1. G THE MATTRESS in FIGURE 6 DEFLATING MATTRESS this instruction sheet Part No 1078996 Rev B 5 INVACARE 8000 PRESSURE MATTRESS REPLACING CUSHION S FIGURE 7 NOTE Refer to WARNINGS in the SAFETY SUMMARY of this instruction sheet Remove patient from bed 2 Remove the therapy pad Refer to INSTALLING REMOVING THERAPY PAD in this instruction sheet 3 Unbutton the EXISTING cushion from the mattress base 4 Repeat STEP 3 for opposite side of the cushion 5 Pull the cushion fitting away from the manifold fitting to remove the cushion from the mattress NOTE Alternating cushions are connected together to form a chamber The mattress consists of two 2 chambers NOTE The cushions in the same chamber as the one removed in STEP 5 will deflate when the fittings are removed 6 Position NEW cushion into mattress 7 Insert cushion fitting into manifold fitting to secure 8 Button the cushion s to secure to the mattress base 9 If more cushions need replacing repeat STEPS 3 8 10 Inflate the mattress Refer to INFLATING THE MATTRESS in this instruction sheet Therapy Pad am Manifold Manifold Fitting Mattress Base Cushion Fitting Button bi Cushion FIGURE 7 REPLACING CUSHION S CARE AND MAINTENANCE The mattress can be cleaned by wiping it with a germicidal disinfectant solution 2 The mattress therapy pad is machine washable Air dry at a cool temperature INVACARE 8000 PRESSURE MATTRESS 6
2. Invacare 8000 Alternating Pressure Mattress Assembly Installation and Operating Instructions Model No CG8000 Yes you can NOTE The above model number contains one 1 Alternating Pressure Mattress and one 1 Power Unit NOTE Check ALL parts for shipping damage In case of shipping damage DO NOT use Contact Carrier Dealer for further instruction SAFETY SUMMARY To ensure the safe installation of the Invacare 8000 Alternating Pressure Mattress and Power Unit these instructions MUST be followed DO NOT install or use this equipment without first reading and understanding this instruction sheet If you are unable to understand the Warnings Cautions or Instructions contact a healthcare professional dealer or technical personnel before attempting to install this equipment otherwise injury or damage may occur DO NOT use the mattress near an open flame or while smoking Inspect power cord plug and air hoses of the Invacare 8000 Alternating Pressure Mattress Power Unit for breaks tears or kinks If damage is found DO NOT use DO NOT under any circumstances cut or remove the round grounding prong from any plug used with or for Invacare products Devices are equipped with three prong grounding plugs for protection against possible shock hazards Where a two prong wall receptacle is encountered it is the personal responsibility and obligation of the customer to contacta qualified electrician and have the two pron
3. Part No 1078996 Rev B LIMITED WARRANTY PLEASE NOTE THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4 1975 This warranty is extended only to the original purchaser user of our products This warranty gives you specific legal rights and you may also have other legal rights which vary from state to state Invacare warrants its product to be free from defects in materials and workmanship for one l year on the pump and six 6 months on the mattress for the original purchaser If within such warranty period any such product shall be proven to be defective such product shall be repaired or replaced at Invacare s option This warranty does not include any labor or shipping charges incurred in replacement part installation or repair of any such product Invacare s sole obligation and your exclusive remedy under this warranty shall be limited to such repair and or replacement For warranty service please contact the dealer from whom you purchased your Invacare product In the event you do not receive satisfactory warranty service please write directly to Invacare at the address on the next page Provide dealer s name address model number date of purchase indicate nature of the defect and if the product is serialized indicate the serial number Invacare Corporation will issue a return authorization The defective unit or parts must be returned for warranty inspection using t
4. SOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE PROVINCIAL LAWS AND REQUIREMENTS Part No 1078996 Rev B 7 INVACARE 8000 PRESSURE MATTRESS Invacare Corporation www invacare com USA Canada One Invacare Way 570 Matheson Blvd E Unit 8 Invacare is a registered trademark of Elyria Ohio USA Mississauga Ontario Invacare Corporation 44036 2125 L4Z 4G4 Canada Yes you can is a trademark of Invacare 800 333 6900 800 668 5324 Corporation Yes you can 2003 Invacare Corporation Part No 1078996 Rev B 05 03
5. g receptacle replaced with a properly grounded three prong wall receptacle in accordance with the National Electrical Code If you must use an extension cord use ONLY a three wire extension cord having the same or higher electrical rating as the device being connected DO NOT place hot objects for example heating pads electric blankets etc on top of the mattress This may weaken the mattress material Use extreme care when handling sharp objects such as needles near the mattress Do NOT inflate mattress ofan OCCUPIED bed The Alternating Pressure Mattress has a weight limitation of 250 Ibs 114 kg The Invacare 8000 APM MUST be installed on medical bed frames or medical grade home care beds with side rails After placing the patient on the mattress verify the side rails provide adequate protection for the patient The side rails MUST be in the raised position whenever a patient is on the bed INVACARE 8000 PRESSURE MATTRESS l Part No 1078996 Rev B INSTALLING THE INVACARE 8000 FIGURES AND 2 NOTE Refer to WARNINGS in the SAFETY SUMMARY of this instruction sheet Do one 1 of the following Power Unit Power Unit Air Hose Quick A Place the Invacare 8000 Power Unit on Disconnect a flat surface near the foot end of the bed B Open the hooks on the back of the Invacare 8000 Power Unit and install on the outside of the bed frame at the foot end of the bed as shown in FIGURE I Mattress Air Foot H
6. he serial number when applicable as identification within thirty 30 days of return authorization date DO NOT return products to our factory without our prior consent C O D shipments will be refused please prepay shipping charges LIMITATIONS AND EXCLUSIONS THIS WARRANTY SHALL NOT APPLY TO PROBLEMS ARISING FROM NORMAL WEAR OR FAILURE TO ADHERE TO THE ENCLOSED INSTRUCTIONS IN ADDITION THE FOREGOING WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PRODUCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DEFACED PRODUCTS SUBJECTED TO NEGLIGENCE ACCIDENT IMPROPER OPERATION MAINTENANCE OR STORAGE PRODUCTS MODIFIED WITHOUT INVACARE S EXPRESS WRITTEN CONSENT INCLUDING BUT NOT LIMITED TO MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE PRODUCTS DAMAGED BY CIRCUMSTANCES BEYOND INVACARE S CONTROL PRODUCTS REPAIRED BY ANYONE OTHER THAN AN AUTHORIZED INVACARE DEALER SUCH EVALUATION SHALL BE SOLELY DETERMINED BY INVACARE THE FOREGOING WARRANTY IS EXCLUSIVE AND IN LIEU OF ANY OTHER WARRANTIES WHATSOEVER WHETHER EXPRESSED OR IMPLIED INCLUDING THE IMPLIED WARRANTIES OF MERCHANTIBILITY AND FITNESS FOR A PARTICULAR PURPOSE AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PERSUANT TO THE TERMS CONTAINED HEREIN THE APPLICATION OF ANY IMPLIED WARRANTY WHAT
7. ose Quick 2 Remove existing mattress from bed Erd Disconnect 3 Place Invacare 8000 mattress on top of bed with the cushions facing up and air hoses at the foot end of the bed FIGURE l INSTALLING 4 Insert mattress air hose quick disconnects THE INVACARE 8000 into the power unit air hose quick disconnects as shown in FIGURE I Outside Bed Frame 5 Slide the transport mat inbetween the cushions and the mattress base as shown in FIGURE 2 NOTE The transport mat will ensure the patient has support should the cushions deflate 6 Inflate the mattress Refer to INFLATING THE MATTRESS in this instruction sheet INFLATING THE MATTRESS FIGURE 2 NOTE Refer to WARNINGS in the SAFETY SUMMARY of this instruction sheet NOTE Before inflating ensure that ALL cushion fittings both air hose quick disconnects and both CPR plugs are secured otherwise inflation will not occur Plug the power cord on the power unit into wall outlet 2 Turn the power switch to the ON position NOTE The green light on the power unit will be illuminated when in the ON position 3 Turn the knob CLOCKWISE on the power unit to the MAX setting NOTE It will take approximately 30 45 minutes for the mattress to be fully inflated at the maximum p s i NOTE Keep the hose quick disconnects attached to the mattress and power unit otherwise deflation will occur 4 Install the therapy pad to the mattress Refer to INSTALLING REMOVING THERAPY PAD in
