Home
        User Manual - CkHealth.de
         Contents
1.     User Manual for Vital Signs Monitor    3 3 Sensor Placement and Connection    3 3 1 CO  Sensor Connection    3 3 1 1 Sidestream CO  Sensor Connection    Exhaust AF   vent           Ir    eS  Baie   a i JA   AS 2 C07    sample cell  Receptacle           AR Sensor Cable          Sidestream  CO  Sensor    Sample cell    Figure 3 3 Demonstration for Sidestream CO  Sensor Connection    1  Take out the CO  Sensor and insert the CO  Sensor Cable into the connector labeled    CO     on the connector  panel of the monitor     2  The sample cell of the sampling cannula must be inserted into the sample cell receptacle of the CO  Sensor  as shown in Figure 3 3  A    click    will be heard when the sample cell is properly inserted  Then connect to  airway tube  After finishing sensor connection  and make sure that the air input end is exposed to room air  and away from all sources of CO  including the ventilator  the patient   s breath and your own  Next  turn on  the CO  switch at CO  Setup Screen and then wait 2 minutes for the sensor warm up     3  Default Tubing Configuration    sampling dehumidification sampling sample     adapter tubing tubing   filter       cell  fi j i       a Fi           Figure 3 4 Sampling line kit  Single patient use     Figure 3 5 Extending airway tube for connecting to sampling tube  Single patient use     User Manual for Vital Signs Monitor    Figure 3 6 Wye Connector  4  Optional sampling cannula kits     1  T connector sampling cannula kits         On
2.    Te CREATIVE  MEDICAL       Vital Signs Monitor    PC 900A    User Manual    Shenzhen Creative Industry Co   Ltd     User Manual for Vital Signs Monitor    This Manual is written and compiled in accordance with the IEC 60601 1 Medical electrical equipment Part    General requirements for safety  and MDD 93 42 EEC  It complies with both international and enterprise  standards and 1s also approved by State Technological Supervision Bureau  The Manual is written for the  current PC 900A Vital Signs Monttor     The Manual describes  in accordance with the Vital Signs Monitor s features and requirements  main  structure  functions  specifications  correct methods for transportation  installation  usage  operation  repair   maintenance and storage  etc  as well as the safety procedures to protect both the user and equipment  Refer    to the respective chapters for details    The Manual is published in English and we have the ultimate right to explain the Manual   Version of This Manual  Ver1 1   Revised date  May 25  2010    All rights reserved     Marks in the Manual   6    Warning  must be followed to avoid endangering the operator and the patient   A Attention  must be followed to avoid causing damage to the monitor        Note  some important information and tips about operations and application     3502 2530002     I     User Manual for Vital Signs Monitor    Instructions to User    Dear Users     Thank you very much for purchasing our product  Please read the following info
3.   1  when battery is used to supply power  user should not charge the battery until the low battery    alarm rings   After line haul or long time storing  using battery may not start the monitor     please charge the battery      2  The battery should be charged for 10 to 15 hours     9 1 4 Service    If the monitor has functional malfunction or is not working  please contact the local dealer or our company     and we are to offer the best solution as soon as possible for your satisfaction  Only qualified service engineer    specified by the manufacture can perform the service  Users are not permitted to repair 1t by themselves     9 2 Cleaning  Sterilization and Disinfection    boob oboe eme    Kept the monitor from dust     It is recommended to clean the outer shell and screen of the monitor to keep it clean  Only non corrosive    cleanser such as clear water is permitted     Use the cloth with alcohol to wipe the surface of the monitor and transducers  and dry it with dry    and clean cloth or simply air dry     The monitor can be sterilized and disinfected  please clean it first       Switch off the monitor and disconnect the power cable before cleaning       Do not let the liquid cleanser flow into the connector jack of the monitor to avoid damage       Clean the exterior of the connector only     Dilute the cleanser    Do not let any liquid flow into the shell or any parts of the monitor   Do not let the cleanser and disinfectant stay on its surface    Do not perform hi
4.   alarm  range     1 100         SpO  Lo     low limit of SpO  alarm  range     0 99         PR 2     pulse rate alarm switch     2    indicates PR alarm is on      X     indicates PR alarm is off      PR Hi     high limit of PR alarm  range     22 250    bpm        PR Lo     low limit of SpO  alarm  range     0 248    bpm        gt       gt        PI display        on    means PI display is enabled     off    means PI display 1s disabled     4 4 3 Nurse Call    Output level    Source  ALM    H Im    L    Duration  Probe off 01 20 14 50    Figure 4 10 Nurse Call Setup Screen       ee    User Manual for Vital Signs Monitor       Output level     two options    low    or    high    output levels are available   When the calling system in hospital works in    Normal Open    mode     low level    should be selected   When the calling system in hospital works in    Normal Close    mode     high level    should be selected       Source     three kinds of alarm sources can trig the nurse call  high level alarm  medium level alarm    and low level alarm  multi optional   If you don   t make choice  nurse call signal will not be sent out      Duration     two options    pulse    or    continuous    output modes are available        continuous     the continuous mode of output means the nurse call signal will keep until the selected    alarm source es  disappear  1 e  the signal will last from starting alarm to stopping alarm        pulse     the output nurse call signal is pulse signa
5.   lt  gt      lt  gt        Balance     balance gas   adjust the concentration of balance gas  Use this setting to correct for the    compensation of the gas mixture administered to the patient  Range     0 100           Agent     Anesthetic agent   select whether the anesthetic gas 1s supplied to patient  two options      on    or    off     Use this setting for ensuring the accuracy CO  monitoring when the anesthetic gas 1s    supplied to patient        Apnea     Use this setting to set the no breaths detected time out  This time out is the time period in  seconds following the last detected breath at which the monitor will signal no breaths detected  At    this time all the measuring values will be cleared  Range     10 60   s  the default is    20   s      22     User Manual for Vital Signs Monitor    Operation Instructions   1  Press     A    key or    V     key to move cursor to select parameter    2  Press    Ml    key to confirm and active this parameter setting    3  Press    A    key or     W    again to adjust or modify parameter value   4  Press    MW    key again to confirm and save the setting     5  After setting  press     mi      key to return to upper level screen     4 4 2 SpO  Setup    PR MM Hi  180  Lo  40     PI display    Probe off 01 19 30 07       Figure 4 9 SpO  Setup Screen  Screen Description    lt  gt     SpO  A     SpO  alarm switch      amp     indicates SpO  alarm is on      IX  gt  indicates SpO  alarm is off      SpO  Hi     high limit of SpO
