Omron M3 User's Manual

image

Contents

1. Do not subject the main unit and the cuff to extreme temperatures humidity moisture or direct sunlight Do not fold the cuff or tubing tightly Do not inflate the arm cuff over 299 mmHg Do not disassemble the unit Do not subject the unit to strong shocks or vibrations for example dropping the unit on the floor Do not use volatile liquids to clean the main unit Do not wash the arm cuff or immerse it in water Do not use petrol thinners or similar solvents to clean the arm cuff Do not carry out repairs of any kind yourself If a defect occurs consult your OMRON retail outlet or distributor as mentioned on the packaging e O Qs The unit should be cleaned with a soft dry cloth Use a soft moistened cloth and soap to clean the arm cuff Calibration and Service The accuracy of this blood pressure monitor has been carefully tested and is designed for a long service life It is generally recommended to have the unit inspected every two years to ensure correct functioning and accuracy Please consult your authorised OMRON dealer or the OMRON Customer Service at the address given on the packaging or attached literature 4 Troubleshooting and Maintenance 4 4 Storage Keep the unit in its storage case when not in use 1 Unplug the air tube from the air connector 2 Gently fold the air tube into the arm cuff Note Do not bend the air tube excessively 3 Place the arm cuff
2. LT 4 Troubleshooting and Maintenance 12 4 1 The Icons and Error Messages 12 4 2 Troubleshooting ssse 14 4 3 Maintenance ssssssssss 16 4 4 Storage c ios dioe ded STRE 17 5 Optional Parts ssssse 18 6 Technical Data ee 19 Please read this instruction manual thoroughly before using the unit Please keep for future reference For specific information about your own blood pressure CONSULT YOUR DOCTOR GQ Intetti SV sense a E uO coco omnon CODO d EET LII E oo Ne SQ A i MEMORY 3 044 ea oo Dm ue al a O I e 5 oe E m Important Safety Information Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia or arteriosclerosis Please read this section carefully before using the unit A Warning Indicates a potentially hazardous situation which if not avoided could result in death or serious injury General Usage Always consult your doctor Self diagnosis of measurement results and self treatment are dangerous People with severe blood flow problems or blood disorders should consult a doctor before using the unit as cuff inflation can cause internal bleeding AC Adapter Usage Never plug in or unplug the power cord from t
3. 40 to 180 min Pressure 3 mmHg Pulse 5 of display reading Fuzzy logic controlled by electric pump Automatic pressure release valve 60 Measurements with date and time DC6V 4W 4 AA batteries 1 5V or AC DC adapter Adapter S 9515336 9 optional INPUT AC100 240V 50 60Hz 0 12A Capacity of new alkaline batteries is approx 1500 measurements e Type B Internally powered ME equipment When using only the batteries m 7 Class II ME equipment Optional AC adapter 10 C to 40 C Maximum 30 to 90 RH 20 C to 60 C Maximum 10 to 95 RH 700 1060hPa Approximately 340g without batteries Approximately 120g Approximately 123 w mm x 85 h mm x 141 l mm Approximately 146 mm x 446 mm Medium cuff arm circumference 22 to 32 cm Nylon and polyester Main unit cuff instruction manual storage case battery set guarantee card blood pressure pass Note Subject to technical modification without prior notice 19 6 Technical Data 20 C 0197 This device fulfils the provisions of EC directive 93 42 EEC Medical Device Directive This blood pressure monitor is designed according to the European Standard EN1060 Non invasive sphygmomanometers Part 1 General Requirements and Part 3 Supplementary requirements for electromechanical blood pressure measuring systems This OMRON product is produced under the strict quality system of OMRON HEALTHCARE Co Ltd Japan The Core component for OM
