User Guide Sentino® Filter Dispenser
Contents
1. Lower spool with retainer clip O N O 10 Refill pack holder 11 Power entry socket located on back of the unit 12 Product identification label Power 13 Transformer power entry plug 14 Transformer input interface 15 Power cord EU UK and NA plugs are included in each instrument 4 Life Sciences P nea g YE Pall Lorporatio Sentino Dispenser r N Symbols Referenced 4 oee manual for instructions The Sentino Filter Dispenser a product of Pall Corporation is in conformity with the requirements of C the European Directive 2004 108 EC and associated amendments with regards to Electromagnetic Compatibility and the European Directive 2006 42 EC and associated amendments with regards to Machinery See EC Declaration of Conformity section X Reuse recycling recovery and disposal instruction Safety Precautions To protect personal safety and ensure correct use of this equipment be sure you read and understand this guide before use The following terms are used DANGER A To designate a condition which MAY involve mortal danger or serious injury CAUTION To designate a condition which MAY involve serious injury or destruction of the equipment involved NOTE i To designate useful information to improve performance or a condition which MAY involve faulty operation of the equipment thus causing poor performance Universal Precautions Appropriate care s
2. Life Sciences User Guide Sentino Filter Dispenser Product Number 13184 Filtration Separation Solution sw Table of Contents anal ku a NM DAN er r mr mga rrrryWW wa rar rarRrzz 3 Bi ge ERSS aa nl a en tttm 3 mese ain TO j sya a5k544 55425555554445 k a kak e Uum 4 hwa e DD e arwo AXOrPe o o_ _e_e e_ o 5 h z sanal sa NNN nines E A EEA sxaccenssiessbcnsdanteeedesnadesedesesscesseseadaesacaesara 5 iSc eie Y DO nye keyen ee KA OE t On Re SENA r N eee Ke T TR RENTUR 6 DOL N L l ls eat n eske ko xaki cuya n kk Sel wn de k ku EI B n KO KUNA Ka aw SEN KG AGA AVA A A Keyan N ieee Se naw KEWA SA K N kn 6 WOULD OO pieles pre Mese PUTET NN CTETUR 8 jut eed NR RT AEII RM c 8 ale D 9 nikl wa r EN A A A ETE A PA AE FEIN MM UI RM MEME M FEMME UR 9 DIS DOS TI G NO L N Ll TRE 9 ECG Decirom OT COMIC RETE ear yk k pa k xa w n eha k o UT 10 2 Life Sciences Introduction Application and Intended Use The Membrane Filter MF Technique is a method of analyzing aqueous samples for microbial contamination A fluid sample is pulled through a membrane filter and the filter is placed on growth media and incubated After the prescribed incubation period colonies are counted and reviewed for appropriate morphology Pall Life Sciences
3. disposal of used equipment www pall com lab 10 Life Sciences EC Declaration of Conformity CE Product Description Sentino Membrane Dispenser Product Part Number 13184 The Sentino Membrane Dispenser a product of Pall Corporation is in conformity with the requirements of the European Directive 2006 42 EC and associated amendments with regards to Machinery the European Directive 2004 108 EC and associated amendments with regards to Electromagnetic Compatibility and the European Directive 2006 95 EEC and associated amendments with regards to Low Voltage safety and the restriction of the use of Hazardous Substances RoHS Directive 2011 65 EU This declaration is based upon testing the compliance of the product against the limits of the following harmonized standards Emissions and Immunity EN 61326 2013 Electrical equipment for monitoring control and laboratory use EMC requirements Part 1 General requirements and assessing the compliance of the product to these harmonized standards EN 61010 1 2010 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory use EN 60204 1 2006 Safety of Machinery Electrical Equipment of Machines Part 1 General Requirements EN 12100 1 2003 Safety of Machinery Basic Concepts General Principles for Design Part 1 Basic Terminology methodology EN 12100 2 2003 Safety of Machinery Basic Concepts General Principles for Design Part 2 T