6.  0 69mmHg  0 25mmHg 70 150mmHg   H  Flow rate  50ml min  10 ml min  Sidestream     5 2 SpO  Monitoring and Pulse Rate Monitoring    A  Sensor  dual wavelength LED  B  SpO  measurement range  3546 9946  C  SpO  measurement accuracy   75    99   2   50    74   3   D  Pulse rate measurement range  30bpm 240bpm  Pulse rate measurement accuracy   2bpm or 2  whichever is greater   F  Performance under low perfusion condition    The measurement accuracy still keeps the above specification while the blood perfusion index 1s as low  as 0 4   Display range of Perfusion Index  0 2  20     5 3 Recorder    A  Recording speed  25mm s    206     User Manual for Vital Signs Monitor    B  Recording speed accuracy   lt 5     5 4 Power Requirement    AC power supply voltage  100 250VAC       AC power frequency  50 60 Hz  Fuse specification  T3 15AL 250V   5 X 20mm     Internal power supply  12VDC  rechargeable     NU  O W p       Internal power supply  rechargeable sealed lead acid battery  12V 2 3AH    5 5 Environmental Requirement    Operating Environment       Ambient temperature range  5  C   40  C  Ambient relative humidity range  30    80   Atmospheric pressure  70kPa  106kPa    Transport and Storage Environment       Ambient temperature range   20  C   60  C  Relative humidity  10   90   Atmospheric pressure  53kPa  106kPa    5 6 Classification    Complied Safety standard IEC 60601 1    The type of protection against electric shock Class I equipment   The degree of protection against ele
7.  User Manual for Vital Signs Monitor    Chapter 1 Overview    1 1 Features     lt  gt  End tidal Carbon Dioxide  EtCO    Inspired Carbon Dioxide  InsCO    Respiration Rate  RR    SpO  and Pulse Rate  PR  are displayed by big  bright numerical LEDs      lt  gt  CO  waveform  SpO  plethysmogram and system parameters are displayed on dot matrix LCD    screen      lt  gt  CO  measuring module with advanced technology for quick and accurate EtCO  and RR    measurement      lt  gt     Unique SpO  measuring technique ensures sensitive and accurate SpO   PR and Perfusion Index     PI  measurement   EtCO  and SpO  trend graph display for latest 12  24 or 96 hours   Audible  amp  visible alarm with 3 levels of alarm events     Nurse call output 1s available                Built in printer is optional to print out waveforms  and text information     1 2 Product Name and Model    Name  Vital Signs Monitor  Model  PC 900A    1 3 Intended Use    This monitor is multi functional equipment designed for monitoring the vital physiological signs of the  patient  It is used to monitor continuous CO  and SpO   and report the End tidal Carbon dioxide  EtCO     inspired CO   InsCO    respiratory rate  RR  value  pulse oxygen saturation value  SpO    pulse rate  PR     and perfusion index  PI  of the intubated and non intubated patient     This equipment is applicable for use in hospitals in long term realtime monitoring of CO  and  SpO  during anesthesia   recovery  It can be used in various depart
8.  air Connectffr 7   BP    Adapter  Figure 3 7     2  Nasal Sidestream Cannula Kits    Figure 3 8   3  Oral Sidestream Cannula Kits       Figure 3 9    User Manual for Vital Signs Monitor    3 3 1 2 Mainstream CO  Sensor Connection        Sensor cable    CO  sensor     gt     Figure 3 10 Demonstration for Mainstream CO  Sensor Connection    1  Take out the CO  Sensor and insert the CO  Sensor Cable into the connector labeled    CO     on the connector  panel of the monitor    2  Snap the CO  sensor onto the airway adapter as shown in Figure 3 10  A    click    will be heard when the  airway adapter is properly inserted     3  Position the airway adapter in the patient   s respiratory circuit  as close to the patient as possible  between  the endotracheal tube and the ventilator circuit  Next  turn on the CO  switch at CO  Setup Screen and then    wait 2 minutes for the sensor warm up         Sens  Mainstream SS  p     CO  Sensor   ang ro P 2    or Cable        a      To Patient    Figure 3 11    Safety Introductions for CO  Monitoring    6 CO  Sensor is a precision measuring part  please use it correctly and store it properly     6 Precautions for electrostatic discharge  ESD  and electromagnetic interference  EMI  to    and from other equipment     6 Failure of Operation  If the CO  Sensor fails to respond as described in this user  manual  DO NOT use it until approved for use by qualified personnel        DO NOT position the sensor cables or tubing in any manner that may cau
9.  display panel  and power supply block  etc      gt  The CO  module measures EtCO  and InsCO  and calculates RR      gt  Besides supplying power to CO  module  the power supply board for CO  collects all the data  from the CO  module  including EtCO   InsCO   RR and CO  waveform etc      gt  The SpO  module detects PR and SpO  via SpO  probe  and then calculates their values and  provides SpO  plethysmogram and Pl as well      gt  The main control unit is in charge of LED and LCD display  keyboard input  data storage   printing  and networking function     User Manual for Vital Signs Monitor    Chapter 3 Installation and Connection  3 1 Appearance    3 1 1 Front Panel                10  11    12  13    14  15                    a   CI   ez    2007 08 09  11  30  25    16    21     AP    FAN        B                      Figure 3 1 Front panel illustration    Description   1 Alarm indicator    Alarm Indicator  Red flashing High priority alarm Exceeding the limits or low voltage       Orange flashing Medium priority alarm SpO  probe off  No CO  sensor  Greenlight   Normal fo    10    11    12    13    14    15    16    17    18    User Manual for Vital Signs Monitor    SpO   display SpO  value  Unit      SpO       EtCO   display EtCO  value    InsCO   display InsCO  value    RR  display respiration rate measuring value  Unit     rpm       PR  display pulse rate measuring value  Unit     bpm  beats per minute                   Bar graph of pulse intensity    LCD panel    EtCO  a
10.  is 12 hours  the upper trend graph will display SpO  trend curve for last 12 hours      lt  gt     cursor on     enable the display of cursor on trend graph  i e  the vertical cursor line displayed in    trend graph  so the user can move the cursor to inspect the SpO  value at the given time      lt  gt     SpO      indicate that the trend graph beside it is SpO  trend  Let the cursor stay here and press     key to confirm  then press    A    key or    V    key again to select trend graph type         SpO       SpO  trend graph       JEtCO      EtCO  trend graph  as shown in Figure 4 6     21 50 01 50 05 50    12 hours  cursor on    Probe off 01 17 44 59       Figure 4 6 CO  Trend Graph    4 3 2 Operation Instructions    l  Press    A    key or    VW    key again to select trend graph type    EtCO     or    SpO          19     User Manual for Vital Signs Monitor    2  Then press     A    key or    W     key to highlight    trend length    or    cursor on    selection   3  Press    MW    key to confirm     4  Press    A    key or    W     key again to select value of trend length  12 24 96 hours  if the selecting box    stays in    trend length    option  or to move the cursor 1f the selecting box stays in    cursor on    option   Instructions for viewing the trend curve     e Select    cursor on    and press    MW    key to confirm  and    cursor on    becomes    cursor off      then you  can press    A    key or    Y    key to move the vertical cursor  the list box below wi
