        User Manual - Lactate Scout
         Contents
1.    Explanation of symbols  Symbols on containers and device labels  on instructions included in the packaging and in the operating instructions    Meaning    In vitro diagnostic   Product meets the requirements of the Directive regarding in vitro diagnostics 98 79 EC  Manufacturer   Follow the operating instructions  read the package information leaflet   Intended only for single use   Expiration date   Batch designation   Storage temperature range   Wireless data transfer with Bluetooth   module    Do not expose to direct solar radiation    Hazardous waste  do not dispose of in household waste    Direct current          34    Accessories    Purchasing accessories  The following accessories are available at EKF diagnostic GmbH or at specialist retailers   as listed on the product website www lactatescout com     BigPack 72   3 x 24 Lactate Scout  sensors in sensor containers  BigPack 48   2 x 24 Lactate Scout  sensors in sensor containers  BigPack 24   24 Lactate Scout  sensors in sensor container    Lactate Scout Test Solution 2 mmol L  Set of 5 x 2 5 ml bottles with 1 8     2 2 mmol L lactate     Lactate Scout Test Solution 5 mmol L   10 mmol L  Combi pack of 2 5 ml bottles  2 x 5 units   with 4 5     5 6 mmol L and 8 9     11 1 mmol L lactate     Lactate Scout Test Solution 10 mmol L  Set of 10 x 2 5 ml bottles with 8 9     11 1 mmol L    SurgiLance Safety Lancets grey  G23    Additional product information  literature and references  can be found online at www lactatescou
2.    Sales and Service   EKF diagnostic GmbH  Ebendorfer Chaussee 3  39179 Barleben    Manufacturer    SensLab GmbH   An EKF Diagnostics Company  Bautzner Str  67   04347 Leipzig    Phone   49  0  39203 785 0  Fax   49  0  39203 785 24  www ekfdiagnostics com  info ekfdiagnostics com    Phone   49  O  341 234 18 41  Fax   49  O  341 234 18 40  www senslab de  info senslab de    ar       DIAGNOSTICS  SCIENCE MADE SIMPLE    Part number 7023 9012 2350  Rev  2 0 11 2013       User Manual       PE p  DIAGNOSTICS  SCIENCE MADE SIMPLE    Copyright   2013 SensLab GmbH    No part of this publication may be  reproduced  transmitted  transcribed   stored in any retrieval system or  translated into any human or computer  language by any means or in any form   without the prior written permission of  Senslab GmbH     Part Number  7023 9012 2350  Rev  2 0  11 2013   Lactate Scout  User Manual  Valid from serial no  XXXXX02135    CE 0123 IVD  FDA   C    This product has been tested to the requirements  of CAN CSA C22 2 No  61010 1  second edition   including Amendment 1  or a later version of the  same standard incorporating the same level of  testing requirements     Manufacturer          An EKF Diagnostics Company    SensLab GmbH   Bautzner Str  67   04347 Leipzig   Germany   Phone    49  0341   234 18 41  Fax    49  0341   234 18 40    www senslab de    Sales and Support     P    ooo d  DIAGNOSTICS  EKF diagnostic GmbH  Ebendorfer Chaussee 3  D 39179 Barleben  Germany    Service   49  0  3
3. 2  LOAD  Load  for load values   3  AFL  Afterload  for afterload values     It is possible to begin with any of the step test  phases  however it is not possible to change  later   e g  a preload cannot be selected after  the load         16    The number of the lactate concentrations  recorded per step test phase correlates to  the number of conducted tests  The count of  tested values occurs separately for every step  test phase and is automatically displayed on  the screen to the right of the respective step  test phase symbol  4   One test is intended  for every load step of the step test phase      LOAD         Upon inserting a new sensor  the last selected  step test phase is preset and is displayed  together with the next digit for the continued  test  Now you can perform this test or change  by scrolling up to the following phases or exit  the step test mode     To exit the step test mode  turn the scroll  wheel directly following the test while the  sensor  5  is still inserted until    End     6   appears on the display  Confirm by pressing  the scroll wheel  Exiting is also possible prior  to a new test  In this case  after achieving test  readiness  7   scroll to    End     8  and briefly press  the wheel  The Lactate Scout  is now in the  individual test mode once again  blinking drop      Attention  One hour after the last step test  measurement  the measuring device automatically  returns to the individual test mode        Performing step tests    Afterload ti
