1.        BE    u    User Manual arsos   soft    dexos    light wds    Item No  111 540 DL wds new    Ulcer Pressure Relief System    Assembly and Operation instructions    Power unit  dexos   light wds    Ce       pressure relief products    Health  Security  Independence           a Aca      List of contents   of   List of contents         Foreword page 3  1 Safety Instructions page 4  2 Product overview arsos   soft   dexos   light wds page 4  3 Introduction page 5  4 Explaination of indications and functions page 5  5 Installation arsos   soft page 6  6 Installation dexos   light wds page 6  7 Other settings page 10  8 Auxiliary functions page 10  9 Cleaning instructions   Service page 12  10 Troubleshooting page 13  11 Indications   Contra indications page 14  12 Technical data page 15    13 air foam combinations page 15       User manual arsos   soft   dexos   light wds new   Version 1 1 JS       BE A      Foreword      Dear client   You are now holding the user manual of the arsos   soft therapeutic support  system of ADL   GmbH in your hands     arsos   soft is a mattress replacement system  divided into three segments  with 28  independent removable air foam cells    This air foam technology enables the mattress by means of air support to adjust  itself to the weight and convenience of the patient    The functions are based upon known mechanisms  The system is equipped with   a soft foam mattress with reinforced edges  Further arsos   soft is a dynamic   or   when neede
2.       arsos   soft   dexos   light wds is an air mattress system  meaning that no extra  treatment nursing mattress is required  This system is suitable for patients with a  body weigth from 20   150 kg     Please note that these weight indications are applicable for patients in a laying  position  Other positions will result in another distribution of weight    The core of soft foam will ensure adaptation to the shape of the human body  which results in an as large as possible surface and maximum pressure relaxation   The reinforcements of the foam edges are an advantage for the sense of feeling  and the mobility of the patient  Touching the bed base is prevented by the 11 5 cm  foam core in the cell of the arsos   soft modules     This also helps to ensure compliance and conformity to DIN EN 1970 2000 and  DIN EN 60601 2 38 because even with full air pressure in the cells the safety  function of the bed rails remains unchanged       On off switch  Power      2  Pressure setting  Pressure   higher    y lower     Cycle time  Cycle      Mode  Static Alternate   6 Alarm button  Alarm     4 Explaination of indications and functions                  User manual arsos   soft   dexos   light wds new   Version 1 1 JS       a A      5 Installation arsos   soft     5 Installation arsos   soft   arsos   soft    Position the arsos   soft on the bed base frame and fix fixation straps  to the bed   to the    up and down  moving parts of the bed   If not  properly attached to the bed can r
3.    In case of technical problems please contact your appointed dealer    or direct to ADL   GmbH         Dealer stamp        ADL   Medical B V  International  Edisonstraat 6   5051 DS Goirle   The Netherlands   Phone    31  0  13 5772 088  Fax    31  0  13 5772 648    E mail  info adlmedical nl  Internet  www adlmedical nl    Qualit  tsmanagement  Wir sind zertifiziert    Ragalm  lige trewillige    berwachung nach EN ISO  13485 2003   AC 2007         Copyright by    ADL GmbH   Muenster  Version 02 2011    ADL   GmbH  Lise Meitner StraBe 3   48161 Muenster   Germany   Phone    49  0 2534 6224 0  Fax    49  0 2534 64270    ADL   GmbH   Production   and service centre Erfurt  Thoereyer Strasse 1  99334 Ichtershausen    Germany  Phone   49  0 36202 7519 0  Fax    49  0 36202 751912    E Mail  info adl gmbh de  Internet  www adl gmbh de       User manual arsos   soft   dexos   light wds new   Version 1 1 JS    
4.  the power supply again and  switching on the unit  is sufficient to continue with the same settings as before     5  Reset function   By pressing the buttons alarm   and cycle     the device can be put back to  factory settings  All settings are lost  The factory settings are alternating pressure  mode with pressure at 20 mbar  cycle time at 5 minutes  audible and visual alarm           User manual arsos   soft   dexos   light wds new   Version 1 1 JS    a Ace    9   9 Cleaning instructions   Service   instructions     9 Cleaning instructions   Service         e During use the mattress and tubes should be cleaned weekly using a damp soft  cloth and mild detergent or non phenol germicidal solution     Never use any detergents containing aldehyds     After changing to another patient the mattress should be disinfected by a  accredited company approved by the Robert Koch Institute  Please ensure  conformity to the European guideline for medical devices   directions for  preparation of medicine products and directions for hospital hygiene and  prevention of infections from the Robert Koch Institute or other guidelines   they rule    The PU cover can be washed to a maximum temperature of 95   C and be prepared  after a thermal process  e g  Ottalin   Peracet   process     When disinfecting the mattress  a maximum temperature of 75   C should be kept   In case of an arsos   soft Mattress with an alternating power unit  this should be  cleaned with an damp cloth and soap detergent 
