 7 3 Factory Default Parameters The instrument retains all original factory settings in non volatile memory At any time you can cause the instrument to change back to these default settings This is particularly useful if you feel that your calibration standards may have been compromised or if someone has erred in their use of the instrument In these situatio
7. 6 Otherwise insert the 10000 NTU calibration standard into the sample well and calibrate as before When the instrument has completed the calibration these blocks will quit flashing and the lower row of the display should now indicate that the 100 0 NTU standard is the next standard to be used for calibration 0008 HUJ Rev 1 9 11 6 Insert the 100 0 NTU calibration standard into the sample well and calibrate as before When the instrument has completed the calibration the Cal and the Store blocks will quit flashing and the lower row of the display should now indicate that the 10 0 NTU standard is the next standard to be used for calibration 939 I HUJ 7 Insert the 10 0 NTU calibration standard into the sample well and calibrate as before When the instrument has completed the calibration the Cal and the Store blocks will quit flashing and the lower row of the display should now indicate that the 0 02 NTU standard is the next standard to be used for calibration 8 Insert the 0 02 NTU calibration standard into the sample well and calibrate as before When the instrument has completed the calibration the Cal and the Store blocks will quit flashing and the lower row of the display should now display the word END indicating that the instrument has completed the calibration phase and you can can now exit into the normal mode for measurement CAL ENI 9
8. A SN AE oooO EE EESE E 13 6 3 Selecting Display Resol tilses c oeo eons Eoo S ETTET OEE 14 6 4 Configuring the Security Code israel EN E eoo eE ESS TTO TETE EAST SESEO S 14 6 5 Setting up the RS232 Port Printing Function 6 6 Setting the Calibration Interval c ssssiscccsssecccesessensesvessustsvonstenesonssessvessnenstononstcoessnswtvsnssdedvosontesnssoustovasshsuessisnsedhebendacsooonsuccensasstes 6 7 Setting the Date Year Month and Day s ssssessossoseseeseessooscoscoscossosesseoseosoosooscossossossossossosooscossossossoseosscseosoossossossossossssseseesess 17 6 8 Setting the Time Hours and Minutes eere eee eere eese tenen ene tn ne snentn sn sn etn nere zena tasas ento seenen tasa een ee seeneneeseeneneeneo 18 6 9 Exiting th Setup MOTE siriene sussceosencesnsnducnsscnsncebesntete r esoo ei SE E n EO ria 19 7 TroublesHoOR Bg saisan e aa a a ENS RE E a E a e Ae 20 TA System Warning MeSSATE S ssissrenissuioieessssdecsessosesivdvonsesesnsetesosisniedodesstesescontadososeiecssosbiceesbeseetesognbasesasaceebsesoiebesesstesasionesdosesedeeutes 20 7 2 System Error Message S ssscccssssnsssoscseiscrssssacesdossensedsgnsatosssscuetbaswiedssessasdesoonsadcessosastess PARERI EP Eroro o oS SIRIA 23 7 3 Factory Default Parameters riesi aria turpe repentini sinti enai esq i bed Soio rs inoar rO eoo EDES SEE niger 23 8 Routine Maintenance pira lidi Lala ia ain ae 23 8 1 Cuyette Cleaning and Caressa eose tirare
9. OWNERS MANUAL HF Micro 1000 Laboratory Turbidimeter 0 10 000 NTU HF scientific inc 3170 Metro Parkway Ft Myers FL 33916 Phone 239 337 2116 Fax 239 332 7643 Email info Q hfscientifc com Website www hfscientific com Catalog No 22615 2 04 Rev 1 9 DECLARATION OF CONFORMITY Application of Council Directive 73 23 EEC Standard to which Conformity is Declared Product Safety Immunity amp EMI UL3101 1 EN61326 1998 CSA C22 2 No 1010 1 92 EN61000 3 2 2000 EN61010 1 1993 A2 1995 EN61000 3 3 1995 Manufacture s Name HF scientific inc Manufacture s Address 3170 Metro Parkway Fort Myers Florida 33916 7597 Importer s Name Importer s Address Type of Equipment Laboratory Turbidimeter Model No Micro 1000 Catalog Numbers 20013 20014 Micro 1000 IR 20015 20016 I the undersigned hereby declare that the equipment specified above conforms to the above Directive and Standard Place Fort Myers Florida USA afi 2 Signature Robert J Maley President Rev 1 9 i l SPECIFICATION S adii e pete to lied 1 2 A PC seteadec 2 2 1 Unpacking and Inspection of the Instrument and Accessories eese ee re reenin ee ene een enses tn ne snenio ne snenio re sues en en enne 2 PAPA User ail du Clu Re DE ieS 3 Mri M 4 4
10. RMEIME Calibration Standard available from HF scientific inc 10000 NTU PRiMEIME Calibration Standard available from HF scientific inc tA dO It is well known that diluted Formazin is unstable If you choose to use Formazin to calibrate the instrument ensure that you are using a fresh stock suspension of Formazin to achieve the accuracy quoted for the instrument A Formazin Stock Solution Kit is available from HF scientific inc Catalog No 50040 The HF scientific inc PRIMEIIME primary calibration sets Catalog No 19927 or 19928 are more stable than Formazin and have a minimum shelf life of 12 months If you use the stable HF calibration standards to calibrate the instrument review the expiration date to ensure that the standards have not expired The calibration standards must be inverted well mixed prior to use to ensure the homogeneity and repeatability of the suspension for measurement 5 2 Indexing the Calibration Standard s The United States Environmental Protection Agency US EPA recommends that cuvettes used for instrument calibration or sample measurement be indexed To comply with this recommendation each HF scientific inc calibration standard is supplied with an indexing ring for quick and repeatable indexing of the calibration standards To index a calibration standard perform the following two steps The EPA recommends that on line turbidimeters be calibrated with the primary standard at least once every three
