Home
        Tektronix 11A52 User's Manual
         Contents
1.     5 0 mV to 9 95 mV div 0 02 div  rms    10 0 mV to 99 5 mV div 0 015 div  rms    100 mV to 995 mV div     1 0 015 div  rms    1 0 V to 10 0 V div 0 015 div  rms     DC Drift with Temperature 20 uV   C  or less  at any sensitivity        Channel Isolation At least 50 1 display ratio     any Coarse  volts div setting  DC  600 MHz        Common Mode Rejection Ratio At least 20 1  DC to 50 MHz  10 div   reference signal on each input        Probe Compatibility The 11A52 is compatible with Level 1 and  2 TEKPROBEs     IDC drift can be calibrated out by invoking a calibration  Enhanced Accuracy  at any specific  operating temperature       Amplitude  div  x V div  driven channel       Display ratio   eccrine  Error amplitude  div  x V div  undriven channel     TEKPROBE is Tektronix    name for the interface used with probes designed for the 11000 series  of oscilloscopes and plug in units  TEKPROBEs have output connectors with one or more spring  loaded coding pins  Two categories of TEKPROBEs are     Level 1  A level 1 probe uses analog encoding to indicate the probe s scale factor to the plug in  unit       Level 2   A level 2 probe uses an EEPROM to store data about the the probe s transfer units  scale  factor  and output voltage scale factor  Such data are serially encoded  then stored in the  EEPROM  The probe data is intended to be read once at instrument power up or when the  probe is first connected to a plug in unit  that is  at probe power up      11A52 User Reference
2.    Numeric representation    Sets the input impedance of the specified channel    Sets the deflection factor of the specified channel  Sensitivity  is a channel specific command which does not apply to    compound waveforms     Quoted string data     11A52 User Reference Supplement    _ Section 3    Operating the 11A52 in  11400 series Mainframes       Display On Off    The 11452 has only two front panel controls   the display on off buttons   Pressing a display on off button will cause its channel  and signal  to be  displayed or removed from the display  the function toggles   When a channel is  displayed  its label  CH 1 or CH 2  will be lighted     Other Functions    All other 11A52 functions are controlled within the 11401 or 11402 oscilloscope   and their operation is described in the 11401 and 11402 User Reference manual   Table 3 1 shows where to find information about operating the 11A52     TABLE 3 1  Functions and Where They are Described  Function Described under Heading     Vertical Offset   Vertical Size and Position   Vertical Size  sensitivity  Vertical Size and Position     Coupling  AC  DC  Off  Selecting Channel Coupling     Select 11A52 Channel Displaying Waveforms  an alternative to pressing  the 11A52 display on off button    Input Impedance Selecting Channel Input Impedance   Offset Vertical Position   BW Limit Selecting Channel Bandwidth Limit   Display Polarity Vertical Description        Under Waveform Control in Section 2  Operating Information     
3.   indicating what type of probe is connected to the  input  If a Level 1 TEKPROBE is connected  the  query response is    Level 1     If a Level 2  TEKPROBE is connected  the query response is     Level 2  lt probe_type gt    lt serial_number gt       When neither Level 1 nor Level 2 TEKPROBE is  connected  the query response is    NONE        L  C  and R mean Left  Center  and Right plug in  compartments  1 and 2 mean Channel 1 and Channel 2     Sets the specified channel input coupling    Sets the specified channel offset    Sets the HF Limit  bandwidth  of the specified channel   Numeric representation    Sets the input impedance of the specified channel    Sets the deflection factor of the specified channel   Sensitivity is a channel specific command which does not    apply to compound waveforms     Quoted string data     3 11    Operating the 11A52 in 11400 series Mainframes    3 12 11A52 User Reference Supplement    Section 4  Specification       Performance  Conditions    Circuit    Display  Deflection Factor   Sensitivity     The specifications that follow apply when the instrument is in the condition of  Enhanced Accuracy  Enhanced Accuracy is initiated by pushing the E A button  on the specific host mainframe after the system has reached thermal equilibrium   which requires 20 minutes warmup  Enhanced Accuracy is indicated on the CRT  display and remains in effect as long as the mainframe internal temperature  change is less than 5   C from the temperature at which 
4.   options  special switch and or strap settings and operating system     VI  SECTION E  Describe the problem completely  Include any information which might help in evaluating the error with the  PR  If you have determined a procedure to avoid the error condition  please include this procedure  If this  problem prevents you from accomplishing any useful work with the product  please state this fact  Be sure to  include with the PR any information  programs  listings  hard copies  etc   which will help us duplicate your  problem     VIL SECTION F  This section is for use by Tektronix Lab Instruments Marketing Support personnel  DO NOT WRITE IN  THIS SPACE     VIH  Mail all copies of the Problem Report to     TEKTRONIX  INC    LAB INSTRUMENTS MARKETING SUPPORT  P O  BOX 500  DEL STA 39 327   BEAVERTON  OREGON 97077       INTERNAL USE ONLY   DO NOT WRITE BELOW THIS LINE        PR LOG    
5.   r scccssscsscesronscecsvesancnnsensonersensonsecssacscsscessseecesesseescnssers bod       11A52 User Reference Supplement iii    Contents       Appendix A   Glossary    Appendix B   Instrument Documentation    Illustrations    1 1  2 1  3 1  3 2    4 1    Tables    2 1  2 2    3 1  3 2    4 1    4 2  4 3    iv    Installing a plug in unit in a mainframe oscilloscope                1 5  Location of AC coupling capacitors in plug in amplifiers           cceeen 2 4  Location of AC coupling capacitors in plug in amplifiers          seeoee 3 4  Sequence of decisions used to select a new waveform   in response to probe ID button    me 3 8  Dimensions Of 11A52        ccsssesscesscenesesscncnsenenssscneseneonsnsonsnsoesasaenoncesooseneserasses 4 8  Functions and Where They are Described       s sscsescsessenorennsossransensssssnrarenes 2 2  Channel Parameter Commands  for 11300 series oscilloscopes     s s  2 9  Functions and Where They are Described sen 3 1  Channel Parameter Commands  for 11400 series oscilloscopes               3 10  Electrical Characteristics en smenenneeneeeeenes 4    Environmental Characteristics sense 4 6  Physical Characteristic     sseeneeaeeeenaneneneenenese 4 7    11A52 User Reference Supplement    Section 1  General Information    This Supplement to the mainframe User Reference manual and the Incoming  Inspection Procedure manual are standard accessories to the 11A52  For a list of  related manuals  see Appendix B        Contents of this Supplement    T
6.  Control knob  ImVto995mV     1V 25E 6 Coarse  0 25 div   100 mV to 995 mV    10 V   250E 6 Fine  0 025 div   1Vto10V  100 V   2 5E 3  BW   lt NRx gt  Bandwidth    24E6 20E6   gt 24E6 to  120E6 100E6   gt 120E6 500E6  11401  600E6  11402    IMPEDANCE   lt NRx gt  Impedance   lt 1E3   50      SENSITIVITY    Sensitivity  1E 3 to 1 99E 3                  Resolution  step size   10E 6    2E 3 to 4 98E 3 20E 6  5E 3 to 9 95E 3   50E 6  10E 3 to 19 9E 3 100E 6  20E 3 to 49 8E 3 200E 6  50E 3 to 99 5E 3 500E 6  100E 3 to 199E 3 1E 3  200E 3 to 498E 3 2E 3  500E 3 to 995E 3 5E 3   1 to 1 99 10E 3  2 to 4 98 20E 3  5 to 10 50E 3    UNITS   lt qstring gt   query only     The 11A52 answers a Units query with a units  status message  which indicates the units of  conversion of a probe connected to its Ch N  input     The numbers listed are those available at the input connectors  Connecting an attenuating probe  will change the value by the probe attenuating factor  e g  a 10X probe wiil sunset the value  10  to  100      3 10 11A52 User Reference Supplement    Header    CH lt L1  L2  C1   C2 R1  or R2 gt     Legend   CH lt L1  L2  C1   C2  R1 or R2 gt     COUPLING  OFFSET  BW   NRx  IMPEDANCE    SENSITIVITY    qstring    11A52 User Reference Supplement    Operating the 11A52 in 11400 series Mainframes    Table 3 2  cont   Channel Parameter Commands   for 11400 series Oscilloscopes     Link Argument       PROBE   lt qstring gt   query only     This query only link returns a quoted string