4. 9 203 78514  Email  support ekf diagnostic de    www ekfdiagnostics com    Intended use and safety guidelines    The Lactate Scout  may only be used  with the Lactate Scout sensors to  determine the lactate concentration in  capillary whole blood  The Lactate Scout  sensors are only for single use     Only for external application outside the  body   In vitro diagnostic application     Lactate Scout  is suitable for self testing    It can also be used by qualified staff in  various medical faculties  Test results  should only be used as guidance in making  decisions for further treatment of patients  following a suitable period of evaluation     Keep Lactate Scout  and accessories out of  the reach of children  choking hazard due to  small parts      The principal risk of infection is in the  handling of blood samples when measuring  lactate concentrations     e Only undamaged and unused lancets  and sensors may be used  Used sensors   lancets and cellulose cloths pose a risk of  infection  These must be safely disposed of  as household waste following use     e Protective gloves must be worn when  performing tests on third parties     e The sterile single use lancets may only  be used to draw capillary blood on the  fingertips or earlobes     Unpacking   Unpack the system and check that the  following components are present and  undamaged in the pack  If there is anything  missing contact customer support     Lactate Scout  Analyzer   1 5V AAA LRO3 batteries  2x   User manua
5. Scroll up and down to view menu  options  Press the wheel to access the menu  Scrolling  Select option or search    Pressing  Activate confirm selection or  deactivate cancel selection  Turn on or off by pressing for 2 seconds      Measurement chamber  4  with opening for  the collection of the blood sample     Connecting contacts  5  must be inserted  completely into the insertion slot of the  measuring device with the black contact  surface facing up       Batteries are included with your Lactate Scout      Two LRO3   AAA  1 5 V  batteries must be  inserted prior to the initial use of your Lactate  Scout   Alkaline or zinc carbon batteries can  be used  Only use batteries of the same type   If the time to replace the batteries exceeds  one minute  the date and time setting will be  lost and must be re entered  see page 20   The  stored data and settings will be kept  however     The Lactate Scout  must be turned off to  replace the batteries  Gently  push the battery  cover outward  Remove the used batteries   When inserting the new batteries  ensure that  the polarity is correct        The batteries must  be inserted in the same direction  Push the  battery cover back until it snaps into place     Attention  In the case of prolonged periods    of non use and when flying  the batteries must  be removed           Please note  Used batteries must not  be disposed of with household waste   Return them to the shop or any public  collection point        Getting started    Code s
6. The resting values  depend considerably on the physiological  condition of the subject as well as the previous  nutritional intake  In some cases  the value can  be increased due to illness     After exercise  the lactate value can be more  than 20 mmol L  particularly in those engaged in  intensive professional sports depending on the  duration and intensity     Turning the device off   The device is turned off in the test mode by  removing the sensor  Otherwise  the Lactate  Scout  will turn off automatically after five  minutes     Deactivate the step test mode   In order to prevent accidental switching   into the step test mode or if step test  measurements should always be recorded as  single measurements the step test function  can be deactivated using the Lactate Scout  Assistant Software Ver  1 0 6  The setting can  be re activated in the same way  For detailed  instructions see Lactate Scout Assistant  Software manual     Selecting and modifying settings    Settings are made in the  Settings and storage  mode     The settings are only accessible when  no sensor is inserted  Press scroll wheel for two  seconds until a signal sounds and the display  with all elements and symbols appears     Upon release the first submenu    set CODE     symbol is shown  Move between submenus by  scrolling the wheel  Accessibility of the submenu  is indicated by blinking     Briefly press the scroll wheel to activate a  submenu  Scroll to perform the desired settings  in the submenu  Pr