4. Disposal of used batteries should be carried out in accordance with the national regulations for the disposal of batteries 2 Preparation 2 2 Setting the Buzzer Date and Time 1 Press the button 2 Set the Buzzer I 4 I gt l on ce 1 Push or p buttons to EE F E Ge select On or Off 0 s EB C 2 Push the button to confirm OC sl On Off 3 Set the monitor to the correct date and time before taking a measurement for the first time l u 4 ani oe GID 4 Push 4 or gt buttons to change R 280037 14 n C 2 Push the button to confirm Year a i SELA 8 m C ee on isi ie yon vic 9 Next setting appears Day 4 r Y 4 Press the O I START button to store the setting Notes If the batteries have been removed for 30 seconds or more the date and time setting will need to be reset If the date and time are not set appears during or after measurement open Open both the front and rear covers to read following The alphabet and number in the cover page correspond to those in the body page 3 1 Applying the Arm Cuff Remove tight fitting clothing or tight rolled up sleeve from your upper arm Do not place the cuff over thick clothes Insert the air plug into the air jack Put your arm through the cuff loop Position the arm correctly The bottom edge of the cuff should be a 1 to 2 cm abo
5. and main unit in the storage case Do not store the unit in the following situations If the unit is wet Locations exposed to extreme temperatures humidity direct sunlight dust or corrosive vapours Locations exposed to vibrations shocks or where it will be at an angle 17 5 Optional Parts Medium Arm Cuff Large Arm Cuff AC Adapter S Arm circumference 22 32 cm Arm circumference 32 42 cm I LU CM2 9513256 6 Model HEM CR24 CL2 9513255 8 Model HEM CL24 Adapter S 9515336 9 CM1 9997578 9 CL 1 9996760 3 Using the Optional AC Adapter 1 Insert the AC adapter plug into the AC adapter jack on the rear side of the main unit 2 Plug the AC adapter into an electrical outlet To disconnect the AC adapter unplug the AC adapter from the electrical outlet first and then remove the AC adapter plug from the main unit 18 6 Technical Data Product Description Model Display Measurement Method Measurement Range Accuracy Inflation Deflation Memory Rating Power Source Battery Life Applied Part Protection Against Electric Shock Operating temperature Humidity Storage temperature Humidity Air pressure Console Weight Cuff Weight Outer Dimensions Cuff Dimensions Cuff Material Package Content Digital Automatic Blood Pressure Monitor OMRON M3 HEM 7200 E V HEM 7200 E8 V LCD Digital Display Oscillometric method Pressure 0 mmHg to 299 mmHg Pulse
6. Check following components Manufacturer OMRON HEALTHCARE Co Ltd V rifier les composants suivants Controleer de volgende onderdelen 53 Kunotsubo Terado cho Muko Kyoto Pr fen Sie folgende Teile des Lieferumfangs llpoBepbre cnenyrouie KoMnrioHeHTbI 617 0002 JAPAN Controllare i componenti indicati di seguito 14G Usa oa giai iCompruebe los siguientes componentes EU representative OMRON HEALTHCARE EUROPE B V Scorpius 33 2132 LR Hoofddorp THE NETHERLANDS REP www omron healthcare com Production facility Oficina de producci n OMRON HEALTHCARE MANUFACTURING VIETNAM CO LTD Site de production Productiefaciliteit Binh Duong Province VIETNAM Produktionsstatte l ipousBoncrBeHHoe nonpa3neneuue Stabilimento di produzione Subsidiary Empresa filial OMRON HEALTHCARE UK LTD Succursale Dochteronderneming Opal Drive Fox Milne Milton Keynes MK15 ODG U K Niederlassung AjouepHaa KomnaHna OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH Italiano Consociata Ax cl Ls uaa John Deere Str 81a 68163 Mannheim GERMANY www omron medizintechnik de OMRON SANT FRANCE SAS Digital Automatic Blood Pressure Monitor 14 rue de Lisbonne 93561 Rosny sous Bois Cedex FRANCE Uniquement pour le march fran ais OMRON Service Apr s Vente N Vert 0 800 91 43 14 Model M3 consommateurs omron sante fr 2 www omron sante fr contact n stru cti on M anu al Made in Vietnam Fabricado en Vietnam CX Intelli Fabriqu
7. RON blood pressure monitors which is the Pressure Sensor is produced in Japan 6 Technical Data Important information regarding Electro Magnetic Compatibility EMC With the increased number of electronic devices such as PC s and mobile cellular telephones medical devices in use may be susceptible to electromagnetic interference from other devices Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation Medical devices should also not interfere with other devices In order to regulate the requirements for EMC Electro Magnetic Compatibility with the aim to prevent unsafe product situations the EN60601 1 2 2007 standard has been implemented This standard defines the levels of immunity to electromagnetic interferences as well as maximum levels of electromagnetic emissions for medical devices This medical device manufactured by OMRON HEALTHCARE conforms to this EN60601 1 2 2007 standard for both immunity and emissions Nevertheless special precautions need to be observed Do not use mobile cellular telephones and other devices which generate strong electrical or electromagnetic fields near the medical device This may result in incorrect operation of the unit and create a potentially unsafe situation Recommendation is to keep a minimum distance of 7 m Verify correct operation of the device in case the distance is shorter Further documentation in a