5. AP    Patients with  lt  10 kg  and  gt  50 kg body weight    Instable fractures   loss of sensitivity   pressure ulcers stage Ill   IV   according to EPUAP    with High to Very High Risk   according to Braden    Patients with  lt  10 kg  and  gt  50 kg body weight    MD   eee    User manual arsos   soft   dexos   light wds new   Version 1 1 JS    a Ace    12 Technical data       Mattress Pump   arsos   soft dexos   light wds  Item No  100 540 111 203 wds neu  Material Polyurethane ABS  Dimensions  cm  200 x 90 x 11 5 23 x 15x 8 5  Weight ca  16 3 kg ca  1 5 kg  Warranty  2 years 2 years  Electrical 230 V 7 W  connections  Protection class Il  Fuse 230 V 500 mA slow  Cleanng Wipe disinfectable with Wipe disinfectable    adequate detergents      wi     p  Warranty only valid in case of production or material failures     13 air foam combinations    Item No  Description dexos   light dexos   autoplus    111 500 arsos   light  stage I ll stage     Ill stage     IV   up to 150 kg     111 540 arsos   soft  stage I ll stage     Ill stage     IV   20   100 kg     111 550 arsos   plus  stage I Il stage     Ill stage     IV   up to 250 kg     111 510 arsos   filius  stage I ll stage     Ill stage     IV   10   70 kg           User manual arsos   soft   dexos   light wds new   Version 1 1 JS       Cover  111 542  PU PA    200 x 95 x 12  ca  0 2 kg    95   C washable   dryer can be  used to 50   C    lenos   light  stage     IV  stage     IV  stage     IV  stage     IV          _ Aa    
6. d   static air filled mattress with an 11 5 cm safety foam layer  which prevents the patient to sink to the bed base level in case of a system  disorder     When changing treatment of the patient  the foam mattress can be converted   into a dynamic system through connecting the mattress to different control power  units  e g  dexos   light wds or dexos   autoplus     The composition of the system can be considered a system in a system making  adaptations possible to suit a wide variety of diseases    This module concept also combines more advantages concerning the safety aspects  as the protection of the bed side rails when large cell systems are being used   Higher bed side rails  which head contain again safety problems by falling heights   do not become any longer necessarily  as the employment of the arsos   soft  mattress  offer appropriate defaults that are kept in accordance with DIN EN  1970 2000 and DIN EN 60601 2 38     With the writing of this user manual the product management of the ADL   GmbH  has set itself the goal  to create a good and understandable aid and guidance for  the employment of the system  If nevertheless questions remain  please ask your  provider for direction and assistance     Your    ADL GmbH          User manual arsos   soft   dexos   light wds new   Version 1 1 JS       a Ao       1   1 Safety Instructions     1 Safety Instructions         The system may not be used in presence of open flames  heaters a o s  Protect the  system against humi
7. dity  Operate it only in dry areas  No combustible gasses of  liquids may enter into the system     EXPLOSION DANGER     Warning  Note  The compressor is under power  as soon as the power supply  plug is put in  Before opening the housing of the compressor the power plug   is to be unplugged  Repairs must be performed only by qualified personnel    The compressor may only to be used for it s purpose  In case of damage to the  compressor  the power supply plug must be unplugged immediately  A damaged  power supply cord should be replaced immediately  Only room temperatures   of 0   35   C provide assurance to good compressor operation  The use of a  pressure relieving mattress will not eliminate the need for a turning routine   for the patient  Do not remove  except for repairs  any cells from the system     When applicable  batteries and accumulators should be disposed on a proper way   These materials are not suitable to be disposed with normal  house garbage        2 Product overview arsos   soft   dexos   light wds    Units Parts list Item No   1 Mattress with 28 Cell tubes  6 fixation straps 111 540 DL wds neu  and Bi elastic washable PU cove  1 dexos   light wds  power unit  incl  power cord  1 Service manual 999 190  1 User Manual arsos   soft   dexos   light wds 112 015 GB  Spare PU cover 111 542    Spare power unit dexos   light wds   incl  power cord 111 203 DL wds       User manual arsos   soft   dexos   light wds new   Version 1 1 JS       Mm An      3 Introduction
8. e drops below the set level  the red alarm light is blinking  again for a long time  Together with the optic alarm  the power unit has an  acoustic alarm  The acoustic sound can be switched off and on with the  alarm button             User manual arsos   soft   dexos   light wds new   Version 1 1 JS    a Ace       There are no standard settings  The most effective pressure level can be set  with the pressure setting button   The scale is divided into 8 parts  After the power unit is turned on  the standard  pressure is set in the middle of the scale  about 30 mbar  Pressure can be  lowered by repeated pressing on the left button    W lower  to about  5 mbar  By repeated pressing on the right button    A higher   a maximum  pressure of 60 mbar can be set    Try to reach an as large as possible contact surface  This will reduce the pressure  at the patient s skin to a minimum  If you need an indication for a  pre adjustment you may use the following table     Indication mbar pressure to body weight    No  of lights body weight The weight intervals only give a rough indication for   T Blue lich Sm  HET  pre adjustment and check of the setting  It can  never be an exact measure for the most effective   2 Blue lights contact area of the body  How far the body sinks   3 Blue lights into the mattress is not only determined by the   5 weight  but also by length and shape  It is imperative   ZUPBIFLS to check the pre adjustment with a  hand check     5 Blue lights Try to move your ha
