Omron Healthcare INTELLISENSE HEM-705CP User's Manual

image

Contents

1. 3 Replace the battery cover BATTERY REPLACEMENT XI Low Battery Indicator When the Low Battery Indicator appears on the display screen remove all the batteries Replace with four new batteries at the same time Long life alkaline batteries are recommended NOTE Measurement values stored in the memory will not be deleted during battery replacement After replacing the batteries reset the date and time Refer to page 9 Setting the Date and Time CONNECTING THE PRINTER 1 Remove the printer cord from around the printer unit 2 Connect the printer to the blood pressure monitor Insert the cable plug with the g indicator facing upward CAUTION Inserting the cable plug in any other position will damage the monitor LOADING THE PRINTER PAPER This printer uses OMRON thermal paper part number 90TRP Use only this thermal paper or thermal paper of similar size width 38mm roll diameter not to exceed 26mm NOTE For information on where to purchase printer paper visit our website at www omronhealthcare com or call customer service at 1 800 634 4350 1 Open the printer cover in the direction as indicated by the arrow symbols in the illustration 2 Peel back the glued edge of the printer paper Cut off the first 4 inches of the paper with a pair of scissors 3 Insert the edge of the printer paper in paper feeding slot Press the FEED button until the paper extends approximately 3 inches
2. 4 Insert the paper through the printer cover Make sure the paper extends out from the top 5 Close the printer cover SETTING THE DATE AND TIME The Blood Pressure Monitor automatically stores up to 28 measurement values with the date and time Set the monitor to the current date and time before taking a measurement for the first time If the current date and time is not set a default date and time will display m a gS Ou D oo 1 SETTING THE YEAR When the batteries are installed the 12 00 AM will display Lt a M Ta Cy 2 33 xN Press and hold the SET button for 2 seconds to SET ADJUST adjust the year O O DATE TIME SET The year flashes on the display E l j F x a ani uu i 2 SETTING THE YEAR The year can be set between 2001 and 2030 When the display reaches 2030 it will return to 2001 SET ADJUST Press the ADJUST button to advance by increments O O of one year DATE TIME SET Press the SET button to set the current year uu The month flashes on the display 7 x NOTE Press and hold the ADJUST button to advance the value quicker 10 SETTING THE DATE AND TIME 3 SETTING THE MONTH Press the ADJUST button to advance by increments of one month Press the SET button to set the current month The day flashes on the display H y play 3 1 C i 4 SETTING THE DAY Press the ADJUST button to advance by increments of one da
3. Ran out of paper Install new roll of paper 22 QUICK REFERENCE GUIDE To ensure a reliable reading avoid eating smoking and exercising for 30 minutes before taking a measurement You should try to measure your blood pressure at the same time each day Avoid taking measurements during stressful times 1 Remove any tight fitting clothing from your upper arm Sit in a chair with your feet flat on the floor and place your arm on a table so the cuff is at the same level as your heart Put your arm through the cuff loop The bottom of the cuff should be approximately 1 2 above the elbow The Green Marker on the cuff should lie over the brachial artery on the inside of the arm Tube should run down center of arm even with your middle finger Bottom Edge 1 2 Inch Cuff Tube Green Marker Wrap the cuff firmly in place around your arm using the closure strip 23 QUICK REFERENCE GUIDE 5 Press the ON OFF button Display symbols appear 6 Press the START button The cuff starts to inflate automatically Remain still and do not talk during the measurement NOTE To stop the inflation or measurement push the START STOP button The monitor will stop inflating start deflating and will turn off 7 When the measurement is complete the arm cuff completely deflates Your blood pressure and pulse rate are displayed NOTE Wait 2 3 minutes between measurements The wait time allows the arteries