11. Routine Operaio ous iet eeu eni a n tice else dudum Se HG oca eu tUi ona 5 4 1 Measurement Principle amp Methodology eere eene e tenen eese enne en ee sne nen ee sae nen re seenen ee ene en en sn ees en sn ees en ene n essa sn ne 5 4 2 Grab Sample Measurement Routine Measurement eere eere eese tenen ene tn ee ene nen en seenio ee tne nen ea sn en en re sn sensn ene n en en enne 6 4 3 Pour Through Sample Measurement Optional Accessory e eeeee eese eese ee ene tn ee enenen en seenen tuens nen re sues en ea sn enean suene 7 4 4 Flow Through Sample Measurement Optional Accessory eese scese see eene e tete reeee store see etere ae ste neene tensa te stenennee 7 4 5 Dry Air P rge Syste eserine 8 4 6 Using the Security Access Feature Good Manufacturing Practice GMP ssssesesessecscssoscocososeosososeoesseseessosescososeososee 8 5 Calibration Procedendo nut en depo fane 8 5 1 Calibration Standards 5 2 Indexing the Calibration Standard s 5 3 Calibrating the Instrumentene E G M 10 6 Instrument Configuration SETUP Mode iere heidi tea dedos ia d deed d cide diodes 15 GAs Setting the SCA er Oooo soe iSo ta iie 13 6 2 Setting the Ratios MOE RES dene A
12. ake sure that the light bulb icon on the back of the lamp module is upright Press the module into the instrument until you hear it click firmly into place 8 3 Battery Replacement The Micro 1000 contains a nomrechargeable 10 year life lithium battery for backup of the real time clock This battery is not field replaceable Failure of this battery will not impede the use of the instrument however some time related activities such as timed printouts will not operate Contact the HF scientific inc Service Dept for replacement of this item Note The side tabs may need to be pressed outward until they click to secure the new lamp module 5 If the two side tabs on the lamp module do not click the lamp module securely in place check that the power wire is not obstructing the lamp module 6 Turn on the instrument and follow the instructions in section 4 1 to re calibrate the instrument with the new lamp module The instrument must be re calibrated after lamp module replacement 7 Resume normal operation after recalibrating the instrument Rev 1 9 24 9 Contacting the HF scientific inc Technical Service Department For technical assistance or to order replacement parts please contact the HF Technical Services Department or Customer Service Department HF scientific inc 3170 Metro Parkway Fort Myers Florida 33916 7597 Phone 239 337 2116 Fax 239 332 7643 Email info hfscientific com 10 Accessories and Replacement Parts List
13. c c NE Sample Flow Rate 0 5 1 5 gpm 2 6 L min 0 5 1 5 gpm 2 6 L min 60 psi max pressure 414 kPa 60 psi max pressure 414 kPa Shipping Dimensions 12 x 14 x 12 12 x 14 x 12 35 6cm x 30 5cm x 30 5cm 35 6cm x 30 5cm x 30 5cm Shipping Weight 7 lb 3 18 kg 7 lb 3 18 kg The specifications found in EPA method 180 1 are essentially the same as the specifications set out in method 2130B of the Standard Methods for the Examination of Water and Wastewater 19th edition and the specifications set out in ASTM Standard Method D1889 94 This turbidimeter meets or exceeds the specifications set forth in these methods Instrumental accuracy measured under controlled laboratory conditions at 25 C 77 F Rev 1 9 1 2 Overview The Micro 1000 process turbidimeter allows you to measure the turbidity of your process water The White Light Micro 1000 has been designed to meet the design criteria specified by the US EPA on turbidity measurement The Infra Red Micro 1000 Micro 1000 IR was designed to meet the design criteria specified in ISO 7027 and DIN 27027 for the measurement of the turbidity of a sample 2 1 Unpacking and Inspection of the Instrument and Accessories The table below indicates the items that you should find in your turbidimeter shipment Remove the Calibration Kit blue case found in the foam pack The Package of 3 Cuvettes blue case and the instrument from the packing carton Carefully