9. er unit should not be covered in  any way  as thermal damage can occur     2  Connect the tubes to the power unit by connecting the tube connectors    The connector must give a clear clicking sound  Check with a short pull at the  tube if the connecter is rightly adjusted   Please check rubber O rings on regular intervals        BED User manual arsos   soft   dexos   light wds new   Version 1 1 JS       na cs    Connection scheme  arsos   soft mattress with power unit dexos   light wds  Group A   Cells pair 1 2  7 8  13 14  19 20  25 26   Group B   Cells pair 3 4  9 10  15 16  21 22  27 28   Group C   Cells pair 5 6  11 12  17 18  23 24        3   Check if the tubes are not bent or clamped between other parts     4   Put the power supply plug in and switch on the unit with the power button     When the lights at the display are lighting up  the unit is operational  The  power unit is pre adjusted so at the start a pressure is building up in the  system right between the minimum and maximum level    The unit starts up in alternating mode with a 5 minute cycle     65  The red alarm light   is flashing and the system will start to be filled with air   The red alarm light is put out of circuit when the process is completed    After 15  20 minutes the system is filled and you can put the patient on the  mattress  The red alarm light should be put out of circuit when the patient is  laying on the mattress  It has no meaning when the alarm now and then briefly  flashes  When pressur
10. esult in damage to the bed and mattress   G  Place the mattress in such a way that the tubing s are located at the  foot end of the bed   Verify the CPR function is closed  CPR   Quick deflating function   In systems with a valve this should be in horizontal position  in systems with a  plug closure  have all plugs put firmly in  The CPR is designed to deflate the  mattress immediately       The dual stretch and breathing cover is to be fixed to the bed base with the  zipper on both sides   Before putting the patient on the mattress you can fill air into the system with a  hand pump  As this is a 3 chambers system  there will be 3 tubes at the foot end    7  After placing the patient on the mattress the power unit must be connected  to the mattress   Cover the mattress only with a cover normally used within nursing environments     Remove the existing nursing mattress     Only use covers which are properly sized and sufficiently elastic in order   to prevent the  hammock effect   see page 7   Creases in the sheets can be  smoothed from time to time  but do not tuck the sheets underneath the  mattress as the tension of the sheets can increase pressure     6 Installation dexos   light wds    Q  Before hanging the power unit at the foot end of the bed you have to mount the  bracket at the back of the power unit  Slide the bracket from below up into the  fixation slot  Hang the power unit at the foot end of the bed  but you can also  put it under the bed  Please note that the pow
11. ime     2  The cycle time on the power unit can be varied from 5 to 20 minutes   By pressing button     Cycle      cycle time can be changed from standard  5 minutes with equal steps of 5 minutes to a maximum of 20 minutes   Within one cycle all three cell groups are inflated and deflated once     8 Auxiliary functions    1  Key lock  The keyboard has a key lock that is activated after about 5 minutes     This is to prevent an unintentional displacement of the set values    The key lock is released by simultaneous actuation of the following buttons   alarm   and pressure     and pressure      After this the settings can be  changed     2  Brightness of the display   The brightness of the display can be set in 3 stages with the key combination  alarm   and static alternate     The starting situation is the brightest  By actuating the key combination it sets the  display less bright        User manual arsos   soft   dexos   light wds new   Version 1 1 JS       Pm An    3  Power failure alarm   When the power system to which the unit is connected fails  there is a power failure  alarm  The unit beeps and the red alarm light flashes  The acoustic alarm can be  disabled with the alarm   and power   button  When the power supply is  restored  the alarm will also be disabled     4  Memory function   The unit has a memory function  This means  when the power is interrupted   the unit will remember the set values up to about 24 hours    Thus  after transfer of a patient  connecting to