4. minute Accuracy Calibration Pressure 3 mmHg Pulse 5 of reading Inflation Automatic by electric pump Deflation Active electronic control valve Rapid Pressure Release Active electronic control valve Pressure Detection Capacitive pressure sensor Measurement Method Oscillometric method Pulse Wave Detection Capacitive pressure sensor Power Source 4 AA batteries or AC adapter Battery Life Approximately 300 uses Operating Temperatures Humidity 50 F to 104 F 10 C to 40 C 30 to 85 RH maximum Storage Temperatures Humidity 4 F to 140 F 20 C to 60 C 10 to 95 RH maximum Main Unit Weight Approximately 13 4 oz 380g not including batteries Main Unit Dimensions Approximately 4 4 1 x 7 w x 2 4 b 115 mm x 177 mm x 71 mm Cuff Dimensions Approximately 5 1 2 x 19 140 mm x 480 mm Cuff Circumference Fits arm circumference 9 to 13 Contents Arm cuff printer paper roll instruction manual Optional Accessories AC Adapter model number HEM ADPT1 Sold Seperately Small Adult Cuff Arm circumference size 7 9 Large Adult Cuff Arm circumference size 13 17 UPC Code 0 73796 70536 7 NOTE These specifications are subject to change without notice PRINTER UNIT Power Source 4 AA batteries not included Battery Life Approxim
5. the same time each day for consistency 8 Wait 2 3 minutes between measurements The wait time allows the arteries to return to the condition prior to taking the blood pressure measurement You may need to increase the wait time depending on your individual physiological characteristics KNOW YOUR UNIT Main Unit AC Adapter Jack Adjust Date Time Heart Symbol Systolic Blood Pressure Set Datele OMROI AUTOMATIC BNQOD PRESSURE MONITOR MODEL HEM 705 SYSTOLIC 1 mmHg is mA nse SET ADJUST Printer 3 Printer Jack DATE TIME SET Cuff Jack ON OFF Button DIASTOLIC Diastolic Blood Pressure START Button Pulse Rate Memory Button Date Time Display Components Printer Printer Cover Printer Cable Plug Power Operation Light Paper Feed GRAPH ALLDATA DATA STOP OO omron 1 Print Current Data Stop Printing Print Graph Print All Data Arm Cuff Air Tube p j RN Air Plug z Connects to Air Jack BATTERY INSTALLATION REPLACEMENT Main Unit Printer Unit 1 Press the w OPEN indicator on the battery cover and slide the cover off in the direction of the arrow 2 Install 4 AA size batteries so the positive and negative polarities match the polarities of the battery compartment as indicated
6. A MEASUREMENT 4 When the measurement is complete the arm cuff completely deflates Your blood pressure and pulse rate are displayed 5 Press the ON OFF button to turn the monitor off NOTE The monitor will automatically turn off after wor five minutes Bs INSTRUCTIONS FOR SPECIAL CONDITIONS If your systolic pressure is known to be more than 220mmHg press and hold the START button until the monitor inflates 30 to 40 mmHg higher than your suspected systolic pressure NOTE e Do not apply more pressure than necessary e The monitor will not inflate above 299 mmHg USING THE MEMORY FUNCTION Every time you complete a measurement the monitor stores the blood pressure and pulse rate in the memory The Blood Pressure Monitor automatically stores up to 28 measurement values When 28 sets of measurement values are stored in the memory the oldest record is deleted to save the most recent measurement values TO DISPLAY THE MEASUREMENT VALUES The date and time is alternately displayed with the measurement values 1 Press the m button down while the Heart Symbol 4 is displayed The most recent measurement values will be displayed q gt f 9308 00 RBI NOTES e Press the button repeatedly to display the next values e Press and hold the button to display the values faster 2 Press the START STOP button to turn the power off TO DELETE ALL VALUES STORED IN THE MEMORY You cannot partially delete reco