7. ambient temperature  ensure that the  temperature sensor is not touched by hand  or covered    The ambient temperature of the device can  be displayed for 2 seconds by pressing the  scroll wheel briefly in test mode        Attention  Ensure that the measuring device  and sensors do not vary considerably from   each other in temperature prior to a test  e g   for winter sports or after storing the sensors in  the refrigerator   If necessary  allow sufficient  time for the temperature of the sensors and the  measuring device to adjust           Getting started    10    Care and cleaning    The Lactate Scout  measuring device is an  electronic precision measuring device and must  be handled and stored with care  It must be  protected from liquids  moisture  prolonged  solar radiation and excessive effects of cold and  heat  below 0  C and above  50  C      Heavy mechanical loads  improper handling  and contamination can partially or completely  restrict functionality  Ensure that no liquid or  dirt enters the inside of the device through the  openings  insertion slot for sensor  scroll wheel  and battery housing      If blood  test solution  dust or other  contaminants are present on the housing  it  must be cleaned using a soft  lint free towel  or cellulose cloth  e g  paper towel   which  has been pre soaked in a mild cleaning agent   e g  water with a minimal amount of washing   up liquid   and then wiped dry  Medical   staff should adhere to the regulations of the  medical 
8. been  inserted in the device so that the blood can    be absorbed  3      Once the measurement chamber of the  sensor is completely filled  an acoustic  confirmation signal sounds and the test  procedure will be activated     LAC    and a    circular symbol with a progress bar appear  on the display as well as the assigned memory  space number  4      After ten seconds  a second acoustic signal  sounds and the test result is displayed in  the measurement unit    mmol L     5     This is stored automatically along with   the date and time of the test     Attention  There is a risk of erroneous   segmented or insuffiecient  filling of the  measurment chamber of the sensor  This may    occur if the subject or user is moving when a  sample is taken  To avoid this  ensure that the tip  of the sensor remains in contact with the drop of  blood whilst the measuring chamber is filling        Caution  The safety lancet is intended for    single use and must be disposed together with  the used sensor  household waste            Performing step tests    Settings for step tests    Lactate Scout  allows for the recording of   values for step tests  The general procedure for  preparation  sampling and measurement is the  same as the individual test  see pages 12   15      After inserting a sensor and confirmation   of test readiness  blinking drop  select the  appropriate step test phase for the subject   The step test phases are displayed as follows        PE Preload  for resting values   
9. ess scroll wheel briefly to  confirm the setting  Exit the settings menu by  pressing the scroll wheel for two seconds      set CODE   enter the correct Code for the  sensors in use  for detailed description see  page 8      set   DUR   at selection of this submenu the  current step duration appears  Set desired step  duration using the scroll wheel as described  above  Maximal duration of one step is 10    20 minutes  Minimal setting is 10 seconds      COUNT DOWN   by activation of this submenu  the count down for the previously selected  step duration is started  A short signal tone will  sound after each minute and at every second  during the last 10 seconds of a step  An initially  filled black circle will empty quarterly after 25    50   70  and 100  of the step duration  The  elapse of the step duration is indicated by brief  consecutive signal tones and blinking    0 00        CLOCK   activation of this submenu starts   the stop clock immediately  Full minutes are  indicated by a brief signal tone  The displayed  circle fills consecutively with one black quarter  every 2 5 minutes  After 10 minutes the device  switches back to the    CLOCK    menu  The stop  clock can be stopped during the 10 min interval  by briefly pressing the scroll wheel  A signal will  sound and the time will be stored together with  the next lactate test result      set   Time Date   set correct hour by turning the  scroll wheel  confirm by briefly pressing the scroll  wheel  Repeat consecutive 
10. est solution bottle  After initial  use the test solution bottle must be firmly closed  for further storage and must be used within 3  months     Preparing the functionality test  Allow the test  solution  the sensors and the device to adjust to  ambient temperature before performing the test     Open the test solution bottle and wipe off the  opening  Discharge the first droplet and press  out the next droplet onto a clean non absorbent  surface     Activate the submenu Ctrl D S as described  under settings  In the display the letters  dSC   are shown blinking  Insert sensor  The code  display appears briefly and switches to the  blinking drop symbol indicating that the device   is ready to test  At the same time the target value  is displayed with the measurement unit  mmol L      Turn the scroll wheel to select the appropriate  target concentration of your test solution  Do  not confirm by pressing the scroll wheel  Touch  the drop of test solution with the sensor like  when performing a regular measurement  see  instructions page 14 15     The test starts and the round progress display  is shown together with  dSC   An acoustic  signal indicates that the test is completed  The  measured lactate concentration is displayed     27    For test solutions of 10 mmol L the result is  evaluated by the device      OK   the test result is within the specified  limits  see label of the test solution bottles  If the  test result is outside the specified range a plug  connector test i
