Home

Stratos®Pro A4... MSCONDI - User Manual

image

Contents

1. of interest Output current mS cm mS cm 200 4 200 f 100 4 Output current 0 T 100 4 20 mA 4 20 mA 57 Current Output 1 Output current curve bilinear Press menu key Select CONF using 4 gt keys press enter Select parameter set using 4 gt press enter Select OUT1 menu using keys press enter All items of this menu group are indicated by the OT1 code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter Exit Press meas key until the meas mode indicator is displayed J UN I I UI e Current range De Process variable LIN biLIN LOG output Current start Current end Bilinear Vertex X Bilinear Vertex Y Time averaging filter Output current during error message Output current during HOLD Output current for HOLD FIX for Cond only Menu item Action Choices Output current Select using 4 v keys LIN curve Press enter to confirm Linear characteristic biLIN Bilinear curve LOG Logarithmic curve Current start Enter value using v Entered value applies to and current end 4 keys selected process variable range If the adjusted range is exceeded the device Press enter to confirm automatically switches to the next higher range Autorange Bilinear curve Enter value using 4 v Entered value applies to Vertex X Y 4 keys selected vertex of bilinear curve Corner X proces
2. 144 Calibration entry of cell factor 102 Calibration mode 97 Calibration operating mode 32 Calibration product 100 Calibration zero point 103 CD ROM 3 Cell factor input for calibration 102 Certificates extra booklet 3 Certificates overview 19 CIP Clean in Process 55 Commissioning 12 Compensation introduction 8 Compensation settings 68 Compulsory marking 13 Concentration curves Concentration curves Concentration curves Concentration curves Concentration curves 01 sodium chloride solution NaCl 132 02 hydrochloric acid HCI 133 03 sodium hydroxide solution NaOH 134 04 sulfuric acid H2SO4 135 05 nitric acid HNO3 136 Concentration measurement ranges 131 Configuration Configuration Configuration Configuration Configuration Configuration Configuration Configuration Configuration Configuration Configuration Configuration Configuration Configuration Configuration Configuration Configuration Configuration 32 alarm 74 cleaning cycles 54 concentration solution selection 52 CONTROL input 72 controller 88 current output 1 56 current output 2 66 individual data for copy 46 limit function 78 menu groups 39 output current during Error and HOLD 64 overview 41 Sensocheck 74 sensor 50 tag number 94 temperature compensation 68 time and date 94 145 Configuration WASH contact 92 Configure output current range 1 56 Configure output current range 2
3. Concentration measurement not possible in this range Conductivity versus substance concentration and process temperature for hydrochloric acid HCI Source Haase Sauermann D cker Z phys Chem New Edition Vol 47 1965 133 03 Sodium hydroxide solution NaOH 10 d 03 gt 1 600 1 400 1 200 1000 800 600 400 200 0 X mS cm 4 10 100 C An 90 C 80 C 70 C 60 C ia 50 C 40 C Ko 0 Concentration measurement not possible in this range 40 50 c 96 by wt Conductivity versus substance concentration and process temperature for sodium hydroxide solution NaOH 134 04 Sulfuric acid H SO 06 09 4 04 4 09 gt 06 mS cm 1 800 m 15 6 C 1 600 odo e 1 400 1 200 1000 800 600 400 200 cy D 60 80 100 c 96 by wt o Concentration measurement not possible in this range Conductivity versus substance concentration and process temperature for sulfuric acid H SO Source Darling Journal of Chemical and Engineering Data Vol 9 No 3 July 1964 135 05 Nitric acid HNO 08 lt 05 08 mS cm 1 200 x I ee 50 C 1 000 05 40 C Bag Il 30 C 800 ES N 600 LC 10 C 0 C 400
4. If the display is set to the message text you can search for a particular message using the 4 keys Press to display the date and time Press meas to return to measurement Extended logbook Audit Trail via TAN By using the a v keys you can scroll backwards and forwards through the extended logbook entries 000 199 000 being the last entry Display CFR Audit Trail also records function activations CAL CONFIG SERVICE some Sensoface messages and opening of the enclosure Displaying the currently measured values sensor monitor Select MONITOR using 4 gt press enter to confirm Use the 4 keys to select the desired parameter from the bottom line of the display R COND G COND RTD TEMP I INPUT Option The selected parameter is shown in the main display Press meas to return to measurement 111 112 Menu item Version Here you find the data you require for requesting a device specific Option Display of device type software hardware version and serial number for all device components Use the 4 keys to switch between software and hardware version Press enter to proceed to next device component Service In the Service mode you can access the following menus MONITOR Displaying currently measured values OUT Testing current output 1 OUT2 Testing current output 2 RELAY Testing the function of the 4 relays CONTROL Testing the controller function IRDA Activating and c
5. Meas value x Kc Controller gain 96 Measuring range Neutral zone Y 0 Tolerated deviation from desired value With the setting 1 mS cm for example a deviation of 0 5 mS cm from the desired value does not activate the controller Proportional action Gradient K 96 100 96 7 2 Kc 500 96 Kc 200 Kc 100 96 Controller output Y 50 96 Kc 50 96 Xy MR 0 2 MR 0 4 MR 0 6 MR 0 8 Measuring T range MR Deviation Process variables Cond Conc SAL 86 Pulse Length Pulse Frequency Controller Pulse length controller PLC The pulse length controller is used to operate a valve as an actuator It switches the contact on for a time that depends on the controller output The period pulse length is constant A minimum ON time of 0 5 sec is maintained even if the controller output takes correspond ing values Y 0 Off Output signal relay contact of pulse length controller ON time Y 20 96 fa ON time Y 80 96 1 pp 0 p Pulse length l Pulse frequency controller PFC The pulse frequency controller is used to operate a frequency controlled actuator metering pump It varies the frequency with which the contacts are switched on The maximum pulse frequency pulses min can be defined It depends on the actuator The contact ON time is constant It is automatically calculated from the user defined maximum pulse frequency Output signal rela
6. 0100 I TIME 0000 9999 SEC 0000SEC D TIME 0000 9999 SEC 0000SEC HOLD MODE Y LAST Y OFF Y LAST 44 Configuration Configuration Choices Default WSH Selected in text line WASH WASH PARSET A B WASH WASHCYCLE 000 0 999 9h 000 0 h WASH TIME 0000 9999 SEC 0060 SEC CONTACT N O N C Select fixed parameter set PARSET FIX PARSET FIX A or switch between A B CNTR INPUT fixed parameter via control input or manu MANUAL set A ally in measuring mode FORMAT 24h 12h Po TIME hh mm 00 23 00 59 Po 12h TIME hh mm 00 11 00 59 AM PM DAY MONTH 01 31 01 12 Te YEAR 2000 2099 Po 45 Configuration Original for Copy Two complete parameter sets are stored in the EEPROM As delivered the two sets are identical but can be edited Note Fill in your configuration data on the following pages or use them as original for copy 46 Parameter Parameter set A Parameter set B SNS Sensor type sc SNS RTD type SNS Cell factor n SNS Transfer ratio SNS Measuring mode SNS Measuring range ze SNS Concentration determination SNS Temperature unit S tz SNS CIP counter Ss SNS SIP counter a OT1 Current range OT1 Process variable OT OT1 Current start OT OT1 Vertex X bilinear curve only OT1 Vertex Y bilinear curve only OTI Filter time Lin bilin log output Current end ar OT1 22 mA error current OT1 HOLD mode OT1 HOLD FIX c
7. C caiman determination d NaCl 0 2696 by wt O C _ 0 28 by wt 100 C 02 HCI 0 1896 by wt 20 C 0 18 by wt 50 C 03 NaOH 0 13 by wt 0 C 0 24 by wt 100 C 04 H SO 0 26 by wt 17 C 0 3796 by wt 110 C 05 HNO 0 30 by wt 20 C 0 30 by wt 50 C 06 H SO 94 99 by wt 17 C 89 99 by wt 115 C 07 HCl 22 39 by wt 20 C 22 39 by wt 50 C 08 HNO 35 96 by wt 20 C 35 96 by wt 50 C 09 H SO 28 88 by wt 17 C 39 88 by wt 115 C 10 NaOH 15 50 by wt 0 C 35 50 by wt 100 C 121 Sensor standardization Permissible cell factor Permissible transfer ratio Permissible zero offset Input of cell factor with simultaneous display of selected process variable and temperature Input of calibration solution conductivity with simultaneous display of cell factor Product calibration for conductivity Zero adjustment Temperature probe adjustment 00 100 19 9999 cm 010 0 199 99 cm X 0 5 mS Sensocheck Monitoring of primary and secondary coils and lines for open circuit and of primary coil and lines for short circuit Delay Approx 30 s Sensoface Provides information on the sensor condition zero point Sensocheck Sensor monitor Direct display of measured values from sensor for validation resistance temperature Calibration data 122
8. Output current Select ON 22 mA for during error message Error message or OFF using 4 v keys Press enter to confirm DEE AiL e eah conr Output current LAST During HOLD the during HOLD last measured value is maintained at the output FIX During HOLD a value to be entered is main tained at the output Select using 4 v Press enter to confirm Output current for Only with FIX selected HOLD FIX Enter current which is to flow at the output during HOLD Enter value using v 4 keys Ut HOLT Press enter to confirm Output signal during HOLD Output current Output signal for HOLD mA FIX setting 21 0 mA 21 Choices ON OFF LAST FIX 00 00 21 00 mA Output signal for HOLD LAST setting HOLDactive HOLDactive 65 Current Output 2 Output current range Process variable Press menu key Select CONF using 4 gt keys press enter Select parameter set using 4 gt press enter Select OUT2 menu using 4 gt keys press enter All items of this menu group are indicated by the OT2 code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter Exit Press meas key until the meas mode indicator is displayed Current range La Process variable LIN biLIN LOG output Current start Current end Bilinear Vertex X Bilinear Vertex Y Time averaging filter Output current during error message
9. Output current during HOLD Output current for HOLD FIX Menu item Action Current range Select 4 20 mA or 0 20 mA range using v keys d eu LI E Te RANGE Press enter to confirm Process variable Select using keys Cond Conductivity TMP Temperature Press enter to confirm Choices 4 20 mA 0 20 mA Cond TMP Begin 0 C End 100 C All the following adjustments are made as for current output 1 see there 67 Temperature Compensation Selecting the compensation method TC process medium J UN I I UI e Press menu key Select CONF using 4 gt keys press enter Select parameter set using 4 gt press enter Select CORRECTION menu using 4 gt keys press enter All items of this menu group are indicated by the COR code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter Exit Press meas key until the meas mode indicator is displayed Temperature compensation ve Current input external temp measurement Current start Current end Menu item Temperature compensation Temperature compensation process medium Enter reference temperature de REF TEMP Action Select desired compensa tion using 4 w keys OFF Temperature compensa tion switched off LIN Linear temperature com pensation with entry of temperature coefficient nLF Temperature compensa tion for natural wate
10. Relay 2 active Controller output 100 0 Relay 2 active Meas value setpoint w Relay 2 active Controller output 0 100 Relay 1 active Meas value setpoint IrDA communication Select IRDA using 4 gt press enter to confirm When IrDA communication is active the device remains in the HOLD mode for reasons of safety HOLD j Further operation is performed via IrDA End communication by pressing meas Exception Firmware update must not be interrupted 115 116 Menu item Assigning passcodes In the SERVICE CODES menu you can assign pass codes to DIAG HOLD CAL CONF and SERVICE modes Service preset to 5555 When you have lost the Service passcode you have to request an Ambulance TAN from the manufac turer specifying the serial number of your device To enter the Ambulance TAN call the Service func tion and enter passcode 7321 After correct input of the ambulance TAN the device signals PASS for 4 sec and resets the Service passcode to 5555 Reset to factory settings In the SERVICE DEFAULT menu you can reset the device to factory settings Caution After a reset to factory setting the device must be reconfigured completely including the sensor parameters Option request Communicate the serial number and hardware software version of your device to the manufacturer These data can be viewed in the Diagnostics Version menu The transaction number TAN
11. lt 10 C amm 20 C 0 20 40 60 100 80 c 96 by wt Concentration measurement not possible in this range Conductivity versus substance concentration and process temperature for nitric acid HNO Source Haase Sauermann D cker Z phys Chem New Edition Vol 47 1965 136 137 Error Handling Alarm condition The display backlighting turns red The alarm icon WW is displayed The complete measured value display blinks e ERR xxx is displayed in the lower menu line Press the info key to view a short error text The error text appears in the lower menu line e The main display reads InFo Parameter errors Configuration data such as current range limit values etc are checked during the input If they are out of range e ERR xxx is displayed for 3 sec e the display backlighting flashes red e the respective maximum or minimum value is shown e input must be repeated If a faulty parameter arrives through the interface IrDA HART an error message will be displayed ERR 100 199 the faulty parameter can be localized by pressing the info key Calibration errors If errors occur during calibration e an error message will be displayed Sensoface If the Sensoface becomes sad e the display backlighting will turn purple the cause can be seen by pressing the info key the calibration data can be seen in the Diagnostics menu 138 ERR 99 ER