4. If they do not contact Pall Technical Service see Service section Films not tracking properly on the spool Initial load not aligned properly Verify dispenser refill pack is loaded properly see Preparing System for Use section Tear off web at perforation and reload per instructions If the above corrective actions do not remedy the problem contact Pall Technical Service see Service section Maintenance Do not immerse Ay The dispenser is designed only for use with Pall membrane refill packs i Visit www pall com sentino for video on system preparation and operation Cleaning Body and Plates e Surfaces may be wiped down with a towel moderately soaked in the following common laboratory sanitizing solutions 70 isopropyl alcohol IPA 10 bleach Clearing Positioning Sensor e Use compressed air to blow dust from the positioning sensor Life Sciences Service Save original packaging for service and warranty repairs Pall does not recommend any service be performed on this device other than by the manufacturer Contact Pall Life Sciences for technical support warranty and service inguiries at 800 521 1520 inside USA and Canada or LabCustomerSupport pall com Specifications Materials of Construction Housing Material Polycarbonate ABS blended polymer Keypad and Label Polyester Bottom Plate Anodized aluminum Upright Anodized aluminum Spools Anodized aluminum Guide Plates Ano
5. Sentino Filter Dispenser is designed to aseptically dispense a single disc of membrane at the press of a button The membrane can easily be transferred using forceps onto a platform to be used in MF Tech nique As the membrane is dispensed the packaging materials neatly wind onto spools for disposal when the end of the pack is reached The dispenser s small foot print is designed to take up less bench top space and the small membrane refill pack results in less packaging for disposal The Sentino Filter Dispenser is designed for laboratory use only and should not be used in a manner other than indicated Ordering Information Sentino Filter Dispenser Part Number Description 13184 1 Power Transformer 1 European power cord with CEE 7 7 plug 1 UK power cord with BS1363 plug 1 NA power cord with NEMA 5 15P 1 User Guide CD Accessories Dispenser Pack efills Part Number Description Pkg 68123 0 2 um Supor membrane 1000 pkg 68121 0 45 um GN 6 Metricel membrane 1000 pkg 68124 0 45 um Metricel Black membrane 1000 pkg 68125 0 8 um Metricel Black membrane 1000 pkg www pall com lab Product Terminology Keypad 1 Main power indicator 2 Membrane dispense button 3 Battery status indicator 4 Battery charge indicators Body 5 Upper spool with retainer clip slides horizontally Off spool as shown a Retention groove b Retainer clip pin c Retainer clip knob Positioning sensor Guide plates Upright support
6. dized aluminum Retainer Clips Stainless steel with anodized aluminum knobs Box Retainer Stainless steel Foot Pads Rubber Dimensions Length 22 0 cm 8 7 in Width 12 0 cm 4 7 in Height 20 0 cm 7 9 in Weight 0 9 kg 2 0 Ibs Temperature Range Operation 15 30 C 59 86 F Storage 10 35 C 14 95 F Battery Voltage 12 0 V Type NiMH nickel metal hydride DC Supply Voltage 24 0 V AC DC Power Transformer 24 VDC 1 6 Amp 100 240 VAC 47 63 Hz Input Interface IEC 320 C8 pin socket Output Interface Power Entry Plug 2 5 mm barrel jack Power Cord C7 connector and NEMA 5 15P grounded plug C7 connector and European CEE 7 7 plug C7 connector and UK BS1363 plug Disposal Instructions Reuse recycling recovery and disposal instruction The materials used in this unit are ROHS compliant The presence of this label on a product means that the product contains electrical or electronic materials and therefore it should not be disposed of as unsorted waste but instead be mm treated separately The presence of these materials may if not disposed of properly have potential adverse effects on the environment and human health Within the European Union users are urged to recycle such products when being replaced with a newer version or when they have outlived their useful lives However as the legislation and facilities vary throughout the member states please visit www pall com weee to arrange for proper
7. echnical principles and specifications Signed Date 21 July 2014 Dr N Moss Director of Instrumentation Technology For and on behalf of Pall Europe Limited 5 Harbourgate Business Park Portsmouth England PO6 4BQ Life Sciences General Limited Warranty Warranty Limitation of Liability and Remedies 1 There is no warranty of merchantability or fitness for any particular purpose with respect to any of the products nor is there any other warranty expressed or implied except as provided for herein 2 For a period of 12 months from the date of delivery from Seller the Warranty Period Seller warrants that products manufactured by Seller when properly installed and maintained and operated at ratings specifications and design conditions will be free from defects in material and workmanship 3 Seller s liability under any warranty is limited solely in Seller s discretion to replacing F O B original ship point repairing or issuing credit for products which become defective during the Warranty Period Purchaser shall notify Seller promptly in writing of any claims and provide Seller with an opportunity to inspect and test the product claimed to be defective Buyer shall provide Seller with a copy of the original invoice for the product and prepay all freight charges to return any products to Seller s factory or other facility designated by Seller All claims must be accompanied by full particulars includi