11. etting   The first time the Lactate Scout  is used   a two digit code must be set  This code is  printed on the label of the sensor vial       Turn on the Lactate Scout  device by  pressing the scroll wheel for two seconds   The request to set the code appears on the  display  blinking     set CODE     1       After pressing the scroll wheel again briefly   the code setting menu is activated and the  left code number will blink  2   Next  set the  left digit listed under    code    on the sensor  container by scrolling up or down and finally  confirm by briefly pressing the scroll wheel      To enter the right code digit  which will now  be blinking  3   repeat this process  The  device is now coded  4     The finalisation of coding is confirmed by a   signal tone  To complete the process  insert   a sensor or turn off the device     Attention  Prior to using a new or other sensor  container  it is necessary to always check whether the    code on the label corresponds to the current code in  8 the device  If necessary  the latter must be reset        Temperature sensor    In order to be able to achieve equally reliable  test values under different climatic conditions   the Lactate Scout  is equipped with a  temperature sensor  It is located underneath  the insertion slot for the sensors     The functionality of the temperature sensor  impacts the establishment of test readiness  as well as the achieved test values  Therefore   contamination must be avoided  To ascertain  the 
12. facility in this case     Never insert foreign objects or sensors  other than Lactate Scout sensors into the  insertion slot     Disposal    In the case of blood lactate measurements it   is impossible to exclude the possibility that  the device will come into contact with blood   As a result  there is a risk of infection  Dispose  of the device with household waste after  removing the batteries     The device does not fall under the application  of the European Directive 2002 96 EC  regarding electrical and electronic equipment      Attention  If the device housing is opened  by the user  this will invalidate all warranty and    liability claims  Disinfecting agents containing  isopropanol will result in the permanent  tarnishing of the display window              11    Taking a measurement    Inserting a sensor  Remove a sensor from the dessicant vial     The sensor must be inserted into the insertion  slot of the device with the black connection  contact surfaces facing up  see right   1      Regardless of whether the device was on or off  before  the test mode is automatically started  upon inserting the sensor     After a brief  one second  display of the  code  a blinking drop appears on the display  indicating that the device is ready to test  2          2304 13 20       13    Attention  Store the sensors only in the original  container and within the specified temperature  range of  18  C to  8  C  Only remove the sensors    for the purpose of testing  Close the conta
13. for minute  day and  year  Format  hh mm and dd mm yy       OFF          Exit  settings  gt     mode                               Enter    settings  mode             ee  o N  Meow  eer        Scroll wheel    21     set        settings for data transfer  see instructions  page 23 25      set  a  the speaker volume can be selected  using the scroll wheel as described above    The signal intensity is indicated by the speaker  symbol being full with three hemispheres  loud    full with one hemisphere  medium  or empty   muted      set   CTRL D S   settings for performing a  functionality test  see instructions page 26 27     mem   to view stored test results the memory  submenu must be activated  The last result is  displayed together with the storage position  number and if available with the step test data  achieved during the test  The storage position  number is showed intermittent with the date and  time of the test result  The memory can hold up  to 250 test results  If more values are recorded  the oldest value will be overwritten first     22    DELICE E    Transferring data to the PC    The Lactate Scout  is capable of the wireless  transfer of data to a PC using    Bluetooth           The data transfer menu is activated in the setting  and storage mode  Press the scroll wheel for two  seconds and scroll until    set    and a Bluetooth     symbol         appears blinking on the display  1    The menu is activated by briefly pressing the  scroll wheel so that    OFF    a