12. via RS 485 Measuring task Conductivity temperature Sensor SE 670 Caution Connection to RS 485 interface The measuring module slot must be empty o UU E EH Ss SEES au MEE DE oe X S gt ee fee fe Ou a 1 12 3 14 5 c v bel Ei c 2 all EI oOo OF j2 2 9 2 gy S e al 0 ZzW t U m o vo E v9 Vi The SE 670 sensor is connected to the RS 485 interface of the device When the SE 670 sensor is selected in the Configuration menu the default values are taken as calibration data They can then be modified by calibration Stratos Pro A2 MSCONDI is intended for connecting a SE 670 sensor via RS 485 interface It does not provide a measuring module 22 23 Protective Wiring of Relay Contacts Relay contacts are subject to electrical erosion Especially with inductive and capacitive loads the service life of the contacts will be reduced For suppression of sparks and arcing components such as RC combinations nonlinear resistors series resistors and diodes should be used T Typical AC applications with inductive load 1 1 Load 2 RC combination e g RIFA PMR 209 Typical RC combinations for 230 V AC capacitor 0 1 uF 630 V resistor 100 0 1W 3 3 Contact wo genie N 24 Typical Protective Wiring Measures A B C A1 A2 A3 B1 B2 B3 C1 C3 B2 y 5 DC application with inductive load AC DC applications with
13. you will then receive is only valid for the device with the corresponding serial number Release of options Options come with a transaction number TAN This TAN must be entered and confirmed using enter to release the option 117 Operating States Operating status N Si O O UT Measuring EN EN Crown MEM NEN Cal solution Cano a W Cell factor P CAL Product cal S1 nas NEN NN Product cal S2 CAL RTD Temp adjustm ES pedi c D ParSet A CONF ParSet B Bei o i Passcodes Monor MM NM MONITOR SERVICE OUT 1 E WI SERVICE OUT 2 Ep MR SERVICE RELAIS wo MEN CONTROL 118 E a 3 0 D NP GG 8 I SESCH EEN cei SO x89 33 I EH IN 77 77 77 77 Time out 60 5 No No No No 20 min 20 min 20 min 20 min 20 min 20 min 20 min 20 min 20 min 3 38 Ba3 BH2 28 8 SERVICE m EX 20 IRDA min Cleaning fct Ei No HOLD input No Explanation LJ as configured Last Fix or Last Off active manual 119 Product Line and Accessories Order Code Stratos Pro A 4 Channel 1 Channel2 TAN Example A 4 1 PH 4 wire 20 254V AC DC A 4 BI E Communication Without HART retrofittable via TAN A Version number Version 1 Approvals General Safety ATEX IECEX FM CSA Zone 2 Cl 1 Div 2 Meas channel 1 Meas channel 2 Memosens pH ORP Digit
14. 66 Configure output signal during HOLD 65 Connection example 22 Contact alarm configure 75 Contact alarm schematic 10 Contact WASH configure 92 Contact WASH schematic 10 Contacts service life 24 CONTROL for flow measurement 37 CONTROL for parameter set selection 37 CONTROL input 36 Control inputs CONTROL 72 Control inputs HOLD 35 Control inputs overview 9 CONTROL floating switching input 72 Controller 85 Controller characteristic 85 Controller configuration 88 Controller controller equations 86 Controller controller functions 85 Controller controller test 115 Controller controller type 89 Controller manual specification of controller output 115 Controller neutral zone 91 Controller pulse frequency 89 Controller pulse length 89 Controller setpoint 89 Current output 1 56 f Current output 2 66 Current start end 57 D Data logger display entries 111 Data logger explanation 8 Date and time display 105 Date and time use 95 146 Device self test 110 Diagnostics functions 32 Diagnostics mode 108 Diagnostics calibration data 109 Diagnostics device self test 110 Diagnostics sensor monitor 111 Diagnostics software version 112 Dimensions 16 Display backlighting 27 Display colors 31 Display device type 112 Display test 110 Display icons and colors 27 Display select main display 30 Disposal 2 Documentation package contents 3 E EEPROM test device self test 110 Electric
15. For the solutions listed above the device can determine the sub stance concentration from the measured conductivity and tempera ture values in by wt The measurement error is made up of the sum of measurements errors during conductivity and temperature measurement and the accuracy of the concentration curves stored in the device We recommend to calibrate the device together with the sensor e g directly to concentration using the CAL CELL method For exact temperature measurement you should perform a temperature probe adjustment For measuring processes with rapid temperature changes a separate temperature probe with fast response should be used When measuring processes such as dilution or intensification of CIP solutions Clean In Place it is helpful to switch between the param eter sets for measuring the process medium and for measuring the CIP solution 131 Concentration Curves 01 Sodium chloride solution NaCl 09 9 X mS cm 100 C 20 C 10 C 0 C 5 30 c 96 by wt Concentration measurement not possible in this range Conductivity versus substance concentration and process temperature for sodium chloride solution NaCl 132 02 Hydrochloric acid HCI 07 z eh 0 X mS cm 1 200 1 000 800 C 5 TE 600 r 10 C 400 20 C 200 0 0 10 20 30 e 96 by wt 40
16. Select process variable Limit 1 switching characteris tics function Limit 1 contact type Limit 1 setpoint Limit 1 hysteresis Limit 1 delay Use of relays See D Menu item Action Choices Limit 1 hysteresis Select hysteresis using 0 50 96 full scale v 4 keys Press enter to confirm Limit 1 delay The contact is activated 0 9999 SEC with delay deactivated 0010 SEC without delay Adjust delay using 4 v 4 keys Press enter to confirm Limit Lo Limit Hi Signal Hysteresis E Setpoint Setpoint Hysteresis votadas dab tuat RR QNO iid 1 1 Contact Contact Wd Ll 0 0 81 Limit Function Relay 2 Wn gt ie YS H Press menu key Select CONF using 4 keys press enter Select parameter set using 4 gt press enter Select REL1 REL2 menu using 4 gt keys press enter All items of this menu group are indicated by the RL2 code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter Exit Press meas key until the meas mode indicator is displayed Select process variable enter Limit 2 switching characteris tics function Limit 2 contact type Limit 2 setpoint Limit 2 hysteresis Limit 2 delay x7 X 7 Menu item Select process variable CHANNEL Limit 2 function FUNCTION Limit 2 contact type CONTACT Limit 2 setpoint LEVEL Limit 2 hystere
17. can change the input range using the left right arrow keys 10 9999 96 0100 96 0 9999 SEC 0000 SEC 0 9999 SEC 0000 SEC Y LAST Y OFF 91 WASH Contact Controlling a rinsing probe or signaling the parameter set 1 Press menu key 2 Select CONF using 4 gt keys press enter 3 Select parameter set A using 4 gt keys press enter 4 Select WASH menu using 4 gt keys press enter 5 All items of this menu group are indicated by the WSH code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter 6 End Press meas key until the meas mode indicator is displayed Function enter Cleaning interval Cleaning duration Contact type TE Menu item Action Function Select WASH contact function using 4 vw keys Press enter to confirm Only with WASH Adjust value using 4 v 4 keys Press enter to confirm Only with WASH Adjust value using 4 v 4 keys NGH WASH TU Press enter to confirm Contact type Only with WASH N O normally open contact N C normally closed contact Select using v keys Press enter to confirm Cleaning Relax time approx 20 s gt la Choices WASH PARSET A B WASH Control of rinsing probes With PARSET A B selected the contact signals Parameter set A open contact Parameter set B closed contact 000 0 999 9 h 000 0 h 0 9999 SEC
18. capacitive load Connection of incandescent lamps Inductive load Free wheeling diode e g 1N4007 Observe polarity Contact Capacitive load Resistor e g 8 O 1W at24V 0 3A Contact Incandescent lamp max 60 W 230V 30W 115 V Contact N WARNING Make sure that the maximum ratings of the relay contacts are not exceeded even during switching 25 User Interface Keypad Le OLY Su A IX tji SA IrDA transmitter receiver Display Keypad Rating plate enclosure bottom AUN Key Function meas Return to last menu level Directly to measuring mode press 2 s e Measuring mode other display info e Retrieve information Show error messages enter Configuration Confirm entries next configuration step e Calibration Continue program flow menu e Measuring mode Call menu Arrow keys e Menu Increase decrease a numeral up down e Menu Selection Arrow keys Previous next menu group left right Number entry Move between digits 26 C NA WI WW NW DW WW NW WI WWW 4 JAN IW JANG TAN pun ee ee ae 1 Temperature 2 Sensocheck 3 Interval response time 4 Sensor data 5 Notused 6 Limit message Limit 1 or Limit 2 amp 7 Alarm 8 Service 9 Parameter set 10 Calibration 11 Memosens 12 Waiting time running memo S ENS e LAN TAN DAN WANT JAN law eee JAN TANT JAN ae VAN TAN JANI AN 19 13 Info available 14 Hold mode active 15 Main display 16 Secondary d
19. confirm Time averaging filter To smoothen the current output a low pass filter with adjustable filter time constant can be switched on When there is a jump at the input 100 96 the output level is at 63 96 after the time interval has been reached The time interval can be set from 0 to 120 sec If the time interval is set to 0 sec the current output directly follows the input Please note The filter only acts on the current output not on the display or the limit value During HOLD the filter is not applied This prevents a jump at the output Display 22 mA Time interval 0 120 s HOLD AN 22 mA OUT2 0 4 20 mA Time interval 0 120 s 63 Current Output 1 Output current during Error and HOLD gt W is UJ N gt Zn UI e P ad Press menu key Select CONF using 4 gt keys press enter Select parameter set using 4 gt press enter Select OUT1 menu using 4 gt keys press enter All items of this menu group are indicated by the OT1 code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter Exit Press meas key until the meas mode indicator is displayed Current range er Process variable LIN biLIN LOG output Current start Current end Time averaging filter Output current during error message Output current during HOLD Output current for HOLD FIX Menu item Action
20. displays each with meas 2 Display of tag number TAG 3 Display of time and date 105 Display Remark With activated controller you can also step through the following displays by pressing the meas key When no key has been pressed for 60 sec the device returns to the standard display Main display Controller output Y Secondary display Setpoint Depending on configuration setting conductivity or temperature Please note After prolonged power outage gt 5 days the time display is replaced by dashes and cannot be used for processing Enter the correct time 106 107 Diagnostics In the Diagnostics mode you can access the following menus without interrupting the measurement CALDATA SENSOR SELFTEST LOGBOOK MONITOR VERSION viewing the calibration data viewing the sensor data starting a device self test viewing the logbook entries displaying currently measured values displaying device type software version serial number Access to diagnostics can be protected with a passcode SERVICE menu Please note HOLD is not active during Diagnostics mode Action Activate Diagnostics Select diagnostics option Exit 108 Key dh 9 menu meas Remark Press menu key to call the selection menu Display color changes to turquoise Select DIAG using 4 gt keys press enter to confirm Use 4 gt keys to select from CALDATA SENSOR SELFTEST LOGBOOK MONITOR VERSI
21. mode indicator is displayed Select process variable Limit 1 switching characteris tics function Limit 1 contact type Limit 1 setpoint Limit 1 hysteresis Limit 1 delay Use of relays N 2 Menu item Use of relays Select process variable Limit 1 contact response Action Select in the text line using 4 keys e Limit function LIMITS e Controller CONTROLLER Press enter to confirm Select desired process variable using 4 keys Press enter to confirm Select desired function using 4 v keys LoLevel active if value falls below HiLevel active if value exceeds setpoint Press enter to confirm N O normally open contact N C normally closed contact Select using v keys Press enter to confirm Enter setpoint using 4 v 4 keys Press enter to confirm Choices LIMITS CONTROLLER Note Selecting CONTROLLER leads to Controller menu group CTR Cond TMP Lo LEVL Hi LEVL Limit 1 icon I N O N C Within meas range 79 Limit Function Relay 1 Press menu key Select CONF using 4 keys press enter Select parameter set using 4 gt press enter Select REL1 REL2 menu using 4 gt keys press enter All items of this menu group are indicated by the RL1 code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter Exit Press meas key until the meas mode indicator is displayed
22. number Manual activation of HOLD mode e g for sensor replacement The signal outputs behave as configured e g last measured value 21 mA Calibration with calibration solution Calibration by input of cell factor Zero calibration Product calibration Adjustment of temperature probe Configuring parameter set A Configuring parameter set B Display of measured values for validation simulators Current source output 1 Current source output 2 Relay test Controller manual specification of controller output Activating the IrDA interface Specifying access codes for operating modes Reset to factory setting Enabling an option via TAN 33 HOLD Mode The HOLD mode is a safety state during configuration and calibration Output current is frozen LAST or set to a fixed value FIX Alarm and limit contacts are disabled The HOLD mode is indicated by orange display backlighting HOLD mode display icon HOLD Output signal response LAST The output current is frozen at its last value Recommended for short configuration procedures The process should not change decisively during configuration Changes are not noticed with this setting FIX The output current is set to a value that is noticeably different from the process value to signal the control system that the device is being worked at Output signal during HOLD Output curr