8. adapter optional designed for this unit Use of unsupported adapters may damage and or may be hazardous to the unit Plug the AC adapter into the appropriate voltage outlet Do not use a multiple tap Do not use the AC adapter if the unit or the power cord is damaged Turn off the power and unplug the power cord immediately Battery Usage If battery fluid should get on your skin or clothing immediately rinse with plenty of clean water Use only four AA alkaline or manganese batteries with this unit Do not use other types of batteries Do not insert the batteries with their polarities incorrectly aligned Replace old batteries with new ones immediately Replace all four batteries at the same time Remove the batteries if the unit will not be used for three months or more Do not use new and used batteries together General Precautions Do not apply strong shocks and vibrations to or drop the unit and arm cuff Do not take measurements after bathing drinking alcohol smoking exercising or eating Do not forcibly bend the arm cuff or bend the air tube excessively When removing the air tube pull on the air plug at the connection with the main unit not the tube itself Do not inflate the arm cuff when it is not wrapped around your arm Do not wash the arm cuff or immerse it in water Read and follow the Important information regarding Electro Magnetic Compatibility EMC in the Technical Data Secti
9. ccordance with EN60601 1 2 2007 is available at OMRON HEALTHCARE EUROPE at the address mentioned in this instruction manual Documentation is also available at www omron healthcare com Correct Disposal of This Product Waste Electrical amp Electronic Equipment This marking shown on the product or its literature indicates that it should not be disposed of with other household wastes at the end of its working life To prevent possible harm to the environment or human health from uncontrolled waste disposal please separate this from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources Household users should contact either the retailer where they purchased this product or their local government office for details of where and how they can take this item for environmentally safe recycling Business users should contact their supplier and check the terms and conditions of the purchase contract This product should not be mixed with other commercial wastes for disposal This product does not contain any hazardous substances 21 22 7 Some Useful Information about Blood Pressure What is Blood Pressure Blood pressure is a measure of the force of blood flowing against the walls of the arteries Arterial blood pressure is constantly changing during the course of the heart s cycle The highest pressure in the cycle is called the Systolic Blood Pressure the lowe
10. d the reading is accurate and reliable 3 Using the Unit 3 3 Taking a Reading Notes To cancel a measurement press the O I START button to turn off the unit and to release the air in the arm cuff Remain still while taking a measurement 1 Press the O I START button The cuff will start to inflate automatically The buzzer will sound when measurement is completed vco If your systolic pressure is more than 220 mmHg After the cuff starts to inflate press and hold the O I START button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure WSTART WINFLATE WDEFLATE WEND J mnm m t Notes The monitor will not inflate above 300 mmHg Do not apply more pressure than necessary Notes During measurement the buzzer if set to on will beep in rhythm with your heartbeat Wait 2 3 minutes before taking another blood pressure measurement Waiting between readings allows the arteries to return to the condition prior to taking the blood pressure measurement A Warning Self diagnosis of measured results and treatment are dangerous Please follow the instructions of your doctor 3 Using the Unit 2 Remove the arm cuff 3 Press the O I START button to turn off the monitor The monitor automatically stores the measurement in its memory It will automatically turn off after five minutes Important Recent research suggest
