Monarch Thermometer 1071-8041-210 User's Manual

image

Contents

1. Instruction Manual 3 MONARCH INSTRUMENT cE Monarch 314 Datalogging Humidity Temperature Meter 15 Columbia Drive Amherst NH 03031 2334 USA Phone 603 883 3390 Fax 603 886 3300 E mail support monarchinstrument com 1071 8041 210 Website www monarchinstrument com Monarch Instrument 2002 all rights reserved IAN Safeguards and Precautions IAN 1 Read and follow all instructions in this manual carefully and retain this manual for future reference 2 Donotuse this instrument in any manner inconsistent with these operating instructions or under any conditions that exceed the environmental specifications stated 3 Do not touch or manipulate the sensor 4 Do not expose the sensor to direct light This may cause false readings 5 Do not expose the sensor to static electricity 6 Making measurements of high or low temperature can be dangerous Keep the hand holding the temperature probe well away from the object being measured 7 This instrument is not user serviceable For technical assistance contact the sales organization from which you purchased the product or Monarch Instrument directly LIMITED WARRANTY SELLER warrants hardware products to be free from any defect in materials or workmanship for a period of one 1 year from date of shipment to BUYER SELLER s entire liability and BUYER s sole and exclusive remedy resulting from any defect in workmanship or material in the
2. F 1 to increase or decrease the number 888p 4 Press TIME to move to the next item The adjusting order is year month day hour minute m d hem 8888 8888 5 Press TIME again after the last setting to complete the process To abort during a setup process press the button Time Function Once setup pressing the TIME button displays time as follows top of the LCD year bottom left of the LCD month and day bottom right of the LCD hour and minute Press the TIME button or any other button to exit this mode This operation will not interrupt the recording and MAX MIN operation MODEL 314 DATALOGGING HUMIDITY TEMPERATURE METER 7 3 DataLogger Select DataLogger from the menu to load recorded data from the humidity meter A progress bar will show how many bytes should be loaded and how many bytes have been received When data is loaded successfully three new windows appear as below Data Sets Graph Tabular Data Sets Window Displays how many data sets were loaded and the detail information for each data set start date start time recording rate and data length Click on any data set to choose the set for graph and tabular window For other operating instructions please refer to the online help while executing TestLink 6 0 MODEL 314 DATALOGGING HUMIDITY TEMPERATURE METER Setup TestLink Humidity DataLogger RS232 Interface Software The TestLink package contains 1 Tw
3. that the display data is being held The Maximum value is now being displayed The Minimum value is now being displayed Indicates that the instrument is recording When flashing the memory is full Indicates year is displayed in the main window Thermocouple type indication Relative Humidity indication Indicates the value below is month and day Indicates which temperature channel is being displayed Indicates the value below is hour and minute hon MAX MIN REC Indicates the value below is minute and second B 7 5 0 Negative temperature indication r RH Centigrade or Fahrenheit indication Bx B B B ge Feature Locations 1 Dust mask 2 Sensor probe 3 T2 channel Type K thermocouple probe input 4 LCD display 5 ON OFF button 6 Time display button Record control button 8 MAXMIN function control button 9 HOLD button 10 C F control button 11 Type K offset calibration screw 12 Digital output connector RS232 3 AC power adapter connector Tripod connector Battery cabinet cover MODEL 314 DATALOGGING HUMIDITY TEMPERATURE METER 4 0 Operating Instructions 4 1 4 2 4 3 4 4 4 5 4 6 4 7 Preparation for Measurement Measurement of ambient humidity and temperature T1 may be made with the probe clipped to the instrument or hand held remotely For measurement of a second temperature plug a Type K thermocouple into the socket marked T2 taking care