8. ery status indicator will extinguish when the battery is fully charged Unplug from main power and press membrane dispense button to check battery charge indicators Plugged into main power source blue battery status indicator not illuminated only left most battery charge indicator LED illuminated blue Battery failing to charge Contact Pall Technical Service See Service section Battery charge indicator lights not illuminating Plugged into main power source Unplug from main power and check for illumination when dispenser button is pushed Not dispensing a disc of membrane or spools are barely rotating Battery discharged Continue using power transformer plugged into AC mains power until battery is fully charged Dispensing two discs at a time Obstruction in positioning sensor Verify dispenser refill pack is loaded properly with rectangular cut outs on the left see Preparing System for Use section Tear off web at perforation and reload per instructions Use compressed air to remove dust from the positioning sensor spools are not rotating Obstruction in one or more spool Check that spools are free of foreign objects Verify dispenser refill pack is loaded properly see Preparing System for Use section Tear off web at perforation and reload per instructions Check that spools allow adjustment in both directions using retainer clip knobs to gently rotate the spools
9. es c After each use place the dispenser back on charge d If the dispenser is not used on a regular basis leave the unit charging or charge at least once a week Filter Refill Pack Installation The Sentino Filter Dispenser is designed to accommodate Pall dispenser refill packs 6 Life Sciences Loading 1 Remove lid flaps of the membrane refill box Remove U shaped cut out for viewing end of pack when loaded in the dispenser 2 Slide box into recessed holder on the dispenser with the product label facing left and the U shaped cut out facing right The refill web should now be aligned such that the clear film is facing up as the web feeds into the dispenser and the rectangular cut outs are on the left closest to the up right support 3 To minimize the length of web required for loading use knobs on retainer clips to adjust upper spool with a retention groove at the 2 o clock position and lower spool with a retention groove at the 4 o clock position 4 Remove retainer clips from upper and lower spools 5 Insert the fan folded media between the membrane guide plates as shown and pull 17 filters between the guide plates 6 Separate the film layer from the paper layer wrap the clear film in a counter clockwise direction over the retention groove positioned at the 2 o clock position Each refill pack has two extra discs for loading the dispenser Discard the two discs exposed during loading 7 Replace the upper
10. hould always be taken when using electrical equipment around aqueous solutions in the laboratory Opening or disassembly of product enclosure may involve electrical shock Only authorized personnel should disassemble product Do not immerse Only use parts approved by the manufacturer For laboratory use only Keypad buttons on this product intended for finger touch only Only use recommended solutions for cleaning sanitization Return product in original packaging for warranty repair VVarranty valid only when used in accordance with this User Guide gt b www pall com lab Keypad and Symbols MAIN POWER INDICATOR e A green LED will illuminate when the dispenser is plugged into the power transformer and VAC power source This LED will not be illuminated when operating on battery power DISPENSE BUTTON Press the dispense button to dispense a single disc of membrane BATTERY STATUS INDICATOR When plugged into the power transformer and VAC power source this symbol will indicate the battery is charging by an illuminated blue LED When the battery is fully charged the blue LED will turn off BATTERY CHARGE INDICATOR 09090 When operating on battery power three blue LED lights will illuminate to indicate the relative level of battery charge when the dispense button is pressed e All three LEDs will illuminate after pressing the Dispense Button when the battery is fully charged or near full charge e wo LEDs will