23. page How do get to the function Stratos Pro A2 Ad pH S z gt I ZOHEZAS m m Ai ie 5 S Documents Supplied Specific Test Report CD ROM Complete documentation e User manuals Safety instructions e Certificates Short instructions Safety Information In official EU languages and others EC Declarations of Conformity Certificates e ECEx e ATEX e FM e CSA e NEPSI e GOST Short Instructions First steps after installation Operation e Menu structure e Calibration Error messages and recommended actions In German English French Russian Spanish Portuguese Italian Swedish and Dutch More languages on CD ROM and on our website www knick de Contents Documents Supplied E 3 Introduction EE EROR 7 Intended D 7 Safety Inform tlon an 12 Safety Precautions for Installation 13 Overview of Stratos Pro A4 MSCONDI eere 14 Assembly ee 15 Package Content 15 Mounting Plan Dimensions een nn 16 Pipe Mounting Protective Hood sene 17 CHE TVG CIA CUES een 18 E HIEN d EEN 19 Installation Instructions aeien ea 19 Rating Plates Terminal Assignments sss 20 Power Supply Signal D aita ee 21 Wiring Example SE 670 via RS 485 esses 22 Protective Wiring of Relay Contacts sss 24 User Interface Keypad rere eese tente e
24. pressing enter or meas Relay test manual test of contacts Select RELAIS using 4 gt press enter to confirm Now the status of the 4 relays is frozen The 4 digits of the main display represent the respective states from left to right REL1 REL2 ALARM WASH Select one of the 4 relays using the 4 gt keys close 1 or open 0 using the a vw keys The selected position blinks The lower display line shows the relay function e g WASH or ALARM End by pressing enter The relays will be re set corresponding to the measured value Press meas to return to measurement Menu item Controller test manual specification of controller output This function is used to start up control loops or check the actuators For bumpless changeover to automatic operation exiting this function configure an l action component reset time Controller characteristic The lower display shows the currently adjusted controller output Yp 100 96 Controller output Yp Pdl Specify new value for controller output Yp Enter sign and value in the main display using the v 4 keys Press enter to confirm Relay 1 Ka Neutral zone Yp 0 Deviation Xw mS cm The new value is taken into the lower display Setpoint EX Relay 2 Press enter or meas to return to the service menu Return to measurement Press meas once more 100 The arrows indicate which relay valve is active amp
25. 00 19 99 K 00 00 K REF TEMP 000 0 199 9 C 025 0 C I INPUT only with TEMP EXT selected The l input option must be enabled via TAN TEMP EXT I INPUT 0 20 mA 4 20 mA 4 20 mA C BEGIN4 mA 50 250 C 000 0 C END 20 mA 50 250 C 100 0 C F BEGIN4mA 58 482 F END 20 mA 58 482 F Control input CNTR_IN IN CONTROL Parameter set PARSET switchover PARSET or flow measurement FLOW FLOW FLOW ADJUST 12000 0 20000 pulses liter pulses liter Alarm ALARM ALA DELAYTIME 0 600 SEC 0010 SEC SENSOCHECK ON OFF OFF FLOW CNTR ON OFF OFF ON FLOW MIN 005 0 L h 0 99 9 L h FLOW MAX 025 0 L h 0 99 9 L h These menu items appear only if selected Hysteresis fixed at 596 of threshold value 43 Configuration Choices Default Switching outputs Rel1 Rel2 REL Selected in text line LIMITS LIMITS CONTROLLER RL1 CHANNEL Cond TMP FLOW Cond FUNCTION Lo LEVL Lo LEVL Hi LEVL CONTACT N O N C N O LEVEL Within meas range HYSTERESIS 0 50 96 full scale DELAYTIME 0000 9999 SEC 0010 SEC RL2 CHANNEL Cond TMP FLOW Cond FUNCTION Lo LEVL Hi LEVL Hi LEVL CONTACT N O N C N O LEVEL Within meas range HYSTERESIS D 50 96 full scale DELAYTIME 0000 9999 SEC 0010 SEC CTR CHANNEL COND TMP Cond TYPE PLC PFC PLC PLC PULSE 0000 0600 SEC 0010 SEC LENGTH PFC PULSEFREQ 0000 0180 P M 0060 P M SETPOINT Within meas range DEAD BAND 0 50 96 full scale P GAIN 0010 9999
26. 00 1000 mS cm Decades Start value End value 60 Menu item Action Logarithmic curve Select using 4 v keys of output current Press enter to confirm Enter value using 4 v 4 keys Enter value using 4 v 4 keys Press enter to confirm Press enter to confirm Choices LOG Logarithmic curve biLIN Bilinear curve LIN Linear characteristic Start value of logarithmic output curve End value of logarithmic output curve Possible start and end values for the logarithmic curve S cm 0 001 mS cm 0 01 mS cm 0 1 mS cm S m 0 001 S m 0 01 S m 0 1 S m 1 0 S m 10 0 S m 100 S m 61 Current Output 1 Adjusting the time interval of the output filter J UN I I UI e Press menu key Select CONF using 4 keys press enter Select parameter set using 4 gt press enter Select OUT1 menu using keys press enter All items of this menu group are indicated by the OT1 code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter Exit Press meas key until the meas mode indicator is displayed Current range er Process variable LIN biLIN LOG output Current start Current end Time averaging filter Output current during error message Output current during HOLD Output current for HOLD FIX Menu item Action Choices Time averaging filter Enter value using 4 v 0 120 SEC 4 keys 0000 SEC Press enter to
27. 0060 SEC N O N C Aum 9 S HOLD M e Cycle interval Jl 93 Time and Date Tag Number N Ye 8 UI Press any arrow key Select CONF using 4 gt keys press enter Select parameter set A using 4 gt keys press enter Select CLOCK or TAG using 4 gt keys press enter All items of this menu group are indicated by the CLK or TAG code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter End Press meas key until the meas mode indicator is displayed Time format ET Time P Day and month Year Tag number Time and Date Control of the calibration and cleaning cycles is based on the time and date of the integrated real time clock In measuring mode the time is shown in the lower display When using digital sensors the calibration data is written in the sensor head In addition the logbook entries cf Diagnostics are provided with a time stamp Please note After prolonged power outage gt 5 days the time display is replaced by dashes and cannot be used for processing Enter the correct time There is no automatic switchover from winter to summer time Be sure to manually adjust the time Tag Number TAG You can enter a designation for the point of measurement tag number in the lower display line Up to 32 digits are possible Pressing meas repeatedly in the measuring mode indicates th
28. 8 mm x 138 mm to DIN 43 700 Approx 1200 g 1600 g incl accessories and packaging 3 knockouts for M20 x 1 5 cable glands 2 knockouts for NPT 2 or rigid metallic conduit Terminals conductor cross section max 2 5 mm 1 Acc to EN 60746 at nominal operating conditions 3 Plus sensor error Calibration Solutions Potassium Chloride Solutions Conductivity in mS cm Temperature Concentration rc 0 01 mol l 0 1 mol l 1 mol l 0 0 776 7 15 65 41 5 0 896 8 22 74 14 10 1 020 9 33 83 19 15 1 147 10 48 92 52 16 1 173 10 72 94 41 17 1 199 10 95 96 31 18 1 225 11 19 98 22 19 1 251 11 43 100 14 20 1 278 11 67 102 07 21 1 305 11 91 104 00 22 1 332 12 15 105 94 23 1 359 12 39 107 89 24 1 386 12 64 109 84 25 1 413 12 88 111 80 26 1 441 13 13 113 77 27 1 468 13 37 115 74 28 1 496 13 62 29 1 524 13 87 30 1 552 14 12 31 1 581 14 37 32 1 609 14 62 33 1 638 14 88 34 1 667 15 13 35 1 696 15 39 36 15 64 1 Data source K H Hellwege Editor H Landolt R B rnstein Zahlenwerte und Funktionen volume 2 part volume 6 129 Sodium Chloride Solutions Conductivity in mS cm Temperature Concentration rq 0 01 mol l 0 1 mol l Saturated 0 0 631 5 786 134 5 1 0 651 5 965 138 6 2 0 671 6 145 142 7 3 0 692 6 327 146 9 4 0 712 6 510 151 2 5 0 733 6 695 155 5 6 0 754 6 881 159 9 7 0 775 7 068 164 3 8 0 796 7 257 168 8 9 0 818 7 447 173 4 10 0
29. 839 7 638 177 9 11 0 861 7 831 182 6 12 0 883 8 025 187 2 13 0 905 8221 191 9 14 0 927 8 418 196 7 15 0 950 8 617 201 5 16 0 972 8 816 206 3 17 0 995 9 018 2112 18 1 018 9 221 216 1 19 1 041 9 425 221 0 20 1 064 9 631 226 0 21 1 087 9 838 231 0 22 1 111 10 047 236 1 23 1 135 10 258 241 1 24 1 159 10 469 246 2 25 1 183 10 683 251 3 26 1 207 10 898 256 5 27 1 232 11 114 261 6 28 1 256 11 332 266 9 29 1 281 11 552 272 1 30 1 306 11 773 2774 31 1 331 11 995 282 7 32 1 357 12 220 288 0 33 1 382 12 445 293 3 34 1 408 12 673 298 7 35 1 434 12 902 304 1 36 1 460 13 132 309 5 1 Data source Test solutions calculated according to DIN IEC 746 3 2 Data source K H Hellwege Editor H Landolt R Bornstein Zahlenwerte und Funktionen volume 2 part volume 6 130 Concentration Measurement Ranges Substance Concentration ranges NaCl 0 26 by wt 0 C 0 26 by wt 100 C Configuration 91 HCI 0 18 by wt 20 C 22 39 by wt 20 C 0 18 by wt 50 C 22 39 by wt 50 C Configuration _02 07 NaOH 0 13 by wt 0 C 15 50 by wt 0 C 0 24 by wt 100 C 35 50 by wt 100 C Configuration _93 10 HSC 0 26 by wt 17 C 28 77 by wt 17 C 94 99 by wt 17 C 0 37 by wt 110 C 39 88 by wt 115 C 89 9996 by wt 115 C Configuration _04 09 06 HNO 0 30 by wt 20 C 35 96 by wt 20 C 0 30 by wt 50 C 35 96 by wt 50 C Configuration gs 08
30. Calibration date cell factor l input TAN Start end of scale Characteristic Measurement error Current input 0 4 20 mA 50 Q for external temperature signal Configurable 50 250 C 58 482 F Linear lt 1 current value 0 1 mA I HOLD input Function Switching voltage Galvanically separated OPTO coupler Switches device to HOLD mode 0 2 V AC DC HOLD inactive 10 30 V AC DC HOLD active CONTROL input Galvanically separated OPTO coupler Function Selecting parameter set A B or flow measurement Parameter set A B Control input 0 2V AC DC Parameter set A 10 30 V AC DC Parameter set B FLOW Pulse input for flow measurement 0 100 pulses s Message via 22 mA alarm contact or limit contacts Display 00 0 99 9 I h 1 0 4 20 mA max 10 V floating galv connected to output 2 Process variable Characteristic Overrange Output filter Measurement error Start end of scale Bilinear Vertex X Y Minimum span Conductivity resistivity concentration salinity or temperature Linear bilinear or logarithmic 22 mA in the case of error messages PT filter time constant 0 120 s lt 0 25 current value 0 025 mA Configurable within selected range Configurable within selected range LIN 5 96 of selected range LOG 1 decade 123 Output 2 Process va
31. Conc 96 Sal 96o X Xxx mS cm xx xx mS cm Xxx x mS cm x xxx S M xx xx S M 51 Sensor Select Concentration determination temperature unit Ww N gt SE CA UI e Press menu key Select CONF using 4 gt keys press enter Select parameter set using 4 gt press enter Select SENSOR menu using keys press enter All items of this menu group are indicated by the SNS code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter Exit Press meas key until the meas mode indicator is displayed Sensor type emer Temperature probe AL AJ Cell factor Transfer ratio Measuring mode Measuring range Concentration determination Temperature unit Cleaning cycles Sterilization cycles Menu item Concentration determination Action For conc measurement only Select desired concentra tion solution using 4 v see appendix for ranges Press enter to confirm Select C or F using a v keys Press enter to confirm Choices 01 NaCl 02 HCI 10 NaOH C F 53 Sensor ISM only Adjust Cleaning cycles sterilization cycles Press menu key Select CONF using 4 gt keys press enter Select parameter set using 4 gt press enter Select SENSOR menu using keys press enter All items of this menu group are indicated by the SNS code Press enter to select menu edit using arr
32. In that case the device displays the actual value and internally corrects the measurement error of the sensor During calibration the device is in HOLD mode During calibration the device remains in the HOLD mode until it is stopped by the operator Configuration The analyzer must be configured for the respective measurement task In the Configuration mode you select the connected sensor the measuring range to be transmitted and the conditions for warning and alarm messages During configuration the device is in HOLD mode Configuration mode is automatically exited 20 minutes after the last keystroke The device returns to measuring mode Service Maintenance functions current source relay test controller test IrDA operation passcode assignment reset to factory settings enabling of options TAN 32 Menu Structure of Modes and Functions main display selectable v CONF ii SERVICE Access via code factory setting 5555 Y after 60 s Y after 60 s if configured Pressing the menu key down arrow opens the selection menu Select the menu group using the left right arrow keys Pressing enter opens a menu item Press meas to return Display of calibration data Display of sensor data Self test RAM ROM EEPROM module Logbook 100 events with date and time Display of direct uncorrected sensor signals Display of software version model designation serial