4. HE SOFTWARE BUYER agrees that any specific right or remedy granted to BUYER hereunder with respect to any breach or default by SELLER shall be in lieu of all other rights and remedies otherwise available to BUYER at law or in equity as the result of such breach or default regardless of whether based on contract tort strict liability or other theory of liability IN NO EVENT SHALL SELLER BE LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL CONSEQUENTIAL OR PUNITIVE LOSSES OR DAMAGES INCLUDING BUT NOT LIMITED TO LOSSES OR DAMAGES FOR ANY LOST PROFITS OR LOST DATA AS THE RESULT OF ANY BREACH OR DEFAULT BY SELLER WITH RESPECT TO THE HARDWARE OR SOFTWARE EVEN IF SELLER HAS BEEN ADVISED OR MADE AWARE OF THE POSSIBILITY OF ANY SUCH LOSSES OR DAMAGES AND REGARDLESS OF WHETHER THE CLAIM IS BASED ON CONTRACT TORT STRICT LIABILITY OR OTHER THEORY OF LIABILITY This limited warranty does not extend or apply to consumables including but not limited to lamps and batteries if applicable or equipment instruments or accessories which are warranted separately by the original manufacturer of these items DECLARATION OF CONFORMITY MODEL 314 DATALOGGING HUMIDITY TEMPERATURE METER Table of Contents Section Page Monarch Instrument 3 Division of Monarch International Inc 15 Columbia Drive Amherst NH 03031 USA 1 0 Introduction ooo cc cece ccc sscssessssesesesessssessssasessasesessasesessesescscessseateusscesessavens 1 declares that the pr
5. d turn the power on Release these buttons and all the segments on the display will flash 2 Within 3 seconds of releasing these buttons press the C F and TIME buttons simultaneously to enter the calibration mode or the instrument will revert to the normal operating mode 3 Upon entering the calibration mode the humidity reading will flash and CAL1 will appear on the second display 4 Insert the humidity probe into the standard humidity cavity of 32 8 RH 25 C Allow the system to stabilize for 20 minutes and then press the MAX MIN button to create the calibration data If the instrument recognizes the value is out of tolerance it will sound 2 beeps and remain in the CAL1 mode If the instrument recognizes the value is within tolerance new calibration data is created and the instrument will advance into the CAL2 mode as indicated on the second display 5 Insert the humidity probe into the standard humidity cavity of 75 3 RH 25 C Again allow the system to stabilize for 20 minutes and then press the MAX MIN button to create the calibration data If the instrument recognizes the value is out of tolerance it will sound 2 beeps and remain in the CAL2 mode If the instrument recognizes the value is within tolerance new calibration data will be written into the memory and the calibration is complete 5 2 T1 Temperature Calibration 1 Turn the unit off Press and hold the MAX MIN HOLD and REC buttons While t
6. e that Auto Power Off is disabled and the will not be displayed 4 12 Low Battery Condition When the battery voltage is at or below the minimum for proper operation the symbol will show on the LCD indicating that the battery must be replaced 4 13 Digital Output The Digital Output is a 9600 bps N 81 serial interface RX is a 5V normally high input port TX is a 5V normally high output port Press HOLD interval Press REC 7 or C F 4 to increase or decrease the number Press HOLD to advance to the next item Cl ey eS Press HOLD again after the last setting to complete the process To abort during a setup process press the button LA re Do not touch or manipulate the sensor Do not expose the sensor to direct light This may cause false readings Do not expose the sensor to static electricity 1 2 3 MODEL 314 DATALOGGING HUMIDITY TEMPERATURE METER Appendix Thermocouple Probe Specifications Model Range Tolerances Description _ _ TP K01 50 C to 200 C 2 2 C or 0 75 Teflon insulation Maximum insulation Wire probe 58 F to 392 F 3 6 F or 0 75 temperature 260 C 500 F TP K01 i EAN Probe for general condition measurements especially lt 5 for complex and hard to reach places i C g Calibration Procedure 5 1 Humidity Calibration 1 Turn the unit off Press and hold the MAX MIN HOLD and REC buttons While these three buttons are presse