11. illuminated after pressing the Dispense Button when the battery is partially charged e One LED will illuminate after pressing the Dispense Button when the battery power is low and the dispenser should be placed on charge before further use Note The dispenser can be operated on main power as the battery recharges Operation o Save original packaging for service and warranty repairs Device Set Up Connecting Power Supply and Charging 1 Position the Sentino Filter Dispenser on a laboratory bench 2 Attach power entry plug from power transformer into the power socket Connect power cord to the input interface on the power transformer then plug power cord into appropriate electrical outlet having 100 240 VAC supply 3 Confirm Main Power Indicator is illuminated and Battery Status Indicator is illuminated blue to indicate battery is charging 4 Operation of the dispenser is optimal when used on battery power however unit can be used while plugged into main power 5 Charging instructions for use on battery charge a Fully charge dispenser before first use b After charging use the dispenser on battery power for the work day During use monitor the Battery Charge Indicator As the battery depletes the number of blue LEDs illuminated will decrease from 3 to 2 to 1 When the Battery Charge Indicator reaches one LED illuminated place the unit on charge NOTE The dispenser can be operated on main power as the battery recharg
12. ng system operating conditions if applicable 4 In no event shall Seller be liable for any product altered outside of the Seller s factory by someone other than oeller or for a product subjected to misuse abuse improper installation application operation maintenance or repair alteration accident or negligence in use storage transportation or handling 5 In no event will Seller be liable for any damages incidental consequential or otherwise whether arising out of or in connection with the manufacture packaging delivery storage use misuse or non use of any of its products or any other cause whatsoever Pall Standard Terms and Conditions apply to the purchases of all Pall products and services which Terms and Conditions for each region can be found at www pall com under the tab Find Products Terms and Conditions of Sale or obtained by written request to the local Pall office for your region Life Sciences Visit us on the Web at www pall com lab E mail us at LabCustomerSupport pall com Corporate Headquarters Port Washington NY USA International Offices Pall Corporation has offices and plants throughout the world in Argentina Australia Austria Belgium Brazil Canada China France Germany India Indonesia Ireland Italy Japan Korea Malaysia New Zealand 1 800 521 1520 USA and Canada Norway Philippines Poland Russia Singapore South Africa Spain Sweden Switzerland Taiwan Thailand LabCustomerSu
13. pport pall com bos u cul and VEJ N Distributors in all major industrial areas of the world To locate the Pall office or distributor nearest you visit www pall com contact The information provided in this literature was reviewed for accuracy at the time of publication Product data may be subject to change without notice For current information consult your local Pall distributor or con tact Pall directly 2014 Pall Corporation Pall ALD Metricel Sentino and Supor are trademarks of Pall Corporation indicates a trademark registered in the USA Filtration Separation Solution is a service mark of Pall Corporation Filtration Separation Solution sm 7 14 PDF GN12 8126c PN89965C
14. retainer clip and secure the film by catching the end of the film between one pin of the retainer clip and the retention groove on the spool To avoid jamming it is only necessary to capture the film with one side of the retention clip 8 Repeat with the paper and lower spool but wrap the paper layer in a clockwise direction 9 Adjust any slack in the refill web by manually turning the spools using the retainer clip knobs Do not press the membrane dispense button until ready to use the filter 10 Press membrane dispense button to dispense a single disc of membrane At any point at idle operation the retainer knobs can be used to adjust the tension of the refill web on the spools www pall com lab T Troubleshooting Problem Plugged into main power source and green main power indicator not illuminated Plugged into main power source but blue battery status indicator not illuminated Cause Incorrect power transformer Incorrect power source Poor connections Incorrect power transformer Corrective Action Use power transformer supplied with dispenser Check that power source supplies 100 240 VAC Check that all connections are securely pushed together Use power transformer supplied with dispenser Incorrect power source Check that power source supplies 100 240 VAC Poor connections Check that all connections are securely pushed together Battery is fully charged Batt