7.  Supplement 4 5       TABLE 4 2    Environmental Characteristics    Characteristic    Ambient temperature   External to main frame     Operating within specs   Nonoperating  Humidity  Operating    and Nonoperating    Altitude   Operating   Nonoperating    Vibration    Operating  installed on  Flexible Extender    Shock  Nonoperating  not  installed in mainframe     Bench Handling  operating  and nonoperating     Packaged Product Vibration  and Shock    Vibration and Bounce  of Packaged Product    Drop of Packaged Product    Electromagnetic Compatibility    Information    0   to  50   C   mainframe ambient    40 to  75   C   Five days  per MIL T 28800C  Type El     class 5 as described in 3 9 2 2 and  4 55 1 2 2     To 4 570 m  15 000 ft       To 15 200 m  50 000 ft       MIL T 28800C  Sec  4 5 5 3 1  type IH     class 5     MIL T 28800C  Sec  4 5 5 4 1  type III       class 5     MIL T 28800C  Sec  4 5 5 4 3  type HI   class 5     Meets ASTM D999 75  Method A   NSTA Project 1A B 1      Meets ASTM D775 61  Paragraph 5   NSTA Project 1A B 2      MIL  STD  461B     FCC Part 15  Subpart J  Class A   VDE 0871 6 78  Class B     11A52 User Reference Supplement    Specification    Table 4 3  Physical Characteristics  Characteristic Information  Weight  max  1 Ib  14 oz   851 grams      Weight of Packaged Product  max    4 Ib  12 oz   2 2 kg      Dimensions  max  Refer to Figure 4 1     Recommended Probes    Tektronix recommends the P6231 Probe for use with the 11A52  It has the  foll
8.  connectors can be displayed or    removed from the display by pressing the display on off buttons adjacent to the    input connectors     All other 11A52 functions are controlled through the host mainframe  Such  mainframe controlled functions are       Sensitivity  Coarse and Fine  over a range of 1 mV to 10 V division   e Vertical Offset        Sane AC  DC  or Off      Bandwidth Limit  100 MHz or 20 MHz      Display Polarity  non or inverted       Trigger Polarity  normal or inverted     e Combination of Display Channels  See Operating Information section of  mainframe User Reference manual     e Combination of Trigger Channels  same as Display Channels  see Operating  Information section of mainframe User Reference manual        11A52 User Reference Supplement   i    ii    11A52 User Reference Supplement      Contents       Section I   General Information    Contents of this Supplement sense sacle   11A52 Service Manual sn wee 1 2  Plug in to Mainframe Compatibility encens 1 2  Initial Inspection    AET IA E EEEN wins 1 3  Operating Temperature sn nel  Installing and Removing the 11A52 sense 1 4  Instrument Options    nn siennes dO  Packaging for Shipment sn O NE EN 1 6    Section 2   Operating the 11A52 in 11300 series Mainframes    Display On Off sun RE  Selecting Coupling  HF Limit  and Impedance 2 1  Other Functions serum RTS ae  Position and OSE sisaccccsesvesccsenseasseceduonencsons entucosdestaubesensennsedspatin ccbavecnssthyoeieiobeay 7  DE Circuit Loading sim Ha
9.  functions  e g   on and off      11A52 User Reference Supplement A 1    Appendix    A 2    11A52 User Reference Supplement    _ Appendix B   Instrument    Documentation       Introducing the 11301 and 11302 Programmable Oscilloscopes  This book helps the  new user learn the concepts and operation of the 11301 and 11302 programmable  oscilloscopes        or    Introducing the 11301A and 11302A Counter Timer Oscilloscopes  This book helps the  new user learn the concepts and operation of the 11301A and 11302A counter   timer oscilloscopes     o Introducing the 11401 and 11402 Digitizing Oscilloscopes  Intended for new users   this book explains the concepts and operation of the 11401 and 11402 digitizing  oscilloscopes      gt  The 11301 and 11302 User Reference Manual contains detailed information about  measurements the instruments can make  and instructions about how to set the  oscilloscope to perform such measurements  Also  installation instructions   specifications  description of options  and details about using the 11301 and 11302  with the GPIB and RS 232 C interfaces are included here     or    The 11301A and 11302A User Reference Manual contains detailed information about  measurements the instruments can make  and instructions about how to set the  oscilloscope to perform such measurements  Also  installation instructions     specifications  description of options  and details about using the 11301A and  11302A with the GPIB and RS 232 C interfaces are included he
10. 11A52 User Reference Supplement 3 1    Operating the 11A52 in 11400 series Mainframes    Vertical Offset    BW Limit      The 11401 and 11402 oscilloscopes use offset to position the trace on the screen   In YT  signal vs  time  mode offset determines vertical position  but when a  plug in unit provides the X component of an XY display  offset controls  horizontal position     Offset is an input signal related control  and its units are those of the input  signal   usually volts     Offset subtracts a precision voltage from the input signal  To access the offset  function  touch the double ended arrow at the left edge of the screen  Control of  offset is then assigned to the Lower Control knob     In the 11401 and 11402  changing the vertical deflection factor magnifies or  compresses the displayed signal about screen center  Therefore  to examine a  portion of the input signal in greater detail  use Vert Offset  the Lower Control  knob  to position the area of interest to screen center  Next  use the Vert Size  the  Upper Control knob  to adjust the display to the size needed for detailed  examination     Two four pole  24 dB octave  bandwidth limit  low pass  filters are available for  each 11A52 channel  The purpose of these filters is to reduce the amplitude of  unwanted noise or interference occurring at frequencies above the frequency of  the signal of interest  The user has a choice of cut off   3 dB  frequencies  either  20 MHz or 100 MHz  independently for each chann
11. 3  1 to 1 99 10E 3  2 to 4 98 20E 3  5 to 10 50E 3  UNITS   lt gqstring gt   query only     The 11A52 answers a Units query with a units  status message  which indicates the units of  conversion of a probe connected to its Ch N  input     1 The numbers listed are those available at the input connectors  Connecting an attenuating  probe will change the value by the probe attenuating factor  e g   a 10X probe will change the    value  10 to  100         11A52 User Reference Supplement    2 9    Operating the 11A52 in 11300 series mainframes    Header    CH lt L1  L2  C1   C2  Ri or R2 gt    cont     Legend CH lt L1  L2  Cl   C2  R1 or R2 gt   COUPLING  OFFSET  BWHI  NRx  IMPEDANCE    SENSITIVITY    qstring    2 10    Table 2 2  cont   Chanel Parameter Commands   for 11300 series Oscilloscopes     Fa    PROBE   lt qstring gt   query only        This query only link returns a quoted string  indicating what type of probe is connected to the  input  If a Level 1 TEKPROBE is connected  the  query response is    Level 1     If a Level 2  TEKPROBE is connected  the query response is     Level 2  lt probe_type gt    lt serial_number gt       When neither Level 1 nor Level 2 TEKPROBE is  connected  the query response is    NONE        L  C  and R mean Left  Center  and Right plug in  compartments  1 and 2 mean Channel 1 and Channel 2   respectively    Sets the specified channel input coupling    Sets the specified channel offset     Sets the HF Limit  bandwidth  of the specified channel
