SIA, SIB, SIC, SID User's Manual
Contents
1. Current 4 to 20mA and 0 to 1mAdc Voltage 0 to 1Vdc and 0 to 5Vdc Response speed 0 5s 90 response Input impedance 4 to 20mA input 102 0 to 1mAdc input 200Q 1 to 5Vdc 1 to 5Vdc and 0 to 5Vdc 1M Input adjustment range 10 FS for both ZERO and SPAN Indication Indication method Red green or yellow LED bar dots color selectable Flashing display SIA and SIB only when input is complete Number of bar dots 101 Effective scale length 100mm Indication range 0 to 100 FS Indication accuracy 1 FS 1 digit Alarm setting unit SIC and SID Setting range High low High limit value H 100 FS to low limit value 1 FS limit Low limit value L high limit value 1 FS to 0 High high High high limit value HH 100 FS to high limit value 1 FS limit low low High limit value H high high limit value 1 FS to low limit value limit 1 FS Low limit value L high limit value 1 FS to low low limit value 1 FS Low low limit value LL low limit value 1 FS to 0 Setting accuracy 1 FS 1 digit Output and rating Non voltage contact SIC 2C SID 2c 2a 125Vac 0 5A and 30Vdc 2A resistive load Contact life 500 000 cycle min at rated load Power failure compensation Settings are safely stored in nonvolatile RAM Ambient temperature 0 to 50 C 0 to 45 C General
2. No CP UM 1221E SIA SIB SIC SID Smart Indicator with LED Bar Graph User s Manual Thank you for purchasing the SIA SIB SIC and SID This manual contains information for ensuring the correct use of the SIA SIB SIC and SID It also provides necessary information for installation mainte nance and troubleshooting This manual should be read by those who design and maintain equipment that uses the SIA SIB SIC and SID Be sure to keep this manual nearby for handy reference Yamatake Corporation RESTRICTIONS ON USE This product has been designed developed and manufactured for general purpose application in machinery and equipment Accordingly when used in applications outlined below special care should be taken to implement a fail safe and or redundant design concept as well as a periodic maintenance program e Safety devices for plant worker protection e Start stop control devices for transportation and material handling machines e Aeronautical aerospace machines e Control devices for nuclear reactors Never use this product in applications where human safety may be put at risk NOTICE Be sure that the user receives this manual before the product is used Copying or duplicating this user s manual in part or in whole is forbid den The information and specifications in this manual are subject to change without notice Considerable effort has been made to ensure that this manual is free fr
3. VI Basic Model Power Supply Input Number Voltage No 1 Display Colour No 1 Display Range No 2 Display Colour No 2 Display Range Additional Processing SIA 1 display SIB 2 displays SIC 1 display High low limit alarm SID 1 display High high limit low low limit alarm 100V 110Vac 50 60Hz 200V 220Vac 50 60Hz 120Vac 50 60Hz 40Vac 50 60Hz 0 1 N 24Vdc 4 to 20mAdc 0 to 1mAdc 0 to 1Vdc 1 to 5Vdc lt I lt Ir oO 0 to 5Vdc R Red G Green Y Yellow See Range Code select Table R Red G Green Y Yellow See Range Code select Table gt Range Code select Table Selection None Inspection certificate provided Tropicalization treatment wl 0 Tropicalization treatment Inspection certificate provided lt Complying with the traceability certification Description U S T V WY Unit Model Z ENN Mantissa Model ppmhPaj Pa kPa MPa Index Model 8 8 4 0 penprarageamra Tt poo me E eer e taty 5 0 Range limits Example tll VI d 1 SIA 8 C Fi AgaX T Range 300 ele a A Unit es 2 3 x10 2 Index X 0 low limit When the range is low limit value to high limit value However the a
4. specifications Power consumption AC power 5VA AC power 5VA 24Vdc power 2W_ 24Vdc power 4W Insulation resistance 50MQ min by DC500V megger between circuit and case 10MQ min by DC500V megger between circuit and case Dielectric strength 1500Vac for 1 min between circuit and case AC power 1500Vac for 1 min between power and input terminal 24Vdc power 1500Vac for 1 min between power and input terminal between output contact and case between output contact and power and between output contact and circuit 2000Vac for 1 min between circuit and case AC power 1500Vac for 1 min between power and input terminal 24Vdc power 500Vac for 1 min between power and input terminal 500Vac for 1 min between input No 1 and input No 2 Mass weight Approx 500g Approx 590g Approx 400g Vibration resistance 4 9m s max 10 to 60Hz Shock resistance 490m s max Case and cover materials Case ABS resin Cover acrylic resin Nameplate ABS resin Standard color Scale black colored aluminum Bezel and case moss green Munsell 2 5GY3 1 Mounting Flush mounted on indoor panel Mounting angle 0 to 90 in any direction Standard accessories Mounting spacers 2 units Additional treatment Selectable With inspection certificate and or tropicalization SI Series Model Selection Guide Mod 0 0 m M M M IV V
