Milwaukee SM806 User's Manual
Contents
1. Milwaukee USER MANUAL PORTABLE pH EC TDS Temperature Meter MODEL SM806 K RANGE HOLD SET CAL Combined Meter Range sh 000 to 14 00 pl 20 Sem Hemp 00 w 800 C 320 io 1400 T Milwaukee GENERAL OPERATION The meter is supplied com plete with a 9V battery Remove the battery com partment cover on the back of the meter and in stall the battery while pay ing attention to its polarity Connect the MA851D 1 probe to the meter Turn the instrument on by pressing the ON OFF key At start up the LCD _ shows the percentage of the remaining battery life for a few seconds and 56 then the current measure ee fi ment To select the temperature measurement unit C or F press and hold the ON OFF key until TEMP and the current temperature unit are displayed on the secondary LCD Use the SET button to select the unit and then press the ON OFF key a couple of times to return to normal mode To activate the HOLD function keep pressed the HOLD key The mea sured value will be fro zen on the display and the HOLD message ap pears on the secondary LCD To set the measurement range pH uS ppm press the RANGE key The chosen mode will be stored until next change Before taking any measurement make sure that the meter has been calibrated the CA
2. the desired buffer set with the SET key 7 01 pH BUFF for standard solutions pH 4 01 7 01 10 01 or 6 86 pH BUFF for NIST solutions pH 4 01 6 86 9 18 Press the ON OFF key again to exit e Remove the protective cap then immerse the probe in the first buffer solution e Press and hold the ON OFF key until CAL is displayed on the lower LCD Release CH the button and the message me 5 7 01 pH USE or 6 86 pH 7 USE for NIST buffer set will 34 4 ae be displayed e For a single point pH calibration place the probe in any buffer from the selected set e g pH 4 01 or 7 01 6 86 or 10 01 9 18 The meter will automatically recognize the buffer value store the calibration point and return to normal mode e If using pH 7 01 or pH 6 86 after recog nition of the buffer press the ON OFF but ton to return to normal mode e For a two point pH calibration place the probe in pH 7 01 buffer or pH 6 86 if the NIST buffer set was selected ie T The meter will recognize the buffer value and then display pH 4 01 USE e Place the probe in the second buffer pH 4 01 or 10 01 or if using NIST pH 4 01 or 9 18 When the second buffer is rec ognized the LCD will display OK for 1 second and the meter will return to normal mode e The CAL tag is displayed on the LCD to indicate that the meter is calibrated SECOND BUFFER EC
3. IRONMENT 0 to 50 C 32 to 122 F 100 RH max BATTERY TYPE 1 x 9V 6F22 BATTERY LIFE approx 300 hours AUTO OFF after 8 minutes DIMENSIONS 200 x 85 x 50 mm WEIGHT 260 g with battery ISM806R1 11 03
4. L tag is displayed on the left lower corner of the LCD After measurements switch the meter off by pressing the ON OFF key The OFF mes sage appears on the LCD release the button Store the electrode with a few drops of stor age solution in the protective cap Note We reserve the right to modify design construction and appearance of our products with out advance notice TAKING pH MEASUREMENTS e Always remove the electrode protective cap before taking any measurement If the electrode has been left dry soak the tip bottom 2 5 cm in M10000 rinse solution for a few minutes to reactivate it e Select the pH range with the RANGE key e Immerse the tip 2 5 cm of the probe into the sample and stir gently e Read the pH value when the clock symbol stops blinking pH CALIBRATION PROCEDURE A Preparation Buffer solutions 1 pH 7 01 6 86 MA9007 MA9006 2 pH 4 01 MA9004 for acidic calibration pH lt 7 or pH 10 01 9 18 MA9010 MA9009 for alkaline range pH gt 7 Use two beakers for each buffer solution one beaker for rinsing the probe the other for calibration In this way the cross contamina tion between solutions is minimized B Procedure Enter the pH mode with the RANGE key Select the calibration buffer set press and hold the ON OFF key until the LCD shows TEMP Press again this key and the BUFF message will appear then select