7. Diastolic Reading Column Column Pulse Reading Month s mmHg ian se Column D T SYS DIA PUL Date Qos 13 75 67 Oldest reading stored 23 11 17 69 64 Time Hour _ 145 13 1i 63 3 14 10 67 66 SF 12 66 64 15 7 10 61 64 20 12 06 66 65 3 12 10 65 64 10 12 08 61 67 12 14 11 56 62 14 7 07 64 64 2115 11 64 64 Most recent reading 2212 21 62 66 nee 4 6 12 64 63 AV 12 65 65 Average of all stored readings Month s The month or months will print For example If the blood pressure readings were stored during the months of March through June then the printout will indicate 3 6 Date The day s date will print For example If the measurement was stored on June Ist then the printout will indicate 1 under the column D Time Hour The hour in military time will print For example If the measurement was stored at 3 00pm then the printout will indicate 15 under the column T 18 USING THE PRINTER 3 TO PRINT A GRAPH OF ALL MEASUREMENT VALUES Press the GRAPH button The graph will list all of the readings stored in the memory The stored data will print as illustrated in the following example Month s mmHg 50 100 150 200 Oldest reading stored If reading is lower than 50 mmHg or higher than 200 mmHg 4 will appear F TW Y l IN Diastolic Reading m Systolic Reading _ Most recent reading Average diastolic reading of Avera
8. OMRON INSTRUCTION MANUAL shie Automatic Blood Pressure Monitor with Printer Model HEM 705CP ESPA OL ENGLISH TABLE OF CONTENTS Before Using the Monitor BIVENS reese deaeenae sean basse eee ons 3 Important Safety Notes cs lt ctasscnansnanscnstqnsnadvaceansirgacisaadexsnncavtacnsndversenaees 4 Before Taking a Measurement j ccciccscasievascasesosasssenseaseasdscestadennnsonatvancs 5 Operating Instructions Kow Your UMi coisas actnetsae cae senn aeei nee eaea e a aedd reana 6 Battery Installation 4 uicescdvin vesnanetrannenasananadvasennceansieaanexaencosiaaeanavetswiades 7 Co necting the Printer Seay cenccvedscaencsaneacahetetnapsetaacnuaddeactouasteeiddeantncent 8 Loading the Papet Sa sct cscs ros eeresanne ee nsaeetnqensauires aces Raer a t E EASRA R ASEA SEE i iE 9 Setting the Date and Time s ssnnsssesseesesseosseesreseesseesresenssesseserssesse 10 Applying the Arm Cuff 5 lt cccasiccssennas ccsiaccossvaretcadsdnasseaiacqansestacnvdeauntece 12 Taking a Measurement 5 os c505 cssiendeantyieanteeannnevensyaeteaciuentsaasaainenienseanayise 14 Using the Memory Function 5 3 cssacevsacniinesdsbacscqasecaudeasasinicneeas 16 Using the Pinter sssseicenis nenna aeneae ians iei tia 17 Care and Maintenance Care and NIA STRAMI CEs cascast cccsicascaasynasbaentaasssniy sadeadteesteamspaaseacsbnddetady ea 20 Error Indicators and Troubleshooting Tips cccceeeeeseeesceeeteeees 21 Quick Reference Guide ss onnnssns
9. ately 300 uses current data Paper Usage Approximately 200 uses for 1 roll current data Printer Weight Approximately 6 oz 170 g not including batteries Printer Dimensions Approximately 4 1 x 2 w x 2 b 123 mm x 72 mm x 50 mm NOTE These specifications are subject to change without notice 27 Made in China Manufactured by OMRON Distributed by OMRON HEALTHCARE INC 1200 Lakeside Drive Bannockburn Illinois 60015 1243 www omronhealthcare com 2005 OMRON HEALTHCARE INC 705CPINST Rev B
10. eries Refer to X page 7 for battery installation TROUBLESHOOTING TIPS PROBLEM CAUSES AND SOLUTIONS Replace all worn batteries No power Check the battery installation for proper pee appes Ou tue uma placement of the battery polarities Blood pressure varies constantly Many factors including stress time of day Measurement values appear too how you wrap the cuff may affect your high or too low blood pressure Review the sections Before Taking a Measurement and Taking a Measurement 21 TROUBLESHOOTING TIPS PRINTER UNIT PROBLEM CAUSES SOLUTIONS No power Worn batteries Replace all 4 AA batteries Reference page 7 Incorrect battery Check the battery installation or installation proper placement of the battery polarities Printer not connected Connect printer to BP Monitor p Insert cable plug in correct No data will print Cable plug not inserted a correctly or damaged position Reference page 8 Contact Customer Service for repair information Make sure the date and time is No measurement values set up in the monitor Take a are stored in the memory reading then print Reference page 17 Not all data stored in Worn batteries Replace all 4 AA batteries the memory will print Reference page 7 POWER ACTIVE Light is blinking Printer paper will Printer paper jammed Remove jammed paper and install not feed roll again Reference page 9