12. larm light   is flashing and  the acoustic alarm  sounds    Cause    Connectors are not  accurately locked  Tubes  are twisted or kinked     Air cell defect  Tubing  defect  Pressure not  correctly set     No power available   Power supply plug  defect  power cord  defect  safety fuse  defect     Defect air cell   defect tube    CPR valve open   connectors not  properly connected    Measure    Check connection and  position of the tubes     Check cells and tubing  connections  Check if  the tubings are  properly functioning   Check if cells are in  proper position    Check pressure setting     Check power supply  plug and partition  socket  Check the  safety fuse     Check cells  tubes    CPR valve and  connectors    check if pressure at the  outlet of the power  unit is within tolerance           Bun      11 Indications   Contra indications   Contra indications       Function    Alternating  pressure with  dexos   light wds    Static  soft bedding  with  dexos   light wds    Indication    Decubitus prevention and  therapy support up  till stage III   according to EPUAP    with High Risk   according to Braden    Patients with   lt  50 body weight    Pressure ulcers prevention and  therapy support up  to stage Ill   according to EPUAP    with high risk   according to Braden    Pain  muscular hypertonia       Counter indication    Instable fractures   loss of sensitivity   pain  muscular hypertonia    e g  contractions  spasms   pressure ulcers stage Ill   IV   according to EPU
13. nd under the knee joint  the   6 Blue lights lumbar and neck lordosis in order to check if   7 Blue lights supporting material is present  When the patient    sinks to far in the mattress the so called Hammock     8 Blue lights   45     55 mbar ca   gt  80 kg  effect occurs  This should be prevented at all times     But it is imperative to check the pre adjustment with a    hand check     Move your hand under the knee joint  the lumbar and neck lordosis  In order  to check if supporting material is present    When the patient sinks too far in the mattress the so called  hammock effect   occurs  This should be prevented at all times    When pressure is to high there will be a lack of supporting material underneath  the lumbar and cranial lordosis and the knee joints  The situation can be  improved by pressing button    W lower     When pressure is too low  hammock effect   press button    A higher    Please note that status of the air cells is to be checked daily and whenever  required adjustment is needed    Each time when changing the pressure level you have to wait for the same time  interval of the set cycle time  before the right level has been reached  Only  after this you can check the contact area again        User manual arsos   soft   dexos   light wds new   Version 1 1 JS              i   j  a  gt  B   D 5    BD w    Q        EELS Principle of alternating pressure       cet grove 6  every second cells pair deflated  showing group B          Cell group           Hamm
14. ock effect   to be avoided at any time   Ni                            High     00000000000 ak  O   Cell group A O   Cell group B     Cell group    in the sacral area     Optimal support of the patient   Knee joints  lumbar and cranial lordosis are supported             Cell group A    O   Cell group B        Cell group C       min  3 4 cm        O     Patient may not    Cell group A   a       srr Static pressure sink to the base     Every cell has equal pressure  of the bed         Cell group C          User manual arsos   soft   dexos   light wds new   Version 1 1 JS    mm Aa    7   7 Other settings     7 Other settings         Static     1  In case of a contra indication for alternating pressure the system can be  switched to static mode by pressing button    static alternate   The light will  change from blue to green  static     In static mode all cells of the system will be filled with air to the set pressure  level  In static mode all cells are active  therefore it is necessary to reduce the  set pressure with about 5 10 mbar  1   2 steps on the scale      To switch back to alternating mode  again press button    static alternate    The light will change from green to blue  alternate     In alternate mode 2 out of 3 cells are filled with air to the set pressure level   One cell remains without pressure    Don t forget to increase the pressure when switching over to alternating  pressure with the same difference used to reduce  switching over to static  mode     Cycle t
15. or a mild non  aldehyds cleaning  detergent    Disinfection with cleaning on locations is also possible  when instruction for use  and process time are properly followed     Recommended disinfectants    BACILLOCID RASANT  BODE   BACILLOL AF  BODE   BACILLOCID SPEZIAL  BODE    KOHRSOLIN FF  BODE   KOHRSOLIN  BODE   MICROBAC FORTE  BODE   DISMOZON  PUR  BODE   INCIDIN FOAM  ECOLAB   INDUR DES  ECOLAB      Check with your supplier that the disinfectant does not damage plastic  surfaces  PUR  PA  ABS  PVC      Prevent contact on electrical components with damp or moisture    Cleaning cloths should be damp  not wet    Carefully check no moisture remains in the system    Execute a functions control procedure before using the system again   E Check your cleaning process on effectiveness on a regular basis     zz ee In    User manual arsos   soft   dexos   light wds new   Version 1 1 JS       BE AL       ADL   GmbH prescribes safety related revisals of all electric powered parts after at    least 2 years     The rules for accident prevention BGV A2 will remain unaffected   The revisals may only be conducted by especially trained personal     10 Troubleshooting    Module    arsos   soft    dexos   light wds        n    User manual arsos   soft   dexos   light wds new   Version 1 1 JS    Problem    Mattress is not or  insufficiently  inflating     The patient sinks  to deep into the  mattress     The power unit  does not operate   blue lamp at  the switch is not  illuminated      Red a
    