14. ce every three months You may select a predetermined calibration interval for automatic prompting for calibration if you exceed the selected calibration interval the Cal block will flash until the instrument is re calibrated During calibration the instrument performs several system self diagnostics As such several warning messages may be displayed If the instrument detects an irregularity with the instrument detectors or lamp a warning message will be displayed on the lower row of the display on exit from the calibration mode If this occurs please attempt to rectify the warning message yourself or contact the HF scientific inc technical services department In any event the instrument will continue to monitor the turbidity of the process water with a decreased accuracy until the error is rectified In addition if the instrument remains in the calibration mode for longer than 5 minutes with no activity it will time out and return to the automatic measuring mode Note The instrument must be re calibrated after lamp replacement Note 5 1 Calibration Standards HF scientific inc recommends that you use the following materials during calibration to achieve the accuracy stated in this manual 0 02 NTU PRMEIME Calibration Standard available from HF scientific inc 10 0 NTU PRIMEIME Calibration Standard available from HF scientific inc LOO NTU PRMEIME Calibration Standard available from HF scientific inc 1750 NTU P
15. cord the turbidity of the sample displayed after the entire 500 ml sample has been poured into the assembly the flow to drain has ceased and after the reading has stabilized 4 Ifyou have turned off the printer output in the customer setup section pressing the EJ key at this time will output data to the RS232 port 5 Repeat steps 2 through 4 for all of your samples The sample cuvette used in the pour through assembly is identical to the standard sample cuvettes supplied with the instrument Clean the cuvette on a periodic basis according to your experience with the type and turbidity of the sample found in your facility 4 4 Flow Through Sample Measurement Optional Accessory Install the flow through assembly according to the instruction sheet that accompanies the assembly The following steps describe how to continuously measure the turbidity of a sample stream using the instrument fitted with the optional flow through assembly 1 Turn on the instrument The instrument will be in the normal automatic mode Allow instrument to warm up for 30 minutes 2 Initiate flow of the sample to the instrument and allow 2 minutes for the flow to stabilize prior to recording the measured turbidity 3 Begin to record the turbidity readings at the desired intervals 4 If you have selected timed printer output in the customer setup section you should be able to monitor the process in real time using PC data acquisition software Rev 1 9 7 The
16. e detector is used to generate a more accurate measure of turbidity should there be any fluctuation in lamp intensity via voltage spikes or lamp ageing Finally two detectors are used to measure transmission of light for the extended turbidity range for ratio measurement and for determination of FAU s see below for definition In the course of normal operation the embedded software measures the intensity of light at each detector During this operation several advanced fuzzy logic techniques are used to determine if air bubbles or large particles are passing through the path of light at the detector If an anomaly is detected the algorithm identifies it and repeats the measurement prior to reporting the turbidity levels By monitoring all detectors the instrument is able to perform system self diagnostics it is able to ensure that all data collected from the detectors is consistent Then the calculations for Rev 1 9 3 turbidity are made according to the methods described in the specifications section of the instrument section 1 During normal operation the AUTO block will be highlighted with the current scale displayed on the lower row of the display and the measured reading on the upper row of the display In certain instances during normal operation the instrument will display a row of dashes across the upper row of the display This indicates that either the instrument is performing an auto ranging function
17. ected the proper day press the EJ key 6 8 Setting the Time Hours and Minutes After pressing the EJ key the Time block will be displayed and you will see the time displayed on the upper row of the display in 24 hour format The number flashing corresponds to the hour Select the correct hour by pressing either the E or keys to change the displayed hour When you have selected the proper hour press the EJ key After pressing the J key the Time block will remain displayed and the second number on the upper row of the display will be flashing this number corresponds to minutes Select the correct minutes level by pressing either the C or W7 keys to change the displayed minutes When you have selected the proper minutes level press the EJ key Note The number of seconds in the minute will be reset to zero when the EJ key is pressed after selecting the proper minutes level Rev 1 9 18 6 9 Exiting the setup mode After you have pressed the EJ key for selecting the time the word END will be shown on the lower row of the display SETUP ENI You may also reach this point by pressing either the q or D key until the word END appears on the lower row of the display At this point simply press the EJ key to return to the normal mode Rev 1 9 19 7 Troublesho oting 7 1 System Warning Message s The instrument generates warning messages to provide you with specific diagnostic information ab