11. ge systolic reading of stored measurements stored measurements l 7 CARE AND MAINTENANCE To keep your digital blood pressure monitor in the best condition and protect the unit from damage follow the directions listed below e Keep the monitor in the storage case when not in use e Do not forcefully bend the arm cuff or air tube Do not fold tightly e Clean the monitor with a soft dry cloth Do not use any abrasive or volatile cleaners Do not attempt to clean the cuff Never immerse the monitor or any components in water e Store the monitor in a safe and dry location Do not subject the monitor to extreme hot or cold temperatures humidity and direct sunlight e Avoid subjecting the monitor to strong shocks such as dropping the unit on the floor e Remove the batteries if the unit will not be used for three months or longer Always replace all the batteries with new ones at the same time e Use the unit consistent with the instruction provided in this manual Use only authorized parts and accessories 20 ERROR INDICATORS SYMBOL CAUSE CORRECTION Remove the arm cuff Refer to m Monitor did not Applying the arm cuff on page 12 me detect pulse rate Wait 2 3 minutes Take another measurement j j Cuff under inflated Remove the arm cuff Read a Taking a Measurement on page 14 Wait 2 3 minutes r Take another measurement g Cuff over inflated Z b IOE EAE Replace the four batt
12. ice by calling 1 800 634 4350 for the address of the repair location and the return shipping and handling fee Information for warranty service is available on our website at www omronhealthcare com Enclose the Proof of Purchase Include a letter with your name address phone number and description of the specific problem Pack the product carefully to prevent damage in transit Because of possible loss in transit we recommend insuring the product with return receipt requested ALL IMPLIED WARRANTIES INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR PARTICULAR PURPOSE ARE LIMITED TO THE DURATION OF THE APPLICABLE WRITTEN WARRANTY ABOVE Some states do not allow limitations on how long an implied warranty lasts so the above limitation may not apply to you OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER INCIDENTAL CONSEQUENTIAL OR INDIRECT COSTS EXPENSES OR DAMAGES Some states do not allow the exclusion or limitation of incidental or consequential damages so the above exclusions may not apply to you This warranty gives you specific legal rights and you may also have other rights which may vary from state to state FOR CUSTOMER SERVICE Visit our web site at www omronhealthcare com Call toll free 1 800 634 4350 26 SPECIFICATIONS MAIN UNIT Model HEM 705CP Display LCD Digital Display Measurement Range Pressure 0 to 299 mmHg Pulse 40 to 180
13. ise emissions from digital apparatus as set out in the interference causing equipment standard entitled Digital Apparatus ICES 003 of the Canadian Department of Communications Cet appareil num rique respecte les limites de bruits radioel ctriques applicables aux appareils num riques de Clase B prescrites dans la norme sur le materiel brouilleur Appareils Num riques ICES 003 dict e par le minister des communications Changes or modifications not expressly approved by the party responsible for compliance could void the user s authority to operate the equipment 25 WARRANTY Your HEM 705CP IntelliSense Automatic Blood Pressure Monitor excluding the monitor cuff is warranted to be free from defects in materials and workmanship appearing within 5 years from the date of purchase when used in accordance with the instructions provided with the monitor The monitor cuff is warranted to be free from defects in materials and workmanship appearing within one year from the date of purchase when the monitor is used in accordance with the instructions provided with the monitor The above warranties extend only to the original retail purchaser We will at our option repair or replace without charge any monitor or monitor cuff covered by the above warranties Repair or replacement is our only responsibility and your only remedy under the above warranties To obtain warranty service contact Omron Healthcare s Customer Serv