11. e each day to get an 50 accurate indication of any changes in 6 1 9 7 blood pressure Blood pressure is Time of day typically low in the morning and Example fluctuation within a day male 35 years old increases from afternoon to evening It is lower in the summer and higher in the winter 23 7 Some Useful Information about Blood Pressure Classification of Blood Pressure by the World Health Organization The World Health Organization WHO and the Systolic blood pressure International Society of Hypertension ISH developed Cape the Blood Pressure Classification shown in this figure Severe hypertension This classification is based on the blood pressure S Keone values measured on people in a sitting position in T TeWUR outpatient departments of hospitals in T9 Normal blood ee no universally accepted definition of hypotension However 120 optimal blood those having the systolic pressure below 100 mmHg are assumed target value Blood as hypotensive 80 85 90 100 110 ery 24
12. en Vietnam Geproduceerd in Vietnam M sense Hergestellt in Vietnam CyenaHo Bo BbeTHame All for Healthcare IM HEM 7200 E V 05 11 2011 Prodotto in Vietnam pui aie 5329530 9D Thank you for purchasing the OMRON M3 Digital Automatic Blood Pressure Monitor The OMRON M3 is a compact fully automatic blood pressure monitor operating on the oscillometric principle It measures your blood pressure and pulse rate simply and quickly For comfortable controlled inflation without the need of pressure pre setting or re inflation the devices uses its advanced Intellisense technology Intended Use This product is designed to measure the blood pressure and pulse rate of people within the range of the designated arm cuff following the instructions in this instruction manual It is mainly designed for general household use Please read the Important Safety Information in this instruction manual before using the unit Important Safety Information 1 1 Overview esee 3 2 Prep ratlon ine inen 4 2 1 Installing Replacing the Batteries 4 2 2 Setting the Buzzer Date and Time 5 3 Using the Unit eene 6 3 1 Applying the Arm Cuff 6 3 2 How to Sit Correctly 7 3 3 Taking a Reading 8 3 4 Using the Memory Function 10
13. he electric outlet with wet hands E Battery Usage If battery fluid should get in your eyes immediately rinse with plenty of clean water Consult a doctor immediately A Caution Indicates a potentially hazardous situation which if not avoided may result in minor or moderate injury to the user or patient or damage to the equipment or other property General Usage Do not leave the unit unattended with infants or persons who cannot express their consent Do not use the unit for any purpose other than measuring blood pressure Do not disassemble the unit or arm cuff Use only the approved arm cuff for this unit Use of other arm cuffs may result in incorrect measurement results Make sure that the air tube is not wrapped around other parts of your body when taking measurements at night This could result in injury when the air pressure in the air tube is increased Do not leave the cuff wrapped on the arm if taking measurements during the night This could result in injury Do not inflate the arm cuff over 299 mmHg Do not use a mobile phone or other devices that emit electromagnetic fields near the unit This may result in incorrect operation of the unit Do not operate the unit in a moving vehicle car airplane To inflate the cuff manually refer to Section 3 3 If the cuff is over inflated it can cause internal bleeding Important Safety Information AC Adapter Usage Use only the original AC
14. on Read and follow the Correct Disposal of This Product in the Technical Data Section when disposing of the device and any used accessories or optional parts L Al Open the rear cover page to read following The alphabet in the rear cover page correspond to those in the body page Main unit Display BO I START 4 button Memory button Up Down buttons Buzzer and Date Time setting Q button Air Jack Battery compartment AC adapter jack for optional AC adapter Arm cuff DArm cuff Medium cuff arm circumference 22 32cm Coloured Marker Air tube DAir plug Display Systolic blood pressure Diastolic blood pressure Heartbeat symbol Flashes during measurement Memory symbol Deflation symbol Average value symbol Movement error symbol Date Time display D Blood pressure level indicator Pulse display W Buzzer symbol Qo Cuff wrapping guide Battery low symbol irregular heartbeat symbol 2 Preparation 2 1 Installing Replacingthe Batteries Re 9 o 1 Remove the battery cover o e 2 Insert four AA batteries as le o indicated in the battery compartment and then replace the battery cover Notes If the battery low symbol x appears on the display turn the unit off then replace all batteries at the same time The measurement values continue to be stored in memory even after the batteries are replaced