7. es 4 and therefore conforms in accordance with 89 336 EEC EMC Directive The testing of this product was performed by GesTek EMC Lab in January of Appendix Thermocouple Probe Specifications 5 2000 Ref No 99C049E 5 0 Calibration Procedure o oioioiooococccccccccccccccscscssssssssssssesessssesssesestesesceceseseeseassceusess 5 5 1 Humidity Calibration 5 2 T1 Temperature Calibration 0 ccccccccceceeceeeeeeeeeeeeeeeeeeaeeeeeaeeeseaaeeeseeeeeseeeeeeaae 5 5 3 Notes on Humidity and Temperature Calibration cccccceceeseeeeeeeeeeeneeeeeeee 6 6 0 Setup TestLink Humidity DataLogger RS232 Interface DOPE W ATO enpreni e T cig cetns Sh dean AE E 7 7 0 Running A k 5 11 rr een ES ST EESES Oi al 2 February 2000 Importer Amherst NH Alan Woolfson VP Engineering Authorized Signature Woolfson VP Engineering Authorized Signature TA Open Testhinkiencncntiitiedianitain ania Se aie ea Lee ae 7 2 Real Time Tabular and Real Time Graph V3 DataLogger idani danna ieee ie ee evened a et doe eee i ae MODEL 314 DATALOGGING HUMIDITY TEMPERATURE METER 1 0 Introduction This instrument is a dual channel recording digital Humidity Temperature meter providing simultaneous display of ambient humidity and temperature on the first channel and a wide range temperature only measurement on the second channel The first channel incorporates a polymer capacitive humidity sensor and a se
8. hardware product covered by this limited warranty shall be limited to and fully discharged by the SELLER s option of replacement or repair of such item without charge The limited warranty provided in this clause is in lieu of all other warranties expressed or implied arising by law or otherwise ALL IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE EXCLUDED This limited warranty shall not be modified except by an arrangement signed by both parties specifically referencing this clause SELLER warrants that any software supplied will operate in accordance with the documentation or manual supplied therewith in all material respects when used in strict compliance with such documentation or manual Notwithstanding the foregoing BUYER acknowledges that since software is complex and therefore may have defects BUYER s sole and exclusive remedy for any such defects or breach of this warranty shall be to require SELLER within a reasonable period of time to provide all reasonable programming services to correct programming errors in the software Except as provided above SELLER MAKES AND BUYER RECEIVES FROM SELLER NO EXPRESS OR IMPLIED WARRANTIES OF ANY KIND WITH RESPECT TO ALL OR ANY PORTION OF SOFTWARE AND BUYER HEREBY AGREES AND ACKNOWLEDGES THAT IT ACCEPTS THE SOFTWARE IN AS IS CONDITION SELLER HEREBY EXPRESSLY EXCLUDES ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE WITH RESPECT TO T
9. hese three buttons are pressed turn the power on Release these buttons and all the segments on the display will flash 2 Within 3 seconds of releasing these buttons press the C F and HOLD buttons simultaneously to enter the calibration mode or the instrument will revert to normal operating mode 3 Upon entering the calibration mode the temperature reading will flash and CAL1 will appear on the second display 4 Insert the probe into a temperature controlled chamber at 0 C 32 F and allow the system to stabilize for 20 minutes Press the MAX MIN button to create the calibration data If the instrument recognizes the value is out of tolerance it will sound 2 beeps and remain in the CAL1 mode If the instrument recognizes the value is within tolerance new calibration data is created and the instrument will advance into the CAL2 mode as indicated on the second display 5 Insert the probe into a temperature controlled chamber at 40 C 104 F and again allow the system to stabilize for 20 minutes Press the MAX MIN button to create the calibration data If the instrument recognizes the value is out of tolerance it will sound 2 beeps and remain in the CAL2 mode If the instrument recognizes the value is within tolerance new calibration data will be written into the memory and the calibration is complete MODEL 314 DATALOGGING HUMIDITY TEMPERATURE METER 5 3 Notes on Humidity and Temperature Calibrations 1 W