12. An offset voltage of less than  one volt will be maintained as the sensitivity is changed over the entire range of  1 mV division to 10 V division  A selected offset of greater than one volt is  beyond the offset range for the most sensitive settings and will be reset to one  volt when the sensitivity is increased to any value between 99 5 mV and   1 mV division     Two four pole  24 dB octave  bandwidth limit  low pass  filters are available for  each 11A52 channel  The purpose of these filters is to reduce the amplitude of  unwanted noise or interference occurring at frequencies above the frequency of  the signal of interest  The user has a choice of cut off   3 dB  frequencies  either  20 MHz or 100 MHz  independently for each channel  The trigger  auxiliary  trigger  and display signal bandwidths for a channel are always the same  The   auxiliary trigger is the signal sent to the right plug in compartment        11A52 User Reference Supplement   2 3    Operating the 11A52 in 11300 series mainframes    DC Circuit Loading    AC coupling capacitors are connected differentiy in the 11A52 and 11A71 than in  the 11A32 and 11A34  Figure 2 1 shows this difference     2 4       CH    114652  11A71    500  11432  11434       C6114 203    Figure 2 1  Location of AC coupling capacitors in plug in amplifiers     In the 11A52 and 11A71  the coupling capacitor isolates both the input  termination and the amplifier from external DC voltages  Avoid inadvertently  selecting DC Coupling if 
13. D section of the Waveform  Acquisition subsection in the mainframe User Reference manual     Pressing the probe ID button can initiate one or more of the following  operations     1  Present a new display  or if that channel is already displayed  select the  existing trace  Pressing the ID button of a probe connected to an  undisplayed left or center compartment channel will cause that channel to  be displayed  unless doing so would exceed the maximum number or  traces  Unlike the 11A52 display on off button  pressing the probe ID  button a second time will not remove the display     Probe ID button presses for displayed channels will do two things  a  select    the trace s  using this channel  and b  momentarily brighten all traces  using this channel     11A52 User Reference Supplement   2 7    Operating the 11A52 in 11300 series mainframes    2  Stored settings can be sequentially recalled  Pressing the probe ID button  can cause a sequential recall of stored settings  The stored settings feature  must be enabled using the Probe ID Utility menu  Settings must be stored  as explained under STORE RECALL Major Menu in the 11301 and 11302  User Reference Manual  or the 11301A and 11302 A User Reference Manual     3  The mainframe can Autoset to accommodate the input signal introduced  by the probe  The Autoset feature can be enabled or disabled using the  Probe ID Utility menu  Autoset is the automatic setup of vertical deflection  factor  triggering  and sweep speed to produc
14. N TO NEXT  WAVEFORM IN LIST   AFTER SELECTION   KNOBS CONTROL ID N  PARAMETERS    NOTE BOTH THE VERTICAL DESCRIPTION AND THE HORIZONTAL    DESCRIPTION MUST BE CHECKED FOR THE OCCURRENCE OF ID N     C61 14 303    Figure 3 2  Sequence of decisions used to select a new waveform in response to  probe ID button     11A52 User Reference Supplement       Operating the 11A52 in 11400 series Mainframes        Front Panel Error Messages    Message     Cause     Message     Cause     Internal DAC overflow on channel N of LEFT  CENTER  or RIGHT  plug in unit  where N   1 or 2     Some plug in unit detected that a requested setting overflowed an  internal DAC  Such overflow usually indicates defective hardware   In this situation  the plug in unit sets the DAC to the limit nearer  the requested setting       Bad Level 2 probe checksum on channel N of LEFT  CENTER  or  RIGHT plug in unit  where N   1 or 2     Some plug in unit detected that a Level 2 TEKPROBE had failed or  been improperly connected        11A52 User Reference Supplement    Operating the 11A52 in 11400 series Mainframes    GPIB and RS 232 C Commands and Syntax    The following commands set parameters of a specified channel   Table 3 2  Channel Parameter Commands   for 11400 series Oscilloscopes     Header Link Argument    TF                        CH lt L1  L2  C1  a  DC   OFF  C2 R1  or R2 gt   loser      lt NRx gt   Resolution Step Size  Volts  div Offset   via RS 232 C  GPIB    via front panel  Range jor Numeric Entry  
15. Report  PR  for each problem     II  SECTION A  Fill in the instrument configuration table  including all plug in units  modules  and probes installed in the  oscilloscope  The information can be found in one of the menus under the UTILITY major menu     Instrument  Write the instrument name  e g   11301  11401  11A32  etc       Section  Microprocessor Subsystem   If the instrument has more than one section write the section name   For plug ins  write in which mainframe slot they are located  e g   left  center  right      ROM  Version   Write the rom version number for each instrument and section  Instruments with more than  one section will have more than one version number     ID   Serial Number   Write the serial number for each instrument  At the factory the ID number is pro   grammed to be the instrument   s serial number  If this value has been changed  please write the serial number  physically attached to the instrument  mainframes  front panel  plug in units  top rail      IE  SECTION B  Use the complete company mailing address  Include the name and phone number of the person reporting the  error  Also  be sure to fill in the name of the person submitting the PR     IV  SECTION C  Check the reason for the report and whether the problem is reproducible  We cannot fix a problem when we  cannot reproduce the problem condition     V  SECTION D  Give a complete description of the system configuration on which the problem occurred  Include related  peripherals  interfaces
16. TE K User Reference Part No  070 6114 01  Supplement Product Group 47    11A52  Two Channel  Amplifier    Serial Number    Please check for CHANGE INFORMATION at the rear of this manual     Second Edition AUG 1988 Tekt ronix    COMMITTED TO EXCELLENCE    INSTRUMENT SERIAL NUMBERS    Each instrument has a seria  number on a panel insert  tag  or stamped on the chassis  The first number or  letter designates the country of manufacture  The last five digits of the serial number are assigned sequen   tially and are unique to each instrument  Those manufactured in the United States have six unique digits   The country of manufacture is identified as follows     B000000 Tektronix  inc   Beaverton  Oregon  USA   100000 Tektronix Guernsey  Lid   Channel Islands   200000 Tektronix United Kingdom  Ltd   London   300000 Sony Tektronix  Japan   700900 Tektronix Holland  NV  Heerenveen  The Netherlands    Copyright   Tektronix  inc   1987 1988  All rights reserved  Tektronix products are covered by U S  and  foreign patents  issued and pending  TEKTRONIX  INC  SCOPEMOBILE and i are registered  trademarks  Printed in U S A     11A52 Features and Functions       Features   DC to 600 MHz bandwidth  in 11402        Dual trace    Calibrated sensitivities from 1 mV to 10 V division    50 Q input impedance      High resolution  calibrated DC offset  0 25 division increment  coarse  0 025  division increment  fine       Fast overdrive recovery  Functions Signals applied to the CH 1 and CH 2 input
17. ance Requirement       Display Accuracy See Deflection Factor Accuracy  which  Offset precedes this characteristic     Range  Resolution depend on  Deflection Factor  as follows     Between 1 mV and 99 5 mV  div   1 V  Coarse and fine resolution are 0 25 div    250 pV  and 0 025 div   25 uV   respectively     Between 0 1 V and 0 995 V  div   10 V  Coarse and fine resolution are 0 25 div    2 5 mV  and 0 025 div   250 uV   respectively          Between 1 V and  10 0 V div   100 V  Coarse and fine resolution are 0 25 div      25 mV  and 0 025 div   2 5 mV  respectively     Display High Frequency Limit    3 dB point    and Calculated Rise Time of  Had ne mespense Display  Auxiliary   amp  Trig signals   0   to   35   C   Type of Mainframe    Volts Division 11301 A    11302 A    11401 11402    350 MHz   400 MHz   500 MHz   600 MHz  1 0 ns 0 9 ns 0 7 ns 0 6 ns    5 mV 9 95 mV 300 MHz   350 MHz   350 MHz   400 MHz  1 2 ns 1 0 ns 1 0 ns 0 9 ns   2 mV   4 98 mV 250 MHz   250 MHz   250 MHz   250 MHz  1 4 ns 1 4 ns 1 4 ns 1 4 ns   1 mV 1 99 mV 200 MHz   200 MHz   200 MHz   200 MHz  1 8 ns 1 8 ns 1 8 ns 18 ns    High Frequency  3 dB    100 MHz Limit 100 MHz 430      20 MHz Limit 20 MHz  30      Low Frequency  3 dB point 1 kHz maximum from 50 Q source   VDC  lt 20 V      11A52 User Reference Supplement 4 3    Specification       Circuit    Display    Step Response     0   to 35      Input    4 4    C       Impedance  AC coupled     TABLE 4 1  cont    Electrical Characteristics    Charact