5. the power high voltage and load lines e The above conditions also apply to internal panel wiring 2 Terminal connection 1 Use solderless terminals that take M3 screws 2 The instrument is not equipped with a power switch Mount one separately if necessary 3 Connect wiring according to the terminal connection diagram Make sure that the cables have been connected correctly Terminal connection diagram SIA type SIB type SIC type SID type ee ME i Input Input No 1 _ Input Input Input No 2 is pa nput No i i r Ha H side output HHa HHside H side output aie i Hb output Power supply Power supply ia 24Vde 24Vde Ha Power supply Power supply a 100V 110V 100V 110V oe LU La LLa 200V 220V KAT 200V 220V i i 120V 50 60Hz 120V AC 50 60Hz Lb L side output Llp LLside L side output 240V 240V i i output cC Power supply 24Vdc joovitiov Power supply 24Vde 200v 220V 100W 110V 120V AC 50 60Hz 200V 220V 240V 20V AC 50 60Hz 240V Polarilty of 24Vdc power supply Polarilty of 24Vdc power supply 4 ALARM SETTING AND ZERO SPAN ADJUSTMENT 1 Location For SIA and SIB types The zero and span control knobs are accessible by removing the nameplates acrylic cover and scale plate Adjust the knobs according to the instructions in Section 4 3 For SIC and SID types First remove the nameplates and acrylic cover Set the alarm and adjust the zero and span knobs accordin
6. bsolute 2 SIB 8 C RA32X high and low limit values must be the same am S Unit 5 ae 0 Index W limit azbil 5 x10 Yamatake Corporation Advanced Automation Company 1 12 2 Kawana Fujisawa Kanagawa 251 8522 Japan URL http www azbil com Printed on recycled paper 07 Specifications are subject to change without notice Printed in Japan 1st Edition Issued in July 1998 7th Edition Issued in May 2007 M
7. econds 4 The PV indicating value can be erased by selecting the PV indicating value clear mode 2 3 Setting of high limit H and high high limit HH 1 Select either the high limit H or high high limit HH mode by pushing M The LED should flash 2 Set either a high or high high limit value by moving the setting bar up A or down V 3 Load a set point by pushing SET The LED should go out 3 2 4 Setting of low limit L and low low limit LL 1 Select either the low limit L or low low limit LL mode by pushing M The LED should flash 2 Set either a low or low low limit value by moving the setting bar up A or down V 3 Load a set point by pushing SET The LED should go out 2 5 About the hysteresis of the alarm contacts The hysteresis of the SIC and SID is different from the operating differential of a general temperature controller It goes through the process described below For a high limit alarm 1 When the set point the PV value the relay contact turns ON 2 In the next instant the internal set point decreases 5 FS 3 After 100ms the internal set point increases 0 5 FS from the above 2 state and comparison is made between the internal set point and the PV value If the internal set point gt PV value the contact turns OFF 4 Afterwards at 100ms intervals the internal set point increases every 0 5 up to the value of internal set point 1 0 FS and each time the inte
8. g spacers to the right and left sides of the instrument body 2 Insert the instrument body into the panel cutout 3 Mounting screws are located at the top and bottom behind the two nameplates With a phillips head screwdriver turn the screws clockwise to raise the fixing bracket the secure the instrument body in place 4 To remove the instrument from the panel simply turn the screws counterclockwise 5 Choose installation Choose a location that meets the following requirements Operating conditions Ambient temperature SIA SIB 0 to 50 C SIC SID 0 to 45 C Ambient humidity Vibration resistance Shock resistance 80 RH at 40 C 4 9m s 10 to 60Hz 490m s Cautions e Do not mount the instrument in an environment where these is flammable gas e Do not use the instrument in a highly corrosive atmosphere e Distance the instrument from any potential source of electrical interference large capacity electromagnetic switches high frequency generating units SCR units or the like 3 WIRING 1 Cables 1 For the input signal use a shielded cable suitable for measuring instrument use 2 Use a 600V vinyl insulated cable IV cable conforming to JIS C3307 for the 100 110Vac power supply Cautions e Leave space of more than 50cm between the input signal cable and the power line e Do not run the input signal cable and the power line in the same conduit or in close proximity e Distance the input signal cable from