14. for performing the test carefully   ensure correct selection of test solution  level  If the error continues to appear please  contact Lactate Scout  Service           29    Er5   blinking    Tem perature error    Ensure that the ambient temperature is          temperatur The temperature is outside the operating range    within the specification of the device    symbol    Er6 Electronic error inside the device Turn the device off and restart  If the error  continues to appear please contact Lactate  Scout  Service    Error when Too large distance between Bluetooth module   Limit the distance  avoid obstacles or    transferring  data       and PC or obstacles in the way  disturbance of  signal by strong elektromagnetic fields  e  g   caused by electronic devices or transmitter masts    software error       interfering transmitters  reinstall software             30       Troubleshooting    Date bar    Blinking date time bar on the display  The date time settings are lost     Set date and time       Battery  symbol   blinking    Battery error  The batteries are nearly empty     Replace batteries       bat    Battery error  The batteries are empty or expired or low quality    batteries have led to corrosion of the battery contacts     Replace batteries  Use only quality  batteries  Contact Lactate Scout  Service  in case of corroded battery contacts        Code control    The display    set CODE    blinks after inserting a sensor    The set code must be checked and and if  neces
15. iner  again immediately  Never leave open  Protect  against solar radiation  prolonged temperatures  above 25  C and moisture  Temperatures outside  the specified range during transportation and  storage may influence the function of the sensors  and lead to faulty results     Sensors  once removed from the container  must  be used immediately  The sensors must be kept at  room temperature with the measuring device at  least 20 minutes  at least two hours when stored  in the freezer  in the closed container prior to  testing  The expiration date of the sensors on the  label of the sensor container must be observed   Following the initial opening of the container  the  sensors must be used within 90 days        Taking a measurement    14    Blood sampling and analysis   Produce a drop of blood with a sterile single  use lancet approved for capillary blood  sampling  Follow the instructions for use of  the lancet  The site where blood is drawn  must be clean and free of sweat  Lactate  accumulates on the skin from perspiration  and leads to falsely elevated results  Hands   or earlobes  must be disinfected and washed  with fresh  hygenically safe water  Dry hands  thoroughly prior to testing  Avoid pressing on  the puncture site too extensively as there is a  risk that sweat and or tissue fluid can merge  with the blood droplet and falsify the test  result  The drop of blood must not run     Bring the fingertip with the droplet of blood  close to the tip of the sensor that has 
16. l    READ INSTRUCTIONS PRIOR TO USE    Only use the device for the purpose described in this manual    Do not use any other accessories than those supplied by the manufacturer    Do not use the device in case of malfunction or damage    In case of non intended use the safety measures for the device may become ineffective     KEEP INSTRUCTIONS FOR USE    Not conforming to cautions may lead to serious health damage   Not conforming to information marked with    attention    may cause faulty measurements or damage of the device     Contents    1  Getting started    Lactate Scout  analyzer and sensors   Inserting the batteries   Code setting   Temperature sensor   Care and cleaning  Disposal 1    2  Taking a measurement    Inserting a sensor 12  Blood sampling 14       sO CON O amp O    3  Performing step tests    Settings for step tests 16  Afterload timer 18  Timing functions for step tests 18  Test results 17  Turning the device off 19  Deactivate the step test mode 19  Selecting and modifying settings 20  Transferring data to the PC 23  6  Troubleshooting  Functionality check in dSC mode 26  Error description  potential causes and solutions 28    7  Technical specifications    Parameters 32  Explanation of symbols 23    8  Accessories  Purchasing accessories 34      Insertion slot for the sensor      Display window  2  for the display of test   readiness  the blood lactate concentration    warnings and setting options     Scroll wheel  3  for easy operation of the  analyser  