18. ces  If you find instrument damage or deficiency  contact  your local Tektronix Field Office or representative     Operating Temperature    The 11A52 can be operated where the ambient air temperature is between 0   and   50   C and can be stored in ambient temperatures from  40   to  75   C  After  storage at temperatures outside the operating limits  allow the chassis to reach  operating temperature limits before applying power     11A52 User Reference Supplement r 1 3    General Information    Installing and Removing the 11A52    To avoid instrument damage set the mainframe ON STANDBY switch  to STANDBY before installing or removing the 11A52        If the green indicator light remains ON  when the STANDBY position  is selected  the switch has been internally disabled     To remove or reinstall plug in units  use the mainframe PRINCIPAL  POWER SWITCH  rear panel  to shut OFF the power  This will  prevent damage to either the plug in or the mainframe     To once again enable the ON STANDBY switch  refer the instrument  to qualified service personnel     When installed in any 11401 or 11402 plug in compartment  or in the 11301   11301A  11302 or 11302A left or center plug in compartment  the 11A52 will  provide a conventional display     When installed in the center or right plug in compartment of the 11301  113014   11302 or 11302A  the 11A52 will also provide the X  horizontal  part of an X Y  display  or provide a trigger signal for the mainframe time base     To install t
19. d references for each trace on separate _  graticule lines  Position is a screen related function  its units are divisions     The offset function  accessed by pressing the VERTICAL OFFSET button   subtracts a precision voltage from the input signal  Changing offset moves the  trace just as does the position control  However  if the deflection factor is  subsequently changed  the effect is different     Changing the sensitivity will increase or decrease the size of the display around  some screen level  e g   two divisions above graticule center  set by the Position  control  When using a sensitivity that makes the displayed waveform larger than  the screen  the Offset control is used to bring the waveform area of interest to the  screen location established by the Position control     11A52 User Reference Supplement    HE Limit    Operating the 11A52 in 11300 series mainframes    The Position control has a range of plus and minus four divisions from graticule  center  but the Offset control has a range defined in volts  Offset can be as much  as 1000 divisions at 1 mV division  The Offset control range is one volt for all  sensitivities from 1 mV to 99 5 mV division  but increases to 10 volts for  sensitivities from 100 mV to 995 mV division  For sensitivities from 1 volt to   10 volts division  Offset control range increases to 100 volts     The 11A52 attempts to maintain the user selected offset voltage even though the  offset range changes due to a change in sensitivity  
20. e a meaningful display  e g    two to five divisions of vertical deflection and two to five repetitions of the  input signal  For more information  refer to Autoset in the 11301 and 11302  User Reference Manual  or the 11301A and 11302A User Reference Manual     4  The mainframe can automatically measure the selected trace  Automatic  measurements of the selected trace can be initiated by pressing a probe ID  button  The Automatic Measurements feature can be enabled or disabled  using the Probe ID Utility menu  Such measurements are peak to peak   maximum  middle  and minimum voltages  frequency  period  pulse width   and duty cycle  For full information about automatic measurements  see  Measure in the 11301 and 11302 User Reference Manual  or the 11301A and  11302A User Reference Manual     5  An interrupt to the GPIB and RS 232 C can be generated  Pressing a  probe ID button will cause the mainframe to produce an SRQ to the GPIB  and RS 232 C  For more information  refer to the GPIB RS 232 C section of  the 11301 and 11302 User Reference Manual  or the 11301A and 11302A User  Reference Manual     Front Panel Error Messages    Message     Cause     Message     Cause     Message     Cause     Internal DAC overflow on channel N of LEFT  CENTER  or RIGHT  plug in unit  where N   1 or 2     Some plug in unit detected that a requested setting overflowed an  internal DAC  Such overflow usually indicates defective hardware   In this situation  the plug in unit sets the DAC to the 
21. el  The trigger  auxiliary    _ trigger  and display signal bandwidths for a channel are always the same  The    _ 3 2    auxiliary trigger is the signal sent to the right plug in compartment     11A52 User Reference Supplement    Operating the 11A52 in 11400 series Mainframes        C  DC Circuit Loading    AC coupling capacitors are connected differently in the 11A52 and 11A71 than in  the 11A32 and 11A34  Figure 3 1 shows this difference     In the 11A52 and 11A71  the coupling capacitor isolates both the input  termination and the amplifier from external DC voltages  Avoid inadvertently  selecting DC coupling if your circuit will be damaged by the resulting 50 Q  termination to ground  See Caution below  The coupling capacitor in the 11A32  and 11A34 does not isolate the input termination from external DC voltages     When AC Coupling is selected  the 11A52 coupling capacitor allows the user to  test low impedance circuits that are elevated in potential by up to 100 volts DC  without loading the circuit at DC  Passive attenuating probes such as the  P6056 or P6057 eliminate the DC loading and lower the low frequency    3 dB point     In the Off Coupling mode the AC coupling capacitor precharges in about three  seconds to the average DC voltage of the circuit being tested  Always use the  Off mode when connecting the 11A52 input to a circuit that may have more than  25 volts  the 11A52 peak input voltage rating  present  This will precharge the  coupling capacitor  Wait a 
22. erent plug in units due to probe  cable length differences are normally removed by deskewing the probes  and the associated arnplifier channel  Samples are still taken simultan   eously  but the display is manipulated to remove time differences  Signals  correlated to the trigger signal are correctly restored to their proper time  relationship  but there is no way for signals unrelated to the trigger to be  corrected with this deskewing technique  High frequency rejection of  signals unrelated to the trigger is not improved by deskewing     The solution to both problems is to invoke averaging  The averaging process  suppresses all signals and noise unrelated to the trigger        11A52 User Reference Supplement   3 5    Operating the 11A52 in 11400 series Mainframes    Overdrive Recovery    Overdrive occurs when any 11A52 channel is driven out of its linear range of  approximately  15 divisions     The 11A52 has extraordinarily good overdrive recovery  and this feature may be  used to greatly extend measurement resolution  For example  suppose a signal  changes from  1 7 V to  0 8 V in 1 ns  The 11A52 could be used to determine if  the signal stabilized immediately at  0 8 V or if perhaps had some small  aberration following the transition  By setting the 11A52 offset to  0 8 V and the  sensitivity to 1 mV division  aberrations of just 0 1  of the original transition  will be 2 5 divisions in amplitude  0 1  of 2 5 V is 2 5 mV or 2 5 divisions at   1 mV division      Any ampl