33. N PARSET A Max 30 V AC DC o1 CONTROL input e g process control system Stratos Pro A4 Parameter set A active 0 2 V AC DC Parameter set B active 10 30 V AC DC 39 Parameter Set A B Manual Selection Signaling via WASH Contact Display 40 Action Remark To switch between Manual selection of parameter sets parameter sets must have Press meas been preset in CONFIG mode Default setting is a fixed parameter set A Wrong settings change the measurement properties PARSET blinks in the lower line Select parameter set using and keys Select PARSET A PARSET B Press enter to confirm Cancel by pressing meas The active parameter set can be signaled using the WASH contact If configured correspondingly the WASH contact signals Parameter set A open contact Parameter set B closed contact Configuration Sensor SENSOR SNS OTHER RTD TYPE CELL FACTOR TRANS RATIO Choices Default SE 670 SE 670 SE 655 SE 656 SE 660 OTHER 100PT 1000PT 1000PT 30 NTC XX XXx 01 980 XXX Xx 120 0 These sensors appear in the menu selection but can only be used with a measuring module installed Stratos Pro A4 MSCONDI is intended for connecting the SE 670 sensor via RS 485 interface It does not provide a measuring module For information on retrofitting and the respective costs please contact the manufac
34. O HIGH Measuring range of con ductance exceeded 3500 mS ERR 13 TEMPERATURE RANGE Temperature range violation ERR 15 SENSOCHECK Sensocheck ERR ep OUTPUT LOAD Load error ERR 61 OUTPUT 1 TOO LOW Output current 1 0 3 8 mA ERR62 OUTPUT 1 TOO HIGH Output current 1 20 5 mA ERR63 OUTPUT 2 TOO LOW Output current 2 lt 0 3 8 mA ERR 64 OUTPUT 2 TOO HIGH Output current 2 gt 20 5 mA ERR72 FLOWTOO LOW Flow too low ERR 73 FLOWTOO HIGH Flow too high ERR 108 OUT1 INVALID CORNER X Y Bilinear curve ERR 109 OUT2 INVALID CORNER X Y Invalid vertex point 140 Sensoface Sensocheck must have been activated during configuration The smiley in the display Sensoface alerts to sensor problems Gs defective sensor defective cable maintenance required The GB permitted calibration ranges and the conditions for a friendly neutral or sad Sensoface are summarized in the following table Additional icons refer to the error cause Sensocheck Continuously monitors the sensor and leads for short circuits or open circuits Critical values make the Sensoface sad and the correspond ing icon blinks A The Sensocheck message is also output as error message Err 15 The alarm contact is active the display backlighting turns red output current 1 is set to 22 mA when configured correspondingly Sensocheck can be switched off during configuration then Sensoface is also disabled Exception After a calibration a sm
35. ON See next pages for further proceeding Exit by pressing meas Menu item Displaying the calibration data Select CALDATA using 4 gt press enter to confirm Use the 4 gt keys to select the desired parameter from the bottom line of the display LAST CAL CELLFACTOR ZERO The selected parameter is shown in the main display Press meas to return to measurement 109 Display Menu item Device self test To abort you can press meas Le OZY twm We 1 Display test Display of all segments with eH HH H E A1 changing background colors white green red eL EL DRIN Press enter to proceed RU RRRRRRRRRRRE NA gt RAM test Hourglass blinks then display of PASS or FAIL Press enter to proceed LA EEPROM test Hourglass blinks then display of PASS or FAIL Press enter to proceed gt FLASH test Hourglass blinks then display of PASS or FAIL Press enter to proceed 5 Module test Hourglass blinks then display of PASS or FAIL Press enter or meas to return to measuring mode 110 Menu item Displaying the logbook entries Select LOGBOOK using 4 gt press enter to confirm By using the a keys you can scroll backwards and forwards through the logbook entries 00 99 00 being the last entry If the display is set to date time you can search for a particular date using the 4 vw keys Press 4 gt to view the corresponding message text
36. R 98 ERR 97 ERR 96 ERR 95 ERR 100 ERR 101 ERR 104 ERR 105 Info text is displayed in case of fault when the Info key is pressed DEVICE FAILURE CONFIGURATION ERROR NO MODULE INSTALLED WRONG MODULE SYSTEM ERROR INVALID SPAN OUT1 INVALID SPAN OUT2 INVALID PARAMETER CONTROLLER INVALID SPAN I INPUT Error Messages Problem Possible causes Error in factory settings EEPROM or RAM defective This error message only occurs in the case of a total defect The device must be repaired and recalibrated at the factory Error in configuration or calibration data Memory error in device program Configuration or calibration data defective completely reconfig ure and recalibrate the device No module Please have the module replaced in the factory Wrong module Please have the module replaced in the factory System error Restart required If error still persists send in the device for repair Span Out configuration error Span Out2 configuration error Controller configuration error I Input configuration error 139 Info text is displayed in case of fault when the Info key is pressed Problem Possible causes ERR10 gt CONDUCTANCETOO HIGH Measuring range of conductance exceeded gt 3500 mS ERR 11 Display range violation CONDUCTIVITY RANGE Cond gt 1999 mS cm gt 99 99 S m CONCENTRATION RANGE Conc gt 99 9 96 SALINITY RANGE SAL 45 0960 ERR12 CONDUCTANCE TO
37. Stratos Pro A4 MSCONDI User Manual Latest Product Information www knick de ERI Knick gt 81082 Warranty Disposal About This Manual Warranty Defects occurring within 3 years from delivery date shall be remedied free of charge at our plant carriage and insurance paid by sender Sensors fittings and accessories 1 year Subject to change without notice Return of products under warranty Please contact our Service Team before returning a defective device Ship the cleaned device to the address you have been given If the device has been in contact with process fluids it must be de contaminated disinfected before shipment In that case please attach a corresponding certificate for the health and safety of our service personnel Disposal Please observe the applicable local or national regulations concerning the disposal of waste electrical and electronic equipment About this manual This manual is intended as a reference guide to your device You don t have to read the book from front to back Take a look at the Table of Contents or the Index to find the function you are interested in Each topic is explained on a double page spread with step by step instructions on how to configure the desired func tion Clearly legible page numbers and headlines help you to quickly find the information Right page Which settings are provided for this function Left
38. TROL input FLOW MIN FLOW MAX Press menu key Select CONF using 4 gt keys press enter Select parameter set using 4 gt press enter Select ALARM menu using 4 keys press enter All items of this menu group are indicated by the ALA code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter Exit Press meas key until the meas mode indicator is displayed ALARM Delay Alarm Sensocheck Alarm CONTROL input For flow monitoring Max flow alarm For flow monitoring Min flow alarm Menu item CONTROL input Alarm Minimum flow FLOW MIN Alarm Maximum flow FLOW MIN Action The CONTROL input can generate an alarm when assigned to FLOW flow monitoring in the CONF menu FLOW CNTR Flow measurement allows monitoring the minimum and maximum flow pulse counter Specify value Specify value Choices ON OFF FLOW MIN FLOW MAX Default 05 00 liters h Default 25 00 liters h 77 Limit Function Relay 1 Wn gt ie YS H Press menu key Select CONF using 4 keys press enter Select parameter set using 4 gt press enter Select REL1 REL2 menu using 4 gt keys press enter All items of this menu group are indicated by the RL1 code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter Exit Press meas key until the meas
39. al MSPH Memosens pH ORP pH ORP Digital MSPH Memosens pH ORP Oxy Digital MSPH Memosens COND Digital MSCOND Memosens COND COND Digital MSCOND Memosens Oxy Digital MSOx Dual COND 2x2 electrode analog Module CC pH ORP value ISM digital TAN Module PH F Cond 2 4 electrode Module COND Conductivity electrodeless Module CONDI Oxygen ISM digital Traces TAN Module OXY D F TAN options HART SW A001 A Logbook SW A002 B Extended logbook Audit Trail SW A003 C Trace oxygen measurement SW A004 D Current input 2 digital inputs SW A005 E ISM digital SW A006 F Mounting accessories Pipe mount kit ZU 0274 Protective hood ZU 0737 Panel mount kit ZU 0738 120 CONDI input Effective ranges Display ranges Measurement error Specifications Input for electrodeless conductivity sensor SE 670 Conductivity 0 02 2000 mS cm Conductivity 0 000 9 999 mS cm not with SE 660 SE 670 Concentration Salinity Response T90 00 00 99 99 mS cm 000 0 999 9 mS cm 0000 1999 mS cm 0 000 9 999 S cm 00 00 99 99 S cm 0 00 9 99 96 10 0 100 96 0 0 45 0 96 0 35 C Approx 1 s Depending on Memosens Temp compensation OFF Without LIN Linear characteristic 00 00 19 99 K reference temp user defined NLF Natural waters to EN 27888 reference temp 25 C NACL NaCl from 0 to 26 96 by wt 0 120 C reference temp 25
40. al installation 13 Electronic signature 143 Enclosure components 15 Enter cell factor 102 Enter output current value Service mode 114 Enter values 29 Error handling 138 Error messages 139 Explosion protection 127 Extended logbook explanation 143 Extended logbook with TAN 111 F Fault current value 64 FDA 21 CFR Part 11 requirements for measuring device 143 Fix output signal in HOLD mode 34 FLASH test 110 FLOW 73 Flow measurement 72 Flow measurement generate message 37 147 H HOLD mode 34 HOLD behavior of output signal 34 HOLD configure output signal 65 HOLD end 34 HOLD external activation 35 HOLD manual activation 35 HOLD operating status 34 Hysteresis application 81 Info text 139 Information on installation 19 Install enclosure 7 Installation Memosens 7 Installation mounting plan 16 Installation safety information 13 Installation terminal assignments 20 Intended use 7 IrDA 26 IrDA potential errors 138 IrDA schematic diagram 11 IrDA select 115 IrDA specifications 126 K Keypad 26 L LAST output signal in HOLD mode 34 Limit function 1 relay 78 Limit function 2 relay 82 Logarithmic curve 60 Logbook 111 M Main display 30 Mark type of protection 20 meas key for function activation 28 Measured values display 111 148 Measurement 105 Measuring mode selection 50 Measuring mode change displays 105 Measuring mode 28 Memosens cable entry 7 Menu s
41. ceed variable and the temperature are displayed The device shows the calculated cell factor and zero point at 25 C Sensoface is active Use the arrow keysto Exit select HOLD is deactivated e MEAS exit after a short time e REPEAT Press enter to proceed Please refer to the Specifications for the nominal cell factor When measuring in a restricted space the individual cell factor must be determined 102 Zero Calibration in Air with Calibration Solution Action Remark Select Calibration Press enter to proceed Select CAL ZERO calibration method Press enter to proceed Ready for calibration Display 3 sec Hourglass blinks Now the device is in HOLD mode Calibration in air Edit digits until the lower display indicates Zero Calibration with solution Edit digits until the lower display indicates the solution value Press enter to proceed The device shows the cell factor at 25 C and the zero point Sensoface is active Use the arrow keys to Exit select HOLD is deacti e MEAS exit vated after a short e REPEAT time Press enter to proceed 103 Temp Probe Adjustment Action Select Calibration Press enter to proceed Select CAL RTD calibration method Press enter to proceed Measure the tempera ture of the process medium using an external thermometer Enter the measured temperature value Maximum difference 10 K Press enter to proceed Th
42. control panel 4 Span piece 4 x 5 Threaded sleeve 4 x Cutout 138x 138 mm DIN 43700 Fig Panel mount kit accessory ZU 0738 All dimensions in mm 18 Installation Installation Instructions Installation of the device must be carried out by trained experts in accordance with this user manual and as per applicable local and national codes Besure to observe the technical specifications and input ratings during installation Besure not to notch the conductor when stripping the insulation Before connecting the device to the power supply make sure that its voltage lies within the range 20 5 to 253 V AC DC The supplied current must be galvanically isolated If not connect an isolator module All parameters must be set by a system administrator prior to commissioning Terminals suitable for single wires flexible leads up to 2 5 mm AWG 14 Application in hazardous locations The approvals for Stratos Pro A4 B in Ex Zone 2 refer to the type of protection nA Memosens sensors are approved for type of protection nL Combining Stratos Pro A4 B analyzers with Memosens in Ex Zone 2 is therefore not permitted WS Se T For use in hazardous locations see separate Certificates document e ECEx e ATEX e FM e CSA e NEPSI e GOST 19 Rating Plates Terminal Assignments WARNING DO NOT SEPARATE WHEN ENERGIZED 22 21 20 19 18 17 16 15 14 8 E
43. ct parameter set using 4 gt press enter Select ALARM menu using 4 keys press enter All items of this menu group are indicated by the ALA code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter Exit Press meas key until the meas mode indicator is displayed enter D ALARM Delay Alarm Sensocheck Alarm CONTROL input For flow monitoring Max flow alarm For flow monitoring Min flow alarm Menu item Action Choices Delay Enter value using 4 v 0 600 SEC 4 keys 010 SEC Press enter to confirm Select Sensocheck ON OFF continuous monitoring of sensor Select ON or OFF using v keys Press enter to confirm At the same time Sensoface is activated With OFF Sensoface is also switched off Alarm Contact The alarm contact is closed during normal operation N C in It opens in the case of alarm or power outage 18 As a result a failure message is provided even in the case of line breakage fail safe behavior For contact ratings see Specifications Error messages can also be signaled by a 22 mA output current see Error messages and Configuration Output 1 Output 2 Operating behavior of the alarm contact see Operating States table The alarm delay time delays the color change of the display back lighting to red the 22 mA signal if configured and the alarm contact switching 75 Alarm Settings CON