17. lease contact Lactate Scout  Service  Test values Values appear to high too low Repeat the test using new sensor and          new coding  follow the instructions for  performing the test carefully  Perform  functionality test  In case of Er4 message  contact Lactate Scout  Service        31       32   Device weight    Parameter    Type of device    Sample material  Sample volume  Measurement range    Precision    Measurement principle    Test duration   Measurement temperature range  Humidity   Storage temperature   Hematocrit range  Hct    Max  elevation for use   Data storage   Data transfer   Power supply   Battery life       Device dimensions    Technical specification    Specification    Lactate Scout  handheld device for measuring blood lactate   model from SN 0000101115     Fresh capillary whole blood  0 2 ul  0 5 mmol L   25 0 mmol L      3    minimal standard deviation    0 2 mmol L   for Hct range 35     50     4   minimal standard deviation    0 3 mmol L   for extended Hct range    Enzymatic amperometric determination of lactate using  lactate oxidase    10 seconds    5   C    45   C   10   85  relative humidity  up to 2 minutes for sensors    18  C    8  C  sensors in container  or  20  C    50  C  device   20 70    up to 4 000 metres   250 test values with date  time and step test allocations  Integrated Bluetooth   module  Class II module BT 1 2    2x 1 5 V AAA   LRO3 batteries   At least 1000 tests   91 mm  L  x 55 mm  B  x 22 mm  H    80 g  with batteries     
18. mer Using the Countdown function    COUNT  DOWN  preceded by the step duration    DUR      it is possible to set times of up to 10 minutes  and to set a signal as they elapse  page 20      If you select the test mode    AFL    for an  afterload test  the time in minutes and seconds  elapsed from the last measurement in the step  test phase    LOAD    will appear below on the  display instead of the time  In addition  a brief  signal tone sounds after every minute that  passes  After no more than twenty minutes  the  timer on the display will remain at    20 00    and  will only be reset upon restarting a step test   The afterload timer will only be activated if one  or more load tests were performed     Timing functions for step tests    To be able to control and adjust the load  intensity and step length for field tests  more efficiently  the Lactate Scout  offers  supplemental application possibilities in the  setting mode  see chapter 4         With the stop clock    CLOCK     you can monitor  your lap time  maximum of ten minutes per    Display of the expired time following  load step   SEE page 20  the last load measurement via the afterload timer    18    19    Detailed information on how to perform step  tests can be found in the Lactate Scout   Compendium on the Lactate Scout Info CD or on  the website www lactatescout com    Test results   Test results that are in    idle mode    i e  measured  prior to exercise  should be between 0 5 mmol L  and approx  2 5 mmol L  
19. nd the continually  blinking Bluetooth    symbol below are displayed  on the main display  2         Menu for activating the Bluetooth    is inactive  Bluetooth    connection    23    Data storage    By turning the scroll wheel by one position    the Bluetooth    module is connected     On     appears on the main display and the Bluetooth     symbol is surrounded by a black line  3   After  confirmation  briefly press the scroll wheel   the  Bluetooth    module in the device is activated   The Lactate Scout  bluetooth function  can be  used with Windows XP  Windows Vista   Windows 7 and Windows 8     For data transfer to a PC additional software must  be installed on the PC  A generic data export  tool Lactate Scout Data link is availabe as free  download from the website www lactatescout   com   Advanced software solutions like the 4  Lactate Scout Assistant software including   data management and chart functions for   performance analysis are available for purchase   from EKF Diagnostics        l Bluetooth    is active       Display of Bluetooth     symbol after activation    The activated Bluetooth    module is permanently Attention  A detailled description for the  displayed in the setting and storage mode as an initial ora of ib densa Ase connection  outlined Bluetooth    symbol  4   Is available on the website www lactatescout   y  4  com  Please use the Bluetooth    dongle  recommended by us        N    U1    In addition to the possibility of transferring in   the set