18. een highlighted to give further clarification to the instructions Refer to the Table of Contents to easily find specific topics and to learn about unfamiliar terms Rev 1 9 4 Routine Operation 4 1 Measurement Principle amp Methodology This instrument is designed to measure the turbidity of a sample in nephelometric turbidity units NTU In addition the instrument is capable of measuring in several other standardized scales Applicable Instrument s Scale Definition White Light Ratio NTU Turbidity measured via a combination of transmitted and 90 scattered light using the 90 scattered light as the primary detector signal This is used to eliminate effects of colored samples FNU Formazin Nephelometric Units 1 FNU NTU FAU Formazin Attenuation Units turbidity measured via transmitted light according to ISO 7027 and DIN 27027 EBC Turbidity scale of the European Brewery Commission Nephelos A turbidity scale used in microbiological research to report progress of bacterial growth e g immunological assays 6 7 Nephelos 2 1 NTU Nephelo is the Greek word for cloud The system measures these scales using five 5 separate detectors Two detectors are located 90 from the path of the light nephelometric detectors to ensure measurement redundancy and to minimize variations in cuvettes One detector is a reference detector that monitors the light output by the light source both white light and infrared The referenc
19. eseese se se ve siepe i upra debe OEO S ETE E aeS ENAS 24 8 2 Lamp Replacement oomini 24 ROM Battery Replacement der M iaaio 24 Rev 1 9 il 9 Contacting the HF scientific inc Technical Service Department seen 25 10 Accessories and Replacement Parts List eie ta ctae deed re De Pede Sa aed 25 11 Glossary d 26 12 Mapa beo oer CEPT eC ERODE oe eat RRO ET RID Re PES MEE A cng AA 27 Rev 1 9 ill 1 SPECIFICATIONS QMICROI100 MICROI0IR Conforms to specifications set forth in Conforms to specifications set forth in Instrument Conformance EPA method 180 1 Nephelometric ISO 7027 Water Quality Determination Method of Turbidity Light Source Tungsten Filament Lamp Infrared LED Measurement Ratio on or Ratio off Ratio on or Ratio off Options and Ranges NTU 0 10 000 FNU 0 10 000 EBC 0 2 450 NEPHELO FAU 0 Accuracy 0 1000 NTU 2 of reading 0 01 NTU 2 of reading 0 01 NTU 1000 4000 NTU 5 of Reading 5 of Reading 4000 10 000 NTU 10 of Reading 10 of Reading Repeatability 1 of reading or 0 01 NTU 1 of reading or 0 01 NTU PS mmeeerepeae ubihenrigreme O Specifications 2 Security Access Code 2 Security Access Code Operating Temperature 0 C 50 C 32 F 122 F 0 C 50 C 32 F 122 F si ii
20. he wrong standards Transmission detector 2 error Transmission detector 2 failed to report the proper signal levels during calibration the detector circuitry has failed or calibration was performed with the wrong standards Nephelometric detector 90 error The nephelometric detector failed to report the proper signal levels during calibration the detector circuitry has failed or calibration was performed with the wrong standards Reference detector The reference detector failed to report the proper signal levels during calibration the detector circuitry has failed or calibration was performed with the wrong standards Calibration failure on range 0 10 The signal difference obtained between the 0 02 NTU and 10 0 NTU calibration standards was too small to ensure the resolution in the specifications Verify that you used the correct standards during calibration Calibration failure on range 10 100 The signal difference obtained between the 10 0 NTU and 100 0 NTU calibration standards was too small to ensure the resolution in the specifications Verify that you used the correct standards during calibration Calibration failure on range 100 1750 The signal difference obtained between the 100 0 NTU and 1750 0 NTU calibration standards was too small to ensure the resolution in the specifications Verify that you used the correct standards during calibration Multiple warnings have occurred Calibration was performed with the
21. ility to display several levels of resolution number of digits to the right of the decimal place The instrument is capable of displaying turbidity with a resolution of four digits to the right of the decimal place 0 0001 for turbidity readings below 10 NTU If you feel that the last digit or two is not stable or useable then you may suppress these digits by selecting the proper display resolution After setting up the ratio mode functionality the upper row of the display will show the current resolution the lower row of the display will show the letters RES You have the ability to select the resolution shown on the display by pressing either the y or V7 keys to change the resolution to be used When you have selected the desired resolution press the key 6 4 Configuring the Security Code The instrument is equipped with a security access option If this option is turned on the user must input the access code to gain access to calibration or instrument configuration The security key will be highlighted on the display and the word S C will be shown on the lower row of the display Rev 1 9 14 The upper row of the display will indicate the operational status of the security access option on or off You can change the operational status of this option using either the C or V7 keys Once you have set the desired status press the E key to accept it If you selected to turn on the security access code you wi