11.  lights     3  The SpO  sensor should not be used at a location or limb tied with arterial or blood pressure cuff or    receiving intravenous injection     4  Do not fix the SpO  sensor with adhesive tape  or else it may result in venous pulsation and    consequential inaccurate measurement result of SpO    5  Make sure the optical path 1s free from any optical obstacles like adhesive tape     6  Excessive ambient light may affect the measuring result  It includes fluorescent lamp  dual ruby    light  infrared heater  and direct sunlight etc   7  Strenuous action of the subject or extreme electrosurgical interference may also affect the accuracy     8  Please do not use the SpO  sensor when having the MRI  or burn may be caused by faradism   7 2 3 Clinical Limitations    1  As the measure is taken on the basis of arteriole pulse  substantial pulsating blood stream of subject  is required  For a subject with weak pulse due to shock  low ambient body temperature  major  bleeding  or use of vascular contracting drug  the SpO  waveform  PLETH  will decrease  In this    case  the measurement will be more sensitive to interference     2  For those with a substantial amount of staining dilution drug  such as methylene blue  indigo green  and acid indigo blue   or carbon monoxide hemoglobin  COHb   or methionine  Me Hb  or  thiosalicylic hemoglobin  and some with icterus problem  the SpO  determination by this monitor    may be inaccurate     3  The drugs such as dopamine  procaine  
12.  of rubber roller lock at the left 90  upwards to unlock it  refer to the following figure  with mark        3  Cut one end of the paper into triangle  and load the paper from the underside of the rubber roller   4  Turn the roller clockwise to get the paper rolled  and put the paper roll into the compartment    5  Pull the paper out of paper slot on the shield    6  Move the tab of the rubber roller lock 90  downwards to lock tt    7  Put the shield back in position and secure it     Operation procedures for taking out printing paper roll     1 2 steps are the same with the 1 2 steps mentioned above for loading printing paper   3  Roll the loading roller anti clockwise and pull the paper out     4 5 steps are the same with the 6 7 steps mentioned above for loading printing paper     a Rewind  G     Formfeed       Figure 3 14 Loading and taking out printing paper    3 3 4 Battery Installation    1  Ensure that the monitor is not connected to AC power supply and the monitor is turned off     2  Open the battery cover and place the battery in the direction as shown in Fig  3 15 to insert the  battery into any one of battery compartments  Do not insert battery with their polarities reversed     3  Move the battery baffle to secure battery     e    User Manual for Vital Signs Monitor    4  Close the battery cover     Battery    Battery baffle  Battery cathode    Underside view    Rubber gallus    Battery cover       Figure 3 15 Battery Installation  Note   e Do not insert battery
13.  sensor by means of a pump  This type of system 1s  needed for non intubated patients    When using mainstream CO  sensors  check the window for the patient secretions pooled on periodically   Because that condition may affect the accuracy of the measurement or even make the sensor not work   When using sidestream CO  sensors  there 1s a water trap or a part of the sampling tube with dehumidifying    function  Please periodically check the flow sensor and tubing for excessive moisture or secretion buildup     7 2 The Principle of SpO  Pulse Measurement    7 2 1 Working Principle    This monitor measures the pulse oxygen saturation  SpO gt   and pulse by means of the radiograph of infrared  light and the red light emitted by LED through body s peripheral areas  such as fingers   whereby the  photoelectric detecting circuits will analyze the absorptivity of the oxyhemoglobin and reduced hemoglobin  respectively  and give the photoabsorption rates before and after pulsation  Using the measure of  photoabsorption change due to pulsatory arterial blood flow caused by PLETH waveform  the SpO  can be    obtained      29     User Manual for Vital Signs Monitor    7 2 2 Points to be noted in SpO  and Pulse Measuring    pama      The finger should be properly placed  see the attached illustration of this instruction manual   or else    it may cause inaccurate measurement result     2  Make sure that capillary arterial vessel beneath the finger 1s penetrated through by red and infrared 
14.  terminal with its polarities reversed  or the monitor can not be started     e Please take out the battery before transport or storage     TAS    User Manual for Vital Signs Monitor    Chapter 4 Operations    4 1 Initial Monitoring Screen    Insert the SpO  sensor cable into the connector labeled    SpO      follow the instructions of Section 3 3 1 to  connect the SpO  sensor and its accessories  and then connect to patient well  The LCD will display the initial    monitoring screen  and it is the default display screen as well     PI  3  PLETH Title Area       Main Display Area    mute C D  120 03 11 30 25 Prompt Info Area       Figure 4 1 Default Display Screen    The LCD screen will display the information by different display views  short pressing     mi      key to shift    screen display modes among display views  SpO  plethysmogram screen  default screen   CO  waveform    screen  and trend graph screen  Long pressing     mi      key will enter the setup menu screen  For every display    view  the display area is divided into 3 parts  title area  main display area  and prompt info area  see Figure  4 1   The prompt info area contains 3 segment of information  status or event indication at the left  patient ID    number in the middle  real time clock at the right  also see Figure 4 1      4 1 1 Screen Description    Title area  See Figure 4 1     lt  gt     PI  3      the perfusion index is 3 o  it displays only when    Setup Menu     SpO   PI Display    is  set as    O
15.  the figure below           Figure 3 12 Finger clip SpO  sensor placement    3  If the neonate SpO gt  sensor is used  please follow Figure 3 13 to connect     Sensor adapter  wrapper       Y type sensor    Figure 3  13 Neonate SpO  sensor placement    Safety Introductions for SpO  Monitoring         Continuous use of finger clip SpO  sensor may result in discomfort or pain  especially for  those patients with microcirculatory problem  It is recommended that the sensor should NOT    be applied to the same finger for over two hours     6    SpO  measuring position must be examined more carefully for some special patient  Do NOT    place the SpO  sensor on the finger with edema or fragile tissue       4 Please do not allow the cable to be twisted or bended      lt 12     User Manual for Vital Signs Monitor    D    Do NOT put the SpO  sensor and pressure cuff on the same limb  otherwise the NIBP measuring    will affect SpO  measuring and cause the alarm error   Using nail polisher or other cosmetic product on the nail may affect the accuracy of measurement   The fingernail should be of normal length     Do NOT use the damaged SpO  sensor     kay E E    The SpO  sensor can not be immerged into water  liquor or cleanser completely  because the sensor    has no capability of waterproofness     3 3 3 Loading printing paper    Operation procedures for loading printing paper   1  Press both    OPEN    notches with force on printer shield with two thumbs to open it     2  Move the tab