44. e value Press enter to proceed Now the sample can be measured in the lab Action Remark The device returns to From the blinking measuring mode CAL mode indicator you see that product calibration has not been terminated Product calibration Display 3 sec step 2 Now the device is in When the sample value HOLD mode has been determined open the product cali bration once more The stored value is displayed blinking and can be overwritten with the lab value Press enter to proceed Display of new cell To repeat factor and zero point calibration Select based on 25 C REPEAT then enter Sensoface is active Press enter To end calibration Select MEAS then enter After calibration is After end of calibra ended the device will tion the outputs re switch to measuring main in HOLD mode mode for a short time 101 Calibration by Input of Cell Factor You can directly enter the value for the cell factor of a sensor This value must be known e g determined beforehand in the laboratory The selected process variable and the temperature are displayed This method is suitable for all process variables Display Action Remark Select Calibration Press enter to proceed Select CAL CELL calibration method Press enter to proceed Ready for calibration Display 3 sec Hourglass blinks Now the device is in HOLD mode Enter cell factor The selected process Press enter to pro
45. e corrected tempera ture value is displayed Sensoface is active To end calibration Select MEAS then enter To repeat calibration Select REPEAT then enter After calibration is ended the device will switch to measuring mode 104 Remark Wrong settings change the measurement properties Display 3 sec Now the device is in HOLD mode Display of actual temperature un compensated in the lower display After end of calibra tion the outputs re main in HOLD mode for a short time Measurement Remark From the configuration or calibration menus you can switch the device to measuring mode by pressing the meas key In the measuring mode the main display shows the configured process variable Cond or temperature the secondary display shows the time and the second configured process variable Cond or temperature The meas mode indicator lights and the active parameter set A B is indicated A B is not displayed with parameter set Fix A Pressing the enter key briefly shows the output currents By pressing the meas key you can step through the following dis plays When no key has been pressed for 60 sec the device returns to the standard display 1 Selecting the parameter set if set to manual in the configuration Select the desired parameter set using the 4 arrow keys PARSET A or PARSET B blinks in the lower display line Press enter to confirm Further
46. e tag number Being part of the device configuration the TAG can be read out via IrDA A standardized tag number helps for example to correctly re install a device after repair Menu item Action Choices Tag number Select character using A Z 0 9 lt gt v keys select next digit using GE 4 keys The first 10 characters are seen in the display with THE XOOOOOOOOUX Press enter to confirm out scrolling 95 96 Calibration Please note All calibration procedures must be performed by trained person nel Incorrectly set parameters may go unnoticed but change the measuring properties Calibration can be performed by Determining the cell factor with a known calibration solution taking account of the temperature e Input of cell factor e g for ultrapure water sensors e Sampling product calibration Zero calibration in air or with calibration solution e Temperature probe adjustment Please note 1 If measurements are performed in fittings with A 110 mm be sure to choose a calibration beaker with the same cross section and of the same material metal plastic Selecting a Calibration Mode Calibration adapts the device to the individual sensor characteristics Access to calibration can be protected with a passcode SERVICE menu First you open the calibration menu and select the calibration mode CAL SOL Calibration with calibration solution CAL CELL Calib
47. e you can view the displays for approx 60 sec You must configure the device for the respective measurement task 28 Selecting the Mode Entering Values To select the operating mode 1 Hold meas key depressed 2 s directly to measuring mode 2 Press menu key the selection menu appears 3 Select operating mode using left right arrow key 4 Press enter to confirm the selected mode Selection menu Selected mode blinks To enter a value 5 Select numeral left right arrow key 6 Change numeral up down arrow key 7 Confirm entry by pressing enter 29 Display in Measuring Mode 30 The MAIN DISPLAY is the display which is shown in measuring mode To call the measuring mode from any other mode hold the meas key depressed for at least 2 sec meas key enter key By pressing meas briefly you can step through further displays such as tag number TAG or flow L h These displays are turquoise After 60 sec they switch back to the main display Press enter to select a display as MAIN DISPLAY the secondary display shows MAIN DISPLAY NO Use the UP DOWN arrow keys to select MAIN DISPLAY YES and confirm by pressing enter The display color changes to white This display is now shown in measuring mode Color Coded User Interface The color coded user interface guarantees increased operating safety Operating modes are clearly signaled The normal measu
48. ee tn teen 26 bite p f ees 27 Signal Colors Display Backlighting u 27 Measuring MOLIS scan WR rH ai 28 Selecting the Mode Entering Values sss 29 Color Coded User Interface u 02000000000000000200000020000000000000000000 31 Operating Modes nn 32 Menu Structure of Modes and Functions 33 HOLD de EE 34 Alaf EE 35 Alarm and HOLD Messages seen 36 COP d E 38 Parameter Set AB een 40 Configuration Original for COD E 46 SEDIBO EE 50 Currenp Qutp t Ines 56 Current Output ee 66 Temperature ComMpensatiOn au u 68 CONTROL INput EE 72 Alarm Setting E 74 Limit Function EE 78 Pulse Length Pulse Frequency Controller 87 CO EE 88 WASH CITC EE 92 Time and Date sche ee 94 E ehn 94 Calibration a Or EO RE EE HU CH er E Dv read nado 97 Selecting a Calibration Mode 97 Calibration with Calibration Solution 98 Product Calibrati OM cken 100 Calibration by Input of Cell Factor 102 Zero Calibration in Air with Calibration Solution 103 Temp Probe Adjustment aussehen 104 Measurement 105 Re Take usa 108 IRIS ETT ETT 113 Operating KE E 118 Product Line and Accessories sene 120 SPECHICAUONS EPIIT TQ 121 Calibration Solutions 2 220000000000000000200000000002000000000 000000000 129 Concentration Measurement eee eene e eee tnn nana 131 Concent
49. ent Output signal for HOLD Output signal for HOLD mA FIX setting 21 0 mA LAST setting 21 4 HOLD active HOLD active Terminating the HOLD mode The HOLD mode is ended by switching to measuring mode hold meas key depressed The display reads Good Bye after that the HOLD mode is exited When the calibration mode is exited a confirmation prompt ensures that the installation is ready for operation e g sensor reinstalled located in process 34 Alarm External activation of HOLD SW A005 The HOLD mode can be activated from outside by sending a signal to the HOLD input e g from the process control system Power supply 12 24 V AC DC o input Stratos Pro A4 Process control system HOLD inactive 0 2 V AC DC HOLD active 10 30 V AC DC Manual activation of HOLD The HOLD mode can be activated manually from the HOLD menu This allows checking or replacing a sensor for example without provoking unintended reactions of outputs or contacts Press meas key to return to selection menu Alarm When an error has occurred Err xx is displayed immediately Only after expiry of a user defined delay time will the alarm be registered and entered in the logbook During an alarm the display blinks the display backlighting turns red Error messages can also be signaled by a 22 mA output current see Configuration The alarm contact is activated by alarm and power fail
50. et B the values are the same as in parameter set A Set B 49 Sensor Select Sensor type temperature probe cell factor transfer ratio measuring mode range 1 2 3 4 D ie e UI Press menu key Select CONF using 4 gt keys press enter Select parameter set using 4 gt press enter Select SENSOR menu using keys press enter All items of this menu group are indicated by the SNS code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter Exit Press meas key until the meas mode indicator is displayed Sensor type Ee Temperature probe E Cell factor Transfer ratio Measuring mode Measuring range Concentration determination Temperature unit Cleaning cycles Sterilization cycles Menu item Sensor type Action Select sensor type SE 670 using 4 e keys default setting Press enter to confirm Only with OTHER Select type of tempera ture probe using 4 v keys Press enter to confirm Enter cell factor using av 4 gt keys Press enter to confirm Enter transfer ratio using v 4 gt keys Press enter to confirm Select desired measuring mode using 4 v keys Press enter to confirm For cond measurement only Select desired range using 4 v keys Press enter to confirm Choices SE 670 SE 655 SE 656 SE 660 OTHER 1000PT 100PT 30 NTC 01 980 XX XXx 120 00 XXX Xx Cond
51. iley is always displayed for confirmation Please note The worsening of a Sensoface criterion leads to the devaluation of the Sensoface indicator Smiley becomes sad An improvement of the Sensoface indicator can only take place after calibration or removal of the sensor defect 141 Display Problem Status y Sensor defect amp Wrong or defective sensor or O excessive cable capacitance see also error message Err 15 l Temperature Temperature outside range for TC conc sal 142 FDA 21 CFR Part 11 Conformity with FDA 21 CFR Part 11 In their directive Title 21 Code of Federal Regulations 21 CFR Part 11 Electronic Records Electronic Signatures the American health agency FDA Food and Drug Administration regulates the production and processing of electronic documents for pharmaceutical devel opment and production This results in requirements for measuring devices used for corresponding applications The following features ensure that the measuring devices of this Series meet the demands of FDA 21 CFR Part 11 Electronic Signature Passcodes Access to the device functions is regulated and limited by individually adjustable codes Passcodes see SERVICE This prevents unauthor ized modification of device settings or manipulation of the measure ment results Appropriate use of these passcodes makes them suitable as electronic signature Audit Trail Every manual change of device settings can be a
52. inear or logarithmic An output filter can be pro grammed the fault current value can be specified See page 56 et seq Relay contacts 2 relay contacts for limit values Adjustable for the selected process variable hysteresis switching behavior MIN MAX limit con tact type N O N C and delay page 78 Alarm An alarm can be generated by Sensocheck flow monitoring or current failure page 74 Wash cleaning function This contact can be used for controlling a rinsing probe or for indicating the active parameter set page 92 Configurable as pulse length or pulse frequency controller page 88 10 Typical Application of Stratos Pro A4 MSCONDI ywr uN Gi gg 11 A ddns 19MOd uwul Xey SET Hulues gt wey DO Josuag JDM N Sd O uju xew wely yseMm O NJeA LUT s 9e U0D dwar KEISER indui WS SW vw Oc 7 p O eoejiajul SuasouJa Josuas GNOD IGNODSW vV Old SOjeHS LEVH m K ddns 1aMod WOO OJ UO Safety Information Be sure to read and observe the following safety instructions The device has been manufactured using state of the art technology and it complies with applicable safety regulations When operating the device certain conditions may nevertheless lead to danger for the operator or damage to the device See also the following documents page 3 Safety Inst
53. inter connected AC lt 250V lt 3A lt 750VA DC 30V 3A 90W N C or N O 0000 9999 s As desired within range User defined PID process controller Setpoint specification Neutral zone Proportional action Integral action Derivative action Controller type Pulse period Max pulse frequency Output via Rel1 Rel2 relay contacts see limit values Within selected range Max 50 96 of selected range Controller gain Kp 0010 9999 96 Reset time Tr 0000 9999 s 0000 s no integral action Rate time Td 0000 9999 s 0000 s no derivative action Pulse length controller or pulse frequency controller 0001 0600 s min ON time 0 5 s pulse length controller 0001 0180 min pulse frequency controller Real time clock Power reserve Different time and date formats selectable 5 days Display Main display Secondary display Text line Sensoface Mode indicators Alarm indication LC display 7 segment with icons colored backlighting Character height approx 22 mm unit symbols approx 14 mm Character height approx 10 mm 14 characters 14 segments 3 status indicators friendly neutral sad face meas cal conf diag Further icons for configuration and messages Display blinks red backlighting 125 Keypad Keys meas menu info 4 cursor keys enter I HART communica