15. on one reading Notes If the memory is full the monitor will delete the oldest readings When viewing the reading taken without setting the date and time is displayed instead of the date and time Average Value Previous Reading MEMORY View the Average View the Previous View the Readings Value Reading Stored in Memory If the previous reading The Memory number was taken without appears for a second setting the date and before the pulse rate is time the average displayed The newest value is not calculated set is numbered 1 10 3 Using the Unit To delete all the values stored in memory When the memory symbol amp appears first press the Memory button Then while holding it down press the O I START button simultaneously for about 2 3 seconds MEMORY oA O START i Note You cannot partially delete the stored readings 11 4 Troubleshooting and Maintenance 4 1 The icons and Error Messages Error Display Remove the arm cuff Wait 2 3 minutes and then take another measurement Repeat the steps in section 3 3 If this error continues to appear contact your doctor Irregular or weak pulses are detected Movement during measurement Carefully read and repeat the steps in section 3 3 Cuff is not applied correctly Apply the arm cuff correctly Refer to section 3 1 You should replace them with new ones ahead of The batteries are low time Refer to sec
16. s that the following values can be used as a guide to high blood pressure for measurements taken at home Systolic Blood Pressure Above 135 mmHg Diastolic Blood Pressure Above 85 mmHg This criteria is for home blood pressure measurement For professional office blood pressure measurement criteria please refer to Chapter 7 Your blood pressure monitor includes an irregular heartbeat feature Irregular heartbeats can influence the results of the measurement The irregular heartbeat algorithm automatically determines if the measurement is usable or needs to be repeated If the measurement results are affected by irregular heartbeats but the result is valid the result is shown together with the irregular heartbeat symbol CD If the irregular heartbeats cause the measurement to be invalid no result is shown If the irregular heartbeat symbol C is shown after you have taken a measurement repeat the measurement If the irregular heartbeat symbol CD is shown frequently please make your doctor aware of it 3 Using the Unit 3 4 Using the Memory Function The monitor automatically stores the result up to 60 sets It can also calculate an average reading based on the measurements from the last three readings taken within 10 minutes If there are only two readings in memory for that period the average will be based on two readings If there is one reading in memory for that period the average will be based
17. st is the Diastolic Blood Pressure Both pressure readings the Systolic and Diastolic are necessary to enable a doctor to evaluate the status of a patient s blood pressure What is Irregular Heartbeat An irregular heartbeat is a heartbeat rhythm that varies by Normal Heartbeat more than 25 from the average heartbeat rhythm detected Pulse M A A A AA while the unit is measuring the systolic and diastolic blood pi pressure ANININ pressure Irregular Heartbeat If such an irregular rhythm is detected more than twice Short Long during measurement the irregular heartbeat symbol Pulse appears on the symbol when the measurement results are Blood pressure NIIS NUN displayed What is Arrhythmia Arrhythmia is a condition where the heartbeat rhythm is abnormal due to flaws in the bio electrical system that drives the heartbeat Typical symptoms are skipped heartbeats premature contraction an abnormally rapid tachycardia or slow bradycardia pulse 7 Some Useful Information about Blood Pressure Why is it a Good Thing to measure Blood Pressure at Home Many factors such as physical activity 200 Upper curve systolic blood pressure anxiety or the time of day can Lower curve diastolic blood pressure influence your blood pressure A single measurement may not be sufficient for SU an accurate diagnosis Thus it is best to d try and measure your blood pressure at KR RES mmHg the same tim