14. rds stored in the memory All records will be deleted 1 Press and hold the memory button and START button simultaneously for more than 2 seconds RP gt Ez USING THE PRINTER The HEM 705CP printer unit has the capability to print all the of measurement values stored in the memory of the Blood Pressure Monitor NOTES e The correct date and time must be set up on the monitor before taking a reading If the date and time is not set the numerical values will not appear on the print out e The printer unit must be connected to the monitor The monitor does not have to be ON to operate the printer unit e The printer unit will automatically shut off when the printing is completed e To stop printing press the DATA STOP button at any time 1 TO PRINT THE MOST RECENT MEASUREMENT VALUES Press the DATA STOP button The numeric results of the most recent reading will print Month Date 7 9 6 36 AM Time Systolic Reading ISys Diastolic Reading 111mmH g Ise Readi DIA 67mmHg Pulse Rea 1ng PULSE 77 min USING THE PRINTER 2 TO PRINT ALL THE MEASUREMENT VALUES Press the ALL DATA button The numeric printout will list all of the readings stored in the memory starting with the oldest reading to the most recent reading followed by the average of all the stored readings The stored data will print as illustrated in the following example Systolic Reading
15. ressure Follow the instructions of your healthcare provider M Operate the unit only as intended Do not use for any other purpose MI The unit is intended for use in measuring blood pressure and pulse rate in adult patient population Do not use on infants MI Do not use a cellular phone near the unit It may result in operational failure MI Do not plug or unplug the adapter power cord with wet hands MI Changes or modifications not approved by Omron Healthcare will void the user warranty Do not disassemble or attempt to repair the unit or components BEFORE TAKING A MEASUREMENT To ensure a reliable reading follow these recommendations 1 Avoid eating smoking and exercising for 30 minutes before taking a measurement Rest for at least 15 minutes before taking the measurement 2 Stress raises blood pressure Avoid taking measurements during stressful times 3 Measurements should be taken in a quiet place 4 Remove tight fitting clothing from your left arm 5 Sit in a chair with your feet flat on the floor Rest your left arm on a table so that the cuff is at the same level as your heart 6 Remain still and do not talk during the measurement 7 Keep a record of your blood pressure and pulse readings for your physician A single measurement does not provide an accurate indication of your true blood pressure You need to take and record several readings over a period of time Try to measure your blood pressure at
16. s snugly around your arm The cuff should make good contact with your skin NOTE You should be able to fit your index finger between the cuff and your arm easily so you can pull the cuff off and on Relax your arm and turn your palm upward Be sure there are no kinks in the air tubing TAKING A MEASUREMENT The monitor automatically determines your ideal inflation level In rare circumstances when a higher inflation may be necessary the monitor automatically re inflates the cuff up to 30mmHg higher than the initial inflation and restarts the measurement The monitor detects the pulse during inflation Do not move your arm and remain still during the measurement 1 Press the ON OFF button All display symbols appear on the screen The cuff starts to inflate automatically NOTE To stop the inflation or measurement push the START STOP button The monitor will stop inflating xu g start deflating and will turn off 2 Press the START button As the cuff begins to inflate the monitor automatically determines your ideal inflation level Because this Y monitor detects the pulse even during inflation do not m move your arm and remain still until the entire Ll measurement completes o Nag 3 Inflation stops automatically and the measurement is started 20 gt 6 Inflation Deflation As the cuff deflates decreasing numbers appear on the display The Heart Symbol flashes at every heartbeat TAKING