16. E LEE OEE T IEO 7  a E A eh eg Z   3 3 SENSOR  PLACEMENT AND CONNECTION zu  ee 8  Hl COS CAS OF CONMECIION Geniessen 8  IN ONNECL A AHA 11  o E KO BN 13  IA BANGIN ASTA NANO nee Ga NGA 13  CHAPTER 4 OPERA TIONS 28  Ba A AA ANNA 15  A INITIAL MONITORING SCREEN ia 15  ATP ORCOS CAPO AA ASA ue ee ie A 13   hd OPTa NOn AINSIFUCHONS atada 16   A 2 COs MONITORING SERBE NE  85 0588 a KG SANG BNG BAG LOG AGA e LAG LANG NA EN NG 17  SES ATI E PO ae ee  o E E E A E NR 17  ALLA DET AMON INSITUCHONS  lt A AA A lada 18   43 TREND GRAPH DISPLAY nee 18  ADSL CIC AO COM NA NAAN ANN aaa 19  ALA CIATION AA AA AA 19   44 SETUP MENU SCREEN ANNA AA 21  44 1 CO SCNID naaa bet i itch trate AS aria ata A A A A 21  AN ANA AA See 23  ANS CO AR 23  NANA een 24  HI Ballen IN Osa ee ee ee A ee 23  44 0 DAGA Naa iaa 25   AUS a Ba Al SEN Nas nn ee ae 25  CHAPTER 5 TECHNICAL SPECIFICATIONS           0000000ss0000000000000000000000000000 000000000 0000000000000 0000000000000 26     IIl     User Manual for Vital Signs Monitor    Sal CO MONITORI ee NAA ANA ada ada 26  5 2 SPO  MONITORING AND PULSE RATE MONITORING   ccocccnoccnnocnnoconnncnnaronnncnnarononcnnanonnnrcnnoconarcnunccnaronnns 26  EE RT ORI a ae estate cia 26  S2 OW ERR EI OUTRE MEIN Trois locas 27  5  ENVIRONMENTAL REQUIREMENTS ie 27  DOC ICAT O N ee ee seen AA AA AA AA AA 21  5 7 DEFAULT ALARM LIMITS AND SETTING RANGES ariete dd dida adas 24  CHAPTER 6 PACKAGING AND ACCESSORIES                ccssccsssccsscccssccsscccscccsccccscccsccccscccsccesccc
17. N       lt      PLETH     Mark of SpO  plethysmogram  when    PLETH    displays in title area  the main display  area will be SpO  plethysmogram and this display screen 1s the default screen   Main display area  See Figure 4 1      When SpO  sensor is placed on the patient and connected to the monitor well  a trace of sweeping waveform     plethysmogram  will be displayed in the main display area  as shown in Figure 4 1      If the SpO  sensor is disconnected from the monitor or off from the patient  the plethysmogram will become a  base line in main display area and    Probe off    will appear at the left side of prompt info area  as shown in  Figure 4 2       15     User Manual for Vital Signs Monitor    Probe off 01 17 44 59  Figure 4 2 Probe Off       Prompt Info Area  See Figure 4 1       lt  gt     Status or event indication segment     This segment will display the probe status  CO  sensor status  alarm silence counting down timer   over limit warning and other error messages for technical warning  If more than one event occurs or    more status appears  the indication message will be displayed alternately at this segment        mute C D  XXX     the counting down timer of alarm silence 1s XXX seconds  This prompt    message appears only when the alarm silence is enabled     Note  refer to Appendix I for detailed explanations     Patient ID segment      03     Patient ID number   Real time clock segment        11 30 25     the current time     4 1 2 Operation Instructi
18. ber  01 100 adjustable     Note  If the patient ID is changed  all the history data will be cleared  that means SpO  trend graph and  EtCO  trend graph will become empty     4 4 6 Date Time    yy  07  mm os  aaja     hh  10 mm  15   ss   20      Probe off 01 12 17 50       Figure 4 13 Data Time Setup Screen   lt  gt     yy07 mm09 dd 21     date setting     07 09 21    shows the date is September 21      2007      lt  gt     hh10 mm15 ss 20     time setting     10 15 20    shows the time is10 15 20   4 4 7 Recover Default Settings    On Setup Menu screen  press     A    button or    W     button to shift cursor to    Default     and then press    MW       button  all the setting parameters will be reset to factory default setting value     _25     User Manual for Vital Signs Monitor    Chapter 5 Technical Specifications    5 1 CO  Monitoring    Technology  Infrared absorption method   Mode of Sampling  Sidestream or Mainstream  Measuring parameters  Respiration Rate  EtCO  and InsCO     Jaw    CO  Response Time   Sidestream   lt 3seconds  includes transport time and rise time      Mainstream   lt 60ms  rise time     E  Typical measurement range   EtCO  0 150mmHg  InsCO  3 50mmHg    Respiration Rate 2 150bpm  Sidestream  or 0 150bpm  Mainstream   F  CO  Accuracy  0 40mmHg  2mmHg    41 70mmHg  5  of reading  71 100mmHg 8  of reading  101 150mmHg 10  of reading   NOTE  Gas temperature at 25  C for Sidestream     Gas temperature at 35  C for Mainstream  G  CO  Resolution  0 1mmHg 
19. ctric shock Type BF applied part    5 7 Default Alarm Limits and Setting Ranges    Alarm range Default  Parameter    LISO pm  RR    0 149  rom  R       Lo   0 160  mmHg 70mmHg  ih Low  0 159  mmHg 10mmHg  Lo  O        27     User Manual for Vital Signs Monitor    Chapter 6 Packaging and Accessories    6 1 Packaging    The product is packed in high quality corrugated cartons with foam inside to protect the equipment against    damage in the shipping and handling process     Weight  Details see the indication on the outer package     Dimension  360 L x320 W x410 H   mm     6 2 Accessories       For Mainstream    For Sidestream       Mainstream sensor    Sidestream Sensor                                  One set One set     CAPNOSTAT 5    LoFlo C5    Airway adapter One piece Sampling Line Kit One piece  SpO  probe One piece SpO  probe One piece  Power cord One piece Power cord One piece  Grounding wire One piece Grounding wire One piece  User manual One copy User manual One copy  Warranty One copy Warranty One copy  Quality Certificate One copy Quality Certificate One copy  Dust cover One piece Dust cover One piece  Packing list One copy Packing list One copy             Note  The accessories are subject to change  Detailed items and quantity see the Packing List      28        User Manual for Vital Signs Monitor    Chapter 7 Working Principle  71 CO  Measurement    7 1 1 Working Principle    The principle 1s based on the fact that CO  molecules absorb infrared light energy of s
20. eesceees 28  OL PACKAGING sauren res 28  02 ACCESSORIES een NANANA idad 28  CHAPTER 7 WORKINGPRINGIPEE 3NA KANINANG IBAAN NANG 29  Lal NMS IR IM Weta ence tetera a e NA Gn a DA Sana ae 29  Dodd INVOCA AP AA 29  74 2 Mainstream Vs  Sidestream SQMPUine crus sziacsunisdacswasatacsun a psn akan baa aka anakan bkas aba 29   7 2 THE PRINCIPLE OF SPO gt  PULSE MEASUREMENT     2202000 0004000 29  RL Working  Principle aa ANA AAA 29  7 2 2 Points to be noted in SpO  and Pulse Measuring               000000200000aaaaaaaaaaaaasasaa saaan 30  PERES IR WI NT LANG LI BA AS A AA MA MATA 30  CHAPTER 8 TROUBLESHOOTING 5000 ecc 31  Bal INO DISPEAYON THE SCREEN a 31  02 NOE TCOBAND SROs RESULT datada isos 31  8 3  BLANK PRINTED  PAPER aa 31  BESSER O 31  CHAPTERS  MAINTENANCE NATN 32  al SERVICE ANDE XA MIN ATION 20 ii EGG 32  OL AN TONTON ON e e aE ao 32  PLSEROMMEMANIeNaN Ce    ss ss AA  32   0  1 5 B  llery  Maintenance a NG dd 32  O  a 6x A Bt A ee 33   9 2 CLEANING  STERILIZATION AND DISINFECTION             ccscccsscccsscccscccssccesccesccesccessccescceusceuseceescenseceuscss 33  9 3 CLEANING  STERILIZATION AND DISINFECTION OF ACCESSORIES          ccccssccsscccsccescceescccscccusccesscensces 34  FS TORE AAA greta ce ewe se eee 34  9 3  ERANSPORTATION ste adidas 34  APPENDIX ss rennen 35  EPROMPFINEORMA  TIONEXPEANATIONS O AA AN NATA MA NINANG pA GANG ER 35   II TYPICAL PRESSURES AND CO  READINGS AT ALTITUDES            c  ccecccseccscccssccesccessccesceesccescceusccessceusces 36    IV    