5. TDS OPERATIONS e The user can select the EC TDS conversion factor CONV value as well as the B coef ficient BETA for automatic temperature com pensation e Enter the EC or TDS mode with the RANGE key then press and hold the ON OFF key until TEMP is displayed on the secondary LCD Press this key AG 5 again to show the current con TONY version factor CONV e Press the SET button to change the value then press again this key to show the cur rent temperature coefficient BETA e Press the SET button to change the value then press the ON OFF key return to normal operation e For taking EC or TDS measurements immerse the probe in the solution to be tested e Select either EC or TDS mode with the RANGE key e Stir gently and wait for the reading to stabi lize the clock symbol stops blinking EC TDS CALIBRATION PROCEDURE e Clean the probe with M10000 solution Enter the EC mode with the RANGE key e Immerse the probe in MA9030 12 88 mS cm calibration solution e Press and hold the ON OFF key until CAL is displayed on the LCD Release the bution _ ic and the message 12 88 mS is nak USE will appear e The meter will perform automatic calibration and then display OK for 1 second before returning to normal mode e The calibration is now complete and the CAL tag is displayed on the LCD to indicate that th
6. e meter is calibrated BATTERY REPLACEMENT The meter shows the remaining battery percent age every time it is switched on When the battery level is below 5 the battery symbol on the bottom left of the LCD lights up to indicate a low battery condition The meter is also provided with BEPS Battery Error Prevention System which automatically switches the meter off if the battery level is so low to cause erroneous readings It is recommended to replace ees immediately the battery Turn the meter off remove the battery compartment cover from the rear of the meter and re place the rundown 9V battery with a new one Install the battery while paying attention to its polarity and re attach the cover NOTES e f reading is over range the display will flash the closest full scale value e To clear a previous calibration press ON OFF after entering the calibration mode The lower LCD will display ESC for 1 second and the meter will return to normal measurement mode The CAL symbol on the LCD will disappear and the meter is reset to the default calibration WARRANTY This instrument is warranted against defects in materi als and manufacturing for a period of two years from the date of purchase Probe is warranted for 6 months The probes are warranted for a period of six months If during this period the repair or replacement of parts is required where the damage is not due to negl
7. igence or erroneous operation by the user please return the parts to either distributor or our office and the repair will be effected free of charge Damages due to accidents misuse tampering or lack of prescribed maintenance are not covered OPTIONAL ACCESSORIES MA9004 pH 4 01 buffer solution 230 ml bottle MA9006 pH 6 86 buffer solution 230 ml bottle MA9007 pH 7 01 buffer solution 230 ml bottle MA9009 pH 9 18 buffer solution 230 ml bottle MA9010 pH 10 01 buffer solution 230 ml bottle MA9030 12 88 mS cm cal solution 230 ml bottle MA9015 Probe storage solution 230 ml bottle MA9016 General cleaning sol 230 ml bottle M10000B Rinse solution 20 ml sachet 25 pcs MA851D 1 Amplified pH EC TDS Temperature probe with DIN connector and 1 m 3 3 cable SPECIFICATIONS RANGE pH 0 00 to 14 00 pH EC 0 00 to 20 00 mS cm TDS 0 00 to 10 00 ppt Temp 0 0 to 60 0 C or 32 0 to 140 0 F RESOLUTION pH 0 01 pH EC 0 1 mS cm TDS 0 01 ppt Temp 0 1 C or 0 1F ACCURACY 25 C 0 01pH 2 FS EC TDS 0 5 or 1 TYPICAL EMC DEV 0 02 pH 2 FS EC TDS 0 5 or 1 F TEMPERATURE Automatic from 0 to 60 C COMPENSATION B adj from 0 0 to 2 4 C pH CALIBRATION 1 or 2 point with automatic buffer recognition EC CALIBRATION Automatic 1 point EC TDS CONVERSION adj from 0 45 to 1 00 FACTOR PROBE included MA851D 1 amplified pH EC TDS Temperature probe with DIN connector and 1 m 3 3 cable ENV
Download Pdf Manuals
Related Search
Related Contents
Leadingm=。vati。n))) 東芝白熱灯器具取扱眠明書 保呂用} MANUAL DE INSTRUCCIONES MANUAL DE INSTRUÇÕES M4号館(RI棟) 利用手引き User Manual Manuel de l`utilisateur IJP382B User Manual English PMID80BK_User_Manual カタログダウンロード Philips CitiScape Strada SHL5100BL USB KVM Extender User Manual Guid l`utilis Guide de l`utilisateur Copyright © All rights reserved.