17. senssnssnsesssiesseessersseressstessenssensseressns 23 PCC Statement sariini e Ea EAE E EE Erain 25 Warranty Information a is5iens eniiaa Seasanenstadonncdesnsnantesgivnadsansenarbaanieessinaae 26 Specifications seisen ae ena Eni AERE ERER E TE AREENA EE 27 INTRODUCTION Thank you for purchasing the Omron HEM 705CP Intellisense Automatic Blood Pressure Monitor with Printer Fill in for future reference DATE PURCHASED SERIAL NUMBER Staple your purchase receipt here Your new digital blood pressure monitor uses the oscillometric method of blood pressure measurement This means the monitor detects your blood s movement through your brachial artery and converts the movements into a digital reading An oscillometric monitor does not need a stethoscope so the monitor is simple to use The HEM 705CP comes with the following components e Monitor e Standard Cuff 9 13 arm circumference e Printer e Printer Tape e Instruction Manual printed in English and Spanish The HEM 705CP Intellisense Automatic Blood Pressure Monitor with Printer is intended for home use SAVE THESE INSTRUCTIONS IMPORTANT SAFETY NOTES To assure the correct use of the product basic safety measures should always be followed including the precautions listed below V Read all information in the instruction book and any other literature included in the box before using the unit MI Contact your physician for specific information about your blood p
18. to return to the condition prior to taking the blood pressure measurement You may need to increase the wait time depending on your individual physiological characteristics 24 FCC STATEMENT Note POTENTIAL FOR RADIO TELEVISION INTERFERENCE for U S A only This product has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC rules These limits are designed to provide reasonable protection against harmful interference in a residential installation The product generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If the product does cause harmful interference to radio or television reception which can be determined by turning the product on and off the user is encouraged to try to correct the interference by one or more of the following measures e Reorient or relocate the receiving antenna e Increase the separation between the product and the receiver e Connect the product into an outlet on a circuit different from that to which the receiver is connected e Consult the dealer or an experienced radio TV technician for help POTENTIAL FOR RADIO TELEVISION INTERFERENCE for Canada only This digital apparatus does not exceed the Class B limits for radio no
19. y Press the SET button to set the current day J The hour flashes on the display ac fo e T N 5 SETTING THE HOUR Press the ADJUST button to advance by increments of one hour Press the SET button to set the current hour The minutes flash on the display 1 ool 6 SETTING THE MINUTE a Press the ADJUST button to advance by increments of one minute Press the SET button to set the current minute 4 The unit will be ready to take a reading N e TE 3 7 Press the ON OFF button to turn the display off APPLYING THE ARM CUFF 1 Make sure the air plug is securely inserted in the main unit 2 Remove tight fitting clothing from your upper arm 3 Sit in a chair with your feet flat on the floor Place your arm on a table so the cuff is level with your heart APPLYING THE CUFF ON THE LEFT UPPER ARM 4 Put your arm through the cuff loop The Green Marker on Bottom Bdge 1 2 inch the cuff should lie over the brachial artery on the inside l p an of the arm The air tube runs oE ape RE ree down the inside of your arm The bottom of the cuff should be approximately 1 2 above your elbow Cuff Tube i Green Marker APPLYING THE ARM CUFF Pull the cuff so the top and bottom edges are tightened evenly around your arm When the cuff is positioned correctly press the sewn hook material FIRMLY against the pile side of the cuff Make certain the cuff fit

Download Pdf Manuals

image

Related Search

Related Contents

TDSHーBA 仲 一 保管用    Appareil photo numérique Mode d`emploi - Support  AlphaVUE User`s Manual - Birmingham Data Systems  Single Zone VAV Systems  Philips Food processor HR7770/00  Mad Catz Turtle Beach Ear Force XLS  Infinity TSS-SAT500 User's Manual  Senseo Senseo HD7873/51 coffee maker  Lien vers le diaporama  