21. er or the condition of the patient     2  Probe off  Please check the connection of the probes   Note  In case of trouble of this machine in the service  follow the instructions below to eliminate the    problem first  If the attempt fails  contact the dealer in your local area or the manufacturer  Do not    open the cabinet without permission     Aj    User Manual for Vital Signs Monitor    Chapter 9 Maintenance    9 1 Service and Examination    9 1 1 Daily Examination    Before using the monitor  the checks below should be carried out     Check the monitor for any mechanical damage   Inspect the exposed parts and the inserted parts of all the leads  and the accessories     Examine all the functions of the monitor that are likely to be used for patient monitoring  and    ensure that it is in good working condition   Make sure that the monitor is grounded properly     Pay close attention to the fluctuation of the local power supply voltage  A manostat is    recommended when necessary     In case any indication of damage about the function of the monitor is detected and proven  it is not    allowed to apply 1t to the patient for any monitoring     9 1 2 Routine Maintenance    After each maintenance or the yearly maintenance  the monitor can be thoroughly inspected by qualified    personnel  including function and safety examinations     NZ    NZ    If the hospital fails to carry out a satisfactory maintenance program about the monitor  it may    get disabled and harm the patie
22. gh pressure sterilization to the monitor     Do not put any parts of the monitor in the liquid     33    User Manual for Vital Signs Monitor     amp  Donot pour the disinfector on its surface while sterilization   9 3 Cleaning  Sterilization and Disinfection of Accessories    It is recommended to use a cloth dampened with isopropyl alcohol 70   a 10  aqueous solution of  sodium hypochlorite  bleach   a 2  gluteraldegyde solution  ammonia  mild soap or disinfectant  spray cleaner to clean the accessories  including sensor  cable and plugs      6 Do not use damaged accessories     Accessories can not be entirely immerged into water  liquor or cleanser     6 Do not use radial  steam or epoxyethane to disinfect accessories     9 4 Storage    If the equipment will not be used for long period of time  wipe 1t clean and keep it in the packaging  which    shall be kept in a dry and good ventilation place free from dust and corrosive gases     9 5 Transportation    This monitor should be transported by land  vehicle or railway  or air in accordance with the contractual    terms  Do not hit or drop it with force     34     User Manual for Vital Signs Monitor    Appendix    I Prompt information explanations    Prompt Info Response    Check that the sensor is properly plugged in  Reinsert the  sensor  If error persists  contact the local dealer or  manufacturer for servicing     Need CO  zero To clear  check airway adapter and clean if necessary  If this  does not correct the error  perf
23. he upper alarm limit of respiration rate  Range     1 150    rpm     RR Low     Setting the lower alarm limit of respiration rate  Range     0 149   rpm     Gain     CO  waveform scale       x1 2     waveform reduced to half of nominal scale      x1     nominal waveform scale    x2     waveform with doubled scale       Baro     barometric pressure   set ambient barometric pressure  It can be determined by  barometer or the ambient altitude  Altitude can be used to determine the typical barometric pressure  if a barometer is not available  refer to Appendix Typical Pressures and CO  Readings at Altitudes  for details  Range     400 850   mmHg     Note  When the cursor stays at    Baro    item  press    W     key to move cursor to next item at another    CO     Setup    screen view as shown in Figure 4 7B CO  Setup      lt  gt         gt  gt      Page down icon  When cursor stays in the Baro filed  press     Y    key to enter CO  Setup II    Screen        Zero     CO  Zero   The CO  zero must be carried out in a drafty place  Set the option as    on    and  press    W    key to conform  and then the result will be displayed on the screen several seconds later   When the CO2 zero 1s done successfully  the set of this 1tem will automatically change to    off    and    the corresponding information will display in prompt info area     Note  you should perform CO  zero only when    Need CO  zero    prompts on the screen  Otherwise  1t 18    unnecessary to perform this operation    
24. hift cursor to corresponding functional group setting     2  Pres   W     key to confirm and enter into corresponding functional parameter setup screen   3  Pres    key  under the setup menu or its submenu  the SpO  waveform will be printed   4  Press     m     to exist from Setup Menu Screen     Note  At Setup Menu Screen or its submenu screen  when pressing   Ey   key  the default display screen    will be printed     The following will cover each functional parameter   s setting up     4 4 1 CO  Setup    CO  Setup CO  Setup       Probe off 01 19 30 07 Probe off 01 19 30 09    Figure 4 8A CO  Setup I Figure 4 8B CO  Setup II    Screen Description      gt     Switch     choosing the mode of CO   It is recommended that the switch is turned on only when    21     User Manual for Vital Signs Monitor    there 1s a need to monitor CO  parameter  This can not only reduce the power consumption and also  extend the life of the CO  module        Unit     choosing the EtCO  and the minimum InsCO   It can be set as kPa  mmHg  and          Et A      EtCO  alarm switch     A    indicates EtCO  alarm is on      X     indicates EtCO  alarm 1s  off        Et High     Setting the upper lower alarm limit of EtCO   Range     1  160   mmHg     Et Low     Setting the upper lower alarm limit of EtCO   Range     0 159 mmHg       RR A      respiration rate alarm switch     422    indicates respiration rate alarm is on      X     indicates    respiration rate alarm is off       RR High     Setting t
25. is in demo status  The entire signal 1s generated  During DEMO from the monitor and waveforms and values are displayed for  demo     Low battery Low bullt in battery voltage  Please recharge the battery     235           User Manual for Vital Signs Monitor    II Typical Pressures and CO  Readings at Altitudes    EtCO  Reading       236     Manufacturer     Shenzhen Creative Industry Co   Ltd   Address     2 F  Block 3  Nanyou Tian an Industry Town  Shenzhen  GD  518054 P R  China  Tel  486 755 2643 3514    Fax  486 755 2643 0930  E mail  market creative sz com 1 01 1    Website  www creative sz com    
26. isplay area  If the CO  sensor cable is disconnected  from the monitor  the CO  waveform will become a base line in main display area  as shown in  Figure 4 3      Prompt Info Area   lt  gt  Status or event indication segment     If the CO  sensor cable 1s disconnected from the monitor  the CO  waveform will become a base line     17     User Manual for Vital Signs Monitor    in main display area and    No CO2 sensor    will appear at the left side of prompt info area  as shown in  Figure 4 3      Note  CO  sensor needs to warm up  Prior to CO  measurement  please wait for sensor warm up     Otherwise  the measuring result may be inaccurate     4 2 2 Operation Instructions    ip 4 OF     key  press this key to shift display mode   a RS Ey     key  press it to print a trace of CO  waveform  press it again to stop printing   a Ss x     key  press 1t to enable disable alarm silence      gt     A    key or    VW    key  press it to change CO  waveform scale  It can also be set in the item of     Menu Setup    CO  Setup Gain         18     User Manual for Vital Signs Monitor    4 3 Trend Graph Display    Short pressing     mi  Display    key to shift the screen view to trend graph display screen  as shown in Figure  4 5     Trend graph    01 50 05 50  Trend length    Prohe off 01 17 44 59    Figure 4 5 SpO  Trend Graph       4 3 1 Screen Description     lt  gt     12 hours     the trend length of trend graph  three options     12        24    or    96    hours  when the  selection