23. eristic Performance Requirement    Overdrive Recovery Time                   1 mV to 99 5 mV div    for signals up to  2 V peak      lt 20 ns to within  0 1  of signal   0 2 division           100 mV to 995 mV div    for signals up to  20 V peak      lt 20 ns to within  0 1  of signal     0 2 division         1 V to 10 V div   lt 20 ns to within  0 1  of signal     for signals up to  25 V peak   0 2 division    Maximum Input Voltage  5 Vrms  0 5 W  or 0 25 watt second pulses  AC or DC Coupled not exceeding 25 V peak   Maximum DC Input Voltage   100 V  DC peak AC    AC Coupled    Signals of more than 25 V peak amplitude must be connected with the  input coupling set to OFF so that the input coupling capacitor is  precharged     Input Disconnect Threshold 4 Vrms minimum to 6 V rms maximum   Power Down Condition 50 Q terminations disconnect when 11A52  is powered down     Impedance  DC coupled  50 Q within 0 5   VSWR  lt  1 25 1 from DC  to 500 MHz           502  1  in series with nominally 2 2 UE  with 500 kQ to ground  VSWR  lt  1 25 1  from 100 kHz to 500 MHz     Input Bias Current 4 0 LA maximum     Cable losses greatly increase the overdrive recovery time as cables are lengthened  See the  discussion on pages 2 6 and 3 6     11A52 User Reference Supplement    Specification       A x TABLE 4 1  cont         Electrical Characteristics    Circuit Performance Requirement  Miscellaneous Typical Noise     1 0 mV to 1 99 mV div 0 087 div  rms    2 0 mV to 4 98 mV div 0 04 div  rms
24. f perhaps had some small  aberration following the transition  By setting the 11A52 offset to  0 8 V and the  sensitivity to 1 mV division  aberrations of just 0 1  of the original transition  will be 2 5 divisions in amplitude  0 1  of 2 5 V is 2 5 mV or 2 5 divisions at 1  mV   division      Any amplifier will ultimately reach an equilibrium value after an input step   although its accuracy will determine how far that equilibrium value is from the  correct value   The 11A52   s ability to settle quickly to within a very small  fraction of its equilibrium value is exceptional  The time it takes the 11A52 to  settle to within a stated fraction of the equilibrium value is its overdrive  recovery time     Measuring the overdrive recovery time of an 11A52 takes some care and can  lead to some surprising results  An interesting experiment is to use a very flat  pulse generator  such as the Tektronix PG 506  to pulse the 11A52  Connect the  generator   s fast rise output to the 11A52 input through a short  one foot or less   coaxial cable  Adjust the pulse amplitude to 1 volt  Set the 11A52 sensitivity to 1  mV division  Each division now represents 0 1  of the input signal  Trigger the  oscilloscope and observe the recovery of the 11452 using 50 ns division sweep  rate  Now increase the cable length by about three feet  for example  use a 42   inch length of RG 58 cable  Tektronix part 012 0057 01  and observe the new  waveform     The waveform change is due to skin effect loss i
25. few seconds before selecting AC Coupling     The coupling capacitor is discharged upon removal of the external circuit     Always use caution when working with voltages in excess of 25 volts     When the input coupling is set to DC  a 50 Q termination resistance is  connected directly from the 11A52 input connector to ground  Take  care that the circuit connected to the 11A52 input will not be damaged  by the 50 Q load     Two ways of unintentionally invoking DC coupling in the 11A52 are     1  Invoking Autoset  because the Autoset process starts by searching  for a DC voltage  and    2  Recalling a stored control setting that dictates DC Coupling   Switching coupling to DC when more than 25 V is present at the  11A52 input will exceed the peak input voltage specification  and may    damage the 11A52 input relay  A damaged relay could cause an error  in calibration     11A52 User Reference Supplement 3 3    Operating the 11A52 in 11400 series Mainframes    CH    11452  11A71    500  11A32  11434    Figure 3 1  Location of AC coupling capacitors in plug in amplifiers     06114 2083       3 4 11A52 User Reference Supplement    Operating the 11A52 in 11400 series Mainframes    Adding and Subtracting Waveforms    The 11401 and 11402 mainframes do not allow the addition or subtraction of live  waveforms without restriction  The reason for this lies in the manner in which  samples are taken and used by the mainframe  Each plug in compartment has  an associated sampler in the 11401 
26. he 11A52 in any 11000 series oscilloscope mainframe  set the  mainframe ON STANDBY switch to STANDBY  Align the grooves in the top  and bottom of the 11452 with the guides in the mainframe plug in  compartment  then insert the 11A52 into the mainframe until its front panel is  flush with the front panel of the mainframe     To remove the 11A52 from a mainframe  set the mainframe ON STANDBY  switch to STANDBY  Then pull the release latch  see Figure 1 1  to disengage the  unit from the mainframe  and pull the 11A52 straight out of the plug in  compartment     1 4 11A52 User Reference Supplement    General Information    TOP GROOVE  MAINFRAME    RELEASE  LATCH    C6114 101       Figure 1 1  Installing a plug in unit in a mainframe oscilloscope     11A52 User Reference Supplement 1 5    General Information    Instrument Options    Option 25 includes two P6231 probes     Packaging for Shipment    1 6    If this instrument is to be shipped by commercial transportation  we recommend  that it be packaged in the original manner  The original carton and packaging  material should be saved and reused for this purpose     NOTE    Package and ship plug in units and mainframes separately     If the 11A52 is to be shipped to a Tektronix Service Center for service or repair   attach a tag to the instrument  On the tag  include the following information     e Name and address of the instrument owner     e Name of a person at your firm who can be contacted about the instrument     Complete in
27. hen  operation 2 or 3  which involve a new waveform and waveform selection   will occur     If the Sequence Settings option is enabled  front panel settings can be  sequentially recalled via the Probe ID function  The settings must be stored  as explained under Storing and Recalling Waveforms in the 11401 and 11402  User Reference Manual     11A52 User Reference Supplement 3 7    Operating the 11A52 in 11400 series Mainframes    2  If the Probe ID Function  in the Probes pop up menu of the Utility major  menu  is set to Waveform Select New Wfm  a waveform will be selected   Figure 3 2 shows the sequence of decisions used to select a new waveform     3  If the Probe ID Function is set to Wim Select New Wim  amp  AutoSet  then    the decisions shown in Figure 3 2 will be made  and the resulting selected  waveform is autoset     Events 1  2  and 3 are mutually exclusive  Event 4 may be combined with any  one of them     4  If the IDPROBE SRQMASK is enabled  an appropriate event will be  returned to the GPIB and RS 232 C ports     PROBE ID N ACTUATED    SINGLE CHANNEL  WAVEFORM EXISTS  CORRESPONDING TO  1D N     SINGLE COMBINATION  WAVEFORM EXISTS  WITH IDNA  COMPONENT     MULTIPLE WAVEFORMS  EXIST WITH ID N  A COMPONENT     DEFINE WAVEFORM WITH  ID N AS DEFAULT  WAVEFORM    SINGLE CHANNEL  WAVEFORM SELECTED  CORRESPONDING TO  ID N  KNOBS CONTROL  ID N PARAMETERS    COMBINATION  WAVEFORM SELECTED  CORRESPONDING TO  ID N  KNOBS CONTROL  iD N PARAMETERS    TOGGLE WAVEFORM  SELECTIO
28. his Supplement contains the following four sections     Section 1   GENERAL INFORMATION  describes mainframe to plug in unit  compatibility  explains how to install and remove the 11A52 from a mainframe   outlines any options available for the 11A52  and gives details about packaging  for shipment     Most 11A52 functions  and their use  are described in the Operating Information  section of the mainframe User Reference manuals  These functions are those  common to all amplifier plug in units  Examples of common functions are offset   sensitivity  input impedance  where selectable   coupling  and bandwidth limit   Only functions unique to the 11A52 11300 series or 11A52 11400 series  combinations are explained in the following sections     Section 2   OPERATING THE 11A52 IN 11300 SERIES MAINFRAMES  explains  how to operate those 11A52 functions not covered in the 11301 and 11302 User  Reference Manual  or the 11301A and 11302A User Reference Manual     Section 3  OPERATING THE 11A52 IN 11400 SERIES MAINFRAMES  explains  how to operate those 11A52 functions not covered in the 11401 and 11402 User  Reference Manual     Section 4   SPECIFICATION  gives detailed specifications of all 11A52  mainframe oscilloscope combinations     11A52 User Reference Supplement 1 1    General Information    11A52 Service Manual    The 11A52 Service manual is for use by qualified service personnel  only  To avoid personal injury  do not perform any service other than  that contained in the Operato