22. inspect all items to ensure that no visible damage has occurred during shipment If the items you received do not match your order please contact your local distributor or the HF scientific inc Customer Service department immediately Note Extra care should be taken when unpacking opening and handling the calibration standards and sample cuvettes in the Accessory Kit surface scratches or finger smudges on the cuvette surface may cause measurement errors Handle these items by the top area of the cuvette only Rev 1 9 2 2 2 User Interface The user interface of the instrument is designed with a 6 key touch pad and a single user display The six keys of the touch pad and their functionality are described below Figures la and 1b are depictions of the front and rear views of the instrument Optical Well Display Lamp Module Touch Pad Figure 1a Front view of the Instrument Optical Well Lamp Module ISVDC Air Purge Out Figure 1b Rear view of the Instrument Rev 1 9 3 The ON OFF key key is used to turn the instrument ON or OFF The enter key 13 is used to accept information The up and down arrow keys G and A are used to set and reset numerical values and to change instrument modes The left and right arrow keys q and D are used to scroll through lists In addition the rear panel of the instrument is designed to provide ready access to the power plug connector the RS 232 serial port and the dr
23. ll need to input an access code to be used for entering calibration and for entering the configuration mode The security key will be highlighted on the display and the word S C will be shown on the lower row of the display while the current access code for entering calibration is shown on the upper row of the display C followed by 4 numbers ISETUP If you wish to change the code you may change each of the four numbers one at a time Use either the G or V7 keys to select the first number in the code then press the EJ key to move to the next number Follow this process through the four numbers in the access code When you are satisfied with your selection press the EJ key to accept the number The access code will change from the calibration access code to the configuration user setup access code S followed by 4 numbers Select the access code for the configuration setup mode as you did with the calibration mode Rev 1 9 15 6 5 Setting up the RS232 Port Printing Function After pressing the EJ key the Print block will be lit and the lower row of the display will indicate the state of the RS232 port UFF This feature allows you to select the function of the serial port on the instrument There are two options to choose from Off and Timed Output Select the RS232 option by pressing either the GS or V7 keys When you have selected the proper function press the IJ key Off When the RS232 port is tu
24. normal operation except when you are performing an instrument calibration This section describes how you can customize your instrument You can enter the SETUP mode of the instrument by pressing either the C or key until the SETUP block is highlighted At this point press the EJ key and you will either enter the SETUP mode or be prompted to enter the security code Note To skip the selection of any parameter simply press either the q or D key to continue on to the next or previous section 6 1 Setting the Scale After pressing the J key the Scale block will be lit and the current scale measured units will be displayed on the lower row of the display Press either the or V7 keys to change the units that will be displayed during normal automatic measurement mode When you have selected the desired units press the EJ key 6 2 Setting the Ratio Mode After pressing the EJ key the Ratio block will be lit and the lower row of the display will indicate the state of the Ratio Mode in the instrument Rev 1 9 13 This feature is designed to decrease the sensitivity of the instrument to dissolved materials that may adversely affect a turbidity reading e g dissolve material creating color Press either the Cy or V7 keys to change the operational state of this option in the instrument When you have selected the desired state press the EJ key 6 3 Selecting Display Resolution The instrument is equipped with the ab