8. peat STEP 3 until the patients body sinks 1 3 into the mattress INVACARE 8000 PRESSURE MATTRESS 4 Part No 1078996 Rev B Power Unit Knob p itch APR MAX Setting FIGURE 5 PLACING PATIENT ON MATTRESS AND ADJUSTING FIRMNESS DEFLATING MATTRESS FIGURE 6 NOTE Alternating cushions are connected together to form a chamber The mattress consists of two 2 chambers DETAIL A STANDARD DEFLATION i Power Unit Remove therapy pad Refer to INSTALLING REMOVING Power Unit Air Hose THERAPY PAD in this instruction sheet Quick Disconnect 2 Disconnect the two 2 power unit air hose quick disconnects from the two 2 mattress air hose quick disconnects as shown in DETAIL A Mattress Air Fold the mattress in half to force the air out Hose Quick Disconnect 4 Before re inflation is possible connect the two 2 power unit air hose quick disconnects to the two 2 mattress air hose quick disconnects 5 Inflate the mattress Refer to INFLATING THE MATTRESS in this instruction sheet RAPID CPR DEFLATION I Pull the two 2 CPR straps labeled PULL CPR away from mattress NOTE Using the PULL CPR straps will disconnect the CPR plugs from the manifold which allows the two 2 chambers to deflate 2 Before re inflation is possible install the CPR plugs on the CPR straps into the fittings on the manifold 3 Inflate the mattress Refer to INFLATIN
9. the bed frame 5 Repeat STEPS 4 for the remaining nine 9 double D rings NOTE Each SIDE contains three 3 double D rings and each END contains two 2 double D rings for a total of ten 10 double D rings 6 Adjust the mattress to the proper firmness Refer to PLACING PATIENT ON MATTRESS AND FIGURE 4 SECURING MATTRESS BASE ADJUSTING FIRMNESS in this TO BED FRAME instruction sheet PLACING PATIENT ON MATTRESS AND ADJUSTING FIRMNESS FIGURE 5 NOTE Refer to WARNINGS in the SAFETY SUMMARY of this instruction sheet Place patient on the mattress 2 Check the level that the patient sinks into mattress NOTE The pressure of the power unit can be adjusted to provide maximum effectiveness and pressure relief for the patient NOTE The pressure can be adjusted with a range from MIN to MAX NOTE Every other cushion inside the mattress is connected together known as a chamber The mattress consists of two 2 chambers NOTE The power unit will cycle every eight 8 minutes It will alternate inflating and deflating the two 2 air chambers NOTE The patients body should sink approximately 3 into the mattress 3 If the level is unacceptable do one 1 of the following A To increase firmness turn knob towards MAX setting B To decrease firmness turn knob towards MIN setting NOTE After adjusting the knob wait ten 10 minutes before rechecking the level the patient sinks into mattress 4 Re
10. this instruction sheet INVACARE 8000 PRESSURE MATTRESS 2 Part No 1078996 Rev B Power Unit Air Hose Quick Disconnect Cushion Fittings Transport Mat Cushion Mattress Air Hose Quick Disconnect Power Unit Knob CPR Plugs MAX Setting Power Switch Setting FIGURE 2 INFLATING THE MATTRESS INSTALLING REMOVING THERAPY PAD FIGURE 3 INSTALLING I Position therapy pad over mattress 2 Insert one 1 of the ten 10 straps into open latch as shown in FIGURE 3 3 Pull strap until therapy pad fits tight around mattress Push down on latch to secure 5 Repeat STEPS 2 4 for remaining nine 9 straps 6 Secure mattress base to the bed frame Refer to SECURING MATTRESS BASE TO BED FRAME in this instruction sheet REMOVING Remove patient from bed 2 Pull up on one 1 of the ten 10 latches that secure the therapy pad FIGURE 3 INSTALLING REMOVING to the mattress THERAPY PAD 3 Remove strap from latch Repeat STEPS 2 3 for remaining nine 9 latches 5 Remove therapy pad from mattress Part No 1078996 Rev B 3 INVACARE 8000 PRESSURE MATTRESS SECURING MATTRESS BASE TO BED FRAME FIGURE 4 Wrap one 1 of the ten 10 straps on the mattress around the bed frame Double D rings 2 Insert strap into the two 2 double D rings 3 Feed strap back through only one 1 of the double D rings 4 Pull the strap tight to secure the mattress base to