54. ion by pressing enter Remark Display of measured variable Sensoface is active After end of calibra tion the outputs remain in HOLD mode for a short time After display of GOOD BYE the device automati cally returns to measuring mode 99 Product Calibration Calibration by sampling For product calibration the uncompensated conductivity mS cm S m is used During product calibration the sensor remains in the process The measurement process is only interrupted briefly Procedure 1 The sample is measured in the lab or directly on the site using a portable meter To ensure an exact calibration sample temperature and process temperature should be the same During sampling the device saves the currently measured value and then returns to measuring mode Then the calibration mode indicator blinks 2 In the second step you enter the measured sample value in the device From the difference between the stored measured value and entered sample value the device calculates the new cell factor If the sample is invalid you can take over the value stored during sampling In that case the old calibration values are stored Afterwards you can start a new product calibration Action Remark Select Calibration Press enter to proceed Select P CAL calibration method Press enter to proceed Ready for calibration Display 3 sec Hourglass blinks Now the device is in HOLD mode Take sample and sav
55. ion via external signal or flow measurement J UN I I UI e Press menu key Select CONF using 4 gt keys press enter Select parameter set using 4 gt press enter Select CNTR IN menu using 4 gt keys press enter All items of this menu group are indicated by the IN code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter Exit Press meas key until the meas mode indicator is displayed CONTROL input function ul PARSET FLOW FLOW ADJUST Menu item Action Choices Select function of Select using 4 keys PARSET CONTROL input Press enter to confirm selecting parameter set A B via signal at CONTROL input Flow for connecting a pulse output flow meter With Flow selected 12000 pulses liter you must adjust the device to the flow meter used Enter value using arrow keys Press enter to con firm In the alarm menu you can configure flow monitoring When you have set CONTROL to FLOW you can specify 2 additional limit values for maximum and minimum flow If the measured value lies outside this range an alarm message and a 22 mA error signal if configured will be generated Display Display Flow measurement in measuring mode Flow measurement sensor monitor 73 Alarm Settings Delay Sensocheck Wn on gt Press menu key Select CONF using 4 gt keys press enter Sele
56. isplay 17 Proceed using enter 18 Not used 19 Diagnostics 20 Configuration mode 21 Calibration mode 22 Measuring mode 23 Sensoface 24 Unit symbols Signal colors display backlighting Red Red blinking Orange Turquoise Diagnostics Green Info Purple Sensoface message Alarm in case of fault display values blink Input error illegal value or wrong passcode HOLD mode Calibration Configuration Service 27 Measuring Mode After the operating voltage has been connected the analyzer auto matically goes to Measuring mode To call the measuring mode from another operating mode e g Diagnostics Service Hold meas key depressed 2 s Sensoface indicator Active sensor status parameter set configuration Time or flow Display indicates OUT1 e g process variable Mode indicator measuring Hold meas key depressed for calling the measuring mode pressing once more switches the display Display indicates OUT2 e g temperature enter key Depending on the configuration one of the following displays can be set as standard display for the measuring mode see page 30 e Measured value time and temperature default setting e Measured value and selection of parameter set A B or flow Measured value and tag number TAG Time and date Output currents e Controller upper display controller output Y lower display setpoint Note By pressing the meas key in measuring mod
57. method The following temperature compensation methods are provided linear by entering a temperature coefficient natural waters nLF NaCl see page 68 Control inputs Input Input HOLD CONTROL Power supply l input The analog 0 4 20 mA current input can be used for external temperature compen sation TAN required See page 70 HOLD floating digital control input The HOLD input can be used for external activation of the HOLD mode see page 35 CONTROL floating digital control input The CONTROL input can be used either for parameter set selection A B or for flow monitoring see page 72 The Wash contact can be used for indicating the active parameter set see next page Current is provided through a universal power supply 24 230 V AC DC AC 45 65 Hz Options Additional functions can be activated by entering a TAN page 116 Signal outputs The device provides two current outputs for transmission of measured value and temperature for example The output curve is adjustable linear bilinear or logarithmic see page 56 et seq Relay contacts Four floating relay contacts are available Cole R1 PID controller Output 1 HART Output 2 Output 2 REL 1 REL 1 2 REL 2 Alarm Alarm Wash Wash Current outputs The floating current outputs 0 4 20 mA are used for transmitting measured values The output curve can be linear bil
58. n Front unit rear unit bag containing small parts e Specific test report Documentation cf p 3 e CD ROM Fig Assembling the enclosure 1 Jumper 3 x 2 Washer 1 x for conduit 6 Sealing insert 1 x 7 mounting Place washer 8 9 1 1 Rubber reducer 1 x Cable gland 3 x between enclosure and nut Filler plug 3 x 3 Cabletie 3 x 0 Hexagon nut 5 x 4 Hinge pin 1 x insertable 1 Sealing plug 2 x for sealing from either side in case of wall mounting 5 Enclosure screw 4 x 15 Mounting Plan Dimensions 148 1 Cable gland 3 x 2 Knockouts for cable gland or conduit 21 5 mm dia 2 knockouts Conduits not included 3 Knockout for pipe mounting 4 X 4 Knockout for wall mounting 2 x Fig Mounting plan All dimensions in mm 16 Pipe Mounting Protective Hood 40 060 EL 4 1 Hose clamp with worm gear drive to DIN 3017 2 x 2 Pipe mount plate 1 x 3 For vertical or horizontal posts or pipes 4 Self tapping screw 4 x Fig Pipe mount kit accessory ZU 0274 All dimensions in mm 199 185 Fig Protective hood for wall and pipe mounting accessory ZU 0737 All dimensions in mm 17 Panel Mounting 1 Circumferential sealing 1 x 2 Screws 4 x 3 Position of
59. nce Immunity to interference 126 EN 61326 1 General Requirements Class B residential area Industry EN 61326 2 3 Explosion protection Stratos Pro A4 B IECEx Ex nA II T4 Ex tD A22 IP5XT 85 C ATEX I 3 G Ex nA II T4 II 3 D Ex tD A22 IP54 T85 C FM C US NI I 2 ABCD TA S ILII 2 FG TA4 Type 4X C 1 2 Ex nA IIC T4 22 Ex tD T85 C Type 4X US 1 2 AEx nA IIC T4 22 AEx tD T85 C Type 4X CSA C US Class LL Div 2 GP A B C D E F G TA Type 4X C Ex nA II T4 DIP IL III 2 EFG Type 4X US AExnAIITA II I1 22 AEx tD 22 T85 C Type 4X NEPSI Ex nA II T4 DIP A22 TA T6 GOST 2ExnAllT4 DIP A22 TA 85 C I Power supply I 24 15 230 10 V AC DC 12 VA 4W AC 45 65 Hz Overvoltage category Il protection class Il I Nominal operating conditions Ambient temperature Transport Storage temperature Relative humidity Power supply Frequency for AC 20 455 C 30 70 C 10 95 not condensing 24 15 230 10 V AC DC DC lt 80V 45 65 Hz 127 Enclosure Mounting Color Ingress protection Flammability Dimensions Control panel cutout Weight Cable glands Connections User defined 2 1 count 4 DC lt 80V 128 Molded enclosure made of glass reinforced PBT PC Wall pipe post or panel mounting Gray RAL 7001 IP 67 NEMA 4X UL 94 V 0 148 mm x 148 mm x 117 mm 13
60. ntil the meas mode indicator is displayed Current range rer Process variable LIN biLIN LOG output Current start Current end Time averaging filter Output current during error message Output current during HOLD Output current for HOLD FIX for Cond only Menu item Current range U EI Ut RANGE Action Select 4 20 mA or 0 20 mA range using v keys Press enter to confirm Select using v keys Cond Conductivity TMP Temperature Press enter to confirm Then select characteristic LIN biLIN LOG Modify digit using a v keys select next digit using 4 keys Press enter to confirm Enter value using 4 v 4 keys Press enter to confirm Remark 4 20 mA 0 20 mA Selectable decades with logarithmic setting LOG S cm 0 001 mS cm 0 01 mS cm 0 1 mS cm 1 0 mS cm 10 0 mS cm 100 0 mS cm 1000 mS cm S M 0 001 S m 0 01 S m 0 1 S m 1 0 S m 10 0 S m 100 S m Entered value applies to selected process variable range If the adjusted range is exceeded the device automatically switches to the next higher range Autorange Entered value applies to selected process variable range If the adjusted range is exceeded the device automatically switches to the next higher range Autorange Assignment of measured values Current start and current end Example 1 Range 0 200 mS cm Example 2 Range 100 200 mS cm Advantage Higher resolution in range
61. o use the checkboxes on the rating plate See also Safety Information chapter O FM3037411 3037411C C US NI I 2 ABCD TA Ta 55 C C DIP II II 2 FG T4 Ta 55 C US SIILIII 2 FG TA Ta 55 C US c V2 Ex nA IIC T4 Ta 55 C 22 Ex tD T85 C V 2 AEx nA IIC T4 Ta 55 C 22 AEx tD T85 C APPROVED H Fig typeax Example of 20 lt Ta lt 55 C an additional CSA 08 2117513 US 1 11 1I 2 ABCDEFG T4 Ta 55 C c Ex nA II T4 Ta 55 C c DIP A22 II III 2 EFG US AExnAII T4 Ta 55 C US ILIII 22 AEx tD 22 T85 C approval plate The specifications refer to the respective device amp Type 4X 20 Power Supply Signal Lines Connect the power supply for Stratos Pro A4 MSCONDI to terminals 21 and 22 24 230 V AC 45 65 Hz 24 80V DC af 1 3V u 2 RS 485 A 3 RS 485 B 4 GND shield 5 input Areas for placing the 6 input ee ron 8 out 2 9 out 1 HART at 10 out 1 HART O peoocoocoocoogooocooo8g 11 hold er LC 12 hold control 13 control 14 REL 1 15 REL 1 2 16 REL 2 17 alarm 18 alarm 19 wash 20 wash 21 power po power Fig Terminals device opened back of front unit 21 Wiring Example SE 670
62. ommunicating via the IrDA interface CODES Assigning and editing passcodes DEFAULT Resetting the device to factory settings OPTION Enabling options via TAN Please note HOLD is active during Service mode Action Key Display Remark Activate a Press menu key to call the selection Service menu Select SERVICE using 4 gt keys press enter to confirm Passcode Enter passcode 5555 for service mode using the a 4 keys Press enter to confirm Display In service mode the following icons _ are displayed LAY HOLD triangle Service wrench Exit meas Exit by pressing meas 113 REL1 114 REL2 SE wt H S lt Menu item Displaying the currently measured values sensor monitor with HOLD mode activated Select MONITOR using 4 gt press enter to confirm Select variable in the bottom text line using 4 gt The selected parameter is shown in the main display As the device is in HOLD mode you can perform validations using simulators without influencing the signal outputs Press meas to return to the service menu Return to measurement Press meas once more Specifying the current at outputs 1 and 2 Select OUT1 or OUT2 using the gt keys press enter to confirm Enter a valid current value for the respective output using 4 v 4 gt keys Press enter to confirm For checking purposes the actual output current is shown in the bottom right corner of the display Exit by
63. onfirm Choices LIMITS CONTROLLER Selecting CONTROLLER leads to Controller menu group CTR Cond TMP PLC PFC 0 0600 SEC 0010 SEC 0 0180 P M 0060 P M pulses per minute Within meas range You can change the input range using the left right arrow keys 89 Controller For description see Controller Functions Neutral zone P I D actions behavior during HOLD Delay Press menu key Select CONF using 4 gt keys press enter Select parameter set using 4 gt press enter Select REL1 REL2 menu using 4 gt keys press enter All items of this menu group are indicated by the CTR code Press enter to select menu edit using arrow keys see next page Confirm and proceed by pressing enter Exit Press meas key until the meas mode indicator is displayed Select process variable Controller type Pulse length Use of relays Dem P Pulse frequency Setpoint Neutral zone Controller P action Controller action Controller D action Behavior during HOLD Menu item Neutral zone HOLD Action Adjust neutral zone using v 4 keys Press enter to confirm Adjust P action using v 4 gt keys Press enter to confirm Adjust action using 4 v 4 keys Press enter to confirm Adjust D action using v 4 keys Press enter to confirm Select response using v keys Press enter to confirm Choices 0 50 96 full scale You
64. ow keys see next page Confirm and proceed using enter Exit Press meas key until the meas mode indicator is displayed Select sensor type enter Select temp probe Ar Select cell factor Select transfer ratio Select measuring mode Select range Concentration determination Temperature unit Cleaning cycles Sterilization cycles Menu item Action Choices CIP SIP ISM only Cleaning cycles Select ON or OFF using ON OFF on off v keys Activates deactivates log ging in extended logbook Press enter to confirm Select ON or OFF using ON OFF on off v keys Activates deactivates log ging in extended logbook Press enter to confirm The cleaning and sterilization cycles are logged to measure the load on the sensor Suitable for biochemical applications process temp approx 0 50 C CIP temp gt 55 C SIP temp gt 115 C Please note A CIP or SIP cycle is only entered into the logbook 2 hours after the start to ensure that the cycle is complete 55 Current Output 1 Output current range Linear Logarithmic Current start Press menu key Select CONF using 4 gt keys press enter Select parameter set using 4 gt press enter Select OUT1 menu using 4 gt keys press enter All items of this menu group are indicated by the OT1 code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter Exit Press meas key u