25. ns you may use the factory default parameters while waiting for new standards or instructions to arrive from HF scientific inc To invoke this option turn the instrument on and then press the key while holding down the q key The instrument will perform several operations taking approximately 2 minutes and then turn off indicating that the memory and all settings have been reset to the factory default parameters After this point you may start to use the instrument again as you normally would 8 Routine Maintenance The instrument has been designed for ease of use and simple operation When not in use ensure that the instrument has been turned off and that a clean sample cuvette fitted with a black light shield cap has been placed in the sample well This will ensure that a minimal amount of dust and or debris will be able to settle on the optics of the instrument Rev 1 9 23 8 1 Cuvette Cleaning and Care Proper measurement of the turbidity of a sample requires the use of a cuvette that is free of marks smudges scratches and any bacterial growth Any typical glass cleaner can be used along with a lint free cloth or tissue to clean the outside of the cuvette The inside of the cuvette can also be cleaned with any typical glass cleaner Once cleaned the cuvette should be rinsed thoroughly 8 to 10 times with clean distilled water to eliminate the possibility of detergent build up and streaking Cleaned and dried cuvettes should be
26. o and from HF scientific inc s factory HF scientific inc is not responsible for damage to the instrument or parts thereof resulting from misuse negligence or accident or defects resulting from repairs alterations or installation made by any person or company not authorized by HF scientific inc HF scientific inc assumes no liability for consequential damage of any kind and the original purchaser by placement of any order for the instrument or parts thereof shall be deemed liable for any and all damages incurred by the use or misuse of the instruments or parts thereof by the purchaser its employees or others following receipt thereof Carefully inspect this product for shipping damage if damaged immediately notify the shipping company and arrange an on site inspection HF scientific inc cannot be responsible for damage in shipment and cannot assist with claims without an on site inspection of the damage This warranty is given expressly and in lieu of all other warranties expressed or implied Purchaser agrees that there is no warranty on merchantability and that there are no other warranties expressed or implied No agent is authorized to assume for HF scientific inc any liability except as set forth above HF scientific inc 3170 Metro Parkway Fort Myers Florida 33916 7597 Phone 239 337 2116 Fax 239 332 7643 Email info hfscientific com Rev 1 9 27
27. of Formazin Polymer This unit of measure is recognized as a measure of the optical clarity of an aqueous sample NTU is the accepted unit of measurement for turbidity Turbidity 1 A measure of the attenuation of a radiant flux as it passes through a liquid media 2 Optical clarity of a liquid 3 a phenomena caused by the presence of undissolved matter in a liquid media Formazin Nephelometric Units FNU see Nephelometric Turbidity Units European Brewery Convention Units EBC A scale related to turbidity where 0 245 EBC s 1 NTU Nephelos A scale related to turbidity where 6 7 Nephelos 1 NTU Rev 1 9 26 12 Warranty HF scientific inc as vendor warrants to the original purchaser of the instruments to be free of defects in material and workmanship in normal use and service for a period of one year from date of delivery to the original purchaser HF scientific inc s obligation under this warranty is limited to replacing at its factory the instrument or any part thereof Parts which by their nature are normally required to be replaced periodically consistent with normal maintenance specifically lamps including fluorescent backlight reagent desiccant sensors electrodes and fuses are excluded Also excluded are accessories and supply type items Original purchaser is responsible for return of the instruments or parts thereof to HF scientific inc s factory This includes all freight charges incurred in shipping t
28. or the sample has not yet stabilized e g air coming out of solution in the form of bubbles The dashes should remain for a brief period of time 45 seconds If the dashes remain for an extended period of time please contact the HF scientific inc Technical Services Department for further assistance and guidance The following sections describe how to use the instrument under normal operating conditions These sections include details on how to take normal routine turbidity measurements using the instrument WARNING NEVER pour liquid directly into the sample well of the instrument always use a cuvette The instrument will accurately measure the turbidity of a sample using only cuvettes with the black light shield on the cuvette provided by HF scientific inc the optional pour through assembly or the optional flow through assembly 4 03 Grab Sample Measurement Routine Measurement The following steps describe how to measure the turbidity of a grab sample using this instrument 1 Turn on the instrument The instrument will be in the normal operating mode Allow instrument to warm up for at least 30 minutes 2 Sample approximately 100 ml of the process stream as you would normally do for turbidity measurement 3 Obtain a clean and dry sample cuvette Rev 1 9 6 4 Rinse the cuvette with approximately 20 ml of the sample water 2 3 of cuvette volume capping the cuvette with the black light shield cuvette top and inverting