10. hile performing the humidity calibration the environment should be maintained at stable conditions i e constant temperature and constant humidity to optimize the accuracy 2 During either calibration pressing the button at any time will return the instrument to the normal operating mode and no calibration data will be changed 3 To restore factory default calibration values press the HOLD and REC buttons simultaneously at any time while in the calibration mode 4 Whenever the probe is inserted into a standard humidity cavity or temperature controlled chamber always allow at least 20 minutes to ensure ambient conditions have stabilized 5 Always begin the humidity calibration process with the 32 8 RH 25 C standard cavity and the temperature calibration process with the 0 C 32 F temperature controlled chamber For maximum efficiency insert the probe into the conditioned environment 20 minutes before applying power to the instrument and starting the calibration process 6 During the calibration all displayed readings are based on pre process calibration data 7 The Auto Power Off function is disabled during calibration but is automatically restored upon completion 8 During the calibration the temperature is fixed on the F scale and it is not selectable
11. miconductor temperature sensor T1 in a detachable probe The second channel employs a Type K thermocouple that complies with the NIST and IEC584 temperature voltage tables for Type K thermocouples The instrument s internal memory can store up to 16312 records in any number of recording intervals It is equipped with a RS232 interface for bi directional communication with a PC 2 0 Specifications Measurement Range Humidity 0 100 RH Temperature 11 20 C 60 C 4 F 140 F T2 200 C 1370 C 328 F 2498 F Accuracy Humidity 2 5 RH at 25 C Temperature 11 0 7 C 1 4 F T2 See table below at 23 5 C Range Accuracy Temperature Coefficient 200 C 200 C 0 3 reading 1 C For ambient temperatures from 0 C 18 C and 200 C 400 C 0 5 reading 1 C 28 C 50 C for each C ambient below 18 C or 400 C 1370 C 0 3 reading 1 C above 28 C add the following tolerance into the 328 F 200 F 0 5 reading 2 F accuracy spec 200 F 2498 F 0 3 reading 2 F 0 01 of reading 0 03 C 0 01 of reading 0 06 F A Note The basic accuracy specification does not include the error of the probe Please refer to the probe accuracy specification for additional details Resolution Humidity 0 1 RH Temperature T1 0 1 C 0 1 F T2 200 C 200 C 0 1 C 200 C 1370 C 1 C 200 F 200 F 0 1 F beyond this range 1 F Sensor Inpu
12. o 3 5 diskettes 2 Custom designed RS232 cable for TestLink System Requirements Windows 95 Windows 98 or Windows NT 4 0 Minimum Hardware Required e 486 100 MHz PC 16 MB RAM e Atleast 5 MB hard disk space available to install TestLink program e Recommended display resolution is 800x600 Install TestLink 1 Close all other applications before installing TestLink software 2 Insert setup diskette 1 in floppy disk drive 3 Choose the Start button on the Taskbar and select Run 4 Type a setup and choose OK to copy TestLink exe executable file and Help file to your hard disk default is c program files testlink Running TestLink 7 1 Open TestLink Select TestLink from Start of Windows Your display should appear as below Real Time Data Sampling Rate Menu and Tool Bar p Real Time Graph O gt Bi direction Control Panel eee Real Time Tabular 7 2 Real Time Tabular Data and Real Time Graph Select Run from the menu or press from the tool bar to begin real time data collection from the humidity meter To change the data interval edit the sampling rate from the box on the right hand side of tool bar see figure above MODEL 314 DATALOGGING HUMIDITY TEMPERATURE METER 4 8 Recording Interval Setup To set the recording interval GSEt 1 Press and hold the MAX MIN button and then power on the meter GEL int 9888 4 8 Recording Data Each momentary press of the REC button
13. oduct 2 0 SDOCITIGATIONS 23 s0 50 2c02 2268 k acts ih A MAA oA ens 1 3 0 Symbol Definitions and Feature Locations 0 0 0 0 0 00 2 Humidity Temperature Meter Monarch 314 4 0 Operating Instructions 3 4 1 Preparation for Measurement ccccccceecceeeeeeeeeeeeeeeeeaeeeeeeaeeeseeaeeeseeeeeeseneeeees 3 AD POWEN Aaa cle cote ai ON E a i A A e E N 3 I amp i i 4 3 Humidity and Temperature T1 Measurement 00 cece cece eee eeneeteeeenee 3 to which this declaration relates is in conformity with the following standards Aca Selecting the Taniperature E ee ee es ee 3 4 5 Data Hold Operation iieii annae a i tect aatan 3 ASG o oos AST At EEE E E A EE NE susan 3 EMC EN55011 1991 4T Time FUNCOM ea Ea e aaa a E e E tack din ECEE E aE OSES 3 4 8 Recording Interval Setup ccccccccccscsecsecsecscsscecsecsscsecsececsecsecsecsessecseeaecaesesneenees 4 EN50081 1 1992 4 9 Recording D ta cnc 2eicih aa cieedk Rae A A sab ei pae aodai NAA 4 EN50082 1 1992 IEC 801 4 10 MAXIMIN Operation c cccsscsssssssesssssseesssssessssssesssssseessstsessssssesessstessssseeseesssesssaseeseen 4 4 11 Auto Power Off 0 0ccccccccceccccccsscesssessesscsecsecesscateessesecesseaeeesecaeesuscausesecaeseaecausensenseease 4 4 12 Low Battery Condition ccccccccccccsscsececsecsccsesecsecsecsscaecacsacsecsecsecsecsessecsesesseenees 4 4 13 Digital Output ii i048 hice ete a a Lanta inna