65. p Sensor selection Select menu item D SE enter Menu item 1 enter Menu item 2 enter r Ai 5 enter Current output 1 OT1 L ont gt WER B r F Ai Current output 2 OT2 Lan gt pen D r F Ai Compensation COR Lon gt CORRECT ON Control input nnn n parameter set or flow IN UM Ue gt measurement B r ai Alarm mode ALA Lon F ALARM y gt r 4 Ai Relay outputs REL LON RELARELE gt gt a r 3 Ai Cleaning WSH LO A WAS H gt a 4 F Ai Setting the clock CLK LO nF CLOCK gt gt 4 r Ai Tag number TAG l L nf m 38 Parameter set A B configurable menu groups The device provides 2 parameter sets A and B By switching between the param eter sets you can adapt the device to different measurement situations for example Parameter set B only permits setting of process related parameters Parameter set A Parameter set B SENSOR Sensor selection OUT1 Current output 1 Current output 1 OUT2 Current output 2 Current output 2 CORRECTION Compensation Compensation CNTR IN Control input ALARM Alarm mode Alarm mode REL 1 REL 2 Relay outputs Relay outputs WASH Cleaning PARSET Parameter set selection CLOCK Setting the clock TAG Tag number External switchover of parameter sets A B You can switch between parameter sets A and B by applying a signal to the CONTROL input parameter setting CNTR I
66. ration CuUrVves eese eene nen tntnntntnatntnun 132 Error Handling E 138 Error MessadulBsS eden elite ear evan eren eire een enden 139 SENSOMACE ET M 141 PDA 21 CFR Part 17 WE 143 Electronic Signature Passcodes sss 143 E Tra tendu td ad teria ta erdt 143 Halo P AMNES 144 e TEE KS ubica iim nad a rera c did aiam 155 PasstOUBS Loves hm iin 156 Introduction Intended Use Stratos Pro A4 MSCONDI is a 4 wire device for measurement of electrical conductivity and temperature in liquids using electrodeless toroidal sensors Fields of application are biotechnology chemical industry environment food processing water wastewater treatment Enclosure and mounting possibilities The sturdy molded enclosure is rated IP 67 NEMA 4X for outdoor use It is made of glass reinforced PBT PC and measures 148 mm x 148 mm x 117 mm HxW x D It is provided with knockouts to allow e wall mounting with sealing plugs to seal the enclosure see page 15 post pipe mounting 40 60 mm L130 45 mm see page 17 e panel mounting 138 mm x 138 mm cutout to DIN 43700 see page 18 Weather protector accessory The weather protector which is available as accessory provides additional protection against direct weather exposure and mechanical damage see page 17 Connection of sensors cable glands For connecting the cables the enclosure provides e 3 knockouts for cable gland
67. ration by entry of cell factor P CAL Product calibration calibration with sampling CAL ZERO Zero calibration CAL RID Temperature probe adjustment 97 Calibration with Calibration Solution Input of temperature corrected value of calibration solution with simultaneous display of cell factor Be sure to use known calibration solutions and the respective temper ature corrected conductivity values see calibration solution tables in the appendix During the calibration procedure the temperature must be kept constant Please note When using an ARF 210 215 flow through fitting you should use the inluded calibration beakers identical dimensions and materials for calibration to prevent calibration errors 98 Action Select Calibration Press enter to proceed Select CAL SOL calibration method Press enter to proceed Ready for calibration Hourglass blinks Immerse sensor in calibration solution Enter the temperature corrected value of the calibration solution us ing the arrow keys see table in the appendix Press enter to confirm Remark Display 3 sec Now the device is in HOLD mode Lower line display of cell factor and temperature Action The cell factor and zero point are displayed The hourglass icon is blinking Use the arrow keys to select Repeat repeat calibration or e Measuring Press enter to confirm With MEAS selected End calibrat
68. riable Characteristic Overrange Output filter Measurement error Start end of scale Bilinear Vertex X Y Minimum span 0 4 20 mA max 10 V floating galv connected to output 1 Conductivity resistivity concentration salinity or temperature Linear bilinear or logarithmic 22 mA in the case of error messages PT filter time constant 0 120 s lt 0 25 current value 0 025 mA Configurable within selected range Configurable within selected range LIN 5 of selected range LOG 1 decade Alarm contact Contact ratings Contact response Response delay Relay contact floating AC lt 250V lt 3A lt 750VA DC 30V 3A 90W N C fail safe type 0000 0600 s Wash contact Contact ratings Contact response Interval Cleaning duration or Parameter set A B Contact ratings Contact response 124 Relay contact floating For controlling a cleaning system AC 250V 3A 750VA DC 30V 3A 90W N C or N O 000 0 999 9 h 000 0 h cleaning function switched off 0000 1999 s For signaling parameter set A B AC 250V 3A 750VA DC 30V 3A 90W Contact open Parameter set A active Contact closed Parameter set B active Limit values Rel1 Rel2 Contact ratings Contact response Response delay Setpoints Hysteresis Rel1 Rel2 contacts floating but
69. ring mode is white Information text appears on a green screen and the diagnostic menu appears on turquoise The orange HOLD mode e g during calibration is quickly visible as is the magenta screen which indicates asset management messages for predictive diagnostics such as maintenance request pre alarm and sensor wear The alarm status has a particularly noticeable red display color and is also signaled by flashing display values Invalid inputs or false pass codes cause the entire display to blink red so that operating errors are noticeably reduced White Red blinking Orange Alarm errors Magenta Turquoise Green Maintenance request Diagnostics Information texts 31 Operating Modes Diagnostics Display of calibration data display of sensor data performing a device self test viewing the logbook entries display of hardware software ver sions of the individual components The logbook can store 100 events 00 99 They can be displayed directly on the device The logbook can be extended to 200 entries using a TAN Option HOLD Manual activation of HOLD mode e g for servicing The signal outputs adopt a defined state Calibration Every sensor has typical characteristic values Calibration is required to supply a correct measured value The device checks which value the sensor delivers when measuring in a known solution When there is a deviation the device can be adjusted
70. rs to EN 27888 NaCl from 0 to 26 by wt 0 120 C Press enter to confirm With linear compensation only Step 1 Enter temperature compensation of the process medium Step 2 Enter reference temperature Enter value using 4 e 4 keys Press enter to confirm Permissible range 0 199 9 C Choices 00 00 19 99 K 69 Temperature Compensation Current input for temp measurement Press menu key Select CONF using 4 gt keys press enter Select parameter set using 4 gt press enter Select CORRECTION menu using 4 gt keys press enter All items of this menu group are indicated by the COR code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter Exit Press meas key until the meas mode indicator is displayed Temperature compensation SCH Current input external temp measurement if enabled via TAN Current start Current end Menu item Action With external temp measurement current input enabled TAN Current range Select desired range using 4 v keys Press enter to confirm Modify digit using 4 v keys select next digit using 4 keys Press enter to confirm Enter value using 4 v 4 keys Press enter to confirm Choices 4 20 mA 0 20 mA Input range 50 250 C 58 482 F Input range 50 250 C 58 482 F 71 CONTROL Input Parameter set select
71. ructions e Certificates CAUTION Commissioning must only be performed by trained personnel autho rized by the operating company Whenever it is likely that protection has been impaired the device shall be made inoperative and secured against unintended operation The protection is likely to be impaired if for example e the device shows visible damage e the device fails to perform the intended measurements e after prolonged storage at temperatures below 30 C or above 70 C e after severe transport stresses Before recommissioning the device a professional routine test must be performed This test must be carried out at the manufacturer s factory Please note Before commissioning you must prove that the device may be connected with other equipment 12 Safety Precautions for Installation The electrical installation shall conform to the national regulations for electrical installations and or other applicable national or local codes or regulations The power supply shall be disconnectable from the device by a two poled circuit breaker Switch and circuit breaker shall be located in close proximity to the equipment and be easily accessible by the OPERATOR They shall be marked as disconnect switch for the device Besure to disconnect the mains supply and any relay contacts which are connected to separate current sources before starting maintenance operations Approvals for application in hazardous loca
72. s variable and Corner Y output current Press enter to confirm see figure below Vertex of bilinear curve Output current Example mA Current range 4 20 mA 20 Current start 0 uS cm Current end 200 uS cm Vertex CORNER X 10 uS cm process variable 124 CORNER Y 12 mA output current Result The output current change in the range 0 10 uS cm is much greater than in the range 10 200 uS cm Process variable 0 10 200 uS cm 59 Logarithmic Curve Nonlinear output current characteristic allows measurements over several decades e g measuring very low values with a high resolution and high values with a low resolution Parameters required Start and end value Possible start and end values The start value must be at least one decade lower than the end value Start value and end value must be specified in the same units either in uS cm or in S m see listing 0 001 mS cm 0 001 S m 0 01 mS cm 0 01 S m 0 1 mS m 0 1 S m 1 0 S m 10 0 S m 100 S m The start value is the next decade value below the lowest measured value The end value is the next decade value above the highest measured value The number of decades results from Number of decades log end value log start value The output current value is defined as follows log measured value log start value uput eunen 16 mA Number of decades 4 mA 2 800 mS cm Measured j lt gt values A 10 1
73. s M20x1 5 2 knockouts for NPT 1 2 or rigid metallic conduit For quasi stationary installations with Memosens sensors we recommend using the M12 device socket accessory ZU 0822 instead of a cable gland it allows simple replacement of the sensor cable without opening the device Sensors The device has been designed for application with the Model 670 electrodeless sensor Knick Display Plain text messages in a large backlit LC display allow intuitive operation You can specify which values are to be displayed in standard measuring mode Main Display see page 30 Color coded user interface The colored display backlighting signals different operating states e g alarm red HOLD mode orange see page 31 Diagnostics functions Sensocheck automatically monitors sensor and cables and the Sensoface function clearly indicates the sensor condition see page 108 Data logger The internal logbook TAN SW A002 can handle up to 100 entries up to 200 with AuditTrail TAN SW A003 see page 111 2 parameter sets A B The device provides two parameter sets which can be switched manu ally or via a control input for different process adaptations or different process conditions For overview of parameter sets table for copy see page 46 Password protection Password protection passcodes for granting access rights during operation can be configured see page 116 TC process medium Selecting the compensation
74. sis HYSTERESIS Limit 2 delay DELAYTIME Limit Lo Hysteresis Action Choices Select desired process Cond TMP variable using w keys Press enter to confirm Select desired function Lo LEVL Hi LEVL using 4 vw keys Press enter to confirm Limit 2 icon a N O normally open N O N C contact N C normally closed contact Select using 4 w keys Press enter to confirm Enter setpoint using v Within meas range 4 keys Press enter to confirm Select hysteresis using 0 50 96 full scale v 4 keys Press enter to confirm The contact is activated 0 9999 SEC with delay deactivated 0010 SEC without delay Adjust delay using 4 v 4 keys Press enter to confirm Limit Hi Setpoint 1 Contact Contact 0 Setpoint Hysteresis 84 Typical applications P controller Application for integrating control systems e g closed tank batch processes PI controller Application for non integrating control systems e g drains PID controller The additional derivative action compensates for measurement peaks Controller characteristic 100 96 Controller output Yp 96 Neutral zone Yp 0 Deviation Xw mS cm Setpoint 100 96 85 Controller equations 1 dy Controller outputY Yp f Ydt Tp E P action action D action with Yp Proportional action Proportional action Yp TN Reset time s Tp Rate time s Yp Setpoint
75. tector 17 Wiring 21 Wiring example SE 670 via RS 485 22 Z Zero calibration 103 152 Index 153 154 Trademarks The following names are registered trademarks For practical reasons they are shown without trademark symbol in this manual Stratos Sensocheck Sensoface Memosens is a registered trademark of Endress Hauser Conducta GmbH and Knick Elektronische Messger te GmbH amp Co KG HART is a registered trademark of the HART Communication Foundation 155 Passcodes In the SERVICE CODES menu you can assign passcodes to protect the access to certain functions Mode of operation Passcode Service SERVICE 5555 Diagnostics DIAG HOLD mode Calibration CAL Configuration CONF Knick Elektronische Messger te GmbH amp Co KG P O Box 37 04 15 D 14134 Berlin Phone 49 0 30 801 91 0 Fax 49 0 30 801 91 200 Internet http www knick de knick knick de TA 212 035 MS KNEO2 20110301 Software version 2 x