29. out the instrument These messages are for your use and do not reduce the performance of the instrument The following sections outline warning messages that could be observed during normal operation 7 1 1 Flashing Cal block A flashing Cal block observed during normal automatic mode indicates that you should re calibrate your instrument The factory default is 30 days HF scientific inc recommends calibration every 30 days The flashing Cal block is only a warning and does not mean that the instrument will stop performing accurately The Cal block will flash until you have re calibrated the instrument 7 1 2 Sample Over Range O r The instrument is capable of reading the turbidity of samples up to 10 000 NTU or equivalent in the other scales If the turbidity of the sample is determined to be higher than 10 000 NTU the instrument will report O r in the upper row of the display Rev 1 9 20 7 1 3 Other warning messages W XX The following table outlines the general warning messages that could arise normally after calibration is performed Rev 1 9 21 WARNING ASSOCIATED MEANING TYPICAL CAUSE Lamp Failure The lamp is no longer working or the light output of the lamp is too low for normal operation Transmission detector 1 error Transmission detector 1 failed to report the proper signal levels during calibration the detector circuitry has failed or calibration was performed with t
30. rned Off you may obtain a printout of the displayed turbidity level each time you press the EJ key in normal mode Timed Output Several automatic printing times are available 5 sec 30 sec 1 min 5 min 10 min 15 min 30 min and 60 min If you selected this option the instrument will print out the average turbidity level time and date for the interval that you select The timing for the interval starts at midnight If you selected to turn off the RS232 port pressing the EJ key will send you to the portion of the code that sets up the date If on the other hand you selected to either turn on the printing function pressing the EJ key will cause the Baud block to be highlighted and you can select the correct baud rate for operation of your printer Select the desired baud rate 1200 2400 4800 or 9600 by pressing either the E or GA keys to change the displayed baud rate The other RS 232 parameters are fixed at 1 stop bit 8 data bits and no parity When you have selected the proper baud rate press the Ed key Note If the security code is enabled and the calibration interval has not been exceeded the letters QC will follow each printout indicating the instrument is functioning properly in a secure fashion If the security code is turned off or if the calibration interval has been exceeded the QC mark will not appear on the printout Rev 1 9 16 6 6 Setting the Calibration Interval After pressing the EJ key
31. sample cuvette used in the flow through assembly is identical to the standard cuvettes supplied with the instrument Clean the cuvette on a regular basis according to your experience with the type of turbidity sample found in your facility 4 5 Dry Air Purge System Cold samples or the continuous flow of cool water through the flow through cuvette may cause condensation on the walls of the cuvette This condensation can cause faulty and erratic readings with the instrument To eliminate this effect a built in dry air purge system has been designed to optimally circulate air through the optical compartment and prevent this condensation This is a particularly important feature when using either the pour through or the flow through option To use connect the regulated dry air source to the air purge fitting on the rear panel of the instrument The fitting is designed for use with 1 8 I D flexible tubing Allow air to flow as long as needed to control condensation dry nitrogen or instrument grade air up to 15 PSIG 103 5 kPa should be used at flow rates between 2 7 SCFH Once installed allow the system approximately 15 minutes to evacuate the chamber before proceeding with your measurements 4 6 Using the Security Access Feature Good Manufacturing Practice GMP The instrument is equipped with a security feature that can be turned on or off when configuring the instrument Two access codes are used in the instrument and configurable by the
32. several times Discard the 20 ml of used sample and repeat the rinsing procedure two more times 5 Completely fill the rinsed cuvette from step 4 with the remaining portion approximately 30 ml of the grab sample and then cap the cuvette with the black light shield Ensure that the outside of the cuvette is dry clean and free from smudges 6 Place the cuvette in the instrument and index the cuvette to the lowest reading the displayed turbidity is updated once a second Once the cuvette is indexed the reading displayed on the instrument s display should be recorded as the sample turbidity see Glossary for more information on indexing a cuvette 7 If you have turned off the printer output in the customer setup section pressing the EJ key at this time will output data to the RS232 port Repeat steps 2 through 7 for all of your samples 4 3 Pour Through Sample Measurement Optional Accessory Install the pour through assembly and index it according to the instruction sheet that accompanies the assembly The following steps describe how to measure the turbidity of a sample using the instrument fitted with the optional pour through assembly 1 Turn on the instrument The instrument will be in the normal automatic mode Allow instrument to warm up for at least 30 minutes 2 Sample approximately 500 ml of the process as you would normally do for turbidity measurement 3 Pour the complete 500 ml sample into the pour through assembly Re