14. to observe the polarity of the thermocouple pins Power Press the button to turn the thermometer ON or OFF When first powered on the display will show how much memory space is available for use Example Display at right indicates 16 000 records of available 6888 memory i Humidity and Temperature T1 Measurement Place the humidity temperature probe in the environment to be measured Be sure to allow sufficient time for the sensors to respond to ambient conditions This time will be dependent on the amount of air circulation through the probe mask See Section 2 0 Response Time Selecting the Temperature Scale When first turned on the instrument defaults to reading in Celsius C To display readings in Fahrenheit F press the C F button To return to readings in Celsius C again press the C F button The instrument remembers the scale setting when last turned off and powers on in that setting the next time Data Hold Function The present reading is held on the display by pressing the HOLD button Pressing the HOLD button again releases the hold function and returns the instrument to continuous reading In the HOLD mode the MAX MIN C F and AREL buttons are disabled as indicated by two consecutive beeps when pressed Clock Setup To set the real time clock 5 E t 1 Press and hold the MAX MIN button and then power on the meter I 2 Press TIME clock _ 3 Press REC t or C
15. ts Channel 1 Detachable probe with humidity and temperature T1 sensors Channel 2 Type K Thermocouple T2 Input Protection at Thermocouple Input T2 60V DC or 24Vrms AC Display Triple 4 Digit Display plus Indicators see Section 3 0 Response Time Humidity 75 seconds in slowly moving air Temperature 40 seconds in slowly moving air T1 Digital Output Bi directional RS232 Software and Cable included Power requirement 9 Volt Battery NEDA 1604 or JIS 006P or IEC6F22 Battery Life Approx 100 hours with alkaline battery Low battery indication Operating Conditions e Operating Temperature and Humidity 0 C 50 C 32 F 122 F 0 90 RH Non condensing e Storage Temperature and Humidity 10 C 60 C 14 F 140 F 0 80 RH Non condensing e Altitude up to 2000 meters 6500 feet Dimensions Meter 186x64x30 mm 7 3x2 5 1 2 inches Probe 190 mm longx15 mm diameter 7 50 6 in Weight Approx 320g 10 8 oz Accessories Type K Wire Thermocouple Probe Battery Carrying Case Humidity Probe Holder Instruction Manual Software RS232 cable Option AC Adapter 9VDC 15 100mA Plug Diameter 3 5mm 1 35mm Notes MODEL 314 DATALOGGING HUMIDITY TEMPERATURE METER 3 0 Symbol Definitions and Feature Locations Ea H MAX MIN REC RH T1 T2 h m ms C F of Low battery indication Battery voltage is not sufficient for proper operation Indicates Auto Power Off is enabled Indicates
16. will alternately start and stop recording To clear the memory power off the meter press and hold the REC button followed by the button holding both on simultaneously for at least 2 seconds Then release both buttons The display will show CLR as shown to the right indicating that the memory has been cleared 4 10 MAX MIN Operation Press the MAX MIN button to enter the MAX MIN mode In this mode both the maximum and minimum values are simultaneously retained in memory and updated with every new data sample The instrument first enters the MAX mode and the MAX symbol and value are both displayed Pressing MAX MIN again advances the display to the MIN symbol and value The next press of the MAX MIN button will cause both the MAX and MIN symbols to flash This indicates that the maximum and minimum values have been updated in memory and the displayed reading is the present value Each successive press of the MAX MIN button circulates the display mode among these options To exit the MAX MIN mode press and hold the MAX MIN button for two seconds In the MAX MIN mode the C F button is disabled as indicated by two consecutive beeps if pressed 4 11 Auto Power Off By default the instrument powers on in the Auto Power Off mode and will automatically shut off 30 minutes after the last key operation or RS232 communication To disable this feature press and hold the HOLD button and then power on the meter Two successive beeps will indicat