29. ifier will ultimately reach an equilibrium value after an input step   although its accuracy will determine how far that equilibrium value is from  the correct value   The 11A52   s ability to settle quickly to within a very small  fraction of its equilibrium value is exceptional  The time it takes the 11A52 to  settle to within a stated fraction of the equilibrium value is its overdrive  recovery time       Measuring the overdrive recovery time of an 11A52 takes some care and can  lead to some surprising results  An interesting experiment is to use a very flat  pulse generator  such as the Tektronix PG 506  to pulse the 11A52  Connect the  generator   s fast rise output to the 11A52 input through a short  one foot or less   coaxial cable  Adjust the pulse amplitude to 1 volt  Set the 11A52 sensitivity to   1 mV division  Each division now represents 0 1  of the input signal  Trigger  the oscilloscope and observe the recovery of the 11A52 using 50 ns division  sweep rate  Now increase the cable length by about three feet  for example    use a 42 inch length of RG 58 cable  Tektronix part 012 0057 01  and observe the  new waveform     The waveform change is due to skin effect loss in the longer cable  What is  surprising is that the skin effect loss persists for over 200 ns even though the  total cable delay is only 5 ns  At 400 ns the loss is still 0 02      This experiment shows the importance of using a short cable to test overdrive  recovery     3 6 11A52 User Reference Su
30. limit nearer  the requested setting     Bad Level 2 probe checksum on channel N of LEFT  CENTER  or  RIGHT plug in unit  where N   1 or 2     Some plug in unit detected that a Level 2 TEKPROBE had failed or  been improperly connected     Requested Coupling not available on Channel N of LEFT   CENTER  or RIGHT plug in unit  where N   1 or 2     Some plug in unit had a Level 2 probe  capable of offset  connected  when either Autoset or Probe Cal was invoked     11A52 User Reference Supplement    Operating the 11A52 in 11300 series mainframes      GPIB and RS 232 C Commands and Syntax    The following commands set parameters of a specified channel     TABLE 2 2    Channel Parameter Commands   for 11300 series Oscilloscopes            A CESR  CH lt L1  L2  C1    COUPLING  ACIDCIOFF    C2  R1 or R2 gt      Resolution Step Size  Volts div Offset   via RS 232 C  GPIB    via front panel  Range jor Numeric Entry   Control knob  ImVto995mV  41V_   25E 6 Coarse  0 25 div          1V to10V    100 mV to 995 mV    10 V       250E 6 Fine  0 025 div      100 V   2 5E 3               BWHI     lt NRx gt  HF Limit    24E6 20E6   gt 24E6 to  lt 120E6 100E6   gt 120E6 350E6  11301  400E6  11302  IMPEDANCE   lt NRx gt  Impedance   lt 1E3 50  SENSITIVITY  Sensitivity Resolution  step size   1E 3 to 1 99E 3 10E 6  2E 3 to 4 98E 3 20E 6  5E 3 to 9 95E 3 50E 6  10E 3 to 19 9E 3 100E 6  20E 3 to 49 8E 3 200E 6  50E 3 to 99 5E 3 500E 6   100E 3 to 199E 3 1E 3  200E 3 to 498E 3 2E 3  500E 3 to 995E 3 5E 
31. n the longer cable  What is  surprising is that the skin effect loss persists for over 200 ns even though the  total cable delay is only 5 ns  At 400 ns the loss is still 0 02   The only way to  eliminate all skin effect loss is to connect the PG506 directly to an 11A52 input  connector       This experiment shows the importance of using a short cable to test overdrive  recovery       2 6 11A52 User Reference Supplement    Operating the 11A52 in 11300 series mainframes    50 Q Overload    When the input voltage substantially exceeds 5 Vrms  the 11A52 will sense the  overload and disconnect the 50 Q input termination  The mainframe will display   the message     Input channel N overload on LEFT CENTER RIGHT plug in  where N   1 or 2     To reconnect the 50 Q input termination  first correct the overload condition   then select AC or DC from the Coupling menu     Active Probes    Using an active probe  such as the P6231  eliminates the possibility of AC  Coupling  Coupling options available when using such a probe are Off and DC     Probe ID    The Probe ID section of the Utility Menu is the means of selecting how any  11300 series oscilloscope responds to an ID button push of recommended  probes  All  or some combination of the following operations may be set to start  in response to probe ID buttons  To display the Utility Menu  press the front   panel UTILITY button  once on 11301 or 11302 oscilloscopes  twice on 11301A or  11302A oscilloscopes   For details  see the Probe I
32. nsen esse 2 4  Adding Waveforms sesse    ei ONS EEN TEN Lan  Overdrive RECOVETY smsnenencmeneeneneeneeeeneneesses 26  50 52 Overload se nscsrensnsnnastnnateiusiannentensanenmese 2 7  Active Probes mens Re Vi  Probe ID summer TR TT LEE ca A    A Front Panel Error Messages TITI CE LEE CCC EC ECE ELLE LE seen ns es none ana anonasses  e see passe ene nes 2 8    GPIB and RS 232 C Commands and Syntaxe 2 9    Section 3   Operating the 11A52 in 11400 series Mainframes    Display On Off nn  Other Functions toes cases uscessrsereannctnssnsunsosesssuisrnssesecondesvsendsbansenceaessrassaseueiastentmeneeneive SL  Vertical Offset syed es castcsssncsatascsiedcascententacensndsanavsaloeecssiauts oniri N a i iaaii S E  DC Circuit Loading mener 973  Adding and Subtracting Waveforms sense 3  Overdrive RECOVETY mm sieste merene 9 6  50 Q Overload         0 EE A E E sana Ga aea 3 7  Active Probes  esssesssserserssee  PEEN A A E E ANNA E E aA  Probe ID sinon ET A TE E T EE R E T 3 7  Front Panel Error Messages      sscsscssesossstsesensensssessersansnnsevensanearsesssesnesnenecesoes 3 9  GPIB and RS 232 C Commands and Syntax    esessessesereversoseseessranrreresesresre 3 10    Section 4   Specification    Performance Conditions    e   rseeseesessesereesresssesesavsavsanereessnnnesevroenresresssserssses ETL  Electrical Characteristics       ssscecsesssrscssasssccsceseesesersscessecesescneneanaes 4nd  Environmental Characteristics enr 4 6  Physical Characteristics een hed   Recommended Probes     
33. ny BW Limit  of 11A52 in 11401 11402  Deflection Factor Mainframes    ee Without Probes    Cont                    Voits  Div DC Offset Acc       2 mV   0 10 div     210 8   001 div      20 mV   0 10 div     With P6231 Probe calibrated from 11401 11402 Calibrator output    USE      0 9    0 01 div    2 mV   0 10 div       1 0    0 01 div    20 mV   0 10 div     Coarse Resolution 1  2  5 sequence     TmV to 99 5 mV   0 15    0 4 mV        100 mV to 995 mV   0 20    4 mV        41Vto10 V   0 20    40 mV                     Volts Div       10 mV to 995 mV   0 15    2 mV           0 15    2 mV        Probe Tip TC term    Fine Resolution Depends on Deflection Factor  as follows     Rotating the Fine control one increment    will change the Deflection Factor by 1  of    the next more sensitive Coarse setting     For example  with deflection factor set to  198 mV  rotating the Fine control  counterclockwise will cause this sequence  of sensitivities  199 mV  200 mV  202 mV   etc  Rotating the Fine control clockwise  from 204 mV div will cause the reverse  sequence  202 mV  200 mV  199 mV  etc         For absolute DC accuracy of single point measurements using Offset  and the DC Offset  Accuracy  DC Balance and AV DC Accuracy terms  Apply the AV DC Accuracy only to the  difference between the Vertical Position setting and the measurement point     4 2 11A52 User Reference Supplement    Specification       TABLE 4 1  cont    Electrical Characteristics    Circuit Characteristic Perform