76. tion Conditions I HART version 6 Digital communication by FSK modulation of output current 1 Device identification measured values status and messages parameter setting calibration records Output current gt 3 8 mA and load resistance gt 250 Q IrDA interface Infrared interface for firmware update FDA 21 CFR Part 11 Access control by editable passcodes Logbook entry and flag via HART in the case of configuration changes Message and logbook entry when enclosure is opened Diagnostics functions Calibration data Device self test Logbook Extended logbook TAN Calibration date sensor parameters Display test automatic memory test RAM FLASH EEPROM module test 100 events with date and time Audit Trail 200 events with date and time I Service functions Sensor monitor Current source Relay test Manual controller IrDA Passcodes Factory setting TAN Display of direct sensor signals Current specifiable for output 1 and 2 00 00 22 00 mA Manual control of the four switching contacts Controller output entered directly start of control process Activating the IrDA function Assigning passcodes for menu access Resetting all parameters to factory setting Enabling optionally available additional functions Data retention Parameters calibration data logbook gt 10 years EEPROM EMC Emitted interfere
77. tions Stratos Pro A4 B MSCONDI see Specifications copies provided in the Certificates document IECEx ATEX FM CSA NEPSI and GOST Terminals Screw terminal suitable for single wires flexible leads up to 2 5 mm AWG 14 Recommended torque for the terminal screws 0 5 0 6 Nm Important Notice The operator must indicate the type of protection When the device provides different types of protection the operator must specify the applied type of protection during installation To do so use the checkboxes on the rating plate i Knick SC91214A KEM 08 0020 ance gt YY ONS Exna Ta EI 14 Vi JII3D ExtDA221P54 T85 C tD A22 IP54 T85 C No 12345 1234567 1006 Electrical data see Knick SC 91214A__ see Ctrl dwg 212 002 100 20 lt Ta lt 55 C PWR 24 15 to 230 10 V AC 45 to 65 Hz lt 12 VA Enclosure Type 4X 24 15 to 80 10 V DC 4W i c Ge anre Additional rating plate at outside bottom of front with check boxes for marking the respective application after installation 13 Overview Overview of Stratos Pro A4 MSCONDI 43V Ch 1 Output 1 HART RS 485 A Output 1 HART RS 485 B Se 2 SI t Output 2 GND Shield Output 2 REL 1 REL 1 2 REL 2 Input Alarm Input 18 Alarm Wash HOLD Wash HOLD Power CONTROL CONTROL Power 14 Assembly Package Contents Check the shipment for transport damage and completeness The package should contai
78. tructure 33 Menu structure of configuration 38 Message via CONTROL input 37 Misuse 12 Module test 110 Mounting plan 16 O Operating mode selection 29 Operating modes 32 Operating states 118 Option request 116 Option request requirements 112 Options overview TAN options 120 Options release 116 Order code 120 Output filter 62 Output signal during HOLD 34 Overview device properties 7 Overview parameter sets 47 Overview terminal assignments 14 P Package contents complete 15 Package contents documentation 3 Panel mounting 18 Parameter errors 138 Parameter set A B introduction 39 Parameter set configure external switchover 72 Parameter set display 105 Parameter set display via WASH contact 40 Parameter set external switchover 39 Parameter set individual configuration data 47 Parameter set switch over manually 40 149 Passcode configure 116 Passcode explanation 143 Passcodes 156 PFC pulse frequency controller 87 Pin assignments 20 Pipe mounting 17 PLC pulse length controller 87 Point of measurement TAG 95 Power supply 21 Power supply mains values 9 Product calibration 100 Product line 120 Protective hood 17 Protective wiring 24 Pulse frequency controller PFC 87 Pulse length controller PLC 87 R RAM test 110 Rating plates 20 Relay 1 78 Relay 2 82 Relay test 114 Release options 116 Reset to factory settings 116 Return of products under warranty 2 S Safet
79. turer see back of this manual MEAS MODE Cond MEAS RANGE Conc Solution TEMP UNIT CIP COUNT SIP COUNT Cond Conc 96 Sal o X Xxx mS cm Xx xx mS cm Xxx x mS cm X xxx S m xx xx S M 01 NaCl 02 HCI NaOH H2SO4 Cond x xxx mS cm 01 NaCl 03 04 05 H2S04 HCI HNO3 09 H2S04 10 NaOH C F eC ON OFF OFF ON OFF OFF 41 OTI Configuration Output 1 OUT1 RANGE CHANNEL Output with Cond only LIN BEGIN 0 4mA END 20 mA BiLIN BEGIN 0 4 mA END 20 mA CORNER X CORNER Y LOG BEGIN 0 4mA END 20 mA TMP BEGIN 0 4mA S END 20 mA TMP BEGIN 0 4mA F END 20 mA FILTERTIME 22mA FAIL HOLD MODE FIX HOLD FIX 42 Choices 0 20 mA 4 20 mA Cond TMP LIN BiLIN LOG XXXX XXXX Default 4 20 mA Cond LIN 000 0 mS cm 100 0 mS cm Input range selected CHANNEL Vertex X BEGIN x CORNER X x END rising BEGIN 2 CORNER X END falling Input range selected CHANNEL Default 12 mA Vertex Y 0 4 mA lt CORNER Y lt 20 mA Decades Decades 50 250 C 50 250 C 58 482 F 58 482 F 0 120 SEC ON OFF LAST FIX 0 4 22 mA 0000 SEC OFF LAST 021 0 mA Configuration Choices Default Output 2 OUT2 OT2 RANGE 0 20 mA 4 20mA 4 20 mA CHANNEL Cond TMP TMP other steps like output 1 Temperature compensation CORRECTION COR TC SELECT OFF OFF LIN NLF NaCl LIN TC LIQUID 00
80. ure see also Configuration Alarm Settings 2 sec after the failure event is corrected the alarm status will be deleted 35 Alarm and HOLD Messages Message Released by Cause Alarm Sensocheck Polarization Cable 22 mA Error Messages Flow CONTROL input Alarm ERR 10 Conductance gt 3500 mS contact opens HOLD HOLD HOLD via menu or input Last Fix CONF Configuration CAL Calibration SERVICE Service 36 Generating a message via the CONTROL input min flow max flow The CONTROL input can be used for parameter set selection or for flow measurement pulse principle depending on its assignment in the Configuration menu When preset to flow measurement CONF CNTR IN CONTROL FLOW an alarm can be generated when the measured flow exceeds a speci fied range CONF ALA FLOW CNTR ON CONF ALA FLOW min specify value default 5 liters h CONF ALA FLOW max specify value default 25 liters h Power supply 12 24 V AC DC o CONTROL I 12 input Stratos Pro AA 37 Configuration The configuration steps are assigned to different menu groups With the left right arrow keys you can jump between the individual menu groups Each menu group contains menu items for setting the parameters Pressing enter opens a menu item Use the arrow keys to edit a value Press enter to confirm save the ads Return to measurement Hold meas key depressed 2 s grou
81. urrent These parameters cannot be adjusted in parameter set B the values are the same as in parameter set A 47 Parameter Set A OT2 Current range OT2 Process variable OT2 Lin bilin log output OT2 Current start OT2 Current end OT2 Vertex X bilinear curve only OT2 Vertex Y bilinear curve only OT2 Filter time OT2 22 mA error current OT2 HOLD mode OT2 HOLD FIX current COR TC SELECT COR Temp coefficient COR Reference temperature COR Current range COR Current start COR Current end IN Parameter set A B or flow IN Flow meter Adjusting pulses liter ALA Delay ALA Sensocheck on off ALA Flow control FLOW CNTR on off ALA Minimum flow hysteresis fixed at 5 96 ALA Maximum flow hysteresis fixed at 5 96 REL Usage RL1 Process variable RL1 Function RL1 Contact response RL1 Setpoint RL1 Hysteresis RL1 Delay 48 Set B Parameter Set A RL2 RL2 RL2 RL2 RL2 RL2 CTR CTR CTR CTR CTR CTR CTR CTR CTR CTR CTR CTR Process variable Function Contact response Setpoint Hysteresis Delay Process variable Controller type Pulse length Pulse frequency Setpoint Neutral zone P gain I time D time HOLD mode Delay Contact response WSH Contact function WSH Wash cycle WSH Wash duration WSH Contact response CLK Time amp Date TAG Tag number These parameters cannot be adjusted in parameter s
82. utomatically docu mented Each change is tagged with a Configuration Change Flag which can be interrogated and documented using HART communica tion Altered device settings or parameters can also be retrieved and documented using HART communication Extended logbook Audit Trail also records function activations CAL CONFIG SERVICE some Sensoface messages cal timer wear and opening of the enclosure 143 Index 21 mA output signal in HOLD mode 34 22 mA output signal in case of fault 140 A A is shown on the display 28 Access codes for operating modes 143 Access codes setup 116 Accessories 120 Activate rinsing probes schematic 10 Activate rinsing probes configuration 92 Alarm and HOLD messages 36 Alarm contact properties 75 Alarm current failure 80 f Alarm fault alarm 138 Alarm flow monitoring 76 Alarm operating status 35 Alarm Sensocheck 74 f Ambulance TAN in event of loss of passcode 116 Application in hazardous locations 19 Application mark rating plate 13 Application rating plates 20 Approvals explosion protection rating plate 13 Approvals explosion protection specifications 127 Assembly enclosure 15 Assembly options 7 Assembly panel mounting 18 Assembly pipe mounting 17 Assembly wall mounting 16 Audit trail 143 Autorange 57 B Backlighting 27 C Calibration data 109 Calibration errors 138 Calibration solutions 129 Calibration calibration solutions 98
83. y contact of pulse frequency controller ON time Pulse frequency Y 20 Pulse frequency Y 80 87 Controller For description see Controller Functions Process variable Controller type Setpoint J UN I I UI e Press menu key Select CONF using 4 gt keys press enter Select parameter set using 4 gt press enter Select REL1 REL2 menu using 4 gt keys press enter All items of this menu group are indicated by the CTR code Press enter to select menu edit using arrow keys see next page Confirm and proceed using enter Exit Press meas key until the meas mode indicator is displayed Select process variable Controller type Pulse length Use of relays p 2 Pulse frequency Setpoint Neutral zone Controller P action Controller action Controller D action Behavior during HOLD Menu item Use of relays Select process variable Action Select in the text line using 4 keys e CONTROLLER Press enter to confirm Select desired process variable using 4 w keys Press enter to confirm Pulse length controller PLC or pulse frequency controller PFC Select using v keys Press enter to confirm Only with PLC Pulse length Adjust using 4 gt keys Press enter to confirm Only with PFC Pulse frequency Adjust using 4 gt keys Press enter to confirm Adjust setpoint using v 4 keys Press enter to c
84. y information 12 Safety information separate booklet 3 Selection menu display menu item 29 Sensocheck activate 74 Sensocheck meaning of icons 141 Sensoface cause of error message 138 Sensoface meaning of icons 141 Sensor connection example 22 Sensor monitor display the currently measured values 111 Sensor monitor in active HOLD mode 114 Sensor type selection 50 150 Serial number display 112 Service 32 Service mode 113 Service controller test 115 Service factory settings 116 Service IrDA communication 115 Service passcode lost 116 Service passcodes 116 Service relay test 114 Service release options 116 Service sensor monitor 114 Service specify current outputs 114 Signal colors 27 Signal colors user prompting 31 Signal lines 21 SIP sterilization cycles 55 Software version display 112 Specifications 121 Standard display settings 30 Start up 12 T TAG 95 TAN options release 116 TAN options required codes 120 Technical data 121 Temp probe adjustment 104 Temperature compensation 69 Temperature detector adjustment 104 Temperature detector select type 50 Temperature unit select 52 Terminal assignments 20 Terminals properties 13 Time and date 95 Time averaging filter 63 Time display 105 Trademarks 155 Type of protection marking 20 151 U User interface 26 W Wall mounting 16 Warranty 2 WASH contact configuration 92 WASH contact signal parameter set 40 Weather pro
85. z r1 Go 91 81 7 6 P1 21 31 2 E10 o gt o a SA D WHICL YE GN BN 228 AA OH 6 AAA g 55 B8 m a ca W aso sr z e or E Tt Ze aa D gt N e E x z e lt 20 4 dod z Tz So So Zo D us lt ul wu o o e on on or rs q ec oc o O F Out LPower L contacts L Digital In 4 LO LOut24 L un J L Rs 485 l DO NOT REMOVE OR REPLACE FUSE WHEN ENERGIZED Fig Terminal assignments of Stratos Pro A4 Knick gt CONDI A401N CONDI O No 76739 0000000 0910 20 lt Ta S 55 C Il Berlin Made in German Power 24 15 to 230 10 V AC 5 to 65 Hz 12 VA 24 15 to 80 10 V DC 4W Ag CE Fig Stratos Pro A4 N rating plate at outside bottom of front Knick SC91214A LJW3G ExnAllT4 Knick gt A4 B No 12345 1234567 1006 20 Ss Ta S 55 C Enclosure Type 4X YY 13D Ex tD A22 IP54 T85 C Electrical data see Knick SC 91214A PWR 24 15 to 230 10 V AC 45 to 65 Hz lt 12 VA a a 15 to 80 10 V DC 4W APPROVED Se G Cg IECEx KEM 08 0020 O Ex nAIT4 C Ex tD A22 IP54 T85 C see Ctrl dwg 212 002 100 AgCE GE Stratos Pro ME rating plate at outside bottom of front Please note The operator must indicate the type of protection When the device provides different types of protection the operator must specify the applied type of protection during installation To do s

Download Pdf Manuals

image

Related Search

Related Contents

Notice de montage et d`installation du chariot Roll`in  La garantie Brabantia  Gebrauchsanweisung Operating Instructions  LCD Monitor Series  DS7400XiV4-FRA - Bosch Security Systems  How to Use MySciLEARN  Sonifier® Cell Disruptor Model SLPe User's Manual  CyberPower UP425UIT User's Manual      

Copyright © All rights reserved.
Failed to retrieve file