33. stored with the black light shield cap on the cuvette and can be stored in a cuvette rack see accessories and replacement parts list During normal operation you may use any typical glass cleaner along with a lint free cloth or tissue to clean the outside of the cuvettes 8 2 Lamp Replacement Periodically the lamp module will require replacement Figure 1 illustrates the location of the lamp module An error message will be illuminated when it is time to replace the lamp see Troubleshooting We recommended that you keep one spare lamp module for each instrument on hand at all times to ensure continuous use of the instrument Before replacing the lamp module ensure that the instrument is tur ned off Once you have turned off the instrument proceed with the following instructions 1 Remove the lamp module from the instrument by squeezing the two side tabs on the module inward while pulling the module out of the instrument Pull the module away from the instrument until the in line power connector is exposed about 6 8 inches 2 Unfasten the connector by holding on to the white in line connector and pulling the in line connector apart When pulling the in line connector apart DO NOT hold on to the wires 3 The new lamp module can now be connected to the instrument using the in line power connector 4 Feed the wire back into the instrument being careful that the wire does not get in the way of the lamp or the lamp holder on instrument M
34. the lower row of the display will have the letters Int printed in it The number in the upper row of the display corresponds to the number of days that you wish to have between scheduled calibrations default is 30 days Select the desired number of days between scheduled calibrations by pressing either the Cy or V7 keys to change the displayed day In normal automatic mode if you exceed this number of days between calibrations the Cal block will flash until you re calibrate the instrument When you have selected the desired calibration interval press the EJ key 6 7 Setting the Date Year Month and Day With the Year block highlighted and the year displayed change the displayed year using either the E or V7 keys When you have selected the proper year press the EJ key to accept the year After pressing the EJ key the Day Month block will be displayed and you will see two numbers on the upper row of the display The number flashing corresponds to the month Select the correct month by pressing either the C or keys to change the displayed month When you have selected the proper month press the EJ key After pressing the EJ key the Rev 1 9 17 Day Month block will remain displayed and the second number on the upper row of the display will be flashing this number corresponds to the day of the month Select the correct day by pressing either the C or V7 keys to change the displayed day When you have sel
35. y air purge connector Figure 2 illustrates all the items that can appear on the display The upper row of the display 1 is used for reporting the turbidity levels and to provide user guidance in the customer setting routine The lower row of the display 2 is used to communicate error messages and user guidance The display has several status indicators 3 that distinguish the operation of the instrument In addition there are several indicators that display the different scales 4 that the instrument will report Other indicators 5 provide guidance when the setup and calibration routines are being used n WW WUWU Figure 2 Display used in the instrument All items that are used on the display are shown in this figure Safety This manual contains basic instructions that you must follow during the commissioning operation care and maintenance of the instrument The safety protection provided by this equipment may be impaired if it is commissioned and or used in a manner not described in this manual Consequently all responsible personnel must read this manual prior to working with this instrument The instrument has a time out feature that automatically returns the system operation to the normal operation mode from any other mode after a five 5 minute period of inactivity Ensure that the instrument has been turned off prior to removing power from the instrument In certain instances WARNINGS Notes or helpful hints have b

Download Pdf Manuals

image

Related Search

Related Contents

Subaru Robin Power Products R1210 User's Manual  PDF版  Cabinet Dimensions SCN60 User`s Manual  Manuel D`UTILISATION - EuroTeknika Implants  取扱説明書  PDF 1334KB  SylvX User's Manual 1.0  

Copyright © All rights reserved.
Failed to retrieve file