34. or 11402  Samples are taken simultaneously  by each sampler  A channel from one plug in unit may be added to or subtracted  from that of another plug in unit accurately by virtue of this simultaneity  Thus  noise and other signals unrelated to the trigger will be canceled when channels  in different plug in units are subtracted     Signals from different channels within a plug in unit are sampled at different  times  Signals that are entirely repetitive and related to the trigger signal may be  successfully added or subtracted in spite of this  However  signals that are not  repetitive or related to the trigger will not be sampled at the same time  and may  not be displayed correctly  Therefore it is preferable that signals to be added or  subtracted be selected from different plug in compartments     The best common mode rejection is obtained with a differential amplifier plug   in unit because its hardware subtracts the signals at its two inputs  passing only  the result to the mainframe       Even when using channels from different plug in compartments  the following  two problems limit the accuracy of sums or differences     1  Each waveform in memory is constantly being updated  and therefore  consists of some points that are older than others  Although samples may  be taken simultaneously  there is a chance that a correct sample may be  replaced in memory with a newer value before the waveform is processed  and displayed     2  Delay differences between channels of diff
35. owing characteristics     P6231 Low Impedance Subminiature 10X Active Probe  The P6231 has a  bandwidth of 1 5 GHz  input impedance of 450 Q  and DC offset of 5 V  controlled by the 11A52  The P6231 is a Level 2 probe        11A52 User Reference Supplement 4 7    Specification       TOP FIGURES ARE IN MILLIMETERS  AND BOTTOM FIGURES ARE IN INCHES        C6114 401    Figure 4 1  Dimensions of 11A52     en ee EAA EAT se tO _      4 8 11A52 User Reference Supplement    Appendix A   Glossary       Deflection factor  Display on off button    Mainframe    Offset    ON STANDBY    Overdrive  Overload    Toggle    The ratio of input signal to response in the 11000 system  The reciprocal of  sensitivity     The front panel  plug in button that designates a channel for display  or removes  a channel from the display     Any 11000 series oscilloscope exclusive of plug in units     A control that allows the user to subtract a precision voltage from the input  signal to vary the position of the displayed signal     The front panel power control on 11000 series mainframes  When set to ON   applies operating power to all circuits  When set to STANDBY  the mainframe  DC power supply supplies power to the counter timer crystal oven  Option 1T      A condition in which amplifiers are driven into a non linear operating range   Typically many divisions off screen     The condition existing when a potentially damaging voltage is applied to the  input connector     To switch alternately between two
36. pplement    Operating the 11A52 in 11400 series Mainframes    50 Q Overload    When the input voltage substantially exceeds 5 Vrms  the 11A52 will sense the  overload and disconnect the 50 Q input termination  The mainframe will display  the message  l    input channel N overioad on LEFT CENTER RIGHT plug in  where N   1 or 2     To reconnect the 50 Q input termination  first correct the overload condition   then select AC or DC from the Coupling menu     Active Probes    Using an active probe  such as the P6231  eliminates the possibility of AC  Coupling  Coupling options available when using such a probe are Off and DC     Probe ID    The Probe ID function is the means of selecting how the 11401 or 11402  responds to an ID button push of recommended probes   The Probe ID  Function is part of the Probes pop up menu  which is selected from the  Utilities major menu   All  or some combination of the following operations may  be set to start in response to probe ID buttons  To display the Utility major  menu  press the front panel UTILITY button  For details  see the Probe Calibration  and ID Function section of the Utilities subsection in the 11401 and 11402 User  Reference Manual     Pressing the probe ID button can initiate one of the following operations     1  If the Probe ID function  in the Probes pop up menu of the Utility major  menu  is set to Sequence Settings  the next front panel setting will be  recalled  If the Probe ID Function is not set to Sequence Settings  t
37. re       The 11401 and 11402 User Reference Manual contains detailed information about  measurements the instruments can make  and instructions about how to set the  oscilloscope to perform such measurements  Also  installation instructions   specifications  description of options  and details about using the 11401 and 11402  with the GPIB and RS 232 C interfaces are included here       The 11301 and 11302 Pocket Reference is a short form reference  The Pocket  Reference gives menus and a brief description of functions selected  External  interface commands are listed with their basic syntax  and a list of error and  warning messages is included     or    The 11301A and 11302A Pocket Reference is a short form reference  The Pocket  Reference gives menus and a brief description of functions selected  External  interface commands are listed with their basic syntax  and a list of error and   warning messages is included        The 11A52 Service Manual includes information needed to check and adjust     perform diagnostic troubleshooting  repair  and otherwise service your  instrument     e The 11A52 Functional Test Procedure manual verifies that the instrument is  functioning properly     11A52 User Reference Supplement  B 1    Appendix    D a lea a ct a at E OEE E E EENE   B 2   11A52 User Reference Supplement    MANUAL CHANGE INFORMATION    At Tektronix  we continually strive to keep up with latest electronic developments  by adding circuit and component improvements to our in
38. re displayed  select a channel by touching the desired  trace description at the top of the CRT     2  On the 11301 and 11302 oscilloscopes press any VERTICAL button once     OFFSET  SIZE  or POS  On the 11301A and 11302A oscilloscopes press any  VERTICAL button twice  This will cause the Control Menu to be displayed     3  Touch the desired function   s label  Successive touches will change its state     11A52 User Reference Supplement     2 1    Operating the 11A52 in 11300 series mainframes    Other Functions    Other 11A52 functions are controlled within the 11300 series mainframe  and  their operation is described in detail in Section 2  Operating Information  of the  11301 and 11302 User Reference Manual  and the 11301A and 11302A User Reference  Manual  Table 2 1 shows where to find information about operating the 11A52     TABLE 2 1  Functions and Where They are Described  Function _ Described Under Heading  Offset  Waveform Acquisition  Vertical Menu   Size  V div  or for X Y Display  Horizontal Menu     Display  Ch 1   Ch 2 Waveform Acquisition  Vertical Menu     Polarity Waveform Acquisition  Waveform Menu   Trigger Selection Trigger Source Major Menu   and Polarity    Position and Offset    All 11300 series oscilloscopes assign position control to the Left Control knob by  pressing the VERTICAL POSition button once  The Vertical Position control  moves the trace as a user convenience  For example  when displaying multiple  channels it may be desirable to set groun
39. rs manual unless you are qualified to do  so  Refer to the Operators Safety Summary and Service Safety  Summary before performing any service    The 11A52 Service Manual contains the following information    Section 1   GENERAL INFORMATION    Section 2   THEORY OF OPERATION    Section 3   MAINTENANCE AND DIAGNOSTICS    Section 4   PERFORMANCE VERIFICATION AND ADJUSTMENT    Section 5   INSTRUMENT OPTIONS      Section 6   REPLACEABLE ELECTRICAL PARTS    Section 7   SCHEMATIC DIAGRAMS  amp  CIRCUIT BOARD ILLUSTRATIONS     Section 8   REPLACEABLE MECHANICAL PARTS     Plug in to Mainframe Compatibility    The 11A52 is designed for use in the 11301  11301A  11302  11302A  11401  11402  and future 11000 series plug in mainframes  11A52 bandwidth varies depending  on the host mainframe  Details about bandwidth are included in Section 4    Specification  of this supplement  and in the Tektronix Corporate Catalog  Refer to  the Tektronix Corporate Catalog for complete compatibility information     1 2 11A52 User Reference Supplement    General Information       This instrument was inspected mechanically and electrically before shipment   and should meet all electrical specifications  First  inspect the 11A52 for physical  damage incurred in transit  Secondly  to verify that the instrument is functioning  properly  perform the Functional Test Procedure  shipped with the instrument   This brief procedure verifies most instrument functions and checks the internal  auto calibration referen