9. g to the instructions in Section 4 3 SlAandSIB types SICandSID types Span control knob Acrylic Nameplate cover Scale plate Sy WEST TETTTry Sy Zero control knob Alarm setting zero span adjustment Fig 7 2 Alarm setting SIC and SID 2 1 Setting indicating Set point operation button High high limit set point Flashes during high high limit setting Lights up during relay action High limit set point Flashes during high limit setting Lights up during relay action Light up during PV input span adjustment Light up during PV indicating value clear erase Low limit set point Light up during PV input zero adjustment Flashes during low limit setting Lights up during relay action Low low limit set point Flashes during low low limit setting Lights up during relay action Moves set point up PV indicating value Mode selector button Modes change in direction of arrow when button is pushed Moves set point down Fig 8 2 2 Selection of setting modes 1 When the M mode selector button is pushed all LEDs either flash or light up and a setting mode can be selected 2 LL L H and HH LEDs flash during mode selection and light up during relay action 3 When the A or W button is pushed the setting mode shifts upward or downward For rapid mode shifting push and hold for at least 3 s
10. om inaccuracies and omissions If you should find an error or omis sion please contact Yamatake Corporation In no event is Yamatake Corporation liable to anyone for any indirect special or consequential damages as a result of using this product 1998 Yamatake Corporation ALL RIGHTS RESERVED CONTENTS GENERAL DIMENSIONS AND INSTALLATION WIRING ALARM SETTING AND ZERO SPAN ADJUSTMENT SPECIFICATIONS AND MODEL NUMBERS 1 GENERAL The Smart Indicator is a compact lightweight highly reliable DIN compatible LED bar graph indicator for process monitoring Features e Bar graph with high intensity custom LEDs e Three bar display colors red green and yellow selectable e Easy alarm setting and zero span adjustment from the front panel 2 DIMENSIONS AND INSTALLATION 1 External dimensions Unit identification Unit mm Mounting spacer and number nameplate Front Side 15 9 4 i W W P E 5 a7 2 LB Fig 1 Fig 2 2 Panel dimensions 3 installation of panel Fixing bracket Installation of single unit Side by side close mounting 35 54 36 X N 441 Mounting spacer 139 5 4 Mounting spacer Over 190 Fig 4 Over 190 N No of units mounted No of mounting spacers Nameplate One spacer on each side of panel either for single unit or multi unit installation 2 Fig 3 4 Panel mounting method 1 Attach the included mountin
11. rnal set point is compared to the PV value If the internal set point gt the PV value the contact turns OFF 5 At 1000ms the internal set point returns to the original set point When SP gt PV value the contact turns OFF 1 At next instant 2 After 100ms 3 After 800ms 4 After 1000ms 5 gt gt gt gt SP SPA p SP SP LEAFE g 1 0 FS Y Internal SP Internal SP y j Contact ON ON ON OFF Note For a low limit alarm the process is reversed 3 Zero and span adjustment For SIA and SIB types amp Span control knob Adjust the knobs to the desired value Value increases when knob is tumed clockwise and decreases when knob is tumed counterclockwise amp Zero control knob Fig 9 For SIC and SID types 1 Zero adjustment See Fig 8 1 Select the zero adjusting mode Z by pushing M The Z lamp should flash 2 Set the desired value by moving the PV indicating bar up or down with A or V 3 Load a set point by pushing SET The LED should go out 2 Span adjustment See Fig 8 1 Select the span adjusting mode S by pushing M The S lamp should flash 2 Set the desired value by moving the PV indicating lever up or down with Aor V 3 Load a set point by pushing SET The LED should go out 5 SPECIFICATIONS AND MODEL NUMBERS Model No No of indicating points Alarm High low limit High high limit low low limit Input type