18. tion 2 1 You should replace them with new ones at once The batteries are exhausted haferto section 12 4 Troubleshooting and Maintenance Error Display Cuff is under inflated Carefully read and repeat the steps listed under section 3 3 Movement during measurement Repeat measurement Remain still and do not talk during measurement Refer to section 3 3 Air plug disconnected Insert the air plug securely Refer to section 3 1 Arm cuff not applied correctly Apply the arm cuff correctly Refer to section 3 1 Clothing is interfering with the arm cuff Remove any clothing interfering with the arm cuff Refer to section 3 1 Air is leaking from the arm cuff Replace cuff with new one Refer to Chapter 5 The arm cuff was inflated above 299 mmHg when inflating the cuff manually Do not inflate the arm cuff above 299 mmHg Refer to section 3 3 Device error Contact your OMRON retail outlet or distributor The irregular heartbeat symbol C may also be displayed with error messages 13 14 4 Troubleshooting and Maintenance 4 2 Troubleshooting Problem The reading is extremely low or high Arm cuff not applied correctly Apply the arm cuff correctly Refer to section 3 1 Movement or talking during measurement Remain still and do not talk during measurement Refer to section 3 3 Clothing is interfering
19. ve the elbow D Marker arrow_ under tube is centred on the middle of your inner arm Close the fabric fastener FIRMLY Notes When you take a measurement on the right arm air tube will be at the side of your elbow Be careful not to rest your arm on the air tube The blood pressure can differ between the right arm and the left arm and therefore also the measured blood pressure values can be different Omron recommends to always use the same arm for measurement If the values between the two arms differ substantially please check with your physician which arm to use for your measurement 3 Using the Unit 3 2 How to Sit Correctly To take a measurement you need to be relaxed and comfortably seated under comfortable room temperature No eating smoking or exercising 30 minutes before taking a measurement Sit on a chair with your feet flat on the floor Sit upright with your back straight c The cuff should be at the same level as your heart d Cuff wrapping guide EN The Cuff Wrapping Guide is a unique feature that indicates if the cuff is not wrapped tightly enough around the arm Even when the Q is displayed a blood pressure reading will be taken Note This reading is NOT reliable due to the incorrect wrapping of the cuff Please wrap the cuff again taking care to wrap it correctly and take the measurement again When the is displayed the cuff is correctly wrapped tightly enough on the arm an
20. with the arm cuff Remove any clothing interfering with the arm cuff Refer to section 3 2 Arm cuff pressure does not rise The air tube is not securely connected into the main unit Make sure that the air tube is connected securely Refer to section 3 1 Air is leaking from the arm cuff Replace the arm cuff with a new one Refer to Chapter 5 Arm cuff deflates too soon The arm cuff is loose Apply the cuff correctly so that it is firmly wrapped around the arm Refer to section 3 1 Cannot measure or readings are too low or too high The arm cuff has not been inflated sufficiently Inflate the cuff so that it is 30 to 40 mmHg above your previous measurement result Refer to section 3 3 4 Troubleshooting and Maintenance Problem Nothing happens when you press the buttons The batteries are empty Replace the batteries with new ones Refer to section 2 1 The batteries have been inserted incorrectly Insert the batteries with the correct polarity Refer to section 2 1 Other problems Press the O I START button and repeat measurement If the problem continues try replacing the batteries with new ones If this still does not solve the problem contact your OMRON retail outlet or distributor 15 16 4 Troubleshooting and Maintenance 4 3 Maintenance To protect your unit from damage please observe the following