20. s started automatically and an  error message    Er3    or    Er4    is displayed  see  error messages      Test solutions can only be measured in dSC  mode  Any measurement of test solution in the  regular test mode will give wrong results     Attention  Never hold the opening of the test    solution bottle directly up to the sensor as its  contents will become contaminated and unusable        Caution  Ingesting or contact of test solution  with the mucous membranes must be avoided     Keep the test solution away from children   choking hazard           28    Troubleshooting    Error description  potential causes and solutions                   Error Description and potential cause Solution   Er1 Sensor error Use a new sensor from a correctly stored  Faulty sensor due to damage  incorrect storage or   container  sample uptake prior to test readiness   Er2 Error when filling the sensor Repeat the test with a new sensor  If  Blood droplet too small or ran  contact time necessary promote circulation  gently  sensor to blood droplet too short  interrupted massage the puncture site   sample absorption or sensor pressed against skin   Er3 Plug connector error Repeat the functionality test with a new  There is a problem with the contacts of the   sensor  If the error continues to appear  lactate sensor please contact Lactate Scout  Service    Er4 Functionality test not successful Repeat the functionality test using new             materials and new coding  follow the  instructions 
21. sary corrected after 24 tests        LO display    Test result too low  The test value is below the test range of the  Lactate Scout     Repeat the test using new sensor and  new coding  follow the instructions for  performing the test carefully  dry hands  carefully prior to testing     Perform functionality test  In case of Er4  message contact Lactate Scout  Service        HI display       Test result too high  The test value is above the test range of the  Lactate Scout        Repeat the test using new sensor and  new coding  follow the instructions for  performing the test carefully  wash hands  carefully prior to testing     Perform functionality test  In case of Er4  message contact Lactate Scout  Service              Turn on Device fails to turn on Replace batteries  Press scroll wheel   gt  2 seconds  If the problem persists  please contact LactateScout  Service    No test No drop symbol displayed Insert sensor until it is noticeable in   readiness Sensor not inserted properly place with the black contact surface    facing up        Signal tones    No acustic signal  Volume was muted in the volume setting menu    Select and activate the volume under  settings          Display Display shut off Restart the device  Happens automatically after 5 minutes of inactivity  Poor display contrast Ensure ambient temperature in operating  Too low ambient temperature or empty batteries range  replace batteries  Display elements missing  Defective electronics or mechanical damage P
22. t com    Item no   7023 3405 0846    7023 3405 0727    7023 3405 1008    7023 6151 0257    7023 6151 0235    7023 6141 0246    7202 4306 0276       
23. ting and storage mode  a data transfer is  also possible in the test mode with an activated  Bluetooth    module  Following the test  the  Bluetooth    module is turned on while the test  values are displayed  5  6  in order to enable a  transfer of data  e g  the last tested values   The  success of this option depends  however  on the  PC software used for presenting the tested data      The procedure for turning off the active  BluetoothTM function is consistent with the  procedure for turning it on as described on  page 24              Display during a test with an  active Bluetooth    module    Display after a test with an  active Bluetooth    module    Attention  To decrease energy consumption   the Bluetooth    module should not be activated  permanently  but rather only for the purpose of  transferring data  The reach of the Bluetooth       connection is no more than 10 metres without    spatial barriers  The functionality of the device  can be influenced by strong electromagnetic  fields  e g  in the proximity of radio towers         Troubleshooting    26    Functionality check in dSC mode   A functionality check should be performed if  there are any doubts regarding the accuracy   of the test result or the correct function of the  device or sensors  Test solutions are available in  following concentrations  10 mmol L  5 mmol L  and 2 mmol L     All test solutions must be stored at 5  C to 25  C   The expiry date for the test solution is stated on  the label of the t
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Maxtor ONETOUCH III 400 User's Manual    Raypak 302B-902B User's Manual  DB500 & DB800 - Dustless Blasting  NETGEAR N300 Wireless Dual Band ADSL2+ Modem Router  LED Weather Station with 433 MHz Remote Sensor  HUD 2010 with Good Faith Estimate User Manual Version 1.3.2  SPN-01 取扱説明書  Zenix-C SEC Column User Manual - Sigma  "user manual"    Copyright © All rights reserved. 