40. s starts by searching  for a DC voltage  and    2  Recalling a stored control setting that dictates DC Coupling     Switching coupling to De when more than 25 V is present at the  11A52 input will exceed the peak input voltage specification  and may  damage the 11A52 input relay  A damaged relay could cause an error  in calibration     _ Adding Waveforms    Before adding waveforms  check that each channel s display is  independently on screen     The 11300 series oscilloscopes allow the addition of any two channels  That is   one channel may be added to a channel from another plug in unit  A simple  restriction applies  Each channel must be in its linear operating range  This is  assured if each channel separately is within the screen area before addition     Those portions of a trace which are off screen will not be valid when brought    back on screen using another channel s input signal or Offset control  This  general restriction applies to any dual channel oscilloscope     11A52 User Reference Supplement 2 5    Operating the 11A52 in 11300 series mainframes    Overdrive Recovery    Overdrive occurs when any 11A52 channel is driven out of its linear range of  approximately  15 divisions     The 11A52 has extraordinarily good overdrive recovery  and this feature may be  used to greatly extend measurement resolution  For example  suppose a signal  changes from  1 7 V to  0 8 V in 1 ns  The 11A52 could be used to determine if  the signal stabilized immediately at  0 8 V or i
41. strument type and serial number  and    e A description of the service required     If the original package is not available or is not fit for use  package the 11A52 as  follows       1     Obtain a corrugated cardboard carton with inside dimensions at least six  inches greater than the instrument dimensions  Use a carton with a test  strength of at least 200 pounds       Fully wrap the 11A52 with anti static sheeting  or its equivalent  to protect    the finish       Cushion the 11A52 on all sides by tightly packing dunnage or urethane    foam between the carton and the instrument  Allow three inches of packing  on each side       Seal the carton with shipping tape or with industrial staples       Mark the address of the Tektronix Service Center and your return address    on the carton in one or more prominent places     11A52 User Reference Supplement    Section 2  Operating the 11A52 in  11300 series Mainframes       Display On Off    The 11A52 has only two front panel controls   the display on off buttons   Pressing a display on off button will cause its channel  and signal  to be  displayed or removed from the display  the function toggles   When a channel i is  displayed  its label  CH 1 or CH 2  will be lighted     Selecting Coupling  HF Limit  and  Impedance    To select the desired setting for Coupling  HF Limit  or Impedance  proceed as  follows     la  If no trace is displayed  press any plug in display on off button to create a  trace       1b  If several traces a
42. struments as soon as they  are developed and tested     Sometimes  due to printing and shipping requirements  we can   t get these  changes immediately into printed manuals  Hence  your manual may contain new  change information on following pages     A single change may affect several sections  Since the change information sheets  are carried in the manual until all changes are permanently entered  some  duplication may occur  if no such change pages appear following this page  your  manual is correct as printed                                    A    SEND TO  TEKTRONIX  INC     LAB INSTRUMENTS MARKETING SUPPORT 11000 SERIES OSCILLOSCOPE    P O  BOX 500  DEL STA 39 327 PROBLEM REPORT  BEAVERTON  OREGON 97077    INSTRUMENT CONFIGURATION       Instrument   ROM  Version  ID    Serial Number     REASON FOR REPORT    Option Information               COMPANY NAME     USER  L Hardware Mechanical Problem  ADDRESS  M Software Firmware Problem  1  O Documentation Problem   PHONE  nn EATEN SION  L  Suggested Enhancement C    IS THE PROBLEM REPRODUCIBLE     SUBMITTED BY  uu DATE  O Yes G No Q Intermittent       SYSTEM DESCRIPTION   Hardware  software  firmware and host related to the problem     ST ERE PRT ERTS R IT STM ME NISRA NO NN CNET TE PET VOX VC NEC EIN NE EL EEE EAN EEE    DESCRIPTION OF PROBLEM     LIST ENCLOSURES        ronix    COMMITTED TO EXCELLENCE    250 065 1 00    Instructions For Completing The Problem Report    F  Please type or print clearly  Use a separate Problem 
43. the calibration was  performed  When the 5   C change does occur the accuracy condition becomes  Not Enhanced  In the Not Enhanced condition those Characteristics that are  temperature sensitive may not remain within the limits of these specifications   The instrument specifications are valid at an ambient temperature of 0   to 50  C   unless otherwise stated     TABLE 4 1  Electrical Characteristics  Characteristic Performance Requirement    Calibrated Range 1 mV to 10 V div     Enhanced DC Accuracy     either polarity  any HF Limit of 11A52 in 11300 series  Mainframes  with on screen cursors        Without Probes                    Volts Div DC Offset Acc     AV DC Acc  DC Balance    1 0    0 04 div    0 20 mV   0 13 div         1 0    0 04 div   26 mV   0 13 div         1 mV to 99 5 mV   0 15    0 4 mV              100 mV to 995 mV   0 20    4 mV            1Vto10V   0 20    40 mV        With P6231 Probe calibrated from 11300 series Calibrator output    Volts Div DC Offset Acc   10 mV to 995 mV   0 15    2 mV     1 For absolute DC accuracy of single point measurements using Offset  and the DC Offset  Accuracy  DC Balance and AV DC Accuracy terms  Apply the AV DC Accuracy only to the  difference between the Vertical Position setting and the measurement point        11A52 User Reference Supplement 4 1    Specification       TABLE 4 1  cont    Electrical Characteristics    Circuit Characteristic Performance Requirement       Display Enhanced DC Accuracy     either polarity  a
44. your circuit will be damaged by the resulting 50 Q  termination to ground  See Caution below  The coupling capacitor in the 11A32  and 11A34 does not isolate the input termination from external DC voltages     When AC Coupling is selected  the 11A52 coupling capacitor allows the user to  test low impedance circuits that are elevated in potential by up to 100 volts DC  without loading the circuit at DC  Passive attenuating probes such as the P6056  or P6057 eliminate the DC loading and lower the low frequency  3 dB point     In the Off coupling mode the AC coupling capacitor precharges in about three  seconds to the average DC voltage of the circuit being tested  Always use the Off  mode when connecting the 11A52 input to a circuit that may have more than 25  volts  the 11A52 peak input voltage rating  present  This will precharge the  coupling capacitor  Wait a few seconds before selecting AC coupling     The coupling capacitor is discharged upon removal of the external circuit     11A52 User Reference Supplement    Operating the 11A52 in 11300 series mainframes    Always use caution when working with voltages in excess of 25 volts     When the input coupling is set to DC a 50 Q termination resistance is  connected directly from the 11A52 input connector to ground  Take  care that the circuit connected to the 11A52 input will not be damaged  by the 50 load     Two ways of unintentionally invoking DC Coupling in the 11A52 are     1  Invoking Autoset  because the Autoset proces
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Bank Audi Merchant Administration User Manual  レイアウト 3  Philips Pressurised steam generator GC8320  Origin Storage 250GB 7200RPM Enigma FIPS Desktop Drive  MegaPro Drums TE3.2 User's Manual  SteamFast SF-210WH Use and Care Manual  CPAT FLEX Operation Manual (ARD4/DRV1/DRV2/ITX1)  Ambulatory Blood Pressure Monitors  Gigabit 1000Base-T  Nouveautés Pierron Education Collège Sciences de la Vie et de la    Copyright © All rights reserved. 
   Failed to retrieve file