27. l which lasts for 1 second  When several alarms    occur at the same time  only one pulse signal will be sent out     Note     Nurse Call function can not be regarded as main alarm notice method  please do not entirely  relay on it  You should combine parameter values with alarm level and patient s clinical    behavior and symptom to determine patient s status     4 4 4 System Setup    bh               Lang  TENG      hackl  te contract  care mode BT SD    01 20 14 50       Figure 4 11 System Setup Screen     VOL     set beeper volume     0 7    level adjustable  the set    0    1 e  no sound      key beep     to turn on off key beep      LANG     language selection  Two options     ENG    for English or    CHN    for Chinese        priority     priority of PR value source displayed on the numerical LED  The default set is PR  1t can    not be set      backlite     turn on off backlight      contract     adjust LCD display contract     0 31    level adjustable        care mode        Demo    shows the demo waveforms and data  In the demo state  all the signals and  data are generated from the patient monitor for demonstration and testing purpose     Real    shows the    real time waveform  1 e  normal monitoring status     BT SD  turn on off the pulse beeping sound      24     User Manual for Vital Signs Monitor    4 4 5 Patient Info    Patient Info    m       Probe off OL 12 14 50    Figure 4 12 Patient Info Screen   lt  gt     ID     change or set current patient   s ID num
28. ll display SpO    EtCO  value and the time value at the point where the cursor stays  Move cursor back and forth  this way  you can view the SpO   EtCO  trend  12 24 96 hours long   Press    MW    key again to exit  trend viewing    e When pressing    A    key or    V    key to move cursor  the moving step 1s variable  The rule is that  the initial step is 1 point  after pressing    A    or    Y    key towards the same direction for 5 times   the step becomes 5 points  and with 5 more pressing the step becomes 10  then 20  No matter what  step is  as long as you press    A    or    Y    key towards the other direction  the step becomes 1 and  towards the other direction     5  press   a mi      key  press this key to shift display mode  s Es     key  press it to print the current displayed trend graph     s x     key  press   t to enable disable alarm silence      20     User Manual for Vital Signs Monitor    4 4 Setup Menu Screen    At any display view screen  long time press     mi  Display    key to shift the screen to Setup Menu screen  as    shown in Figure 4 7  All the functional parameters of the system can be set through Setup Menu     Setup Menu    CL Patient Info    Sp Date   Time  Nurse Call Default    System    Probe off 01 19 56 07       Figure 4 7 Setup Menu Screen  There are 7 functional groups for setting parameters     CO   SpO    Nurse Call  System  Patient Info  Date Time    and Default    on the Setup Menu Screen   1  Press    A    key or    V     key to s
29. ments at hospital  such as  Intensive Care Unit  ICU   Emergency Department  Respiratory Care and Operation Room  It  also adapts to use in monitoring patient during inter hospital transport and intra hospital  transport  for example road ambulance  fixed wing and rotary wing aircraft  and first aid  station     User Manual for Vital Signs Monitor    1 4 Symbols on the Monitor    Pa    nh  lt  nH  gt     Alarm Silence    Print    Up    OK    Down    Setup Menu    AC Power    a A ls    a  r    Qt  amp     DC Power    Type BF applied part    Type CF applied part with    defibrillator protection    Warning  refer to User Manual     Equal potential terminal    Nurse call output    User Manual for Vital Signs Monitor    Chapter 2 Operating Principle    2 1 Conformation    The monitor consists of the main control unit  power supply board for CO   SpO  module  display panel     printer  power supply block  etc  and the related accessories for CO  sensor and SpO  measurement   2 2 Overall Structure    The overall structure of the monitor is shown in Fig  2 1     Operating Power Suppl    COZ Sensor     EUA  Power Supply     nsco2  RR H  Board for CO      LED Display    Pulse Oxygen Central   p 02 Probe EY Saturation Module FT      LED Display   S02  PR  Processing  Networking   Built in Printer    Figure 2 1       PC 900A Vital Signs Monitor is module designed product  it consists of power supply board for CO    CO   module  SpO  module  main control unit  built in printer  optional  
30. monitor     Note  Short time pressing power button for entering the Power Saving Mode screen  then according to    your need to make the device stay in the power saving mode or exit from power saving mode  this  function is optional and needs hardware support         20 SpO   SpO  sensor connector  21 CO   CO  sensor connector  22 TEMP  temperature probe connector  reserved     23 ECG  ECG cable connector  reserved     3 1 2 Side Panel    The built in thermal printer is on the left panel  It is easy for user to print waveform and data     3 1 3 Rear Panel       Figure 3 2 Rear Panel    Introduction to the rear panel     Handle   Fan   Nameplate      FUSE 2XT3 15 A     Fuse holder  Fuse specification  T3 15AL 250V D5 X20mm      AC100 250V      AC power supply socket   Loudspeaker   Mounting hole for hanging the monitor    NET  serial communication port which 1s used to network with central monitoring system  optional            NI Mm YN A UQ N m    Nurse call connector    10 Y  Equipotential ground terminal    _6     User Manual for Vital Signs Monitor    3 2 Installation    3 2 1 Opening the Package and Check    1  Open the package  take out the monitor accessories from the box carefully and place it in a safe    stable and easy to watch position   2  Open the accompanying document to sort the accessories according to the packing list     Inspect the monitor for any mechanical damages      Check all the accessories for any scratch or deformity  especially on connector  wire a
31. nd InsCO  measuring unit indicator  the unit can be set in the item of    Menu    Setup   CO  Setup    unit     Three options     mmHg        kPa        Yo        IK Alarm silence key  Enable disable alarm silence function  When the alarm silence indicator  on the left of keys is on  it means the system is in alarm silence status and it lasts for 2 minutes   When finishing counting down  the system will resume normal alarm status automatically  1f alarm    event occurs at this time the alarm sound will be effective again    Alarm silence indicator  When it is on  it indicates that the monitor stays in alarm silence status   Es Print  print SpO CO  trend and waveforms    A Up  shift cursor forward upward   OK  to perform confirmation or modification   V Down  shift cursor backward downward   Y Display  shift LCD display modes      AC Power indicator    y    DC Power indicator    The monitor is using mains power supply  ON  green  ON  green  and the battery is full  when the battery is  installed      ON  green  The battery is being used  FF    Status Ser The battery   s being used  but battery  O ON  orange  blinking   voltage is low  the beeper also gives warning    The monitor is using mains power supply  ON  green  ON  orange   and the battery 1s being recharged       The battery 1s being recharged while the  ON  green  OFF  monitor 1s off     User Manual for Vital Signs Monitor       19 O  Power button  Press power button for 3 seconds to start the monitor or shut off the 
32. nd probe  parts    If in doubt  please contact the local dealer or our company in case of any problems  We are to offer  you the best solution for your satisfaction     3 2 2 Connecting the Power Supply    1  When powered by AC mains power supply   Make sure that the AC mains power supply is within 100 250VAC with 50Hz or 60Hz       Use the power cord provided by the manufacturer  Insert one end of it to the power port of the  monitor and the other end to the single phase mains power outlet with protected earth       Caution  if necessary  make the monitor grounded properly by the provided grounding wire   2  When powered by built in battery  9 Install battery  refer to Chapter 3 3 4 Battery Installation       Caution  it s better to recharge the battery after it is used up  the charging time should be  13 15 hours long     3 2 3 Starting the Monitor    The system performs self test and enters initial display after switching on the monitor  and the orange alarm    indicator blinks to inform that the user can begin operating it   9 Check all the applicable functions to make sure that the monitor works normally       If the battery is applied please recharge it after using the monitor to ensure sufficient power  storage     A Do not use the device to monitor the patient if the device appears obvious damage or indication of fault     Please contact the local dealer or our company     A After the monitor is switched off  it s recommended to delay 1 minute to start it again     ae
33. nt s safety and health   In case of ECG leads damage or aging  please replace the lead       If there is any indication of cable and transducer damage or they deteriorate  they are    prohibited from any further use     The adjustable units in the monitor such as potentiometer are not allowed to adjust without    permission to avoid unnecessary failures that affect normal application     9 1 3 Battery Maintenance    NZ    Please pay attention to the polarity of battery  do NOT insert it into battery compartment  with reversed polarities     Do NOT use the batteries manufactured by other companies  if being inserted  the device will    may be damaged     In order to avoid damaging the battery  do NOT use other power supply device to charge the  battery     305    n  VIN    p dl dl dl dl  YIN WIN WIN WIN    a    User Manual for Vital Signs Monitor    After battery ageing phenomenon occurring  to avoid explosion risk do NOT throw the    battery into fire     Do not hit or strike it with force   Do not use this battery on other devices   Do not use this battery below  10  Cor above 40  G    Dispose of the battery  the local law should be followed   It is recommended to use the battery once a month to ensure its strong power supply capacity    and long service life  and recharge it after running out of the power     In order to maintain battery supply time and prolong battery lifetime  please use the battery  once a month and do not charge it until it is used up each time     Note
34. ons    Ks mi      key  press this key to shift display mode       Ey     key  press it to print a trace of SpO  plethysmogram  press it again to stop printing       x     key  press 1t to enable disable alarm silence     e    User Manual for Vital Signs Monitor    4 2 CO  Monitoring Screen    Insert the CO  sensor cable into the connector labeled    CO     follow section 3 3 1  s instruction to connect the  CO  sensor and its accessories  and then connect to patient well  next turn on the CO  Switch Short time press       mi  Display    key to shift the screen view to CO  monitoring screen  as shown in Figure 4 3   CO       xl       No CO2 sensor OS  11 30 25      Figure 4 3 CO  Monitoring Screen  No CO  sensor     If the CO  sensor cable is disconnected from the monitor  the CO  waveform will become a base line in  main display area  as shown   n Figure 4 3   When CO  tube is attached on the patient and CO  sensor  cable 1s connected to the monitor well  CO  waveform will be displayed in the main display area  as    shown in Figure 4 4        CO2    03 11 30 25  Figure 4 4 CO  Monitoring Screen    4 2 1 Screen Description    Title area    lt  gt      CO       indicate the current waveform displayed is CO  waveform   Main display area      lt  gt     x1     CO  waveform scale  Three options     x1 2     waveform reduced to half of nominal scale        x1     nominal waveform scale     x2     waveform with doubled scale      lt  gt  CO  waveform will be displayed in the main d
35. orm an adapter zero     One of the following conditions exist   This error occurs whenever the pneumatic pressure is outside  the expected range     CO  module fail       Check airway    Check that the sampling line is not  occluded or kinked      To clear  clean airway adapter if  mucus or moisture is seen  If the  adapter is clean  perform a Capnostat  ZETO         Usually caused when the airway adapter is removed from the  sensor or when there is an optical blockage on the windows  of the airway adapter  May also be caused by failure to  perform sensor zero to when adapter type 1s changed     Check that the sensor is properly plugged in  Reinsert the  sensor     The CO  module cable is not plugged in  Please connect the  No CO  sensor    CO  module to the monitor properly     One of the following conditions exist    Breaths have been detected by the CO  module within the last  20 seconds while CO  zero was attempted    Check that the sensor is properly plugged in  Reinsert or reset  the sensor 1f necessary    Temperature 1s not stable    The set of CO  monitoring switch 1s    off     Please set as    on     in CO  Setup screen     One of the following conditions exist      The sensor is connected unwell            CO  zero fail       Sensor warm up    Sensor under temperature     This condition 1s normal at startup   Temperature not stable       CO  is sleevin The set of CO  monitoring switch 1s    off     Please set as    on     2 ping in CO  Setup screen     The monitor 
36. pecific wavelengths   with the amount of energy absorbed being directly related to the CO  concentration  When an IR light beam  is passed through a gas sample containing CO   the electronic signal from a photodetector  which measures  the remaining light energy   can be obtained  This signal 1s then compared to the energy of the IR source  and  calibrated to accurately reflect CO  concentration in the sample  To calibrated  the photodetector   s response    to a known concentration of CO  is stored in the monitor  s memory     The monitor determines CO  concentration in the breathing gases by measuring the amount of light absorbed  by these gases  FtCO  is display as a numerical value in millimeters of mercury  mmHg   percent      or  kilopascals  kPa   In addition  a CO  waveform  capnogram  may be displayed which is a valuable clinical  tool that can be used to assess patient airway integrity and proper endotracheal tube placement  Respiration    rate 1s calculated by measuring the time interval between detected breaths   7 1 2 Mainstream vs  Sidestream Sampling    Mainstream CO  sensors are placed at the airway of an intubated patient  allowing the inspired and expired  gas to pass directly across the IR light path  The major advantages of mainstream sensors are fast response  time and elimination of water traps    Sidestream CO  sensors are located away from the airway  requiring a gas sample to be continuously  aspirated from the breathing circuit and transported to the
37. prilocaine  lidocaine and butacaine may also be a major    factor blamed for serious error of SpO  measurements     4  As the SpO  value serves as a reference value for judgement of anemic anoxia and toxic anoxia  the    measurement result of some patients with serious anemia may also present as good SpO  value      30     User Manual for Vital Signs Monitor    Chapter 8 Troubleshooting    8 1 No Display on the Screen    Shut down the machine and unplug the power  Use a universal meter to check 1f the outlet has proper voltage   if the power cable is in good condition  and if the power cable is properly connected with this apparatus or    outlet  Remove the fuse from the back cover of this machine  and make sure it is in good condition     8 2 No EtCO  and SpO  Result       1  Check that the CO  module cable 1s properly plugged in  Reinsert or reset the sensor if necessary and    the set of CO  monitoring switch is on   2  Check that the SpO  sensor cable is properly plugged in and the light in SpO  sensor is flash     3  Ifthe problems still exist  please contact the local dealer     8 3 Blank Printed Paper    1  Check whether the printing paper is installed with its face reversed  Please reinstall it and let the    sensitive page face upward     2  If the problems still exist  please contact the local dealer     8 4 System Alarm    1  When the parameter value is higher or lower than the alarm limits  the alarm will ring  Please check    whether the alarm limit value is prop
38. rmation    very carefully before using this device     Read these instructions carefully before using this monitor  These instructions describe    the operating procedures to be followed strictly  Failure to follow these instructions can    cause monitoring abnormity  equipment damage and personal injury  The manufacturer    is NOT responsible for the safety  reliability and performance issues and any monitoring    abnormality  personal injury and equipment damage due to user s negligence of the    operation instructions  The manufacturer  s warranty service does not cover such faults     NZ  NZ  NZ    Do not use this device in an environment with ignitable or inflammable gas   Do not place the monitor in any position that may cause it to fall on the patient   Check the safety and performance of this monitor every time before using it to    ensure it works normally and safely    Ensure the monitor is grounded reliably before using it  if the integrity of  grounding connection is in doubt  the monitor should work by using the built in  battery     6    Equipment connected with this monitor should be IEC 60601 1 complied     6 Turn off the monitor and take away the sensors from the patient during MRI    scanning  Otherwise it may cause burn to the patient and the quality of MRI image  or the measurement accuracy of the monitor may be affected     Although biocompatibility tests had been done to all the applied parts  some  exceptional allergic patients may still have anaphylaxi
39. s  Do NOT apply to those  who suffer from anaphylaxis     All the cables and hoses of the applied parts should be kept away from the patient s  neck to prevent any possible choke of the patient     All the parts of the monitor should NOT be replaced at will  If necessary  please use  the components provided by the manufacturer or those of the same model and  standards as the accessories along with the monitor which are provided by the same  factory  otherwise negative effects concerning safety and biocompatibility  etc  may  be caused     If the monitor falls off accidentally  please do NOT operate it until its safety and  performance have been carefully tested and positive testing results obtained     Dispose of the device and its accessories  the local law should be followed     User Manual for Vital Signs Monitor    Table of Contents   CHAPTER UOVERVIEW usa aran convent ae 1  a AA NA l  12 PRODUCT NAME AND MODE Ltda kakanan a R E kan ana l  1 3 INTENDED USE id 0a ea haha l  1 4 SYMBOLS ON THE MONITOR ii 2  CHAPTER 2 OPERATING PRINCIPLE 2 0 AA 3  2 1 2  ONOR NANO Naa AO NAA kaa 3   2 OVER O ROTURA ANA ANA NAA AA AA ee ee 3  CHAPTER 3 INSTALLATION AND CONNECTION         22000000000000000000000000000000000000000 0000000000000 ns snnnnnnnnne 4  3  AP BE AIAN CB un eisen 4  SL TONG Panel naaa ANNA AA AA eee aaa  4  ML VO LON ARA AAA AAA ec Ei eee     O A L O O      32 INS ALTO Adoos 7  3 2 1 Opening the Package and nl E ee E 7  322 C ONNCCHNS Ihe POWER SUP DIY mawa ksa A N A E 
40. se     10     User Manual for Vital Signs Monitor    entanglement or strangulation   6 Support the airway adapter to prevent stress on the ET tube     6 Reuse  disassembly  cleaning  disinfecting or sterilizing the single patient use CO   airway adapters may compromise functionality and system performance leading to a  user or patient hazard  Performance is not guaranteed if an item labeled as single    patient use is reused     6 Inspect the sidestream on airway adapters and sidestream sampling kits for damage  prior to use  DO NOT use the sidestream on airway adapters and sidestream sampling  kits if they appear to be damaged or broken     6 If the CO  waveform  Capnogram  appears abnormal  inspect the CO  airway adapters  and replace if needed     6 Periodically check the CO  Flow sensor and tubing for excessive moisture  or secretion buildup  Do not use then if there is excessive moisture or exterior    condensation   6 Electric Shock Hazard  The CO  Sensor contains no user serviceable parts     6 Refer service to qualified service personnel  Do not open the sensor cabinet at will  as  electric shock hazard may occur     6 Place the exhaust vent of the CO  Sensor in drafty ambient and do not let anything  block the exhaust vent     6 Always disconnect the CO  Sensor before cleaning  Do NOT use if it appears to have  been damaged  Refer servicing to qualified service personnel     6    DO NOT sterilize or immerse the CO  Sensor in liquids     Es    When changing sampling t
41. ube  it is suggested choosing the default sampling  tube with dehumidifying function  The sampling tube without  dehumidifying function may be easily blocked by excessive moisture   Use  life  ordinary sampling tube  6 12hours  the sampling tube with  dehumidifying function  about 120hours      Es    If the measurement appears abnormity caused by sampling tube block  please replace  it    Es    The total length of the sampling tube and extending airway tube shouldn t  be longer than 3 meters  too long may cause measurement abnormity  If    using T connector sampling cannula kits  please insert the sampling tube  with the tubes upward to avoid the affects of excessive moisture     Es Altitudes are different in different area  so set the Barometric Pressure    setting value as the ambient barometric pressure     Es    When stopping CO  monitor  please disconnected the CO  sensor from the    patient monitor   3 3 2 SpO  Sensor Connection    SpO  sensor is a very delicate part  Please follow the steps and procedures in operating it  Failure to operate it    IT     User Manual for Vital Signs Monitor    correctly can cause damage to the SpO  sensor     Operation procedure   1  Connect the SpO  sensor to the connector labeled    SpO      When unplugging the probe  be    sure to hold the head of the connector and pull it out     2  Ifthe finger clip SpO  sensor is used  insert one finger into the sensor  index finger  middle  finger or ring finger with short nail length  as shown in
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
BAZZ 300LED5W Installation Guide  アナログ&デジタル2系統入力。 - 株式会社 エイ・アイ・ディー  TE DRS-M  Samsung Galaxy A5 Vartotojo vadovas  取扱説明書 ルータ 電子ルータ  DeLOCK 15m USB 2.0 AM/BM  Inspector PI50  Tabletop Controller  User`s Manual - Reliability Direct, Inc.  A/Cシ ステムに彰害を与えす、保書です。 レシーバー ドライヤを交換する    Copyright © All rights reserved. 
   Failed to retrieve file